ilShin DF 8517S vriezer manual ENG
Contents
1. JoyeJode 3 45 Principal of Cascade Refrigeration System The cascade refrigeration system is designed to use different refrigerants which have different boiling points in one system in order to reach ultra low temperatures 60 down to 150 range Therefore a cascade cooling system needs two sepeate cooling systems compressors Each cooling system compressor works simultaneously The first stage Compressor exists to condense the high pressure line of the second stage compressor The second stage comprressor is the main system which maintains the ultra low temperature of the freezer The refrigerant of the first stage compressor is compressed This compressed refrigerant is cooled by the air cooled condenser The condensed refrigerant is dried by a drier The dried refrigerant evaporates in the evaporator by a capillary tube low pressure low temperature The heat exchanger is cooled down to less than 40 C This heat exchanger is now used to condensate the compressed refrigerant coming from the second stage compressor The strainer removes dirt from the evaporated refrigerant Because we use a liquid separator liquid refrigerant is prevented from sucking into the compressor The second stage refrigerant is compressed in the second stage compressor The compressed refrigerant is pre cooled by passing through a pre cooler After passing the oil separator and drier it is condensed in the heat exchanger cooled to 40 C by
2. Volt free contacts The main power switch is located at the back of the freezer After power supply is checked and the unit is earthed open de condenser door By switching the main power switch in ON position the Freezer will use the power and on this time is ready for operation If the freezer is not in use please put the main switch in OFF position 44 Door Alarm The door alarm can be used in combination with the CO2 backup system It is a safety feature to make sure the operator does not get into contact with liquid CO2 LN2 in case of opening the door of the freezer Always connect the LN2 backup system to the door alarm connection at the back of the freezer For safety measure resistance on the contacts with Ohm metre 1 Do contact 1 amp 2 show resistance Connection should be a closed loop normal conditions closed loop door open conditions open loop 2 Open door lid 3 Do contact 1 amp 2 show unlimited resistance Connection should be a open loop 4 5 Remote Alarm The remote alarm is a volt free dry contact which will alarm in case of 1 High temperature alarm 2 Low temperature alarm 3 Power failure Male and female 4 Clean filter alarm alarm plug 1019 COM NO COM NO NC NC Alarm Conditions Normal Conditions In normal conditions relay is always closed which means connection between COM and NC normally closed In case of alarm relay is open and means connection between COM and NO
3. DF358517 3 595x620x18 745x620x18 745 728 18 5 5 Index of Control Panel and Functions Temperature Display 2 Scan Alarm indicator 1 Key Switch 3 Cursor 4 Numeric Indication 6 Voltage 5 Alarm Leds TA BUTTON CONTROLS KEY SWITCH During OFF all functions are stopped During SET all functions are working and it is possible to adjust the temperature During ON the unit is operating as planned and it is not possible to adjust the temperature Function blocking system If you take out the key during ON all functions will be blocked Please keep the key well for the planning of functions This button is to select each temperature mode Only in the case that the KEY SWITCH is set to SET CURSOR This button is to select a position of numerical value NUMERIC This button is to change a numerical value 5 ALARM If the inside temperature of the unit exceeds the alarming temperature range an error message will be shown on the temperature indicator The lamp will be on and the alarm signal will sound Push the alarm button mute then the alarm signal does not sound VOLTAGE This button indicates the voltage Push the button then the voltage will be shown on the Temperature indicator 5 6 How to Operate 1 Turn the KEY SWITCH to the right and set it to SET 2 All functions will be operational The PRESENT led will light and the current inside temperature of the Chamber will be shown on the
4. normally open 4 6 RS232C Port The RS232C Port can be used to monitor temperature and alarms on your PC by using the ilShin software and a RS232 cable This software can monitor Actual Temperature Temperature Setpoint High Temperature alarm setpoint Low Temperature alarm setpoint High alarm Low alarm EE LM CE These parameters can be logged also on your computer More information about the Software installation is described in paragraph 6 2 The diagram below illustrates the wiring between the RS 232C port of the computer and the Deep freezer Rx ze E T A E 3 Com o d a 5232 P1 P2 Freezer Control Board Connection Cable Assembly 100m max 5 Index of parts and functions 5 1 Index of the Parts Chest Model sensor inlet Handle Control panel back aW Optional recorder Locking device Condenser u door wu Cui 5 2 Index of the Parts Up Right Model CO back up inlet Eme on top op unit Control panel Locking device Z Sensor inlet on side of unit Optional recorder Condenser Castor door Ec L 5 3 Index of Inner Sublids for Chest Models Gasket Inner sublids TT n Aa M 4 m eT Ax2 Ax2 Ax2 Ax2 5 4 Index of Inner Compartment Shelves for Upright Models Backup Tube Insert Hole Validation Port 595x470x18 3 595x470x18
5. Hermetically sealed cooling system with very durable compressors Cold air loss is minimized by insulated inner doors Efficient evaporator with maximum contact with walls for fast cooling down 130 mm polyurethane foamed insulation Heavy duty casters with built in adjustable stands Cabinet with round corners for easy cleaning 3 Built in access ports Very user friendly control panel at eye level Easy to clean condenser filter Pressure equalization port allows easy access to cabinet after door closing Quiet running fans and compressors create a very low noise level lt 50 dBa Heated door seals hot gas prevent ice to build up Theory Behind the Perfect Fit System High pressure insulated PU foam Door seal 1 Hot gas line Door seal 2 Plastic strip Seal 3 Door with control panel Insulated inner doors High pressure insulated PU foam The hot gas line prevents building up at the sides of the door and cabinet The plastic strip prevents direct contact between the metal cabinet housing and the door air Direct cold conduction ice is prevented Insulated inner doors prevent cold loss from the individual compartments The unique 3 door seal system ensures a tight door closing with as result a minimum of ice 2 2 Chest Model Specifications 185 45 20 40 230 1 2 211 86 60 80 50 60 1 2x2ea 205 45 20 40 230 1 2 236 86 60 80 50 60 1 2x2ea 230 45 20
