        3 Cleaning & Maintaining the MagPurix 12 system
         Contents
1.   and humidity  is 30 to 8096RH  non condensing     A location that is flat and stable  with no vibration    A location away from direct sunlight  Block the sunlight by  closing curtains or blinds as necessary      A location which is well ventilated and not dusty    A location where the temperature does not go up and down    www ZINEXTS com Page 10 Ver  V1 00    suddenly  Warming a cold room suddenly or moving the  MagPurix 12 system from a room with low temperature to a  warm room may cause condensation inside the device   resulting in abnormal extraction      B A location where the temperature and humidity are kept within  the specified range  far from water taps  water heaters   humidifiers  air conditioners and heaters     m A location far from objects which generate strong magnetic  fields  motors  transformers  TV  audio speakers  magnets   etc    Bringing the MagPurix 12 system close to any type of  magnetic field may cause a malfunction      AN Warning     Do not use the MagPurix 12 system to a location where it is wet or  can be splashed with water  It may cause a device failure  fire  or  electric shock     When relocating the MagPurix 12 system  disconnect the plug from  the outlet first  If the power cable is damaged  this may cause a  device failure  fire  injury  or electric shock     AN Caution     Do not use the MagPurix 12 system in an unstable place such as a  slanted surface or a place subject to vibrations  It may cause injury  or device failure     Do no
2.  MagPurix 12 system    Check that the following are included in the package  Contact your  local representative if you notice any missing part s       ZZINEXTS     MagPurix 12s     Automatic Nuc       Q ty  A  MagPurix Instrument 1  B  Power Cord 1  C  Sample Tube Rack 1  D  Barcode Reader 1  E  User s Manual 1    www ZINEXTS com Page 8 Ver  V1 00    Please make allthe components are free from damages as soon  as you get the system  If any damage found  please contact your  local representative for the instant support     Note     Zinexts   s global warranty does not cover the damages from the  transportation or improper operation     1 2 Contents of Reagent Kits       Reagent Cartridge s     Reaction Chamber s    Tip Holder   Pestle  essential for some Kit types   Filtered tip   Sample tube   Elute tube    ommoomw gt     www ZINEXTS com Page 9 Ver  V1 00    Note       Reagent kits are separated purchase  Please contact your  local agents or representatives to get further info       The contents of reagent kits will be various  Refer to the  handbook of kits enclosed in reagent box for details     1 3 Operating Environment   Condition    Use the MagPurix 12 system in a location that meets the following  conditions        B The space for the allocation of the MagPurix 12 system  system must be remained 5 10 cm from instrument to side  walls     B A location where power can be provided  Be sure to use the  attached AC adapter     B A location where the temperature is 15 to 30 C
3.  Repairs to the MagPurix 12 system should only be performed by  such agencies as are specifically authorized by the ZINEXTS LIFE  SCIENCE CORPORATION     Only original the ZINEXTS LIFE SCIENCE CORPORATION  replacement parts should be used     If the equipment is used in a manner not specified by the  manufacturer  the protection provided by the equipment may be  impaired     About Reagent Kits     N CAUTION     Note     When handling any of the kits  refer to the respective handbook     Reagents in each kit should be handled by observing the safety  information and precautions regarding the kit     Extraction should be performed in an appropriate laboratory or  workplace     The kits are not supplied with the MagPurix 12 system  Select the  desired kit s  and order it  them  separately     www ZINEXTS com Page 6 Ver  V1 00    About Samples     N BIOHAZARD     B Always wear appropriate gloves  a mask  and safety goggles etc   when handling any infectious samples     About Infectious Wastes    B When handling or disposing of infectious materials  follow the  laboratory guideline or the law regarding infectious waste to  perform proper incineration  fusion  sterilization  and or  disinfection     B When you use a third party to dispose of it  outsource this work an  operator licensed to handle medical waste subject to special  control  and give them the manifest of the medical waste at the  same time     www ZINEXTS com Page 7 Ver  V1 00    1 Installation  1 1 Composition of the
