User manual CuroCell Cirrus®
Contents
1. Care of sweden Supporting Life User manual 2015 03 03 CuroCell Cirrus Alternating pressure system User manual item number 95 001054 fa Oo CE as eSa SS Contents Important safety precautions Introduction Product description Assembly Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information 11 12 14 15 16 17 18 20 21 22 23 IMPORTANT SAFETY PRECAUTIONS Read all instructions prior to use WARNING To minimise the risk of fire electric shocks or bodily injury me The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts e g pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Do not use the pump in wet rooms or where there is a risk of the pump coming into contact with water or another liquid Never handle a pump that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorised service technician for servicing 3 DO NOT USE in the vicinity of or in contact with fire sources hot surfaces such as fire burning cigarettes or hot lamps Although the product has undergone fire te
2. products through incineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled It must instead be taken to a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information contact Care of Sweden or your local dealer 13 2 Returns and complaints Complaints are to be reported to the supplier Return of a product shall follow the supplier s instructions and can only be done after the complaint has been reported Damages due to the transportation should be reported to the transportation agent 23 bead Manufactured by Care of sweden Supporting Life Contact Phone 46 771 106600 Fax 46 325 12840 Email info careofsweden se Internet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns vag 4A SE 514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell Cirrus are trademarks of Care of Sweden AB Care of Sweden AB 2015 24
3. information category Heel function gt lt 1 Chlorine X Do not iron lt 2 gt gJ T Vv Foot placement y Y a Recycling Directly on bed base TX Do not wash Q Tumble dry Ry Do not dry clean IP__ IP class Ingress Protection Rating 10 Spare parts Art no CC 925531 CC 925532 CC 925533 CC 925544 CC 925535 CC 925541 CC 925542 CC 925549 CC 925539 CC 925529 CC 901001 CC 901002 CC 901003 CC 901004 CC 901005 CC 8217 CC 8202 If other spare parts are needed please contact Care of Sweden or your local dealer for further assistance Description Top part Steel Top part Steel Top part Steel Top part Steel Top part Steel Top part Steel Top part Steel Top part Steel Top part Steel Top part Steel Size 80x200 cm 80x210 cm 85x200 cm 85x210 cm 90x200 cm 90x210 cm 105x200 cm 105x210 cm 120x200 cm 120x210 cm Air cell 80 cm wide mattress Air cell 85 cm wide mattress Air cell 90 cm wide mattress Air cell 105 cm wide mattress Air cell 120 cm wide mattress Fuse set of 2 Air filter 20 11 Options and accessories 11 1 Options Sturdy transport bag Transport and storage bag Top part CuroCell LF low friction Top part with increased sliding function can be used for example heavy immobile patients 11 2 Accessories Cover Steel Size 80x200 cm 80x210 cm 85x200 cm 85x210 cm 90x200 cm 90x210 cm 105x200 c
4. or unlock the panel 5 ALARM MUTE Press the ALARM MUTE button to temporary turn off the alarm Should the situation not be resolved within 5 minutes the alarm will be activated again Please refer to section 8 to troubleshoot any problems with the alarm A Power failure alarm During a power failure both the sound and light alarm will go on Turn it off by pressing the ALARM MUTE button B Low pressure alarm When the low pressure light is on the pressure in the mattress is below normal Both the light and sound alarm will activate The alarm will sound until the ALARM MUTE button is pressed If the problem remains see chapter 8 for trouble shooting 6 SERVICE This light indicates a technical failure in the pump First check that the air filter is clean and that the pump is not covered by anything such as a pillow or a blanket If the problem remains please contact the manufacturer or your local dealer for further information 7 POWER Power on off 10 3 Assembly Installation When unpacking the mattress check that none of the parts are damaged In the event of damage contact the manufacturer or your local dealer before using the product 1 Place the mattress on the bed base Note e Make the bed with sheets for added comfort e Check that the CPR valve is closed yxy 2 Hang the pump at the foot end of the bed or place it on a flat J a sturdy surface as Route the power cord so that there is no risk of it causin
