1. BioZ Cardio Profile Service Manual 16 After the case has been opened electronic cables and the blood pressure module tubing may be disconnected The cables and tubing have the following attachments Valve control cable BP module tubing Loudspeaker Fan BP module cable SS ON OS A WU PB VW N Disconnecting the BP module cable reguires both sides of the connector to be pressed simultaneously to release Now the backside can be completely removed System reassembly will require additional attention during the following steps The BP module tubing must be reinstalled on the correct valves 2a NIBP module tubing 2b PW module tubing Figure 4a shows in detail the proper fan cable routing to avoid damage to this cable from the metallic shield Damage to this cable may result in a short circuit fault BioZ Cardio Profile Service Manual ARE IZ NIBP 17 11 Replacing the Hard Disk Drive Disconnect the hard disk drive cable Remove the 3 Philips screws shown in the figure to the right Remove the hard disk drive as show in the figure to the right Installing the replacement hard disk drive reguires performing the above steps in the reverse order Make sure that the hard disk drive cable is correctly connected and that the screws are securely tightened BioZ Cardio Profile Service Manual 12 Assembling the System System reassembly reguires the steps fr
2. A C power cord is properly connected e A C source is live e Blown fuses e Defective A C power cord No battery is connected to the system Check if e Battery is properly connected to the system see section 9 for details The battery is defective and should be replaced See section 9 for details Searching for ICG patient cable even though it is connected Check if e ICG cable is properly connected to the system See section 14 5 for more information Searching for ECG patient cable even though it is connected Check if e ECG cable is properly connected to the system 25 14 5 ICG Patient Cable Troubleshooting During the ICG mode of operation the green LED on the ICG patient cable will be constantly illuminated while the red LED will remain unlighted The Cardio Profile ICG waveform should be free of noise and artifacts In case of a problem check the LEDs on the patient cable using the chart below Green Orange Description of function The electronic part of the patient cable is not connected with the power supply cable is disconnected or the system is switched off lo Patient cable is ready to use but the measurement has not been started elo Measurement is running sensor contact is good Bad contact between sensors and patient at least one lead wire is X disconnected or not properly fixed sensors are too dry eventually new sensors are necessary Insufficient contact between sensors and patient at least one l
3. CBMVINCO NOLIN 7 An T 21 BioZ Cardio Profile Service Manual Troubleshooting The BioZ Cardio Profile is designed for reliable operation If you experience any problems with the system there are several things that you should verify prior to calling for service 14 1 System Errors Problem s Possible Causes Potential Solutions System fails to start when connected to the A C power mains System does not start when running on battery System starts but display remains black Unusually long startup boot up time Touch screen does not work as intended e g select wrong function Battery is not recharging Printer does not work Printer makes an unexpected clicking or grinding noise No measurement data is stored to the hard disk drive System not connected to live A C mains Defective blown fuses Battery is drained or defective Hardware failure System is performing a hard disk drive check Touch screen has calibration issue System not connected to A C power mains Battery is defective Printer paper is empty Printer cover is not closed Printer mechanical failure Data storage is deactivated Measurement was too short or did not include enough valid data Hard disk drive is full BioZ Cardio Profile Service Manual Connect A C power cord Check if wall socket is live Replace defective fuses see section 8 Connect A C power cord to recharge battery Check Hardwa
