ROTARY VISCOMETER VP 1000 RM
Contents
1. shearthinning flow behavior Most common pseudoplastic fluids are coatings milk ink and jam Plastic Under static conditions they might have a similar behavior to a solid For a correct evaluation of the fluid it is necessary to reach the yield value to make fluid flow so that product later shows any of the possible material characterizations Newtonian pseudoplastic or dilatant Examples toothpaste chocolate grease Dilatant Viscosity of dilatant fluids increases when shear rate increases It is also called shearthickening flow behaviour Examples solutions of sugar and water and mixtures of sand and water Time dependent fluids Viscosity of dilatant fluids increases when shear rate increases It is also called shearthickening flow behaviour Thixotropic Those fluids in which viscosity and shear stress decrease maintaining a constant shear rate with time Ketchup honey anti drop paints mayonnaise Rheopeptic Those fluids in which viscosity and shear stress increase maintaining a constant shear rate with time Lubricants and some paints types are rheopeptic fluids Stand 1 06 09 13 Page 17 Portable viscometer VP 1000 RM Reference substances and viscosities eet Coatings airbrushed Nanocellulose common 8 000 10 000 additive Page 18 Stand 1 06 09 13 Portable viscometer VP 1000 RM 8 Viscosity table The viscosity scale readings for the different Rotors
2. on the viscometer by pressing for an extended time between 6 Units and 10 seconds the ENTER button After that an introductory display with a symbol of the battery state appears VPR 02 v2 01 Within the period of 2 seconds press one after the other UP and DOWN keys Display will show following Options Select Units Press ENTER to validate Screen will show the current mesuring unit being used Actual Units dPas Press UP or v keys to select between mPas or dPas Once selected press again to validate Instrument will return automatically to the presentation screen and to the measurement mode showing selected measuring unit Stand 1 06 09 13 Page 11 4 2 Measurement Preparing sample Measurement start Waiting for viscosity reading Overrange Finish of measurement Page 12 Portable viscometer VP 1000 RM Pour sample in the delivered measuring vessel Cup no 3 up to the filling mark and attach desired Rotor Locate Cup no 3 onto the base by means of the 3 centring pivots If using a different sample container centre Rotor correctly in sample To start the rotation and trigger the measurement press simply ENTER The Rotor beginns to rotate and the display delivers following data speed 62 5 rpm Rotor reference and question marks meaning that viscometer waits for viscosity reading rpm R3 B dPas 2 2 Viscosity readings in mPas dPas and full scale percen
3. Myr ROTARY VISCOMETER VP 1000 RM VISCOTECH HISPANIA S L Viscotech Hispania S L would like to thank you for the purchase of this equipment and for your confidence in choosing our company We are committed to offer you the best and most reliable service Don t hesitate to contact your local dealer for any question related to this viscometer Your Viscotech Hispania S L team User manual Stand 1 06 09 13 Portable viscometer VP 1000 RM Contents l General information ooo EE SOE Oa 1 LD Introd ction quum M 1 1 2 Intended 186 2 60 HEIDE E arkeseneaaesouneaseeeendacseeeet 1 LS Security Symbols aroser S 2 1 4 Safety not s ooo er I ccmsesidaniecdeannaeusloaadesiaddaaeebationasaseidsdsneessouadatscdvannewes 2 1 5 Certification warranty and documentation Lcceeeeeeee eee eee eene nnne 3 2 Technical data bI AE IN EAE LIAE IA D MAE MDC MM M d o 3 Setting up the viscometet ooo OWEPRERRRRRERREREEEE ERR RREFREERENEEREREERRE E EIN Er TEE 6 3 1 Reception and unpacking of the instrument Lcceeee ee eene eene noon 6 3 2 Scope of delivered parts 4 eoo piesni Axe Esa E o I ER ERERIFETCRUREREU UNUS ECC Ee RES 6 Jo SETUP mS T 3 4 Setting up the ROC Or ooo oO REIS CE a a 8 3 5 Meassuring Vessel Cup no 3 cccccccccccccccccccccccccccccccccccccccccssscccccccccccscssees 9 BE ODS TACO f x 10 4 1 Initial turn on and configur