5. since our products are designed to work well together Please contact Care of Sweden or your local dealer for accessories or more information on our products 12 Warranty 12 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 12 2 Service and repairs This CuroCell mattress is a medical device in accordance with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance Like other technical devices a pump that is properly cared for works better and lasts longer Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty A warranty claim must be submitted to Care of Sweden before the product is returned Please contact the manufacturer or your local dealer if you have questions regarding warranty and service 12 3 Life expectance This product s expected life is estimated to be 5 years 22 13 Other information 13 1 Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell
6. the mattress and use of the Evac function Please read section 1 2 for more information To ensure proper function of the Evac cover it is important to regularly check the quality of the mattress Check for damages that could cause problems when evacuating ee BHRHEHB ERE HBR HERR ERR 2 Product description 2 1 Pump and mattress CuroCell Cirrus has eight alternating air cells and these cells are encased in HR foam high resilient foam On top of the HR foam closest to the user is a thin layer of visco elastic foam 1 Mattress 2 CPR rapid deflation 3 Pump connector 4 Pump 1 Air cells 2 Foam core 3 CPR rapid deflation 4 Tubes 2 2 Pump The CuroCell Cirrus pump applies alternating pressure that is adapted to the weight of the user The air cells inflate and deflate with different cycle times Cross section of mattress 1 2 3 1 Quick connector unit air 3 Fuses 2 pcs 6 Bed hooks 2 Front panel 4 Off On 0 1 7 Air filter 5 Power cord 2 3 Pump functions control panel 2 Hy i i Gye le Time 4 C omisar i Theda 6 1 COMFORT SETTING Suitable air pressure is set using the comfort dial Guidelines for user weight can be found in the table on the pump 2 CYCLE TIME There are four selectable cycle times Press the button to select cycle time 3 PROGRAM THERAPY Change the mode by pressing the button the chosen mode is indicated by the green light The mod
7. 0 90 non condensing L fAthmospheric pressure 700hPa 1060hPa _ Classification Class l Type BF PA MATTRESS SPECIFICATION Model Cuecas Dimensions BxLxH L 80 85 90 105 120x200x20 cm Weight ft kg width 85 cm NOTE The manufacturer retains the right to modify the product specification at any time WARNING Follow the operations and maintenance instructions carefully in order to prevent bodily injuries as well as damages to the product or other equipment m m m m m m m m m Ba 9 2 Labelling The mattress are labelled with product name production date article number size care and wash instructions EAN bar code as well as space for own labelling The product also displays which side is up down and head foot 18 9 3 Symbol Definitions Article No Read the user manual pa Not in domestic refuse follow recycling instructions aVola 0 130kg Recommended user weight when using Evac function Machine wash at 70 C me Do not tumble dry Wipe off Temperature range Manufacturer GROY anna RL Ss Ynna Group of function gt x S lt S lt Zi Recommended user weight Counteract shear force Machine wash at 95 C Class Il equipment Double Insulated i The mattress should only be used lying in the lengthwise direction on the mattress 19 XXXX XX XX Year Month Day CE CE marked acc to MDD 93 42 EEC Category Patient
8. RRR RRR RRR RRR RRR RRR RRR RRR RRR 1 Introduction CuroCell air mattresses operate according to the alternating pressure principle a technique that is well proven and been used for a long time Alternating pressure is when the pressure regularly alternates between the mattress cells to replicate body movement Variable pressure cycles are set up where there is little or no pressure This is optimal for skin improvement or to prevent skin damage Always read the user manual prior to use 1 1 General information This CuroCell mattress is a medical device in accordance to MDD 93 42 EEC According to statutory regulations made by the authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving there products We would be very grateful for all information involving accidents or incidents relating to our products by reporting them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell Cirrus combines alternating pressure with foam and can be used in all types of health environments to prevent pressure ulcers and as an aid in the treatment of category 4 1 It is suitable for lightweight and sensitive people or to people with pain The mattress has a built in heel function for a
