Home

Philips C3 Patient Monitor C3 Service Guide Models: 862474, 862478

image

Contents

1. Pin Signal Direction 1 not used 2 Rx data input 3 Tx data output 4 DTR output 5 Signal Ground input output 6 DSR input 7 RTS output 8 CTS input 9 Alarm Out output Nurse Call Pin 9 of the RS 232 serial interface connector provides an Alarm Out signal Any time there is any alarm condition active in the monitor and ifthe Nurse Call Signal option in the Setup Menu is On pin 9 goes to plus RS 232 level voltage gt 5 VDC The alarm delay shall be lt 0 5 seconds If in the Setup Menu the Nurse Call Signal option is Off pin 9 is always at the minus RS 232 level voltage In order to make use of the Alarm Out signal pin 9 should be connected to a high impedance circuit gt 1000 and protected against transient voltages 10 4 RS 232 Interface 11 Training This chapter provides a complete flexible training program that addresses the needs of technical personnel who service the monitor at all levels of involvement Training 11 1 Introduction Introduction The training itself is self paced This chapter guides you through the entire service training program All of the information you require is in this chapter or referred to in this chapter Levels of Involvement Preventative If you are performing only preventative maintenance on the monitor you must review the Maintenance following sections of this chapter Only e Overview Preventative Maintenance Phone If you are providing phone supp
2. Message Cause of Failure Remedy Loss of Pulse from SpO SpO cannot find a pulse Make sure the SpO patient cable is connected and the sensor is correctly attached to the patient Low Battery The battery has less than 30 minutes of Recharge the battery charge left NBP Artifact Incomplete measurement due to patient motion or physiology 1 Check the patient 2 Check cuff application and orientation NBP Blocked Hose NBP tubing blocked measurement not possible Check that the tubing is not obstructed or pinched NBP No Cuff Monitor unable to detect NBP cuff 1 Check to see that cuff is present 2 Check cuff and tube connections for leaks NBP Overpressure This INOP arises when the NBP cuff Remove the cuff from the pressure increases above overpressure safety limits patient Make sure the rubber tube to the NBP cuff is not kinked NBP Time Out NBP measurement took too long to complete Recorder Out of Paper New paper needed Replace paper as instructed in the section titled Recorder Paper Replacement in the C3 Instructions for Use guide SpO Cable Sensor Disconnect Monitor cannot detect the SpO cable or sensor Check that the SpO gt patient cable is connected and the sensor is properly attached to the patient SpO Equip Malf SpO hardware is faulty Verify failure and replace SpO module See Removing the
3. CO Sensor Warm up The sensor has not reached operating temperature Allow the sensor to warm up CO Occlusion The FilterLine or exhaust tub is blocked to the extent that a measurement sample cannot be taken The outlet is occluded at Power On This INOP also appears after the Pump Op Time is reset Check the FilterLine and exhaust tube then disconnect and reconnect the FilterLine If the INOP is still displayed use a new FilterLine If Pump Op Time has been reset perform a flow check and store the flow in Service Mode select Store Flow CO Overrange The CO value is outside the measurement range CO Purging The monitor is purging the FilterLine This occurs when an occlusion is detected in the line or airway adapter If the occlusion is not removed by purging a CO Occlusion INOP displays Check for an occlusion and remove If necessary replace the FilterLine CO Auto Zero An AutoZero is in progress An AutoZero will be done at the following intervals 5 minutes after connecting the FilterLine hour later 12 hours later and every 12 hours from there on ECG Leads Off Not all leads are connected Make sure the ECG trunk cable is connected the leads are connected to the electrodes and the electrodes are attached If INOP still displays use a new lead set 6 6 Troubleshooting Part 2 Isolating and Solving Monitor Problems
4. Mounting Standoffs 100230V 50 60 Hz TIAL 250V O E PHILIPS C3 PATIENT MONITOR Manufacti ilips for Philips Medical Systems 1000 Minuteman Road Andover MA 01810 Site Preparations 2 7 Warnings Cautions and Safety Precautions Relating to Wall Mount Installation Warnings Cautions and Safety Precautions Relating to Wall Mount 2 8 Site Preparations Installation Make sure that you have read all applicable instructions before attempting to install the wall mount Wall mounts that are intended to support monitors must be capable of supporting four times the weight of the monitor when properly installed If wall mounts are to be installed on plaster board walls we recommend that the mounts be installed using steel or plywood plates sufficient to distribute the load over a large section of the wall Lead expansion bolts are neither adequate nor safe for this purpose It is the responsibility of the hospital institution or its designated representatives to determine that the wall is adequate to safely mount monitoring instrumentation This includes the selection of and proper installation of the appropriate fasteners and mounts In a new construction or in a remodelled construction where the wall can be completely or partially removed a 16 gauge steel stud should be mounted for the purpose of attaching the wall channel This design safety facto
5. 5 Select your language from the following screen then select YES to place the monitor into that language After language is selected the main monitoring Nederlands T rkce o enl Norsk Svenska Suoni Polski Cesky 5 14 Configuring the Power up Defaults Menu 6 Troubleshooting This chapter explains how to troubleshoot the monitor if problems arise Tables are supplied that list possible monitor difficulties along with probable causes and recommended actions to correct the difficulty Use this chapter when you suspect you have a problem with your monitor or whenever you are referred to this chapter from another part of the Service Guide Troubleshooting 6 1 General Troubleshooting Safety Information General Troubleshooting Safety Information Warning Do not attempt to troubleshoot the monitor if there is any sign of burning or smoking coming from the monitor Call the Philips Response Center or your local Philips representative immediately Caution Only qualified service personnel should open the monitor housing remove and replace components or make adjustments If your medical facility does not have qualified service personnel contact the Philips Response Center or your local Philips representative 6 2 Troubleshooting Objectives Concepts Trouble shooting Error Codes Objectives In order to meet this chapter s objectives you should be able to diagnose and isolate
6. Anyone who connects additional equipment to the signal input port or signal output port configures a medical system and is therefore responsible to ensure that the system complies with the requirements of system standard IEC Standard 60601 1 1 If in doubt contact the Philips Response Center or your local Philips representative Note The monitor and its accessories must be tested by qualified service personnel at regular intervals to verify proper operation according to the procedures of the user s institution Other important safety information is located in this Service Guide where appropriate 1 2 Overview Introduction Introduction This manual contains information for servicing the C3 patient monitor subsequently referred to as the monitor throughout this manual Only qualified service personnel should service this product Before servicing the monitor carefully read the C3 Instructions for Use guide for a thorough understanding of operation C3 Patient Monitor Description The purpose and function of the C3 family of patient monitors is to monitor ECG heart rate non invasive blood pressure NBP functional arterial oxygen saturation SpO gt respiration rate temperature and carbon dioxide CO for adult and pediatric patients in hospital areas and hospital type facilities such as clinics The monitors may be used by clinical users during hospital transport The intended users of the C3 patient mon
