Veterinary GrPro™ 2.1 Control Unit User`s Manual

image

Contents

1. NOTE Ur is the AC mains voltage prior to application of the test level EN 16 Table 3 for RF Electromagnetic Immunity GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY environment IMMUNITY TEST Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 NOTE 1 At 80 M NOTE 2 These g absorption and r IEC 60601 TEST LEVEL 3Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz H COMPLIANCE LEVEL 3Vrms 3 V m z and 800 MHz the higher frequency uidelines may not apply in all situations eflection from structures objects and people The Veterinary GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the Veterinary GRPro 2 1 should assure that it is used in such an ELECTROMAGNETIC ENVIRONMENT GUIDANCE Portable and mobile RF communications equipment should be used no closer to any part of the Veterinary GRPro 2 1 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 VP 80 MHz to 800 MHz d 1 2 VP 80 MHz to 800 MHz d 2 3 VP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts VV according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electrom
2. WARRANTY LIMITATIONS AND EXCLUSIONS THESE WARRANTIES REPLACE ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE COOLSYSTEMS MAKES NO EXPRESS WARRANTIES BEYOND THOSE STATED HERE COOLSYSTEMS DISCLAIMS ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE SOME JURISDICTIONS DO NOT ALLOW THE EXCLUSION OF IMPLIED WARRANTIES SO THIS LIMITATION MAY NOT APPLY TO YOU ALL EXPRESS AND IMPLIED WARRANTIES ARE LIMITED IN DURATION TO THE LIMITED WARRANTY PERIOD EN 22 NO WARRANTIES APPLY AFTER THAT PERIOD SOME JURISDICTIONS DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS SO THIS LIMITATION MAY NOT APPLY TO YOU LIMITATIONS OF LIABILITY COOLSYSTEMS RESPONSIBILITY UNDER THIS OR ANY OTHER WARRANTY IMPLIED OR EXPRESS IS LIMITED TO REPAIR OR REPLACEMENT AS SET FORTH ABOVE THESE REMEDIES ARE THE SOLE AND EXCLUSIVE REMEDIES FOR ANY BREACH OF WARRANTY COOLSYSTEMS IS NOT RESPONSIBLE FOR DIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY BREACH OF WARRANTY OR UNDER ANY OTHER LEGAL THEORY INCLUDING BUT NOT LIMITED TO LOST PROFITS DOWNTIME GOODWILL AND DAMAGE TO OR REPLACEMENT OF EQUIPMENT AND PROPERTY SOME JURISDICTIONS DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES SO THE ABOVE LI
3. 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor If you are not sure who the local distributor is call Game Ready Customer Service at 1 510 868 2100 and we will refer you to one You must have returned the Warranty Registration card to CoolSystems within thirty 30 days from the date of purchase to qualify for warranty service If you qualify for warranty service from CoolSystems you will be issued a Returned Material Authorization RMA number When you return the Veterinary GRPro 2 1 Control Unit to CoolSystems you must write the RMA number on the outside of the package CoolSystems will not accept returned Veterinary GRPro 2 1 Control Units without an RMA number on the package If you return the Veterinary GRPro 2 1 Control Unit to CoolSystems you must assume the risk of damage or loss during shipping You must use the original packaging or the equivalent CoolSystems may require you to verify in writing that you are the original purchaser of the Veterinary GRPro 2 1 Control Unit CoolSystems may elect to replace or repair the Veterinary GRPro 2 1 Control Unit with either a new or reconditioned product The returned product shall become CoolSystems property upon receipt The replacement Veterinary GRPro 2 1 Control Unit is warranted under this written warranty and is subject to the same limitations and exclusions for the remainder of the original warranty period THIS WARRANTY IS NOT TRANSFERABLE
4. Any solvent based cleaners Doing so will damage the plastics and will void your warranty Abrasive materials Doing so will damage the plastics and will void your warranty Any petroleum based lubricants Doing so will damage the o rings and will void your warranty If lubrication is required the use of silicon spray is recommended WRAPS Gently remove Heat Exchanger from the Sleeve Hand wash the Sleeve in cold water using a mild detergent or antibacterial soap Do not use fabric softener Hang to dry If needed the external surface of the Heat Exchanger may be cleaned by wiping down with commercial non bleach cleaning wipes or hand washed using a very small amount of mild detergent or antibacterial soap DO NOT MACHINE WASH Refer to the Wrap Use Guide accompanying individual Wraps for more information EN 9 CARRY BAG The Carry Bag should be cleaned using a soft cloth or brush and a mild detergent Febreze or the equivalent can be used on the Bag if desired If the Carry Bag has a biological material on the surface Steri Fab may be used to decontaminate those surfaces Be sure to test any product on a small portion of the Bag to ensure that it will not cause damage Note To operate the Veterinary GRPro 2 1 System you do not need to remove it from the Carry Bag Simply unzip the Bag s main compartment and end panel Fill the reservoir with ice and water Attach the Connector Hose and the AC Adapter to the end panel of
