1. MELAG Operating Manual For Autoclaves of the Pro Class Vacuklav 23 B Vacuklav 31 B as of software version 4 06 Please read this Operating Manual before you start operation of the autoclave The instructions contain important safety precautions Make sure to keep the Operating Manual near to the autoclave The Operating Manual is part of the product Subject to technical changes 0535 MELAG Medizintechnik oHG E mail info melag de GeneststraBe 6 10 www melag de 10829 Berlin Rev 2 11 2537 Germany Doc KBA 2 GB 23B 31B _v2 doc Instruction Mains cable and power plug Setting up installation commissioning Preparation and sterilization of textiles and instruments Program termination Removal of the items to be sterilized Maintenance Errors Application area Sterilization tasks Safety instructions To operate the autoclave pay attention to the following safety instructions given below and in the individual chapters of this Operating Manual Do not sterilize any liquids with this autoclave Never damage or change the mains cable or power plug Never operate the autoclave if the mains cable or power plug are damaged Never pull on the mains cable to take it out of the socket Always take hold of the power plug itself Only have the autoclave set up installed and commissioned by people authorised by MELAG Only operate the autoclave in areas which are not subject to explosi
2. ambient pressure If the program was successfully finished an appropriate notification appears on the display Manually terminate program You can manually terminate a running program in any of the phases with the KEY S If you terminate a program before the beginning of drying then the items to be sterilized are still unsterile In the drying phase the expired drying time is shown in alternation with the message Immed Removal Press Stop is displayed By pressing the KEY S the program is cancelled during the drying The termination is confirmed with Drying stopped on the display MELAG If a printer or another output medium is connected to the autoclave and the Immediate output is Set to yes a reference to Drying stopped Is issued to the logging output Operating Manual Vacuklav 23 B 31 B Remove sterilized items After the request Quit with button at the end of a program press the KEY After unlocking the door you can open the door and remove the sterilized items m Check the wrapping of the sterilized objects for damage when removing them from the autoclave Danger lf a wrapping is damaged wrap the object again and re sterilize it Non observance can result in unsterile instruments This endangers the health of the patient and the practice team If you remove the sterilized items from the autoclave directly after the end of the program it can happen that slight quantities of condensate moisture a
3. filled out by the person responsible and a copy sent to MELAG Prepare items to be sterilized A significant prerequisite for safe disinfection and sterilization of sterilizing materials is the appropriate preparation i e cleaning and maintenance of the sterilizing materials according to the manufacturer s instructions Furthermore the materials cleaning agents and processing procedure employed are of significance lt s Notice If possible sterilize textiles and instruments separately from each other in separate sterilization containers or sterilization packages This way you achieve improved drying results Load the autoclave Only if the autoclave is properly loaded can the sterilization be effective and the drying deliver good results lt s Notice Use perforated trays such as the trays from MELAG Only this way can condensate run off If you use closed materials or shell halves to bear the items to be sterilized you risk bad drying results Please note that the use of tray insertions made of paper can perhaps likewise lead to worse drying results Employ only packaging materials and systems sterile barrier systems which satisfy the DIN EN ISO 11607 1 Standard e g Standard tray cassettes or transparently wrapped Warning sterilization packages Closed sterilization containers Preferably use sterilization containers made of aluminium Aluminium conducts heat well and thus accelerates drying Stack sterilizat
4. time to time At the given point in time the autoclave outputs an appropriate message Fill up the storage tank Fill up the internal storage tank with fresh purified feed water of the respective quality You can use the removable funnel insert to conveniently fill fresh feed water into the right chamber of the storage tank up to the MAX mark The internal storage tank holds maximum 5 litres This quantity of feed water suffices for up to 7 sterilizations Emptying both chambers of the storage tank Attach the drain hose on a quick fitting connection left of the wastewater to the right of the feed water until this perceptibly engages Drain off the water into a container with at least 5 litres holding capacity To remove the drain again press the grey release knob at the quick fitting connection The hose detaches itself automatically from the connection Warning spring tension Please stand to the side of the connection when removing the hose MELAG Operating Manual Vacuklav 23 B 31 B Prerequisites for initial start up Trays and accessories must be removed from the chamber after switching on the autoclave and before initial Start up The feed water supply must be secured The power supply of the autoclave must be assured Record of installation and setting up mS NOTICE As documentation of proper setting up installation and setting up as well as for your warranty claim the record of installation must be