9. Use the comfort buttons 1 10 to set a suitable air pressure based on the table on the pump Note The pump always starts in the Autofirm mode When the mattress has reached the required pressure the pump is set automatically to the previously used setting or to the settings made in the Autofirm program Please refer to section 2 3 for more details NOTE raised pressure should be increased 1 2 levels 12 kg 0 40 50 60 70 80 90 100 110 120 140 150 160 170 180 190 230 oj jxjajaj swin 1 F h O e We recommend the Seat inflation mode for all situations where the head of the bed is e When only parts of the mattress are under load e g if the patient is an amputee the e Always check the set pressure with a hand check see 4 4 for more information 4 4 Hand check Hand check is used to ensure that the mattress is properly inflated This should be done regularly once per shift and after changing the position in the bed or changing the comfort settings Insert your hand between the air cells directly underneath the user s pelvic region use a vertical hand There should be a clearance of 2 3 fingers between mattress base and the patient If not you need to adjust the pressure setting 4 5 CPR Rapid deflation of the air cells To empty the air cells pull the CPR Cardio pulmonary rescue band placed between mattress and pump In case of an emergency where CPR is necessary fo
10. dded relief of heels The mattress is placed directly on the bed base The recommended user s weight is up to 230 kg For the Evac function the maximum user weight is 130 kg Important For users between 180 230 kg the recommended use is changed to treatment of pressure ulcers up to category 2 Cleaning of the mattress shall be made between replacements of each user We recommend that position changes are made regularly The time interval must be evaluated by the responsible personnel depending on the patient s status diagnosis and general condition NOTE e When used together with positioning pillows read the pillow user manual carefully for correct positioning of the patient e f you are using this product with an evacuation sheet it is the user s responsibility to ensure that it is safe to use for evacuation e Be careful when using bed rails or other protection on the bed so that the mattress does not become trapped or damaged e The mattress must not be lifted with a person lying on it or used for any other type of transport other than a possible evacuation using an evacuation sheet e This mattress is unsuitable for x ray examinations because of the risk of blurred images or artefacts that may lead to diagnostics error 1 National Pressure Ulcer Advisory Panel European Pressure Ulcer Advisory Panel and Pan Pacific Pressure Injury Alliance Prevention and Treatment of Pressure Ulcers Quick Reference Guide Emil
11. ean area which is suitable for washing 5 Wipe off the cells all tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 6 Place the air cells back into the foam core Check that all the tubes and connectors are secure and tight before closing the cover 7 Wipe off work surfaces with a mild cleaning solution for this purpose such as alcohol with or without tensides lsopropanol 70 oxidating solutions or chlorine max 1 If cells for any reason have loosened from the tubes these must be replaced according to drawing in chapter 2 1 FOAM CORE Clean the affect area with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Press out the water carefully Let the foam core dry in a warm ventilated area without direct exposure of sunlight The foam core needs to be completely dry before it is used again NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 6 Storage Store the mattress on a flat surface or carefully rolled folded with the zipper closed Always store at room temperature in a dry area Make sure the mattress is stored where it will not be ex
12. ent in the bed should not be at the very top of the mattress There could be a risk of hitting the head in the bed when pulling down Pulling in the stairs e Start with pulling an empty mattress down the stairs in order to practice the technique e When training with a person on the mattress a string or rope can be fasten in the handles at the head end Another person can then walk behind the mattress and help out if problems should occur e It is advisable to practice in the same stairs where an evacuation can be necessary General e The main purpose of the training is not to reach high speed but to learn the correct technique Training covers e Because of the heavy strain on the equipment when evacuating always check the seams and handles at the end of training We recommend that equipment used for training is marked with Training equipment and is only used for this purpose Explanations e If the Velcro straps are not used there is a risk of the patient sliding rolling down from the mattress e Pulling at the foot end reduces the strain on the puller which means faster and safer evacuation is possible e Pulling in the stairs is the most difficult part of the evacuation and should be practiced extra carefully e The arms of the patient shall be inside of the Velcro straps This way reduces the risk of the user grabbing hold to something out of fear NOTE There is a difference in maximum user weight between use of
