Home
        Biomek NX Multichannel Quick-Start Guide - QB3
         Contents
1.              Method1  BiomekXMC  BiomekNX MC ETC  0 00 02                      Figure 2 3  Instrument Setup step with labware added    Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       2 6    Naming Labware    It is helpful to name  labware  particularly  when you use lots of  labware on the deck  You  can assign a name that  identifies the contents of  the labware or a  descriptive name that fits  the work being done in  the laboratory        Beckman Coulter  Inc     3     14     15                Learning to Create Methods for the Biomek   NX    From the Deck Layout Display  double click the reservoir on P4 to display  Labware Properties  Figure 2 4      tesy  Labware Type  Reservoir Maximum Yolume  110000 uL  Bar Code  E  Labware contains a  known x  volume   100000 a uL of liquid type   MERS      Sense the liquid level the first time a well with Unknown or Nominal volume is accessed    from the Liquid         Sense the liquid level every time a well is accessed  from the Liquid         Show Labware Volumes       OK   Cancel      Figure 2 4  Labware Properties for Reservoir on P4    In Name  enter Resv in Labware Properties   In Labware contains a  select Known from the pull down menu     In volume  enter 100000  This means that you know you have 100 000  microliters of liquid in the reservoir     In liquid type  select Water from the pull down menu  Labware Properties for  the Reservoir should look like Figure 2 4     Choose OK to close La
2.         Method1  BiomekNX MC  BiomekNX MC ETC  0 00 03    Figure 2 7  Transfer step inserted into the Method View    Beckman Coulter  Inc        ring to Create Methods for the Biomek   NX 2 9       2 4 2 Configuring Tip Handling    Tip Handling    Configuring Tip To configure tip handling     Handling is not always 1  In Tip Handling  make sure Load is checked  Figure 2 7    required because the          BONE ENANA ESA 2  Click on the AP96_200uL tip box on TL1 in the Current Deck Display to load    AP96_200uL tips as part of the Transfer step  This change is displayed in the    loads clean tips i   i  Load field of the Tip Handling configuration     automatically during a  Transfer step  Refer to  the Biomek Software  User    s Manual  Chapter  15 3 1  Configuring Tip    Handling  for   ae    information on other tip   Your tips are configured for your liquid transfer  so click the up arrow next to    handling options  such as Tip Handling to collapse the configuration  This allows more room for the step  Wash tips in and configuration  Figure 2 8   The text displayed reflects your tip handling    Change tips between configuration   destinations       Make sure unload them is selected in the next field     Check Change tips between sources       Uncheck Change tips between destinations     nH wn A U          Biomek   Software   Method1   New     File Edit Project Instrument Execution Options Help   De eB RS eM oe aly wo  B       Start Use pod   Pod1 v  for transfer     ARA  a
3.     23 2004 2 16 00 PM  Figure 2 17  Deck confirmation prompt for Quick Start method       Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide                  2 18 Learning to Create Methods for the Biomek   NX    INGA    5  Choose OK  The prompt closes and the method continues  View the method on  screen as it proceeds in the simulator  The method can also be followed in the  Method View as steps are highlighted in green as they are executed     Note  If you are running this method on your instrument instead of the  simulator  make sure that the physical deck layout matches the expected deck  layout shown in the deck confirmation prompt before you choose OK     You can proceed directly to Chapter 3  Using the Editors While Creating Methods  to  learn about the editors you   ll use for customizing your methods        Beckman Coulter  Inc           Using the Editors While  Creating Methods       3 1 Overview    In this chapter you will be given information and instructions for using the editors in  the Biomek Software  The editors allow you to customize a method to your special  requirements  You can go beyond the predefined project items and instrument  configurations  manually control the pod  restore instrument settings and project files   and configure external devices using specific editors  The editors are accessed from  the Project or Instrument menus located on the toolbar of the Biomek main editor   Figure 3 1      Read the information in thi
4.    COULTER    Biomek   NX Multichannel  Laboratory Automation  Workstation    Quick Start Guide    Beckman Coulter PN 987893  Revision AB    September 2004  B ule Coulter  Inc  4300 N  Harbor Boulevard  Fulle rton  CA sae 3100  Copyright    2004 Bec invited   ulter  Inc  Printed in U S A     Except as provided in writing signed by an officer to Beckman Coulter  Inc   this system and any related  documentation are provided    as is    without warranty of any kind  expressed or implied  including that the system is     error free     This information is presented in good faith  but Beckman Coulter  Inc  does not warrant  guarantee  or  make any representations regarding the use or the results of the use of this system and related documentation in terms  of correctness  accuracy  reliability  currentness  omissions  or otherwise  The entire risk as to the use  results  and  performance of this system and related documentation is assumed by the user     Except as expressly provided herein  Beckman Coulter  Inc  makes no other warranty  whether oral or written   expressed or implied  as to any matter whatsoever  including but not limited to those concerning merchantability and  fitness for a particular purpose  nor is freedom from any patent owned by Beckman Coulter  Inc  or by others to be  inferred     LIMITATIONS OF LIABILITY   Beckman Coulter  Inc  shall not be liable  to any extent whatsoever  for any damages resulting from or arising out of  the use or performance of this sy
5.    Overrides Technique  wa  amp        Destination  Dest1    JECFIat96 x  at  Dest x   100 uL of  Water x   JV Auto Select Customize         Technique   Standard    ve ay  amp                Stop when finished with destinations  Dispense up to 1 time per draw  Transfer Detai Is             collapsed                 Method1  BiomekNx MC BiomekNxX MC ETC  0 00 03          Figure 2 10  Transfer step with Destination configured    Note  The method view changes to reflect the amount transferred when the next step  is configured     The Transfer step configuration is now complete  Proceed to the next section to  configure a Move Labware step in your method     Beckman Coulter  Inc        fing to Create Methods for the Biomek   NX 2 13       2 5 Configuring a Move Labware Step          The Move Labware step uses the gripper to move Gripper  labware from one position to another on the deck or from    Th has t  one ALP to another  Move Labware can also be used to S EEPE Tas WS  mechanical fingers that  remove labware from the deck or to an external hardware df he vod  device without halting the Biomek instrument or the exten rom the po to  method  grip labware  Gripper  fingers grasp labware  Note  Refer to the Biomek Software User s Manual  along the long side   Chapter 15 4  Move Labware Step  Refer to the Biomek    NX Multichannel  To configure a Move Labware step to move your Dest1 Laboratory Automation  labware at position P5 to another deck position  Workstation User 5  1  Inser
6.   Figure 3 7  Deck Editor    2  Select New Deck  Choose a name for this deck appears  Figure 3 8      Choose a name for this deck       Please enter a name     em      Figure 3 8  Choose a name for this deck       3  Enter QuickStartDeck and choose OK  QuickStartDeck appears     4  Click and hold the mouse button on TrashRight from the ALP Types List   Locations capable of supporting that ALP are indicted in the Deck View by  dashed boxes     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide              3 10 Using the Editors While Creating Methods       5  Drag and drop TrashRight to any location capable of supporting it  Figure 3 9    A TrashRight ALP has been added to the deck  This deck is now the default deck  and is available in the Instrument Setup step     yo    x E fF   al   al Pp    NewDeck Delete Deck Rename Deck OpenDeck Clear Deck Renumber Delete ALP Properties Save                TrashRight ALP  added to the deck              Coords  cm    0 303  13 436  fis 935    A    Figure 3 9  QuickStartDeck with TrashRight added          6  Select Open Deck  Select a Deck appears  Figure 3 10      Select a Deck             Figure 3 10  Select a Deck    7  Choose Deck1 to make it the default deck since it is the deck you used when you  created the quick start method and choose OK     8  Choose Save to exit the Deck Editor     Beckman Coulter  Inc           Using the Editors While Creating Methods 3 11    3 6 Manually Controlling the Pod    If you