5. Inc FSP030 RCAM in order to be compliant with IEC EN 60601 1 2 section 6 1 and 6 2 EMC requirements The use of accessories transducers and cables other than those specified and sold by the manufacturer of the Veterinary GRPro 2 1 System as replacement parts for internal components may result in increased emissions or decreased immunity of the system The Veterinary GRPro 2 1 System should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the Veterinary GRPro 2 1 System should be observed to verify normal operation in the configuration in which it will be used The essential performance of the Veterinary GRPro 2 1 System is Compression Cycle High cyclic 5 75mm Hg Med cyclic 5 50 mm Hg Low cyclic 5 15 mm Hg NO Pressure wrap shall be vented to atmosphere The cooling temperature of the circulating ice water will be adjustable between 34 F 1 C and 50 F 10 C as long as the ice water in the ice box is supplied with sufficient amount of ice EN 18 TROUBLESHOOTING ERROR WHAT DOES IT MEAN Air Pressure Sensor The Control Unit has detected a problem calibrating the air pressure circuit on start up WHAT CAN 1 DO This is most likely to occur if you have restarted the System with an inflated Wrap attached Detach the Wrap pressing it flat to expel the air accumulated inside and try again Self Test Error Air Pump The Control Unit has detected an
6. all connections are secure Test outlet with another electrical device Press the power button again If the Control Unit still will not turn on call Customer Service Control Unit won t reach Target Temperature or temperature is unstable First completely fill the reservoir with ice then fill the remaining space with water to administer therapy on the coldest se unable to achieve your de try the following steps e f you are still unable to temperatures be sure tting If you are still sired target temperature achieve WARMER he Temperature Knob is turned to full warm use less ice and reduce the amount of water if necessary If you are still unable to temperatures confirm achieve COLDER hat the Temperature Knob is set to full cold and that the reservoir is full of ice replenishing requently and stirring the reservoir to break up large formations of ice if necessary Make sure there are no kinks in the Wrap or Connector Hose Reapply the Wrap making sure to follow all accompanying application instructions Finally if you still cannot achieve the coldest temperature desired adding more water may also allow the Control Unit to reach colder temperatures To achieve this effect water may be added to the top of the ice level CAUTION By overfilling the reservoir as in the last step the temperature control feature of the System will not work and the Control Unit will be administer
7. pages 12 amp 13 of this manual Set your target temperature displayed in the Target Temperature window by turning the knob Turning the knob clockwise towards 3 snowflakes will Q provide more cold therapy Turning the knob counter clockwise towards 1 snowflake will provide less cold therapy ii EN 7 Manual You can choose to operate in either Manual Mode or Program Mode E Manual Mode allows you to adjust time and pressure settings as you choose Program Mode allows you to choose one of six programs that provide treatment for a set time then sleep providing no treatment for a set time continuously at a specific pressure setting refer to list of available programs below The unit automatically starts in Manual Mode Program Manual Mode Set the time in five minute increments by pushing the buttons Set the pressure by pushing the pressure button You can select A from 4 pressure settings No pressure Low Pressure 5 15mmHo o Medium Pressure 5 50mmHo o High Pressure 5 75mmHg Program Mode Push the program button to enter Program Mode In Program Mode the unit will operate continuously according to the selected program You will need to replenish ice and water as previously indicated in step 2 You can select from the following programs Push the Program Button to scroll through the available programs Program 1 30 minutes on 30 minutes sleep No pressure Progra
8. 1 Control Unit should be plugged into a grounded electrical outlet prior to operation e Position the Control Unit to minimize the risk of tripping over the Control Unit Connector Hose or power cord Precautions e Failure to properly follow the instructions of this manual and those of your medical provider may interfere with or prevent delivery of appropriate therapy e To avoid risk of electrical shock unplug the Control Unit from the electrical outlet prior to filling the Control Unit with ice and water Push the door release button to open the ice box door EN 5 Fill entire reservoir with ice Then add water to the reservoir up to 1 below lid Place the Control Unit in the location where you plan to use it The Control Unit should only be used in the upright orientation as shown The Control Unit will leak if placed on its side Connect the AC Adapter to the receptacle on the end panel of the Control Unit then plug the AC Adapter into a grounded electrical outlet The power indicator light orange color on the Control Unit should illuminate Press the power button The screen should light up and the Control Unit should beep twice The power indicator light should turn from orange to fe N green If you do not see these indicators refer to Control Unit will not turn on on page 21 of the Troubleshooting Guide in this manual AC Adapter Note The Control Unit should be turned o
