1.       2013 OMRON HEALTHCARE  INC     Made in Vietnam 2298802 0D    
2.   Change Confirm    3  Press the START STOP button to turn the monitor off        Notes       If the batteries have been replaced  the date and time setting will need to be  reset       If the date and time are not set             appears during or after measurement     14    USING THE DEVICE    Applying the Arm Cuff    Remove tight fitting clothing or tight rolled up sleeve from your left upper arm   Do not place the arm cuff over thick clothes     1  Insert the air plug into the air jack  securely              2  Wrap the arm cuff firmly in    place around your left upper  arm        The bottom edge of the arm cuff  should be 1 2 inch  1 to 2 cm  above  the elbow  Air tube is on the inside  of your arm and aligned with your  middle finger         1 2 inch   1 2 cm     15    USING THE DEVICE    3  Secure closed with the       J  fabric fastener       Notes     When you take a measurement on the right  arm  the air tube will be at the side of your a  elbow  Be careful not to rest your arm on the  air tube       The blood pressure can differ between the right arm and the left arm  and  the measured blood pressure values can be different  OMRON  recommends to always use the same arm for measurement  If the values  between both arms differ substantially  please check with your physician  which arm to use for your measurements     16    USING THE DEVICE    How to Sit Correctly   To take a measurement  you need to  be relaxed and comfortably seated   under comfortable room tempera
3.  Movement Error Symbol  fi    The movement error symbol is displayed if you move your body during the  measurement  Please remove the arm cuff  and wait 2   3 minutes    Take another measurement  remain still during measurement     Average Value Symbol    i    The average value symbol is displayed when you press and hold the memory  button for more than 3 seconds  The most recent average value appears on the  display screen     10    KNOW YOUR DEVICE    Blood Pressure Level Indicator       2013 ESH ESC Guidelines for the management of arterial hypertension  Definitions of hypertension by office and home blood pressure levels    Systolic Blood Pressure 2 140 mmHg 2 135 mmHg  Diastolic Blood Pressure 2 90 mmHg  gt  85 mmHg    These are from statistical values for blood pressure             11    KNOW YOUR DEVICE    Before Taking a Measurement   To help ensure an accurate reading  follow these directions    1  Avoid bathing  drinking alcohol or caffeine  smoking  exercising and eating for  30 minutes before taking a measurement  Rest for at least 15 minutes before  taking the measurement    2  Stress raises blood pressure  Avoid taking measurements during stressful  times    3  Measurements should be taken in a quiet place    4  Remove tight fitting clothing from your arm    5  Sit on a chair with your feet flat on the floor  Rest your arm on a table so that  the arm cuff is at the same level as your heart    6  Remain still and do not talk during the measurement    7  Keep a re
4.  bend or pull    Do not twist  Do not bundle during use    Do not pinch  Do not place under heavy objects     A  Wipe the dust off from the power plug    A  Unplug monitor when not in use    A  Disconnect the power plug before cleaning    A Use only an OMRON AC adapter designed for this device  Use of  unsupported adapters may damage and or may be hazardous to the device     Battery Usage    A  Do not insert the batteries with their polarities incorrectly aligned     A Use only 4    AA    alkaline or manganese batteries with this device  Do not use  other types of batteries  Do not use new and used batteries together   A Remove the batteries if the device will not be used for three months or more     IMPORTANT SAFETY INFORMATION    General Precautions     Do not forcibly crease the arm cuff or the air tube excessively      Do not press the air tube while taking a measurement    e To unplug the air plug  pull on the air plug at the connection with the monitor   not the tube itself      Do not drop the monitor or subject device to strong shocks or vibrations      Do not inflate the arm cuff when it is not wrapped around your arm      Do not use the device outside the specified environment  lt may cause an  inaccurate reading    e Dispose of the device  components and optional accessories according to  applicable local regulations  Unlawful disposal may cause environmental  pollution     KNOW YOUR DEVICE    Contents   Monitor  arm cuff  battery set  instruction manual  quick sta
