Home
EC9850 Operation Manual
Contents
1. 2 11 Analyzer Rear Panel illustration 2 2 Automatic zero and span 78 3 15 B Bavarian command 4 8 4 15 Bavarian 1 4 11 C Cable Connections 4 6 Calibration Menu Manual sees 2 24 Calibration Menu Timed sess 2 22 Calibration 3 19 Calibration requirements eene 2 42 Calibration 1 2 14 3 2 Calibration multipoint esee 3 3 Calibration multipoint with over ranging 3 14 Command set 9800 4 7 4 16 Command set 4 8 4 15 Communications multidrop 4 13 Connections 2 7 Connections recorder and 85 2 2 Connections sample 2 6 Connections serial sess 4 5 Connections Zero air 2 7 D DAS 9 2 2 Data Logging RESP 2 34 Date and time setting sene 2 13 Diagnostic Mentai near 2 30 Diagnostic Menu 22 2 2 30 Digital communication sse
2. an e san esso e anos epos sean seen E sean e seo P tense sane eaa 2 1 2 MECHANICAL INSTALLATION sten Aca Te I e dde Ac he die eid 2 1 ZL Selecting a Location ie DH ee ORO Ee 2 1 2 1 2 Connections etis tietveteiisetiecteia te Pe o REO Rae Evae Pea dee 2 1 2 2 AC IPOWER CONNECTIONS ette rte tee eee trei reel mi o 2 8 2 2 1 Display Adjustments iiec rite e te e Er ad 2 8 2 29 t Ee eee fade ete toes sade ee beer e e e ere de e ees 2 10 2 3 OPBRATI N4 ROG I rU ID E SE A 2 11 2 3 1 General Operation Information 2 11 2 3 2 Using the Menu and Making Entries eese eene enne nnne ene 2 12 2 33 Setting the Date and Time gait tar ee ORE P i et an exei e preis 2 13 2 4 ANALYZER CALEIBRATI ON edite e rr ee e dese PE eR Ces idee dee e ee Tree ee Ges nde Pee edes eee 2 14 Z 4 I PrectSiom Checks v eR Rn n Nu 2 14 2 472 AULOM AIC See edenda aee mtt eddie NN dead 2 14 2 4 3 Manual iis iih elle ed DI e e RU 2 14 2 4 4 Analyzer Calibration Instructions eese eese ener teen rennen rennen nter enne nnne 2 15 2 9 MENUS AND SCREENS sos ote impete hime tope eae s et dre e dp cer eed de Pu e ee EN a eee atu ety 2 16 2 3 LI Primary E ROUEN ER
3. 98507600 Rev D 3 13 EC9850 SO ANALYZER OPERATION MANUAL Equation 3 3 A total flow rate of 0 783 slpm is required Always be sure that the required flow rate exceeds the analyzer demand at the output manifold Consult the permeation device distributor for further details 3 2 3 4 Procedure Follow the procedure given in section 3 2 1 using the permeation system to generate the SO concentrations required in steps 5 and 11 3 3 Calibration Requirements When Over Ranging Is Employed If you are utilizing the over ranging feature of the analog outputs use the following steps in conjunction with the procedure in section 3 2 to calibrate the instrument Choose the desired upper range limit for the normal monitoring range RANGE Choose and set the desired upper range limit for the higher over range OVER RANGE A value between 2 and 5 times the range value is recommended Set OVER RANGING to DISABLED to deactivate the over ranging feature Temporarily set the range to equal the chosen over range Check the zero and set the span as described in section 3 2 Generate several concentration standards and determine the slope intercept and linearity of the higher over range Reset the range to the normal monitoring range Generate several concentration standards including zero air and determine the slope intercept and linearity of the lower range Note Once the span has been set on the higher range
4. 2 6 3 Analog Output Calibration Procedure This procedure is appropriate for connecting the EC9850 analyzer to a strip cart recorder Data logger DAS or to a Digital voltmeter DVM 1 Goto the INTERFACE MENU and choose ANALOG OUTPUT MENU 2 Select RANGE and enter the desired range by selecting the appropriate digits Press Enter to confirm your choice 3 Set the output type according to the termination selected for the discrete I O connector The choice will be either CURRENT or VOLTAGE If current output is desired and the 50 pin board is installed set the selection jumper to CURRENT and de select all voltage ranges If current output is desired and the 50 pin board is not installed no hardware change is required If voltage output is desired and the 50 pin board is installed set the selection jumper to voLTAGE If voltage output is desired and the 50 pin board is not installed an external termination resistor is required This resistor must be 50 ohms per full scale voltage desired 50 ohms 1 V full scale 500 ohms 10 V full scale etc 4 If voltage output type was selected choose the desired orrsET and press Enter If current output type was selected choose the desired output range and press Enter 5 Select ZERO ADJUST and adjust the analog output to the selected offset position for zero concentration i e if 10 OFFSET is selected position the recorder pen or DAS at 10 of full scal
5. 83 Instrument Status 84 Reference Voltage 85 Calibration State 0 MEASURE 1 CYCLE 2 ZERO 3 SPAN 86 Primary Raw Concentration before 984X background and gain 87 Secondary Raw Concentration before 984X background and gain 88 984 Background Concentration before gain 89 Calibration Pressure 90 Converter Efficiency 91 Multidrop Baud Rate 92 Analog Range Gas 1 93 Analog Range Gas 2 94 Analog Range Gas 3 95 Output Type Gas 1 0 Voltage Current 96 Output Type Gas 2 0 Voltage Current 97 Output Type Gas 3 0 Voltage Current 98 Voltage Offset Current Range Gas1 02096 or 0 20mA 12596 or 2 20mA 2 10 or 4 20mA 99 Voltage Offset Current Range Gas2 02096 or 0 20mA 1 5 or 2 20mA 221096 or 4 20mA 100 Voltage Offset Current Range Gas3 02046 or 0 20mA 1 5 or 2 20mA 2 10 or 4 20mA 101 Full Scale Gas 1 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL 102 Full Scale Gas 2 103 Full Scale Gas 3 104 Zero Adjust Gas 1 105 Zero Adjust Gas 2 106 Zero Adjust Gas 3 107 Negative 10V Supply 108 50Pin IO ANINI 20mV resolution analog input 0 5V 109 50 Pin IO ANIN2 20mV resolution analog input 0 5V 110 Instrument State 111 CO Linearisa
6. The choice of INTERNAL Or EXT ERNAL valves that will operate during a zero span check The EZS valve option must be installed to use EXTERNAL 98507600 Rev D 98507600 CHAPTER 2 INSTALLATION AND OPERATION SPAN COMP A choice of ENABLED Or DISABLED This function automatically corrects span readings to the expected value See Chapter 3 for additional information Note For U S EPA designated use SPAN must be set to DISABLED SO2 SPAN RATIO A microprocessor generated field that displays the value by which the span reading is multiplied to correct it to the calibration value This value is only applied if SPAN COMP is ENABLED BACKGROUND A choice of START DISABLED or ENABLED If START is selected then when Enter is pressed the instrument begins an auto zero background cycle If DISABLED is selected then the instrument will not run the normal auto zero background cycle If ENABLED is selected then the instrument will run the normal auto zero background cycle Note For U S EPA designated use the background cycle must not be disabled BACK INTERVAL Set time in HOURS 2 4 6 8 12 24 A microprocessor controlled field that indicates when an autozero cycle will begin The results of the background are stored in the EVENT LOG SO2 CAL PRESSURE This is the measured ambient pressure during the last SO2 calibra
7. ETX characters it will cause the response lt NAK gt BAD STX ETX PAIR lt CR gt lt LF gt 4 4 3 12 Enhanced Protocol Idiosyncrasies g g 4 12 The string DA CR is a valid command The DA command will function without an I D 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION o The data type must be sent on 9800 style commands but it is not checked against the actual parameters o The number of data parameters must be sent on 9800 style commands but it is not checked against the actual parameters 4 4 4 Establishing Communications The first step in establishing communications with the EC9850 is to connect a computer or terminal to one of the instrument s RS232 serial ports as specified in section 4 2 1 The default serial configuration for either serial port is 2400 8 N 1 2400 baud 8 bits no parity and one stop bit If you need to change the serial configuration from the default use the INTERFACE MENU Once the instrument has been connected place the instrument in Command mode by entering the INTERFACE MENU via the front panel and selecting comman as the INTERFACE MODE Then using a communication package such as Hyper Terminal establish communications with the instrument To test the communication connection type and press the Enter key The complete alphanumeric set recognized by the EC9850 should be displayed on the computer followed by END
8. Device response lt ACK gt if the unit under test is able to perform the command lt NAK gt if not Command PINID Function Programs the 50 pin device ID 98507600 Rev D 4 21 EC9850 SO ANALYZER OPERATION MANUAL 4 22 Format PINID DEVICE I D gt 1 1 BBBBBBBB TERMINATOR where BBBBBBBB 15 the desired bit pattern in binary format to be programmed into the device ID The most significant bit is on the left least significant bit on the right for example 10100101 would correspond to a device ID of A5 hex or 165 decimal Device response lt ACK gt Command REMOTE Function Puts the instrument in the VT 100 compatible terminal mode All of the menus become available to a remote controller through the serial port The remote PC an ANSI terminal may also be used should be configured as follows Windows Terminal mode Hyper Terminal accessory terminal emulation VT 100 communications settings 9600 or whatever the current instrument host baud rate is 8 bits 1 stop no parity An ANSI terminal should be configured as follows WYSE WY 60 or WY 75 VT 100 emulation full duplex WYSE WY 50 Not recommended no ANSI mode Recommended baud rate is at least 4800 baud The following remote terminal keys are now active using numeric keypad with NUM LOCK enabled on remote terminal Key Key Label Function Enter Enter Enter 7 Home Exit 9 Pg Up Page up 8 Up arrow
9. SAMPLE MEASURE Instrument operational must be calibrated if this is the first power up sequence Note The EC9850 will re run the above startup routine whenever power has been lost for greater than two minutes If power is lost for less than two minutes the analyzer will return to its previous settings without the startup routine 2 10 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION 2 3 Operation The operation section describes the actions necessary to operate the instrument first in general then in specific terms In section 2 5 the menu headers are shown as they appear on the display screen The illustration is followed by explanatory information regarding the menu entries or choices The entire menu tree is shown in Figure 2 6 2 3 1 General Operation Information operator responses needed to operate the EC9850 are performed by pressing the 6 keys available on the front panel to the right of the display screen The key functions are described on the following page AA Cb Figure 2 5 Analyzer Keyboard The key functions are listed below o Up arrow key A Moves the cursor to the previous menu item or in an input field moves the cursor to the next choice or increments the digit in a numerical field Down arrow key Moves the cursor to the next menu item or in an input field moves the cursor to the next choice or decrements the digit
10. VALVE 42 Zero air INT VAVLE 3 Span gas AUX VALVE 2 connections Not used at this time EXT MEASURE Externally supplied EXT ZERO GAS sample stream Externally supplied zero air EXT SPAN GAS Externally supplied VALVE SEQUENCING span gas Set to either ON or OFF ON is used for automatic valve control of zero span cycles OFF is used for operator control of valves VALVE SEQU be set to oN whenever the primary screen is displayed ENCING Will automatically 2 29 EC9850 SO ANALYZER OPERATION MANUAL 2 5 11 Diagnostic Menu DIAGNOSTIC MENU MULTIDROP PORT TEST NO WATCHDOG TEST NO DISPLAY TEST NO ANALOG OUTPUT TEST NO Figure 2 18 Diagnostic Menu The DIAGNOSTIC MENU is information used to diagnose problems or suspected problems The settings return to the previously set conditions when the operator leaves this menu MULTIDROP PORT TEST Sends test of all printable characters to the Multidrop rear serial ports WATCHDOG TEST Disables strobes to the watchdog timer The system resets when this test is executed DISPLAY TEST A series of 6 tests are available to check the working order of the display screen Once the test is selected press the Select key to verify that the alternate pixels are visible Press the Pg Up key to exit The available tests are STRIPE 1 Causes the screen to show a series of very closely spaced vertical lines STRIPE 2 Sh
11. no further adjustments should be made on the lower normal monitoring range Reset OVER RANGING to ENABLED to reactivate the over ranging feature 98507600 Rev D CHAPTER 3 CALIBRATION 3 4 Automatic Zero Span Checks AZS 3 4 1 Over time the calibration of the instrument may change slightly drift causing error in the measured values Accordingly good quality assurance practice requires that the calibration of the EC9850 be checked periodically and if necessary that the instrument s zero and span be adjusted to restore accurate calibration U S EPA Definitions Section 12 of the QA Handbook for Air Pollution Measurement Systems defines two types of calibration checks a Level 1 check and a Level 2 check A Level 1 zero and span calibration check is an authoritative assessment of the analyzer calibration using an SO span gas standard that is certified traceable to an SRM or CRM and the results of the Level 1 check can be used to adjust the analyzer zero and span to restore accurate calibration A Level2 zero and span check is an informal calibration check often with an uncertified SO standard used to monitor the day to day relative readings of the analyzer The results of a Level 2 check must not be used to adjust the analyzer calibration but may indicate the immediate need for a more authoritative Level 1 calibration check 3 4 2 AZS Outline When used with a certified traceable SO span standard and extern
12. the remote mode window the keyboard will not work The keyboard will be reac tivated when one of the buttons on the Applet is selected with the mouse NOTE To access the Hidden menu press the H key then click the refresh button with the mouse immediately 4 7 3 Download The download option gives the user the possibility of downloading the logged data to their hard drive EC9841A NOx Analyser Software Version 1 13 04 Start Date OPEN DATE 1 1 2003 v hh v mm v Start Date PICKER End Date v Download all data logged since Start Time ALL DATA Data Type All Logged Data X DATA SINCE Date Format yy mm dd nh mm v TYPE AND d File Path stlec9800 t Browse y SM ATE Download ime Rew FILE PATH AND uL FILE NOME Figure 4 9 Download Data window Starting date The starting date of the data to be downloaded can be selected either using the drop down scrolls in the first two fields or by using the Start date button to open a date picker End date The end date of the data to be downloaded can be selected either using the drop down scrolls in the first two fields or by using the Start date button to open a date 4 34 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION picker This option is only possible if the Download all data logged since Start time is not ticked Download all data since Tick the box next to Download all data logged since Start Time to ensure a
13. which ever is greater with Kalman filter active o Analog output 0 25 ppb or 0 1 of analog output full scale whichever is greater 1 1 3 Lower Detectable Limit o Measurement process Less than 0 5 ppb or 0 2 of concentration reading whichever is greater with Kalman filter active o Analog output 0 5 ppb or 0 2 of analog output full scale whichever is greater 1 1 4 Zero Drift o Temperature dependent 0 1 ppb per C o Time dependent at fixed temperature o 24 hours Less than 1 0 ppb o 30 days Less than 1 0 ppb 1 1 5 Span Drift o Temperature dependent 0 1 per C o Time dependent at fixed temperature o 24 hours 0 5 of reading o 30 days 0 5 of reading 1 2 98507600 Rev D CHAPTER 1 DESCRIPTION 1 1 6 Lag Time Less than 20 seconds 1 1 7 Rise Fall Time 95 of Final Value Less than 120 seconds with Kalman filter active 1 1 8 Linearity Error 1 of full scale from best straight line fit 1 1 9 Precision 0 5 ppb or 1 of reading whichever is greater 1 1 10 Sample Flow Rate 0 50 slpm A series 0 60 slpm B series 1 1 11 Sample Pressure Dependence A 5 change in pressure produces less than 1 change in reading 1 1 12 Temperature Range o 5 to 40 C 41 to 104 F o U S EPA designated range 15 to 35 C o Eignungsgepr ft Range 5 to 40 C 1 1 13 Power o 99 to 132 VAC 198 to 264 VAC 47 to 63 Hz o U S EPA designated range 105 to 125 VAC 60 Hz or
14. 4 Discrete contre i anon aot ande 4 1 Display adjustment seen 2 8 E Enhanced 4 12 98507600 EC9850 SO ANALYZER OPERATION MANUAL Bvent EOg SCIeen eeepc taie 2 37 Event Log Screen illustration ssse 2 37 Exhaust connections cescceseeeseeeseceeeeeeeseeeeeeteees 2 7 I Illustrations 50 Pin Connector Board with Sample Choices essere 2 4 Illustrations Analog Output 2 34 Illustrations Analog Output Menu Current Output 2 32 Illustrations Analog Output Menu Voltage Output 2 33 Illustrations Analyzer Front Panel 2 9 Illustrations Analyzer 2 11 Illustrations Analyzer Rear Panel 2 2 Illustrations Cable 4 6 Illustrations Diagnostic Menu 2 30 Illustrations Event Log Screen 2 37 Illustrations Instrument Menu s 2 18 Illustrations Instrument Status Screen 2 37 Illustrations Interface 2 31 Illustrations Main Menu eee 2 18 Illustrations Manual Calibration Menu 2 24 Illustrations Measurement Menu 2 20 I
15. 4 24 4 31 USEPA Equivalent 1 4 V Valve Test derer herein 2 29 Valve Test Menu illustration 22 2 2 29 W Wartmup uctor tete 2 10 Z Zero air 2 7 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL APPENDIX A USB PROTOCOL PARAMETER LIST Note parameters are for all EC9800 analyzers and may not be applicable to an individual ana lyzer Description Notes 0 Internal Valve 1 0 1 1 Internal Valve 2 0 1 2 Internal Valve 3 0 1 3 External Measure Valve 0 Closed 1 4 External Zero Valve 0 1 5 External Span Valve 0 Closed 1 Open 6 Aux Valve 1 0 1 7 Aux Valve 2 0 1 8 Aux Valve 3 0 1 Open 9 Valve Sequencing 0 Off 1 10 LCD Contrast POT O Lightest 99 Darkest 11 1 Measure coarse pot for all analysers except 9841A which is chassis fan speed 12 PRE POT 2 Measure Fine 981X 9820 9830 9841 9842 Bench Fan Speed 9841A Reference zero 9850 13 PREPOT3 Input for all except 9850 which is measure gain 14 PRE POT 4 981X 984X Test Measure 9820 9830 test_reference 9850 reference gain 15 PRE POT 5 981X Lamp Adjust 9820 9830 9850 test measure 984X hi
16. 50 Pin I O Outputs The status outputs are active low open collector Darlingtons The status outputs can be used to drive relays or with the use of external pull up resistors as a voltage indication of on off conditions The internal 12 V pin 50 or an external power supply may used as the relay or indicator power source Current through the outputs should be kept as low as possible ideally around 1 mA If an external supply is used it should be less than 50 VDC and the current sunk by each output should be 50 mA If the internal 12 V supply is used the total current drawn must be kept to less than 50 mA or damage to the analyzer will result 4 4 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION Figure 4 1 Status Output Connections 4 2 Serial Control Two modes of operation are available using the serial interface These modes are Terminal and Command In Command mode a library of commands becomes available These are listed at the end of this chapter In Terminal mode the instrument communication is through the analyzer menu structure 4 2 1 Serial Connections 98507600 Rev D The EC9850 has two tristate RS232 ports on the rear of the analyzer The tristate RS232 causes all instruments not addressed to turn off their transmission capability until the next activation command is received Communication among devices is defined in terms of Data Terminal Equipment DTE and Data Communication Equipment DCE per the EIA standar
17. BIT 7 P4 BIT 6 P3 BIT 5 P2 BIT 4 Pl BIT 3 Spare Driver 1 BIT 2 BIT 1 98507600 Rev D 9850 SO ANALYZER OPERATION MANUAL BIT 0 Reference Fail 46 50 PIN IO bits 40 47 BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 P6 BIT 0 P5 47 50 PIN IO bits 48 55 BIT 7 Status 2 LED BIT 6 Status 1 LED BIT 5 Sys Fail LED BIT 4 HeartBeat LED BIT 3 BIT 2 BIT 1 BIT 0 48 50 PIN IO bits 56 63 BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0 Status 3 LED 49 PGA Gain 0 7 50 Primary Gas Concentration 51 Secondary Gas Concentration 52 Calculated Gas Concentration 53 Primary Gas Average 54 Secondary Gas Average 55 Calculated Gas Average 56 Instrument Gain 57 Main Gas ID 58 Aux Gas ID 59 Decimal Places 60 Noise 61 Gas 1 Offset 62 Gas 3 Offset 63 Flow Temperature 64 Lamp Current 65 Digital Supply 66 Concentration Voltage 67 High Voltage 68 Ozonator 0 Off 1 On 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL 69 Control Loop 70 Diagnostic Mode 71 Gas Flow 72 Gas Pressure 73 Ambient Pressure 74 Analog Supply 75 Cell Temperature 76 Converter Temperature 77 Chassis Temperature 78 Manifold Temperature 79 Cooler Temperature 80 Mirror Temperature 81 Lamp Temperature 82 Lamp Temperature