7.   HEE eR AS 3 9  3 6 Manually Controlling the Pod        nnannananaa nanana 3 11  3 7  Quick Start Guide Summary       ooon nannan enna 3 13    Beckman Coulter  Inc           Introducing the Biomek   NX  Laboratory Automation  Workstation       1 1 Introduction    Welcome to the Biomek   NX Laboratory Automation Workstation     This quick start guide is designed to introduce you to the Biomek NX Multichannel  instrument and Biomek Software operating system  This quick start guide includes       Getting to Know the Biomek   NX Laboratory Automation Workstation  this  chapter      an overview of the hardware and operating software        Learning to Create Methods for the Biomek   NX  Chapter 2      an  overview of how to create methods using the Biomek Software        e Using the Editors While Creating Methods  Chapter 3      an overview of  how to use the editors in Biomek Software to accommodate unique method  requirements        Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide          1 2       Beckman Coulter  Inc     Introducing the Biomek   NX Laboratory Autome    1 1 1 How to Use This Quick Start Guide    This quick start guide is intended to help you start using the Biomek NX instrument  and Biomek Software immediately  Read through this first chapter to acquaint  yourself with the system  proceed to Chapter 2 and complete the tutorial steps for  building a small method  then use Chapter 3 to complete the tutorial steps for using  the edito
8.   ror in the Dispense operation of the  Defa Template  pipetting template  Cannot move relative to liquid leveli K               Method1   Biomeknx Mc  BiomekNx MC Error trying to dispense to P5  coming from P4   Error in the Dispense operation of the    Default       Beckman Coulter  Inc     Figure 2 12  Quick Start method with error       2 7    Methods and  Project Files    Methods are saved as  elements of a project file   not as separate method  files  Saving a method  saves only the current    status of the method  It  does not create a revision  or check in project items  as revisions  Refer to the  Biomek Software Users  Manual  Chapter 6   Understanding and  Using Project Files              ling to Create Methods for the Biomek   NX       Saving a Method    After you   ve clicked on Finish step  you can save or  check in your method  You will just save it for now     1  Choose File gt Save from the toolbar  Save Method  AS appears     Save Method As    Enter method name     Cancel      Figure 2 13  Save Method As       2  Enter Quick Start Method and choose OK  The  method editor now looks like Figure 2 14        Biomek   Software   Quick Start Method  Development  x    File Edit Project Instrument Execution Options Help    PLAC ORS YBBoo gl wo  K B          Start    Instrument E Instrument Setup  Q Transfer 100 pL frc       Move Labware fror    od  lg Finish    Setup    Comment       Vv Clear current instrument setup of all labware  after the method completes    Vv 
9.  Nme o    fj AccuFrame    8 Podi Port   simulate bat         Devices  A DeviceControllerO Pod Type  span Pod x   g7 ShakerALPO  oF ShakerALP1  oF StirrerALPO  oF StirrerALP1  47  TiploaderO  67 Tiploader1     Digital Devices  B E Simulator  OP Camera  E Materials  f Stacker Carousels             BiomekNx Span       Figure 2 15  Hardware Setup    2  In Port  select Simulate     3  Choose Accept to save changes and close Hardware Setup  Biomek Software  is now running in simulation mode     Beckman Coulter  Inc           Learning to Create Methods for the Biomek   NX 2 17    4  Click on the green arrow button on the toolbar  The simulation display is  automatically launched and the step configuration area is replaced with a three   dimensional model of your Biomek NX  Figure 2 16   The deck confirmation  prompt  Figure 2 17  appears on top of the main editor        Biomek   Software   Quick Start Method  Development   File Edit Project Instrument Execution Options Help    Dee BRS BRB oo    8                   k    Instrument   Instrument Setup  a Transfer 100 L fre      Move Labware frorn  IB Finish          ft   2l   l  Quick Start Method BiomekNX MC  BiomekNx MC  ETC  0 00 00    Figure 2 16  Biomek NX simulator displayed             Biomek   Software    The left pod should have no tips loaded   Does the Biomek   Software deck match the above layout  including the labware and their locations     If yes  choose OK to continue the method   If no  choose Abort to stop the method 
10.  Tip and Labware Types         e Technique Browser     allows you to customize techniques for your  specific needs and save them for use in future methods  refer to the Biomek  Software Users Manual  Chapter 9  Understanding and Creating  Techniques     e Liquid Type Editor     allows you to create new liquid types or to modify    existing liquid types  refer to the Biomek Software User s Manual  Chapter  8  Understanding and Creating Liquid Types            e Pipetting Template Editor     allows you to create or modify pipetting  templates that configure the precise actions during pipetting operations   refer to the Biomek Software Users Manual  Chapter 10  Using the  Pipetting Template Editor         3 2 2 Instrument Menu    The editors on the Instrument menu  Figure 3 1  allow instrument configuration to  be changed  They include the following     e Hardware Setup     if you change the configuration of your instrument  it  is used to install  configure  and remove other devices  refer to the Biomek    NX Multichannel Laboratory Automation Workstation User s Manual   Chapter 3 1 1  Accessing Hardware Setup         e Deck Editor     used to define and change deck configurations  refer to the  Biomek Software User 5 Manual  Chapter 5  Preparing and Managing the  Deck            Note  Refer to Section 3 5  Creating a New Deck  in this quick start guide  to follow step by step instructions to create a new deck layout in the  software     e Device Editor     external devices may
11.  Volume field     Note  The volume is measured in microliters  uL  automatically and indicates  the amount of liquid that will be dispensed  100 uL  into each of the 96 wells  so   in this case  you   re dispensing a total of 9600 uL       Right click on the large tip illustration and choose Measure from Bottom       Set the dispense height in the large tip illustration to 1 00 mm from bottom     using the same technique as you used for setting the aspirate height     Note  If you accidentally open too many destination configurations  right click  on the title above the graphic of the plate in the configuration  Click on Delete  from the popup menu and the entire configuration goes away     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       2 12 Learning to Create Methods for the Biomek   N         The destination labware is now configured  and the main editor looks like Figure 2   10  The Transfer step is now configured to transfer 100 uL from the source reservoir  Resv to the destination microplate Dest1        Biomek   Software   Method1   New     Fie Edit Project Instrument Execution Options Help    PEALE ELI CETACEI                      Start      Instrument Setup  Q Transfer 100 uL frc    B Finish       Changes to  reflect the  amount  transferred  when the next  step is  configured                 Use pod   Pod1 w   for transfer     V Load AP96_200uL tips  change between sources  and unload them when finished        1 00 mm from bottom  
12.  about the possibilities  designed into the Biomek NX           In this chapter  you    ll learn       Creating a New Method  Section 2 2        e Configuring the Instrument Setup Step  Section 2 3          Configuring a Transfer Step  Section 2 4        e Configuring a Move Labware Step  Section 2 5        e Using the Finish Step  Section 2 6        e Saving a Method  Section 2 7   e Running a Method  Section 2 8        Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       2 2    2 2    If using Accounts  amp   Permissions       You must have Develop  Methods permission to  create methods  Refer to    the Biomek Software   User 5 Manual  Chapter 2   Using Accounts  amp   Permissions           Learning to Create Methods for the Biomek            Creating a New Method    In this chapter  you will build a simple method that  transfers liquid between labware  In building this  method  you will     e Populate the deck with labware using an  Instrument Setup step     e Transfer liquid using a Transfer step  This  requires that a Tip Loader ALP be installed on  the deck     e Move labware via the gripper using a Move  Labware step           Tip Loader ALP  The Tip Loader ALP  loads disposable tips  onto a 96 well head or a  384 well head mounted  on a Multichannel Pod   Refer to the ALPs User s  Manual  Chapter 23  Tip  Loader ALP        You launched Biomek Software in Chapter 1  you are now going to create a small    method     1  Drag and drop an Inst