5.  into a digital reading   An oscillometric monitor does not need a stethoscope so the monitor is simple  to use     Intended Use   This device is a digital monitor intended for use in measuring blood pressure  and pulse rate in adult patient population  The device detects the appearance of  irregular heartbeats during measurement and gives a warning signal with the  measurement result     Cli  Please read this instruction manual thoroughly before using the device   Please keep for future reference  For specific information about your own  blood pressure  CONSULT YOUR PHYSICIAN     IMPORTANT SAFETY INFORMATION    A Warning  Indicates a potentially hazardous situation which  if not   avoided  could result in death or serious injury     General Usage    A DO NOT adjust medication based on measurement results from this blood  pressure monitor  Take medication as prescribed by your physician  Only a  physician is qualified to diagnose and treat High Blood Pressure    A This device is not intended to be a diagnostic device    A Consult your physician before using the device for any of the following  conditions  common arrhythmias such as atrial or ventricular premature  beats or atrial fibrillation  arterial sclerosis  poor perfusion  diabetes  age   pregnancy  pre eclampsia  renal diseases    Note that PATIENT motion  trembling  shivering may affect the measurement  reading    A Do not use the device on the injured arm or the arm under medical  treatment    A Do not apply the arm 
6.  transmitters as determined by  an electromagnetic site survey   should be less than the    compliance level in each frequency range      Interference may occur in the vicinity of equipment marked  with the following symbol      o     Note1  At 80 MHz and 800 MHz  the higher frequency range applies   Note2  These guidelines may not apply in all situations  Electromagnetic propagation is affected by  absorption and reflection from structures  objects  and people     a Field strengths from fixed transmitters  such as base stations for radio  cellular  cordless  telephones  and land mobile radio  AM and FM radio broadcast  and TV broadcast cannot be predicted  theoretically with accuracy  To assess the electromagnetic environment due to fixed RF transmitters   an electromagnetic site survey should be considered  If the measured field strength in the location in  which the OMRON BP742N including AC adapter is used exceeds the applicable RF compliance  level above  the OMRON BP742N including AC adapter should be observed to verify normal  operation  If abnormal performance is observed  additional measures may be necessary  such as  reorienting or relocating the OMRON BP742N including AC adapter      gt  Over the frequency range 150 kHz to 80 MHz  field strengths should be less than 3 V m        36    GUIDANCE AND MANUFACTURER S DECLARATION    Recommended separation distance between portable and mobile RF communications  equipment and the OMRON BP742N including AC adapter    OMRON 
7.  within 10 minutes    Notes    e If there are only 2 measurement values in the memory for that period  the  average will be based on these 2 values    e If there is 1 measurement value in the memory for that period  this is displayed  as the average    e If the memory is full  the monitor will delete the oldest value      When viewing the measurement value taken without setting the date and time              is displayed instead of the date and time     To View the Measurement Values Stored in Memory  1  Select your User ID  1 or 2    2  Press the E button     The Memory number appears for a second  before the pulse rate is displayed   The newest set is numbered  1                  display       20    USING THE DEVICE    3  Press the    or  gt  button to view the values stored in  memory    lt   To view the older values   gt  gt    To view the more recent values       To View the Average Value  1  Select your User ID  1 or 2      2  Press and hold the    button for  more than 3 seconds              Notes      If the previous measurement was taken without setting the date and time   the average value is not calculated      If there are no measurements results stored in the memory   the screen to the right is displayed        21    USING THE DEVICE    To Delete All the Values Stored in Memory  The values stored in the memory are deleted by USER ID     1  Select your User ID  1 or 2    2  Press the E button  while the memory symbol appears     3  While holding the Ez  button dow