7. Perform the following sequence a Press the Home button b Press the Pressure Tests button c Press the Pressure Leak Test button d Press the Setup button e Press 3 0 0 then Enter f Set the Cuff to Internal Confirm that the simulator test screen is active and displays Leak Test Press the Volume button to ensure that both valves are closed Perform an offset adjustment so that the simulator and monitor both display a pressure of 0 mmHg or kPa by doing the following a Press the Contrast button on the monitor s front panel b If needed zero the simulator see To Zero the Simulator on page 4 16 Press the NBP button on the monitor s front panel to activate the pump Hold the button until the monitor displays a pressure of approximately 250 mmHg 33 3 kPa Allow 15 20 seconds for the pressure to stabilize Record the pressure displayed on the monitor P1 Testing the Monitor 4 17 Preventative Maintenance Tests Step Action 8 Start a one minute timer After one minute record the pressure displayed again P2 The pressure should drop by no more than 6 mmHg 0 8 kPa during the one minute period Calculate P1 P2 and document the leakage test value X2 9 Press and hold the Volume button until the monitor displays a pressure of 0 mmHg or kPa 10 If no further NBP tests are to be conducted turn the monitor Off Normal monitoring operation retur
8. In addition to the acquisition of the QRS waveform complex the ECG input and subsequent signal processing computing circuitry perform a number of other functions e They detect a lead off condition if one of the electrode connections is disrupted They detect the presence of pacemaker signals within the QRS waveform complex of the ECG e They generate a synchronization pulse for external use with defibrillators The Defib Sync Pulse output is available at a connector in the rear panel The patient s respiration is detected by using two of the three leads of the ECG electrodes and cable A low level excitation signal is applied to these leads and the variation of the thoracic impedance caused by the breathing is sensed and processed for display and measurement Training 11 9 Theory of Operation and System Architecture NBP Processing SpO Processing CO Processing Temperature Processing 11 10 Training The NBP processing uses an oscillometric technique to provide needed measurements at selected intervals This technique uses an inflatable sphygmomanometer cuff similar to those used by clinicians in routine measurements A motorized pump inflates the cuff to approximately 180 mmHg initially at which point the pressure effectively stops the flow of blood Then under monitor control the pressure in the cuff is gradually reduced while a pressure transducer detects the air pressure and transmits the parameter signal
9. 40493D 40493E M1980A M1981A M1S10A M1550C M1611A M1613A M1615A MI879A MI878A M1877A M1876A MI875A MI874A MI576A M1575A M1574A M1573A M1572A MIS71A M1577A M1578A M1579A 40401E 40401D 40401C 40401B 40401A M1598B M1599B M2526A M2525A M2524A M2528A M2522A M2521A M2520A M1920A M1922A M1921A M4816A M4817A Accessories from companies other than Philips Medical Systems carry CE markings appropriate to the accessory Additional accessories not identified above fall outside the definition of a medical device The Former Agilent Technologies Healthcare Solutions Group is now a part of Philips Medical Systems Some accessories may still be branded with the Agilent name United States Canada Philips Software License Terms ATTENTION Authorized EU representative Philips Medizinsystems B blingen GmbH Hewlett Packard Note United States Federal Law restricts this device to sale by or on the order of a physician This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme a la norme NMB 001 du Canada USE OF THE SOFTWARE IS SUBJECT TO THE PHILIPS SOFTWARE LICENSE TERMS SET FORTH BELOW USING THE SOFTWARE INDICATES YOUR ACCEPTANCE OF THESE LICENSE TERMS IF YOU DO NOT ACCEPT THESE LICENSE TERMS YOU MAY RETURN THE SOFTWARE FOR A FULL REFUND IF THE SOFTWARE IS BUNDLED WITH ANOTHER PRODUCT YOU MAY RETURN THE ENTIRE UNUSED PRO
10. When recording test results these are always associated with a particular monitor by means of the serial number The serial number is 10 characters and is located on the back of the monitor amp e ES Serial Number qe ER E USC3xxxxxx q Montand iri ie for e CE LS Anis IPX1 AN amp e Passwords Some of the test may require that you enter the Power Up Default Menu located in the Setup Menu To access Power Up Default Menu you must enter a password The password is 2 1 5 Testing the Monitor 4 9 Visual Test Visual Test Inspect the system for obvious signs of damage Also check external leads and accessories What to record on the service record V P or V F Power On Test Step Action Switch the monitor On Observe whether the system boots up successfully without displaying any error codes The main monitoring screen should display What to record on the service record PO P or PO F 4 10 Testing the Monitor Functionality Assurance Tests Functionality Assurance Tests Performance Assurance Test Power On Self Test The following assurance checks are recommended to verify proper operation daily before the monitor is used to monitor a patient During functionality assurance checks verify the overall operation by completing the following Performance Assur
11. 00000000 Y 6 8 E f A e Empty Battery Perfusion Indicator Bar Heart Rate Powering Down Heart Rate taken from ECG Heart Rate taken from NBP Heart Rate taken from SpO Respiration Stat Mode Audio Pause Setup Menu ECG Size Bar RF Interference rt Qiu E P E T1AL 250V IPX1 Battery Compartment RS 232 I O Equipotential Grounding Post Defib Sync Attention consult accompanying documents Alternating Current Fuse Type Date of Manufacture Y year X Month Drip proof ETL Mark Sales and Support Offices Please call your local Philips Medical Systems sales office listed in your telephone directory or a Philips Medical Systems regional office listed below for the location of your nearest sales office or for information on how to contact the Philips Response Center CORPORATE HEADQUARTERS Philips Medical Systems Netherlands B V Postbus 10 000 5680 DA Best Netherlands CORPORATE HEADQUARTERS Philips Medical Systems 3000 Minuteman Road Andover MA 01810 United States 800 934 7372 CANADA Philips Medical Systems Canada 281 Hillmount Road Markham ON L6C 283 800 291 6743 EUROPE MIDDLE EAST AND AFRICA Philips Medizinsysteme B blingen GmbH Cardiac and Monitoring Systems Hewlett Packard Str 2 71034 B blingen Germany Fax 49 7031 463 1552 xi xii LATIN AMERICA HEADQUARTERS Philips Medical Systems 1550 Sawgrass Cor
12. 12 volt 4 5 Ah sealed lead acid Battery Operating Time 3 hours fully charged battery at 25 C without CO under normal one NBP measurement every 15 minutes all other modes active conditions 45 minutes if printing continuously 1 hour fully charged battery at 25 C with CO under normal conditions 30 minutes if printing continuously AC Mains 100 to 240 VAC 50 60Hz 1A max Environmental Parameter Specification Mechanical Shock IEC 60068 2 27 100g 6 msec 3 axes 18 total shocks non operating Mechanical Vibration IEC 60068 2 6 Sinusoidal 10 58 Hz 0 15 in displacement 58 150 Hz 2g acceleration 4 min sweep 20 sweeps axis non operating Water Resistance IEC 60529 Classification IPX1 Protected against vertically dripping water Thermal Operating 0 to 50 C 32 to 122 F Storage 20 to 60 C 4 to 140 F Humidity Operating 5 to 95 RH non condensing Storage 5 to 95 RH non condensing Specifications 12 3 Measuring and Displaying Parameters Measuring and Displaying Parameters ECG 12 4 Specifications Parameter Specification Heart Rate Range 20 250 b min Heart Rate Accuracy 5 b min Bandwidth Normal Monitoring 0 5 Hz 40 Hz Bandwidth Extended Low Frequency Range use 0 05 Hz 40 Hz selectable Leads 3 lead user selectable Display Sweep Speeds 12
