Home
Additions and corrections to the User`s Manual for the COOLSCOPE
Contents
1. Modify 4 5 Replicating a set of observation conditions saved on the CF CARD in P 46 as indicated below Add change the underlined text and figure TULLE LLL TTT 3 Click on the RECALL button A list of Recall folders is displayed List of Recall folders NOTE All clear the Memory before reading out the CF card appears if any of the observation conditions remain undeleted from the COOLSCOPE memory The message will disappear automatically 4 Select the name of the folder in which the desired set of observation conditions is saved Reading of stored observation conditions starts Click the CLOSE button to close the Recall folder list and halt the reading operation 5 Click on the NUMBER button with a number corresponding to the set of observation conditions to be replicated The set of observation conditions for the clicked NUMBER button is replicated In the absence of a specimen slide in the COOLSCOPE the image saved on the CF CARD is replicated TALLEY Additions and corrections to the User s Manual for the COOLSCOPE lt 13 gt Modify 3 1 NetWork Settings submenu in P 62 as indicated below Add change the underlined text and figure TALL Postback Seat ties fy Ereakilas iiri Internal Server a bakle FOP cl a External Server IP akra emar i i Liker Ih Ss ma 12 cho 1 1 Submenu buttons 2 Return to Window button 3 Enable Ne
2. an asterisk and register authorized IP addresses by the procedures given below Up to eight IP addresses Registering IP addresses authorized to connect 1 Insert a CF card into the PC 2 Create a folder named NET on the CF card 3 Create a file and list of IP addresses authorized to connect Example of data 192 168 10 2 192 168 10 10 192 168 10 36 192 168 21 You can use the wild card character In this example the wild card character represents all values from 1 through 254 4 Assign a name with the ip extension to the text file and save it to the NET folder Example of filename Autho ip 5 Insert the CF card into the COOLSCOPE 6 Connect to the COOLSCOPE from the external PC using Telnet and execute a w_IPA command to register the IP addresses Refer to 3 5 Using Telnet Commands to Operate on the COOLSCOPE in Chapter VIII TLLTTLTLLTLTTLTLTE LTT LTLL TT TEEL LT TTTTTT Additions and corrections to the User s Manual for the COOLSCOPE lt 17 gt Modify 3 Date and Time in 3 2 Additional Settings submenu in P 65 as indicated below Change the underlined text TILT TLL 3 Date and Time Enter the date and time here NOTE The timer stops if the COOLSCOPE is left unpowered for an extended period Check the date and time prior to use if the power has not been turned ON for a month or more ILLITE TELLTALE lt 18 gt Modify 3 2 2 Set
3. settings are reloaded the next time power is switched on TALL TL TLL TLL lt 8 gt Add the following text to 3 Capturing an Image in P 38 TLL LTTE NOTE e When a scale is displayed on the screen the scale is also added to the saved image However the scale cannot be displayed when the image size is 2560 x 1920 TALLEY lt 9 gt Add the following text to 2 2 Closing opening the aperture in P 42 TTL NOTE e Right click to enable adjustment in 5 increments range 25 to 100 TALULA TLL Additions and corrections to the User s Manual for the COOLSCOPE lt 10 gt Add the following text to 4 1 Enabling the saving of observation conditions to the CF CARD in P 44 TTL NOTE If you use the POWER OFF button to switch off the COOLSCOPE the observation save enable disable conditions are saved These settings are reloaded the next time power is switched on TALL TLT LTT TTTTELTTTTE lt 11 gt Modify 4 2 Saving the current observation conditions in P 44 as indicated below Change the underlined text TALL TT OILIOLLA 1 Click on the MEMORY button A number appears on the macro image At the same time the RECALL button w with the same number is displayed in the Memory area Numbers are assigned automatically incremented by 1 TALL ELTL OILIOLLA OLLOLLU LLLA Additions and corrections to the User s Manual for the COOLSCOPE lt 12 gt