9. CA 94520 In the United States call toll free 1 888 GameReady 1 888 426 3732 Outside of the United States 1 510 868 2100 www gamereadyvet com DESCRIPTION OF THE VETERINARY GR PRO 2 1 SYSTEM The following items are included in your Veterinary GRPro 2 1 System Power Cord AC Adapter Welcome Packet lt lt fas the a s Manual and other information about your System Connector Veterinary Hose GRPro 2 1 Control Unit Heat Exchanger Wrap sold separately Sleeve A Wrap comprised of an inner Heat Exchanger and an outer a System to begin treatment Each Wrap is sold separately and is not included in the System Volume Pressure Indicator Power Pressure Backlight Indicator Actual o Temperature Power Ns Volume Clas Program Program Temperature Pause Knob Play Ga i Pause Seep in Coldest Time Time Wrap Temp Target C F Celsius Fahrenheit Tempere Subtract Add Time Time Manual Mode TABLE OF CONTENTS Detailed Instructions for Use 2 Modes of Operation 2 Buttons 2 Adjusting Temperature 3 Display 3 Operating the System 5 Storage 8 Cleaning 9 Maintenance 10 Accessories 11 Indications for Use 11 Contraindications 12 General Warnings and Cautions 12 Specifications 13 Safety Electrical Classification 14 Electromagnetic Compatibility 15 Troubleshootin
10. GAME G2 READY VETERINARY GRPro 2 1 CONTROL UNIT User s Manual MODEL NUMBER 550450 This User s Manual is also available online at www gamereadycanine com and www gamereadyequine com under Products THE GAME READY DIFFERENCE THE GAME READY DIFFERENCE Game Ready was built upon the concept that the principles behind spacesuit technology could be used to develop an injury treatment system As a designer for the National Aeronautics and Space Administration NASA and an inductee into the U S Space Foundation s Hall of Fame our founder helped pioneer the technology involved in controlling pressure and temperature within a flexible garment In 1998 Game Ready began working with athletes athletic trainers veterinarians and engineers to expand breakthrough discoveries into the realm of sports medicine REGISTER YOUR PRODUCT Please complete your Warranty Registration for both the Veterinary GRPro 2 1 Control Unit and the Wraps online at www gamereadyvet com The Wrap registration card is packaged with the Heat Exchanger of each Wrap You must register within 30 days from the date of purchase in order to receive warranty service Further warranty information can be found on page 22 Let us know what you think of our products by calling us at the numbers below or emailing us at info gameready com We d love to hear from you GAME GR READY Brought to you by CoolSystems Inc 1800 Sutter Street Suite 500 Concord
11. MITATIONS OR EXCLUSIONS MAY NOT APPLY TO YOU THIS LIMITED WARRANTY GIVES YOU SPECIFIC RIGHTS AND YOU MAY ALSO HAVE OTHER RIGHTS THAT VARY FROM JURISDICTION TO JURISDICTION WARRANTY REGISTRATION REGISTRATION Please complete the Warranty Registration for both the Veterinary GRPro 2 1 Control Unit and the Wraps online at www gamereadyvet com You must register within 30 days from the date of purchase in order to receive warranty service You will need the following information to complete your Warranty Registration The Control Unit s model number REF and its serial number SN These numbers are located on the label on the bottom of the Control Unit Simply go to www gamereadyvet com click on the Product Registration link at the top of the page fill out and submit your information EXTENDED WARRANTIES Extended Warranties are available for the Veterinary GRPro 2 1 System For details and information in the U S call Game Ready Customer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor CE EN 23 GAME G2 READY Brought to you by CoolSystems Inc 1800 Sutter Street Suite 500 Concord CA 94520 1 888 GameReady 1 888 426 3732 1 510 868 2100 www gamereadyvet com Game Ready Control Unit User s Manual PN 703520 Rev A 2012 CoolSystems Inc All rights reserved
12. agnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Kp range applies Electromagnetic propagation is affected by a Field strengths telephones and field strength in compliance leve from fixed transmitters such as base s and mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured the location in which the Veterinary GRPro 2 1 is used exceeds the applicable RF above the Veterinary GRPro 2 1 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the Veterinary GRPro 2 1 b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m ations for radio cellular cordless EN 17 Table 4 for RF Immunity RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE VETERINARY GR PRO 2 1 The Veterinary GRPro 2 1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Veterinary GRPro 2 1 can help prevent electromagnetic interference by maintaining a