4. VP 1000 RM is battery operated and work at a fixed speed 62 5 rpm Viscosity readings are delivered in two different units milli Pascal seconds mPas and deciPascal seconds dPas see conversion information in page 16 of this User Manual Measuring principle is based on a rotor immersed in the fluid to be tested and rotating at a constant speed fluid s resistance to the rotation is measured to obtain direct readings of viscosity in mPas and dPas Each Rotor included in basic delivery has a different measuring range Those viscosity values measuring range are only valid if measuring vessel Cup no 3 is used to obtain viscosity readings It is also possible to use any available vessel In this case viscosity results will be only comparative The most outstanding feature of the portable viscometers VP 1000 RM is their user friendliness This characteristic makes them an ideal tool to carry out measurements in the own premises during the manufacturing process for batch control for QC check or in laboratories research centres whenever quick and reliable viscosity readings are needed The VP 1000 RM viscometer has been developed specifically to fulfil viscosity readings in all kinds of fluids Only skilled or specially trained personnel must operate the viscometer Everyone who works with the viscometer VP 1000 RM must strictly follow the security precautions and observe the safety rules of the laboratory It is not allowed to oper
5. al dealer by giving the serial number of the instrument the number of the user manual and the date of issue This user manual is available in the following languages English and Spanish Errors and omissions excepted Subject to amendment and improvement without further notice Page 3 Manufacturer VISCOTECH HISPANIA S L Page 4 Vidriers 21 43700 EI Vendrell SPAIN Tel 34 977 66 80 20 Fax 34 977 66 80 30 viscotech viscotech es www viscotech es Portable viscometer VP 1000 RM Stand 1 06 09 13 Portable viscometer VP 1000 RM 2 Technical data Portable viscometer VP 1000 RM Alkaline batteries 4x AA LR6 pt Rechargeable batteries 4 x AA R6 Batteries capacity Alkaline 1 900 mAh Rechargeable 2 500 mAh Batteries autonomy 24 30 hours non stop use fixed speed 62 5 rpm selected Rotor R1 R3 dynamic viscosity mPas dPas full scale range full scale percentage battery level indicator Displayed data mere 100 compatible with RION VT 04F and HAAKE VT 2 Compatibility Plus Rotor amp Rotors no 1 2 and 3 fixed speed at 62 5 rpm Rotor 3 0 3 13 dPas 30 1 300 mPas with measuring vessel Cup no 3 Viscosity range Rotor 1 3 150 dPas 300 15 000 mPas with Rotors no 1 2 and 3 with measuring vessel Cup no 3 Rotor 2 100 4 000 dPas 10 000 400 000 mPas with measuring vessel Cup no 3 ee Measuring vessel Cup no 3 included in basic deliver
6. ate the viscometer for any other purpose as the one described in this chapter No right to claim warranty will be granted if any of the above norms are disregarded Page 1 Portable viscometer VP 1000 RM 1 3 Security symbols 1 4 Safety notes Page 2 Installation and operation of the VP 1000 RM viscometer are user friendly and very easy if instructions of this manual are strictly followed Be aware that any use of this equipment out of the scope described by the manufacturer may compromise the security of the operator Besides this and in order to stress the points that may represent any risk both for the operator and the viscometer itself this manual uses following symbols and messages DANGER Wherever this symbol along with the message DANGER appears in this manual it indicates a potential personal injury hazard or damage on the viscometer For your security strictly observe instructions provided CAUTION Wherever this symbol along with the message CAUTION appears in this manual it indicates that damages on the viscometer may occur if instructions are not observed For the proper operation of the equipment instructions must be strictly observed INFORMATION This symbol refers to specific details of the viscometer which must be specially considered because of their relevance Apart from the instructions contained in this manual everyone who operates
7. ation Lecce ee eeeeeeeeeeeess sesso eee esee eeeoe 10 4 2 Measurement oo o o EIER I Ier MR E EPI IO PEN ET P QI IO iN Nds 11 5 About viscometer s calibration eee e eee eee ee eee eee eese ses ee ee 14 5 1 Viscome ter s verification dore hives sS eod ned ee s E ER SERERE M a PE RED SiS NEM EUR 14 5 2 How to perform the auto Zeroing function ccceeeee eee eee eere noon 14 6 Accessories loptional s vodici esos vos aude ese coa a osa d os dave Pss RU rese uus 15 a ih idein mr m 16 Tob DES CLIP TION i5 onsec ve EAS ERN EQUI SNR ARA TEROPRME AR DNA EUIAERECKU anna A UAR EE eared 16 Tae Related concepts io ese LII I puteo eER E Ee ERATUM Eu 16 8 VISCOSITY table ovis ee euni a Desa Doo e ea E PD Peu genes 19 9 Problems and solut0hs ooo RO EEREKNOEERESTUEEI EE OUI ST PER P OPI SE ERIS 20 Stand 1 06 09 13 Portable viscometer VP 1000 RM 1 General information 1 1 Introduction 1 2 Intended use Stand 1 06 09 13 The VP 1000 RM viscometer of the MYR Group is a portable viscometer for quick flexible comparative viscosity measurements of a wide range of products The VP 1000 RM is not a hand held viscometer The idea of portability is here based in the light weight of the instrument 1 8 kg that permits transportation very easily The instrument is ready to use just unpack set and press a button Viscometer
