TROUBLESHOOTING Note: The information in this
Contents
1. PSs DPSS Lasers TROUBLESHOOTING Note The information in this section is intended to be used as an aid to the user in assessing problems with the operation of the laser and as a guide to assist experienced users in resolving routine problems If there is any question about the procedures contained in this section or associated hazards call DPSS Lasers Inc Customer Service Caution These procedures could result in exposure to hazardous laser radiation or bright light sources Procedures should be performed only by persons who are familiar with the hazards involved and with the use of appropriate safety equipment WARNING Do not remove any covers from the laser unless specified 1 Unauthorized removal of covers from the laser will void all warranties 2 There are no user serviceable components within the laser head or power supply 3 Inexperienced handling of the components in the laser can cause severe damage TROUBLE SHOOTING GUIDE PROBLEM POSSIBLE CAUSES SOLUTION Power On light not on Key switch on but no status LED s on Fault light on at first turn on DPSS User s Manual 3500 SMPS 780 023 Rev A Line cord not connected Main power switch at rear of power supply not on Cables not connected between power supply and laser head Cables J2 amp J3 interchanged at head or power supply Interlock open Cooling system not turned on System recently turned on and water water cooled units only has not s
2. is connected to REMOTE connector Verify that RF BNC cable is connected at power supply and laser head Verify that shutter is fully open Verify that ambient temperature is within acceptable operating limits Verify that Cooling system is maintaining set temperature Refer to Service Appendix for laser alignment procedure Allow laser to warm up for one hour prior to use Verify that ambient temperature remains within acceptable operating limits Verify that cables J2 and J3 are well connected at power supply and laser head Verify that Cooling system is maintaining set temperature Check optics in optical path for contamination or damage Turn of laser system and clean laser output window according to cleaning instruction in Service Appendix If necessary an aperture can be placed one foot away from the laser head to block this green emission Strong UV emission when q e Electronic malfunction e Contact DPSS Service switch is DISABLE Representative DPSS User s Manual 3500 SMPS 16 of 16 780 023 Rev A Created on 02 11 03 5 38 PM If the problem experienced cannot be resolved by this troubleshooting guide please contact a DPSS Lasers Inc Customer Service Representative at either your local distributor or directly through the DPSS Lasers Inc Customer Service Department DPSS User s Manual 3500 SMPS 17 of 17 Created on 02 11 03 5 38 PM 780 023 Rev A
3. m running with laser READY and q switch ENABLE Low output power after one hour warm up Power oscillation after turn on Power oscillation after several hours of operation Poor beam quality Scattered green emission with UV output Interlock open Cooling system water not flowing water cooled units only Laser head over temperature Power supply over temperature Shutter not open REMOTE connector not connected RF BNC cable not connected at power supply or laser head Beam path partially blocked Ambient temperature too high Cooling system not functioning properly Laser misaligned Damaged or contaminated optics internal to the laser head It is normal for the laser output power to oscillate slightly during the first hour of operation Ambient temperature changing dramatically Cables J2 or J3 not seated well Cooling system malfunction Contaminated or damaged optics external to the laser head Contamination or damage to laser output window It is normal for some green light to be emitted with the UV output e Check all user installed interlocks e Open Cooling system reservoir and verify bubbling action of water Verify that laser baseplate is at or near room temperature Verify that the power supply chassis is at or near room temperature Ensure that there is adequate ventilation for the Cooling system and the power supply Open shutter on laser head Verify that either JUMPER or alternate control
4. tarted flowing yet Flow switch interlock connector water cooled units only not plugged in at power supply or laser head Water hoses interchanged at Cooling system water cooled units only Cooling system pump malfunction water cooled units only Laser head over temperature Power supply over temperature 15 of 15 Connect line cord to power supply and suitable AC power source Turn on main power switch at read of power supply Connect all cables between power supply and laser head Verify that all cables are well seated and there are no bent pins in the connectors Verify that J2 amp J3 are correctly connected Check all user installed interlocks for proper operation Turn on Cooling system Wait 30 seconds and press STANDBY button to clear fault Open Cooling system reservoir water cooled units only and verify that water is bubbling Verify that the flow switch interlock cable red twisted pair is connected at power supply and laser head Reverse water hose connections at Cooling system water cooled units only Touch laser baseplate to verify that it is close to room temperature Touch power supply enclosure and verify that it is close to room temperature Ensure that ambient cooling air is Created on 02 11 03 5 38 PM not recirculating through power supply or Cooling System Fault light on after laser is operating No pulsed laser output after 5 minutes assumes no faults detected and syste
Download Pdf Manuals
Related Search
Related Contents
器具の取外し方法 自 ございます。 Kettler Rivo M Baumatic BTM25SS Convection Microwave Oven 取扱説明書(0.2MB) Origin Storage 1TB 2.5" 7.2k SATA Page 1 Page 2 金沢大学文学部論集 行動科学・ 哲学篇 第20号 2000 東京都船形学園 Operating instructions Programmable multifunctional transmitter SMT395 User Manual - Sundance Multiprocessor Technology Ltd. User Guide - Constructionline Copyright © All rights reserved.
Failed to retrieve file