Home
        Tissuelyser User Manual - UTA
         Contents
1.      F03  displayed    Specified oscillation If the specified oscillation frequency is exceeded    frequency exceeded the TissueLyser II switches itself off and    F03    is  displayed  Contact QIAGEN Technical Services  to request instrument repair  The speed sensor  may be defective        F04    displayed  Transparent hood not To protect the operator  the drive motor of the  lowered TissueLyser II will only start if the transparent    hood has been lowered  If the  START  key is  pressed while the hood is raised     F04    is  displayed and the drive motor does not start  If  the hood is raised while the TissueLyser II is in  operation  the drive motor switches itself off and     F04    is displayed  To cancel the    F04    display   press the  STOP  key     TissueLyser Il User Manual 04 2010    n    r       33    Troubleshooting Guide    Comments and suggestions   F05  displayed    Keys held down for too If one or more keys on the keypad are pressed   long and held down for more than 15 seconds  the  drive motor of the TissueLyser Il switches itself off  and    F05    is displayed  This is a safety  mechanism to prevent unwanted changes in the  operation of the TissueLyser Il should an object  land on the keypad  To cancel the    FO5    display   switch the TissueLyser Il off and then on again        F06    displayed    Speed sensor defective  f one of the magnets controlling the speed of the  TissueLyser ll is missing or defective  the  instrument switches itself off and
2.   FO6  is  displayed  Contact QIAGEN Technical Services  to request instrument repair      F07  displayed    Solenoid switch If there is a defect in the cable for the solenoid   defective switch  which monitors the raising of the  transparent hood  the TissueLyser Il switches itself  off and    F07    is displayed  Contact QIAGEN    Technical Services to request instrument repair                                                                                                         E    34 TissueLyser Il User Manual 04 2010    Appendix A       Appendix A    Technical data    QIAGEN reserves the right to change specifications at any time   Environmental conditions    Operating conditions    Power 100 240 V AC  50 60 Hz  150 W    Mains supply voltage fluctuations are not to exceed 10  of  nominal supply voltages    Overvoltage Il    category   Air 5 40  C  41 104  F     temperature   Relative Maximum 80    humidity  for temperatures up to 31  C  linearly decreasing down to a  relative humidity of 50  at 40  C    Place of For indoor use only   operation   Altitude Up to 2000 m  6500 ft     Pollution level 2    Transportation conditions    Air  25  C to 60  C   13  F to 140  F  in manufacturer s package  temperature   Relative Maximum of 75   noncondensing    humidity      The instrument can also be operated at 4  C under noncondensing conditions     TissueLyser Il User Manual 04 2010 35    Appendix A       Storage conditions    Air 5 40  C  41 104  in manufacturer s package  tem
3.  or           key causes the disruption time to  increase or decrease rapidly without any pauses  If 99 minutes is exceeded   the disruption time starts from 10 seconds again     TissueLyser Il User Manual 04 2010 27    Operating Procedures    For the optimal disruption time for your application  see the TissueLyser       Handbook    7  Press the  START  key on the keypad  The TissueLyser Il will now operate at  the frequency specified in step 5  In addition  the  Time min sec  display  will count down the disruption time  i e   the display shows the time  remaining   If necessary  the disruption time can be adjusted while the  TissueLyser ll is in operation    Note  The TissueLyser II will not start if the transparent hood has not been  lowered   WARNING   Moving parts          A    Do not leave the TissueLyser Il unattended during operation        8  When the remaining time on the  Time min sec  display reaches zero  the  TissueLyser Il stops operating  The display resets to display the disruption  time for the previous disruption procedure carried out     To pause the TissueLyser II before the remaining time reaches zero  press  the  STOP  key on the keypad  To continue operation of the TissueLyser Il  from the time remaining  press the  START  key     To cancel operation of the TissueLyser Il  press the  STOP  key twice  The  TissueLyser Il then goes into standby mode  When the  START  key is  pressed again  the display is reactivated and the disruption time for the  prev
4.  placed in a 10 ml jar  containing a grinding ball and shaken at high speed     TissueLyser Adapter Sets  Grinding Jar Sets  and sample tubes can be precooled  to minimize degradation of nucleic acids during sample disruption  To find out  which procedures require precooling  see the TissueLyser Handbook     TT   TissueLyser Il User Manual 04 2010 13    General Description       3 2 Features of the TissueLyser II system    TissueLyser Il    The TissueLyser ll provides high speed shaking with oscillation frequencies of 3   30 Hz  180 1800 oscillations per minute   Two clamps at the front of the  instrument hold either a TissueLyser Adapter Set or a Grinding Jar Set  Figures 1  and 2   A safety mechanism prevents the instrument from operating until the  transparent hood is lowered over the TissueLyser Adapter Set or Grinding Jar  Set     Power is switched on and off at the power switch located at the back of the  instrument  Oscillation frequency can be adjusted in steps of 0 1 Hz using a  keypad  Figure 3  located on top of the instrument  Disruption time  from   10 seconds up to a maximum of 99 minutes  can also be set using the keypad   Figure 3   The disruption parameters for a particular application  i e    oscillation frequency and disruption time  can be stored in the memory of the  TissueLyser Il  Up to 9 disruption parameters can be stored                    Figure 1  TissueLyser II  closed   The transparent hood is lowered  and the clamps at the  front of the instrume
