Home
        OPERATOR`S MANUAL SUPPORTED TORQUE CALIBRATION BEAM
         Contents
1. ORE THAN 1 PERSON        2 1 Line up the bearing support with holes in the middle of the bed     Beam    y Bearing Support       FIGURE 3   Location of Beam  amp  Bearing Support    2 2 Position of beam is established by fitting bearing support with 2 off M8 shoulder bolts supplied  Tighten  to 10 N m     2 3 Fix using 6 off M12 bolts supplied  Tighten to 80 N m     2 4 Fit 2 off guide plates to bed  Tighten 4 off M8 bolts to 10 N m     3  Levelling Gearbox  3 1 Remove M6 end stop screw from pedestal bed     3 2 Line up the levelling gearbox with the slots in the pedestal bed then slide on   Ensure drive square input is facing beam as shown in figure 4   se  Levelling Gearbox       OSE  2 End Stop  Screw       FIGURE 4   Location of Levelling Gearbox    3 3 Fit M6 end stop screw to pedestal bed     4  Hangers    TIP  The hangers are supplied ready to attach to the beam   Ensure the correct hanger is fixed to the correct end of the beam   If the hanger assembly is taken apart  DO NOT mix parts with the other hanger assembly     4 1 Bolt hanger marked    XX    to beam end marked    XX      Bolt hanger marked    YY    to beam end marked    YY      Choose a location  A  B  C  D  E  F  G or H  for each beam  the required location is set in the  Operating Instructions     4 2 Place weight platform under hanger to be loaded  as shown in figure 5        FIGURE 5     Hanger Weight Platform    OPERATING INSTRUCTIONS    WARNING  KEEP HANDS CLEAR OF MOVING PARTS DURING USE        Transdu
2. PORE 128383   Tel   65 6841 1371   Email singapore norbar com au    NORBAR TORQUE TOOLS  SHANGHAI  LTD  E Building   5F  no  1618 Yishan Road    Minhang District  Shanghai   CHINA 201103   Tel   86 21 6145 0368   Email sales norbar com cn    nag  NORBAR TORQUE TOOLS INDIA PVT  LTD  Plot No A 168  Khairne Industrial Area   Thane Belapur Road  Mahape   Navi Mumbai     400 709  INDIA  Tel   91 22 2778 8480  Email enquiry norbar in    WWW  norbar com       
3. SUPPORTED TORQUE CALIBRATION BEAM       Part Number 34320   Issue 4   Original Instructions  English        CONTENTS    Part Numbers Covered by This Manual  Safety    Introduction  Parts Included  Accessories    Features and Functions    External Effects  Temperature Compensation  Gravitational Effects  Buoyancy Effects    Set up Instructions  Fixing Pedestal   Beam  amp  Bearing Support  Levelling Gearbox  Hangers    Operating Instructions  Locating Transducer  Locating Hangers  Exercising   Calibration    Maintenance   Beam Repair  amp  Recalibration  Weight Recalibration  Specifications    Trouble shooting    Glossary of terms    Oo NNN     amp K BOW  amp     N NOOA        mma  23o    11    11    12    12    12    PART NUMBERS COVERED BY THIS MANUAL       21842 500   5000 500   5000    SAFETY    It is recommended that safety shoes are worn  Use safe manual handling techniques when lifting     INTRODUCTION    This calibration beam is designed for the static calibration of torque transducers  Torque is generated by the  application of a known force  weights  at a known radius from the centre of rotation of the torque transducer   Reaction is taken via a levelling gearbox so the beam can be kept horizontal     Parts Included             Hangers  Weight  Platform   N  Pedestal A  FIGURE 1   Parts Included   nG  Quantity Quantity  Beam with bearing support 1 Guide plate 2  Pedestal with bed 1 Hangers 2  Levelling gearbox 1 Weight platform 1  Jacking plate 2 Raw bolts 4  Torque 
4. air     Weight Recalibration    It should be possible to obtain local recalibration of weight sets  Please contact Norbar if further assistance is  required     11    SPECIFICATIONS    Absolute maximum torque  110   for each beam location    Recommended temperature  20  C     2  C    Humidity  30  to 70     Location hole accuracy      0 01  of nominal     TROUBLE SHOOTING  Problem   Solution O OOOO    Can not achieve zero torque  Levelling gearbox causing small load  remove levelling gearbox     GLOSSARY OF TERMS  rem Meaning    Bearing Support Centre of beam rotation    Class M1 Weight classification    Hanger Device to hold weights    Pedestal Main support for calibration beam   Weight Platform Device to rest weights on between tests   Levelling Gearbox Device to ensure beam is level     PAA   IS   NORBAR TORQUE TOOLS LTD  Beaumont Road  Banbury   Oxfordshire  OX16 1XJ   UNITED KINGDOM   Tel   44  0 1295 270333   Email enquiry norbar com       2    NORBAR TORQUE TOOLS PTY LTD  45   47 Raglan Avenue  Edwardstown   SA 5039   AUSTRALIA   Tel   61  0 8 8292 9777   Email enquiry norbar com au    NORBAR TORQUE TOOLS INC  36400 Biltmore Place  Willoughby   Ohio  44094   USA   Tel   1 866 667 2279   Email inquiry norbar us       NORBAR TORQUE TOOLS  NZ  LTD  B3 269A Mt Smart Road   Onehunga  Auckland 1061   NEW ZEALAND   Tel   64 9579 8653   Email nz norbar com au       Norbar Torque Tools Ltd 2007    NORBAR TORQUE TOOLS PTE LTD  194 Pandan Loop    07 20 Pantech Business Hub  SINGA
