Contents

1. Practical Information Maintenance Do not sterilize the stimulator Clean using a soft cloth and an alcohol based solvent free cleaning product Use only a minimum amount of liquid when cleaning the Empi Elite Do not attempt to repair the stimulator or any of its accessories Never dismantle the Empi Elite or the charger containing high voltage parts because of risk of electric shock Empi declines all responsibilities for any damages or consequences resulting from unauthorized attempts to open modify or repair the stimulator This may only be done by persons or repair services authorized by Empi Your stimulator does not require calibration Each Empi Elite stimulator is always tested and validated prior to distribution Its characteristics do not vary under normal conditions Nonetheless as the Empi Elite is a high quality electrical instrument its lifespan depends on the use that is made of it and the care and maintenance it receives 16 during its lifetime If your stimulator contains parts that seem worn or defective please contact your nearest Empi customer service center regarding a replacement unit Storage and transportation conditions The Empi Elite contains rechargeable batteries This is why it must be stored and transported in accordance with the following conditions Storage and transportation temperature from 20 C 4 F to 45 C 113 F Maximum relative humidity 7596 Atmospheric p
2. For Knee treatment typical electrode For Shoulder treatment typical electrode placements are placements are For Hip treatment typical electrode placements are Empi Elite User s Manual 25 Made in Tunisia for Empi SEmpi Empi Inc 205 Hwy 22 East DJO Clear Lake SD 57226 USA GLOBAL 651 415 9000 800 328 2536 360439 Rev A 2014 Empi Inc MOTION IS MEDICINE
3. Do not recharge the battery in a confined space There is a risk of fire or electric shock Keep the Empi Elite and its accessories out of reach of children Do not share electrodes with other persons All users should have their own set of electrodes to prevent undesirable skin reactions Do not allow any foreign bodies soil water metal etc to penetrate the Empi Elite the battery compartment and the charger Empi Elite User s Manual Do not use the Empi Elite while driving operating machinery or during any activity in which involuntary muscle contractions may put the user at risk of injury Do not use the stimulator at altitudes of over 9 842 feet 3 000 meters ZN 1 Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner 2 Safety of powered muscle stimulators for use during pregnancy has not been established 3 Use with caution on patients in the presence of the following a When there is a tendency to hemorrhage following acute trauma or fracture b Following recent surgical procedures when muscle contraction may disrupt the healing process c Over the menstruating or pregnant uterus d Over areas of the skin which lack normal sensation e Suspected or diagnosed heart problems f Suspected or diagnosed epilepsy 4 Some patients may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical co
4. Technical characteristics Device connector 6 pin Electrode connector female Snap Length 1500 mm Protection index Protection IPXO Neuro stimulation All electrical specifications are given for an impedance of 500 1000 ohms per channel Channels Four independent and individually adjustable channels that are electrically isolated from each other and earthed Pulse shape Constant rectangular current with pulse compensation to eliminate any direct current component to prevent residual polarization at skin level Maximum pulse intensity 120 mA Pulse intensity increments manual adjustment of stimulation intensity from O to 999 energy in minimum increments of 0 5 mA Pulse width 200 to 380 us Maximum electrical charge per pulse 96 microcoulombs 2x48 uC compensated Standard pulse ramp up time 3 us 20 80 of maximum current Pulse frequency 1 to 120 Hz 17 EMC Table Electromagnetic Compatibility The Empi Elite needs special EMC precautions and must be installed and started according to the EMC information supplied in this manual All RF wireless transmission systems could affect the Empi Elite The use of accessories other than those recommended by the manufacturer may result in stronger emissions or reduce the immunity of the Empi Elite The Empi Elite should not be used beside or stacked on top of any other equipment If you must use it side by side or on top of another system you sho
5. doubt as to the use of the Empi Elite close to another medical device seek advice from the manufacturer of the latter or from your doctor Do not use the Empi Elite if you are connected to a high frequency surgical instrument as this could cause skin irritation or burns under the electrodes Do not use the Empi Elite stimulator in areas where unprotected devices are used to emit electromagnetic radiation Portable communications equipment can interfere with electrical medical equipment Do not use the Empi Elite in water or in a humid atmosphere sauna hydrotherapy etc Do not use the Empi Elite in oxygen rich environments Do not use in areas where there is a risk of explosion Sudden temperature changes can cause condensation to build up inside the stimulator To prevent this allow it to reach ambient temperature before use Do not disconnect any leadwires during a session while the stimulator is switched on Switch the stimulator off first Never connect leadwires to an external power supply There is a risk of electric shock Never recharge the stimulator without first disconnecting the leadwires Never use the Empi Elite or the charger if itis damaged case cables etc or if the battery compartment is open There is a risk of electric shock Disconnect the charger immediately if the Empi Elite beeps continuously if there is abnormal heating or smell or if smoke comes from the charger or the Empi Elite