5.  replacement product is being supplied by  QIAGEN  provides free recycling of its WEEE marked electronic equipment in  Europe  If a replacement product is not being purchased from QIAGEN   recycling can be provided upon request at additional cost  To recycle electronic  equipment  contact your local QIAGEN sales office for the required return form   Once the form is submitted  you will be contacted by QIAGEN either to request  follow up information for scheduling collection of the electronic waste or to  provide you with an individual quote     TissueLyser Il User Manual 04 2010 37    Appendix A       EC declaration of conformity    Name and address of the company  QIAGEN GmbH  QIAGEN Strasse 1  D 40724 Hilden  Germany  We herewith declare under our sole responsibility that the product  TissueLyser II  meets all applicable requirements of the following European Directives  Low Voltage Directive  LVD  2006 95 EC  Electromagnetic Compatibility Directive  EMC  2004 108 EC    and the relevant harmonised standards    IEC 61010 1 2001  IEC 61326 1 2005    Hombrechtikon  11 January 2010          Pit Muggli  Director Business Excellence    QIAGEN Instruments AG       EE ns   38 TissueLyser Il User Manual 04 2010    Appendix B       Appendix B    Liability clause    QIAGEN shall be released from all obligations under its warranty in the event  repairs or modifications are made by persons other than its own personnel   except in cases where the Company has given its written consent to 
6. 293 422 911   Fax 01293 422 922   Technical 01293 422 999 106 e    USA   Orders 800 426 8157   Fax 800 718 2056   Technical 800 DNA PREP  800 362 7737  QIAGEN                     Sample  amp  Assay Technologies    
7. 79  Canada   Orders 800 572 9613   Fax 800 713 5951   Technical 800 DNA PREP  800 362 7737   China   Orders 86 21 3865 3865   Fax 86 21 3865 3965   Technical 800 988 0325  Denmark   Orders 80 885945   Fax 80 885944   Technical 80 885942  Finland   Orders 0800 914416   Fax 0800 914415   Technical 0800 914413  France    Orders 01 60 920 926   Fax 01 60 920 925   Technical 01 60 920 930   Offers 01 60 920 928  Germany   Orders 02103 29 12000   Fax 02103 29 22000   Technical 02103 29 12400  Hong Kong   Orders 800 933 965   Fax 800 930 439   Technical 800 930 425  Ireland   Orders 1800 555 049   Fax 1800 555 048   Technical 1800 555 061  Italy   Orders 800 789 544   Fax 02 334304 826   Technical 800 787980  Japan    Telephone 03 6890 7300   Fax 03 5547 0818   Technical 03 6890 7300  Korea  South    Orders 080 000 7146    Fax 02 2626 5703   Technical 080 000 7145  Luxembourg    Orders 8002 2076   Fax 8002 2073   Technical 8002 2067  Mexico   Orders 01 800 7742 639   Fax 01 800 1122 330   Technical 01 800 7742 436  The Netherlands   Orders 0800 0229592   Fax 0800 0229593   Technical 0800 0229602  Norway   Orders 800 18859   Fax 800 18817   Technical 800 18712  Singapore   Orders 1800 742 4362   Fax 65 6854 8184   Technical 1800 742 4368  Spain   Orders 91 630 7050   Fax 91 630 5145   Technical 91 630 7050    Sweden   Orders 020 790282   Fax 020 790582   Technical 020 798328    Switzerland   Orders 055 254 22 11   Fax 055 254 22 13   Technical 055 254 22 12 e Q e    e  UK   Orders 01
8. April 2010       TissueLyser Il User Manual    For disruption of up to 48 or 192 biological    samples       QIAGEN       Sample  amp  Assay Technologies    QIAGEN Sample and Assay Technologies    QIAGEN is the leading provider of innovative sample and assay technologies   enabling the isolation and detection of contents of any biological sample  Our  advanced  high quality products and services ensure success from sample to  result     QIAGEN sets standards in     Purification of DNA  RNA  and proteins  Nucleic acid and protein assays  microRNA research and RNAi    Automation of sample and assay technologies    Our mission is to enable you to achieve outstanding success and  breakthroughs  For more information  visit www giagen com     Contents          Contents  1 Safety Information 5  1 1 Proper use 2  1 2 Electrical safety 6  1 3 Environment 7  1 4 Biological safety 8  1 5 Chemicals 9  1 6 Waste disposal 9  1 7 Mechanical hazards 10  2 Introduction 11  2 1 About this user manual 11  2 2 Intended use of the TissueLyser Il 11  3 General Description 13  3 1 TissueLyser ll principle 13  3 2 Features of the TissueLyser ll system 14  3 3 Applications 20  4 Installation Procedures 21  4 1 Unpacking the TissueLyser Il 21  4 2 Installation requirements 21  4 3 Transporting and repacking the TissueLyser ll 23  4 4 Installation of a TissueLyser Adapter Set or Grinding Jar Set 24  4 5 Replacing the fuses 25  5 Operating Procedures 27  5 1 Sample disruption 27  5 2 Storing disruption 
9. II  see Appendix A  page 35      TissueLyser Il User Manual 04 2010 9    Safety Information       1 7 Mechanical hazards    WARNING   Moving parts  A Do not leave the TissueLyser II unattended during operation                    To ensure safety  the TissueLyer II will not operate until the transparent hood is  closed        WARNING   Moving parts    A Damage or personal injury can occur if the TissueLyser  Adapter Set or Grinding Jar Set is not securely attached     Ensure that both adapters or grinding jars are of similar  weight  Do not operate the TissueLyser Il with only a single  adapter or grinding jar installed                 10 TissueLyser Il User Manual 04 2010    Introduction       2 Introduction    Thank you for choosing the TissueLyser II system  We are confident it will  become an integral part of your laboratory     Before using the TissueLyser Il  it is essential you read this user manual carefully  and pay particular attention to the safety information  The instructions and  safety information in the user manual must be followed to ensure safe operation  of the TissueLyser Il and to maintain the TissueLyser Il in a safe condition     2 1 About this user manual    This user manual provides information about the TissueLyser ll in the following  sections    Safety Information   Introduction   General Description   Installation Procedures   Operating Procedures   Maintenance Procedures    Eg ACE E    Troubleshooting Guide  Appendices    The appendices contain th
10. ad mill designed for rapid and efficient  disruption of up to 48 or 192 samples at the same time  A wide variety of  biological samples can be processed  including human  animal  and plant  tissues as well as yeast and bacteria  Each sample is simultaneously disrupted  and homogenized by high speed shaking with a bead in a sealed tube  bead  milling      3 1 TissueLyser Il principle    Efficient sample disruption is a prerequisite for nucleic acid and protein  purification procedures  Incomplete sample disruption can lead to significantly  reduced yields and can increase the risk of clogging when using purification  columns or magnetic particles  The TissueLyser Il thoroughly disrupts and  simultaneously homogenizes biological samples in the presence of lysis buffer   Plant tissues can alternatively be disrupted and homogenized in the absence of  lysis buffer     Tubes containing sample and bead are placed in a TissueLyser Adapter Set and  shaken at a very high speed  causing the samples to be disrupted and  homogenized by the beating and grinding effect of the beads  The tubes can be  discarded after use  which prevents cross contamination and provides time  savings  as there is no need to clean tubes after disrupting samples  In addition   there is no cross contamination during sample disruption as each tube is  securely sealed with its own lid     The TissueLyser II can also disrupt and homogenize large samples when used in  combination with a Grinding Jar Set  Samples are