4.  expect      2  Remove the Cartridges and plastic disposables from the  instrument and then dispose of them     www ZINEXTS com Page 23 Ver  V1 00    3 Cleaning  amp  Maintaining the MagPurix 12 system    Two types of maintenance are performed on the the MagPurix 12  system Instrument as listed in the table  For details on each type of  maintenance  see below     Maintenance Type Performed by Schedule    Routine  Cleaning of Sample Tray After each use  Cleaning of instrument Bi weekly    body  outside     Cleaning of instrument Bi weekly    body  inside   Service  Engineer    B Clean the Sample Tray with mild detergent and rinse with  deionized water  Allow the parts to dry before use  Clean and  disinfect the platform surface by wiping with deionized water  followed by 70  ethanol        B Clean the instrument body by removing dust gently with a dry   soft cloth  If the outside of the MagPurix 12 system is heavily  soiled  or if any samples that may cause infection are adhered  to the outside of the the MagPurix 12 system  wipe with a soft  paper tissue  soaked with 0 5  sodium hypochlorite solution  or ethanol     www ZINEXTS com Page 24 Ver  V1 00    4 Troubleshooting     Probie  ease  sown    Instrument Problems    No power  the LCD  Screen remains blank  when the power is  turned on    LCD Screen turns on  when the power is on  but the self testing  program does not run  Protocol stops after an    initial start    Bubbles formed during    extraction    Presence of buffer
5.  in    the Cartridge Tray    Leakage from Filtered  Tips or uneven liquid  handling between  Filtered Tips    Blockage of tips    www ZINEXTS com    AC power cord is not    connected    Forget to remove the  packing stuff from    instrument    Cartridge s    Plasticwares  Reaction  Chamber  Tip Holder   Filtered Tip Pestle   Sample Tube  Elute  Tube  incorrectly  loaded on the the  MagPurix 12 system  Problem with motion    sensors    Missed adding sample  or sample volume is  lower than the    recommended volume    Motor movements may  not be smooth   incorrect placement of  plasticware  or leakage    from tips    Air leakage on the  Filtered Tip       Use the correct cords     MagPurix Reagent Kits     proper motor movements     Check AC power cord connections at both ends  or    Turn off the instrument and remove the packing stuff     Technical problem Contact your local representative or agent    Turn off the power and then turn it on again to stop the  program  The system will move back to initial state     Re load them according to the instructions shown in    Note  you could not resume the protocol after stop    and you may lose your samples     Turn off the power and remove all samples and  plasticwares  Contact your local representatives   Be sure to add the sample to tubes prior to starting the  protocol  To ensure proper mixing of reagents in the  tip and prevent bubble formation during mixing  make  sure the sample volume is at least the recommended    volume list
6. MagPurix 12s     Automated Nucleic Acid Purification System     Instruction Manual       UMO1001    www ZINEXTS com Page 1 Ver  V1 00    TABLE OF CONTENTS    Page    INTRODUCTION  see 3  SAFETY INFORMATION    u  uhodux ine EN uExhE a 4  1 0  INSTALLATION  2    8  1 1 Composition of the MagPurix 12 System                                         8  1 2 Contents of Reagent Kits                          eeeeeeeeeeseeeeeeeeeenenn 9  1 3 Operating Environment   Condition                               eeeeseeesssss 9  1 4 Unpack the MagPurix 12 System                                eeeeeeeeeeeses 11  Eo VS U VIEW P                n 14  2 0  GETTING STARTED u  u 15  2 1  Lurningtr the DOW CM nern 15  Bee FOO al lO A A EE E E EE 16  2o EXV ACUO 6 e rer 16  2 4 After Extraction Operation            cccccccccsecccsececeeeeeeeeseeseeeseeeeseeeeas 23  3 0 Cleaning  amp  Maintaining the MagPurix 12 Instrument                      24  40 TFOUDIESNOOUING Ee TEILT 25  5O SpeciCatlONS sonens ee ae 27  IAP PEIN DI oei see 28    www ZINEXTS com Page 2 Ver  V1 00    INTRODUCTION    Thank you for purchase of the MagPurix 12 system which is a fully  automated  standalone robot that can purify nucleic acids within 30 45  minutes  With advanced magnetic bead separation technology  it enables  you to have high quality extraction results  Moreover  the most  user friendly interface makes users free from troublesome parameter  settings and maintenance     This guide contains important information regard