6. Use the channel C1 button to choose the language you wish to use Use the channel C2 button to adjust the contrast of the screen Use the channel C3 button to adjust the volume Use the channel C4 button to adjust the backlighting On backlighting always on Off backlighting always off Auto backlighting activated whenever a button is pressed Empi Elite User s Manual Selecting a body area This menu allows the patient to select which body area she he would like to treat There are 3 different possible body areas selectable SHOULDER HIP and KNEE BODY ZONE SHOULDER Press the On Off A button to go to the previous screen Use the channel C1 button to choose a body area e Press the channel C4 button to validate your selection and go to the treatment selection menu The corresponding symbol on the display is a black case with NEXT written in it During the stimulation session 1 Adjusting stimulation energies Selecting a treatment To choose a treatment consult your medical professional It is also particularly useful to consult chapter VI Programs SHOULDER am ENDURANCE STRENGTH ma A we E ePress the On Off A button to go to the previous screen Use the channel C1 button to choose treatment program Press the channel C4 button to validate your selection and go to the treatment selection menu The correspondin
7. E Socket for the battery charger G Disconnect all leadwires before inserting the charger pin E Belt clip socket G Rechargeable battery compartment Your kit contains A 1 stimulator B 1 charger C 1 set black snap connectors lead wires for electrodes color coded blue green yellow red D 2 bags of large snap electrodes 2x4 in 5x10cm E 2 bags of small snap electrodes 2x2 in 5x5cm F 1 belt clip G Rechargeable battery pack H 1 patient information packet I 1 user manual J 1 electrode placement guide Empi Elite User s Manual 9 Connections 1 Connecting the cables The stimulator cables plug into the 4 sockets on the front bottom of the device Four lead wires can be connected simultaneously to the four channels of the device Both the sockets and the lead wires are color coded to simplify use and facilitate identification of the di erent channels blue channel 1 green channel 2 yellow channel 3 red channel 4 2 Connecting the electrodes Each lead wire has a red snap positive lead and a black snap negative lead Connect the lead by snapping the lead on the electrode Once you snap on the lead the lead should stay directly on the electrode Note use the electrode placement pictures to correctly position the positive and negative leads 3 Electrode placement Please refer to the following pictogram for the electrode placements Install yourself comfortabl
8. Fev n3 Xo woods 15 Strength oa s s ss m oo om RR R on ROW A 16 Practical Information 2 2 0 00 llle 16 Maintenance t n a DT MCCC 16 Storage and transportation conditions 0 00 0 16 Disposal isa samara Ra SR ae Rom mh RO AAA eA ed 17 Schere E WT CC 17 Patents e io Po e bow e uou E ende Eod BA Se eee Ard ay i e E ei 17 Technical characteristics ee 17 EMC Table Electromagnetic Compatibility lll 18 Warranty PTT 22 Ordering Information o 24 Electrode Placement Examples 0 0 0002 eee eee ee 25 2 Empi Elite User s Manual Forward Read this User Manual carefully before you start using your Empi Elite device The manufacturer strongly recommends careful reading of the Warnings and Cautions and Safety Guide Sections of this User Manual www djoglobal com website helps you for your first steps with the device How to get help To order the Empi Elite device for your patients contact either your local Empi DJO Global representative or Empi Clinic Services at 800 325 5663 Empi Inc 205 Hwy 22 East Clear Lake SD 57226 USA Caution Federal law restricts this device to sale by or on the order of a practitioner licensed by the law of the State in which he she practices to use or order the use of the device Warnings and Cautions uN Contraindications Do not use the Empi Elit
9. or consequential damages lost profits or medical expenses caused by any defect failure malfunction or otherwise of the product regardless of the form in which any legal or equitable action may be brought against Empi Inc e g contract negligence or otherwise the remedy provided in paragraph A above shall constitute purchaser s sole remedy In no event shall Empi Inc s liability under any cause of action relating to the product exceed the purchase price of the product This Warranty gives the purchaser specific legal rights and Purchaser may also have other rights which vary from state to state Some states do not allow limitations of how long an implied warranty lasts so the above limitation may not apply to the Purchaser Empi Elite User s Manual 23 Ordering Information To reorder call Empi at 800 328 2536 Product Number Empi Elite Stimulator 199729 E Small Snap Electrodes 199730 199725 2x2 in 5x5 cm 199725 Large Snap Electrodes 2x4 in 5x10 cm 199730 Battery Charger 683026 683026 Snap Connectors 601131 lead wires 199729 Rechargeable Battery 941213 or Pack 941210 Belt Clip 949000 601131 24 Empi Elite User s Manual Electrode Placement Examples Please refer to the following pictogram for electrode placement Install yourself comfortably in the position shown in the pictogram Always follow your doctor s recommendations for the electrode placement and for the intensities applied
10. parameters and this is known as electro muscular stimulation EMS Motor nerve stimulation EMS In voluntary activity the order for muscular work comes from the brain which sends a command to the nerve fibers in the form of an electrical signal This signal is then transmitted to the muscular fibers which contract The principle of electrostimulation accurately reproduces the process observed during a voluntary contraction The stimulator sends an electrical current impulse to the nerve fibers exciting them This excitation is then transmitted to the muscular fibers causing a basic mechanical response muscular twitch The latter constitutes the basic requirement for muscular contraction This muscular response is completely identical to muscular work controlled by the brain In other words the muscle cannot distinguish whether the command comes from the brain or from the stimulator The parameters of the Empi Elite programs number of impulses per second contraction time rest time total program time subject the muscles to different types of work according to muscular fibers In fact different types of muscular fibers may be distinguished according to their respective contraction speed slow intermediate and fast fibers Fast fibers will obviously predominate in a sprinter while a marathon runner will have more slow fibers With a good knowledge of human physiology and a perfect mastery of the stimulation para