6. Q15 19 D16 21 LD1 5 RL2 3 R43 56 CN1 4 5 AC POWER INPUT 220V AC DS2Y S DC12V 200mwvy 660R c3 4700uF2 T P12 1N4004 gt gt CN2 PIN7 4700uF T cs 104 Ss ees Ee i V ALL PART Ui 18 Li 2 Q1 19 TNR1 2 01 21 Fi 2 PCi 6 LD1 5 21 VR1 4 51 R1 65 RL1 3 RA1 2 C1 40 CN1 1 5 CF1 19 CN2 3 U1 TP1 P5 Gu L1 100uH 1A DIGITAL TOR C7 T wm ANALOG gt 5 2 C8 09 C10 C11 Cn oy C12 10uF 16V 100uFKAL05 5V ais ICL7660CPA C13 010 C14 1OuF 6V 1N5817 10uFM6V N DIGITAL GND ANALOG GND DEEP FREEZER CONTROLLER POWER SUPPLY PART MAIN BOARD mw lxi ENABLE WE deje quc 1 0 1 pes rej RB5S INT 5 PR V 4 3 ar eS Sa _ Jen A ve Aman RCO SCAN mr Mi SET POINT RC CURSOR a mt COLD ALARM RBO NUMERIC a nta WARM ALARM _ RC3 ALARM RESET ALARM RB5 INT SET LM LIGHT RC4 LIGHTON nu Ma REMOTE POWER RC5 REMOTE POWER _ Tn VOLTAGE BOOSTER RB4 DISPLAY SEL Ma DEFROST TEMP LINE VOLT j RL1 SET ENABLE PRESENT FND1 LD8 16 Size Document Number ISE 120D 2 i ill l SRE I Tce 2 lI EER ee EE eee ii Z Z Z ee DEEP FREEZER MAIN BOARD ISE 120D g 0 12 0
7. condenser filter is cleared of dust every month b We recommend that ice is removed from the door seals every month c We recommend a maintenance visit by your local dealer on an annual basis We wish you a quick and trouble free installation IMPORTANT WARRANTY AND ROOM TEMPERATURE This unit is equipped with the WarmMark indicator near serial tag 1 In case this WarmMark is removed from the product warranty will be terminated 2 In case the room temperature climbs above 33 for a long period of time a red colour gradually moves through the window of the WarmMark Indicator Every window which is indicated red will decrease warranty period with 1 year Contents Quick Installation Guide 1 Dimensions of the freezer models 1 1 Chest Model 1 2 Upright Model 2 Specifications 2 2 Chest Model Specifications 2 3 Upright Model Specifications 2 4 Characteristics of Temperature DF85 Series 2 5 Characteristics of Temperature DF90 Series 3 Installation 3 1 Installation Place 3 2 How to Install 4 Power supply power switch amp external connections 4 1 Power Supply 4 2 How to Earth 4 3 Main Power Switch 4 4 Door Alarm 4 5 Remote Alarm 4 6 RS232C Port 5 Index of Parts and Functions 5 1 Index of the Parts and Functions Chest Model 5 2 Index of the Parts and Functions Upright Model 5 3 Index of Inner Sublids for Chest Models 5 4 Index of Inner Compartments Shelves for Upright Models 5 5 In
8. description Freezers DF9014 Manufacturer ilShin Lab Co Ltd LZ A M gt ilShin Europe B V Krypyonstraat 11 6718 WR Ede The Netherlands Mr J W van Rheenen Researcher Consultants Europe bv Binnenveer 2 JUN ddress 1381 BT Weesp 762 pcode and place 1381 BT Weesp the Netherlands Harmonisation with the Safety regulations e Directive 98 37 EC Lh 4 igh E NEN EN 1050 NEN EN 292 1 r Msi S lt NEN EN 1050 NEN EN 292 1 ll NN NEN EN 292 2 EN 60204 NAK A us NEN EN 292 2 NEN EN 60204 Z NEN EN 50081 2 NEN EN 50082 2 x Ed s NEN EN 50081 2 NEN EN 50082 2 SA 20 za Product description A n NG Product description y gt Freezers A Product name Freezers DF9020 t PF15007 Manufacturer ilShin Lab Co Ltd Ibamri 82 6718 WR Ede The Netherlands Mr J W van Rheenen Researcher 1381 BT Weesp the Netherlands Harmonisation s with the Safety regulations NEN EN 1050 NEN EN292 I X INTE NEN EN 292 2 NEN EN 60204 2 W NEN EN 292 2 NEN EN 60204 NEN EN 50081 2 NEN EN 50082 2 S ie NEN EN 50081 2 NEN EN 50082 2 54 iIShin Evrope Corporate identity of ilShin TRUST ilShin is a combination of words originating from the traditional Korean language The first part of the word il means ONE The second part of the word Shin means TRUST This explains our main value of the company ilShi
9. i gt lt H D UJ 2 2 gt KEY S W we 3 gt 7 E k rmp Su E L E E x 23 3 lt 0 6 go E 53 1 2 3 4 5 6 ig 8 9 10 11 12 13 14 15 16 17 18 19 20 21 ma 220V AC AC POWER SOURCE PB 220V z PC SERIAL PORT WIRING n x 25 PINS os lt jPIN14 6 7 59 ec FREEZER CONTROLLER BOARD WIRING 7 mbe ISE 120D 42 7 3 Refrigerational Trouble Shooting 7 3 1 Compressor not Running Repairing the compressor not in working over one day Necessary tools Manifolds Refrigerant R408A R409A R290 R23 Welding machine R _ R408A Below Normal t High pre 60 80 psi Soak Sight R409A Low pre 60 80 psi Leak Detector Welding on the Manifolds leaked part Vacuum pump 2nd pressure Below e R290 Low pre 70 80 psi Soak Sight Yes Leak Detector nn E Welding on the leaked gt Manifolds check pressure part Vacuum pump Yes High normal High normal Clean condenser High pre 1307150 psi High pre 140 160 psi Refill R409A little Low pre 5 5 psi Low pre 5 5 psi Yes Below Normal pressure Clean condenser If ambient 20C chamber temp 80C Refill R408A little 1st high 130 150 psi lo
10. is open for a long time or the amount of the test sample is too big Please check whether the door is open When you put the test sample please do it in small quantity and in a short period 33 7 2 2 Electrical Trouble Shooting Yes Yes No Yes Yes Yes Yes Yes Yes 34 Yes Yes Yes Yes Yes 35 RS 232C 2 PCB120B TXD PCB 3 PCB1208 RXD PCB H1 Ze 1 2 f rio 3 4 Bc 5 7 PT100 gl 8 12 9 REF1 10 REF2 11 Voltage Booster 12 124 13 15 CH4 349423 1 2 1 gel 8 2 3 25 AS 5 7 MAIN 513 8 9 sl 19 12 T C ral 2 Batt Sensor PT 100 K Alarm Sensor Condensation Temp SAAT Housing 4P Ref 1 TROut Put Ref 2 TROutPut 24V Vol tage Booster TR Out Defrost ALARM Out ALARM ALARM RELAY A RELAY B TXD GND RELAY com i 1st compress START CAP EN
11. 15 CF16 CF17 CF18 CF19 104 104 104 104 104 104 104 104 104 gt U3 8 Qi D13 14 VR1 4 R4 27 C15 28 CF1 10 CN1 3 1 bridge ref 2 A D ref V REF 2 5V V REF 2 5V R26 T 1 2K ie P5 2 2 000V B 2K CT6P R25 2 250K i V 1 C 1040 R27 10mV 1 TP4 ce C26 d 104 FILM 3 NA R24 C24 D p5 3 5 104 u CPU PORT RTD 9 7 9 NS 10 BR 782C 3P H temp L line voltag DISP SELLECT U1S P19 RMS TO D C A C INPUT TRANS 1 5V TAP gt 5 Y A C INPUT TRANS 1 5V TAP gt 1 1V 220V ior gt al m C22 c23 10 16 104 m x SE eT DEEP FREEZER CONTROLLER ANALOG PAR ANALOG GND Size Document Number a j N B ISE 120D m gt DIGITAL GND 8 5 A eee eee ee eee Pee be Tee GE ey Ss aa Pa Pees 38 VOLTAGE BOOSTER gt R53 i U I AS CEN DAS O A Ae Fi CLEAN lt PC1 TLP521 gt R44 330R FREEZER 1 gt R47 330R FREEZER 2 gt gt A R50 330R 330R gt DEF ROST E 3 D16 1N4148 1959 016 C1959 Q17 C1959 Q18 C1959 Q19 C1959 R43 2 2K P24 R46 390R 3W ed dr LP RL2 UVD110P DC24V UVD110P DC24V 4 24V COMM CLEAN BR 762C 6P DEF ROST 3 ALARM 20 gt GND 24V BR 762C 8P Document Number ISE 120D FREEZE FREEZE2 R49 15 VOLTAGE BOOSTER mem 24V PC1 6