13. es order is Autofirm Alternate Static Alternate seat inflation Static seat inflation A Autofirm Autofirm is used to fill the mattress with air completely for firm support The pump always starts in this mode When the correct pressure is reached it changes automatically to the previous setting or to the pre setting outlined below Presetting features for Autofirm process 1 Program Therapy Select the required program 2 Comfort setting Set the required air pressure 1 10 3 Cycle time Select from 10 15 20 or 25 minutes Note The preset Cycle Time Comfort Level and Therapy mode will be illuminated to indicate the settings during the Autofirm process Autofirm LED is also illuminated After mattress is inflated Autofirm LED will be turned off automatically B Alternate In this mode the pressure alternates between the cells and there are four different cycle times to choose from C Static When Static is chosen the alternating process is terminated and the pressure between the cells evens out The function can be used in association with patient care repositioning transferring in and out of bed The function is turned off by pressing the button again or it will automatically go back to alternating mode after 20 minutes D Seat inflation This mode gives extra support when sitting up and can be set for both alternating and static mode 4 PANEL LOCK Press down the Panel lock button for about 2 seconds to lock
14. evacuation The straps are folded in the protective pockets on the side of the mattress with the ends sticking out to ensure that they are easy to pull out Te Le ECOL CEO ECCT ee eee ECL CeCe Cee er OC rr cor Cert WARNING Person to perform an evacuation needs to have the necessary training Always check the equipment cover and mattress before commencing to make sure an evacuation can be carried out safely It is the user s responsibility to ensure that it is safe to use for evacuation 5 Evacuation instructions Down from the bed e Inform the patient about what will happen e Lower and lock the bed in position e Disconnect the tubes from the pump e Pull out and fasten the Velcro straps e Use the outer handles at the foot end of the mattress and turn the mattress out from the bed Slowly pull the patient down on to the floor tf SV On an even surface e Pull on the outer handles at the foot end of the mattress be A e Working position should be a little bit leaning back be Down the stairs e Walk backwards until the main part of the mattress has slide pass the top stair e Turn the back to the patient and look in the moving direction e Hold one hand on the stair handle and pull with the other hand on the middle handle of the mattress e Walk down the stairs It is important to maintain speed without running Training practise To bear in mind when training Pulling down from the bed e The pati
15. formed by authorised technicians Do not get the pump wet or submerge it in any liquid PUMP Wipe of the pump with a damp cloth and a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 If using a different agent choose one that does not affect the pump housing MATTRESS COVER Covers can be wiped off with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner 87 O u Bat BS Multi layer covers should be divided before washing Wash with like colours Chlorine and phenol based cleaners could have a negative impact on the PU surface zy and should be avoided If chlorine is used we recommend a mixture of max 1 INNER COVER SOFT o O Ill Bx X Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 14 5 2 Reconditioning PUMP See section 5 1 for instructions MATTRESS 1 Remove the covers See instructions in chapter 5 1 for cleaning covers 2 Disconnect the tubes from the pump 3 Carefully pull the air cells out from the foam core 4 Place the air cells and the tubes on a cl
