Home

RF300 large loop antenna - Laplace Instruments Ltd

image

Contents

1. Figure 8 3 3 a The configuration of these loops is as shown in fig 9 and fig 10 The connectors on each end of the loop are mated with the sockets either side of each loop transducer The retaining ring on each connector needs to be screwed firmly to the socket but take care These rings will bind and be difficult to turn unless the loop is aligned properly Force is not necessary simple jiggle the loop into alignment and the ring will turn easily Figure 9 Page 10 Laplace Instruments Ltd Screw loops connectors into transducer sockets Do not force adjust alignment to ease tightening BNC outlet ector Figure 10 3 3 b Install the first loop on the stand This is the horizontal orientation This task requires at least 2 persons The loop will simply rest in the slots at the top of the corner posts Locate the current transducer close to one of the corner posts It will be found that if the loop does not lie flat by rotating it relative to the stand and or inverting it an optimum position can be found which results in the flatest loop Note that there is no requirement for a perfectly flat loop Some vertical displacement round the loop will not affect results Figure 11 Figure 12 M6 x 12 screws Laplace Instruments Ltd 3 3 c The second loop is now fitted Figs 12 and 13 show the details This loop is fitted inside the first horizontal loop orientated s
2. 00 20 00 10 00 0 00 0 01 0 10 1 00 10 00 100 00 Frequency MHz Antenna factor Antenna factor RF300 s n 9108 15 00 10 00 5 00 a 0 00 o 5 00 S a 10 00 E 15 00 c X 20 00 25 00 30 00 0 01 0 10 1 00 10 00 100 00 Frequency MHz Page 27 Laplace Instruments Ltd Page 28 Laplace Instruments Ltd Appendix B Changes to standards CISPR15 and CISPR16 2009 Page 29 Laplace Instruments Ltd Technical Report Laplace Instruments Ltd January 2010 RF300 Large loop antenna an analysis of changes to the standards These changes are due to the amendments to CISPR15 and CISPR16 The latest versions are now CISPR15 2006 A2 2009 and CISPR16 1 4 2007 A1 2008 The key changes related to LLAs are 1 Sections related to the construction and specification of LLAs are moved from CISPR15 to CISPR16 Note that in the new CISPR15 the requirements for the LLA are referred to Section 4 7 1 in the new CISPR16 which does not exist It seems that the reference should be to Section 4 6 1 The same error is repeated in CISPR16 which again refers in Annex C to the non existent section 4 7 2 The definition of the calibration data has been re defined CISPR15 a e di J Annex AnnexC A ntenna S stem onnectors lause C3 onstruction of loop nformation previously included with diagrams now included in text Note low for inner conductor is required Clause B3 Positioning of the LAS equire
3. is not affected by the environment 4 1 Shipping list Quantity Item Loop Main table antenna 1 Main table ew Central pillar pa 1 boop antenna adaptor 5m BNC BNC calble m Qty 4 Legs Central pillar Q eL Grip Washer Vertical screw adaptor I3 5m BNC BNC cable Figure 17 Page 15 Laplace Instruments Ltd 4 2 Assembly Mount the table on the 4 legs These just slot into the recess on the underside of the table Locate the table in the centre of the RF300 with the centre in line with the central pillar of the LLA Slot the central pillar of the calibratrion assembly into the hole in the centre of the table Figure 18 The assembly should now be as shown in fig 19 Figure 19 7 Page 16 Laplace Instruments Ltd 4 3 a Horizontal axis a Locate the washer Ll over the top spigot i on the cenral pillar then locate the loop antenna on this top Spigot Figure 20 4 3 b Vertical axis Locate the vertical adaptor over the top Spigot Add the washer to the horizontal stub then the loop antenna Retain both items with the grip screw The calibration antenna can be rotated to check the several angles as required by the standard and the whole top assembly rotated through 90 degrees to Figure 21 reposition to the other vertical axis Page 17 Laplace Instruments Ltd 5 0 Calibration The calibration techn
