1. GAYMAR AIR EXPRESS LAL P1500 LOW AIR LOSS THERAPY SYSTEM e REF C1500 Control Unit e REF M1500 Mattress Operator s Manual P N 11882 000 6 03 Air Express LAL Before you begin Table Of Contents Section Description Page 1 0 2 2 0 SYMBOLS FEST DRY RFC FR FRYN FOTY FFR HF RF NN FRY DF 2 3 0 Indications for USec nonna LLY LY 2 4 0 Contraindications stisni ieee ier EEEE ENEE Y RYN 2 5 0 E iaae Ee 3 6 0 5 4 7 0 Fealur6S eu ia GR saad GU RC CN GR ND Er ER GN ROC n 4 8 0 Operator Control Panel ee 5 9 0 Setting Up the System YY YY LL ALL LL FY Y nennen 5 10 0 Instructions for Use e Y CLYN 6 11 0 Transferring Patients In and out 0 san 7 12 0 CPR Cardio Pulmonary Resuscitation 0 ceeeeeeeeeeeee ees 7 13 0 System Cleaning Care and Storage cccecececeeneeeaeeeeeeeeeees 8 14 0 TPOUDIGSNOOUNG 42 4 52 9 15 0 Filter InstructionS na a a DG yD 10 16 0 Specifications Control Unit Yr YY YY YY Y YY Y Y Y YN 11 INTRODUCTION This Owner s Manual was designed to assist both caregivers and patients when using the P1500 Low Air Loss Therapy System It can be used as a continual information resource and we suggest that you have it available to answer most of the guestions that you may have regarding the set up cleaning and routine maintenance of the system Before you b
2. Gaymar customer service at 1 800 828 7341 Filter Removal Instructions Air Express LAL 15 0 Filter Instructions 15 1 Filter Removal Instructions Step 1 Using a finger or other suitable tool such as a pen pry up the end of the filter through the large vent opening Step 2 Next grasp the filter between the thumb and index finger and pull slowly from the housing 15 2 Filter Installation Instructions Step 1 Insert the first inch of the filter into the large vent opening in the rear housing Step 2 Next using a series of short strokes with a pen or similar tool move the filter completely up into the rear enclosure 10 Air Express LAL Specifications 16 0 Specifications Control Unit C1500 Enclosure Dimensions 10 x 12 x 4 1 2 25 cm x 30 cm 11 cm Weight 9 5 pounds 4 3 kg Power Cord Detachable 14 18 AWG minimum with ground wire Overcurrent Protection Primary Two 32mm x 6 3mm fuses 5A 250V T L Secondary One 32mm x 6 3mm fuses 15A 250V T L Input 120 VAC 60Hz 5A Operating Ambient 60 to 80 F 16 to 26 C Temperature Range Classification Class grounded equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Type BF equipment UL 544 CSA C22 2 NO 125 IPXO enclosed equipment without protection against ingress of water Continuous operation Electromagnetic CISPR 11 Classified a
3. accompanying documents Type BF eguipment AN Dangerous voltage 5 Protective earth pe Latex Free 3 0 Indications for Use This device is intended to assist in treating and preventing pressure ulcers 4 0 Contraindications Air support treatment is not recommended when spinal stability is a concern Air Express LAL Safety Precautions N 5 0 Safety Precautions Review the following SAFETY PRECAUTIONS prior to using the P1500 System e Risk of electric shock Refer servicing to qualified service personnel EXPLOSION HAZARD DO NOT USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS e Disinfect the P1500 system between patient installations Failure to disinfect may risk cross contamination and infection e Check patient at least every 8 hours or once per shift to assure proper system operation e Deflate mattress before performing CPR or CPR will be ineffective CAUTION e Some medical conditions may not respond to treatment of this type Patient s skin condition should be inspected regularly Consult physician if any redness or skin breakdown occurs e For grounding reliability plug only into a properly grounded outlet e Make certain all mattress straps are secured to the bed frame to prevent mattress from sliding and causing patient injury e Plexus mattresses are not intended to be AND DO NOT FUNCTION AS a patient fall safety device SIDE RAILS MUST BE USED WITH THE PLEXUS MATTRESS TO PREVENT FALLS unle
4. assembly loosely and roll the deflated air mattress starting at the head end Use the straps for containment Place in plastic bag for storage 14 0 Troubleshooting Symptom Corrective Action Mattress does not inflate or is too soft Make certain the air hose is not kinked cracked or split Verify the power switch is illuminated signifying the control unit has power Verify the hose quick connect is fully inserted Verify that all 16 air cushions are connected to the air manifold Inspect and or clean intake air filter on the control unit Loss of power Make certain the power receptacle has power and the light on the front panel is illuminated and the patient comfort setting is 1 or greater MAINTENANCE There is only one operator maintenance item for the P1500 system The air filter must be cleaned every 30 days Remove the filter see filter removal instructions located near the mattress hose connector Wash the filter with soap and water and rinse thoroughly Allow the filter to dry completely and reinstall Filters that are brittle or crumble should be replaced Replacement filters can be ordered under P N 20228 Failure to clean the filter can result in damage to the control unit and or inadequate patient support Periodic testing for function and or repairs should be referred to qualified personnel in accordance with P1500 service manual Factory repairs can be arranged through your local Plexus dealer or by contacting
