Home
Eclipse 5™ with autoSAT®
Contents
1. gt _ Figure 7 Eclipse 5 Power Cartridge The Power Cartridge Battery may be charged in two ways a place the Power Cartridge in the Eclipse 5 and connect the Eclipse 5 to AC or DC power or b place the Power Cartridge in the optional Desktop Charger PN 7112 SEQ Operating and servicing the Power Cartridge will ensure longer life and higher performance It is important to follow the tips and recommendations when storing and operating the Eclipse 5 on the Power Cartridge PN 20631679 RevD 21 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT The Eclipse 5 Power Cartridge operation time may be affected by several factors such as bolus size breathing rates ambient temperature age of power cartridge and use over time The table below describes the typical operating time for a new Power Cartridge If the Eclipse 5 is used in Pulse Dose Mode there will be longer operating time The patient should consult their physician for a prescription for Pulse Dose Mode TYPICAL NEW POWER CARTRIDGE OPERATING TIMES At nominal temperature P Battery Duration eae neil pratt of 25 C 77 F 0 5 LPM 44 hours 1 0 LPM 3 7 hours 16mL 1 0 5 4 hours 2 0 LPM 2 0 hours 32mL 2 0 5 1 hours 3 0 LPM 1 3 hours 48mL 3 0 4 9 hours 64mL 4 0 4 0 hours 80mL 5 0 3 7 hours 96mL 6 0 3 5 hours 128mL 7 2 5 hours 160mL_ 8 2 0 hours 192mL 9 1 7 hours Battery Consumption Char
2. Replace HEPA filter CHECKING AND REPLACING THE POWER CARTRIDGE Typically the Power Cartridge will require no routine maintenance beyond cleaning by the patient and calibration CAIRE recommends that the Power Cartridge be replaced if there is physical damage to the cartridge enclosure or connector or if the battery s recharge capacity after calibration is 20 less than published values at its given flow rate ECLIPSE 5 MONTHLY RUN TIME PROCEDURE 1 Power on Eclipse 5 using AC Power DC Power or Power Cartridge 2 Allow unit to run for a minimum of 2 hours This procedure is performed to ensure and prolong the life of your Eclipse 5 This is also an ample opportunity to calibrate the power cartridge of the Eclipse 5 PN 20631679 RevD 38 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WIH autoSAT ANNUAL MAINTENANCE PROCEDURES The following section lists procedures that are necessary to maintain the Eclipse 5 Service should only be performed by a qualified technician To perform periodic maintenance the only tools that should be necessary are 1 Phillips Screwdriver Wire cutting pliers Small cable ties ESD Mat or approved ESD system Figure 11 Maintenance Tools REMOVING THE UNIT COVER 1 Disconnect power supplies and remove Power Cartridge before removing the unit cover 2 Place the unit horizontally on the front cover 3 Using a Phillips 1 screwdriver remove a total of
3. The Eclipse 5 should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary the Eclipse 5 should be observed to verify normal operation in the configuration in which it will be used Guidance and Manufacturer s Declaration electromagnetic emissions The Eclipse 5 is intended for use in the electromagnetic environment specified below The customer or the user of the Eclipse 5 should assure that it is used in such an environment Emissions Test Compliance Electromagnetic environment guidance RE emissione The Eclipse 5 uses RF energy only for its internal function Group 1 Therefore its RF emissions are very low and are not likely to EN 55011 cause any interference in nearby electronic equipment RF emissions Class B EN 55011 The Eclipse 5 is suitable for use in all establishments including Harmonic emissions IEC Class A domestic establishments and those directly connected to 61000 3 2 the public low voltage power supply network that supplies Voltage fluctuations buildings used for domestic purposes flicker emissions IEC Complies 61000 3 3 PN 20631679RevD 9 C eclipse autoSAT Personal Ambulatory Oxygen System Provider Technical Manual Guidance and manufacturer s declaration electromagnetic immunity The Eclipse 5 is intended for use in the electromagnetic environment specified below The customer or the user of the Eclipse 5 shoul
4. autoSAT POWER CARTRIDGE BATTERY CONSERVATION FEATURE While in Pulse Dose Mode the Eclipse 5 is always monitoring for breath detection After 15 seconds of no breath detected the Eclipse 5 delivers Continuous Flow at the last Continuous Flow setting The system and display are still in Pulse Dose Mode and the green Delivery Mode Indicator is blinking fast indicating you are receiving a Continuous Flow After 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath If a breath is detected then the Eclipse 5 will go back to Pulse Dose Mode If a breath is not detected the Battery Conservation function will continue 15 seconds Continuous Flow 15 seconds Pulse Dose Mode The Eclipse 5 will stay in this modality until a breath is detected This power management feature of the Eclipse 5 greatly conserves the operating life of the battery extending therapy time and patient mobility ADJUSTING THE PULSE DOSE MODE SENSITIVITY SETTING The oxygen delivery trigger sensitivity may be adjusted to satisfy patient requirements by selecting setting 1 2 or 3 1 most sensitive 3 least sensitive The default setting for units coming from CAIRE Inc is 2 To adjust the Pulse Sensitivity perform the following steps 1 Have the patient sit quietly with the nasal cannula properly fitted 2 Put the Eclipse 5 into Pulse Dose Mode at the patient s prescribed flow setting 3 Press the No Smoki
5. Fig 48 Cable locations PN 20631679 Rev D 74 echip Personal Ambulatory Oxygen System Provider Technical Manual Ce Se autoSA 1 4 Unscrew the 2 Cooling Fan screws PN 6961 21 SEQ holding the fan in place Remove the Fan PN 1074 SEQ Set aside for later installation 1 5 Remove the Exhaust tube in Figure 49 Set aside for later installation Fig 49 Exhaust Tube 1 6 Pull the Power Manager pcb 1 out of the case Gently push the compressor box away from the Power Manager pcb about 1 8 Slide the board and the cabling out from under the edge of the compressor for better access Disconnect the two wire harnesses 9V batt amp ATF harness and the Control board ribbon cable shown in Fig 50 Remove the Power Manager pcb T a we 9V Battery Harness G l a gt Control Board Ribbon Cable G s eS ATF Gearmotor Harness Fig 50 Wire harnesses and control board ribbon PN 20631679 Rev D 75 1 7 Unbox the SP5932 4 SEQ and account for all parts Fig 51 Fig 51 New Power Manager 1 8 Position the Power Manager PCB PN SP5932 4 SEQ 1 out of the case as shown in Fig 52 Gently push the Compressor Box away from the Power Manager pcb by 1 8 Connect the 2 wire harnesses Ext Power amp ATF harness and the Control Board Ribbon Cable Note Route the wiring under the edge of the compressor Position the Power Manager PCB into the alignment slots on the back case Finall
6. 1 0 Pulse Dose Setting 45 Watts 6 0 Pulse Dose Setting 95 Watts Battery Charging 50 Watts Nominal Power Cartridge operating time Continuous Flow 2 0 hours 2 0 LPM 1 3 hours 3 0 LPM Note Times will decrease with higher bolus size Pulse Dose 12 BPM breath rate ambient temperature power cartridge Setting 2 0 5 2 hours age and use over time Setting 6 0 3 5 hours Continuous Flow Indication Expressed in liters per minute LPM Loss of Power Hot Power Cartridge Low Power Cartridge Warm Power Cartridge Low Oxygen Output O Flow Outside Normal Limits Audible Alarm Indicators Unit Malfunction e Low 9V battery Back Up Alarm Power 9V Internal Battery Air Inlet HEPA Compressor Intake Device Classification IEC Class Type B Applied Part IPXO PN 20631679RevD 6 CLL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITis autoSAT Pulse Dose Mode Specifications Pulse Settings 1 0 to 6 0 in 8mL increments up to 96 mL 7 9 in 32mL increments up to 192 mL Trigger Sensitivity Adjustable between settings of 1 most sensitive to 3 least sensitive Adjustable Bolus Rise Time Adjustable settings of Fast factory setting Medium or Slow Cannula pressure has dropped below the trigger point typically between 0 135 0 37 Trigger Criteria cm of H O of negative pressure At least 1 4 seconds has passed since the last pulse began Minimum time between 1 25 seconds max 3
7. Disconnect the braided tubes 5a Install port caps 6 Unscrew the 3 screws M4x16 Pan Head Machine Screw P N 6961 416 SEQ and remove the screws and washers M4 Flat washer P N 6985 04 SEQ as shown below Figure 33 Remove ATF Module from Eclipse 5 Case 7 Lift the ATF Take off the 3 grommets Disconnect the wire harness from the gearbox of the ATF 8 Remove the ATF 9 Install 3 grommets into the mounting slots on the ATF as shown below Position the ATF into the bottom cover but DO NOT install the mounting screws yet Route the 16 Silicone Tube under the ATF as shown on Figure 35 on next page Figure 34 ATF Module Grommet Installation PN 20631679 RevD 64 C E F f p SA4L Personal Ambulatory Oxygen System Provider Technical Manual autoSAT 16 Silicone Tube Routing Figure 35 Tube Routing under ATF Module 10 Place the ATF in the Unit Case and remove the caps from the ATF Ports Install the Braided Tubes into the ATF pressure and vacuum ports as shown on Figure 32 Do not use oil or grease if the tubing is difficult to install Secure joints with cable ties WARNING DO NOT use any petroleum based or other lubricants A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source NOTE Inspect all tubing before
8. Wing Nut 3 Unscrew the Clear HEPA filter and discard filter amp small o ring Figure 21 Unscrewing the HEPA Filter 4 Install the new small o ring and HEPA filter Figure 22 Installing the new HEPA Filter PN 20631679 RevD 45 E F f p SA4L Personal Ambulatory Oxygen System Provider Technical Manual autoSAT 5 Locate the large o ring on the product tank replace it with the new large o Ring included in the PM kit 5022 SEQ Screw the new HEPA filter into the carriage holder wing nut topped housing Insert and screw the new HEPA filter wing nut topped housing into the carriage holder product tank and finger snug in place do not over tighten Re attach tubing and secure with a cable tie Figure 23 O Ring REMOVE AND REPLACE THE COMPRESSOR INTAKE FILTER Replace the compressor intake filter annually or more often as needed The compressor intake filter may become clogged depending upon the amount of contaminants in the air smoke dust dirt pollen etc and may need to be changed more frequently The most common cause of low concentration and eventual system failure is a dirty or clogged compressor inlet filter Environmental conditions usually determine the effective life of a filter CAUTION Operating the Eclipse 5 with a clogged compressor intake filter may ES reduce performance and lead to system damage or premature failure NOTE Always cut the heads of cable ties to avoid damaging t
9. 120 16 2 0 9 213 32 3 0 10 288 48 4 0 11 349 64 5 0 12 400 80 6 0 13 443 96 4 15 512 128 17 565 160 19 606 192 Pulse Profiles The fundamental approach to triggering and controlling the oxygen bolus in Pulse Dose Mode is as follows The User may select a pulse trigger sensitivity in settings ranging from 1 3 Bolus volume should be prescribed by a physician and may range from 16mL up to 192mL The pulse will be triggered when the system meets all of the following criteria The cannula pressure has dropped below the trigger point typically between 0 135 and 0 37 cm H 0 At least 1 seconds has passed since the last pulse began PN 20631679 Rev D 17 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT PROVIDER MODE FUNCTIONS Providers can access provider mode software functions using the Control Panel All provider mode information is displayed on the LCD The software shall advance the following Eclipse 5 display mode when the No Smoking icon is pressed Alarm Code ALRM Displays most recent alarm code Additional alarm codes will also be displayed if present by continuing to press the key Pulse Sensitivity PS Hours of Operation HRS 9 volt Battery Status 9V Rise Time BOL Control Board Software Revision CS Power Manager Software Revision PM ALARM CODE While in Continuous Flow Mode pressing the No Smoking icon displays the cur
10. Cap the three ATF module ports using tight fitting vinyl caps or vinyl electrical tape CAUTION Once the braided hoses are disconnected from the ATF Module the ATF Module is exposed to the atmosphere Cap the ATF Module ports immediately in order to minimize exposure to debris and humidity which may result in ATF Module damage To ATF Module Vacuum Port To ATF Module Pressure Port Figure 58 Removal of Compressor Pressure and Vacuum Hoses from ATF Module PN 20631679 Rev D 80 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual aUutoSAT 6 Turn the unit over onto a padded ESD safe surface and remove the 4 screws M4x16 Pan Head Machine Screw PN 6961 420 SEQ and washers M4 Flat Fender Washer PN 3568 SEQ shown in Figure 59 Screws Washers 4 Places Figure 59 Remove Compressor Box screws 7 Remove the Compressor Box There are no field serviceable parts inside the Compressor Box Do not attempt to disassemble or modify the Compressor Box in the field 8 To install the Compressor Box lay the new Compressor Box in the unit Turn over the unit 9 Position the Compressor Box into the Bottom Case with the compressor box screw holes aligned with the 4 brass eyelets Screw in the 4 screws and 4 washers as shown in Figure 59 Apply small amount of Loctite 425 on the threaded tips of the screw before installing 10 Install the braided tubes into the ATF pressure and vacuum ports
11. Cartridge Exhaust Vent Battery Power Cartridge Latch BACK PN 20631679RevD 5 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual a ENT ECLIPSE 5 OXYGEN SYSTEM SPECIFICATIONS Oxygen Concentrator 19 3 x 12 3 x 7 1 inches DIMENGE 49 0cm x 31 2cm x 18 0cm Weight Eclipse 5 15 0 pounds Power Cartridge 3 4 pounds Flow Settings 0 5 to 3 0 LPM 0 5 liter increments Continuous Flow measured in Liters Per Minute LPM Pulse Dose measured in mL Settings 1 0 6 0 16 96mL 8mL increments Settings 7 9 128mL 160mL 192mL Continuous Flow Accuracy 10 or 200ml min whichever is greater Oxygen Concentration 87 95 6 for all flow settings Maximum System Pressure 15 psig 103 5 kPa Oxygen Output Pressure 5 0 psig 34 5 kPa nominal Green Light Normal Operation Yellow Light Warning or Caution less than 85 3 Red Light Flashing Abnormal Operation less than 70 5 Oxygen Concentration Status Indicator Nominal Sound Level 3 0 LPM Continuous Flow 48 dB A 3 0 Pulse Dose Mode 40 dB A Operating Environment Temperature 50 F to 104 F 10 to 40 C Humidity 10 to 95 Non condensing 82 4 F 28 C Maximum Dew point Storage Environment Temperature 4 F to 140 F 20 to 60 C Humidity Up to 95 Non condensing Altitude 0 13 123 feet 0 4 000 meters Nominal Power 0 5 LPM Continuous flow 48 Watts 3 LPM Continuous Flow 145 Watts
12. Manager software using the Control Panel Refer to the Level 1 Maintenance section for further information Example CB394111 Control Board PCB PN 3941 with Rev 1 1 software PM593211 Power Manager PCB PN 5932 with Rev 1 1 software PN 20631679 RevD 18 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT SERVICE MODE FUNCTIONS Factory maintenance or service updates may sometimes be required on the Eclipse 5 Factory and qualified factory trained technicians can access service mode software functions by using the Service Port located on the back of the unit The Service Port is not for patient use Figure 4 Eclipse 5 Service Port ECLIPSE 5 DATA ACQUISITION TOOL EDAT EDAT is the world s first computer based data retrieval service tool for oxygen concentrators EDAT is proprietary to CAIRE Inc and is comprised of both hardware and software components EDAT is CAIRE Inc s global solution for your service and support needs EDAT software is capable of sensing and displaying historical event logs and values of components and sensors within the Eclipse 5 system This can be used to determine system faults or user errors and communicate data for troubleshooting or documentation related to service and updates EDAT hardware attaches or Plugs into the Eclipse 5 Service Port and into a computer s USB port The EDAT software set allows a provider to set up a hub and spoke service center
