1. Fivela traseira Correias traseiras 2b J y Mirage Swift Il NASAL PILLOWS SYSTEM Back straps Sangles arri res Correas posteriores Illustrations Sheet Fiche d illustrations Hoja de Figuras Folheto de Ilustra es B Correias traseiras Top buckle Boucle sup rieure Presilla superior Fivela superior Note For complete instructions read the text section in conjunction with the illustrations on this sheet Remarque pour les instructions compl tes veuillez lire le texte du guide avec cette fiche D d illustrations Nota Para tener instrucciones completas lea el texto en conjunto con las figuras de esta hoja Nota Para obter as instru es completas leia a sec o de texto em conjun o com as Top straps Sangles sup rieures Correas superiores ilustra es deste folheto Top straps Sangles sup rieures Correas superiores Correias superiores Correias superiores PM PPWS Y 8007 Component of 60893 2 RESMED Mirage Swift II NASAL PILLOWS SYSTEM Thank you for choosing ResMed s Mirage Swift II Nasal Pillows System The nasal pillows system is light flexible and quiet It is designed to minimize contact with your face thus ensuring that you feel comfortable during therapy This user guide provides you with the information you need for the correct use of your system Intended Use The Mirage Swift ll is intended for multipatient use by adult patients gt 66 lb 30 kg p
2. MA Do not pull on the short tube or the flow generator tubing as they may get damaged To disassemble the Mirage Swift ll for daily cleaning follow steps 1 3 To disassemble the Mirage Swift Il for weekly cleaning follow steps 1 5 1 Detach the short tube assembly To do this hold the elbow and pull it away from the seal ring C 1 Note The short tube assembly cannot be taken apart 2 Remove the cap by pulling it away from the seal ring C 2 3 Detach the headgear ring from the pillow sleeve assembly Simply pull the headgear rings from the ends of the frame C 3 4 Disassemble the pillow sleeve assembly To do this slide off the clip C 4a and then peel the pillow sleeve off the frame C 4b Note When sliding the clip hold it at one end not in the middle 5 The seal rings do not need to be removed from the headgear C 5 Cleaning the Mirage Swift Il in the Home Notes Wash your hands before cleaning the mask Wash the headgear before first time use as the dye may run It is important to always handwash the headgear The headgear may be washed without being disassembled Daily Handwash the pillow sleeve assembly short tube assembly and cap in warm approx 86 F 30 C water using mild soap or detergent Rinse the components well and allow them to air dry out of direct sunlight and heat The system components have been tested using Dove Unscented beauty bar Note Ensure that the vent holes are not
3. Remontage du masque narinaire Mirage Swift ll Montaje del sistema Mirage Swift Il Como voltar a montar a Mirage Swift II Headgear ring Anneau du harnais Elbow Coude Codo Cotovelo Anillo del arn s Anel do arn s Seal rings Bagues d tanch it Juntas An is de veda o Cap Bouchon Tapa Tampa para limpeza ResMed logo faces away from headgear Logo ResMed tourn dans la direction oppos e au harnais El logotipo de ResMed queda en sentido contrario al arn s O logotipo da ResMed est virado para fora do arn s PM POWSON 8007 O PM PSINSY 8007 O N Note The headgear does not need to be taken apart for cleaning Headgear reassembly instructions are provided here as additional information Remarque il n est pas n cessaire de d monter le harnais pour le nettoyage Les instructions de remontage sont fournies titre indicatif Nota El arn s no necesita ser desmontado para su limpieza Las instrucciones para el montaje del arn s se proporcionan nicamente como informaci n adicional RESMED Nota N o necessita de desmontar o arn s antes de proceder limpeza As instru es de montagem do arn s s o oferecidas como informa es adicionais E Reassembling the Headgear Remontage du harnais Montaje del arn s Como voltar a montar o arn s Back straps Sangles arri res Correas posteriores 6 E Back buckle Boucle arri re Presilla posterior