4. Press the Print button 5 Cardio Yascular Lab E Ioj x Sofware Yersion Cardio Vascular Lab Professional v2 5 14 Apr 7 2011 14 23 07 GOC 945 BIOS v6 4 Device Nr 10625001 o Serial number FARE SEA Operating system Linux version 2 6 34 tc enabled CPU 496 MHz Unicode 36 0 7 Lib Yersion libadm 1 2 10b160 settings directory Iniccomo confi Data directory data rs4data udata mounted on Hardware USE Human interface device Generic USB KB USE Bia Cardio Profile 10625001 AccuCtrl v4 29 High current charging max 223 minutes 21 74 536 844h R USE Biaz Cardio Profile 10625001 ECG module ECG Board R1 3 Fwwv3 21 Build 53 cable disconnected USE Bia Cardio Profile 10625001 ICG Patient cable MeasureBox RA 3 Fw v2 17 Build 127 SerialMoti wr USE Bio Cardio Profile 10625001 MIBP Dievice SunTech Si Lida 428 Sash V220 HW0847 SM MACAT Ai USE Bia Cardio Profile 10625001 Pulse extension NiccomoBoard RB 2 Fw w4 29 Build 163 2Valves USE Bia Cardio Profile 10625001 Spo2 Device Mell v1 0 0 0 ready sensor disconnected USE Bia Cardio Profile 10625001 Thermal printer ready F1 F4 Fi Fe Device NIEP sa Show Format Br config Test Partitions USB stick The integrated strip chart printer will create the below printout Wed Apr 13 14 46 18 2011 Device Nr 10625001 Serial number 77 3 Software Version Cardio Vascular Lab Professional v2 5 14 Apr 7 2011 14 2
5. is quite easy to cause unwanted overpressure errors using the hand bulb Procedure 1 Connect a manometer volume and the hand bulb to the Module using T adapters 2 Inflate SLOWLY when adding pressure over 200 mmHg 3 Apply various pressures between 0 and 280 mmHg to the Module with the hand bulb Verify that the Module pressure is equal to the manometer pressure 3 mmHg If the Module pressure does not agree with the manometer pressure then a calibration is required Please contact our service support for assistance BioZ Cardio Profile Service Manual 10 6 2 LeakTest International standards for automated NIBP devices require that air leakage within the pneumatic system must not exceed 6mmHg minute During manufacturing acceptable air leakage is less than 3 mmHg minute when using a 500 mL rigid volume Both of these pass criteria will not affect the performance or accuracy of the NIBP module so we suggest using the 6mmHg minute pass criteria with a 500 mL rigid volume Equipment e Calibrated Manometer optional may use pressure values supplied by NIBP module e Pneumatic T Adapters e Volume 500 mL rigid bottle a blood pressure cuff is not recommended e Hand Bulb e Connection Tubing e Timer i e Stop Watch NOTE 1 Volume is added to the pneumatic system so that the user has more control of the pressure increments when using the hand bulb If no volume is added it is quite easy to cause unwanted overpressure errors usi
6. system use the information in this manual to help correct the problem If the problem is not covered here contact SonoSite Technical Support as follows Technical Support USA Canada 1 877 657 8118 Technical Support fax 1 425 951 6700 Technical Support e mail service sonosite com SonoSite website www sonosite com Select Support 3 Warnings and Cautions For maximum safety observe the following warnings and cautions WARNING The Bio Z Cardio Profile is intended for use only within hospitals and other healthcare facilities e g outpatient clinics physician s office that provide patient care WARNING To avoid the risk of an electric shock this device must be connected to supply mains with protective earth WARNING Before opening the system ensure that the system is switched OFF and the power cable is disconnected from the A C power mains WARNING The Bio Z Cardio Profile must be connected by a three wire power cable to an electrical outlet that has been properly installed and grounded WARNING The monitor is not protected against explosive substance Do not use the Bio Z Cardio profile in the presence of flammable anaesthetics or gases WARNING Before replacing the external fuse on the unit disconnect the unit from the main power supply by pulling the plug out of the wall socket WARNING Before cleaning or disinfecting the measuring unit disconnect the unit from the main power supply by pulling the pl
7. 3 07 GCC2 95 BIOS v6 4 Operating system Linux version 2 6 34 rtc enabled CPU 498 MHz Unicode 23 5 C Data directory data rs4data data mounted on USB BioZ Cardio Profile 10625001 AccuCtri v4 29 High current charging max 238 minutes 20 3V 36 9Wh R 0 USB BioZ Cardio Profile 10625001 ECG module ECG Board R1 3 Fwv3 21 Build 53 cable disconnected USB BioZ Cardio Profile 10625001 ICG Patient cable MeasureBox R5 3 Fw v2 17 Build 127 SeriaiNo 1445353037031001 USB BioZ Cardio Profile 10625001 NIBP Device SunTech SW LM3 389 Saf SM V220 HW 0847 SN MXC8055355S Cycle 41 USB BioZ Cardio Profile 10625001 Pulse extension NiccomoBoard R6 2 Fw v4 29 Build 163 2Valves USB BioZ Cardio Profile 10625001 SpO2 Device Nelli v1 0 0 0 ready sensor disconnected USB BioZ Cardio Profile 10625001 Thermal printer ready BioZ Cardio Profile Service Manual 6 Connecting an external printer to the system will create a system identification printout as shown in the figure below Date Wed Apr 13 14 57 32 2011 Device Nr 10625001 Serial number FRM Software Version Cardio Vascular Lab Professional v2 5 14 Apr 7 2011 14 23 07 GCC2 95 BIOS v6 Operating system Linux version 2 6 34 rtc enabled CPU 498 MHz Unicode 36 0 C Printer dev usblp0 300 dpi Settings directory niccomo conf Data directory data rs4data data mounted on Hardware USB Printer Deskjet 6940 series HP Ser
