Home
Operating Instruction for Moisture Analyzer PMX series USA
Contents
1. F6 Zeroing Tarring of analyzer indication jan for directional keys Scrolling arrow up a Exit to menu one level higher mininin Enter setting chosen parameter Scrolling arrow down for ENTER PRINT and ESC keys Enter Confirm the value fee Esc Escape and leave function without changes Esc 19 8 USER MENU OF MOISTURE ANALYZER User menu is activated by pressing SETUP key In menu it is possible to calibrate the analyzer set transmission parameters set the date and time and their format and other usage parameters of program In this part of manual program possibilities of moisture analyzer and methods of setting and change of particular parameters will be described 8 1 Calibration of moisture analyzer 8 1 1 Calibration of moisture analyzer If the moisture analyzer is used as laboratory analyzer make the calibration before test Calibration determines relations between value shown by the analyzer and standard mass calibration mass which is load and corrects indication if it is necessary Additional equipment of the moisture analyzer is F2 weight to make the calibration Type Calibration mass PMX 50g 1071101 Setup 13 47 56 P1 gt 01 gt Ext calibr wkkwk function ki 03 cara sasannan function Enter MENU group P1 Calibration Setup 04 Tomp calibr k function a 05 Print report 1 on Scroll
2. If first digit is flashing it can be changed Introduce new external weight using functional keys Confirm calibration mass Analyzer will start the calibration giving successive commands Enter on display Analyzer will inform about determining mass of empty pan and next to place the calibration mass of declared mass Put declared mass on the pan and press key Enter to confirm The instruments returns to menu to group P1 Calibration According to previous point move to weighing mode gi 1 External mass must be 34 of max capacity 8 1 3 Temperature calibration Calibration of drying chamber is a process which task is to calibrate temperature sensor of moisture analyzer To carry out the temperature calibration of moisture analyzer place the control thermometer in drying chamber hole as show on the picture below hole thermometer SAN pd ext calibi eawaaeae fou Enter MENU P1 Calibration 02 User calibr deseteteteteteie function 03 Calibr test sxaaxeax function SCFOll to parameter 04 Temp calibr 04 gt Temp calibr k e function 05 Print report 1 on Calibration Factory Calibration T 4 1 point 07 59 Calibration Factory Calibration T I heal ate ta Press the navigation key Calibration Factory Calibration T Set temp value C 4 130 8 1 4 Calibration test During calibration test value of calibration mass is compared to present result
3. I Li Li Li 13 46 42 0 standard 1 printout 2 11 printout 3 1 printout 4 i 1 4 I 11 Operator can set up 5 different versions of printouts which are sent to external device printer computer through socket RS 232 one STANDARD with data set in the procedures GLP four NOT STANDARD freely programmed 30 8 6 1 Standard printout The instrument sends after pressing the ENTER PRINT key through RS 232 display load mass and declared variable in submenu P2 GLP Exemplary printouts in weighing mode 62 690 g Date 22 10 2001 Date 22 10 2001 Time 13 04 23 Time 13 16 49 User WILK User WILK Project XW 456 Project XW 456 62 690 g Balance 10 62 685 g Question mark before load mass indicates that result of weighing is not stable There are three parts in the printout e Headline with information about chosen drying procedure and start mass of the sample e Line partial time of drying and result of the moisture e Footer with the final mass total drying time and final result 8 6 2 Nonstandard printouts Rules about creating not standard printouts 1 user can create own 4 printouts 2 for each printout give text number which starts number e g Printout 1 Start 1 and text number which finishes the printout e g Printout 1 Stop 40 In that case text contents is printed from1 to text 40 3 then write in text contents to lines 1 40 It is recommended to use the PC
4. Manual switch off With this switch off criterion the measurements process continues until you stop it with the keys START STOP Timed switch off With this switch off criterion the measurement lass until the present drying time has elapsed 13 6 Analysis of drying profile In the first type the drying profile is asymptotic The amount of moisture lost assumes a constant value and no longer changes after drying time With this drying profile repeatable determinations of the moisture content are always simple 68 The measurement result then corresponds exactly to the constant value of the asymptote It is also correspondingly easy to find a suitable switch off criterion In the second type drying runs quickly at the start and then flattens out The moisture content never assumes a constant value The causes of such a drying profile can be as follows The sample exhibits thermal decomposition the decomposition products other volatile components can lead to superimposed profiles owing to their slower vaporization than water The difficult to volatilize components lead to a slow continuous decrease in weight Measurements results of such a drying profile can be optimized eLowering the temperature can slow down the decomposition reaction eThe selection of a suitable switch off criterion can allow recognition of the end of the analysis at the desired break point of the drying curve e The selection of a constant drying time often
5. difficult vibrations draughts set up filter as slowly or very slowly parameter 01 Filter for 4 or 5 Efficiency of filter working is different for different range of filter working Filter works slower during achieving weighing mass and faster if mass is in set range of filter parameter range of filter working is accessible from level of service menu user has no access to it 8 4 3 Setting the autozero operation Programmable function AUTOZERO controls and corrects zero indication automatically Parameters which define the range and time of autozero are set up by l the manufacturer and there is no access to it for the user 28 AUTOZERO function is turn on if the measurement starts at precise zero Turn off the autozero function depends on the unstable result if there is no load on the pan This function is very helpfull during slowly pouring the substance Switch on off AUTOZERO function in the parameter P4 02 8 4 4 View of the temperature inside the drying chamber 10 11 01 Setup 13 Scroll to parameter P4 03 Temperature P4 gt 01 Filter 3 no 02 Autozero tion 03 gt Temperature fun 04 Negative 0 dis Press the Setup key Temperature inside wa Ba the drying chamber is displayed To return a NON to previous display press Esc key 22 C 8 4 5 Negative Parameter determining method of data indication on the display Depending on used display set the parameter in a appropriate wa
6. group P6 Printouts and press the F2 key to enter menu group Then by means of cursors scroll to demanded parameter and pressing the F2 key to activate procedure of introducing text By means of keyboard write the text 35 max 8 marks and confirm by the Enter key Repeat the operation for rest of the texts 8 6 4 Selection of printouts According to point 5 2 6 enter menu group P6 Printouts For parameter P6 01 Printout Nr select one of the values standard printout 1 printout 2 printout 3 printout 4 khOND o After selection suitable value confirm by the ENTER key 8 6 5 Setting non standard printouts To set up e g Printout 1 Enter group of the parameters P6 Printouts according to point 5 2 6 In the parameter 01 Printout Nr introduce 1 Printout 1 according to point 13 2 2 and in the parameter 02 Printout 1 start declare beginning of the text number of the text which starts the Printout 1 You can introduce the values from 1 to 80 After introduce number of beginning of the printout introduce number of the text which finishes the printout To do that Enter setting the parameter 03 Printout 1 stop and declare end of the text number of the text which finishes Printout 1 You can introduce values from 1 to 80 After confirmation and return to weighing mode and recording the changes each press the PRINT key prints out declared text 8 6 6 Non standard printouts in drying mode Scheme of creati
7. www radwag com i gt DIN EN ISO 9001 2000 CERTYFIKAT Nr 71 100 C20
8. 2 a 58 Press navigation key to display command with the question Press ENTER key to print all parameters and return to previous window Program number 1 Name FLOUR 100 Profile Standard Dry temp 120 C Switch off Automatic 3 Result Moisture content M Print int 30s Target None T coeficient 0 00 Program library Program P01 gt Name FOLUR 100 Profile Standard Dry temp 1120 C Switch off Automatic 3 Result Moisture content M Print int 130s Target None T coeficient 1 0 00 Default H Print bi Pressing Esc key will not print the ina parameters The software returns to previous display without any changes 10 12 Settings record to moisture analyzer memory Remember that all changes of the parameters are recorded in transitory memory of the instrument To record them permanently return to main display and use procedure of record 59 Procedure Program library gt Program gt P01 gt Name FOLUR 100 Profile Standard Dry temp 1120 C Switch off Automatic 3 Result Moisture content M Print int 30s Target None T coeficient 1 0 00 Default H Print bi e Press Esc The software displays command with the Save program question for the operator 2 Pressing the ENTER key will record the lt changes and display the library Program library i a TERES E P02 Program 2 P03 Program 3 P04 Program 4 P05 Program 5
9. 35 Text 27 3 a lolo 3 x Ne o gt 36 Text 28 37 Text 29 38 Text 30 Pai J a fe o 0 3 5 o s o x x l 39 Text 31 Selection of the text Enter paramaters group P6 Printouts 34 18 11 02 Setup 13 46 42 P6 gt 01 gt Printout No 0 standard 02 Header No 1 printout 2 03 Line No 11 printout 3 04 Footer No 05 Pr 1 start 06 Pr 1 stop 07 Pr 2 start 08 Pr 2 stop 09 Pr 3 start 10 Pr 3 stop 11 Pr 4 start 12 Pr 4stop 13 String 1 1 1 I 45 String 79 1 i printout 4 Scroll to the text from text 1 to text 80 Pressing the navigation key enter the setting the parameter if the parameter field was empty pulsing sign appears at the beginning If the text was in the field first mark of this line flashes Method of entering the texts a from balance keyboard Move up through collection of the digits letters and marks o 1 Move up through collection of the digits letters and marks o 1 Move to RIGHT side next pressing moves in RIGHT flashing sign if we do not write in the sign we introduce spaces in the text by pressing this sign Select sign to change by moving in LEFTside next pressing the button causes cancelling flashing sign and makes the previous sign flashes Confirm entered text ee PACE b by means of PS keyboard Pressing the F2 key give you access to main menu press F3 or scroll to
10. absorbent action Pre drying of the glass fibre filter and storage in which desiccators is necessary only for highly precise measurement result Skin forming and temperature sensitive substances The use of a glass fibre filter can be useful for temperature sensitive and skin forming substances In this case the sample to be dried is covered by the filter and receives a new surface This shields the surface of the sample against direct IR radiation Gentler heating of the samples is based on convection rather than on IR radiation Sugar containing substances Samples containing a large amount of sugar tend to caramelize on the surface In such cases ensure thin laser is applied Also select a moderate temperature Sample application on drying pan Loose materials They are dried in natural form Decomposition of the sample reduces smaller dispersions between following measurements Sample mass cannot be too big Ensure application of a thin and uniform layer thickness Correc Incorrect Liquids Semi liquid substances are dried in natural form Big amounts of fat in some substances hampers procedure of determining moisture In such cases use additional components e g quartz sand tissue paper filter Before the actual drying dry additional components in order to minimize their moisture Liquid Ja Hid Porous substance Liquid Solids 232222 Depending on structure of solid dense or loose determine
