        GE Dinamap Dinawin.dll Usage Instructions
         Contents
1.      lt Result name  Body_temperature  gt    lt Value gt 98 6 lt  Value gt    lt Units name  F  gt  lt  units gt     World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1 414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790     lt Method gt predicted lt  Method gt     lt Time_stamp year  2004  month  7  day  30  hour  13  minute  2  second  04  gt  lt  Time_stamp gt    lt  Result gt    lt Result name  Pulse  gt     lt Value gt 84 lt  Value gt     lt units name  BPM  gt  lt  units gt     lt Method gt pulse_ox lt  Method gt     lt Time_stamp year  2004  month  7  day  30  hour  13  minute  15  second  23  gt  lt  Time_stamp gt    lt  Result gt     lt  Vitals gt     Additional Information on Results XML  e Vendor tag  Monitor Manufacturer     General Electric Healthcare      e Model tag  PCDATA within this tag indicates the model name or number of the monitor     e Result tag  This is the parent tag of all elements that  taken together  describe the result of a single vital sign  measurement     Name attribute  Contains CDATA indicating the name of the vital sign     Possible names   Systolic_blood_pressure  Diastolic_blood_pressure  Mean_arterial_pressure  Oxygen_saturation  Body_temperature   Pulse    e Value tag  This contains PCDATA representing the value of the measurement  It shall be text for the decim
2.    REQUIRED   gt            lt  ELEMENT Value   PCDATA  gt    lt  ELEMENT Units  HPCDATA  gt    lt  ATTLIST Units name  F C mmHg BPM     REQUIRED gt    lt  ELEMENT Time_stamp   PCDATA  gt    lt  ATTLIST Time_stamp year CDATA  REQUIRED  month CDATA  REQUIRED  day CDATA  REQUIRED  hour CDATA  REQUIRED  minute CDATA  REQUIRED  second CDATA  REQUIRED     gt    lt  ELEMENT Method   PCDATA  gt    lt  ATTLIST Method name CDATA  REQUIRED gt   Page 8 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1 414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790    Workflow Example     0  coo OY GT Ee oY i  A    e e  e Oo    12     Host loads DinaWin dll at startup   Host calls getBufferLength to determine memory allocation for the returned XML string   Host calls checkReadiness to verify that the monitor is connected and powered   Host prompts user to identify the patient   Host prompts user to measure the patient s vitals with the monitor   Host prompts user to hit a button to retrieve the patient s vitals from the monitor   Host calls getState when the retrieve button is hit   Host parses the returned XML string and displays the data on the screen with the patient s identity  Host prompts the user to verify the information and to hit a send save button for confirmation    Host sends save
3.  Linked Library  DLL  is provided to enable a host application running on a PC with Microsoft  Windows   xP 2K to interface with a GE Medical Systems Information Technologies DINAMAP   PRO Series 100 400  Monitor or DINAMAP   ProCare Monitor without requiring knowledge of the low level protocol  This product may be  used to extract the most recent values for blood pressure  saturation  and predictive temperature     Warnings   Recommendations    These warnings recommendations are being supplied as mitigations to the hazards identified in the risk analysis of  this software     End User Warnings   Recommendations    GE Medical Systems Information Technologies strongly recommends that the following warnings be supplied to  the end user of this software     The specified adapter and cables  See    Adapter and Cables    in the System Requirements section below   electrically isolate the monitor from the host PC and MUST be used to protect the patient and clinician from  electrical shock     The system time settings on the host PC must be properly adjusted to assure that the time stamps on the  parameter values are accurate     If not implemented as an automatic process in the host application  the resetMonitor function must be  manually initiated by the end user between each patient to reduce the likelihood of associating the data with  the wrong patient  The user must also be warned if this function fails     The blood pressure remote operations feature of the monitor must be di
4.  with DinaWin DLL using single threaded interface where each request  will be processed and returned in the same order that is was transmitted     Operating Mode    Request Response Mode   In this mode DinaWin DLL waits for a request from the host application before sending a  response  This involves the DLL listening for incoming commands and responding to valid ones     Troubleshooting    This section provides a basic troubleshooting guide     Connections   e Check the connections at the monitor  at both ends of the ILC adapter  and at the PC     e Assure that the opposite end of the cable connected to the monitor is connected to the end of the ILC marked     Dinamap   Monitor         Monitor Configuration     e Assure that the monitor   s baud rate is set to 9600   See the service manual for detailed instructions      Power     e The ILC is powered from the monitor s host communications port  Verify that 4 5Vdc 5 5Vdc is measured  between pin 4    and pin 1    on the monitor   s comm  port  If the voltage is not present  call GE Medical  Systems technical support  800  558 7044  Option 2     Loop Back Test     e if using a USB RS232 adapter  short pins 2  amp  3 on the USB RS232 adapter   s RS232 connection    e Start the HyperTerminal application on the PC and configure it for the port being used to connect the monitor  and a baud rate of 9600    e lf typed characters are echoed  check the connections to the ILC or the ILC may be connected backwards    e If typed characte