13. atch the temperature you have selected TIP The maximum amount of cold is dictated by the amount of ice in the reservoir and the setting of the temperature adjustment knob You may need to stir or replenish the ice to achieve coldest temperatures Ice will melt faster in the acute phases of injury as the Game Ready System is removing greater amounts of heat from the treatment site an ala DISPLAY Status bar Off Mute Vol Vol q AW a Manual f Z Manual Li gt Indicates selected mode Program Program Pause il Pause ii Indicates unit is paused Play gt Play gt Indicates unit is running Other Icons Error A Indicates an error See troubleshooting pages 19 21 for error codes Sleep O Indicates sleep mode time Battery Aut Indicates remaining battery capacity Temperature Actual Temperature Approximate temperature of the water leaving the Control Unit Target Temperature Indicates the target temperature setting The Control Unit will automatically match the Actual Temperature and Target Temperature as closely as possible EN 3 EN 4 Sleep time Displays the amount of run time count down bars remaining When the System Runiti d b is running in 70 _ Run time count down bars Program Mode JU o4 Displays the amount of sleep time remaining l When the System pesevst cee ime is running in 40 isp is e amount of run tim Manual Mode re
14. cing and repair please contact Game Ready Customer Service from the U S call toll free at 1 888 426 3732 from outside the U S call 001 510 868 2100 Opening the case will void the Game Ready warranty To avoid the risk of electrical shock always turn off the System and disconnect the power cord from its electrical outlet when not in use or before adding or emptying ice and water Do not use any adapter other than that provided by Game Ready Use of other adapters may result in electrical shock and will void the Game Ready warranty EN 12 CAUTIONS Caution This product is for veterinary use only and is not indicated for human use Caution Always attend the animal while the Game Ready Veterinary System is being used in the case that the animal needs to be quickly disconnected from the Control Unit e Do not operate the System without any water in the ice box e Never operate the System with damaged power cords or Connector Hoses or other mechanical damage or if the unit is otherwise not fully operational e Do not pour hot water into the ice box The System was not designed to operate and has not been tested with hot water e Do not pick up the Control Unit by the lid Carry using the handle only e Do not use other manufacturers wraps with the Game Ready System e Do not operate the Control Unit without a Connector Hose attached e Be careful not to trip over the System s power cords and Connector Hose or use t
15. ect Wrap from Connector Hose e Make sure there is water in the ice box Verify that the ice box filter is not clogged Make sure there are no kinks in the Wrap or Connector Hose Reapply the Wrap snugly making sure to follow all application instructions accompanying the Wrap Turn the Control Unit off and on again Disconnect and reconnect the hose from the Control Unit and the Wrap EN 20 ERROR gt T WHAT DOES IT MEAN Self Test Error Fluid Circuit The Control Unit has detected an electronic problem in the fluid circuit on start up Temperature Calibration Error The Control Unit has detected a failure in the Temperature Control circuit or is being operated outside of the advised temperature range 33 120 degrees Fahrenheit WHAT CAN I DO Turn off the Control Unit e Turn the Control Unit back on and resume treatment If the problem persists contact Customer Service e Be sure you are operating and storing the System within the advised temperature range of 33 120 degrees Fahrenheit e Turn the Control Unit off e Fill the Control Unit with ice and water Turn the unit back on again and push Play Pause e Repeat this process up to three times If the problem persists contact Customer Service Control Unit will not turn on Press the power button If no orange or green light is illuminated make sure the AC Adapter is plugged into a working electrical outlet and
16. electronic problem in the air circuit on start up Disconnect the Wrap from the Control Unit Turn the Control Unit off and on again without a Wrap connected Reconnect the Wrap and resume treatment f the problem persists contact Customer Service S e Dry Pump The Control Unit has detected a dry pump In order to prevent possible damage to the fluid pump the unit will stop therapy Note that if you are using new Wraps he water in the reservoir may have been depleted and causing this error Make sure here is adequate water in the reservoir based on the fill line indicator label Refer to instructions for priming a Wrap below Verify that the ice box filter is not clogged refer to filter maintenance instructions on page 10 Make sure there are no kinks in the Wrap or Connector Hose Turn the Control Unit off and on again Disconnect and reconnect the hose from the Control Unit and the Wrap verifying that an audible click is heard at both connection points Prime the Wrap using the following steps Select No Pressure Attach the hose to the unit and the wrap Lay the Wrap open and flat next to or lower than the Control Unit not on the body Run the System for 2 minutes Prime the Control Unit using the following steps Disconnect the hose from the Control Unit Now look at the Wrap connection location on the Control Unit On the top valve push the white pron