16. g ye someone to trip being run over by the bed wheels or getting Ros f pinched when the bed is raised or lowered 3 Connect the air tubes from the mattress to the side of the Ora Ey pump There is an audible click sound when the quick SY connectors are properly attached 4 Check that the switch on the pump is in the O position Off Connect the plug to an approved electrical socket 220 V 11 4 Operation e f Note Always read the user manual prior to use 4 1 Start up 1 Set the power switch on the side of the pump to 1 On and start the pump by pressing the power button on the front panel 2 Air starts to fill up the mattress It takes about 20 minutes until the mattress is ready for use Note The pump always starts in Autofirm mode l E Oe A 3 The light for Low pressure is lit until the mattress has reached the set pressure When the light goes out the system changes mode to previous settings and the mattress are ready to be used 4 2 Program selection There are four different programs Please refer to section 2 3 for more details As a standard setting we recommend an alternating mode and a 10 minute cycle time 4 3 Comfort settings The air pressure in the mattress can be adjusted according to the patient s desire for comfort ALWAYS consult the prescriber or other authorised personnel for the appropriate setting The mattress is now ready for use
17. llow the local instructions Note Make sure that the CPR connector is properly connected after use 4 6 Alarms When the pressure drops to an abnormally low level the Low Pressure diode lights up together with an audio alarm Please refer to section 8 for troubleshooting The alarm is deactivated by pressing the ALARM MUTE button on the pump s control panel If the problem is not resolved within 5 minutes the alarm will start again The diode remains on until the pressure level is remedied and reset The pump is also fitted with a Power Failure alarm An alarm sounds and flashes when the pump is disconnected from the electric power Turn off the alarm by pressing the ALARM MUTE button WARNING 4 7 Transport function If the patient must be moved in bed the tubes can be connected together detached from the pump and the mattress put in transport mode We recommend that this function is only used for short periods of time 13 If the air pressure remains low check the mattress and tubes for damage Repair if necessary or contact the manufacturer or your local dealer for help 5 Cleaning and reconditioning It is important to follow the instructions below before using the product for another user Consult the Clinician responsible or the supplier for help and instructions when unsure 5 1 Cleaning WARNING Do not open the pump housing risk of electric shock Service and maintenance must be per
18. m 105x210 cm 120x200 cm 120x210 cm Steel Evac cover Size Bottom part Steel CC 925111 CC 925112 CC 925114 CC 925115 CC 925117 CC 925118 CC 925120 CC 925131 CC 925122 CC 925133 Top part Steel Complete cover Steel CC 925631 CC 925632 CC 925633 CC 925644 CC 925635 CC 925641 CC 925642 CC 925649 CC 925639 CC 925651 Bottom part Evac Inner cover Soft CC 928011 CC 928012 CC 928013 CC 928010 CC 928015 CC 928020 CC 928018 CC 928029 CC 928010 CC 928031 Complete Steel Evac 80x200 cm CC 926041 CC 926210 CC 925841 80x210 cm CC 926042 CC 926211 CC 925842 85x200 cm CC 926043 CC 926212 CC 925843 85x210 cm CC 926044 CC 926213 CC 925844 90x200 cm CC 926045 CC 926214 CC 925845 90x210 cm CC 926046 CC 926215 CC 925846 105x200 cm CC 926047 CC 926216 CC 925847 105x210 cm CC 926048 CC 926217 CC 925848 120x200 cm CC 926049 CC 926218 CC 925849 120x210 cm CC 926050 CC 926219 CC 925850 Top part CuroCell LF low friction Art no Size Art no Size CC 916210 80x200 cm CC 916215 90x210 cm CC 916211 80x210 cm CC 916216 105x200 cm CC 916212 85x200 cm CC 916217 105x210 cm CC 916213 85x210 cm CC 916218 120x200 cm CC 916214 90x200 cm CC 916219 120x210 cm Art no Description Size 58 900905 Transport bag 105 120 cm 58 900906 Transport bag extra high 80 90 cm 58 900911 Transport bag extra high 105 120 cm 21 11 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden
19. p side is set to 1 Alarm is on and Check that the quick connectors on the air tubes are properly mattress is losing air connected to the side of the pump Check that the CPR valve is closed Check that all tube connectors are properly seated and connected Open the mattress and check for leaks from air cells or tubes The patient is bottoming out The pressure might be incorrectly set Adjust according to the weight table and wait a few minutes Check again using a Hand Check see section 4 4 Check that the CPR valve is closed Some cells have less air This is normal since the cells are alternately inflated and deflated when the system is set to alternating mode Pump is noisy Check that the pump hangs correctly on the bed Resonance vibrations are felt can occur in parts of the bed Remove the pump and listen whether this makes a difference Can be resolved by putting the pump on a flat steady surface or by placing a towel between the pump and bed If your questions are not answered by the above information please contact Care of Sweden or your local dealer 17 9 Technical description 9 1 Technical specification PUMIP SPECIFICATION Model o o Groce Vari Voltage peoveaov 50 60hz3 S Current consumption OFA Fusetype TAL 250M Cycletime Pf 15 200 25min o Dimensions BxLxH P9Ixu7x0m S Weight Rg po Btorages ts8c50 C ooo Bhippings t58C 70 C Po Humidity Operations 10 90 non condensing Pd Storage 1
