        40-6900 22x Builders Level Service Manual
         Contents
1.        method                Prepared By  Tim Wojo    Revision  1 030309    Page 7 of 7    i   ae    
2.  Compression Spring 1   6 9 AP2009   Horizontal Micro Motion knob 1 6 18 AP2016   Ejector Rod 1                         1  Remove horizontal micro motion loop  6 14  from Telescope Assembly  6  by continuously turning the horizontal micro motion  locking knob  6 15  counter clockwise until it is comes out    2  Remove Vial Seat Assembly  6 8  by removing the 2 retainer screws  6 7     3  Remove horizontal micro motion loop  6 9  from Telescope Assembly  6  by removing retaining screw  6 10  then continuously  turning the horizontal micro motion loop  6 9  counter clockwise until it is comes out    4  Remove horizontal veneer  6 12  from Telescope Assembly  6  by removing 2 retaining screws  6 13     5  Remove Eye Piece Assembly  6 11  by loosening set screw  6 6  with 1 5 mm Allen wrench    6  Remove objective glued lens  6 2  from Telescope Assembly  6  by removing objective press loop 6 1   Be very careful not to  damage the lens as 6 1 is unscrewed    7  Remove Focus Lens Assembly  6 3  by loosening set screw  6 6  with 1 5mm Allen wrench and remove lens focus knob  assembly  6 4   Once removed Lens focus assembly  6 3  is removed by pushing it from the front towards the back until it    freely slides out        Prepared By  Tim Wojo Revision  1 030309 Page 3 of 7            omens al sho    CARER LEVEC   EEE  40 6900 Service Manual    1 3 Base Assembly  AP1997                                                    Item   JLT Part Description Qty  8 1 AP2017   C ring 1  8 2 AP2
3.  Using a 1 5mm Allen wrench  loosen calibration set screws 6 6 and 6 16   Rotate the eyepiece seat to make the crosshair center on the reticle of instrument on the same level with crosshair center on  the reticle of collimator tube  Then rotate the eyepiece tube to make the horizontal hair on the reticle of intrument level   Now the middle horizontal hair on the reticle of the instrument coincide with the horizontal line on the reticle of the  collimator tube     4  Tighten the two set screws 6 6 and 6 16 and restore rubber cover 3  3  to its original position          Eyepiece tube    _    Eyepiece seat    Figure G       Prepared By  Tim Wojo Revision  1 030309 Page 6 of 7    i  ir    omama ul sho  Thin ee NU    LASER LEVELS PRO    3 0 Trouble Shooting Guide    40 6900 Service Manual                            No  Symptom Cause Corrective Action  i The horizontal micro motion knob  The horizontal locking mechanism is Replace the horizontal locking  doesn   t work damaged or defective mechanism   The upper gear on the focus tube has a lot  2  The focus knob doesn   t work Replace the focus tube  of wear  3  Fail to level the bubble The bubble is broken Replace the bubble  4  The cross hair is obscure The position of the eyepiece is not correct  Adjust the position of the eyepiece  5  There is something on the reticle  The reticle is dirty Clean the reticle  Bubble is not centered in any   Recalibrate according to the above  6 The bubble level screw is loose       direction   
4. 018   Compression Spring 1  8 3 AP2019   Spacer 1  8 4 N A N A 0  8 5 AP2020   Horizontal Graduated Circle 1  8 6 AP1579   2 5M Split Washer 3  8 7 AP2021   Base Module 1  8 8 AP1449   M2 5 x 6 Cross slot Plate Screw 3  8 9 AP2022   Press Board 1   8 10 AP2023   Bottom Plate 1  8 11 AP2024   Foot Screw 3   8 12 AP2025   Bottom Module 1   8 13 AP1022   M3 x 6 Cross slot Plate Screw 3                         1  Remove Horizontal graduated circle  8 5  from Based Module  8 7  by removing C ring retainer  8 1   compression spring   8 2   and spacer  8 3  respectively  A ring pliers is a very useful tool in removing the C ring   2  Remove the 3 foot screws  8 11  from Base Module  8 7  by removing 3 split washers  8 6  located at the top each foot  screw  8 11   Once the 3 split washers  8 6  have been removed  rotate each foot screw counter clock wise to unscrew  them from the Based module  8 7    Remove Press Board  809 from Bottom plate  8 10  by removing 3 retaining screws  8 8     4  Remove the Bottom module  8 12  from the Bottom plate  8 10  by removing the 3 retaining screws  8 13         Prepared By  Tim Wojo Revision  1 030309 Page 4 of 7             omens al sho    sasen LEVES   EEE  40 6900 Service Manual    2 0 Calibration    Calibration is a process that is used to correct for accuracy and or functional errors above and beyond those stated in published  specifications  This product is fairly easy to verify and calibrate as there are only two adjustments that can be m
5. ade  Each item    discussed is shown below     2 1Vial Verification Calibration    Mount the instrument on the tripod  Turn the telescope to  the position as shown in figure A  and line up the    telescope bubble in alignment with foot screws A  amp  B           Note that the bubble moves always in the same direction             as your left thumb    Cres                 If the bubble is not centered  first grasp screws A  amp  B so       that both thumbs are moving in opposite directions        either toward each other or away from each other  in Figure A Figure B  order to make the bubble centered  Then turn the   telescope to the position as shown in figure B  turn the   foot screw C to make bubble centered  Now  the instrument c Cc    should be leveled up  but to be certain  double check  a       Rotate the instrument by any angle  and the bubble should       always be centered  Otherwise  please make calibration for Pa X       the bubble     Ec                                          If the bubble leans to the right side  adjust screws A  amp  B as rm J    w bae S  figure C shows  in order to remove one half the error  P gs B       Remove the other half of the error by adjusting the two   bubble level screws as shown in figure E  i e  loose one screw while Figure C Figure D  tightening the other screw  The instrument is calibrated when the bubble remains center as the telescope    is rotated throughout a 360   rotation  If the bubble leans to the left  Bubble level scre
