Home
        OPERATOR`S MANUAL
         Contents
1.     Alarm Summary    42    L AUDIBLE ALARM OFF button has not been pressed  after alarm occurs     e Alarm condition present      Audible alarm    Flashing display    e Alarm condition ceases to exist  non latching option       Audible alarm stops    Display stops flashing    e Alarm condition ceases to exist  latching option     Audible alarm and      Flashing display continues until AUDIBLE  ALARM OFF button is pressed    Start Up and Use    2 AUDIBLE ALARM OFF button has been pressed  after alarm occurs     e Alarm condition still present    Ee      Noaudible alarm for silence duration    AUDIBLE ALARM OFF indicator lights    Flashing display continues    e Alarm condition ceases to exist      Noaudible alarm    Noflashing display    NoAUDIBLE ALARM OFF indicator    3  Audible alarm silence duration set to OFF   e Noalarm condition is present      No audible alarm    Audible alarm off reminder at 3 minute intervals    AUDIBLE ALARM OFF indicator flashes    e Alarm condition occurs      Noaudible alarm     Audible alarm off reminder at 3 minute intervals     AUDIBLE ALARM OFF indicator flashes     Display flashes continuously for out of limit  parameter    ADJUSTING ALARM LIMITS    Overview    When the N  3000 is first turned on  alarm limits will be set to  their power on default values  Power on default alarm limits  can be changed by qualified personnel with the N 3000 in the  configuration mode described in the N 3000 service manual   You can change alarm limits from de
2.     e Accuracy can be affected by     incorrect sensor application and use      significant levels of dysfunctional hemoglobins   intradermal or intravenous dyes  bright light   excessive patient movement  venous pulsations   electrosurgical interference      placement of a sensor on an extremity that has a  blood pressure cuff  arterial catheter  or  intravascular line    e Observe all instructions  warnings  and cautions in  this manual and in the sensor directions for use     EMI INTERFERENCE    The N 3000 has been tested and found to comply with  the limits for medical devices to the EN60601 1 2 1993   EN60601 1 2 1994  Medical Device Directive 93 42 EEC   These limits are designed to provide reasonable  protection against harmful electromagnetic  interference  EMI  in a typical medical installation  In  addition  the device contains alarms which notify the  user when EMI is detected  although detection cannot  be assured in all cases      57    Troubleshooting and Maintenance    58    Because of the proliferation of radio frequency  transmitting equipment and other sources of electrical  noise in the health care environment  for example   cellular phones  mobile two way radios  electrical  appliances   it is possible that high levels of EMI  interference due to close proximity or strength of a  source  may result in disruption of performance of the  N 3000     The N 3000 generates  uses  and can radiate radio frequency  energy and  if not installed and used in accordance
3.    Start Up and Use    To adjust neonatal alarm limits  switch to neonatal mode  then  follow the procedures in the    Adjusting Alarm Limits     paragraph     ysi 6uR    PATIENT TREND DATA    N 3000 patient trend data is available to be displayed as  waveforms or in a tabular format on a stacked and active  N 3200  Refer to the N  3200 operator s manual to display and  print N 3000 trend data  Additionally  the data is available for  viewing on a PC attached to the N 3000 serial port  Contact  Nellcor Puritan Bennett   s Technical Services Department or  your local Nellcor Puritan Bennett representative for serial  port protocol information     Data stored includes SpO2 percent  pulse rate  pulse  amplitude  and alarm status  The N 3000 uses one of three  selectable formats to record samples of parameters of patient  data  Format 1 is the default format  The configuration mode   as detailed in the N  3000 service manual  can be used by  qualified service personnel to change the default format  setting     Format 1    Patient data is sampled for 10 seconds and the average of the  measurements during that period is recorded  Trend data is  available for the last 24 hours of actual monitoring during the  previous 5 days  If an alarm occurs anytime during that period   it is recorded    Format 2    Patient data is sampled for 20 seconds and the maximum and  minimum values during that period are recorded  Trend data is  available for the last 32 hours of actual monitoring during 
4.   3000  If a failure occurs  the N  3000 will display an error  code as indicated in the Troubleshooting and Maintenance  section     N 3000 Turn On Procedure    32    The safe and proper working condition of the N 3000 can be  verified through the completion of the power on self test in the  following procedure     1  Ensurethat the SpO2 cable with sensor is connected to the  N 3000  Press the ON STANDBY button  You will hear  three consecutively higher pitched beeps  and the POWER  ON indicator will light     2  All indicators light for a few seconds     Caution  If any indicator or display element does not  light  do not use the monitor  Instead  contact qualified  service personnel  your local Nellcor Puritan Bennett  representative  or the Nellcor Puritan Bennett   s  Technical Services Department     If segments of the digital displays or other indicators do  not light  notify service personnel and do not use the  N 3000     3  Digital displays individually light in a scanning test  pattern  Following a successful self test  you will hear a 1   second tone and all indicators will light while the tone  sounds     If a self test fails  you will hear an alarm and you may see  an error code in the digital display  If this occurs  press the  ON STANDBY button to turn the unit off  Attempt to  restart  If the unit again fails the self test  refer to the  Troubleshooting and Maintenance section     Start Up and Use    Note  The bottom segment of the pulse amplitude  indicator 
5.   Biocompatibility testing has been conducted  on NPB s sensors in compliance with ISO 10993 1  Biological  Evaluation of Medical Devices Part 1  Evaluation and Testing   which suggests cytotoxicity  sensitization and irritation or  intracutaneous reactivity testing be performed on devices  falling into this category  The sensors have passed the  recommended biocompatibility testing and are therefore in  compliance with ISO 10993 1     ELECTRICAL CLASSIFICATION  IEC 601 1      Protection Class    Class    per I E 601 1  clause 2 2 4    Degree of Protection    Type CF  per I E 601 1  clause 2 2 26    Enclosure Degree of Protection    Ordinary  IP X0     63    Specifications   Mode of Operation  Continuous   Emissions Compliance    EN55011 E missions Classification  CISPR II  Group    Class B    Battery    Type  Rechargeable  sealed lead acid internal battery    Operating Time  4 hours minimum from a new  fully charged battery in stand   alone configuration  age and usage affect battery performance     Recharge Period  14 hours for full charge  6 hours for 1 hour operating time    Expected Battery Life   Age and usage patterns will affect the capacity of the N 3000  battery  If the monitor is operated primarily from auxiliary  power supplies  its battery will maintain nominal capacity for  at least two years     In situations of intensive battery use   where the monitor is  operated frequently in portable applications  for example   battery life can be as short as six months   
6.   Factory Alarm Default Settings                cesceeseeeee cee 65  Factory General Default Settings                cccseeeeeeee 66   Hertormance cececeaaeeeeeeeeseeanaeeeeeeeeeqenaees 66  Measurement Range           c ceeeeeeeeeeeeeeeee caeeeeenaeeeeeenas 66  Pulse Rate Display Update Frequency             sssessseneen 66  Alarm Limit Range   Adult Pediatric and Neonate         66  Alarm Characteristics          c cccccceccecesseeese eener eerren 67  AEEHISEN EPEE EEEE EER TEE T 67   Environmental Conditions             c cceeceeeeeeeeeee ceeeeeeeeeeeeeees 68  Transport and Storage  in shipping container               68  Transport and Storage  not in shipping container         68  OPOAUON RE 68   Appendix  Principles of Oximetry                ccccccccce cette A 1   Pulse Oximetry Overview ceeeeeeeeeteeeeees A 1   Automatic Calibration    s es ceeeeseteeesaeeseeeeees A 2   Functional Versus Fractional Saturation        0    ceeeee A 2   Measured Versus Calculated Saturation               0 ceeeeeee A 3    vii    Ee    Contents    FIGURES  1 N 3000 Front Panel  North American  AA 7  2 N 3000 Front Panel  International          0  cesses eee 8  3 N 3000 Rear Panel  8  4 N 3000 Right Side Panel  North American   9  5 N 3000 Right Side Panel  International                 ccceeee 9  6 N 3000 Top Panel oecniinnisuniues niini A NA 10  7 N 3000 Left Side Panel  10  8 Connecting to the SPS External Power Supply                 19  9 Connecting Power Con    19  10 Connecting to the
7.   In these applications it is recommended that the battery  operating time be tested every six months and the battery  replaced if the capacity is less than two hours of operating  time     Avoiding deep discharge and operating from auxiliary power  supplies where possible will prolong battery life and preserve  capacity    N 3000 Input Voltage    15V      DC     Specifications  External Power Supplies    Input Voltage Current   Model SPS N or N1  100 120 VAC  500 mA  max   50 60 Hz  Model SPS I or 11  100 240 VAC  1 3 amps  50 60 Hz  Model PSS lor 1I  100 240 VAC  1 3 amps  50 60 Hz    Ee    Output Voltage  15V      DC     PHYSICAL CHARACTERISTICS   Dimensions  Height  when used standalone  6 8 cm  2 68 in    Height  excluding docking pedestal  5 4 cm  2 13 in    Width 23 9 cm  9 41 in    Depth 14 7 cm  5 79 in     Weight  1 8 kg  4 0 Ib     FACTORY SETTINGS   Factory Alarm Default Settings    Adult Pediatric Neonate    SpO2 Upper Alarm Limit  100  95   SpO2 Lower Alarm Limit  85  80   Pulse Rate Upper Alarm   Limit  170 bpm 190 bpm  Pulse Rate Lower Alarm   Limit  40 bpm 90 bpm    65    Specifications  Factory General Default Settings    Operating M ode   Pulse Beep Volume     Audible Alarm Volume     Audible Alarm Silence Duration     Alarm Silence Reminder   Latching Alarms    Serial Port Baud Rate   Trend Format     PERFORMANCE  Measurement Range    SpO2  0 100     Pulse Rate  20 250 bpm    Pulse Rate Display Update Frequency    Adult Pediatric   57 5 dB A  at 1 meter   
8.   measured by an SpO2 sensor      The monitor is intended for use on adult  pediatric  and  neonatal patients in all hospital areas  hospital type facilities   and home environments  It may be used during intra hospital  transport when powered by its internal battery     The N  3000 can operate as a standalone monitor or it can be  connected to  stacked with  other Nellcor Symphony  instruments  such as the N 3100 blood pressure monitor and  N 3200 display printer  When used with the N  3200  display printer  the instruments can display and print out  plethysmographic waveforms and SpO2 and pulse rate tabular  data     WARNING  The N 3000 is intended only as an adjunct in    patient assessment  It must be used in conjunction with  clinical signs and symptoms        ABOUT THIS MANUAL    This manual explains how to set up and use the N 3000 pulse  oximeter  Important safety information relating to general use  of the N 3000 appears before this introduction  Other  important safety information is located throughout the text  where appropriate  Read the entire Safety Information  section before you operate the monitor     In addition to the safety section  this manual includes the  following sections     e Controls  Indicators  and Symbols shows the monitor   locates all controls  indicators  and symbols  and explains  their functionality     Ee    Introduction    Setup explains how to set up the monitor and connect it to  other Ndlcor Symphony instruments and accessories     Sen
