        Appendix Quick Installation Guide for TXR Generators
         Contents
1.                                                   Table   Generator     PROGENCY TRUE SPEED BUCKY  pre wired    Terminal Strip   TB3  on board in center   Wire Terminal Strip Wire Color   of cabinet  Color   Left Side of Strip  marked table    B1 Blue 1 Blue   B2 Orange 2 Orange   B3 White 3 White   B4 Yellow N Yellow   B8 Brown L Brown   Any labeled earth ground in cabinet Green Ground Green   Terminal Strip     TXR 96  mounted      Table Lock Printed Circuit Board   right side of control cabinet    Left Table Tab Red 24 VAC Red   Right Table Tab Black 24 VAC Black  Note  Red and Black continue through bucky  junction box  Both terminate at table lock printed  circuit board    Spare Purple Spare Purple   Wall Stand   Generator     PROGENCY TRUE SPEED BUCKY  pre wired    Terminal Strip   TB3 Wire Terminal Strip Wire Color   Right Side  marked wall  Color   B1 Blue 1 Blue   B2 Orange 2 Orange   B3 White 3 White   B4 Yellow N Yellow   B8 Brown L Brown   Any labeled earth ground in cabinet Green Ground Green   Terminal Strip     TXR 96  mounted on Wall Stand Lock Printed Circuit Board   right side of control cabinet    Left Table Tab Red 24 VAC Red   Right Table Tab Black 24 VAC Black  Note  Red and Black continue through bucky  junction box  Both terminate at wall stand lock  printed circuit board    Spare Purple Spare Purple                Technical Pamphlet W 2009 7 20 31              Tingle X Ray Products  1    Technical Pamphlet    TXR Standard Frequency Generators   Pictur
2.                2  tube 2   Bucky 2   no lon Chamber        1 0 0   tube 1   No Bucky   no lon Chamber     No Bucky          AP 0015R2    11    HF Series Generators  Appendix   Quick Installation Guide for TXR Generators       6  Continue Configuration for the remaining Workstations     T  Exit from configuration by pressing the  200  and    800    pushbuttons  down simultaneously for at least two seconds until you hear an audible  double beep  This sound signals that you have successfully saved the  workstations you had selected  Configuration of the Workstations is  now completed     2 3 CONFIGURING THE TUBE TYPE    T To configure the X ray Tube type  it is necessary to enter into Calibration  Mode by pressing the  200  and    800    push buttons simultaneously  The  light of the selected workstation blinks  Bucky 1  Bucky 2 or No Bucky                    kVp mAs mA    EG                Time    off On            Bu    AEC Film Speed  g         FocusS FocusL    Ready X Ray    AE    PREP EXPOSE                   Density            faa   WARNING                   T ACA A             MEET    RE CLTC NC REDE      2  Press the  mA Increase  push button until you reach the maximum mA  capacity  Press the  mA Increase  push button twice more to reach the  Extended Memory Code    E02     that will appear on the mA Display     Density 3  In order to configure the X ray Tube type and model  press the         or          Density push buttons until one of the Tube Type numbers listed bel
3.           REGION1 VIEW3 PROJECT1    MINIMUM THICKNESS  3 cm  DEFAULT THICKNESS  5 cm    THICKNESS STEP  cm    ENTER TECHNIQUE          28    AP 0015R2         Series Generators       Appendix   Quick Installation Guide for TXR Generators    10  Modify the Radiographic Technique Values  In order to modify the  Radiographic Technique values for a specific Region View Projection   press the Display Selector close to    ENTER TECHNIQUE        d g  REGIONI VIEW3 PROJECT1 m  g  MINIMUM THICKNESS  3 cm       DEFAULT THICKNESS  5 cm    Enter Selection and  go to next screen THICKNESS STEP  cm Go to previous screen    6 ENTER TECHNIQUE  lt             The Display will show the Region  View  Projection and the Default  Thickness  highlighted  for the selected Region View Projection  Press   Enter  to edit the APR technique for the displayed Thickness  or  before pressing    Enter    modify the Thickness with the    Up Down  Edit Arrows  ifthe APR technique to be modified corresponds to another    Thickness value   REGIONI VIEW3 PROJECT1      Chahge    TECK  I   Thickness value  CANCEL        Cancel and go to    previous screen       After pressing  Enter   the Console automatically displays all the  radiographic parameters  kVp  mA  mAs  Time   selections  Workstation  and AEC  and SID indication related to that Region View Projection     Now  the operator can modify and store the Radiographic Parameters   Workstation  AEC Selections and SID indication  36   40   60  or 72   selected w
