1. Oximeter Plus inc OPERATOR S MANUAL VER2 0C16 Fingertip Pulse Oximeter General Description Thank you for purchasing the Oxi Go QuickCheck pulse oximeter This oximeter can be used to measure your Blood Oxygen Saturation in SpO2 and your pulse rate It should be used for Spot Monitoring only and not for continuous monitoring The Oxi Go QuickCheck Finger Unit spot check oximeter is not a medical device and should only be used by healthy individuals who are performing non medical activities such as sports and aviation It is ideal for use during sports activities mountain climbing and piloting airplanes It is sold without a prescription only to consumers who do not require a medical prescription It is not to be used for wellness applications Package Contents Oxi Go QuickCheck With 2 Way Display 1 One Oximeter 2 One lanyard 3 Two AAA batteries 4 One user manual Product Operation Scope 1 Oximeters are sensitive to motion artifacts Keep hands still while taking a reading N Oximeters require sufficient blood flow to obtain proper readings Poor blood circulation can result in inaccurate readings If your hands are cold or you have poor circulation warm your hands by rubbing them together or use another method before attempting to obtain a reading A tourniquet blood pressure cuff or other blood flow hindrances may also result in inaccurate readings w Fingernail polish or acrylic nai
2. SpO 80 99 2 70 79 38 69 no definition PR 30 99BPM 2 BPM 100 235 BPM 2 Operation Temperature 41 F 104 F Storage Temperature 4 F 131 F Ambient Humidity lt 80 in operation lt 93 in storage 6 Automatic power off The oximeter will power off automatically after 8 seconds when the finger is removed 7 It is best to store the product in a place where ambient temperatures range from 20 C to 55 C 4 F 131F 8 It is recommended that the product be kept in a dry place A damp ambient might affect its lifetime and even might damage the product Declaration EMC of this product comply with IEC60601 1 2 standard The materials which user can come into contact with are non toxici and comply with ISO10993 1 15S010993 5 and 1SO10993 10 Guidance and manufacture s declaration electromagnetic emissions for all EQUIPMENT and SYSTEMS Guidance and manufacture s declaration electromagnetic emission The Pulse Oximeter is intended for use in the electromagnetic environment specified below Emission test Compliance Electromagnetic environment guidance RF emissions Group 1 The Pulse Oximeter uses RF energy only for its CISPR 11 internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emission Class B The Pulse Oximeter is suitable for use in all CISPR 11 establishments including domestic establis
3. fully before use 7 The pulse oximeter has no SpOz alarms it is not for continuous monitoring as indicated by the symbol In addition to items described in the Product Operation Scope inaccurate measurements may be caused by FACTORS INCLUDING BUT NOT LIMITED TO 1 Autoclaving ethylene oxide sterilizing or immersing the sensors in liquid 2 Significant levels of dysfunctional hemoglobin such as carbonxy hemoglobin or methemoglobin 3 Intravascular dyes such as indocyanine green or methylene blue 4 SpO measurements may be adversely affected in the presence of high ambient light such as direct sunlight In bright light conditions cover the sensor area if necessary 5 Excessive user movement 6 High Frequency electrosurgical interference and defibrillators 7 Venous pulsations 8 The user has hypotension severe vasoconstriction severe anemia or hypothermia 9 Fingernail polish or false fingernails may cause inaccurate SpO readings 10 Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components including batteries Technical Specifications 1 Display mode LED Measuring range for SPO2 70 99 Measuring range for PR 30 235 BPM PR display mode bargraph Low power indication LX Battery standard Two AAA 1 5V Alkaline Battery Power consumption Less than 40mA Resolution 1 for SoO2 and 1BPM for Pulse Rate a S o p Measurement accuracy
4. hments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Possible Problems and resolutions Problems Possible reason Solution Abnormal 1 Finger is not inserted correctly 1 Take a new reading as per Spozor PR 2 User s Oxyhemoglobin value is instructions above readings too low to be measured 2 Possible oximeter failure 3 Seek medical attention 1 Retry by reinserting the finger 2 Try not to move Spo or PR 1 Finger might not be inserted deep reading is enough unstably 2 Finger is trembling or user s body is in moving The 1 Batteries power low or no 1 Please replace batteries Oximeter batteries installed 2 Please reinstall the batteries cannot be 2 Batteries might be installed 3 Please contact with local powered on __ incorrectly customer service center 3 The Oximeter might be damaged Indication 1 The product is automatically 1 Normal light is powered off when no signal is 2 Replace the batteries suddenly off detected longer than 8 seconds 2 Lower power 1 Low power 1 Change new battery Error3 or E 2 Receiving tube being shielded or 2 Please contact with local rror4 damaged together with broken customer service center Displayed connector 3 Please contact with local on screen 3 mechanical Misplace for customer service center 4 Please contact with local customer service center receive emission tube 4 amp circuit malf