5. cer Levelling Gearbox          7  gt     ps           o      o      E a  haet    FIGURE 6     Location of Transducer    TIP  This is a general procedure to use the calibration beam   Use this procedure along with any specific standard being followed     Locating Transducer    1  Connect transducer to display unit required  Zero reading on display     2  Mount the transducer between beam  amp  levelling gearbox using appropriate adaptors  see figure 6  Slide    the levelling gearbox so the transducer drive squares are fully engaged     Allow the required warm up time for transducer  amp  display  a minimum twenty  20  minutes is  recommended     Locating Hangers       FIGURE 7     Beam Location Holes    1  The beam has many location holes for the hangers  see figure 7  The required location depends on the  torque required  amp  the weights used  Use the following table to determine the hanger location     Absolute  Torque i  eae maximum torque      e J x Distance   for overload test       A 1 52400 4 448 22 1000 5000 Ibf ft 110   B 1 34885   4 448 22 1000 6000 N m 110   C 1 12404   4 448 22 1000 5000 N m 110   D 0 899236   4 448 22 1000 4000 N m 110   E 0 786832   4 448 22 1000 3500 N m 110   F 1 21920 4 4 448 22 1000 4000 Ibf ft 110   G 1 00000      i    110   H 0 91440 3 4 448 22 1000 3000 Ibf ft 110         For use with metric weights     2  Bolt the hangers to the required location   Ensure that the correct hanger is located at each end of the beam     3  Place weight plat
6. ement can be taken     Q Tm Mm O    After the last increment remove all weights   Remove levelling gearbox from transducer     A final zero result can be measured  Ea  FIGURE 8   Beam Spirit Level    NOTE  If the transducer needs adjusting refer to appropriate operator   s manual or service manual     NOTE  For reverse calibration the weights are applied to the opposite hangers  The transducer will  need to be exercised in the reverse direction before calibration     10    MAINTENANCE    Beam Repair  amp  Recalibration    Norbar Supported Calibration Beams are designed for transducer calibration at low levels of uncertainty   They are robust for normal handling by technicians  but if dropped or otherwise damaged  may lose their  calibration integrity  It is therefore essential that the beam is recalibrated after any such incident     Recalibration involves establishing the beam length from the centre of the driving square to a locus of points  on the beam end   See original calibration certificate for further details   This is difficult to achieve with  conventional length standards  although use of a Co ordinate Measuring Machine may give satisfactory  results  Planned calibration intervals will depend on usage  and should be decided by the laboratory     Norbar offers a full repair and recalibration service  with traceable certification of status before and after  repair  Due to the complex nature of the beams  please contact Norbar in the event of any damage requiring  rep
7. form  figure 5  under hanger to be loaded     NOTE  The hangers must be loaded equally  so weights are added in pairs     Exercising  1  Itis recommended to exercise the transducer before calibration  this can be done by loading to full scale  then removing the weights three  3  times     The display output should be watched during this process to avoid overloading     The exercising procedure is     A  Apply torque by adding the required weights to the hangers  then winding the handle on the levelling  gearbox until full scale torque is displayed     B  Decrease torque by un winding the handle on the levelling gearbox until the display reads the  minimum value and weights are firmly seated on weight platform     C  Repeat steps A  amp  B two more times   Remove all weights     E  Ensure transducer display is set to zero before calibration   The levelling gearbox may need to be removed from the transducer to get a true zero     Calibration    1  The transducer is loaded in the required increments   Gentle application of the weights is advised to maintain a steady reading   At each torque increment the procedure is   A  Place weights onto hangers  loading the twin hangers equally   B  Wind the handle on the levelling gearbox to lift the weights clear of the weight platform     C  Level beam using the levelling gearbox   Ensure the beam spirit level reads correctly  see figure 8     Waiting for the reading to stabilise   Take torque reading from display     The next torque incr
8. ording to the details below     The Coefficient of linear expansion for steel is 12 x 10     C or  0 000012 per   C   The formula for calculating the change in effective beam length is     Change in length    Initial beam radius x Coefficient of linear expansion x Change in temperature in   C  from 20  C Nominal     Example  1 metre radius beam used at 24   C    Change in length   Initial beam radius x  12 x 10    x 4   Change in length   1 x  12 x 10    x 4   Change in length   4 8 x 10     Change in length   0 000048 metre   New effective beam radius   Initial beam radius   Change in length  New effective beam radius   1 000048 metre    Example  1 metre radius beam used at 16   C    Change in length   Initial beam radius x  12 x 10    x  4   Change in length   1 x  12 x 10    x  4   Change in length    4 8 x 10     Change in length    0 000048 metre   New effective beam radius   Initial beam radius   Change in length  New effective beam radius   0 999952 metre    Gravitational Effects    It is very important that the gravitational value for the Laboratory is established  The effect of not doing this in  the UK could be a variation in the force produced by the weight of up to approximately 0 05   which is five  times the 0 01  tolerance of the weight  Outside of the UK this variation in force could be significantly more     It is therefore strongly recommended that you establish the local value of gravity  g  for your Laboratory and  use weights that have been calibrated at 