8. AND OMRON HEREBY DISCLAIMS  ANY OTHER WARRANTIES  EXPRESS OR IMPLIED  INCLUDING IMPLIED  WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR  PURPOSE  IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE  IMPOSED BY LAW  IF ANY  ARE LIMITED IN DURATION TO THE PERIOD  OF THE ABOVE EXPRESS WARRANTY     OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER  SPECIAL  INCIDENTAL  CONSEQUENTIAL OR INDIRECT COSTS   EXPENSES OR DAMAGES     This warranty provides you with specific legal rights  and you may have other  rights that vary by jurisdiction  Because of special local requirements  some of  the above limitations and exclusions may not apply to you    FOR CUSTOMER SERVICE   Visit our web site at  www omronhealthcare com  Call toll free  1 800 634 4350    32    GUIDANCE AND MANUFACTURER S DECLARATION    OMRON Automatic Blood Pressure Monitor  Information for accompanying documents in the scope of IEC60601 1 2 2007  Model  BP742N including AC adapter       Important information regarding Electro Magnetic Compatibility  EMC     With the increased number of electronic devices such as PC   s and mobile  cellular   telephones  medical devices in use may be susceptible to electromagnetic interference  from other devices  Electromagnetic interference may result in incorrect operation of the  medical device and create a potentially unsafe situation  Medical devices should also not  interfere with other devices     In order to regulate the requirements for EMC  Electro Magnetic Compati
9. BP742N including AC adapter is intended for use in an electromagnetic environment in  which radiated RF disturbances are controlled  The customer or the user of this OMRON BP742N  including AC adapter can help prevent electromagnetic interference by maintaining a minimum  distance between portable and mobile RF communications equipment  transmitters  and the  OMRON BP742N including AC adapter as recommended below  according to the maximum    output power of the communications equipment   Separation distance according to frequency of  transmitter in meter  Output Power of Transmitter    in Watt 150 kHz to 80    MHz  d 1 2 VP    80 MHz to 800 MHz   800 MHz to 2 5GHz    For transmitters rated at a maximum output power not listed above  the recommended separation   distance d in meters  m  can be estimated using the equation applicable to the frequency of the   transmitter  where P is the maximum output power rating of the transmitter in watts  W  according   to the transmitter manufacturer    Note  At 80 MHz and 800 MHz  the separation distance for the higher frequency range applies    Note  These guidelines may not apply in all situations  Electromagnetic propagation is affected by  absorption and reflection from structures  objects  and people        37    Manufactured for   OMRON HEALTHCARE Co   Ltd   53  Kunotsubo  Terado cho  Muko  Kyoto  617 0002 JAPAN    Distributed by    OMRON HEALTHCARE  INC    1925 West Field Court Lake Forest  IL 60045 U S A   www omronhealthcare com 
10. OMRON     INSTRUCTION MANUAL    5 series  Blood Pressure Monitor    Model BP742N    ESPANOL           ENGLISH    TABLE OF CONTENTS    Before using the Monitor  IO UC iii A 3  Important Safety InformatiON     oooooocnncccnnnocccnnocccnnonccconanncnonnnnnnanacanonnccnnnnos 4    Operating Instructions  Know Your Device    cocina ride  Display symbols fea  Before Taking a Measurement                cccceeeeteeeeeeeeeeeeteeeeeeaee 12   Preparation  a A  Battery Installation  Setting the Date and Time                      Using the Device  cocos tir ti nie  Applying the Arm Cuff     15  How to Sit Correctly       2 17  Taking a Measurement            17  Using the Memory Function                 ccccceeeeeeeeeeeeeeeeeteeeeeneees 20  Care and Maintenance  Error Messages and Troubleshooting               cccccceecceeeeeeeeeeeeeeeeteeeeeeaee 23  Error Messages a  TOUS IN tee eee cee es ee ele 25  Maintenance and Storage  Maintenance    ses  O  Optional Medical Accessories    ooocconocccinocccconacccnnnncccnnnnncnonnnos 27  Specifications a  FCC Statement niasin cca AA 31  Limited  Warranty  iveco caidas rr abana nes 32  Guidance and Manufacturer   s Declaration               ccc cece tee eeeeeeeee 33    INTRODUCTION    Thank you for purchasing the OMRON   BP742N Blood Pressure Monitor   Your new blood pressure monitor uses the oscillometric method of blood  pressure measurement  This means the monitor detects your blood movement  through your brachial artery and converts the movements