12. 240 boxes 558x145x283 DF8524 20 racks 5 deep 5 high Part UR0070 5 deep 3 high Part UR0071 500 boxes 300 boxes 692x145x283 145 x 136 x 51 145 x 136 x 88 6 2 Aluminum Storage Racks for Chest Freezers Al Amount of racks view from above Type of racks for 2 inch boxes Type of racks for 3 inch boxes Max of 2 inch Max of 3 inch External Rack Dimensions mm D x W x H Internal Rack Dimensions mm D x W x H DF9007 12 racks 12 high Part CROO62 8 high Part CROO63 144 boxes 96 boxes 146x146x660 For 2 inch boxes each opening For 3 inch boxes each opening DF9010 12 high Part CROO62 8 high Part CRO063 216 boxes 144 boxes 146x146x660 DF9014 12 high Part CROO62 8 high Part CROO63 288 boxes 192 boxes 146x146x660 Storage Racks for Chest Freezers DF9017 30 racks 12 high Part CROO62 8 high Part CRO063 360 boxes 240 boxes 146x146x660 145 x 138 x 51 145 x 138 x 78 DF9020 ilShin E rope 36 racks 12 high Part CROO62 8 high Part CRO063 432 boxes 288 boxes 146x146x660 6 3 Fiberboard Storage Boxes and Dividers Easy to remove cover Thick Plastified Cardbox Material External dimensions External dimensions of 2 inch boxes model PP0060 133x133x49 mm DxWxH External dimensions of inch boxes model PP0061 133x133x77 mm DxWxH
13. 40 50 230 1 2 261 86 60 80 50 60 1 2x2ea 255 45 20 4 500 230 1 2 286 86 60 80 50 60 1 2x2ea 285 45 20 40 59g 230 1 593 311 86 60 80 50 60 1x2ea 2 3 U pright Model Specifications 200 230 50 60 1 2x2ea 200 230 50 60 1 2 1 2x2ea 484 305 350 380 200 230 50 60 1 2 1 2x2ea 380 410 200 230 50 60 1 2 General Remarks Usage Temperature We recommend to run the ilShin Freezers DF85 and DF90 Series with a setpoint of 200 230 50 60 Foamed In place Poly urethane Foamed In place Poly urethane Textured steel or stainless steel optional Textured steel or Stainless steel optional 80 C Our freezers are produced tested and calibrated on this temperature setpoint With this setpoint the freezer is performing in most optimal conditions and guarantees the longest lifetime for compressors In case of using set points with colder temperature will decreases compressor lifetime and does not have any affect on better conditions for your samples For DF35 and DF45 Series we recommend a setpoint of 40 C 2 4 Characteristics of Temperature DF85 Series Pull down curve for DF8514 for testing cooling speed Model DF amp 514 Temperature C w Temperature Testing environment Powered by Ambient temperature 258C 2 9C Sal saa Wirele
14. Deep Freezer Series User Manual version 3 0 07 08 07 ilShin Europe B Kryptonstraat 11 l h r 6718 WR EDE The Netherlands uro pe info ilshineurope com www ilshineurope com tel 31 318 64 64 81 fax 31 318 64 60 34 Failure to read understand and follow the instructions in this manual may result in damage to the unit injury to operating personnel and poor equipment performance Quick Installation Guide 1 Recommended room conditions a Recommended room temperature between 5 C and 28 C If the room temperature is normally gt 30 C please install suitable air conditioning b Make sure the floor is level and even c Make sure the condenser grid is not blocked adequate air circulation is required d Install the two spacer pins on the back of the freezer to allow heat dispersion behind the freezer fig 1 2 Turning Freezer on gt always start up the freezer empty a Push the power plug firmly into the power socket please check you have the correct voltage b Make sure the mains power switch at the back of the freezer is switched ON Switched upwards fig 2 Fig c Turn the key switch on control panel to ON position 3 Recommended set points a set point 80 C b cold alarm 90 C c warm alarm 70 Fig 2 These set points are factory set and gives the freezer the optimum conditions for prolonged life 4 Recommended maintenance a We recommend the
15. Internal Dimensions Internal dimensions of 2 inch boxes model PP0060 124x124x45 mm DxWxH Internal dimensions of inch boxes model PP0060 124x124x70 mm DxWxH Possible dividers Dividers are also made of plastified cardbox Height of all dividers is 35 mm Model Description PP1049 Storage box divider 7 x 7 PP1064 Storage box divider 8 x 8 PP1081 Storage box divider 9 x 9 PP1100 Storage box divider 10 x 10 21 6 4 CQ Back Up System optional 6 4 1 Introduction of Backup System model LC0070 A Backup System ensures an extra safe environment for your samples even in case the Freezer has a long term power failure or in case the freezer has a technical problem The backup system in combination with a liquid bottle cylinder can remain a temperature around 70 C for several hours For example for reference only bottle cylinder with 50 litres Ambient of 20 C Freezer capacity of 282 litres 1 2 loaded with products Temperature inside the cabinet will be stabile on 70 C for approximately 18 hours 6 4 2 Index of the parts Conia System s i gt gt P a P d liquid Coe man ir E EM liquid T c A k qa L 22 Check if following items are present In case the system is ordered together with an ilShin Freezer following parts are already fitted to the freezer Part b part e
16. L E 2 1ST RELAY e 9 e e 2nd RELAY S S R 15 251 compress RUN CAP START CAP 7 2 3 Electrical Schematic DF85 and DF90 N 7 N N j Circuit SK Breaker 0 220235 MAGNET CONTACT 2a2b eee ranisfonner TERMNAL BLOCK Electric Box Door S W PCB pin amp x o 3 36 m DISPLAY amp KEY BOARD a2 s KRC101M I i i T 3 VSS cou a 5 P5 1 21 104 z R32 T4HCO4 74HC04 51R eau ex BUZZER FREEZER 1 WO PART Fi 2 5 1 lt REMOTE POWER 621 DIP SWITCH SET 1 INTERNAL SET ON SET 2 DISPLAY 0 1 C ON USE MES os p deu e 3 VOLTAGE BOOSTER ON USE 4 LINE VOLTAGE OFF 220V PC SERIAL PORT 25P PING PIN20 CN3 Ui4 US Uie U17 CF11 CF12 CF13 CF14 CF
17. L452B cascade system Approx 10psi 3 5 bar suction 1 2 HP L262B single stage pressure system suction 1 HP L452B single stage system 2 x Ve HP L262B cascade system pressure Opsi suction 0 psi suction 155psi pressure 2 x 1 HP L452B cascade 1psi system suction 47 7 4 Registration of Service Before you register for the service please read the following information 1 The unit does not operate at all O Please check the power supply 230V plug connection O Please check whether the temperature is set in a right way O Please check whether the unit is too close to a heating object O Please check whether the rear side of the unit is too close to the wall O Please check whether the condenser is blocked by dust Remedy wipe off the dust by soft brush O Please check whether a big amount of test sample is inserted within a few hours Remedy this is a temporal phenomenon please just wait 2 Loud running noise O Please check whether the floor is even O Please check whether the noise is coming from the compressor The noise of the compressor is bigger than that of household refrigerators O Do you hear a turbo sound This unit is equipped with a fan motor which is used for air cooling system when the fan motor is running you can hear some noise O Do you hear a knocking sound This sound originates from the parts of the unit when they are expa