9.  0cc an 53  TFroubleShooting sarts renet eas ee d  ses Reer EEE RRE   58  Error COGS  EH   58  EEN 54  EEE DEE ee a a ee 54  EEE OB Piiale aaraa aeaaea aana eataa a iara enkaa 54  Error Condgons         c cccccccccccecsseceeees ceeeeeeseeeeessaeeeeeeaaes 54  EMI  Interferences  EN 57  Compliance with Test Limits for a Class B Digital Device    59  Periodic Safety Checks              cccccceeeeeeeneeee ceeeeeaeeeeeeeeeeees 59  ele 60  Gleanings EEN 60  Battery Maintenance AAA 60  Obtaining Technical Assistance       0       cccceeeeceeseeee eee 61  Returning System Components       sssesseesseeseeesieesrree nrsrrnees 61  SPOCCiTICALIONS EE 63  General asea EEA AEAN 63  SONSOMS o reee ee e eetas ER EEEE EE EERS 63    vi    Contents    Electrical Classification            cc cccccecseeeeesee ceeeeeeeeeeeeeeeeseees 63  Protection Class          cccccceesceceeeeeseeeee caeeeeeeeeseaeeesaeeeenees 63  Degree Of Protection ledene i a ea 63  Enclosure Degree of Protection                    seceeeeeeeeee eee 63  Mode of Operation  2       cccccceccceeceeeteee ceseeeeeeeeseeeeenaees 64  Emissions Compliance               ccscceeeceeeeeeee ceeeeeneeeeees 64  attente EES E   64  N 3000 Input Voltage  isseire ienen 64  External Power Supplies              cccecccseeeseeeeseeeeneeeeees 65   Physical Characteristics            c cccccecessceeeeeeeeceeeesseeeeneeeeees 65  DIMENSION sek cecil ve Mar ath A tate has estate cele ba 65  NN EE ee ttn nie et sa one ei ete nts 65   Factory  e E   65
10.  Indicators  and Symbols    Control knob                      Audible Alarm Off button    Figure 6  N 3000 Top Panel    Speakers                      Handle    Figure 7  N 3000 Left Side Panel  N 3000 SYMBOLS    Rear Panel    EK    New Patient Neonatal Button    Data Interface    Attention  Refer to Manual    10    Right Side Panel    SpO2     w    Controls  Indicators  and Symbols    SpO2 Cable Input    Ee    Type CF Equipment    SPS EXTERNAL POWER SUPPLY SYMBOLS           gt     Input Voltage  Output Voltage    Equipotential  Ground  Terminal  Caution  Shock Hazard    Power Available Indicator    DESCRIPTION OF CONTROLS          PRINT    Pi Hai    The ON STANDBY button is used to turn the   N  3000 on or off  Three consecutively higher   pitched beeps sound when turning the N  3000 on   and three consecutively lower pitched beeps  sound when turning the N  3000 off     The NEW PATIENT NEONATAL button on the  rear panel is used to clear stored patient data or  to switch between adult pediatric and neonatal  modes     The AUDIBLE ALARM OFF button is used to  silence current audible alarms for the selected  alarm silence duration period     The PRINT button is used when the N  3000 and    the N  3200 display printer are stacked to cause  the N  3200 to begin printing current N 3000 data     The UPPER ALARM LIMIT button is used to  view and set upper alarm limits     11    Controls  Indicators  and Symbols    G  N    The LOWER ALARM LIMIT button is used to  view and set lower a
11.  PSS External Power Supply             0      20  11 Removing the N 3100 Top Cover 22  12 Connecting oeh 2100    22  13 Removing the N 3000 From the N 3100     0       eeeeeeeee 23  14 Serial Port Interface  0        ec eceeceeeeeeeeeee ceeeeeeeeeeeeseeeeeseeeeeas 24  15 Connecting SpO2 Sensor to SCP 10 Cable                0008 29  16 Monitoring Conditions and Alarm Responses            00s100    37  17 Selecting Smart Pulse Rate Alarm Lime  46  18 Setting Audible Alarm Silence Duration to On    48  A 1 Oxyhemoglobin Dissociation Curve            cceesseseseeeeeee ees A 3  TABLES   1 Compatible Nellcor Puritan Bennett Sensors           eeeeeeeae  28  Audible Alarm Off Duration    40    viii    SAFETY INFORMATION    General Safety Information    ysi 6uR    GENERAL SAFETY INFORMATION    This section contains important safety information related to the  general use of the Nelicor Symphony N 3000 Pulse Oximeter   Other important safety information appears throughout in sections  that relate specifically to the precautionary information  Read all  text surrounding all precautionary information     Important  Before use  carefully read this manual   accessory directions for use  all precautionary  information in boldface type  and specifications     WARNING  Explosion hazard  Do not use the N  3000  pulse oximeter in the presence of flammable anesthetics    WARNING  The N 3000 is to be operated by qualified  personnel only  Before use  carefully read this manual     accessory dire
12.  at the end of the alarm silence duration     To deactivate the feature before the alarm silence duration has  expired  press the AUDIBLE ALARM OFF button twice within  2 seconds  Again  three beeps indicate that this feature has  been deactivated  The AUDIBLE ALARM OFF indicator is  extinguished     While in the all alarm silence mode  no alarms will sound  during the alarm silence duration  including those of a higher  priority  The appropriate parameter displays or indicators   however  will flash  Similarly  if any alarm condition is  corrected and then recurs during the alarm silence duration  no  alarms will sound  but the appropriate parameter displays or  indicators will flash     41    Start Up and Use    Latching Alarm Option    Qualified service personnel can configure your system to have  latching alarms  The factory default is non latching alarms     Latching alarms   The parameter display flashes and an audible alarm sounds  even after the alarm condition no longer exists  until you press  the AUDIBLE ALARM OFF button     Note  If the N 3000 is configured for latching alarms and the  AUDIBLE ALARM OFF button is pressed during a  high  or medium priority alarm  the parameter display  will stop flashing when the alarm condition is no  longer present     Non latching alarms   The parameter display flashes only until the alarm condition  no longer exists  The audible alarm sounds until the condition  no longer exists or until you press the AUDIBLE ALARM OFF  button 
13.  changed by qualified service personnel using the  configuration mode described in the service  manual for the instrument needing to be changed     51    Start Up and Use    BATTERY OPERATION    52    In case of loss of external power  the N 3000 operates on its  internal battery that provides at least 4 hours of operation  from a new  fully charged battery when operating  independently of the N  3100 and N  3200  When operating in  the stacked configuration with the N  3200  either the N 3000  battery or N  3200 battery provides power to the stack   depending on which battery has the higher charge voltage   When stacked with the N 3100  battery life varies  depending  primarily on the frequency of N  3100 blood pressure  measurements     Recharge the N  3000 after extensive battery use or after the  BATTERY IN USE BATTERY LOW indicator has been  flashing  If the monitor is not recharged  the monitor may fail  to operate or stop operating suddenly     When the N  3000 is connected to an external power supply or  is stacked with an N  3200 connected to AC power  or stacked  with an N 3100 connected to an active SPS or PSS external  power supply  the self contained battery automatically  recharges  whether the N  3000 is on or off  A complete  recharge of a fully drained battery requires 14 hours     Caution  If the N 3000 is to be stored for a period of 24  months or longer  notify service personnel to remove  the battery from the monitor prior to storage  Recharge  the batter
14.  detected and the data display is routinely updated     ysi 6uR    Pulse search mode  during which the PULSE SEARCH  indicator lights  the data display alternates between data  and dashes  and the data evaluation period is ongoing  The  data display is updated when a qualified pulse is detected     Loss of pulse mode  during which an alarm sounds  the  PULSE SEARCH indicator flashes  and the data displays  flash zeroes     Alarm Descriptions    The N  3000 has five levels of audible alarms     1     High priority alarm  Indicated by a high rate  high   pitched  pulsing tone  A high priority alarm sounds after  loss of pulse is detected  refer to the    Management of the  Loss of Pulse Alarm    paragraph in this section      Medium priority alarm  Indicated by a medium rate   medium pitched  pulsing tone  A medium priority alarm  sounds and the corresponding parameter display flashes  when any measured patient parameter moves outside the  set alarm limits  A medium priority alarm also sounds  during battery operation  when 15 minutes or less battery  capacity remains     Low priority alarm  Indicated by a low rate  low pitched   pulsing tone  A low priority alarm sounds during the  following conditions     e When an SpO 2 cable has disconnected  this alarm  function is inactive from the time the N  3000 is  turned on until the cable is connected     e 50 seconds after loss of pulse is detected in the  presence of continuous motion    Microprocessor failure alarm  Indicated by 
15.  in the N  3000 service manual  The setting OFF can  not be selected as a power on default setting     To view the current audible alarm silence duration  press and  hold the AUDIBLE ALARM OFF button for less than 3 seconds   The setting is displayed similarly to that shown below     Sp02  PULSE RATE  min       L6e   sec     WARNING  If an alarm condition  except those for low  battery  power failure and microprocessor failure     occurs while the audible alarm silence period is set to  OFF  the only alarm indication will be visual displays  related to the alarm condition        You can set the audible alarm silence period to 30  60  90  or  120 seconds or to OFF  To adjust the audible alarm silence  period  press and hold the AUDIBLE ALARM OFF button   Within 3 seconds  after pressing the AUDIBLE ALARM OFF  button  turn the control knob until you see the desired alarm  silence duration in the PULSE RATE display  The displayed  setting takes effect when you release the AUDIBLE ALARM  OFF button     When operating in the stacked configuration  setting the  audible alarm silence period for one instrument causes any  other active instrument in the stack to be set to the same  period     47    Start Up and Use    Shortcut to Select OFF for Audible Alarm Silence Duration    To quickly select OFF for the N 3000 audible alarm silence  period  simultaneously press the AUDIBLE ALARM OFF  button and the UPPER ALARM LIMIT button as shown in  Figure 16        Figure 18  Setting Audible A
16.  mounting adapter such as the Ndlcor  Symphony DB 1 mounting kit  the N 3000 may be  attached to an IV pole  bed rail  wall mount  or headboard     2  Placethe SPS power supply on a flat surface near the  N 3000 so the power supply cable will reach the N  3000     Caution  For AC operation  use only the Nedlicor  Symphony SPS N1  North America   SPS I1   international   PSS 1  North America  or PSS 1I   international  external power supplies  Use of other  power supplies may result in damage to or improper  operation of the N 3000     3  Plug the SPS power supply cable into the socket on the  right side of the N 3000  The connector will engage only  one way  with the cable toward the rear of the monitor as  illustrated in Figure 8     18    Setup    SpO2 connector  N 3000 right side External power supply    Equipotential  terminal                  Ee                                  External power  supply cable    Figure 8  Connecting to the SPS External Power Supply    4  Connect the Nellcor Puritan Bennett power cord to the  external power supply as shown in Figure 9     Power available indicator       Nellcor Puritan Bennett  hospital grade  power cord    External power supply    Figure 9  Connecting Power Cord    WARNING  In the USA  do not connect to an electrical  outlet controlled by a wall switch because the device  may be accidentally turned off     5  Plug the power cord into a properly grounded AC outlet   6  Useonly the original hospital grade AC power cord  provi
17.  oxygen transducer OXI A N    lt 3 or  gt 40 kg   reusable with disposable non OXI P I   3 40 kg  sterile shee     Durasensor    oxygen transducer Kg  100A ee   reusable  non sterile     Nellcor Puritan Bennett Bae 10 a g  reflectance oxygen transducer  Taaa oa sterile     Dura Y    multisite oxygen  transducer non sterile  reusable     For use with Dura Y sensors   Ear clip  reusable  non sterile     Pedi Check    pediatric spot check  clip  reusable  non sterile     oxygen transducers   sterile  single use only   lt 3 or  gt 20 kg    3 to 40 kg   gt 30 kg    BIOCOMPATIBILITY TESTING    28    Biocompatibility testing has been conducted on Nellcor Puritan  Bennett sensors in compliance with ISO 10993 1  Biological  Evaluation of Medical Devices  Part 1  Evaluation and Testing   The sensors have passed the recommended biocompatibility  testing and are therefore in compliance with   SO 10993 1     Sensors and Accessories    CONNECTING SPO2 SENSORS    Use only notched Nellcor Puritan Bennett sensors  such as  those indicated in Table 1  which are compatible with the  SCP 10 cable     Connect the sensor to the SCP  10 cable  locking it into place as  illustrated in Figure 15  Notice that the sensor connector and  SCP 10 connector are shaped to fit together only one way  with  the NELLCOR name on the upper side of the connector     OXISENSOR II SCP 10  connector       Figure 15  Connecting SpO2 Sensor to SCP 10 Cable    29    Ee    Sensors and Accessories  PERFORMANCE CONSIDERATIO