18. Flow Control Block COOLER TEMP Temperature of the cooled PMT block Additional information on the SYSTEM TEMPERATURES screen is included in the EC9850 Service Manual 2 5 21 System Faults Screen SYSTEM FAULTS CELL TEMPERATURE SAMPLE GAS FLOW FLOW BLOCK TEMP A D INPUT 12 VOLT SUPPLY REFERENCE VOLTAGE LAMP CURRENT COOLER STATUS HIGH VOLTAGE Figure 2 28 System Faults Screen The SYSTEM FAULTS display provides a start pass or fail indication for various parameters which are continually monitored These parameters must be within acceptable operating ranges in order to display Pass If the instrument is in startup mode START will be displayed Additional information on the 5 5 FAULTS screen is included in the EC9850 Service Manual 2 6 Analog Output Analog output connections are described in section 2 1 2 1 above Before setting up the recorder and DAS analog outputs decide what offset and over ranging choices to make A brief explanation of these terms follows then the setup procedure is given The setting of the analog output and over range has no impact on the measurement range of the analyzer it only affects the analog output scaling The flow block temp is only present in the A series analyzer 98507600 Rev D 2 39 EC9850 SO ANALYZER OPERATION MANUAL 2 6 1 Offset and Live Zero At any selected output range the operator may want to observe negative signal in
19. O board include 0 to 10 V 0 to 5 V 0 to 1 V and 0 to 0 1 V The 50 pin I O board is optional in the EC9850 B series Data collection and recording is available for either a data acquisition system such as a datalogger or a strip chart recorder A DB50 connector is also included for digital input control and digital output status The EC9850 also features internal data storage capabilities The instrument includes an over range feature that when enabled automatically switches the analog output to a preselected higher range if the reading exceeds 90 of the nominal range When the reading returns to 80 of the nominal range the analyzer automatically returns to that range The US EPA has designated the EC9850 sulfur dioxide analyzer as an Equivalent Method Included in section 1 2 are the operational parameters necessary when using the analyzer in this mode 1 1 EC9850 SO ANALYZER OPERATION MANUAL 1 1 Specifications Note All specifications are referenced to STP standard temperature and pressure 1 1 1 Range o Display Autoranging 0 to 20 ppm Resolution 1 ppt selectable units and decimal places o Analog output 0 full scale from 0 0 05 ppm to 0 20 ppm with 0 5 and 10 offset o Autoranging between 2 user specified full scale values o U S EPA designated range Any full scale range between 0 0 05 ppm and 0 1 0 ppm 1 1 2 Noise RMS o Measurement process 0 25 ppb or 0 1 of concentration reading
20. PER eI Pao TRS gad ga eter etd deret VERE E 1 2 REZ Noise RMS ecuun deos vae ee doa 1 2 1 1 3 Lower Detectable Lintit cccccccccccccccsssssccccessssnsscecececsessnseaccececsessaeaecececeensauaecececsenseaecececcsensnsaaecececeenenanee 1 2 EIS 1 2 TJI SPAN ate iH de teer ep t ei a dece taies 1 2 PUN FOIT R O AEAEE NE E EE 1 3 1 1 7 Rise Fall Time 95 of Final eese ette teet entente nannten EE EES 1 3 LO TAN CAPIEY FTV OP s eode eee reali testet et ier ira EE 1 3 JO PFeCISIOW eS AOS E mamie duoi edt 1 3 TIO Sample Flow R te seit ge eter edle tate eda cesi eris 1 3 LIII Sample Pressure Dependent i t EU e b ee ee ta 1 3 EDAD Temperature Range e AR eR e A ew ce aha ya tl aise ces PR eR ee Ree ERI UR REA EARS 1 3 TEL D3 POWER 1 3 UR E 1 3 1l II5 Analog O tput erue aee edidi ra tee de e Fe ai PEU nent oe eR nee 1 3 Ded LOD QUAL OUND UE calc e e e o eene at o ttd t e ees 1 4 1 2 U S EQUIVALENT oves eerte e sive o epe deep 1 4 2 0 INSTALLATION AND
21. Rev D CHAPTER 4 DIGITAL COMMUNICATION Found New Hardware Wizard Welcome to the Found New Hardware Wizard This wizard helps you install software for Ecotech 9810 Analyser If your hardware came with an installation CD lt 6 or floppy disk insert it now What do you want the wizard to do Install the software automatically Recommended Install from a list or specific location Advanced Click Next to continue Next gt Cancel Figure 4 3 Screenshot of menu which appears when USB is connected 4 Insert the CD containing the Ecotech 9800 Analyzer Driver into the CD drive The computer should recognize the CD and continue with the installation after a few seconds If it does not click the Next button after loading the CD Note A dialog box similar to that in Figure 4 4 may appear If it does click the Continue Anyway button Hardware Installation The software you are installing for this hardware Ecotech 9800 Analyser has not passed Windows Logo testing to verify its compatibility with Windows XP Tell me why this testing is important Continuing your installation of this software may impair or destabilize the correct operation of your system either immediately or in the future Microsoft strongly recommends that you stop this installation now and contact the hardware vendor for software that has passed Windows Logo testing Continue Anyway STOP Installation Figure 4
22. SEDRVRI 48 Spare Driver 1 AGND 1 14 16 Ground reference for analog outputs DGND 12 PGND 13 34 Ground reference for digital inputs or outputs CGND 49 Chassis ground 12V 50 12 50 mA maximum 3 4 Unused 4 1 1 1 Notes 1 All outputs are open collector active LOW 2 Analog output 3 is unused 3 Analog output 1 is instantaneous gas concentration 4 Analog output 2 is unused 5 Not valid 98507600 Rev D 4 3 EC9850 SO ANALYZER OPERATION MANUAL 6 An error is flagged if the lamp current is below 20 mA or above 50 mA 7 Not valid 8 Span ratio out of range is defined as calibration gain changing below 75 or above 12546 gain change 9 Not used 10 An error is flagged if the 12 volt supply voltage is below 11 1 volts or greater than 14 3 volts 11 An error is flagged if the high voltage reading differs by more than 25 of the expected value as determined from the high voltage pot setting 12 This signal is the logical OR of FLOWFAIL LAMPFAIL CHOPFAIL CVFAIL COOLERFAIL HEATERFAIL REFFAIL PS FAIL and HV FAIL 4 1 2 50 Pin Inputs The DOZERO and posPaN controls pins 5 and 6 are TTL compatible inputs with internal 4 7K ohm pull up resistors These inputs are active low and can be driven to ground by dry contact relays open collectors or TTL compatible ICs The logic levels for control inputs are standard TTL levels They are low 0 8 V 2V high 5 V 4 1 3
23. SPAN 3 Allow the analyzer to sample the span gas until a stable reading is obtained typically 15 minutes 4 Verify this stable reading against the know calibration concentration 5 Typically if it is within 5 then a calibration is not required 6 Ifa calibration is required continue with the following procedure in section 2 4 4 If not return the CAL MODE tO MEASURE 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION 2 4 4 Analyzer Calibration Instructions 98507600 Rev D Note This procedure is a quick guide to single point span calibration of the EC9850 analyzer For complete gas preparation multipoint calibration instructions please refer to the multipoint calibration information in Chapter 3 Calibration With a stable supply of calibration gas connected to the Auxiliary port of the analyzer verify that in the CALIBRATION MENU CALIBRATION is set to MANUAL and CAL MODE to SPAN From the primary screen start the calibration sequence by pressing either the Up or Down arrow A or Y until the display prompts START MANUAL CALIBRATION Pressing the lt Select gt key will allow you to choose from No SPAN or ZERO Confirm that the display reads span and press Enter 4 A backlit cursor will be displayed on the SO concentration display Use the Select key to move the position of the backlit cursor and the Up and Down arrow keys to increment and decrement the value of the back
24. SPAN BACK INTERVAL HTGH VOTTAGR CALIBRATION SO2 CAL PRESSURE SPAN COMP S02 SPAN RATIO PREPROCESSOR POTS FLOW CONTROL ZERO BACKGROUND FLOW CONTROL POTS FAN SPEED CONTROL BACK INTERVAL VALVE TEST MENU PUMP SPEED FINE 502 CAL PRESSURE PUMP SPEED COARSE MULTIDROP PORT TEST GAS FLOW OUTPUT TEST MENU WATCHDOG TEST GAS PRESSURE DIAGNOSTIC MENU DISPLAY TEST CALCULATION FACTORS ANALOG OUTPUT EASUREMENT GAIN INT VALVE PRES TEMP FLOW COMP INSTRUMENT GAIN DN D MAIN INSTRUMENT MENU DIAGNOSTIC MODE CORRECTION Se CONTROL LOOP BACKGROUND AUX VALVE ZERO OFFSET EXT MEASURE Lp CALIBRATION MENU RANGE EXT ZERO GAS ANALOG OUTPUT MENU OUTPUT TYPE EXT SPAN GAS DATA LOGGING MENU OFFSET VALVE SEQUENCING ERASE LOGGED DATA TESEOMENU IAIN GAS ID FULL SCALE INTERFACE MODE ZERO ADJUST INTERFACE MENU ULTIDROP BAUD OVER RANGE DATA LENGTH OVER RANGING STOP BITS TREND SELECT MENU AVERAGE DATA bora EVENT LOG NPTWABDE ANADTAP TOTAL CE Ber ES CONCENTRATION AVG CONCENTRATION SAMPLE FLOW INSTRUMENT STATUS REFERENCE GAS FLOW SYSTEM TEMPERATURES GAS PRESSURE ZERO OFFSET mA REF VOLTAGE SPAN COMPENSATION CONC VOLTAGE ANALOG SUPPLY SAMPLE PRESSURE DIGITAL SUPPLY HIGH VOLTAGE LAMP CURRENT VERSIO CELL TEMP CELL TEMPERATURE CHASSIS TEMP SAMPLE GAS FLOW FLOW TEMP FLOW BLOCK TEMP COOLER TEMP REFERENCE VOLTAGE A D INPUT 12 VOLT SUPPL
25. T T 0 100 200 300 400 500 600 Calculated Figure 3 2 Data point formatting menu with Excel chart 3 9 EC9850 SO ANALYZER OPERATION MANUAL Add Trendline 2 x Type Options Trendline name 6 Automatic Linear Actual C Custom Forecast Forward fo 44 Units Backward fo zi Units Set intercept 7 Di Figure 3 3 Trendline menu in Excel Select Add Trendline Linear should be selected under the TYPE tab and enter Options Select the tick boxes that will display the equation and the R value on the chart Figure 3 3 Clicking OK will return to the chart and display the required data necessary to determine the effectiveness of the calibration Calculated Measured D 1 MULTIPOINT CHECK 100 99 200 201 300 300 400 400 500 503 Offset y 1 0043x 0 0952 R 0 9999 Measured Calculated Figure 3 4 Excel chart showing equation with required criteria Using the linear regression equation y mx b from the chart where y instrument concentration ppm x calculated value ppm m gradient gain b z y intercept offset 98507600 Rev D CHAPTER 3 CALIBRATION Note The R value is a correlation factor that relates to the similarity between the data points Values close to 1 indicate a linear relationship whereas a value close to zero will show a random distribution of data The calibration is accepted if o The gradient m falls between 0 98 and 1
26. Up 2 Down arrow Down 6 Right arrow Select Format REMOTE DEVICE I D gt TERMINATOR Device response lt ACK gt then clearscreen then menu display 98507600 Rev D 98507600 CHAPTER 4 DIGITAL COMMUNICATION Command RESET Function Reboots the instrument software reset Format RESET DEVICE I D gt TERMINATOR Device response lt gt Command GETDATA Function Used to collect logged data from an analyzer Format This command takes two different formats depending on the transmission state TO begin with the following format must be used GETDATA DEVICE I D gt 2 1 lt START TIME END TIME DATA TYPE gt TERMINATOR Where START TIME is the date time of the first piece of data to collect and END TIME is the date time of the final data to collect Both must be in the following format YY MM DD SPACE HH NN If END TIME is omitted then all data since START TIME is returned Year must be 03 or greater Where DATA TYPE I to only receive instantaneous logged data A to only receive Averaged logged data EMPTY to receive both instantaneous and averaged logged data After the request has been issued data will be returned in the same packet format as is documented for USB data requests After each packet the following com mand should be i
27. amp ecotech environmental moni itoring solutions EC9850 A amp B Series Sulfur Dioxide Analyser User Manual www ecotech com FRONT MATTER EC 9850 Quick Start Guide Step 1 Installation e Inspect analyzer for damage before turning on Service Manual 1 1 e Select an appropriate location Operation Manual 2 1 1 e Connect Gas lines Operation Manual 2 1 2 2 e Connect Analog Output Cables Operation Manual 2 1 2 1 e Connect RS232 Cables Operation Manual 4 2 1 e Check the mains power selection switch 115 or 230 VAC Operation Manual 2 2 Connect AC Mains Power Step 2 Start up Operation Manual 2 2 Set Service Switches Turn On power Service Manual 1 1 2 Operation Manual 2 2 e The Display should read 9850 SO2 Analyzer e Adjust the Display Contrast if required Operation Manual 2 2 1 e Verify that the software is running by observing the Ecotech Globe rotating in the bottom left hand corner of the display step 3 Operation Verify Instrument warm up and operation mode Service Manual 2 2 e Set the correct time and date Operation Manual 2 3 3 If using RS232 configure the Interface menu Operation Manual 2 5 12 e Check SYSTEM FAULTS menu All PASS Operation Manual 2 5 21 Verify other menu settings Service Manual 4 2 Step 4 Calibration Perform a quick single point calibration Operation Manual 2 4 e Setup and Calibrat
28. be used See output connections information in Chapter 4 MAIN GAS ID The ID address of the analyzer when Multidrop RS232 communications is used 98507600 Rev D 2 31 EC9850 SO ANALYZER OPERATION MANUAL INTERFACE MODE This establishes the RS232 communication mode Choices are COMMAND and TERMINAL T ERMINAL uses the menu structure and COMMAND uses the 9800 Serial Command Set MULTIDROP BAUD The communication rate for RS232 DB9 connectors on the rear panel The available rates are 1200 2400 4800 9600 19200 and 38400 DATA LENGTH Sets the number of data bits for the RS232 port The available choices are 7 BITS and 8 BITS STOP BITS Sets the number of stop bits for the RS232 port The available choices are 1 BIT and 2 BITS PARITY Sets the parity for the RS232 port The available choices are NONE EN and ODD zd COMM PROTOCOL Sets the communication protocol for serial transmissions using the 9800 Serial Command Set The available choices are ORIGINAL BAVARIAN and ENHANCED See the serial communications information in Chapter 4 2 5 14 Analog Output Menu The ANALOG OUTPUT MENU contains settings that relate to the recording devices For a detailed explanation of the analog output see section 2 6 The setting of analog output ranges has no impact on the measurement range of the analyzer it only affects the analog output scaling 2 5 14 1 SO Current Outpu
29. command USER ID byte bit 4 Used in conjunction with the PINID command USER ID byte bit 5 Used in conjunction with the PINID command USER ID byte bit 6 Used in conjunction with the PINID command USER ID byte bit 7 Used in conjunction with the PINID command USER ID byte bit 8 Used in conjunction with the PINID command Active output indicates that the sample flow is less than 0 1 slpm Active output indicates that the lamp has failed Note 6 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION Signal Name Number Function CHOEERID 37 Active output indicates that the chopper has failed Note 7 SPAN_OOR 38 Active output indicates that the span ratio is out of range Note 8 SPAREOCI 39 Spare open collector output 1 40 Active output indicates that a system heater has failed Note 9 SPAREOC2 41 Spare open collector output 2 PIES 42 Active output indicates that the system has been put into the Optic Test mode BCT SI 43 Active output indicates that the system has been put into the Electric Test mode bert 44 Active output indicates that the 12 volt supply voltage has gone out of range Note 10 HV PATI 45 Active output indicates that the PMT high voltage supply has failed Note 11 SYSFAIL 46 The sum of all failures in the instrument Note 12 POWER_ON 47 Active output indicates that power to the analyzer is on
30. enter the Interface Menu Click up or down until Interface Mode is highlighted 98507600 Rev D 4 29 EC9850 SO ANALYZER OPERATION MANUAL 4 30 6 Click select 7 Click up or down to change the interface mode to Command 4 6 3 Settings Open the Settings dialog box by either clicking the button choosing the Comm Settings menu option or by pressing F2 Click on one of the icons on the left of the dialog box to access that tab 4 6 3 1 Output Tab This function sets the options for the text file the program downloads data to Output file Enter the path and filename of the text file that the EC9500 Communicator will write acquired data to Clear the text box if a text file is not required If the file exists Choose Append to have the data added to the end of an existing file choose overwrite to have an existing file overwritten or choose Prompt to have the user prompted before writing to an existing file Date format Choose the date time format or the user can select their own in which to record the date and time of the analyzer data 4 6 3 2 Connection Tab This function sets the options for the communication connection between the computer and the analyzer Connection type Choose the type of connection to communicate with the Direct Serial Connection or Modem Connection or USB The choice changes the other options available in this tab 4 6 3 2 1 Direct Serial Connection Port Choose the COM port on th
31. heat transfer medium circulated around the device housing eg a jacketed condenser West or Liebig type The flushing Zero air or nitrogen passes through a heat exchanger e g a coil of copper tubing 98507600 Rev D CHAPTER 3 CALIBRATION contained in the constant temperature chamber before passing over the device to adjust its temperature to that of the device 3 2 3 3 Reagents 3 2 3 3 1 Dilution Gas Zero Air A regulated source of clean dry zero air for diluting the SO gas effluent from the permeation device Use a source that is capable of providing air flow up to approximately 20 slpm 3 2 3 3 2 SO Permeation Device The permeation device should be traceable to an SRM or CRM An SRM SO permeation device such as SRM 1626 from NIST is suitable for the 0 500 ppm range of this analyzer The SO from a permeation device is used as calibration source according to the following formulas PxKy _ 0 382P F Fr SO Jour Equation 3 2 Where Ky 0 382 ul ug for SO at 25 C and 760 Torr total flow rate of gas after the mixing chamber F F in slpm P permeation rate of device in ug minute SO ou 4 concentration in total flow rate stream in ppm The information is used in the following example P 0 820 ug minute at 30 C If you want a concentration of 400 ppb 0 400 ppm then solve the following 0 382xP 0 382 0 820 SO Jour 0 400 0 783 slpm T
32. in a numerical field o Select Selects the menu choice or selects the field for input g Pg Up Moves the cursor to the previous page or screen 0 Exit Leaves a field without making a change or returns the cursor to the main screen 0 Enter 1 Confirms a menu item or a field selection to the microprocessor 98507600 Rev D 2 11 9850 SO ANALYZER OPERATION MANUAL 2 3 2 Using the Menu and Making Entries The EC9850 analyzer is programmed with a series of menus that allow the operator to view parameters such as those generated by the microprocessor or to enter digital parameters when appropriate or to select from among the choices displayed The cursor is displayed as a movable highlighted area of text Letters appear as the opposite of the rest of the text on the screen 2 3 2 1 Screen Fields Screen fields that allow input are of two types o Choice fields Contain a fixed series of choices in a wraparound scrolling format o Digit fields Fields of programmable digital parameters in either wraparound scrolling or non wraparound scrolling format To select from among the choices in a choice field first press the lt Select gt key to designate the field then use the Up and Down arrows to highlight the desired selection When the desired selection is displayed press the lt Enter gt key to confirm the entry To set digits in a digit field first press the lt Select gt key to designate the
33. network adapter to load the files get into the command window and use the following instruction tftp i instrument IP address PUT file path gt cobox rom X2 Wait for 20 seconds and then use the following instruction to load the COB file tftp i instrument IP address PUT file path gt Ecotech cob WEBI Notice that If you do not have a TFTP client there is a demo TFTP client applica tion available at www weird solutions com A Users Guide is included and can be used as reference for using the weird solutions tftp client Use the following parameters e Use the instruments IP address for the TFTP server e Select Upload for the Operation e Select Binary for the Format e The path and name of the cob file e For the Remote File Name use X2 e Click the Upload Now button to start the operation Note Be very careful to set the TFTP application to do a binary transfer when upgrading over the network 4 36 98507600 Rev D INDEX 5 50 Pin Connector Board with Sample Choices Rear illustration eese 2 4 9 9800 command 4 7 4 16 A Analog Output Menu eee 2 32 2 34 Analog Output Menu illustration 2 34 Analog Output Menu Current Output illustration 2 32 Analog Output Menu Voltage Output illustration 2 33 Analyzer Front Panel illustration 2 9 Analyzer Keyboard
34. of the digital pots when DISABLED is selected the microprocessor does not control the digital pots and the user can manually adjust the digital pots When CONTROL LOOP is ENABLED the microprocessor will take control of the pots at the point at which the pots were last set CONTROL LOOPS will be set to ENABLED when the primary screen is displayed 98507600 Rev D 2 25 EC9850 SO ANALYZER OPERATION MANUAL TEST MEASURE Software controlled pot that is used by technicians when troubleshooting or verifying correct instrument performance This option only appears when the diagnostic mode is setto OPTIC ELECT or PREAMP 502 Gas concentration reading during diagnostics This option only appears when the diagnostic mode is setto OPTIC ELECT or PREAMP 2 5 7 Output Test Menu OUTPUT TEST MENU PREPROCESSOR POTS FLOW CONTROL POTS VALVE TEST MENU The OUTPUT TEST M Figure 2 14 Output Test Menu The FLOW CONTROL ENU reports readings for digital potentiometers and valves POTS menu does not appear on the B series analyzer 2 5 8 Preprocessor Pots Screen PREPROCESSOR POTS MEASURE COARSE ZERO REFERENCE ZERO MEASURE GAIN REFERENCE GAIN TEST HIGH MEASURE VOLTAGE ADJUST ADJUST VOLTAGE 4 016 VOLTS 0 400 PPM VOLTAGE 2 327 VOLTS VOLTAGE 700 VOLTS Figure 2 15 Preprocessor Pots Screen Preprocessor pots are electronically controll