5. external surfaces of the Control Unit hoses and mattress Allow contact time according to manufacturers instructions The solution contact time is what makes disinfection effective Wipe down the mattress with a clean dry cloth to remove any excess disinfectant Top sheets of mattress may also be laundered between patient uses or as required to maintain good patient hygiene Fill the washing machine with warm water 70 140 F or 21 60 C Add one cup of laundry detergent Place no more than four top sheets in a single extra large load capacity washing machine When wash cycle is complete remove promptly from machine and ensure all excess water is drained from load Place dryer on LOWEST heat setting or AIR FLUFF if available until dry Verify top sheet is completely dry before placing under patient 5 If individual air cells of mattress become soiled clean and disinfect as described above or simply replace air cell with a clean replacement Single air cell replacement can be successfully achieved with patient remaining on the mattress Air Express LAL Storage Troubleshooting 13 2 Storage Control Unit 1 Check the AC power cord and plug for abrasions or excessive wear 2 Coil power cord up and place both cord and Control Unit in plastic bag for storage Mattress 1 Check the air manifold of the mattress for kinks or breaks Replace if necessary 2 To deflate disconnect hose assembly from Control Unit 3 Coil the hose
6. 5 Connect the free end of the tubing 3 to the respective mating connector located on the side of the pump 6 This quick connect fitting can be secured by simply mating the two ends together in a linear but opposing direction 6 Check that the hose is secure by gently pulling Ensure the air hose is not kinked or tucked under the mattress Cover P1500 air mattress with the PLEXUS top sheet and attach to the mattress D rings provided Plug power cord into a properly grounded outlet Unit will enter STAND BY mode when dial is set to 0 Settings greater than 1 will automatically initiate the units operation 9 Place patient on the air mattress Note The control unit can inflate the mattress with the patient laying on it To assist in patient ingress egress as well as normal nursing procedure the air mattress can be maximally inflated by fully rotating the comfort adjust dial in a clockwise direction NOTE To reduce the risk of skin break down MAX Inflate should not be maintained for greater than 30 minutes Instructions for Use Air Express LAL 10 0 Instructions for Use Check patient at least every 8 hours or once per shift to assure proper system inflation 10 1 Use With M1500 Mattress 1 For proper inflation and patient comfort adjust pressure by rotating dial initially to MAX After maximum inflation has been attained sequentially reduce the pressure setting i e 9 8 7 etc until four fingers stacked v
7. egin N Important Before using the P1500 System please read and understand this manual and all safety precautions prior to each application Only gualified medical service personnel should attempt to repair this device A Service Manual is available for use by qualified personnel For assistance contact your local dealer If additional assistance is needed contact Gaymar s technical service department International Customers Direct 716 662 8636 Fax 716 662 0730 Domestic Customers Toll Free 800 828 7341 Fax 716 662 8795 Warranty Symbols Indications for Use Contraindications Air Express LAL 1 0 Warranty The Control Unit is warranted free of defects in material and workmanship for a period of two 2 years The Mattress is warranted free of defects in material and workmanship each for a period of one 1 year The Control Unit and Mattress are warranted under the terms and conditions of the Gaymar warranty in place at the time of purchase A copy of the warranty is available upon request Gaymar disclaims all implied warranties including but not limited to the implied warranties of merchantability and of fitness for a particular purpose Control Units may be returned to the factory for servicing Contact Gaymar Customer Service for return authorization prior to return Please contact Gaymar Customer Service if you have warranty questions Toll Free 800 828 7341 2 0 Symbols N Attention consult