20. posed to thrusts impact or other things that can damage it Avoid direct exposure of sunlight If the mattress is to be stored more than 3 months we recommend it is stored on a flat surface not folded or rolled up 15 7 Maintenance 7 1 General 1 Check the power cord and plug for damage and scratches 2 Check that the mattress cover is intact and that the cover and air cells are correctly assembled 3 Check that the air flow alternates between each outlet on the side of the pump 4 Check for damaged or pinched tubes and connectors Contact the manufacturer or your local dealer for spare parts 7 2 Changing fuses d eo Pull the plug out of the electrical socket before replacing fuses Remove the cover from the fuses with a screwdriver Pull out the fuse holder Fit a new fuse of the same type T1AL 250V and screw the same type and screw the protective cover back in place There are two fuses Unscrew the round safety plate on the back of the pump Lean the pump backwards in order for the filter to fall out from the holder Place the new filter in the filter holder and replace the safety plate If the pump is used in a dirty environment the filter should be checked regularly 16 8 Troubleshooting Problem Solution Pump does not start Check that the power cord is connected to an approved electrical socket Check the fuses Make sure the power switch on the pum
21. sting it could become damaged if it comes into contact with fire sources 4 Position and use the product in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails Never use the product if the pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality through a hand check see section 4 4 Also check that the comfort setting is correct as indicated in the table on the pump 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress are protected with a hygiene cover Please try to avoid the simultaneous use of multiple hygiene cover with this mattress Because of the risk vapour permeable can be destroyed 9 The hygiene cover does not allow liquid or air to penetrate but is vapour permeable Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Prevent damage to the hygiene cover from sharp objects m m e If the cover is equipped with side handles these are for lifting the mattress All other use takes place under your own liability and is not covered by the product guarantee Lae eee eee RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR Ui Lae eee eRe RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR RRR
22. y Haesler Ed Cambridge Media Osborne Park Western Australia 2014 1 3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well planned and well structured working environment be used In our guideline Action plan you will find examples which can be applied and put into practice when using our products Mattress guide This practical guide gives you as a care giver help in choosing the correct mattress from the Care of Swedens range CuroCell Cirrus is a mattress in function group C This information can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to info careofsweden se Evac cover A cover with protected zipper on all sides divisible top and bottom part i Functional handles for evacuation moving and transport Colour grey black 0 130kg 1 4 Evacuation instructions Evac is a mattress cover with a built in function for evacuation of bed bound users In case of fire or other emergency there is no time to think about what to do next Therefore it is very important that the safety equipment as well as the evacuation procedures are well planned and known in advance The Evac cover is equipped with three handles on the head and foot end On the sides there are handles with Velcro straps These straps are to be used to fixate the user to the mattress before commencing the
Download Pdf Manuals
Related Search
Related Contents
Atlantic 144 DVD User's Manual Cat 5 High-Density Dual Rail LCD KVM Switch KL1508 Porter-Cable Model 548 User's Manual ATH-BT08NC 取扱説明書 Panasonic KX-TG5622 Telephone User Manual LinkMotion for IPG YAG Laser CD3200 Thyristor Unit 取扱説明書 - Changer & Dresser Corp. DEO FUEL GUÍA DEL USUARIO Copyright © All rights reserved.