17. environment The Veterinary GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the Veterinary GRPro 2 1 should assure that it is used in such an flicker emissions IEC 61000 3 3 EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT GUIDANCE RF emissions CISPR 11 Group 1 The Veterinary GRPro 2 1 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The Veterinary GRPro 2 1 is suitable for Harmonie einissions Class A use in all establishments including domestic IEC 61000 3 2 establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic Voltage fluctuations Complies purposes EN 15 Table 2 for Transient Electromagnetic Immunity GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY The Veterinary GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the Veterinary GRPro 2 1 should assure that it is used in such an environment IMMUNITY TEST Electrostatic discharge ESD IEC 61000 4 2 IEC 60601 TEST LEVEL 6 kV contact 8kV air COMPLIANCE LEVEL 6 kV contact 8 kV air ELECTROMAGNETIC ENVIRONMENT GUIDANCE Floors should be wood concrete
18. g 19 Warranty 22 EN 2 DETAILED INSTRUCTIONS FOR USE Do not use this device without your veterinarian s specific recommendations for the frequency and duration of your treatments Be sure to read further safety warnings on pages 12 13 of this User s Manual MODES OF OPERATION Manual Mode The System automatically starts in this mode and allows the user to adjust treatment time and pressure settings Program Mode This mode allows the user to choose one of six treatment programs that provide therapy for a set time then sleep no treatment for a set time continuously at a specific pressure setting Drain Mode Allows a user to connect a Hose with a special Hose Adapter purchased separately to the unit enter Drain Mode and have the unit empty the water out of the ice box through the Hose Drain Mode can be accessed by pressing the program key and toggling through all six programs To empty water in the System while in Drain Mode press the program key until you reach Drain Mode attach the Hose Adapter to the Connector Hose place the Hose Adapter over a sink and press the Play button Drain Mode 1 will run the Control Unit s fluid pump for up to six minutes long enough rat to fully empty the ice box Drain Mode is indicated by the following icons BUTTONS Power Use this button to turn the Control Unit on and off Program Use this button to select one of the available Programs or to return to Manual M
19. g in so that it is flush with the metal connector Make sure you are not fully covering the opening in the prong Push start and water should squirt out of the valve Restart the System Over Pressure Indicates that the Control Unit has exceeded the target air pressure e Turn the Control Unit off and back on Make sure the Wrap is applied securely around the body Try using a different Wrap and Hose to isolate which component may be producing the error For example an error which occurs with a Utility Wrap but not with any other Wrap may indicate that the Utility Wrap itself is causing the error not the Control Unit EN 19 WHAT DOES IT MEAN Under Pressure The Control Unit can t reach its target maximum compression This often indicates that there is a leak in the pneumatic compression circuit either in the Connector Hose Wrap or Control Unit Or it may occur because the hook and loop fastener on your Wrap has worn out WHAT CAN I DO Make sure the Wrap is applied securely Try using a different Wrap and Hose to isolate which component may be producing the error For example an error which occurs with a Utility Wrap but not with any other Wrap may indicate that the Utility Wrap itself is causing the error not the Control Unit If using a Dual Hose make sure you have two Wraps attached Deflation Error The Control Unit has detected that the Wrap has not properl
20. gnificant vascular impairment in the affected region e g from prior frostbite diabetes arteriosclerosis or ischemia e Who have acute paroxysmal cold hemoglobinuria or cryoglobulinemia GENERAL WARNINGS AND CAUTIONS Thank you for choosing the Game Ready System To assure you have the best experience possible please be sure to read the entire User s Manual prior to use of the product WARNINGS Improper placement or prolonged use of Game Ready could result in tissue damage such as frostbite During the course of Game Ready therapy the treatment provider should monitor the tissue surrounding the treated region or the extremities of the treated limb for any abnormal appearance increased swelling or pain If any of these signs present or any changes in skin appearance occur such as increased redness discoloration or other noticeable skin changes you are advised to discontinue use and consult a veterinarian The Game Ready Wrap is not sterile do not place directly against open wounds sores rashes infections or stitches May be applied over dressing The Game Ready wraps are available in multiple configurations but are not intended for all possible uses Use caution when selecting a particular wrap for use on anatomic regions other than that for which the wrap is indicated and closely supervise the animal s status e To avoid the risk of electrical shock do not remove any panels from the Control Unit For all servi