35. operating ranges in order to display PAss 2 5 3 Instrument Menu INSTRUMENT MENU MEASUREMENT MENU DATE 15 AUG 03 TIME 18 57 PASSWORD UNLOCKED ERASE MEMORY NO MAIN GAS ID 050 Figure 2 9 Instrument Menu 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION The items in the INSTRUMENT MENU address instrument settings needed to initiate operation DATE The date format is day month year TIME Set in 24 hour format Setting the time resets the seconds internally to zero for synchronization with an external clock PASSWORD See section 2 7 ERASE MEMORY Memory can be erased in two different ways either RAM which does not wipe some settings or SETTINGS which will erase everything and reset all settings to default If you do not wish to erase all setting select NO when the following message is displayed THIS WILL ERASE SYSTEM GAINS ARE YOU SURE NO The word xo is highlighted in this warning Scrolling to vEs and pressing Enter will erase the memory in the analyzer Caution If the analyzer memory is erased all user configured parameters will return to their default values In addition all instrument calibration will be lost so the analyzer will have to be fully recalibrated This feature is provided for service and for preliminary configuration purposes Please do not choose this selection during normal operation MAIN GAS ID The ID address of
36. temperature TEMP2 converter temperature TEMP3 chassis temperature TEMP4 flow temperature TEMP5 cooler temperature TEMP6 mirror temperature TEMP7 lamp temperature TEMP8 ozone generator lamp temperature TEMP9 IZS temperature TEMP10 manifold temperature 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION All temperatures are in floating point format The temperature field is null if it does not apply to the analyzer type Command DTREND Function Dumps the requested trend buffer to the serial port Format DTREND DEVICE I D gt 1 1 lt PARAMETER gt TERMINATOR where PARAMETER GASCONC for the last 100 instantaneous SO2 readings GASAVG for the last 100 averaged SO2 readings PRESSURE for the last 100 cell pressure readings Low for the last 100 flow readings EF for the last 100 reference readings PANCMP for the last 100 span compensation readings from AZS cycles ERO for the last 100 zero readings from AZS cycles A Device response INDEX lt SPACE gt PARAMETER lt CR gt lt LF gt where INDEX is the index into the trend buffer 0 oldest point Formatted as an integer PARAMETER is the requested data in floating point format Command DZERO Function Commands the unit under test to enter the zero mode and stay there Format DZERO DEVICE I D gt TERMINATOR
37. the flow control board FLOW CONTROL POTS FLOW CONTROL ZERO FAN SPEED CONTROL PUMP SPEED FINE PUMP SPEED COARSE GAS FLOW 0 48 SLPM GAS PRESSURE 585 6 TORR Figure 2 16 Flow Control Pots Screen FLOW CONTROL ZERO A pot that sets electronic zero for differential flow The pot value is stored in EEPROM on flow controller board FAN SPEED CONTROL Software controlled pot that sets chassis fan speed PUMP SPEED FINE Software controlled pot that controls the pump speed which controls sample flow rate PUMP SPEED COARSE Software controlled pot that controls the pump speed which controls sample flow rate GAS FLOW Instrument generated information from the flow controller GAS PRESSURE Atmospheric pressure instrument generated information 2 28 98507600 Rev D 2 5 10 98507600 Valve Test Menu CHAPTER 2 INSTALLATION AND OPERATION VALVE TEST MENU INT VALVE 1 OPEN INT VALVE 2 CLOSED INT VALVE 3 CLOSED AUX VALVE 2 CLOSED EXT MEASURE CLOSED EXT ZERO GAS CLOSED EXT SPAN GAS CLOSED VALVE SEQUENCING ON Figure 2 17 Valve Test Menu The vALVE TEST M ENU allows the valves to be set oPEN or CLOSED according to the operator s choice To manually operate the valves VALVE SEQUENCING needs to be turned off See the pneumatic diagram in the EC9850 Service Manual for details on plumbing INT VALVE 1 Sample stream INT
38. the following may also be used NATA NIST Traceable Temperature sensor o NATA NIST Traceable Barometric sensor o NATA NIST Traceable SO Gas cylinder with dual stage regulator o Laptop Computer with Microsoft Excel o Bios DryCal flow calibrator o Ecotech portable zero air source o Ecotech portable Dilution Gas Calibrator 3 7 EC9850 SO ANALYZER OPERATION MANUAL 3 8 3 2 2 3 Procedure Connect the flowmeter to the analyzer sample inlet and measure the sample flow reading Record the average of flow readings and ensure that it meets the specifications of section 1 1 10 1 Setup the dilution gas calibrator to obtain a SO gas concentration of 80 of the EC9850 s operating range as discussed in section 3 2 1 4 Perform a manual calibration at this point Note Record the analyzer instruments gains before and after the calibration Setup the dilution gas calibrator to obtain Zero air and ensure that the EC9850 has a zero reading lower than 5ppb NOTE Perform a background otherwise and repeat step 2 Note Do not make any further span adjustments during the remainder of the calibration Record the new span values in the spreadsheet Generate 5 additional concentrations by decreasing Fso or increasing Fp sure the total flow exceeds the analyzer s total flow demand For each concentration generated calculate the exact SO concentration using Equation 3 1 Note The recommended multipoint span concentrations a
39. this manual 98507600 Rev D 3 EC9850 SO ANALYZER OPERATION MANUAL Internationally Recognized Symbols Used on Ecotech Equipment IEC 60417 No 5016 IEC 60417 No 5017 IEC 60417 No 5021 IEC 60417 No 5032 a 22 60417 5041 5 7000 0434 5 3864 3 6 Electrical fuse Earth ground terminal Equipotentiality Alternating current Caution hot surface Caution refer to accompanying documents Caution risk of electric shock 98507600 Rev D FRONT MATTER Safety Requirements g To reduce risk of personal injury caused by electrical shock follow all safety notices and warnings in this documentation This equipment should always be used with a protective earth installed The EC9850 is compliant with the requirements of EN61010 1 A2 1995 Safety Requirements for Equipment for Measurement Control and Labo ratory Use If the equipment is used for purposes not specified by the manufacturer the protection provided by this equipment may be impaired Replacement of any part should only be carried out by qualified personnel only using parts specified by the manufacturer Always disconnect power source before removing or replacing any components Equipment Rating g g g g g 100 120 220 240V 10 50 60 Hz 250 VA max FUSE 5 3 15A T 250V wiring must be in accordance with local norms and be carried out by experienced personnel Env
40. 0 R Recorder connections 2 2 S Sample gas connections eee 2 6 Serial command control sess 4 7 Serial connections essent 4 5 Serial control Ree en 4 5 Serial Interface Connection Diagrams illustration 4 6 Serial Port ee gero eer 2 30 2 31 4 7 Serial terminal control 4 7 Setting the date and time 2 13 SO2 Analyzer Calibration Using Cylinder Gas Dilution illustration 3 4 SO2 Calibration Using a Permeation Tube illustration ee 3 12 SO2 Current Output Menu 2 32 502 Voltage Output 2 33 Specifications 1 2 neg eR gon 1 2 Status Output Connections illustration 4 5 Strip Charts Illustrating Offset illustration 2 40 System Faults screen 2 39 System Faults Screen 2 39 System Temperatures screen 2 38 System Temperatures Screen illustration 2 38 T Terminal control serial sees 4 7 Test Menu tree ote Eee ts 2 25 Test Menu 2 25 Timed Calibration 2 22 Timed Calibration Menu illustration 2 22 U USB 4 1
41. 02 o The intercept b lies between 0 3 o The correlation is greater than 0 9995 Reject the calibration if the above criteria are not met If the calibration fails perform a leak check check zero air scrubbers and consult the EC9850 service manual for troubleshooting assistance 3 2 3 Procedure Using the Internal Permeation Device 3 2 3 1 Principle In a permeation device an easily liquefied gas such as SO is condensed inside an inert container all or part of which is constructed from a polymeric material often Teflon Gas escapes from the container by dissolving in and permeating through the polymer walls at a temperature dependent rate The rate of gas effusion in ug minute at a constant temperature can be established by gravimetric determination of the weight loss of the permeation device over a known period of time In this calibration procedure accurately known concentrations of SO are produced dynamically by diluting the effusion from an SO permeation device with various flow rates of zero air 3 2 3 2 Apparatus A diagram of a typical permeation calibration system is seen in Figure 3 5 In addition to the components used in the cylinder gas dilution method a permeation calibration system uses a constant temperature chamber a controlled temperature housing that contains the SO permeation device The chamber is flushed continuously with purified dry zero air or nitrogen 98507600 Rev D 3 11 EC9850 SO
42. 200 to 240 VAC 50 Hz 1 1 14 Weight 25 kg 55 1 1b A series 21 3 kg 47 Ib B series 1 1 15 Analog Output Menu selectable current output of 0 20 mA 2 20 mA and 4 20 mA 98507600 Rev D 1 3 EC9850 SO ANALYZER OPERATION MANUAL o Jumper selectable voltage output of 100 mV 1 V 5 V and 10 V with menu selectable zero offset of 0 5 or 10 The 9850B requires an op tional 50 pin I O board to enable voltage output 1 1 16 Digital Output g g g g Multidrop RS232 port shared between analyzers for data status and con trol Service RS232 port gives front panel access to a local or remote user USB port connection on the rear panel provides data transfer and control DB50 with discrete status user control and analog output 1 2 U S EPA Equivalent Method The EC9850 series sulfur dioxide analyzers are designated under U S EPA regulations as equivalent method EQSA 0193 092 Using the EC9850 under U S EPA designation as an equivalent method as defined in 40 CFR Part 53 requires operation under the following conditions Range Any full scale range from 0 to 0 050 ppm and 0 to 1 0 ppm Ambient temperature 15 to 35 C Line voltage 105 to 125 VAC 60 Hz or 200 to 240 VAC 50 Hz Flow rate 0 50 SLPM A series Flow rate 0 65 SLPM B series Pump Internal Pump A Series Ecotech external pump or equivalent B Series see section 2 1 2 3 Filter A 5 micron PTFE filter must be ins
43. 3 3 3 2 2 Procedure For 5 Point Multipoint Calibration eese eene nennen 3 7 3 2 3 Procedure Using the Internal Permeation Device eese nennen 3 11 3 3 CALIBRATION REQUIREMENTS WHEN OVER RANGING IS EMPLOYED eene ener 3 14 3 4 AUTOMATIC ZERO SPAN CHECKS AZS cccscccesssscecsessececesseecesaececsesaececsceeceeaeeecsesaececseseeeeeaeeecsesaeeeesenaeeess 3 15 3 4 1 U S Definitions iiit t HERE santa eei eE Eae EEEN ar e Ensi NE STE E Saai er ena en 3 15 2 4 2 Outline e ess tede tti pe ttu eta dt ep egt nds e Rer mat ee bet 3 15 NI SPATIUM 3 17 3 4 4 Description of the AZS PrOCeSS iae ete rs ear ete sb FREE 3 19 3 5 CALIBRATION REFERENCES 5 3 19 4 0 DIGITAL COMMUNICATION eere esee entes entes sensa stesso seta sosta seta sse ta sens 4 1 4 DISCRETE CONTROL 75 ste oltra diste e Ce E D Ir DDR 4 1 4 1 1 50 Pin Functional Specification eese eene nnne ener ener nnne eterne nnne 4 1 4 1 2 50 Pin VO Inp ts iiis cut ite eite ARIES FG E ETE Eee EU nei 4 4 4A SSO VO Outputs EE 4 4 4 2 SERIAL CONTROL eere Re aan e n ee a on en umma P 4 5 4 2 1 Serial Connect ns EA S AE SREK EAE ET TORNE 4 5 4 2 2 CADE CONNECTIONS EE 4 6
44. 4 Dialog Box which may appear during installation 98507600 Rev D 4 25 EC9850 SO ANALYZER OPERATION MANUAL 5 installation should now proceed When complete click the Finish button The driver installation is now complete 4 6 EC9800 Communicator Software The EC9800 Communicator software is supplied on CD with the EC9850 series analyzer and allows the user to communicate with the analyzer by direct serial connection modem or USB The two functions of the program are to e Download recorded data Data Acquire mode e Remotely access the analyzer s control panel Remote Terminal mode To set the 9800 Communicator s output connection and analyzer properties use the settings dialog box Refer to section 4 6 3 4 6 1 Data Acquire Mode Data Acquire mode enables the user to download recorded data from the analyzer to a text file 4 6 1 1 Using Data Acquire Mode 1 Ensure that all Settings are correct Refer to section 4 6 3 2 Under the Mode menu tick the Data Acquire option 3 Onthe Comm menu select Start 4 Inthe dialog box that appears enter the start date time for the data in dd mm yy hh mm format For example enter 30 11 2003 14 20 for 2 20PM on 30 November 2003 5 In the dialog box that next appears enter the end date time in the same format The EC9800 Communicator will now retrieve the data To stop downloading before all data has been retrieved select Stop on the Comm menu 4 26 98507600 Re
45. 4 3 SERIAL TERMINAL CONTROE visas teo nme ino nmt E e DO t P pm cous rt peers 4 7 4 4 SERIAL COMMAND CONTROL eerte teer e RNC e o tree i E HER EE TIE IH ERE 4 7 24 1 9800 Command Set Format a e eee a E eet bbb etes a 4 7 4 4 2 Bavarian Network Command Set Format eese ener nnne nnne 4 8 4 4 3 Protocol Definition and Selection iino ie eene enne GA nnne trennen rennen ene 4 10 4 4 4 Establishing Communications eee eee eese eee enne nete eene trennen enne nete teen enne enne 4 13 44 5 Serial Command Sets e itte ute oi ee tere dee ea eeu eve 4 15 4 5 US B COMMUNICATION ciet trit ree rhet te iH he E Mu o T e ph eei ee 4 24 98507600 Rev D 9850 SO ANALYZER OPERATION MANUAL 4 5 1 Installing the Driver on d PC aa eda ener eie nee e rii ede 4 24 4 6 EC9800 COMMUNICATOR 4 26 461 Acquire 4 26 4 6 2 Remote Terminal Mode et abre ey Ea e ERE TY 4 28 26 338 CLINGS cse eti dedere ete cose eb erit e e erede ehm epe RIT 4 30 7 6 4 Keyboard SNOTICUIS TS aie e ete eet e e RU ei eee 4 31 4 7 NETWORK INTERFACE OPTIONAL esses 4 32 47 C
46. ANALYZER OPERATION MANUAL Figure 3 5 SO Calibration Using a Permeation Tube Temperature control is the primary concern in using an SO permeation device as a standard SO source For example a change in temperature of about 0 5 C effects a change in the permeation rate of the device of about 4 For this reason it is important that the temperature of the device be maintained at a constant value within 0 1 and that it be closely monitored when the device is in use Generally the SO permeation device is housed in a temperature controlled glass container that has an entrance and exit port at opposite ends a glass thermometer accurate to 30 05 C can be placed beside the device to monitor its temperature A small fixed zero air or nitrogen flow about 100 SCCM that is maintained at the same temperature as the permeation device flushes the SO out of the device housing into a mixing chamber where the SO is diluted with clean dry zero air A valve e g a three way stopcock placed at the exit of the device housing may be used to divert the SO stream to a vent when clean air is required at the manifold for making the necessary zero adjustments to the analyzer To maintain the temperature of the permeation device within 0 1 C of the desired value the device and housing may be placed either physically inside a constant temperature chamber as shown in Figure 3 5 or they can be located outside the constant temperature chamber with the
47. C9850 SO ANALYZER OPERATION MANUAL 4 16 is selected If the Bavarian protocol is selected the response will be 10 pad zeroes BCC1 first byte of the block check calculation BCC2 second byte of the block check calculation The block check calculation is performed by clearing the block check number An iterative EXCLUSIVE OR is performed on this number with every character in the message from the lt 5 gt to the lt ETx gt inclusive The resulting value is converted into a two digit pseudo hex number and sent out as Bcc1 and 2 Command PI Function Bavarian network command that sets the device ID and serial number of the analyzer Format lt STX gt PI lt DEVICE I D gt lt SP gt lt INSTRUMENT SERIAL NUMBER gt lt ETX gt lt BCC1 gt lt BCC2 gt Command ST Function Bavarian network command that sets the instrument mode to zero span or measure or runs a background cycle Format lt STX gt ST lt DEVICE I D gt lt SP gt COMMAND lt ETX gt lt BCC1 gt lt BCC2 gt where COMMAND M for measure N for zero for span s to start background cycle 4 4 5 2 9800 Command Set Note The TERMINATOR can be either a CR or lt LF gt The DEVICE I D Three Digit Instrument I D in ASCII Format Command ABORT 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION Function Commands the addressed device
48. D TERMINATOR Device response STAT1 STAT2 STAT3 STAT4 STAT5 STAT6 STAT7 STAT8 STAT9 STAT10 STAT11 STAT12 lt CR gt lt LF gt where gas flow ground offset ozone generator flow 5 5 2 gas pressure STAT3 reference voltage STAT4 concentration voltage STAT5 analog supply STAT6 digital supply STAT7 STATS 4 19 EC9850 SO ANALYZER OPERATION MANUAL 4 20 STAT9 high voltage STAT10 lamp current mA STAT11 ambient pressure STAT12 Startup flag 1 in startup mode Startup flag and ground offset are integers all other numbers are in floating point format The status field is null if it does not apply to the analyzer type Command DSPAN Function Commands the unit under test to enter the span mode and stay there Format DSPAN DEVICE I D gt TERMINATOR Device response ACK if the unit under test is able to perform the command lt NAK gt if not Command DTEMPS Function Dumps the SYSTEM TEMPERATURES MENU variables to the serial port Format DTEMPS DEVICE I D gt TERMINATOR Device response TEMP1 TEMP2 TEMP3 TEMP4 5 TEMP6 TEMP7 TEMP8 9 TEMP10 lt CR gt lt LF gt where TEMP1 cell
49. D CHAPTER 2 INSTALLATION AND OPERATION DECIMAL PLACES Set the number of decimal places in which the data is displayed on the screen 0 1 2 3 40f 5 Note The screen is able to display up to 7 characters of data including the decimal place for each reading AVERAGE PERIOD Set time in HOURS 1 4 8 12 Or 24 MINUTES 1 3 5 10 15 or 30 This establishes the period for average computations This field is a wraparound field FILTER TYPE Sets the time constant of the digital filter Choices are No FILTER 300 SECONDS 90 SECONDS 60 SECONDS 30 SECONDS 10 SECONDS Or KALMAN adaptive Note The Kalman filter is the factory default setting and must be used when using the instrument as a U S EPA equivalent method The Kalman filter also gives the best overall performance for this instrument NOISE The standard deviation of the concentration The manner in which this is done is as follows 1 Take a concentration value once every two minutes 2 Store 25 of these samples in a first in last out buffer 3 Every two minutes calculate the standard deviation of the current 25 samples This is a microprocessor generated field and cannot be set by the operator Note The noise reading is only valid if zero air or a steady concentration of span gas has been supplied to the analyzer for at least one hour ZERO OFFSET ZERO calibration correction factor User can manually set the offset between 10 00 P
50. D ERR 2 37 2 5 19 Instrument Status Screen EAE ies Eaa o seio des GISI ea to enne tete enne ben 2 37 2 5 20 System Temperatures 5 2 38 2 5 21 System Faults Screen t ei eee e REC EE SAKES ESETRE 2 39 2 6 ANALOG OUTPUT een reo nnne date de petat e ie eem ene 2 39 2 6 1 Offset and EiVe ZerQa 5 2 40 2 6 2 Over Range Adjustment siisii i irienn 2 40 2 6 3 Analog Output Calibration Procedure 2 42 2 6 4 Calibration Requir tments iste err d rH anny rep E HE PUER Eo re UE Cesena ns 2 42 2 1 PASSWORD PROTECTIONS eret ertet tee ter ere RENNES Ht Iesu 2 43 2 74 Rules of Operation eae ia e e e eh e Ug ER eR RR eens ER e ire ea ttd 2 43 2 7 2 SAMPLE Sessiohtui cis ce ipea d e ERE Ree RR Dee eT PER Eu ER Teen eet etd 2 43 Un UOn 3 1 3 3 1 3 1 1 Analyzer Calibration Instructions eee eese eee eene 3 2 3 2 MULTIPOINT CALIBRATION 5 2 oet Th roO AE b ete RE D rd e 3 3 3 2 1 Procedure Using Cylinder Gas Dilution Method esee eene
51. During the last 5 minutes the instrument takes readings every second and averages them to yield a value to be stored as the zero calibration value At the end of 8 minutes the zero air valve is closed and the span gas valve is opened admitting span gas for 8 minutes During the first 3 minutes no data is taken During the last 5 minutes the instrument takes readings every second and averages them to yield a value to be stored as the span calibration value If span COMP 18 ENABLED this is the value which is used to correct all subsequent readings to the calibration The zero air valve is switched on for 1 minute to purge the cell of span gas The Sample valve is then activated for an additional 6 minutes to allow the cell to come back to monitoring concentration At the end of 25 minutes monitoring resumes including putting data in the average etc The data averages are not updated during calibration 3 5 Calibration References 98507600 Rev D 1 Quality Assurance Handbook for Air Pollution Measurement Systems Volume IL Part 1 EPA 454 R 98 004 U S Environmental Protection Agency Environmental Monitoring Systems Laboratory Research Triangle Park NC 27711 1998 EC9850 SO ANALYZER OPERATION MANUAL This page is intentionally left blank 3 20 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION 4 0 Digital Communication The EC9850 has three methods of digital communication serial communication using RS232 si