4. and the leak rate Parts of the Mirage Swift Il See section A on the illustrations sheet Fitting the Mirage Swift Il Note Refer to your flow generator manual for details on settings and operational information See section B on the illustrations sheet Put the headgear on your head and place the nasal pillows below your nose Make sure the top strap passes over the top of your head Make sure the ResMed logo on the pillow sleeve faces outward away from your face B 1 Insert the nasal pillows into your nostrils Gently tighten the headgear straps first the back straps B 2a and then the top straps B 2b Alternate the adjustment of each set of straps so that the headgear panels rest between the eye and the ear The pillow sleeve should rest comfortably above your upper lip Adjust the angle of the nasal pillows To do this rotate the pillow sleeve assembly forward or backward so that the nasal pillows sit comfortably inside your nostrils B 3 Tube Retainer optional Use the tube retainer if you prefer to have the short tube assembly secured to the headgear in the up position B 4 The short tube assembly can be attached in four different ways B 5 Note The short tube assembly and cap can be connected to either side of the pillow sleeve See Disassembling for Cleaning and Reassembling the Mirage Swift 11 for instructions on how to remove and attach the short tube assembly and cap Connect
5. 4 JP 1243027 JP 1243025 JP 1242455 JP 1242456 JP 1243026 JP 1243475 JP 1303347 JP 1303348 JP 1312331 JP 1312517 JP 1312518 NZ 405576 NZ 407158 NZ 407159 NZ 407143 NZ 407144 NZ 407145 NZ 407155 NZ 407160 NZ 407161 NZ 407162 NZ 407163 NZ 407164 NZ 408220 NZ 408221 NZ 408222 US D542912 US D545961 US D546441 US D549323 US D556899 US D556900 US D557409 US D557410 US D557800 US D561893 US D562729 US D562976 Other designs pending Mirage Mirage Swift AutoSet CS 2 and VPAP Adapt SV are trademarks of ResMed Ltd and Mirage Swift AutoSet CS and VPAP are registered in U S Patent and Trademark Office Dove is a trademark of Unilever 2008 ResMed Ltd 60893 Eng 2 08 09 pe A Parts of the Mirage Swift Il Composants du Mirage Swift II Partes del Mirage Swift II Componentes do Mirage Swift II Nasal pillows Masque narinaire Almohadillas nasales Almofadas nasais Pillow sleeve Coussin Manguito de la almohadilla Manga da almofada Vent holes Orifices de ventilation Orificios de ventilaci n Respiradouros Elbow Coude Codo Cotovelo Rotate pillow sleeve assembly for comfort and seal Faites Frame Support pivoter le corps du masque pour obtenir un ajustement confortable et Armaz n Arma o Tube retainer optional Bague de fixation du ResMed logo fa es outward Logo Remed tourne vers Adjust back straps Ajustez les
6. RESMED sem Mirage Swift II NASAL PILLOWS SYSTEM User Guide English ResMed Ltd Manufacturer 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp US Designated Agent 14040 Danielson Street Poway CA 92064 6857 USA ResMed UK Ltd EU Authorized Representative 96 Milton Park Abingdon Oxfordshire 0X14 4RY UK ResMed Offices Australia Austria Belgium Brazil China Finland France Germany Greece Hong Kong India Ireland Italy Japan Malaysia Mexico Netherlands New Zealand Norway Portugal Singapore South Africa Spain Sweden Switzerland UK USA see www resmed com for contact details Mirage Swift Il Nasal Pillows System Protected by patents AU 785376 DE 20321468 4 DE 20321469 2 DE 20321470 6 DE 20321471 4 DE 20321472 2 EP 1314445 NZ 562417 NZ 562418 NZ 562419 NZ 553756 NZ 553757 NZ 553822 NZ 553824 NZ 553825 US 6581594 US 6823865 US 7159587 US 7318437 Other patents pending Protected by design registrations AU 301936 AU 301937 AU 301938 AU 301939 AU 301940 AU 301941 AU 301942 AU 301943 AU 302271 AU 302804 AU 311482 AU 311483 AU 311484 AU 311485 CN 200430102460 2 CN 200430102461 7 CN 200430102462 1 CN 200430102463 6 CN 200430102464 0 CN 200430102465 5 CN 200430102466 X CN 200430102467 4 CN 200430102468 9 CN 200630138503 1 CN 200630138504 6 CN 200730158310 7 EP 259056 EU 604632 HK 0412683 5 JP 1243021 JP 1243024 JP 1243272 JP 1243273 JP 124327