13. 2 0 m 8120 1351 Power Cord UK 900 2 3 m 8120 4475 Power Cord Australia 901 2 0 m 8120 2104 Power Cord Switzerland 906 2 0 m 8120 2956 Power Cord Denmark 912 2 0 m 8120 4211 Power Cord South Africa India 917 2 0 m 8120 5342 Power Cord Japan 918 2 5 m 8120 5182 Power Cord Israel 919 2 0 m 8120 6869 Power Cord Argentina 920 2 0 m 8120 6980 Power Cord Chile 921 2 0 m 8120 8376 Power Cord China 922 2 0 m Spare Parts 8 9 Exchange Unit Part Numbers Exchange Unit Part Numbers Exchange Unit Part Number Description 453563344311 Thermal Recorder for C3 Patient Monitor GSI 453563480441 Exchange Recorder for C3 Patient Monitor GSI 453563480381 C3 Standard Color Monitor 453563480401 C3 Standard Color Monitor with Recorder 453563480411 C3 Sedation Color Monitor 453563480431 C3 Sedation Color Monitor with Recorder 8 10 Spare Parts 9 Packing for Shipment This chapter explains how to pack the monitor for shipment for any reason Packing for Shipment 9 1 General Instructions General Instructions To ship the monitor for any reason follow the instructions in this section To pack the monitor for return disconnect all cables It is not necessary to return sensors patient cables NBP tubing and cuff or power cord Pack the monitor carefully Failure to follow the instructions in this section can result in loss or damage not covered by the warranty If t
14. Monitor Applications and Algorithms 0 0 eee o 11 13 Contents 6 12 Specifications ia A bal Hardware Specifications coi a Re gun e eU REGN Ue 12 2 SER ALS he AL Ghee et ULL Bes 12 2 Electrical 2252 ento tte idi e Mat da ALSO uL 12 3 Environmental ai a tue aes ove RL pe tU ee 12 3 Measuring and Displaying Parameters oooooooooooorrrr eee 12 4 ECG icant A ee OE ee GU RE EE A CDI IIBER a 12 4 ECG Standards 3 023 et ite rehash tte tod ael iet tr s fe 12 5 Respiration ic EE 12 6 NBP Non Invasive Blood Pressure 0 000 c cece ccc eh 12 7 Temperature ii Oe eee ped een OLE A et 12 7 NjeSP PPM 12 8 poA Em 12 9 Trends E A a LESE n es Dres 12 10 Contents 7 Contents 8 1 Overview This chapter provides a brief overview of the C3 patient monitor It also provides a list of the monitor s features and parameters Overview 1 1 General Safety Information General Safety Information Warning This monitor is not intended for neonates If you connect the monitor to any instrument verify proper operation before clinical use Refer to that instrument s Instructions for Use guide for full instructions Accessory equipment connected to the monitor s data interface must be certified according to IEC Standard 60950 for data processing equipment or IEC Standard 60601 1 for electromedical equipment All combinations of equipment must be in compliance with IEC Standard 60601 1 1 systems requirements
15. STAT Series of consecutive measurements for 5 minutes a Systolic and Diastolic blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method and mean pressure measurements determined with this device are equivalent to those obtained by an intra arterial blood pressure measurement device within the limits prescribed by the American National Standard Electronic or automated sphygmomanometers Temperature Parameter Specification Technique Probe Range 15 to 45 C 59 113 F Accuracy 0 1 C Specifications 12 7 Measuring and Displaying Parameters SpO 12 8 Specifications Parameter Specification Range Saturation 0 to 100 Pulse Rate 30 to 300 b min Accuracy SpO 0 69 not specified M1191A M1192A 2 5 70 to 100 M1194A 4 0 70 to 100 Accuracy with Philips disposable sensors M1902B M1903B M1904B and with Nellcor disposable sensors OxiCliq A P N 1 3 70 to 100 Pulse Rate 2 or 1 b min whichever is greater Display Update Period l sec Calibration functional a The specified accuracy is the root mean square RMS difference between the measured values and the reference values Complies with EN 864 ISO 9918 Measuring and Displaying Parameters Parameter Specification Range 0 to 98 mmHg 0 to 13 kPa
16. nonvolatile Data interval 20 seconds stored data points is the average over 20 second interval Graphical Trend Format One graph per vital sign Display range 2 hours scrollable Vertical scaling Heart Rate 50 100 0 250 25 125 b min NBP 50 100 0 150 0 300 mmHg SpO 0 100 60 100 80 100 CO 0 100 0 60 20 50 mmHg Respiration Rate 0 20 0 50 0 150 b min Temperature C 15 45 33 41 35 39 C Temperature F 55 115 91 107 96 102 F Tabular format One table for all variables Six fields per row time and five vital signs Display interval Per NBP measurement or 15 minutes for no NBP or 20 seconds during alarm condition
17. 100 mOhm Test current 25 Amps applied for 5 to 10 seconds We recommend to flex the main cable during the test in order to identify potential bad contact or damage of the earth wire Safety test according to EN60601 1 Clause 18 Report largest value What to record on this service record S 1 Por S 1 F 4 40 Testing the Monitor Electrical Leakage Earth Leakage Current Patient Safety Tests The following tests verify the electrical leakage of the monitor Earth Leakage Current Enclosure Leakage Current Patient Leakage Current e Patient Source Current with Mains Voltage on the Applied Part Warning Perform all leakage tests any time the unit is opened This test is in compliance with IEC 60601 1 Earth Leakage Current In locations where mains voltage is 100 120 volts the applied voltage is 132 volts In locations where mains voltage is 220 240 volts the applied voltage is 264 volts The applied AC frequency should be the same as the local mains 50 or 60 Hz All measurements shall be made with the power switch in both On and Off positions 1 Connect the monitor AC plug to the electrical safety analyzer as recommended by the analyzer operating instructions 2 Perform test as recommended by analyzer operating instructions Earth leakage current is measured under various conditions of the AC mains and protective earth conductor For each condition the measured leakage current must not exceed tha
18. 5 25 and 50 mm sec Pacemaker Detection Indicator on waveform display user selectable ECG size sensitivity 0 5 1 2 4 mV cm Lead off Detection detected and displayed Input Impedance gt 5 MQ CMRR Common Mode Rejection Ratio gt 90 dB at 50 or 60 Hz Input Dynamic Range 5 mV AC 300 mV DC Defibrillator Discharge lt 5 sec per IEC 60601 2 27 Recovery lt 8 sec per AAMI EC13 1992 ECG Standards Measuring and Displaying Parameters Standards Meets the performance standards of ANSI AAMI EC13 1992 Instead of a 1 mV standardizing voltage section 3 2 2 9 a fixed 1 cm reference bar is always present in the ECG display along with the ECG size setting expressed in mV cm The following information references particular sections of ANSI AAMI EC13 1992 Respiration leads off sensing waveform 3 1 2 1 b A 85 kHz square wave is used to inject 50 uA p p differentially between RA and LL Tall T wave rejection 3 12 1 c T wave of 0 8 mV amplitude will not affect heart rate determination Heart rate averaging 3 1 2 1 d Averages six of the most recent eight detected R R intervals excluding the longest and shortest of the eight intervals Response to irregular rhythms 3 1 2 1 e a Ventricular bigeminy 80 b min rhythm b Slow alternating ventricular bigeminy 30 b min c Rapid alternating ventricular bigeminy 80 to 10
