Symphony Technical and Maintenance Guide
Contents
1. Program card isn t completely inserted Push card in all the way See Technical Bulletin No 109A Batter low Battery needs charging soon will beep every 20 seconds 15 minutes of pumping time remaining from when the indication started Be sure power cord is plugged into back of pump See Technical Bulletin No 106A BatLery flat Battery is completely discharged Charge the batteries but you may continue to pump See Technical Bulletin No 101A Motor overload Motor is drawing too much current Send Symphony pump to Medela service center if this happens repeatedly Configuration Software configuration Replace program card if experiencing error See Technical Bulletin No 109A Error Potential fault in electronics of gear control Send Symphony pump to Medela service center if this happens repeatedly Power fault Potential fault in electronics of power supply Send Symphony pump to Medela service center if this happens repeatedly Menu Codes The following are Menu Codes and descriptions indicating historic error activity and steps you can take to address the issue Directions on how to read Menu Codes 1 Turn the pump on 2 Push the On button again while turning the vacuum control knob 3 To see Menu Codes turn the vacuum control knob and refer to the Menu Codes chart listed below Menu Codes Description Solution Unplug and plug back in Reboot pump
2. see below Replace program card and reboot Send to Medela Inc for repair if this does not solve the issue Motor Fault 4 Motor Overload Return to Medela Inc for repair if this happens repeatedly Reboot pump see below Verify proper program card orientation Correct as needed Invalid Card Replace program card and reboot Return to Medela Inc for repair if this does not solve the issue Install program card Verify that card is pushed in all the way Replace program card and reboot No Card Charge for full 12 hours Seep AC Return to Medela Inc for repair if this does not solve the issue Charge for full 12 hours Pump will charge during use if plugged in Battery Low Return to Medela Inc for repair if this does not solve the issue Verify pump plugged in properly at both ends Plug pump into another outlet Return to Medela Inc for repair if this does not solve the issue Power Fault Reboot pump Replace program card and reboot Return to Medela Inc for repair if this does not solve the issue Configuration WONMNH_ wWwNhMH MOH NaJ ONAJ AON Quick Fixes The following is a list of potential issues and steps you can take to readily address the issue 1 Press the let down button and hold While holding the let down button press the on off button Release both buttons when pump reads pleas
3. Symphony Vacuum Settings 120 CPM Vacuum Range 50 200mmHg 78 54 CPM varies Vacuum Range 50 250mmHg Refer to page 3 for extensive values Chart for vacuum settings Helpful Tips e Vacuum levels will vary based on weather and elevation Refer to the pump service instructions for vacuum levels based on elevation e Dropping or damaging the vacuum gauge may cause incorrect readings e This gauge is intended as a reference tool only not as a means of accurate measure If you need further assistance with troubleshooting low suction with your Symphony please refer to Technical Bulletin No 103A Inspect amp Replace Power Cord In the event that the Symphony 4 5 Insert the new cord through the breastoump power cord becomes cord holder damaged this technical bulletin will guide you through the steps to replace 6 Loop the cord around inside of the power cord If you continue to have screw post problems replacing your power cord please contact Customer Service at 7 Place the cord holder back on the 800 435 8316 pump and tighten the screw Symphony power cord has been Tools Required repaired and is ready for use 2 5 Allen Wrench if oump does not have a thumb screw Symphony Power Cord Item 9280043 Warning Please be sure to unplug the Symphony prior to troubleshooting Directions 1 Inspect power cord for damage ie 1a Damaged Cord 1b Damaged Plug 1c Bent Prong 1d S