11. Ur  in Ur  an uninterruptible power  for 5 sec  for 5 sec  supply     Surge  IEC 61000 4 5    Power frequency magnetic   50 60 Hz  magnetic fields should be at levels   field characteristic of a typical   IEC 61000 4 8 location in a typical commercial  or hospital environment     Power frequency    Note  Ur is the A C  mains voltage prior to application of the test level        35    GUIDANCE AND MANUFACTURER S DECLARATION    Guidance and manufacturer   s declaration   electromagnetic immunity  OMRON BP742N including AC adapter is intended for use in the electromagnetic environment  specified below  The customer or the user of this OMRON BP742N including AC adapter should assure  that it is used in such environment     Compli   Immunity test coat ance Electromagnetic environment   guidance    Portable and mobile RF communications equipment should be  used no closer to any part of the OMRON BP742N including  AC adapter including cables  than the recommended  separation distance calculated from the equation appropriate to  the frequency of the transmitter     Conducted RF  3 V rms Recommend separation distance  IEC 61000 4 6   150 kHz to  80 MHz d 1 2 4P    Radiated RF  3 V m d  1 2 4P  IEC 61000 4 3   80 MHz to 80 MHz to 800 MHz  2 5 GHz d 2 3 VP  800 MHz to 2 5 GHz    where P is the maximum output power rating of the  transmitter in watts  W  according to the transmitter  manufacturer and d is the recommended separation distance  in meters  m     Field strengths from fixed RF
12. bility  with the  aim to prevent unsafe product situations  the IEC60601 1 2 standard has been  implemented  This standard defines the levels of immunity to electromagnetic  interferences as well as maximum levels of electromagnetic emissions for medical  devices     Medical devices manufactured by OMRON Healthcare conform to this IEC60601 1 2 2007  standard for both immunity and emissions     Nevertheless  special precautions need to be observed   The use of accessories and cables other than those specified by OMRON  with the    exception of cables sold by OMRON as replacement parts for internal components  may  result in increased emission or decreased immunity of the device     The medical devices should not be used adjacent to or stacked with other equipment   In case adjacent or stacked use is necessary  the medical device should be observed to  verify normal operation in the configuration in which it will be used     Refer to further guidance below regarding the EMC environment in which the device  should be used     The MEDICAL ELECTRICAL EQUIPMENT BP742N including AC adapter needs  special precautions regarding EMC and needs to be installed and put into service  according to the EMC information provided in this documentations     The Essential Performance of the BP742N including AC adapter is to measure a blood  pressure and a pulse rate and using the memory function     The BP742N including AC adapter may be interfered with by other equipment  even if that  other equ
13. cord of your blood pressure and pulse readings for your physician   A single measurement does not provide an accurate indication of your true  blood pressure  You need to take and record several readings over a period  of time  Try to measure your blood pressure at the same time each day for  consistency     12    PREPARATION    Battery Installation  1  Remove the battery cover              2  Insert 4    AA    batteries as  indicated in the battery  compartment           3  Replace the battery cover    Notes      When the low battery symbol  7   appears on the display  turn the monitor  off  then replace all batteries at the same time  Long life alkaline batteries  are recommended      The measurement values continue to be stored in memory even after the  batteries are replaced      The supplied batteries may have a shorter life      Dispose of the device  components and optional accessories according to  applicable local regulations  Unlawful disposal may cause environmental  pollution     13    PREPARATION    Setting the Date and Time    Set the monitor to the correct date and time before taking a measurement for  the first time     1  Press the    button           2  Push    or    to change the year     Push    button to confirm the year and then the month flashes  Repeat the  same steps to change the month  day  hour  and minute                            Hour   f Minute  Change Confirm mat ao y   gt    Forward a Confirm  Ca D ak 00 qa  4   Back Change Confirm Cr IO     
14. cuff on the arm while being on an intravenous drip or  blood transfusion    A Consult your physician before using the device on the arm with an arterio   venous  A V  shunt    A Do not use the device with other medical electrical  ME  equipment  simultaneously    A Do not use the device in the area the HF surgical equipment  MRI  or CT  scanner exists  or in the oxygen rich environment    A The air tube or the AC adapter cable may cause accidental strangulation in  infants    A Contains small parts that may cause a choking hazard if swallowed by  infants     IMPORTANT SAFETY INFORMATION    A Caution  Indicates a potentially hazardous situation which  if not  avoided  may result in minor or moderate injury to the user or  patient or damage to the equipment or other property     General Usage    A Always consult your physician  Self diagnosis of measurement results and   self treatment are dangerous    A Consult your physician before using the device for any of the following   conditions      If you have had a mastectomy      Do not take measurements more than necessary  It may cause bruising  due to blood flow interference      People with severe blood flow problems or blood disorders as cuff inflation  can cause bruising     A Remove the arm cuff if it does not start deflating during the measurement    A Do not use this device on infants or persons who cannot express their  intentions    A Do not use the device for any purpose other than measuring blood pressure    A  Use onl