18. POSITION 2 Re attach the chart hub knob and screw securely by hand against the chart Press and hold the change chart button 3 again for approximately one 1 second until the pen begins to move back onto the chart and then release the button Check to make sure that the pen is marking on the chart paper If it is not then carefully adjust the pen arm to establish contact with the paper 6 6 2 Replacement of pen Recorders that are equipped with fibre tipped cartridge pens will have a cartridge that is colour coded red to designate pen number one 1 The pen cartridge is securely fastened to the metal pen arm using a special U clip tab For ease of replacement it is suggested that the two 2 screws that hold the pen arm be loosened and the pen cartridge and metal pen arm be removed as an assembly Refer to the following figure next page for the location of the pen arm screws 28 BATTERY PEN ARM SCREWS STRAP BATTERY HOLDER 9 VOLT BATTERY PEN CAP PEN CARTRIDGE i m 0 0 rp POSITION 27 Unsnap the plastic U clip tab of the pen cartridge from the metal pen arm remove and discard the old pen cartridge Replace the new cartridge by opening the hinge and snapping it securely around the metal pen arm Refer to the figure above 6 6 3 Pen arm calibration To check and or adjust the recording pen calibration to the outer most temperature graduation of the
19. R 231003 25 27 A Product description Freezers NEN EN 1050 NEN EN 292 1 NEN EN 292 2 NEN EN 60204 NEN 3008122 NEN EN 50082 2 EC Declaration of Conformity NR 231003 35 Product description Freezers DF4514 Manufacturer ilShin Lab Co Ltd 6718 WR Ede The Netherlands Mr J W van Rheenen Researcher Consultants Europe bv Binnenveer 2 1381 BT Weesp NEN EN 1050 NEN EN 292 1 NEN EN 292 2 NEN EN 60204 NEN EN 50081 2 NEN EN 50082 2 Product description N 2 Ss ter IEA BS Freezers DF4510 Manufacturer ilShin Lab Co Ltd Thamri 82 6718 WR Ede The Netherlands Mr J W van Rheenen Researcher Consultants Europe bv Binnenveer 2 1381 BT Weesp the Netherlands NEN EN 1050 NEN EN 292 1 NEN EN 292 2 NEN EN 60204 NEN EN 50081 2 NEN EN 50082 2 lt J EC Declaration of Conformity NR 231003 45 Product description Freezers DF4517 Manufacturer ilShin Lab Co Ltd 1 82 Yangjukun Kyongkido Iu N maith Korea AN Wy NEN EN 1050 NEN EN 292 1 NEN EN 292 2 NEN EN 60204 ISO 9001 2000 4 NEN 50081 2 NEN EN 50082 2 IN BY x lt EC Declaration of Conformity NR 231003 55 Product description Freezers DF4520 Manufacturer ilShin Lab Co Ltd 6718 WR Ede The Netherlands Mr J W van Rheenen Researcher Consultants Europe by Binnenveer 2 Binnenveer 2 Y 1381 BT Wees
20. Warm alarm Cold alarm 6 5 3 System Setup Minimum requirements 486 PC with W98 installed Copy the file ilshin exe in a folder of your own choice and start the program by double clicking on the ilshin icon Once the program is started it checks for a connected freezer in operation on either COM1 or COM2 If no freezer is found the program enters a simulation mode In case of normal operation a configuration file with user settings is loaded and the values which are received from the freezer are displayed If no configuration file is found default user settings are loaded When during normal operation the freezer is disconnected or turned off the text Freezer Off is displayed until the normal operation is restored 26 6 5 4 User Setup Via the menu Control gt User Settings a Dialog Box is opened with the following information Userld UserName for instance the location or contents of the freezer Manufacturer TypeNumber Serial Number Logging Interval Pathname to folder on harddisk where the logfile is stored For instance c or cNlshin etc 5 NO ending backslash This information is effected when the button OK is checked When Close is checked without checking OK before the user settings remain unchanged The user settings are stored in the configuration file on the harddisk when the program terminates 6 5 5 Temperature Logging Logging is switched on or off by checking LogOn or LogOff i
21. chart press and hold the change chart button 3 until the pen begins to move off of the chart Once the pen has moved off of the chart again press and hold the change chart button 3 until the pen begins to move back onto the chart The pen should briefly stop at the outer most temperature graduation of the chart before continuing onto the chart to begin recording If the pen does not stop exactly at this location on the chart it can be adjusted or calibrated by using the left 1 or right 2 arrow buttons When the pen moves back onto the chart and briefly stops you will have approximately five 5 seconds in which to adjust the pen s position using the left and right arrow buttons Each time the chart paper or fibre tip pen cartridge is changed you should make sure that each pen stops at the outer most temperature graduation of the chart paper Otherwise this pen offset will cause the unit to record an incorrect temperature on the chart 29 6 6 4 Temperature recorder calibration check This recorder has been accurately calibrated at the factory Before making any adjustments this instrument should be in service for 24 hours Thereafter if any adjustment is required perform the following procedure 1 Place a Certified Test Thermometer s in a solution bottle s alongside the recorder s sensor probe s 2 Once the temperature has levelled out compare the position of the pen on the recorder to the test thermometer s
22. dex of Control Panel and Functions 5 6 How to Operate 6 Accessories 6 1 Aluminium Storage Racks for Upright Freezers 6 2 Aluminium Storage Racks for Chest Freezers 6 3 Fiberboard Storage Boxes and Dividers 6 4 Back Up System optional 6 4 1 Introduction of Backup System model LC0070 6 4 2 Index of the Parts 6 4 3 How to Install the Backup System 6 4 4 How to set the Temperature of the Backup System 6 4 5 Backup System Safety Remarks 6 5 Data Acquisition System Optional 6 5 1 Hardware Setup CO DOIN O N N N PN PNP PNP PNP P PO VP a o 2 oL ol ol ol o ell O O OI OI PB DIA DAN O alta BPW YN 6 5 2 Alarms 6 5 3 System Setup 6 5 4 User Setup 6 5 5 Temperature Logging 6 6 Chart Recorder Optional 6 6 1 Changing the chart paper 6 6 2 Replacement of pen 6 6 3 Pen arm calibration 6 6 4 Temperature recorder calibration check 6 6 5 Battery backup 6 6 6 Battery location and replacement 6 6 7 Chart range selection 7 Service 7 1 General Troubleshooting 7 2 Electrical Trouble Shooting 7 2 1 PCB ISE 120C ERROR CODE 7 2 2 Electrical Trouble Shooting 7 2 3 Electrical Schematic DF85 and DF90 7 3 Refrigeration Trouble Shooting 7 3 1 Compressor not running 7 3 2 Compressor is running but wrong temperature 7 3 3 Refrigeration drawin