4. 13 Laplace Instruments Ltd 3 4 Mounting the EUT Within the loop construct a wooden stand or table to suit the products to be tested This is not included with the RF300 because it needs to be matched to individual customers requirements The stand should hold the product roughly central within the loop 3 5 Connections Connect the loop transducers to the switch unit as shown in fig 16 The 3 transducer cables are identified by having thick RF absorber filters along their length Fit the cables so that these filters are nearest to the switch unit The switch unit is intended for floor mounting or may be mounted on a Suitable table if preferred The short co ax cable acts as a patch cable to switch each input one at a time to the output The output from the switch unit is connected direct to the analyser or receiver Loop transducers Horiz Vert Vert X Y Z Ferrite absorbers are at this end of cables Select loop using link cable To analyser Figure 16 Page 14 4 0 4 0 Laplace Instruments Ltd 4 0 Calibration loop RF300C The RF300C calibration loop is manufactured to comply with CISPR16 Full details of the use of this loop are provided in this standard This item RF300C is available as a separate option to go with the RF300 LLA The RF300 is supplied fully calibrated but the calibration loop is available to those who wish to check the calibration at regular intervals and to ensure that the calibration
5. 6 Switch the UUT on Check the levels of signal over the background levels using the techniques used for conventional radiated testing as described in the user guide The levels displayed are fully compensated for the RF300 characteristics Page 19 Laplace Instruments Ltd and can be compared directly with the limit lines 7 ALWAYS check for compression overload when taking measurements The amplitude of the reading on the screen cannot be used directly as a guide for compression because the readings have been adjusted for the RF300 characteristics If it is suspected that the signals are too strong even with the attenuator switched in install the 10dB attenuator in the input lead to the switch unit This additional attenuator can be automatically compensated for in the software by adjusting the pre amp settings on the screen 10dB attenuator Figure 22 8 If the background and the product is quiet and especially at frequencies above 20MHz use OdB attenuation Page 20 Laplace Instruments Ltd APPENDIX A CALI BRATI ON DATA 1 CISPR16 ideal plot 2 Calibration data 3 RF300 actual sensitivity plots and RF300 correction plot Antenna Factor Page 21 Laplace Instruments Ltd This Page intentionally blank Page 22 Laplace Instruments Ltd EN55015 Standard antenna response rmi 2 o E S 5 E a lt q al a o TT co a 9 o swyogp 10394 uonepijeA Freq MHz Page 23 Lap
6. ACE INSTRUMENTS LTD Tudor House Grammar School Road North Walsham Norfolk NR28 9JH UK Tel 44 0 16 92 40 20 70 Fax 44 0 16 92 40 49 10 Web www laplace co uk E tech laplace co uk Page 32
7. LAPLACE INSTRUMENTS LTD RF300 Large Loop Antenna Van Leen Loop 2 metre diameter Mk II A USER MANUAL Ver 2 00 October 2011 Serial number XXXX Laplace Instruments Ltd This Page intentionally blank Page 2 Laplace Instruments Ltd Index 1 0 Introduction 2 0 Packing List 3 0 Assembly 4 0 Calibration loop 5 0 Calibration 6 0 Operation Appendix A Calibration data B Latest changes to standards CISPR15 amp CI SPR16 Page 3 Laplace Instruments Ltd This Page intentionally blank Page 4 Laplace Instruments Ltd 1 0 Introduction The RF300 large loop antenna has been developed to meet the requirements of EN55015 2006 A2 2009 section 9 9 1 which refers to CISPR16 1 4 2007 Al 2008 section 4 6 1 This specifies the limits for magnetic field induced current for luminaires with lamp operating frequencies in excess of 100Hz Construction of the RF300 is as detailed in annex C of EN55016 This antenna is fully compliant with the standard and details of the calibration factors are included in this manual These antenna factors should be added to the results to obtain correct measurements that can be compared directly with the limits Test setup and limits should be conducted as required by EN55015 section 4 4 The design of the RF300 is such that it can be readily erected and provide a rigid and stable configuration for accurate and repeatable measurements A calibration loop RF300C is availabl
8. e as an option if on site calibration is deemed necessary The RF300C allows the calibration loop to be used on all three axes and rotated within each as required by the standards Page 5 Laplace Instruments Ltd 2 0 Packing List On receipt of the RF300 check the contents of the packages against the diagram shown in fig 1 Qty 31 Grip PBB ae wm OU 31 grip nuts db db Ba d Plate C Qty 8 M6 nuts 1123 Qty 3 loops Qty 2 Plate B Qty 2 Plate A hi Qty 4 Corner posts T aty 2 tiewraps 3 Loop transducers Qty 3 loop leads i PPPP Central Saddle clips pillar Tl 10dB attenuator OO ee oo Switch unit oo with cable Qty 1 5m cable O JO JO JO O O JO JO oo oo oo oo Qty 4 Base extensions Identify each component and check for shortages Page 6 Laplace Instruments Ltd Quantity Item Plates A Plates B Plates C Plates D Loop antennas Blue Corner posts Base extensions Central pillar A HB FF FW 0 FN N Saddle Clips UJ E Grip screws M10 w E Grip nuts M10 M6 x 12 screws and nuts Loop transducers current sensors Switch Unit Short BNC BNC cable patch cable BNC BNC cable 5m output cable BNC BNC cable 1 5m loop leads 10dB attenuator N fF W kB HR LE UU Oo Tiewraps Pa