15. inst  electric shock E   Class II ME equipment  Optional AC adapter     29    SPECIFICATIONS    Notes     e These specifications are subject to change without notice      In the clinical validation study  the 5th phase was used on 85 subjects for  determination of diastolic blood pressure    e This device has not been validated for use on pregnant patients     30    FCC STATEMENT    FCC CAUTION  Changes or modifications not expressly approved by the party responsible for  compliance could void the user   s authority to operate the equipment     Note    This equipment has been tested and found to comply with the limits for a Class B   digital device  pursuant to part 15 of the FCC Rules  These limits are designed to   provide reasonable protection against harmful interference in a residential   installation  This equipment generates  uses and can radiate radio frequency   energy and  if not installed and used in accordance with the instructions  may   cause harmful interference to radio communications  However  there is no   guarantee that interference will not occur in a particular installation  If this   equipment does cause harmful interference to radio or television reception  which   can be determined by turning the equipment off and on  the user is encouraged to   try to correct the interference by one or more of the following measures    e Reorient or relocate the receiving antenna    e Increase the separation between the equipment and receiver      Connect the equipmen
16. ipment complies with CISPR EMISSION requirements        33    GUIDANCE AND MANUFACTURER S DECLARATION    Guidance and manufacturer   s declaration   electromagnetic immunity    OMRON BP742N including AC adapter is intended for use in the electromagnetic environment  specified below  The customer or the user of this OMRON BP742N including AC adapter should  assure that it is used in such environment       Emissions test   Compliance   Electromagnetic environment   guidance  The OMRON BP742N including AC adapter uses RF    RF emissions energy only for its internal function  Therefore  its RF  CISPR 11 emissions are very low and are not likely to cause any    interference in nearby electronic equipment   RF emissions  CISPR 11 ClassiB    Harmonic The OMRON BP742N including AC adapter is  emissions Class A suitable for use in all establishments  including  IEC 61000 3 2 domestic establishments and those directly connected    Voltage to the public low voltage power supply network that  fluctuations flicker   supplies buildings used for domestic purposes   oe Complies  emissions  IEC61000 3 3       34    GUIDANCE AND MANUFACTURER S DECLARATION    Guidance and manufacturer   s declaration   electromagnetic immunity    OMRON BP742N including AC adapter is intended for use in the electromagnetic environment  specified below  The customer or the user of this OMRON BP742N including AC adapter should  assure that it is used in such environment     r IEC 60601 i Electromagnetic  Immunity 
17. n  press the  START STOP button for    more than 3 seconds       start  eG       Note  You cannot partially delete the values stored in the memory   All values for the user you select will be deleted                             22    ERROR MESSAGES AND TROUBLESHOOTING       Error Messages    Irregular heartbeats  are detected     Movement during  measurement     The batteries are low     The batteries are  exhausted     Air plug  disconnected     Arm cuff is applied  too loosely     Air is leaking from  the arm cuff     Remove the arm cuff  Wait 2   3 minutes  and then take another measurement   Repeat the steps in    Taking a  Measurement     If this error continues to  appear  contact your physician     Carefully read and repeat the steps in     Taking a Measurement        You should replace the batteries with  new ones ahead of time   Refer to    Battery Installation        You should replace the batteries with  new ones at once   Refer to    Battery Installation        Insert the plug securely   Refer to    Applying the Arm Cuff        Apply the arm cuff tighter   Refer to    Applying the Arm Cuff        Replace the arm cuff with a new one   Refer to    Optional Medical Accessories           23    ERROR MESSAGES AND TROUBLESHOOTING  rs IC ICI    Repeat measurement  Remain still  and do not talk during measurement   Movement during Refer to    Taking a Measurement      measurement and  g the arm cuff has not  If    E2    appears repeatedly  inflate  been inflated the arm c