5. I 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1414 355 5000 or 1 800 558 5120  US only   Fax    1414 355 3790       Revision B    Checking the Installed DLL Version    1  Use Windows explorer to navigate to DinaWin installation folder   Program Files  Vista Vitals Plugins   and  use the mouse right click on DinaWin dll file     2  Then click on the    Version    tab to view DinaWin    File version    on the first line from the top     Monitor Connections and Settings  1  Connect the cable  683235  between the monitor s communications port and the ILC adapter  001926      2  Connect the cable  683242  between the ILC adapter  001926  and the RS232 port on the PC or the RS232 port  on a USB RS232 adapter   NOTE  IF using a USB RS232 adapter  the adapter must be configured for COM1 COM10  If the adapter  requires a port greater than COM10  call GE Medical Systems technical support  800  558 7044  Option 2     3  Configure the monitor for a baud rate of 9600  Refer to the monitor   s service manual for instructions on how  to configure the baud rate     Programming Reference    The host application issues commands to the DinaWin DLL  The DLL then issues appropriate commands to the  monitor  The monitor returns results to the DLL and the DLL then returns appropriately formatted results to the host  application     If the host application requests vital signs measurements  the DLL responds with a small XML document repre
6. ROM drive     2  Double click on    Setup exe    to proceed to the following Welcome dialog box     i  DinaWin    Welcome to the DinaWin Setup Wizard    The installer will guide you through the steps required to install DinaWin on your computer     WARNING  This computer program is protected by copyright law and international treaties   Unauthorized duplication or distribution of this program  or any portion of it  may result in severe civil  or criminal penalties  and will be prosecuted to the maximum extent possible under the law        3        i DinaWin    Select Installation Folder    The installer will install Dina Win to the following folder     To install in this folder  click  Next   To install to a different folder  enter it below or click  Browse      Folder     C  Program Files Vista Vitals Plugins      Install Dina Win for yourself  or for anyone who uses this computer     O Everyone     Just me       4  Select the destination folder  or accept the default destination folder  Program Files  Vista Vitals Plugins  and  leave    Just me    if only you will use this DLL or select    Everyone    if other users will use the DLL     Page 4 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1414 355 5000 or 1 800 558 5120  US only   Fax    1414 355 3790    5  When you are done cl
7. al  number of the value  None of these measurements require scientific notation and it is not intended that the  program encode the binary value of the number in any way  It shall be represented as human readable text     e Units tag  This contains PCDATA indicating the units that apply to the value of the measurement  The DTD  specifies these units     Possible units tags    F   Temperature in Fahrenheit   C   Temperature in Celsius   mmHg   Blood pressure in millimeters of mercury  BPM   Pulse rate in beats per minute       Percent of blood oxygen saturation    e Method tag  This contains PCDATA that represents the method used to produce the result     Possible method tags   Predicted   The temperature operational mode  Monitor mode is not supported    pulse_ox or blood_pressure   The parameter that derived the reported pulse rate     e Time_stamp tag  The time tag for the results is calculated from the age of the result and the PC system time   which is independent of the monitor   s system time  Therefore  the monitor s time is not required to be  synchronized with the PC   s system time     Page 10 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1 414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790    Multi Threading    The host application is expected to interface