13. Product Tube Assy to the ATF output barb secure with a cable tie Attach the other end of the Product Tube to the Bottom port on the Product Tank See Figure 39 5 Connect the Output tubing to the Top port on the Product Tank and secure with cable ties See Figure 41 Add cable tie to both top and bottom parts Figure 41 Connect output tubing CC 4 p SA Personal Ambulatory Oxygen System Provider Technical Manual autoSAT Electronics REMOVE AND REPLACE THE CONTROL BOARD ASSEMBLY The Eclipse 5 contains electrostatic sensitive components Do not open or handle except at a static free workstation 1 Remove the Unit Cover as described in the section Remove and Replace the Unit Cover Air Inlet Power Temperature p Manager PC Board Z Proportional Valve Connector Figure 42 Control Board Removal ATF Module not shown for clarity 2 Remove the 2 screws M3x10 Plastite Screw P N 6950 310 SEQ holding the Control Board Assembly to the Unit Case as shown in Figure 42 Use a magnetized Phillips 1 screwdriver to remove the screws This greatly reduces the chances of dropping a screw in the Unit Case 3 Lift the Control Board Assembly and disconnect the Inlet Air Temperature Data Communications Port Power Manager PC Board Communications Proportional Valve wire harness and Buzzer wire harnesses as shown in Figure 42 PN 20631679 Rev D 70 E F f ps Personal Ambulatory Oxygen System Prov
14. ae enna eae pee nee err arr rae 26 Locating the Eclipse 5 for Proper Use and Ventilla UOM ess eR E EAEE 27 Whe Users Ma Udlisssremesicrissiemeniiiueniinienen 27 Showing Panel Buttons Indicators Alerts GING Ali S ca ois eaten eanaess 27 Showing Power Cartridge Power Level s s s 28 Selecting the Proper Flow Mode and Setting 28 Selecting Continuous Flow Mode Wusssssssesssseeseensseeens 28 Selecting Pulse Dose Mode ccessssssssssssssesesssssessesnssseee 28 Power Cartridge Battery Conservation FOU ee a E E ene eens 29 Adjusting the Pulse Mode Sensitivity Setting 29 Adjusting Rise TIME ncsictcacacuniancmcneekeoannats 30 Connecting the AC Power Supply cssssssseesssssssessssseens 30 Connecting the DC Power SUPDIY wssssssesssseceessseeens 31 Active Lifestyle Training ssssssssssssssesssssesesssssessssssssessssssseens 31 Attach the Universal Cart neessessssssesessssssscsessssssssessssssees 32 Using Around the HOUSE sssssssssssssecssssnssssssecsesssssnsee 32 Traveling By Vehicle ccitssancasminmiinsiaataninimancian 32 Traveling by All caisawatsecinnmiemniaiamemasien 34 Traveling by Cruise Ship sceciacisrtaneccc cantante 34 Traveling by Tra Mieresccsnemussismnanaminisioas 34 Eclipse 5 MainteMannce seessssscsessssssessssssssssssssssesesnsssssssenses 35 Weekly Maintenance Patient 35 Clean the Air Inlet Filter sssri 35 Clean and Care for the Tubing and Cannula 35 Clean the Cabinet and C
15. as operation time flow settings and environmental conditions Service is required when oxygen purity and or flow rates cannot be maintained REMOVE AND REPLACE THE COMPRESSOR BOX There are no field serviceable parts inside the Compressor Box Do not attempt to disassemble or modify the Compressor Box in the field 1 Remove the Unit Cover as described in the section Remove and Replace the Unit Cover The Eclipse 5 contains electrostatic sensitive components Do not open or handle except at a static free workstation 2 Flip unit over onto a padded ESD safe surface Using a small flat tip screwdriver to pry open the cover at each pry point remove the Case Bottom Cover as shown in Figure 55 Case Bottom Cover Figure 55 Case Bottom Cover Removal PN 20631679 Rev D 78 C SC ecl OSAT 3 Disconnect the 3 wire harnesses shown in Figure 56 Compressor Signal Wire Harness Personal Ambulatory Oxygen System Provider Technical Manual i NI T ionn F Paa 4 Blower Wire gt 1 Harness Figure 56 Compressor Box electrical connections 4 Remove the Exhaust Tube as shown in Figure 57 Verify Enga During Installation Figure 57 Exhaust Tube Removal PN 20631679 Rev D 79 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT 5 Cut the three cable ties shown in Figure 58 Disconnect the two braided hoses from the pressure and vacuum ports of the ATF
16. as shown on Figure 58 on previous page Secure joints with cable ties 11 Use a cable tie to secure the silicone tube to the braid tube to ensure that there is NO kinking on the bend as shown on Figure 58 on previous page Hand tighten the cable tie Inspect the tubing to ensure the tubing is not kinked 12 Hold onto the Compressor Box and lift the Bottom Case into a standing position Insert the Exhaust Tube into the slot of the Exhaust Duct as shown on Figure 57 Ensure that it is secure without any gaps 13 Plug the 3 harnesses into the Power Manager PCB as shown on Figure 56 14 Install the Case Bottom Cover as shown in Figure 55 Install the Unit Cover as described in the section Remove and Replace the Unit Cover as outlined on page 39 Nn Oo A W EDAT may be utilized to record Eclipse 5 device status and dates of service PN 20631679 RevD 81 E F f p SAV Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT MAINTENANCE AND REPLACEMENT PARTS Preventative Maintenance Parts Replacement Parts List Description CAIRE Ine Part Number Description SCARE Inc PartNumber 8076 SEQ DC Power Suppl 5942 SEOQ Foot bottom 6956 SEQ Handle 6963 SEQ Inlet Filter 7028 SEQ Membrane Panel Overlay Eclipse 5 SP20618461 Nut Outlet 2777 SEQ Outlet Assembly Oxygen 3922 SEQ Power Manager PCBA Eclipse 5 Not Field Replaceable SP5932 4 SEQ Power Cartridge 7082 SEQ Product Tank Assembl SP4
17. cartridge and external AC or DC power supply 2 Re attach the Eclipse 5 to external power or install a fully charged power cartridge in the Eclipse 5 Allow warm power cartridge to cool outside the concentrator for 30 minutes 3 Re charge the power cartridge using AC or DC power 4 Replace the power cartridge if another is available 5 If the condition persists contact Chart Technical Support for service 1 Remove the power cartridge and disconnect the AC or DC power supply from the Eclipse 5 The red light should disappear and the FAIL message will disappear from the screen when the power sources are removed Wait approximately 20 seconds reconnect power sources and attempt to power on the Eclipse 5 again 2 If the Eclipse 5 does power on monitor it to determine if the FAIL message occurs again If the FAIL message occurs again contact Chart technical support for service 3 If the Eclipse 5 does not power back on advise patient switch to an alternate source of oxygen and contact Chart Technical Support for service Other Alarm Conditions The Eclipse 5 is unable to detect the user s breathing effort See page 17 for pulse mode information 1 Pulse mode may not work correctly if the user has any condition that causes blocked re stricted nasal passages Illness deviated septum etc Switch to continuous flow 2 Ensure cannula is no longer than 7 feet in length 3 Ensure that there are no kinks or blockage in cannula tu
18. consecutive breaths breaths While in Pulse Dose Mode the Eclipse 5 is always monitoring for breath detection After 15 seconds of no breath detected the Eclipse 5 delivers Continuous Flow at the Response to Missing Breaths last Continuous Flow setting After another 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath The Eclipse 5 will stay in this modality until a breath is detected See Power Cartridge Conservation Feature page 26 AC Power Supply and Pulse Dose Bolus Size Power Cartridge Battery DE Power Supply Setting 15 mL Max Breath Max Breath z Rate Rate A 1 0 16 40 40 T 2 0 32 40 40 es 3 0 48 40 40 4 0 64 40 31 gt 5 0 80 37 25 2 6 0 96 31 20 v 7 128 23 15 3 8 160 18 12 Q 9 192 15 10 NOTE Bolus volume decreases as breath rate exceeds published range Power Accessory Specifications AC Power Supply DC Power Supply Input Voltage 100 240VAC 50 60 Hz Input Voltage 11 5 16VDC Input Power 245 260 VA Output Voltage 26 VDC Output Voltage 28 VDC Output Power 150W Max Output Power 200W Power Cartridge Battery Output voltage 14 8 VDC f Quantity 2 97 5 W hrs batteries Capacity me Each containing 7 92 grams equivalent Lithium content Nominal Power Cartridge Life 80 Capacity after 500 Charge Discharge cycles 1 8 to 5 0 hours dependent on flow set
19. securing with a cable tie 11 Mount the ATF with 3 M4x16 screws and 3 washers as shown on Figure 33 on previous page Apply Loctite 425 Thread Locker on the threaded tips of the screws before installing Tighten screws but DO NOT squash the grommets 2 in lbs maximum If needed install spacer 20623336 underneath the ATF mount closest to the product tank with a longer screw 6961 418 SEQ to ensure that the ATF is level NOTE Over tightening may cause excess vibration CAUTION Remove caps on ATF ports ONLY before connecting the tubes DO NOT leave the ports open unless ready to install the tubes Also place the removed caps from the new ATF and place on the ports of the replaced ATF PN 20631679 Rev D 65 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT 12 Remove caps from ATF Module 13 Connect the silicone tube from the Product Tank into the product port of the ATF and secure joints with cable ties as shown on Figure 31 14 Use cable tie to secure the silicone tube to the braid tube to ensure that there is NO kinking on the bend as shown on Figure 32 Hand tighten the cable tie Inspect the tubing to ensure the tubing is not kinked 15 Perform product test prior to installing cover and again after cover replacement 16 Installation of the Unit Cover on the unit is described in the Remove and Replace of the Unit Cover PN 20631679 RevD 66 Removal of Product Ta
20. supporting ribs tighten the 2 screws to the unit case Lift Bracket 1 8 above Ribs Figure 43 Control Board Installation 9 Connect the Inlet Air Temperature Data Communications Port Power Manager PC Board Communications Proportional Valve wire harness and Buzzer wire harnesses as shown on Figure 42 10 Refer to Flow Calibration in EDAT User Manual PN 5419 SEQ to calibrate Eclipse 5 with new Control Board 11 Installation of the unit cover on the unit is described in the Remove and Replace of the Unit Cover PN 20631679 Rev D 7 1 C CC f p SA Personal Ambulatory Oxygen System Provider Technical Manual autoSAT REMOVE AND REPLACE THE BUZZER WIRE HARNESS 1 Unplug the buzzer from the Control Board PC Board 2 Remove the Control Board 3 To remove the buzzer from the case cut the cable tie grasp the buzzer body with a pair of pliers and rotate the buzzer to break the adhesive joint 4 To install a new buzzer apply a ring of cyanoacrylate adhesive Super Glue around the inlet hole as shown in Figure 44 Apply Yer Adhesive Figure 44 Bonding the Buzzer to the case 5 Place buzzer as shown in Figure 44 Ensure that the buzzer opening is centered over the grill hole Hold in place until adhesive cures Figure 45 Buzzer Installation 6 Install the Control Board Assembly Secure the wire harness with Cable as shown on Figure 45 Insure that there is a slack on the wire harne
21. the battery out of the compartment located at the bottom right corner of the unit Battery Compartment Figure 17 Removing the 9 volt Battery 3 Disconnect the 9 volt battery harness 4 Remove and replace the 9 volt battery Figure 18 Disconnecting the 9 volt Battery 5 Reconnect the battery harness Ensure that the terminals are aligned correctly 6 Place the battery back into the compartment 7 Install battery foam pad if necessary PN 20631679 Rev D 43 C E F f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual autoSAT REMOVE AND REPLACE HEPA FILTER Replace the HEPA filter annually or more often as needed To replace the HEPA filter follow these steps WARNING DO NOT use any petroleum based or other lubricants A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source NOTE Always cut the heads of cable ties to avoid damaging the tubing Internal Filters HEPA Filter replacement 1 Cut cable tie and Disconnect the tubing at the top of the filter carriage Figure 19 Tubing on Filter Carriage PN 20631679 Rev D 44 CLL f p SA4L Personal Ambulatory Oxygen System Provider Technical Manual autoSAT 2 Unscrew the wing nut that holds the HEPA filter Figure 20
22. to the Eclipse 5 Personal Ambulatory Oxygen System with autoSAT Technology Setting up and training your patient to use the Eclipse 5 has never been easier You can expect your patients and care providers to easily learn how to use the device by following the directions in this section While setting up and training a patient be sure to point out the advantages of the Eclipse 5 For example m Slim and sleek appearance m Easy to use controls m Quiet operation m Lower electric bills m Self monitoring alarm system More consistent FiO at higher breath rates After completing each training procedure ask your patient if he or she has any questions Proper training of your patients will result in fewer service calls improved compliance and increased patient satisfaction PRE DELIVERY CHECK LIST Before delivering the device check and log the status of the following m Parts Inventory Verify that each Eclipse 5 is provided to the patient with the following items v Users Manual v AC Power Cord v Quick Start Guide v AC Power Supply v Eclipse 5 Passport v DC Power Supply v Universal Cart lt Power Cartridge Battery v Cannula v Air Inlet Filter and Spare E Power Sources Insert and check the following for proper operation v AC Power Supply with Power Cord v DC Power Supply v Power Cartridge Battery E Electrical System Use the Control Panel to check and or adjust the following v Pulse Dose Setting v 9V Batter
23. 10 screws M3x12 Pan Head SEM Screw P N 6974 312 SEQ from the back of the unit WARNING Keep hands out of moving parts Disconnect power supplies and remove Power Cartridge before AN removing the unit cover ESD Safety procedure must be in place PN 20631679 Rev D 39 E F f ps Personal Ambulatory Oxygen System Provider Technical Manual autoS INE CAUTION The Eclipse 5 contains electrostatic sensitive components Do not open or handle except at a static free workstation Do not remove cover without ESD protection Figure 12 Removing screws to open the Front Cover 4 Turn over and place the unit horizontally on the back cover 5 Remove the front cover pop the bottom end off first and rotate it towards the handle Be careful not to tear the Control Panel ribbon cable that is attached to the front cover Rotate front cover up towards handle Figure 13 Opening the Front Cover PN 20631679 Rev D 40 CLL f p SA4V Personal Ambulatory Oxygen System Provider Technical Manual WITis autoSAT 6 Disconnect the cable from the Control Panel as follows Grasp the circuit board firmly between your fore fingers and thumb Grasp the head of the Membrane Panel Overlay cable in your other hand c Firmly pull the cable away from the board Pull the cable straight out Do not rock the connector back and forth This may damage the pins of the header Oo ow Do not disconnect the Membrane Panel O
24. 378 SEQ Rubber Gasket Channel Seal 4106 2 SEQ Screw M2x10 Pan Head Machine Cooling fan 6961 210 SEQ Screw M3x10 Plastite CS pcb Product Tank 9V Batt harness 6950 310 SEO Screw M3x12 Pan Head SEM Top to Bottom case BBB PCB 6974 312 SEQ Screw M4x16 Pan Head Machine ATF to Bottom case 6961 416 SEO Screw M4x16 Pan Head Machine Compressor to Bottom Case 6961 420 SEQ Tubing Silicone 3 16 ID SP6981 SEO Tubing Silicone 2mm ID SP4101 SEQ Tubing Braided 5 16 ID SP3534 SEO Washer M3 Flat 6985 03 SEO Washer M4 Flat 6985 04 SEO Washer M4 Flat Fender 3568 SEQ Wire harness 9V 1076 SEQ Wire Harness Service Port 7012 SEQ Wire Harness External Power Eclipse 5 4063 SEO Kit Preventative Maintenance 5022 SEQ PN 20631679 RevD 82 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT OPTIONAL ACCESSORIES Visit us at www sequal com for more information about optional accessories There are many different types of oxygen tubing cannula and humidifiers The following items are recommended by CAIRE Inc for use with the Eclipse 5 Salter Labs Humidifier Part Number 7600 or equivalent f your physician has prescribed an optional humidifier follow the manufacturer s instructions for use Attach the humidifier to the oxygen outlet port of the Eclipse 5 Use of optional humidifiers not recommended for the Eclipse 5 may impair performance of the device and may void the warranty