4. and for double pumping by using the appropriate accessory kit Refer to the separate accessory kit user s instructions for further details IMPORTANT SAFEGUARDS WARNING To reduce the risk of burns electrocution fire or injury to persons 1 Separation from the power outlet is only assured through the disconnection of the plug and socket connection 2 This equipment is not suitable for use in the presence of a flammable anesthetic mixture 3 Use only in a dust free environment 4 Do not place in or drop motor unit into water or other liquid 5 Do not reach for any electrical device if it has fallen into water Unplug immediately 6 Never operate an electrical device if it has a damaged cord or plug if it is not working properly if it has been dropped or damaged or if it has been dropped into water 7 Only use the electric cord that comes with the Symphony Breastoump 8 Make sure the voltage of the device is compatible with the power source Symphony Technical Description A Safety Instructions SSI w Medical Equipment N5 9 With Respect to Electrical Shock Fire and Mechanical Hazards Only CLT US In Accordance With IEC 60601 1 CAN CSA 22 2 No 601 1 22PL Protection class Il ol Double Insulation UN Follow Instructions a A Alternating Current Type B Direct Current Vacuum 50mmHg 250mmHg 7kPa 33kPa Power Supply Conventional Outlet 100V 240VAC 50
5. and remove each diaphragm with flathead screwdriver 6 7 Clean area with soft cloth and Tools Required Cavicide or like product by spraying on cloth before replacing diaphragms Flathead Screwdriver Replacement Diaphragms Kit 8 Item 8007199 Cavicide or similar product 9 Apply equal pressure downwards to Soft Cloth snap diaphragm in place Align new diaphragms with openings 10 Install screws with flathead screwdriver Warning Please be sure to unplug Do not over tighten Symphony prior to troubleshooting Directions Inspect 1 2 Press gently on diaphragm An undamaged diaphragm will roll along the edge which does not affect suction If the rolled edge appears to be torn replace If holes or tears appear in diaphragm suction can be reduced and diaphragm needs to be replaced If you experience other issues regarding ie suction reference Technical Bulletin 103A Warning e Stop when screws stop turning e Do not over tighten or internal sensor may get damaged If the above steps do not correct the issue please contact Customer Service at 800 435 8316 13 Low or No Suction In the event that your Symphony breast pump is producing low or no suction this technical bulletin will guide you through the steps to troubleshoot your issue If you continue experiencing issues with low or no suction please contact Customer Service at 800 435 8316 Tools Required Technic
6. experience values outside of the normal limits reference refer to Technical Bulletin Low or No Suction on pages 14 15 Recommendations for Replacement Over the life of the Symphony it is possible that certain items may have to be replaced The more preventative maintenance that is conducted on the Symphony the less parts will have to be replaced The following are recommended guidelines for when certain parts should be replaced Recommendations for Replacement Repair Membranes As needed after inspection Refer to Low or No Suction Technical Bulletin on pages W 14 15 for detailed instructions Diaphragms As needed after inspection or low suction Refer to Inspect Replace Diaphragms on pages 12 13 OD for detailed instructions 6 000 hrs or as needed 6 000 hrs or as needed As needed Refer to Battery Care amp Maintenance Technical Bulletin on pages 10 11 for detailed instructions y ce Main Boards As needed If pump does not turn on and after troubleshooting Refer to Error Codes Technical Bulletin on pages 8 9 Housing If broken or separated Refer to Correct Separated Housing Technical Bulletin on pages k 28 29 for detailed instructions Must be replaced or repaired at Medela How to Check Pump Total Running Hours Checking Pump Total Running Hours 1 Turn the pump on 2 GO to service menu by pushing the On button again while ON turning the vacuum control kn
7. please send back to Medela for repairs Caf Housing Inspect for damage cracks or if housing is separated Refer to Correct Separated Housing Technical Bulletin on pages 28 29 for detailed instructions A Rubber Feet Inspect for damage or missing rubber feet 4 total If any are missing contact Medela Customer Service to order Part 8000596 A Pump Diaphragms Inspect for damaged diaphragms Refer to Inspect Replace Diaphragms Technical Bulletin on pages 12 13 for detailed instructions Cat Power Cord Inspect power cord for damage Refer to Inspect Replace Power Cord on pages 26 27 for detailed instructions Mf If you have a Symphony Plus check battery status by referring to Battery Care amp Maintenance Technical Bulletin on pages 10 11 Performing a vacuum check on your pump ensures your Symphony is operating correctly for breastopumping mothers Follow the instructions in the Proper Use of Vacuum Gauge Technical Bulletin on pages 24 25 for detailed instructions The following values will guide you through gauging your Symphony vacuum levels The following values should read Stimulation Expression O M q min max max min Elevation OFF 4 166 207 41 6600 47 187 234 47 3300 50 200 250 50 1600 53 213 266 53 Sea Level 20 mmHg Stimulation mode changes to Expression mode after 2 minutes unless the button is pushed If you