4. M 307E o3 cr 2 supplement Nikon Additions and corrections to the User s Manual for the COOLSCOPE Thank you for purchasing a Nikon product This booklet contains important material which you should read together with the the User s Manual for the COOLSCOPE Additions and corrections to the User s Manual for the COOLSCOPE lt 1 gt Modify 7 Unpacking and Unclamping in P 7 as indicated below Change the underlined text TULL 7 Unpacking and Unclamping 1 Check the contents to ensure that the package contains the following COOLSCOPE Hex wrench _Pewereerad lt lt deleted User s manual Compact flash card To avoid discoloration and fading of printed characters do not use any organic solvents e g alcohol ethanol thinner on painted plastic and printed areas Use of a silicon cloth is recommended To clean the display and other peripherals please refer to their respective instruction manuals ILLITE lt 2 gt Modify 7 Unpacking and Unclamping in P 8 as indicated below Add the underlined figure MALT q a l Enlarged i Section B Section C The cover 1 on the right side is detached TATTLE lt 3 gt Modify 2 Names and Functions of Buttons Observation buttons in P 26 as indicated below Add the underlined text TUTTLE 2 lee MAG button Changes the magnification P 32 Right click first to change the optical zoom magnification the
5. focus is complete e The color of the progress bar will be the same as the color selected for the scale Progress bar TLL lt 6 gt Add the following text to 4 2 Changing magnifications in P 32 TULLE LTTE TTT NOTE Right click first to change the optical zoom magnification then to perform autofocus This function is enabled only when the optical zoom magnification is adjusted from low to high magnification If the magnification is changed from high to low right clicking will not activate autofocus TATTLE TLL LTTE Additions and corrections to the User s Manual for the COOLSCOPE lt 7 gt Modify 2 Specifying the Image Storage Format Storage Medium and File Name in P 36 as indicated below Add change the underlined text TLL On delivery the COOLSCOPE has the following settings set for image storage format storage medium and filename No Item Initial setting 1 Storage format Image file format BMP Image size 1280 960 Image quality FINE 2 Storage medium Compact Flash Card Folder name F File name IMAGE Change the image storage format storage medium and or file name as necessary The settings remain in effect until the power switch of the COOLSCOPE is turned off If you use the POWER OFF button to switch off the COOLSCOPE the current settings for image file format image quality and storage medium are saved These
6. lowing image the COOLSCOPE network setting is set to Read Only Refer to 3 1 NetWork Settings submenu in Chapter VII TALL E LTTE Ts Additions and corrections to the User s Manual for the COOLSCOPE lt 21 gt Modify 2 Selecting intervals for updating the micro image in 3 3 4 Working with the WEB screen in P 78 as indicated below Change the underlined text and figure ILLITE TLL TTT TTT 2 Selecting intervals for updating the micro image Select intervals for updating the micro image Eight options off 5 sec 10 sec 30 sec 1 min 2 min 5 min and 10 min are available to choose from for updating intervals The updating intervals are set at off when the WEB screen opens 1 Click on the INTERVAL TIMER list box and select the desired intervals from the list IntervalTimer e TALL L TT L TT TELTT TT T LT lt 22 gt Modify 2 Selecting intervals for updating the micro image in 3 3 4 Working with the WEB screen in P 78 as indicated below Change the underlined figure LLL 3 Starting pausing updating of the micro image 1 Click on the RUN button to start micro image updating Titer al T imer or Furi 2 Click on the RUN button again to pause updating NOTE The RUN button becomes a Pause button when updating of the micro image starts TELLTALE LTTE lt 23 gt Add the following text to 4 Cha