11. adapter or grinding jar installed                 4 5 Replacing the fuses  The fuses in the TissueLyser II can be replaced as follows   M Unplug the power cord connected to the TissueLyser II     M The fuse compartment is located between the power switch and the power  cord socket  Figure 12   Press at the side of the fuse compartment to  release it     M Replace the fuses  The following are required  two MT 3 15 A fuses     M Reinsert the fuse compartment and lock it into place     TissueLyser Il User Manual 04 2010 25    Installation Procedures     eee                                                                                                                                           ae IIMA                                                                       Figure 12  Backside of TissueLyser II     Power switch     Fuse compartment   8 Power cord  socket   5  RS 232 port  for service personnel    3 Safety information plate   j Voltage rating  plate     j Safety information plate     26 TissueLyser Il User Manual 04 2010    Operating Procedures    5 Operating Procedures    This section describes how to operate the TissueLyser Il system  Before  proceeding  you should familiarize yourself with the features of the TissueLyser II   Section 3  and have properly installed the instrument  Section 4      5 1 Sample disruption    The procedure below describes how to disrupt samples using the TissueLyser II   For detailed information about sample disruption and protocols  see 
12. d press it lightly into the indentation in  the clamp  C      gre LLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLILLOLLLLULLIZLIILIILLZ  ZLLLLLLLLLLLLLALLLALALZLSLLLILLLLLNLunM  24 TissueLyser Il User Manual 04 2010    Installation Procedures    M Turn the handwheel  B  clockwise with two fingers until the adapter or  grinding jar is seated in the clamp and level  and does not move freely   Continue to rotate the handwheel clockwise until 6 8 easily audible clicks  are heard  the locking pin  A  rises and falls with each audible click     The engaged locking pin prevents automatic opening of the adapter or  grinding jar  If the locking pin cannot be pulled upward to release it  do  not force it upward using a hammer or other tool  as damage may occur   To release the locking spin  turn the handwheel slightly clockwise  the  locking pin should then be freely moveable     M To remove the adapter or grinding jar  lift the locking pin  A  and turn the  handwheel  B  in the opposite direction        CAUTION   Damage to the instrument    A Use of adapters and plasticware other than those described in  this user manual may lead to damage to the clamps of the  TissueLyser Il              WARNING   Moving parts    A Damage or personal injury can occur if the TissueLyser  Adapter Set or Grinding Jar Set is not securely attached     Ensure that both adapters or grinding jars are of similar  weight  Do not operate the TissueLyser Il with only a single  
13. e carried out at room temperature or after freezing the grinding  jars in liquid nitrogen  For more information  see the product sheet supplied    with the Grinding Jar Set                       Figure 6  Grinding Jar Set  S  Steel  Each 10 ml jar is made of stainless steel and contains    a 20 mm diameter stainless steel grinding ball     Grinding Jar Set  Teflon      The Grinding Jar Set  Teflon  cat  no  69986  allows disruption of 2 large  samples in parallel using Teflon grinding balls  Figure 7   Sample disruption  can be carried out at room temperature  For more information  see the product    sheet supplied with the Grinding Jar Set           Z             Figure 7  Grinding Jar Set  Teflon  Each 10 ml jar is made of Teflon and contains a 20 mm    diameter Teflon grinding ball     18       TissueLyser Il User Manual 04 2010    General Description       TissueLyser Single Bead Dispensers    TissueLyser Single Bead Dispensers dispense individual beads into any sample  container  Figure 8   They are available in 2 sizes     M Tissuelyser Single Bead Dispenser  5 mm  cat  no  69965   Dispenses  beads of 5 mm diameter  and contains a reservoir that can hold  approximately 150 beads    M Tissuelyser Single Bead Dispenser  7 mm  cat  no  69967   Dispenses  beads of 7 mm diameter  and contains a reservoir that can hold  approximately 45 beads    The dispensers can be cleaned with water or detergent  For more information   see the product sheet supplied with the TissueLyser Si
14. e clamps of the  TissueLyser ll        CAUTION   Damage to the instrument    A Avoid spilling water or chemicals onto the TissueLyser Il   Damage caused by water or chemical spillage will void your  warranty                 In case of emergency  switch off the TissueLyser Il at the power switch and  unplug the power cord from the power outlet     1 2 Electrical safety    WARNING   Electrical hazard    A Any interruption of the protective conductor  earth ground   lead  inside or outside the TissueLyser Il or disconnection of  the protective conductor terminal is likely to make the  TissueLyser ll dangerous        Intentional interruption is prohibited     Lethal voltages inside the TissueLyser ll    When the TissueLyser ll is connected to line power  terminals  may be live  and opening the TissueLyser Il or removing parts  is likely to expose live parts                 To ensure satisfactory and safe operation of the TissueLyser Il  follow the advice  below     MB The line power cord must be connected to a line power outlet that has a  protective conductor  earth ground      M Do not adjust or replace internal parts of the TissueLyser Il   M Do not operate the TissueLyser II with any covers or parts removed     M   If liquid has spilled inside the TissueLyser II  switch off the TissueLyser II   disconnect it from the power outlet  and contact QIAGEN Technical Services        6 TissueLyser Il User Manual 04 2010    Safety Information       M If the TissueLyser II becomes elect