8. BioZ Cardio Profile Service Manual 60197 Bio Z Cardio Profile Service Manual P15738 01 Document revision 1 0 Copyright 2011 medis Medizinische Messtechnik GmbH Ilmenau Germany BioZ Cardio Profile Service Manual BioZ Cardio Profile Service Manual Table of contents 4 1 4 2 5 1 5 2 6 1 6 2 10 11 12 13 13 1 13 2 14 14 1 14 2 14 3 14 4 14 5 15 BioZ Cardio Profile Service Manual Use Of the Manaos tiamina 1 Explanation Of used symbols anno 2 Contact IT ONIN AU ON asa aria iii ia i i i i a is 3 Warnings and C utiOns nise caca ieacaeeoaacete isa aleea acea de as altele ea aceasi altele n ate sssksassss lss sssss 3 Identification of System and Software Version 00 0 0 00 5 Device NUDO ANR RR a a iii i RR IRI i a o a IRI INCI NR 5 System Identification Printout i 6 Default Hardware Configuration ccccccsssssssssssssssssssssscsccscssscsssssssssssssssssssccccees 8 Ea Cave NO Scene 8 AS o A AA 9 Blood Pressure Module Verification cccssssssssssssssssssssssscscscssssssssssssssssccccccsees 10 Pressure Transducer Verification nono aaa 10 O a i i PO o A 11 Touchscreen Callibrationn ccccccccccsssssssssssssssssssssssssssssssssssssssssssscccccsscssssssssssssssses 12 External Fuse Re
9. Nr MY81GCSOKJ04Q9 USB BioZ Cardio Profile 10625001 AccuCtri v4 29 High current charging max 226 minutes 21 4V 36 9Wh R 0 USB BioZ Cardio Profile 10625001 ECG module ECG Board R1 3 Fw v3 21 Build 53 cable disconnected USB BioZ Cardio Profile 10625001 ICG Patient cable MeasureBox R5 3 Fw v2 17 Build 127 SerialNo 4535303703 100i USB BioZ Cardio Profile 10625001 NIBP Device SunTech SW LM3 389 Saf SM V220 HW 0847 SN MXC805555M5 Cycle 41 USB BioZ Cardio Profile 10625001 Pulse extension NiccomoBoard R6 2 Fw v4 29 Build 163 2Valves USB BioZ Cardio Profile 10625001 SpO2 Device Nell1 v1 0 0 0 ready sensor disconnected USB BioZ Cardio Profile 10625001 Thermal printer ready Please send a copy of this system identification printout to us by e mail or fax BioZ Cardio Profile Service Manual 5 Default Hardware Configuration 5 1 Hardware Info Screen Connect all eguipment to the Cardio Profile ICG patient cable ECG patient cable NIBP cuff connect to A C power and go to the Hardware Info Screen see section 4 2 Default entries The following entries should be stated in the Hardware Info Screen Battery AccuCtrl vx xx High current charging The battery is being charged additional information additional Liss additional formation AccuCtrl vx xx Initializing The battery is initializing e g additional information after startup and should change to one of the entr
10. compartment using a screw driver Use care to not damage the case A WARNING Make sure the replacement fuse is of the same type and value Use of a substitute fuse type or value may result in safety risks or damage to the system 1 0 Amp Slo Blo See the Parts List for more information If the replacement fuses continue to fail then the system reguires inspection by an authorized service department BioZ Cardio Profile Service Manual 9 Battery Pack Replacement The battery pack can be disconnected or replaced by opening the battery compartment cover located on the bottom of the system This compartment cover entry allows access to the battery without having to remove the system enclosure This procedure requires the removal of 4 Philips screws securing the battery access cover Access into the battery compartment will allow the battery cable to be disconnected or the removal replacement of the battery pack 1 Battery pack replacement requires performing the above steps in the reverse order Upon battery pack reinstallation ensure that the battery cable is connected and routed properly as to not be pinched or cut by the internal hardware or the battery compartment cover Secure this cover with the 4 screws removed earlier BioZ Cardio Profile Service Manual 14 10 Opening the System A WARNING Turn OFF the power switch and disconnect the A C power cord from the system Place