18.  stack   press the docking release button on the rear pane  of the  bottom instrument and slide the top instrument off the bottom  instrument                          Figure 13  Removing the N 3000 From the N 3100    23    Setup    SERIAL INTERFACE    24    WARNING  To ensure patient electrical isolation     connect only to other equipment with electrically  isolated circuits        The N  3000 serial interface provides patient data output using  EIA 232 or RS 422 signal levels and is located on the rear  panel  When data is requested by a computer connected to the  serial interface port  requested data is checked to ensure  integrity and then sent out through the port in a proprietary  format     The serial port is functional only when the N 3000 is operating  on AC power  Use of the port by a qualified programmer  requires an ElA 232 Nellcor Puritan Bennett cable  purchased  separately  to connect the N  3000 with the computer as  illustrated in Figure 14  The cable and directions for use are  available by contacting Nellcor Puritan Bennett Technical  Services or your local Nellcor Puritan Bennett representative                                                                                                                       Figure 14  Serial Port Interface    This output port can also be configured to provide an alarm  active function instead of serial data     Noe When the N  3000 is configured for the alarm active  function  the serial data interface connector is  ded
19.  the rear pane  of the N  3200     21    Setup    2  Press the docking release button on the rear of the N 3100  or N 3200 to remove the top cover and expose the docking  connector  N  3100 illustrated in Figure 11      Docking release button   on rear panel           NELLCOR Symphony N 3100       Figure 11  Removing the N 3100 Top Cover    3  Slide the N 3000 onto the N 3100 or N  3200 until the  docking connectors firmly engage and you hear a    click      See Figure 12                                   SYSTOLIC DIASTOLIC  mmHg mmHg                                           Figure 12  Connecting to the N 3100    4  Operate the N 3000 as indicated in the Start up and Use  section of this manual     22    Setup    Note  TheSTACKED indicators on the N 3000  the  N 3100  and N  3200 in a stack will light while the  instruments are stacked  communicating  and  power is turned on  If the STACKED indicator  does not light  refer to the Troubleshooting and  Maintenance section     Ee    Read the N 3100 or N  3200 operator   s manual before using the  stacked configuration     REMOVING THE N 3000 FROM THE N 3100 OR N 3200    Note  When operating on battery power  you may remove the  N 3000 from the N  3100 while the N 3000 is ON   However  the N 3100 must be off prior to removal from  the stack in order to avoid a power failure alarm  If a  power failure alarm occurs  press the N  3100  AUDIBLE ALARM OFF button to silence the alarm     To remove a Ndlcor Symphony instrument from the
20.  with these  instructions  may cause harmful interference with other  devices in the vicinity     Disruption may be evidenced by erratic readings  cessation of  operation  or incorrect functioning of other devices  If this  occurs  the site of use should be surveyed to determine the  source of this disruption  The following actions may be taken to  eliminate the source     e Turn equipment in the vicinity off and on to isolate the  source of interference     e  Reorient or relocate the N  3000 or the receiving device     e     Increase the distance between the N 3000 and the other  equipment     e Select alternative power receptacle  A C power outlet    If assistance is required  contact Nellcor Puritan Bennett   s    Technical Services Department or your local Nellcor Puritan  Bennett representative     Troubleshooting and Maintenance  COMPLIANCE WITH TEST LIMITS FOR A CLASS B DIGITAL DEVICE    This equipment has been tested and found to comply with the  limits for a Class B digital device  pursuant to part 15 103 of  the FCC Rules  and CISPR II  Class B   These limits are  designed to provide reasonable protection against harmful  interference in a residential installation  This equipment  generates  uses and can radiate radio frequency energy and  if  not installed and used in accordance with the instructions  may  cause harmful interference to radio communications  However   there is no guarantee that interference will not occur ina  particular installation   If this equi
21. ANCE    For technical information and assistance  or to order a service  manual  contact Nellcor Puritan Bennett   s Technical Services  Department or your local Nellcor Puritan Bennett  representative  The service manual includes information  required by qualified service personnel when servicing the  N 3000     RETURNING SYSTEM COMPONENTS    If it is necessary to return the N 3000 or the external power  supply for service  call Nellcor Puritan Bennett s Technical  Services Department or your local representative for shipping  instructions  Have all equipment serial numbers available  when calling     To pack the N  3000 for return  disconnect the sensor  sensor  cable  and external power supply  It is not necessary to return  the sensor  To pack the SPS power supply  disconnect the  power cord  Pack items to be returned in their original shipping  carton if available  If not  use a suitable carton with  appropriate packing material to protect the item or items  during shipping     61    SPECIFICATIONS    General m  Electrical Classification a  Physical Characteristics e  Factory Settings   Performance    Environmental Conditions  Environmental Protection    GENERAL  Designed to meet safety requirements of     UL 2601 1  CSA C22 2 No  601 1 M 90  EN60601 1 1995   ISO 9919  EMC per EN60601 1 2 1993  ISO 10993 1   Sensors      Sensors  Nellcor Puritan Bennett   s oxygen transducers  sensors  can be  categorized as surface devices contacting skin for a limited  duration of time
22. MOTION indicator lights when patient  motion is detected by the N 3000     NEONATAL    or  NEO    LINKED    zO     amp     Controls  Indicators  and Symbols    The NEONATAL indicator lights when the  N 3000 is in neonatal mode     Ee    The LINKED indicator lights when data is being  transmitted from the N 3000 to a remote  monitoring station     The POWER ON indicator to the right of the  ON STANDBY button lights continuously while  the N  3000 is on     The AUDIBLE ALARM OFF indicator lights  steadily to signify that the current audible alarm  has been temporarily silenced  If the alarm  condition ceases prior to the end of the alarm  silence duration  the light goes out     While the condition exists  the indicator remains  lit for the selected duration or until a higher  priority alarm condition arises  in which case the  AUDIBLE ALARM OFF indicator is no longer  illuminated and an alarm sounds  The indicator  flashes when the audible alarm silence duration  is set to    OFF        When the N  3000  or an attached N 3100  has the  audible alarm silence duration set to    OF F     during an alarm condition  the AUDIBLE   ALARM OFF indicator flashes and the audible  alarm off reminder will sound at 3 minute  intervals  The latter occurs only if the N  3000  or  any attached N  3100  has had the audible alarm  off reminder enabled  using the service mode as  described in the N 3000 service manual      The BATTERY IN USE BATTERY LOW indicator  lights continuously when the bat
23. NS    WARNING  Pulse oximetry readings and pulse rate  signals can be affected by certain ambient    environmental conditions  sensor application errors   and certain patient conditions        Use only Nellcor Puritan Bennett sensors and sensor cables   Select an appropriate sensor  apply sensor as directed  and  observe all warnings and cautions presented in the directions  for use accompanying the sensor     High ambient light sources such as surgical lights  especially  those with a xenon light source   bilirubin lamps  fluorescent  lights  infrared heating lamps  and direct sunlight can interfere  with the performance of an SpO2 sensor  To prevent  interference from ambient light  cover the sensor site with  opaque material  Failure to take this action in high ambient  light conditions may result in inaccurate measurements     If patient movement presents a problem  try one or more of the  following remedies to correct the problem     e Verify that the sensor is properly and securely applied   e Movethe sensor toa less active site   e Usean adhesive sensor that tolerates some patient motion  e Usea new sensor with fresh adhesive backing    30    START UP AND USE    Power on and Self Test   Clearing Data   Neonatal Mode   Patient Trend Data   Description of Alarms   Adjusting Alarm Limits   Selecting    Smart    Pulse Rate Alarm Limits  Adjusting Audible Alarm Silence Duration  Adjusting Audible Alarm Volume  Adjusting Pulse    Beep    Volume  Technical Staff Adjustable Fe
24. Troubleshooting and Maintenance    BATTERY IN USE indicator lights steadily while the  N 3000 is connected to its external power supply and  AC outlet     e Make sure that the external power supply is properly  connected to the N 3000 and to the hospital grade  power cord     e Confirm that the Power Available indicator on the  external power supply is on  If it is not  and  connections are good  the external power supply may  be defective  Notify your service personnel     e Check to see if power is available to other equipment  on the same AC circuit     The PULSE AMPLITUDE indicator seems to  indicate a pulse  but the digital displays show  zeroes     e Check the patient    e Excessive patient motion may be making it impossible  for the N 3000 to acquire a good pulse signal  If  possible  keep the patient still  Check whether the  sensor is applied securely and properly and replace if  necessary  or move the sensor to a new site    e Thesensor or cable may be damaged  replace it     e Thesignal may be too low to allow the N  3000 to  measure oxygen saturation and pulse rate     55    Ee    Troubleshooting and Maintenance    56    4  SpO2or pulse rate values change rapidly  pulse  amplitude indicator is erratic     Check the patient     Excessive patient motion may be making it impossible  for the N 3000 to acquire a good pulse signal  If  possible  keep the patient still  Check whether the  sensor is applied securely and properly and replace if  necessary  or move the sens
25. You turn the instrument off and then back on where limits  revert to power on defaults    e You use standard alarm setting procedures to select new  pulse rate alarm limits    e You reactivate the Smart pulse rate alarm limits setting  as described above    e You see an error code displayed indicating current limits  have been lost    ADJUSTING AUDIBLE ALARM SILENCE DURATION    46    WARNING  Do not silence an audible alarm or decrease       its volume if patient safety could be compromised     Alarms can be silenced for a preset period called the audible  alarm silence duration  The audible alarm silence duration  starts when you press the AUDIBLE ALARM OFF button     Start Up and Use    If an alarm of the same or lower priority occurs during the  audible alarm silence duration  it will not be heard until the  end of the duration  If the alarm condition is still present or a  new alarm occurs after the audible alarm silence duration has  elapsed  an audible alarm will sound     Ee    An alarm with a higher priority will sound regardless of  whether the duration has ended  Also  if alarm condition ceases  to exist before the audible alarm silence duration has ended   any alarm condition will cause an alarm to sound     The power on default audible alarm silence duration setting is  in effect unless it is changed  The power on default set at the  factory is 60 seconds  The power on default setting can be  adjusted by service personnel using the configuration mode as  indicated