15. e following   M Technical data  E Warranty terms    2 2 Intended use of the TissueLyser Il    The TissueLyser Il is intended for molecular biology applications  This product is  neither intended for the diagnosis  prevention  or treatment of a disease  nor has  it been validated for such use either alone or in combination with other products   Therefore  the performance characteristics of the product for clinical use  i e    diagnostic  prognostic  therapeutic  or blood banking  are unknown     The TissueLyser Il is not designed for continuous operation  but for single 8 hour  shifts with 30  operating time  QIAGEN has successfully tested the TissueLyser Il  with the following adapter sets and grinding beads     M TissuelLyser Adapter Set 2 x 96  with one 3 mm diameter tungsten carbide  bead or one 5 mm diameter stainless steel bead per collection microtube    M TissuelLyser Adapter Set 2 x 24  with one 3 mm diameter tungsten carbide  bead or one 5 mm or 7 mm diameter stainless steel bead per  microcentrifuge tube       TissueLyser Il User Manual 04 2010 11    Introduction       Damage to the clamps of the TissueLyser II    The clamps of the TissueLyser II are designed for use with the following QIAGEN  adapters and consumables     M TissueLyser Adapter Set 2 x 24  cat  no  69982  containing standard 2 ml  microcentrifuge tubes    M TissueLyser Adapter Set 2 x 96  cat  no  69984  containing Collection  Microtubes  racked   cat  no  19560  sealed with Collection Microtube  Cap
16. els on the TissueLyser II     Risk of personal injury and material damage    Only perform maintenance that is specifically described in this    user manual           Servicing    The TissueLyser II is supplied with a warranty that lasts for 12 months  starting  from the date of shipment  The warranty includes all repairs due to mechanical    breakdown     6 1 Regular maintenance    The TissueLyser ll should be wiped after use using a soft cloth moistened with a  suitable cleaning agent  see below   To remove residual cleaning agent and  avoid instrument damage  the TissueLyser II should then be immediately  wiped with a soft cloth moistened with water  Be sure to switch off the  TissueLyser Il at the power switch and unplug the power cord from the power  socket before cleaning     The components of the TissueLyser Adapter Set or Grinding Jar Set should be  cleaned after use using a suitable cleaning agent  see below   rinsed with  distilled water  and then dried using paper towels     If the TissueLyser II is contaminated with infectious material  it should be    decontaminated  The TissueLyser Il should also be decontaminated before  shipping  and a decontamination certificate must be completed to certify that    the instrument has been decontaminated  To request a decontamination  certificate  contact QIAGEN Technical Services     30    TissueLyser Il User Manual       04 2010    Maintenance Procedures       Cleaning agents    The following disinfectants and detergents are 
17. f the TissueLyser II and avoids invalidation of your guarantee     4 2 Installation requirements    Removal of the transport security device    A transport security device is located on the underside of the TissueLyser II  this  is comprised of 2 screws  see Figure 10   Using a screwdriver  remove the   2 screws  Do not discard the screws  as they will need to be reinstalled if the  TissueLyser I  needs to be transported in the future     oo N  TissueLyser Il User Manual 04 2010 21    Installation Procedures          Transport security device             Figure 10  Transport security device  The transport security device  comprised of 2 screws  indicated by the arrows  is located on the underside of the TissueLyser II     CAUTION   Damage to instrument    A Operation of the TissueLyser II with the transport security   device installed  or transport of the TissueLyser II without the  transport security device installed  can lead to damage to the  mechanical components of the instrument                 Site requirements    Place the TissueLyser Il on a flat  stable surface  and ensure that there is  sufficient space around and under the instrument for ventilation  The TissueLyser  Il is for indoor use only under the following conditions     M Ambient temperature within the range of 5 40  C  41 104  F   M Up to 2000 m  6500 ft   above mean sea level  MSL     M Maximum relative humidity of 80  for temperatures up to 31  C  linearly  decreasing down to a relative humidity of 50  a
18. ious procedure is displayed     switch     When the TissueLyser ll is no longer in operation  switch it off at the power    5 2 Storing disruption parameters    The TissueLyser ll has an internal memory which has 9 storage locations to  store disruption parameters  i e   disruption time and oscillation frequency   To  store disruption parameters  follow the procedure below     1     Switch on the TissueLyser Il     on    appears in the    Memory    display   Press the    PROG    key repeatedly until the desired storage location in the    memory is displayed  i e      P1    to  P9       Press the  SET  key  This causes all displays to flash   Set the disruption time and oscillation frequency using the         and       keys    Note  To cancel the operation  press the    PROG    key  The disruption  parameters selected will not be stored     Press the  SET  key again  The display stops flashing  and the disruption       2   3   4   on the keypad   5   parameters are stored   28    TissueLyser Il User Manual 04 2010       Operating Procedures    To retrieve disruption parameters from the memory  press the    PROG    key  repeatedly until the desired storage location  i e    P1  to  P9   is displayed   Then press the  START  key to start operation of the TissueLyser Il     5 3 Displaying the accumulated operating time    To display the total amount of time that the TissueLyser Il has been operated   follow the procedure below     1  When the TissueLyser ll is in standby mode  sim
19. ive bacteria    For information about specific applications and protocols for sample disruption   see the TissueLyser Handbook        20 TissueLyser Il User Manual 04 2010    Installation Procedures    4 Installation Procedures    4 1 Unpacking the TissueLyser II    Before unpacking the TissueLyser II  check whether the cardboard container is  damaged  In case of damage  contact the transporter of the container  The  TissueLyser Il must not be knocked  shaken  or thrown during transport   otherwise its electronic and mechanical components may be damaged     Make a note of how the TissueLyser ll is packed  If returning the instrument to  QIAGEN for repair  it needs to be repacked using the original packaging   When lifting the TissueLyser II  place both hands below the base of the  instrument     After unpacking the TissueLyser Il  check that the following items are supplied   M Tissuelyser II   M Power cords for different countries   M Tissuelyser I  User Manual   MB TissuelLyser Handbook   If anything is missing  contact QIAGEN Technical Services     Check that the TissueLyser II is not damaged  If anything is damaged  contact  QIAGEN Technical Services  Make sure that the TissueLyser II has equilibrated  to ambient temperature before operating it  The TissueLyser Il should be stored  dry when not in use     Retain the original packaging in case you need to return the TissueLyser II to  QIAGEN for repair  Using the original packaging minimizes damage during  transportation o