11. arrow key and PO1 Name FLOUR 100 replace the black arrow next to the il 1 key bi ewo parameter Switch off Switch off gt Automatic 3 Criteria of setting are described in see point 9 2 10 6 Setting type of displaying and printing results User can define kind of the printout during the tests One of four method can be selected M D R g Procedure Ly Program library Program P01 gt Name FLOUR 100 Profile Standard Press the down arrow key and replace the Dry temp 120 C pary Automatic 3 black arrow next to the parameter Result Result gt Moisture content M Print int 130s Target None If you select option Chart Scheme result M appears 54 10 7 Setting time of printout This parameter defines frequency of sending result during determining moisture Procedure Program library Program PO1 Name FLOUR 100 Press the down arrow key and replace Profile Standard A Dry temp 1120 C black arrow next to the parameter Print Switch off Automatic 3 int Result gt Moisture content M Print int 130s Target None If you select option Chart Scheme result M appears 10 8 Setting size of tested sample Parameter defines mass of the sample User chooses one of three options lack necessarily optionally Procedure Lt Program library Program P01 gt Name FLOUR 100 Profile Standard P
12. in the function of deviation under 1000 reading unit Calibration factor of drying chamber id different to factory factor of the chamber in more than 20 Value of the parameter apart from admissible range Inadmissible value of the parameter Errors in data transmission to the computer or the printer Wrong date overcrossed time 6 1 Transgressed admissible time to realize one operation e g setting to zero 18 MEMORY OF PERFORMED MEASUREMENTS Moisture analyzer enables memorizing Max 100 recently performed measurements containing following data such as date measurement time name profile drying temperature drying time start mass final mass and final result Measurements are recorded in pile After memorizing 100 measurements next one 101 measurement will be added in position 99 and whole list will be moved one position upper This will cause deleting of measurement from position 00 To enter in measurements memory Weighing b LI Set program Start fe SK Set AA an 0 0005 From weighing state press Display key Window containing particular measurements will appear Results a 00 28 06 05 10 19 26 01 28 06 05 11 15 30 02 129 06 05 09 45 30 Tea 03 129 06 05 09 56 45 Tea Data are arranged according to schedule of their performing if measurement was performed acc to Program from Data Base then next to date and time of measurement name of program is displayed To print data
13. influenced by the sample characteristics Hard brittle samples are mainly reduced in size by pressure impact or friction action cutting action Whatever the operating principle of a mill may be for subsequent moisture determination there must be no loss of moisture during the milling operation The quantitative recovery of the mill chamber should also be simple and complete Use of quartz sand To ensure an optimum drying process samples should always have as large an area as possible Results of substances which form crusts e g glucose syrup or pasty substances e g butter can be considerably improved by mixing with quartz sand Sample pans with a large volume and relatively high walls are needed for this Pasty fat containing and melting substances For pasty fat containing and melting substances use a glass fibre filters is advantageous to increase the surface area of the sample The liquid contained in the substance is uniformly and extensively distributed in the interstices between the fibres throughout the available area Pre drying of the glass fibre filter and storage in a desiccators is necessary only for highly precise measurement results 63 Liquid substances Liquid substances often tend to form drops on the sample pan owing to the surface tension of the liquid The use of commercial glass fibre filter shortens the drying time The glass fibre filter distributes the liquid sample over which wide area as result of its
14. l ilon 06 Contrast kkae eee function 07 Balance Id 12564 i 10101 Setup 13 47 56 P7 gt 01 Password o 02 gt Prog library 1 disabled 03 Beep 1 enabled 04 Language 4 English 05 Backlight tion 06 Contrast parii function 07 Balance Id 112564 After move to main display of the program with record changes in menu user can realize drying without selecting the drying program After set up the parameter P7 02 Programs Base on the value 0 user cannot use data base of drying programs recorded in the analyzer memory After pressing the key software automatically moves to setting basic parameters of drying 9 2 Way of setting the parameters Press key Test Menu A new display now appears with first view of shortened menu declaring parameters of drying profile of drying name of the program now appears above the picture and 4 With navigation key activate selection of profile in up left corner of the display Press either up or down keys and scroll to the desired program Standard Ramp Fast or Step 42 Press navigation key and set up the parameters of the program Next to first of parameters black triangle appears on the display Activate parameter to change it starts flashing Press arrow keys and set up the value Confirm with ENTER key value stops flashing If you selected ramp or step program set up other parameters If you sel
15. mass must be picked after examinatio ns and regarding to the quality of the examined substance Close the drying chamber after pictograph appears on the screen Drying starts automaticall y The moisture analyzer does the drying process in Drying 09 28 30 0 03 25 1120 c iie 0 2355 amp Display 09 28 30 2 ETE 84 C 57 Yo 03 120 C HIN 3 19 09 03 End results oe 5 265g gt 0 03 25 gt 4 378g 2 AMJ 305 2 57 vou End 49 set parameters Information about the process appear on the screen Printout headline is sent to communicati on port RS232 according to set options of printout parameters group P6 Printouts During drying actual results of the measurement and information about the process appears on the display testing time temperature in the chamber etc During drying results are sent to communication port RS232 automatically of printout lines according to set options of the printout group of the parameters P6 Printouts in set periods of time Press DISPLAY key and switch between kind of displayed result After the measurement the summation of the test and final result appear on the display And footer is sent to communication port RS232 according to set options group of the parameters P6 Printouts 10 DECLARING THE DRYING PROCEDURES AND USING THE PROCEDURES LIBRARY In this section you will learn how t
16. menu To change the method press DISPLAY button Drying time temperature in the chamber drying profile result in M D R diagram lost of mass in g are expose on the display 19 0903 Drying 09 28 30 EI 0 03 25 1120 120 c n fk aw man 30s 2 57 sm pi Jx 10903 Drying 09 28 30 bE 0 03 25 1120 c 12c ti AM E i 10 54 D amp 10903 Drying 09 28 30 bie 0 03 25 1120 12b c ti AM bio 2 10 ur amp So Drying 0 03 25 1120 c H A 0 2355 amp Changes are confirmed after return to weighing mode with procedure of confirming changes Se 25 s4 c a re BIO e DE 120 C 120 C AMJ Ti 30s n min After entering all changes press few times ESC until display indicates Save Enter Esc 10 12 45 Save setup Enter Esc Van 7 2 By means of computer After displaying the command choose one of possibilities ENTER save changes ESC without record keyboard PS type Each key on the analyzer keyboard has its equivalent on the PS keyboard according to below scheme for functional keys Description Balance Fl Switch on off drying process in selected program Start Stop Enter main menu L Setup 18 F3 Enter choice menu and select drying program Eq F4 Change of type of displayed data about drying sn F5 Transmission data to external device PRINT
17. of weighing It is just checking and the result have no influence on changes of factors After displaying results press key ESC Calibration test a Move to submenu P1 Calibration b Scroll to function Calibration test c Press key Right d The instrument makes test Check results Cal load on the pan Act calibration mass in moisture analyzer memory Dif difference between these two values e Return to weighing mode in point 6 of instruction manual 8 1 5 Temperature calibration Parameter enebling to carry out temperature sensor calibration in drying chamber 8 1 6 Printout of the calibration report After each calibration report can be printed The report is printed on external printer or sent to computer and recorded in the file to archivize P1 08 Print report 1 on report is printed P1 08 Print report 0 off without printing report If parameter is set up 1the report is generated and sent automatically 23 Example of the calibration report Contents of the printout depends on the set External calibration report parameters in the group P2 GLP see point Date 09 01 2004 8 2 of the manual The variables with values Time 12 24 10 i Balence ld 112284 ki set on 1 tak are printed l Calibr 50 000 g Report contains also calibration mass Old 49 999 g k defined during actual calibration description Difference 0 001 g Calibr Userld OR 13L
18. parameters User can set up various parameters depending on needs The parameters decides about work in group P7 Inne e g signal beep during pressing the key contrast on the display etc 8 7 1 Enter to group menu P7 Globals Enter to submenu group P7 Globals 10 11 01 Setup 13 47 56 P7 gt 01 Password 0 02 gt Prog library 0 disabled 03 Beep 1 enabled 04 Language 1 English 05 Backlight 1i on 06 Brightness kkkkkkk function 07 Contrast kkkkkkkk function 08 Analyzer ld 12564 08 Software rev IMBA 0 11 09 Par printout kkkkkkkk ifunction 10 Par receive kkkKKEKK function 11 ID setting xkkKKEKK function 12 ID autoprint 0 off 8 7 2 Setup the parameters Setting up the parameters can be realized the same as in previous menu groups 01 Password protection of the possibilities of entering set up the software 02 Prog library switch on off access to the library with programs 03 Beep decision if each key is signalling by beep signal on the analyzer keyboard 04 Language select language in description of the parameters Polish or English 05 Backlight backlight on the graphic display backlight improves readability of the displays 06 Brightness changes brightness of display backlit after enter to function by means of keys on keyboard it is possible to change brightness of display 38 07 Contrast change the contrast on the display after enter the function display appears and by means of keys chan