7. See section D on the illustrations sheet 1 Make sure both seal rings are properly fitted to the headgear rings D 1 English 2 Assemble the pillow sleeve assembly To do this a Wrap the pillow sleeve around the frame Then insert the pillow sleeve edges into the groove of the frame D 2a b Hold the clip at one end and then slide it onto the frame to secure the pillow sleeve in place D 2b Note If the clip does not slide easily check that the pillow sleeve edge is fitted flush to the groove of the frame 3 Attach the headgear ring to the pillow sleeve assembly To do this firmly push the headgear rings onto the ends of the frame D 3 When attaching the pillow sleeve assembly make sure the headgear panels are facing out the nasal pillows are pointing upwards the ResMed logo faces out away from the headgear Make sure the pillow sleeve assembly is properly fitted to the headgear before going on to the next step 4 Attach the short tube assembly to your preferred side of the pillow sleeve assembly To do this push the elbow into the seal ring until it clicks into place D 4 5 Insert the cap into the other seal ring until it clicks into place D 5 6 The assembled system should look like this D 6 Make sure the ResMed logo on the pillow sleeve faces out away from the headgear Additional Information Reassembling the Headgear See section E on the illustrations sheet If taken apart the headgear has four p
8. arts A Left strap B Right strap C Back buckle D Top buckle 1 Lay out the four parts of the headgear as shown in E 1 The panels must face up Note Before threading the straps make sure that the buckles are positioned as shown in E 2 and E 3 2 Thread the back straps through the back buckle part C as shown in E 2a and E 2b 3 Thread the top straps through the top buckle part D as shown in E 3a and E 3b Disposal The Mirage Swift ll Nasal Pillows System does not contain any hazardous substances and may be disposed of with your normal household refuse 6 Technical Specifications Pressure Flow Curve 60 Pressure Flow 5 cm H20 L min 5 40 4 20 E 8 29 3 12 37 E 16 3 5 o E gt 20 49 0 Mask Pressure cm H 0 Note As a result of manufacturing variations the vent flow may vary from the values listed in the table Dead Space Information Dead space is the empty volume of the pillow sleeve and short tube assembly up to the swivel The dead space of the Mirage Swift II for all pillow sleeve sizes is 91 mL Standards CE designation in accordance with EC directive 93 42 EEC class lla EN ISO 14971 ISO 5356 1 ISO 10993 1 and EN ISO 17510 2 Therapy Pressure 4 to 20 cm H20 Resistance Drop in measured pressure between the swivel and pillow sleeve average for three pillow sleeve sizes at 50 L min 0 4 cm H20 at 100 L min 1 6 cm H20 English Symb
9. blocked Use a soft bristle brush to clean the vent Weekly Handwash the headgear tube retainer and all components in warm approx 86 F 30 C water using mild soap or detergent Rinse the components well and allow them to air dry out of direct sunlight and heat a AP CAUTION Do not use solutions containing bleach chlorine alcohol aromatics moisturizers antibacterial agents or scented oils to clean any part of the system or air tubing These solutions may cause damage and reduce the life of the product Exposing any part of the system or tubing to direct sunlight or heat may cause deterioration If any change in performance or visible deterioration of a component is apparent cracking crazing tears etc the component should be discarded and replaced Reprocessing the Mirage Swift Il between Patients The Nasal Pillows System should be reprocessed when used between patients Cleaning disinfection and sterilization instructions are available from the ResMed website www resmed com masks sterilization americas If you do not have Internet access please contact your ResMed representative Reassembling the Mirage Swift Il A CAUTION Components of the Mirage Swift II pillow sleeve assembly pillow sleeve frame and clip see section A on the illustration sheet are not compatible for use with Mirage Swift components Combining these components from the two systems may result in rebreathing of exhaled air