52. E 2 17 Ze ZAM GI MON BEE pee me teu IR 2 18 2 3 3 Instrument EE 2 16 2 5 4 Measurement 22 2 20 2 599 Calibration Mens sed cese ORES aR RN 2 21 RON A KAI ANI LA AA PRENTE A E E EEE derer tod repe EE as Ga 2 25 98507600 Rev D FRONT MATTER 2 3 7 Output Test Menu isi eee eh v e epe E Ce dau eater HERE EE ore 2 26 2 23 0 Preprocessor Pots Screen so oai eee dite e e ER 2 26 2 5 9 Flow Control Pots Screen A series only cessisse eene tenete entente 2 28 2 20 Valve Fest MenM iiic eee err GTP EC RERO MEE EUR E Ee Ces 2 29 2 9 T Di enostic Menu 5s uod eerte e Reset opem ie t ae via AN e ta gens HERRERA 2 30 2 9 12 Calculation f CiOrs s iu eii e tette e e Ages pe e 2 31 2 09 13 Interface Menu is RR t ER Re A EREIGNIS Cu eene eR eth 2 31 2 25 14 Analog Output Menu e re eb Une Oeo day 2 32 2 345 DataLoeeine Mena at a rt e rcu ee RB 2 34 2 946 Network Adaptor Meriu s oe bo eet di e Ban E Roco blc p e CERE Re Se 2 36 Zio Trend Select Menton e eet te LRU ata aa EOE Bs ROE OR ER ERS 2 36 225216 Event Log Screens iso cs e eR er devon d d A R
53. GING DISABLED Figure 2 22 Analog Output Menu Voltage Output RANGE Set upper range limit in digits to desired SO concentration See section 2 6 This value cannot exceed the ovER RANGE value 98507600 Rev D 2 33 EC9850 SO ANALYZER OPERATION MANUAL 2 5 15 OUTPUT TYPE Setting must match the jumper settings on the 50 pin I O board if installed CURRENT VOLTAGE OFFSET Choices are 0 5 and 10 Recorder or DAS output will reflect this FULL SCALE X XX a correction factor for full scale setting Used when calibrating the analog outputs ZERO ADJUST X XX a correction factor for the zero setting Used when calibrating the analog outputs OVER RANGE Set to desired over range value This value cannot be set below the RANGE value See section 2 6 This is the alternate scale the recorder or DAS indicates when over ranging is active and enabled When 90 of the set range is reached this auto range is effective When 80 of the original range is reached it returns to the original range OVER RANGING Set to ENABLED Or DISABLED to turn the over ranging feature on or off Data Logging Menu DATA LOGGING MENU AVERAGE DATA OFF INST DATA OFF TOTAL CAPACITY 0 0 DAYS ERASE LOGGED DATA NO Figure 2 23 Data Logging Menu The DATA LOGGING MENU contains settings that relate to the internal data recording facilities of the EC9850 This data can latter be retrieved usin
54. IL E LAMPFAIL HEATERFAIL OUT OF SERVICE VOL FLOWFAIL COOLERFAIL HV FAIL SPANON SYSFAIL CVFAIL PS FAIL F CHOPFAIL REFFAIL ZEROON VOL LAMPFAIL HEATERFAIL OUT OF SERVICE SPANON FLOWFAIL COOLERFAIL HV FAIL SYSFAIL CVFAIL PS FAIL CHOPFAIL Indicates that the chopper has failed LAMPFAIL Indicates that the lamp has failed FLOWFAIL Indicates that the sample flow is less than 0 1 slpm SYSFAIL Indicates one or more components have failed HEATERFAIL Indicates that a system heater has failed COOLERFAIL Indicates that a cooler has failed CVFAIL Indicates that a converter has failed ZEROON Indicates that the instrument is in the Zero mode 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL OUT OF SERVICE Out of service switch has been activated on analyzer HV FAIL Indicates that the PMT high voltage supply has failed PS FAIL Indicates that the 12 volt supply voltage has gone out of range GRAV Measuring in gravimetric units i e MG M3 VOL Measuring in volumetric units i e PPM Example If a failure status is received as C022 then the failures of the instrument as determined by this code are C FLOWFAIL Indicates that the sample flow is less than 0 1 slpm SYSFAIL Indicates one or more components have failed 0 No Failure 2 2 OUT OF SERVICE Out of service switch has been activated on analyzer 2 2 VOL Measuring in volumetric units i e PPM 98507600 Rev D
55. Inst 03 12 01 10 12 0 001012 ppm 3 min M C200 Avg 1212 01 10 15 n neto nara Anni conn IET Download Complete COM1 Closed Figure 4 5 Acquired Data completion screenshot 2 The same data is displayed in the output text file as set on the Output tab of the Settings dialog box with the fields delimited by commas A description of each field follows below 4 27 EC9850 SO ANALYZER OPERATION MANUAL Field On screen In text file Date Time The date time in the format As for on screen selected in the Output tab of the Settings dialog box when the data in that row were recorded Data Up to three channels of analyzer As for on screen data with column headings as set by the analyzer Unit The unit for the analyzer data Codes representing the data units Period The repetition period For As for on screen with averaged data the repetition the period in minutes period is also the averaging period Function The function status of the analyzer As for on screen status at the time of measurement Failure The failure status of the analyzer As for on screen status at the time of measurement Type Inst instantaneous data I instantaneous data Avg averaged data A averaged data 4 6 2 Remote Terminal Mode Remote Terminal mode can be used to access the analyzer s control panel remotely 4 6 2 1 Starting a Remote Terminal Mode Session 1 Ensure that all Settings ar
56. O ANALYZER OPERATION MANUAL 2 4 Analyzer Calibration When the EC9850 analyzer is powered on for the first time the analyzer must be calibrated to ensure accurate SO measurements The analyzer does not require recalibration after further power interruptions or resets However if the instrument is transported to a new location or maintenance work is performed the instrument may require re calibration To determine weather the instrument requires a calibration a precision check can be performed as discussed in the following sections 2 4 1 Precision Checks A precision check is a Level 2 calibration as discussed in section 3 4 This means that the instrument is only checked against a know calibration source and is not adjusted A precision check can be performed either manually or automatically 2 4 2 Automatic Most modern air quality monitoring systems have data acquisition systems which can automatically initiate and record the results of a daily precision check The means by which the analyzer is externally controlled is via the 50 PIN IO connection or via the RS232 multidrop connection Refer to section 4 0 for more details on interfacing to these ports 2 4 3 Manual A manual precision check can be initiated as follows 1 Connect a source of span gas to the analyzer through the Auxiliary port see chapter3 for instructions on preparing calibration gas 2 From the CALIBRATION MENU set CALIBRATION tO MANUAL and CAL MODE to
57. OF MULTI DROP PORT TEST 4 4 4 1 Multidrop Communications 98507600 Rev D The term multidrop is a idiomatic contraction of the term multiple drops It is a term used to denote a parallel connection of multiple RS232 transceivers In this scheme all receivers share the same receive line that comes from a single master Likewise these multiple transceivers share the same transmit line which goes back to a single master This strategy is a method of attaching multiple slave units instruments to a single master computer In the multidrop strategy each unit is given an identification number L D which is sent with each command from the master When a unit recognizes its unique I D it processes the command and responds appropriately The integrity of this method relies on a strict enforcement of the following rules o Each unit in the multidrop must have a unique I D that is programmed into the unit before attaching to the network o After a command is sent by the master the master must then wait for a response Only after a reasonable time out period should the master send another command o The multidrop master must include a time out mechanism in the event that the I D sent with the command is garbled Clearly a lt gt on a bad I D is not possible for the units in this scheme 4 13 EC9850 SO ANALYZER OPERATION MANUAL 4 14 The master must correlate the unit response with I D sent in the command to kno
58. Operation Manual A amp B Series Revision History Rev Date Summary Affected Pages A Jan 2004 New Release for new Microprocessor Board amp series Combined Based on original manuals B Feb 2004 Changes to menu options and structure June 2005 Changed to updated EC manual format and updated safety information C 1 November Added Calculation factors menu 2 30 2005 Added C tick information updated CE information 3 D April 2007 Updated specifications language and links within All 98507600 Rev D pdf manual created EC9850 SO ANALYZER OPERATION MANUAL NOTE The photograph on the binder of this manual is of the south coast of Australia during Bushfires in 2003 The photograph is courtesy of Earth Sciences and Image Analysis Laboratory NASA Johnson Space Center Photo Reference ISS006 E 19897 Notice The information contained in this manual is subject to change without notice and does not represent a commitment on the part of the Ecotech Pty Ltd Ecotech reserves the right to make changes in construction design specifications and or procedures that may not be reflected in this manual Copyright 2007 rights reserved Reproduction in any form is prohibited without the written consent of Ecotech Pty Ltd This manual is furnished on the express condition that the information herein will not be used for second source procurement or purposes directly or indirectly detrimental to the inter
59. PB 2 5 5 Calibration Menu The CALIBRATION MENU contains entries used to choose calibration gases or to perform automatic zero span cycles The choice of TIMED or MANUAL calibration displays a slightly different screen TIMED calibration generates a zero span check 98507600 Rev D 2 21 EC9850 SO ANALYZER OPERATION MANUAL that occurs at a chosen interval without operator intervention MANUAL calibration allows for operator controlled calibration Only one choice TIMED Or MANUAL applies at any given time 2 5 5 1 Timed Calibration The following screen appears when CALIBRATION TIMED is selected CALIBRATION Designates CALIBRATION MENU CALIBRATION TIMER INTERVAL STARTING HOUR CYCLE TIME SO2 TIMED SPAN CALIBRATION SPAN COMP SO2 SPAN RATIO BACKGROUND BACK INTERVAL SO2 CAL PRESSURE TIMED 24 HOURS 0 15 MINS 10 000 PPM INTERNAL ENABLED 1 0000 ENABLED 24 HOURS 750 0 TORR Figure 2 11 Timed Calibration Menu TIMER INTERVAL The number of hours between the zero span checks STARTING HOUR The hour when the first zero span check will be performed CYCLE TIME ED Of MANUAL calibration control The period 1 to 59 minutes of the zero amp span steps during a timed calibration SO2 TIMED SPAN Digital setting of the span concentration the operator expects the instrument to read during an AZS cycle An entry is required only if SPAN COMP is ENABLED CALIBRATION
60. TION 2 1 Mechanical Installation Note Before installation the unit should be checked to ensure that the instrument arrived undamaged The 9850 Service Manual contains initial installation inspection instructions 2 1 1 Selecting a Location Select a location for the analyzer where temperature variation dust and moisture are minimal The location should be well ventilated and should allow convenient access to the operator controls and front panel display The analyzer can operate in a range of 5 to 40 C without risk of damage 2 1 1 1 Rack Mount or Enclosed Location The analyzer is supplied as a bench top version with rubber feet or with the chassis slides to convert it to a rack mount version The optional rack mount version is 24 61 cm deep and fits into a 19 48 3 cm RETMA instrumentation rack The front panel will protrude slightly Refer to the instructions provided with the rack mount kit for assembly into a rack Caution The rack mount version requires a properly ventilated rack enclosure The temperature inside enclosures that are not properly ventilated may rise as much as 15 C above the ambient air temperature This may force the analyzer to operate outside of specifications Optimum operation is obtained at operating temperature of 20 to 30 C inside the rack enclosure For ventilation calculations use a heat dissipation rating of 150 watts or 512 Btu per hour After the analyzer has been mounted
61. TION REFERENCE ZERO OFFSET SPAN COMPENSATION SAMPLE FLOW SAMPLE PRESSURE Figure 2 24 Trend Select Menu Each graph is displayed as an x y plot with the x axis zero being the current time and the most distant number being the most historic data 2 36 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION 2 5 18 Event Log Screen EVENT LOG 1 BACKGROUND CYCLE OCCURRED AT 00 01 15 0 03 2 ZERO FLOW OCCURRED AT 17 02 08 AUG 03 3 BACKGROUND CYCLE OCCURRED AT 17 02 07 AUG 03 4 Figure 2 25 Event Log Screen The EVENT Loc displays notations of key events such as autozero and calibration or specific error conditions for up to 100 occurrences This screen is a first in last out type screen The first entry is the latest occurrence You can scroll through the events using the Up or Down arrow keys A or 2 5 19 Instrument Status Screen INSTRUMENT STATUS GAS FLOW 0 65 SLPM GAS PRESSURE 586 6 TORR REF VOLTAGE 2 501 VOLTS CONC VOLTAGE 2 237 VOLTS ANALOG SUPPLY 11 909 VOLTS DIGITAL SUPPLY 4 977 VOLTS HIGH VOLTAGE 700 VOLTS LAMP CURRENT 34 794 VERSION 1 03 0005 EXIT Figure 2 26 Instrument Status Screen INSTRUMENT STATUS is information continuously generated by microprocessor for various parameters GAS FLOW Calculated gas flow Will indicate 0 00 if the flow transducer senses flow has gone to Zero GAS PRESSURE Current Gas pressure shoul
62. Y LAMP CURRENT COOLER STATUS HIGH VOLTAGE Figure 2 6 Menu Structure 2 16 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION Note The values shown in the illustrations are examples only Your display will be affected by the settings you choose 2 5 1 Primary Screen 98507600 Rev D GAS SO X XXX PPM AVG INSTRUMENT GAIN 1 000 OPERATIONAL MODE INSTRUMENT FAULT 12 14 15 AUG 03 MAIN MENU Figure 2 7 Primary Screen When power is applied the screen displays the Ecotech logo for a few seconds It then identifies the analyzer and the notation MAIN MENU appears in the lower right corner In the lower left hand corner there is the Ecotech Globe rotating indicating that the program is running After the warmup period the operation mode is designated at the left of the screen and the current gas measurements for the analyzer are indicated as shown in Figure 2 7 Instrument faults will be reported on the status line which appears one line below the instrument state display The following rules apply governing the information displayed on this line If there are no failures the status line is blank If there is a single failure that failure is displayed on the status line i e ZERO FLOW HEATER FAULT etc The status line will clear when the fault clears If there are multiple failures the failure at the top of the failure list will be displayed on the status line When this failure cle
63. ainst the actual parameters The number of data parameters must be sent on 9800 style commands but it is not checked against the actual parameters 4 4 3 9 Enhanced Protocol This protocol is provided to allow easier and more robust interfacing between 9800B instruments and a computer Every command with a valid I D will respond with either ACK or lt NAK gt Bavarian commands also respond with either lt ACK gt Or lt NAK gt although this is outside the normal Bavarian Network protocol Note This protocol selection should not be used in Bavarian network applications 4 4 3 10 Command Acknowledgment g For all valid 9800 and Bavarian commands an ASCII lt acx gt character is returned For commands that request data the data will be sent after ack char acter 4 4 3 11 Negative Command Acknowledgment g Any detected error will respond with the ASCII xax character followed by an error message Due to the constraints of the multidrop environment the unit I D must be received intact for a NAK response to be sent An invalid command will cause the response NAK UNKNOWN COMMAND lt CR gt lt LF gt An invalid command format will cause the response lt NAK gt BAD COMMAND FORMAT lt CR gt lt LF gt A bad block check on a Bavarian command will cause the response lt NAK gt BAD BLOCK CHECK lt CR gt lt LF gt If a Bavarian command is sent without a set of matching sTx and
64. al to the accuracy and stability of the analyzer 98507600 Rev D 9 EC9850 SO ANALYZER OPERATION MANUAL This page is intentionally left blank 10 98507600 Rev D CHAPTER 1 DESCRIPTION 1 0 DESCRIPTION 98507600 Rev D The EC9850 sulfur dioxide SO analyzer is an ultraviolet UV fluorescence spectrometer designed to continuously measure low concentrations of SO in ambient air The 9850 analyzer comprises an optical sensor assembly an analog electronic signal preprocessor module microprocessor based control and computation electronics and a pneumatic system that samples ambient air by point monitoring The EC9850 A series has a built in charcoal scrubber that provides SO free zero air to the analyzer the instrument is designed to monitor the fluorescence background signal by periodically sampling SO scrubbed air This results in the virtual elimination of zero drift This feature is an optional extra in the 9850 B series where an external scrubber is normally used In addition to temperature and pressure compensation the analyzer adjusts the span ratio based on a known concentration of gas used to span the analyzer This feature is not automatically implemented and must be selected by the operator Analog and digital outputs are available for data monitoring The operator can select analog output as either current or voltage output Current ranges are 0 to 20 mA 2 to 20 mA or 4 to 20 mA Voltage outputs with the 50 pin I
65. al zero standard EC9850 automatic zero span AZS feature may be used to automatically carry out a Level 1 calibration check on a periodic basis Further when the 5 in the instrument is ENABLED the EC9850 automatically and continually compensates subsequent concentration measurements for any minor calibration drift as follows 44 X ISO Equation 3 4 Where SO pean the corrected instrument concentration reading based on the span compensation ratio obtained during the previous AZS cycle SO luucowe the instrument concentration reading without compensation 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL faz the span compensation ratio determined during the previous AZS cycle the default value of fazs is 1 000 until the first AZS cycle is carried out During an AZS cycle the EC9850 measures the concentration of the span gas provided to the span gas port This measurement reading should equal the actual concentration of the span gas standard If it does not the instrument sets 5 5 and calculates a new f 7 as follows 180 SO luncome 75 Equation 3 5 Where SO stp the certified concentration of the span gas standard at the span gas port The new value of is then used to compensate subsequent measurement readings until the next AZS cycle The zero gas reading during an AZS cycle is used fo
66. alibration and MEASURE mode Check the optional external zero scrubber by initiating an auto zero BACKGROUND cycle Exit and return to the primary screen Following the auto zero cycle allow the analyzer to sample zero air until a stable response is obtained If the response goes negative by more than 0 596 of full scale the zero scrubber may need to be replaced Record the final stable zero air response as Zsoo Note The EC9800 analyzer family does not have zero and span pots that are physically like those on traditional instruments Zero is set automatically on all units Span can be set manually by adjusting the instrument readings or by adjusting the instrument gain Adjust the zero air flow and the SO flow from this standard SO cylinder to provide a diluted SO concentration of approximately 80 of the full scale range of the analyzer Ensure that the total air flow at the output manifold exceeds the demand by at least 0 5 slpm to ensure that no ambient air is pulled into the manifold vent The exact SO concentration is calculated from SO xF SO Jour E 2 50 Fp Fo Equation 3 1 Where SO our diluted SO concentration at the output manifold ppm 50 concentration of the undiluted SO standard ppm Fyo flow rate of the SO standard corrected to 25 C and 760 Torr 760 Torr 101 kPa slpm 98507600 Rev D CHAPTER 3 CALIBRATION F flow rate of the dilution air correc
67. ands even in the absence of a lt STx gt character The DA command will function without a serial I D The string DA CR is a valid command The zero padding on the response to the DA command contains six ASCII zeros instead of the standard ten ASCII zeros The data type must be sent on 9800 style commands but it is not checked against the actual parameters The number of data parameters must be sent on 9800 style commands but it is not checked against the actual parameters 4 4 3 5 Bavarian Protocol This protocol is intended to correct the idiosyncrasies in the original protocol as noted in section 4 4 3 1 as they apply to the Bavarian network This protocol selection strictly applies the Bavarian network protocol to all commands 4 4 3 6 Command Acknowledgment g g For 9800 style commands that provide a data response the response itself is the acknowledgment For 9800 style commands that do not provide a data response no ac knowledgment is returned For Bavarian Network commands no acknowledgment is returned 4 4 3 7 Negative Acknowledgment For 9800 commands and for Bavarian Network commands no negative command acknowledgment is sent 4 4 3 8 Bavarian Protocol Idiosyncrasies o 98507600 Rev D The string DA CR is a valid command EC9850 SO ANALYZER OPERATION MANUAL The DA command will function without an I D The data type must be sent on 9800 style commands but it is not checked ag