4.    Mn CLR PORE    mh f           SW1 Jumper          T1 Terminal Strip          2 AP 0015R2         Series Generators       Appendix   Quick Installation Guide for TXR Generators    4  TXR products require 24 VAC for locks  At top of the Generator cabinet  a 24 VAC Distribution Board has been added to connect these locks     rte        us      s   E   S Rn    E  BELA A Em  4 5 6 7 8  24VAC OVAC  24VDC OVDC  24VDC OVDC    Terminals 1 and 2 are in coming 24VAC  Collimator is connected to terminals 3 and 4       5  Remove Allen screws that secure hinged door on MODS side           AP 0015R2 3    HF Series Generators  Appendix   Quick Installation Guide for TXR Generators       6  The Terminal Strip 4TS2 is located on Module 4  back of Module 3  of the  hinged door  Connect the X ray Tube Rotor wiring as follows     WIRE FROM X RAY TUBE CONNECTION IN 4TS2    BLACK       RED       WHITE       THERMAL       THERMAL       GROUND            f X ray Tube have not a shielded Ground wire  terminal 4TS2 6 is not used     4TS2 Terminal Strip          T  Carefully connect the J5 side of the Serial cable to the Console and then  route the Handswitch wire  Connect the J1 side of the Serial cable from  the Console to the Generator Cabinet     Connection of M  A  Connection of Serial Cable  Serial Cable du TOREM UT Ics at Generator Cabinet  at Console P t      located on upper back           4 AP 0015R2         Series Generators  Appendix   Quick Installation Guide for TXR Generators       8  
5.   3 contains eight Dip Switches  for factory use      Swicth  2       Swicth  3       Swicth  1             Configuration of Switch  1       For use with 60 Hertz Line Supply  All four Dip Switches must  be placed in the down  OFF  position        For use with 50 Hertz Line Supply  The first Dip Switch must be  placed in the up  ON  position  The second  third and fourth Dip  Switches must be placed in the down  OFF  position     Configuration of Switch  2  The first  third and fourth Dip Switches  must be placed in the up  ON  position  The second Dip Switch must be  placed in the down  OFF  position     Configuration of Switch  3  All eight Dip Switches must be placed in  the down  OFF  position        AP 0015R2    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    2 2 CONFIGURING THE WORKSTATIONS       With the Generator OFF  make sure Dip 3 of Switch  2 on the ATP  Console CPU Board  Module 4 side  is placed in the closed  ON  position   Service Mode enabled      Turn ON the Line Power and then the Console power  Wait for the self  testing to finish  then enter in configuration mode by maintaining pressed  Power ON     200    and    800    simultaneously for at least two seconds until  all of the workstations push buttons are illuminated     FocusS FocusL    Ready X Ray    Density      Si    S        PREP EXPOSE           WARNING  m  P  ATENCION  cs    Results you should see     A display of numbers will be visible on the kVp  mAs  mA and
6.   Section Page  1 INSTALLATION         rrr RR rev                           eee 1  2 CONFIGURATION 21    l   i kg              RR E Ruhm ee 9  2 1 Proper Dip Switch Selections on the ATP Console CPU Board              9  2 2 Configuring the Workstations                                           10  2 3 Configuring the Tube Type                                             12  3 CALIBRATION          ee t                     eed eee eee 15  3 4 Previous TaSKS    oss odere SEDI ae S oe Ros See aeos Sae eR s 15  3 2 Digital kV Loop                                                          16  3 3 Digital mA Loop Closed              sssssesssseseee eh 17  3 4  Auto calibration of Digital mA Loop Open                                17  3 5  FinalChecks           ee cee                              GR EXC ROS UE A pde 22  4 APR EDITING                    obs eus 23  5 BUCKY INTERFACE CONNECTIONS                                           31       AP 0015R2 i    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators       ii AP 0015R2         Series Generators  Appendix   Quick Installation Guide for TXR Generators       SECTION 1 INSTALLATION    THIS DOCUMENT PROVIDES A STEP BY STEP GUIDE AND IS  LT  WARNING NOT INTENDED TO TAKE THE PLACE OF THE SERVICE  MANUAL  AS STEPS ARE COMPLETED CHECK OFF THE    LIST TO THE LEFT OF EACH NUMBER OR PARAGRAPH     1  Place the Power Module close to its permanent position  For easy  interconnection and calibration place the 
7.   Series Generators       SECTION 3    3 1 PREVIOUS TASKS    Appendix   Quick Installation Guide for TXR Generators    CALIBRATION    Before calibration  keep in mind that     To calibrate and measure the kVp output  a Non Invasive kVp Meter is  needed     Place and center a Non Invasive kVp Meter on the X ray Tube output at  the required SID  refer to the Non Invasive kVp Meter documentation      To calibrate and measure the mA or mAs  a mAs Meter needs to be  plugged to the banana plug connection at the top of the HV Transformer     The cross bar  link plate  should not be connecting the  banana plug  Turn the Generator OFF to connect the mAs  meter if you did not connect one at the beginning     Remember to re install the cross bar between the banana  plug connections after removing the mAs meter     Remove the cross bar  link   and plug the mAs meter here           AP 0015R2    15    HF Series Generators  Appendix   Quick Installation Guide for TXR Generators       3  Before calibration procedure  verify that the Dip Switches of Switch  2 of  the HT Controller Board  A3000 xx  are in the    OFF    position  This board  is located on Module 3 at the hinged door            Switch  2 on the  HT Controller Board        3 2 DIGITAL KV LOOP OPEN    Extended Memory Location E06 contains the calibration factor for Digital kV  Loop Open     Note   gt  The Value in E06 Memory Location is only related to the Generator  performance  it is not related to the X ray Tube s  or another  