18. omatically turn off after 2 minutes           Note  Wait 2 3 minutes before taking another measurement  Waiting  between measurements allows the arteries to return to the condition  prior to taking a measurement     Using the Guest Mode   The monitor stores measurement values for 2 users in the memory    The guest mode can be used to take a single measurement for another user    No measurement values are stored in the memory when the guest mode is selected     1  Press and hold the START STOP    button for more than 3 seconds    The USER ID symbol and the Date Time   display will disappear  E  E      MM      A   i ERORE y    A                    2  Release the START STOP button when the Date Time  display turns off     The arm cuff will start to inflate automatically     A DO NOT adjust medication based on measurement results from this blood  pressure monitor  Take medication as prescribed by your physician  Only a  physician is qualified to diagnose and treat High Blood Pressure     A This device is not intended to be a diagnostic device     19    USING THE DEVICE    A Always consult your physician  Self diagnosis of measurement results and  self treatment are dangerous     A  Inflating to a higher pressure than necessary may result in bruising where  the arm cuff is applied     Using the Memory Function   The monitor automatically stores the results up to 50 sets for each user  1 and 2     lt can also calculate an average value based on the last 3 measurement values   taken
19. r from the electrical  outlet first  and then remove the AC adapter plug from the monitor     28    SPECIFICATIONS                Model BP742N _  REF  HEM 7131 Z  Display LCD digital display   Pressure  0 to 299 mmHg  Measurement range Pulse  40 to 180 beats   min    Pressure   3 mmHg or 2  of reading  Accuracy Pulse    5  of display reading  Inflation Fuzzy logic controlled by electric pump  Deflation Automatic pressure release valve  Measurement method  Oscillometric method  IP classification IP 20   4    AA    batteries 1 5V or optional AC adapter  Power source  INPUT AC100 240V 50 60Hz 0 12A   Battery life Approximately 1000 measurements  using new alkaline batteries     Operating temperature    no o o o  humidity 50  F to 104  F  10  C to 40  C    15 to 90  RH    Storage temperature   o o o o  humidity   air pressure  4  F to 140  F   20  C to 60  C    10 to 95  RH   700 to 1060 hPa    Weight Monitor   Approximately 10 5 8 oz   300 g  not including batteries  9 Arm cuff   Approximately 6 oz   170 g     Monitor   Approximately 4 1 4     w  x 3 1 8     h  x 5 1 2     I    107 mm x 79 mm x 141 mm        Dimensions Arm cuff   Approximately 5 3 4    x 23 1 2     air tube  29 1 2     145 mm x 594 mm  air tube  750 mm     Cuff circumference 9    to 17     220 to 420 mm   Memory Up to 50 per user  Contents Monitor  arm cuff  battery set  instruction manual  quick start guide  Applied part   Type BF   Internally powered ME equipment  When using only the batteries   Protection aga
20. rt guide    Monitor           AQ  ISITG      XZ  A  Display F  Memory button  B  Up Down buttons G  Battery compartment  C  USER ID selection switch H  Air jack  D  Date Time setting button    AC adapter jack  E  START STOP button  for optional AC adapter   Arm cuff   J  Arm cuff   Arm circumference 9      17     22   42 cm    K  Air plug  L  Air tube       KNOW YOUR DEVICE             Display    ape  E    U    ES  oe i  M  USER ID symbol S  Average value symbol  N  Systolic blood pressure T  Date Time display  O  Diastolic blood pressure U  Movement error symbol  P  Low battery symbol V  Irregular heartbeat symbol  Q  Heartbeat symbol W  Blood pressure level indicator   Flashes during measurement   X  Deflation symbol   R  Memory symbol Y  Pulse display   Memory number    KNOW YOUR DEVICE    Display symbols    Irregular Heartbeat Symbol       When the monitor detects an irregular  rhythm two or more times during the Normal Heartbeat    measurement  the irregular heartbeat Pulse A AAAAA  symbol will appear on the display with the Blood pressure NANNININ  measurement values  l r Heabel  An irregular heartbeat rhythm is defined a Long  as a rhythm that is 25  less or 25  more Pulse   than the average rhythm detected while Blood pressure NA J SU_S    the monitor is measuring the systolic and   diastolic blood pressure     If the irregular heartbeat symbol displays with your measurement results  we  recommend you consult your physician  Follow the directions of your physician    