19. place indicator next to required measurement and press Setup key Results 00 28 06 05 10 19 26 01 28 06 05 11 15 30 02 129 06 05 09 45 30 Tea 03 129 06 05 09 56 45 Tea Measurement data will be displayed arrangement as on printout Results z 02 gt Name Tea Profile i Standard Dry temp 1 160 C Total time 0 43 20 Start weight 5 6783 g Final weight 15 3218 g Final result 5 26 M To print these data press PRINT key Data will be sent to RS232 port 81 19 PRINCIPLE OF USE OF MOISTURE ANALYZER FOR DRYING TEMPERATURE ABOVE 160 C For measurement temperatures in range of 161 C 250 C time of keeping the temperature during the measurement is determined in proportion with range of 1 hour for 161 C 20 min for 250 C During drying in temperature of 250 C Max temp is kept for 20 min then program automatically decreases the temperature without stopping the drying to 160 C Time of decreasing the temperature to 160 C is 20 min For rapid profile of Max drying temperature is 30 but no more than Max temperature set in factory menu For step profile time of keeping the temperature for particular steps is limited to 20 min Number of instruction LMI 02 01 10 1 1 ENG US 82 MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS A RADWAG 26 600 Radom Bracka 28 Street POLAND Central phone 48 48 384 88 00 phone fax 48 48 385 00 10 Sales Department 48 48 366 80 06
20. provides good measurement results eKeep the initial weight of the sample constant 10 20 14 COOPERATION WITH PRINTER OR WITH COMPUTER Printing to an external printer or computer will occur each time the PRINT key is pressed Signal with present presentation with 10 measurement units on display is sent to computer or printer The speed i of transmission is set up for 9600 bit s If external device printer computer needs different speed of transmission change parameter of speed in the menu see point 13 in instruction manual 14 1 Schemes of connection cables Analyzer socket DB 9 F printer KAFKA socket WM 560 3 TxD 1 RxD 5 GND 3 GND 7 8 zwarte Analyzer socket DB 9 F computer socket DB 9 F if there is no control of sending data 2 RxD 3 TxD 3 TxD 2 RxD 5 GND 5 GND 4 6 closed 4 6 closed 7 8 closed 7 8 closed 69 Analyzer socket DB 9 F computer socket DB 9 F and control sending data 2 RxD 3 TxD 3 TxD 2 RxD 4 DTR 4 5 GND 5 GND 6 Tara 6 7 RTS 7 CTS 8 CTS 8 RTS 9 Print 9 Analyzer socket DB 9 F computer socket DB 25 F 2 RxD 2 TxD 3 TxD 3 RxD 5 GND 7 GND 7 8 closed 4 5 closed 6 20 closed 15 LIST OF COMMAND COMPUTER MOISTURE ANALYZER Function Command Function Command Function Command Function Comm
21. substances absorb heat The sample can be overheated and if the temperature is 120 C dark color sample absorbs heat and reach temperature about 135 C In these cases define correcting factor of drying temperature Defining this factor is done by means of additional equipment set with the control thermometer and probe T coefficient 1 Program all parameters of drying 2 Distribute sample evenly and start drying process 3 After testing note the result 4 Take second sample and distribute it in thicker coating 5 Slip the probe through the cover and dip it in the sample press ON OFF key on the control thermometer 6 Start drying process 7 Check temperatures on the keyboard and control thermometer 8 After period of time for first sample read out two values 9 Count factor according to formula T measured T set 56 Setting the parameter M i Program library gt Program P01 gt Name Program 1 Profile Standard Dey iip 1420 C In the menu press the down arrow key and Switch off Automatic 3 replace the black arrow next to the paai i Mp ure content M Parameter T coefficient Target None T coeficient bi 0 00 Default i Print By means of functional keys of the moisture analyzer or PC keyboard select value of the factor Set up according to user s menu 10 10 Default settings of program Parameter enables setting all parameters as default determined by programmer Method of setting
22. test selected are now available to be run again You do not have to enter any new data In top part of the screen name of the program and number appear At the left side pictographs of parameters appear on the screen Further procedure is the same as for test in short drying menu see point 9 3 of instruction manual 12 PREPARATION OF SAMPLE FOR DRYING This section provides you with essential information on optimum results during tests In this section rules of how to select parameters of drying depending on kind of substance are presented 12 1 Moisture measuring in moisture analyzer During heating the sample loses its mass evaporates The moisture is measured The RADWAG moisture analyzer contains 2 parts precise analyzer and drying chamber Measurement in halogen moisture analyzer is faster and no additional calculations need to be done the result is displayed during test Independently of method of determining moisture also sample preparation and selection suitable parameters of test have special influence Sample size Kind of substance Drying temperature Drying time Speed of measurements can be enlarged after optimum setting parameters of the instrument depending on examined substances Optimum temperature and drying depends on kind and size of sample and expected precision of measurement The selection of these parameters can be done after experimental tests 12 2 Sampling and sample preparation Cha
23. to enter to choice of drying modes Pressing the Test Menu button accesses the Test library menu Library contains up to 100 tests All tests which are previously entered can be recalled and the parameters for the recalled tests when selected will be repeated How to escape menu without recording changes Press Esc key How to enter main menu of moisture analyzer After pressing Setup Access to user menu of analyzer is available How to exit the setup menu without making changes Press Esc key twice How to select the menu option Press either the up or down arrows keys and scroll to the desired options 10 11 01 P1 gt 01 Ext calibr 02 gt User calibr 03 Calibr test 04 Printreport Setup How to enter the chosen function in menu KERR function pies function Iekeeekskeek l function 13 47 56 tion After scrolling to desired function press button SETUP How to change parameters and functions ramus Bae Press the UP DOWN buttons to scroll through values of parameters from software or active mark changes flashing f eg in Name Press either the LEFT RIGHT arrows buttons and scroll to desired sign e g in name which we want to change parameters How to confirm changes in functions and Press PRINT ENTER button to confirm How to change displayed data during drying There are five methods of presentation results during and after drying in the moisture analyzer
24. to parameter 01 Ext calibr Setup mamas Press Right key External Calibration Command to take the weight off the pan is Clear the pan displayed pan must be empty After empty Enter pan press ENTER key tosis The analyzer shows command it External Calibration determines mass of empty pan and bie lacs then following command for user to put weight 50 0 g 4 05 01 04 Setup 10 12 45 External Calibration Load weight 50 0 g Enter a After the calibration the instrument returns to submenu P1 Calibration b Press ESC key many times until appears 20 Setup Save setup Enter Esc c After the calibration the instrument returns to submenu P1 d Calibration e Press ESC key many times until appears 8 1 2 User calibration sonwou Setup 13 47 56 P1 gt 01 Ext calibr eeeee function 02 gt User calibr eekekee function User Calibration Clear the pan 050104 Setup User Calibration Weight measuring Setup 101245 User Calibration Load weight 50 0 g Enter 3 User Calibration Weight measuring Jis User s calibration must be done by means of F2 calibration mass Enter menu P1 Calibration 03 Calibr test axkknkan function SCFOll to parameter 02 User calibr 04 Tomp calibr function 05 Print report ton Press Right key The analyzer shows command to introduce Key in weight value ol calibration mass 50 0000
25. weighing pan than weighing result should only be read if stability marker mu is displayed Do not put magnetic materials on weighing pan of a moisture analyzer This can defect the measuring set of the device Do avoid dynamic loading of weighing pan and do not exceed maximal capacity of the device Please consider applied tare mass when calculating mass of tested object Never use moisture analyzer in explosive area Moisture analyzer is not designed and adjusted to operate in explosive conditions Do not perform any modifications to the device 2 3 Principles of safety operation Use of PMX moisture analyzer conversely to safety principles and user manual may be hazardous to operator s health and life It is obligatory to acknowledge with safety principles listed in user manual use moisture analyzer onto to determine humidity content in samples and determination of sample mass Any other use of moisture analyzer may be dangerous either to the device or the user before switching on the moisture analyzer make sure that the nominal power of the device specified on its data plate is compatible with the supply in the mains to which moisture analyzer is connectable change of halogen can be carried out only in authorized service protect moisture analyzer against contacts with liquids as the area around moisture analyzer gets heated do not put any inflammable objects or substances in close distance to device substance contai
26. 06 project print off 12 0085 g P2 07 Id print fon P2 08 last cal print off 8 3 Setting time and date Parameters to set up date and real time clock in memory and to display and print them How to enter menu group P3 Date Time Enter menu P3 Date Time 10 11 01 13 47 56 P5 gt 01 gt Date format i 0 DA MO YR 02 Timeformat 0 24 hour 03 Time x function 04 Date kxkx function 05 Disp time 1 jon 06 Disp date l 1jon 8 3 1 Setting date format Press navigation key and come to submenu 01 Date format Set up one of the values 1 date format 2 date format Confirm by ENTER key 8 3 2 Setting time format Month Day Year Day Month Year Press Right key to enter menu 02 Time format Set up one of the values 1 time format 12 2 Confirm by pressing key ENTER hour time format 24 hour Format 12 hours It is distinguished by letters PM or AM on the printouts 26 8 3 3 Setting time Saus L 101101 Hours gt 08 Minutes 47 Seconds 56 With Right key enter to paramater 03 Time setting Set current time Hours gt 08 Scroll to value which is Minutes 47 changed Seconds 56 si a Hour Minutes Seconds Enter setting value first digit is flashing r Hours gt 08 Minutes 47 Seconds 56 4888 Set up new value using keys Selection of digit change or Value of digit change e n Confirm set value p
27. 74 disposable pan disposable pan pan bracket pan bracket pan handle pan handle chamber cover chamber cover chamber basis chamber basis insert Open the cover and remove all components disposable pan the pan handler and cross draft shield Use a lint free cloth for cleaning tetra chamois leather etc and mild cleaning agents Although the housing is extremely rugged and resistant to solvents never use abrasive cleaning agents or solvents After cleaning dry all subassemblies Do not let any dirt or liquids to interior of moisture analyzer Ensure that no liquid enters the interior of the instrument Replace the components after cleaning according to shown scheme Cleaning the temperature sensor To ensure correct measurements pay attention to cleanness of temperature sensor Clean the instrument with taking all precautions To clean it use smooth cloth and mild Do not use abrasive materials and dissolvent It can damage the sensor 75 temperature sensor Do not touch the halogen and do not damage them Ifit necessary halogen covers can be also cleaned Clean with using only soft and textiles Do not touch the filaments 76 17 2 Replacement of fuses After plug the moisture analyzer in autotext pictograph appears on the display no reaction on the display one of the fuses may be burnt Remember before start any repairing especially exc