8. symphony qito Sua Battery Check if battery model 9 O 3 ae E i ME as If the above steps do not correct the issue please contact Customer Service at 800 435 8316 1 Charge pump for 12 hours NOTE Display switches from charging to Symphony 2 0 when battery reaches 80 charge It takes the full 12 hours to get 100 charge Turn the pump on Go to the service menu by pushing the On button again while simultaneously turning the vacuum control knob clockwise until S N is displayed Release the power button and continue to turn the vacuum control knob until locating Hres resettable hours then press the let down button to reset it back to zero Let the pump run until the batteries are drained and the screen reads Battery flat The pump turns off automatically after 30 mins and may need to be restarted to fully drain the batteries Once the batteries are drained plug the pump into a wall outlet and go back to Hres on the service menu to check how long the pump ran The time will give an indication of the battery s condition A new battery lasts for about 60 minutes of pumping time 11 Inspect amp Replace Diaphragms Symphony diaphragms need to be Replace inspected periodically to ensure proper suction levels If you experience issues replacing the diaphragms please contact Customer Service at 800 435 8316 3 Remove diaphragm screws with flathead screwdriver 4 5 Gently pry
9. 60 Hz 0 5 A DC Connection 12 VDC Use only with 25w Medela Vehicle Lighter T 2 5A e Adapter 67173 Battery 2x6V 1 2 Ah Type Yuasa NP 1 2 6 Battery optional Battery DC powered Charging time 12 hours Pa Full battery pumping time approx 60 minutes Ea ESD Use ESD protection when servicing pump Disposal This breastoump is made of technical plastics and metals that cannot be separated It must be disposed of in accordance with local regulations Dimensions With Stand Without Stand Length 16 1 8 5 Width 16a 6 1 Height 35 4 9 6 Weight With Battery 6 lbs 7 oz Without Battery 5 lbs 1 oz Environmental Conditions for Operation Temperature 41 F to 104 F Relative Humidity 20 to 80 Atmospheric Pressure 70 kPa to 106 kPa Environmental Conditions Transport and Storage Temperature 4 F to 122 F Relative Humidity 20 to 95 Atmospheric Pressure 70 kPa to 106 kPa Standards FDA GMP ISO 9001 ISO 13485 CE 93 42 EEC Electromagnetic or Other Interference Complies with IEC 60601 1 2 Unlikely to interfere with other electrical equipment Warranty 3 years for the unit from the date of delivery excluding pump set and batteries The manufacturer is not liable for any damage or consequential damage caused by incorrect operation inappropriate usage or use by unauthorized persons Routine checks servicing and battery exchange may only be carried out by Med
10. Use of Vacuum Gauge sssssnnsssunnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn mnnn 24 25 Inspect amp Replace Power Cord cssscssecesscssecnscensensecnsenneeesensenssennenaes 26 27 Correct Separated HOUSING c nscensecesecensccnsecnsecenseonssonssenssecnsenensenaes 28 29 NS AIAG EE AE E E A A E E E A 30 31 Terms of Guarantee asssssnnnnnnnnnnnnnnnnnnnnnnnunnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn 32 The Medela Symphony Breastoump has been specially developed to meet the most stringent demands where hygiene and mobility are concerned The operating principle of the Medela Symphony is to simulate the natural physiological conditions encountered in breastfeeding The pump s kit is separate from the pumping mechanism and protected from overflow by a specially designed membrane Kit components which could come in contact with the milk are part of the accessories and can be very easily disconnected for cleaning and sterilization A carrying case is available offering ample space for the pump and the accessory kit There are two different models of Symphony with the following general specifications Symphony Breastpump article no 0240108 one knob control for both vacuum and suction cycles E powered by regular electric outlet 100 240 VAC Symphony Breastpump Plus article no 0240208 same as above 0240108 but with 2 internal rechargeable 6 volt batteries e used for single