5. ed Door seal is sticking Switch the autoclave on and pull vigorously at the door to the sealing surface After pauses carry out the tests described below depending on the pause duration Testing in daily operation Vacuumtest Weekly in routine operation at initial start up after pauses longer than two weeks and in case of error Select with the KEY P the program Vacuum test and start by pressing the KEY S The chamber is evacuated until the pressure for the vacuum test is achieved An equilibration time of five minutes and a measurement time of ten minutes follow The rise in pressure in the chamber is measured within the measurement time The evacuation pressure and the equilibration time or measurement time are shown on the display If the leak rate rises too high that means 1 3 mbar there is shown an appropriate notification on the display Bowie amp Dick test The Bowie amp Dick test serves the verification of the steam penetration of porous materials such as textiles For this purpose it is especially recommended when sterilizing larger quantities of textiles Select with the KEY P the program Bowie amp Dick test and start with the KEY S Helix test body system MELAcontrol Pro The test body system MELAcontrof PRO is an indicator and batch monitoring system It consists of a test specimen the helix and an indicator strip If you sterilize instruments of the category Critical B the MELAcontroP PRO tes
6. ion containers Stack on top of each other if possible only sterilization containers of the same size and with respectively identical spacer feet where the condensate can run off on the side at the walls Soft sterilization wrapping Soft sterilization packages can be sterilized both in sterilization containers on trays or standing when using a package holder do not place several soft sterilization wrappings flat on top of each other on the trays or in a sterilization containers Multiple wrapping The autoclave functions on the fractionated pre vacuum method This permits the use of multiple wrapping Mixed loads For sterilizing mixed loads observe the following instructions Always place textiles on the top Sterilization containers at the bottom Unwrapped instruments at the bottom Transparently wrapped sterilization packages and paper packaging on the top exception in combination with textiles on the bottom The heaviest loads at the bottom Operating Manual Vacuklav 23 B 31 B Stack transparently wrapped sterilization packages if possible edgewise in such a way that alternately paper side lies on paper side and foil side on foil side if this is not possible then with the paper side facing downwards Select program Select by rotating between the initial state and the desired program with the program selection button P P Programs Wrapping Especially Sterilization Operating Drying Load suitable for a
7. on hazards The electric connections and connections for feed water and used water may only be completed by a specialist technician Follow the instructions of the textile and instrument manufacturers for preparing and sterilizing textiles and instruments Only use wrapping materials and systems which are suitable for steam sterilization according to the manufacturer s information Take heed that when opening the door after terminating a running program hot water steam can escape out of the chamber depending on when the program was terminated Depending on the time of the program termination the load might not be sterile Observe the clear instructions shown on the display of the autoclave If necessary sterilize the affected objects after rewrapping Never open the door by force Use a tray lifter to remove the tray Never touch the sterilized items the chamber or the door with unprotected hands The parts are hot Check the wrapping of the sterilized objects for damage when removing them from the autoclave If a wrapping is damaged wrap the object again and re sterilize it Have the maintenance done only by authorized persons If repeated error messages occur while operating the autoclave turn the device off and notify your specialist dealer Only have the autoclave repaired by authorised persons Proper use The autoclave is intended for use in the medical field e g in clinics and in physician and dental practices Acco
8. rding to DIN EN 13060 this autoclave is considered a Class B sterilizer It is Suited as a universal autoclave for demanding sterilization tasks For instance you can sterilize narrow bore instruments and transmission instruments wrapped or unwrapped and larger quantities of textiles Operating Manual Vacuklav 23 B 31 B Device views Front side Back side one Od 1 2 a l 2 3 Wu 4 Front side below with opened door 5 7 18 wets 17 1 Operating and display field 11 Emergency overflow hose 2 Door pivots opens to the left 12 Cooler 3 Sliding closure grip 13 Purified feed water inlet for MELAdem 4 Mains switch fast screw connection for hose 8x1 5 Front foot of the unit adjustable 14 Power cable 6 Tank cap 15 Serial data and printer connection RS232 7 Slot for optional upgrade with the safeguard 16 Device fuses 2x 16A gRL combination EN1717 17 Reset button for overheating protection 8 Sterile filter 18 Connection for emptying the storage tank feed water 9 Safety valve 19 Connection for emptying the storage tank waste water 10 One way drain optional Control panel The control panel consists of a two line alohanumerical LED display and four membrane keys A 2 line LED display for program statu
9. re found on the sterilized items Smaller quantities of water which may be are found on the upper side of the paper bags and clear plastic sterilization packages are unobjectionable if they dry out within 30 minutes after removal from the autoclave Storage of sterilized items The maximum storage life depends on the packaging and the storage conditions For sterilized items which have been wrapped in compliance with the standards dust protected storage is assumed this time period is up to six months Sealed against dust e g in a closed instrument cabinet Protected from damage on smooth surfaces Protected from excessive fluctuations in temperature Protected from moisture e g alcohol disinfectant Storage duration in accordance with the type of package wrapping Batch documentation The batch documentation is indispensable as documentation for the successfully completed sterilization process and as an obligatory measure of quality assurance The capacity of the internal storage suffices for about 40 logs If the internal log memory is full when starting the next program the respectively oldest log is automatically overwritten You can store the logs of the completed programs on the following output media and archive them accordingly Log printer MELAprinf 42 MELAflash CF Card Writer on CF card MELAnet Box Computer e g with software MELA view MELAwin modem Please learn from the operating manual of the re