6. omama al sho    CARER LEVEC   EEE  40 6900 Service Manual    7    oo  Wy        LA ER LEVELS PRO  40 6900 22x Builders Level          Service Manual                                                       Item Description Pages  1 0 Overall Instrument Assembly 2  1 1 Main Assembly 2  1 2 Telescope Assembly 3  1 3 Base Assembly 4  2 0 Calibration 5 6  2 1 Bubble Verification Calibration 5  2 2 Line of Site Verification Calibration 6  3 0 Trouble Shooting Guide 7  Prepared By  Tim Wojo Revision  1 030309 Page 1 of 7    omama al sho    LASER LEVELS PRO    1  Overall Instrument Assembly    This optical level is a highly accurate instrument  Out side of a few customer    adjustments  outlined in the owners manual   all adjustments service  operations are internal to the instrument and to be performed only by    authorized service personnel     Authorized personnel should adhere to the    guidelines described within this service manual for all repairs and or service    work  It should be note that procedures in this manual should be referred to    based on the specific situation     1 1 Main Assembly  40 6900                                                       Item   JLT Part Description Qty   1 AP1990   Rubber Cover 1 1  2 AP1991   Rubber Cover 2 1  3 AP1992   Rubber Cover 3 1  4 AP1993   Compression Spring Screw 1  5 AP1994   Horizontal Compression Spring 1  6 AP1995   Telescope Assembly 1  7 AP1996   Horizontal Compression Spring Base 1  8 AP1997   Base Assembly 1  9 AP1998   Wa
7. ve Ring 1  10 AP1999   Retainer 1  11 AP2000   M4x8 Cross Sunk Screw 1          40 6900 Service Manual          Remove Rubber 1   1   rubber cover 2   2   and rubber cover 3   3  by hand     2  Remove horizontal compression spring base   7  and horizontal compression spring   5  by loosing horizontal compression    spring screw   4      3  Remove telescope assembly   6  from Base Assembly   8  by removing Retaining screw   11   Retainer   10   and Wave    Ring   9         Prepared By  Tim Wojo    Revision  1 030309    Page 2 of 7    t ISIN    BCL easyPDF  Printer Driver    Click here to purchase a license to remove this image    Eat    omama al shoo    sasen LEVEC   EEE  40 6900 Service Manual    1 2 Telescope Assembly  AP1995                                           Item   JLT Part Description Qty   Item   JLT Part Description Qty  6 1 AP2001 Objective Press Loop 1 6 10 AP2010   M1 6x3 Cross Plate Screw 1  6 2 AP2002   Objective glued lens 1 6 11 AP2011   Eye Piece Assembly 1  6 3 AP2003   Focus Lens assembly 1 6 12   AP2012   Horizontal vernier 1  6 4 AP2004   Focus knob assembly 1 6 13 AP1449   M2 5 x 6 Cross slot Plate Screw 2  6 5 AP2005   M4 x 6 Inner Hexagon Notched 1 6 14   AP2013   Horizontal micro motion loop 1  Tightening Screw   6 6 AP2006   M3 x 6 Inner Hexagon Notched 2 6 15 AP2014   Horizontal Micro motion 1  Tightening Screw Locking Knob   6 7 AP2007   M4x12 Inner Hexagon Column head 2  6 16 N A N A 0  screw   6 8 AP2008   Vial Seat Assembly 1 6 17   AP2015  
8. w       side  turn screws A  amp  B as shown in figure D  and repeat the steps as    described above     Figure E       Prepared By  Tim Wojo Revision  1 030309 Page 5 of 7    i  Py    omens al sho    sasen LEVES   EEE  40 6900 Service Manual    2 2 Line of Site Verification Calibration    Once Vial calibration has been verify achieved  it is necessary to verify calibrate the line of site  Essentially this assures that the  stadia lines with in the telescope  cross hairs  are on the same plane as your reference  This can be done a number of ways     e By purchasing an automatic level collimator         Leveling accuracy usin  e By comparing the test instrument to a known good reference laser  g y g    an automatic level       theotolite  etc     e Using the long shot  short shot method                      SS  Regardless of the method  calibration adjustments are the same  This manual 239 o    SS            3 ACA  only discusses verification using an automatic level  ail T                          Setup the instrument as shown in figure F  level the bubble  and focus the lens                   until you can clearly see the stadia lines inside the collimator tube  If the               center horizontal hair on the stadia lines of the instrument is not coincident  with the horizontal reference line on the of the collimator tube  the instrument Figure F    should be calibrate using the method below     Remove rubber cover 3  3  to expose calibration set screws 6 6 and 6 16    2 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual de instalação e configuração    Quality Control System (QCS) - Government Home Page  Handleiding snackwerkbanken      Copyright © All rights reserved. 
   Failed to retrieve file