13.  and  approved by an expressly authorized officer of the Company     Parts replaced during the warranty period are warranted to the end of the instrument  warranty     Note  Performance characteristics and specifications are only warranted when  Beckman Coulter replacement parts are used     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       Table of Contents    Warranty and Returned Goods Requirements              00  cece cece eens iii    Table of Contents             00 cee eee ee ee eee v    1 introducing the Biomek   NX Laboratory Automation Workstation    1 1 TOU NOHO i  2 3 6 605 aaue eera ened aaa a a aes 1 1  1 1 1 How to Use This Quick Start Guide    0 0    0 0    ccs 1 2  1 2  Bi    mek   INK pc ee cca ae aos KAA RE RA eta RSA ARAM EEEE EREEREER EEA 1 3  1 2 1 Automated Labware Positioners  ALPs               0 000 cee eeae 1 4  1 2 2 Light Curtain and Other Protective Barriers                  004  1 5  1 3 Biome  k SOUWAC 24 ccvesadenctusedenn Ce ei oies i ead hua taneaa eee eden 1 6  1 4 Starting the  Biomek   NX win ee sadeea seen aba eked peed desu ead eee ns 1 7  1 4 1 Turning On the Biomek NX  i dacecaeiadaunad pea daundGes beens 1 7  1 4 2 Launching the Software 2444 4 Ses Ss  aa wean dh heed dn beees 1 7  1 4 3 Homing All Axes of the Biomek NX Pod                  00 000 1 8  2 Learning to Create Methods for the Biomek   NX  2 1 OVETVIEW aitia RiGee hha ag Geeks eRe a a e ae e 2 1  2 2 Creating a New Method        00  cnc ce
14.  be integrated to your Biomek NX  In  order to use the integrated device  it must be configured through the Device  Editor  refer to the Biomek Software User s Manual  Chapter 22 4 1   Configuring Devices Using the Device Editor         e Manual Control     allows you to control the movement of the bridge  head   and gripper independently of a method  It is used to control the pod when  framing the deck or gripper and when recovering from errors  refer to the    Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide          3 4 Using the Editors While Creating Methods    Biomek   NX Multichannel Laboratory Automation Workstation User 5  Manual  Chapter 5  Manually Controlling the Biomek   NX in Biomek  Software      Note  Refer to Section 3 6  Manually Controlling the Pod  in this quick   start guide to learn how to manually move a pod              Import Export Utility     allows you to import and export instrument settings   decks  framing tools  and pod settings  as import files   imp   refer to the  Biomek Software User s Manual  Chapter 3  Using Instrument Files and  Settings         Beckman Coulter  Inc        Using the Editors While Creating Methods 3 5       3 3 Creating New Labware in the Labware Type Editor    The Labware Type Editor is used to create labware definitions so you can use  labware that is not predefined in Biomek Software     Labware  Definitions      Assume that you have new labware that is very similar to a BCFlat96  but some  
15.  labware  during operation    A Beckman Coulter  Service Engineer frames  the instrument during  installation  but it may be  necessary to repeat the  framing procedure if  ALPs or devices are  added  moved  or  removed from the deck   or if the head on the pod  is changed  Refer to the  Biomek   NX  Multichannel Laboratory  Automation Workstation  User 5 Manual   Appendix 4 Framing          Using the Editors While Creating Methods 3 9    Creating a New Deck    A deck in Biomek Software refers to the software representation of the physical  Biomek NX deck  The Deck Editor is used to define and change deck configurations   and stores deck information used for multiple methods  A deck also stores framing  information and device associations for each position on the deck     Note  Refer to the Biomek Software User s Manual  Chapter 5  Preparing and  Managing the Deck        Assume you have added a Trash ALP to your instrument deck and now must create  another deck in the Deck Editor that reflects that addition to the physical deck     To create a new deck in the Deck Editor   1  Choose Instrument gt Deck Editor  Deck Editor appears  Figure 3 7         x   t E    my       NewDeck Delete Deck Rename Deck OpenDeck ClearDeck   Renumber Delete ALP Properties Save Cancel                BarcodeReaderRight  FourByThree  MagBeadALP  OneByFive  OneByOne  OneByThree  OrbitalShakerALP  PositivePosition amp LP                         Coords  cm   fo s03  13 436 fis 935                       
16.  need to manually control the Biomek NX instrument outside of a method   including moving the bridge  head  and gripper  you    II use the Manual Control editor   Manual Control is also used to control the pod when framing the deck or gripper and  when recovering from errors  Refer to the Biomek   NX Multichannel Laboratory  Automation Workstation User 5 Manual  Chapter 5  Manually Controlling the  Biomek   NX in Biomek Software for additional information           Note  Manual Control is used also for controlling individual ALPs and Device  Controller outside a method  refer to the ALPs User 5 Manual         Using the following instructions  you   ll use Manual Control to move the pod to a  different position on the deck  You   ll move the pod to the Z max height first to make  sure you don   t hit anything on the deck and then you   ll move the pod to a position     To move the pod using Manual Control     1  Choose Instrument gt Manual Control  Manual Control appears  Figure 3   11   Here is where you choose the position to which you want to move the pod     s  a tt         Advanced Controls Home All Axes Get Version Stop Exit          Click on a position to move the pod to it                 Figure 3 11  Manual Control    Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       3 12 Using the Editors While Creating Methods       2  Select Advanced Controls and highlight Pod1  Advanced Manual Control   Pod1 appears  Figure 3 12      Advanced Manua
17.  round  flat bottomed wells    Preview    The shape and general description of the There are no errors in this labware type  SHSSSSSSSESE   labware  definition        Figure 3 5  NewLabwareQuickStart    Note  Each piece of labware in Labware Types has its own mini editor similar  to Figure 3 5 to allow specifications to be revised and saved     5  In Height  enter 1 46 and choose Save  You have created a new piece of  labware in the Labware Type Editor     6  Choose Exit to close the Labware Type Editor     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       3 8    3 4    Revisions    Changes made to any  item in a project file are  saved but do not become  revisions until checked  in  Once checked in  a  new revision is created  for the project items   Revisions can never be  lost  If a project item is  later modified or deleted   any revision of the  project item may be  restored     Validated Methods    When Accounts  amp   Permissions is enabled   methods may be  validated  A validated  method is a revision of a  method that is checked  in  approved with an  electronic signature  and  protected from further  modification  Revisions  of project items required  to run the validated  method are also checked  in and protected from  further modification   This ensures that  validated method runs are  reproducible  Refer to the  Biomek Software User    s  Manual  Chapter 12 15   Checking Out a Method           Beckman Coulter  Inc     Using the Ed
18.  tell the software what labware and ALPs are on the  Avoidance deck for obstacle avoidance  see sidebar   You   ll learn more about the Deck Editor in  When building a method    the next chapter  refer to Section 3 5  Creating a New Deck    you must ALWAYS use  the Deck Editor and       Instrument Setup step  to tell the software what    labware and ALPs are on  the deck and the position  each occupies  If this is  not done  the pod could  collide into these objects  as it moves around the  deck during a method  run           Biomek   Software   Method1   New     File Edit Project Instrument Execution Options Help   D ee BRS seeBlor alr no  x B    Start Deck   Deckt x  JV Pause to confirm setup     Pause for bar code input   Serena   Instrument Setup Labware Category    lt Any gt  7  IV verify Pod Setup Configure              7  BGS    Setup  a AB84M Ap384 3C AP96_20C AP96_20C AP36_20  AP96_20  BCDeep  BCDeep9 BCFIatIE BCFulRe    OSeBEa80    z CostarCe   CostaFl  CostarFl  i il Greineri  Greinerit  p Collar_36r CollarSpa CostarDer Drainable FilterHole il iner                          ET  BCSPECol       Comment                          Method1  BiomekWXMC  BiomekNX MC ETC  0 00 02    Figure 2 2  Instrument Setup step in the Method View    Beckman Coulter  Inc            Learning to Create Methods for the Biomek   NX 2 5        To configure the Instrument Setup step     position  TL       1  In Deck  make sure the default deck layout  Deck1  is selected  The Instrument  Setup s