8.  Spot  second group      EACH TIME THAT AUTO CALIBRATION IS ACTIVATED IN A   n  WARNING mA STATION  OR IN    E01    MEMORY LOCATION   ALL THE  FILAMENT CURRENT NUMBERS OF THE SELECTED FOCAL    SPOT ARE AUTOMATICALLY SET TO    344     SO A NEW  COMPLETE CALIBRATION OF THE FILAMENT CURRENT  NUMBERS FOR THIS FOCAL SPOT WILL BE REQUIRED        AP 0015R2 17    HF Series Generators  Appendix   Quick Installation Guide for TXR Generators       Auto calibration starts with the minimum available mA station for the  selected Focal Spot at 50 kV and follows with the other combinations of  mA stations for the selected Focal Spot at 80 kV  120 kV and 40 kV     1  For this procedure  set Dip Switch 4 of Switch  2 on the HT Controller  Board back in the  ON  position  Digital mA Loop Open   Filament  Current Constant      1  Select the  No Bucky  workstation  Direct      2  Enter calibration mode by pressing the  200  and  800  push buttons  simultaneously and select one of the configured mA stations for the  Small Focal Spot  Light will be blinking for workstation selected if you  are in Service Mode          Off On        g  Bucky  ij Bucky No               Film Speed                       5                             X Ray    Density          PREP EXPOSE    7x  WARNING          CEA ATENCION  con                 18 AP 0015R2    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    3  Enter Auto calibration mode by pressing the  Power On  and  kV  increase  
9.  or Exposure Controls to stop momentarily the Auto calibration  procedure  The alarm will stop when the X ray Tube begins to cool and  recovers the Heat Units capacity  Exposures can be performed again  even though code    777    is shown on the Console     At this point  it is recommended to wait until the Heat Units available are  close to 8096 of the X ray Tube capacity without making any exposure     The Generator will try to calibrate each kV   mA combination for ten  10   attempts  maximum   If calibration is aborted  after ten attempts   code   888  will flash on the Console until the  Power On  push button is  pressed to exit Calibration  Calibration might also abort due to space  charge during calibration of the lowest kV at the highest mA stations for  the Focal Spot selected  so code  777  will be flashing on the Console  until the  Power On  push button is pressed     IF AUTO CALIBRATION IS ABORTED  CODE  888  OR     777      CONTINUE AUTO CALIBRATION FOR THE OTHER  FOCAL SPOT  CHECK AT THE END OF AUTO CALIBRATION  WHICH KV   MA COMBINATIONS HAVE NOT BEEN  AUTO CALIBRATED FOR EACH FOCAL SPOT  THESE  COMBINATIONS HAVE THE FILAMENT CURRENT NUMBER  SET TO  344    MANUALLY CALIBRATE THESE KV   MA  COMBINATIONS AS EXPLAINED IN THE SERVICE MANUAL   CALIBRATION CHAPTER    MANUAL CALIBRATION OF  DIGITAL mA LOOP OPEN         When Auto calibration is successfully performed  code    999    will be  flashing on the Console until press the  Power On  push button for 2  seconds
10.  previous screen  or press the Display Selector close to  CANCEL  to  go back to previous screen without saving the changes     d    MINIMUM THICKNESS  mes   Edi    DEFAULT THICKNESS  Thickness    CANCEL  lt a Cancel and go to    previous screen    THICKNESS STEP           AP 0015R2 27    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    Note   gt     Minimum Thickness  MT  setsthe minimum value in centimeters used  for the selected Region View Projection        Default Thickness  DT J setsthe default value in centimeters according  to the operator preferences        Thickness Step  TS    sets the number of centimeters between each  thickness value  Each Region View Projection may have a maximum of  twenty  20  Steps     The maximum range for a Thickness value is 99 cm  which means  that a Minimum Thickness value too high or a Thickness Step too  large will not be allowed if the resulting value is over 99 cm     Calculate if the Thickness values are allowed by applying the formula    MT   19  x TS    where   MT   Minimum Thickness  19   Maximum number of Steps minus one  TS   Thickness Step value    For example  in the figure below  the formula is  3 19  x 2   44 cm    In which    3    is the Minimum Thickness     19    is the number of Steps  20 1   and    2    is the Thickness Step value     44    is the resulting maximum  Thickness but if the Thickness Step value is 8  it would not be allowed  as the range is over 99 cm   84 19  x 8   168