8. gehealthcare com    DinaWin Software  Instructions for Use    DinaWin Software Version 1 2 0 3    PURPOSE   Intended USC siones ron r See es ro r ea ni leds cedetededgesminys 2  Warnings   Recommendations            sseseesesessessesseessesreoseoseessesreoseoseessessroseoseoseessesseoseeseeseeesee 2  System Requirements issu  ct evzevabetetenotssets vation oto vakovadeeshouotssobayaton skbatovabovatovabssabavalenolovestvovecetens 3  RELEASE NOTES ninen sen ast sarcastic eo iee e eS tee rO ES E Eees EO SS EE OnE E ES paa a 3  Software Instdllqnon nsstctetatatotinioted atetobehededatehalnianateiuanes 4  SOMWGL SC REMOUG 42scssd c deccderesetetsen tects svasasssestecscsantecscscacccacsentscuarebenauseomen caeseraen samanseunsuassensns 6  Checking the Installed DLL VERSION  cssstsicssascsaedscadhtadadsdscsdscaci  ndatsdadedosududadasstadadadodoandsdstecans 7  Monitor Connections and Settings    eesesscsssesesesesesesescssssssssesesesesescsssssssceceeeseeceees 7  Programming RETCFENCE yivictscatuceteivecteseseastovetcnesoedeiotebagusnceseperetessaetedsbebesedrarigsnbesueverserscniens 7  TOUS STOO LING cst Oia AEEA E AOA 11  Page 1 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790    Purpose   Intended Use    This Dynamically
9. ick    Next  gt     to proceed to the following Confirm Installation dialog box     i DinaWin    Confirm Installation    The installer is ready to install Dina Win on your computer     Click  Next  to start the installation        6  Click    Next  gt     to proceed to the following Installation Complete dialog box     i DinaWin    Installation Complete    Dina win has been successfully installed     Click  Close  to exit        7  Click    Close    to finish the DinaWin installation     8  Verify that the DinaWin dll file was created in the installation folder specified during Step 4     Page 5 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1414 355 5000 or 1 800 558 5120  US only   Fax    1414 355 3790    Software Removal  1  Insert the DinaWin CD into the CDROM drive     2  Double click on    Setup exe    to proceed to the following Welcome dialog box     i  DinaWin    Welcome to the DinaWin Setup Wizard    Select whether you want to repair or remove Dina Win     O Repair Dinawin    3     if DinaWin    Installation Complete    Dina win has been successfully removed     Click  Close  to exit        4  Click    Close    to complete the removal     Page 6 of 11 2031886 001    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  W
10. n Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1 414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790    System Requirements  Operating System   e   Microsoft Windows   XP or Windows   2000    Free Disk Space   e 100KB    Adapter and Cables     e 001926 Isolated Line Converter  ILC  Adapter  e 683235 Cable  Monitor to ILC Adapter   e 683242 Cable  PC to ILC Adapter     Release Notes  Version 1 2 0 3    e   DinaWin will not return monitor values older than the system time when resetMonitor   was called  monitor  values older than the system time when the DLL was loaded  or monitor values older than 30 minutes     e    resetMonitor   intentionally does not clear the displays on the monitor   e DinaWin does not return the serial number of the monitor     e   By default  DinaWin sequentially searches COM1 COM10 for the monitor  When the DLL loads  DinaWin  checks for a configuration file that can change the default COM port search range to COM 255 and initial port  to search     Page 3 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1 414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790    Software Installation  1  Insert the DinaWin CD into the CD
11. rs are not echoed  the USB RS232 adapter may not be properly installed or the adapter is  configured to use a COM port greater than COM10  If a COM port greater than COM10 is required  call GE  Medical Systems technical support  800  558 7044        Page 11 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1 414 355 5000 or 1 800 558 5120  US only   Fax    1 414 355 3790    
12. s the information when the send save button is hit      Host calls resetMonitor to prevent the data in the monitor from being retrieved the next time getState is    called    Repeat steps 4 11 for each patient    Example Results Document    This is an example of the results document that is returned by the getState command      lt  xml version  1 0  encoding  ASCII   gt    lt Vitals gt     Page 9 of 11 2031886 001 Revision B     lt Vendor gt General Electric Healthcare lt  Vendor gt    lt Model gt Pro Procare lt  Model gt    lt Result name  Systolic_blood_pressure  gt    lt Value gt 120 lt  Value gt    lt Units name  mmHg  gt  lt  units gt      lt Time_stamp year  2004  month  7  day  30  hour  13  minute  12  second  30  gt  lt  Time_stamp gt      lt  Result gt     lt Result name  Diastolic_blood_pressure  gt    lt Value gt 80 lt  Value gt    lt Units name  mmHg  gt  lt  units gt      lt Time_stamp year  2004  month  7  day  30  hour  13  minute  12  second  54  gt  lt  Time_stamp gt      lt  Result gt     lt Result name  Mean_arterial_pressure  gt    lt Value gt 103 lt  Value gt    lt Units name  mmHg  gt  lt  units gt      lt Time_stamp year  2004  month  7  day  30  hour  13  minute  13  second  12  gt  lt  Time_stamp gt      lt  Result gt     lt Result name  Oxygen_saturation  gt    lt Value gt 95 lt  Value gt    lt Units name      gt  lt  units gt      lt Time_stamp year  2004  month  7  day  30  hour  13  minute  15  second  05  gt  lt  Time_stamp gt      lt  Result gt