21. t into an outlet on a circuit different from that to which  the receiver is connected    e Consult the dealer or an experienced radio TV technician for help     31    LIMITED WARRANTY    Your BP742N Automatic Blood Pressure Monitor  excluding the arm cuff and  batteries  is warranted to be free from defects in materials and workmanship  appearing within 2 years from the date of purchase  when used in accordance with  the instructions provided with the monitor  The arm cuff is warranted to be free  from defects in materials and workmanship appearing within 1 year from the date  of purchase when the monitor is used in accordance with the instructions provided  with the monitor  The above warranty extends only to the original retail purchaser     We will  at our option  replace without charge any monitor or arm cuff covered  by the above warranty  Replacement is our only responsibility and your only  remedy under the above warranty     To obtain warranty service contact Customer Service by calling 1 800 634 4350  for the address of the inspection center and the return shipping and handling fee     Enclose the original printed receipt  Include a letter  with your name  address   phone number  and description of the specific problem  Pack the product  carefully to prevent damage in transit  Because of possible loss in transit  we  recommend insuring the product with return receipt requested     THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN  CONNECTION WITH THIS PRODUCT  
22. test Compliance level A A  test level environment   guidance    Floor should be wood  concrete   or ceramic tile  If floors are  covered with synthetic material   the relative humidity should be  at least 30        2 kV for power  2 kV for power Mains power quality should  supply lines supply lines be that of a typical    1 kV for  1 kV for input  commercial and or hospital  input output lines   output lines environment     Electrostatic  discharge  ESD   IEC 61000 4 2     6 kV contact  6 kV contact   8 kV air  8 kV air    Electrical fast  transient burst  IEC 61000 4 4     1 kV line s  to  1 kV line s  to Mains power quality should be  line s  line s  that of a typical commercial   2 kV line s  to earth    2 kV line s  to earth   and or hospital environment      lt 5   Ur   gt 95   dip    lt 5   Ur   gt 95   dip   Mains power quality should  in Ur  in Ur  be that of a typical  for 0 5 cycle for 0 5 cycle commercial and or hospital        i t  If the user of  40   Ur  60   dip  40   Ur  60   dip   E    Vronmen  Voltage dips  short  in Ur  in Ur  the OMRON BE 742N    A  gt  including AC adapter  interruptions and for 5 cycles for 5 cycles requires continued operation  voltage variations on    power supply inputlines 70   Ur  30   dip 70   Ur  30   dip during power mains  IEC 61000 4 11 in Ur  in Ur  interruption  it is  for 25 cycles for 25 cycles recommended that the     gt  OMRON BP742N including   lt 5   Ur  95   dip   lt 5   Ur  95   dip   AC adapter be powered from    in 
23. ts in a clean  safe location    e Do not use any abrasive or volatile cleaners    e Do not wash the device and any components or immerse them  in water    e Do not use gasoline  thinners or similar solvents to clean the  device     ON bs ly e  IAS la sly Le   e                   e Use a soft and dry cloth  or a soft and moistened cloth and  neutral soap to clean on the monitor and the arm cuff    e Changes or modification not approved by the manufacturer  will void the user warranty  Do not disassemble or attempt to  repair the device or components  Consult Customer Service     26    MAINTENANCE AND STORAGE    Storage  1  Unplug the air plug from the air jack           2  Gently fold the air tube into the arm  cuff     Note  Do not bend or crease the air tube  excessively        Do not store the device in the following situations     If the device is wet     e Locations exposed to extreme temperatures  humidity  direct sunlight   dust or corrosive vapors such as bleach     Locations exposed to vibrations  shocks or where it will be at an angle     Optional Medical Accessories    Arm cuff AC Adapter  Arm circumference    9      17     22   42 cm        CD WR17 HEM ADPTW5   Model  HEM RML31     27    MAINTENANCE AND STORAGE    Using the Optional AC Adapter          1  Insert the AC adapter plug 2 ts  into the AC adapter jack on T       the rear side of the monitor   1  2  Plug the AC adapter into an EA  electrical outlet  vs    To disconnect the AC adapter  unplug the AC adapte