11. e department prior to troubleshooting We can assist to identify a possible failure and recommend the best and most efficient method to resolve the issue s For the best possible support we recommend a short description of the issue s observed and all possible data from the system See Chapter 4 Indentification of System and Software Version These conventions are used in this service manual AWARNING describes precautions necessary to prevent injury or loss of life e A Caution describes precautions necessary to protect the product Numbered steps must be performed in a specific order e Bulleted lists present information in list format but do not imply a sequence BioZ Cardio Profile Service Manual 1 1 Explanation of used symbols Associated instructions provide important information to prevent injury loss of life or product damage Do not dispose with residual waste oka Device type BF Defibrillator proofed Device type BF is ae Insulation class Equipotential connector Caution electrostatic sensitive devices USB Connector Device Powered On Stand By Connector for external keyboard Network connector Connector for external VGA monitor jose Connector for serial interface BioZ Cardio Profile Service Manual 2 Contact Information Ouestions and comments are encouraged SonoSite is interested in your feedback regarding the service manual If you encounter difficulty with the
12. ead wire is disconnected or not properly fixed sensors are too dry new sensors are necessary xt Patient cable has power but the software cannot access the cable software has not been started or is not ready for measurement Legend O LED off xy LED flashing 6 LED on BioZ Cardio Profile Service Manual 26 15 Parts List P15787 01 BioZ Cardio Profile Service Manual 27 N LM
13. et the NIBP module it must have the power cycled or it needs to be set to its sleep mode and awakened Not all NIBP module customers have the ability to use the sleep mode a Open the hand bulb valve to exhaust the air from the pneumatic system 7 Subtract P2 from P1 and this is the leak rate per minute BioZ Cardio Profile Service Manual 11 7 Touchscreen Calibration Recalibration of the touch screen may be necessary when 1 Pressing a touchscreen button does not activate that function but the click sound is heard 2 Pressing a touchscreen button activates a nearby function To calibrate the touchscreen please perform the following steps 1 Power OFF then power ON the system The calibration function can only be performed on a freshly restarted system 2 When the patient data screen is displayed touch the screen on any random blank area and continue pressing the screen for approximately 5 seconds until the calibration menu is displayed see next picture 3 The calibration menu will prompt you to touch every corner of the screen display The touched positions will be saved as the calibrated data and now the touchscreen operation will respond properly to the keystrokes entries Please touch the point Touchscreen calibration BioZ Cardio Profile Service Manual 8 External Fuse Replacement The system is protected by 2 fuses which may be removed as shown in the figure to the right Open the fuse
14. for details see section 14 5 Prep the patient s skin with soap and water before applying sensors Be sure patient is laying still motion such as coughing and talking will produce instability in the waveform Remove any jewelry that may interfere with the lead wires Attach fresh sensors Reposition the sensors Remove the ECG cable from the system and restart the measurement Connect ECG patient cable to the patient Invert the ECG waveform Connect the ECG patient cable to the system and at the patient and restart the measurement 23 14 3 Blood Pressure Errors Problem s Possible Causes Potential Solutions Weak or No Cuff is in wrong position Reapply Cuff and retake BP Oscillometric Signal Cuff is loose on patient s arm ensuring that Artifact Erratic Oscillometric Signal Exceeded Measurement Time Limit Pneumatic Blockage Cuff Overpressure Inflate Timeout Air leak or Loose Cuff Blood pressure measurement does not start Excessive clothing between The arrow on the cuff is cuff and patient s arm Cuff is the wrong size Patient is not motionless and relaxing their arm Patient has weak and or low pulse Hose cuff and or fittings are blocked Hose or cuff is crimped and or folded Hose cuff and or fittings are blocked Hose or cuff is crimped and or folded Hose cuff and or fittings have a leak Cuff is loose Hose cuff or fittings are not connected properly Input e