19. 2 1  Automated Labware  Positioners  ALPs    and for integrating additional  devices  such as detectors and washers  A gripper is  installed on the Multichannel Pod for grasping and  moving labware           Note  Refer to the Biomek   NX Multichannel  Laboratory Automation Workstation User s Manual   Chapter 2  Multichannel Pod  for detailed information on  the Multichannel Pod and interchangeable heads     bee  1 tt    Pein     i    Zz          Multichannel Pod and he  with extended gr    A       KZ    Lf    FANS ASS          The deck is the work surface of  the Biomek NX instrument and  positions ALPs to be accessed  by the Multichannel Pod     CES    1 3           Multichannel Pod          The Multichannel Pod is  a full microplate  replication tool  incorporating a gripper  and interchangeable  heads to accommodate a  variety of functions              Heads          Interchangeable heads  aspirate and dispense  liquid using disposable  tips  The tip  compatibility and  maximum aspirate and  dispense volumes vary  for each head                    Power switch             Light curtain                Figure 1 1  Biomek NX with Multichannel Pod    Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide          1 4 Introducing the Biomek   NX Laboratory Automati    1 2 1   Automated Labware Positioners  ALPs     Automated Labware Positioners  ALPs  are interchangeable structures that are  installed on the Biomek NX deck in suitable positions to accommod
20. Clear current device setup of all labware  after the method completes    v Unload disposable tips from all pods  after the method completes    JV Clear all global variables after the method completes       No Reporting    2 15          Check In  Checking in a method is  similar to saving a  method  but in addition  creates a new revision of  the method  checks in all  project items of the  project file  and creates  new revisions for any  items that have changed   You will check in the  method later in Section  3 4  Checking In a  Method and Project  File                           Quick Start Method  BiomekNXMC   BiomekNX MC ETC  0 00 42         Figure 2 14  Quick Start method saved and named       Move on to Section 2 8  Running a Method  to run your saved Quick Start method     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide             2 16 Learning to Create Methods for the Biomek          2 8 Running a Method    Next you will run the method you just created using the simulator  The simulator  allows you to watch the method on your screen without actually running the method    on an instrument     To run your saved method     1  If you do not see the Simulation Display button on the toolbar  Figure 2 15    from the Instrument menu  choose Hardware Setup  Biomek Hardware    Setup appears     Biomek   Hardware Setup p      Reconnect ft Home All Axes     Add Device s   Remove Device   Y Accept  gt  lt  Cancel    EE F Biomek   NX  SN  None  Serial
21. To retain the dimensions of predefined labware in Biomek Software  it is  important to copy and then edit the copy of the labware type rather than simply  editing the predefined labware to meet the new specifications        Copy Labware Type x     Please enter a name                 Figure 3 3  Copy Labware Type    3  Enter NewLabwareQuickStart and choose OK  The copy of BCFlat96 is  created with the specified name and appears in Labware Type Editor  Figure 3   4      Labware Types    ne     b     x     Filter   lt any gt  xl    New Copy Delete Rename Edit Exit    Greiner1536ConeF Greineri536FlatPE Greiner384ConePF Greiner384ConePP Greiner384FlatPP Greiner384  Lid    Greiner384Therma Greiner36ConeHalf GreinerS6ConePP GreinerS6ConePS Greinerg6Flat    Greiner6Roun 7 a rah FERA FE oon GreinerS6RoundPP GreinerS6RoundPS Greiner36Thermal      Copy of BCFlat 96 renamed to    NewLabwareQuickStart     atse45qu  Greinershallows34       IMReservoir96          sessseses  SOSeoeeoese    Figure 3 4  NewLabwareQuickStart in Labware Type Editor    Beckman Coulter  Inc        Using the Editors While Creating Methods 3 7       4  Double click the NewLabwareQuickStart graphic  NewLabwareQuickStart  appears  Figure 3 5      NewLabwareQuickStart    Basic Information x Y    Miscellaneous Span fi 2 7762 fe  5471 cm    Movement Information    Orbital Shaker Height fi 4224 em  Ordering Information    Stacking Colors Edit       Wells_1    Bitmap FO es    Description  Beckman Coulter microplate with 96
22. When creating a new specifications are slightly different  You must create a new labware type in the  labware type  it is Labware Type Editor to accommodate the difference  Refer to the Biomek Software  important that the User 5 Manual  Chapter 7  Creating and Modifying Tip and Labware Types    labware dimensions are TESRENASNEN E  entered precisely per the Note  Do not run a method outside of the simulator using the labware you create in  americans this tutorial  the specifications may not be accurate for the physical piece of labware  you use in this method  see sidebar         specifications  An  incorrect labware  definition could result in  a number of problems  1  From the Project menu  choose Labware Type Editor  Labware Types  including collisions  appears  Figure 3 2     gripper errors  or poor   pipetting performance     To create new labware in the Labware Type Editor        Labware Types       ap   Filter    New Copy Delete Hename Edit Exit       AB334WellReaction  4P384 30uL    e  eeeee tee  LULIT TT TTT    BCUpsideDownTil       CostarCone36Rour    Collar 36mm CollarSpacer_23mm    Pm    CostarDeep96Squat CostarFlat384Squz CostarFlat384Squ  GranableRefilabler FilterHolder    EE i     Figure 3 2  Labware Types    AP9I6_20uL_Barier       Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide          3 6    Using the Editors While Creating Methods          2  Click on BCFlat96 and choose Copy  Copy Labware Type appears  Figure 3   3      Note  
23. ate  Dest1      Note  For information on the extensive selections available for configuring the  Transfer step  refer to the Biomek Software User    s Manual  Chapter 15 3   Configuring Transfer and Combine Steps        2 4 1    To insert a Transfer step into your method     Inserting a Transfer Step    Drag and drop a Transfer step from the Step Palette into the Method View just  below the Instrument Setup step  Notice that the Current Deck Display shows  the current position of all labware on the deck  Figure 2 7      You will complete three sections to configure the Transfer step for this method     e Section 2 4 2  Configuring Tip Handling       e Section 2 4 3  Configuring Source Labware       e Section 2 4 4  Configuring Destination Labware          Biomek   Software   Method1   New   File Edit Project Instrument Execution Options Help    PLACII CEIC                xel       Instrument      Start    Instrument Setup    Use pod  Pod1 x  for transfer         Tip Handling    Setup    JV Load  AP96_200uL    tips and   unload them z  when the transfer is done   J Wash tips in p ater z   p Ads of 7110    J Change tips between sources     S  oz  i Finish    IV Change tips between destinations        Click here to add a source           Stop when finished with  Destinations x   Dispense up to f1   time per draw      Aspirate at most fo pL per transfer for repeated dispensmg       Transfer Details    Advanced         Comment             Current Deck  Display                     
24. ate a wide variety  of labware  ALPs are either active or passive     e Passive ALPs hold labware in place or act as receptacles for method by   products or consumables  Figure 1 2      e Active ALPS contain mechanisms that may use power and or air sources for  various mechanical operations  such as the Orbital Shaker and Stirring ALPs   refer to the ALPs User s Manual            Figure 1 2  Example of a passive ALP    Note  Refer to the ALPs User s Manual for detailed information on the ALPs on  your system        Beckman Coulter  Inc           nek   NX Laboratory Automation Workstation 1 5    1 2 2 Light Curtain and Other Protective Barriers    WARNING  To reduce the risk of personal injury  safety shields and light  curtain must be in place before operating  Safety shields and light  curtain prevent entry into the work area during machine movement     WARNING  Dark non reflective material affects the sensitivity of the light  curtain and adversely impact its effectiveness  Typical lab dress  such  as lab coats and latex gloves  do not degrade light curtain operation   however  it is advisable to test the impact of all lab attire on light curtain  sensitivity before operating the Biomek NX      gt   amp     The Biomek NX is equipped with protective barriers   e Front light curtain  e Side protective shields  e Overhead canopy    When an object larger than approximately 1  in diameter penetrates this protective  zone  the instrument shuts down immediately  stopping all pod a