10. irage Swift II immediately The exhaust vent flow is lower at lower system pressures see section Technical Specifications for Pressure Flow Curve details If the system pressure during expiration is too low there may not be enough exhaust vent flow to completely flush the air delivery hose of exhaled air before the start of the next inspiration Discontinue using the Mirage Swift II if you frequently experience nasal irritation and consult your physician The Mirage Swift Il must be cleaned and disinfected sterilized if it is used between patients See section Reprocessing the Mirage Swift Il between Patients for details Do not modify the components of this system Intentional modification of any of the components may cause injury or reduce therapy f oxygen is used with the flow generator the oxygen flow must be turned off when the flow generator is not operating Explanation of the Warning When the flow generator is not in operation and the oxygen flow is left on oxygen delivered into the ventilator tubing may accumulate within the flow generator enclosure Oxygen accumulated in the flow generator enclosure will create a risk of fire This warning applies to most types of flow generators Please refrain from smoking while oxygen is in use Note At a fixed flow rate of supplemental oxygen flow the inhaled oxygen concentration will vary depending on the pressure settings patient breathing pattern interface selection
11. ivel Piece pivotante Pieza giratoria Pe a girat ria Short tube assembly elbow short tube and swivel Ensemble tuyau court coude tuyau court et pi ce pivotante Conjunto del tubo corto codo tubo corto y pieza giratoria Conjunto de tubo curto cotovelo tubo curto e pe a girat ria Headgear Harnais Arn s para la cabeza Arn s Top buckle Boucle sup rieure Presilla superior Fivela superior Top strap Sangle sup rieure Correa superior Correia superior Panel Partie lat rale Panel Painel Headgear ring Anneau du harnais Anillo del arn s Anel do arn s DO e Seal ring x2 Bague d tanch it x2 Back strap Sangle arri re Correa posterior Correia traseira Back buckle Boucle arri re Presilla posterior PM PPWS Y 8007 Junta x2 Anel de veda o x2 Fivela traseira Elbow Coude Codo Cotovelo Seal rings Bagues d tanch it Juntas An is de veda o Cap Bouchon Tapa Tampa Pillow sleeve Coussin Manguito de la almohadilla Manga da almofada Headgear ring Anneau du harnais Anillo del arn s Anel do arn s Slide clip Faites glisser la barrette Deslice el broche Deslize o clipe Pillow sleeve Coussin Manguito de la almohadilla Manga da almofada Slide clip Faites glisser la barrette Deslice el broche Deslize o clipe D Reassembling the Mirage Swift Il
12. nty English
13. ols System and Packaging A Caution consult accompanying documents LOT Lot number REF Part number User Guide Indicates a Warning or Caution and alerts you to a possible injury or A explains special measures for the safe and effective use of the device Ordering Information Part No Item System 60512 Mirage Swift Il Nasal Pillow System AMER Replacement Parts 60541 Pillow sleeve Small 1 pk 60542 Pillow sleeve Medium 1 pk 60543 Pillow sleeve Large 1 pk 60523 Headgear buckle 4 pk 60524 Seal ring 10 pk 60525 Cap 1 pk 60526 Headgear 1 pk 60536 Frame with Clip and Cap 1 pk 60528 Short tube assembly 1 pk 60529 Short tube assembly 10 pk 60530 Tube retainer 3 pk 60537 Frame System with Pillow sleeve excluding headgear Small 60538 Frame System with Pillow sleeve excluding headgear Medium 60539 Frame System with Pillow sleeve excluding headgear Large 60540 Clip 1 pk Part No Item Accessories 61121 Distal pressure port 1 pk 61124 Distal pressure port 10 pk 60549 Main pressure port 1 pk 60550 Main pressure port 10 pk Limited Warranty ResMed warrants that your ResMed system including its components frame pillow sleeve headgear tubing and other accessories Product shall be free from defects in material and workmanship for a period of ninety 90 days from the date of purchase by the initial consumer For further details refer to the Warra