21. he product where these are likely to trip an animal e The Game Ready Control Unit is a veterinary medical device Handle it with the same care as you would a laptop computer Do not drop it kick it or otherwise abuse it unnecessarily Such abuse will void all Game Ready warranties Do not place the AC Adapter or battery pack inside the ice box for storage or transport In addition to the precautions listed above additional warnings and safety precautions are posted throughout this manual Read and carefully follow these instructions prior to operating the System SPECIFICATIONS Size 16 25 length x 7 75 width x 9 25 height 413 x 197 x235 mm not including Carry Bag Weight 7 3 lb empty approximately 18 Ib full of ice and water Pressure level cycles from 5mm Hg up to 75mm Hg AC power 100 240 V 50 60 Hz 1 6A DC input 12V 2 5 A EN 13 SAFETY ELECTRICAL CLASSIFICATION Protection against electric shock 5 1 The Veterinary GRPro 2 1 System is considered to be Class protective earth when connected to the FSP Group Inc model FSP 030 RCAM power supply The Veterinary GRPro 2 1 System is considered to be Class double insulated when connected to the GlobTek Model GTM 21097 5012 power supply Protection against harmful ingress of water 5 3 This product provides ordinary protection against ingress of water Degree of safety in the presence of flammable anesthetics or oxygen 5 5 Not suitable for use in an o
22. ing therapy at full cold Consult the Warnings on page 12 to make sure you take proper steps to minimize the risk of injury Need more help with a problem In the U S call Game Ready Customer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor EN 21 WARRANTY MANUFACTURER S WARRANTY CoolSystems warrants that the Veterinary GRPro 2 1 Control Unit if properly used will be free from defects in material and workmanship for a period of two 2 years after the date the Veterinary GRPro 2 1 Control Unit was purchased If the Veterinary GRPro 2 1 Control Unit which is the subject of this Limited Warranty malfunctions during the warranty period for reasons covered by this Limited Warranty CoolSystems at its options will e REPAIR the Veterinary GRPro 2 1 Control Unit OR e REPLACE the Veterinary GRPro 2 1 Control Unit with another Veterinary GRPro 2 1 Control Unit THIS LIMITED WARRANTY AND ANY IMPLIED WARRANTIES THAT MAY EXIST UNDER STATE LAW APPLY ONLY TO THE ORIGINAL PURCHASER OF THE GRPRO 2 1 CONTROL UNIT AND ARE NON TRANSFERABLE Extent of Limited Warranty This limited warranty does not cover damages due to external causes including without limitation accident usage not in accordance with product instructions misuse neglect alteration or repair How to Obtain Warranty Service To obtain warranty service in the U S call Game Ready Customer Service at 1
23. m 2 30 minutes on 30 minutes sleep Low pressure Program 3 30 minutes on 30 minutes sleep Medium pressure Program 4 30 minutes on 60 minutes sleep No pressure Program 5 30 minutes on 60 minutes sleep Low pressure Program 6 30 minutes on 60 minutes sleep Medium pressure Program d Drain Mode Please refer to page 2 in this manual for details When you are done using the System for the day e Unplug the AC Adapter and the Connector Hose from the Control Unit e Push the door release button to open the door e Carefully pour out the ice and water Wipe off any excess water Store the Control Unit with the lid fully open to allow the interior to dry and preserve the reservoir seal Store the Control Unit in the Carry Bag or in another safe location Remember that your GRPro 2 1 Control Unit is a valuable piece of equipment and should be treated with great care like a laptop computer Storage Temperature Conditions 33 120 F 1 50 C Relative Humidity 10 95 non condensing Caution Do not keep in extreme cold or hot temperatures below 33 F or above 120 F or below 1 or above 50 C Do not leave in a hot or freezing car Do not leave the Control Unit in direct sunlight The UV light may damage or discolor the Control Unit EN 8 CLEANING CONTROL UNIT The exterior of the Control Unit and the visible interior surfaces of the reservoir can be cleaned with a soft cloth and one of the follo
24. minimum distance between portable and mobile RF communications equipment transmitters and the Veterinary GRPro 2 1 as recommended below according to the maximum output power of the communications equipment RATED MAXIMUM OUTPUT SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER POWER OF TRANSMITTER 150 KHZ TO 80 MHZ 80 MHZ TO 800 MHZ 800 MHZ TO 2 5 GHZ d 1 2 4P d 1 2 4P d 2 3 yP 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 73 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people WARNINGS Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the User s Manual Portable and mobile RF communications equipment can affect Electrical Equipment The Veterinary GRPro 2 1 System has to be powered with the AC adapter models GlobTek GTM 21097 5012 or FSP Group
25. n prior to attaching a Wrap to electrical outlet mmm EN 6 Connect the larger end of the Connector Hose with the red button to the Control Unit Make sure you hear it click To disconnect simply press the red button and remove the connector from the Control Unit Disconnect Button Connector Hose to Wrap gt 4 Apply the selected Wrap consult the Use Guide accompanying each Wrap for application instructions Caution e The Wrap is not sterile Do not place directly against open wounds sores rashes infections or stitches The Wrap may be applied over dressing e To ensure best fit be sure the Wrap is completely deflated prior to each application 8 Connect the smaller end of the Connector Hose with a blue or gray button to the Wrap Make sure you hear it click To disconnect simply press the blue or gray button and remove the connector from the Wrap Disconnect Button If you have any problems with the set up of your Veterinary GRPro 2 1 System in the U S call Game Ready Customer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor Warnings e Follow the treatment recommendations of your veterinarian for the use of this device e Improper placement or prolonged use of the Veterinary GRPro 2 1 System could result in tissue damage e For additional warnings and precautions please refer to