25. 5022 1998 A1 2001 A2 2003 EN61000 6 1 2001 EN61000 4 Compatibility 2 1995 A1 1998 EN61000 4 3 2002 Any CSA CUS mark for the Eclipse 5 system does not encompass operation with the DC Power Supply Madel 5942 PROVIDER SUPPORT POLICY Safety Objective As a manufacturer our organizational goal is to provide customer support and assistance to the highest level of excellence Customers are Providers which include Dealers Distributors and Agents Support includes but is not limited to troubleshooting and Return Material Authorizations RMA Business Hours are Monday Friday 8 00am 5 30pm EST CAIRE Inc can only support customers who are recognized as Providers Dealers Distributors and or Agents These partner ships are qualified as having an existing account or are in the process of credit application completion All patient or end user inquiries including but not limited to RMA warranty or serial number questions must be handled by their Provider Provider Support Policy CAIRE Inc is unable to provide direct assistance clinical advice or recommendations to a patient or end user Providers have sole responsibility in assisting their patients PN 20631679RevD 8 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT ELECTROMAGNETIC COMPATIBILITY This equipment has been tested and found to comply with the limits for medical devices to the IEC60601 1 2 Electromagneti
26. CLL f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT ALARM CONDITIONS AND ALARM CODES Use the table below to decode Eclipse 5 alarm conditions If other alarm codes are displayed by the Eclipse 5 contact Chart Technical Support for assistance Note The following table is intended as a guide for the provider not the user ALARM GREEN YELLOW RED AUDIBLE DISPLAY INDICATOR INDICATOR INDICATOR ALARM Warming Up Please Wait Warming Up Please Wait 3 0 or flow setting we wom Low Power Cartridge 010 Warm Power Cartridge 002 Low 9V Battery po fe Oz Concentration lt 85 008 Oz Concentration lt 70 Flow Rate Error Blocked FLOWRATE OFF Flashing Flow CONDITION WHAT TO DO Wait The system typically takes 3 5 minutes to reach id specified performance Wait The system typically takes 3 5 minutes to reach specified performance Nothing The system is operating properly at the specified flow rate oS 7a nm Flashing 2 Beeps Plug into external power or replace with a charged Power Cartridge Flashing 1 Beep Plug into external power or replace with a charged Power Cartridge 3 Beeps on POST Replace the internal 9 volt battery OFF OFF OFF OFF OFF Flashing OFF Perform Annual Filter PM Maintenance If condition persists service is required A Perform Annual Filter PM Maintenance If condition persists Flash
27. DO NOT use a humidifier in the Pulse Flow Mode The Eclipse 5 will not detect inspiratory effort The device will alarm and default to the Continuous Flow Mode for continuing operation after 15 seconds CAIRE Humidifier Adapter Part Number 7116 SEQ If your physician has prescribed an optional humidifier you may need to use the CAIRE Humidifier Adapter Follow the instructions for use Attach the Humidifier Adapter to the oxygen outlet port of the Eclipse 5 and then to the humidifier Attach the cannula or oxygen tubing to the humidifier outlet Salter Labs Oxygen Supply Tubing Part Number Series 2000 or equivalent The internal diameter should be no less than 3 16 0 48 cm Connect the oxygen tubing to the outlet port of the humidifier or directly to the oxygen outlet port of the Eclipse 5 if you do not use a humidifier Connect the other end of the tube to the nasal cannula if oxygen supply tubing is not already attached to the cannula Tubing not specified for use with this Eclipse 5 may impair the performance of the device Salter Labs Oxygen Cannula Part Number 1600 Series or equivalent Your physician will have prescribed a cannula to deliver oxygen In most cases they are already attached to the oxygen tubing If not follow the instructions included with the cannula to attach it to the oxygen tubing Use of an oxygen cannula not specified for use with this Eclipse 5 may impair the performance of the device PN 20631679 RevD 83 LV
28. Eclipse 5 with autoSAT Personal Ambulatory Oxygen System PAOS PROVIDER TECHNICAL MANUAL eclipse toSAT LF CAIRE SEQUAL General Information ccccccccccsssscssssscccssscssccees 4 Warning and Caution Statements 4 Introduction to the Eclipse 5 Oxygen System s 5 Eclipse 5 Oxygen System Specifications 04 6 Pulse Dose Mode Specifications sss 7 Power Accessory Specifications eessssnsssssessesenseceenesestneseeen 7 AC Power Supply ssssssssssssssesesososeoossesonosenososaseonesonosestsosasnrernonoseses 7 DC Power SUD IV ssssssssssssessstssstrsssonosestststtoornonesesentrnsnenonrnososnnene 7 Power Cartridge Battery 7 Independent Safety Testing cccssssssssssssssssssssssssssssssssees 8 Provider Support PONCY s issaccsciacansmalhinnansndnaiions 8 Electromagnetic Compatibility essssesessssssesssssssesesssssees 9 Guidance and Manufacturer s Declarations 9 How The Eclipse 5 Works csscsssccssssssccssssssscessees 13 MOJUTO rerainane TRES 3 ATF Concentrator Module ssssssisisiriinsiiisn 3 Compressor and Compressor ENClOSUME csssssssssseeseneen 4 Power Distribution 2c ascisecscsseissiisdsncrbungasiscudadsieimieecmnaintes 4 CONTO BOI Ree ene ee eee ene TE eee er ee ee 4 510110 M21 eee ee emer 5 User Controls amp System Status INdICatOrS nesses 5 Continuous Flow MOE ssssssssssessssssssesensssssessnssssssssnsssssesen 7 Pulse Dose MOE wcs
29. Field service reps can travel to a patient s home troubleshoot and transmit the data to either in house tech support personnel or CAIRE Support personnel can diagnose provide solutions and print documentation for equipment records EDAT also makes Eclipse 5 software upgrades easy and accessible For more information on EDAT contact CAIRE Inc Technical Service at 1 800 482 2473 and reference EDAT PN 5535 SEQ PN 20631679 RevD 19 EA F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT POWER SUPPLIES The Eclipse 5 may operate from either the AC or DC Power Supply or the Power Cartridge When power is available from an external supply the Eclipse 5 will draw from the external source rather than depleting the Power Cartridge Connection to external power is indicated when the External Power Present Indicator located on the Control Panel is illuminated AC POWER SUPPLY The AC Power Supply is a universal input type capable of accepting 100 240VAC 50 60 Hz It is capable of delivering up to 200W of 28VDC output The input cord requires a grounded receptacle Country specific cords or universal power adapter PN 5591 SEQ may be used with the AC Power Supply as the power supply is equipped with a universal input receptacle When used in a medical care facility international safety standards require the use of hospital grade AC power cord with the Eclipse 5 Figure 5 Eclipse 5 AC Power Supply The E
30. L lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT CAIRE INC CUSTOMER SERVICE CONTACT INFORMATION If you need any additional assistance contact CAIRE Inc ie By mail CAIRE Inc 2200 Airport Industrial Drive Suite 500 Ball Ground GA 30107 USA By telephone 800 482 2473 By E mail techservice usa chart ind com www sequal com AUTHORIZED EUROPEAN UNION REPRESENTATIVE Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany E mail techservice europe chart ind com www sequal com SeQual ATF and Eclipse 5 are registered trademarks of CAIRE Inc PRO Check Elite is a trademark of Salter Labs Inc Lysol is a registered trademark of Reckitt Benckiser UK Loctite is a registered trademark of Loctite Corporation USA CE 0029 Copyright 2013 Chart Industries CAIRE Inc reserves the right to discontinue its products or change the prices ma terials equipment quality descriptions specifications and or processes to its products at any time without prior notice and with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable PN 20631679 Rev D 84
31. M STATUS INDICATORS CONT Definition Definition No Smoking Icon button Do not smoke near unit Providers can access provider mode software functions using the Control Panel All provider mode information is displayed on the LCD The software shall FAA approved for use aboard passenger aircraft advance the following Eclipse 5 display mode when the No Smoking icon is pressed see pg 15 ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priority awareness condition or Caution Use of the unit may continue and refer to the Troubleshooting Pulse Mode Operation Table for the proper response A flashing yellow indicates a medium priority alert A prompt response is necessary ALARM Red Indicator High Priority Alarms Indicates a high priority alarm condition An immediate response is necessary Refer to Device operating normally power button the Troubleshooting Table Flow Setting Indicator This is the main focus on your control panel The home care provider will correctly set the prescribed flow for either the Continuous Flow Mode LPM and or the Pulse Dose settings mL Do not get wet Each time you power the device ON the previous mode and or setting has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approximately 20 of the total bat
32. Manual WITia autoSAT CAUTION Be sure to accurately determine the amount of current the vehicle AN accessory outlet is rated to supply Avoid placing the Eclipse 5 in direct sunlight Do not store the Eclipse 5 in a vehicle where the device may be subject to extreme temperatures Extreme heat or cold may impair operation and damage the device AN and degrade the Power Cartridge When using the Eclipse 5 in a vehicle check the cannula to make sure it is not pinched or occluded PN 20631679 RevD 33 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT TRAVELING BY AIR TRAVEL APPROVED The Eclipse 5 is an FAA approved portable concentrator A new US Department of Transportation regulation regarding portable oxygen concentrators took effect on May 13 2009 Under this regulation every FAA approved portable concentrator is now authorized for use during any commercial flight that departs or arrives in the USA regardless of whether the airline itself has approved the device or not When traveling by air instruct your patient to remove the Power Cartridge from the Eclipse 5 before checking the unit as checked baggage or cargo Transport of the Eclipse 5 as cargo or checked baggage with the Power Cartridge installed is prohibited by international air cargo regulations The Power Cartridge may be installed if the patient is using the Eclipse 5 as a carry on item If the airline allows u
33. NIT COVER CAIRE Inc recommends providers have EDAT PN 5535 SEQ or a spare Control Panel PN SP20618461 that can be attached to the unit and the unit run for a minimum of 20 minutes prior to re installing the cover to check for a smooth and quiet calibration to be replaced by operation To re install the cover follow the steps using the reverse order Ensure that the Membrane Panel Overlay cable plug is correctly lined up with circuit board connection and that there are no twists in the ribbon cable Figure 26 Membrane Panel Ribbon Cable Installation Align Front Cover over Control Board and then gently position front cover Verify that the black rubber gasket channel seal and strain relief are aligned and fit properly Do not over tighten cover screws 5 in Ibs at maximum After the cover has been reinstalled but before installing screws on the cover re connect power and start unit Let the unit run for 20 30 minutes Whenever the cover of the Eclipse 5 is removed proper functionality should be checked by following the Test Procedure described next PN 20631679 Rev D 48 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT TEST PROCEDURES PURITY AND FLOW RATE TEST PROCEDURE PREFERRED METHOD It is recommended that the Eclipse 5 be tested for oxygen concentration and flow performance The CAIRE recommended test setup is shown on the following diagram Oxygen monitors may or m
34. Oxygen System Simplified Block Diagram Control Panel Ribbon Cable litle Figure 29 Top Case Components PN 20631679 Rev D 61 ii i T bai gt Control i Board 4 ATF PONa m Module l LS a K a Compress y Intake Filter Compressor Box Product Tank Power Manager PC Board A e a Figure 30 Bottom Case Components E F f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual autoSAT Oxygen Circuit REMOVE AND REPLACE THE ATF MODULE There are no serviceable parts inside the ATF Module Do not attempt NOTE to disassemble or modify the ATF Module 1 Remove the Unit Cover as described in the section Remove and Replace the Unit Cover 2 Cut the cable tie as shown below and disconnect the silicone tube that goes into the product port of the ATF this tube comes from the Product Tank 3 Install port cap Silicone Tube to Product Tank Figure 31 Removal of ATF Module Product Hose 4 Cutthe cable tie that secures the silicone tube to the braid tube at shown below To ATF Module Vacuum Port To ATF Module Pressure Port Figure 32 Removal of ATF Module Pressure and Vacuum Tubes PN 20631679 RevD 63 CLL f p SA4V Personal Ambulatory Oxygen System Provider Technical Manual autoSAT 5 Cut the cable ties to the braided tubes that connect into the ATF pressure and vacuum ports as shown in Figure 32
35. USE AND VENTILATION Ask your patient where they would like to set up the device Whenever possible the Eclipse 5 should be in the same room as the patient for convenience and assurance that the patient can adequately hear and respond to Eclipse 5 alerts and alarms While unpacking and setting up the device tell your patient about these important cautions and warnings WARNING Locate the Eclipse 5 in a well ventilated space that provides adequate airflow Ensure that furniture draperies or clothing will not impede air circulation Avoid placing the unit over a floor heat register or against a ON baseboard heating system Do not use in the presence of flammable anesthetics solvents aerosols or flammable cleaning agents Avoid high pollutant environments CAUTION Some patients are highly mobile and may use the device under varying circumstances Make sure your patient or patient caregiver completely understands the basic precautions to safely locate the device After completing this training procedure ask your patient if s he has NOTE any questions THE USERS MANUAL Give your patient a copy of the Eclipse 5 Users Manual and advise them to read prior to operating Open the manual and briefly review the Table of Contents including subheadings Show your patient the icons and symbols and explain what each means CAUTION Advise your patient to thoroughly read the Eclipse 5 Users Manual and keep the manual in a saf
36. acity of the Power Cartridge Under certain conditions the Eclipse 5 may not have enough external power available to charge the batteries at the full rate In this case the charging rate will be limited to the available power As the Power Cartridge accumulates charge the charging current required will eventually fall When the charging current for each Power Cartridge has fallen below 5 of the rated capacity charging is complete and the Power Cartridge charger is shut off The software will only charge when the Power Cartridge temperature is at or below 45 C The software will always run the cooling fan whenever the charger is enabled The Power Cartridge charger is disabled and the cooling fan is set to maximum whenever the Power Cartridge temperature exceeds 45 C The Power Cartridge charger is disabled when temperature is less than 0 C All lithium ion batteries self discharge at very low rates when not in use Eclipse 5 Power Cartridges are shipped from CAIRE Inc in a partially charged state nominally 40 When stored in a cool dry location the Power Cartridge can sit unused for up to 12 months without appreciable self discharge occurring Nevertheless CAIRE recommends a first in first out rotation of Power Cartridge inventory for maximum Power Cartridge service life PN 20631679 Rev D 24 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITH autoSAT Training The Patient INTRODUCTION Welcome
37. acturer PN 20631679 RevD 35 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT MONTHLY MAINTENANCE PATIENT CARE FOR THE POWER CARTRIDGE The Power Cartridge battery in the Eclipse 5 requires special care to assure a longer life and the highest level of performance The CAIRE Power Cartridge is the only approved Power Cartridge recommended for use with the Eclipse 5 The following are generic guidelines for the Power Cartridge Avoid high temperatures Avoid cold temperatures Do not drop Power Cartridge No not poke objects into contacts Power Cartridge Cleaning Use a damp not soaking wet cloth or sponge First spray the cloth or sponge with a mild detergent and then clean the Power Cartridge case and the latch WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury or harm A N DO NOT tamper with or try to repair the Power Cartridge There are no serviceable parts inside Power Cartridge Storage The Power Cartridge should be stored in a cool and dry location CALIBRATING THE POWER CARTRIDGE In order to maintain the Power Cartidge your patient should fully discharge the Power Cartridge once a month In order to ac curately maintain the Eclipse 5 Power Cartridge periodic full discharge of Power Cartridge to zero is recommended This serves to calibrate the Power Cartridge Status Indicator To do this the patient will need an AC Power S