25. bware Properties     Double click the labware on P5 and name it Dest1 in Labware Properties       In Labware contains a  select Known     In volume  enter 50       In liquid type  select Water from the pull down menu  Labware Properties for    the BCFlat96 should look like Figure 2 5       Choose OK to close Labware Properties     Labware Properties g  Name   Pest1 Labware Type   BCFlat96 Maximum Volume  362 76 uL    Bar Code     Labware contains a  known x  volume   50 a uL of liquid type   Water 7         Sense the liquid level the first time a well with Unknown or Nominal volume is accessed  from the Liquid         Sense the liquid level every time a well is accessed  from the Liquid         Show Labware Volumes       OK   Cancel      Figure 2 5  Labware Properties for Dest1    Confirm that Pause to confirm setup  is checked  This will result in a  confirmation prompt being displayed later when you run your method in Section  2 8  Running a Method        Leave the Verify Pod Setup option as is  This setup is used to inform the    software of the current state of the pod  for example  if the pod already has tips  loaded        arning to Create Methods for the Biomek   NX 2 7       Your Instrument Setup step is now configured and the main editor should look like  Figure 2 6     Note  To learn more about the Instrument Setup step  refer to the Biomek Software  User s Manual  Chapter 15 2  Instrument Setup Step           Biomek   Software   Method1   New  i  Fie Edit Project I
26. ca eens aces eendenensenneneeues 2 2  2 3 Configuring the Instrument Setup Step               00 0  2 4  24 Configuring a Transfer Step  424 scas4 nes as ba wet heursi esd duarsaarweas 2 8  2 4 1 Inserting a Transfer Step         docs dee eeaitenaneeaidenaianked 2 8  2 4 2 Configuring Tip Banding  y 0 00ve14  sau 64 ananuna 2 9  2 4 3 Configuring Source Labware                00 cece cece eens 2 10  2 4 4 Configuring Destination Labware             2 0 0 0 000 cece eee 2 11  2 5 Configuring a Move Labware Step           0 0  csc e cece ence ee esenees 2 13  2 6    Using the Finish Stopes    lt scs lt die ded dee ad AS Abee oe dee eRe ee dd eee de de 2 14  2 6 1 Handling Configuration Errors       0 0 0    cece cece ee eee enee 2 14  2 7 Saving    Method crcr cache a aeee nae egeednn UREE ER EDERE REEERE EA 2 15  2 8  R  nnngaMethod srecrsceirrisrerer Rees edna Ekr G re EEE AEE Re 2 16    Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       3 Using the Editors While Creating Methods    3 1 OVETVIEW errad peder iara er ea eN Eea REA A Aer h 3 1  3 2 General Description of Editors      2 0 0 0    ccc cette nee 3 2   3 2 1 Project  Mees srererg Bakers An aR ES eh E EES 3 3   3 2 2 Instrument Men    peser eaweneeundawt eos eet ek eee beaten eed eae 3 3  3 3 Creating New Labware in the Labware Type Editor                        3 5  3 4 Checking In a Method and Project File     2 0 0 2    eee 3 8  3 5 Creating a New Deck        nnnunnnnn nnne con CaO  
27. ditor showing steps  available for insertion in  a method  They are  located on the left of the  main editor  Use the  Step Palette Builder to  add other step palettes   Refer to the Biomek  Software User    s Manual   Chapter 29 5  Using the  Step Palette Builder for  more information                      jiomek   NX Laboratory Automation Workstation 1 7    1 4 Starting the Biomek   NX    To start the Biomek NX  you will follow the instructions in this chapter to turn on the    Accounts  amp  instrument  launch Biomek Software  and home all axes     Permissions  When enabled  Accounts   4 44 Turning On the Biomek NX     amp  Permissions is an  integrated set of features   To turn on the Biomek NX instrument   built into the software to  assist users in complying  with electronic signature 2  Flip the switch to the On position  The indicator light comes on in solid green   requirements for closed   systems  The   functionality for 1 4 2  Accounts  amp  Permissions  is included in the Biomek  Software installation and    provides control over 2  If Accounts  amp  Permissions is enabled  see sidebar   a dialogue box appears   who can create and run requesting you to log in using the user name and password assigned to you by  methods  and who may your system administrator  Enter the required information  Your account must  only run validated have Develop Methods  Develop Projects  and Setup Instrument   methods  Refer to the permissions to complete all parts of this quick start 
28. e left mouse button and move the hand up and down until the  Auto Select is chosen in depth is as close to 1 00 mm from bottom as you can get   a pipetting step  such as  Transfer  each time a  method is created   Biomek Software OR  automatically selects the  technique most  appropriate for the This will be your aspirate height for this transfer   pipetting operation     4  Right click on the large tip illustration next to the microplate graphic in the  configuration and choose Measure from the Bottom     6  To adjust the height precisely to 1 00 mm  click on the tip and use the up or down  arrow keys on your keyboard while holding the Alt key     Right click on the graphic  then select the height from the menu that appears        The source labware configuration is complete and the main editor now looks like  Figure 2 9  Yov   ll now configure the destination labware to complete the transfer        Biomek   Software   Method1   New   File Edit Project Instrument Execution Options Help     D  s ORS Leo een nxe     Start Use pod  Pod1  gt   for transfer     ea   Instrument Setup Y Load AP96_200uL tips  change between sources  and unload them when finished              Setup    Q Transfer Source  Resv    ig Finish E  Reservoir x  at  Resv x   Using liquid type  water x   IV Auto Select Customize         1 00 mm from botto  Overrides Techni          Stop when finished with  Destinations x  Advanced             Dispense up to  1   time per draw   C Aspirate at most jo uL per tra
29. eckman Coulter  Inc  policies governing returned goods apply to this  product  Subject to the exceptions and upon the conditions stated below  the Company  warrants that the products sold under this sales agreement shall be free from defects in  workmanship and materials for one year after delivery of the products to the original  Purchaser by the Company  and if any such product should prove to be defective  within such one year period  the Company agrees  at its option  either  1  to correct by  repair or at the Company   s election by replacement  any such defective product  provided that investigation and factory inspection discloses that such defect  developed under normal and proper use  or  2  to refund the purchase price  The  exceptions and conditions mentioned above are as follows     a  Components or accessories manufactured by the Company which by their  nature are not intended to and will not function for one year are warranted  only to reasonable service for a reasonable time  What constitutes a  reasonable time and a reasonable service shall be determined solely by the  Company  A complete list of such components and accessories is maintained  at the factory     b  The Company makes no warranty with respect to components or accessories  not manufactured by it  In the event of defect in any such component or  accessory  the Company will give reasonable assistance to Purchaser in  obtaining from the manufacturer   s own warranty     c  Any product claimed to be d
30. efective must  if required by the Company  be  returned to the factory  transportation charges prepaid  and will be returned  to Purchaser with transportation charges collect unless the product is found  to be defective  in which case the product must be properly decontaminated  of any chemical  biological  or radioactive hazardous material     d  The Company shall be released from all obligations under all warranties   either expressed or implied  if any product covered hereby is repaired or  modified by persons other than its own authorized service personnel  unless  such repair by others is made with the written consent of the Company     e  Ifthe product is a reagent or the like  it is warranted only to conform to the  quantity and content and for the period  but not in excess of one year  stated  on the label at the time of delivery     It is expressly agreed that the above warranty shall be in lieu of all warranties of  fitness and of the warranty of merchantability  and that the company shall have no  liability for special or consequential damages of any kind or from any cause  whatsoever arising out of the manufacture  use  sale  handling  repair  maintenance  or  replacement of any of the products sold under the sales agreement     Representatives and warranties made by any person  including dealers and  representatives of the Company  which are consistent or in conflict with the terms of  this warranty  shall not be binding upon the Company unless reduced in writing