19. CO2 TEST CO2 Pressure in mmHg ambient cell 745 740 Pump Voltage Volts 3 0 CO2 mmHg 0 00 Status Set Barometric Pressure mmHg 745 Flow Rate Check and Calibration CO2 Calibration 07 16 2002 15 13 Pump Op Time 58 Return 5 12 Configuring the Power up Defaults Menu Diagnostic Mode The CO Test Screen is described below Menu Item Function CO Pressure in mmHg ambient cell Displays the real time value of the ambient pressure and the measured cell pressure The ambient pressure value will be the same as the Barometric Pressure value This value CANNOT be changed or reset Pump Voltage Volts Takes the correct amount of air from outside of the monitor This value CANNOT be changed or reset CO mmHg Displays the real time value of CO This value CANNOT be changed or reset Set Barometric Pressure mmHg Select a value which matches the reference value received from a reliable source airport weather station hospital weather station Confirm that the ambient pressure value is the same as the barometric pressure value Flow Rate Check and Calibration Adjust CO flow rate by Decrease Flow or Increase Flow After desired flow rate is obtained select Store Flow Highlight and press Return to exit pop up menu CO Calibration Displays the date mm dd yyyy and time that the last calibration occurred To calibrate CO you must first set the percentage of CO th
20. Disposing To avoid contaminating or infecting personnel the environment or other equipment make of the sure you disinfect and decontaminate the monitor appropriately before disposing of it in Monitor accordance with your country s law for equipment containing electrical and electronic parts For disposal of parts and accessories such as ECG and SpO where not otherwise specified follow local regulations regarding disposal of hospital waste For disposal of the lead acid battery follow local regulations for safe disposal of lead Warning Explosion Hazard Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide Sl Unpacking the Monitor The box comes with the following e the C3 Instructions for Use guide e the C3 patient monitor an AC power cord In addition you should receive all of the accessories and options that you have ordered If there is anything missing contact your Philips Medical Systems representative immediately Checking the Shipment Examine the carton carefully for evidence of damage in transit If you discover any damage contact the carrier immediately Retain all packing material If you have to return the monitor contact the Philips Response Center or your local Philips representative for shipping instructions To pack the monitor for return disconnect all cables Pack the monitor in its original shipping carton If this is una
21. Leakage Current Patient Leakage Current with Mains Voltage on the Applied Part At least once every 2 years after any repairs if the monitor has been dropped or opened 4 6 Testing the Monitor a Only when in use as part of hospital protocol safety checks Test Map Test Map The Test Map shows which tests are required in which situations Service Event When performing Test Block Required complete these test Installation Visual Power On Monitor exchange Visual Power On Monitor opened Power On Pneumatic Leakage Test All Safety tests Battery replacement Power On Battery Performance All Safety tests Speaker replacement Power On Pneumatic Leakage Test All Safety tests Backlight tube replacement Power On Pneumatic Leakage Test All Safety tests NBP pump replacement Power On Pneumatic Leakage Test All Safety tests Preventative Maintenance Tests SpO module replacement Power On Pneumatic Leakage Test All Safety tests SpO Tests CO module replacement Power On Pneumatic Leakage Test All Safety tests CO Tests Power supply module replacement Power On Pneumatic Leakage Test Safety Tests Patient monitoring I O module replacement Power On Pneumatic Leakage Test All Safety tests All Performance tests Communications mod
22. NBP patient monitoring input connector 4 Place the monitor in Diagnostic Mode with the NBP test screen active Pressure Transducer Accuracy Test Pneumatic Leakage Test Preventative Maintenance Tests This test verifies the pressure accuracy of the monitor s pressure transducer Step Action Confirm that the simulator displays Leak Test see steps 1 and 2 in the previous table Press the Volume button to ensure that both valves are closed Perform an offset adjustment so that the simulator and monitor both display a pressure of 0 mmHg or kPa by doing the following a Press the Contrast button on the monitor s front panel b If needed zero the simulator see To Zero the Simulator on page 4 16 Press Start on the simulator Allow 15 20 seconds for the pressure to stabilize The pressure displayed on the monitor and on the simulator should be within 5 mmHg 0 7 kPa of one another Document the difference between the simulator value and the monitor value X1 Press the Stop button on the simulator to stop the test Press and hold the Volume button until the monitor displays a pressure of 0 mmHg or kPa If no further NBP tests are to be conducted turn the monitor Off Normal monitoring operation returns the next time the monitor is turned On This test verifies the integrity of the pneumatic system Step Action Turn the simulator on
23. No response when pressing the buttons on the front panel Keypad malfunction or keypad connector is loose 1 If only one button does not work verify that the monitor is On 2 Check the keypad connector 3 Replace the keypad according to the instructions in Removing the Keypad on the Front Panel on 7 6 Training 11 11 Disassembly Audible alarm does not sound Speaker malfunction or speaker has a loose connector Main PCB module malfunction 1 Check speaker connector 2 Replace speaker according to the instructions in Replacing the Speaker on 7 9 Replace the Main PCB module according to the instructions in Removing the Main PCB Module on 7 12 No sound from monitor Tone is switched off or low 1 Turn the volume up by pressing the Volume button 2 Power cycle monitor Tone should sound when monitor boots up 3 Check cable connections to the speaker No tone alarms Speaker will not work Audible alarms are not switched on Cable disconnected Defective speaker Make sure audible alarms are switched on Reconnect cable Replace speaker The wheel is rotated no highlight appears on the display screen and or the monitor does not respond to wheel presses Main PCB module malfunction Replace the Main PCB module according to the instructions in Removing the Main PCB Module on 7 12 Whenever the monitor has been ope
24. OPEM e aos 3 10 Why is Battery Conditioning Necessary 20 00 c cece eee eee eens 3 10 When Should Battery Conditioning be Performed 0 00 00 eee 3 10 How to Condition a Battery ooooooooooororrrr rr 3 10 Battery INOP Messages i ti A AAA e 3 11 4 Testing the Monitor ri bas as EL Testing Safety Information seess r oree e e E t eee t ene n ene 4 2 Objectlyesss thee ALES S Ls lees ido eet teet es 4 3 IB 4 3 Functionality Assurance ente soe aS pha wR RR 4 3 Preventative Maintenance 0 ence ene he 4 3 Performance and Safety Tests 0 0 n eens 4 3 Testing Checklist eee A Re GRE EA 4 4 Test Reporting e ee AREE 4 5 Recommendations for Testing Frequency ooooooooorr eh 4 6 Test Maps dede et ehe V idee ede oU dee e se ed TC Oodd eere E 4 7 Equipments ud setup E IM REI ee dte te doas etm rU HUP eA Ig de 4 8 Seral NUMDETS en onn tar pte Shihan ste iG nie ES Reg aS 4 9 Passwords si 0 326 bet tae A UB A VR 4 9 Visual Testa i2 so ttes tee ta AP AR UM e 4 10 Power On Test cee ted nein ce ER pce e reU etr iore toe ep ANDRE AI d 4 10 Functionality Assurance Tests 2 0 2 0 cece cece tenet tenn eens 4 11 Contents 2 Performance Assurance Test 4 11 Poweron Self Test at A A A ROR AS 4 11 A o A E e e eee ig 4 12 Volume Control Test lllo etn a RnR 4 12 Preventative Maintenance Tests ei hed A a QU TE Re 4 14 Documenting NBP Test Results 0 0 0 cece en 4 15 Pneumatic