38. adequate clearance around the DC Power Supply and the Eclipse 5 Ensure the DC Power Supply and Eclipse 5 are secured in the vehicle so that is does not become a projectile in the event of a sudden stop A N DO NOT operate the Eclipse 5 on DC power while the vehicle engine is off In such a scenario the Eclipse 5 could quickly drain the vehicle s battery CAUTION omon When operating an Eclipse 5 5 with the DC power supply the power cartridge will only recharge if there is sufficient excess voltage from A N what the concentrator requires to operate When the Eclipse 5 is running from an external AC or DC power supply the External Power Present Indicator will illuminate E The Eclipse 5 will always operate from external power when external power within the specified range is available ACTIVE LIFESTYLE TRAINING Small lightweight and easily moved about the Eclipse 5 is ideally suited to an active lifestyle The AC and DC supplies enable recharge of the Power Cartridge during longer excursions and even extended overnight travel To train your patient on mobility and ambulation follow the procedures below PN 20631679 RevD 31 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT CAUTION Always check to see that the Air Inlet and the Exhaust Vent are not blocked and the Air Inlet Filter is dry and clean before using your AN Eclipse 5 Do not drop the Eclipse 5 or Eclipse 5 power su
39. air dry thoroughly 5 Re insert the filter in the cabinet 6 Maintenance may be easier for the patient when a second filter is provided Figure 10 Cleaning the Air Inlet Filter CAUTION The filter should be completely dry before using it again Excess moisture may impair proper operation or damage the unit Replace the Air Inlet Filter annually If the Eclipse 5 is operated in a dusty environment the filter may need to be cleaned or replaced more often Do not operate the Eclipse 5 for more than 30 minutes without a filter installed CLEAN AND CARE FOR THE TUBING AND CANNULA Provide your patient instructions on cleaning disinfection and or replacement information for the tubing and cannula CLEAN THE CABINET AND CONTROL PANEL AND POWER SUPPLIES To clean the cabinet Control Panel or power supply do the following 1 Turn OFF the Eclipse 5 and disconnect from AC or DC power before any cleaning or disinfection activity 2 Use mild detergent and water solution 3 Use a damp not soaking wet cloth or sponge 4 Spray or wet the cloth or sponge with the mild detergent solution DO NOT spray the cabinet Control Panel or power supplies 5 Wipe down the cabinet Control Panel or power supplies 6 To disinfect the Eclipse 5 use Lysol Brand II disinfectant or equivalent Spray or wet a cloth or sponge with the disinfectant DO NOT spray the cabinet Control Panel or power supplies Proceed as directed by the manuf
40. al AC or 3 Ensure that the power verification lights are on DC power to charge the There will be a green light on the power supply transformer box battery There will be a green light on the Eclipse 5 concentrator control panel in the shape of a power The Power Cartridge bat cord tery is not charging 4 If power verification LEDs are not lit remove all connections of the power supply for 20 seconds and reconnect 5 If the condition persists contact Chart Technical Support for service 1 Remove power cartridge and re install to ensure that it is secure in the concentrator The Eclipse 5 is not 2 Inspect the Eclipse 5 for a solid yellow light when the battery is installed or a flashing battery properly communicating icon when attempting to charge If either of these symptoms is present continue to step 3 with the power cartridge 3 If possible install a separate power cartridge in the Eclipse 5 If the replacement cartridge to charge it works properly then the original cartridge needs to be replaced 4 If this condition persists contact Chart Technical Support for service The Eclipse 5 beeps 3 times N A The internal battery that 1 Internal 9V battery needs replacement Perform preventative maintenance or contact Chart when it is first powered on powers the alarms islow Technical Support for service WARNING WARNING Do not modify this equipment without authorization of the manufacturer PN 20631679 RevD 58
41. al over temperature conditions If any of these conditions exist output power will be interrupted and the LED will turn off These three conditions are self resetting and output power will resume when protection circuits fall back into acceptable operating ranges The Eclipse 5 will run at all flow settings when being operated with the DC power supply in a vehicle including 3 0LPM on continuous flow and all pulse settings 1 9 The Eclipse 5 will charge the power cartridge while being operated with the DC power supply while running at 2 0LPM or below with continuous flow The power cartridge can charge at all pulse flow rates however charging functionality may stop depending on the breathing rate of the user If the Eclipse 5 power cartridge is being charged the battery symbol will display with a waterfall animation However if the power cartridge is not charging then the battery symbol will remain steady and display the current charge capacity DO NOT use the DC supply to power the Eclipse 5 once the vehicle s engine is turned off as this may drain the car s battery POWER CARTRIDGE BATTERY The Power Cartridge allows operation away from AC or DC power The Power Cartridge used with the Eclipse 5 contains Lithium ion battery cells similar to those used in laptop computers and cell phones The Eclipse 5 Power Cartridge contains a quantity of two 97 5 Watt hour battery packs Each battery pack contains 7 92 grams of equivalent lithium content
42. ansporting the Eclipse 5 54 Storing the Eclipse 5 sssssssssissssisisssisrrsrerrsrssrrrsssr 54 Dis Can ING eee E 54 Troubleshooting Service and Repair Procedures sssssoocsssoocesssoocesssoocesssocsesss 55 System Troubleshooting and Alarms wuss 56 Alarm Conditions and Alarm Codes wesc 59 Malfunction CodeS s s ssssssssessissssisrssrsrsrrsrsrsrrersrsernrsrssrsrsrssrsrserns 60 System Schematics and Diagrams scesssssseesesseeesssseeen 61 Oxygen CIC Uit paces cccaprscensttiares enn a 63 Remove and Replace the ATF Module nsss 63 Remove and Replace the Product Tank Assembly PN 4378 SEQ Eclipse 5 wes 67 BIECHONICS aiena ea a E 70 Remove and Replace the Control BO GC Assembly crgssieocenienannesiaiise 70 Remove and Replace the Buzzer Wire Harness 72 Control Board Connector Diagram s s s 73 Remove and Replace the Power Manager Printed Circuit Board s 74 COMPE SSO denna 78 Remove and Replace the Compressor BOX 78 Maintenance and Replacement Parts 82 Preventative Maintenance Pafts sssssssssssississisin 82 Replacement Parts LIST sssc croscsctsasusssttessevessaseaonsvasssvarscotanies 82 Optional ACCeSSOrieS acaccsccssascssssessessctdvacetcececesienest stesenietdes 83 CAIRE Inc Customer Service Contact Information ssssssssscsessssssssssseccsoossssssesosossossssse 84 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WH autoSAT GENERAL INFORMATION Th
43. ay not have an internal pump to draw samples of oxygen to be measured Placement of the oxygen monitor in the test setup depends whether they have an internal pump Only one oxygen monitor is needed An oxygen monitor such as Salter Labs PRO Check Elite or equivalent may be used Connect the circuit per the diagram shown below Turn the concentrator on and set the target flow on the Eclipse 5 to 3 LPM continuous flow Allow the Eclipse 5 to stabilize can take up to 10 minutes Verify the O concentration and O flow rate are within specification shown in Table 1 ECLIPSE Flow Meter O Monitor Vent to Unit alternate location Atmosphere O Monitor with Pump preferred location Figure 27 CAIRE Recommended Test Setup to verify performance of Oxygen Concentrator PURITY AND FLOW RATE TEST PROCEDURE ALTERNATE METHOD Attach a calibrated oxygen monitor to the oxygen outlet port in accordance with oxygen monitor manufacturer s recommendations Some sensing equipment may restrict the actual flow rate and provide inaccurate concentration read ings If testing is performed with the oxygen monitor in the alternate location pictured above ensure that the oxygen monitor does not significantly restrict oxygen flow PN 20631679 RevD 49 Personal Ambulatory Oxygen System Provider Technical Manual cr eclipse autoSAT ASSEMBLY AND ALARM VERIFICATION TESTS To e
44. bing 4 If a humidifier bottle is connected remove the humidifier bottle and connect cannula directly to outlet port Note Pulse mode will not work with a humidifier bottle installed 5 If the condition persists contact Chart Technical Support for service PN 20631679 RevD 57 CLL f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual aUutoSAT Other Alarm Conditions Continued Symptom Alarmcode Possblecause Yourhcn OOOO O O 1 Verify that the outlet is providing power Do not connect the power supply to a dimmer circuit or a power strip 2 Check that cable connections on power supplies are secure to the wall vehicle and con centrator If using the ACor DC power supply check the cable connection on the power supply transformer as well 3 Ensure that the power verification lights are on There will be a green light on the power supply transformer box There will be a green light on the Eclipse 5 concentrator control panel in the shape of a power cord 4 If power verification LEDs are not lit remove all connections of the power supply for 20 seconds and reconnect 5 If the condition persists contact Chart Technical Support for service Power Cartridge is draining The Eclipse 5 is not while plugged in to AC DC receiving power from the power supply attached power supply 1 Connect to AC or DC power to recharge the battery Verify that the battery charges for 2 5 hours and the batt
45. c Compatibility standard These limits are designed to provide reasonable protection against harmful interference in a typical medical installation This equipment generates uses and can radiate radio frequency energy and if not installed according to the instructions may cause harmful interference to other devices in the vicinity There is however no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the User is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device Increase the separation distance between the equipment Connect the equipment into an outlet on a circuit different from that which the other device s are connected Consult with CAIRE s Technical Support Department for help Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cables other than those specified with the exception of transducers and cables sold by the Manufacturer of this device as replacement parts for internal components may result in increased Emissions or decreased Immunity of the Eclipse 5
46. clipse 5 AC Power Supply does not contain a fan When in use the AC Power Supply should be located in a well ventilated area Located near the output cord there is a green LED that is illuminated when the AC Power Supply is supplying 28VDC power If the LED is not illuminated there is no input power available In addition the Power Supply contains protection circuits for output over current input over voltage and internal over temperature conditions If any of these conditions exist output power will be interrupted and the LED turns off However these three conditions are self resetting such that output power will resume when protection circuits fall back into acceptable operating ranges DC POWER SUPPLY The DC Power Supply is intended for use with DC accessory outlets such as those found in motor vehicles Input voltage to the DC Power Supply is 11 5 16 VDC Sized to be compatible with most passenger vehicle electrical systems it is capable of delivering up to 150W of 26VDC output to the Eclipse 5 PN 20631679 Rev D 20 EJ F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT Figure 6 Eclipse 5 DC Power Supply Located near the output cord there is a green LED that is illuminated when the DC Power Supply is supplying 26VDC power If the LED is not illuminated there is no input power available The Power Supply contains protection circuits for output over current input over voltage and intern
47. d assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electromagnetic 6 kV contact 6 kV contact Floors should be wood concrete or ceramic environment tile If floors are covered with synthetic mate guidance rial the relative humidity should be at least 30 8 kV air 8 kV air IEC 61000 4 2 Electrical fast tran 2 kV for power supply 2 kV for power supply sient burst lines lines Mains power quality should be that of a typi cal commercial or hospital environment IEC 61000 4 4 Surge 1 kV line s to line s 1 kV for input output lines N A 1 kV line s to line s Mains power quality should be that of a typi IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power frequency 50 60 Hz mag netic field IEC 61000 4 8 2 kV line s to earth lt 5 U gt 95 dip in U for 0 5 cycle 40 U 60 dip in U_ for 5 cycles 70 U 30 dip in U_ for 25 cycles lt 5 U gt 95 dip in U for 5 sec 2 kV line s to earth lt 5 U gt 95 dip in U for 0 5 cycle 40 U 60 dip in U_ for 5 cycles 70 U 30 dip in U for 25 cycles lt 5 U gt 95 dip in U for 5 sec NOTE U is the a c mains voltage prior to application of the test level cal commercial or hospital environment Mains
48. d key press by the user PN 20631679 RevD 16 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual autoSAT CONTINUOUS FLOW MODE Continuous Flow Mode delivers a constant flow of oxygen to a patient by means of tubing and a nasal cannula at rates between 0 5 LPM and 3 0 LPM Within the Eclipse 5 concentrated oxygen is stored in a 500ml product tank at pressures in the range of 5 to 9 psi This pressure gives Continuous Flow Mode the capability to deliver the indicated flow rate to the patient even if extension tubing is used up to 50 feet long In addition the Eclipse 5 Continuous Flow Mode is fully compatible with humidifier use with NC not exceeding 7 ft PULSE DOSE MODE The Eclipse 5 Pulse Dose Mode delivers a measured bolus of oxygen at the very beginning of each inspiration The approach is based on the manner in which gas is absorbed into a patient s airway Eclipse 5 users may select pulse dose volume delivery rates Regardless of setting the pulse profile is a simple square wave pulse based on a 16 mL volume Refer to Pulse Profiles table below Pulse durations are no less than 100 milli seconds at the beginning of the inspiration There are three Rise Time settings Slow Medium and Fast The table below is for the FAST Rise Time setting factory default is Fast Flow Pulse Peak Flow Pulse Duration Bolus Setting LPM volumetric milli second Volume ml 1 0 8
49. determined and set by a trained clinician Adjusting the Rise Time will not affect the chosen volume of oxygen delivered to the patient 16 192 mL CONNECTING THE AC POWER SUPPLY The Eclipse 5 operates from external power when connected to a power outlet To connect the Eclipse 5 to the AC Power Supply follow these steps 1 Insert the AC supply plug tip into the Eclipse 5 2 Insert the power cord into the AC Power Supply 3 Plug the power cord into a grounded outlet 4 The power supply LED displays green to indicate that the AC Power Supply is drawing power The External Power Present Indicator on the Control Panel illuminates o If the Power Supply Status LED is not illuminated after inserting the plug into a grounded outlet check to make sure the Power Supply and power cords are securely plugged into the Eclipse 5 WARNING Ensure adequate clearance around the AC Power Supply The AC Power Supply is universal input but the AC power cord AN is appropriate to specific country s electrical service Ensure that power cord is appropriate to country s electrical service CAUTION DONOT connect the Eclipse 5 to an extension cord or electrical outlet controlled by a switch When removing the AC Power Supply from the Eclipse 5 remove the plug from the AC outlet before removing the AC Power Supply plug from the Eclipse 5 International safety standards require the use of hospital grade AC power cords 3893 SEQ