28. AB ibrati remember h Project ld 15 01 04 calibration mass remembered by the analyzer description Old deviation of the calibration description Difference so difference between these masses 8 2 Printouts for GLP procedures Group of parameters which enables to introduce user s name and project and data declaration which are printed in reports and standard printouts All information can be introduced using analyzer desk or PS keyboard Use PS keyboard It is easier and faster method 8 2 1 User Enter menu group P2 GLP 401101 P5S gt 01 gt User Wilk i 02 Project WX 235 n 03 Time print I 0 off 04 Date print l 0 off 05 User print I 0 off 06 Project print 0 off 07 Id print Hi 0 loff 08 Lastcalprint 0 off Introducing user s name Using the analyzer keyboard Pressing the navigation key give you access to the parameter 01 User where you can introduce name or number of the operator as series of digital or letters not more than 8 signs Use the direction keys All marks and digits are accessible in closed cycle Select the right one Move into right side and write next mark in After write all marks in press the key ENTER Using PS keyboard 24 Pressing the F2 key enter main menu press the F3 key and scroll to the parameter dark triangle 01 User Pressing the key F2 give you access to setting the parameter firts mark of the number is flashing and pulsing indicator if the number was no
29. ER key The moisture analyzer returns to first display twl Print int 4 4 FE Press ENTER key once again The moisture analyzer returns to main display on level you can determine moisture in the sample Set program Start b gt dK Set et amo 0 0005 9 3 Way of performing the drying according to set values of shortened drying menu Adjustme nts for drying procedure are Set program Start presented 0 000 g ae a n hs at left side of the display To start Fa drying press the key Weighing 45 Preparing Set the pan Tare b l SK Set ans 0 0005 46 Start Stop Program show following command s which enable to operate the instrumen t Act according to them It ensures correct preparatio n of the sample and testing the sample The moisture analyzer displays command to prepare the sample pan Put the sample pan in the handler on the pan support Mass of the pan appears on the display To ensure stable conditions during tarring closing the drying chamber is advisable Tare Press key ZERO TAR E and write sample mass to memory This sets moisture analyzer to zero Next message will appear Place the specimen sample in 47 b D e 3NJ Prepare the sample Close Ba sas 041 o 0 00 01 84c 120 0 00 un amp 48 the sample pan Sample
30. Operating instruction Number of instruction LMI 02 01 10 11 ENG US Moisture analyzer PMX series MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG 26 600 Radom Bracka 28 Street POLAND phone 48 48 38 48 800 phone fax 48 48 385 00 10 Sales Department 48 48 366 80 06 www radwag com 8 2 8 3 TECHNICAL PARAMETER s cccsssessseesseeeseeesseesseessenesnneeenees 5 BASIC INFORMATION c cccsceesseeeeseeeeeeeeeseeeenseeeeeeeeesseeseseeeenseees 6 PUIPOSG Of UI SO wooo ei ee i deve cad 6 Usage COnditions cc0 sc va ea nieve n pa ik m n n kn kana ga de k 6 Principles of safety OpEratiON wrurreeeervrorreeosoortenasononnaasononnannn 7 Warani koki ke beke koka ni Oh vers s n e kek a ok DUR feet ee ab DO a a eee din nin 7 Supervision over metrological parameters of the moisture analyzer 8 Data in the user manual sssr ce eect ee eeeeeeeeeeeeeeeeneeeeeeeneeeeeeeneneeeeenenees 8 Operation competence ceccceceeeceeeeeeeeeeeeeaeeeesaeseeneeseeeetaeeteneeeeaes 8 TRANSORT AND STORAGE ccccseesssecesseeeeseeeeeseeeseeesesnenenseeeees 8 Delivery CNOCK sheets tas aie teneb a ke ko konbe ki santo soi pe TERRAE ei E EED 8 PACKAGING vii io eee yn a kn oi ab dot e pan sa ooo beni ea ian elas trase 8 UNPACKING INSTALLATION AND DEVICE CHECKING 9 Installation place and operation place rrrrrreeevrrreooosooronnsnnn 9 UNPACKING kk a ob e ok pako na n ia a o
31. P06 Program 6 Pressing the Esc key will display the library without the changes Program library P02 Program 2 P03 Program 3 P04 Program 4 POS Program 5 P06 Program 6 11 USING THE LIBRARY To accelerate test user can select program from the library of the instrument and do not have to set up parameters of drying for each substance separately User selects 60 profile with settled parameters in the library e g name of dried product without necessity of changing it if other sample is selected CAUTION To use data base with procedures during determining moisture set up in user menu in group of parameters P7 Inne parameter P7 02 Prog library value 1 Procedure of turn parameter on is described in point 8 of instruction PROCEDURE b D Set program D My 30s 0 Weighing 0 000 o Start Program library P02 Program 2 P03 Program 3 P04 Program 4 P05 Program 5 P06 Program 6 Program library P02 Program 2 P03 Program 3 P04 Program 4 P05 Program 5 P06 Program 6 e 19 09 03 Weighing b P01 FLOUR 100 a 3 _ Set program M a 30s 0 Start 0 000 5 61 Press Test Menu key The program list is displayed Press either up or down arrow keys and scroll to desired option of program Press key ENTER and return to main screen The moisture analyzer is ready to work All of the parameters associated with the
32. STRUMENT IMMEDIATELY TI CR LF tare the instrument immediately after sending command if it is possible in this moment SWITCH ON CONTINUOUS TRANSMISSION IN BASIC UNIT CO CR LF break continuous transmission in basic unit SWITCH ON CONTINUOUS TRANSMISSION IN BASIC UNIT C1 CR LF start continuous transmission in basic unit GIVE THE NUMBER OF THE INSTRUMENT NB CR LF send factory number GIVE THE RANGE OF WEIGHING FS CR LF send value of max capacity in basic unit by the instrument GIVE THE VERSION OF THE SOFTWARE RV CR LF sends name of the program GIVE OR CHANGE DATE IN THE INSTRUMENT PD CR LF sends date and changed date GIVE OR CHANGE TIME IN THE INSTRUMENT PD CR LF sends settled time and changed time GIVE CURRENT WORK MODE PM CR LF sends code of current work mode DISPLAY SETUP PS CR LF sends all setup printout of the parameters SIGNAL BEEP B CR LF signal beep in the instrument Py Ane Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command SEND LAST ERROR CODE ER CR LF sends last code in the instrument DISPLAY HEADLINE DH CR LF sends in headline upper bargraph on the display series of sings CANCELL HEADLINE CH CR LF cancel upper bargfaph i
33. and Function Command Function Command RESET INTERFACE R CR LF zero current made commands and return to factory setting SEND ALL COMMANDS FROM THE INSTRUMENT PC CR LF sends information about all commands in the program from the instrument SEND THE RESULT IN BASIC UNIT S CR LF sends result in basic unit after stability from the instrument SEND THE RESULT IN BASIC UNIT IMMEDIATELY SI CR LF sends result in basic unit from the instrument SEND THE RESULT IN CURRENTLY USED UNIT SU CR LF sends the result in current used result after stability SEND THE RESULT IN CURRENTLY USED UNIT IMMEDIATELY SUI CR LF sends the result in current used unit from the instrument 70 Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command Function Command SET THE MOISTURE ANALYZER TO ZERO Z CR LF set to zero the instrument after stability SET THE MOISTURE ANALYZER TO ZERO IMMEDIATELY ZI CR LF set to zero the instrument immediately after send command if it is possible in this moment TARE IF IT IS STABLE T CR LF set to zero the moisture analyzer after stability TARE IMMEDIATELY TI CR LF tare the instrument immediately after sending the command if it is possible in this moment TARE THE IN
34. ast drying program Ramp drying program Step drying program STANDARD profile Standard drying is suitable for precise determination of the moisture content of most substances It is the most often used profile of drying RAMP profile Ramp drying is selected if substances are not stable on exposure to the full heating power of the halogen radiator of the start In ramp drying sensitive samples are prevented from decomposition by the gentle heating Ramp drying can also be used successfully with substances which form a skin FAST profile The fast drying profile is suitable for samples with a moisture content between 5 and 15 In fast drying the radiator power exceeds the set temperature value during the first minute following the start of drying This compensates the endothermic heat of vaporization and accelerates the drying process The sample must contain sufficient moisture during the first minute to cool it STEP profile The step drying has a use similar to that of fast drying The duration of excessive temperature stage and the temperature value are freely selectable It is used primarily with samples with a moisture content greater than 15 67 Temp profile Final Standard temp Ambient temp 1 Start Temp profile Final Ramp dryin temp Ambient temp ke Start Time Temp profile Final Fast dryin temp Ambient temp Start Time Temp profile Final Step dryin temp Ambien
35. drying weighing counting is much faster and easier It is different method of measure which needs individual for each substance selection of the parameters e temperature of drying e sample mass e time of drying If series of measurements are made in the same temperature first result should be omitted The first measurement sometimes has an error because of thermal stabilization of drying chamber and the moisture analyzer The sample should be distributed evenly on all of the surface coarse grained substances must be disintegrated before the measurement For substances which evaporate water set up short time of sampling For the substances which evaporate water slowly select the time experimentally If you test the substance which covers impervious coat use sand or tissue paper during drying ST ADDITIONAL EQUIPMENT OF ANALYTICAL BALANCES 1 Anti vibration table 2 Thermal printer v KAFKA 3 PC keyboard type PS 2 17 CARE AND MAINTENANCE In this section you will learn how to keep your moisture analyzer in good condition and how to replace expendable parts halogens safety valve 17 1 Components cleaning of moisture analyzer To continue to obtain precise measurements it is advisable to clean the interior components at regular intervals Please note the following instructions for cleaning your instrument a Remember to disconnect the instrument from the power supply before cleaning
36. eceeeeeceeseeseeeeeseeeeeeeeeeeeees 53 10 5 Setting method of switch Off Arying cceeeeseeeeeeeeeeteeeeseeeeeeees 54 10 6 Setting type of displaying and printing results e eeeeeeeeeons 54 10 7 Setting time Of PrintOUt eee eee e cece eeeneeeeeaeeeeeeeseeeesaeeeeaaeeeeeees 55 10 8 Setting size of tested SAMple eccceeeeeeeeeeeeeeeeeeeeeeeaeeeeeeteneees 55 10 9 Setting Correction fACtOM ceceeceeeeeeeeeceeeeeeeaeeeeeeeceeeessaeeesneeeeneees 56 10 10 Default Settings Of program ceeeeeeeseeeeeeeeeeeeeeeteeeeeeeeeenaeees 57 10 11 Settings Printout eee ceceeeeeseeceeeeeceeeeeeeaeeeeeeeseeeeesaeeesaeeeeneeee 58 10 12 Settings record to moisture analyzer MEMOFIY v eeeeeeeeooroen 59 11 USING THE LIBRARY cccssccseeeeeeeeeeseeeseseeeenseeeseeeeeseaneneeeeeees 60 12 PREPARATION OF SAMPLE FOR DRYING s cssssseeeeees 62 12 1 Moisture measuring in moisture analyZel eeceeeeeeeeeeeeeeeees 62 S 12 2 Sampling and sample prepafatiON vurreererreerrrreerassaooonoone 62 13 SELECTION OF DRYING PARAMETERG cccssccssesseeresees 65 13 1 Selection of optimal mass Sample c ccceeseeeeeeeeeeeeeeeteeeeeeees 65 13 2 Influence of sample mass on results repeatability 65 13 3 Drying temperature eee ee ceeeee cence ee eeeeeeeeeeeeeeeeseeeeeeeteneaeereeeeaeess 66 13 4 Selection of the drying Progra