8. ented above a two inch convoluted foam base The material used in their construction is a low shear nylon which aids in the reduction of shear forces The air cushions are designed with orifices to provide continuous low air loss to aid in the prevention of skin maceration Covering the entire mattress assembly is a low shear producing 70 Denier nylon top sheet The top layer of the top sheet is coated with a vapor permeable polyurethane film This coating creates a monolithic osmotic membrane impermeable to liquids air and bacteria though still permeable to water vapor The bottom layer of the top sheet is a quilted soun bonded polyester fiberfill which exhibits less frictional resistance to nylon than that of the patient s skin The significance of this is that when the patient moves the top sheet tends to move relative to the air cushions rather than relative to the patient A secondary feature of this quilting is that it acts to provide a diffusion layer by which water vapor can be drawn away from the patient Concentrations of moisture will diffuse throughout the top sheet thus resulting in efficient evaporation and subsequent reduced skin maceration 7 0 Features 7 1 P1500 CONTROL UNIT FEATURES Compact light weight and easy to use Durable plastic housing with bedside bracket for easy placement on the foot board of bed Push button connector for quick disconnection and rapid deflation as needed for CPR Dialed Soft Firm Comfort Cont
9. ertically clearance exists between the underlying foam and the patients pelvic region 2 For mattress deflation when CPR is needed disconnect the hose connector from the controller Mattress will be fully deflated in less than 15 seconds P1500 Diagram Air Express LAL Transferring Patients CPR 11 0 Transferring Patients In and Out of Bed 1 For patient transfer turn dial to the MAX INFLATE position This will maximally inflate the mattress to provide a firm and stable support surface Upon completion of procedure return the control unit to the appropriate patient comfort setting 12 0 CPR Cardio Pulmonary Resuscitation To deflate mattress for CPR 1 Disconnect the quick connect coupling of the mattress hose from the Control Unit Mattress will deflate in 15 seconds 2 Proceed with CPR procedures Cleaning Care Storage Air Express LAL 13 0 System Cleaning Care and Storage 13 1 Cleaning Disconnect the AC power cord from the wall outlet before attempting to clean the Control Unit Do not heat or steam autoclave any component of the system 1 To clean use soap water and a clean cloth to wipe down the Control Unit power cord hoses and mattress Do not use abrasive cleaners on the mattress Wipe dry with a clean dry cloth Note Blood and other body fluids must be thoroughly cleaned from all surfaces before applying disinfectants 2 Apply any EPA approved disinfectant to the
10. rol P1500 LOW AIR LOSS MATTRESS FEATURES e Horizontal tubular cell construction e Individual air cushions designed for maximum support area and pressure distribution e Dimensions 80 x 35 x 8 when fully inflated Air Express LAL Operator Control Panel Setup 8 0 Operator Control Panel NIRISEXPRESS LAL P 1500 3 4 Low Air Loss Mattress System do 1 Power Indicator An amber light indicates AC power is present to the controller the controller is in standby A green light indicates system is on and running 2 Comfort Control Dial This dial is used to adjust the patient comfort setting In the position shown in the figure above the control unit is in standby 3 Comfort Setting Indicator Indicator will be near current set point 9 0 Setting Up the System 1 Remove standard mattress from bed frame 1 Refer to diagram of mattress and bed 2 Replace standard mattress with the P1500 mattress 2 Make sure to orient mattress so the air tubing 3 exits the mattress at the foot of the bed Loop each of the four black nylon straps 4 to the bed frame and secure with the D Rings provided Suspend control unit from the footboard using the two hooks located on the back of the unit as shown 5 If no footboard exists place unit on a flat surface near the foot end of the bed Be careful not to position the unit on the floor in such a manner that it may become a hazard to foot traffic
11. s Class B Group 1 ISM equipment Compatibility 11 Air Express LAL GAYMAR GAYMAR INDUSTRIES INC 10 Centre Dr Orchard Park New York 14127 2295 Phone 1 800 828 7341 716 662 2551 Outside the USA 716 662 8636 Authorized CE representative Europe INVATECH GmbH amp Co Wandsbeker K nigstra e 27 29 D 22041 Hamburg Germany
12. ss determined unnecessary based on the facility protocol or the patient s medical needs as determined by the facility IN WHICH CASE THE USE OF OTHER SUITABLE PATIENT SAFETY MEASURES ARE RECOMMENDED e Use minimal layers of sheeting and incontinence pads Too many layers between the patient s skin and the support surface will reduce the effectiveness of the system _ Do not pull linens tightly over mattress Tight sheets can cause hammocking and reduce effectiveness of treatment e Do not block the pump intake on the enclosure Any obstruction to airflow particularly a dirty filter will cause excessive heat and greatly shorten the life of the blower Keep unit away from radiators or other heat sources e Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion Portable and RF communications equipment can affect medical electrical equipment Description Features Air Express LAL 6 0 Description The P1500 is a portable low air loss mattress replacement designed to economically provide therapeutic benefit to patients suffering from or at risk of developing pressure ulcers The system consisting of a Control Unit an Air Flotation Mattress and a Top Sheet is designed to provide pressure management therapy and patient comfort The mattress base consists of sixteen transverse air cushions Each is over six inches tall inflated height and ori