50. e Eclipse l hours then rolls over to 00 000 5 will drop out of Provider Mode PN 20631679 RevD 5 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT ELECTRICAL SAFETY TEST This is required only for the Eclipse 5 Oxygen System that is used in a hospital or institutional environment This is not required for home care use To test the basic electrical safety of the Eclipse 5 AC Power Supply CAIRE recommends using an LKG 601 Electrical Safety Analyzer Netech Corporation Hicksville NY or equivalent to verify that the current leakage to ground is within appropriate limits 1 Plug the AC Power Supply into the electrical safety analyzer Disconnect the AC Power Supply from the Eclipse 5 unit 2 Plug the electrical safety analyzer into a wall outlet 3 Follow the analyzer manufacturer s instructions for measuring both the forward and reverse earth leakage current only Verify that forward and reverse current leakage to ground is lt 250 uA 100 115VAC applications and lt 500 uA 220 240VAC applications 4 Disconnect the AC Power Supply from the electrical safety analyzer 5 If the AC Power Supply measures leakage current greater than the criteria in step 3 please return it to CAIRE for service 6 EDAT may be utilized to record Eclipse 5 device status and dates of service CLEANING THE ECLIPSE 5 Clean inside the unit as needed using a small vacuum cleaner or brush to re
51. e place for future reference SHOWING PANEL BUTTONS INDICATORS ALERTS AND ALARMS Open the Users Manual to the section on UNDERSTANDING THE CONTROLS AND FRONT PANEL and show your patient where each button is located on the Eclipse 5 Control Panel and how each button and indicator operates Go over the alerts and alarms and how to handle alert and alarm conditions Detailed instructions for each button and indicator can be found in the Eclipse 5 Users Manual PN 20631679 Rev D 27 EA F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT SHOWING POWER CARTRIDGE POWER LEVEL The display on the Control Panel shows the amount of Power Cartridge capacity available and waterfalls when charging Point out the table showing typical new Power Cartridge duration of use time in the Users Manual WARNING The display gives an approximate level of remaining battery power If the patient is dependent on oxygen therapy the users should always have a backup oxygen supply or access to AC or DC Power Supply in the event of loss of battery power SELECTING THE PROPER FLOW MODE AND SETTING The recommendations of the qualified clinician responsible for the patient s care should always be followed when instructing patients on the use of Continuous or Pulse Dose Mode WARNING Do not attempt to prescribe oxygen level settings Warn your patient that only a qualified clinician or physician is qualifi
52. e sure there are at least 3 inches The yellow light is flashing 008 The Eclipse 5 has detected between the back of the Eclipse 5 and any obstructions furniture curtain etc There is no audible alarm low oxygen levels 3 If operating in the car be sure the back of the Eclipse 5 is facing outward in the seat 4 Perform the preventative maintenance and replace the HEPA Air Inlet and Compressor Intake filters 5 If the condition persists contact Chart Technical Support for service PN 20631679 RevD 56 C eclipse autoSAT Personal Ambulatory Oxygen System Provider Technical Manual Red Light The red light is flashing The alarm is giving 3 beeps every 2 minutes The red light is solid The alarm is giving a constant beep The Eclipse 5 is not delivering oxygen and it will not power on The red light is solid The display screen says FAIL The Eclipse 5 is not produc ing delivering oxygen and will not power on While in pulse mode the compressor speeds up and the pulse mode LED flashes quickly for 15 seconds FAIL Xx The Eclipse 5 has detected low oxygen levels The Eclipse 5 has lost power temporarily while running on external power and needs to be re set The Eclipse 5 has lost power due to a depleted power cartridge charge or an over heated power cartridge The Eclipse 5 has experienced a system malfunction 1 Ensure the air intake filter is not clogged or restricted Clean a
53. eat the Power Cartridge above 140 F 60 C The Power Cartridge may present a risk of fire or explosion and will void the warranty CAUTION A Store the Power Cartridge in a cool dry place when not in use CAUTION DO NOT leave the Eclipse 5 or the Power Cartridge in a vehicle or in the trunk during a hot or cold day CAUTION The Eclipse 5 system can only work with a CAIRE Power Cartridge Use of another Power Cartridge or Battery may damage the unit present a risk of fire or explosion and will void the warranty Figure 8 Power Cartridge Status Gauge While the Eclipse 5 Power Cartridge allows the concentrator to operate at its full range of capabilities the primary purpose of the Power Cartridge is to allow a patient to ambulate while they are moving between stationary power sources The Eclipse 5 s Power Cartridge like all lithium ion batteries is susceptible to permanent damage from excessive heat Exposure to excessive heat may significantly shorten the service life of the Power Cartridge Each Power Cartridge contains multiple temperature sensors to monitor battery cell temperature The amount of heat the Power Cartridge can safely endure varies depending on how the Power Cartridge is being used During discharge the Eclipse 5 software will alarm when internal battery cell temperature exceeds 60 C and will shut the system down if internal battery cell temperature exceeds 70 C While the Power Cartridge is charging so
54. ed to perform the necessary tests to prescribe treatment SELECTING CONTINUOUS FLOW MODE When delivered from CAIRE Inc the Eclipse 5 is set to Continuous Flow Mode Refer your patient to the Users Manual for instructions on how to select Continuous Flow Mode Demonstrate the procedure on the device Set the prescribed flow using the increase or decrease flow rate buttons CAUTION Continuous Flow Mode is required in order to use a humidifier bottle Continuous Flow Mode without a humidifier bottle is required in order use 50 foot tubing SELECTING PULSE DOSE MODE Pulse Dose Mode delivers a measured bolus of oxygen pulse at the beginning of each inspiration Refer to the Users Manual for instructions on how to select Pulse Dose Mode Demonstrate the procedure for your patient Set the prescribed flow using the increase or decrease buttons Verify that the patient is able to adequately trigger the oxygen delivery while speaking sitting quietly and walking If the patient is not able to satisfactorily trigger oxygen delivery the Pulse Sensitivity may need to be adjusted CAUTION e Donot use a humidifier when in Pulse Dose Mode e Use only a 7 foot cannula when in Pulse Dose Mode Do not use a tubing extension Review with the patient the breath rate capabilities of Pulse Dose Mode for their prescription See Table on page 7 PN 20631679 RevD 28 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WH
55. ery icon is full and not flashing before use 2 Refer to the actions for Power Cartridge is Not Charging Symptom below The power cartridge was not fully charged The Eclipse 5 was not op erating at the anticipated The Power Cartridge bat flow rate tery charge is not lasting as long as it should 1 Ensure that the prescribed flow rate is being used and that pulse or continuous flow is being used as required 1 Attempt to drain the battery cartridge completely Do this by running the Eclipse 5 on bat tery power until the machine shuts off completely Then remove the battery and allow it to cool The power cartridge isnot for approximately 30 minutes After this time re insert the battery in the Eclipse 5 and recharge performing to specifica it using ACor DC power Perform the Battery Calibration procedure on page 36 tions 2 Time the battery duration and compare it to the chart on page 22 3 If the battery duration is less than 80 of the run time listed in the chart on page 22 after the battery calibration it is recommended to replace power cartridge 1 Verify that the outlet is providing power Do not connect the power supply to a dimmer circuit or a power strip if using AC power supply 2 Check that cable connections on power supplies are secure to the wall vehicle and concentra tor If using the AC power supply check the cable connection on the power supply transformer as The Eclipse 5 is not well receiving extern
56. es a user interface consisting of a membrane panel keyboard Liquid Crystal Display LCD external power present indicator Power Cartridge capacity indicator alarm status indicators and an audio transducer The user interface informs the user of the system status and allows the user to set the desired flow rate and flow mode Figure 3 User Control Panel USER CONTROLS AND SYSTEM STATUS INDICATORS The Eclipse 5 control panel displays important operating information This section will help you understand this op erating information ON OFF Button Green Indicator This button powers the device ON or OFF The Green Indicator is illuminated when the device is ON Use these buttons to change prescribed settings Delivery Mode Button and Indicator The button toggles between Continuous Flow and Pulse Dose Mode The Pulse Dose Mode activates autoSAT Technology as the patient s breath rate changes the Eclipse 5 servo controls the unit to provide a consistent bolus size The Pulse Dose Mode allows a significant increase in the operating time while powered by the battery When Pulse Dose Mode is activated the green Delivery Mode Indicator illuminates and a pulse of oxygen is delivered with each inspiratory effort When in Continuous Mode the LED is off A Increase or Decrease Setting Buttons PN 20631679 Rev D 15 LVL f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual autoSAT USER CONTROLS AND SYSTE
57. esssssssssscesssssnsssssssesesssssnsssseessussnsssssssessssssesee 7 Provider Mode FUNCION S s ssssesssssssessssssonosorosesnstrnrsonoseseesaeres 8 Alam COE ge REE EEEO 8 Pulse Mode Sensitivity Adjustment wees 8 Hours of Operation s ssssssssssressssosssesostsestrorsonesesesesosereenonesenene 8 ON Battery STETU S sssusa 8 Rise Time Bolus DeliVery sssssssssesssssesesessssessssseeses 8 Software VEISION sscnscaninciaisasanminawancmaiios 8 Service Mode FUNCTIONS scssvessieciacnscassvctsatvccintasenesaincinveeeanie 9 Eclipse 5 Data Acquisition Tool EDAT essssssesee 9 Power SUPpli S s ssssssssssesosssesesssssesersostsostntsnsooeosenosostntroroenosenenosaseres 20 AC Power SUD BIV wccan ccceauiteccnisinamieuhaoachaleieiaics 20 DC Power Seca cenro base lacicdeebesteeacde 20 Power Cartridge ese crccttaccrereonnaaieuan cadena 21 Typical New Power Cartridge Operation Time 21 Charging AlgOrIthM eacesseccenesacreccebseicorentsbesssoeetteoecercheeecteersx 24 Training The Patient ssooessssocesssoocessssocesssoosessosoe 25 WRG ENON O e re EEE EN 25 Pre Delivery Check LIST sccsccessuramanicednrctessaccaniqanaatet 25 Indications for US sssssssssssssssosesesososasoresnsnosesonososoonsnroonosasesesnseons 26 ContrainNdications wiscdskitcaconmencandwecendnanosmune 26 Basic COncEptITA INING scccsizesttscoesetsS cacsscsacestansbarensei easbeasassnescists 26 Safety Guidelines and Operational Safety Warmings Ca Uton ote
58. from the adsorbent material using a vacuum pressure cycle Oxygen flows from the product tank through a HEPA filter and past a sensor that measures flow and concentration A flow control valve regulates the flow of concentrated oxygen presented to the patient The process is continuously repeated during operation PN 20631679 RevD 13 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITH autoSAT COMPRESSOR AND COMPRESSOR ENCLOSURE The Eclipse 5 Compressor is a two cylinder variable speed wobble piston compressor driven by a highly efficient Brushless DC BLDC motor When air flows into the Compressor enclosure it passes through an air intake filter muffler that muffles sound and filters out impurities Using one cylinder the compressor takes in filtered air and delivers it to the ATF Module under pressure The second cylinder draws a vacuum on the ATF module and exhausts nitrogen rich gas to the exhaust vent Using a multifaceted approach sound heat and vibration generated by the compressor are mitigated by the compressor enclosure Vibration and structure borne noise are addressed by the dual axis gimbal that supports the compressor and the tubing that connects the compressor to the ATF module The rigid walls of the compressor enclosure and the sound adsorbing foam that lines it diminish the radiated noise The centrifugal blower mounted within the compressor enclosure serves to efficiently draw cooling air
59. ftware will interrupt charger operation when the internal battery temperature exceeds 45 C or temperature is less than 0 C In both of these cases when internal battery cell temperature falls within these limits the Eclipse 5 will resume normal operation Operation near these temperature limits will not damage the Power Cartridge but are in place to ensure that the service life of the Power Cartridge is preserved Heat in the Power Cartridge is generated during discharge and can also be the result of operating the Eclipse 5 in high ambient temperatures The amount of internally generated heat varies with the flow setting higher flows induce greater Power Cartridge heating While high ambient temperatures are typically the result of operation on a hot day running the Eclipse 5 with inadequate ventilation can also add additional heat Always ensure that the Eclipse 5 is operated in a well ventilated space the air intake filter is clean and intake and exhaust vents are unobstructed PN 20631679 RevD 23 EA F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT CHARGING ALGORITHM The charging algorithm is performed by the Power Manager software and involves three basic decisions 1 When to start charging 2 How fast to charge 3 When to stop charging Charging begins when Power Cartridge voltage falls below 16 0 volts The charging current is limited by the charger capability and the rated cap
60. g AC or DC power every 2 minutes 3 Replace the power cartridge if another is available 4 If the condition persists contact Chart Technical Support for service The yellow light is flashing The alarm is giving 2 beeps every 30 seconds 1 Clean and replace cabinet inlet filter 2 Ensure that the cannula is not kinked or blocked If used with a humidifier bottle ensure that it is filled properly and not creating a blockage 3 Ensure that the Eclipse 5 has proper ventilation It needs to be at least 3 inches from any The yellow light is flashing The Eclipse 5 has detected a surface to ensure the vents aren t blocked The alarm is giving 1beep 0 flow restriction causing low 4 Remove any external tubing attachments to verify that any attached devices are not causing every 2 minutes or blocked flow a flow restriction 002 010 The Eclipse 5 power cartridge 1 Plug Eclipse 5 into AC or DC power supply and recharge or replace power cartridge with fully voltage is low charged power cartridge 20 5 If the Eclipse 5 has been recently opened for maintenance verify that no tubing is kinked between the circuit board and the oxygen outlet 6 If the problem persists advise patient to switch to alternate source of oxygen and contact Chart Technical Support for service 1 Ensure the air intake filter is not clogged or restricted Clean and replace the filter if neces sary 2 Ensure the Eclipse 5 is in a well ventilated area Mak
61. he tubing PN 20631679 Rev D 46 E F f ps Personal Ambulatory Oxygen System Provider Technical Manual autoS INE To replace the compressor intake filter follow these steps 1 Cut the cable tie on the silicone tubes attached to the compressor intake filter Figure 24 Removing the Compressor Intake Filter 2 Remove the silicone tubes attached to each end of the filter body 3 Install a new filter by pushing each tube completely over the barb on the filter body Verify flow direction is correct 4 Reconnect silicone tubes to the filter Ensure that the feed tubes are not blocked crimped or kinked upon completion of the installation or the unit will alarm for low oxygen purity after the warm up cycle is complete Figure 25 Installing the Compressor Intake Filter 5 Re install two cable ties on silicone tubes PN 20631679 Rev D 47 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITi autoSAT After replacing the Compressor Intake Filter check the following Verify proper seating of the filter in the Eclipse 5 The arrow on the filter body should point toward the 9 volt battery Ensure that the inlet tube is inserted securely into its hole in the compressor box and is not pinched After filter is replaced and unit cover reinstalled proper functionality should be checked by following the Test Procedures described in this manual REINSTALLING OR REPLACING THE U