31. er  Inc             Learning to Create Methods for the Biomek   NX 2 3    ZF    As you can see  the following two steps appear automatically when you drop the first  step into the Method View        e Start     the first step in a method  allows Variables  Start creation of global variables that are applicable   A variable is a value that  to the entire method  Refer to the Biomek  Software Users Manual  Chapter 12 2 1   Configuring the Start Step     has been assigned a  name that may be  referenced repeatedly          a e Finish     the last step in a method  provides during a method  Refer  Finish options to clear the Biomek NX deck  clear the   to the Biomek Software  devices from the deck  unload tips from the pod    User s Manual  Chapter  and clear all global variables  Refer to the 13  Using Variables and  Biomek Software User s Manual  Chapter Expressions in a  12 2 2  Configuring the Finish Step  Method                 Proceed to Section 2 3  Configuring the Instrument Setup  Step to configure the Instrument Setup step you just inserted        Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide           2 4 Learning to Create Methods for the Biomek   N        2 3 Configuring the Instrument Setup Step    CAUTION  An inaccurate Instrument Setup may result in pod and  A N labware collisions  or in inappropriate pipetting     When creating Biomek methods  it is important to use the Instrument Setup step   Obstacle along with the Deck Editor  to
32. guide  The Biomek Software  Biomek Software User 5 main editor appears  Figure 1 3     Manual  Chapter 2    Using Accounts  amp  Note  Ifyou are logging into Biomek Software for the first time  you must enter  Permissions  a new password  Remember your new password  you will need to enter it each  time you access Biomek Software     1  Locate the power switch on the right side of the instrument  Figure 1 1      Launching the Software    To launch Biomek Software     1  Choose Start gt All Programs gt Beckman Coulter gt Biomek Software           Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       1 8 Introducing the Biomek   NX Laboratory A       Biomek   Software    Fie Edit  Project   Instrument  Execution Options Help  Ag ji  pel     Decaan ee                        K          Instrument  Setup    Other editors are available from the  Project and Instrument menus     Step Configuration Area  When a step is highlighted in the               Transfer         Combine             Method View  The method is built  step by step in this  area              Comment       I       Basic Step  Palette   Displays the steps  available for  insertion into a  method  More steps  are available  through additional  step palettes  accessed from the       Method View  the corresponding  configuration appears in this area                 Current Deck Display  Changes to display the status of the deck  upon completion of the previous step                 Options men
33. itors While Creating Methods           Checking In a Method and Project File    Checking in a method is similar to saving a method  but also creates a revision record  of the method  checks in all project items of the project file  and creates new revisions  for any items that have changed     You will check in your quick start method  along with the new labware you just  created     To check in your method with the new labware     1  From the File menu  choose Check In  Check In appears with the new labware  type     Select the project items to check in   Project Item Modified Time    E M g Labware Classes    Change       La NewLabwareQuick        E  New    mes      Figure 3 6  Check In    2 3 2004 2 47 00 PM       2  Choose OK to check in the new revisions of project items     Notice that the method name in the Biomek main editor title bar changes to Quick   Start Method  Revision 1  Development  If Accounts  amp  Permissions is enabled   other dialogues appear and more steps are required for checking in methods  Enter a  reason and your password to check in the method and project   refer to the Biomek  Software User s Manual  Chapter 12 10  Checking In a Method         Note  Ifusing Accounts  amp  Permissions  methods may also be validated by choosing  Validate from the File menu  see sidebar         3 5    Framing    Framing teaches Biomek  Software the location of  ALPs and devices on the  deck  The software uses  this framing information  to move the pod and  manipulate
34. l Control  Pod1    Multichannel   gt  Vector Buld  e  gt     Home D  z      Back    Home Z  xY  Extend Gripper Left 4m   vam   gt  Right      Retract Gripper       Move 2 Max D Units  sie  lt i gt    cm    Auchan Maes Squeeze Aspirate   UnSgueeze Dispense CT pl                   M Current Position       36 895 cm Y  25 788 cm Z  22 cm D  0 2 cm     Movement Vector    x fo cm Y fo cm Z jo cm D jo cm  M AutoClear   Clear   Speed 100   o   Stop                     Figure 3 12  Advanced Manual Control for a Multichannel Pod         Select Move Z Max  The pod moves to the highest configured height     Select Close  Advanced Manual Control  Pod1 closes       From Manual Control  click on P5  The pod moves to position P5 on the deck     NH Wn A U      You can click on additional positions to move the pod to that position  When you  are finished moving the pod with Manual Control  choose Exit to close Manual  Control     7  From the toolbar  chose Close to close the quick start method   OR    From the File menu  choose Close  The method closes and you are now ready to  create a method of your own     Beckman Coulter  Inc        Using the Editors While Creating Methods 3 13       3 7 Quick Start Guide Summary    Congratulations  You   ve just completed this quick start guide   You   ve completed    e a small method   e steps using the editors   e steps creating some new project items   e steps to manually control the Biomek NX    If you   ve used the links to other manuals  you   ve a
35. lso learned where to find hardware  information on the Biomek NX  ALPs  and more information on Biomek Software     Enjoy your new Biomek NX Multichannel Laboratory Automation Workstation     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       
36. mp  Instrument Setup    Load AP96_200uL tips  change between sources  and unload them when finished                          Setup     a Transfer Click here to add a source   lg Finish Use this arrow  to collapse  and expand  Tip Handling  view                       Stop when finished with  Destinations x       Dispense up to  1   time per draw      Aspirate at most fo uL per transfer for repeated dispensing        Transfer Details       Comment                 Method1  BiomekWXMC BiomekNx MC ETC  0 00 03                Figure 2 8  Transfer step with Tip Handling collapsed    Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       Learning to Create Methods for the Biomek   NX       2 4 3 Configuring Source Labware    To configure the labware named Resv as source labware in the Transfer step   1  Click on Click here to add a source     2  Click on Resv labware sitting on P4 in the Current Deck Display     Note  The information you supplied during Instrument Setup is displayed in  the source labware configuration       3  Leave the Auto Select checkbox selected to use the technique chosen by  Techniques Biomek Soft  iomek Software   Techniques allow  pipetting settings to be  saved globally and used  in any method  Biomek 5  To set the aspirate height for the depth the tip descends into the reservoir  place  Software includes several the mouse cursor over the tip illustration  When the cursor turns into a  predefined techniques  If hand  hold down th
