1. Elapsed Time Vv 7 Confirm the selection SET MUTE EXTERNAL FAN 7 4 9 Confirm the change Revision 120614 of 14 06 12 Page 11 of 21 ISO 9001 ISO 13485 E Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano L O pe p Ca Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 I 1 1 pica Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench Aspiration speed adjustment ARSPEED The fan s speed is visualized on the display In order to change it use the increase decrease keys positioned over the MENU key Fan self switching off WORKING TIMER The programmable time MAX TIME can change between 0 DISABLED default value and 24 hours In order to visualize alternately set time and elapsed time press SET MUTE key ELAPSED TIME is automatically updated by the software Normally it is automatically zeroed after the fan s self switching off but it can be zeroed by the operator by pressing at the same time the increase decrease keys placed over MENU key The count happens only if the fan is in operation Filter life LIVE FILTER Visualizes sets the life of the filter in order to effect the relative replacement The programmable time MAX TIME can change between 0 DISABLED and 999 hours The default preset time is equal to 600 hours In order to visualize altern
2. Lighting 230V 50 60Hz 1 Kw 6 3A with delayed fuses Placed at a height of 1200 mm it can be on the right or on the left depending on the position of the aspiration collector it is on the opposite side compared with the collector Not necessary Remote exit from card See paragraph 8 Fuses replacement on page 16 Not necessary The bench is provided with 200 mm collector on its upper side in order to allow the connection to the hospital s centralized canalization system The collector is placeable on request on the right or on the left N 1 Neon N 2 for models 19 500 2000 30 Watt protection IP20 placed in the upper frontal side Structural features Structure made entirely of satinized thickness 12 10 AISI 304 stainless steel with smooth surface and rounded edges in compliance with the regulation in force 6 7 mm accident prevention satinized stratified side glasses in order to avoid draughts near the working surface and contain fumes Base with doors Antistatic and antislip adjustable feet allow to get a perfect horizontal position of the working surface Noise level lt 56 dB A ISO 9001 CSQ MED ISO 13485 Revision 120614 of 14 06 12 Page 5 of 21 bE f Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano L O a d I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 www bio optica it Fax Assistenza Contabilit 39 02 2
3. Nous declarons sous notre propre responsabilit que Il dispositivo medico diagnostico in vitro the in vitro diagnostic medical device das Medizinprodukt fur die In vitro_Diagnostik AE CEIR BENON TOSO A00 le dispositif m dical de diagnostic in vitro della classe Altro of class Other der Klasse Sonstiges produrti de la classe Autre soddisfa tutte le disposizioni della direttiva 98 79 CE e successive modifiche ed integrazioni che lo riguardano meets all the provisions of the directive 98 79 EC and following amendment which apply to it allen Anforderungen der Richtlinie 98 79 EG entspricht die anwendbar sind remplit toutes les exigences de la directive 98 79 CE et modification qui lui sont applicables EN 375 Norme nazionali o armonizzate applicate EN 980 Applied harmonised standards and National standards ENISO 14971 Angewandte harmonisierte Normen nationale Normen EN 60601 1 2 Normes harmonis es et normes nationales EN 61010 1 EN 61010 2 101 Edizione in vigore alla data di emissione del documento Current ed at document date Aktuelle Ausgabe am belegdatum Edition actuelle a la date du document BIO OPTICA MILANO SPA Legale Rappresentante Carlo Sbona Revision 120614 of 14 06 12 cso Page 21 of 21 MED ISO 9001 ISO 13485
4. be used for the purpose for which it was designed that is as bench for histology for laboratory use Any other use is to be considered improper and therefore hazardous 4 The instrument must only be used by authorized and professionally qualified technician 5 The electrical safety of this instrument can be guaranteed only if it is correctly connected to an efficient earth circuit as indicated by current electrical safety regulations It is necessary to check this fundamental safety prerequisite and if in doubt ask to check the circuit The instrument is provided with a power supply cable having 2 wires ground tap that have to be connected to the power supply socket 6 Do not remove the chassis or parts of it during operation Switch off the instrument and disconnect the power supply cable before opening it This operation must to be effected only by authorized and professionally qualified technician 7 To eliminate instrument malfunctioning risks do not work near strong magnetic fields and do not use transmitters such as cellular phones near the instrument In case of serious malfunctioning switch off the instrument and contact the Technical Assistance Service 8 All waste material both infectious and radioactive deriving from the appliance working cycle must be disposed in compliance with the regulation in force This appliance is marked from this symbol in compliance with EU directive 2002 96 CE regarding electric and electron