23. display 3 Push the SCAN button once to check the current setpoint he SET POINT led will light To adjust SET POINT use the CURSOR button and NUMERIC button 4 Push the SCAN button once more to check low alarm setpoint he COLD ALARM led will light To adjust the COLD ALARM use the CURSOR button and NUMERIC button The COLD ALARM should be set lower than the SET POINT temperature e g In the case that the SET POINT is 70 C the COLD ALARM is 75 5 Push the SCAN button once more to check the high alarm setpoint gt The WARM ALARM led will light To adjust the WARM ALARM use the CURSOR button and NUMERIC button WARM ALARM should be set higher than the SET POINT temperature e g In the case that the SET POINT is 70 C the WARM ALARM is 65 6 Push the SCAN button once more to turn on the PRESENT lamp showing the current inside temperature of the deep freezer again 7 Now everything is set Turn the KEY SWITCH to the right once more to ON Take out the KEY by pulling it to the front Please keep the KEY carefully in order to change temperature setpointsin the future As long as the key is taken out it is impossible to change setpoints It prevents changing setpoints byunauthorised people or by mistake In this way thesamples storedare protected 8 To check the voltage push the VOLT button gt he current incoming voltage will be shown Standard every ilShin is equiped with vollage co
24. emperature of the CO back up system How to set temperature Remarks Push 1 SET SWITCH In case the temperature goes up because of a power failure or technical To change the set temperature hold the the problem the temperature set to the button 1 SET SWITCH and use a small BACK UP SYSTEM will be observed screw dryer to turn the 3 TEMP ADJ to the Thus the damage on the sample in the left and right to set the temperature unit can be limited to a minimum Release the SET SWITCH at the chosen temperature The PRESENT led will ight and the current temperature of the chamber will be shown The temperature is standard factory set to 60 C 6 4 5 CO Backup System Safety Remarks The BATTERY SWITCH should be OFF during moving storage or POWER OFF When the BATTERY SWITCH is on for a long time with power OFF the battery will be discharged and should be replaced When you replace the used CO cylinder please use the CO cylinder with Siphon Tube not general cylinder When the door is open during injection of CO2 the injection should be stopped automatically to protect the user For safety for our Upright Freezers we recommend to use our backup system always in combination with our heated equalization valve to prevent over pressure in the cabinet To check this freezer has a heated pressure equalization valve please check inside of the door for the pictured valve For safety for our Ch
25. en very cold samples Liquid Nitrogen are placed into the freezer HI T Warm Alarm If the temperature is higher than the Warm Alarm Setpoint This can happen when very warm samples or many new samples are placed into the freezer at once When you put in test samples please do it in small quantity and in a short period The HI T alarm also sounds in case the door of the freezer was open for some time Also please check the ambient temperature When the ambient temperature is above 30 or the ventilation of the condenser is bad Please keep the room temperature lower by ventilation or clean the dust in the condenser Please make sure the Freezer is clear from all sides of the wall LO V Low Voltage If the input voltage is lower than 205V Please don t use other electrical appliances and take some actions to boost the voltage Maybe the line voltage is OK and the main board needs calibration of voltage CLEAN Clean Filter alarm When the ambient temperature is above 30 C or the ventilation of the condenser is bad Please keep the room temperature lower by ventilation or clean the dust in the condenser Please make sure the Freezer is clear from all sides of the wall BATE Operation condition of charging battery In the case that there is a power failure If the power failure takes long please use CO Backup oystem or move the test sample to a safe place LOAD Abrupt temperature increasing In the case that the door of the unit
26. est Freezers we recommend to use our CO backup system always in combination with a basic equalization valve to prevent over pressure To check this freezer has a basic equalization valve please check backside of the chest for the pictured valve For safety for our Upright Freezers we recommend to use our 1004C Large Door Handle with Integrated Lock to prevent pressure breaking off standard door handle 25 6 5 Data Acquisition System Optional This Software dispays a monitoring panel which shows _ Present Temperature SetPoint Se Warm Alarm Limit TA k Cold Alarm Limit L _ N TL 6 5 1 Hardware setup 1 Connect the 9 Pins connector of RS232 cable to the COM1 or COM2 2 Connect the 9 Pins of RS232 cable to the Jack of rear side of the freezer more technical details about the wiring and location of the RS232C Port is described in paragraph 4 5 RS232C port 6 5 2 Alarms Exceeding of alarm limits is shown by a blinking display window of the alarm in question An audible alarm is given as well The alarm condition can be acknowledged with a left mouse click on the blinking window or by hitting the spacebar The window will stay red Warm Alarm or blue Cold Alarm as long as the alarm condition exists The window will keep blinking until the alarm is acknowledged even when the temperature is returned within normal limits In this case the blinking colour will be light grey
27. g and principal of cascaderefrigeration system 7 3 4 Charging Quantity and Type of Refrigerants 7 3 4 1 Refrigerant Quantity for Chest Models 7 3 4 2 Refrigerant Quantity for Upright Models 7 4 Registration of Service 8 Warranty and contact information 8 1 International Warranty 8 2 Conditions and terms of sale 8 3 Contact information 9 EC Declaration of Conformity 26 26 27 27 28 28 28 29 30 30 30 31 32 32 33 33 34 36 43 43 44 45 47 47 47 48 49 49 49 49 50 1 Dimensions of the freezer models 1 1 Chest Model Exterior Dimensions Interior Dimensions Model No Exterior Dimensions nterior Dimensions Capacity L WxDxH WxDxH _ RECON NUN GE Sm _ m _ The operating temperature of Model DF45 is 20 C to 45 C and for the Model DF90 40 C to 86 In addition if the last two numbers of the model are the same the same operating temperature is applicable 1 2 Upright Model WxDxH es EZ E SER The operating temperature of Model DF35 is 20 C to 45 C and for the Model DF85 this is 50 C to 86 In addition if the last two numbers of the model are the same the same operating temperature is applicable 2 Specifications 2 1 O 11 12 ey ae Deep Freezer Charicteristics