11. sole obligation in the case of any breach of its warranties set forth in Paragraph A in the Warranty section above shall be at Empi Inc s option to repair or replace the Product without charge to Purchaser or to refund the purchase price of the Product In order to recover under this Warranty Purchaser must send Empi Inc written notice of the defect setting forth the problem in reasonable detail prior to expiration of the Warranty Period and within 30 days of discovery of the defect Upon Empi Inc s written request and authorization Purchaser shall return the Product to Empi Inc freight and insurance prepaid for inspection Notice and return shipment shall be sent to Empi Inc at 47492 Hwy 22 Clear Lake South Dakota 572206 USA or to an Empi Inc Authorized Service Center To locate the appropriate Service Center outside of North America or to request shipment approval contact Empi Inc directly Empi Inc will not be responsible for damage due to improper packaging or shipment If Empi Inc determines in its sole reasonable discretion that the Product contains defective workmanship or materials Empi Inc will refund to the Purchaser the purchase price for the defective product or return the repaired Product or a replacement thereof to Purchaser the purchase price for the defective product or return the insurance prepaid as soon as reasonably possible following receipt of the Product by Empi Inc Empi Inc determines in it
12. the device dry Keep the device away from sunlight Do not wear electrode or controller during Magnetic Resonance Imaging MRI scans as this may result in metal overheating and causing skin burns in the area of the patch For prescription only ETL CLASSIFIED Intertek ETL Classified US 9900900 Electronic Testing Lab indicates product meets US safety standards This device complies with UL STD 6060101 Conforms to UL Council Directive 2002 96 EC concerning Waste Electrical and Electronic Equipment WEEE Indicates a requirement not to dispose of WEEE as municipal waste Contact your local distributor for MS information regarding disposal of the unit and accessories 70 C 158 F TRANSPORT amp STORAGE 40 C 40 F Date of Manufacture Reference number part number Lead wires comply with the Performance Standard for electrode lead wires 21 CFR part 898 E Empi Elite User s Manual Operating Instructions You are strongly advised to carefully read Warnings and Cautions section pages 3 6 described at the beginning of this manual prior to using your stimulator Description of the stimulator A On Off button B i button used to increase stimulation energies in several channels simultaneously C 1 2 3 4 buttons for each of the D 4 stimulation channels D Sockets for the 4 leadwires E Channel 1 blue Channel 2 green Channel 3 yellow Channel 4 red E
13. SEmpi Empi Elite COMPLETE ELECTROTHERAPY SYSTEM USER S MANUAL Read this manual carefully before operating the Elite Q DIO GLOBAL MOTION IS MEDICINE Visit us at www djoglobal com Table of Contents Foreword z nui a a a ita ia ai Gee nas Warnings and Cautions e Contraindicationss uv IRE a a Wa a a a a a E Warnings s s a anca ria a a XP ae a Precautions s s 3 ox ec yof y om hok e ooh Eo 93 33 e RR f RR deos Precautions when using the electrodes llle Adverse Reactions 22k o RR ERR hakt RR US Where Never to Apply the Electrodes llle Indications for Use ee Intended Use is tito EA eo EDO A QA We ee SK S Wsage QuidelineS iue ceed be mom o cR DEER ERS a we o 1 Choice of the Appropriate Muscle Program less 2 Planning of Stimulation Sessions ee ee 3 Electrode positions ed sooo a Domo RR x xx EO osos m E Ros 4 Stimulation positions sse o o or RR RS 5 Adjusting stimulation energies o e e e How Does Electrostimulation Work lll ln Motor nerve stimulation EMS leer Benefits of electrostimulation 2 le Explantion of Symbols ce Operating Instructions o eee ee ee Description of thestimulator o o e ee Your kit contalns eco ar Rue oe e A Ee C ec ds Gonnectio
14. e Level 6 kV at the contact Electromagnetic Environment Guidance Floors must be wood concrete or ceramic tile ESD 8 kV in air IEC 61000 4 2 8 kV in air If floors are covered with synthetic material the relative humidity should be at a minimum of 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power Not Applicable supply lines System batterypow 1 kV for input ered output lines 1 kV differential Not Applicable mode System batterypow 2 kV common mode ered The quality of the power supply should be that of a typical commercial or hospital environment Surge 1 IEC 61000 4 5 The quality of the power supply should be that of a typical commercial or hospital environment Voltage dips short interruptions lt 5 UT gt 95 dip in Not Applicable UT for 0 5 cycle System batterypow 40 UT 60 dip in ered UT for 5 cycles 70 UT 30 dip in UT for 25 cycles lt 5 UT gt 95 dip in UT for 5 sec The quality of the power supply should be that of a typical commercial or hospital environment If the Empi Elite user requires continuous operation during mains power cuts we recommend that the Empi Elite is powered by a UPS ora battery and voltage variations on power supply input lines IEC 61000 4 11 Power frequency 50 60Hz magnetic field IEC 61000 4 8 Magnetic fields at the mains frequency should be at the level of a representative
15. e if you have one or more of the following medical conditions 1 This device must not be used on persons with cardiac pacemakers defibrillators or other implanted electronic devices Cardiac demand pacemakers Epilepsy Following acute trauma or fracture Critical ischemia of lower limbs Abdominal or inguinal hernia Cancerous lesions Pregnancy do not use on abdominal region 9 Serious arterial circulation problems in lower limbs 10 Serious infectious disease ANOA FW Nm Empi Elite User s Manual 11 Heat or fevers 12 Capillary fragility or capillary damage 13 Sensitivity problems or unable to express yourself Osteosynthesis equipment The presence of osteosynthesis equipment metallic equipment in contact with the bone pins screws plates prostheses etc is not a contraindication for the use of Empi Elite programs The electrical currents of the Empi Elite are specially designed to have no harmful effect on osteosynthesis equipment AN i e Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner The Empi Elite is to be used under the supervision of a physician Never use it in ways other than recommended in the Operating Instructions Chapter of this User Manual The long term effects of chronic electrical stimulation are unknown Stimulation should not be applied Over the carotid sinus nerve particularly