5. p A e via San Faustino 58 e I 20134 Milano Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 Ica Opi Ha Opi Ial usar P F Ti i Dio sd User and service manual ReceivingBench 11 ELECTRIC WIRING DIAGRAMS 19 500 2000 i a it Di H 1 U OF We EH PT MOT e n IH CPA IA x BB BBEH Set persrstae JivERTER SKA D00 S ks CONTROL TEGHHIILES Zio Spiro usi l ITER IERI NELL La N FTS Co O iiy lt t rT q O lt t i ga Co O N T 2 a gt CC MED N Qu O N oO DI O O ISO 13485 ISO 9001 ci Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano 1 L O a p i I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 ja inopia Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench Milano 14 giugno 2012 12 DICHIARAZIONE DI CONFORMITA DECLARATION OF CONFORMITY KONFORMITATSERKLARUNG DECLARATION DE CONFORMITE Nome e indirizzo della ditta BIO OPTICA Milano S p A Name and address of the firm Via S Faustino 58 Name und Adresse der Firma 20134 MILANO Nom et adresse de l enterprise C F P IVA 06754140157 Dichiariamo sotto la nostra responsabilit che We declare under our sole responsibility that Wir erkl ren in alleiniger Verantwortung dass
6. via San Faustino 58 e 1 20134 Milano ZIO LCA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 a www bic optica it I Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench Description of keys and switches AIRSPEED Display Visual and acoustic alarm SET Parameters setup change Key MUTE Alarms silencing key Values change keys PN fa Increases the values De Decreases the values Key to access the functions U V lamp ignition switch this function is not active Neon switch Halogen spotlights switch this function is not active Electric frontal glass ascent descent key this function is not active Fan ignition switch ae lt z x ad n lt Green general ignition switch Emergency switch Important If this key is kept pressed accidentally the bench doesn t work io Revision 120614 of 14 06 12 or i Page 10 of 21 MED ISO 9001 ISO 13485 E 4 fi T Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano L O a p LCA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 li nn v bio optica Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench Example of parameters change EXTERNAL FAN l Press 4 Confirmn the 5 Seleer 3 Function selection 6 Choose the function Max Time
7. 6 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench Working surface features Weldings free thickness 12 10 AISI 304 stainless steel surface provided with liquids containing rim Large aspirated basin to collect incidental liquid leaks covered by two removable perforated tops three for models 19 500 2000 holes of 7 mm which serve as a supporting surface Dimensional features in mm Overall dimensions Working usable dimensions 19 500 1000 1000x750x1200 100 kg 900x650x900 19 500 2000 2000x750x1200 190 kg 1900x650x900 Revision 120614 of 14 06 12 Cso Page 6 of 21 MED ISO 9001 ISO 13485 ci Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano 1 L O a p i I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 ja inopia Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 2 INSTALLATION During the unpacking and positioning manoeuvring proceed cautiously in order to avoid possible damages Apply the two glass parts in the proper side seats In case of connection to a canalization system for fumes expulsion outside the bench must be positioned so that it is possible to carry out the discharged air 3 ELECTRIC CONNECTIONS Before connecting the instrument to the electric system using the provided cable verify the adequacy of the electric line the existence of ground tap
8. CING THE FUSES 1 If the instrument doesnt start check that the electric cable has been connected properly there is current and the two fuses under the instrument s socket aren t burnt If necessary change them with fuses of identic value SOCKET INPUT N 2 delayed fuses placed in the fuses holder socket of 6 3 A T6 3AH250V Switch off the appliance and unplug the cable from the socket Make a light pressure on the little case covering the fuses if necessary using a little screwdriver change them and close the case 2 If the green switch is on but the bench doesn t work check and if necessary replace the delayed fuse placed inside the control panel of 3 15 A 5x20 mm T3 15AH250V Revision 120614 of 14 06 12 csQ Page 16 of 21 MED ISO 9001 ISO 13485 ci Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano 1 L O a p i I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 ja opinni Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 10 CLEANING AND MAINTENANCE Cleaning A proper standard of hygiene and the routine maintenance of the laboratory are necessary to keep the instrument in good working conditions The operator is responsible for the proper cleaning of the bench 1 Use only products suitable for steel surfaces Do not use brushes with hard bristles nor abrasive sponges or detergents in order to