20. l is intended to supplement  not supersede  the  normal safety requirements prevailing in the user   s country     1 1 Proper use       WARNING   Risk of personal injury and material damage    A Improper use of the TissueLyser II may cause personal injuries  or damage to the instrument  Do not attempt to move the  TissueLyser ll during operation     The TissueLyser Il should only be operated by qualified  personnel who have been appropriately trained     Servicing of the TissueLyser II should only be performed by  QIAGEN or an authorized agent  service technicians   Do not  carry out any modifications of the instrument     QIAGEN shall not be held responsible for any injuries or  damages resulting from nonobservance of the instructions and  safety information in this user manual                 The TissueLyser II should only be used for the applications described in the  TissueLyser Handbook  The TissueLyser Il should only be used in combination                                         aay  TissueLyser Il User Manual 04 2010 5    Safety Information       with accessories approved by QIAGEN  see Section 3 2  page 14   Do not  continue to use the TissueLyser ll or TissueLyser accessory if it is damaged     Perform the maintenance as described in Section 6  QIAGEN charges for  repairs that are required due to incorrect maintenance        CAUTION   Damage to the instrument    A Use of adapters and plasticware other than those described in  this user manual may lead to damage to th
21. ld be used only for the  TissueLyser Adapter Sets and Grinding Jar Sets     M If solvents or saline  acidic  or alkaline solutions are spilt on the TissueLyser  Il  wipe the spilt liquid away immediately    M Follow manufacturer s safety instructions for handling cleaning agents    W Follow manufacturer s instructions for soaking time and concentration of  cleaning agents  exposure for longer than the recommended soaking time    can damage the TissueLyser II  TissueLyser Adapter Sets  and Grinding Jar  Sets        TissueLyser Il User Manual 04 2010 31    Maintenance Procedures          WARNING    A       Toxic fumes    Do not use bleach to clean or disinfect used labware  Bleach in  contact with salts from the buffers used can produce toxic  fumes                 A       CAUTION   Damage to the instrument  A Do not use bleach  solvents  or reagents containing acids   alkalis  or abrasives to clean the TissueLyser Il   CAUTION   Damage to the instrument       Avoid spilling water or chemicals onto the TissueLyser Il   Damage caused by water or chemical spillage will void your  warranty           6 2 Regular safety checks    Every 6 months  the limit switch and the brake should be checked for correct    operation     To check the limit switch and the brake     M Remove the TissueLyser Adapter Set or Grinding Jar Set from the clamps of  the TissueLyser II     Switch on the TissueLyser II   Set the oscillation frequency to 30 Hz   Press the    START    key     Raise the transpare
22. ngle Bead Dispenser                    Figure 8  TissueLyser Single Bead Dispenser  The dispenser dispenses individual beads  into any sample container     TissueLyser Bead Dispensers  96 Well    TissueLyser Bead Dispensers  96 Well dispense 96 beads in parallel into  Collection Microtubes  racked   Figure 9   They are available in 2 sizes     M TissueLyser 3 mm Bead Dispenser  96 Well  cat  no  69973   Dispenses  beads of 3 mm diameter  and contains a reservoir that can hold  approximately 1000 beads    M Tissuelyser 5 mm Bead Dispenser  96 Well  cat  no  69975   Dispenses  beads of 5 mm diameter  and contains a reservoir that can hold  approximately 300 beads    The dispensers can be cleaned with water or detergent  For more information   see the product sheet supplied with the TissueLyser Bead Dispenser  96 Well        TissueLyser Il User Manual 04 2010 19    General Description  OE LL SSS             2  2  DU  gt   2 a        a    G             Figure 9  TissueLyser Bead Dispenser  96 Well  The dispenser dispenses 96 beads into a  rack of collection microtubes     Beads   The following beads for the TissueLyser II are available from QIAGEN   M Stainless Steel Beads  5 mm  cat  no  69989    M Stainless Steel Beads  7 mm  cat  no  69990    M Tungsten Carbide Beads  3 mm  cat  no  69997     3 3 Applications   The TissueLyser ll enables disruption and homogenization of   Human and animal tissues   Human and animal cells   Plant tissues    Yeast    Gram positive and gram negat
23. nt are empty     Note  The maximum oscillation frequency may drop down to 25 Hz  depending  on local fluctuations in the mains voltage        14 TissueLyser Il User Manual 04 2010    General Description                Set is installed           Figure 2  TissueLyser II  open   The transparent hood is raised  and a TissueLyser Adapter    auuu aa  TissueLyser Il User Manual 04 2010    15    General Description                      Figure 3  Keypad of the TissueLyser II  A  Memory  display  with    PROG    and  SET  keys  for storing disruption parameters   3  Frequency 1 s  display  with       and         keys for  adjusting oscillation frequency     Time min sec  display  with       and         keys for adjusting  disruption time   5     START    and  STOP  keys for starting and stopping operation of the  TissueLyser Il     TissueLyser Adapter Set 2 x 24    The TissueLyser Adapter Set 2 x 24  cat  no  69982  allows disruption of   48 samples in parallel using standard 2 ml microcentrifuge tubes  e g    Eppendorf Safe Lock micro test tubes    The adapter set contains 2 adapters   with each adapter comprised of a top plate  a bottom plate  and a tube holder  for holding 24 microcentrifuge tubes  Figure 4   Sample disruption can be  carried out at room temperature or after storing the adapter set at  80  C for at  least 2 hours  The adapter set can be cleaned with detergent  microbicides  or  up to 96  ethanol  For more information  see the product sheet supplied with  the adapte