8. harge of batteries is not enough to perform accurate measurements viscometer will inform user and will turn off automatically Low Battery Shutting Down Page 13 Portable viscometer VP 1000 RM 5 About viscometer s calibration The instrument delivered has been calibrated with standard Newtonian oils by the manufacturer before shipping We recommend carrying out calibrations on a regular basis according to operation frequency For the correct calibration of the viscometer please contact your local dealer 5 1 Viscometer s verification M TES RITE TED t We recomend to check the proper functionallity of the viscometer on a regular basis using our Newtonian viscosity standards available on demand Viscotech calibration oils are calibrated at two 2 different temperatures 20 and 25 C and allow a more accurate correlation between temperature and viscosity When carrying out the verification of the viscometr we suggest to work with Rotor no 3 if the complete set of Rotors is available and measuring vessel Cup no 3 Calibration standard suggested is of 1 000 mPas Once the viscometer began to rotate and after allowing a mimimum of 5 readings the value on the display should correspond with the one indicated in the certificate of the standard oil considering also allowances viscometer s accuracy standard accuracy involved Due to the great influence that temperature has on viscosity measureme
9. ion shown in the illustration on the left counter clockwise Once it has been inserted submerge the Rotor into a fluid until it reaches the notch that indicates the top immersion level allowed In doing so please take care not to hit neither the Rotor nor the pivot shaft with the vessel to avoid distortion of Rotor perpendicularity Rotors included in standard delivery have a thread connection to pivot shaft Optionally it is possible to order a coupling between Rotor and pivot shaft so that connection is done through a hook If this option is preferred a slight concentric movement of the Rotor during measurement is considered totally normal Be very careful when you perform this operation in order to prevent the Rotor from bending or the pivot shaft from suffering damages Both the Rotor and the screw must be in perfect conditions and extremely clean Rotors are made of stainless steel Each Rotor carries the reference name engraved on its head Stand 1 06 09 13 Portable viscometer VP 1000 RM 3 5 Measuring vessel Stand 1 06 09 13 To avoid that rotation stress might drag measuring vessel Cup no 3 containing sample to be analyzed viscometer features a very simple anchorage system three holes in the viscometer base where the fixing pins of the measuring vessel are inserted For a better accuracy of viscosity reading it is advisable to use the measuring vessel Cup no 3 incl
10. nts we strongly recommend to perform the viscometer s verification test with the oil at the temperature specified in the oil certificate 5 2 How to perform the auto zeroing function Page 14 If after turning on the viscometer and initiating the rotation of the Rotor without immersing it in a fluid the reading on the display does not indicate zero you should proceed as follows e Turn on the viscometer by pressing for a few seconds the ENTER button You will enter the presentation screen e Within the period of 2 seconds press the DOWN arrow v and ENTER consecutively in order to access to the AutoZERO option After this following information appears on the display Options AutoZERO Press ENTER to enter the following screen Stand 1 06 09 13 Portable viscometer VP 1000 RM Remove Spindle Press lt ENTER gt Don t forget that the AutoZERO option must be carried out without Rotor and with the viscometer placed on a levelled surface Press ENTER to start the AutoZERO process which will allow you to modify the zero value During this process following message appears on the display AutoZERO wait When the process has successfully finished the display will automatically return to the presentation screen and to the measurement mode 6 Accessories optional Battery charger R6 AA Ni Cd Ni Mh Rechargeable batteries Rotor rack Special coupling for hook s