9. HARGE OUTSIDE THROUGH AN ASPIRATED CANALIZATION SYSTEM Important In case of non aspirated canalization the length of the draining pipe must not be higher than 3 metres the diameter must be equal to 200 mm Revision 120614 of 14 06 12 cso Page 7 of 21 MED ISO 9001 ISO 13485 oe o Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano LO pee p Ica Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 e wivw bio optica it Fax Assistenza Contabilita 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 5 REPLACEMENT OF FILTER The control panel is provided with a digital timer preset at 600 hours filters middle life When this value is reached a visual and acoustic alarm enters into action to advise the operator to change the filter For the replacement proceed in the following way 1 Switch off the instrument and unplug the cable form the socket 2 Wear individual protection devices in compliance with the regulation in force 3 Remove the frontal panel unscrewing the four Knobs unscrew the screws of the case and remove the internal anel O 4 Extract the consumed filter and prefilter and replace them with identical filter and prefilter 5 Replace the internal panel screw the screws of the case and replace the frontal panel screwing the four knobs 6 Dispose the consumed filter and prefilter in compliance with the regulation in for
10. and the compliance of the electric system with the law in force Important Do not use any extension or adapter and do not modify the provided cable 4 ASPIRATION SYSTEM The bench is provided with N 1 anti spark three phase electrofan N 2 for models 19 500 2000 adjustable through inverter preinstalled N 1 synthetic fibres prefilter preinstalled to retain possible impurities and improve the filter s efficiency N 1 filter for formalin preinstalled whose replacement happens from the lower side simply and cleanly The bench is provided with 200 mm collector on its upper side in order to allow the connection to the hospital s centralized canalization system for fumes discharge outside The electrofan allows fumes aspiration from the perforated supporting surface and the frontal grid The air is purified passing through prefilter and filter and can be discharged in this way 1 FAN PREFILTER 2 FILTERS DISCHARGE IN THE SAME ROOM Important In case of discharge in the same room it is necessary to install two filters in order to guarantee an effective removal of the polluting substances In this way it is guaranteed the filter s efficacy of 100 as the residence time Tr of the aspirated flow will be higher than 0 1 second 2 FAN PREFILTER FILTER DISCHARGE OUTSIDE THROUGH A NON ASPIRATED CANALIZATION SYSTEM draining pipe not longer than 3 metres with diameter of 200 mm 3 FAN PREFILTER FILTER DISC
11. ately set time and elapsed time updated automatically by the software press SET MUTE key When elapsed time reaches the maximum value MAX TIME a visual and acoustic alarm enters into action to advise the operator to change the filter because it is saturated In order to silence the acoustic alarm press SET MUTE key The red LED instead disappears only after zeroing ELAPSED TIME effect this operation only after changing the filter in compliance with the safety regulations Revision 120614 of 14 06 12 csQ Page 12 of 21 MED ISO 9001 ISO 13485 i _f Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano I O or LCA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 ren bio optica it Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench How ELAPSED TIME of the function LIVE FILTER is zeroed After changing the filter zero ELAPSED TIME by pressing at the same time the increase decrease keys placed over the MENU key 1 Press 3 FUNCELON 5 Select 4 Confirm the Elapsed Time selection SET MUTE CS Max Time amr Elapsed Time by pressing LIVE FILTER LIVE U V LAMP Sy Contirm the 7 Zero the value change by pressing at the same time 6 Elepsed time 999 CSQ Revision 120614 of 14 06 12 Page 13 of 21 MED ISO 9001 ISO 13485 i Bio Optica Milano S p A e via San F