25. SpO2 Module on page 7 14 SpO Erratic SpO measurement is erratic 1 Make sure the SpO sensor is correctly placed 2 If this does not solve the problem make sure the sensor is working SpO Interference The level of ambient light is so high that the transducer cannot measure the pulse or the cable is picking up interference 1 Cover the SpO sensor so that it does not get as much ambient light 2 If this does not solve the problem make sure the sensor cable is not damaged SpO No Transducer Monitor does not detect SpO sensor Make sure the SpO sensor is connected Troubleshooting 6 7 Part 2 Isolating and Solving Monitor Problems Message Cause of Failure Remedy SpO Noisy Signal patterns Excessive patient movement or electrical interference are causing irregular pulse Try to reduce patient movement or to relieve the cable strain on the sensor SpO Non Pulsatile Pulse is too weak or is not detectable or the application site is too thin Change the application site of the sensor or stimulate circulation at the current site SpO Transduc Malf Sensor is malfunctioning 1 Change the SpO sensor as soon as possible 2 Return the faulty sensor to your biomedical department Temperature Probe Disconnect reading Monitor is not detecting temperature Check that the temperature probe is properly c
26. System Functionality 2 0 0 0 0 cece ee 4 16 To Zero the Simulator 0 0 cette teens 4 16 Pressure Transducer Accuracy Test 0 0 0 0 e cee ce cect eee nee nee 4 17 Pneumatic Leakage Test 2 0 0 ccc cect s 4 17 Inflation Rate Testi do Se P EN ee SE 4 18 OverPressurev Testing ue cai a dis ean A plata 4 19 Detlation Rate Test tance A A a Eee Oe 4 20 Basic Pneumatic Leakage BPL Test 0 00 4 21 Performance Procedures RS a ee R 4 22 Battery Performance Test 0 0 0 ce ccc eee tenet enna 4 22 Temperature Tdi eet Cher gue vs ro x URP EI Roe NI ba E 4 23 ECG Respiration Performance 0 0 ccc cette nent eens 4 25 ECG Tests ocx be Se vm 4 25 Respiration Test e Let ues ip ds 4 26 SpO2 Performance srir ch ere Rete ecd t hha tte aS ls CQ e Rees 4 27 Dynamic Operating Range Test 2 0 00 c cece ee 4 27 LED Excitation lest it ee ee 4 28 COZ Performance iin Lud et eee etu e te t an tients 4 29 Documenting CO2 Test Results 0 0 eect E 4 30 Barometric Pressure Check and Calibration 0 0 000 eee eee eee ees 4 31 Leakage Check 2 2601 nis Se hee Sahay Sate oe a te es 4 32 Pump Checo e Bek ek ets aan ed oA 4 33 Flow Rate Check and Calibration 0 000 ce cece ccc eee 4 33 CO2 Gas Measurement Calibration Check eese 4 34 Calibration Verification 0 0 cece n ene n ees 4 35 Reset Pump Operating Time Counters 0 00 cece cette es 4 35 Serial Interface and Nurse Ca
27. V219 a 123456 Cycle 369 Return When in the System Information screen the wheel can be rotated to select any of the changeable items If one of those items is selected a press of the wheel displays a popup menu The first item is Make No Change the second item is Reset to Zero This screen displays several system related items They are as follows 5 6 Configuring the Power up Defaults Menu Diagnostic Mode Menu Item Function Monitor On Time Displays the number of hours rounded to the nearest hour that the Main PCB has been operational This value CANNOT be reset Backlight On Time Displays the number of hours rounded to the nearest hour that the LCD Backlight has been operational This value can be reset to zero An example is when a technician changes the backlight or installs a new LCD Recorder On Time Displays the number of hours rounded to the nearest hour that the C3 recorder has been operational This value can be reset to zero An example is when a technician installs a new recorder Battery Deep Discharges Displays the number of deep discharge cycles seen by the battery The monitor records a deep discharge cycle when the battery volt age reaches 5 6 volts At this voltage a Low Battery alarm is issued This value can be reset to zero An example is when a technician installs a new battery System Software Version Displays the revision level of the system software The revisio
28. and decontaminated See the section titled Electrical on page 12 3 for all electrical specifications The C3 patient monitor is classified as Class I Equipment with an internal power source according to IEC 60601 1 CSAC22 2 601 1 UL 2601 1 which means that it is a monitor included in the protective grounding protective earth system of the room by way of grounding contacts in the power plug Warning Disconnect the monitor from the AC source by unplugging the power cable from the AC power connector located on the rear of the monitor The On Standby button does not disconnect the monitor from the AC mains supply To eliminate potential differences between difference pieces of equipment the monitor must have a separate connection to the equipotential grounding system One end of the equipotential grounding cable potential equalization conductor is connected to the equipotential grounding post on the rear of the monitor and the other end is connected to one point of the equipotential grounding system All combinations of medical equipment with non medical equipment must comply with IEC 60601 1 1 Warming 0 Accessory equipment connected to the monitor s data interface must be certified according to IEC Standard 60950 for data processing equipment or IEC Standard 60601 1 for electromedical equipment All combinations of equipment must be in compliance with IEC Standard 60601 1 1 systems requirements Unpacking the Monitor
29. as the following may appear cardioversion vfib cardioversion late 4 38 Testing the Monitor Patient Safety Tests Patient Safety Tests Ground Integrity Philips safety tests meet the standards of and are performed in accordance with IEC 60601 1 Clause 19 EN60601 1 Second Edition 1988 Amendment 1 1991 11 Amendment 2 1995 03 The C3 patient monitor is a Class I device It requires a protective earth ground wire Keep this in mind when performing the following test procedures There are two categories of safety tests e Ground Integrity Electrical Leakage Table 4 3 Ground Integrity Test or Inspection to Perform Expected Test Results Protective Earth See Safety Test Diagram Protective Earth on page 4 40 With mains cable Maximum impedance x lt 100 mOhms Testing the Monitor 4 39 Patient Safety Tests Figure 4 1 Safety Test Diagram Protective Earth Test or Inspection to Perform Instrument under test L N e N L Ri el i me Applied part TIR Ohm 6v Insulating pad ENG 50 Hz 25 Aor 1 5 Ir If equipotential connection present measure also with yellow green E P conductor connected Measures impedance of Protective Earth PE terminal to all exposed metal parts of IUT which are for safety reasons connected to the Protective Earth PE Max
30. cee cece teens 7 12 Removing the Patient Monitoring I O Module 0 0 eee eee eee 7 13 Removing the SpO2 Module 0 0 cee cece cette eee nee 7 14 Removing the CO2 Module oococcooccccccocc teen ene 7 15 Removing the NBP Pump 2 0 cece eee t p EE nee 7 16 Removing the Power Supply Module 0 0 cc cee eee cece 7 17 8 Spare Parts GA ce git A ta Ad ete ee we pile veda RA ACA paced ce ae OTL Small Parts Kit i pe hai ae ea A EA eee a DU NUR Pipe et 8 2 ES Top Aseve Assembly ii tet acl aa ae aen aa 8 3 Front Panel Assembly cio Sic Ca Be A ab es ERI RR 8 4 Rear Panel Assembly 0 0c cece cence he 8 6 Power Cords uode Rue A ee 8 9 Exchange Unit Part Numbers 000 e e 8 10 9 Packing for Shipment 5 6 3 6 dsb ek da del General Instructions 1v ide a A Sere eed ER E 9 2 Repacking the Original Carton 2 0 0 eh 9 2 Repacking in a Different Carton 2 0 1 eee m 9 2 10 RS 232 Interface coxis coe ches e eiere a IL General RS 232 Safety Information 20 0 cece cece tenes 10 2 About th RS 232 Interface cosas a lee e 10 3 Cable Connections ysa area a 10 3 Nurse o nca mE 10 4 Contents 5 Ti Drain 22222 ee nee A ee Rae ee a a ow ee eden ue LIE Introd ction cuc A RERO A AA eae hea joe ey 11 2 Leyelsof Involvement 0 2 4 och ose A E t 11 2 Preventative Maintenance Only 0 ccc cette tte 11 2 Phone Support or Service Sch