68. ars the next failure on the list will be displayed The entire list of failures is displayed on the SYSTEM FAULTS screen Instrument gain displayed above the operational mode indicates the relationship between the calibration concentration and a measured gas concentration within the analyzer It is an essential parameter for the calibration of the analyzer and is an important requirement for system audits 2 17 EC9850 SO ANALYZER OPERATION MANUAL When the primary screen is displayed and the cursor highlights the words MAIN lt 2 5 2 Main Menu ENU press the lt Select gt or lt Enter gt key to enter the MAIN MENU MAIN MENU INSTRUMENT MENU CALIBRATION MENU TEST MENU INTERFACE MENU TREND SELECT MENU EVENT LOG INSTRUMENT STATUS SYSTEM TEMPERATURES SYSTEM FAULTS Figure 2 8 Main Menu Except for the last four menus each of the menus listed in Figure 2 8 has one or more levels of menu items contained within the selection The EVENT 106 is log created by the microprocessor to indicate deviations in the operating parameters This item can be used to determine the cause of system problems The INSTRUMENT STATUS and SYSTEM TEMPERATURES screens constantly update readings that apply to the operation of the instrument The SYSTEM FAULTS screen is a pass fail indication of various parameters which are continually monitored These parameters must be within acceptable
69. aution Sample and zero air connections to the EC9850 should be maintained at ambient pressure with any excess flow vented to the atmosphere The EC9850 requires at least 1 00 slpm 0 65 slpm sample plus 50 overflow of particulate filtered 5 micron dry noncondensing sample furnished at all times A 5 micron inlet filter is necessary to meet USEPA requirements which is already installed in the A series analyzer Tubing used for sample gas and exhaust connections must be 1 4 inch OD and 1 8 to 3 16 inch ID The recommended ID is 5 32 inch A segment of clean Teflon tubing should be purchased to connect the sample source to the sample inlet Only use lines and fittings made of stainless steel Teflon or glass Instructions for tubing connections with Kynar fittings o Cut the tubing squarely and remove any burrs o Insert the tubing through the back of the nut until it reaches the tube stop in the fitting o Tighten the nut finger tight plus 1 1 2 to 2 turns A squeaking sound when tightening the nut is normal o All nuts should be re tightened when the system reaches operating temperature o 98507600 Rev D 2 1 2 CHAPTER 2 INSTALLATION AND OPERATION 2 3 Exhaust Connections 1 2 3 1 A Series When making exhaust connections locate the exhaust outlet away from the sample inlet and occupied enclosed areas Connect a 1 4 inch OD line from the exhaust port to an exhaust manifold that vents outside of o
70. ccupied areas Lines and fittings of materials other than those cited above can be used for these connections 2 1 2 3 2 B Series 2 1 98507600 Rev D Connect the exhaust port of the analyzer to a vacuum pump capable of 4 slpm at 15 Hg 50 kPa vacuum minimum capacity The exhaust of the pump should be connected to a manifold to vent the exhaust gas away from occupied areas If desired a charcoal exhaust scrubber may be placed in the exhaust line to remove 90 Optional exhaust pump and exhaust scrubber are available from Ecotech Caution Flow in the EC9850B is calculated assuming critical pressure across an orifice thus the exhaust vacuum must be maintained at less than 1 2 atmosphere approximately 15 Hg 50 kPa at sea level to keep the orifice critical It is advisable that the user install a vacuum gauge on the exhaust line and periodically check that sufficient vacuum is being maintained If pump performance deteriorates below this level flow indications from the analyzer will no longer be valid 2 4 Zero Air Connection B Series only Caution Sample and zero air connections to the EC9850B should be maintained at ambient pressure with any excess flow vented to the atmosphere Ecotech recommend that only a charcoal scrubber be connected to the aux port This optional zero air scrubber is available from Ecotech Pty Ltd 2 7 EC9850 SO ANALYZER OPERATION MANUAL 2 2 AC Power Connection Verify that the
71. configuration of the machine by the user This option prevents the user from configuring the EC9800 menus by creating an individual password This feature allows the user to exclude changes to the front panel menus by locking them through a user specified password 2 7 1 Rules of Operation The password must be a four digit number o o o 98507600 Rev D After a memory erasure the analyzer will default to UNLOCKED The user must enter a four digit number to lock the analyzer The same four digit number is used to unlock the analyzer as well Once the analyzer is locked the user may navigate through the menus but cannot select a field for data entry Each time the user wishes to lock the analyzer the password must be en tered The password is only previous passwords are not valid while the analyzer remains locked remembered On the INSTRUMENT MENU there is a new entry labeled PAsswonp that dis plays the status of the menu as either UNLOCKED Or LOCKED 2 7 2 Sample Session 1 At INSTRUMENT MENU there is a field labeled PAsswonp This should display the status UNLOCKED 2 Select the field labeled PAss 0000 WORD The status UNLOCKED Will be replaced by 3 Using the select and arrow keys scroll to the desired numbers to represent the password 4 When the desired password will disappear and the Lock now locked appears press the Enter key Th
72. d RS232 9850 SO ANALYZER OPERATION MANUAL 4 2 2 Cable Connections Figure 4 2 Serial Interface Connection Diagrams 4 6 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION 4 3 Serial Terminal Control If the EC9850 is operated in the Terminal mode a terminal connected to one of the RS232 ports will produce the same results as pressing the six front panel keys with the exception that the same characters sent to the LCD instrument display will also be sent to the terminal The terminal keys will map into the front panel keys as follows Key Key Label Function Enter ENTER enter 4 HOME exit g UP ARROW up 2 DOWN ARROW down 6 RT ARROW select The Terminal mode must not be used if the multidrop port is daisy chained to other instruments The mode may be changed using the INTERFACE MENU through the INTERFACE MODE menu selection When in Terminal mode this choice may be made manually or through the serial port The mode may be changed from Command to Terminal through the serial port using the REMOTE command For information on required communication parameters refer to the REMOTE command in section 4 4 5 2 4 4 Serial Command Control When in the Command mode two command sets are available These are the 9800 command set and the Bavarian Network command set The 9800 command set is recommended for general use The Bavarian Network command set was set to support a specialized network in Bavar
73. d be a little below current barometric pressure 98507600 Rev D 2 37 REF VOLTAGE EC9850 SO ANALYZER OPERATION MANUAL Reference voltage as measured on the detector PCA in the A series and the pressure preamp PCA in the B series This voltage is indicative of the UV lamp intensity CONC VOLTAGE Voltage from the preprocessor proportional to the fluorescent signal from the reaction cell This voltage ANALOG SUPPLY is represents the actual measurement of gas 12 volt primary power supply DIGITAL SUPPLY 5 volt microprocessor power supply HIGH VOLTAGE PMT power supply high voltage reading LAMP CURRENT UV lamp power supply current VERSION Indicates the current firmware version installed in the Microprocessor Additional information on the INSTRUMENT STATUS screen is included in the EC9850 Service Manual 2 5 20 System Temperatures Screen CELL TEMP SYSTEM TEMPERATURES 50 7 DEG C CHASSIS TEMP 28 1 DEG C FLOW TEMP 49 9 DEG C COOLER TEMP 10 3 DEG C Figure 2 27 System Temperatures Screen The SYSTEM TEMPERATUR ES screen displays information continuously generated by the microprocessor CELL TEMP Temperature of the reaction cell 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION CHASSIS TEMP Temperature of air inside the chassis measured on the microprocessor PCA FLOW TEMP ONLY IN A SERIES Temperature of the
74. dications Moving the zero indication up the scale to a specific point creates a live zero thus allowing the recorder or DAS to show negative as well as positive indications The adjustment used to create a live zero is OFFSET For example a 10 offset moves the zero indication to the point where 10 would normally be indicated The full reading available on the recorder paper or DAS would then be 10 to 90 of full scale See Figure 2 29 Figure 2 29 Strip Charts Illustrating Offset Signal adjustments for zero and instrument gain to align the output with the user s recorder or other measurement device can be made in the ANALOG OUTPUT MENU in the FULL SCALE and ZERO ADJ fields These adjustments may be necessary due to tolerance buildup power supply variation etc in either the analyzer or the measurement device 2 6 2 Over Range Adjustment Over ranging is also enabled from the ANALOG OUTPUT MENU The OVER RANGE setting is the auxiliary range the operator chooses to track the data should the data exceed full scale of the original range The setting of over range has no impact on the measurement range of the analyzer it only affects the analog output scaling 2 40 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION With over ranging enabled as the concentration reaches 90 of the full scale value for the selected output range the software generates a positive spike that takes the indicator from the 90 positi
75. e To make the adjustment watch the recorder paper or DAS while you increment or decrement the zero adjustment correction factor that is displayed Press Enter to confirm your setting 6 Select FULL SCALE and adjust the analog output to 100 on the recorder paper or DAS To make this adjustment watch the recorder paper or DAS while you increment or decrement the full scale correction factor that is displayed Press Enter to confirm your setting 7 Select ovER RANGE and set to a range that is higher than the Range chosen at the top of the screen When the digits reflect the desired over range press Enter 8 Select OVER RANGING and choose either ENABLED Or DISABLED Press Enter 2 6 4 Calibration Requirements To make your data acceptable to the regulatory authorities and to pass required periodic audits you must calibrate the instrument before any data is collected for 2 42 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION use in a monitoring program The calibration procedure is included in Chapter 3 of this manual Most regulatory requirements also include establishing a calibration verification program If your organization does not have the staff to perform this task Ecotech s Service personnel can for contact details 2 7 Password Protection provide assistance See the front of this manual A password protection option was designed in order to solve the problem of altering the
76. e Figure 2 3 EC9850 SO ANALYZER OPERATION MANUAL Figure 2 3 50 Pin Connector Board with Sample Choices Rear Select the output for your application using the following steps Remove the 50 pin connector board from the rear panel of the analyzer Place the jumpers on the pins that correspond to the desired printed selections on the front of the board If current is selected only the jumpers selecting current make contact with both rows of pins The other jumpers are offset as shown in Figure 2 3 If a current output is selected the range must also be chosen from the menu when the instrument is operating The compliance voltage for the current output is 12 V A terminating resistor of 600 ohms or less should be used for measurement errors no greater than 1 If voltage output is selected both the REc and DAs outputs are factory set for 10 volts full scale Other full scale outputs of 5 V 1 V and 0 1 V can be selected Select the full scale output for REC and pas When using voltage output the source resistance for both REC and pas outputs is 1000 ohms The recorder and DAS input resistance should be greater than 500K ohms for a measurement error no greater than 1 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION 3 Connect the recorder or DAS wires to the appropriate terminal block The wire positions are OUT positive or signal COM ground or low SHLD shielded cable Caution To prevent ground loop
77. e RS232 line An example of this is shown here DCONC CR Caution Using this I D will result in a response from all analyzers connected to the serial line An example of an 9800 command with a parameter would be the trend dump command DTREND used here with an instrument I D of 134 DTREND 134 1 1 GASAVG lt CR gt 4 4 2 Bavarian Network Command Set Format All Bavarian Network commands follow the command format as specified in this section The specific Bavarian commands and their function are described in section 4 4 5 1 4 8 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION Bavarian Network Command Format lt STX gt lt TEXT gt lt ETX gt lt BCC1 gt lt BCC2 gt Where lt STX gt ASCII Start Of Transmission 02 hex lt TEXT gt 2 ASCII text maximum length of 120 characters lt ETX gt ASCII end of transmission 03 hex lt gt ASCII representation of block check value MSB lt BCC2 gt ASCII representation of block check value LSB The block check algorithm begins with 00 Hex and exclusive OR each ASCII character from srx to lt ETX gt inclusive This block check value is then converted to ASCII format and sent after the lt ETx gt character 4 4 2 1 Examples The following is an example of a valid Bavarian data request for an instrument that has an LD of 97 lt 5 gt 097 lt gt The block check calculation is best shown the fo
78. e computer where the serial cable is connected Connect the other end of the serial cable to the analyzer Baud rate 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION Choose the baud rate that has been set on the analyzer 4 6 3 2 2 Modem Connection Connect using Choose from the list of modems detected from the computer Phone Enter the phone number to which the analyzer is connected 4 6 3 2 3 USB Connection Analyzer Select the analyzer to communicate with from a detected list 4 6 3 3 Analyzer Tab This function sets the analyzer information for the analyzer being communicated with Analyzer ID If the user has multidropped multiple analyzers onto the one communication line enter the ID of the analyzer to communicate with Average Data Tick this box to download the averaged data that has been generated by the analyzer Instantaneous Data Tick this box to download the instantaneous data that has been generated by the analyzer 4 6 4 Keyboard Shortcuts The following are keyboard shortcuts that can be used in general operation of the program o F2 Display the Settings dialog box o F5 Start communicating with analyzer o F6 Stop communicating with analyzer 98507600 Rev D 4 31 EC9850 SO ANALYZER OPERATION MANUAL 4 7 Network Interface optional The EC9800 network interface is an optional feature that can be added to an EC9800 instrument It adds the possibility to connect the analyser to a
79. e correct Refer to section 4 6 3 2 Under the Mode menu choose the Remote Terminal option 3 Onthe Comm menu select Start 4 The screen should replicate the analyzer s display similar to Figure 4 6 The user now has access to the analyzer control panel with the buttons at the bottom of the screen replicating the buttons on the front panel of the analyzer If a blank screen appears terminate the connection as per step 6 below and reconnect 4 28 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION 9800 Data Downloader Bl x Comm Mode 8 Connected Figure 4 6 Remote Terminal 4 6 2 2 Ending a Remote Terminal Mode Session Controlling the analyzer by remote terminal automatically sets the analyzer to Remote mode Note It is advisable that the user always returns the analyzer to Command mode at the end of the remote terminal session so that other users may download data To end the remote terminal session 1 Set the Interface Mode option on the Interface Menu to Command For detailed instructions on how to do this see section 4 6 2 3 2 Terminate the connection by selecting the Stop option on the Comm menu 4 6 2 3 Setting the Analyzer to Command Mode During a Remote Terminal Session 1 Click exit repeatedly to display the analyzer s start up window Main Menu should be highlighted 2 Click enter to enter the Main Menu Click up or down until Interface Menu is highlighted 4 Click enter to
80. e password ED message will take its place The analyzer is 2 43 EC9850 SO ANALYZER OPERATION MANUAL 5 Scroll through the instrument menus From this point forward it is impossible to select any alterable fields 6 Return to the INSTRUMENT MENU and select the PASSWORD field 7 The LOCKED message will disappear and 0000 appears in its place 8 Using the Select and arrow keys scroll the numbers of the password entered previously 9 When the password is displayed press the Enter key The password will disappear and be replaced by the message UNLOCKED 10 The analyzer is unlocked and the menu configuration can be altered 2 44 98507600 Rev D CHAPTER 3 CALIBRATION 3 0 Calibration 3 1 Overview This chapter consists of the following o a general discussion of analyzer calibration o a description of the multipoint calibration procedure o a description of automatic zero span AZS setup o a discussion of the AZS feature The EC9850 sulfur dioxide analyzer is a precision measuring device that must be calibrated against a known source of sulfur dioxide that is traceable to a National Institute of Standards and Technology NIST standard Formerly NIST was the National Bureau of Standards or NBS In general terms the calibration process consists of the following steps 1 Establish a reliable and stable calibrating source 2 Provide a satisfactory connection between the calibrati
81. e the Analog Outputs if applicable Operation Manual 2 6 3 e f necessary perform a leak check Service Manual 3 3 13 e f necessary perform a flow calibration Service Manual 3 5 If necessary perform a Multipoint calibration Operation Manual 3 2 Step 5 Data Validation e Verify the results from your data acquisition system agree with the readings of the EC 9850 SO2 analyzer e Verify that the analyzer responds to automatic calibration sequences The analyzer is now operating correctly 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL Table of Contents MANUAL HISTORY 2 25 5 tatione tame bam omi amu RID 1 2 INTERNATIONALLY RECOGNIZED SYMBOLS USED ON ECOTECH 4 SAFETY REQUIREMENTS eite ERE RE ee di OH ER E ERE BREITER 5 FACTORY SERVICE eripe elo ue hib eie celles a ipi tin c breed ie ERR 6 CLAIMS FOR DAMAGED SHIPMENTS AND SHIPPING DISCREPANCIES 02 0 0000 7 SERVICE AND SPARE eee emper eee itd ectetur eiut eot ee epe d ede ded ede 8 WARNING edente eode e en etii ig ute Mte e eus 9 1 0 DESCRIPTION e P 1 1 LAS PECIFICATIONS eere Pede ngu eate reete 1 2 IA Ranges aee eere ie
82. econdary ID of 004 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION 4 4 5 Serial Command Sets This section describes the Bavarian Network and 9800 command sets available on the EC9850 using the instrument Command mode 4 4 5 1 Bavarian Protocol Command Set 98507600 Rev D Command DA Function Bavarian network command that returns the current instantaneous concentration Format lt STX gt DA lt DEVICE I D gt lt ETX gt lt BCC1 gt lt BCC2 gt Device response lt STX gt MD 01 lt SP gt lt kkk gt lt SP gt lt nnnn ee gt lt SP gt lt ss gt lt SP gt lt ff gt lt SP gt lt mmm gt lt SP gt 000000 0000 lt SP gt lt ETX gt lt BCC1 gt lt BCC2 where kkk instrument ID nnnn ee instantaneous gas concentration in ppb or mg m3 ss Status byte for both channels with the following bit map positive logic unused D1 out of service D2 instrument is in zero mode D3 instrument is in span mode 4 unused D5 unused D6 units 1 ppm 0 mg m3 D7 instrument is performing a background cycle ff failure byte for both channels with the following bit map positive logic DO flow sensor failure D1 instrument failure 2 unused unused D4 lamp failure D5 temperature sensor failure 6 unused D7 unused mmm instrument serial number 0000 the response will be 6 pad zeroes if the original or enhanced protocol 4 15 E
83. ed digital potentiometers used to adjust the operation of the preprocessor board Each pot is set with digits 0 99 in a non wraparound scrolling field MEASURE COARSE ZERO Software controlled pot for the electronic zero of the measure channel 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION REFERENCE ZERO Sets the zero for the reference channel and is controlled by the microprocessor MEASURE GAIN Sets gain for the measure channel and is controlled by the microprocessor board REFERENCE GAIN Software controlled pot for the gain of the reference channel TEST MEASURE Software controlled pot that is used by technicians when troubleshooting or verifying correct instrument performance HIGH VOLTAGE ADJUST Software controlled pot to adjust the PMT high voltage LAMP ADJUST Software controlled pot that sets UV lamp current REF VOLTAGE Reference voltage as measured on the detector PCA in the A series and the pressure preamp PCA in the B series This voltage is indicative of the UV lamp intensity 502 PPM Gas concentration reading CONC VOLTAGE Voltage from the preprocessor proportional to the fluorescent signal from the reaction cell This voltage represents actual gas measurement HIGH VOLTAGE PMT power supply high voltage reading 98507600 Rev D 2 21 EC9850 SO ANALYZER OPERATION MANUAL 2 5 9 Flow Control Pots Screen A series only FLOW CONTROL POTS are potentiometers used to adjust the functions of