12. austino 58 e I 20134 Milano L sr Ica Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 F awai bio optica htu J Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench Working with external fan EXTERNAL FAN This function is used only if the fan is not installed inside the bench but is situated outside This parameter can have only two values DISABLED default value and ENABLED The filter s life computation system does not to change in case of external fan Software features During the parameters visualization change the selected function continues working normally The parameters changes are executed by the software in real time The parameters and timers state are saved automatically on internal EPROM every hour and for every exit from the screen of a parameter by pressing the MENU key Alarms function When a determinate function reaches the maximum value MAX TIME a visual and acoustic alarm enters into action and the display shows the function in alarm WARNING WARNING LIVE FILTER lt ALARM In order to silence the acoustic alarm press SET MUTE key The red LED instead disappears only after zeroing ELAPSED TIME effect this operation only after changing the filter in compliance with the safety regulations Revision 120614 of 14 06 12 C SQ Page 14 of 21 MED ISO 13485 Yi Bio Optica Milano S p A e via Sa
13. avoid ruining the surface 2 Wear anti cut gloves over or under the common gloves to avoid possible accidents 3 The bench is made of AISI 304 stainless steel resistant against a lot of solvents and acids except for muriatic acid that might cause a rust effect 4 In case of accidental pouring of dangerous or corrosive material remove every trace of substance immediately The prolonged contact of such substances with steel might stain it and in extreme cases corrode it 5 The control panel keys are sealed by a membrane which prevent liquids or dirt from entering It is recommended however to clean periodically the panel surface with a soft and damp cloth and avoid the contact with hydrocarbons and chemical substances xylol toluol ecc which might damage it Maintenance Daily Remove possible accidental leaks of liquids on the bench Half yearly Control that the filter isn t obstruct by formalin salts or dust and that isn t consumed Effect the replacement if necessary The bench is however provided with visual and acoustic alarm in case of filters saturation See paragraph 6 Control panel Filter life on page 12 Annually Effect the following main operations Verify the wholeness of stainless steel structure power supply cable and aspiration tube Replace filter and prefilter apart form the use of the bench Inspect the possible aspiration tube Effect the electric controls inspecting control panel fan lig
14. ce In case of accidental infiltration of liquids or foreign bodies in the filter area contact the Technical Assistance Service immediately and do not use the bench Filter Prefilter CEA Mii Revision 120614 of 14 06 12 CSQ Page 8 of 21 MED ISO 9001 ISO 13485 DE j i l E i Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano LITO I p LCA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 www bio optica it Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 6 CONTROL PANEL Control panel with polycarbonate soft touch keyboard and digital display for the working parameters visualization and modification Main functions General ignition switch Emergency switch Start stop switch for aspiration working Germicidal U V lamp switch this function is not active Neon switch Adjustable halogen spotlights switch this function is not active Aspiration speed regulation Red LED visual alarm and BUZZER acoustic alarm for filters saturation Acoustic alarms silencing key Working parameters change key Electric frontal glass ascent descent key this function is not active Bench ignition and switching off programming Filters change programming Revision 120614 of 14 06 12 cso Page 9 of 21 MED ISO 9001 ISO 13485 NT Bio Optica Milano S p A
15. fi eee Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano i 0 p LCA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 if furan bio opica Fax Assistenza Contabilita 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench RECEIVING BENCH Codes 19 500 1000 19 500 2000 CE IvD TJU Revision 120614 of 14 06 12 CSQ Page 1 of 21 MED ISO 9001 ISO 13485 bz E Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano 1 L O a p I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 invi Bio copie Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench USE PRECAUTIONS Before using the instrument read carefully the instructions and warnings contained in this manual and keep it for further reference They supply important indications regarding the functions and safety for installing using and maintaining the instrument Bio Optica Milano S p A cannot be held responsible for any damage caused by improper or incorrect use and by the non observance of any of the prescription provided in this manual and by the safety regulations in force 1 After unpacking make sure that the instrument is complete and not damaged by transport 2 Before connecting the instrument to the power supply make sure that its rating corresponds to that of the power supply 3 This instrument must only