31. conducted turn the monitor Off Normal monitoring operation returns the next time the monitor is turned On 4 20 Testing the Monitor Preventative Maintenance Tests Basic The purpose of this test is to verify the integrity of the NBP pneumatic system after the Pneumatic monitor has been opened This includes all external and internal tubing connections Leakage BPL Test No simulator is required for this test and results are not required to be reported Step Action 1 Attach the NBP cuff to the NBP tubing and the tubing to the NBP patient monitoring input connector 2 Wrap the cuff around itself and place it on a table for the test DO NOT place the cuff on your arm 3 Turn the monitor On and enter the NBP test screen located within Diagnostic Mode 4 Press the Volume button to close valves 5 a Press and hold the NBP Start Stop switch until the monitor s screen reads 250 mmHg or 33 3 kPa b Wait 15 20 seconds to allow the pressure to stabilize 6 a Note the value on the screen P1 then start a timer b After one minute note the value on the screen P2 c Calculate the difference between the two values P1 P2 The value should be less than or equal to 6 mmHg 0 8 kPa 7 Press and hold the Volume button until the screen shows the pressure has released and the value is 0 mmHg or kPa 8 Turn the monitor Off Note If the test fails ensure the integrity of the cuff and t
32. hardware failures to the level of the field replaceable part In addition this chapter describes how to troubleshoot problems using error codes Caution Only qualified service personnel should open the monitor housing remove and replace components or make adjustments If your medical facility does not have qualified service personnel contact the Philips Response Center or your local Philips representative This chapter is divided as follows Part 1 Troubleshooting Checklist This part provides check for the obvious hints and tips Part 2 Solving Monitor Problems This part provides the following e Tables that list monitor symptoms as well as causes and remedies for monitor faults Techniques for troubleshooting the monitor before you disassemble it e Checks you can perform to isolate problems down to a specific replaceable part for example the power supply Part 3 Using Support Functions This part provides information on how to use error codes as a support feature This is the essential diagnostic step to be taken prior to any repair These are the types of messages that generate and display to inform the user when the monitor has been unable to perform an operation Troubleshooting 6 3 Part 1 Troubleshooting Checklist Part 1 Troubleshooting Checklist Checks for Obvious Problems Checks Before Opening the Monitor 6 4 Troubleshooting If any unit is not functioning properly your firs
33. or 13 CO whichever is lower Flow Rate 50 7 5 ml min Warm up Time 180 seconds max typically 30 seconds Gas Sampling Delay Time 2 7 seconds Resolution Numeric 1 0 mmHg 0 1 kPa Wave 0 1 mmHg 0 01 kPa Rise Time 190 ms for adult mode measured with sample line for humidified ventilation and airway adapter for adult Calibration Interval 4 000 operating hours Auto Zero Interval lt 2 duty cycle Increases if the temperature change by 8 C from the previous Auto Zero interval or the pressure changes by 45 mmHg Leak Tightness lt 40 mbar min with 30 vacuum Accuracy 0 20 minutes 0 to 38 mmHg 4 mmHg 38 to 99 mmHg 12 of indicated value gt 20 minutes 0 to 38 mmHg 2 mmHg 38 to 99 mmHg 5 of indicated value 0 08 for every 1 0 mmHg above 40 mmHg a Accuracy applies for respiration rate up to 80 b min For respiration rate above 80 b min the accuracy complies with EN 864 ISO 9918 4 mmHg or 12 of indicated value whichever is greater for End tidal CO values exceeding 19 mmHg The above accuracy is maintained to within 4 for the following gas mixtures CO 0 13 Ny 0 97 5 O5 0 100 N50 0 80 HO Dry to saturated Specifications 12 9 Measuring and Displaying Parameters Trends 12 10 Specifications Parameter Specification Type Graphical and tabular Memory Storage 12 hours
34. signal The nurse call signal reacts when a low medium or high level alarm is activated The rear panel connector for the Defib Sync Pulse is keyed so that the connection of a cable can be detected by the processor When a connection is detected the processor software initiates the generation by hardware of a TTL compatible pulse capable of driving 1 TTL load over a three meter cable with less than 200 pF capacitance The defib pulse is triggered by the detection of the R wave in the QRS sequence of the ECG wave form complex The pulse signal is active for 100 10 ms The speaker is capable of providing 73 dBA of volume at a distance of one meter during alarm conditions The processor drives the speaker in different patterns as specified for the different alarm priorities and conditions Refer to the C3 Instructions for Use guide for descriptions of alarm responses Recorder ECG Processing Respiration Processing Theory of Operation and System Architecture The optional recorder module is installed in the right panel of the monitor Refer to the C3 Instructions for Use guide for printing procedures It provides users with the capability to obtain hard copy records of selected vital sign information Basic control of the recorder is implemented by two push button controls on the recorder s front panel The Continuous button is used to obtain continuous recordings of the real time waveforms displayed in the top two graphic frame
35. the Main PCB Module on 7 12 Fluid dripping from the monitor LCD Display damaged or cracked Replace LCD Display Decrease in light intensity on display Brightness decreases from top to bottom or bottom to top Backlight tube is worn out Replace backlight tube according to instructions in Replacing the Backlight Tube on page 7 9 No tone alarms Speaker will not work Audible alarms are not switched on Cable disconnected Defective speaker Make sure audible alarms are switched on Reconnect cable Replace speaker The wheel is rotated no highlight appears on the display screen and or the monitor does not respond to wheel presses Main PCB module malfunction Replace the Main PCB module according to the instructions in Removing the Main PCB Module on 7 12 6 10 Troubleshooting Part 3 Using Support Functions Part 3 Using Support Functions Serviceable Hardware Error Codes When the monitor detects an error condition the monitor shows an error code on the display screen If such an error occurs during monitoring operation an audible alarm tone also sounds Press the Alarm Silence button to terminate the audible alarm tone When an error code appears on the display a number in hexadecimal representation indicates the nature of the error Additionally Diagnostic Mode can be used to gain access to an error code record stored in non volatil