11.  rename other Views        g  REGION1 VIEW  VIEW1 VIEW2 VIEW3 VIEW4  Edit name  VIEW5 VIEW6 VIEW7 VIEW8 Cancel and go to    previous screen    ENTER VIEW cime             Press the Display Selector close to  CANCEL  to go back to previous  screen without saving the changes or press the Display Selector close  to  ENTER VIEW  to go to step 6        AP 0015R2    25    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    Enter Selection and  go to next screen    71    Enter Selection and  go to next screen    71       Projection  Once the desired Projection  highlighted  is chosen with the   Edit Arrow  buttons  if the operator requires to modify the Projection  go  to step 7  If not  press the Display Selector close to    ENTER  PROJECTION    to go to step 8  or    BACK    to return to the previous    d d  d    PROJECT1 PROJECT2 PROJECT3    screen     PROJECT4 PROJECT5  Go to previous screen    ENTER PROJECTION    Modify the Projection name  Press  Enter  to edit the name of the  selected Projection  The first letter of the name is highlighted  the     Up Down Edit Arrow  buttons will change the letter in alohanumerical  order  plus underscore and blank   the    Right Left Edit Arrow  buttons  will scroll in the word  Once the name is written  press    Enter    to save the  new Projection name  If necessary  repeat this process to rename other    Projections        REGION1 VIEW3 PROJECT       PROJECT1 PROJECT2 PROJECT3  Edit name    Cancel and go
12.  the PC APR Editor    Refer to Section 4 of the Operating Manual      Proceed in the following way     APR 1  With the APR deactivated  press simultaneously  APR ON OFF    and   Ll   m  Enter  buttons  All the LEDs of the Edition Module buttons lights up   Enter    On Off    Edition Module with    Edit Arrow     buttons and  Enter  button    d  REGIONI Messen  oi REGION3 REGION4 m  d    REGIONS REGION6 REGION7 REGIONS    Display Selector D ENTER REGION       The Edition Module is operated with the    Edit Arrow    buttons  the    Enter     button and the Display Selectors     Press the    Edit Arrow    buttons to scroll in the APR Display  When  pressing the Display Selector on the left lower corner  the APR Display    will show sequentially the selected Region   View   Projection atthe upper  line        AP 0015R2 23    HF Series Generators  Appendix   Quick Installation Guide for TXR Generators       2  Body Region  Once the desired Body Region  highlighted  is chosen  with the  Edit Arrow  buttons  if the operator requires to modify the  Region name  go to step 3  If not  press the Display Selector close to   ENTER REGION  to go to step 4     g g  g  REGION1 EKAL REGION3 REGION4    Enter Selection and  go to next screen        vL ENTER REGION    REGION5 REGION6 REGION7 REGION8       3  Modify the Region name  Press    Enter    to edit the name of the selected  Region  The first letter of the name is highlighted  the    Up Down  Edit Arrow    buttons will change the letter 
13.  time  displays  The visible numbers will either be O  1  2  etc     The Console is equipped to operate in five different modes   workstations   The three push buttons atthe upper left provide a way to  program for different operations  See the following tables for proper  selections        10    AP 0015R2         Series Generators       4     Appendix   Quick Installation Guide for TXR Generators    Selectthe first workstation to be configured  Bucky 1  Bucky 2  No Bucky   by pressing the respective push button  only this push button blinks and  the console shows one of the following values     DISPLAY FUNCTION VALUE DESCRIPTION    No configured workstation       15 Value SELECT THE TUBE Tube 1       Tube 2       Direct  No Bucky        Bucky 1       SELECT THE DEVICE   WORKING MODE     2nd Value Bucky 2       Standard Tomo           Standard RF       No AEC       lon Chamber 1       SELECT THE ION  3rd Value CHAMBERS BEING lon Chamber 2  USED  ONLY WITH AEC        lon Chamber 3       lon Chamber 4    Notes   Some values may not be configurable  depending on Generator model    1  Only when the Tomo is controlled by the Generator                  Set the new value by pressing the Density         or         buttons located at  the bottom left corner ofthe Operator Console  Whenthe first value is  O    the second and third value are    0        The following table is an example of a 2 tube System with 2 Buckys and  no lon Chambers     Bucky 1    1  Quy 1   no lon NE       2 2 0