16. g symbol on the display is a black case with START written in it When you start a treatment you will be prompted to increase the stimulation energies This is critical to the success of any program The energy level should be based on the guidance of your prescribing practitioner Program time in minutes and seconds The animated rising convection symbol represents the warming up sequence Session progression scale For details on how it works see the following Program progression Press the On Off A button to place the unit in Pause mode The Empi Elite beeps and the symbols of the four channels flash the four channels are at 000 energy You must increase the stimulation energy so that the stimulation can start To do this press the buttons C1 C2 C3 or C4 for the relevant channels until the desired setting is reached If you want to increase the energy level of all four channels simultaneously press the i button B located below the On Off button A Press the i button twice to increase the levels in the first 3 channels and 3 times to increase the levels in the First 2 channels When you press the i button the associated channels are overlightened in black Empi Elite User s Manual 11 2 Program progression Stimulation actually starts when the stimulation energy has been increased The examples reproduced below show the general rules Session sequences Warm up Rema
17. in patient with a known sensitivity to the carotid sinus reflex Transthoracically across the thoracic cavity or chest in that the introduction of electrical current into the heart may cause cardiac arrhythmias Transcerebally across the brain The effects of stimulation on the brain are unknown On the neck or mouth Contractions and spasms of the muscle of the airway may cause difficulty in breathing or close the airway passages Stimulation on the neck may also have adverse effects on the heart rhythm or blood pressure Over swollen infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc Over or in proximity to cancerous lesions Over pulled muscles Using the stimulator on a previously extended muscle may further full such muscle The higher the stimulation intensity the higher the risk to further over extend such muscle Do not use the Empi Elite on persons with capillary fragility to avoid excessive capillary damage Some persons may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium gel Never carry out an initial stimulation session on a person who is standing The first five minutes of stimulation must always be performed on a person who is sitting or lying down In rare instances people of a nervous disposition may experience a vasovagal reaction This is of psychological origi
18. ining time in minutes and seconds before the end of the program UU om Bud 012 345 678 323 Work period Relaxation Time bar showing contraction time and active rest time is shown only during the work sequence The different energies reached during the contraction phase are shown by a series of black bar graphs Active rest phase energies are shown by checkered bar graphs Active rest phase stimulation energies are automatically set at 5096 of contraction energies These can be modified during the rest phase Once modified they will be totally independent of the contraction energies 3 Program pause mode Press the On Off button A to interrupt the program momentarily To restart simply press the channel 4 C4 button The session will resume at 8096 of the energy levels that were in use prior to the interruption 4 Skip function Endurance amp Strength program only The skip function allows you to end the current program phase and move on to the next one For skip function to work the unit must be in Pause mode Press the On Off button A to return to the previous list of programs Use the channel 3 C3 button to skip a particular sequence of a program When you do this the unit will beep and the cursor will move to the beginning of the next sequence 12 e Use the channel 4 C4 button to resume the program at the chosen sequence NOTE The skip function is not a
19. le red connection A negative pole black connection An electrode must be connected to each pole Note It is possible and normal to have an electrode arrangement that leaves one electrode connection free from a cable Each lead should be connected to the proper electrode positive and negative ends Depending on the characteristics of the current efficiency can be optimized in certain programs When working with a muscle stimulation program program involving muscle contractions it is important to connect the electrode to the positive pole red connection and place the positive electrode on the motor point of the muscle It is crucial to choose the right size electrodes large or small and correctly position these on the muscle group you want to stimulate to ensure the efficiency of the program Therefore always use the size of electrodes shown in the pictures Unless you have other specific instructions by your medical professionals always follow the placement directions in the pictures Empi Elite User s Manual Where necessary look for the best possible position by slowly moving the positive electrode over the muscle until you find the point that will produce the best contraction or the most comfort for you 4 Stimulation positions This position will vary depending on the position of the electrodes the muscle group you wish to stimulate and the program you are using For programs involving powe