26. ode See page 8 in this manual for more information on Programs Play Pause Use this button to start or pause a treatment Add Time Use this button to add time in Manual Mode does not work in Program Mode You can add up to 90 minutes Treatment must be paused in order to add or subtract time Subtract Time Use this button to reduce time in Manual Mode does not work in Program Mode You can subtract up to 90 minutes Treatment must be paused in order to add or subtract time Pressure Selection Use this button to select one of four pressure settings No Pressure Low Pressure 5 15 mmHg Medium Pressure 5 50 mmHg and High Pressure 5 75 mmHg Pressure selection is not available in Program Mode Treatment must be paused to change pressure settings Volume Use this button to select the option of sound or no sound Push to mute sound Alarms will still sound even with Volume off C F Button Use this button to select either Celsius or Fahrenheit on the temperature display DOO O 0 000 Backlight Use this button to turn the backlight on or off To adjust the temperature being applied during treatment simply turn the temperature knob towards 3 snowflakes for the maximum amount of cold or towards 1 snowflake for the least amount of cold Notice that as you adjust the knob the Target Temperature on ADJUSTING TEMPERATURE the display will change The System will automatically adjust to m
27. or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast 2 kV for power 2 kV for power ains power quality should be that transient burst supply lines supply lines of a typical commercial or hospital IEC 61000 4 4 1 kV for input 1 kV for input environment output lines output lines Surge 1 kV line s to 1 kV line s to ains power quality should be that IEC 61000 4 5 line s ine s of a typical commercial or hospital 2kVline s to 2kVline s to environment earth earth Voltage dips short lt 5 Ur lt 5 Ur ains power quality should be interruptions and gt 95 dip in Ur gt 95 dip in Ur that of a typical commercial or voltage variations for 0 5 cycle or 0 5 cycle hospital environment If the user of on power supply 40 Ur 40 Ur the Veterinary GRPro 2 1 requires input lines 60 dip in Ur 60 dip in Ur continued operation during power IEC 61000 4 11 for 5 cycles or 5 cycles mains interruptions it is recommended oe ee ably Paar bee A cy power 30 dip in Ut 30 dip in Ur for 25 cycles or 25 cycles supply or a attery lt 5 Ur lt 5 Ur gt 95 dip in Ur gt 95 dip in Ur for5s or5s Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
28. the Control Unit and plug the AC Adapter into a grounded electrical outlet MAINTENANCE The reservoir filter should be inspected cleaned and or replaced as necessary Be A I identify the filter within Using two fingers grasp Slide the filter out the ice reservoir and squeeze the two protruding prongs Rinse debris from the filter and be sure there are no obvious signs of eS damage We If you have questions or would like to order a new filter in the U S call Game Ready EON Customer Service at N 1 888 426 3732 To replace the filter or install You will feel and 1 510 868 2100 a new filter first ensure that hear the filter snap from outside of the the filter is oriented properly back into place U S please contact with the plastic tab facing your local distributor up If the filter is not oriented properly the protruding tab will prevent the filter from sliding back into place EN 10 Other than maintaining the reservoir filter no product service should be performed by the user If your Veterinary GRPro 2 1 System requires service in the U S call Game Ready Customer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor ACCESSORIES The Veterinary GRPro 2 1 Control Unit can be used with any of the following accessories e Any Game Ready Wrap wraps made by any o
29. ther manufacturer CANNOT be used with this System Game Ready supplied Power Supply GlobTek Model GTM 21097 5012 or FSP Group Inc model FSP 030 RCAM and Hospital Grade Power Cord Game Ready supplied Connector Hose Game Ready Carry Bag Game Ready Drain Mode Adapter INDICATIONS FOR USE e Follow the recommendations of your veterinarian for duration and frequency of use for this device The Veterinary GRPro 2 1 System combines cold and compression therapies It is intended to prevent injuries treat post surgical and acute injuries to reduce edema swelling and pain where cold and compression are indicated EN 11 CONTRAINDICATIONS Compression Therapy vasopneumatic compression using the Game Ready System or any compression therapy device should not be used in patients e Who are in the acute stages of inflammatory phlebitis in the affected region e Who have any history or risk factors for deep vein thrombosis or pulmonary embolus including prolonged bed or stall rest in the affected region to be treated with this therapy e Who have significant arteriosclerosis or other vascular ischemic disease in the affected region e Who have a condition in which increased venous or lymphatic return is not desired in the affected extremity e g carcinoma e Who have decompensated hypertonia in the affected region Cryotherapy using the Game Ready System or any cryotherapy device should not be used in patients e Who have si