11. nvironment According to this transmitters such as mobile phones must not be operated near to it In case of malfunctions and or service work please turn off the instrument and contact your local dealer 1 5 Certification warranty and documentation Certification Warranty Documentation Stand 1 06 09 13 Viscotech Hispania S L certifies that this instrument has been tested and checked carefully Its technical data has been verified before shipment to be in accordance with the published specifications The instrument complies with applicable international safety regulations This MYR product is warranted against defects in material and workmanship for a period of 2 two years Parts which prove to be defective during the warranty period will be repaired or replaced free of charge No other warranty is expressed or implied Unauthorized modification or repair by third party persons will void the warranty The warranty will expire in case of improper or wrong use of the instrument and in case the warning and precautionary messages are not observed Viscotech Hispania S L is not liable for any occurring damage Once the warranty period has expired we recommend signing a service contract in order to guarantee the perfect functionality of the instrument For more information on this subject please contact your local dealer This user manual will be delivered together with each instrument Further copies can be ordered at the loc
12. overlap partially Depending mainly on the properties of the sample fluid viscosity readings obtained with different Rotors can be also different To evaluate correctly results it is necessary to note down Rotor number and viscosity result obtained Viscometer VP1000 R has 3 measuring ranges depending on the Rotor used The viscosity values stated both dPas or mPas are only valid when the measuring vessel cup no 3 is used with the Rotors ROTOR Measuring range Measuring range mPas Measuring Sample dPas Cup Vessel capacity 0 3 13 dPas 30 1 300 mPas 170 ml 3 150 dPas 300 15 000 mPas 170 ml 100 4 000 dPas 10 000 400 000 mPas t 170 ml Stand 1 06 09 13 Page 19 Portable viscometer VP 1000 RM 9 Problems and solutions Problem The viscometer does not work The instrument does no read zero without fluids Viscosity speed is unstable and inaccurate Page 20 Check the battery level Carry out the AutoZERO function Perform a calibration check using calibration standards Check if the Rotor rotates steadily Check if the sample temperature is stable Check the rheological features of the sample that must be analyzed Stand 1 06 09 13
13. peed from which a transfer of mass between layers occurs Result is an apparently greater shear stress and an erroneously high viscosity reading Turbulent flow is characterized by a sudden and notorious increase in viscosity above a certain speed Newtonian fluids Viscosity in Newtonian fluids at a given temperature remain constant regardless of viscometer model spindle and speed being used The most common Newtonian fluids are water and thin motor oils Page 16 Stand 1 06 09 13 Portable viscometer VP 1000 RM Non Newtonian fluids This kind of fluids does not show a lineal relation between shear stress and shear rate Different working conditions have asa result different viscosity values Apparent viscosity is defined as the result of a fluid analysis This result can be reproduced in another viscometer only if analysis is carried out maintaining identical working conditions and following a defined working process Variables below influence results Viscometer model Dimensions of sample container Filling level Sample temperature Rotor Rotating speed Duration of test time dependant fluids Generally speaking each modification in the working method and working process will indefectibly lead to variations in final analysis results There are different behaviors within the non Newtonian fluids Pseudoplastic Samples whose viscosity decreases when increasing shear rate It is also called