37. ected standard or fast program set up following steps of drying process Profile asi i eM J 30s Manual 43 To set up other drying parameters press key ENTER The moisture analyzer returns to first display Press the down arrow key and scroll to the following option Finish Press functional keys selection of settings and change of the parameters the same as in previous version and select program of finishing measurement and set up the parameters After setting way of finish move to setting of next parameters of drying process Press ENTER key to set up the other parameters The moisture analyzer returns to first display Press the down arrow key and scroll to the following option Result Press functional keys selection settings and changes the parameters are done in the same way as in previous step and select method of displaying result Press functional keys selection settings and changes the parameters are done in the same way as in previous step and select method of displaying result Press the down arrow key and scroll to following option Printout time It is period of time in seconds The moisture analyzer transfers present result to external device 44 bro Print int bro Print int 8 8 E w a 30s e Perform the measurement after set up the printout To finish setting up the parameters and determine moisture press ENT
38. ected to it After connecting the devices plug in the moisture analyzer to power supply 213 5 KEYBOARD OF MOISTURE ANALYZER www radwag com Start Stop key switch on off process of drying according to chosen program Esc key resignation from introduced changes exit to one level higher in menu ga a Group of navigation keys change of parameters value moving in menu Q Print Enter key sending display state to external device pa Print or confirmation of chosen value or function Enter TARA key zeroing the indications On Off key switch on off display of moisture analyzer n After switch off the display other subassemblies are 0 A f supplied moisture analyzer will be in standby mode 14 6 Display Display key changes type of data displayed during and J after drying process 7h Setup Setup key enter to main menu EE E Jere Test menu key starting choice of drying modes menu Vo A MENU OF MOISTURE ANALYZER User menu of moisture analyzers is divided into two modes First one is available after pressing SETUP key This menu permits calibration time and data format temperature communication setting and print criteria Second one is available after pressing TEST MENU key This menu permits setting up parameters which includes test temperature profilling timed information and other parameters A test library is also inc
39. ed package and device immediately after receipt and evaluate whether its free from any external defects 3 2 Packaging Please keep on stock all the elements of the packaging in case future transport of a moisture analyzer is needed Only the original packaging of the moisture analyzer is suitable for transportation purposes Before packing please unplug the device from mains and take out all moveable parts pan covers etc parts of moisture analyzer should be put into original packaging so that they are protected while transport 4 UNPACKING INSTALLATION AND DEVICE CHECKING 4 1 Installation place and operation place Moisture analyzer should be stored and used in places free from vibration and shakes free of wind and breeze dust free and not higher than 2000 meters above sea level Operation place should be situated in place with good air circulation especially around the device 20 cm around the moisture analyzer and 1 meter from top of the device Room temperature should equal 10 C 40 C Relative humidity of operation room should not exceed 80 in 31 C and decrease linearly to 50 of in 40 C Moisture analyzer should be placed on a wall console or a stable table free of vibration and shakes and distant from heat sources Special attention should be paid to weighing magnetic materials as balance is equipped with a strong magnet 4 2 Unpacking Gently take the moisture analyzer from package take
40. enees 31 8 6 2 Nonstandard printouts 0 ccceeeeeeeeeeeeeeeeeeeneeeeeenaeeeeeeaaes 31 8 6 3 Witting textS Niassa ta kankan eka kan ENRETA RT kt da e ena eee ENA 32 8 6 4 Selection of printouts iirrevrrrreerrreeerteearsassassasnonononn 36 8 6 5 Setting non standard printouts s seese 36 8 6 6 Non standard printouts in drying Mode 36 8 7 Setup the parameters cccececeee cesses ceeeeeeeaeeeeeaeseeeeeseaeeesaeeeeneeee 38 8 7 1 Enter to group menu P7 Globals ccceeeeeeeseeeeeeeeees 38 8 7 2 Setup the parameters ccccccceceeeeeeeeeeeeceereeeeeeesneeeeneees 38 9 MAKING THE DRYING WITH USE OF SHORTENED MENU 40 9 1 Way of shortened menu activation eeeeererrereererrenassanonnannn 40 9 2 Way of setting the parameters eeeeeeeeeeeeeeeeeeeaeeeeeenaeeeeeeaaes 41 9 3 Way of performing the drying according to set values of shortened NOMEN ke iT tot de ven an kaa pon er k akt e AR band RARA TARA n r kon kre Deka errr err 45 10 DECLARING THE DRYING PROCEDURES AND USING THE PROCEDURES LIBRARY ccesscsseteeseeeeeeeeesenesesneesneneeeeeeeeseaesaseaeenseeeeees 50 10 1 Defining the drying procedure eeeceeeeeeeeeeeeeeeeeeeeeeeeteeeeeees 50 10 2 Writing in name of software 0 ee eect eeeeteeeeeeeeeeeeeteeeeeeeeeeeaeens 51 10 3 Selection of drying profile 2 eee eeeeceeeeeeeeeeeeeeeeseeeeeseaeeesaeeeenees 52 10 4 Setting the drying temperature cccec
41. ey Transgression max measure range of the instrument Transgression max measure range of the instrument No divisions of converter Transgressed max amount of divisions of the converter Transgression admissible value zero or tare range for the instrument Tare outside the range Result gt 10 Max Result gt 4 Max The difference gt 1 Max The sample mass lt 1g The sample mass outsider the range Rev lt 1000 Div Difference gt 20 Outside range Faulty value Incorrect password Record error Parity error Table error suspended transmissionCTS suspended transmissionXOFF Incorrect date 2 6 2 7 2 8 2 9 2 10 2 11 2 12 2 13 3 1 3 2 3 4 4 1 4 2 4 3 4 4 4 5 5 1 Transgression admissible value tare range for the instrument Difference of values of initiative zero above 10 of factory value check after start zero Difference between start mass pan handler etc for the moisture analyzer after start and start mass recorded in the memory is bigger than 4 turn the instrument on with load on the pan Difference between calibration mass currently measured and calibration mass recorded in the memory is bigger than 1 Value of sample mass below 1 g Value of the sample mass below value settled in the parameter Sample necessarily in data base of drying programs Value of reference mass
42. ficient 1 0 00 Default Print To activate program data press navigation key After pressing the key window with saved data in chosen program will be displayed According to described way of moving in menu and making the changes in settings enter required parameter changes Setting the parameters must ensure correct process and reliable results of measurement 10 2 Writing in name of software Name enables to identify and use the program easily depending on tested substance It can contain 12 alphanumerical signs Method of entering the name Program library gt Program P01 gt Name gt FLOUR 100 Profile Standard Dry temp 1120 C Switch off Automatic 3 Program library Program P01 gt Name gt CORNFLOUR 100 Profile Standard Dry temp 1120 C Switch off Automatic 3 Indicated by black arrow parameter can be changed Press navigation key and activate edition field to change First sing of the new name starts flashing Using functional keys or PC keyboard write the new name of the program e g name of the substance If you write new sign the flashing sign disappears Confirm new name and press ENTER Last sign stops flashing 51 A After write the new name set following parameters of the program 10 3 Selection of drying profile There are four programs available when the moisture analyzer is running a test standard fast ramp step In each of these programs parameter
43. ge contrast on the display 08 Analyzer ld information about factory number of the analyzer 09 Software rev information about factory number of the analyzer 10 Parameters printout activation this function causes printout set parameters in users menu User gives number of the parameters which are printed 05 01 04 Setup 40 12 45 aka 6 aes After enter into the parameter of printing arcane re display with question appears Setup 43 47 56 a a After the confirmation starts printing 10 41 04 _ Printing through socket RS 232 actual sent set parameters ke E en 100 11 Par receive Activation the function causes receive all parameters from computer through RS 232 After enter the function display appears 10 11 01 Receiving 13 47 56 jo Setup receiving After receive settings the analyzer informs user about how many parameters were accepted and changed and how many declared faulty and not accepted by the program Printing and receive the parameters in very comfortable and fast way to introduce new settings After printing actual parameters to file in the computer user changes the parameters fast and easily Value of these parameters must be corrected and after record the changes in the file new corrected setting must be sent to the analyzer program After recording changes in the parameters the analyzer accepts new settings This procedure requires good knowledge of the parameters and comp
44. hange the fuse plug out the moisture analyzer The moisture analyzer is equipped with two fuses Before start exchange take power cable off at the rear of the casing Tabular fuse ZKT 3 15A T4L 250V protects electronic systems of moisture analyzer Replacement of fuse 1 Screw out the fuse socket from mounting 2 Remove fuse from socket 3 Replace the fuse for new one We 4 Screw in socket with new fuse 17 3 Troubleshooting Problem display remains dark after switching ON Probable causes no line voltage power cable not connected power line fuse blown instrument faulty Problem the measurements takes too long Probable cause you have selected an unsuitable switch off criterion Try experimentation to arrive at a suitable switch off criterion Problem The measurements are not repeatable Probable causes The samples are not homogeneous i e they have different compositions The more inhomogeneous a sample the larger the amount of sample needed to obtain a repeatable result The drying time is too short Extend the drying time select a suitable switch off criterion You have selected a temperature that is too high and the sample has oxidized Lower the drying temperature The sample boils Lower the drying temperature The temperature sensor is contaminated or faulty Clean the temperature sensor The support on which the instrument is standing is not stable Use a stable
45. ka e pann dra e ek e vip bn konnen a ve pa poo as ken 9 SOUNVA Lb r k a koi anana kana piba entak aba v e oo aaa kise pie konen a r pei pan 12 List of standard elements of the dEliVETY rveerrrreeooronnen 12 Cleaning of moisture analyzer h rrvirrreeroreeorterrrrtasaasssanassanonnn 13 Connecting to MAINS iya seza ra tan e pea kapa aaan taa nn teat 13 Connection of additional equipment Li rrevrereeetrreerertserasenonoen 13 KEYBOARD OF MOISTURE ANALYZER c cceseeesseeeeseeeeeees 14 MENU OF MOISTURE ANALYZER 0 2sccssseeceseeeseseeeseeeenseeeeeees 15 MOVING IN USER MENU ccecsseceseeesseeeesneeenseeeeeneeeseaeenseeeeeees 16 Menu Operatioune tak oun ko bon kone rese koni ye ped oka da kana 16 By means of computer keyboard PS type ccceeeeeeeeeeeeeeeeees 18 USER MENU OF MOISTURE ANALYZER cccesecsseeeseeeeeees 20 Calibration of moisture analyzer eeeeeeceeeeeeeeeceeeee sete eeeeaeeeeeeeees 20 8 1 1 Calibration of moisture analyzer rrerrrreeroroeonosonone 20 8 1 2 WS Gr CaliDTatlOn wc vin cee Sete ceie dive gs here et ev pese iv panes a r SES 21 8 1 3 Temperature Calibration cceceeeseeeeeeneeeeeeeeneeeeeteteeeeeene 22 8 1 4 Calibration testien asa aaae a aae Tat 23 8 1 5 Temperature calibration ssasssseessrnesserresrsrnesrennenrennesnnna 23 8 1 6 Printout of the calibration report cc eeeeeeeeeeeeeeeeeees 23 Printout