84. erature Manifold Tem perature and Cooler Temperature o Instrument Status is where all the voltage gas flow gas pressure and ambient pressure parameters are shown This group shows all of the parameters of the Instrument Status menu with in the instrument 4 32 98507600 Rev D CHAPTER 4 DIGITAL COMMUNICATION 4 7 2 Remote Mode The remote mode allows the user to access the current menu screen in the ana lyser and remotely control the analyser through the menus The buttons on the left substitute the Up Down Select Pg Up Back Exit and Enter keys within the analyser ECOTECH World class Air amp Water Monitoring REMOTE MODE Current Readings 1075 PPM 075 AVG Remote Mode 075 PPM Download 075 AVG Contact Us 000 PPM 000 AVG ADJUST HIGH VOLTAGE START UP SEQUENCE ACTIVE INSTRUMENT GAIN 0 961 14 nFc n4 Applet Screen started Figure 4 8 Remote Mode Window The keyboard can also be used to control the menu with the following keys Table 1 Keyboard key and there commands Keyboard Key Menu Function Up Key Up Down Key Down Left Key Pg Up Back Right Key Select 98507600 Rev D 4 33 EC9850 SO ANALYZER OPERATION MANUAL Home Key Exit Function Key 5 F5 Refresh IMPORTANT When the window is left and an another program is used the other program takes control of the keyboard In this situation when you return to
85. ests of Ecotech 2 98507600 Rev D FRONT MATTER MARK DECLARATION Declaration of Conformity Sulfur Dioxide Analyzer Scope of Declaration The declaration applies to Sulfur Dioxide Analyzers as manufactured by Ecotech P L and which may be sold in the following configurations Description 98501000 100 Sulfur Dioxide Analyzer 98507000 T Sulfur Dioxide Analyzer Trace 98507000 1 Sulfur Dioxide Analyzer B Series 98507000 H1 Sulfur Dioxide Analyzer High Level 98507000 2 Sulfur Dioxide Analyzer with IZS Ecotech certifies that this product operates in compliance with the following standards EN 61326 1 Electrical Equipment for measurement control and laboratory use EMC Requirements Edition 1 1 with amendment 1 plus amendment 2 0 Immunity Requirements EN61326 1 IEC 61000 4 11 Voltage Interrupts IEC 61000 4 11 Voltage Dips IEC 61000 4 3 Radiated RF electromagnetic field immunity test IEC 61000 4 4 Electrical fast transient burst immunity test IEC 61000 4 5 Surge immunity test IEC 61000 4 6 Immunity to conducted disturbances induced by radio frequency fields o Electromagnetic compatibility EN61326 1 Annex A CISPR 22 and CISPR 16 2 CISPR 16 1 and CISPR 16 2 EN 61010 1 Safety requirements for electrical equipment control and laboratory use Section 19 of 60204 1 Insulation Resistance Check Residual Voltage Check Earth Continuity The equipment must be operated as per the directions given by Ecotech P L in
86. field and to highlight the different digits in the field When the cursor indicates the digit you wish to change press the Up or Down arrow key until the desired digit appears Go to the next digit by pressing lt Select gt When all digits of an entry are correct press the lt Enter gt key to confirm the entry Caution The lt Select gt key does not confirm an entry You must press the lt Enter gt key 2 3 2 2 Microprocessor Generated Information Some fields such as those on the INSTRUMENT STATUS and the SYSTEM TEMPERATURES screens contain information generated by the microprocessor The operator cannot affect the readings in these fields If you find that the cursor will not enter a field the field contains microprocessor generated information 2 3 2 3 Exiting Without Making a Change If you decide not to make a change during this process simply press the lt Exit gt key and the values will return to the previous entries 2 12 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION 2 3 3 Setting the Date and Time 98507600 Rev D Before the instrument can be calibrated or collect data for regulatory use the time and date must be set Go to the INSTRUMENT MENU and select DATE and TIME If these are not already set use a 24 hour clock setting for time and set the date in the day month year format See section 2 5 for instructions on programming menu entries 2 13 EC9850 S
87. g the Ecotech data downloading software mentioned in section 4 6 AVERAGE DATA If the average data is set to off no average data is recorded If it is set to on then the average data displayed on the primary screen is recorded The averaging period of this data is set in the MEASUREMENT MENU 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION INST DATA The INST DATA option allows you to select either off where no data is recorded or record instantaneous data with the following intervals 1 HOUR 30 24 MINUTES 10 MINUTES 5 MINUTES 3 MINUTES OR 1 MINUTE 98507600 Rev D 2 35 EC9850 SO ANALYZER OPERATION MANUAL TOTAL CAPACITY When either of the above are set to on the amount of free memory available for data logging will be displayed in days This indicates how much data can be stored before the earliest data will start to be overwritten Inst Data Total Capacity min days ERASE LOGGED DATA When yes is selected and enter is pressed all the logged data will be erased 2 5 16 Network Adaptor Menu The Network Adaptor Menu allows the user to enter or change the address Netmask and Gateway NETWORK ADAPTER MENU I P ADDRESS 0 0 0 0 5 0 0 0 0 GATEWAY 0 0 0 0 2 5 17 Trend Select Menu TREND SELECT MENU is the graphic display of the parameters listed TREND SELECT MENU CONCENTRATION AVG CONCENTRA
88. gh voltage adjust 16 PRE POT 6 9850 high voltage adjust 17 PRE POT 7 9850 lamp adjust 18 VREG POT I Flow control zero 19 VREG POT 2 20 VREG POT 3 21 VREG POT 4 22 VREG POT 5 Fan speed control 23 VREG POT 6 Pump speed fine 24 VREG POT 7 Pump speed coarse 25 Analogue input 0 26 Analogue input 1 27 Analogue input 2 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL 28 Analogue input 3 29 Analogue input 4 30 Analogue input 5 31 Analogue input 6 32 Analogue input 7 33 Analogue input 8 34 Analogue input 9 35 Analogue input 10 36 Analogue input 11 37 Analogue input 12 38 Analogue input 13 39 Analogue input 14 40 Analogue input 15 41 50 PIN IO bits 0 7 BIT 7 Span Out of Range BIT 6 Span On BIT 5 Copper Fail BIT 4 Zero On BIT 3 Lamp Fail BIT 2 Out Of Service BIT 1 Flow Fail BIT 0 Span Cycle 42 50 PIN IO bits 8 15 BIT 7 Pump On BIT 6 Range 1 BIT 5 Startup BIT 4 Heater Fail BIT 3 Range 0 BIT 2 IZS On BIT 1 Spare 1 BIT 0 ZeroCycle 43 50 PIN IO bits 16 23 BIT 7 Power On BIT 6 Sys Fail BIT 5 High Voltage Fail BIT 4 Power Supply Fail BIT 3 Electric Test BIT 2 Optical Test BIT 1 Range 2 BIT 0 PPm Metric 44 50 PIN IO bits 24 31 Really User ID 45 50 PIN IO bits 32 39
89. gnals Universal Serial Bus USB or discrete control over the 50 pin I O Discrete control is limited to lines which either command a specific operation or indicate an operation is in progress Serial communication allows access to the menu structure using a terminal and also includes a library of other specific operations USB provides a simple way to monitor the current state of the analyzer and download data that has been logged to the internal FLASH memory 4 1 Discrete Control Two control inputs are available through the 50 pin I O connector They are DOZERO and These inputs will place the analyzer in either Zero mode or Span mode respectively the analyzer will remain in the selected mode while the input is active When these inputs are made active the analyzer will actuate the valve drivers selected in the CALIBRATION MENU for CALIBRATION INTERNAL EXTERNAL other discrete connections are status outputs from the analyzer 4 1 1 50 I O Functional Specification The 50 pin connector on the back of the instrument will have functions assigned to pins per the following table Note 1 Signal Name Number Function 10073 2 Analog current output 3 Note 2 DOZERO 5 EXTERNAL INPUT TO PUT THE INSTRUMENT INTO THE ZERO MODE DOSEAN 6 External input to put the instrument into the span mode OVERANGE1 7 Active output indicates that analog output 1 has gone int
90. h AZS cycle Note Use of sPAN COMP ENABLED is not allowed under U S EPA designation at this time Caution Setting an incorrect 502 TIMED SPAN value with SPAN COMP ENABLED will cause all subsequent readings to be incorrect A built in check verifies that the observed value is not different from the calibration value by more than 25 If the difference is greater than 25 no updated correction will be made and a message is sent to the EVENT LOG setting the CALIBRATION ERROR flag Note The zero value is a reference value only Regardless of the state of the span comp option the analyzer does not correct for shifts in the zero The number displayed after so2 SPAN RATIO is the factor by which the instrument gain is multiplied to cause the display and output to agree with the span gas and will always be between 0 75 and 1 25 You cannot set this number This value is reset to 1 000 any time that the span is set manually via the front panel The assumption is that front panel adjustment is an instrument calibration thus preventing compound adjustments 98507600 Rev D CHAPTER 3 CALIBRATION 3 4 4 Description of the AZS Process The instrument will initiate a full zero span cycle starting at the prescribed hour The valve to admit zero air will be opened and the sample valve closed The instrument will allow the cell to fill with the gas for 8 minutes During the first 3minutes no data is taken
91. ia Additionally three communication protocols are provided to allow the user to specify the different handshaking based on their requirements 4 4 1 9800 Command Set Format 98507600 Rev D 9800 commands follow the command format as specified in this section The specific 9800 commands and their functions are described in section 4 4 5 9800 Command Format lt ccccccccc gt lt III gt lt D gt lt NN gt lt PPPPPPPPPPPP gt lt T gt EC9850 SO ANALYZER OPERATION MANUAL Where lt CCCCCCCCC gt command in ASCII 1 to 9 characters SUBE three digit instrument ID in ASCII format Sm single digit data type in ASCII optional number of parameters in ASCII 1 to 9 optional lt PPPPPPPPPPPP gt parameter in ASCII 1 to 15 characters optional lt I gt termination lt CR gt Or lt LF gt For commands that do not have parameters the format is the subset lt CCCCCCCCC gt lt III gt lt T gt For commands with multiple parameters the parameters are separated by the comma delimiter and the termination character follows the last parameter lt CCCCCCCCC gt lt III gt lt D gt lt NN gt lt PPPPPPPPPPPP gt lt PPPPPPPPPPPP gt lt T gt 4 4 1 1 Examples An 9800 command with no parameters would be the concentration request DCONC used here with an instrument I D of 001 DCONC 001 lt CR gt If no device I D is programmed the I D can be used to address any analyzer connected to th
92. ient for this operation 3 2 1 4 Procedure 98507600 Rev D 1 Assemble a dynamic calibration system such as the one shown in Figure 3 1 All calibration gases including zero air must be introduced into the sample inlet of the analyzer system The EC9850 instrument sample inlet is labeled Inlet and is located on the rear panel 2 Ensure that all flowmeters are properly calibrated under the conditions of use if appropriate against an authoritative standard such as a soap bubble meter or wet test meter All volumetric flow rates should be corrected to 25 C and 760 torr 760 torr 101 kPa A discussion of calibration of flowmeters is given in Calibration Reference 1 3 5 EC9850 SO ANALYZER OPERATION MANUAL 3 6 3 Ensure that the analyzer is properly connected to the output recording device as described in Chapter 2 If necessary go to the INTERFACE MENU select the ANALOG OUTPUT MENU field and select the appropriate settings for the recording device Offsetting the analyzer s zero indication OFFSET and ZERO ADJ to 5 of scale is recommended to facilitate observing negative zero drift Exit and return to the primary screen Adjust the calibration system to deliver zero air to the output manifold The total air flow must exceed the total demand of the analyzer s connected to the output manifold by 0 5 slpm to ensure that no ambient air is pulled into the manifold vent Enter the CALIBRATION MENU and select MANUAL c
93. ironmental Conditions Relative humidity 10 to 80 Temperature 5 to 40 degrees C Pollution degree 2 Installation category II Maximum altitude 2000m Instruments suitable for use in a sheltered environment only Never operate this equipment in the presence of flammable liquids or vapors as this could cause a safety hazard 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL Factory Service We strive to provide efficient and expedient service when an instrument or component is returned for repair Your assistance can help us to better provide the service you need To ensure that we process your factory repairs and returned goods efficiently and expeditiously we need your help Before you ship equipment to our factory please call our Service Response Centre at 461 1300 364 946 This enables us to complete the necessary paperwork and process your equipment correctly when it reaches our facility When you call please be prepared to provide the following information Your name and telephone number Your company name with shipping address The number of items being returned The part number of each item The model number or a description of each item The serial number of each item if applicable A description of the problem you are experiencing if factory repair is needed or the reason you are returning the equipment eg sales return warranty return etc The original sales order number or invoice number related to
94. ld should be of sufficient diameter to ensure an insignificant pressure drop at the analyzer connection The system must have a vent designed to ensure atmospheric pressure at the manifold and enough inlet flow to prevent ambient air from entering the manifold 3 2 1 3 Reagents 3 3 2 1 3 1 SO Concentration Standard The cylinder of SO in air or nitrogen must contain the appropriate concentration of SO suitable for the selected operating range of the analyzer under calibration Typical concentration standards are in the range of 20 to 100 ppm for dilution to the analyzer s concentration range The assay of the cylinder must be traceable either to a National Institute of Standards and Technology NIST 50 in air Standard Reference Material SRM or to an NIST EPA approved commercially available Certified Reference Material CRM A recommended protocol for certifying SO gas cylinders against either an SO SRM or a CRM is given in Section 12 of Calibration Reference 1 SO gas cylinders should be recertified on a regular basis as determined by the local quality control program 2 1 3 2 Dilution Gas Zero Air Zero air is described as air free of contaminants which will cause a detectable response in the SO analyzer The zero air should contain 0 1 ppb SO A procedure for generating SO free air is given in Calibration Reference 1 The Ecotech GasCal 1000 calibrator and 8301 Zero Air source has SO free air capability suffic
95. lit digit until the span calibration gas concentration value is displayed When the desired concentration is displayed press Enter Then move the backlit cursor to the INSTRUMENT GAIN field The instrument gain is automatically calculated by the analyzer Press Enter to confirm this value Press Exit to return to the primary screen The concentration on the primary screen should now read the same as the concentration of the calibration gas Note The auto zero function of the EC9850 eliminates the need for a traditional zero calibration This completes the span calibration of the EC9850 analyzer 2 15 9850 SO ANALYZER OPERATION MANUAL 2 5 Menus and Screens This section illustrates the various menus and screens for the EC9850 A series analyzer The B series menu structure will be very similar in structure yet will not have some of the options that are available in the A series These will be made clear throughout the relevant sections A short description of each menu and screen is provided The entire menu structure is shown in Figure 2 6 below UNIT SELECTION ROREM MENN MEASURE COARSE ZERO TIME AVERAGE PERIOD RT MEASURE GAI 2222 FILTER REFERENCE GAIN ERASE MEMORY NOISE TESTO MASUR MATA GBS TR AERO OS HIGH VOLTAGE ADJUST CALIBRATION TIMED CALIBRATION MANUAL LAMP ADJUST TIMER INTERVAL CAL MODE REF VOLTAGE STARTING HOUR CALIBRATION 502 CYCLE TIME BACKGROUND CONC VOLTAGE SO2 TIMED
96. ll data is downloaded from the start date until the last reading Data Type The data type to be downloaded can be selected between Instantaneous data Av erage data or all logged data Date Format The Date format can be changed with the order of time date month and year be ing interchangeable to your preference File Path In this field the file name and path where the data is to be saved should be speci fied either by typing the path name or by using the browse button When typing the file name put the extension either txt or csv NOTE By default the path is c test EC9800 txt if you do not have a folder named test and se lect download this would cause an error When all the options are set the download can start by selecting the Download Button A dialog window will prompt to inform that the download data process may take several minutes after that the status bar will show the current data being written to the file Status Downloading 2004 12 14 11 08 Figure 4 10 Status Downloading bar When the downloading process is finished a dialog window pops up to inform that the download has been successfully completed The file should then be saved in the current directory 98507600 Rev D 4 35 EC9850 SO ANALYZER OPERATION MANUAL 4 7 4 Firmware Update for the Network Interface The firmware in the network adapter consist of two files the cobox rom and Ecotech cob both of this files need to be loaded into the
97. llowing table 5 00 ot00 _ o f o wuow 3 o wuiw weno The binary value 0011 1010 corresponds to the hex value 3A This value in ASCII forms the last two characters of the data request message Please note that the I D of 97 is sent as the sequence 097 All I D strings must have 3 digits and the user should always pad with ASCII zero characters This is an example of a valid command to put the unit in the manual span mode if the instrument has an I D of 843 lt 5 gt 5 843 lt gt 52 98507600 4 9 EC9850 SO ANALYZER OPERATION MANUAL 4 10 The block check operation is best shown with the following table 3 nne o F000 0000 wei _ The binary block check value is 0101 0010 which is the hex value 52 as shown at the end of the command string 4 4 3 Protocol Definition and Selection There are three protocol selections available for the EC9850 via the INTERFACE MENU These are provided so the user may select the appropriate protocol for their desired application The first protocol designated original should be used when upgrading software in analyzers that are already in serial networks The original protocol is provided for back compatibility as it completely duplicates the protocol already in the field The second protocol provided is Bavarian The Bavarian protocol should be used with the Bavarian Network Command Set for an
98. llustrations Optional 50 Pin Connector Board Front iae ertet er dr Ye nee 2 3 Illustrations Output Test 2 26 Illustrations Over Range as Seen on a Strip Chart e EU E RF ES ene 2 41 Illustrations Preprocessor Pots Screen 2 26 Illustrations Primary Screen sssss 2 17 Illustrations Serial Interface Connection Diagrams 4 6 Illustrations SO2 Analyzer Calibration Using Cylinder Gas Dilution eee 3 4 Illustrations SO2 Calibration Using a Permeation i pila eeiam E 3 12 Illustrations Status Output Connections 4 5 Illustrations Strip Charts Illustrating Offset 2 40 Illustrations System Faults Screen 2 39 Illustrations System Temperatures Screen 2 38 Illustrations Test 2 25 Illustrations Timed Calibration Menu 2 22 Illustrations Valve Test Menu 2 29 Installation ee tertie 2 1 Instrument 4 14 Instrument Menu sese 2 19 Instrument Menu 2 2 18 Instrument Status 2 37 Instrument Status Screen illustration 2 37 Interface 2 31 Interface Men
99. make the pneumatic and electrical connections 2 1 2 Connections All pneumatic connections must be secure to ensure accurate operation of the analyzer The following information describes connection techniques for pneumatic and electrical connections Figure 2 1 below shows the rear panel of 98507600 Rev D 2 1 EC9850 SO ANALYZER OPERATION MANUAL the analyzer with associated connections Notice the Network connection is optional DISCRETE VO Figure 2 1 Analyzer Rear Panel 2 1 2 1 Recorder and DAS Connections Caution The EC9850 analyzer electrical ground is isolated from earth ground To avoid possible ground loops all electrical devices connected to the analyzer should have floating inputs not connected to earth ground 2 1 2 1 1 The 50 I O PCA The 50 pin I O connector board plugs into the discrete I O connector and provides voltage and current outputs to drive a strip chart recorder REC and a data acquisition system Das The outputs are illustrated in Figure 2 2 The 50 pin I O PCA is optional for the EC9850 B series analyzer 2 2 98507600 Rev D 98507600 CHAPTER 2 INSTALLATION AND OPERATION Figure 2 2 50 Pin Connector Board Front The output is jumper selectable as Current see Figure 2 3 Range is set using the menu in a later step o Voltage with selectable ranges of 0 to 0 1 V 0 to 1 V 0 to 5 V and 0 to 10 V Se