7. n to perform autofocus Note This function is enabled only when the optical zoom magnification is adjusted from low to high magnification TUTTLE Additions and corrections to the User s Manual for the COOLSCOPE lt 4 gt Modify 2 Names and Functions of Buttons Observation buttons in P 26 as indicated below Add the underlined text TALI 6 APERTURE button Adjusts the aperture coarse movement P 42 Right click to adjust the aperture fine movement TALLTTLTLTTTTELTLTTEL TLL T TTT LLT TT TTTEELTT TT TTLT lt 5 gt Add the following text and figure to 3 Bringing the Image into Focus Automatic Focusing in P 31 TTL NOTE e When the AF button is used for the first time following specimen slide replacement always select 5 x as the magnification Note also that the AF button must be used with the specimen positioned at the center of the field of view e If itis difficult to bring the image into focus display a high contrast area of the image at the center of the field of view and then click on the AF button again Note that right clicking on the AF button enables confirmation of the AF area For more information please refer to 1 9 Displaying the AF area on the micro image in Chapter VI e During autofocusing a progress bar is displayed on the micro image to indicate the process status When the bar reaches the end auto
8. nging the observation position in 3 3 4 Working with the WEB screen in P 78 TATTLE NOTE This function is disabled if the COOLSCOPE network setting is set to Read Only TULLE LLL TTT 12 Additions and corrections to the User s Manual for the COOLSCOPE lt 24 gt Modify 5 Changing magnifications in 3 3 4 Working with the WEB screen in P 79 as indicated below Add change the underlined text and figure TTT 5 Changing magnifications 1 Click on the MAG button Magnification F Changing magnifications automatically adjusts the lighting brightness and aperture s 10x 20x 40x NOTE Changing magnifications from the PC automatically updates the micro image This function is disabled if the COOLSCOPE network setting is set to Read Only TALL ELTLLTTLTLTLL LLL T TTT LL TTT LLULLU lt 25 gt Modify 6 Bringing the image into focus in 3 3 4 Working with the WEB screen in P 79 as indicated below Add change the underlined text and figure TALITY 6 Bringing the image into focus The image is brought into focus with 5x selected as the magnification 1 Click on the AF button The image is automatically brought into Facies AF focus NOTE When the AF button is used for the first time following specimen slide replacement always select 5 x as the magnification Note also that the AF button must be used with the specimen positioned a
9. ns and corrections to the User s Manual for the COOLSCOPE lt 29 gt Add the following text and figure to 3 3 4 Working with the WEB screen in P 82 TTT 16 Setting autofocus to be performed immediately after optical zoom magnifications When the AF function is enabled autofocus activates automatically after any changes in optical zoom magnification Note that this function is enabled only when the optical zoom magnification is adjusted from low to high magnification 1 Click the AF check box to enable the AF function lie LAF 2 To cancel click the AF check box once again to uncheck NOTE After autofocusing is completed the micro image is automatically updated TULLE TLL TTT LTTE LTT lt 30 gt Modify 3 4 1 Using FTP Commands to Get COOLSCOPE Image Files from a PC in P 80 as indicated below Add change the underlined text TULL 3 Execute the cd RAMIBOX command Eight file names that are virtual names will be displayed e 1MG1280 bmp e IMG1280 jpg e 1MG640 bmp e IMG640 jpg e 1MG320 bmp e 1IMG320 jpg e MACRO bmp e MACRO jpg The time of day for the files displayed at this time is O hours 0 minutes The actually acquired files are given the time of day when you issued the Get command TULLE 15 Additions and corrections to the User s Manual for the COOLSCOPE lt 31 gt Modify Setting External Servers of the COOLSCOPE Network Settings in P 91 as indicated bel
10. ow Change the underlined figure TLL TLL TLL LTTE Wetiiork eisirean Retkork Settings G crate rh H ai ety l FTE Se daska Melee Doried unar O Amic t m IF Internal Server T stirs sim nt CEE EA E E ler IK On HS am 10 chor Fa od ee ma 0 ch Enables disables FTP client i function i bnbl Ft e O om erir External Server Ip airs mE Sevan ib oO User Ti annem Ler 12 che I Sets the IP address of ee pe e a the server ara aer aacroapugmnggpgmoim aces ce aes TTL lt 32 gt Modify 6 Transporting Clamping in P 110 as indicated below Add the underlined figure TELLTALE Enlarged The cover 1 on the right side is detached TUTTLE TLL TLL TTT 16
