OPTICAL MEASURING INSTRUMENTS ML9001A

image

Contents

1. Calibration coefficient Adjustable Recorder output 1 V full scale linear output Range select Manual selection and automatic ranging Measurement mode Continuous and modulated light Wavelength sensitivity correction Automatic correction in 1 nm steps Data memory Max 1000 data via GPIB Dimensions and mass 213 W x 88 H x 250 D mm lt 4 kg For product ordering information see pages 4 8 www anritsu com 45 OPTICAL MEASURING INSTRUMENTS Anritsu Sensor Model MA9411A A1 MA9412A MA9611A Wavelength range 0 38 to 1 15 um 0 75 to 1 7 um Element Si photodiode InGaAs photodiode Active area diameter 9 5mm Input type Direct to photodiode Connector Connector 40 W x 32 H x 62 73 D mm 61 W x 42 H x 110 D mm Dimensions and mass lt 400 9 lt 800 g 40 W x 32 H x 65 D mm lt 400 g Model MA9612A MA9711A A1 MA9712A MA9714B Wavelength range 0 75 to 1 7 um 0 75 to 1 8 um Element InGaAs photodiode Ge photodiode Cooled Ge photodiode Active area diameter 5mm Input type Connector Direct to photodiode Connector l 61 W x 42 H x 110 D 40 W x 32 H x 62 73 D 42 W x 47 H x 110 D 47 W x 61 H x 128 D Dimensions and mass mm lt 800 g mm lt 400 g mm lt 500 g mm lt 800 g Overall Model
2. 3 MA9412A Optical Power Sensor 0 38 to 1 15 um MA9612A Optical Power Sensor 0 75 to 1 7um m gt Sensor connecting cord A _ MA9412A Optical Power Sensor 0 75 to 1 8 um 4 Ami ML9001A Optical Power Meter MA9611A Optical Power Sensor 0 75 to 1 7 um y Sensor connecting cord B a ial bh Oar Gil m Lof Fe oun MA9411A Optical Power Sensor 0 38 to 1 15 um MA9711A Optical Power Sensor 0 75 to 1 8 um gt Adapters option gt hy MAQ805A Attenuator Ly er gt H MA9001B MA9411A 9711A Connector Adapter gt Optical Power Sensor MA9306A Attenuator R Ed gt lt j MP92B Aa o Sensor Fiber cord with plug Connector Adapter MA9611A Optical Power Sensor mounted on MA9005A i 2 s on a gt Pigtail fiber MA9013A Fiber Adapter MA9412A 9612A Optical Power Sensor gt Ez gt lt j MA9805A Attenuator _ Pr i MA9411A 9711A MP93A Fiber Adapter MASODAA AgAPIET E Optical Power Sensor Ls a MA9306A Attenuator p MA9712A Plastic fiber MP916A Fiber Adapter MP94A Adapter Optical Power Se
3. MA9411A A1 MA9412A MA9611A Optical ponor Continuous light 70 to 10 dBm 4 0 1 nW to 10 mW 90 to 0 dBm 4 1 pW to 1 mW 70 to 3 dBm 5 0 1 nW to 2 mW measuremen jande Modulated light 70 to 7 dBm 0 1 nW to 5 mW 90 to 3 dBm 6 1 pW to 0 5 mW 80 to 0 dBm 7 10 pW to 1 mW Absolute accuracy 5 8 0 5 to 0 95 um 5 9 1 0 to 1 6 um 23 dBm Measurement Li it ii accuracy fight 23 C 23 0 15 dB 0 15 dB 9 0 15 dB ABin as elerance 0 45 dB for 70 to 60 dBm 0 45 dB for 90 to 80 dBm 0 45 dB for 70 to 60 dBm Resolution W W REL display 0 1 to 1 dBm display 0 01 dB dB REL display 0 001 dB Power 100 115 120 200 220 Vac 12 240 Vac 50 60 400 Hz lt 40 VA Operating temperature 0 to 50 C EMC EN55011 1991 Group 1 Class A EN50082 1 1992 Safety EN61010 1 1993 Installation Category I Pollution Degree I Model MA9612A MA9711A A1 MA9712A MA9714B 100 to 3 dBm 5 40 to 10 dBm 5 60 to 10 dBm 5 47 to 23 dBm apuca porer Continuous light 0 1 pW to 2 mW 0 1 uW to 10 mW 1 nW to 10 mW 20 nW to 200 mW mi 7 _ 7 7 i 13 range Modulated light 90 to 0 dBm 60 to 7 dBm 70 to 7 dBm 57 to 20 dBm 1 pW to 1 mW 1 nW to 5 mW 0 1 nW to 5 mW 2 nW to 100 mW Absolute accuracy ore oy 9 4 5 1 3 um 4 5 1 55 pum 4 23 dBm Soe ie Le um 15 0 95 to 1 5 um 5 0 95 to 1 6 um 5 0 95 to 1 6 pm 5 Measurement ET 0 ET He accuracy Linearity continu