100. n must perform the periodic maintenance procedures in the EC9850 Service Manual especially checking the particulate filter The INSTRUMENT STATUS and SYSTEM TEMPERATURES screens in the EC9850 Service Manual give the ranges for correct operation of the instrument Note Calibration should only be performed when the instrument is stable and has been powered up for at least two hours 3 2 1 Procedure Using Cylinder Gas Dilution Method 3 2 1 1 Principle A certified standard cylinder of SO diluted as necessary with zero air is used to obtain the various calibration concentrations needed 3 2 1 2 Apparatus The major components and configuration of a typical calibration system are shown in Figure 3 1 98507600 Rev D 3 3 EC9850 SO ANALYZER OPERATION MANUAL 3 4 Figure 3 1 SO Analyzer Calibration Using Cylinder Gas Dilution 3 2 1 2 1 Flow Controllers A device capable of adjusting and regulating flow rates Flow rates must be regulated to 1 3 2 1 2 2 Flowmeters A calibrated flowmeter capable of measuring and monitoring flow rates with an accuracy of 2 of the measured value 3 2 1 2 3 Pressure Regulator for Standard SO Cylinder The regulator must have a nonreactive diaphragm and internal parts and a suitable delivery pressure 3 2 1 2 4 Mixing Chamber A chamber designed to provide thorough mixing of SO and diluent air 3 2 1 2 5 Output Manifold 98507600 Rev D CHAPTER 3 CALIBRATION The output manifo
101. n the CALIBRATION MENU the second lower CALIBRATION prompt requires the designation of INTERNAL Or EXTERNAL This choice should be EXTERNAL only if the optional EZS valve assembly is installed and gas is attached to the Zero and Span ports 1 Enter the CALIBRATION MENU 2 Atthe first upper CALIBRATION prompt select TIMED 3 At the TIMER INTERVAL prompt set the number of hours between timed calibrations Typical settings are 23 or 24 hours EC9850 SO ANALYZER OPERATION MANUAL 4 At the STARTING HOUR prompt enter the hour of the day when AZS is to commence 5 Atthe prompt CYCLE TIME enter the number of minutes required for the span and zero steps to run 6 Atthe so2 TIMED SPAN prompt enter the concentration value of the span gas to be used 7 Atthe second lower CALIBRATION prompt choose INTERNAL unless the EZS valve option is installed and you want to use the external calibrator gas Using the INTERNAL choice the span gas is connected through the Span bottom port of the internal valve manifold and the output of the zero air scrubber is used for zero air If EXTERNAL is chosen the span is connected to the Span port of the EZS valve manifold and user supplied zero air is connected to the Zero port of the EZS valve manifold 8 At the SPAN prompt select ENABLED if you want the instrument span adjusted to agree with the span gas after eac
102. network and access directly using a web browser Within the web browser multiple users have the possibility of seeing current readings which are updated every 5 sec onds control the analyser remotely and download data 4 7 1 Current Readings The current reading option gives the possibility of seeing current parameter val ues in the analyser such as temperatures concentration and status The values on the screen are auto refreshed every 5 seconds and multiple users are able to view the current values simultaneously ECOTECH World class Air amp Water Monitori T CURRENT READINGS EC9841A NOx Analyser Standard Current Readings RematuMods 75 08 0 0 75 08 Download Primary Gas Concentration Secondary Gas Concentration Calculated Gas Concentration Conte UE 75 97 9 0 75 07 Primary Gas Average Secondary Gas Average Calculated Gas Average 0 96 Instrument Gain Gas Concentrations instrument Temperature Instrument Status Annet Reainns started E Figure 4 7 Current Readings window The parameters are grouped in three main categories Gas Concentration Instru ment Temperatures and Instrument Status o The gas concentrations group shows the current concentration average con centration and instrument gain o The Instrument temperature group shows the relevant current temperatures for the instrument e g the standard EC9841 NO analyser would show Cell Temperature Conversion Temperature Chassis Temp
103. o SRM CRM sources section 12 98507600 Rev D CHAPTER 3 CALIBRATION Note Use of sPAN COMP ENABLED is not allowed under U S EPA designation at this time A Level 2 span check with SPAN COMP DISABLED does not require certification of the span gas used during AZS and the result of such a check may not be used to correct the data but merely serves to indicate that the analyzer is functioning properly If the AZS is used for a Level 2 span check the sPAN comp must be set to DISABLED A Level2 AZS cycle should be initiated immediately after multipoint calibration so that a valid reference point can be determined As described in this section two modes of operation are possible In the Internal mode the zero air scrubber is used as the source of zero air In the External mode the user must supply zero air through the 7 inlet of the optional EZS valve In the External mode the Outlet port is connected with a short piece of tubing to the Inlet port In either mode the user must supply the span gas In the Internal mode the span gas must be connected through the analyzer to the Span port bottom port of the internal valve manifold In the External mode the span gas is connected to the Span port of the optional EZS valve manifold These gases zero and span must be provided to the analyzer at atmospheric pressure such as through a manifold as shown in Figure 3 1 3 4 3 AZS Setup 98507600 Rev D Note I
104. o over range OVERANGE2 8 Active output indicates that analog output 2 has gone into over range OVERANGE3 9 Active output indicates that analog output 3 has gone into over range PNA LNT 10 Unused analog input 1 ANAIN2 11 Unused analog input 2 TOUTI 15 Analog current output 1 Note 3 IOUT2 17 Analog current output 42 Note 4 98507600 Rev D 4 1 EC9850 SO ANALYZER OPERATION MANUAL 4 2 Signal Name SPANCYL OUTSERV ZEROON SPANON ZEROCYL IZSON STARTUP PPM MET FLOWFAIL LAMPFAIL Number 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 35 36 Function Active output indicates that the instrument is in the Span or Span Hill mode Active output indicates that the Out of Service switch is in the out of service position Active output indicates that the instrument is in the Zero mode Active output indicates that the instrument is in the Span mode Active output indicates that the instrument is in the Zero or Zero Fill mode Active output indicates that Internal Zero Span has been selected Note 5 Active output indicates that the startup sequence is active Active output indicates that the instrument is in mg M USER ID byte bit 1 Used in conjunction with the PINID serial command USER ID byte bit 2 Used in conjunction with the PINID command USER ID byte bit 3 Used in conjunction with the PINID
105. on source and the analyzer 3 Calibrate the analyzer against the calibration source Multipoint calibration is used to establish the relationship between analyzer response and pollutant concentration over the analyzer s full scale range Zero and span checks are frequently used to provide a two point calibration or an indication of analyzer stability and function Regulations generally required that the analyzer be recalibrated any time it is moved or serviced or whenever the analyzer characteristics may have changed This includes changing the instruments units from volumetric to gravimetric Regulatory agencies establish the time intervals at which the analyzer must be calibrated to ensure satisfactory data for their purposes Important Use of the EC9850 SO analyzer as U S EPA designated equivalent method requires periodic multipoint calibration in accordance with the procedure described below In addition the instrument must be set to the parameters indicated in Chapter 1 98507600 Rev D 3 1 EC9850 SO ANALYZER OPERATION MANUAL 3 2 3 1 1 Analyzer Calibration Instructions Note This procedure is a quick guide to span calibration of the EC9850 analyzer intended for operators who are familiar with gas analyzers and preparation of calibration gas For complete gas preparation and multipoint calibration instructions please refer to section 3 2 Connect a source of span calibration gas to the analyzer through the Inlet po
106. on to the 100 position The output data is then scaled for the full scale chosen for over range As the output drops back to 80 of the original full scale the software generates a negative spike from the displayed value to zero The output then reverts to the original range See Figure 2 30 for an example of over range on a typical strip chart recorder Figure 2 30 Over Range as Seen on a Strip Chart Recorder The range value should generally be set first However because the range value must be less than the currently selected over range value it may be necessary to increase the over range value to the desired setting first The over range value is limited to be equal to or greater than the currently selected range value For practicality it is recommended that the over range be set to a value between 2 and 5 times the range value For example if the desired monitoring range is 0 2 ppm the over range should be set between 0 4 and 1 0 ppm Certain precautions must be taken when over ranging is enabled to ensure that pollutant concentration measurements are reported correctly When a data acquisition must interface with the analog output of the instrument some means must be provided to indicate which range is in effect during all measurements The user should monitor pin 7 on the 50 pin I O connector which is an open collector output indicating analog output 1 is in over range 98507600 Rev D 2 41 EC9850 SO ANALYZER OPERATION MANUAL
107. ontrast Down arrow Y and lt 5 lect for lighter contrast Backlight The backlight brightness is fixed to maximum and cannot be adjusted Hold the key combinations until the desired contrast appears on the dis play 2 9 EC9850 SO ANALYZER OPERATION MANUAL Note Pressing the Up or Down arrow key while not simultaneously pressing the Select key when the main screen is displayed causes the screen query START MANUAL CALIBRATION If this happens while adjusting the display press the Exit key Note The display is sensitive to the ambient air temperature analyzer temperature The appearance of the display will vary with changes in these conditions 2 2 2 Warmup When the instrument is initially powered up several components in the instrument are automatically configured by the microprocessor and an automatic zero is run This process requires about 30 minutes During the startup period several messages are displayed on the initial screen These indicate the progression toward normal operation Initial Screen Message Instrument Activity Lamp turned on and allowed to stabilize PMT high voltage set AUTO ZERO ADJUST Coarse zero of measurement channel with zero air flowing ELECTRONIC ZERO ADJUST ROUND FILL Cell filling with zero air BACKGROUND MEASURE Zero reading from measurement cell Final determination of system zero Cell filling with sample air
108. ows a series of vertical lines in alternate positions to those is STRIPE 1 CLEAR Clears the screen of all pixels FILL Fills the screen of pixels CHECK 1 Causes the screen to show a checkered pattern made up of single pixels CHECK 2 Displays a checkered pattern in alternate spaces to CHECK 1 ANALOG OUTPUT TEST 2 30 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION Sends a 0 1 Hz sawtooth waveform to the selected analog output device to test its functionality There are 6 analog outputs to choose from 41 to 6 analog outputs 41 to 3 are available via the 50 PIN IO connector 2 5 12 Calculation factors CALCULATION FACTORS INSTRUMENT GAIN 1 0592 P T F CORRECTION 1 0390 BACKGROUND 0 0012 ZERO OFFSET 0 0000 PPB EXIT Figure 2 19 Calculation Factors Menu The Calculation factors screen is a non editable screen which provides the values used to calculate different aspects of measurement and calibration 2 5 13 Interface Menu INTERFACE MENU ANALOG OUTPUT MENU DATA LOGGING MENU MAIN GAS ID 050 INTERFACE MODE COMMAND MULTIDROP BAUD 2400 DATA LENGTH 8 BITS STOP BITS 1 BIT PARITY NONE COMM PROTOCOL ORIGINAL NETWORK ADAPTOR MENU Figure 2 20 Interface Menu The INTERFACE MENU is used for interfacing analog recording instruments and programming RS232 parameters The following are used only when one or more of the serial ports are to
109. power selection switch on the rear panel and the power cord and fuse are appropriate for your use Move the switch right or left so the appropriate voltage rating is visible on the switch Figure 2 1 shows the voltage selection switch Warning Power is supplied to the analyzer through a three pin power plug The ground must not be defeated and an adequate ground must be connected to the instrument both for proper performance and for the safety of operating personnel The warranty on the analyzer applies only if the analyzer is properly grounded If it is not properly grounded and electric power is applied in violation of the National Electric Code Ecotech assumes no responsibility for any injury or damage to personnel or property Warning Be sure to check that the mains power selection switch is at the correct setting before turning the instrument on Failure to do so may result in damage to the power supply Connect the power plug to the power receptacle and press the power switch to the ON position on the rear panel Also make sure that the DC POWER switch on the front secondary panel is switched to ON 2 2 1 Display Adjustments Adjust the display contrast by simultaneously pressing two keys on the front panel see Figure 2 4 2 8 98507600 Rev D 98507600 CHAPTER 2 INSTALLATION AND OPERATION 4 RACK MOUNT OPTION 2 Figure 2 4 Analyzer Front Panel Contrast Up arrow A and Select for darker c
110. problems connect the shield of the cable at the analyzer only not at the recorder or DAS For additional information regarding output see section 2 6 2 1 2 1 2 Current Output Connections When using the EC9850 without the 50 pin I O PCA the analyzer still provides current outputs to drive a strip chart recorder or DAS These outputs are present on the discrete I O connector at the following pins Pin Discrete I O Connector Comet DGND Gnd 1 12 14 or 16 If a current output is connected the range must also be chosen from the menu when the instrument is operating The compliance voltage for the current output is 12 V A terminating resistor of 600 ohms or less should be used for measurement errors no greater than 1 2 1 2 1 3 Voltage Output Connections The current output mentioned above can be converted to a voltage output by adding a terminating resistor across the output This resistor must be 50 ohms per full scale voltage desired 50 ohms 1 V full scale 500 ohms 10 v full scale etc Following is a list of typical output ranges and required terminating resistance 98507600 Rev D 2 5 EC9850 SO ANALYZER OPERATION MANUAL Desired Output Volts Terminating Resistance Ohms 10 Volts 500 Ohms When using voltage output the source resistance is 1000 ohms The recorder or DAS input resistance should be greater than 500K ohms for a measurement error no greater than 1 2 1 2 2 Sample Gas Connections C
111. r reference purposes only The reading is not used to compensate future values regardless of whether COMPENSATION iS ENABLED Or DISABLED The analyzer automatically runs a zero cycle background at midnight using the external zero air and compensates future readings based on this value Use of a Level 1 span calibration with sPAN COMP ENABLED and CALIBRATION TIMED Of CAL MODE CYCLE adjusts the instrument gain so the output of the instrument agrees with the concentration expected for span gas The previously determined multipoint calibration curve is used to verify that the analyzer output is linear Note that a Level 1 span calibration requires external zero and span standards connected via the EZS valve assembly It is recommended that the SO gas cylinder or permeation device be checked against the previous instrument calibration curve immediately after the generation of the calibration curve refer to section 12 of Calibration Reference 1 It is also recommended that the concentration of this pollutant source be between 70 and 90 of the upper range limit of the analyzer and previous calibration curve Subsequent use of this pollutant source with AZS and compensation enabled adjusts the span of the instrument to agree with the previous calibration line Specific guidelines are contained in Calibration Reference 1 for use of Level 1 span checks section 12 and certification of gas cylinders or permeation devices t
112. re 20 40 60 80 10046 of Full Scale minimum of ILPM Record the concentration and the analyzer s response for each concentration after a stabilization period of 15 minutes per point Plot the analyzer s responses versus the corresponding SO concentrations and draw or calculate the calibration curve as discussed in the following section 3 2 2 4 Calculating Multipoint Calibration Results 3 2 2 4 1 Manual Calculations Used when Excel not available 98507600 Rev D 98507600 CHAPTER 3 CALIBRATION Determine the percent difference between instrument response and the calculated concentration using the following equation Instrument Response Calculated Concentration 100 Percent Difference Calculated Concentration If the difference between values is less than 1 then the instrument is within specifications Otherwise a Leak Check and or service is required 3 2 2 4 2 Using Microsoft Excel to Display Multipoint Calibration Results Produce a X Y scatter plot of the data with the calculated SO concentration in the X axis and the instruments response concentration in the Y axis Right mouse click on any data point to bring up the data formatting menu shown in Figure 3 2 Calculated Measured MULTIPOINT CHECK Format Data Series Chart Measured co e e Source Data Add Trendline Clear 0 Lo T T T
113. ribed below Each of the 4 units of the code represent a column below the description within the box of the corresponding unit explains the failure status of various components if any and more detailed descriptions are outlines below the table Unit 1 Digit 274 Digit 377 Digit 4 Digit 0 NO FAILURE NO FAILURE NO FAILURE GRAV 1 CHOPFAIL REFFAIL ZEROON GRAV 2 LAMPFAIL HEATERFAIL OUT OF SERVICE VOL 3 CHOPFAIL REFFAIL ZEROON VOL LAMPFAIL HEATERFAIL OUT OF SERVICE 4 FLOWFAIL COOLERFAIL HV FAIL GRAV B CHOPFAIL REFFAIL ZEROON GRAV FLOWFAIL COOLERFAIL HV FAIL 6 LAMPFAIL HEATERFAIL OUT OF SERVICE VOL FLOWFAIL COOLERFAIL HV FAIL 7 CHOPFAIL REFFAIL ZEROON VOL LAMPFAIL HEATERFAIL OUT OF SERVICE FLOWFAIL COOLERFAIL HV FAIL 8 SYSFAIL CVFAIL PS FAIL GRAV SPANON 9 CHOPFAIL REFFAIL ZEROON GRAV SYSFAIL CVFAIL PS FAIL SPANON A LAMPFAIL HEATERFAIL OUT OF SERVICE VOL SYSFAIL CVFAIL PS FAIL SPANON B CHOPFAIL REFFAIL ZEROON VOL LAMPFAIL HEATERFAIL OUT OF SERVICE SPANON SYSFAIL CVFAIL PS FAIL S FLOWFAIL COOLERFAIL HV FAIL GRAV SYSFAIL CVFAIL PS FAIL SPANON D CHOPFAIL REFFAIL ZEROON GRAV FLOWFAIL COOLERFAIL HV FAIL SPANON SYSFAIL CVFAIL PS FA
114. rt see the remainder of this section for instructions on preparing calibration gas Allow the analyzer to sample the gas until a stable reading is obtained typically 15 minutes From the primary screen start the calibration sequence by pressing either the Up or Down arrow or Y until the display prompts START MANUAL CALIBRATION Pressing the Select key will allow you to choose from No SPAN or ZERO Confirm that the display reads span and press Enter 4 A backlit cursor will be displayed on the SO concentration display Use the Select key to move the position of the backlit cursor and the Up and Down arrow keys to increment and decrement the value of the backlit digit until the calibration concentration value is displayed When the desired concentration is displayed press Enter Then move the backlit cursor to the INSTRUMENT GAIN field The instrument gain is automatically calculated by the analyzer Press Enter to confirm this value Press Exit to return to the primary screen Note The auto zero function of the EC9850 eliminates the need for a traditional zero calibration This completes the span calibration of the EC9850 analyzer 98507600 Rev D CHAPTER 3 CALIBRATION 3 2 Multipoint Calibration Either of the two procedures described in this chapter may be used to calibrate the EC9850 SO2 analyzer Before beginning a multipoint calibration of the instrument a qualified service technicia
115. s are in floating point format The STATUS woRD indicates the instrument status in hex using the following format Bit 15 MSB SYSFAIL Bit 14 FLOWFAIL Bit 13 LAMPFAIL Bit 12 CHOPFAIL Bit 11 CVFAIL Bit 10 COOLERFAIL Bit 9 HEATERFAIL Bit 8 REFFAIL Bit 7 PS FAIL Bit 6 HV FAIL Bit 5 OUT OF SERVICE Bit 4 instrument is in zero mode Bit 3 instrument is in span mode Bit 2 instrument is performing a background Bit 1 selected CLEAR gt MG M3 Bit O LSB reserved Command DEVENT Function Dumps the Event Log message buffer to the serial port 4 18 98507600 Rev D 98507600 CHAPTER 4 DIGITAL COMMUNICATION Format DEVENT DEVICE I D gt TERMINATOR Device response XX Message XX lt CR gt lt LF gt OCCURRED AT HH MM DD MON YY lt CR gt lt LF gt The last 100 messages are reported xx is the index into the event log message buffer 99 oldest point reported first 0 newest point reported last The message field is null if no message exists Command DGAIN Function Dumps instrument gain data to the serial port Format DGAIN DEVICE I D gt TERMINATOR Device response INSTRUMENT GAIN lt CR gt lt LF gt Command DINSTR Function Dumps the INSTRUMENT STATUS MENU variables to the serial port Format DSTATUS DEVICE I