14. pindles Newtonian certified calibration oils 600 ml Stand 1 06 09 13 Page 15 Portable viscometer VP 1000 RM 7 About viscosity 7 1 Description Myr viscometer VP 1000RM are rotational and portable viscometers to determine the viscosity of a wide range of fluids The principle applied by this viscometer is equivalent to that of the rotational viscometers a Rotor turning at a constant speed is immersed in the sample fluid to be analyzed The resistance to Rotor movement caused by the viscosity of fluid being analyzed is displayed directly in mPas dPas Depending on the viscosity the resistance exerted by a fluid opposing the motion of a Rotor varies depending on the size of the Rotor The viscometer has been calibrated to deliver readings in mPas dPas according to speed and Rotor type used 7 2 Related concepts Viscosity Viscosity is a distinctive property of the fluids It is the measure of internal friction of a fluid when a layer of this fluid is forced to move in relation to another layer Viscosity is a value highly dependant on temperature The standard units for dynamic viscosity measurements are mPa s S I or cP C G S 1lmPas miliPascal second 1cP centiPoise 1 dPas deciPascal second 100 mPas 1dPas 1P Poise Laminar flow It is the ideal movement between layers without transfer of mass from one to the other It is the base to calculate dynamic viscosity Turbulent flow There is a certain s
15. re battery operated The basic delivery scope comprises 4 alkaline batteries 4xAA LR6 6V installed in the unit and 4 additional batteries for replacement The unit can work both with alkaline batteries from type AA LR6 and with rechargeable ones from type AA R6 In continuous operation batteries have a lifespan of 24 to 30 hours Battery level is indicated on the display by means of a symbol VPR 02 v2 01 Batteries are positioned at the back side of the viscometer To replace them proceed as follows E 22 cE Meds ic Sean 6 DC 411 5 AA LRG Bar He VP DOLZ IP20 Mis EHEER EE 0 u NR EE Remove the screws as indicated in the illustration Remove the cover plate of the battery housing and replace the batteries Please pay attention to the correct polarity when doing this operation To replace the cover in its original position place correctly the cover and tighten the screws again On the cover plate there is a label with the most relevant data of the viscometer Stand 1 06 09 13 Page 7 Portable viscometer VP 1000 RM 3 4 Setting up the Rotor Page 8 All Rotors included in the standard delivery must be smoothly immersed into the fluid to avoid bubble formation below the surface To assemble the Rotor hold firmly the pivot shaft with one hand and screw up the Rotor onto it using the other hand To screw up the Rotor turn it in the direct
16. tage will be displayed afterwards If user operates with a different container than the measuring vessel Cup no 3 delivered with instrument it is absolutely necessary to centre Rotor in the sample container before operate If following display appears during the measurement process you have exceeded the maximum reading s limit overrange In this case change the Rotor considering full scale of each one For a correct selection enter the Rotor selection screen to check FSR for each one It is advisable to operate with viscometer in the best measuring range this is between gt 10 and lt 90 of full scale range Nevertheless viscometer shows viscosity readings within the whole range To finish the measurement process press ENTER again to stop the drive The measurement result along with the full scale percentage will remain on the display To start another measurement proceed as described before Stand 1 06 09 13 Turn off the viscometer Stand 1 06 09 13 Portable viscometer VP 1000 RM Press the ENTER key for an extended time to turn off the viscometer By doing this following information appears on the display Power Off Shutting Down Don t forget to switch off the instrument by pressing for a few seconds the ENTER button when you have finished the measurements Otherwise the viscometer will remain active and batteries will run out In the event that c
17. the portable viscometer VP 1000 RM is obliged to know and observe the security and hygienic measures of the laboratory where this device is operated Everyone who should operate and set up the viscometer must have completely read and understood these instructions before starting the operations Operator security may be at risk if instructions of this manual are not observed Please observe the following general precautions during operation of this instrument Failure to comply with these precautions violates safety standards and the intended use of the instrument Viscotech Hispania S L is not liable for misuse of the instrument and failure to comply with basic safety requirements Stand 1 06 09 13 Battery operated Danger in explosive environment Hazard of malfunction Portable viscometer VP 1000 RM The viscometer works with batteries It is strictly forbidden to remove the instrument hood during operation Component changes and settings must only be fulfilled by specialized personnel Use only original spare parts when changing any component Battery replacement must be carried out after switching off the unit and removing the Rotor from the shaft The instrument must not be operated in the presence of flammable gases It is also forbidden to expose the instrument in environments where dangerous gas concentrations may occur To avoid damages on the instrument it must be only operated in a controlled electromagnetic e