14.  to  revious screen    PROJECT4 PROJECT5    p  ENTER PROJECTION CANCEL ET           Press the Display Selector close to  CANCEL  to go back to previous  screen without saving the changes or press the Display Selector close  to  ENTER PROJECTION    to go to step 8        26    AP 0015R2         Series Generators       Appendix   Quick Installation Guide for TXR Generators    8  Once the Display Selector close to  ENTER PROJECTION  has been  pressed  the APR Display shows       The name of the Region View Projection  upper line         The Thickness values for Minimum Thickness  Default Thickness  and Thickness Step that might be applied to the selected Region    View   Projection     If required  these Thickness values can be edited as described below     g g  REGION1 VIEW3 PROJECT1 m  g  MINIMUM THICKNESS  3 cm    DEFAULT THICKNESS  5 cm    Enter Selection and  go to next screen THICKNESS STEP  cm Go to previous screen       ENTER TECHNIQUE  lt             9  Modify the Minimum Thickness  the Default Thickness or the  Thickness Step  If the operator decides to modify a Thickness value   follow this step  if not  go to next step           Press the  Edit Arrow  button to highlight the corresponding line and  press  Enter  button to modify the selected Thickness      MT    minimum   DT    default  or TS    thickness step  appears  atthe right corner of the Display  then select the desired number with the   Edit Arrow  buttons and press  Enter to save the changes and go back  to
15.  to exit from Auto calibration mode  The    999    code will  disappear when you have successfully exited Calibration     To calibrate the Large Focal Spot  select one of the configured mA  stations for the Large Focal Spot by pressing the mA         push button   Repeat steps 2   5     Before starting the exposures  it is recommended to wait until the Heat  Units available are closed to the 8096 of the X ray Tube capacity        20    AP 0015R2         Series Generators       10     Table 3 1  mA Calibration Numbers    Appendix   Quick Installation Guide for TXR Generators    After performing both procedures  for Small and Large Focal Spots    select in calibration mode each combination of the available mA stations  at the kV break points  40  50  80 and 120 kV   Press the    Reset     push button to read on the kV Display the new value of the Filament  Current Number stored for each combination and write down the new  values in the Data Book     Note that the highest mA station for Small Focal Spot may have numbers  larger than the lowest mA station for Large Focal Spot  This is normal     Repeat the above calibration process for the second tube   Exit calibration mode     Set Dip Switch 4 of Switch  2 on the HT Controller Board back inthe   OFF  position  Digital mA Loop Closed      FILAMENT CURRENT NUMBERS AT kV BREAK POINT       mA STATION    50 80                                                                Note   Some Generator models do not contain all the mA statio
16. Add insulating compound or insulating oil to both High Voltage  Transformer receptacles  Connect both High Voltage cables  Anode and  Cathode  from the X ray Tube to the HV Transformer receptacles     9  The Terminal Strip 3TS1 is located on Module3  out side of hinged door    Verify or connect a jumper between terminals 3TS1 22 and 3TS1 23 to  bypass the Door Interlock     Terminal Strip 3TS1          10  THIS POINT ONLY APPLIES TO LINE POWERED GENERATORS     Remove the plastic cover over the Line Fuses and connect the Line  Power cables to the right side of the Fuses at L1  amp  L2 for Single Phase  Generators and also at L3 for Three Phase Generators     Plastic Cover over Line Fuses 1 Input Transf  6T2          PREQQ UCION  Esto    GND Stud       L1 Input Line       L2 Input Line          L3 Input Line       Single Phase    Neutral Stud   optional              AP 0015R2 5    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    11  THIS POINT ONLY APPLIES TO BATTERY POWERED GENERATORS        a  Connector 6J10 on the right side of Generator Cabinet  Module 6   is unplugged for safety during transport  Plug in it     Connector 6J10       b  Connect the Line Power cable to Line Connector  wire 1 to left   wire 2 to middle  ground wire to right   These connections are  color coded     Line Connector       Circuit Breaker          6 AP 0015R2         Series Generators       Appendix   Quick Installation Guide for TXR Generators    c   TurnOnthe C
17. Appendix    Technical Publication  AP 0015R2    Quick Installation Guide  for TXR Generators    HF Series Generators         Series Generators  Appendix   Quick Installation Guide for TXR Generators       REVISION HISTORY    REVISION DATE REASON FOR CHANGE  0 JAN 7  2004 First edition  1 OCT 16  2007 General Update  amp  Improvement  2 AGO 06  2009 Dip Switch configuration of HT Controller    This Document is the English original version  edited and supplied by the manufacturer     The Revision state of this Document is indicated in the code number shown at the bottom of this page     ADVISORY SYMBOLS    The following advisory symbols will be used throughout this manual  Their  application and meaning are described below     DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT   7    ANGER   IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS     PERSONAL INJURY OR DEATH     DATA     Advise of conditions or situations that if not heeded or  avoided could cause personal injury or damage to equipment  or data     ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT  0 WARNING HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL  INJURY  OR CATASTROPHIC DAMAGE OF EQUIPMENT OR     CAUTION    Note   gt  Alert readers on pertinent facts and conditions  Notes represent  information that is important to know but which do not necessarily  relate to possible injury or damage to equipment        AP 0015R2         Series Generators       Appendix   Quick Installation Guide for TXR Generators    TABLE OF CONTENTS