7. S com Page 13 Ver  V1 00    1 5 Overall View  B Front view with Sliding Door closed    LCD Screen  Control Panel    Sliding Door                   Moving Handler    B Front view with Sliding Door opened    UV Lamp    Operational Area    www ZINEXTS com Page 14 Ver  V1 00    B Back View    Power Switch       Power Inlet 1 x Barcode Reader  Port  amp  1 x USB Port    2 Getting Started    Biohazard     Always wear appropriate gloves  a mask  and safety goggles during  any biohazardous operations in extraction process  Even when  touching the device after any operation with a biohazard risk  wear  appropriate gloves and a mask since the device may be  contaminated     Important     Before starting extraction  put on appropriate gloves  a mask  and  safety goggles if required by the operation  In the operation from  preparing samples to extraction completion  be careful not to  contaminate the samples with sweat  saliva  etc     2 1 Turning on the power    www ZINEXTS com Page 15 Ver  V1 00     1  Make sure that the power cable are connected securely to the  the MagPurix 12 system      2  Connect the plug of the power cable to the outlet        2 2 Preparation    The following preparations are required for extraction operation   B liems to be prepared    Gloves    Mask    Safety Goggles    Reagent Cartridge       Reaction Chamber     Tip Holder       Pestle  Option      Filtered tip     Sample tube     Elute tube     www ZINEXTS com Page 16 Ver  V1 00    B For the preparation of s
8. amples  refer to the handbook of  each kit        Provided with the Reagent Kits     2 3 Extraction     1  Turn the power switch on  the system will process self testing   and then go to steady mode        Note     The system will block main functions before the completion  of self testing process      2  Open the sliding door and remove the sample rack from the  instrument              www ZINEXTS com Page 17 Ver  V1 00     3  Take out Reaction Cartridges  and all plastics disposables   Reaction Chamber  Tip Holder  Filtered Tip  Pestle  Sample  Tube  Elute Tube  from the Reagent Kit box  You need one  Reaction Cartridge and one set of plastic plastic disposables  per sample     www ZINEXTS com Page 18 Ver  V1 00     4  Place reagent cartridges        Important     Cartridge should be placed in a flat and stable table  when setting up the Reagent Cartridge     Set Cartridges in the order of the number from left to  right     Make sure that Cartridges are inserted in to the  Cartridge Tray tightly  You will hear a click if the  cartridge is placed correctly on the tray     You can load 1 12 cartridges on the tray depending on  the number of samples that you wish to process      5  Place Reaction Chamber s         6  Place Tip Holder s     www ZINEXTS com    Page 19 Ver  V1 00                     MXN S E C     um LS  FEKKSA memet  i    ne CS          9  Place Sample Tube and Elute Tube to Sample Rack on the  bench    www ZINEXTS com Page 20 Ver  V1 00        10  Load the sampl
9. e s  to Sample Tube       Note     Pretreatments are essential for some sample types before  loading to Sample Tube  Please refer to the handbook of  reagent kits for details     Make sure the caps of Elute Tube are open as the figure  shown above     www ZINEXTS com Page 21 Ver  V1 00     11    Place Sample Rack on the instrument platform             Note       Use two hands to handle the Sample Tray       Make sure the Sample Tray be placed correctly in the  instrument     12  Close the Sliding Door     13  Scan the protocol barcodes to select purification protocol   sample volume and elute volume    Note       There is one protocol barcode paper enclosed in the  reagent kit box        Protocol s name  sample volume and elution volume will  be shown on LCD screen after protocol barcode is  scanned      14  Follow the instructions displayed on LCD screen to double  check the operating steps being completed before program  running      15  Push    Enter    to confirm  Instrument will start to run the  protocol program automatically until whole processes are  completed     Note     www ZINEXTS com Page 22 Ver  V1 00      Ittakes from 30 to 45 minutes to complete the extraction  according to reagent types  Find the detailed time  consumption by checking the handbook of your reagent    Kits     2 4 After Extraction Operation   1  Take out the Elute Tubes  The extraction yields in them     Note     You can apply quality checking or do downstream study  or storage them as what you