5. ls obstruct the light transmission and may also result in inaccurate readings Your finger must be clean for proper reading The Oximeter must be clean for a proper reading If a reading is difficult to obtain switch to another finger or to the other hand N oe ot Ss There are a number of other conditions which may lead to an inaccurate reading including but not limited to recent medical tests that included an injection of dyes use of arterial catheters a weak pulse low levels of hemoglobin in the blood low perfusion the quality of your pulse elevated levels of dysfunctional hemoglobin the strength and type of light that you are in while using the Oximeter and the existence of cell phones radios and fixed transmitters within certain ranges of the oximeter during use 0 The Oximeter will not alert you if your readings are out of normal range Operating Instructions Battery Installation 1 Install two AAA batteries into the battery compartment minus signs in the compartment may be caused to the oximeter Match the plus and If the polarities are not matched damaged 2 Slide the battery door cover horizontally along the arrow shown as below Do not force Install batteries in right polarity Incorrect placement may cause damage to the bracket Remove the batteries if the Oximeter will not be used for long periods of time 3 Battery indicator symbol on front panel display will light when battery vol
6. tage is too low for normal operation of the Oximeter Replace batteries when indicator symbol lights 4 Dispose of batteries as per local ordinances 2 Place one of your fingers into the rubber opening of the Using the Oximeter Oximeter your finger should touch the bottom portion 1 Open the clamp as shown in the picture gt before releasing the clamp When you put your finger into the Oximeter your nail must be facing upward towards the display portion of the oximeter Press the switch button one time on front panel to turn the oximeter on 4 Keep your hands still for the reading Do not shake your finger during the test It is recommended that you do not move your body while taking a reading Read data from display screen There may be times when the signal strength is to low see heart rate bar graph indicator below for a proper reading Please switch to another finger or hand and follow instructions in this manual for low profusion blood flow 7 When no signal or low signal is detected the oximeter will power off automatically in 8 seconds Maintenance 1 Clean the oximeter and sensor with a soft cloth dampened with isopropyl alcohol Do not pour or spray any liquids onto the oximeter and do not allow any liquid to enter any openings in the device Allow the oximeter to dry thoroughly before reusing The rubber inside of the Oximeter is composed of medical grade rubber It is non toxic and is not harmful
7. to the skin 2 The pulse oximeter requires no routine calibration or maintenance other than replacement of batteries Brief Description of Front Panel Fower Switch Low Voltage Lamp Pulse bar graph Oxygen Saturation Pulse Rate The height of the bar graph indicates the intensity of the pulse and signal strength You should have at least 3 bars for a proper signal Two display modes By pressing the power on button during measurement you can get two display modes The display can be positioned to read towards you or facing away from you for others to read Display Facing Towards You Display Facing Away From You Keep the oximeter away from young children Small parts such as the battery door batteries etc may be hazardous if swallowed 2 The lanyard may cause strangulation in conditions that may cause it to twist around the neck Precautions for use 1 Do not use the pulse oximeter in situations where alarms are required The device has no alarms 2 Explosion hazard Do not use the pulse oximeter in an explosive atmosphere 3 The pulse oximeter is intended only as a non medical device 4 Check the pulse oximeter sensor application site frequently to determine the positioning of the sensor adequate circulation and skin sensitivity of the user 5 Do not stretch the rubber while applying the pulse oximeter sensor This may cause inaccurate readings or skin blisters 6 Read the manual care
8. unction Symbol Definitions Symbol Definition Type BF applied part N Attention consult accompanying documents SpO 2 Arterial Oxygen saturation v BPM Pulse rate BPM Low power indication No SpO Alarm SpO02 C Power switch SN Serial No Limited 1 year warranty Note The illustration used in this manual may differ slightly from the appearance of the actual product Manufactured for Oximeter Plus Inc 18 Milburn Lane Roslyn Heights NY 11577 Phone 516 626 6226 Made in China