36. the assembly level For this purpose they should review this high level information Please be aware that Philips only stocks assembly level parts The monitor is a full function monitor for use on adult and pediatric patients The functions performed by the system include monitoring patient ECG heart rate respiration rate blood pressure blood oxygen saturation carbon dioxide and temperature In addition to monitoring and displaying the status of these physiological parameters the instrument performs various microprocessor programmed analytical functions such as e Creating both visual and audible alarm signals when set limits are violated Creating and displaying warning messages when conditions are detected that would degrade or prevent valid measurements Creating and displaying trend waveforms or tabular data Providing a synchronizing pulse for defibrillator operation Providing input to an optional recorder for printout of current or trend waveforms or tabular data The monitor is essentially a battery powered instrument An internal charging unit is designed to accept an external AC line voltage The charger uses the external power source to maintain a float voltage source available from the batteries The functions are represented graphically in the following figure Each section of the System Block Diagram is described briefly in the text that follows the illustration This is followed by more detailed desc
37. to the NBP input circuitry As the pressure is reduced blood flows in the previously occluded artery and changes the measurements made by the transducer The point at which oscillation increases sharply is defined as systolic pressure As the cuff continues to deflate oscillation amplitude increases to a maximum and then decreases The peak oscillation amplitude is defined as the mean arterial pressure The point at which the system detects a rapid decrease in oscillation is defined as the diastolic pressure Measurement of oxygen saturation in the blood uses a specrophotometry technique It is based on the facts that oxyhemoglobin and deoxyhemoglobin differ in their absorbtion of red and infrared light and that the volume of arterial blood in tissue changes during the pulse Using these facts a pulse oximeter passes red and infrared light into an arteriolar bed and measures changes in light absorption during the pulsatile cycle The light sources are red and infrared light emitting diodes LEDs while the detection is accomplished by a photo diode To identify the oxygen saturation of arterial hemoglobin the monitor uses the pulsatile nature of arterial flow During systole a new pulse of arterial blood enters the vacular bed and both blood volume and light absorption increase During diastole blood volume and light absorption reach their lowest point The measurement is based upon the difference between maximum and minimum absorption fo
38. 3 b min d Bi directional systoles 90 to 105 b min Heart rate meter response time 3 12 1 f a Change from 80 to 120 b min 4 to 5 sec b Change from 80 to 40 b min 6 to 8 sec Time to alarm for tachycardia 3 12 1 g Waveform 4 a Amplitude Average Time to Alarm 0 5mV 8 6sec 10mV 7 0 sec 20mV e 3 0 sec Waveform 4 b Amplitude Average Time to Alarm 10mV 8 0 sec 20mV e 24 sec 40mV 24 sec Specifications 12 5 Measuring and Displaying Parameters Respiration 12 6 Specifications Standards Pacemaker pulse With the exceptions noted below the monitor will reject all single and rejection double pacemaker pulses either 150 or 250 msec apart of amplitudes 2 to 3 1 4 1 700 mV from 0 1 to 2 0 msec with and without under overshoot 3 1 4 2 Following are pacer pulse conditions that the monitor will reject but which are less than the maximum range listed in the standard Pacer pulse width 2 msec Pacer amplitude lt 400 mV Double pulse either 150 or 250 msec apart without over undershoot no QRS Pacer pulse width 2 msec Double Pacer amplitude lt 500 mV pulse either 150 or 250 msec apart without over undershoot ineffective pacer Pacer pulse width 2 msec Single Pacer amplitude lt 125 mV or Double pulse either 150 or 250 msec apart with 2 mV over undershoot no
39. 6 Turn the monitor Off 4 24 Testing the Monitor ECG Respiration Performance ECG Test Performance Procedures This section includes tests for both ECG and Respiration Tools Needed for both ECG and Respiration testing ECG leads M1605A AAMI or M1615A IEC ECG trunk cables M1540C ICU AAMI or M1550C ICU IEC Dynatech Nevada MedSim 300 or equivalent simulator The accuracy of the monitor s ECG measurements is 5 bpm In the following procedure add the tolerance of the simulator to the acceptable range of readings Step Action Verify that the monitor is turned Off Connect the ECG leads to the appropriate jacks on the ECG simulator Connect the leads to the ECG trunk cable Connect the cable to the ECG patient monitoring input connector Set the ECG simulator as follows Set To Heart Rate HR 30 bpm Lead select II Press the On Standby button to turn the monitor On After the power up sequence verify the following monitor reactions a After at least five heartbeats the monitor displays a heart rate of 30 5 bpm b The audible alarm sounds and the Heart Rate frame flashes indicating heart rate is below the default lower alarm limit Press the Alarm Silence button twice on the monitor s front panel to invoke the Silence Reset mode Increase the heart rate setting on the ECG simulator to 240 bpm After at least five heartbeats verify that th
40. 8 Testing the Monitor Over Pressure Test Preventative Maintenance Tests This test verifies the functionality of the over pressure relief system of the monitor Step Action Turn the simulator on Perform the following sequence a Press the Home button b Press the Pressure Tests button c Press the Pressure Relief button d Press the Setup button e Press 3 3 5 then Enter Confirm that the simulator is active and displays Relief Valve Test Press the Volume button to ensure that both valves are closed Perform an offset adjustment so that the simulator and monitor both display a pressure of 0 mmHg or kPa by doing the following a Press the Contrast button on the monitor s front panel b If needed zero the simulator see To Zero the Simulator on page 4 16 Press the Start button on the simulator The peak value X4 shown on the simulator should be between 270 mmHg and 330 mmHg 35 9 kPa and 43 9 kPa The pressure should return to 0 automatically on the monitor and simulator Both conditions should be true for the monitor to pass this test If no further NBP tests are to be conducted turn the monitor Off Normal monitoring operation returns the next time the monitor is turned On Testing the Monitor 4 19 Preventative Maintenance Tests Deflation Rate This test verifies the deflation rate of the monitor Test Step Action 1 Turn the simulator o
41. A ee Sa RO ee no ee ey 11 2 LO Materials 222 i pc tech crete coh o iet e 11 3 Essential Materials luis it eae Ss gk Saco eae cece ses 11 3 Optional Materials 5 oo oe Let estere e Gs cabled p ERE etl beh 11 3 OVEIVIEW outer Hatin RR RP A A Lona as Canepa 11 3 SUPPLY catenin SE yea 11 4 Wnit Ex changers asta dentes Me ee 11 4 Bench Repairs sel td ter es 11 4 Theory of Operation and System Architecture 00 eee cece eee eee 11 5 System Overview 5 esas eerte A erin Sdn eon Me ents 11 5 Block Diagraiy ic oe By eR APE Ghee RS Pe ee eee 11 5 Isolated Front End 1 tl et pe pe oie 11 7 NBP Eront End iii iate ker xa epe tid E ia tie EDU Ree eg 11 7 POWER Systema cae DECUS ea les ERS i hg RECTE SLVR A NN BOR ees 11 7 Micro processor Memory and Control 00 cece cect ees 11 7 Display 224 a o re aut e M e e ld e ae eats ad See el dats 11 7 Keypads is A e test este equ et peta e 11 7 Navigation Wheel oS ii gd ges 11 8 RS aaa es 11 8 Detib Sync Pulse te ER hoe 11 8 Speaker O IS a teg 11 8 RECO it E 11 9 EGG ProCeSSIDg 5 onRPOSE C RR E E IER A A did 11 9 Respiration Processing 0 cece cette cette s 11 9 NBP Processing A A A eG tut 11 10 SpO2 Processing A A oa e de dat elit 11 10 72 Processing vo o Sy Uer rien HR SU ee ds 11 10 Temperature Processing 0 cee eect cette tenet teens 11 10 Troubleshooting teer A a ena oo ial Eee Se CS 11 11 Disaster ie Noes saver hae do an det de lOs a ia 11 12