16. hting Cleaning and general working Revision 120614 of 14 06 12 cso Page 17 of 21 MED ISO 9001 ISO 13485 bz E Bio Optica Milano S p A via San Faustino 58 e I 20134 Milano 1 L O Lr p I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 wenn Bio copie Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 10 MAIN COMPONENTS E o fi om Revision 120614 of 14 06 12 oe Page 18 of 21 MED ISO 9001 ISO 13485 Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 Ica Opi Ha Opi Ial usar P F Ti i Dio sd User and service manual ReceivingBench 11 ELECTRIC WIRING DIAGRAMS 19 500 1000 IJ 30 7 o TI DEI HIS lh LT Ls zi eA EHEH EL 37 70 gt ol 12 2 gt 5 2 gt gt 5 UW MEOW EROT o e ani COMPAKHDER Ek BB BEBEE EEM ET PREEG S CTCH ibi taken from the rta mne platerei det parareter INVERTER SPA 1200075 by CONTROL TECHHIGLES Mth PIO form T GAEE Ti di metin Eplica ELECTRIC WIRE DIJAGRSM EEE BLA N TS Co O ses lt t Tr q O lt t a Co O N T 2 a gt CC MED N Un O OY O DI Ae O ISO 13485 ISO 9001 Bio Optica Milano S
17. ic appliances waste This mean that the instrument at the end of its useful life must be collected separately from other refuse The user must deliver it to the special differentiated refuse collection centres that are predisposed by the public authority 9 The contents of this manual is subject to change without further notice 10 Please find enclosed the declaration of conformity fi om Revision 120614 of 14 06 12 e Page 2 of 21 MED ISO 9001 ISO 13485 Ya i Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano ali LO LF p Ica Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 nu bio pica Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 11 Graphic symbols indicated on the label positioned near the instruments power supply socket Symbol for CATALOGUE NUMBER Symbol for SERIAL NUMBER SN 2 Symbol for ALTERNATING CURRENT FaF Symbol for FUSE j Symbol for CONSULT THE INSTRUCTIONS lil Symbol for EC MARK CE Symbol for IN VITRO DIAGNOSTIC MEDICAL DEVICE Symbol for DISPOSAL OF ELECTRIC AND ELECTRONIC EQUIPMENT A Symbol for DATE OF MANUFACTURE Symbol for MANUFACTURER Revision 120614 of 14 06 12 i CSQ Page 3 of 21 MED ISO 9001 ISO 13485 io Opfica vana io opica i i 2 Technical features Installation 3 Electric connections 4 Aspirat
18. ion system 5 Replacement of filter 6 Control panel 7 Replacing the neon lamp 8 Replacing the fuses 9 Cleaning and maintenance 10 Main components 11 Electric wiring diagram 12 Declaration of conformity ISO 9001 Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 INDEX Revision 120614 of 14 06 12 Page 4 of 21 User and service manual ReceivingBench Page 5 Page 7 Page 7 Page 7 Page 8 Page 9 Page 15 Page 16 Page 17 Page 18 Page 19 Page 21 cso MED ISO 13485 1 13485 vana io Opi Ial Bio Optica Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 1 TECHNICAL FEATURES Aspirated bench designed to hold and store disposable containers during reporting of histological samples preserved in formalin Provided with fumes aspiration system from the supporting surface and the frontal side predisposed for fumes discharge outside Electrical connections Power supply Rated output Nominal current Power supply socket Emergency power supply Remote exit Fuses Other connections Water connections Fumes discharge
19. n Faustino 58 e I 20134 Milano Pap LO SF p I CA Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 ja opinni Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 7 REPLACING THE NEON LAMP 1 Switch off the instrument and unplug the cable form the socket 2 Unthread the two plastic blocks placed at the ends of the lamp holder by pressing the lever and at the same time by pushing outwards the block Picture A 3 Remove the plastic transparent protection and extract the lamp rotating it of 90 clockwise or anticlockwise Picture B 4 Insert the new lamp and block it rotating of 90 Verify the ignition by pressing the switch on the control panel In case of non ignition check that the switch situated on the back side of the lamp is on 5 Replace the plastic protection and the two blocks by pushing them inwards Picture C 6 Dispose the lamp in compliance with the regulation in force Picture A Picture B Picture C m Cs Revision 120614 of 14 06 12 CSQ i Page 15 of 21 MED ISO 9001 ISO 13485 k i F x E l Bio Optica Milano S p A e via San Faustino 58 e I 20134 Milano L O pe p Ca Tel 39 02 21 27 13 1 e Fax Acquisti Export 39 02 21 54 155 _ 1 1 optical Fax Assistenza Contabilit 39 02 26 41 74 48 e Fax Vendite 39 02 21 53 000 User and service manual ReceivingBench 8 REPLA
Download Pdf Manuals
Related Search
Related Contents
design makes objects timeless, fashion corrupts Axis P1354 TX - MEDIA TV Video Systeme Kramer Electronics 621R Icom iF60 User's Manual 世界一。 - mono shop VA2465SH-1, Multiple Models User Guide, Spanish/Español 20ー2年ーー月~ー2月大坂 消費者庁後援 独立行政法人 製品 MasterMatrix 885 Instruction Manual for DOHC 4-VALVE HEAD PISTON / CYLINDER Copyright © All rights reserved.