9. ge 7 Laplace Instruments Ltd 3 0 Assembly 3 1 Stage 1 Central sub assembly Establish a clear area with a level floor clear of metallic structures and having a 3m clearance height 1 a Secure Plates A to the central pillar using the supplied grip screws Note that the plates have tapered holes to mach the taper on the grip screws E so ensure that these 6 Central pillar plates are assembled correct way round ge 1 1 b Secure Plates B as shown in fig 3 Plates A Figure 2 Plates B Figure 3 1 c Place the central post assembly in the centre of the prepared area and add the 4 base E extensions each using 2 grip screws Horizontal strut Repeat in 4 places Figure 4 Page 8 Laplace Instruments Ltd 3 2 Stage 2 Corner post assembly 0 Corner 0 post 2 a Add plate C to the base of each corner post using 2 grip screws Repeat for the other corner posts Note that the plate is fitted on the opposite side to the 2 mounting holes at the top of the post NO 2 b Add the 2 Plates D to each corner post as shown In Fig 6 Plate C Figure 5 Plates D Figure 6 2 c Secure the corner posts to the outer end of each base extension gt Join posts to horizontal struts Figure 7 Page 9 Laplace Instruments Ltd 3 3 Assembly of Loops The frame should now look like that shown in Figure 8 The next steps are to add the loops
10. ique is fully described in CISPR16 It requires the use of a signal generator covering the range 9KHz to 30MHz with a 50ohm output that is calibrated for 1v rms emf See appendix A The calibration data for each loop is virtually identical The following details therefore apply to all three axes although full data for each axis is given in the appendix For reference the ideal curve is shown in graph 1 This is taken directly from EN55016 annex C Figure C8 The antenna factor data for the RF300 are also given This antenna factor correction data effectively converts the output from the RF300 measured in dBuV directly to dBUA The data can be used with any analyser or receiver capable of EMC measurements The estimated measurement uncertainty is 3dB For a detailed explanation of the calibration of the RF300 and discussion of changes to CISPR15 EN55015 and CISPR16 see appendix B If using the SA1002 or SA3000 analyser The RF300 is listed under the normal Inputs menu Just select this device and the software automatically applies the appropriate correction factors so that the trace reads correctly in dBuA and can be compared directly with the EN55015 limits Further details are given in section 6 1 Page 18 Laplace Instruments Ltd 6 0 Operation Details of the measurements and limits are given in EN55015 section 4 4 A copy of this should be consulted if performing compliance tests The UUT should be mo
11. lace Instruments Ltd This Page intentionally blank Page 24 Calibration Laplace Instruments Ltd Calibration data 9108A Average 37 00 36 00 48 00 48 00 57 00 57 00 68 00 67 67 71 00 71 00 72 00 72 00 69 00 69 00 64 00 64 00 63 00 63 00 61 00 60 33 56 00 56 00 55 00 55 00 52 00 52 00 Calibration sources Laplace Instruments Ltd RF300C calibration loop to CISPR16 annex C5 Laplace Instruments Ltd SA1002 Spectrum analyser s n 1010 Calibration Laplace 16 03 2009 Marconi TF2016A signal generator s n 118032 004 Calibration Checked against Marconi TF2019A s n 118449 169 Advantest R4131D calibrated by Schaffner 20 11 02 and checked against Marconi TF2019A s n 118449 169 Marconi TF2019A serial number 118449 169 calibrated by Industrial Calibration Ltd 22 06 07 Page 25 Laplace Instruments Ltd This Page intentionally blank Page 26 Laplace Instruments Ltd Actual output plots for the loops Appendix B Output Output RF300 s n 9108 80 00 70 00 60 00 S 50 00 2 a o 40 00 2 g 6 30
12. ment for minimum distance to nearby objects not included in new ISPR16 Clause B4 lause C4 alidation ew definition for validation factor see ee TG Figure B1 General view view O pei fe Re Figure B4 Figure C8 alidation factor onverted from dBuA to Zizzi See below Figure C1 Figure C11 Sensitivity vs diameter vs diameter Figure C2 Figure C10 onversion factors between ii Factors el magnetic field with electric urrent and magnetic field field added Factors for distance 30m strength at a distance removed Page 30 Laplace Instruments Ltd Note 1 The details of the LLA were given in Annex B of CISPR15 These are now transferred to Annex C of CISPR16 Most of the content has remained the same but Table 1 summarises the changes Note 2 CISPR15 gave the verification data as a plot of loop current in dBuA vs frequency for the standard test signal 1V open circuit voltage with a source impedance of 50ohm This seems to be a straightforward method especially as the limits are quoted in dBuA so it s a direct correlation between the calibration loop and the limits CISPR16 is essentially the same information but presented differently It specifies the relationship between the source voltage 1V as specified above and the output current in the loop as measured by the current probe Note that the current probe has a transfer characteristic of 1V A The relationship between volts and current is ohms hence the use of dB ohm