10. ed in the handbook supplied with the    Perform preventive maintenance annually to ensure    Swap the air leaked Tip with new one    Too much starting Decrease the amount of starting material  Use the    Page 25    Ver  V 1 00    material causing recommended amount of starting material as listed in    clumps or aggregates the Reagent Kit manual  Handbook      DNA Quality Problems    Decrease the amount of starting material used   Be sure to add Proteinase K during lysis  if included in the  Incomplete lysis protocol   Make sure that the sample is completely immersed in the  Lysis Buffer       Be sure to process sample immediately after collection or  Poor quality of starting      store the sample at appropriate temperature  The yield and  materia  Low DNA yield the quality of DNA isolated depend on the starting material   During shipping  some magnetic bead solution may adhere  Insufficient amount of to the sealing foil of the cartridge  To collect any bead  magnetic beads added   solution from the foil  tap the cartridge to deposit the bead  solution at the bottom of the well   Ensure that the lysate does not contain any particulate  Clogged Tips resulting    material that can clog the tip sprout  If needed  centrifuge  in DNA loss ES  the sample prior to the MagPurix purification     otore cartridge containing the beads at room temperature     Do not freeze the cartridge as the beads may be irreparably  No DNA Magnetic beads stored  damaged     recovered or handled improper
11. ing the safe use of the  MagPurix 12 system  Please read this manual carefully before you start to  run the system at the first time  especially for Safety Information     If there is any question about how to install or operate it  please contact  our certified distributors   agents or email to our technical support center    support zinexts com       MagPurix Magnetic Bead Purification Process             www ZINEXTS com Page 3 Ver  V1 00    SAFFETY INFORMATION    The meanings of safety precaution marks are as follows    N WARNING      WARNING  indicates a dangerous condition that may lead to death  or serious injury      N BIOHAZARD      BIOHAZARD  indicates exposure to a biological agent or condition  that may be infectious or harmful      N CAUTION      CAUTION  indicates the possibility of danger that may lead to minor  or moderate personal injury      CAUTION  may also indicate an accident of property damage   Important      Important  shows the important notes for usage  as well as prohibited  actions     Note      Note  indicates the notes  procedures that should be obeyed and  Supplementary information for use     For your safety and that of others  follow the guidelines provided in the  following pages concerning the use of the MagPurix 12 system     www ZINEXTS com Page 4 Ver  V1 00    About Instrument     N WARNING     Ignoring the following notations may lead to fire or electric shock     In countries other than Taiwan  US and Canada  use a power cable  that mee
12. ly  Make sure that the beads are in solution at all times and do    not dry  Dried beads are non functional                 Remove any magnetic beads using a magnetic separator or  Magnetic beads present    Eluate centrifuge the sample in a microcentrifuge for 1 minute at  in the eluate  containing DNA maximum speed     is discolored DNA contaminated with   Minimize the amount of blood or blood stained sample used  heme  s 20 ul blood spot for forensics sample      To prevent bubble formation during mixing  make sure the       Bubbles formed during       gt  sample volume is at least the recommended volume listed  mixing steps  in the manual supplied with Reagent Kits     DNA is sheared Purified DNA    Aliquot purified DNA and store at 4  C  short term  or  20  C    or degraded repeatedly frozen and     long term   Avoid repeated freezing and thawing   thawed    DNA contaminated with   Maintain a sterile environment while working  i e  wear  DNases gloves and use DNase free reagents      www ZINEXTS com Page 26 Ver  V1 00       5 Specifications    The MagPurix 12 system    Was T 2777    Instrument Dimensions        CE  LV directive  73 23 EEC   EMC directive  89 336 EEC   EMC  EN61326 1  Safety  EN60101 1    www ZINEXTS com Page 27 Ver  V1 00    APPENDIX    B Replacement Optional Parts    Barcode Reader A01001  Power Cord A01002    Filttered Tip A02001  Sample Tube A02002  Elute Tube A02003       B Product Guarantees      The guaranty period shall be one year from the deliver
13. stem in its original shipping box or some other  equivalent box     www ZINEXTS com Page 29 Ver  V1 00    