24. nt hood  The drive motor should switch itself off  and       F04    should appear in the display     M Lower the transparent hood  The drive motor should not switch itself on    again     If there is a defect  contact QIAGEN Technical Services to request repair of the    instrument     TissueLyser Adapter Sets and Grinding Jar Sets may become worn  depending  on the frequency of use and the samples disrupted  This may lead to damage to  the clamps of the TissueLyser II  Check the TissueLyser Adapter Sets and  Grinding Jar Sets regularly for wear and tear  and replace if necessary     32       TissueLyser Il User Manual 04 2010    Troubleshooting Guide    7 Troubleshooting Guide    7 1 Error codes displayed by the TissueLyser II    Comments and suggestions     FO1    displayed    Drive motor of When the drive motor is overloaded  the  TissueLyser II TissueLyser II switches itself off to allow the drive  overloaded motor to cool  A display on the TissueLyser Il    alternates between    F01    and the remaining  cooling time in minutes  e g    03  indicates 3  minutes   After the cooling time has elapsed   switch the TissueLyser II off and then on again  before continuing to operate the instrument      F02  displayed    Drive motor of If the drive motor does not start after the  START  TissueLyser ll not key has been pressed  contact QIAGEN  starting Technical Services to request instrument repair     The drive motor may be blocked  or the motor o  speed sensor may be defective 
25. pair  please contact QIAGEN to request  a    Declaration of Decontamination    form  After completing and signing this  form  place it in an envelope which should then be attached to the external  surface of the cardboard container  In addition  place a copy of the form inside  the container     To repack the TissueLyser Il  first reinstall the transport security device using the  two original screws supplied with the instrument  for details  see Section 4 2    Then return the TissueLyser II to its original packaging by following the  unpacking steps in reverse order     ME  TissueLyser Il User Manual 04 2010 23    Installation Procedures    4 4 Installation of a TissueLyser Adapter Set or Grinding  Jar Set  The TissueLyser ll is not supplied with any accessories  TissueLyser Adapter Sets    and Grinding Jar Sets need to be ordered separately  to order the TissueLyser Il  or its accessories  visit www giagen com Products TissueLyserll aspx                  Figure 11  Left hand clamp of the TissueLyser II  1 Locking pin  H Handwheel   Indentation in clamp     Attach a TissueLyser Adapter Set or Grinding Jar Set to the TissueLyser Il as  follows     M Pull the locking pin  A  upward out of its slot  and rotate it by 90   Figure  11   This releases the locking device     MB   Turn the handwheel  B  counterclockwise until the maximum clamping  range is reached     M Rotate back the locking pin  A  by 90   until it engages with its slot     M Insert the adapter or grinding jar  an
26. parameters 28  5 3 Displaying the accumulated operating time 29  5 4 Displaying the software version 29  6 Maintenance Procedures 30  6 1 Regular maintenance 30  6 2 Regular safety checks 32  7 Troubleshooting Guide 33  7 1 Error codes displayed by the TissueLyser II 33  TissueLyser Il User Manual 04 2010 3    Contents       Appendix A 35  Technical data 35  Environmental conditions 35  Waste Electrical and Electronic Equipment  WEEE  37  EC declaration of conformity 38   Appendix B 39  Liability clause 39    4 TissueLyser Il User Manual 04 2010    Safety Information       1 Safety Information    Before using the TissueLyser ll system  it is essential that you read this user  manual carefully and pay particular attention to the safety information  The  instructions and safety information in the user manual must be followed to  ensure safe operation of the TissueLyser Il and to maintain the TissueLyser ll in a  safe condition     The following types of safety information appear throughout this user manual     WARNING   The term WARNING is used to inform you about situations  A that could result in personal injury to you or other persons        Details about these circumstances are given in a box like this  one              CAUTION   The term CAUTION is used to inform you about situations that  A could result in damage to the TissueLyser II or other   equipment    Details about these circumstances are given in a box like this   one                 The advice given in this manua
27. perature   Relative Maximum of 85   noncondensing    humidity    Mechanical data   Dimensions Height  266 mm  up to 524 mm with the hood raised   Width  371 mm  Depth  461 mm    Mass Approx  26 kg  without TissueLyser Adapter Set or Grinding  Jar Set installed    Oscillation 3 30 Hz   frequency   Noise The noise characteristic values are also influenced by the   characteristic properties of the sample to be disrupted    values Sound power level Lwa   71 4 dB A     Emission value related to workplace LA    61 dB A     36 TissueLyser Il User Manual 04 2010    Appendix A       Waste Electrical and Electronic Equipment  WEEE     This section provides information about disposal of waste electrical and  electronic equipment by users in the European Union     The European Directive 2002 96 EC on WEEE requires proper disposal of  electrical and electronic equipment when it reaches its end of life  The crossed   out wheeled bin symbol  see below  indicates that this product must not be  disposed of with other waste  it must be taken to an approved treatment facility  or to a designated collection point for recycling  according to local legislation   The separate collection and recycling of waste electronic equipment at the time  of disposal helps to conserve natural resources and ensures that the product is  recycled in a manner that protects human health and the environment     QIAGEN accepts its responsibility in accordance with the specific WEEE  recycling requirements and  where a
28. perform  such repairs or modifications     All materials replaced under this warranty will be warranted only for the  duration of the original warranty period  and in no case beyond the original  expiration date of original warranty unless authorized in writing by an officer of  the Company  Read out devices  interfacing devices and associated software  will be warranted only for the period offered by the original manufacturer of  these products  Representations and warranties made by any person  including  representatives of QIAGEN  which are inconsistent or in conflict with the  conditions in this warranty shall not be binding upon the Company unless  produced in writing and approved by an officer of QIAGEN     TissueLyser Il User Manual 04 2010 39    Notes    40 TissueLyser Il User Manual 04 2010    Notes    TissueLyser Il User Manual 04 2010 41    Notes    42 TissueLyser Il User Manual 04 2010    Trademarks  QIAGEN    QIAGEN Group   DNA ExitusPlus     AppliChem   Gigasept    Lysetol    Mikrozid    Sch  lke  amp  Mayr GmbH   DECON QUAT     Veltek Associates  Inc    Teflon   E     du Pont de Nemours and Company         2009 2010 QIAGEN  all rights reserved     eee  www qiagen com  Australia   Orders 1 800 243 800   Fax 03 9840 9888    Technical 1 800 243 066  Austria   Orders 0800 28 10 10   Fax 0800 28 10 19   Technical 0800 28 10 11  Belgium   Orders 0800 79612   Fax 0800 79611   Technical 0800 79556  Brazil   Orders 0800 557779   Fax 55 11 5079 4001   Technical 0800 5577