37. nd head operations   Some ALP operations  such as shaking or stirring  continue  refer to the ALPs User 5  Manual   When the instrument is sitting idle or in the paused mode  no violations are  registered when the protective zone is penetrated     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide          1 6 Introducing the Biomek   NX Laboratory Automati    1 3 Biomek Software    Biomek Software is designed to  Method          e Do a substantial amount of the method building    work for you     Allow you to take as much direct and precise  control over the method building process as you  want        A method is a series of  steps controlling the  operations of your  Biomek NX           The main editor that appears when the software is launched  Figure 1 3  is used to  build and configure methods        Beckman Coulter  Inc     e Each step is dragged and dropped from a Step    Palette into the Method View  refer to Chapter  2  Learning to Create Methods for the Biomek         NX      When a step in the Method View is highlighted   the corresponding configuration options appear  in the Step Configuration Area  refer to Chapter  2  Learning to Create Methods for the Biomek         NX      The method is customized to the desired level of  precise control via the editors accessed from the  Project and Instrument menus  refer to  Chapter 3  Using the Editors While Creating  Methods            Step Palettes  Step Palettes are panes  in the Biomek main  e
38. nsfer for repeated dispensing     Y Transfer Details                    Method1  BiomekXMC BiomekNx MC ETC  0 00 03                Figure 2 9  Transfer step with source labware configured    Beckman Coulter  Inc        Transfer Details    Transfer details include a  stop condition for  transfer operations  a  repeat pipetting  configuration  and a    maximum tolerance in  timing between transfer  operations  Refer to the  Biomek Software User   s  Manual  Chapter 15 3 4   Configuring Transfer  Details           2 4 4       Learning to Create Methods for the Biomek   NX 2 11        Configuring Destination Labware    To configure the labware named Dest    as destination labware in the Transfer step     1     Before you configure the destination labware  click the down arrow in Transfer  Details located just above the Current Deck Display to collapse that section  You  will not be making any changes to the options in this section       Click on Dest  sitting on P5 in the Current Deck Display  This one operation    accomplishes the same task as steps 1 and 2 of Section 1 5 2 Configuring Source  Labware         Note  Notice that the source labware configuration is collapsed and replaced  with a text summary of the setup  If you want to reopen the source configuration   click anywhere in the collapsed area       The Volume field is highlighted  which allows you to designate the amount of    liquid to be dispensed  For this tutorial  you    Il transfer 100 uL  so  type 100 into  the
39. nstrument Execution Options Help    PEIC CCORI CEIC     amp    Start Deck  Joeckt 7  JV Pause to confirm setup     Pause for bar code input   ie   Instrument Setup Labware Category    lt Any  gt  IV Verify Pod Setup Configure         B Finish                                                                    BCSPECol    Collar 36r CollarSpa  CostarCe CostarDe Costal CostarFl  Drainable FilterHolc Greinerl  Greiner1t                   Method1   BiomekNx MC  BiomekNX MC ETC  0 00 02                Figure 2 6  Instrument Setup step with labware named    Proceed to the next section to configure a Transfer step in your method     Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide           2 8       2 4    Transfer Step    A Transfer step loads  tips  aspirates and  dispenses liquid  and  unloads tips  This  eliminates the need to  insert four separate steps  to complete a liquid  transfer  However  there  are also steps in Biomek  Software that allow these  steps to be performed  individually using the  Intermediate Step Palette   refer to the Biomek  Software User   s Manual   Chapter 16  Using the  Intermediate Step  Palette         Learning to Create Methods for the Biomek   N             Configuring a Transfer Step    The Transfer step transfers liquid from a single source to one or more destinations   You are going to insert a Transfer step into the method and configure it to move  liquid from your source reservoir  ReSv  to your destination micropl
40. oximately 0 41     Note  Your ETC may vary slightly  That   s OK  Refer to the Biomek Software User s  Manual  Chapter 12 2 2  Configuring the Finish Step        2 6 1 Handling Configuration Errors    If there is a step configuration error in your method  the step with the error is  highlighted in red  Figure 2 12  and the ETC is displayed as Failed when you click  on the Finish or any step that occurs after the one with the error  The error message  also is displayed in the status bar or when hovering the mouse over the highlighted  step  Use the displayed message to correct the error and reconfigure the step     Note  Refer to the Biomek Software User s Manual  Chapter 25  Handling and  Preventing Errors           Biomek   Software   Method1   New   File Edit Project Instrument Execution Options Help    ALCELI CELERE           Instrument  Setup    Transfer         Combine    Comment    Start    Instrument Setup    Clear current device setup of all labware  A UCAC oN    after the method completes    N Move Labware from P Vv Unload disposable tips from all pods    i        Vv Clear current instrument setup of all labware  after the method completes    after the method completes    IV Clear all global variables after the method completes       Error Message  The error message  is displayed in the  status bar and as a  tool tip                       ETC  Displayed as  Failed when    there is an error     feporting                  Error trying to dispense to PS  coming from P4 
41. rs for customizing a method     To understand the depth of operating opportunities designed into this system  as well  as to familiarize yourself with the suite of comprehensive user   s manuals made  available at installation  links are provided throughout this guide to specific sections  in the Biomek   NX Multichannel Laboratory Automation Workstation User s  Manual  the Biomek Software User    s Manual  and the ALPs User s Manual              In addition to the instructions  graphics  and links to other sources  you will see in this  guide additional supportive information in the following forms        Definition    Important Concept These boxes contain  r brief definitions of terms  These boxes emphasize and concepts that may    important concepts be specific to the  relevant to the Biomek Biomek NX  or  NX instrument and    generally unfamiliar   Biomek Software  Refer             to the links in this quick   start guide or the    Biomek   NX  Multichannel Laboratory  Automation Workstation  User s Manual or the  Biomek Software User 5  Manual for more  information on these  important concepts                    1 2       iomek   NX Laboratory Automation Workstation       Biomek   NX    The Biomek NX  Figure 1 1  is a multiaxis instrument  designed with an open architecture to allow  expandability of the system  The design provides a  foundation for interchangeable specific use components   such as the 96 and 384 Channel Heads and a variety of  ALPs  refer to Section 1 
42. rument Setup step from the palette into the Method View   A new method is created and assigned the default name Method1  New   Figure    2 1      Note  The Instrument Setup Step tells Biomek Software where you have    physically placed labware on your deck           Biomek   Software   Method1   New      l0  x        Fie Edit Project Instrument Execution Options Help    PECI ELI DEIC                o                  amp    Start Deck   Deckt x   aa   Instrument Setup Labware Category    lt Any gt  x   Setup           Step Configuration Area   verify p  When a step is highlighted in the  4 Method View  the corresponding  configuration appears in this area     IV Pause td                     Q ig Finish E E    mee    Method View  The method is built  step by step in this  area                             BCUpsid  CostarCe   CostarFl  Costal  Drai il Greineri  Greiner  5 Collar 36r CollarSpa CostarDe  Drainable FilterHolc i    Brainard GuninavIe rainards Greiner    rainart Grainarat GrainarGt Arainarak                                 Basic Step  Palette  Displays the steps   available for  insertion into a  method  Additional    steps are available     through other step  palettes     Current Deck Display  Displays the status of the    P8   deck upon completion of  the previous step           Comment                           Methodi  BiomekXMC BiomekNX MC ETC  0 00 02                            Figure 2 1  Biomek Software main editor for multichannel Biomek NX    Beckman Coult