9. s   1  Fixing Pedestal   2  Beam  amp  Bearing Support   3  Levelling Gearbox   4  Hangers   NOTE  The set up instructions must be followed in the sequence shown   Items required  Spanner set   Allen key set     Spirit level  recommended accuracy    0 5 mm m   4 x Suitable fixing bolts  amp  Drill  to suit fixing bolts      1  Fixing Pedestal  1 1 Ensure the floor is flat  clean and capable of supporting a weight of 480 Kg   1 2 Position pedestal to allow access around the calibration beam when assembled   NOTE  Fixing bolts NOT included   Customer must supply 4 off M10 Fixing bolts to suit their concrete floor   Pull out force on fixing bolts must exceed 2 5 KN   Ensure the concrete floor is capable of taking 30 N mm7        Tighten fixing bolts to achieve a minimum of 6 KN tension in the bolts     1 3 Drill concrete floor to accept the fixing bolts   Fit fixing bolts to hold pedestal to floor  do not fully tighten yet     1 4 Locate jacking plates  supplied  under jacking screws  as shown in figure 2        FIGURE 2   Pedestal Base with Fixing Holes  amp  Jacking Screws    1 5 Place a spirit level on the machined surface of the pedestal bed and adjust jacking screws to ensure  pedestal is level  For correct calibration it is essential the pedestal is level in all directions     1 6 Fully tighten the raw bolts to the required torque     1 7 Check pedestal bed is level     2  Beam  amp  Bearing Support    WARNING  THE BEAM  amp  BEARING SUPPORT IS HEAVY   THIS TASK REQUIRES M
10. that gravitational constant     Norbar will supply weights calibrated to gravitational constants specified by the customer  However  if the  customer does not specify a value for    g     the value can be calibrated at an estimated gravitational constant  for the customer s location  This will be agreed with the customer prior to calibration     Buoyancy Effects  The Norbar system uses calibrated weights to generate a downwards force     This means that Archimedes principle applies  i e  air pressure under the weights causes an upwards force   This reduces the effective force generated by the weights and therefore the mass must be increased to allow  for this     Under standard conditions  i e  Air density 1 2 kg m  and 20  C and working in conventional mass terms  the  increase required is by a factor of 0 015   Weights purchased from Norbar will already have this factor taken  into account     Weights that are calibrated to standard procedures do not have this factor taken into account because the air  buoyancy affects both sides of the mass balance and can be ignored  It is important that weights used for  torque transducer calibration are adjusted for air buoyancy     It should also be noted that the double ended beam design employed by Norbar means that each half of the  beam is balanced with regard to buoyancy of the beam  This is a significant advantage over single arm  counterbalanced systems     SET UP INSTRUCTIONS    The set up of this beam covers the following item
11. wrench 1 Calibration certificate 1  M8 Bolt  25354 90  2 M12 Bolt  25356 35  6     Bearing support to pedestal   Bearing support to pedestal   M6 Bolt  25353 25      Guide plate to bed  4 Operator s Manual 1  Accessories  Part Number  21469 NAM Set of 20 x 50 Ibf weights to give 5000 N m   5000 Ibf ft    FEATURES AND FUNCTIONS    Transducer Compatibility    The calibration beam is ideally suited to Norbar torque transducers  but may be used on other  manufacturer s equipment     Beam Alignment    A self locking gearbox provides an easy method of keeping the beam horizontal once loaded     Traceability    The beams and weights are guaranteed within a tolerance of accuracy and traceable certification is provided     Torque Accuracy    The beam and its associate stand have been robustly constructed to ensure only torque is applied to the  transducer under test  Any torque losses  through bending or bearing loss  seen by the transducer have  been minimised  Weight accuracy is required to be equal to or better than 0 01  which approximates to  class M1     Clockwise   Counter Clockwise Operation    The symmetrical beams allow for both clockwise and counter clockwise calibration     EXTERNAL EFFECTS    Temperature Compensation    The calibration beam should be used in a temperature controlled environment of 20  C     2  C   If the beam must be used outside these limits  the temperature must be stable  within 1  C change per hour   and the effective length of the beam calculated acc
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  Copyright © All rights reserved. 
   Failed to retrieve file