28. gh temperature alarm and damage to the compressor system 4 In case the condenser door is Please always make sure the condenser blocked by an object door is kept clear of any object his unit uses a condenser which is air cooled In case the air circulation intake is blocked by an object the compressor is not cooled effectively and this can cause high temperature alarm and damage to the compressor system 5 In case the freezer is installed Please always make sure the freezer is directly to the wall always at least 15 cm free from all sides his unit uses a condenser which is air cooled Warm air should have change to circulate In case the unit is too close to a wall this freedom is blocked This will cause the compressor is not cooled effectively and This can cause high temperature alarm and damage to the compressor system 6 Under heavy humidy circumstances Ventilate the space properly by using a It can cause a technical malfunction in the ventilator or by opening the windows electrical applications and short circuit An application which can reduce moisture will be helpful 32 7 2 Electrical Trouble Shooting 7 2 1 PCB ISE 120C D ERROR CODE aaa Meaning Situation Line Voltage is low BATE LOAD Rise Temperature Over 15 C Fast F S Over F S Over F S Over F S Over Pt Sensor Short Error indication LO T Cold Alarm If the temperature is lower than the Cold Alarm Setpoint This can happen wh
29. he chart and the selected range will be saved into the recorder s permanent memory Note The correct range for your purchased ilShin is factory set Changing of chart range is not necessary 31 7 Service 7 1 General Troubleshooting 1 When the room temperature is above If the room temperature around the unit is 30 C 30 C or warmer please open the condenser In case the room temperature is door and the window s of the room to above 30 C for longer time the optimize the ventilation and cooling Please condenser cannot cooled optimal This can note you need an air conditioned room in cause high temperature alarm and damage these circumstances to the compressor system 2 When the room temperature is below When room temperature is below 0 0 C heat the area room and keep the The oil in the compressor will be temperature minimal 5 C condensed It can have negative effect on the operation of the compressor Some electrical parts will not operate optimal anymore 3 In case there is excessive dust In case there is excessive dust around the around the unit unit open the condenser door once a month This unit uses a condenser which is air and remove and clean dust filter Also cooled In case of excessive dust around remove dust from the condenser with a soft the unit dust will pile up at the front of the cloth condenser This will cause the compressor is not cooled effectively and this can cause hi
30. ice list are based on 230 V 50Hz power supply Continuing research and improvements may result in specification changes Errors and misprints excepted 8 3 Contact information In case service or sales back up is needed please do not hesitate to contact your local ilShin Dealer Or contact the European ilShin Headquarter ilShin Europe Kryptonstraat 11 6718 WR EDE The Netherlands info ilshineurope com www ilshineurope com tel 31 318 64 64 81 fax 31318 64 60 34 49 9 CE Declaration of Conformity EC of Conformity NR 231003 75 EC Deana of Conformity NR 231003 115 Product Freezers DF3514 Product name Type Manufacturer HShin Lab Co Ltd Ibamri 82 flan mee ag x E 129 7 m NEN EN 1050 NEN EN 292 1 NEN EN 292 2 NEN EN 60204 NEN EN 50081 2 NEN EN 50082 2 NEN EN 292 2 EN 60204 NEN 50081 2 NEN EN 50082 2 Model DF351 4 EC of Conformity ra EC fiestavation of Conformity Product description Y JI Product description Freezers Manufacturer ilShin Lab Co Ltd Ibamri UN N Leaf NEN EN 1050 NEM EN2924 f jn NEN EN10S 2924 NEN EN 292 2 EN 60204 V ayy es NEN EN 292 2 NEN EN 60204 NEN EN 50081 2 NEN 500822 NANA e 8 NEN 50081 2 EN 50082 2 Model DF3517 Model DF 3520 50 IF is NN on of Conformity WIN NR 231003 105 EC Declaration of Conformity t N
31. mpensation in order to protect the compressor system In case the input voltage is lower than 180V the operation of the unit will be stopped to protect the vital parts of the unit The alarm functions and temperature indication are still operating Factory Installed and Recommended Setpoints SET POINT SET POINT 80 C 40 DF90 Series COLD ALARM DF45 Series COLD ALARM DF85 Series 90 DF35 Series 50 WARM ALARM WARM ALARM 70 C 30 C There can be differences in the standard temperature of SET POINT COLD ALARM AND WARM ALARM depending on the model 18 6 Accessories 6 1 Aluminium Storage Racks for Upright Freezers Model Amount of racks view from afront Type of racks for 2 inch boxes Type of racks for 3 inch boxes Max of 2 inch Max of 3 inch External Rack Dimensions mm D x W x H Internal Rack Dimensions mm D x Wx H DF8510 12 racks 3 deep 5 high Part UROO66 3 deep 3 high Part UROO67 180 boxes 108 boxes 425x145x283 For 2 inch boxes each opening For 3 inch boxes each opening DF8514 16 racks 3 deep 5 high Part UROO66 3 deep 3 high Part UROO67 240 boxes 144 boxes 425x145x283 19 DF8517 16 racks 4 deep 5 high Part UROO68 4 deep 3 high Part UROO69 320 boxes 192 boxes 558x145x283 DF8520 20 racks 4 deep 5 high Part UROO68 4 deep 3 high Part UROOG9 400 boxes
32. n We have a strong trust in the quality of our innovative products The main goal of our fast and reliable worldwide service team is to maintain satisfied customers As a result also our customers keep the trust in continuous innovation ilShin Delivering quality through service Your local ilShin Dealer Version 3 0 7 August 2007 Copyright ilShin Europe The Netherlands 2007
33. n the control menu The logging status or interval is shown in the low right corner prep E Warm Alarm 60 0 Logging Cold Alarm Poi Log didius AHB DEEPFREEZER Interval In the User settings dialog box Control gt UserSettings the log interval can be set from 1 minute to 60 minutes The log information is written to a textiles named username date txt in the folder which is specified in the User Settings If an invalid pathname is specified the current directory will be used The format of the logfile is also compatible with an MS Excel worksheet The used format for the date is daymonthyear for instance 311203 On changing of the date at midnight new log file with the new date in the name is created 27 6 6 Chart Recorder Optional 6 6 1 Changing the chart paper Press and hold the change chart button 3 for approximately one 1 second until the pen begins to move to the left of the chart and then release the button Wait until the pen has moved completely off of the chart To remove the chart paper unscrew counter clockwise the chart hub knob at the centre of the chart Remove the old chart paper and position the new one so that the correct time line coincides with the time line groove on the chart plate Refer to following figure for the location of the time line groove BATTERY CHANGE CHART BATTERY BATTERY HOLDER ha m 5 0 5 rp