20. meters of the various programs muscular work can be directed very precisely towards the desired goal muscular reeducation relaxation of muscle spasm increased blood flow maintaining or increasing range of motion etc Electrical pulse A Excitation Transmission v p E Motor of the excitation nerve Stimulated muscle Elementary Mechanical Response Twitch Benefits of electrostimulation Electrostimulation is a very effective way to make your muscles work e with significant improvement of different muscular qualities e without cardio vascular or mental fatigue e with limited stress on the joints and tendons Electrostimulation thus allows a greater quantity of work by the muscles compared with voluntary activity To be effective this work must involve the greatest possible number of muscular fibers The number of fibers working depends on the stimulation energy The maximum tolerable energy should therefore be used The user controls this aspect of stimulation The higher the stimulation energy the greater the number of muscular fibers that are working and therefore the more significant the progress achieved Explanation of Symbols The following symbols are used either in this user manual on the device packaging or on the device label They may also appear on an accessory Consult the instructions for use Follow instruction for use Type BF applied parts Keep
21. n and is connected with a fear of the muscle stimulation as well as surprise at seeing one of their muscles contract without having intentionally contracted it themselves A vasovagal reaction causes heart to slow down and blood pressure to drop which can make you feel weak and faint If this does occur all that is required is to stop the stimulation and lie down with the legs raised until the feeling of weakness disappears 5 to 10 minutes Never allow movement resulting from muscular contraction during a stimulation session You should always stimulate isometrically this means that the extremities of the limb in which a muscle is being stimulated must be firmly fixed so as to prevent the movement that results from contraction Do not use accessories manufactured by other companies on the Empi Elite DJO LLC is not responsible for any consequence resulting from using products manufactured by other companies The use of other accessories or cables may result in increased emissions or decreased immunity of the Empi Elite The Empi Elite should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary the unit should be observed to verify normal operation in the configuration in which it will be used Do not use the Empi Elite within X meter see EMC table of short wave or microwave devices as this could alter the currents generated by the stimulator If you are in any
22. n on the screen SHOULDER ca MASSAGE ENDURANCE STRENGTH If the battery icon is only 1 3 blackened this means that power is running low At this stage the device should be recharged Don t start a new program when the battery level is below 1 3 as it may not be possible to complete a program If the symbol normally displayed above the channel 4 button C4 is not visible and if the battery icon is flashing this means that the battery is completely discharged It is no longer possible to use the device Recharge it immediately Empi Elite User s Manual 2 Recharging Remove all leadwires from the stimulator before recharging the battery To recharge it use the charger supplied with your device and connect it to the front of the device then plug the charger cable in the charging socket E and then plug the charger to an electrical outlet The charge menu shown below will automatically appear on the screen The duration of the charging operation is shown on the screen a complete charge may take 10 hours The battery icon is animated while the battery is recharging When fully charged the icon will be full and the total time taken to recharge the battery will flash on the screen Simply disconnect the charger Empi Elite will turn off automatically Charging Socket Plugging the charger cable into the charging socket 13 Problems and solutions 1 Open lead detection The Empi Elite beep
23. nS gt s d rs erd Ro Romo oko wok OE E a a BO RES Rx E E 3e AUR 1 Connecting the cables e 2 Connecting the electrodes o oo ln 3 Electrode placement 2 22 222 cz i ed x moe RR RE 4 Connectingthecharger s s s s e sa saso o ee a Preliminary settings i i ra e a ea kG ee REOR a o y dtc 3S 8 POE UR ED UL Selecting abodyarea lt lt ira o aa RR ko my s Selecting a treatment o o o o tts During the stimulation session f s s ra dalra ee 1 Adjusting stimulation energies lll 2 Program progressio de ey ae Se CR ay a Race id Ar M A OUS e 3 Program pause mode 2 2 22 4 Skip function Endurance amp Strength program only a a oa 5 End of program 0 ek aaa Row RR RE E RE E RR RR ee E Esos Empi Elite User s Manual Consumption andrecharging istit aosta o ee ee ala iR 13 Techarge loveless mk Rh A UE A REIR di A 13 2 Recharging 9 s sud Rock xo A RA OR E Ec Rs RU 13 Problems andsolutions ee 14 1 Openleaddetection et ee ee teers 14 2 Stimulation is not producing the usual sensation 14 3 The stimulation effect causes discomfort lll 14 4 Thestimulatorisnotworking e e eo 14 Programs15 Massage 2 2 64 445253 Pow eee Bae eS OES SAE X Xd Sede rcm Wo vL ae 15 Endurance mcs Roperdbenre e y Wow oo xge wo mox cR boROR RoOROR
24. nductive medium The irritation can usually be reduced by using an alternate conductive medium or alternate electrode placement 5 Using the Empi Elite on muscles gradually to avoid over extend muscles A gradual regimen is recommended 6 Use only with the battery charger lead wires and electrodes supplied by Empi Skin irritation may occur in persons with particularly sensitive skin or lacking normal sensitivity Precautions when using the electrodes Only use electrodes supplied by Empi Other electrodes may have electrical properties that are unsuitable for the Empi Elite stimulator Empi Elite User s Manual Do not use electrodes with a surface lt 19 cm as there will be a risk of suffering a burn injury Caution should always be exercised with current densities gt 2mA cm Always turn off the stimulator before moving or removing any electrodes during a session Do not place the electrodes in water Do not apply solvents of any kind to the electrodes For best results wash and clean the skin of any oil and dry it before attaching the electrodes e Attach the electrodes in such a way that their entire surface is in contact with the skin For obvious reasons of hygiene each user must have his or her electrode set Do not use the same electrodes on different people Never use a set of adhesive electrodes for more than 15 sessions as their bonding power deteriorates over time and optimal con