15. he password request go to the password entry screen leave the password entry line blank and push the keyboard Enter Y The system will not prompt for a password request upon power ON BioZ Cardio Profile Service Manual 6 Blood Pressure Module Verification If you have any doubts in the measured blood pressure values or you think the cuff is deflating too quickly perform the following Pressure Transducer Verification or the Leak Test to check the function of the NIBP module If the system fails in one of the performed tests then the Cardio Profile and the blood pressure supplies must be sent to service for further testing 6 1 Pressure Transducer Verification International standards for automated NIBP devices require that the maximum static pressure accuracy shall be 3 mmHg This is a stringent requirement and all test equipment must be in excellent working order to properly perform this test It is important to verify the calibration before changing it Although experience has shown that the transducers rarely need to be re calibrated the annually pressure verification is required Equipment Calibrated Manometer Pneumatic T Adapters Volume 500 mL bottle or regular sized cuff wrapped around a solid object is suggested Hand Bulb Connection Tubing NOTE Volume is added to the pneumatic system so that the user has more control of the pressure increments when using the hand bulb If no volume is added it
16. ies above within some seconds ECG module ECG Board Rx x Fw vx xx Build xx cable connected version and cable is stated ICG ICG Patient cable MeasureBox Rx x Fw vx xx The connected ICG patient cable Build xxx SerialNo xxxx with Serial Number and Firmware is stated see section 4 2 for more information NIBP Device SunTech SW LMx xxx Saf SM The connected NIBP module with Vxxx HW xxxx SN xxxx Cycle xxx Serial Number and Firmware is stated Pulse extension NiccomoBoard Rx x Optional Typically not installed Thermal printer ready USA There can be additional entries depending on the connected devices e g USB sticks printers etc Using the information in the Hardware Info Screen for troubleshooting is addressed in Section 14 Troubleshooting BioZ Cardio Profile Service Manual 8 5 2 HIPAA Password The HIPAA Password configuration is performed by going to Options F5 select Password and type in the selected password Push the keyboard Enter Y key to save the entry The system will now require this password during each power ON or when the system is at idle for more than 5 minutes The operator will not be prompted for this password during a patient study x C Cardio Yascular Lab Pat ID Fxaminatinn Tue 06 21 2011 00 54 Please enter Password re Fe 1 F5 Fa F10 F41 Exit zelas RENE See e Start Back Alerts patient patient values To remove t
17. ng the hand bulb Also leak rates are exaggerated with extremely small volumes NOTE 2 Blood pressure cuffs have large variations in size and materials which can cause large amounts of variation in the leak test results To limit this effect we strongly urge to only use a 500 mL rigid volume Test Method 1 Connect a manometer volume and the hand bulb to the Module using T adapters 2 When using a hand bulb make sure the exhaust valve is closed 3 Increase the pressure to approximately 250 mmHg using the hand bulb or the pump on the module It is difficult to slowly increase the pressure in the pneumatic system with the pump so a hand bulb is preferred NOTE When applying high pressures take special care to increase the pressure at a rate that will not cause unwanted overpressure errors 300 mmHg 4 Start the timer and wait 60 seconds for the pneumatic system to reach its pressure equilibrium point All pneumatic systems have some degree of expansion and rebound which are affected by the several factors including volume material types and material durometers 5 After the waiting period record the pneumatic pressure level P1 and wait another 60 seconds and record the pneumatic pressure level again P2 NOTE Safety circuitry on the module only allows the pressure in the pneumatic system to remain above 10 mmHg for 180 seconds When this safety time limit is exceeded the valves will open and the module will be disabled To res