13. o that it fits in the lower slot in the central pillar and is held by the saddle clips at the top of the corner posts fig 13 Arrange the loop so that the transducer is within 20 cm of the central pillar Adjust the position of the loop in the saddle clips to give best circular shape of the loop Figure 13 Laplace Instruments Ltd 3 3 d The final loop is now fitted INSIDE the other two See fig 14 This fits in the upper slot in the central pillar and under the other vertical loop at the top Again arrange the loop so the transducer is close to the central pillar 3 3 e The loops can now be adjusted for best shape p s and position Generally the C vertical loops will tend to sag into the lower half of Figure 14 the antenna These are held up in position by using the saddle clips on each corner post to take some on the weight of the loop The top of the vertical loops can be adjusted for best shape and the two loops fastened together with the tiewraps as shown in fig 15 With care the tiewraps can be adjusted to produce reasonable circles for each loop Note that the exact shape is not critical Deviation from perfect circles is inevitable but this has no significant effect on antenna performance EN55015 states in section 7 2 that even the position of the UUT in the antenna is not critical v T lt Figure 15 Arrangement at top of loops Use 2 tywraps in series for adequate length Page
14. s as the validation factor The result is therefore a conversion factor scaled in dB Q to convert current to voltage CISPR16 defines the validation factor dB Q 20 log Vs li where Vs is the source voltage and li is the loop current Vs 1V 1 000 000uV Under old CISPR15 for li o 100KHz 46dBuA 200uA So the new CISPR16 value is 20 log 1000000 200 74 dB Q and Old CISPR15 for li o 30MHz 29dBuA 29uA So the new CISPR16 value is 20 log 1000000 29 91 dB Q These calculations confirm the relationship between the CISPR15 plot and the CISPR16 validation factor The plots in the standards assume a current probe with a 1V A transfer function Such probes are active but provide a flat frequency response The RF300 uses passive probes which have a non flat frequency response This is not important if the probe is inside the calibration loop and has a linear transfer function with amplitude These factors hold true for the probe that is used So the RF300 antenna uses an antenna factor correction to produce a calibration that agrees with the validation factor This antenna factor is supplied with each antenna and is equivalent to the correction factors as supplied with all EMC antennas test cells LISNs and other types of transducer Using the antenna factor data with the RF300 enables the output to be compared directly with the limits as specified in EN55015 Page 31 Laplace Instruments Ltd LAPL
15. unted on a wooden frame or table in the centre of the antenna The position is not critical Connecting cables to the EUT should leave the volume enclosed by the loops in such a way as to be kept away from the loops to avoid capacitive coupling See Fig C6 in CISPR16 Each axis loop should be measured in turn Each should meet the requirements of the standard The loops are individually selected by connecting the short patch lead to the appropriate input socket as shown in fig 11 6 1 Measurement with the SA1002 or SA3000 and EMCEngineer Windows software 1 Select the RF300 item under the input menu The vertical scale will indicate units of dBuA If this item is not available you need a later version of the software Contact your supplier 2 Under the limits menu select the EN55015 2m loop antenna limit line 3 Connect the switch unit to the analyser input 4 With the UUT switched off check the background signal level At frequencies below 1MHz the background can be very strong If strong signals do exist check that the analyser is not in compression by changing the input attenuator setting on the analyser and comparing scans Apart from the base line the traces should overlay If this happens then it is OK to use the analyser with that attenuator setting If not increase the attenuation and try again 5 If the background is strong it is advisable to either find a quieter location or screen the room