46. keyboard 4 Not standard printouts can cloud over each other etc Printout 1 Start 1 Printout 1 Stop 40 Printout 2 Start 20 Printout 2 Stop 40 37 Not standard printout can contains Variables depending on the users needs mass date Number of the project temperature of the chamber etc Stable texts in user s menu Composed not standard printout contains not more than 640 marks written as 80 texts 8 marks each from the parameter String 1 to String 80 User can compose 4 not standard printouts 8 6 3 Writing texts in Variables in the printouts Printout of singular mark K_ Kind of the last calibration N Actual net mass in basic unit l Deviation in the last calibration d_ Actual date Yo1 Code 1 t Actual time 2 Code 2 i Number of the analyzer 3 Code3 r Number of the program 4 Code 4 P Project number 5 Code 5 U User s number 6 Code 6 k Date and time of the last calibration Variables only in drying function C Temperature of drying chamber YoM Drying result moisture D Drying result dry R Drying result moisture dry N Drying result actual net mass in the basic unit Yoc Drying time S Start mass of the sample E Final mass of the sample p Parameter of chosen drying procedure 32 Special marks used in creating special printouts Singu
47. lar mark c CRLF r CR n LF t Tabulator s Move to next string 0 Finish the printout Each text String 1 89 String 80 contains max 8 marks letters digits special marks spaces To write the sentence which contains many words and marks create them with neighbouring texts with 8 marks each User can use special marks mentioned below to include variables depending on the needs Example 1 The maximum mass of the load cannot exceed 11 250 g To write this sentence use 46 grouped marks in neighbouring lines of the text Write 8 marks in each text untill the sentence finishes Parameter no Jext 1 2 3 4 5 6 7 8 19 Text 10 T h e m a x i 20 Text 11 m u m m a sS s 21 Text 12 o f t h e l 22 Text 13 l o a d c a n 23 Text 14 n o t e x c e 24 Text 15 e d 1 1 A 2 5 25 Text 16 0 g 33 Example 2 Zak ad Mechaniki Precyzyjnej RADWAG Date Time Load mass lt actual work mode gt Enter setting the following texts and write 8 marks in each of them untill you finish writting printout in Text Parameter no a w 4 5 25 Text 17 a IN 26 Text 18 n 27 Text 19 c 28 Text 20 29 Text 21 oo 2 EH 30 Text 22 31 Text 23 lololas ke 32 Text 24 33 Text 25 allaj mme 34 Text 26 0 2jaj o m N x
48. logen Hitting Halogen module IR emiter Drying modes standard quick stepped mild Auto switch off options time defined moisture stabilisation time optimal manual Operating temp 10 C 40 C Power supply 120V 50Hz AC 420W Display graphic backlit Pan size 90mm h 8mm Net pe 5 1 7 2kg Packaging size 455 x 315 x 302 mm Drying chamber dimensions 120 x 120 x 20 mm Dimensions BASIC INFORMATION CAUTION Utilization of PMX moisture analyzer conversely to safety provisions and user manual may be hazardous to health and life Never use the moisture analyzer is explosive area Moisture analyzer is not prepared to use in areas with danger of explosion The interior part of drying chamber may heat up to 400 C nominal power supply of moisture analyzer is from 120 to 230 V thus it is necessary to use the device according to safety principles The symbol visible on the top of the moisture analyzer cover as presented below is a warning against hot surface 2 1 Purpose of use Moisture analyzer is a device used for determination of relative humidity of small samples of various materials determination of dry mass in small samples of various materials and determination of mass of weighed objects It guarantees fast and precise determination of water content in tested sample and through application of graphic display it provides easy use and operation Moisture ana
49. luded which holds data for up to 100 tests which can be recalled and run at any time P1 Calibration P1 01 Ext calibr P1 02 User calibr P1 03 Calibr test P1 04 Temp calibr P1 05 Print report P2 GLP P2 01 User P2 02 Project P2 03 Time print P2 04 Date print P2 05 User print P2 06 Project print P2 07 ld print P2 08 Last cal print P3 Date Time P3 01 Date format P3 02 Time format P3 03 Time P3 04 Date P3 05 Disp time P3 06 Disp date P5 RS 232 P5 01 Baud rate P5 02 Parity P5 03 Data bits P5 04 Stop bits P5 05 Handshake P5 06 Print on stable P5 07 Printer type P6 Printouts P6 01 Printout No P6 02 Header No P6 03 Line No P6 04 Footer No P6 05 Pr 1 start P6 06 Pr 1 stop P6 13 String 1 P6 92 String 80 P7 Globals 15 P7 01 Password P7 02 Prog library P7 03 Beep P7 04 Language P4 Readout 7 MOVING IN USER MENU P4 01 Filter P4 02 Autozero P4 03 Temperature P4 04 Negative P7 05 Backlight P7 06 Brightness P7 07 Contrast P7 08 Balance Id P7 09 Software rev P7 10 Par printout P7 11 Par receive P7 12 ID settings P7 13 ID autoprint It can be done by means of keyboard PS keyboard commands sent from computer to the analyzer 7 1 Menu operation Program library Program 2 Program 3 Program 4 Program 5 Program 6 Fa 10 10 03 Setup P1 gt Calibration P2 P3 P4 P5 P6 P7 GLP Date Time Readout RS 232 Printouts GE 16 How
50. lyzer e Insert for base of the drying chamber e Drying chamber cover 12 Pan holder Cross Disposable pan Power cable User manual 4 5 Cleaning of moisture analyzer Moisture analyzer should be cleaned with use of a damp cloth Gently rub dirty places Weighing pan must be removed from the moisture analyzer for cleaning Cleaning weighing pan when it is installed may cause damage of moisture analyzer mechanism Remember about disconnecting the moisture analyzer from power supply before the cleaning 4 6 Connecting to mains Moisture analyzer can be connected to mains only by means of original supply wire which is basic equipment of balance Rated voltage given on data plate should be consistent with mains rated voltage Supply wire can be connected only to socket with ground contact Switch on supply of moisture analyzer plug of supply wire should be connected to the socket on the back of analyzer casing Display of moisture analyzer will indicate name and number of program next display will indicate 0 000 g for balances with accuracy of 1 mg or 0 0000 g for balances with accuracy of 0 1 mg If indication is different from zero press TARE key 4 7 Connection of additional equipment Before connection of additional equipment or its change printer computer PC disconnect the moisture analyzer from power supply Only additional equipment which is recommended by the manufacturer of moisture analyzer can be conn
51. lyzer series PMX can be used to determine humidity content of multiple materials At the initial stage of measurement the device precisely determines the mass of object placed on its weighing pan Following this there is fast heating of the sample with halogen or IR lamps This causes vapouration of humidity from the tested sample While sampling the moisture analyzer is continuously checking the decline of mass and after calculation it displays current indications on the display of the balance Compared to conventional methods of humidity content determination in various materials application of moisture analyzer series PMX significantly shortens measurement time and simplifies testing procedure Moisture analyzer allows for setting of multiple parameters which influence testing of a sample like temperature time drying profiles etc 2 2 Usage conditions CAUTION If drying process is active do not allowed to open drying chamber Moisture analyzer is equipped with a halogen lamp which is very powerful heat source 6 Thus user should pay special attention no to touch those elements of moisture analyzer that get hot while drying procedure disposable pan handle of pan and internal cover of drying chamber Please remember that some elements may become dangerous if heated poisoning vapours danger of ignition or explosion Moisture analyzer can not be used for dynamic weighing Even if small masses are added to taken off from
52. m cccccceeeeeeseeseeeeeeeeeeeseeeeneees 67 13 5 Selection of drying time cee cececeeeeeeeeeeeeeeae scence seeeeeseaeseeeaeenenees 68 13 6 Analysis of drying profile eee eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeneeaees 68 14 COOPERATION WITH PRINTER OR WITH COMPUTER 69 14 1 Schemes of connection CablES ervrrreverrerorrresrarsanonasnonan 69 15 LIST OF COMMAND COMPUTER MOISTURE ANALYZER 70 16 GUIDELINES OF USAGE csseccsseeeeeseeesseeeseeeeeeeeeseeeseneeeenseeeees 73 17 CARE AND MAINTENANCE 0 cceecseseeesseeeeseeeeeeeeeseeesenneeenseeeees 74 17 1 Components cleaning of moisture analyzer reeneeerenroooons 74 17 2 Replacement of USES cuirinn anin e ENa EAE S ERAEN 77 T43 Troyoleshootings ecrire neno iiai i e ari aa 78 17 4 Errordete ction c sec mwn a EN po na AEN ET IRE 78 18 MEMORY OF PERFORMED MEASUREMENTG cseeeee 81 19 PRINCIPLE OF USE OF MOISTURE ANALYZER FOR DRYING TEMPERATURE ABOVE 160 C ccseceseeeeeseeesseenseeeeeeeeesseeseseeeenseeeees 82 1 TECHNICAL PARAMETERS Type PMX 50 1 PMX 50 Max capacity 50 9 Reading unit 0 1 mg 1 mg Accuracy of moisture 0 0001 0 001 reading Tare range 50g 50g Max mass of sample 50 g Repeatability of 0 24 sample do 2g 0 06 sample 2 10g moisture 0 04 sample over 10g Drying temperature max 160 C halogen IR emiter range max 250 C ha
53. moisture is faster or ann flower Size of solid surface decides about S speed of drying and reliability of measure Tieas The surface of solid should be the biggest as it is possible Because solids tittiti CL Sample with correct thickness Dried substance 64 evaporates moisture by external surface also thickness of sample is very important 13 SELECTION OF DRYING PARAMETERS 13 1 Selection of optimal mass sample The sample weight has an influence on both the accuracy of the measurement results and the measurement time With large amounts of samples a great deal of water must vaporize and the moisture determination takes longer To keep the measurement time as short as possible we advise you choose a low weight sample but not so low that attainment of the required measurement accuracy is no longer possible The precision of result High Low Sample weight 0 5g The measurement time Long Short Sample weight 13 2 Influence of sample mass on results repeatability Sample weight influences the repeatability of the moisture analyzer The relation between sample weight and repeatability is shown in the following Repeatability 10 6 0 3 40 15 0 06 0 03 65 The preceding table is based on the assumption that the sample is ideal homogeneous and its moisture can always be separated completely and free from decomposition e g moist quartz sand Deviations always comprise the uncer
54. n the headline MAKE INTERNAL CALIBRATION CL CR LF starts internal calibration BLOCK THE KEYBOARD KL CR LF blocks the keyboard UNBLOCK THE KEYBOARD KU CR LF unblocks the keyboard SWITCH OFF THE ECHO OF THE KEYBOARD EO CR LF switches off sending codes of the keys SWITCH ON THE ECHO OF THE KEYBOARD E1 CR LF switches on sending codes of the keys SWITCH OFF THE INSTRUMENT OO CR LF switches off the moisture analyzer the same as usage ON OFF SWITCH ON THE INSTRUMENT O1 CR LF switches on the moisture analyzer the same as usage ON OFF SWITCH OFF THE AUTOZERO AO CR LF switch off autozero SWITCH ON THE AUTOZERO A1 CR LF switch off autozero GIVE OR CHANGE ID ID n CR LF the moisture analyzer sends settled code or change this code n 1 6 SEND THE PRINTOUT NUMBER n PP n CR LF the moisture analyzer sends programmed printout with number n n 0 4 279 Function GIVE OR CHANGE VARIABLE NUMBER n Command Vn CR LF the moisture analyzer sends variable or change variable with number n n 0 4 Notice Sending command to the instrument which is not in the list or has an error and it is finished with CR LF back sending the command in form If there is no command in the list or has an error and it is finished the command is returned in E S CR LF Spaces should be omitted 16 GUIDELINES OF USAGE Measurement moisture of various substances in comparison to traditional method weighing