25. on Indications for Use Intended Use The Empi Elite is intended for the following conditions Muscle reeducation Prevention or retardation of disuse atrophy Relaxation of muscle spasms Increasing local blood circulation Maintaining or increasing range of motion Usage Guidelines The use guidelines presented in this section should be considered as general rules For all programs it is recommended that you read the usage information and advice presented in chapter VI Programs carefully IF YOU ARE UNSURE CALL YOUR MEDICAL PROFESSIONAL OR EMPI REPRESENTATIVE BEFORE PROCEEDING o 1 Choice of the Appropriate Muscle Program The choice of a program is determined by the injured body parts or joints The appropriate muscle stimulation programs e g Massage Strength or Endurance and frequency of the program s are determined by the medical professional Consult your medical professional to be sure to understand the Empi Elite 2 Planning of Stimulation Sessions Your Medical Professional will instruct you on a protocol intensity electrode placement number of times per day and per week to follow in order to use the Empi Elite most effectively 3 Electrode positions For optimal results use the electrode positions recommended by your medical professional Please also refer to the pictures and pictograms shown in back of the manual Each stimulation cable has two poles A positive po
26. pport Toll free 800 862 2343 Empi Inc 205 Hwy 22 East Clear Lake SD 57226 USA Empi Elite User s Manual Programs The Empi Elite provides three muscle stimulation training programs which correspond to target muscle groups Shoulder Knee Hip of the patient intended for use during rehabilitation These rehabilitation programs are Massage Endurance e Strength You must consult your medical professional for the proper protocol to follow Massage The Empi Elite Massage program is intended to relax muscle spasms reduce the muscular spasticity and enable the target muscles to contract effectively This program is recommended for use as needed before and after surgery as prescribed by your medical professionals MASSAGE 21 minutes Reference name used in Europe Relaxing 1st sequence 2nd sequence 3rd sequence Frequency 7 Hz 5 Hz 3 Hz Time 7 min 7 min 7 min Pulse width Shoulder 250 us Hip 300 us Knee 380 us Endurance The Empi Elite Endurance program focuses on activating an average medium working level on muscle fibers This working level is maintained over an extended period of time 25 minutes per session The Endurance program particularly activates the aerobic metabolism of the fibers during the stimulation session The purpose is to increase the time for the muscle to maintain a medium level of working power or the average power level for extended periods of
27. requency where P is the maximum output power of the transmitter in watts W as set by the transmitter manufacturer NOTE 1 At 80 MHz and at 800 MHz the spacing for high frequency amplitude is applied NOTE 2 These guidelines may not be appropriate for some situations Electromagnetic wave propagation is modified by absorption and reflection due to buildings objects and persons Empi Elite User s Manual 21 Limited Warranty While in the opinion of Empi Inc the use of the Empi Elite the Product has met with some success Empi Inc makes no warranties to the purchaser as to the effectiveness of the product A Empi Inc warrants to the initial Purchaser Purchaser and to no other person that the Product with the exclusion of accessories such as chargers rechargeable batteries electrodes lead wires tape adhesive patches and electrode cream and the component parts thereof distributed or manufactured by Empi Inc shall be free from defects in the workmanship and materials for three years from the initial date of purchase from Empi Inc the Warranty Period B Accessories including but not limited to chargers rechargeable batteries electrodes lead wires tape adhesive patches and electrode cream are excluded from the Warranty and sold AS IS because their structure is such that they may be easily damaged before or during use Limitation of Liabilities and Disclaimer or Warranties A Empi Inc s
28. ressure 700 hPa to 1060 hPa Conditions of use Temperature 0 C 32 F to 40 C 104 F e Relative humidity 30 to 75 e Atmospheric pressure 700 hPa to 1060 hPa e Do not use in areas where there is a risk of explosion Empi Elite User s Manual Disposal To prevent the creation of waste electrical and electronic equipment WEEE and to reduce the amount of waste to be disposed we encourage reuse recycle and other forms of recovery The wheelie bin pictogram with the bar through it means that the equipment cannot be discarded with household refuse but that it must be collected selectively The equipment must be delivered to a suitable collection point for treatment By doing so you will be contributing to the safeguarding natural resources and health Batteries must be disposed of in compliance with relevant national or state regulatory requirements Standards The Empi Elite complies with current medical standards To guarantee your safety the Empi Elite has been designed manufactured and distributed in compliance with the requirements of IEC 60601 1 standard on general safety requirements for electro medical devices IEC 60601 1 2 standard on electromagnetic compatibility and the IEC 60601 2 10 standard on particular safety requirements for nerve and muscle stimulators Patents The Empi Elite incorporates several innovations with patents pending Normalized symbols Empi Elite User s Manual
