Home

Aurora Forensic Sample Clean

image

Contents

1. Profiler Plus range from 100 350 bp thus the Quantifiler is not always a good indicator of suitability for STR amplification It is probable that DNA from Q1 2 was too degraded or fragmented for STR amplification It is also worth noting that because the Q1 2 lysis area included the area previously sampled for Q1 1 the concentration of DNA recovered from Q1 2 might be expected to be less than that for Q1 1 however the distribution of touch DNA on the surface of the object remains unknown Figure 2a shows that amplification of Q1 1 was either below the standard curve or non existent in undiluted and 1 5 diluted PC extracts but was successful with undiluted output from the Aurora Forensic Sample Clean up Protocol These results are consistent with the presence of PCR inhibitors in the PC extracts Figure 1b shows that PCR amplification of the internal positive control IPC was completely inhibited in the undiluted Q1 1 sample even after two rounds of PC extraction and Q1 1 at a dilution of 1 5 amplified over 5 cycles later than the DNA standards resulting in failed STR amplification Although IPC amplification for the 1 5 dilution of the second PC extract was uninhibited there was insufficient DNA present for STR amplification indicating the necessity of removing inhibitors without further diluting the sample When the same PC extract from Round 2 was further purified using the Aurora Forensic Sample Clean up Protocol amplification of the undilu
2. Use deionized water or very weak buffer solutions when resuspending or eluting a sample for use with this protocol DNA in the sample must be free to migrate in response to electric fields Example results To demonstrate the ability of the Aurora to provide highly purified DNA from an inhibited sample genomic DNA from an object from a criminal investigation was analyzed Investigators suspected that this object was used as a weapon in a murder and thus two profiles were expected to emerge those from the victim and the perpetrator In this case phenol chloroform PC extraction was used to obtain DNA from the object followed by a second PC extraction and dilution in an attempt to remove more inhibition after which the Aurora Forensic Sample Clean up Protocol was used To ensure maximum DNA recovery and avoid losses through swabbing the tip of the suspected weapon where touch DNA was believed to have been was submersed in lysis buffer 10 mM Tris HCl pH 8 0 10 mM EDTA 50 mM NaCl 2 SDS and Proteinase K 0 4 mg ml To avoid cross contamination from the blood to touch DNA as much as possible a volume was chosen to cover the end opposite to the victim s blood Q1 1 Figure 1 However as a precaution a second volume of lysis buffer was chosen to cover a larger surface area in case more touch DNA could be retrieved the second extraction Q1 2 was performed after the lysate for Q1 1 had been collected Blood Internal Standards BIS 20 ul
3. as well as the Troubleshooting section in the Aurora user manual for details in resolving this failure mode Yield may also appear low if the sample contains contaminants that are bound to DNA Diluting the input sample and performing multiple SCODA runs may help Post SCODA a dilution series of DNA templates will indicate any remaining PCR inhibition See Failure Mode 2 to address this problem Ordering and support For support for Aurora protocols or cartridges or to order additional cartridges please contact support borealgenomics com 10 SCODA conditions These conditions are pre programmed in the AURORA FORENSIC SAMPLE CLEAN UP PROTOCOL SP file that accompanies this protocol guide and are intended for reference purposes Note that electric current or power values that slightly exceed these expected values may not indicate a problem Cartridge Running buffer 0 25x TBE Sample volume 5 ml Expected sample lt 100 uS cm conductivity Injection Injection voltage 600 V Injection charge 3000 mC Expected current 9 13 mA Expected average power 8 12 W Expected voltage drop lt 10 across the gel Wash 6 Channel SCODA field strength 70 V cm SCODA cycle period 4s Duration 2 5h Wash strength 20 Expected current 20 30 mA Expected power 7 9 W Focus SCODA field strength 70 V cm SCODA cycle period 4s Duration 1 5 hrs Expected curre