11. aintenance Symphony Plus needs battery care and 3 Battery Flat appears when the maintenance If you are unable to resolve battery has been depleted The your Symphony battery issues please breastoump will issue a long beep contact Customer Service at 800 435 8316 and the pump will switch off The battery should be re charged using the above instructions Charging Battery 1 The Symphony Plus should 4 Check battery connections by be charged removing battery caps and e upon receipt of pump Inspecting wires for approximately 12 hours e daily after having run from battery e before battery completely Helpful Hints discharges Extreme temperature charging rates and state of charge play a part in the life span of your battery e every two 2 months when pump is not in use The battery is continually charging when the breastpump is plugged into an outlet even during a pumping session To help maintain the life please do the following e completely recharge the battery after each use Battery Status e do not allow the battery to fully drain on a regular basis e during use the Symphony can run on battery for up to one 1 hour 2 Battery Low warning appears on LCD display when the battery is low The breastoump will beep every 20 seconds The breastpump still has approximately 15 minutes of pumping time left The battery should be re charged using the above instructions symphony
12. al Bulletin No 102A Inspecting and Replacing Diaphragms Flathead Screwdriver Symphony Diaphragm Kit Item 8007199 if needed Symphony Membrane Cap Item 8100769 if needed Symphony Membrane Item 8200094 if needed Vacuum Gauge Item 1977005 if needed Warning Please be sure to unplug Symphony prior to troubleshooting Directions 1 Inspect Symphony Membranes If membranes do not appear to have a rip or tear move to next step If membranes do appear to be worn or have a tear replace prior to next step Inspect Symphony Diaphragms by following Technical Bulletin 102A If diaphragms do not appear to have a rip or tear move to next stop Make sure Symphony lid completely closes before pump is in use Make sure the hard yellow ends of the tubing fit securely into breastshield connectors Make sure the yellow triangular end fits into the membrane cap Make sure yellow valve on breastoump kit is fitted correctly into the breastshield connector Make sure white membrane is lying flat against yellow valve head Check Symphony vacuum levels by following Technical Bulletin 110A Proper Use of Vacuum Gauge If the above steps do not correct the issue please contact Customer Service at 800 435 8316 15 Trolley Repair In the event that your Symphony trolley becomes loose this technical bulletin will guide you through methods to resolve this iss
13. and is ready for use If the above steps do not correct the issue please contact Cus 21 Replace and Load Program Card In the event that your Symphony displays an invalid card or no card error please refer to the following to resolve the issue If you continue to experience issues with your program card please contact Customer Service at 800 435 8316 Tools Required 2 5mm Allen Wrench if oump does not have thumb screw Symphony Replacement Cards if needed 2 0 Card English 67208 French 67209 Spanish 67210 Preemie Card English 67227 French 27225 Spanish 67226 Warning Please be sure to unplug the Symphony prior to troubleshooting Directions 1 Loosen thumb screw or 2 5mm allen head screw Slide cord holder up 2 Remove program card by pulling slowly and gently 3 4 5 10 Gently wipe chip contacts with alcohol wipe or dry towel and re insert into the Symphony Hold half circle in hand with micro chip facing up to insert card If program card still does not function or is missing insert a new Symphony program card Slide cord holder down and tighten screw Plug Symphony into wall outlet and turn Symphony on Press and hold the let down button While holding let down button press on off button until Please wait displays This process can take up to 45 seconds Release buttons Pump will start running Press