29. rful muscular contractions for example Strength and Endurance programs the muscle should always be stimulated in an isometric fashion It is therefore strongly recommended that you must keep the knee bent at 90 degrees and the foot flat on the ground In this way you provide maximum resistance to the movement and prevent any shortening of the muscle during the contraction which could create cramp pains and serious stiffness after the session For the MASSAGE program which does not involve powerful muscular contractions position yourself as comfortably as possible 5 Adjusting stimulation energies In a stimulated muscle the number of recruited fibers depends on the stimulation energy For programs involving powerful muscular contractions for example Strength and Endurance programs you must therefore use the stimulation energy level programmed up to 999 at the limit that you can endure in order to recruit the maximum number of fibers Always consult your medical professionals Empi Elite User s Manual How Does Electrostimulation Work The principle of electrostimulation is to stimulate nerve fibers by means of electrical impulses transmitted by electrodes The electrical pulses generated by Empi Elite stimulators are high quality pulses offering safety comfort and efficiency which stimulate motor points of the target muscle resulting in a muscle contraction The benefits obtained depend on the stimulation
30. romagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Empi Elite are used exceeds the applicable RF compliance level above the Empi Elite should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the Empi Elite b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m 20 Empi Elite User s Manual TABLE 4 RECOMMENDED SPACING BETWEEN A PORTABLE AND MOBILE COMMUNICATION APPLIANCE AND THE EMPI ELITE The Empi Elite is designed for use in an electromagnetic environment in which radiated RF waves are controlled The buyer or user of the Empi Elite can contribute to preventing electromagnetic interference by maintaining a minimum distance between RF portable and mobile communication appliances transmitters and the Empi Elite according to the table of recommendations below and according to the maximum output power of the telecommunication appliance Maximum Transmitter Spacing according to the frequency of the Output Power transmitter m CISPR 11 From 150 From 80 From 800 MHz kHz to 80 MHz to to 2 5 GHz MHz 800 MHz d 2 3JP In the case of whose maximum output power is not shown in the table above the recommended spacing of d meters m can be calculated using the appropriate equation for the transmitter f
31. s and alternatively displays the symbol of a pair of electrodes and an arrow pointing to the channel where a problem has been detected In the example below the stimulator has detected an error in channel 1 Check whether the lead wire is plugged into this channel completely Check whether electrodes are connected to each lead Check whether electrodes are placed on your body correctly Check whether the electrodes are old worn and or the contact is poor try using new electrodes Try using the stimulation cable on a different channel If the cable is still showing a fault replace it 2 Stimulation is not producing the usual sensation Check that all the settings are correct and ensure the electrodes are positioned properly Change the positioning of the electrodes slightly 14 3 The stimulation effect causes discomfort The electrodes are beginning to lose adhesion and no longer provide good contact on the skin The electrodes are worn and need to be replaced Change the positioning of the electrodes slightly 4 The stimulator is not working If an error screen appears while you are using the device write down the menus code and error number in the example below menus code 1 0 and error number 1 0 0 and contact the nearest authorized customer support service or Elite Product Support PLEASE CONTACT YOUR AGENT NER os MENUS 1 0 ERROR 1 0 0 Product Su
32. s sole reasonable discretion that the Product does not contain defective workmanship or materials Empi Inc will return the Product to the Purchaser freight and insurance billed to the Purchaser B This Warranty is voided immediately as to any Product which has been repaired or modified by any person other than authorized employees or agents of Empi Inc or which has been subjected to misuse abuse neglect damage in transit accident or negligence 22 Empi Elite User s Manual C Except as provided in paragraph A the product is being sold on an as is basis all accessories are sold as is and the entire risk as to the quality and performance of the product is with purchaser The warranty provided in paragraph A is intended solely for the benefit of the initial purchaser and Empi Inc disclaims all other warranties express or implied including but not limited to any implied warranties of merchantability and fitness for a particular purpose provided however that notwithstanding the foregoing sentence in the event an implied warranty is determined to exist the period for performance by Empi Inc thereunder shall be limited to the lifetime of the initial purchaser No employee representative or agent of Empi Inc has any authority to bind Empi Inc to any affirmation representation or warranty except as stated in this written warranty policy D Empi Inc shall not be liable to any person for any direct indirect special incidental
33. site located in a typical commercial or hospital environment Portable and mobile RF commu nications equipment should be used no closer to any part of the Empi Elite including cables than the recommended separa tion distance calculated from the equation applicable to the frequency of the transmitter Empi Elite User s Manual 19 Recommended separation distance Conducted RF 3 Vrms Not Applicable IEC 61000 4 6 150 kHz to 80 MHz Radiated RF d 12 P 80 MHz to 800 MHz IEC 61000 4 3 d 2 3JP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol i NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people A Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the elect