62. ider Technical Manual autoS INE 4 Lift the Control Board out of the Unit Cut the green cable tie to the silicone tube that is connected to the sensor as shown in Figure 42 Disconnect the silicone tube that is attached to the sensor Cut the green cable ties to the silicone tubes that are connected to the flow tube as shown in Figure 42 Disconnect the silicone tubes that are attached to the flow tube 5 Pinch or plug the silicone tube coming from the ATF product port and secure with a cable tie 6 Remove the Control Board Assembly 7 To install a new Control Board Assembly cut the cable tie used to pinch the silicone tube coming from the product tank Attach this tube and the tubes from the Proportional Valve and from the Product Tank to the flow tube as shown in Figure 42 Secure with a cable tie Connect the silicone tube to the sensor as shown in Figure 42 Secure with a cable tie Place the 2 screws in the sheet metal bracket as shown in Figure 43 Inspect the tubing to ensure the tubing is not kinked 8 Lower the Control Board Assembly into the Unit Case only to where the aligning pins go through the sheet metal bracket make sure the sheet metal bracket is 1 8 above the supporting ribs this will ensure the 2 screws stay in the bracket and do not fall in the unit case as shown in Figure 43 Start the 2 screws by turning them 2 times into the unit case After starting both screws lower the Control Board Assembly onto the
63. in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator POWER DISTRIBUTION The Power Manager takes external power that comes into the Eclipse 5 from the power supplies or Power Cartridge and monitors and controls power distribution to the rest of the Eclipse 5 system The Power Manager drives the compressor ATF module motor blower and provides power to the Control Board In addition when the unit is connected to an external power source the power manager monitors and controls the recharging of the Power Cartridge CONTROL BOARD The Control Board is at the center of nearly all Eclipse 5 functions The board constantly monitors dynamics such as temperatures pressures product flow and concentration and user input It determines proper compressor and ATF motor speeds needed in order to provide optimum system performance In addition this system supports the operation of the Control Panel and its indicators The Control Board utilizes a proprietary ultrasonic flow and concentration sensor and a flow control valve to accurately control the flow of oxygen in Continuous Flow and Pulse Dose Modes Ultrasonic Flow Concentration Measurement From Product Tank To Output Port Product Pressure Control Board Figure 2 Flow Sensor Board Diagram PN 20631679 RevD 14 C CC 4 p SA Personal Ambulatory Oxygen System Provider Technical Manual autoSAT CONTROL PANEL The control panel provid
64. ing 3 Beeps ae f service is required OFF OFF OFF OFF Check tubing and or humidifier for obstruction Check clean 1 Beep air intake filter If condition persists past 10 minutes service is required Cannot Charge Power Cartridge CHARGER Flashing Check battery connections One Wire ominunias ONEWIRE tion Loss No Inspiration Detected in Pulse Mode only If no inspiratory effort is detected after 15 seconds the Eclipse 5 will switch to Continuous Flow for 15 seconds and continue this sequence until a breath is detected Instruct patient to seek an external power source Remove Power Cartridge and allow to cool to room temperature If Power Cartridge malfunction persists service is required Loss of External Power deans without charged Power Cartridge installed flashing Instruct patient to seek an alternative oxygen supply Flashing ON for 5 min until external power is restored or install a charged Power Cartridge Reset power wait 20 seconds reconnect power restart ON 10 seconds device then silent Instruct patient to use back up oxygen supply Service is required PN 20631679 Rev D 59 System Fault FAIL XX CLL f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual AaUtoSAT MALFUNCTION CODES If a malfunction occurs in the Eclipse 5 the device will stop the Red LED on the front panel will light and the buzzer will sound for 10 seconds and then si
65. is technical manual will familiarize you with Provider specific information regarding the Eclipse 5 Oxygen System Instructions in this manual are intended to help ensure that Providers are familiar with Eclipse 5 system components and system principles of operation Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can be conveyed to patients Providers are made aware of the care diagnostics maintenance and repair of the Eclipse 5 Warning and Caution Statements Safety instructions are defined as follows WARNING Important safety information for hazards that might cause serious Important information for preventing damage to the Eclipse 5 Places emphasis on an operating characteristic or important consideration PN 20631679 RevD 4 C eclipse autoSAT Personal Ambulatory Oxygen System Provider Technical Manual INTRODUCTION TO THE ECLIPSE 5 OXYGEN SYSTEM ECLIPSE 5 easy UNIVERSAL CART ae PN 5991 SEQ Control Panel External Power Receptacl Oxygen eceptacle FAA Approval Icon Power Cord 4 Retainer Rating Label amp Serial Number Location FRONT AC POWER SUPPLY WITH NEMA POWER CORD PN 5941 SEQ DC POWER SUPPLY Y PN 5942 SEQ POWER CARTRIDGE BATTERY PN 7082 SEQ EDAT Service Port Not for Patient Use Air Inlet Filter J Cart Attachment FAA Approval Icon Location _ Power
66. lence The LCD will display one of the following Malfunction Codes Malfunction FAIL code Recommended Action Invalid RESET FAIL 80 Check for secure communication cables Verify functionality of PM and CS PCBs 1 0 Port Failure FAIL 81 Check for secure communication cables Verify functionality of PM and CS PCBs RAM Failure FAIL 82 Check for secure communication cables Verify functionality of PM and CS PCBs FLASH Failure FAIL 83 Check for secure communication cables Verify functionality of PM and CS PCBs EEPROM Failure FAIL 84 Check for secure communication cables Verify functionality of PM and CS PCBs IPC Watchdog Timeout FAIL 90 Check for secure communication cables Verify functionality of PM and CS PCBs Compressor Motor Too Hot FAIL 91 Allow Eclipse 5 to Cool for 30 Mintues Check compressor fan functionality Check to ensure vents P are not obstructed Verify functionality of the Compressor Assy PCB Too Hot FAIL 92 Allow Eclipse 5 to cool for 30 Minutes Verify functionality of PM fan Check to ensure vents are not obstructed Compressor Motor Stalled FAIL 94 Check Compressor wires for damage Verify functionality of PM PCB Verify functionality of the Compressor Assy Seen je Verify functionality of the Battery Bridge PCB Check Battery and Battery Bridge PCB for pin dam Battery cemmununieationiFallite a age Verify functionality of battery Verify functionality of PM pcb IPC Failure FAIL AO Verify functionality of PM PCB Product Press
67. ll not discharge ECLIPSE 5 MONTHLY RUN TIME PROCEDURE 1 Power on Eclipse 5 using AC Power DC Power or Power Cartridge 2 Allow unit to run for a minimum of 2 hours PN 20631679 RevD 36 Use the following checklist as a guide to assist in setup and training a patient on the use of the Eclipse 5 with autoSAT Technology and its accessories Patient Name Eclipse 5 Serial DC Power Supply Serial AC Power Supply Serial Power Cartridge Serial Training Topic Pre Delivery Check List Indications for Use Contraindications Basic Concept Training Advise to read the Users Manual Go over all accessories included with the Eclipse 5 AC DC Power Cord Cart Battery Safety Guidelines and Operational Safety Warnings Cautions Locating the Eclipse 5 Panel Buttons and Indicators Alerts and Alarms Power Cartridge Power Level Selecting Flow Mode and Setting Selecting Continuous Flow Mode ecting Pulse Dose Mode with autoSAT Technology ting the Pulse Dose Made Sensitivity Setting ecting Rise Time ttery Conservation feature Connecting the AC Power Supply Connecting the DC Power Supply Active Lifestyle Training Attach the Cart Changing Power Cartridge Battery Use Around the House Traveling by Vehicle Traveling by Air Traveling by Boat Traveling by Train Eclipse 5 Maintenance Weekly Clean the Air Inlet Filter Weekly Clean and Care for the Cannula Weekly Clean the Cabinet and Control Panel Month
68. ly Care for the Power Cartridge Annually Schedule PM at least once a year Trained By LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT Annual Maintenance Provider INTRODUCTION Properly maintaining the Eclipse 5 will ensure longer life and higher performance Minimum annual maintenance is required CAUTION The Eclipse 5 contains electrostatic sensitive components Do not open or handle except at a static free workstation Do not remove cover without ESD protection ANNUAL MAINTENANCE CHECK LIST Perform the following maintenance procedures at least once a year or more often as needed The frequency of the periodic maintenance should be based on the environment in which the Eclipse 5 is used Replace air inlet filter Check Power Cartridge Inspect AC and DC Power Supply plugs and cords for damage Replace as needed Verify that the AC and DC Power Supplies operate with the Eclipse 5 Readjust the Pulse Dose Sensitivity setting between patients as needed Readjust the Rise Time setting between patients as needed Replace 9V battery Check Universal Cart for functioning wheels and telescoping handle Replace as needed Read and record hour meter Check flow rate purity Pulse Dose Mode and alarm functions Replace compressor intake filter Perform electrical safety test Required only if used in a hospital or institutional setting not required for home use
69. move any accumulation of dust or debris prior to attaching the covers After reinstalling the cover verify that the rubber gasket is installed correctly Use mild detergent solution to clean the cabinet Control Panel and power supplies Turn OFF the Eclipse 5 and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the cabinet Control Panel or power supplies Use a damp not soaking wet cloth or sponge Spray the cloth or sponge with a mild detergent solution to clean the cabinet and power supplies To disinfect the Eclipse 5 use Lysol Brand II disinfectant Proceed as directed by the manufacturer WARNING Unplug Power Cords AC or DC Power Supplies before cleaning the exterior cabinet DO NOT use denatured alcohol or apply liquid spray or aerosol cleaners PN 20631679 Rev D 52 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT PROVIDER SERVICE AND MAINTENANCE RECORD Whenever maintenance or service is performed on an Eclipse 5 unit an entry should be made in the service log for that concentrator or recorded in accordance with your company s standard procedure Whenever the case of the Eclipse 5 is opened the flow rate purity and alarm status should be verified per the Test Procedures in this manual EDAT may be utilized to record Eclipse 5 device status and dates of service NOTE Use the Save As function under the File sec
70. ncrease Decrease Pulse Dose Mode and No Smoking buttons and observe that the Eclipse 5 buttons function normally 5 Press the Delivery Mode button and set the Eclipse 5 to the patient s normal Pulse Dose setting Confirm that without breathing from the unit after 15 seconds of no breath detected the system changes automatically to Continuous Flow Mode The green Delivery Mode light is blinking fast indicating the system is delivering a Continuous Flow After another 15 seconds the system stops delivering Continuous Flow and waits 15 seconds trying to detect a breath The cycle continues until a breath is detected 6 After 5 minutes of operation block the Cannula Fitting Outlet with your finger for 2 minutes and confirm that the visual and audible alarm occurs Unblock the Cannula Outlet Port and let it stabilize until the alarm stops 7 Disconnect the AC Power Supply and allow the Eclipse 5 to run for about one minute Set the Eclipse 5 to the patient s normal Continuous Mode setting Confirm that the Power Cartridge Status Gauge is illuminated and External Power Present Indicator is off 8 Remove the Power Cartridge and verify that the Eclipse 5 alarms and red Alarm Indicator is flashing and the alarm sounds After about 5 seconds re install the Power Cartridge and connect the AC Adapter and observe that the Eclipse 5 automatically restarts Confirm that the External Power Present Indicator is illuminated 9 Turn off Eclipse 5 and unplug
71. nd CAIRE recommends checking those requirements prior to making a trip TRAVELING BY CRUISE SHIP When traveling by watercraft instruct your patient to inform the cruise line that the Eclipse 5 will be used onboard Each cruise line has a Special Needs coordinator that can answer any questions regarding travel and the cruise line s individual requirements Have your patient contact the cruise ship Special Needs Coordinator directly TRAVELING BY TRAIN When traveling by train instruct your patient to inform the rail line in advance that the Eclipse 5 will be used onboard Operate the Eclipse 5 only from the Power Cartridge The patient must ensure that they have an adequate number of spare Power Cartridges to last for the duration of their trip PN 20631679 Rev D 34 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT ECLIPSE 5 MAINTENANCE WEEKLY MAINTENANCE PATIENT Training your patient to maintain the Eclipse 5 properly will lead to longer service intervals and lower maintenance costs Train your patient to perform the following procedures CLEAN THE AIR INLET FILTER The Air Inlet Filter located at the rear of the unit must be cleaned at least once a week To clean the filter 1 Remove the filter from the back of the cabinet 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to
72. nd replace the filter if neces sary 2 Ensure the Eclipse 5 is in a well ventilated area Make sure there are at least 3 inches between the back of the Eclipse 5 and any obstructions furniture curtain etc 3 If operating in the car be sure the back of the Eclipse 5 is facing outward in the seat 4 Verify that the internal filters HEPA Compressor Intake are clean and not restricted If necessary replace filters 5 If the condition persists advise patient switch to an alternate source of oxygen and contact Chart Technical Support for service 1 Remove the power cartridge and disconnect the AC or DC power supply from the Eclipse 5 The red light should disappear when the power sources are removed Wait approximately 20 seconds reconnect power sources and attempt to power on the Eclipse 5 again If the Eclipse 5 does not power on proceed to step 2 2 Inspect the external AC or DC power supply and ensure that its connections are secure at the Eclipse 5 the transformer and the external power outlet 3 If possible attempt to use another electrical AC or DC outlet to power the Eclipse 5 4 Verify that the status LED on the external power supply AC or DC and external power LED on the Eclipse 5 are lit If they are not attempt to use another external AC or DC power supply to see if the problem persists 5 If the condition persists contact Chart Technical Support for service 1 Reset the Eclipse 5 by removing the power
73. ng icon until the PS value is displayed 4 Press the or buttons to raise or lower the sensitivity 1 most sensitive 3 least sensitive Raising the sensitivity will require additional inspiratory effort to trigger the oxygen pulse Lowering the sensitivity will require less inspiratory effort to trigger the oxygen pulse Too low of a setting may result in false triggering 5 Verify that the patient is able to adequately trigger the oxygen delivery while speaking sitting quietly and walking Steps 1 6 may be repeated until the patient is able to adequately trigger oxygen delivery WARNING As with all conserving devices the Eclipse 5 may not be able to detect some respiratory efforts The Eclipse 5 requires a minimum of one minute to stabilize after a change After a setting change always wait at least one minute before determining if another adjustment is necessary PN 20631679 Rev D 29 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WH autoSAT ADJUSTING RISE TIME The adjustable Rise Time feature on the Eclipse 5 was designed for patient comfort The Rise Time feature adjusts flow and speed of bolus delivery and determines how quickly the patient receives their bolus volume while in Pulse Dose Mode The Eclipse 5 offers delivery settings of FAST MEDIUM and SLOW The factory default setting is FAST A MEDIUM or SLOW rise time may be appropriate for certain patients and should be