18. Generator Cabinet on a table  or pedestal  or mount it to the wall using the Wall Support provided     2  Verify that wire marked          on Transformer 6T2 terminal strip is matched  to the incoming power  Identify the Generator type  Single Phase   Three  Phase or Battery Powered     POWER SUPPLY LINE FOR SINGLE PHASE OR BATTERY POWERED GENERATORS    110 VAC or  Stand Alone option    208 VAC 220   230 VAC 240 VAC       Cable  gt    in Transformer 6T2 TB 3 or TB 8 TB 21 TB 4 or TB 5 TB 6 or TB 7    POWER SUPPLY LINE FOR THREE PHASE GENERATORS    220   230   240 VAC 400 VAC 415 VAC 440 VAC 480 VAC  Cable  gt    in Transformer 6T2 TB 4 or TB 5 TB 8 TB 7 TB 9 TB 10          6T2 Terminal Strip          AP 0015R2 1    HF Series Generators  Appendix   Quick Installation Guide for TXR Generators       3  THIS POINT ONLY APPLIES TO BATTERY POWERED GENERATORS        a  Open hinged door to access to Charger Module  Mod 1  by  removing the Allen screws     b  Verify that wire marked     gt     is connected to Transformer 2T1  terminal strip   110  for 110 VAC      c    Remove plastic covers over Charger Board and Line Monitor  Board     d  Plug J1 into the Charger Board  it is unplugged at the factory for  safety     e  Verify Jumper SW1 connection between SW1 5 and SW1 4   110 VAC      f  Re install plastic covers     Plastic Covers Charger Board       Line Monitor Board             j  CAUTION                           F    Gv   gt           Y       7112211   FEEFEERETR EI  d        
19. components installed   this Memory Location is factory  adjusted  Only perform this procedure if the HT Controller Board  and or HV Transformer have been replaced in the unit  For that   refer to the Service Manual   Calibration Chapter        16 AP 0015R2         Series Generators  Appendix   Quick Installation Guide for TXR Generators       3 3 DIGITAL mA LOOP CLOSED    Extended Memory Locations E03 and E05 contain the calibration factor for  Digital mA Loop Closed     Note   gt  Values in E03 and E05 Memory Locations are only related to the  Generator performance  they are not related to the X ray Tube s   installed   These Memory Locations are factory adjusted  Only  perform this procedure if the HT Controller Board and or HV  Transformer have been replaced in the unit  For that  refer to the  Service Manual   Calibration Chapter     3 4 AUTO CALIBRATION OF DIGITAL mA LOOP OPEN    To achieve the most accurate calibration  this procedure has to be  automatically performed by the Generator  Auto calibration      The Calibration procedure should be manually performed        if  Auto calibration is not possible  In this case  refer to the Service Manual   Calibration Chapter   Section  Manual Calibration of Digital mA Loop Open         Auto calibration of the Filament Current Numbers is divided in two separated  procedures related to the mA stations configured for the Small or Large Focal  Spots  Start with the Small Focal Spot  first group  and continue with the Large  Focal
20. es    y  j    Terminal Strip TB3 P e  Terminal Strip TXR 96                ale    t            Rear View     Rear Panel Removed       32 Technical Pamphlet W 2009 7 20           Tingle X Ray Products  LLC  Technical Pamphlet       High Frequency Generators     SHF Series                                                          Table  Generator     PROGENCY TRUE SPEED BUCKY  pre wired   Terminal Strip     3TS1 Wire Terminal Strip Wire Color   located on right side of swing out door    Color  3TS1 5 Blue 1 Blue  3TS1 6 Orange 2 Orange  3TS1 4 White 3 White  3TS1 3 Brown L Brown  3TS1 1 Yellow N Yellow  Any labeled earth ground in cabinet Green Ground Green  Terminal Strip     TXR 96  mounted on Table Lock Printed Circuit Board  top of control unit   Left Table Tab Red 24 VAC Red  Right Table Tab Black 24 VAC Black  Note  Red and Black continue through bucky  junction box  Both terminate at table lock printed  circuit board   Spare Purple Spare Purple                NOTE   Add following jumpers in generator       1  Jump 3TS1 2 to 3TS1 3       2  Jump 3TS1 1 to 3TS1 27                                                             Wall Stand  Generator     PROGENCY TRUE SPEED BUCKY  pre wired   Terminal Strip     3TS1 Wire Terminal Strip Wire Color   located on right side of swing out door    Color  3TS1 12 Blue 1 Blue  3TS1 13 Orange 2 Orange  3TS1 11 White 3 White  3TS1 10 Brown L Brown  3TS1 8 Yellow N Yellow  Any labeled earth ground in cabinet Green Ground Green  Terminal Stri