4. NGE 23 Amphenol 906 type 34 DIAMOND g 3 5 35 1 HP SMA Amphenol 905 type 36 Amphenol 905 type 38 ST 39 DIN 40 SC 41 TOCP172 43 HMS 10 A 45 FC 1 If adapter mounted on MA9412A 9612A repeatability may be reduced 2 For MA9411A For product ordering information see pages 4 8 Ordering information Please specify model order number name and quantity when ordering Model order No Name ML9001A J0313 J0314 F0004 F0007 W0420AE W0420BE MA9411A A1 1 MA9412A MA9413A MA9611A MA9612A MA9711A A1 MA9712A MA9714B 2 MA9001B 8 J0480A 3 J0480B 3 MA9005A 8 MP92B 8 MA9013A S MP916A MP93A MP94A MA9002A MA9805A MA9306A MZ8010A J0007 J0008 B0186 J0617B 4 J0618D 4 J0618E 4 J0618F 4 J0619B 4 Main frame Optical Power Meter Standard accessories for ML9001A Sensor connecting cord A 2m for MA9412A 9413A 9612A 9712A 1 pc Sensor connecting cord B 2m for MA9411A A1 MA9611A and MA9711A A1 1 pc Power cord 2 5 m 1 pc Fuse 0 4 A T400MA250V 2 pcs Fuse 0 8 A T800MA250V 2 pcs ML9001A operation manual 1 copy ML9001A service manual 1 copy Optical power sensors Optical Power Sensor Optical Power Sensor with JO480A connector adapter Optical Power Sensor Optical Power Sensor with MA9005A connector adapter Optical Power Sensor with JO480A connector adapter Optical Power Sensor Optical Power Sensor Optical Power Sensor Optional a
5. ccessories Connector Adapter FC type for MA9411A MA9711A Connector adapter FC type for MA9412A Connector adapter FC type for MA9612A Connector Adapter FC type for MA9611A Connector Adapter FC type for MA9413A MA9712A Fiber Adapter with FC type plug for fibers with 125 um clad dia 0 25 to 1 0 mm jacket dia Fiber Adapter for MA9002A and MP94A for plastic fiber with 1 mm dia Fiber Adapter lt 150 um clad dia 0 8 to 1 0 mm jacket dia Adapter for MA9413A MA9712A used with MP93A Adapter for MA9411A MA9711A used with MP93A Optical Attenuator for MA9411A 10 dB Optical Attenuator for MA9711A 10 dB Optical Sensor Holder securely mounts MA9411A A1 or MA9711A A1 for measuring light traveling through free space GPIB cable 1 m GPIB cable 2 m Front cover Replaceable optical connector FC Replaceable optical connector ST Replaceable optical connector DIN Replaceable optical connector HMS 10 A Replaceable optical connector SC 1 MA9411A1 and MA9711A1 are lateral input sensors 2 Specify one of FC ST DIN SC or DIAMOND HMS 10A When the connector type is not specified FC is supplied 3 The optical connector of the standard product is FC Please specify the option numbers along with model names shown in the tables if you need a different optical connector 4 For MA9714B www anritsu com 47 m OPTICAL MEASURING INSTRUMENTS Anritsu ML9001A with sensor
6. m OPTICAL MEASURING INSTRUMENTS Anritsu OPTICAL POWER METER ML9001A OPTICAL POWER HETER f Anitu A Variety of Optical Sensors such as Si Ge and InGas MLL ATIJ C GPIB The ML9001A is a single channel digital display optical power meter It ensures accuracy and linearity over a wide wavelength range and greatly improves measurement reliability It also has an improved ba sic performance For example measurements can be made over the wide level range from 100 to 20 dBm because internal reflection in the power sensors has been suppressed The ML9001A also has many new functions that make it easier to use than other power meters It can be used for all optical power measurements such as optical fiber loss measurement and optical device performance evaluation Features e Enables high accuracy measurement The ML9001A accurately and automatically calibrates all the power sensors within the specified wavelength range and ensures a 5 accuracy at 23 dBm It also has a 0 15 dB linearity 23 dBm ref erence value The ML9001A extends the guaranteed accuracy range of the measured values and enables high accuracy measurement One power sensor for repeater maintenance and long distance fiber loss measurement The MA9612A Optical Power Sensor has ultra high sensitivity Its mea surement level range is 100 to 3 dBm in the 1 3 um band and it can sense either continuous light or modulated light A