74. nications equipment and the Eclipse 5 The Eclipse 5 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The cus tomer or the user of the Eclipse 5 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Eclipse 5 as recommended below according to the maximum output power of the communications equipment Rated maximum out Separation distance according to frequency of transmitter put power of transmit m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people PN 20631679 RevD 12 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT How The Eclipse 5 Works INTRODUCTION The Eclipse 5 Personal Ambulatory Oxygen System with autoSAT Technology is a portable medical device used to ex
75. nk 1 Follow the steps to Removing the Unit Cover 2 Remove the Front Cover Assembly completely disconnect outlet tubing and front membrane ribbon cable Set aside 3 Locate the Product Tank and cut the cable ties from the ATF to the product tank Bottom and O2 out from the outlet connector Top Remove tubing from ATF to lower product tank barb See Figure 36 Important Cap off the ATF hose barb with a black hose barb cap Figure 36 Cut Cable Ties 4 Locate the 2 screws that hold the securing straps and the product tank in place remove the screws Set aside for re installation Figure 37 Figure 37 Remove Screws 5 Lift out the old product tank Set aside for RMA return or scrap CC f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual autoSAT Install new Product Tank 1 Locate all parts needed Product Tank Product Tube from ATF correct version Securing arm and screws See Figure 38 Figure 38 New product tank 2 Position the product tank with the 2 hose barbs facing in the direction of the ATF Refer to Figure 39 Figure 39 Product tank and hose barbs PN 20631679 Rev D 68 3 Install the securing arm over the product tank and locate the screw hole on the back case cover to secure the product tank Install securing arm and install two screws Figure 40 Install securing arms 4 Locate the correct Product Tube remove black ATF outlet barb cap and attach the
76. nsure proper assembly and functionality of the Eclipse 5 after it has been reassembled the following steps should be followed 1 Install the Power Cartridge into the Power Cartridge compartment of the Eclipse 5 Plug the AC Power Supply into the wall outlet and connect it to the External Power Connector of the Eclipse 5 2 Press the ON button and set the Eclipse 5 to 2 LPM At initial start up Eclipse 5 units light the Green Yellow and Red LED s accompanied by a beep As concentration increases to 70 the Red LED turns off and the Yellow and Green remain on When the concentration reaches normal operating range above 85 the Yellow LED will turn off The Green LED will stay lit indicating normal operation Table 3 shows the normal start up operating conditions for Eclipse 5 Oxygen Systems If LED is not green within 7 minutes there may be a problem with the Eclipse 5 areen eaw Red Indicator Anabi Alarm code Operating Condition Indicator Indicator Alarm off off Blinking Beeps 004 Purity lt 70 NeT Purity between 70 and off Blinking off off 008 35 On off off off 000 Normal Operating Condition 3 Verify that the Power Cartridge is charging as indicated by the Power Cartridge Status Gauge is moving in waterfall fashion scrolling from bottom to top If the Power Cartridge Status Gauge is not moving verify that the Power Cartridge is properly engaged 4 Check the Control Panel by pressing each of the I
77. on distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site sur vey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following sym bol A NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Eclipse 5 is used exceeds the applicable RF compliance level above the Eclipse 5 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the Eclipse 5 Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m PN 20631679 Rev D 11 CLL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITis autoSAT Recommended separation distances between portable and mobile RF commu
78. ontrol Panel and Power Supplies sss 35 Monthly Maintenance Patient css 36 Care for the Power Cartridge 36 Calibrating the Power Cartridge wees 36 Patient Training Checklisteisersisissinusansasum 37 Annual Maintenance Provider e ssesssssessossssosesse 38 ligsicete LU ai Nnriiccsnnanann EEA 38 Annual Maintenance Check List 38 Checking and Replacing the Power Cartridge 38 Eclipse 5 Monthly Run Time CYCle cessssssssssussssseennsee 38 Annual Maintenance Procedures s s s 39 Removing the Unit CGV ET ccssscrssesessscetacniamiecsaasnes 39 Remove and Replace 9 Volt Battery esses 43 Remove and Replace HEPA Filter 44 IVES Al FIER sc siisnnacinaeatian sen aiei anemone 44 Remove and Replace the Compressor TIAtAKE FIILOL ecccscsssseccscsssccscecesescssscscescssssnssccessessnsssseeesenssreess 46 Reinstalling or Replacing the Unit Cover 48 Test Procedu eS scscsdsscsccscsecccsccedssiatescscescessssecetcuacactadbivsutesaachectetteatet 49 Purity and Flow Rate Test Procedure Preferred Method sessssssseesssssseusses 49 Purity and Flow Rate Test Procedure Alternate Method 49 Assembly and Alarm Verification Tests ssssseeseen 50 Record Hours of Operation amp Software Version 5 1 Electrical Safety TOSt ssssssssssssssssesesssssesesssssseesssssseses 52 Cleaning the ECMO SSS acces acenssccanrcctdvalenrnterniaciitecoteues 52 Provider Service and Maintenance RECOM 53 Shipping and Tr
79. pose the device to extreme heat cold or humidity STORING THE ECLIPSE 5 Heat and humidity may degrade performance or severely damage the Eclipse 5 Store the device in a cool dry protected area away from high temperatures moisture and humidity Remove the Power Cartridge when storing the device The Eclipse 5 must be turned on and run for 2 hours each month to ensure proper operation performance Tip Running the power cartridge down completely and recharging it will provide two different features for the Eclipse 5 1 Ensures the battery cartridge is calibrated 2 Running the Eclipse 5 between uses while in storage allows the user to know if service may be needed prior to needing the device DISCARDING Power Cartridge Disposal The Power Cartridge can be recycled and should not be thrown into the trash Contact the local city or town offices for instructions on proper disposal of the Power Cartridge Alternately CAIRE may be contacted for the Power Cartridge disposal Eclipse 5 Oxygen System Disposal Local environmental laws may prohibit disposal of electrical and or electronic equipment suchas the Eclipse 5 AC Power Supply or the DC Power Supply Contact the local city or town offices for instructions on proper disposal of electrical or electronic equipment Alternately CAIRE Inc may be contacted for disposal information PN 20631679 Rev D 54 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia a
80. power quality should be that of a typi cal commercial or hospital environment If the user of the Eclipse 5 requires continued operation during power mains interruptions it is recommended that the Eclipse 5 be pow ered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typical location ina typical commercial or hospital environment PN 20631679 RevD 10 CC 4 p SA Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT Guidance and manufacturer s declaration electromagnetic immunity The Eclipse 5 is intended for use in the electromagnetic environment specified below The customer or the user of the Eclipse 5 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF 3 Vrms Portable and mobile RF communications IEC 61000 4 6 150 kHz to 80 MHz equipment should be used no closer to any part of the Eclipse 5 including cables than the recommended separation distance cal Radiated RF 3V m culated from the equation applicable to the IEC 61000 4 3 80 MHz to 2 5 GHz frequency of the transmitter Recommended separation distance d 1 2 NP d 1 2 JP 80 MHz to 800 MHz d 23JP 800 Mz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separati
81. pplies If dropped or damaged verify unit performance The Eclipse 5 will not detect a cannula that has been disconnected from the Oxygen Outlet Port ATTACH THE UNIVERSAL CART Your patient will enjoy the mobility offered by the Eclipse 5 Universal Cart in and outside the home Show your patient how to perform the following important activities e Mounting the Eclipse 5 to the Universal Cart Extending and collapsing the telescoping handle CAUTION A Do not lift the Eclipse 5 by the cart handle USING AROUND THE HOUSE Your patient may use 50 tubing when operating the Eclipse 5 in the house In order to use extended tubing the device must be in Continuous Flow Mode When a humidifier is used the tubing between it and the patient must not exceed 7 CAUTION When using a humidifier adapter of any kind the Eclipse 5 unit must remain stationary meaning that the unit must NOT by moved or AN transported in any manner TRAVELING BY VEHICLE Ask your patient whether he or she will be using the device to travel by vehicle If so show your patient how to safely use the Eclipse 5 while driving Refer to the previous section on Connecting the DC Power Supply for proper setup of the Eclipse 5 system in a vehicle Whenever possible go to your patient s vehicle to demonstrate this NOTE procedure Figure 9 Eclipse 5 DC Power Supply PN 20631679 RevD 32 E F lipse Personal Ambulatory Oxygen System Provider Technical
82. rent Eclipse 5 alarm code on the LCD This code may be used to help diagnose conditions indicated by the alert and alarm indicators The LCD will show ALRM Refer to the Level 1 Maintenance section for further information PULSE MODE SENSITIVITY ADJUSTMENT During patient setup a qualified clinician may adjust the Pulse Mode sensitivity to best suit patient inspiratory effort The LCD will show PS x PS 1 is the most sensitive setting PS 3 is the least sensitive setting Factory defaut is PS 2 HOURS OF OPERATION The Eclipse 5 hour meter provides valuable information on hours of operation Providers may display the total number of hours of operation using the Control Panel The display counts up to 99 999 hours then rolls over to 00000 The LCD will show HRS xxxxx Refer to the Level 1 Maintenance section for further information 9 VOLT BATTERY STATUS The status of the 9 Volt Battery is provided This is the actual voltage of the 9 Volt battery The LCD will show 9V x x RISE TIME BOLUS DELIVERY During patient set up a qualified clinician may adjust the Rise Time fast medium or slow i e how quickly the set bolus volume is delivered The LCD will display BOL Factory default is BOL FAST SOFTWARE VERSION Providers may occasionally need to obtain the software version on the Eclipse 5 to perform maintenance The provider may display software version numbers for both the Control Board and the Power
83. riods of non operation as a result of electrical power interruption or the need to have the Eclipse 5 serviced by a qualified technician The Eclipse 5 is not appropriate for any patient who would experience adverse health consequences as a result of such temporary interruption BASIC CONCEPT TRAINING The following section is intended to assist the provider to train the patient or caregiver to use the Eclipse 5 Topics in this section should be discussed with each patient prior to release for use of the Eclipse 5 SAFETY GUIDELINES AND OPERATIONAL SAFETY WARNINGS amp CAUTIONS Provider should review all safety guidelines and operational safety Warnings Cautions with each patient In addition provider should carry out a risk assessment prior to installation of the Eclipse 5 to assure proper connection and compatibility with other equipment the patient may be using PN 20631679 Rev D 26 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WIH autoSAT away from the oxygen system and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty Keep the Eclipse 5 at least five 5 feet 1 5 m from heat sources sparking objects or open flames WARNING No Smoking or Open Flames For safety concerns all possible sources of ignition must be kept LOCATING THE ECLIPSE 5 FOR PROPER
84. se of the Eclipse 5 Oxygen System only operate the Eclipse 5 from the Power Cartridge DO NOT use DC or AC Power Adapters on an aircraft The patient must ensure that they have an adequate number of spare Power Cartridges to last for the duration of their trip During taxi take off and landing the Eclipse 5 must be turned off and stowed under the seat or in another approved stow age location as to not block the aisle way or the entry way into the row if the Eclipse 5 will not be used During taxi take off and landing the user must be in a seat location that does not restrict any other passenger s access to or use of any required emergency or regular exit or the aisle s in the passenger compartment of the aircraft if the Eclipse 5 is used User s are not permitted to be seated in an exit row if using the Eclipse 5 If the Eclipse 5 is used when decompression of the cabin occurs and the oxygen cabin oxygen system deploys then the user is to use discontinue use of the Eclipse 5 and use the aircraft supplemental oxygen The Eclipse 5 unit is to be turned off after securing the aircraft supplemental oxygen Prior to travel the user needs to inspect the Eclipse 5 to ensure in good operational condition Visit our website www sequal com to the Travel Approved section for helpful travel tips forms used to assist with traveling and estimated number of Power Cartridges batteries needed for flights Each airline has their own requirements a
85. ss before tightening the Cable Tie Plug in the buzzer PN 20631679 Rev D 72 EA F f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual autoSAT CONTROL BOARD CONNECTOR DIAGRAM Use the figure below as an aid to ensure proper connection of wire harnesses to the Control System printed circuit board Air Inlet ge i TIELA Power Temperature pe S nee ee me Manager PC Board C Communications yy Proportional Valve Connector Buzzer Connector a Figure 46 Control Board Connector Diagram PN 20631679 Rev D 73 CLL f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual aUutoSAT Remove and Replace Power Manager Printed Circuit Board 1 Remove and Replace Combo Power Manager board set WARNING l l Disconnect power supplies and remove Power Cartridge before i removing the unit cover Do not touch exposed circuits during maintenance without ESD protection CAUTION The Eclipse 5 contains electrostatic sensitive components Do not open or handle except at a static free workstation 1 1 Remove the Power Cartridge and unscrew the 4 screws PN 6974 312 SEQ holding the BBB pcb to the case shown in Fig 47 Fig 47 Screws 1 2 Remove the Unit cover as described in section Remove and Replace the Unit Cover 1 3 Disconnect the 5 wire harnesses as shown in Figure 48 Compressor Cable External Power Cable PAY Pma A aAA S Cooling Fan Cable NA
86. t NOTE Battery times will decrease with higher bolus size breath rate ambient temperature Battery age and use over time The Power Cartridge packaged with the Eclipse 5 is not fully charged Before using the Eclipse 5 Oxygen System for the first time the Power Cartridge must be fully charged Store the power cartridge in a cool dry location Do not leave the Eclipse 5 or Power Cartridge in a vehicle or trunk during a hot or cold day When checking the Eclipse 5 as baggage on a commercial airline flight for international travel remove the Power Cartridge and properly package When shipping the Eclipse 5 for any reason remove the Power Cartridge from the Eclipse 5 If the Power Cartridge gets too warm charging will not begin until the Power Cartridge sufficiently cools Consider removing the Power Cartridge to allow for faster cooling The Power Cartridge operating time is longer if the Eclipse 5 is operated in Pulse Dose Mode Refer to Battery Consumption Chart above The typical time to recharge the Power Cartridge to achieve 80 capacity from a fully discharged Power Cartridge is 1 8 hours to 5 0 hours dependent upon the flow setting The capacity of the Eclipse 5 Power Cartridge is determined by electronics and the Eclipse 5 software PN 20631679 RevD 22 C E F f p SA4V Personal Ambulatory Oxygen System Provider Technical Manual autoSAT WARNING WARNING DO NOT tamper with disassemble crush or h