18. om Section 10 Opening the System be performed in the reverse order When reinstalling the metallic back to the enclosure ensure that the connecting slots of the internal shield into the corresponding tabs as shown figure to the right A Caution Carefully check that the fan cable is not damaged from being cut by the internal shielding and the metallic back panel Damage to this cable may result in a short circuit and or subsequent damage to the electronic components A WARNING All necessary function and safety tests must be performed and filed after the reassembly of this system BioZ Cardio Profile Service Manual 19 13 System Overview 13 1 Basic Components The figure below shows the internal view of the system and identifies the following internal components O aa a q E 9 Q q IL i Power supply module protector Strip chart printer Multi channel measurement Input module SpO2 measurement module optional Power management board Power supply module Battery pack Computer board Power connection module Hard disk drive SD ASA ASA gt 00 Y VU BUN A A A 2 BioZ Cardio Profile Service Manual 20 13 2 Block Diagram TVNENTIHENO SKAN 30 INS Dy OA HOHLITY NY 40 LN33N00 NILLH AM SES bd XS HL INOHLM 350d Hd ANY 404 0351138 11 NOd YO NI MOL WEJOJN IHL 20 AN AMIN BON EEN ANY NI 0350 YO 031400 38 100 AW LI 13101 HAND SIDIN 30 ALSO ci ALS ed JHL HY LI NI
19. placements 13 Battery Pack Replacement e dap dea aha kasi ai ki sasas akiai 14 Opening AS I II 15 Replacing the Hard Disk Drive ooooommmmmsssss 18 Assembling the AA oo e o a o 19 System OVE OW ase asis iais ninia pierna cines 20 Bas le COMPONER taa 20 oy el EI 6 19 een 21 TODOS OO CLIN e sasas iat iii i is i ss S S SS 22 OA O ac e D 22 S EC a TONS a 23 BIOOG Pressure Error de sac catei oaie le oaie datat alu ii ata ratata i 24 Hardware Info Screen Failure ENtrieS a aaa aaa aaa aaa aaa aaa 25 ICG Patient Cable Troubleshooting 26 O o LO 27 BioZ Cardio Profile Service Manual vi 1 Useofthe Manual Before repairing the system read completely and carefully the instructions provided in this manual If you do not understand any instructions or need additional information please contact SonoSite Technical Support The intended purpose of this manual is to help biomedical and or clinical engineering personnel understand the functions of the system to identify possible failures and to guide in the repair Before starting an analysis with regards of potential hardware problems it is important to make certain that the system has been applied and operated in accordance with the instructions outlined in the user s manual Because of the complexity of the system and the numerous interactions between hardware and software components when encoutering issues we recommend contacting our servic
20. re Info Screen see section 4 2 and 5 1 for battery status Connect A C power supply cord and restart the system Wait until the check is performed Do not power OFF the system The system will perform this check after every 20 restart Perform the touch screen calibration see section 7 Connect A C power cord to recharge battery Check if the A C wall supply is live Check Hardware Info Screen see section 4 2 and 5 1 for battery status Replace battery see section 9 Replace the paper Close the printer cover Check the Hardware Info Screen see section 4 2 and 5 1 Open the printer s cover and check for any obstructions Check if The data storage option is activated in the Option screen Valid data was recorded The hard drive capacity is at 096 22 14 2 ICG ECG Errors Problem s Possible Causes Potential Solutions An ICG measure e ICG patient cable is not ment cannot be connected started ICG button is not available to select as exam type ECG or ICG has Patient movement excessive artifact Sensors are dry Sensors are not properly applied No heart rate is ECG cable is connected to measured during the system but not to the ICG Measurement patient ECG vector cannot be measured with the ICG patient cable BioZ Cardio Profile Service Manual Connect the ICG patient cable to the system Check the Hardware Info Screen see section 4 2 and 5 1 Check the ICG patient cable LEDs
21. rror Communication error Hardware error BioZ Cardio Profile Service Manual located over the patient s artery just above the elbow joint The cuff is wrapped snugly There isn t excess clothing under the cuff The size is correct for the patient s arm circumference The patient is motionless with their arm relaxed during the measurement cycle Clear hose cuff or fittings of any blockage Replace hose and or cuff Clear hose cuff or fittings of any blockage Replace hose and or cuff Check cuff connection Replace hose cuff and or fittings Check if the measurement was started properly Check the Hardware Info Screen for the NIBP module status see section 4 2 and 5 1 24 14 4 Hardware Info Screen Failure Entries The following entries can be an indicator of a hardware failure Follow the information in the Note section If any Entries remain for the corresponding module then the system with its equipment must be sent to service for further testing Battery AccuCtrl vx xx Running System is running on battery even though it is on accumulator power additional information AccuCtrl vx xx No accumulator found additional information AccuCtrl vx xx Accumulator possibly defective additional information Patient cable Waiting ECG module ECG Board Rx x Fw vx xx Build xx cable disconnected BioZ Cardio Profile Service Manual connect to the wall supply Check if e