26. a shrill     continuous tone  A microprocessor alarm sounds when the  microprocessor has failed     39    Start Up and Use    40       5  Powe failurealarm  Indicated by a shrill  pulsing tone A  power failure alarm sounds when the N  3000 loses power  for any reason  other than when the ON STANDBY button  is pressed      As shown in the    Alarm Summary    paragraph later in this  section  alarm pitch  time between beeps  and the type of visual  indication depend on the alarm condition     Note  Beeps that indicate valid and invalid keypresses  pulse  beep  patient data cleared  and selection on    Smart     pulse rate alarm limits will also be heard during  audible alarms     There will be no upper lower alarm limit  loss of pulse  or  sensor cable disconnect alarms if the N  3000 is turned on  before the cable or SpO2 sensor is attached  These alarms  sound only after a pulse has been monitored     Press the AUDIBLE ALARM OFF button once to silence an  alarm  The alarm will be turned off for an amount of time as  described in Table 2     Table 2  Audible Alarm Off Duration    When AUDIBLE ALARM OFF  button is pressed  alarm will  remain off until   High priority 1  Alarm silence duration setting  expires    Parameter outside of set   1  Alarm silence duration setting  alarm limits expires  or    2  A higher priority alarm occurs  50 sec  after loss of pulse  with continuous motion  SpO 2 cable sensor 1  Thesituation is corrected  then  disconnect  recurs  or  Low battery  
27. any liquid on the    N 3000 its accessories  connectors  switches  or openings  in the chassis        Toclean the N 3000  dampen a cloth with a commercial  non   abrasive cleaner and wipe the top  bottom  and front surfaces  lightly  Wipe the sensor extension cable with a damp cloth  F or  sensors  follow cleaning instructions in the directions for use  accompanying the sensor     If liquid is accidentally spilled on the N  3000  clean and dry  thoroughly before reuse     BATTERY MAINTENANCE    60    Caution  If the N 3000 is to be stored for a period of 24  months or longer  notify service personnel to remove  the battery from the monitor prior to storage  Recharge  the battery when it has not been charged for 6 months  or more     If the N 3000 has not been used for a long period of time  the  battery will need charging  To charge the battery  connect the  N 3000 to an SPS or PSS external power supply  or an N  3200   which is connected to an AC outlet  refer to the Setup section    It may be necessary to briefly turn on the N 3000 to start  battery charging     Troubleshooting and Maintenance    Note  Storing the N  3000 for a long period without charging  the battery may degrade the battery capacity  A  complete battery recharge requires 14 hours     Ee    Nellcor Puritan Bennett recommends that the N  3000 battery  be replaced at 2 year intervals  Refer to the N  3000 service  manual for battery replacement and general service  instructions     OBTAINING TECHNICAL ASSIST
28. atures  Nellcor Symphony Instruments in Stack    Ee    WARNING  The N 3000 is to be operated by qualified  personnel only  for non prescription device   Before use   carefully read this manual  accessory directions for use   all precautionary information  and specifications     WARNING  The N 3000 is intended only as an adjunct in  patient assessment  It must be used in conjunction with  clinical signs and symptoms     WARNING  Pulse oximetry readings and pulse rate  signals can be affected by certain ambient    environmental conditions  sensor application errors   and certain patient conditions     WARNING  For pacemaker patients  the N 3000 may  continue to count the pacemaker rate during  occurrences of cardiac arrest or some arrhythmias  Do  not rely entirely upon N 300 alarms  Keep pacemaker  patients under close surveillance     WARNING  Do not silence the audible alarm or decrease  its volume if patient safety could be compromised        POWER ON AND SELF TEST    WARNING  Ensure that the speaker is clear of any    obstructions  Failure to do so could result in an  inaudible alarm tone        31    Start Up and Use    Automatic Self Tests    When the N 3000 is turned on  it performs automatic self tests   A 1 second beep indicates that self tests were performed  successfully     While in use  the N 3000 performs self tests every 4 to 5  minutes without interrupting normal operation  If these self   tests are successful  you will see no changes in the operation of  the N
29. ce   3 avenue du Canada  Batiment Sigma  LP851 Les Ulis   91975 Courtaboeuf Cedex   France   Tel   33 01 69 82 14 00    Nelicor Puritan Bennett Italia Srl  Via dei Tulipani  3   20090 Pieve Emanuele  M1    Italy   Tel   39 2 90786404    Nellcor Puritan Bennett Finland Oy  Kappelitie 8   02200 Espoo   Finland   Tel   358 9 270 92 900    To obtain information about a warranty  if any  for this product  contact  Nellcor Puritan Bennett Technical Services or your local Nellcor Puritan    Bennett representative     Purchase of this instrument confers no express or implied license under any  Nellcor Puritan Bennett patent to use the instrument with any sensor that is  not manufactured or licensed by Nellcor Puritan Bennett     The following are trademarks of Nellcor Puritan Bennett Inc   Durasensor   Nelicor Symphony  Oxisensor II  Dura Y   Oxismart  OxiCliq  and the Nellcor  Puritan Bennett knob configuration     Covered by one or more of the following U S  Patents and foreign equivalents   4 621 643  4 700 708  4 770 179  4 653 498  5 078 136  5 351 685  4 869 254  and    5 368 026        CONTENTS    Figures   Tables  Safety Information            ccccccccecee cece eenen reenen a 1  General Safety Information             cccccecceeseeeeese ceteeeeeeeeees 1  Measurements eeeeeeeeeaeeeeeeeeeseesenaees 2  N 3000 and Other Equipment eters 3  introduction ue hei nei o beaches beeline eeu DEE es 5  Intended Ulee ceneeeeeeeeeaeaeeeeseeenineeeee aaant 5  About This Manual  5  Controls  I
30. ctions for use  all precautionary  information  and specifications     WARNING  Pulse oximetry readings and pulse rate  signals can be affected by certain ambient  environmental conditions  sensor application errors   and certain patient conditions        CAUTION  When connecting the N 3000 to any  instrument  verify proper operation before clinical use   Both the N 3000 and the instrument connected to it must  be connected to a grounded outlet  Accessory equipment  connected to the monitor   s data interface must be  certified according to IEC Standard 950 for data   processing equipment or IEC Standard 601 1 for  electromedical equipment  All combinations of  equipment must be in compliance with IEC Standard  601 1 1 system requirements  Anyone who connects  additional equipment to the signal input port or signal  output port configures a medical system and is  therefore responsible that the system complies with the  requirements of system standard IEC Standard 601 1 1     Safety Information    Nellcor Puritan Bennett   s oxygen transducers  sensors   can be categorized as surface devices contacting skin  for a limited duration of time  Biocompatibility testing  has been conducted on Nellcor Puritan Bennett   s  sensors in compliance with ISO 10993 1  which suggests  cytotoxicity  sensitization and irritation or  intracutaneous reactivity testing be performed on  devices falling into this category  The sensors have  passed the recommended biocompatibility testing and  are t
31. ctrosurgical unit isin use   but the readings may be inaccurate during use and  shortly thereafter     WARNING  Disconnect the N 3000 and Nellcor sensor  during magnetic resonance imaging  MRI  scanning   Use during MRI could cause burns or adversely affect  the MRI image or the monitors accuracy  Also  to avoid  burns  remove the N 3000 sensor from the patient before  conducting MRI     WARNING  Do not lift the monitor by the sensor cable   lead  or power cord because the cable  lead  or cord  could disconnect from the monitor  causing the monitor  to drop on the patient        17    Setup  PERFORMANCE VERIFICATION    The N 3000 performance can be verified by following the  procedures outlined in the Performance Verification section of  the N 3000 service manual  Qualified service personnel should  perform this procedure before using the monitor for the first  time in a clinical setting     CONNECTING TO AN EXTERNAL POWER SUPPLY    The N  3000 may be used alone or stacked with the N  3100  blood pressure monitor  the N  3200 display printer  or both  If  connected to the N 3200  it is not necessary to connect to an  external power supply     If transporting the monitor  or if AC power is not available  you  may operate the N  3000 on battery power for a limited amount  of time  In that case  it is not necessary to connect to an  external power supply     CONNECTING THE N 3000 TO THE SPS POWER SUPPLY    1  Place the N 3000 on a flat surface near the patient  With  an optional
32. ded by Nellcor Puritan Bennett  Ensure that the    power available indicator on the external power supply  lights     19    Setup  CONNECTING THE N 3000 TO THE PSS POWER SUPPLY  1  Placethe PSS on a flat surface near the patient     2  Connect the PSS to the N  3000 as shown in Figure 10                                      Figure 10  Connecting to the PSS External Power Supply    3  Slidethe monitor onto the PSS until the docking  connectors firmly engage and you hear a    click        WARNING  In the USA  do not connect to an electrical  outlet controlled by a wall switch because the device  may be accidentally turned off     A PlugthePSS into an AC outlet     5  Using the PSS rear panel rocker switch  turn the power  supply on  Ensure that the green LED on the front of the  PSS is lit     BATTERY OPERATION    In case of loss of external power  the N 3000 operates on its  internal battery that provides at least 4 hours of operation  from a new  fully charged battery when operating  independently of the N  3100 and N  3200  When operating in  the stacked configuration with the N  3200  either the N 3000  battery or N  3200 battery provides power to the stack   depending on which battery has the higher charge voltage   When stacked with the N 3100  battery life varies  depending  primarily on the frequency of N  3100 blood pressure  measurements     Recharge the N  3000 after extensive battery use or after the  BATTERY IN USE BATTERY LOW indicator has been  flashing  If the m
33. dure   ss sssssssnesssnnsnesnneeennn meserna 32  Clearing Datared a ad N 33    Ee    Contents    Neonatal MOOG ss chess cccetebsanetheegecee Bieeieskia ege Dee aaar 34  Patient Trend Data               cccccccseseeceeee ceeeeneeaeeeeeeseesenaeeeeess 35  FORMAL EE 35   El E 35  FOMMAat ugesi eege Edge eege eebe terest sett 35  Description Of Alarms  0       ccceeeeeeeeeeenteeeeeeceeeeeenaeeeeenaeeeeeeaes 36  Management of Loss of Pulse Alam  36  Alarm Descriptions   0        ccc eeeeeeeeeeeeeeee ceeeeeeaeeeeeaaeeeeeaas 39  All Alarm Silence           cccceccccecesseceeeee ceeeesnaeeeesssaeeseeaaes 41  Latching Alarm Option            cecceeesceseeeeeenee ceeeeeneeeeneees 42  Alarm SUMMARY EE 42  Adjusting Alarm Limits 20 0 0    ceceeeeeseeeeeee ceeeeeenaeeeeenaeeeeneas 43  Melen esic cadet e i ran eaea Seel duet  Seege   eebiei og eer 43  Selecting Alarm Limit to be Adlueted 44  Alarm Limits Changed Indicator          s sesseeeeeeeeeeeeeee ee 45  Selecting    Smart    Pulse Rate Alarm Limits                c00  45  Adjusting Audible Alarm Silence Duration  46  Shortcut to Select OFF for Audible Alarm Silence  RTL E 48  Adjusting Audible Alarm Volume A 48  Audible Alarm Off Reminder          ssssnnnsennnnnennnne nne naene 49  Adjusting Pulse    Beep    Volume nesre 50  Technical Staff Adjustable Features             0ccecssceecseee a 51  Nellcor Symphony Instruments in Stack      0   ccccceeeeeees 51  Battery Operation    51  Troubleshooting and Maintenance                     
34. e NEW PATIENT   NEONATAL button twice within 2 seconds     When the N  3000 is in neonatal mode  the NEONATAL MODE  indicator lights  the indicator does not light when the N 3000 is  in the adult pediatric mode  To change back to adult pediatric  mode  press the NEW PATIENT NEONATAL button twice  within 2 seconds  The NEONATAL MODE indicator should no  longer be lit    Noe The default power on operating mode can be changed  to the neonatal mode by qualified service personnel  using the configuration mode described in the N 3000  service manual     When you change operating modes  adult pediatric to neonatal   or vice versa   alarm limits return to power on defaults for the  respective modes and previous patient data is cleared from the  displays     When the N  3000 is stacked with the N 3100  the initial power   on default operating mode may be different for the two  instruments  Both instruments in the stack should be in the  same operating mode  neonatal or adult pediatric      If active instruments in the stack arein different operating  modes when power is turned on  press the NEW  PATIENT NEONATAL button twice  2 presses within 2  seconds  on the active instrument that is in the incorrect mode  as needed to place all active instruments in the same operating  mode     Note  The default power on operating mode can be changed  by technically qualified service personnel using the  configuration mode described in the service manual for  the instrument needing to be changed  