4. 3 BOREAL GENOMICS Aurora Forensic Sample Clean up Protocol O 2015 Boreal Genomics Inc All rights reserved All trademarks are property of their owners http www borealgenomics com support borealgenomics com 1 604 822 4111 Table of Contents o A a A A A aa a aadal 4 Example resUltS cd 4 Safety pulida o E A T A EEE 8 Procedure Over Wena laa ia 8 Preparing the Mpio A ad 8 Loading your sample and running the AurOFA ccccnnonoococnnnnncnnnononnnnnnnnnnnnnonnnnnnnnnnnnnnonnnnononnnnnnnnnnnnnnnos 9 Troubleshooting icsi E EN OEA AAA A A AA 9 Ordering and SUDOR a A E e T E 10 SCODA CONdItIONS aih rie tds 11 Introduction This protocol is for removing inhibitors that remain in DNA samples after purification by other methods which is necessary in order to obtain a Short Tandem Repeat STR profile for forensic studies The protocol works with input from phenol chloroform preparations commercial silica based spin columns and other preparative methods that yield DNA with relatively low conductivity see input specifications below This protocol is appropriate for recovering DNA molecules 0 3 50 kb in length The presence of shorter or longer molecules will not impair yield of 0 3 50 kb fragments Input Sample Specifications Volume 5 ml Conductivity lt 100 uS cm when diluted to 5 ml The conductivity of the sample after dilution to 5 ml must be lt 100 uS cm which is similar in conductivity to 0 2x TE or 0 1x TBE
5. and a blank were also prepared and extracted in the same manner as positive and negative controls respectively Object Lysis Buffer Level for Q1 2 Lysis Buffer Level for Q1 1 Figure 1 Illustration of the object processed twice with two different volumes Following an overnight incubation at 56 C for each extraction the object was removed and the collected Q1 1 and Q1 2 lysates were diluted by a factor of 10 with Filtered Autoclaved Water FAD H20 The diluted supernatants were then concentrated to 500 ul in an Amicon Ultra 15 30 KDa centrifugal filter catalogue no UFC903096 at 4 000 g The supernatants were each transferred to a 2 ml heavy phase lock gel Inter Medico 5P2302830 where 400 ul of phenol chloroform isoamyl alcohol 25 24 1 was added and centrifuged at 13 000 rpm for 5 min A final extraction step with N butanol in place of phenol chloroform was performed using the same centrifugal conditions as above The supernatants were then concentrated and washed twice with TE buffer in a Vivacon 500 ETO centrifugal filter Vivaproducts VNO1H42ETO before being eluted in 50 ul of the same buffer SCODA and Speed Vac Purification The samples eluted from PC extraction Round 2 were each diluted to 5 ml with FAD H 0 in the sample chamber of a 210 0001 AA D cartridge the cartridges were loaded onto the Aurora and the Aurora Forensic Sample Clean up Protocol was run In each case a 60 ul output was collected and further concen
6. can continue but yield will be decreased In general yield decreases with increasing conductivity over 100 uS cm Solution To solve this problem adjust the DNA extraction protocol to reduce the amount of salt in the sample Some suggestions are depending on the method employed to elute samples from silica column based methods in nuclease free water or 0 1x TE buffer resuspend DNA pellets in nuclease free water or 0 1x TE buffer and to increase the number of ethanol based washes in precipitation methods 2 Failure Mode PCR reactions remain inhibited even after processing with the Aurora Solution Increasing the time of the wash block may help improve contaminant rejection Some dilution of the Aurora output may still be necessary for best amplification This protocol may also be less effective at rejecting contaminants that are complexed with or bound to DNA The addition of low conductivity additives such as proteinase K prior to injection may help reduce the amount of bound contaminants 3 Failure Mode Yield is too low Solution This failure mode can have several causes In the case where the sample did not contain sufficient DNA try processing more of the sample through the lysis and desalting steps If using a silica based column eluting off the column with larger than recommended volumes can assist in the recovery of small amounts of DNA Yield may be reduced if the sample conductivity is too high See troubleshooting Error 1
7. ctor of 5 The beige bar indicates that DNA amplified below the standard curve b PCR inhibition represented as the delay in IPC threshold cycle C from averaged IPC Cy values of DNA standards QL_1 Aurora D3 1358 Wa 100 200 300 i A A QL_1 Aurora DESITTS D21STI 7 100 200 300 E MTT Q1 1 Aurora D55818 D75820 100 200 300 L 1110100981000 E Figure 3 STR profile recovered from the object after the Aurora Forensic Sample Clean up Protocol Scale and range of Relative Fluorescent Units RFUs and allele calls were removed to protect the identity of those involved in the investigation Safety guidelines Wear gloves during all stages of the protocol Avoid skin contact with all reagents Cartridges are made from non hazardous plastics metal TBE buffer and agarose Appropriate precautions should be taken if hazardous samples are used with the cartridges Procedure overview e Prepare the sample e Prepare the cartridge e Load the sample e Prepare the Aurora e Begin the run Preparing the sample Process samples using your choice of DNA purification methods following required safety procedures Alternatively if your sample meets input sample specifications it may be run directly without pre processing Aurora DNA purification works best with low conductivity samples To maximize DNA yield take steps to keep conductivity of the Aurora input as low as possible These may include eluting in deionized water in