22. the Bio Z Cardio Profile cannot fall down if someone pulls on the cable Caution The Bio Z Cardio Profile is protected against accidental ingress of liquids If liquids ingress into the case disconnect the monitor from the wall outlet Caution Do not expose the cables to mechanical or thermal impact Avoid temperatures above 40 C 100 F BioZ Cardio Profile Service Manual 4 4 Identification of System and Software Version This system may be delivered in different hardware and software configurations In the event of a service request it will be necessary to identify your system from our data base To identify your system and the installed software version we need the following information e Device number e System identification printout 4 1 Device Number The device number is located on the backside of the system as indicated in the figure to the right Type BioZ Cardio Profile Device no 10625001 lt 1 100 240V gt 0 5 0 2A M 50 60Hz 2011 BioZ Cardio Profile Service Manual 5 4 2 System Identification Printout The system identification printout includes all relevant hardware and software information from the system The following steps will create a system identification printout Power ON the system and wait until the patient screen appears Press the Options button a new screen appears Press the Hardware button the next screen appears see figure below
23. the system on its front side as shown in the figure to the right and use a soft underlay so that the display touch screen will not be damaged Before you open the system enclosure it is important that you have disconnected and removed the battery pack as previously described Failure to do so will result in the battery pack falling down and or creating an electrical short Remove the 4 middle screws of the system back panel Now remove the remaining 13 back panel screws The enclosure is now loose and will allow movement upward to clear the system See the figure to the right BioZ Cardio Profile Service Manual Remove only the 10 red marked screws from the backside of the inner metallic shield using a Philips screwdriver Refer to the figure to the right A Caution Do not apply excessive downward force while removing the screws Too much force will cause the press fit nut to separate and drop from its mount on the metallic shield A Attention Removal of the other screws that are not marked red will result in loosening other components such as the fan loudspeaker and blood pressure module Moving the metallic inner panel upward will allow enough space to disconnect electrical cables and the blood pressure module tubing A Caution y 4 5 Uh gt A M m Use caution when moving the internal metallic panel as damage could result to the internal parts and the electrical cables
24. ug out of the wall socket WARNING Do not modify any SonoSite products Any modifications to the device will void the warranty WARNING Systems equipped with a battery must have the battery disconnected to avoid short circuit or contact with developed high voltage up to 900V which is present for the backlight inverter of the TFT display BioZ Cardio Profile Service Manual 3 Caution The system and or software must be maintained repaired modified or altered only by persons authorized by medis Sonosite or with prior approval by medis SonoSite Caution Never gas sterilize the patient cables They are covered with a PVC shroud that can react with the gas sterilization procedure to form a highly toxic ethyl nitrite Sterilization in an autoclave with ultrasound or gamma rays is also not permitted Caution Do not sterilize the Bio Z Cardio Profile parts or accessories Sterilizing the device or accessorries will void the warranty Caution Use only cables or components recommended by SonoSite to connect to the inputs and outputs of the monitor Caution Do not place the measuring unit on a surface which emits heat or expose it to direct sunlight Do not cover the ventilation slits Caution Do not operate in bathrooms or other areas where water damage is possible The Bio Z Cardio Profile is not protected against splashing water Caution The Bio Z Cardio Profile should be placed near the patient and fixed in such a way that