35. etect at least  one qualified pulse during any 10 second period  it enters pulse  search mode  the PULSE SEARCH indicator lights  the  displays alternate between data and dashes  and the data  evaluation period starts     During the 50 second data evaluation period     e Ifthe patient is not moving  for example  because of  cardiac arrest  shock  paralysis  anesthesia  and has no  qualified pulse for 6 seconds  a high priority alarm  immediately sounds  the PULSE SEARCH indicator  flashes  and the data displays flash zeroes     e Ifthe patient is constantly moving  the N 3000 searches  for qualified pulses for up to 50 seconds  Each timea  qualified pulse is detected  even during motion  the data  displays are updated     The N  3000 returns to normal operation as soon as it  detects one of the following  three consecutive qualified  pulses  when there is no motion   five consecutive qualified  pulses  when thereis motion   or a total of 10 qualified  pulses  Otherwise  at the end of the data evaluation  period  the data displays flash zeroes and  if motion is still  present  a low priority alarm sounds  if there is no motion   a high priority alarm sounds     If  during the 50 second data evaluation period  motion  stops and no qualified pulse is found for 6 seconds  a high   priority alarm sounds immediately     Start Up and Use    N 3000 loss of pulse alarm management can therefore be  considered a three tier system     Normal mode  during which qualified pulses are being 
36. fal NELLCOR PURITAN BENNETT     OPERATOR   S MANUAL    Nellcor Symphony     N 3000 Pulse Oximeter    Caution  Federal law  U S   restricts this device to sale by or on the order of a  physician     To contact Nellcor Puritan Bennett   s representative  In the United States   call 1 800 NELLCOR or 510 463 4000  outside of the United States   call your local Nellcor Puritan Bennett representative           1997 Nellcor Puritan Bennett Inc  All rights reserved  035502A  1097       Corporate Headquarters  Nelicor Puritan Bennett Inc  4280 Hacienda Drive  Pleasanton  California 94588  USA   Tel  510 463 4000  1 800 NELLCOR    European Office   Nellcor Puritan Bennett Europe BV  Hambakenwetering 1   5231 DD  s Hertogenbosch   The Netherlands   Tel  31 73 6485200    Asia Pacific Office   Nellcor Puritan Bennett HK Ltd   Room 1602 Evergo House   38 Gloucester Road   Wanchai    Hong Kong   Tel   852 2529 0363    Latin American Office  Nelicor Puritan Bennett Inc   1750 University Drive  Suite 216  Coral Springs  Florida 33071  USA   Tel  954 346 7977    Regional Local Offices  Nellcor Puritan Bennett UK Ltd   10 Talisman Business Centre  London Road   Bicester  Oxfordshire OX6 OJ X  United Kingdom   Ted   44 1869 322700    Nellcor Puritan Bennett Germany GmbH  Black  amp  Decker Strasse 28   65510 Idstein   Germany   Tel   49 6126 5930    Nellcor Puritan Bennett Belgium NV SA  Interleuvenlaan 62 8  Zone 2   B 3001 Heverlee   Belgium   Tel   32 16 400467    Nellcor Puritan Bennett Fran
37. fault values if necessary   as described below  Changes you make will remain in effect  until you change them again or turn the N 3000 off     WARNING  Each time the monitor is used  check alarm    limits to ensure that they are appropriate for the  patient being monitored        43    Start Up and Use    When monitored patient values appear in the digital displays  or dashes are displayed before a measurement is taken  the  N 3000 is in normal  patient monitoring  mode     When either the UPPER or the LOWER ALARM LIMIT button  is pressed  the N 3000 is in view adjust alarm limits mode     While in the view adjust alarm limits mode  current alarm  limits will be displayed  but patient parameter values will not  be displayed     The N 3000 automatically returns from view adjust alarm  limits mode to normal mode if you do not press a limit button  or adjust a setting within 3 seconds  It also returns to normal  mode if you press the AUDIBLE ALARM OFF  or NEW  PATIENT NEONATAL button  Any alarm limit changes made  will take effect     Selecting Alarm Limit to be Adjusted    To select an upper alarm limit to view or adjust  press the  UPPER ALARM LIMIT button  When the UPPER ALARM  LIMIT button is pressed  the upper alarm limit for SpO2 is  displayed  Subsequent pressing of the UPPER ALARM LIMIT  button will cycle through the displays showing upper alarm  limit for pulse rate and back again to the SpO2 display  The  display of an upper alarm limit is indicated by dashes in the  u
38. fter the displayed parameter value  during patient monitoring or when alarm limits  are viewed  The decimal point remains until  you change modes  adult or neonatal   turn the  N 3000 off  or until the value is adjusted back to  the default value     SELECTING    SMART    PULSE RATE ALARM LIMITS    The    Smart    feature automatically sets pulse rate alarm limits  as follows     e Lower Alarm Limit  75  of the patient   s current pulse  rate  at time of setting  or 30 bpm  whichever is larger     e Upper Alarm Limit  150  of the patient   s current pulse  rate or 250 bpm  whichever is smaller     For example  if the patient   s pulse rate is 80 bpm when you  select the Smart feature  the upper alarm limit would be set at  120 bpm  and the lower limit would be set at 60 bpm     Note  Smart pulse rate alarm limits can be set only when the    N 3000 is displaying a pulse rate value other than  zero     45    Start Up and Use          Figure 17  Selecting Smart Pulse Rate Alarm Limits    To select Smart pulse rate alarm limits  press and hold the  UPPER and LOWER ALARM LIMIT buttons simultaneously  for 3 seconds as shown in Figure 15  Three beeps indicate that  the new limits are set  based on a valid pulse  A single     invalid  keypress    beep  after the 3 seconds  indicates that Smart pulse  rate alarm limits were not set for lack of a qualified pulse     Once Smart pulse rate alarm limits are activated  alarm limits  will not change unless one of the following occurs     e 
39. gs may have returned to  power on defaults due to an internal instrument reset     Turn the instrument off and back on again  Check and  if  necessary  readjust alarm limits  Notify service personnel if it  is necessary to reconfigure power on defaults     EEE 058    Alarm and beep volume  audible alarm silence duration  and  operating mode may have returned to power on defaults due to  an internal instrument reset     Turn the instrument off and back on again  Check and  if  necessary  readjust alarm and pulsatile beep volume setting   audible alarm silence duration  and operating mode     Notify service personnel if it is necessary to reconfigure power    on defaults     EEE 081    The N  3000 is unable to interface with the sensor     Check the sensor and SCP 10 cables and the connections  To  clear the error  disconnect and reconnect the sensor     If the error code remains  replace the sensor and or the cable  If  the error code still persists  notify service personnel     Error Conditions    54    L A pulsing  shrill alarm sounds and the N  3000 does  not start up     e Thereis no power available to the N  3000  Press the  AUDIBLE ALARM OFF button to stop the alarm  The  battery has discharged below the shutdown point  and  either AC power is not available or the external power  supply has failed  Notify your service personnel     e The ON STANDBY button may be inoperable  in  which case the N  3000 may attempt to restart until  the battery is discharged or removed     
40. herefore in compliance with I SO 10993 1     Measurements  If the accuracy of any measurement does not seem  reasonable  first check the patient   s vital signs by  alternate means and then check the N 3000 for proper  functioning   Inaccurate measurements may be caused by     e incorrect sensor application or use    e significant levels of dysfunctional hemoglobins   such as carboxyhemoglobin or methemoglobin     e intravascular dyes such as indocyanine green or  methylene blue    e exposure to excessive illumination  such as surgical  lamps  especially ones with a xenon light source    bilirubin lamps  fluorescent lights  infrared heating  lamps  or direct sunlight    e excessive patient movement  e venous pulsations    e placement of a sensor on an extremity with a blood  pressure cuff  arterial catheter  or intravascular line    You can continue to use the N  3000 on a patient during  defibrillation  but the readings may be inaccurate for a  short time     Safety Information    Loss of pulse signal can occur in any of the following  situations     e the sensor is too tight    Ee    e there is excessive illumination from light sources  such as a surgical lamp  a bilirubin lamp  or  sunlight    e ablood pressure cuff is inflated on the same  extremity as one to which an SpO2 sensor is  attached    e the patient has hypotension  severe  vasoconstriction  severe anemia  or hypothermia    e thereis arterial occlusion proximal to the sensor    e the patient isin cardiac arrest 
41. icated to the alarm active function and patient data  will not be available  Refer to the N 3000 service  manual     Setup    The alarm active function  also located in the serial port  connector  may be used to monitor N 3000 alarms from a  remote location  The alarm active function remains active  whether the monitor is operating from an AC source or the  internal battery  The alarm active function allows use of a     nurse call    light  Alarm monitoring also applies to an  attached  active N 3100 or N  3200  Refer to the N 3000 service  manual for information concerning the required serial port  configuration to activate the alarm active function     ysi 6uR    WARNING  The alarm active function will not be  activated by alarms from other stacked instruments  unless all instruments are communicating with each  other  For proper operation  verify that the STACKED  indicator on each stacked instrument is on     WARNING  The nurse call feature should not be used as  the primary source of alarm notification  The audible  and visual alarms of the N 3000 monitor  used in  conjunction with clinical signs and symptoms  are the  primary sources for notifying medical personnel that an  alarm condition exists        25    SENSORS AND ACCESSORIES    Selecting a Nellcor Puritan Bennett Sensor  Biocompatibility Testing   Connecting SpO2 Sensors   Sensor Performance Considerations    Ee    WARNING  Use only one extension cable to increase the  length of the sensor  Use of more than one 
42. iority alarm sounds       PULSE SEARCH PULSE SEARCH indicator flashes    indicator lights  continuously       SpO2 and pulse rate SpO2 and pulse rate flashes zeros  alternates between  previous display    and dashes  oO    0 seconds 10 seconds 16 seconds 60 seconds          6 seconds           50 seconds        It during motion the N 3000 detects a qualified pulse  the values on the display are updated     If motion stops at any time during the pulse search mode  and no qualified pulse is detected   the high priority alarm will immediately sound after 6 seconds     Figure 16  Monitoring Conditions and Alarm Responses       37    Start Up and Use    38    When an N 3000 is stacked with an N  3100 blood pressure  monitor and the N 3100 begins a blood pressure measurement   the N  3000 is notified  Blood perfusion at the SpO 2 sensor on  the same limb as an N  3100 blood pressure cuff will be  momentarily interrupted during a blood pressure  measurement     Therefore  an N 3000 stacked with an N 3100 will be inhibited  from producing a loss of pulse alarm when the N  3100 begins a  blood pressure measurement  This happens even if the N 3000  SpO 2 sensor and the N 3100 blood pressure cuff are on  different extremities  The N 3000 will resume monitoring for  loss of pulse after the N  3100 has completed its measurement   including any needed measurement re attempts     The N  3000 manages Ioss of pulse alarms as follows  and as  illustrated in Figure 16  If the N  3000 fails to d