34. nded or shrunk O Do you hear water stream The unit is equipped with circulation of refrigerant You can hear the refrigerant stream through the circulation It is not a technical problem 3 Alayer of ice has formed in the inside wall of the deep freezer Put the samples with distances for good ventilation O If a sample contains much moisture then put a top on it or keep it in Wrap O Please check whether the door is closed O Please reduce the times to open the door 48 8 Warranty and contact information 8 1 International Warranty The warranty period starts at the moment the unit leaves the production facility After this date five year warranty on the parts is valid During these first 5 years component parts proven to be defective can be send to your local dealer and will be covered by this 5 years warranty on parts only The compressors in the system are covered also by a 5 year warranty Expendable items i e glass filters pilot lights light bulbs and door gaskets are excluded from this warranty Shin will not be liable for any indirect or consequential damages including without limitation damages relating to lost profits or loss of products 8 2 Conditions and terms of sale All prices are ex works from the ilShin Europe Warehouse in The Netherlands excluding insurance seaworthy packing and customs duty Effective prices are those valid on date of confirmation of order All specifications mentioned in the pr
35. p M and place 1381 BT Weesp SP Mf Y w NEN EN 1050 NEN EN 292 1 AOS 77777 4 NEN EN 1050 NEN EN 292 1 NEN EN 292 2 NENEN 60204 S 2 NEN EN 292 2 EN 60204 NEN EN 50081 2 NEN EN 50082 2 x NEN 50081 2 NEN EN 50082 2 7 rmity NR 231003 80 7 Mi ira NR 231003 90 Product mune Type NUT NEN EN 1050 NEN EN 292 1 NEN EN 292 60204 WO NEN EN 292 2 NEN 60204 sowiso 50081 2 EN 50082 2 5 e NEN EN 500812 EN 0082 2 52 Conformity NR 231003 110 Product description 8 i NEN EN 1050 NEN EN 292 NEN EN 2922 NEN EN 60204 NEN EN 50031 2 NEN EN 50082 2 NS N ON NU 29 SSA Ba 227 NR 231003 30 EC Declaration of Conformity vs EC Declaration of Conformity VI i j We we Binnenveer 2 Binnenveer2 1381 BT Weesp Sq Apode and 1381 BT Weesp the Netherlands X N SW My the Netherlands HA WA NEN EN 1050 NEN EN 292 1 ass NEN EN 1050 NEN EN 292 1 NEN EN 292 2 60204 A W WM NEN EN 292 2 EN 60204 Sn S 09001200 50081 2 NEN EN 50082 2 Reg 7977 1 Model DF9007 Model DF9010 5 EC Declaration of Conformity W T al EC Declaration of Conformity NR 231003 40 NR 231003 50 Product
36. part h and part e Also some holes are already drilled a CO Back up system b Injection tube with holes straight one is used in case of Up Right freezer bended one is used in case of Chest freezers c Connection cable for door switch d Black flexible hose e Capillairy tubing f Copper insulated hose g 21 8 connection for cillinder including teflon ring for tight fitting h kit lute for sealing up holes i Connection screws 2x 23 6 4 3 How to install the Backup System 1 Remove the black stoppers on the back side and top of the freezer in case of a chest freezer both stoppers are on the back side 4 2 Remove the protection panel covering the temperature sensor inside the freezer Drill a hole of 13 mm in the stopper which is located on top of the freezer and which was removed in step 3 Drill one hole of 10 mm in the stopper which is located on the back of the freezer and which was removed in step 3 Insert the injection b tube into the freezer Connect the injection tube b to the inner wall of the freezer with the two connection screws i holes are factory drilled 7 Seal insulate the stopper hole with kit h 8 Connect the door switch cable c between the back up system a and the connection on the back of the freezer 9 Put the sensor of the Backup System through the 10 mm hole of the stopper which was removed in step 3 10 Seal insulate
37. reading 3 If an adjustment is required use the left 1 and right 2 arrow push buttons on the recorder to calibrate or move the pen s position on the chart to correspond to the temperature of the solution The arrow buttons must be held for approximately five 5 seconds before the pen will begin to move 6 6 5 Battery backup The green LED light remains a constant green colour indicating that both the battery and the main power to the unit are good If the AC power were to fail or the battery becomes weak then the green LED light will begin flashing indicating that either you have lost the main power to the unit or it is time to replace the battery Having a 9 volt DC battery back up in place will allow the recorder to continue to function normally for approximately 24 hours in the event of a power failure 6 6 6 Battery location and replacement To replace the battery first open the recorder door The battery will be located in the upper right hand corner of the unit Note Use only NEW 9 volt alkaline replacement batteries 30 6 6 7 Chart range selection The recorder that you are using has a range sticker that lists several temperature ranges mounted on the front of the chart plate This recording unit has the option for the user to select from several temperature ranges that are programmed into the recorder Note The chart paper that is used on the recording unit must match the range that is selected for
38. ss Lab www Wirelesslab eu Warm up curve for DF8510 for testing isolation rate Model DF8510 40 Becr EE eS POCHE Ade e ee ee ee SO pac aer Temperature r 100 w Temperature Testing environment Ambient temperature 259C 2 9C Powered by Empty freezer Wir eless Lab ww w wirelesslab eu 2 5 Characteristics of Temperature DF90 Series Pull down curve for DF9010 for testing cooling speed Model DFS010 20 40 Temperature 216 50 eee i ace rears 100 12 00 w Temperature Testing environment Powered by Ambient temperature 25 C 2 0 Empty freezer Wir eless Lab www wirelesslab eu Warm up curve for DF9010 for testing isolation rate Model DF8010 Temperature C ea 8 9 00 AM Temperature Testing environment Ambient temperature 25 2 C Fowered by Wireless Lab www wirelesslab eu 10 3 Installation 3 1 Installation Place 1 Do not subject to direct sunlight Adequate air circulation Not subject to excessive dust Not subject to mechanical vibration or shock Operating temperature 5 C to 3 C Recommended performance conditions are ambient of 18 250 Operating Height below 2 000 m Operating moisture below 80 in case that the max operating temperature is 33 it should be below 57 8 Allowance of supplying voltage change lower