14. e wait Reboot Pump Turn off pump and restart unit Reboot pump Replace program card and reboot pump refer to Technical Bulletin No 109A to replace program card Send pump to Medela Inc if this does not solve the issue Error Messages ROwONM WP Check program card a Chip side up Invalid Card b Chip end inserted into pump first 2 Replace program card Reboot pump with new card See reboot instructions Return pump to Medela Inc for repair if this does not solve the issue If program card is missing a Install new program card refer to Technical Bulletin No 109A to replace program card b Reboot pump with new card See reboot instructions No Card 2 If program card is present a Remove card and reseat b Send to Medela Inc for repair if reseat does not correct Check if the pump is properly connected to the electric supply or that the battery is charged No Power Inspect power cord and replace if it is damaged refer to Technical Bulletin No 111A to replace power cord Inspect fuse and replace if needed refer to Technical Bulletin No 106A to replace fuse Send to Medela Inc for repair if issue is not resolved Check battery connections by removing battery caps and inspecting wires Check Battery if present Refer to Technical Bulletin No 101A for further Battery Care amp Maintenance instructions wp RWND Battery Care amp M
15. ela authorized personnel Symphony Breastpump Controls amp Parts Symphony Diagram Symphony Kit Diagram Kit shown is Symphony Double Pumping Kit 67099S Let down Button Button PersonalFit Breastshield Standard 24mm PersonalFit Breastshield Connector Valve amp Membrane 8117009 8107057 8007213 8100769 8200094 Breastfeeding Information Guide 1907200 10 Symphony Instruction Manual 1907724 oo Sl we OS a 11 Symphony Container Stand 12 Symphony Vehicle Lighter Adapter Call Customer Service at 800 435 8316 for ordering information Visual amp Functionality Checklist The Symphony Breastoump Maintenance amp Technical Guide is a resource and tool to be used to perform maintenance and troubleshoot issues on the Symphony Breastoump A visual and functional check of the Symphony breastoump should be performed for optimal performance on a periodic basis If maintenance or repairs are needed this guide will walk you through the steps If you have any questions please contact Customer Service at 800 435 8316 A Cleaning Use a soft cloth to wipe with warm soapy water Quick Clean Wipes may be used between uses with the same mother Use Cavicide to disinfect between different mothers Refer to Cleaning Technical Bulletin on pages 30 31 for detailed instructions iA Cover Inspect for damage and that cover closes and locks shut If cover is damaged
16. evered Cord 2 Loosen screw and remove the card holder by moving upwards 3 Remove and discard recycle bad cord If the above steps do not correct the issue please contact Customer Service at 800 435 8316 21 Correct Separated Housing In the event that the Symphony Breastoump housing becomes separated or dislodged this technical bulletin will guide you through the steps to resolve the issue If you continue to experience issues please contact Customer Service at 800 435 8316 Tools Required Phillips Screwdriver Warning Please be sure pump is disconnected from the power cord prior to repairing Directions 1 Remove four 4 screws at the bottom of the pump using the Phillips screwdriver Carefully pull cover straight up Component pieces of pump are attached to cover Snap separated halves back together by carefully aligning tabs 5 6 Replace cover back on pump by aligning internal posts into holes Reseat cover making sure to cover housing completely Replace four 4 bottom screws at the bottom of the pump Symphony housing has been repaired and is ready for use oo a KO ae If the above steps do not correct the issue please contact Customer Service at 800 435 8316 29 Cleaning The Symphony breastpump may be cleaned with Quick Clean wipes or a non abrasive detergent for general cleaning However the Symphony should be disin