43. larm Silence Duration to Off    When operating in the stacked configuration  using the  shortcut method to set the N  3000 audible alarm silence  duration to OFF does not affect the audible alarm silence  duration of other instruments in the stack     To return to the previous alarm silence duration selection  or to  the power on default if no change was made   simultaneously  press the AUDIBLE ALARM OFF button and the LOWER  ALARM LIMIT button     ADJUSTING AUDIBLE ALARM VOLUME    48    To adjust the audible alarm volume  for alarms other than  microprocessor and power failure  press and hold the  AUDIBLE ALARM OFF button for at least 3 seconds  without  turning the knob or pressing any other buttons  until you hear  a tone  Observe the audible alarm volume range  expressed in a  value from 1 to 10 in the AUXILIARY display  A    1    represents  the lowest volume  and a    10    represents the highest volume  setting  Turn the control knob to adjust the volume     Sp02  PULSE RATE  min    L decl s     Start Up and Use    When operating in the stacked configuration  the audible alarm  volume level setting for each instrument remains as it was  before being stacked    Setting the audible alarm volume on one active instrument in a  stack  sets the audible alarm volume on any other active  instrument in the stack to the same level     Ee    Audible Alarm Off Reminder    An audible alarm off reminder sounds three beeps at 3 minute  intervals when the audible alarm silence duratio
44. larm limits     When used by itself  the knob adjusts pulse beep  volume on the N 3000  It is also used with other  controls to adjust alarm limits  audible alarm  silence duration  and alarm volume     DESCRIPTION OF VISIBLE INDICATORS AND DISPLAYS    12    Sp02     E     E  E  E                        PULSE RATE  min    OQ min    PULSE SEARCH  or          MOTION  or  d    The Sp02  display shows patient arterial  hemoglobin oxygen saturation level  It is updated  with each pulse  The Sp02  display flashes and  an alarm sounds when Sp0O2  is outside the  alarm limits  During Pulse Search  the display  will alternate between dashed lines and the last  qualified oxygen saturation measurement     The PULSE AMPLITUDE indicator is a  10 segment display that shows the relative pulse  amplitude  The bottom segment remains lit as  long as an SpO2 cable and SpO2 sensor are  connected to the N 3000  As the detected pulse  becomes stronger  more contiguous segments light  with each pulse     The PULSE RATE display shows the pulse rate in  beats per minute  It is updated with each pulse  It  flashes and an alarm sounds when the pulse rate  is outside the alarm limits  During pulse search   the display alternates between dashed lines and  the last detected pulse rate  If the detected pulse  is lost  a flashing zero is displayed and an alarm  sounds     The PULSE SEARCH indicator lights during  power up  prior to initial acquisition of a pulse  signal and during pulse search mode     The 
45. mits  Audible alarm volume   Audible alarm silence duration  Pulse beep volume   Serial port baud rate   Trend format    Values that can be adjusted in the service mode include the  following     e Alarm silence reminder  e Alarm latching    NELLCOR SYMPHONY INSTRUMENTS IN STACK    50    When NdIcor Symphony N 3000 and N  3100 or N  3200  instruments are connected together or    stacked     certain  functions are shared  Active instruments are those that are  connected together and are on     e TheAUDIBLE ALARM OFF button on any active  instrument silences the current audible alarm on all active  instruments in the stack  However  only the AUDIBLE  ALARM OFF button on the N  3000 is accessible when the  instruments are stacked     e Changing the audible alarm silence period and alarm  volume on one instrument in the stack will affect all active  instruments in the stack  unless using the shortcut  method described previously in this section      e When an instrument is added to a stack of instruments  it  will retain its audible alarm silence period and alarm  volume settings until manually changed     Start Up and Use    During operation  when the operating mode  neonatal or  adult pediatric  is adjusted on one instrument  all active  instruments in the stack are automatically set to that  mode    The knob on the N  3000 may be used to perform  adjustments on the N  3100 or N  3200 when a knob related  function is activated     Ee    One Nellcor Puritan Bennett external powe
46. n has been set  to OFF  The AUDIBLE ALARM OFF indicator continues to  flash until the silence duration is changed  The volume of the  audible alarm off reminder is equal to the alarm volume setting  or  if stacked with another instrument  equal to the highest  current audible alarm volume setting for any active instrument  in the stack     Note  Theaudible alarm off reminder may be disabled using  the service mode as indicated in the N 3000 service  manual     ADJUSTING PULSE    BEEP    VOLUME    The N  3000    beeps    with each detected pulse  The beep pitch  varies proportionately with percentage of saturation  rising as  SpO 2  increases toward 100   and falling as it decreases     EN Whilein normal  patient monitoring  mode  turn   E the knob to adjust pulse beep volume  When the  N  3000 is turned off and back on again  the pulse  beep returns to its power on default volume   Pulse beeps will continue during an audible  alarm     49    Start Up and Use    TECHNICAL STAFF ADJUSTABLE FEATURES    When the N 3000 is placed in the configuration mode by  qualified service personnel  they may adjust power on default  values of N 3000 features     Instructions for activating the configuration and service modes  and making these adjustments are found in the N 3000 service  manual  Power on default values that can be adjusted or set in  the configuration mode include the following     Operating mode  adult or neonatal   SpO 2  upper and lower limits  Pulse rate upper and lower li
47. ndicators  and Symbols              00aaaaaanaaaaaaaena aaaea 7  Displays  Controls  Indicators  and Connectors                 7  N 3000 Symbols AAA EE  10  Rear Kamelle ereeschen Eed  10  Right Side Panel  11  SPS External Power Supply Symbols               c cceeeeeeees 11  Description Of Controls            ccccceeeeeeeeeeeeeee eeeeeeeeeeeeeeeeeeaees 11  Description of Visible Indicators and Displays            0 10020   12  Description of Audible Indicators eeeenaes 15  Setup E 17  Performance Verification               cccccccscceeceeee ceeseeteeeeeeeeeeees 18  Connecting to an External Power Supply         sssesseesseeseee 18  Connecting the N 3000 to the SPS Power Supply                18  Connecting the N 3000 to the PSS Power Supply                20  Battery Operation           ccccceececeeeceeeeeee caeeeeeeeeseeeesaeeeeeeteaes 20  Stacking with the N 3100 or N 3200 es 21  Connecting the N 3000 to the N 3100 orhN 2200 21  Removing the N 3000 From the N 3100 or N 3200             23  Serial NGM ACCS wc  haveccewesttsctesand capchor Seed EE 24  Sensors and Accessories ceeeeteeeteneeeeaees 27  Selecting a Nellcor Puritan Bennett Sensor            nssnasnanaaens 28  Biocompatibility Testing          ec eeeeeeeeeesneeeee caeeeeeenaeeeeeeaaes 28  Connecting SpO2 Sensors ceeeeeessteeeees 29  Performance Considerations sesseeeeeeees 30  Start Up and Use eee eeee ce eeeeeeeeeeeeeeeseeeeeaeeeeaas eee 31  Power On and Get Test 31  Automatic Get Tests  32  N 3000 Turn On Proce
48. ns are based on testing the subject monitor  on healthy adult volunteers in induced hypoxia studies across  the specified range  Adult accuracy is determined with  Oxisensor II D 25 sensors  Neonatal accuracy is determined  with Oxisensor    N 25 sensors  In addition  the neonatal  accuracy specification is adjusted to take into account the  theoretical effect of neonatal blood on oximetry measurements     Pulse Rate  20 250 bpm  3 bpm    67    Specifications    ENVIRONMENTAL CONDITIONS    Transport and Storage  in shipping container     Temperature   40   C to  70   C   40   F to  158   F     Altitude B arometric Pressure   390 m to  6 096 m   1 280 ft  to  20 000 ft     1 060 hPa to  500 hPa   31 3 in  Hg to  14 in  Hg     Relative Humidity  15  relative humidity to 95  relative humidity   noncondensing     Transport and Storage  not in shipping container     Temperature   20   C to  60   C   4   F to  140   F     Altitude B arometric Pressure   390 m to  6 096 m   1 280 ft  to  20 000 ft     1 060 hPa to  500 hPa   31 3 in  Hg to  14 in  Hg     Relative Humidity    15  relative humidity to 95  relative humidity   noncondensing     Operation    68    Temperature   5   C to 40   C  441   F to  104   F     Altitude B arometric Pressure   390 m to 43 048 m   1 280 ft  to  10 000 ft     1 060 hPa to  700 hPa   31 3 in  Hg to  20 6 in  Hg     Relative Humidity  15  relative humidity to 95  relative humidity   noncondensing     APPENDIX  PRINCIPLES OF OXIMETRY    Pulse Oximetry O
49. nt Button   Battery in Use Battery Low Indicator  Battery Charging Indicator   Linked Indicator   RF Lock Indicator     Lower Alarm Limit Button   Upper Alarm Limit Button   Leads Off Indicator     Pulse Search Indicator   Patient Motion Indicator       Controls  Indicators  and Symbols                                                    20 19 18                      17 16 151413 1211 10    Figure 2  N 3000 Front Panel  International     1 SpO2  Display 11  2 Pulse Amplitude Indicator 12  3 ECG Heart Rate Indicator   13  4 Pulse Rate Display 14  5   Auxiliary Display 15  q 16  7 Audible Alarm Off Indicator 17  8  On Standby Button 18  9   Power On Indicator 19  10 Stacked Indicator 20      Not used on this model    New patient and neonatal   adult pediatric mode button    Print Button   Battery in Use Battery Low Indicator  Battery Charging Indicator   Linked Indicator   RF Lock Indicator    Lower Alarm Limit Button   Upper Alarm Limit Button   Leads Off Indicator     Pulse Search Indicator   Patient Motion Indicator    Serial interface connector                   Figure 3  N 3000 Rear Panel          Controls  Indicators  and Symbols    SpO2 cable connector   SCP 10     Ee       W    A CONNECTIONS ELECTRICALLY ISOLATED          External power supply or  docking connector    Figure 4  N 3000 Right Side Panel  North American     SpO2 cable connector   SCP 10              External power supply or  docking connector    Figure 5  N 3000 Right Side Panel  International     Controls 
50. o those from an  instrument that measures fractional saturation  fractional  measurements must be converted as follows       s fractional saturation  functional saturation         x100  100      carboxyhemoglobin     methemoglobin        Appendix  Principles of Oximetry  MEASURED VERSUS CALCULATED SATURATION    When arterial saturation  SpOz   is calculated from a blood gas  partial pressure of oxygen  PO2   the calculated value may differ  from the SpO2 measurement of a pulse oximeter  This usually  occurs because the calculated saturation  derived from a blood  gas analyzer  was not appropriately corrected for the effects of  variables that shift the relationship between PO2 and saturation   Figure A 1   pH  temperature  the partial pressure of carbon  dioxide  PCO 2   2 3 DPG  and fetal hemoglobin     t pH EE    Temperature X T    Ee    100               PCO2    2 3 DPG  Fetal Hb    E  lt    pH   2 50 S   Temperature     H   PCO2   E   2 3 DPG        on    50 100  PO2  mmHg     Figure A 1  Oxyhemoglobin Dissociation Curve    A 3    
51. onitor is not recharged  the monitor may fail  to operate or stop operating suddenly     20    Setup    When the N  3000 is connected to an external power supply or  is Stacked with an N  3200 connected to AC power  or stacked  with an N 3100 connected to an active SPS or PSS external  power supply  the self contained battery automatically  recharges  whether the N  3000 is on or off  A complete  recharge of a fully drained battery requires 14 hours     Ee    STACKING WITH THE N 3100 OR N 3200    If you wish to operate in the stacked configuration  use the  following instructions to connect the N 3000 to the N  3100 or  N 3200  These procedures assume that you have read the  appropriate N 3100 or N  3200 operator   s manual    When operating stacked with the N 3100  ensure that both  instruments are in the same operating mode  neonatal or  adult pediatric  and check all configurable settings     If the active instruments in the stack are in different operating  modes when power is turned on  press the NEW PATIENT   NEONATAL button on any of the active instruments twice   two presses within 2 seconds  as needed to place all active  instruments in the same operating mode     CONNECTING THE N 3000 TO THE N 3100 OR N 3200  The N  3000 must be the top unit in the stack   1  Connect the external power supply and power cable to the  N 3100  as indicated in the N  3100 operator   s manual  If    stacking with the N 3200  plug the power cord directly into  the power input connector on