18. uded in basic delivery If a different container is used the viscosity resistance applied to the Rotor will be different leading to deviations in the viscosity results As the viscometer is not equipped with regulating feet the surface where the instrument is placed must be perfectly levelled The filling mark 170 ml in measuring vessel Cup no 3 must be respected to be able to work with complete set of Rotors no 1 2 and 3 Page 9 Portable viscometer VP 1000 RM 4 Operation m Initial turn on and configuration The configuration of the VP 1000 RM viscometer is very simple The mt 7 only parameters that can be modified are the Rotor and the measuring unit The information on the display can be modified VP 1000 using the ENTER button and the two arrows 4 up and v down rv Initial turn on and Turn on the viscometer by pressing for an extended time between 6 configuration and 10 seconds the ENTER button After that an introductory display with a symbol of the battery state appears VPR 02 v2 01 Shortly afterwards you can view the main display with the configuration that was last used with the viscometer rpm STOP R3 E dPas 0 The display includes the following information speed rpm measurement unit mPas dPas Rotor type R1 R2 R3 and full scale percentage The viscometer works always at one speed 62 5 rpm The only parameters that can be modified are the Rotors and the measurement
19. units milliPascal seconds mPas or deciPascal seconds dPas Should you wish to carry out measurements with a different Rotor as the one that appears on the display follow these steps to modify it Press shortly any one of the arrows 4 or v In doing so the display will show following information Rotor E R3 13dPas The Rotor reference last used flashes on the display At the right side you can see the full scale range corresponding to the combination of selected Rotor and instrument speed Page 10 Stand 1 06 09 13 Portable viscometer VP 1000 RM The full scale range refers to the maximum viscosity reading that the viscometer can achieve with selected combination of speed and Rotor This value helps the user to select the correct Rotor according to the viscosity of the sample that must be analyzed Use the arrows 4 or v to set the correct Rotor type between R1 and R3 When the correct type appears on the display press ENTER to validate After that the main display appears again with the new modification introduced If by any reason you omit to press ENTER during 15 seconds when modifying the Rotor the viscometer returns automatically to the main display and will accept the last Rotor selected It is therefore very important to check that the Rotor that has been set up in the viscometer matches with the one that is being showed on the display Selecting Measuring Turn
20. y ea eae 2 of full scale De ee 1 a 17 x 11 x 41 cm Weight 8 from 10 C until 40 C at a max rel humidity of 80 without Operating conditions condensation altitude up to 2000 m M S L for indoor use only Pollution degree Protection category Stand 1 06 09 13 Page 5 Portable viscometer VP 1000 RM 3 Setting up the viscometer 3 1 Reception and unpacking of the instrument Follow these steps after receiving the instrument Please check carefully the packaging box of the viscometer before removing it in order to find out transportation damages Should the packaging box be broken or have suffered any other damage please contact the shipping agency before opening it Once you have taken the instrument out of the box check if damages can be observed Should this be the case please inform the dealer that has delivered you the equipment Remove all packing straps protectors and accessories used during transport Recyclable materials are to be disposed in the containers provided therefore 3 2 Scope of delivered parts e Portable rotational viscometer VP 1000 RM e Set of rotors Rotor no 1 Rotor no 2 Rotor no 3 Material stainless steel e Measuring vessel Cup no 3 Approx capacity 170 ml Material DELRIN e Alkaline additional batteries AA LR6 6V Page 6 Stand 1 06 09 13 Portable viscometer VP 1000 RM 3 3 Setup Portable viscometers VP 1000 RM a