10. s display and parameter display B Function key and D to select set and display special functions press KEY Program Start Stop to unlock door A 14 27 12 0 02bar 25 C C Chamber pressure bar and temperature C D Time hh min ss B C E F E Program selection button P to select the sterilization programs test programs and select or set options submenus of the special function F Start Stop button S to start programs terminate programs Operating Manual Vacuklav 23 B 31 B Conditions for setting up installing and starting up m Only have the autoclave set up installed and started up by people authorized by MELAG Set up the autoclave in a dry and dustproof location The air humidity should be 30 to 60 and the ambient temperature 16 to 26 C Space requirements aa Vacuklav 23 B Vacuklav 31 B The space requirements for the autoclave corresponds to its dimensions plus at least 5 cm on the side and 10 cm at the back side For a fault free operation the autoclave must be set up with a distinct tilt back Starting from a horizontal position check by using a water level at the chamber flange the front feet of the unit must be screwed out by about five ro
11. spective device how to connect the additional devices and which settings you take at the autoclave Set date and time Date and time of the autoclave must be correctly set for proper batch documentation Observe the time re settings in autumn and spring daylight saving time since this is not reset automatically Select the SETUP Menu Function by short simultaneous pressing of the KEYS and Navigate with the KEY up to the display Function Date Time are displayed Press the KEY P in order to confirm Select between the setting possibilities with the KEY or hour minute second day month or year n order to adjust the hour parameter press the KEY P for the confirmation The value can be increased or reduced with the KEYS and n order to store the value confirm with the Key P Press KEY S 2x in order to leave the SETUP Menu Operating Manual Vacuklav 23 B 31 B Pause times Pause times between individual programs are not required since the sterilization chamber is permanently maintained at the required temperature Longer operating pauses After operating pauses which last longer than two weeks carry out a vacuum test and thereafter an empty sterilization with the Quick Program B The following situations can occur after longer pauses conductivity too high Bad feed water Change the feed water or the mixed bed resin pellets for a MELAdem unit Door cannot be open
12. t specimen should be included as a batch control for every sterilization cycle Display water quality You can at all times have the display of water quality shown on the switched on the autoclave also during a running program To do this keep the KEY pressed Errors Error messages are displayed if safe operation or sterilization security is not guaranteed These can appear on the display shortly after switching on the autoclave or while a program Is running Follow the operating procedures shown in the connection with a warning text or an error message on the display After that you can continue the operation of the autoclave Only if the messages appear repeatedly turn to your specialist dealer or the authorised MELAG customer service centre in your vicinity
13. t time 23 B 31 B Single and Mixed loads 134 C 30 min 20 min 5 kg multiple long narrow bore wrapping hollow bodies Only Simple solid 134 C 15 min 5 min 5kg unwrapped instruments simple no textiles hollow bodies Single wrapped Long narrow bore 134 C 28 min 10 min 1 5 kg instruments hollow bodies no textiles Single and Larger quantities of 121 C 45 min 20 min Textiles multiple textiles Ewe Thermo instable goods e g plastic Thermo instable rubber articles Goods 5kg Prion Program Single and Instruments where aj 134 C 45 min 20 min multiple danger of infection wrapping by a virus Is suspected without drying full load in the Vacuklav 23 B and Vacuklav 31 B 5 kg and depending on the load and installation conditions such as the supply voltage Please notice your national regulations of your country Start program Close the door with light pressure against the chamber flange and simultaneously press down the sliding closure grip Press the KEY S in order to start the program Select additional drying Press KEY S and KEY simultaneously at the start of a program If the Quick Program S is started the warning text Warning only unwrapped instruments is shown on the display If the load exclusively contains unwrapped instruments press the KEY S once again to confirm and start the program At the end of a program run the chamber pressure is adapted to the
14. tations for the Vacuklav 23 B and three rotations for the Vacuklav 31 B Connections required Electronical connection Electric circuit with 220 240 V and 50 60 Hz 16 A separate fuse protection Protection from leakage current 30 mA One way drain In order to attach the autoclave to the effluent pipe you can order an upgrade set MELAG Art No 26695 from MELAG The waste water is collected in the storage tank waste water tank and flows through the one way outflow hose into the flush siphon using the MELAG double chamber waste water trap Art No 26635 The drain must be located at least 30 cm beneath the autoclave and installed dip free with continuous descent lt Notice In order to prevent water damage MELAG recommends the use of a leak monitor e g water stop valve from MELAG Art No 01056 Supply with feed water The autoclave works in the gentle yet effective one steam way water system This means that it uses fresh purified feed water for every sterilization run for the steam generation For this purpose the autoclave gets the feed water either from the internal storage tank which the practice team has refilled e g with distilled water from the MELAdesf 65 Or it gets the feed water fully automatically from the water treatment unit MELAdem 40 or MELAdem 47 Only use feed water according to VDE 0510 as feed water If you employ the internal storage tank for the feed water supply you must fill it manually from