52. or 2  A higher priority alarm occurs  microprocessor  power  failure    Start Up and Use    If the alarm condition ceases prior to the end of the alarm  silence duration  the AUDIBLE ALARM OFF indicator is  extinguished  While the condition exists  the indicator remains  lit for the duration of the selected alarm silence period or until  a higher priority alarm condition arises  in which case the  AUDIBLE ALARM OFF indicator goes out and an alarm  sounds     Ee    If the audible alarm silence duration is set to OFF  the  AUDIBLE ALARM OFF indicator flashes and no audible  alarms will occur  except those for low battery  power or  microprocessor failure  and sensor or sensor cable disconnect    Silence duration selection  including OFF  does not affect low  battery  power or microprocessor failure  or sensor or sensor  cable disconnect alarms     All Alarm Silence    Theall alarm silence feature is available only in models with a  software version of 3 4 3 or later  Qualified service personnel  can determine the software version using the procedures in the  configuration mode section of the service manual     Press the AUDIBLE ALARM OFF button twice within 2  seconds to silence all alarms for the alarm silence duration   whether an alarm is currently sounding or not  Three beeps  indicate that this feature has been activated and the AUDIBLE  ALARM OFF indicator lights continuously until the end of the  alarm silence duration  The all alarm silence mode is  automatically exited
53. or is in shock    N 3000 and Other Equipment    When connecting the N 3000 to any instrument  verify  proper operation before clinical use  Refer to the other  device   s manual for full instructions  Accessory  equipment connected to the monitor   s data interface  must be certified according to IEC Standard 950 for  data processing equipment or IEC Standard 601 1 for  electromedical equipment  All combinations of  equipment must be in compliance with IEC Standard  601 1 1 systems requirements  Anyone who connects  additional equipment to the signal input port or signal  output port configures a medical system and is  therefore responsible that the system complies with the  requirements of system standard IEC Standard 601 1 1   If in doubt  consult Nellcor Puritan Bennett   s Technical  Services Department or your local Nellcor Puritan  Bennett representative     Noninvasive blood pressure monitors may interrupt  N 3000 pulse rate measurements  creating false alarms   A stacked N 3100 Nellcor Symphony blood pressure  monitor informs the N  3000 of an inflation in progress   reducing the likelihood of false alarms  It is  recommended that only a stacked N 3100 blood pressure  monitor be used with the N 3000     INTRODUCTION    Intended Use  About this Manual    INTENDED USE    The purpose and function of the Nellcor Symphony N  3000  pulse oximeter is to continuously and noninvasively measure  functional arterial oxygen saturation of arterial hemoglobin   SpO2  and pulse rate
54. or leads to a new site     The sensor may be damp or may have been reused too  many times  replace it     An electrosurgical unit  ESU  may be interfering with  performance       Move the N 2000  cables and sensor as far from the  ESU as possible       Plug the N 3000 power supply and the ESU into  different AC circuits       MovetheESU ground pad as close to the surgical  site as possible and as far away from the sensor as  possible     5  Displayed SpO2 pulse rate does not correlate with  that of an ECG monitor connected to the patient     The patient may have a dysrhythmia  All beats  counted by the ECG monitor may not be perfused     Excessive patient motion may be making it impossible  for the N 3000 to acquire a good pulse signal  If  possible      keep the patient still      check whether the sensor is applied securely and  properly and replace if necessary      move the sensor to a new site  The patient may have a pronounced dicrotic notch     which could cause the pulse rate measurement to  double  Try another sensor site     Troubleshooting and Maintenance    e An ESU may be interfering with performance  Refer  to the discussion above     e Interference  artifact  or patient motion may be  affecting the accuracy of the ECG monitor     ysi 6uR    6 Oxygen saturation  SpO2  measurement does not  correlate with the value calculated from a blood gas  determination     e Calculated saturation values are inherently and  slightly different from pulse oximeter measurements 
55. pment does cause harmful  interference to radio or television reception  which can be  determined by turning the equipment off and on  the user is  encouraged to try to correct the interference by one or more of  the following measures     e  Reorient or relocate the receiving antenna     e Increase the separation between the equipment and  receiver     e Connect the equipment into an outlet on a circuit different  from that to which the receiver is connected     e Consult the dealer or an experienced radio TV technician  for help     PERIODIC SAFETY CHECKS    It is recommended that the following checks be performed  every 24 months     e     Inspect the equipment for mechanical and functional  damage     e Inspec the safety relevant labels for legibility     59    Ee    Troubleshooting and Maintenance    SERVICE    The N  3000 requires no routine maintenance other than  cleaning  battery maintenance and that which is mandated by  your institution     If the monitor has been visibly damaged or subjected to  mechanical shock  for example  when dropped   qualified  service personnel should perform the procedure in the  Performance Verification section of the service manual     If problems cannot be corrected by your service personnel  the  N 3000 should be returned to Nellcor Puritan Bennett for  service  Contact Nellcor Puritan Bennett   s Technical Services  Department or your local representative for return  instructions     CLEANING    WARNING  Do not spray  pour  or spill 
56. pper portion of the displays that are not selected  An example  of a pulse rate upper alarm limit is shown below     Sp02  PULSE RATE  min    To select a lower alarm limit  press the LOWER ALARM  LIMIT button  When the LOWER ALARM LIMIT button is  pressed  the lower alarm limit for SpO2 is displayed     Pressing of the LOWER ALARM LIMIT button again whilein  the view adjust alarm limits mode will display the lower alarm  limit for pulse rate and back to the SpO2 display when pressed  again  The display of a lower alarm limit is indicated by dashes  in the lower portion of the displays that are not selected  An  example of a pulse rate lower alarm limit is shown below     Sp02  PULSE RATE  min    E USCH    Start Up and Use    If the LOWER ALARM LIMIT button is pressed while an  upper alarm limit is displayed  the lower alarm limit for that  parameter will then be displayed     ysi 6uR    While the upper alarm limit or the lower alarm limit is  displayed and before 3 seconds have elapsed  turn the knob on  the top of the N 3000 to adjust the displayed alarm limit     Note  Ifthe SpO2 limits are set to their minimum  20   or  maximum  100   values  the SpO2 alarm is disabled  and will not sound  However  if the pulse rate limits  are set to their maximum  250 bpm  or minimum  30  bpm   the alarms are enabled and will alarm normally     Alarm Limits Changed Indicator    Sporx  If alarm limits have been changed from the  N 3000   s power on defaults  a decimal point will  appear a
57. r supply  connected to AC power can supply power to the N  3000  and N 3100  If the N  3200 is used  it can be directly  connected to AC power and provide power to any  combination of stacked instruments     Pressing the NEW PATIENT NEONATAL button for  3 seconds on either the N  3000 or N  3100 clears data for  both active instruments in the stack     When the N  3100 is connected to the N  3000  it informs  the N  3000 of blood pressure cuff inflation  The N 3000  may then correctly process pulse signals affected by cuff  inflation to prevent a nuisance alarm  N  3000 SpO2 alarms  will be silenced for the entire time the N  3100 is taking a  blood pressure measurement     When the N  3000 is connected to the N  3200  the N 3000  provides data for plethysmographic waveforms  The  N 3000 also provides SpO2 percentage  pulse rate  and  pulse amplitude for display in a tabular format on the  N 3200     When the N  3000 is connected to the N  3100  the initial  power on default operating mode may be different for the  two instruments  Both instruments in the stack should be  in the same operating mode  neonatal or adult pediatric    If the active instruments in the stack are in different  operating modes when power is turned on  press the NEW  PATIENT NEONATAL button twice  2 presses within 2  seconds  on the active instrument that is in the incorrect  mode as needed to place all active instruments in the same  operating mode     Note  The power on default operating mode can be 
58. remains lit as long as the N 3000 is  powered on and connected to a sensor cable and  SpO2 sensor  Other segments may light briefly     Ee    While the SpO2 cable with sensor is attached to the  monitor  or any time the N  3000 is attempting to acquirea  pulse signal  the first and third  first and fifth  or first and  seventh segments of the pulse amplitude indicator may  light briefly as shown below     UNI  a   DR RR  Q  UD    If the SpO2 cable with sensor is attached to the monitor  but not attached to the patient when you first power up  the N  3000  the SpO2  and PULSE RATE displays  indicate    0        Sp02  PULSE RATE  min    Note Other combinations  for example  the SpO2 cable  without sensor  will result in different initial  displays     CLEARING DATA    Caution  Pressing and holding the NEW  PATIENT NEONATAL button 3 seconds or more until  three beeps sound  clears all previously stored patient  data     If the N 3000 is being applied to a new patient  press and hold  the NEW PATIENT NEONATAL button for 3 seconds until you  hear three beeps to indicate that stored patient data is cleared   Alarm limits are not affected     33    Start Up and Use    NEONATAL MODE    34    WARNING  Each time the monitor is used  check alarm    limits to ensure that they are appropriate for the  patient being monitored        The default power on mode set at the factory is the adult   pediatric mode  To change the N  3000 from the adult pediatric  mode to the neonatal mode  press th
59. se is lost because  of deteriorating patient condition  the oximeter should alarm  quickly     However  normal interference can obscure the pulsatile signal   If interference is short lived  no alarm should sound    Such alarms are distracting and potentially time consuming   However  if interference persists  an alarm should notify the  attendant     The N  3000 loss of pulse alarm is designed for use in  environments in which the pulse can be obscured by  interference  Most loss of pulse alarms are attributable to  interference from patient motion which typically is brief     When the pulsatile signal is lost  the N 3000   s ability to  identify interference from motion becomes particularly  important because the response of the monitor is determined  by the presence or absence of interference from patient motion     Start Up and Use    Because a spontaneously moving patient can be assumed to  have a pulse  as long as continuous motion is detected  the                            N 3000 continues to search for the pulse  m  J  Qe  SPORADIC OR LOST PULSE WITH CONTINUOUS MOTION e  Low priority  alarm sounds  PULSE SEARCH indicator lights continuously   PULSE SEARCH  indicator flashes  SpO 2 and pulse rate Data display  alternates between previous flashes zeros  display and dashes  MOTION indicator lights continuously   0 seconds 10 seconds 16 seconds 60 seconds    m       6 seconds        gt        50 seconds       SPORADIC OR LOST PULSE WITHOUT CONTINUOUS MOTION       High pr
60. sensor  extension cable may have an adverse effect on  performance  Do not attach to the sensor port any cable  that is intended for computer use     WARNING  Before use  carefully read the sensor  directions for use  including all warnings  cautions  and  instructions     WARNING  Do not use a damaged sensor such as one  with exposed optical components     WARNING  Use only Nellcor Puritan Bennett sensors  and sensor cables with this monitor  Other sensors or  sensor cables may cause improper N 3000 performance        WARNING  Tissue damage can be caused by incorrect  application or duration of use of an SpO2 sensor     Inspect the sensor site periodically as directed in the  sensor directions for use     WARNING  Pulse oximetry readings and pulse rate  signals can be affected by certain ambient  environmental conditions  sensor application errors   and certain patient conditions        27    Sensors and Accessories    SELECTING A NELLCOR PURITAN BENNETT SENSOR       When selecting a sensor  consider the patient   s weight and  activity  the adequacy of perfusion  the available sensor sites   the need for sterility  and the anticipated duration of  monitoring  For more information  refer to Table 1 or contact  your local Nellcor Puritan Bennett representative     Table 1  Compatible Nellcor Puritan Bennett Sensors    Oxygen Transducer   Model   Patient Size    Oxisensor   I1 oxygen transducer    sterile  single use  3 to 20 kg  10 to 50 kg   gt 30 kg   gt 50 kg    Oxiband   