11. t the center of the field of view Ifitis difficult to bring the image into focus display a high contrast area of the image at the center of the field of view and then click on the AF button again Executing auto focus from the PC automatically updates the micro image This function is disabled if the COOLSCOPE network setting is set to Read Only TTT 13 Additions and corrections to the User s Manual for the COOLSCOPE lt 26 gt Modify 7 Adjusting the image focus in 3 3 4 Working with the WEB screen in P 80 as indicated below Add change the underlined text and figure TH 7 Adjusting the image focus 1 Click on the FOCUS button TR AF gt EA NOTE Adjusting the image focus from the PC automatically updates the micro image This function is disabled if the COOLSCOPE network setting is set to Read Only TULLE TLL TTL lt 27 gt Add the following text to 9 Adjusting the aperture in 3 3 4 Working with the WEB screen in P 80 TELLTALE NOTE This function is disabled if the COOLSCOPE network setting is set to Read Only TULLE lt 28 gt Add the following text to 10 Adjusting the image brightness in 3 3 4 Working with the WEB screen in P 80 TALITY NOTE This function is disabled if the COOLSCOPE network setting is set to Read Only TALLTTLTTLTTELTL TLL TL LLL TTEL LTT TTTELTTTTT 14 Additio
12. tWork 4 Read Only 5 Internal Server 6 Enable FTP Client 10 IP Authority 7 External Server 8 SAVE button a O n aaoqgctngnpnagpgom mAr 9 Default button TLL TELTLTLLTLTLTLLTLLTTTTLTTTEL LT L LTT TTTLLTTTTTTLLTTTE lt 14 gt Modify 4 Read Only FTP WEB disable Write Control in 3 1 NetWork Settings submenu in P 62 as indicated below Change the underlined text TALI 4 Read Only FTP WEB disable Write Control Allows you to write enable or write disable the Web and FTP server To operate the COOLSCOPE from another PC through the Web screen as when magnifications are changed uncheck this box so that the asterisk disappears Note however that this option may be left write disabled if you simply wish to retrieve images Write COOLSCOPE operation disabled Write COOLSCOPE operation enabled TELLTALE TELLTALE Additions and corrections to the User s Manual for the COOLSCOPE lt 15 gt Modify 7 External Server in 3 1 NetWork Settings submenu in P 64 as indicated below Change the underlined text TULLE GateWay Sets the gateway address TLL TLL TLL TTT TLL lt 16 gt Add the following text to 3 1 NetWork Settings submenu in P 64 TLL 10 IP Authority Used to specify IP addresses permitted to connect to the COOLSCOPE To designate IP addresses allowed to connect to the COOLSCOPE display
13. ting Essential ltems Network Settings in P 72 as indicated below Change the underlined figure ILLITE TLL TTT LTTE Footie Setting Enables disables rity FTE Gi dlabla Writs i Cont 5 network operation Sets the IP address of the Iniemal Server 1p atiras COOLSCOPE Eb Hark ler I kua Eod G mle ptr clen m Exlermal Server TP atiras Faa zi Ea OOODODoooe POEOMOOQbDoOOoeOm aces ce ea e ee se i i TULLE LLL Additions and corrections to the User s Manual for the COOLSCOPE lt 19 gt Modify 3 3 3 WEB screen functions in P 77 as indicated below Change the underlined text and figure TLL Micro image Controller 1 RELOAD button 2 INTERVAL TIMER list box 3 RUN button 4 MOVE button 16 AF Check box 5 MAG button 6 AF button 7 FOCUS button 8 DIGITAL ZOOM button 9 APERTURE button 10 AE button 11 DOWNLOAD button 12 TYPE list box 13 SIZE list box 14 QUALITY list box 15 ANNOTATION check box Macro image TTT TELLTALE TLL 10 Additions and corrections to the User s Manual for the COOLSCOPE lt 20 gt Add the following text and figure to 3 3 3 WEB screen functions in P 77 TATTLE TTT 16 AF Check box Allows automatic enabling disabling of autofocusing after optical zoom magnification adjustment NOTE If the Web screen displays the fol
Download Pdf Manuals