87. tery charge When the battery is being charged the charge indicator bars will blink in a waterfall type fashion If the battery is not installed or if it is improperly installed the Power Cartridge battery Status Gauge will not be illuminated Type B Applied Part degree of protection against electric shock External Power Indicator When the Eclipse 5 is properly plugged in and is using the AC or DC Power Supply this indicator will appear on Use no oil or grease the User Control Panel DCVoltage No open or naked flames Drip Proof Equipment IPXO0 The Eclipse 5 enclosure does not provide protection against the harmful effects of the ingress of liquids IPX0 No serviceable parts insde Do not open cover per IEC 60529 F Certified for both the U S and Canadian markets to the Read user manual before operation See user manual for instructions applicable U S and Canadian standards This symbol is to remind the equipment owners to return it toa This device complies with the requirements of Directive recycling facility at the end of its life per Waste Electrical and Electronic 93 42 EEC concerning medical devices It therefore bears Equipment WEEE Directive the CE marking as shown Name and address of manufacturer Authorized representative in the European Community Buzzer An audible alarm or buzzer is used to alert you to the operating condition of the device either a warning or failure and to confirm a vali
88. the AC Power Supply 10 Record results initial and date the Service and Maintenance Record 11 EDAT may be utilized to record Eclipse 5 device status and dates of service PN 20631679 Rev D 50 CLL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITi autoSAT RECORD HOURS OF OPERATION amp SOFTWARE VERSION To help maintain the Eclipse 5 you may obtain the Total Hours of Operation and software version numbers for the Control Board and the Power Manager Systems by following the steps below Data Output Form Example Alarm Code If connected to the AC or DC Power Supplies press the No Smoking icon on the control panel and advance to the following display mode for each time the icon is depressed Pulse Sensitivity Hours of Operation Alarm Code ALRM Displays most recent alarm code Additional alarm codes will also be displayed if present by 9 Volt Battery continuing to press the key Status Volts e Pulse Sensitivity PS e Hours of Operation HRS 9 volt Battery Status 9V Rise Time Rise Time BOL Control Board Software Revision Power Manager Software Revision Control Board Software Revision Power Manager Board Software Revision Log the Results If external power is not available and the power s cartridge is installed start by pressing the ON OFF key for 1 second NOTE The display accrues to 99 999 If no key press is sensed within 5 seconds th
89. ting to achieve 80 capacity power Canmildge nes verde time from a fully discharged Power Cartridge PN 20631679 RevD 7 LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WH autoSAT Independent Safety Testing Eclipse System and Eclipse Concentrator IEC 60601 1 1988 A1 1991 A2 1995 Corrigendum 6 95 mally EN 60601 1 1990 A1 1993 A2 1995 A12 1993 A13 1996 Corrigenda 7 94 FCC 15B Sec 107 amp 109 EN55011 EN60601 1 2 2001 EN6100 3 2 EN61000 3 3 IEC61000 4 2 IEC61000 4 3 IEC61000 4 4 IEC61000 4 5 IEC61000 4 6 IEC61000 4 8 IEC61000 4 11 IEC 60601 1 2 2001 RTCA DO 160 Rev E AC Power Supply Model 5941 Safety IEC 60601 1 1988 A1 1991 A2 1995 Electromagnetic Compatibility Electromagnetic FCC 15B Sec 107 amp 109 EN55011 EN60601 1 2 2001 EN6100 3 2 EN61000 3 3 IEC61000 4 2 Compatibility IEC61000 4 3 IEC61000 4 4 IEC61000 4 5 IEC61000 4 6 IEC61000 4 8 IEC61000 4 11 EN55014 1 DC Power Supply Model 5942 Portions of IEC 60601 1 1988 A1 1991 A2 1995 Safety Electromagnetic FCC 15B Sec 109 EN55011 EN60601 1 2 2001 IEC61000 4 2 IEC61000 4 3 IEC61000 4 4 Compatibility IEC61000 4 6 IEC 60601 1 2 2001 Power Cartridge 2400 PN 7082 IEC 62133 UL60950 1 First Edition UL File MH29443 IEC 60601 1 1988 A1 1991 A2 1995 UN Transportation Tests T1 T8 Electromagnetic EN 61000 6 3 2001 EN5
90. tion and save the file as the SN and or the date serviced example 08D0110xxxx 08012000 Eclipse 5 Serial Number System Checkout Hour meter gi Date Readi Initials Service Performed Comments eading Purity Flow Alarms Table 13 Sample Eclipse 5 maintenance record aanw SS 2 JIROV LVL lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT SHIPPING AND TRANSPORTING THE ECLIPSE 5 When shipping the Eclipse 5 use original packaging if possible Always remove the Power Cartridge and cart from the Eclipse 5 prior to shipping If original packaging material is available repack the Eclipse 5 Power Cartridge cart and power supplies in the designated packaging areas If original packaging material is not available then place the Eclipse 5 in a plastic bag and surround the concentrator with a minimum of two inches of soft foam packing material or bubble wrap Wrap each accessory in a similar manner Place the Eclipse 5 and accessories in an appropriate cardboard box for shipping WARNING DO NOT expose the Eclipse 5 to water The Eclipse 5 enclosure does not provide protection against the harmful effects of liquid ingress Electrical shock or damage to the unit may result When the Eclipse 5 must be transported in a delivery vehicle simply secure the Eclipse 5 and its accessories to prevent damage Do not ex
91. tract oxygen from the atmosphere concentrate it to 87 95 6 and present the oxygen to the patient The device will operate in Continuous Flow Mode or Pulse Dose Mode In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in a bolus at the beginning of each inspiration providing a selectable range setting of 16mL to 192mL The Eclipse 5 operates from either external power or from an internal rechargeable Power Cartridge The system includes a Smart battery charger that recharges the internal Power Cartridge whenever the Eclipse 5 is connected to AC or DC power The system monitors and controls both the power source and the Power Cartridge charger Air Flow In Figure 1 Eclipse 5 System Block Diagram ATF CONCENTRATOR MODULE The Eclipse 5 uses a passive system to separate oxygen from air Air flows into the Eclipse 5 where it is filtered and then enters the compressor Pressurized air flows from the compressor into the ATF Concentrator Module where it is separated into oxygen and nitrogen components The air separation process uses a rotary valve system to force air through a series of pressurized sieve beds Through a process known as vacuum pressure swing adsorption nitrogen molecules are collected on an adsorbent material allowing the concentrated oxygen to be forced through a sieve bed into the product tank The nitrogen molecules are then purged
92. upply to recharge the Power Cartridge and should be able to run the Eclipse 5 from AC power for at least 5 hours NOTE The patient can still use the Eclipse 5 while zeroing the Power Cartridge The patient should follow these steps 1 Install the Power Cartridge in the Eclipse 5 2 Remove external power and operate the Eclipse 5 from the Power Cartridge 3 Allow the Eclipse 5 to completely discharge the Power Cartridge This is best performed at a setting of 2 0 LPM Continuous Flow Ten to fifteen minutes before the end of the discharge cycle the Low Power Cart ridge alert will sound followed by the Power Cartridge Shutdown alarm If using the Eclipse 5 while discharging the Power Cartridge DO NOT deviate from the patient s prescription 4 Press the ON OFF button to silence the alarm 5 Allow the cartridge to cool for a minimum of 1 hour prior to recharging 6 Reattach external power Verify that the External Power Present Indicator illuminates Leave the Power Cart ridge installed Note that the Power Cartridge Status Gauge may not indicate charging immediately When the Power Cartridge reaches proper temperature charging will begin automatically 6 Allow Power Cartridge to fully charge as indicated by the Power Cartridge Status Gauge If the Power Cartridge temperature rises above 45 C 113 F the Power Cartridge will not charge If the Power Cartridge temperature rises above 70 C 158 F the Power Cartridge wi
93. ure Sensor Failure FAIL A1 Verify functionality of CS PCB Breath Pressure Sensor Failure FAIL A2 Verify functionality of CS PCB Ultrasonic Failure FAIL A3 Verify functionality of CS PCB Check for water intrusion from humidifier Product Temperature Sensor Failure FAIL A4 Verify functionality of DB9 cable harness Verify functionality of CS PCB Loss of Product Tank Pressure FAIL A5 Check for leaking cracked HEPA filter or damaged tubing Verify ATF motor is turning Invalid Assembly Option Remove Power to Reset Step 1 Remove all power sources battery and external power Wait 20 seconds Reconnect power to the Eclipse 5 Turn the device to the ON position If resolved continue use Step 2 If unresolved call Chart Technical Support for assistance PN 20631679 Rev D 60 CC lipse Personal Ambulatory Oxygen System Provider Technical Manual autoS INE System Schematics and Diagrams SIMPLIFIED BLOCK DIAGRAM Atara Prose Tare i EOR Eclipse Il System Enclosure amevssuusens rawn 73 135 5S surr re DE sonssasssnassensa UART Liege BLDC FETS amp Driver m z Pewee Manager a ji DeepCore ff 1 wien I fs Pare eS A n Mias i J Cecuts a i FMN power Manager PM Board l To A ERTA Power acana hmla f IN Ve IOW Opt 4 3 100 260 VAC SOMO HI Je Ll S g sss D lt ret 195005 00 16 Amen Figure 28 Eclipse 5
94. utoSAT Troubleshooting Service and Repair Procedures CAUTION The Eclipse 5 contains electrostatic sensitive components Do not open or handle except at a static free workstation Do not remove cover without ESD protection To adequately troubleshoot and repair the product in the field EDAT is recommended NOTE Eclipse 5 Data Acquisition Tool EDAT PN 5535 SEQ PN 20631679 RevD 55 E F f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual WITis a autoSAT SYSTEM TROUBLESHOOTING AND ALARMS PROVIDER TROUBLESHOOTING TABLE po nor IGNORE ALARMS Yellow Light ie yell igiit sansoli 1 Remove power cartridge and re install to ensure that it is secure in the concentrator There is no audible alarm The Eclipse 5 has detected a Tn ie sa x 001 or 100 2 If possible install a separate power cartridge in the Eclipse 5 to verify if the condition occurs This alarm may occur with roblem with the connection y y P with multiple batteries me Edipeo 3 aisle ne tothe power caeuiage 3 If this problem persists contact Chart Technical Support for service battery is charging as well 1 Re attach the Eclipse 5 to external power or install a fully charged power cartridge in the The yellow light is flashing Eclipse 5 Allow warm power cartridge to cool outside the concentrator for 30 minutes re The Eclipse 5 power cartridge The alarm is giving 1 beep iwa 2 Re charge the power cartridge usin
95. verlay cable by pulling on the ribbon cable Membrane Panel Overlay Cable Connection Figure 14 Removing the Membrane Panel Overlay Ribbon Cable 7 Disconnect the silicone tube from the oxygen outlet tube at the top of the cover First cut the cable tie then pull the silicone tube off the oxygen outlet tube Avoid possible eye injury by wearing protective eyewear or shielding the eyes from possible flying debris PN 20631679 Rev D 41 C eclipse WITis Personal Ambulatory Oxygen System Provider Technical Manual autoSAT Cut Cable Tie Figure 15 Disconnecting the Oxygen Outlet Tube 8 Lay the front panel away from Eclipse 5 Inspect PEM Nut anchors in the front cover assembly Perform necessary maintenance Figure 16 Front Cover Removal Once the cover has been removed the procedures listed in this section may be performed PN 20631679 RevD 42 CLL f p ke Z Personal Ambulatory Oxygen System Provider Technical Manual autoSAT REMOVE AND REPLACE 9 VOLT BATTERY Replace the 9 volt battery when the unit beeps three times at the end of power on self test when voltage is less than 7 0 Volts or during annual PM To replace the 9 volt battery follow these steps Disconnect power supplies and remove Power Cartridge before removing the unit cover Do not touch exposed circuits during maintenance without ESD protection 1 Remove the Eclipse 5 front cover assembly 2 Lift
96. when using the Eclipse 5 ina medical care facility Contact CAIRE Inc to obtain a hospital grade AC power cord for the Eclipse 5 if necessary PN 20631679 RevD 30 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual WITia autoSAT CONNECTING THE DC POWER SUPPLY To install the DC Power Supply follow these steps ADC Power Supply allows the system to operate from DC outlets such as those found in motor vehicles CAUTION The DC Power Supply is designed for 12VDC minimum vehicle electrical systems Do not attempt to operate with 6V 24V or other vehicle electrical system 1 Consult your patient s vehicle owner s manual for DC Accessory outlet requirements 2 Secure your Eclipse 5 and DC Power supply in your vehicle Ensure that alerts and alarms are observable 3 Start the vehicle s engine 4 Insert supply plug into DC accessory outlet 5 Attach the Power Supply plug to the Eclipse 5 6 Advise the patient to use Pulse Dose Mode if prescribed by their clinician 7 When the device is properly connected and receiving power from the DC power source a green indicator light on the DC Power Supply will illuminate Ask your patient whether he or she will be using the device to travel by vehicle If so show your patient how to safely use the Eclipse 5 while driving Refer to the previous section on Connecting the DC Power Supply for proper setup of the Eclipse 5 system in a vehicle WARNING Ensure
97. y Status v Continuous Flow Setting v Rise Time v Hour meter v Software Revision v Pulse Dose Sensitivity v Power Cartridge Status Gauge The Eclipse 5 is shipped from CAIRE Inc at default flow settings of 2 LPM Continuous Flow 2 0 Pulse Dose and 2 for Pulse Mode Sensitivity Adjustment and FAST Rise Time You may adjust these settings to your patient s prescription when you deliver and set up the device PN 20631679 RevD 25 E F lipse Personal Ambulatory Oxygen System Provider Technical Manual aUutoSAT INDICATIONS FOR USE The Eclipse 5 is indicated for the administration of supplemental oxygen The device is not intended for life support nor does it provide any patient monitoring capabilities A physician must prescribe a specific oxygen flow rate setting to meet patients individual needs Recommended oxygen flow rates should be adjusted only under the advice of a physician WARNING Federal law restricts this device to sell by or on the order CONTRAINDICATIONS WARNING The Eclipse 5 is not intended for life supporting or life sustaining applications nor does it provide any patient monitoring capabilities WARNING In certain circumstances the use of non prescribed oxygen can be hazardous This device should only be used when prescribed by a AN physician WARNING A Not for use in the presence of flammable anesthetics WARNING As with any electrically powered device the user may experience pe
98. y connect the 9V battery wiring harness between the exhaust amp battery compartment and 3 Compressor wiring harnesses to their connectors Fig 52 Routing Wires 1 8 Install the Cooling Fan PN 1074 SEQ with the 2 screws PN 6961 210 SEQ and lightly tighten as shown in Fig 53 Apply Loctite 425 thread locker on the tips of the threads before installing Note Ensure the cooling fan is blowing toward the circuit board as indicated by the arrow on the side of the fan Fig 53 Cooling Fan placement 1 9 Turn the device over and align the BBB with the holes in the case Screw the 4 screws into the Battery Bridge Board from the Battery compartment side securing it to the case snug screws 1 10 Turn the device back over Install the 2 clips that come with the power manager board set Note There is a right and left clip Large gap is on the right ensure you get both the metal heat sink and pcb corner in the black clip 1 11 Install the Unit s front cover according to the Remove and Replace front cover section This holds the Power Manager pcb corner clips in place no screw required ON F Paa n Fig 54 Power Manager installed E F f p SAV Personal Ambulatory Oxygen System Provider Technical Manual autoSAT Compressor As with any concentrator the compressor is a limited wear component and may require servicing during the lifetime of the device The point of service will be dependent on factors such
Download Pdf Manuals
Related Search
Related Contents
Teledyne LLG User's Manual IPL4000 Sony DSLR-A700 User's Manual Montageanleitung JVC HR-E249E User's Manual Origin Storage 256GB MLC SATA 2.5" ONR User Manual - Central Bank of Ireland PSD-30 - TechnoMad RFX9600 Starter`s Guide H5D-50c Wi-Fi Datasheet Copyright © All rights reserved.
Failed to retrieve file