43. s chapter and complete the tutorial steps provided  After  you have completed the instructions  or if you need additional information along the  way  use the links provided to access the Biomek Software User s Manual        In this chapter  you    ll learn        General Description of Editors  Section 3 2        e Creating New Labware in the Labware Type Editor  Section 3 3   e Checking In a Method and Project File  Section 3 4    e Creating a New Deck  Section 3 5       Manually Controlling the Pod  Section 3 6                 Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide          3 2 Using the Editors While Creating Methods        3 2 General Description of Editors    If using Accounts  amp   Permissions       You must have Develop  Projects and Setup  Instrument permissions  to access the editors on the    Project and Instrument  menus  Refer to the  Biomek Software User 5  Manual  Chapter 2  Using  Accounts  amp  Permissions   for more information           Biomek   Software       Several editors are used to configure Biomek Software to control your Biomek NX  instrument  Editors are accessed from one of two menus     e Project menu  Figure 3 1   OR           Instrument menu  Figure 3 1        File Edit  Instrument   lt    xecutien   Options   Help     De cf       Go     Ee P  xo c AlS m E xel                         Instrun  Sety      Labware  Tran  TIP Type   Techniqu    Liquid Ty  Pipetting    Comb       Includes access to  Includes acce
44. ss to     Project Instrument    Type Editor Hardware Setup  Editor Deck Editor   e Browser Device Editor   pe Editor Manual Control  Template Editor Import Export Utility                   SY    k    Comment                      BiomekNX MC  BiomekNX MC ETC     Beckman Coulter  Inc     Figure 3 1  Biomek main editor    Following is a brief description of the editors available from the two menus displayed  in Figure 3 1  If you would like to learn more about a specific editor  refer to the  referenced chapter for more information     You will work with two of the more commonly used editors and Manual Control in  this quick start guide        Using the Editors While Creating Methods 3 3       3 2 1 Project Menu    The editors on the Project menu  Figure 3 1  allow project items to be created   copied  revised  and renamed  They include the following     e Labware Type Editor     allows you to add new labware types  copy  labware types  delete labware types  rename labware types  and edit labware  type properties  refer to the Biomek Software User s Manual  Chapter 7   Creating and Modifying Tip and Labware Types         Note  Refer to Section 3 3  Creating New Labware in the Labware Type  Editor  in this quick start guide to follow step by step instructions to create a  new piece of labware        e     Tip Type Editor     allows you to add new tip types  remove tip types  and  change tip properties  refer to the Biomek Software User s Manual  Chapter  7  Creating and Modifving
45. stem and related documentation or the procedures specified in this manual   regardless of foreseeability or the form of action  whether in contract  tort  including negligence   breach of warranty   strict liability or otherwise  and including but not limited to damages resulting from loss of data  loss of anticipated  profits  or any special  indirect  incidental or consequential damages  In no event shall Beckman Coulter  Inc    s  liability to the user exceed the amount paid by the user to Beckman Coulter  Inc  hereunder  The user assumes full  responsibility for the results obtained from the use of this system and related documentation and for application of  such results     Beckman Coulter  SAGIAN Operations  Indianapolis  Indiana  www beckmancoulter com       Biomek  Beckman Coulter  and the Beckman Coulter logo are registered trademarks of Beckman Coulter  Inc   Microsoft and MS DOS are registered trademarks and Windows is a trademark of Microsoft Corporation  All other  trademarks appearing in this manual are owned by their respective companies     Copyright 2004 by Beckman Coulter  Inc  All rights reserved  No part of this publication may be reproduced   distributed  or transmitted in any form or by any means  electronic  mechanical  photocopying  recording  or  otherwise  or stored in a database or retrieval system  without the prior written permission of Beckman Coulter  Inc     Beckman Coulter  Inc        Warranty and Returned Goods Requirements    All standard B
46. t a Move Labware step into the Method View  Manual  Chapter 2 3   after the Transfer step  Gripper              2  In Move labware from  select P5 from the drop   down menu     3  In Move labware to  select P8 from the drop down menu  The main editor  should now look like Figure 2 11        Biomek   Software   Method1   New   Fie Edit Project Instrument Execution Options Help    Dee ORS Y DELCE   i Start Using pod  Podt 7   UN rs Instrument Setup Move labware from  P5 x  to    Setup    Q b Transfer 100 uL from      Move the entire stack of labware     1 Move Labware from P C Move stack  leaving the bottom piece of labware at the source position                    Transfer rar  es  Ket C Move the topmost  1 piece of labware from the stack     Finish          Comment       E       Laced     Method1  BiomekXMC  BiomekNX MC ETC  0 00 32                Figure 2 11  Configured Move Labware step    The Move Labware step configuration is complete  Now move on to Section 2 6   Using the Finish Step  to learn how long the method will take to run        Biomek   NX Multichannel Laboratory Automation Workstation Quick Start Guide       2 14 Learning to Create Methods for the Biomek           2 6 Using the Finish Step    Use the Finish step to see how long it will take to run your method     1  Click on the Finish step in the Method View     2  Check the status bar at the bottom of the main editor for a display of the  Estimated Time for Completion  ETC   For this method  the ETC is  appr
47. tep uses a deck layout display as a map for the placement of labware   ALPs  and external hardware devices  such as stacker carousels     CAUTION  Do not place labware other than a tip box on a tip loader    2  Drag and drop the following labware from the available labware list onto the    Creating Labware    Deck Layout Display   HOU wantin use   an AP96_200uL to TL1  labware type that does  not appear in the e a Reservoir to P4  available labware list  use  lt  a BCFlat96 to P5  the Labware Type  Editor to create a new Note  To remove unwanted labware from the Deck Layout Display during  labware type  Refer to setup  drag and drop the labware to the Clear  trash  icon   Section 3 3  Creating  New Labware in the Your Biomek NX main editor should look like Figure 2 3        Labware Type Editor                 Biomek   Software   Method1   New  7 2  File Edit Project Instrument Execution Options Help Available labware list     D  amp  a   Rs Belo a       01 O    Drag the desired labware graphic to the    Start Deck   Decki x    N desired position on the Deck Layout Display   RA E Instrument Setup Labware Category   lt Any  gt  7     Setup  IB Finish                      Hm e    i Collar 36r CollarSpa  CostarCe CostarDe CostarFl  CostarFl  Drainable FilterHole Greinerl  Greinerit          Reservoir L Deck Layout  on P4 Display                         Comment                From the Deck Layout  Display  double click  the labware graphic to  display Labware  Properties          
48. u              1 4 3    MC ETC     Figure 1 3  Biomek Software main editor    Homing All Axes of the Biomek NX Pod    Every time the Biomek NX is turned on  yov   ll have to Home All Axes  This  automatic process initiates the pod and defines the home position from which all  subsequent moves are determined     To home all axes     1  From the Instrument menu  choose Home All Axes  A Warning appears     2  Make sure that all warnings are addressed and choose OK  The pod is homed and  the instrument is now ready for use     Beckman Coulter  Inc           Learning to Create Methods for  the Biomek   NX       2 1 Overview    In this chapter you will be given information and instructions for building a small  method using Biomek Software  A method is a series of steps that control the  operation of your Biomek NX     The method you complete in this chapter represents basic operations and is intended  simply to familiarize you with the Biomek NX  and to create a basic understanding of  method building opportunities     Read the information in this chapter and complete the tutorial steps provided  After  you have completed the instructions  or if you need additional information along the  way  use the links provided to access the suite of user   s manuals provided during  installation  These manuals include Biomek Software User s Manual  Biomek   NX  Multichannel Laboratory Automation Workstation User 5 Manual  and ALPs User s  Manual  Each manual contains comprehensive information
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
AQUASTAR  Continental Electric CE29502 User's Manual  Modbus COMMUNICATION PROTOCOL USER MANUAL    page ZX2 / ZX4    グルーピングについて    Copyright © All rights reserved. 
   Failed to retrieve file