30. training When selecting a Displays pressure program Pressure off Low Med High Actual cr vs wv vw Temperature UW Votume cy G NAONNANA 50 10 Manual P Z a5 pay Program 408 Pause H Og Play gt w Sleep Run Target Time Time Temperature Displays amount of Displays amount of time unit will sleep time unit will run Pressure Of Low Med High off Low Med High d off Low Med High mmi off Low Med High d OPERATING THE SYSTEM To operate your Veterinary GRPro 2 1 System you need e Control Unit reservoir filled with ice and water Optimal performance is achieved by first completely filling the reservoir with ice then adding water to 1 below the reservoir lid e Game Ready supplied power supply e Connector Hose e Wrap includes a Heat Exchanger and Sleeve Notes e The Wrap is comprised of an inner Heat Exchanger and an outer Sleeve The combination of Sleeve and Heat Exchanger is referred to throughout this manual as a Wrap e The Veterinary GRPro 2 1 Control Unit should be placed on a stable surface such as the floor or a table during use e Note that using the System in an environment with a high ambient temperature may affect its ability to provide adequate cooling or may limit the ice life If you will be using the System with a Game Ready optional battery pack sold separately please consult the instructions for use that accompany that battery pack Warning e Your Veterinary GRPro 2
31. wing cleaning agents e Mild detergent e 70 Isopropyl alcohol 100 Simple Green e Quaternary ammonium such as Virex typically only found in a clinical use setting e Cavicide Procedure e Follow the manufacturer s instructions and precautions for the cleaning agent you select Apply the selected cleaner to a soft cloth and wipe down all surfaces of the Control Unit e Allow the Control Unit to dry thoroughly before storing it in the bag e The Control Unit should be cleaned as needed DO NOT USE e Phenolic based disinfectants such as Amphyl typically only found in a clinical use setting e Any solvent based cleaners on the Control Unit Doing so will damage the plastics and will void your warranty e Abrasive materials to clean the Control Unit Doing so will damage the plastics and will void your warranty Caution The Control Unit is not a waterproof device Do not apply a direct stream of any liquid onto the Control Unit submerge the Control Unit or allow any liquid to pool on the surface of the front panel of the Control Unit CONNECTOR HOSE The surface of the Connector Hose can be cleaned using a soft cloth and one of the following e Mild detergent e Steri Fab e 70 Isopropyl alcohol e We do not recommend the use of quaternary ammonium such as Virex or Cavicide DO NOT USE e Phenolic based disinfectants such as Amphy typically only found in a clinical use setting e
32. xygen enriched environment or in the presence of flammable anesthetics Powered by Direct Current Type BF Applied Parts NK Attention Consult accompanying documents before use FEC Electromagnetic interference This equipment has been tested and found to comply with the limits for medical devices in IEC 60601 1 2 2001 These limits are designed to provide reasonable protection against harmful interference in a typical medical installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving device e Increase the separation between the equipment Connect the equipment into an outlet on a circuit different from that to which the other device s are connected Consult the manufacturer or field service technician for help EN 14 ELECTROMAGNETIC COMPATIBILITY ELECTROMAGNETIC COMPATIBILITY INFORMATION ACCORDING TO IEC EN 60601 1 2 Table 1 for Emissions GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS
33. y deflated Turn the Control Unit off If there is air left in the Wrap disconnect the Wrap from the Connector Hose and manually deflate the Wrap by applying pressure to it Reconnect the Hose to the Wrap and reapply the Wrap to the body Turn the Control Unit on and press play pause Make sure the Wrap is applied securely against the body If possible try using a different Wrap and Hose to isolate which component may be producing the error For example an error which occurs with a Utility Wrap but not with any other Wrap may indicate that the Utility Wrap itself is causing the error not the Control Unit gt Pump Performance Error The Control Unit has determined that the fluid pump may be working too hard This could be caused by ice or debris in the fluid circuit In order to prevent possible damage to the fluid pump the unit will stop therapy Turn the Control Unit off and back on again Reapply the Wrap making sure to follow all application instructions accompanying he Wrap Disconnect and reconnect the hose from he Control Unit and the Wrap verifying hat an audible click is heard at both connection points f that does not solve the problem turn the Control Unit off for 20 minutes to let the pump cool down before turning it on again o try again gt Low Flow Control Unit has detected something blocking the water flow Check all hose connections Disconnect and reconn