21. in alohanumerical order  plus  underscore and blank   the    Right Left Edit Arrow  buttons will scroll in  the word  Once the name is written  press    Enter    to save the new Region  name  If necessary  repeat this process to rename other Regions     g    R             g  REGION1 Eixe Koy REGION3 REGION4    Enter Selection and REGIONS REGION6 REGION7 REGIONS  go to next screen    610 ENTER REGION    Press the Display Selector close to    ENTER REGION  to go to step 4        Edit name          24 AP 0015R2         Series Generators       Enter Selection and  go to next screen     103    Enter Selection         go to next screen       u          Appendix   Quick Installation Guide for TXR Generators    Anatomical View  Once the desired Anatomical View  highlighted  is  chosen with the    Edit Arrow    buttons  if the operator requires to modify  the View name  go to step 5  If not  press the Display Selector close to   ENTER VIEW  to go to step 6   or    BACK    to return to the previous  Screen     g g  g  VIEW2    VIEW5 VIEW6    Go to previous screen    ENTER VIEW    Modify the View name  Press    Enter    to edit the name of the selected  View  The first letter ofthe name is highlighted  the    Up Down Edit Arrow     buttons will change the letter in alohanumerical order  plus underscore  and blank   the    Right Left Edit Arrow    buttons will scroll in the word   Once the name is written  press    Enter    to save the new View name  If  necessary  repeat this process to
22. ircuit Breaker  Visually check that all green lights and  LED 051 are lit on the Charger Board        1    LED DS1       Connector J1       Row of Green LEDs       Charger Board       veec000000000606    d  Visually verify that LED DL1  yellow  is flashing on the Line  Monitor Board  2A3   The MAX and MIN LEDs  red  on the Line  Monitor Board check the line coming in  neither should be on     Potentiometer POT1 is factory adjusted  If MAX or MIN LED is  ON  first check that Mains Voltage is according to specifications on  page 1 ofthis guide  If Mains Voltage is correct  adjust POT1  CW  up  CCW lower  measuring between TP 1 and TP 2 on Line  Monitor Board  Potentiometer adjustment should be       Mains Voltage  Potentiometer                                     x 2 5    Nominal Voltage    Potentiometer       Max and Min Lights       Line Monitor Board          AP 0015R2 7    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    This page intentionally left blank        8 AP 0015R2         Series Generators       SECTION 2    Appendix   Quick Installation Guide for TXR Generators    CONFIGURATION    2 1 PROPER DIP SWITCH SELECTIONS ON THE ATP CONSOLE CPU BOARD    Note   gt     ATP Console Board    The ATP Console CPU Board is located on Module 4 side of the  Hinged Door     Locate Switches  1  42 and  3 onthe ATP Console CPU Board from the  illustration     n Switch  1 contains four Dip Switches     Switch  2 contains four Dip Switches     Switch
23. ith the  Up down Edit Arrows          AP 0015R2 29    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    Keep in mind thatall Thickness values of a Projection will share the same  AEC values  Workstations and SID  that means that if one of these  Parameters is modified  it will affect to all Thickness values for that  specific Region   View   Projection     AEC  Workstations and SID are common values for all Thickness in a Projection                 mAs       Time                             FocusS FocusL    REGION1 VIEW3 PROJECT1 THCK 3     eo   Ready X Ray      onc        PREP                              Note F If an APR technique is to be stored with        parameters      suitable back up time  and or mAs  MUST be stored in the  program by the operator     11  Press  Enter  to save in memory the new parameters and selections  assigned to the selected Region View Project and Thickness value     12  Press simultaneously  APR ON OFF    and  Enter  buttons to exit from the  APR edition mode        30 AP 0015R2           Tingle X Ray Products  LLC  Technical Pamphlet  INTERFACE CABLE  Bucky to X Ray Control    A nine conductor  18 AWG cable is provided to supply control and power wiring for both  the bucky and the table or wall stand electric locks  All bucky s are pre wired at the table  or wall stand               Standard Frequency Generators     Models 425  amp  525                                                                          