17. fected between use by different mothers This technical bulletin guides you through the steps to clean or disinfect your Symphony Breastoump If you have questions regarding this technical bulletin please contact Customer Service at 800 435 8316 Tools Required Cavicide 8 fl oz Item 3007009 Soft Cloth Quick Clean Wipes Item 87056 Warning Please be sure to unplug Symphony prior to cleaning Directions 1 For cleaning in between a mother s own sessions To clean the pump use a Quick Clean wipe to wipe over the breastpump Optionally soapy water or a non abrasive detergent may be used 2 a b For cleaning in between different mothers To disinfect the pump we recommend a solution called Cavicide Spray onto a clean cloth and wipe the breastpump making sure not to spray or pour liquid directly onto the pump Do not immerse the pump unit in water do not spray or pour liquid directly onto the pump r N DO NOT SPRAY OR POUR LIQUID DIRECTLY ONTO PUMP If the above steps do not correct the issue please contact Customer Service at 800 435 8316 31 Terms of Guarantee The manufacturer may only be held responsible for inadequate safety reliability and performance of the unit if assembly additional items adjustments alterations and or repairs have been carried out by persons authorized by the manufacturer and the electrical installation within the room concerned
18. medela The Symphony Breastpump Maintenance amp Technical Guide Table of Contents Save These INSWUCTIONS cin secsiecictcse cece nccsesdntntcesiecceascecsacnsnaesencctarecsienbemssenceceane 1 Symphony Technical Description c csscesseessenneenseensennsensenesecnsensenenenseees 2 Symphony Breastpump Controls amp Parts ccccccccsseeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeenees 3 SVD HOMY Diagram aorin anini ie e ea aadi 3 SVINONONY KH DNA eraes sininen rna EENEN EE EE Er E EE in aren 3 Visual amp Functionality Check list ccsccescsssecsecesecnsecnsenssensenseessensensensseeas 4 Vacuum GCC ioc cats caterers etc neste cee ete ew cite nnmnnn nnmnnn nnmnnn nnmnnn 5 Recommendations for Replacement cescescenscescesecsecnsensensensecnsensensensenas 6 How to Check Pump Total RUNNING HOUTrS ccnsccesecenscenseenseeenesenseenseees 7 Error 10 0 eee ne Bn a eee ee ee ene eee een eee en er emer eerie errr 8 9 Battery Care amp Maintenance c csssessecesecnseensenssenseneeenseeseensennsennenaes 10 11 Inspect amp Replace Diaphragm s csssssscesceseensenseeseessensensensenssensensenees 12 13 Low or No oC Oa ascites cece ence seein e scence ewewnaceetecnneeueuaves 14 15 ATONE Y RODI aeaea E EEEE EEEE G 16 17 NO ON senei EEA EEEE EAA EEE 18 19 FSS SAI OVEN sociano anaa aaa aAa aa aO aE 20 21 Replace and Load Program Card csssscssscensccnsenseesecenseonseensseenseees 22 23 Proper
19. o the bottom of the Symphony Flathead Screwdriver small Directions 1 Inspect power cord and replace if damaged Refer to Technical Bulletin No 111A Inspect Replace Power Cord if power cord needs to be replaced 2 Make sure pump is plugged into a wall outlet 3 Make sure short black cord is plugged into back of Symphony 4 Check power outlet for functionality Attempt to use in another power outlet Replace if Damaged 4 9 gt On p X fx f ri IS G a Shorted Fuse If the above steps do not correct the issue please contact Customer Service at 800 435 8316 19 Reseating Cover In the event that the Symphony breastoump cover becomes separated or dislodged this technical bulletin will guide you through the steps to resolve this issue If you continue to experience issues with the cover please contact Customer Service at 800 435 8316 Tools Required Phillips Screwdriver Warning Please be sure to unplug the Symphony prior to troubleshooting Directions 1 Inspect the pump to locate the housing ridge s dislodged outside of cover 2 Remove four 4 screws at the bottom of the pump with a Phillios Screwdriver 3 Pull cover straight up fust enough to clear top of housing Reseat cover over top of the housing Make sure ridge s are not showing Replace four 4 screws at the bottom of the pump Symphony cover has been reseated
20. ob 3 Turn the vacuum control knob until you get Htot hours total This is the total hours the pump has run throughout its life 4 Use these hours to check to see if maintenance Is required per recommendations on page 6 Error Codes There are two ways in which a Symphony Breastoump will communicate that the unit has an issue If the Symphony is experiencing an issue an error code will be displayed on the LCD screen see fig 1 Previous problems are recorded in the Menu function under Err To retrieve a history of errors hold down the On button while turning the vacuum control knob then continue to move the knob until you reach Err see fig 2 If you continue to experience issues with your Symphony please contact Customer Service at 800 435 8316 LCD Display Messages The following describes LCD messages that may be displayed indicating an issue with the unit the potential causes and action steps to correct the issue Potential Issue Actions Invalid card Program card isn t a Symphony card Check if correct program card is used See Technical Bulletin No 109A Program card is incorrectly inserted Check that card is positioned correctly See Technical Bulletin No 109A Program card is damaged and not working Replace program card See Technical Bulletin No 109A Ho card Program card isn t inserted Insert program card See Technical Bulletin No 109A