39. than 1096 In case you install the unit other places than mentioned above the unit cannot operate properly QE une 8919 3 2 to Install 1 Keep the distance between the wall and the backside of the unit 15cm or more please use the standard supplied distance pins 2 Do not install the unit in an uneven position Make the unit fixed with fixing screws on the sleepy floor 4 The ventilation should not be obstructed 4 Power supply power switch amp external connections 4 1 Power Supply 1 To ensure trouble free connection connect the plug firmly in the power socket 2 Dont connect the plug of the unit together with plug of other units in the same socket 3 Make sure that the power supplying cord is not folded and pressed 4 If you use an extension wire it should has capacity above 250V 15A and not longer than 3m 5 This unit is exclusively designed for 208 230V 50Hz please check the operating voltage before use Under 110V transmitter for elevating voltage above 3Kw should be used 4 2 How to Earth 1 For security reasons the unit should be earthed 2 If the power socket is fitted with earth rod 230V please check the outlet for earth before use 3 In case the power socket is not fitted with earth rod 230 V please contact your local ilShin supplier 4 3 Main Power Switch P d Main Power Switch Door alarm CO2 Backup System RS232C Port Remote alarm
40. the first stage cooling system The condensed refrigerant goes through strainer and passing through the second stage capillary tube Bacuse of this capillary tube low temperature low pressure it is evaporated in the evaporator and cools the inner chamber to temperatures down to 86 C Summary e Operation of first compressor e Cooling of first stage heat exchanger down to 40 C e Operation of second stage compressor e Second stage refrigeration is condensing in the heat exchanger of fist stage e Cooling the inner chamber by the second stage refrigeration down to 86 C 46 7 3 4 Charging Quantity and Type of Refrigerants 7 3 4 1 Refrigerant Quantity for Chest Models all data 25 C ambient temperatures Retigerant Approx 4 bar 140psi pressure 5psi suction 150psi 5 5 Pressure HP L262B single stage system 1 HP L452B single stage system oe 155psi pressure 2 x 1 2 HP L262B cascade system Approx 10psi suction 120psi 3 5 bar pressure 0 psi 160psi suction pressure 7 3 4 2 Refrigerant Quantity for Upright Models all data at 25 C ambient temperatures H 290 270 H 290 270 H 290 270 H 290 270 Approx 150psi R 290 270 4 bar 5psi isa ind ii Above tables are for reference only In case of re charging please contact your local ilShin service engineer 2 x 1 HP
41. the recorder Otherwise the pen s position on the chart paper will not correspond to the temperature that is measured Also if the pen moves to the centre or outer edge of the chart and remains there while the unit is powered on then this maybe an indication that the current range that is selected for the unit is not correct The recorder has a built in safety mechanism that will always move the pen to the highest temperature on the chart when the current temperature that is measured is not within the selected range This recording unit has five 5 user selectable ranges programmed into the unit To select from one of the available ranges follow these instructions After the recording unit has been powered on and is operating normally recording temperature press and hold the change chart button 3 until the pen arm begins to move off of the chart Once the pen has moved off of the chart press and hold the left arrow 1 or right arrow 2 button for approximately five 5 seconds and then release the button The green LED light will begin flashing one 1 time if range 1 is selected or will flash two 2 times if range 2 is selected and so on Press the left arrow button 1 to increase the range number or press the right arrow button 2 to decrease the range number that is selected for the recorder When you have finished selecting the range press and hold the change chart button 3 until the pen begins to move back onto t
42. the stopper hole with kit h 11 Insert this sensor into the freezer using the throughput on the back of freezer 12 Click this sensor in the free clip and re install the protection panel to cover the temperature sensors 13 Connect the capillary tubing e to the injection tube b 14 Connect the black flexible hose d to the capillary tubing 15 Connect the black flexible hose d to the OUT 5 of the back up system a 16 Connect copper insulated hose f to IN 4 of the CO back up system a 17 Connect the connection for cylinder g to the cylinder Do not forget to use the Teflon ring for well fitting 18 Connect the copper insulated hose f to the connection for CO cylinder g 19 Open the valve on top of the liquid CO cylinder never use gas 20 Test the connections on leakages with soap water 21 Set the set point of the back up system on 55 C 22 Always put the high alarm of the freezer on a colder set point 65 C 23 Switch on the BATTERY SWITCH on the back of the system 24 Switch the ALARM SWITCH on to activate the alarm signal 24 WARNING PLEASE DO NOT TURN ON THE CO2 BACK UP SYSTEM FOR THE FIRST TIME BEFORE THE UNIT HAS REACH THE TEMPERATURE OF 80C If the CO2 Back up system is turned on before the unit is cold 80C the back up system will start injecting CO2 because it has not reached the right temperature yet 6 4 4 How to set the t
43. w 5 5 psi Cheeking temp amp pressure 2 high 100 110 psi low 5 5 psi 43 7 3 2 Compressor is running but with wrong temperature Necessary tools manifolds refrigerants R408A R409A R290 R23 welding machine Think of the no leaked at the refrigerating machine Power ON Set Point Check After 10 min High High normal Normal pressure pressure Clean condenser High pre 130 140 psi High 140 160 psi Refill R409A little Low pre 5 5 psi Low 5 5 psi After 10 min Below normal pressure Below Normal pressure Clean condenser Refill R408A little High 140 160 psi Low 5 5 psi Checking temp amp pressure Checking temp amp pressure Charging R290 R23 a little If ambient 20C chamber temp 80C 1st high 130 150 psi low 5 5 psi Pressure of 1 2 pressures may differ according 231 high 100 110 psi low 5 5 psi to ambient amp chamber temperature 44 7 3 3 Refrigeration drawing and prinipal of the cascade refrigeration system Jesuepuoo 1 ssel 1 JosseJduuoo oDeis 15 JOo1e nuuoowv 1018186985 eqn Jeuc uoisuedx3 eqn eqn Arejideo JosseJduuoo oDeis puz