14. one end of the main air tubing to your flow generator and the other end to the swivel Turn your flow generator on Notes Flow generator mask setting If available select Swift otherwise select Mirage as the mask option when using the Mirage Swift ll with ResMed flow generators that have mask setting options SmartStop feature This feature may not operate effectively when using the Mirage Swift ll with some ResMed flow generators that have the SmartStop feature This is particularly relevant when using the small size pillow sleeve If you feel air leaks readjust the angle of the nasal pillows to get a better seal You may need to readjust the headgear straps Do not overtighten the headgear straps If you are unable to resolve any leaks you may need an alternative pillow sleeve size Take time to adjust the following to achieve a secure and comfortable fit headgear straps angle of the pillow sleeve assembly position of the short tube assembly English Removing the Mirage Swift Il To remove the Mirage Swift II pull the pillow sleeve assembly away from your nose and lift the headgear off your head You do not need to loosen the headgear straps Disassembling for Cleaning See section C on the illustrations sheet Before disassembling the Mirage Swift ll disconnect the flow generator tubing To do this hold the swivel and pull it away from the flow generator tubing A Ah 4A CAUTION
15. rescribed continuous positive airway pressure or bilevel therapy for use in home hospitals or clinics Note The Mirage Swift ll is not compatible for use with ResMed AutoSet CS 2 and VPAP Adapt SV devices A A CAUTION In the USA federal law restricts this device to sale by or on the order of a physician Medical Information Note The Mirage Swift ll does not contain latex PVC or DEHP materials If you have ANY reaction to any part of the system discontinue use and consult your sleep therapist y IN WARNINGS AND CAUTIONS This system should be used only with flow generators CPAP or bilevel recommended by your physician or respiratory therapist It should not be used unless the flow generator is turned on and operating properly The vent holes should never be blocked Explanation of the Warning Flow generators are intended to be used with special interfaces with connectors which have vent holes vent holes are English located on the frame see section A on the illustrations sheet or separate exhalation devices to allow continuous flow of air out of the interface When the flow generator is turned on and functioning properly new air from the flow generator flushes the exhaled air out through the attached exhalation port However when the flow generator is not operating eg power failure not enough fresh air will be provided through the system and exhaled air may be rebreathed In this situation remove the M
16. sangles Adjust top straps Ajustez les sangles une tanch it efficace Rote el conjunto del manguito de la almohadilla Short tube Tuyau court Sainte ga SAN de RNa se a hacia afuera arri res Ajuste las correas posteriores sup rieures Ajuste las correas superiores para un ajuste c modo y sin fugas Gire o conjunto de manga da tuyau facultative Retenedor del tubo opcional Tubo corto Tubo curto Rs en Cap Bouchon ogoupoda ResMed esta virado para iora Ajuste as correias traseiras Ajuste as correias superiores almofada para uma melhor veda o e maior conforto Retentor do tubo opcional LY Tapa Tampa Optional tube positions Positions possibles pour le tuyau Posiciones posibles del tubo Posi es opcionais do tubo Clip Barrette Broche Clipe Note The seal rings do not need to be removed for cleaning Remarque vous n avez pas besoin de retirer les bagues d tanch it du harnais pour le nettoyage Nota Para la limpieza no es necesario retirar las juntas Nota Os an is de veda o n o necessitam de ser removidos Tube retainer Bague de fixation du tuyau Retenedor del tubo Pillow sleeve assembly pillow sleeve frame and clip Corps du masque PM PINS 8007 O oe coussin support et barrette Conjunto del manguito de la almohadilla manguito armaz n y broche Conjunto de manga da almofada manga da almofada arma o e clipe Retentor do tubo Sw