116. ssued to request the next packet of data GETDATA DEVICE I D 2 1 REQUEST TERMINATOR Where REQUEST 0 to retransmit previous packet logged data 1 to transmit next block of packet data 4 23 EC9850 SO ANALYZER OPERATION MANUAL Device response Refer to command 2 in the USB protocol specification The complete USB packet format is used for the response to this serial command 4 5 USB Communication The USB port is located on the rear of the analyzer This cannot be multidropped with other analyzers but multiple analyzers can be connected to a single USB port on a computer by using a USB hub This connection is ideal for collecting data from a standalone analyzer or using a laptop that may not have a serial port 4 5 1 Installing the Driver on a PC The following are instructions to install the EC9850 analyzer to a computer through the USB connection It will provide efficient communication between the analyzer and computer with the use of the EC9500 Communicator software described in section 4 6 Note Screen shots and instructions below apply to Windows XP but will be similar for any other Windows operating system 1 Turn on computer and log in 2 Connect the analyzer by USB cable to the USB port on the rear of the computer 3 After 10 20 seconds the dialog box shown in should appear If no dialog box appears open the Control Panel and double click Add New Hardware 4 24 98507600
117. sults of the background are stored in the EVENT LOG 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION SO2 CAL PRESSURE This is the measured ambient pressure during the last SO calibration 2 5 6 Test Menu TEST MENU OUTPUT TEST MENU DIAGNOSTIC MENU CALCULATION FACTORS MEASUREMENT GAIN 8 PRES TEMP FLOW COMP ON DIAGNOSTIC MODE OPERATE CONTROL LOOP ENABLED TEST MEASURE 0 502 0 00000 PPM Figure 2 13 Test Menu The TEST MENU includes a series of submenus containing information and control settings for testing and verifying instrument functions The operator may make changes to settings however when the instrument is returned to normal operation the instrument s automatic control function resumes Changes made from this menu are for diagnostic and test purposes only MEASUREMENT GAIN Entries are microprocessor controlled settings of 1 2 4 8 16 32 64 128 PRES TEMP FLOW COMP Either oN or OFF OFF is used when running diagnostics to see fluctuations in readings oN is used to compensate for environmental fluctuations that might affect readings DIAGNOSTIC MODE Allows the operator to choose OPERATE OPTIC ELECT Or PREAMP During measurement set to OPERATE During diagnostic testing set to the system to be diagnosed CONTROL LOOP Allows the operator to choose ENABLED Or DISABLED When ENABLED is selected the microprocessor maintains control
118. t Menu SO2 OUTPUT MENU RANGE 0 500 PPM OUTPUT TYPE CURRENT CURRENT RANGE 0 20 MA FULL SCALE 0 00 ZERO ADJUST 0 00 OVER RANGE 20 00 PPM OVER RANGING DISABLED Figure 2 21 Analog Output Menu Current Output 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION RANGE Set upper range limit in digits to desired SO concentration This value cannot exceed the ovER RANGE value OUTPUT TYPE Setting must match the jumper settings on the 50 pin I O board if installed CURRENT Or VOLTAGE CURRENT RANGE Choices are 0 to 20 MA 2 to 20 MA and4 to 20 MA FULL SCALE X XX a correction factor for full scale setting Used when calibrating the analog outputs ZERO ADJUST X XX a correction factor for the zero setting Used when calibrating the analog outputs OVER RANGE Set to desired over range value This value cannot be set below the RANGE value See section 2 6 This is the alternate scale the recorder or DAS indicates when over ranging is active and enabled When 90 of the set range is reached this auto range is effective When 80 of the original range is reached it returns to the original range OVER RANGING Set to ENABLED Or DISABLED to turn the over ranging feature on or off 2 5 14 2 SO Voltage Output Menu SO2 OUTPUT MENU RANGE 0 500 PPM OUTPUT TYPE VOLTAGE OFFSET 0 FULL SCALE 0 00 ZERO ADJUST 0 00 OVER RANGE 20 00 PPM OVER RAN
119. talled in front of the sample inlet Zero and Span gas must pass through this filter EC9850B series analyzer must be fitted with a zero air scrubber If the units in the MEASUREMENT MENU are changed from volumetric to gra vimetric or gravimetric to volumetric the analyzer must be re calibrated The analyzer must be operated and maintained in accordance with this op eration manual The following menu selections must be used 98507600 Rev D 98507600 CHAPTER 1 DESCRIPTION INTERFACE MENU ANALOG OUTPUT MENU RANGE 0 05 PPM to 1 0 PPM OVER RANGING ENABLED Or DISABL INSTRUMENT MENU MEASUREMENT MENU FILTER TYPE KALMAN CALIBRATION MENU CALIBRATION MANUAL TIMI SPAN COMP DISABLED BACKGROUND not DISABLED TEST MENU PRES TEMP FLOW COMP ON DIAGNOSTIC MODE OPERATE o The Service switch must be positioned to IN The EC9850 series analyzers are U S EPA equivalent with or without the following options items a valve assembly for external zero span EZS o rack mount assembly o 50 pin connector PCA o exhaust scrubber o internal zero span assembly IZS 1 5 EC9850 SO ANALYZER OPERATION MANUAL This page is intentionally left blank 1 6 98507600 Rev D CHAPTER 2 INSTALLATION AND OPERATION 2 0 INSTALLATION AND OPERA
120. ted to 25 C and 760 Torr slpm 6 Sample the concentration until a stable response is obtained 7 From the primary screen press the Up or Down arrow key Respond sPAN to the screen query START MANUAL CALIBRATION by pressing the Up or Down arrow key then lt Enter gt The cursor now appears in the first digit of the concentration field 8 Use the Select and arrow keys to input the SO span point concentration calculated in step 5 Use the Select key to select the digit to be changed and the Up or Down arrow keys to change the value 9 Press Enter to confirm the input value Record the SO concentration and the analyzer s stable response 10 Then move the cursor to the INSTRUMENT GAIN field Record the instrument gain value for future reference Press Exit to return to the primary screen 3 2 2 Procedure For 5 Point Multipoint Calibration 3 2 2 1 Principle Multipoint calibration consists of five 5 concentrations across the instruments operating range The concentration levels are derived to determine the accuracy between calculated and expected values of the analyzer using a simple Excel spreadsheet This procedure is only to be carried out on a 6 monthly basis or after the analyzer has had major repair as per AS3580 4 1 and U S EPA Standards 3 2 2 2 Apparatus 98507600 Rev D The apparatus discussed in section 3 2 1 2 above should be used to perform the Multipoint Calibration In addition to this
121. the analyzer when Multidrop RS232 communications is used 98507600 Rev D 2 19 EC9850 SO ANALYZER OPERATION MANUAL 2 5 4 Measurement Menu MEASUREMENT MENU UNIT SELECTION uG M3 CONVERSION TEMP 0 DEG C DECIMAL PLACES 2 53 AVERAGE PERIOD 1 MINUTE FILTER TYPE KALMAN NOISE 0 204 PPB ZERO OFFSET 0 00 PPB Figure 2 10 Measurement Menu The MEASUREMENT MENU consists of items needed for basic operation and data integrity UNIT SELECTION PPM parts per million nc w milligrams per cubic meter nanograms per cubic meter uc w micrograms per cubic meter PPT parts per trillion or PPB parts per billion Note If the gravimetric units are selected mG m nG V then the conversion factors listed below will apply depending on the CONVERSION TEMP selected To convert 1 PPB Gas to ug m 0 C 20 Zoe Multiply by Note If the units in the MEASUREMENT MENU are changed from volumetric to gravimetric or gravimetric to volumetric the analyzer must be re calibrated in order to meet U S EPA requirements CONVERSION TEMP Sets the temperature that should be used in internal calculations to convert the concentration from volumetric units ppm ppb ppt into gravimetric units mG M pG M nG M in DEGREES CELCIUS 0 20 25 This menu option is only displayed when the gravimetric units are selected 2 20 98507600 Rev
122. the equipment 9 Whether repair work is under warranty or is to be billed and a purchase order number for any work to be billed TU See 95 When you call in our Customer Service Representative will assign a Return Material Authorization RMA number to your shipment and initiate the necessary paperwork to process your equipment as soon as it reaches us Please include this RMA number when you return equipment preferably both inside and outside the shipping container This will ensure that your equipment receives the most prompt attention possible 1 the RMA number is not marked the outside of the shipping container the shipment will be rejected when it reaches our facility and returned at your expense Your assistance in this matter will enable us to serve you better We appreciate your cooperation and support of our products and services 6 98507600 Rev D FRONT MATTER Claims for Damaged Shipments and Shipping Discrepancies Damaged Shipment 1 Inspect all instruments thoroughly on receipt Check material in the container s against the enclosed packing list If the contents are damaged and or the instrument fails to operate properly notify the carrier and Ecotech immediately 2 following documents are necessary to support claims Original freight bill and bill of lading Original invoice or photocopy of original invoice Copy of packing list Photographs of damaged equipment and container You ma
123. tion 2 23 9850 SO ANALYZER OPERATION MANUAL 2 5 5 2 Manual Calibration The following screen appears when CALIBRATION MANUAL is selected CALIBRATION MENU CALIBRATION MANUAL CAL MODE MEASURE CALIBRATION EXTERNAL BACKGROUND ENABLED BACK INTERVAL 24 HOURS SO2 CAL PRESSURE 750 0 TORR Figure 2 12 Manual Calibration Menu CALIBRATION Designates TIMED or MANUAL calibration control CAL MODE A choice of MEASURE normal mode CYCLE zero span SPAN span valve or ZERO zero valve The choice is based on the valve the operator wants to open Selecting CYCLE starts an AZS cycle which is discussed in Chapter 3 CALIBRATION The choice of INTERNAL Or EXTERNAL valves that will operate during a zero span check The EZS valve option must be installed to use EXTERNAL BACKGROUND A choice of START DISABLED or ENABLED If START is selected then when lt Enter gt is pressed the instrument begins an auto zero background cycle If DISABLED is selected then the instrument will not run the normal auto zero background cycle If ENABLED is selected then the instrument will run the normal auto zero background cycle Note For U S EPA designated use the background cycle must not be disabled BACK INTERVAL Set time in HOURS 2 4 6 8 12 24 microprocessor controlled field that indicates when an autozero cycle will begin The re
124. tion Factor A 112 CO Linearisation Factor B 113 CO Linearisation Factor C 114 CO Linearisation Factor D 115 CO Linearisation Factor E 116 Instrument Units 0 PPM 1 2 3 mG M 4 1G M 5 06 117 Background Measure Time In seconds 118 Sample Fill Time In seconds 119 Sample Measure Time In seconds 120 Aux Measure Time In seconds 121 Aux Sample Fill Time In seconds 122 Background Fill Time In seconds 123 Zero Fill Time In seconds 124 Zero Measure Time In seconds 125 Span Fill Time In seconds 126 Span Measure Time In seconds 127 Span Purge Time In seconds 128 Background Pause Time In seconds 129 Background Interleave Factor In seconds 130 Calibration Pressure 2 131 AUX Instrument Gain 132 Background voltage 133 AUX Background Voltage 134 O3 Generator Output PPM 135 O3 Generator On Off 136 Calibration Point 1 PPM 137 Calibration Point 2 PPM 138 Calibration Point 3 PPM 139 Calibration Point 4 PPM 140 Calibration Point 5 PPM 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL 141 Desired Pump Flow SLPM 142 Actual Pump Flow SLPM 143 Set Lamp Current 144 Lamp Current mA 145 Cycle Time Minutes 146 Analog GND Offset Volts 98507600 Rev D EC9850 SO ANALYZER OPERATION MANUAL Appendix B Failure Status descriptions The failure status codes provided by the 9800 downloader are desc
125. to abort the current mode and return to the measure mode Format ABORT DEVICE I D gt TERMINATOR Device response lt ACK gt if the unit under test is able to perform the command lt NAK gt if not Command DAVGC Function Sends the current average concentration data to the serial port Format DAVGC lt DEVICE I D gt TERMINATOR Device response GAS lt SPACE gt STATUS WORD lt CR gt lt LF gt All numbers are in floating point format See the pcoNc command for an explanation of the STATUS WORD Command DAZSC Function Commands the addressed device to perform a zero span cycle The system returns to the measure mode when the cycle has completed Format DAZSC lt DEVICE I D gt TERMINATOR Device response lt ACK gt if the unit under test is able to perform the command lt NAK gt if not Command DCOMM Function Performs a character dump to the serial output when called 98507600 Rev D 4 17 EC9850 SO ANALYZER OPERATION MANUAL Format DCOMM DEVICE I D gt TERMINATOR Device response lt ACK gt if the unit under test performs a successful loopback lt Nak gt if not Command DCONC Function Sends the current instantaneous concentration data to the serial port Format DCONC lt DEVICE I D gt TERMINATOR Device response GAS lt SPACE gt STATUS WORD lt CR gt lt LF gt All number
126. u 2 31 M Main Ment aisen a eite 2 18 EC9850 SO ANALYZER OPERATION MANUAL Main Menu 2 18 Manual Calibration Menu sese 2 24 Manual Calibration Menu illustration 2 24 Measurement 2 20 Measurement Menu illustration 2 20 Multidrop communications esses 4 13 Multipoint calibration sse 3 3 Multipoint calibration with over ranging 3 14 Offset 2 40 Operation nie aee etn ere te 2 11 Optional 50 Pin Connector Board Front illustration 2 3 Original protocol seen 4 10 Output Test MENU erea tente 2 26 Output Test Menu 2 26 Over range adjustment essere 2 40 Over Range as Seen on a Strip Chart Recorder illustration 2 41 OV r r lig Hg os nete nem 3 14 P Password protection eese 2 43 Preprocessor Pots screen 10 2 26 Preprocessor Pots Screen illustration 2 26 8 rente ed 2 17 Primary Screen illustration 2 2 2 17 Protocol selections esee 4 10 Protocol Bavarian 4 11 Protocol enhanced 4 12 Protocol original eee 4 1
127. urrent Readings us rete rid ree eher a bee e e REED SE Lade 4 32 4 72 Remote Mode vi t ERE SR RR e i 4 33 27 3 DOWRLOQ s m e RR OE Reto ne Eee en aseo abe ades ceo 4 34 4 7 4 Firmware Update for the Network Interface esee 4 36 INDEX Corr 1 Pid pO c 4 3 DESCRIPTION eA REOR be EROR ARE 4 3 Pid pg 4 7 98507600 Rev D FRONT MATTER This page is intentionally left blank 98507600 Rev D FRONT MATTER Manual History This manual is the combination of two previous versions which have now been merged into one document to cater for the continuing development of the EC9800 series analyzers The original manuals were o ML9850 Operation Manual PN 98500022 Rev V December 1998 o ML9850B Operation Manual PN 98507005 Rev P July 1999 The scope of this new manual covers the following analyzers o EC9850 Sulfur Dioxide Analyzer A Series PN 98501000 100 o EC9850B Sulfur Dioxide Analyzer B Series PN 98507000 1 Both of the instruments are Manufactured by Ecotech P L in Australia and support the new SMD Microprocessor Board Part number 98000063 4 This manual is current for firmware version 1 33 and above Ecotech Manual ID MAN 0022 Manual PN 98507600 Current Revision D Date Released April 2007 Description EC9850 Sulfur Dioxide Analyzer
128. v D CHAPTER 4 DIGITAL COMMUNICATION Note The analyzer must be in Command mode before the Data Acquire mode can be used If the program was last used in Remote Terminal mode the analyzer may still be in Remote mode See section 4 6 2 2 for further details Note Data Acquire mode only retrieves data already logged by the analyzer To remotely instruct the analyzer to log data use the Remote Terminal mode 4 6 1 2 Viewing the Acquired Data 98507600 Rev D If the communication was successful a table of data similar to the below will be displayed 9800 Data Downloader ini x Comm Mode 03 12 01 08 59 0 001013 ppm 1 min C200 Inst Date Time Gas Uni Period FunctonStatus FaiueStaus Type __ M zi 03 12 01 10 00 0 001012 ppm 1 min M C200 Inst 03 12 01 10 01 0 001012 ppm 1 min M C200 Inst 03 12 01 10 02 0 001012 ppm 1 min M c200 Inst 03 12 01 10 03 0 001012 ppm 3 min M C200 03 12 01 10 03 0 001012 1 min M C200 Inst 03 12 01 10 04 0 001012 ppm 1 min M C200 Inst 03 12 01 10 05 0 001012 ppm 1 min M C200 Inst 03 12 01 10 06 0 001012 ppm 3 min M C200 03 12 01 10 06 0 001012 ppm 1 min M C200 Inst 03 12 01 10 07 0 001012 ppm 1 min M C200 Inst 03 12 01 10 08 0 001012 ppm 1 min M C200 Inst 03 12 01 10 09 0 001012 ppm 3 min M C200 Avg 03 12 01 10 08 0 001012 ppm 1 min M C200 Inst 03 12 01 10 10 0 001012 ppm 1 min M C200 Inst 03 12 01 10 11 0 001012 ppm 1 min M C200
129. w which unit in the multidrop is responding Any command that would cause two units on the multidrop to respond at the same time must be avoided If more than one unit attempts to respond on the common transmit line a data collision will occur destroying both messages 4 4 4 2 Programming Instrument Identifiers Note The Main Gas ID can be set manually in the Instrument Menu or the Interface Menu Refer to section 2 5 3 or 2 5 12 for further details This 1s the preferred method The command is the only command used to set the Instrument I D for a given analyzer The instrument can then be used standalone or as one of several multidrop daisy chain analyzers The format of this command is Where PIXXX YYY lt CR gt xxx is the unit I D yyy is the unit serial number The parameter xxx is the unit I D and must be three characters Unit I D s such as 1 should be programmed as 001 The second parameter vvv is an optional serial number This serial num ber is reported in the response to the Bavarian command but is otherwise unused Only one analyzer at a time may be programmed with an I D Do not issue this command with multiple units on a multidrop 4 4 4 3 Examples g g The string P1001 lt cR gt will program a unit to the I D of 001 The string P1001 123 will program a unit to the I D of 001 with a serial number of 123 The string P1003 004 will program an EC9850 for a main ID of 003 and a s
130. y Bavarian network applications Note specifying the Bavarian protocol still allows the user to access the 9800 command set The third protocol provided is the enhanced protocol The enhanced protocol provides a more robust handshaking environment as specified in section 4 4 3 9 4 4 3 1 Original Protocol This protocol is provided for back compatibility with pervious versions before Version 2 05 of 9800B software There are a number of idiosyncrasies in the original protocol that are preserved to allow existing applications to use upgraded software without modifying their interface 4 4 3 2 Command Acknowledgment n For 9800 style commands that provide a data response the data response itself is the acknowledgment For 9800 style commands that do not provide a data response the ac knowledgment is the returned ASCII string 98507600 Rev D o CHAPTER 4 DIGITAL COMMUNICATION For Bavarian Network commands no acknowledgment is returned 4 4 3 3 Negative Command Acknowledgment g For 9800 commands if a valid Instrument I D is received with an invalid command string the message INVALID COMMAND is sent For Bavarian Network commands no negative command acknowledgment is sent 4 4 3 4 Original Protocol Idiosyncrasies g g g Block check characters are not checked on Bavarian commands The lt 5 gt character is ignored The lt gt character is a valid termination for Bavarian comm
131. y want to keep a copy of these documents for your records also Refer to the instrument name model number serial number sales order number and your purchase order number on all claims Upon receipt of a claim we will advise you of the disposition of your equipment for repair or replacement Shipping Discrepancies Check all containers against the packing list immediately on receipt If a shortage or other discrepancy is found notify the carrier and Ecotech immediately We will not be responsible for shortages against the packing list unless they are reported promptly 98507600 Rev D 7 EC9850 SO ANALYZER OPERATION MANUAL ECOTECEH Service and Spare Parts For world wide customer service amp spare parts contact ECOTECH Address Ecotech Pty Ltd 1492 Ferntree Gully Rd Knoxfield Australia VIC 3180 Phone 61 1300 364 946 Fax 61 1300 668 763 Email Service ecotech ecotech com au Email Spare Parts parts ecotech com au Web www ecotech com au Our Service Response Centre handles product information application assistance factory repair training service maintenance agreements and technical assistance 98507600 Rev D FRONT MATTER WARNING Avoid smoking in the vicinity of the analyzer Due to the complex chemical makeup of tobacco smoke smoke drawn into the sample line may result in incorrect readings Furthermore tobacco smoke has been shown to contaminate converter and scrubber materials critic
Download Pdf Manuals
Related Search
Related Contents
Tracer 第49号 目 次 - 名古屋大学アイソトープ総合センター Xantrex Technology 175 PLUS Power Supply User Manual tactix™ - Tramsoft Copyright © All rights reserved.
Failed to retrieve file