7. nsor qs gt gt ny MA9306A 9805A MA9411A 9711A Light traveling through free space Attenuator Optical Power Sensor y MA9411A1 9711A1 Optical Power Sensor lateral input sensors y MA9712A Optical Power Sensor 48 www anritsu com For product ordering information see pages 4 8
8. ous 0 15 dB 0 15 dB 0 15 dB 0 15 dB 37 to 20 light 23 C 23 0 45 dB for 90 to 80 0 45 dB for 40 to 30 0 45 dB for 60 to 50 dBm 0 45 dBm for 47 to dBm as reference dBm dBm dBm 37 dBm Resolution W W REL display 0 1 to 1 dBm display 0 01 dB dB REL display 0 001 dB Power 100 115 120 200 220 Vac 42 240 Vac 50 60 400 Hz lt 40 VA Operating temperature 0 to 50 C EMC EN55011 1991 Group 1 Class A EN50082 1 1992 Safety EN61010 1 1993 Installation Category II Pollution Degree II 1 Twelve modulation frequencies including 270 Hz and 1 kHz 9 For wavelengths other than 1 3 um specified at 23 5 C 2 FC type connector standard 10 At 23 45 C 3 Only for PC type SM fiber 10 125 um NA 0 1 11 Electromagnetic compatibility 4 At 0 85 um 12 At 1 55 um 5 At 1 3 um 13 At 1 55 um 270 Hz 6 At 0 85 um 270 Hz 14 At 1 55 um 0 dBm 7 At 1 3 um 270 Hz 15 At 0 dBm 8 For wavelengths other than 0 85 um specified at 23 5 C 16 Reference 0 dBm Note When an optical fiber is used performance is guaranteed for a fiber core diameter of up to 62 5 m and an NA of up to 0 29 When any other fiber is used a measurement error may occur 46 www anritsu com For product ordering information see pages 4 8 m OPTICAL MEASURING INSTRUMENTS Anritsu e Optical connector options Option No Optical connector 21 D4 22 RU
9. single MA9612A can measure the near end and far end outputs of a repeater as well as measure long distance fiber losses Specifications e ML9001A Optical Power Meter Indicator e Multi core fiber cable losses easily measured For modulated light measurement the ML9001A has 12 modulation frequencies including 270 Hz This meter can easily be used to measure multi core fiber cable losses by using it with the MG9002A Stabilized Light Source mounts up to 12 light source units e Interchangeable optical connectors The optical connectors of all the power sensors accept adapters This system allows the optical connectors to be interchanged so the ML9001A can be quickly used with various optical connectors Since the inter nal coating of the optical power sensors suppresses reflected light measurement errors are reduced in beam measurement with or without an optical fiber Reduced measurement time The ML9001A has a much better response speed and stability than conventional optical power meters With GPIB it can measure at 30 ms point so the measurement time can be reduced to less than 50 of conventional automatic measurement High performance optical loss test set Stacking the ML9001A with the MG9001A Stabilized Light Source quickly configures a high performance optical loss test set Selecting various light source units and enables the ML9001A to measure all optical losses Display 4 digit W WireL dBm dBiret selectable

Download Pdf Manuals

image

Related Search

Related Contents

AOC MONITOR USER`S GUIDE 2036S  Manual Lavadora Agile Family Mais e Family Stile.indd  TP8-64 - Gladiusnet S.r.l.  hilna g2v1 - NuWaves Engineering    COMPTE RENDU DE LA REUNION Du 25 juin 2013 3ème  114A4262 Models 3130M,3240M Owner`s Manual  利用規約に同意する  Wireless Presentation Adaptor User`s Manual  Ksix B8989FTP05 mobile phone case  