29. r set       Use of other tubes in combination with 7 mm diameter beads may result in the beads getting  lodged in the lids of the tubes     16 TissueLyser Il User Manual 04 2010    General Description                   Figure 4  TissueLyser Adapter Set 2 x 24  Each adapter consists of a top plate  a bottom  plate  and a tube holder for holding 24 microcentrifuge tubes  2 ml      TissueLyser Adapter Set 2 x 96    The TissueLyser Adapter Set 2 x 96  cat  no  69984  allows disruption of   192 samples in parallel using Collection Microtubes  racked  and Collection  Microtube Caps  cat  nos  19560 and 19566   The adapter set contains   2 adapters  with each adapter comprised of a top plate and a bottom plate for  holding a rack of capped collection microtubes  Figure 5   Sample disruption  can be carried out at room temperature or after storing the adapter set at     80  C for at least 2 hours  The adapter set can be cleaned with detergent   microbicides  or up to 96  ethanol  For more information  see the product  sheet supplied with the adapter set                 Figure 5  TissueLyser Adapter Set 2 x 96  Each adapter consists of a top plate and a  bottom plate for holding a rack of capped collection microtubes     TissueLyser Il User Manual 04 2010 17    Grinding Jar Set  S  Steel    General Description       The Grinding Jar Set  S  Steel  cat  no  69985  allows disruption of 2 large  samples in parallel using stainless steel grinding balls  Figure 6   Sample  disruption can b
30. recommended for cleaning the  TissueLyser Il  TissueLyser Adapter Sets  and Grinding Jar Sets     Note  If you want to use disinfectants different from those recommended   ensure that their compositions are similar to those described below     General cleaning    MB Mild detergents   B 70  ethanol   Disinfection    M Ethanol based disinfectants can be used for disinfection of surfaces  e g    25 g ethanol and 35 g 1 propanol per 100 g liquid or Mikrozid   Liquid   Schilke  amp  Mayr GmbH  cat  no  109160     M Disinfectants based on glyoxal and quaternary ammonium salt can be  used for submerging the TissueLyser Adapter Sets and Grinding Jar Sets   e g   10 g glyoxal  12 g lauryldimethylbenzylammonium chloride  12 g  myristyldimethylbenzylammonium chloride  and 5 15  nonionic detergent  per 100 g liquid  Lysetol   AF  Gigasept   Instru AF in Europe  cat  no   107410  or DECON QUAT   100  Veltek Associates  Inc   in the USA  cat   no  DQ100 06 167 01      Removal of RNase contamination     M RNosekKiller  5 PRIME  cat  no  2500080  can be used for cleaning surfaces  and spraying the TissueLyser Adapter Sets and Grinding Jar Sets    Removal of nucleic acid contamination     M DNA ExitusPlus     AppliChem  cat  no  A7089 0100  can be used for  cleaning surfaces and submerging the TissueLyser Adapter Sets and  Grinding Jar Sets     General instructions     M Do not use spray bottles to spray cleaning or disinfectant liquids onto  surfaces of the TissueLyser II  Spray bottles shou
31. rically unsafe  prevent other personnel  from operating it  and contact QIAGEN Technical Services  the TissueLyser  Il may be electrically unsafe when     Mit or the line power cord appears to be damaged     M It has been stored under unfavorable conditions for a prolonged  period     Mt has been subjected to severe transport stresses     If the TissueLyser ll is not in use  switch it off at the power switch  If the  TissueLyser II will not be used for a prolonged period of time  unplug the power  cord from the power outlet     1 3 Environment    Operating conditions       WARNING   Explosive atmosphere    A The TissueLyser Il is not designed for use in an explosive  atmosphere           WARNING   Risk of explosion    A The TissueLyser ll is intended for use with reagents and  substances supplied with QIAGEN kits  Use of other reagents  and substances may lead to fire or explosion        CAUTION   Damage to the instrument    A Direct sunlight may bleach parts of the instrument and cause  damage to plastic parts  The TissueLyser ll must be located out  of direct sunlight                 EE  TissueLyser Il User Manual 04 2010 7    Safety Information    1 4 Biological safety    Specimens and reagents containing materials from humans should be treated  as potentially infectious  Use safe laboratory procedures as outlined in  publications such as Biosafety in Microbiological and Biomedical Laboratories     HHS  www cdc gov od ohs biosfty biosfty htm      Samples    Samples may con
32. s  cat  no  19566     QIAGEN adapters together with the recommended tubes form a component that  is of an optimal thickness  allowing secure tightening of the clamps of the  TissueLyser II without straining the clamping mechanism     Damage to the clamps of the TissueLyser II will occur if other components of  reduced thickness are installed  Therefore  the TissueLyser Il must not be used  with adapters from other suppliers or with QIAGEN adapters containing tubes  not recommended by QIAGEN      Requirements for TissueLyser II users    This table covers the general level of competence and training necessary for  transportation  installation  use  maintenance  and servicing of the TissueLyser  ll        Training and       Task    Personnel    experience       Transportation    Installation  routine  use  and maintenance    Servicing    No special  requirements    Laboratory technicians  or equivalent    QIAGEN or an  authorized agent   service technicians     No special  requirements    Appropriately trained  and experienced  personnel            Use of the TissueLyser Il with non QIAGEN adapters has been tested by other suppliers  e g    Roche recommend use of the TissueLyser II with GS FLX Titanium emPCR Shaker Adapters   LV  which are accessories for the Genome Sequencer FLX Instrument   Use of such adapters  has not been tested by QIAGEN     12    TissueLyser Il User Manual    04 2010    General Description       3 General Description    The TissueLyser II is a flexible be