24. ns listed above        AP 0015R2    21    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    3 5 FINAL CHECKS    Verify that all Configuration and Calibration data have been properly stored in  memory     Enter in calibration mode and check that the values noted for the   Filament Current Numbers  and  Extended Memory Locations  of the  Data Book are the same values displayed and stored in memory  Press  the  Reset  button to read the stored values     Exit from calibration mode and Service mode     Turn the Generator OFF and verify position of Dip Switches of Switch  2  on the HT Controller Board are       Dip Switch 2 in    Off    position  enables Filament and Rotor  Interlocks        Dip Switch 4 in    Off    position  Digital mA Loop Closed      Set Dip Switch 3 of Switch  2 on the ATP Console CPU Board in    Off     position to place the Generator in normal operating mode        22    AP 0015R2         Series Generators  Appendix   Quick Installation Guide for TXR Generators       SECTION 4 APR EDITING    Any APR Technique may be modified to fit the operator requirements  It is  possible to change and store in memory the names  Body Region   Anatomical  View   Projection   SID  Thickness value  range and default value   selections   Workstation and AEC  and radiographic parameters related to selected  Thickness value     Note   gt  Creating or Deleting an Body Region  Anatomical View  Projection  and Thickness can only be performed from
25. ow is    shown on the kVp Display   The entire list of Tube Data is found in your    Service Manual on the last page of the Configuration Section      Normally one of the following three Tube capacities is chosen     TUBE NUMBER TUBE CODE  ID  FOCAL ROWER RATINGS  5   MODEL   kVp Display   mAs Display  SPOT LS  kW  HS  kW                 001 139 TOSHIBA E7239X   133  003 140 TOSHIBA E7242X   187  004 090 TOSHIBA E7252X 26   73 300    12 AP 0015R2            Series Generators           Appendix   Quick Installation Guide for TXR Generators    Verify that the Tube code  ID  shown on the mAs Display is the same as  the tube code listed in the table above or in the Tube Table of the Service  Manual     Press the    Rese     Push button to save and permanently store the X ray  Tube configuration  You will once again hear an audible beep to confirm  that you have successfully saved your selection     AEC Film Speed    we     Density       EA           WARNING             ATENCION           FocusS Focus L        Ready X Ray                     PREP EXPOSE          Note F At this point  if your selected Bucky is still blinking  you are already  in Service Mode and ready for Autocalibration  If your selected  Bucky is not blinking  press    200    and    800    simultaneously for at  least two seconds to enter Service Mode   AP 0015R2 13    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    This page intentionally left blank        14 AP 0015R2       
26. p     TXR 96  mounted on Wall Stand Lock Printed Circuit Board  top of control unit   Left Table Tab Red 24 VAC Red  Right Table Tab Black 24 VAC Black  Note  Red and Black continue through bucky  junction box  Both terminate at wall stand lock  printed circuit board   Spare Purple Spare Purple                NOTE   Add following jumpers in generator       1  Jump 3751 9 to 3751 10          2  Jump 3TS1 8 to 3TS1 27       Technical Pamphlet W 2009 7 20 33              Tingle X Ray Products  1    Technical Pamphlet    High Frequency Generators   Pictures        amp         KS M    S E       REXU   gt     Terminal Strip TXR 96 A                Terminal Strip 3TS1          Cover   Removed    34 Technical Pamphlet W 2009 7 20    
27. push buttons simultaneously for 2 seconds  After releasing  both push buttons  code  222  will be flashing on the Console waiting for  confirmation to then enter Auto calibration mode     Press the    kV increase  and  Power         push buttons again until code   222  disappears on the Console to confirm Auto calibration mode     Auto calibration is activated after releasing both push buttons  At this  moment  the Generator will check the mA stations available for the Small  Focal Spot  When 50 kV at 10 mA is displayed in the display windows   the Unit is ready for Auto calibration          mA    BB                 Focus S FocusL       Tm    PREP EXPOSE                  Density  Z    T  WARNING  v           ATENCION  ren       4  Check that the Heat Units available for the X ray Tube are 100    kV Display shows    H        on the Console      5  Keep the Handswitch push button fully pressed or use the Exposure  Controls on the Console to perform continuous exposures     Auto calibration can be stopped momentarily by releasing  the Handswitch push button or the Exposure Controls  Do  not exit from Auto calibration before the procedure has been  completed        AP 0015R2    19    HF Series Generators       Appendix   Quick Installation Guide for TXR Generators    If the Heat Units available for the X ray Tube are less than 40    exposures are inhibited momentarily and code    177    will be flashed on the  Console accompanied by an alarm  In this case  release the Handswitch 
    