55. ng not standard printout for function of drying is the same as for weighing function The only one characteristic which differs it from the printout is separate definition of three parts of the printout see standard printout for the drying Headline Line Footer Example of definition of drying printout Headline Printout 2 Line Printout 3 Footer Printout 4 36 18 11 02 Seti 13 46 42 P6 01 Printout No i o standard 02 Header No i 2 printout 2 03 Line No I 31 printout 3 04 Footer No l 4 printout 4 05 Pr istart 1 06 Pr 1stop 1 l 07 Pr 2 start l 101 08 Pr2stop 20 09 Pr 3start 24 10 Pr 3 stop l 22 11 Pr 4 start 2 3i 12 Pr 4stop I 351 Printout 2 Printout 3 22 String 10 i Dr yi ng si 33 String 21 C M c Ni 23 String 11 tart cDal 34 String 22 We Ni 24 String 12 ite Yd Ve Ti I 25 String13 ime tol l si f 26 String 14 Drying_p 27 String 15 jarameter 28 String 16 is c c pl 29 String 17 i e c Stai 30 String18 rt _weigh 31 String 19 It YN EE 32 String 20 AW ANN Printout 4 35 String23 IStop weit 36 String 24 1ght Yo NA Gi 37 String 25 ITime dry 38 String 26 ling C c 39 String 27 leFinal 40 String 28 Iresult Yi 41 String 29 IM c ecDry 42 String 30 ling stop 43 String 31 ile eSign 44 String 32 lature 45 String33 i i 46 String 34 FO AS A 47 String 35 A SAA EN 8 7 Setup the
56. ning toxic or caustic vapour should be tested in a chamber which absorbs the vapours samples of substances that produce inflammable vapours if heated should have relatively small mass and drying process should be performed in low temperatures please remember that aggressive substances may cause corrosion to the device 2 4 Warranty Warranty does not cover below mentioned cases e Not observing the regulation listed in user manual e Use of the moisture analyzer conversely to its application e Any modifications of moisture analyzer e Mechanical defects and defects caused by media liquids and natural wearing off e Improper placing of defects of electrical network mains e Overloading of measuring mechanism of the moisture analyzer 2 5 Supervision over metrological parameters of the moisture analyzer Metrological features of the moisture analyzer should be checked by the user in set period of time This period depends on environmental factors of moisture analyzer s location kinds of performed processes and quality system introduced byte user 2 6 Data in the user manual Read the user manual carefully before switching on the device even if the user is experienced with this kind of devices 2 7 Operation competence Moisture analyzer should only be operated and supervised by trained personnel with experience in usage of this kind of devices TRANSORT AND STORAGE 3 1 Delivery check Please check deliver
57. o define new drying procedure and how to make a measurement in defined procedure remembered in the moisture analyzer memory Each procedure contains Name Profile Dry Temperature Switch off Kind of displaying result Frequency of printing data Size of the sample Correcting factor 10 1 Defining the drying procedure To use data base with recorded procedures switch on in user s menu in group of the parameters P7 Inne parameter P7 02 Prog library value 1 Switch on the parameter is described in point 8 7 2 of the manual Before start work with data base introduce data to the moisture analyzer According to these data tests are done Way of entering to setting data in procedures 19 09 03 Program library P02 P03 P04 POS P06 Program 2 Program 3 Program 4 Program 5 Program 6 Weighing b D la 3NJ Set program D My 30s 0 50 Start 0 000 s Press key TEST MENU and display listofremembered procedures in memory CAUTION In option POO last program is displayed or program chosen for drying Press the arrow keys and replace the arrow on list of programs P02 IProgram 2 P03 Program 3 P04 Program 4 P05 Program 5 P06 Program 6 Program library gt Program P01 gt Name gt Program 1 Profile Standard Dry temp 1 120 C Switch off Automatic 3 Result Moisture content M Print int 130 s Target None T coe
58. off plastic cartoon and foil transport protections and gently put the device in previously prepared operation place Install components of moisture analyzer according to below schema Montage schema for parts of moisture analyzer disposable pan pan bracket pan handle chamber cover chamber basis 10 Inst all cov er of dryi ng cha mb er Inst all wei ghi ng pan han dle Inst all pan sup port er Put disp osa ble wei ghi ng pan in pre par ed plac Setting the pan supporter Al When installing weighing pan suppeorter pay special attention to positioning of the weighing pan mandrel cross The mandrel is equipped with special cut for unique positioning of the handle towards the cross which eliminates their contact and friction Positioning of weighing pan cross After placin g the weighi ng pan cross on the mandr el turn the cross slightly SO that the cutting on the mandr els are situati on in correct positio n When turning the mandr el use most gentle move ments so that mecha nism of the moistu re analyz eris not damag ed 4 3 Setting Before switching on the device please level the moisture x analyzer using the feet situated at the back of its casing The O bubble of the level should be situation in the centre of the ring OK 4 4 List of standard elements of the delivery e Moisture ana
59. parameters as default Program library Program P01 gt Name Program 1 Serena ae In the menu press the down arrow key and Switch off Automatic 3 i replace the black arrow next to the Ga we eonan parameter Default Target None T coeficient 1 0 00 Default bi Print If you press navigation key command to confirm chosen function to restore default values appears Program library gt Program Set defaults Pressing the ENTER key will set up all e parameters of the program for primary values and return to main display 57 PO1 10 11 library Program Name Program 1 Profile Standard Dry temp 1120 C Switch off Automatic 3 Result 1 Moisture content M Print int 30s Target None T coeficient 1 0 00 Default bi Print L E Settings printout Pressing Esc key will make resignation and returns to previous display without any changes Parameter enables printing of inserted program settings on connected printer Procedure EJ Program library Program gt PO1 Name Program 1 Profile Standard Dry temp 1120 C Switch off Automatic 3 Result Moisture content M Print int 130s Target None T coeficient 1 0 00 Default i Print bi Press the down arrow key and scroll to parameter Print in menu _ Program library gt Program Print program
60. racteristic preparation and size of the sample are all important contributing factors in increasing speed and quality of the measurement process Sampling and the sample preparation have a great influence on the reproducibility of the measured units It is also important that the sample being investigated is a representative part of the total amount of the sample under test 62 Repeatability High Low The sample reparation Various Always the same The final result of a moisture determination depends on a carefully thought out sample preparation The part of sample used for analysis must be representative of the total quantity The sample preparation includes work processes such as sampling sample division size reduction homogenization and others All of these processes should be carried out as quickly as possible and without loss or uptake of moisture The appropriate standards and directions must be consulted to determine the method of sampling as this is depended upon the product consistency and the amount used Number of samples An increase in the number of samples always lead to an improvement in the statics reliability of the analysis results The size depends on the homogeneity of the test material the accuracy of the test material the accuracy of the measurement method and the desired accuracy of the measurement result Mechanical size reduction Sample division is usually accomplished by specific types of mills
61. rature Nisam M TEC In step profile set up the temperatures of the level 1 and 2 and drying temperature and periods of heating in following average temperatures After set up profile and the parameters move to programming following parameters of the program 10 4 Setting the drying temperature The drying temperature is repeated parameter of setting option Profile Procedure Ly Program library Program P01 gt Name FLOUR 100 Profile Standard Dry temp bi 120 C Switch off Automatic 3 Press the down arrow key and replace black arrow next to parameter Dry temperature Drying temperature Using functional keys of moisture analyzer Or PC keyboard set required value of temperature Setting can be made according to previously described way of moving in user menu After setting and saving new value of temperature this temperature will be also inserted in profile settings as valid one 53 10 5 Setting method of switch off drying Switch off can be Automatic 1 change 1mg 10s Automatic 2 change 1mg 25s Automatic 3 change 1mg 60s Automatic 4 change 1mg 90s Automatic 5 change 1mg 120s Time defined max time 9h 59min Manual after pressing the key Defined give change of mass Am 0 1 9 9mg and change of time At max 2 55 sec Test enables choice of parameters for auto switch off for sample Procedure Le Engram Ratan an In menu press the down
62. ress the down arrow key and replace Dry temp 120 C Sh od l Automatic 3 black arrow next to the parameter Sample Result Moisture content M Print int 130s Target gt None Press navigation key and move to following display to choose option PO1 Target gt i None select the suitable Select them according to described method in user s menu After selection suitable profile set up other parameters drying temperatures and Program library gt Program Target Using functional keys or PC keyboard time according to displayed data for selected profile 55s Program library Program Target PO1 gt Target gt Optional i j Target weight 15 000 g In option Optional set up sample mass Tolerance 110 Yo and tolerance of this mass If you choose this mass the moisture analyzer informs user to put suitable amount of substance on the pan In other case the instrument starts testing the sample even if its mass is not compatible to settings Program library gt Program Target P01 gt Target gt Obligatury j i Target weight 15 000 9 In option Obligatory set up sample mass Tolerance 110 and tolerance for this mass If you select option program the moisture analyzer informs user to put suitable amount of substance on the pan In this option program does not start testing sample until settled amount of substance is on the pan 10 9 Setting correction factor Dark color