24. ture   Avoid bathing  drinking alcohol or  caffeine  smoking  exercising or eating  30 minutes before taking a  measurement       Sit on a chair with your feet flat on the floor    e Sit upright with your back straight    e Sit with your back and arm being  supported      The arm cuff should be placed on your arm  at the same level as your heart        Taking a Measurement   Notes      To stop the measurement  press the START STOP button once to deflate the  arm cuff    e Remain still and quiet while taking a measurement     The monitor is designed to take measurements and store the measurement  values in the memory for 2 people using USER ID 1 and USER ID 2              17    USING THE DEVICE    2  Press the START STOP button     The arm cuff will start to inflate automatically           y START TINELATE X DEFLATE Y COMPETE     nage nage 7300 7390  0   781   Se  3   LI   I H  i    T  t I  lle 8 le 1 X y       If your systolic pressure is more than 210 mmHg  After the arm cuff starts to inflate  press and hold the START STOP  button until the monitor inflates 30 to 40 mmHg higher than your  expected systolic pressure     Notes     The monitor will not inflate above 299 mmHg     A  Inflating to a higher pressure than necessary may result in bruising  where the arm cuff is applied        18    USING THE DEVICE    3  Remove the arm cuff   4  Press the START STOP button to turn the monitor off     The monitor automatically stores the measurement result in its memory   It will aut
25. uff manually until it is  sufficiently  30 to 40 mmHg above your previous  measurement result   Refer to    Taking a Measurement        The arm cuff was  inflated exceeding the   Do not touch the arm cuff and or bend  maximum allowable  the air tube while taking a  pressure  and then measurement  Do not inflate the arm  deflated automatically   cuff more than necessary   when inflating the arm   Refer to    Taking a Measurement      cuff manually   F Repeat measurement  Remain still   Movement during     and do not talk during measurement   measurement  Tali      Refer to    Taking a Measurement      Clothing is interfering Remove any clothing interfering with  with the arm cuff eames      Refer to    Applying the Arm Cuff        Device error  Contact Customer Service        24    ERROR MESSAGES AND TROUBLESHOOTING    Troubleshooting    Replace all batteries with new ones   Check the battery installation for  proper placement of the battery  polarities    Refer to    Battery Installation        No power   No display appears on the monitor     Blood pressure varies constantly   Many factors including stress  time of  day  and how you wrap the cuff  may  affect your blood pressure  Review the  sections    Before Taking a  Measurement    and    Taking a  Measurement        Measurement values appear too high  or too low        25    MAINTENANCE AND STORAGE       Maintenance   To protect your device from damage  please observe the   following      Store the device and the componen
26. y the approved arm cuff for this device  Use of other arm cuffs may  result in incorrect measurement results    A Do not use a mobile phone or other devices that emit electromagnetic fields   near the device  This may result in incorrect operation of the device    A Do not disassemble the monitor or arm cuff  This may cause an inaccurate  reading    A Do not use in a location with moisture  or a location where water may splash  on the device  This may damage the device    A Do not use the device in a moving vehicle  car  airplane     A  Read    If your systolic pressure is more than 210 mmHg    of this instruction  manual  if your systolic pressure is known to be more than 210 mmHg   Inflating to a higher pressure than necessary may result in bruising where  the arm cuff is applied     IMPORTANT SAFETY INFORMATION     AC Adapter  optional  Usage   A Do not use the AC adapter if the device or the power cord is damaged  Turn  off the power and unplug the power cord immediately     A Plug the AC adapter into the appropriate voltage outlet  Do not use in a  multi outlet plug     A Never plug in or unplug the power cord from the electric outlet with wet  hands     A  Fully insert the power plug into the outlet     A When disconnecting the power plug from the outlet  do not pull the power  cord  Be sure to pull from the power plug safely     A When handling the power cord  take care not to do the following     Do not damage  Do not break it    Do not tamper with it  Do not forcibly
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
FP Iss 4 template dual power un  RSTC1000 HMI/PLC Design Guide WEB CONTROL  Jardins de particuliers  Notice d`information salariés OCIANE SANTÉ  Gamme HYGIELAND - Hygiena Boutique  Acumulador compacto  please read and understand this introduction before using  SP DP PR PP SP DP PR PP SP DP SP DP PR PP AVERAGES  Audi。C。mm イヤホン巻き取りポケットフンオを お買い上げ  Halo H273RICAT1D Instructions / Assembly    Copyright © All rights reserved. 