34. tact is very important for both user comfort and efficacy Some people with very sensitive skin may experience redness under the electrodes after a session Generally this redness is completely harmless and disappears after some time from a few minutes up to a few hours Never start another stimulation session in the same area if the redness is still visible Adverse Reactions 1 Skin irritation and burns beneath the electrodes have been reported with the use of stimulation electrodes applied to the skin Note Some people with very sensitive skin may experience redness under the electrodes after a session Generally this redness is completely harmless and disappears after some time from a few minutes up to a few hours Consult your doctors if the redness of the skin does not disappear after a few hours 2 Headache and other painful sensations have been reported during or following the application of electrical stimulation applied to the head face and near the eyes Where Never to Apply the Electrodes e On the head or any area of the wi o On the neck or any area of the throat On both sides of the thorax simultaneously front and back sides or lateral sides Ch e On or in the vicinity of skin A lesions or eruptions of any RS kind wounds swelling burns YN irritation eczema etc e Over the abdominal region during menstruation periods e On skin areas lacking of normal sensati
35. time ENDURANCE 25 minutes Reference name used in Europe Disuse atrophy level1 Warm up Contraction Active rest Final recovery phase 74 cycles Frequency 6 Hz 35 Hz 4 Hz 3 Hz Duration of ramp up 15s 15s 0 5s 15s Duration of phase 2 min 6s 7s 3 min Duration of ramp down 2s 0 75s 0 5s 3s Pulse width Shoulder 2 50 us Hip 300 us Knee 380 us Empi Elite User s Manual 15 Strength The Empi Elite Strength program imposes a high and instantaneous power working level on muscle fibers These contractions are separated by long periods of rest The result is an average medium power working level 20 minutes This program is intended to increase the maximum strength of muscle isometric contraction This program targets the muscle fibers that are typically afflicted with immediate atrophy after injury or and surgery This program is recommended for use only after having completed at least 10 sessions of the previous Endurance program as prescribed by your medical professionals STRENGTH 20 minutes Reference name used in Europe Reinforcement level1 Warm up Contraction Active rest Final recovery phase 52 cycles Frequency 6 Hz 75 Hz 4 Hz 3 Hz Duration of ramp up 1 5s 15s 0 5s 15s Duration of phase 2 min 4s 10s 3 min Duration of ramp down 2s 0 75s 0 5s 3s Pulse width Shoulder 250 us Hip 300 us Knee 380 us
36. uld check that the Empi Elite works properly in the chosen configuration Specifications TABLE 1 RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER CONCERNING ELECTROMAGNETIC EMISSIONS The Empi Elite is intended for use in the electromagnetic environment specified below The customer or the user of the Empi Elite should assure that it is used in such an environment Emission Tests CISPR 11 RF Emissions CISPR 11 RF Emissions Harmonic Emissions IEC 61000 3 2 Voltage Fluctuations Emission Oscillations IEC 61000 3 3 18 Compliance Group 1 Class B Electromagnetic Environment Guidance The Empi Elite uses RF energy only for its internal opera tion Consequently its RF emis sions are very low and are unlikely to interfere with any adjacent electrical device The Empi Elite is suitable for use in any establishment including a private dwelling and a place connected directly to the low voltage mains supply which powers residential buildings Not Applicable Not Applicable Empi Elite User s Manual TABLES 2 8 3 RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER ELECTROMAGNETIC IMMUNITY The Empi Elite is designed for use in the electromagnetic environment stipulated below The customer or the user of the Empi Elite must ensure that it is used in this recommended environment Immunity Test Electrostatic discharge IEC 60601 Test Level 6 kV at the contact Complianc
37. vailable for all programs lt should also be noted that this function can reduce the effectiveness of a program if used to shorten the work sequence excessively 5 End of program At the end of each session a small flag will be displayed on the screen and a short melody will be played To switch off the stimulator press the On Off button A e gg oo py LM 000 ooa boo poA For the ENDURANCE and STRENGTH programs when a muscle work has been performed the end session screen will also provide the average value AVG of the stimulation level and the maximum MAX level reached for each channels The skip Empi Elite User s Manual icon will also be monitored if the program has been completed by using the skip function lt lt TS i m MAR 607 866 Baa BBB Mil ooa poo Bee Bag Consumption and recharging Never recharge the stimulator without first disconnecting the leadwires Always use the charger supplied by Empi to recharge the batteries The Empi Elite runs on rechargeable batteries The battery s life depends on the programs and stimulation energy used We strongly recommend you fully charge the battery before using your Empi Elite for the first time as this will improve the length of usable time and life span If you do not use your device for a long period of time please recharge the battery regularly 1 Charge level The charge state of the battery is indicated by a small battery ico
38. y in the position shown on the pictogram Always follow your doctor s recommenda tions for the electrode placement and for the intensities applied This section is located at the end of this manual 4 Connecting the charger Empi Elite has rechargeable batteries and can be recharged Remove any leadwires connected to the stimulator before recharging it To recharge them use the charger supplied with your device and connect it to the front of the device then plug the charger in a socket Never try to plug leadwires simultaneously with the charger We strongly recommend you fully charge the battery before using it for the first time as this will improve its life span 10 Preliminary settings Before using the unit for the first time you should select the working language of the device which is displayed on the options screen Follow the instructions below Afterwards for the greatest comfort the Empi Elite offers you a number of setting options operating language selection display contrast setting adjusting backlighting and volume setting To change any of these settings bring up the options screen by holding the On Off button on the left of the stimulator for a few seconds when the device is being turned on SETTINGS Lem ESPA OL FRANCAIS Ch E a 65m Ea A Press the On Off button to turn on your stimulator or to bring up the option screen It is now ready for use with the settings you selected B