63. reviously changed digit stops flashing Repeat previous actions for following values After set up new time press key ENTER The analyzer returns to submenu P3 Date Time and the hour changes After set up new time return to weighing mode according to point 7 of instruction manual 8 3 4 Setting date Press Right key to set parameter 04 Date CAO Setup 13 47 56 ARL Year p 2004 Month 01 Day 05 agy According to previous point 8 3 3 set up present date Then return to weighing mode according to point 7 1 of the manual 8 3 5 Determination if date and time should be displayed User decides if date and time is displayed If suitable values of parameters are set date and time are displayed The values can be changed according to point 6 If parameter 1 date and time are displayed on upper bargraph of screen 8 4 Setting parameters of work for the moisture analyzer The moisture analyzer can be adjusted to difficult conditions of work by means of suitable parameters in group lt P4 Readout gt autozero displaying last digit 8 4 1 Menu group P4 Readout Enter menu group P4 Readout 10 11 01 Setup 13 47 56 P4 gt 01 gt Filter 3 normal 02 Autozero 1 on 03 Temperature funkction 04 Negative 0 disabled 8 4 2 Filter setting Depending on conditions of work set up filter For perfect conditions set up filter as very fast value of parameter 01 Filter for 1 but if the conditions are
64. s are referred to drying temperature and time needed to achieve set temperature The program must be selected according to examined substance Selection of the program is described in further section of the manual Way of setting the profile Test menu Program library gt Program P01 gt Name y FLOUR 100 Press the down arrow key and replace H 1 ai azi black arrow next to parameter Profile Switch off Automatic 3 Press navigation key and move to next display i Using functional keys or PC keyboard Pi library gt Profile m H teen renn select the profile Select the profile Dry temp 120 C according to described method of moving in user s menu After selection suitable profile set up the 2 other parameters drying temperature and lt time according to data for chosen profile Program library Program gt Profile i A P01P Profile gt Standard In standard profile set up drying Dry temp 1420 C temperature PO1 Profile gt Rapid Program library Program gt Profile 1 Dry temp 1120 C 52 Program library Program gt Profile P01 gt Profile gt Gentle Ramp time 5 min Dry temp 1120 C Program library Program gt Profile PO1 Profile gt Step Temperature 1 50 c Time 1 13min Temperature 2 100 C Time 2 1 3 min Dry temp 120 C In ramp profile set up the drying gen temperature and period of time to achieve set 6 tempe
65. s for GLP procedures cceccceceeeeeseeceeeeeeeteeeetaeeeeneeeeaes 24 8 2 1 SOM seke potanta eye Bout yok a po epa depose ian pea ta p pe bote 24 8 2 2 POO Che ot kwe pena teka ae SO eben ot atak pena tee ones di sa yes TRAA 25 8 2 3 Parameters concerning content of GLP procedures 25 Setting time and Cates ia ceed onnu aea pena aa ni pen pas 26 8 3 1 Setting date format u 0uiiirrevrrrrovroroerteananssanassaononanonn 26 8 3 2 Setting time fONMAL 0iureiirreerrrroeoorosortnaransasaassaononanone 26 8 3 3 Sating M E is castes cvceceaeesccadhyevenceuescesievunepeciadaswenine unease 27 8 3 4 Setting dates sise kose petri on bye n katab yon bay Poz s aaa EEA 27 8 3 5 Determination if date and time should be displayed 28 8 4 Setting parameters of work for the moisture analyZerf 28 8 4 1 Menu group P4 ReadOUt r rereeeerrrreooerorrenassononnannn 28 8 4 2 Filter SOUNO AEE AE AE E A ba pants EE OE 28 8 4 3 Setting the autozero operation eseeeeeeeeeeeesreerese 28 8 4 4 View of the temperature inside the drying chamber 29 8 4 5 NOGAUVO niista i E AEE AE S 29 8 5 Functions connected with use of RS 232 ccccceeeeeeseeeeteeeees 29 8 5 1 Set up values of parameters erreen 29 8 6 PP AOU NS knn no ded np AA EA ak pk ok aa ann po senate n da t pani TAARA 30 8 6 1 Standard printout 0 ccccccceeeceeeeeeeeeeeeeeeeseeeeeseeeeeneeee
66. support The surroundings are very unstable vibrations draughts etc Change the surrounding into suitable 17 4 Error detection The moisture analyzer utilizes a suitable command on graphic display to indicate an error Command has error number and short description of the reasons of error List of errors in moisture analyzer Command on display Error Error description number Error of control sum 1 1 Error in data transmission 78 Error A D Error 1 3 AD Th L Error 1 3 AD Th H Error 1 3 AD Tc L Error 1 3 AD Tc H Exceed range Exceed range A D Null A D Full Tara Zero outside the range 1 2 1 3 1 3 1 3 1 3 2 1 2 2 2 3 2 4 2 5 Converter error Number of A D divisions from the thermometer in the drying chamber 0 temperature sensor defected heating process of the drying chamber will not be initiated return to weighing mode by pressing ESC key Number of A D divisions from the thermometer in the drying chamber 1024 temperature sensor defected heating process of the drying chamber will not be initiated return to weighing mode by pressing ESC key Number of A D divisions from the thermometer of the solenoid actuator 0 temperature sensor defected return to weighing mode by pressing ESC key Number of A D divisions from the thermometer of the solenoid actuator 0 temperature sensor defected return to weighing mode by pressing ESC k
67. t temp ke Start Time 13 5 Selection of drying time Drying time is determined by choose suitable criterion of drying A switch off criterion is understood to mean the condition which must be met for the moisture analyzer to switch off automatically and the drying The automatic switch off is after settled time independently of loss of mass It is used for substances which decomposes and their mass is not stable The second type recognizes the end of drying automatically The integrated analyzer continuously determines the weight loss of the sample during drying If the weight loss less than 1 mg per unit of time is less than the set level the drying is stopped and the final result is displayed Automatic switch off User has possibility of few types of drying switch off Switch off can be Automatic 1 change 1mg 10s Automatic 2 change 1mg 25s Automatic 3 change 1mg 60s Automatic 4 change 1mg 90s Automatic 5 change 1mg 120s Time defined max time 9h 59min Manual after pressing the key Defined give change of mass Am 0 1 9 9mg and change of time At max 2 55sec Test enables choice of parameters for auto switch off for sample Auto switch off free defined criterions The auto switch off criterion is based on a user defined mean weight loss per unit of time or a user defined mean M loss per unit of 60s As soon as this drops below the present value the measurement is automatically ended
68. t written in by means of the keyboard write new name of the user User and confirm by pressing the key Enter Using the keys Esc and Enter move into weighing mode 8 2 2 Project Name of the project contains max 8 marks Project name should be written in the same way as user s name according to point 8 2 1 8 2 3 Parameters concerning content of GLP procedures In the standard printout set up following data which are printed P2 03 Time print measurement time and reports of real time clock P2 04 Date print measurement date and reports of the real time clock P2 05 User print user s name in the parameter P2 01 User P2 06 Project print project name in the parameter P2 02 Project P2 07 ID print analyzer number given by the producer P2 08 Last caribration print printout with the information about last calibration date and time kind deviation The parameters of the data which are printed must be set on 1 on EXAMPLES OF PRINTOUTS FOR DIFFERENT SETTINGS GLP P2 03 time print on P2 04 date print on P2 05 user print on P2 06 project print on P2 07 ld print on P2 08 last cal print on 25 Date 16 01 2002 Time 13 12 30 User R Wilk Project Sprawd Balance 11111111 Last calibration 16 01 2002 13 02 External calibration Difference 0 0001 g 130 0500 g P2 03 time print off Date 16 01 2002 P2 04 date print on User R Wilk P2 05 user print on Balance 11111111 P2
69. tainty which depends on the sample and the repeatability of the instrument In practice measurement differences appearing within a measurement series can consequently be larger than the values of the halogen moisture analyzer shown in the table 13 3 Drying temperature The drying temperature exerts a controlling influence on the measurement time Value depends on kind of the material Too low temperature causes evaporation of water partly and extension of drying time unnecessary Too high drying temperature causes combustion of material chemical decomposition overheating the sample Drying temperature is determined by norms If not the temperature must be determined experimentally The following procedure is suggested for selecting the temperature e Estimate moisture content of the sample Determine the decomposition temperature of the sample by experiments Compare the measurement result with the reference method if one exists With samples which have Monsironot swen a high moisture content it is possible to shorten the Len9 measurement time by selection of the step or rapid drying program Short Here the greatest part of the existing moisture is Drying temperature separated at an elevated temperature The drying temperature is then lowered and kept constant until the end of drying 66 13 4 Selection of the drying program There are four programs which control the temperature profile Standard drying program F
70. uter operation 39 12 ID setting codes are introduced to the analyzer memory 13 IDautoprint defines if code introduced in the analyzer memory is sent automatically through the socket RS 232 9 MAKING THE DRYING WITH USE OF SHORTENED MENU This section provides you with essential information on making measurement In shortened procedures There are parameters to set up in menu procedure Profile with parameters temperature time Finish drying Displaying result Frequency of printouts 9 1 Way of shortened menu activation To use shortened procedures menu switch on P7 Globals in user s menu in parameter s group parameter P7 02 data base of the programs value 0 E 10 10 03 Setup 10 10 10 P1 Calibration D E After pressing Setup key window with P4 Readout parameters group is displayed P5 RS 232 P6 Printouts P7 Globals A y o 10 10 03 Setup 10 10 10 P1 Calibration ke aik Using functional keys set cursor next P3 Date Time P4 Readout to group of parameters P7 Inne P5 RS 232 P6 Printouts P7 gt Globals Procedure 40 2 Setup 13 47 56 P7 01b Password 0 J 02 Prog library 1 8enabled 03 Beep 1 1 enabled 04 Language Hi 1 English 05 Backlight ijon 06 Contrast lee function 07 Balance ld 12564 L Ba 10101 Setup 13 47 56 P7 gt 01 gt Password i o 2 Prog library 1 enabled 03 Beep i 1 enabled 04 Language 4 English 05 Backlight
71. y 8 5 Functions connected with use of RS 232 User can set up the parameters for correct communication with the computer or printer Enter menu group P5 RS 232 10 11 01 13 47 56 P5 gt 01 gt Baud rate i 3 9600 02 Parity n 0 none 03 Date bits 118 bits 04 Stop bits i 0 i 1 bit 05 Handshake n 0 none 06 Printon stab 1 1 enabled 8 5 1 Set up values of parameters By means of functional keys set up values of parameters see point 7 of the manual moving in user menu which have influence on correct communication analyzer computer or analyzer printer 29 Number and parameter name Value of parameter Number and parameter name Value of parameter 01 Baud rate 2400 4800 9600 19200 05 Handshake 0 none 1 RTS CTS 2 XON XOFF 02 Parity none even odd 06 Print on stable 0 disabled 1 enabled 03 Date bits 7 bits 8 bits 07 Printer type 0 Stand 1 Kafka 2 Epson 04 Stop bits 1 bit 2 bits N N NP 20O0ON 0O 8 6 Printouts 118 11 02 P6 01 Printout No Header No 02 03 04 05 06 07 08 09 10 11 12 13 j l 45 Line No Footer No Pr 1 start Pr 1 stop Pr 2 start Pr 2 stop Pr 3 start Pr 3 stop Pr 4 start Pr 4 stop String 1 String 79 Setup Li LI LI I Li I I LI LI Li Li Li I Li LI I
Download Pdf Manuals
Related Search
Related Contents
Projet Gestactiv WARNING - MyToolService.com T_707_manual_out9W VIDAS® 25 OH Vitamin D TOTAL pédiculose outils d`intervention en milieu scolaire Limpiador dieléctrico en aerosol no inflamable para equipos Manual do Usuário (Gerenciador Financeiro) Copyright © All rights reserved.
Failed to retrieve file