61. sors and Accessories describes Nellcor Puritan Bennett  sensors and accessories used with the monitor     Start up and Use explains how to operate the monitor   Troubleshooting and Maintenance provides information  about servicing the monitor and obtaining technical  assistance     Specifications lists technical specifications for the monitor   default alarm limit settings  and performance ranges     Appendix  Principles of Oximetry presents a clinical  explanation of the how pulse oximetry works        CONTROLS  INDICATORS  AND SYMBOLS    Displays  Controls  Indicators  and Connectors    N 3000 Symbols    SPS External Power Supply Symbols    Description of Controls  Description of Visible Indicators and Displays    Description of Audible Indicators    Ee       DISPLAYS  CONTROLS  INDICATORS  AND CONNECTORS    Figures 1 through 5 show the front  rear  side  and top views of  the N  3000 and identify displays  controls  and connectors           E    LI   ttt    HULSE RA min  N    PULSE SEARCH  1                woar E3                                               18                   17 1615 14131211 10    Figure 1  N 3000 Front Panel  North American     SpO02  Display    Pulse Rate Display  Auxiliary Display  Neonatal Mode Indicator    On Standby Button  Power On Indicator  Stacked Indicator    OANOARWD         CH      Not used on this model    Pulse Amplitude Indicator  ECG Heart Rate Indicator      Audible Alarm Off Indicator    11  12  13  14  15  16  17  18  19  20    Pri
62. stablished     Controls  Indicators  and Symbols    DESCRIPTION OF AUDIBLE INDICATORS    Audible indicators listed previously in this section have  symbols associated with them  The following list represents  tones for which there is no accompanying symbol or button     Valid keypress  Invalid keypress  Power on self tests passed       Smart    Pulse Rate Alarm  Limits Set    Pulse    beep       Ee    1 high pitched beep   1 low pitched beep   1 one second medium pitched beep  3 medium pitched beeps    Beeps that are synchronous with  each measured pulse    15    SETUP    Performance Verification   Connecting to an External Power Supply  Connecting the N 3000 to the SPS Power Supply  Connecting the N 3000 to the PSS Power Supply  Battery Operation   Stacking with the N 3100 or N 3200   Connecting the N 3000 to the N 3100 or N 3200  Removing the N 3000 from the N 3100 or N 3200  Serial Interface    Ee    WARNING  In the USA  do not connect to an electrical  outlet controlled by a wall switch because the device  may be accidentally turned off     WARNING  Do not subject the N 3000 to extreme  moisture  such as direct exposure to rain  Such  exposure may cause inaccurate performance or device  failure    WARNING  As with all medical equipment  carefully  route patient cabling to reduce the possibility of patient  entanglement or strangulation     WARNING  The N 3000 is not defibrillator proof   However  it may remain attached to the patient during  defibrillation or while an ele
63. tery is in use   When 15 minutes or less battery capacity  remains  the indicator flashes and a medium  priority alarm sounds     13    Controls  Indicators  and Symbols    14    WARNING  If the unit is turned off while the BATTERY  IN USE BATTERY LOW indicator is flashing and then  turned on again without recharging  the indicator may    take up to 30 seconds to begin flashing  even though the  battery has less than 15 minutes of use left  Always  recharge the battery after turning off an N 3000 with a    flashing                   PRINT       BATTERY IN USE BATTERY LOW indicator     The BATTERY CHARGING indicator is always  lit when the N 3000 is connected to AC power and  turned on  When the N 3000 is connected to AC  power and the monitor is turned off  the  BATTERY CHARGING indicator will remain lit  for approximately 14 hours and then turns off  If  there are any power interruptions  the 14 hour  charging period begins again when power is  restored     An illuminated BATTERY CHARGING indicator  is not necessarily an indication that the battery  contains less than a full charge  It is merely used  as a timer to indicate that the battery has been  continuously charging for less than 14 hours     The N  3000 may be used with a less than fully  charged battery but with a corresponding  decrease in operating time from that charge     The STACKED indicator lights when the N  3000  is stacked with the N 3100  the N 3200  or both  and communication between the instruments is  e
64. the  effects of nonpulsatile absorbers such as tissue  bone  and  venous blood     ysi 6uR    Appendix  Principles of Oximetry    AUTOMATIC CALIBRATION    Because light absorption by hemoglobin is wavelength  dependent and because the mean wavelength of LEDs varies   an oximeter must know the mean wavelength of the sensor   s  red LED to accurately measure SpO2  During manufacturing   the mean wavelength of the red LED is encoded in a resistor in  the sensor     During monitoring  the instrument   s software reads this  resistor and selects coefficients that are appropriate for the  wavelength of that sensor   s red LED  these coefficients are then  used to determine SpO 2  This resistor is read when the monitor  is turned on  periodically thereafter  and each time a new  sensor is connected     Additionally  to compensate for differences in tissue thickness   the intensity of the sensor   s LEDs are adjusted automatically     FUNCTIONAL VERSUS FRACTIONAL SATURATION    A 2    This monitor measures functional saturation     oxygenated  hemoglobin expressed as a percentage of the hemoglobin that  can transport oxygen  It does not detect significant amounts of  dysfunctional hemoglobin  such as carboxyhemoglobin or  methemoglobin  In contrast  CO oximeters such as the  L 482   report fractional saturation     oxygenated hemoglobin  expressed as a percentage of all measured hemoglobin   including measured dysfunctional hemoglobins  To compare  functional saturation measurements t
65. the  previous 5 days  If an alarm occurs anytime during that period   it is recorded     Format 3    Patient data is sampled for 5 seconds and the last  measurement during that period is recorded  Trend data is  available for the last 12 hours of actual monitoring during the  previous 5 days  An alarm is recorded only if the alarm was  active during the last measurement     35    Start Up and Use    The sample period for all three formats does not necessarily  have to be continuous  When the N  3000 is removed from the  patient  the sample period is stopped and then continued when  the N  3000 is again applied to the patient  For all three  formats however  any sample data that is over 5 days old is  discarded     Note  Only the data from the last format used is available   For example  if format 1 was used continuously for an  hour  then format 2 was used for 2 hours  only the 2  hours of format 2 data is available     DESCRIPTION OF ALARMS    Management of Loss of Pulse Alarm    36    Nalcor Symphony   N 3000 Oxismart     signal processing is  designed to reduce the adverse effects of normal interference  associated with motion  low patient perfusion  and spurious  electrical or optical signals  Figure 16 illustrates monitor  response in sporadic or lost pulse situations with and without  continuous motion     Loss of pulse presents an alarm management challenge for any  oximeter   sometimes it is caused by a clinical emergency   sometimes by signal distortion  When the pul
66. verview   Automatic Calibration   Functional versus Fractional Saturation  Measured versus Calculated Saturation    PULSE OXIMETRY OVERVIEW    The N  3000 uses pulse oximetry based on two principles  that  oxyhemoglobin and deoxyhemoglobin differ in their absorption  of red and infrared light  i e  spectrophotometry   and that the  volume of arterial blood in tissue  and hence  light absorption  by that blood  changes during the pulse  e  plethysmography    A pulse oximeter determines SpO2 by passing red and infrared  light into an arteriolar bed and measuring changes in light  absorption during the pulsatile cycle  Red and infrared low   voltage light emitting diodes  LEDs  in the oximetry sensor  serve as light sources  a photo diode serves as the photo  detector     Because oxyhemoglobin and deoxyhemoglobin differ in light  absorption  the amount of red and infrared light absorbed by  blood is related to hemoglobin oxygen saturation  To identify  the oxygen saturation of arterial hemoglobin  the monitor uses  the pulsatile nature of arterial flow  During systole  a new  pulse of arterial blood enters the arteriolar bed  and blood  volume and light absorption increase  During diastole  blood  volume and light absorption reach their lowest point  The  monitor bases its Sp02 measurements on the difference  between maximum and minimum absorption  i e    measurements at systole and diastole   By doing so  it focuses  on light absorption by pulsatile arterial blood  eliminating 
67. volume setting of 4    62 5 dB A  at 1 meter   volume setting of 5    60 seconds   On   Off   19 200 bits per second    10 second averaged   format 1     The PULSE RATE display updates in less than 2 5 seconds  with a 1 second change in SpO2 derived pulse rate from 30  pulses per minute  ppm  to 200 ppm     Alarm Limit Range   Adult Pediatric and Neonate    SpO 2 Upper Alarm Limit   SpO 2 Lower Alarm Limit   Pulse Rate Upper Alarm Limit   Pulse Rate Lower Alarm Limit     66    20  100   20  100   30 250 bpm  30 250 bpm    Specifications  Alarm Characteristics    Audible Characteristics    Ee    Pulse Repetition  Alarm Pitch Pulse Width   Interval  Priority   30 Hz      20 msec      20 msec     High oz  Medium nm  Low 500 Hz 3600 msec    Rise fall time for pulses is 16 milliseconds  3 milliseconds        Flashing Display Characteristics    During high or medium priority alarms  the  5 inch  red   front panel display with a parameter outside the alarm limits   flashes at the following rates     High priority   300 milliseconds ON  150 milliseconds OF F  Medium priority   750 milliseconds ON  600 milliseconds OF F    Accuracy  SpO2  Adults 70  100   2 digits  0 69  unspecified  Neonatal 70  100   3 digits  0 69  unspecified    Accuracies are expressed as plus or minus    X    digits   saturation percentage points  between saturations of  70 100   This variation equals plus or minus one standard  deviation  1SD   which encompasses 68  of the population  All  accuracy specificatio
68. y when it has not been charged for 6 months  or more     Caution  Follow local government ordinances and  recycling instructions regarding disposal or recycling  of device components  including batteries     TROUBLESHOOTING AND MAINTENANCE    Troubleshooting D  EMI Interference oy  Compliance with Test Limits for a Class B Digital Devece e  Periodic Safety Checks   Service    Obtaining Technical Assistance  Returning System Components    WARNING  The cover should be removed only by  qualified service personnel  There are no internal user   serviceable parts     WARNING  Do not spray  pour  or spill any liquid on the  N 3000 its accessories  connectors  switches  or openings  in the chassis     WARNING  If you are uncertain about the accuracy of  any measurement  check the patient   s vital signs by  alternate means  then make sure the monitor is       TROUBLESHOOTING  Error Codes    When the N  3000 detects an error condition  it displays an  error code similar to the illustration below     PULSE RATE  min    Sp02  K RI Sp02     If the number is other than one of those listed below  turn the  instrument off and back on again  If the error code reappears   record it  press the AUDIBLE ALARM OFF button to silence   the alarm  if present  and notify your service personnel     Error codes that begin with a    0    are user correctable  The  following N  3000 error codes are user correctable     53    Troubleshooting and Maintenance    EEE 055    SpO2 alarm limits and other settin
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual de Instalación PROTECT 4  Nady Systems TMP-1 MICPREAMP Stereo Amplifier User Manual  HIGH CAPACITY IN A SMALL SPACE  spiral arch design fireplace screen pare  こちらをクリック    Copyright © All rights reserved. 
   Failed to retrieve file