13. sabled to prevent inadvertent control  of the blood pressure parameter  See the monitor s user s manual for instructions to disable this feature     Monitor mode temperature values are not reported by the communications protocol  so the end user must be  warned not to expect these values when using this software     The user must be warned not to connect multiple monitors to the same PC  since the DLL will only report  values from the first monitor it finds     The user must be warned not to connect any other device to the same PC because DinaWin   s search feature  may interfere with the other device s operation     Implementation and System Warnings   Recommendations    A confirmation step must be added that allows the end users to confirm that the values were accurately  transferred from the monitor  The confirmation step reduces the likelihood of associating the wrong data with  the current patient     The resetMonitor function reduces the likelihood of associating the data with the wrong patient by  disallowing data older than the PC s time when the function was called  GE Medical Systems Information  Technologies recommends that the host application automatically call this function when a new patient is  identified     The host system must provide anti virus protection and restrict remote access to prevent malicious or  unintended interference with the operation of this software     Page 2 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Informatio
14. senting  the results of the request  The XML document conforms to the Document Type Definition  DTD  supplied in this  specification     The XML document shall not include any result sections that were not measured or not supported by the monitor  If   for example  temperature  SpOz  and pulse are taken but not blood pressure  the result sections pertaining to blood  pressure are to be omitted altogether as opposed to sending a value of 0     Commands    e checkReadiness    Description  This command determines whether the software can communicate with the monitor and  whether the monitor can appropriately respond    Parameters  None    Return type  Boolean  32bit signed integer  O FALSE  O TRUE     Return values  True if the software can properly communicate with the monitor  false otherwise     e getBufferLength    Description  This command returns the XML message buffer in a maximum length of 8bit characters   Parameters  None    Return type  Boolean  32bit signed integer    Return values  The  unchanging  maximum size of the buffer the DLL has allocated for XML return strings     Page 7 of 11 2031886 001 Revision B    World Headquarters  GE Medical Systems Information Technologies  Inc  8200 West Tower Avenue  Milwaukee  WI 53223 USA    2006 General Electric Company going to market as GE Healthcare  All rights reserved   Tel    1414 355 5000 or 1 800 558 5120  US only   Fax    1414 355 3790    e getState    Description  This command is a request to retrieve the current val
15. ues from the monitor  The values are  returned in an XML formatted buffer per the DTD specification section of this document    Parameters  None    Return type  Pointer    Return value  A pointer to the beginning of the buffer containing the string that is an XML document  conforming to this specification and containing the information about the current state of the monitor  i e    the vital signs currently in the monitor  The string is represented by a null terminated character array and is  encoded as ASCII text     e  resetMonitor    Description  The monitors retain the last temperature and blood pressure values for 90 120 minutes  This  function prohibits monitor values older than the function   s invocation time from being returned with the  getState function  This provides the host application a means to avoid confusing readings from a prior patient  with the current patient s readings  This function must be called between patients    Parameters  None   Return type  Boolean  32bit signed integer  O FALSE  0 TRUE    Return value  True if the monitor was reset  false otherwise     DTD for Vital Signs Results  Below is the XML DTD      lt  xml version  1 0 gt     lt  ELEMENT Vitals  Vendor Model Result   gt     lt  ELEMENT Vendor   PCDATA  gt     lt  ELEMENT Model   PCDATA  gt     lt  ELEMENT Result  Value units Time_stamp  gt     lt  ATTLIST Result name Systolic_blood_pressure   Diastolic_blood_pressure   Mean_arterial_pressure   Oxygen_saturation   Body_temperature Pulse
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
22mm WRX Rear Sway bar Upgrade  Philips Dustbin CRP770  La Cité - Lacite DZ  BroadWorks Assistant–Enterprise User Guide  ELECTRIC STOVE  Tascam 102MKII User's Manual  dossier de presse  Samsung GT-C3630 User Manual  Kelly KHB High Power Brushless Motor Controller User`s Manual    Copyright © All rights reserved. 