21. on off Q button to turn off Symphony 2 0 or Preemie Standard will be displayed in LCD window Symphony is ready for use symphony mD amp a Q s Be symphony symphony symphony w 10 symphony Sarto ne O 3 If the above steps do not correct the issue please contact Customer Service at 800 435 8316 23 Proper Use of Vacuum Gauge This technical bulletin will guide you through the steps of checking the vacuum levels of your Symphony using a vacuum gauge If you are unable to resolve the issue please contact Customer Service at 800 435 8316 Tools Required Vacuum Gauge Item 1977005 Symphony Double Pumping System Item 67099 Symphony Lactina Double Pumping System Item 67116 Directions 1 Insert the vacuum gauge into the hole in the rubber stopper 2 Pull the tab on the rubber stopper to ensure proper fit onto gauge post The tab should be positioned in the back of the vacuum gauge 3 Insert the rubber stopper with gauge into breastshield 4 Ensure a complete seal by firmly pressing on all edges of the stopper 5 Connect only one Symphony membrane at a time onto diaphragm before you begin testing Turn on the Symphony to measure the vacuum level Both sides must be tested Each side should be within 20mmHg or the mom will notice a difference 6 After you have finished checking the vacuum levels grasp the gauge and stopper tab to remove it
22. satisfies IEC requirements in the course of preparation E the unit is used as specified in the instructions for use E accessory kits used on Symphony Breastoumps are authorized by Medela Inc Symphony Preemie gyn Y a mede e Significantly more milk in less time when premature infants need it most Because every drop counts Symphony Preemie Preemie offers two breastoumping pattern programs that have been shown in clinical trials to initiate milk supply in mothers of premature infants and produce significantly more milk in fewer minutes spent pumping In addition Preemie has been shown to allow mothers of premature infants to achieve as much milk as mothers of exclusively breastfeeding term infants 530ml by day 6 of breastoumping medela lt Medela Inc P O Box 660 1101 Corporate Drive McHenry IL 60051 0660 Phone 800 435 8316 Fax 815 363 1246 Email customer service medela com www medela com Symphony and Medela are registered trademarks and Preemie Quick Clean and Micro Steam are trademarks of Medela Inc Cavicide is a registered trademark of Micro Aseptic Products Inc 1907555 E 0909 2009 Medela Inc
23. ue If you continue to experience issues with your Symphony trolley please contact Customer Service at 800 435 8316 Tools Required 3 0mm Allen Wrench 13mm Socket Wrench Adjustable Wrench Trolley Repair Kit Item 6007116 if necessary Directions 1 2 Remove Symphony from trolley by removing four 4 screws with 3 0mm Allen Wrench 3 4 Turn trolley on side and hold nut by using adjustable wrench Nut should be Nylok Nut provided in Repair Kit 6 7 Tighten rod by using 13mm socket wrench while holding nut which is located on bottom of trolley with adjustable wrench See image 4 Place Symphony back on trolley and tighten four 4 screws with 3 0mm Allen wrench Your Symphony trolley is now tightened and ready for use If the above steps do not correct the issue please contact Customer Service at 800 435 8316 17 No Power In the event that your Symphony 5 Remove fuse holder from bottom of breastoump does not have power this the Symphony using a small flathead technical bulletin will guide you through screwdriver and gently turn clock wise methods to resolve the issue If you continue to experience issues with loss 6 7 If the fuse appears shorted remove of power please contact Customer shorted fuse item 9310035 Service at 800 435 8316 8 9 Replace with new fuse Item 9310035 Tools Required 10 Insert fuse holder black round cap Fuse Item 9310035 back int