8. nt 30 45 mA Expected power 9 11 W 11
9. stead of buffer or repeating wash steps depending on the purification method Additionally it may be possible to improve the yield of silica columns by eluting in larger than recommended volumes for subsequent concentration with SCODA Dilute the extracted product to 5 ml with a low conductivity buffer like 0 01x TBE 0 01x TE or nuclease free deionized water Final sample conductivity must be lt 100 uS cm Running more conductive samples will decrease yield Loading your sample and running the Aurora Please refer to the Aurora Reusable Cartridge Handling Manual 106 0014 for detailed instructions on how to load your samples into the Aurora Reusable Cartridge 211 0004 AA D prepare the Aurora instrument and run the Aurora protocol Please select the 106 0008 BA D FORENSIC DNA CLEAN UP PROTOCOL file when asked to select the protocol sp file appropriate for your application Troubleshooting Please see the Aurora user manual for more information about troubleshooting machine faults 1 Error The Aurora control software warns that the sample is too conductive Running high conductivity samples will result in lower yields and may cause gel damage and other issues during the run The Aurora instrument will give the warning Injection Conductivity test failed Sample conductivity is too high Injection might fail for highly conductive samples Conductivity for a 5 ml sample should be lt 100 uS cm Immediate Remedy The run
10. ted sample was successful even after subsequent concentration by speed vac Note that analysis of the output from Aurora for Q1 1 at a 1 5 dilution was not performed due to limited amounts of material available for STR amplification Visually the brown pigmentation seen in Q1 1 after PC extraction and the fact that organic matter covered the suspected weapon suggests that the contaminant responsible for PCR inhibition may be humic acid Figure 3 is an electropherogram of the mixed partial profile that was obtained from sample Q1 1 using the Aurora Forensic Sample Clean up Protocol Amplification was unsuccessful for 1 5 and 1 10 dilutions of the Q1 1 PC extracts These extracts were likely too inhibited and or contained insufficient DNA for STR amplification consistent with results from qPCR The STR profiles for the Q1 1 BIS and negative control extracts were compared confirming that no cross contamination had occurred data not shown a DNA Recovery Quantifiler Total DNA ng o o N w 2 m 1 1 PC Round1 PC Round2 Aurora Q1 1 b PCR Inhibition IPC IPC Delay of Cycles 1 1 PC Round1 PC Round2 Aurora Q1 1 Figure 2 Quantifiler Human DNA Quantification Results a Quantifiler Human DNA quantities of Q1 1 from two phenol chloroform extracts PC Round 1 and 2 and the Aurora Forensic Sample Clean up Protocol Undiluted samples are denoted as 1 1 while 1 5 indicates that the sample was diluted by a fa
11. trated by speed vac Vacufuge Eppendorf at room temperature This was only possible because the Aurora Forensic Sample Clean up Protocol enables successful PCR amplification from very highly concentrated DNA samples due to its thorough removal of PCR inhibitors Finally the lyophilized samples were re suspended in 15 ul TE buffer for quantification and STR amplification Quantification and STR Profiling Quantification and STR profiling of DNA was performed using the Applied Biosystems Quantifiler Human DNA Quantification Kit catalogue no 4343895 and the AmpFISTR Profiler Plus PCR Kit catalogue no 4303326 respectively 25 ul PCR reactions were carried out for both assays and amplified according to manufacturer specifications STR products were separated and visualized on an ABI Prism 310 Genetic Analyzer catalogue no 310 00 100 120 W with POP 4 polymer catalogue no 402838 Electrokinetic injection and electrophoresis run times were set to 5 s and 24 min respectively allowing for the resolution of the 400 bp peak of the GeneScan 500 ROX sizing standard catalogue no 401734 Data Collection Software v3 1 and Genemapper ID v3 2 Applied Biosystems were used for data collection and analysis Although quantification suggested that sufficient amounts of DNA were obtained for Q1 2 for PCR amplification no profile was obtained data not shown The amplicon size in the Human Quantifiler Kit is 62 bp while those from

Download Pdf Manuals

image

Related Search

Related Contents

Proline Plus 5 stage Reverse Osmosis  Numero - Amped Wireless  ダウンロード - ノーベル・バイオケア・ジャパン  Fiche sécurité Wironit  EXs16 Funk and Soul Brass Manuel d`utilisation  Vision Fitness X1500 User's Manual  Product User Manual  「わんにゃんらいふ 携帯型自動環境見守り計&超音波トレーナー」1月27  remarque  

Copyright © All rights reserved.
Failed to retrieve file