33. sruption process        Always wear safety glasses  gloves  and a lab coat     The responsible body  e g   laboratory manager  must take  the necessary precautions to ensure that the surrounding  workplace is safe and that TissueLyser ll operators are not  exposed to hazardous levels of toxic substances  chemical or  biological  as defined in the applicable Material Safety Data  Sheets  MSDSs  or OSHA   ACGIH   or COSHH  documents     Venting for fumes and disposal of wastes must be in  accordance with all national  state  and local health and  safety regulations and laws                   OSHA  Occupational Safety and Health Administration  United States of America      t ACGIH  American Conference of Government Industrial Hygienists  United States of  America        COSHH  Control of Substances Hazardous to Health  United Kingdom      Toxic fumes    If working with volatile solvents  toxic substances  etc   you must provide an  efficient laboratory ventilation system to remove vapors that may be produced     WARNING   Toxic fumes    A Do not use bleach to clean or disinfect used labware  Bleach  in contact with salts from the buffers used can produce toxic  fumes                    1 6 Waste disposal    Used labware  such as sample tubes  may contain hazardous chemicals or  infectious agents from the sample disruption process  Such wastes must be  collected and disposed of properly according to local safety regulations  For  information on how to dispose of the TissueLyser 
34. t 40  C       22 TissueLyser Il User Manual 04 2010    Installation Procedures       CAUTION   Damage to instrument    A Operation of the TissueLyser II at a temperature and humidity   outside the range specified may lead to damage to its  electronic and mechanical components as well as a reduction  in instrument performance                 Power requirements    The power line to the TissueLyser II should be voltage regulated and surge   protected  Make sure that the voltage rating of the TissueLyser II  see the plate  on the back of the instrument  is compatible with the AC voltage available at  the installation site     Grounding requirements    To protect operating personnel  the National Electrical Manufacturers  Association  NEMA  recommends that the TissueLyser Il be correctly grounded   earthed   The instrument should be plugged into an AC power outlet that has a  ground  earth  connection     4 3 Transporting and repacking the TissueLyser II  Carry the TissueLyser Il with both hands below the base of the instrument     Please retain all original packaging for the duration of the warranty period and  beyond  A charge may be incurred in the event of damage during shipment if  the original packaging is not used for the return of the TissueLyser II to QIAGEN  for repair        CAUTION   Damage to instrument  A Transport of the TissueLyser II using packaging other than the    original packaging may lead to instrument damage        Before returning the TissueLyser Il for re
35. tain infectious agents  You should be aware of the health  hazard presented by such agents and should use  store  and dispose of such  samples according to the required safety regulations        WARNING    A          Samples containing infectious agents    Some samples used with the TissueLyser ll may contain  infectious agents  Handle such samples with the greatest of  care and in accordance with the required safety regulations     Always wear safety glasses  2 pairs of gloves  and a lab coat     The responsible body  e g   laboratory manager  must take  the necessary precautions to ensure that the surrounding  workplace is safe  and that TissueLyser ll operators are  suitably trained and not exposed to hazardous levels of  infectious agents as defined in the applicable Material Safety  Data Sheets  MSDSs  or OSHA   ACGIH   or COSHH     documents     Venting for fumes and disposal of wastes must be in  accordance with all national  state  and local health and  safety regulations and laws       OSHA  Occupational Safety and Health Administration  United States of America    t ACGIH  American Conference of Government Industrial Hygienists  United States of    America        COSHH  Control of Substances Hazardous to Health  United Kingdom      TissueLyser Il User Manual 04 2010       Safety Information       1 5 Chemicals    WARNING   Hazardous chemicals    A Some chemicals used with the TissueLyser II may be  hazardous or may become hazardous after completion of the  sample di
36. the  TissueLyser Handbook     1  Make sure that the TissueLyser Adapter Set or Grinding Jar Set is securely  installed on the clamps of the TissueLyser II  see Section 4 4  page 24  and  that the transparent hood is lowered        WARNING   Moving parts    A Damage or personal injury can occur if the TissueLyser  Adapter Set or Grinding Jar Set is not securely attached     Ensure that both adapters or grinding jars are of similar  weight  Do not operate the TissueLyser Il with only a single  adapter or grinding jar installed                 P9    Make sure that the TissueLyser II is connected to a power outlet     ES    Switch on the TissueLyser Il using the power switch     4  Check that  on  appears in the  Memory  display  If necessary  press the   PROG  key repeatedly until  on  is displayed     5  Using the         or         key under the  Frequency 1 s  display  set the  oscillation frequency     Briefly pressing the         or     key changes the oscillation frequency by  0 1 Hz  Continuous pressing of the         or           key causes the oscillation  frequency to increase or decrease rapidly without any pauses  The  selectable range of frequencies is 3 30 Hz     For the optimal oscillation frequency for your application  see the  TissueLyser Handbook     6  Using the         or         key under the  Time min sec  display  set the  disruption time   Briefly pressing the         or     key changes the disruption time by 1 second   Continuous pressing of the        
37. ultaneously press the    SET     key and the           key under the  Frequency 1 s    display     2   bS  appears in the  Memory  display  In addition  the accumulated  operating time for the TissueLyser II is displayed in hours and minutes in the   Frequency 1 s  and  Time min sec  displays     3  Press the  STOP  key to cancel the display     5 4 Displaying the software version    To display the software version installed on the TissueLyser Il  follow the  procedure below     1  When the TissueLyser ll is in standby mode  simultaneously press the    SET     key and the         key under the    Frequency 1 s  display    2   S  appears in the  Memory  display  In addition  the software version is  displayed in the  Frequency 1 s  display    3  Press the  STOP  key to cancel the display     TissueLyser Il User Manual 04 2010 29    6 Maintenance Procedures    Perform the following regular maintenance procedures  i e   after each daily  use  to ensure reliable operation of the TissueLyser Il     Maintenance Procedures          CAUTION    A          Damage to the instrument    Do not use bleach  solvents  or reagents containing acids     alkalis  or abrasives to clean the TissueLyser Il           If solvents or saline  acidic  or alkaline solutions are spilt on the TissueLyser Il   wipe them away immediately     Do not autoclave any part of the TissueLyser Il  TissueLyser Adapter Set  or  Grinding Jar Set        WARNING    A          Risk of electric shock    Do not open any pan
    