14. t use the MagPurix 12 system in direct sunlight or close to a  heating device  It may shorten the life of the MagPurix 12 system  or  cause a trouble     Operating Conditions    ems   Conditions                Temperature    C  During operation  15 30 00       www ZINEXTS com Page 11 Ver  V1 00    Max  wet bulb During operation 29  non condensing   temperature    C  During down time 29  non condensing   Temperature gradient    C hr  12 or less  non condensing     Humidity gradient  RH day  30 or less  non condensing   Altitude  operating  m  1600 or less       1 4 Unpack the MagPurix 12 system     1  Open the packing box and take out the instrument and related  accessories     Important     The MagPurix 12 system has a weight of more than 45  kgs  It should be lift and moved by two persons     Hold the moving handler of the instrument from two sides  to move it out from the box           B Do not hold the plastic outer covering   B Do not hold the front panel     B Do not hold the door     AN Caution     Improper handling of the movement of the MagPurix 12  system will lead to instrument damages     www ZINEXTS com Page 12 Ver  V1 00     2  For correct and safe use of MagPurix 12s  install itin a  location that is close to electrical outlet and has enough space  for installation and operation     Important     Keep the shipping box  They will be needed again when  transporting the MagPurix 12 system      3  Connect Barcode Reader with the Instrument     EU If       www ZINEXT
15. ts your country   s standard  or contact your local  distributor     Use the AC adapter provided     Do not use the MagPurix 12 system with voltage other than the  voltage specified on the device     Do not overload the power outlet with too many devices     Do not use the MagPurix 12 system with a damaged power plug or  a loose socket     If there is dust on the prongs of the power plug or on the plug  socket  remove it with a dry cloth     When you disconnect the plug from the outlet  be sure to hold the  power plug itself  Do not pull the power cable     For maintenance  disconnect the power plug from the outlet   Do not touch the power plug when you hear the crash of thunder   Do not pour any liquid on the MagPurix 12 system     Do not place any objects containing liquid on the MagPurix 12  system  Doing so may cause a device failure  fire  or electric shock     B Inthe event the device overheats  starts to smoke or smells   strange  immediately unplug the power cable    N CAUTION    B Never attempt to remodel the MagPurix 12 system without the  manufacturer s permission  Doing so may cause fire or electric  shock    B Do not place or drop objects on the MagPurix 12 system  Also    refrain from bumping or knocking it  as doing so may cause a  failure or malfunction of the MagPurix 12 system     www ZINEXTS com Page 5 Ver  V1 00    If any liquid materials are left inside the device  wipe it up a soft  paper tissue  etc  Otherwise  the MagPurix 12 system may be  damaged    
16. y of the  product       Any part which fails to function properly under normal use   provided that all warnings and cautions in this manual are  observed  during the guarantee period will be repaired by the  ZINEXTS LIFE SCIENCE CORPORATION free of charge         f the problem is due to the causes listed below  some charge  may be applied for the repair even within the guarantee  period     B Any failures due to improper use or that are affected  by other devices except for those approved by the  ZINEXTS LIFE SCIENCE CORPORATION    B Any malfunctions or damages during transportation or  due to dropping of the product by a user     www ZINEXTS com Page 28 Ver  V1 00    B About Services    Before you ask for device repairs  refer to    4 Troubleshooting    to  check the problems  If the problems cannot be solved  contact your  local representative  or the relative customer support     B Notes for Transporting   Shipping the MagPurix 12  system    When transporting shipping the MagPurix 12 system to a new  location  perform the following tasks     Warning       Before carrying the MagPurix 12 system  disconnect the  plug from the outlet  If the power cable is damaged  this  may cause a device failure  fire  injury  or electric shock       When transporting shipping the MagPurix 12 system  be  sure to perform the following tasks     Remove all attached parts  Sample Rack  Barcode  Reader  from the MagPurix 12 system       After completing the above tasks  pack the MagPurix 12  sy
    