42. DUCT FOR A FULL REFUND PHILIPS SOFTWARE LICENSE TERMS The following License Terms govern your use of the accompanying Software unless you have a separate signed agreement with Philips Medical Systems License Grant Philips Medical Systems grants you a license to Use one copy of the Software Use means storing loading installing executing or displaying the Software You may not modify the Software or disable any licensing or control features of the Software If the Software is licensed for concurrent use you may not allow more than the maximum number of authorized users to Use the Software concurrently Ownership The Software is owned and copyrighted by Philips or its third party suppliers Your license confers no title to or ownership in the Software and is not a sale of any rights in the Software Philips third party suppliers may protect their rights in the event of any violation of these License Terms Copies and Adaptations You may only make copies or adaptations of the Software for archival purposes or when copying or adaptation is an essential step in the authorized Use of the Software You must reproduce all copyright notices in the original Software on all copies or adaptations You may not copy the Software onto any public network No Disassembly or Decryption You may not disassemble or decompile the Software unless Philips prior written consent is obtained In some jurisdictions Philips consent may not be required fo
43. EG Bi cies AS 6 3 Part 1 Troubleshooting Checklist llle 6 4 Checks for Obvious Problems e 6 4 Checks Before Opening the Monitor 00 cece e 6 4 Part 2 Isolating and Solving Monitor Problems 2 0 0 0 c cece cent eens 6 6 INOR Messa cess uN Sern DAR Ee 6 6 Isolating the Defective Component 0 0 nunne 6 8 Part 3 Using Support Functions 0 0 0 0 0 ccc cette teen eens 6 11 Serviceable Hardware Error Codes 0 cece eee co 6 11 Error Code Catriel s 6 12 Other Error Codes 22 dee ERR cI es 6 13 7 Disassembly a4 i 3 E ee ee EE Nees eee ese ado Introductorio rs aati ek ah da 7 2 Tools Required iii A IER edd eae et EE 7 2 Disassembly Safety Information 00 0 cece rr 7 3 Disassembly Procedures ti d p a RE a EEE Sea 7 4 Closed Case Disassembly Procedures 00 ccc cece eese 7 4 Removing the Battery 0 0 0 ccc ect eee teen eens 7 4 Removing the Navigation Wheel 0 0 0 cece eee teens 7 5 Removing the Keypad on the Front Panel 0 cece ee 7 6 Contents 4 Removing the Optional External Recorder 7 6 Separating the Front from the Rear Case 00 ccc tte eh 7 7 Front Case Disassembly iii a ee Be uer pde vg I IHRE ER 7 8 Replacing the Speaker 2 0 1 ec cent teenies 7 9 Replacing the Backlight Tube 7 9 Rear Case Disassembly 2s BUS UIS a T ag hn Nigh se See R a N 7 10 Removing the Main PCB Module 0 0
44. Monitor to the Wall Channel Attaching the Monitor to the Wall Channel 2 10 Site Preparations Step Action 1 Slide the Tilt Swivel mount into the Wall Plate as illustrated by the above diagram 2 Make sure the latch snaps into the hole on the upper right hand comer of the wallplate 3 Maintaining the Monitor This chapter describes how to clean your monitor and how to keep your monitor in the best working condition Maintaining the Monitor 3 1 Maintenance Safety Information Maintenance Safety Information Warning Follow local governing ordinances and recycling plans regarding disposal or recycling batteries and other device components If the battery shows signs of damage or signs of leakage replace it immediately Do not use a faulty battery in the monitor Never use a monitor that is monitoring a patient to perform battery conditioning Caution Do not immerse the monitor in liquid or use caustic or abrasive cleaners Do not spray or pour any liquid on the monitor or its accessories Do not allow any liquid to penetrate connectors or openings in the monitor s chassis 3 2 Maintaining the Monitor Objectives Concepts Light Maintenance Battery Conditioning Objectives In order to meet this chapter s objectives you should be able to perform light maintenance and preventative maintenance for the monitor through the following tasks I
45. Philips C3 Patient Monitor C3 Service Guide Models 862474 862478 Part Number 989803129451 Printed in the U S A March 2003 Edition 1 IDA A LL LU PHILIPS About this Manual Proprietary Information Warranty Copyright Printing History Additional Documents This document contains proprietary information which is protected by copyright All Rights Reserved Reproduction adaptation or translation without prior written permission is prohibited except as allowed under the copyright laws Philips Medical Systems 3000 Minuteman Road Andover MA 01810 1085 978 687 1501 Publication number 989803129451 Printed in USA The information contained in this document is subject to change without notice Philips Medical Systems makes no warranty of any kind with regard to this material including but not limited to the implied warranties or merchantability and fitness for Philips Medical Systems shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Copyright 2003 Philips Electronics North America Corporation Microstream is a registered trademark of Oridion Medical Inc New editions of this document incorporate all material updated since the previous edition Update packages may be issued between editions and contain replacement and additional pages to be merged by a revision date at the bo
46. Procedure 8 453563480091 Front Housing No Disassembly Procedure Spare Parts 8 5 Rear Panel Assembly Rear Panel Assembly Red to Terminal Black to Terminal 13 Figure 8 4 1 of 3 Item No Part Number Description ae 9 453563480211 Spare Recorder Blanking Plate page 7 6 10 453563480011 AC Connector Power Entry Module No Disassembly Procedure 11 453563480171 RS 232 Port No Disassembly Procedure 12 453563480041 Battery Housing page 7 4 13 453563480071 Battery Cover page 7 4 14 453563480031 Battery page 7 4 8 6 Spare Parts Figure 8 5 2 of 3 Rear Panel Assembly Page Ref for Item No Part Number Description Disassembly 15 453563480191 Side Panel with CO page 7 15 NotShown 453563480181 Side Panel without CO No Disassembly Procedure 16 453563480121 NBP Assembly page 7 16 16 453563480131 NBP Cable page 7 16 17 453563480141 Patient Monitoring I O Module page 7 13 Spare Parts 8 7 Rear Panel Assembly 8 8 Spare Parts Figure 8 6 3 of 3 Page Ref for Item No Part Number Description Disassembly 18 453563480161 Power Supply Board page 7 17 NotShown 453563480061 Communication Board page 7 17 Power Cords Power Cords Part Number Description 8120 5429 Power Cord US 903 2 4 m 8120 1689 Power Cord Europe 902
47. QRS Pacer pulse width 2 msec 250 Pacer amplitude lt 300 mV msec spaced double pulse with 2 mV over undershoot effective pacing Pacer pulse width 2 msec 150 Pacer amplitude lt 220 mV msec spaced double pulse with 2 mV over undershoot ineffective pace Pacer pulse width 2 msec 250 Pacer amplitude lt 120 mV msec spaced double pulse with 2 mV over undershoot ineffective pace Parameter Specification Technique Trans thoracic impedance Range 3 to 150 breaths min Accuracy 3 breaths min Leads RA to LA Display Sweep Speeds 6 25 12 5 25 mm s Lead Off Condition detected and displayed Measuring and Displaying Parameters NBP Non Invasive Blood Parameter Specification Pressure Standards Meets performance standards of ANSI AAMI SP10 1992 Technique Oscillometric Cuff Pressure Display 10 to 300 mmHg Pulse Rate Range 40 to 200 b min Blood Pressure Accuracy Mean error and standard deviation per ANSI AAMI SP10 1992 Pulse Rate Accuracy Greater of 2 BMP or 2 Initial Cuff Inflation 180 mmHg Subsequence Cuff Inflation Prev SYS 30 mmHg Blood Pressure Measurement Range Systolic 60 to 250 mmHg Diastolic 40 to 220 mmHg Mean Arterial Pressure 45 to 235 mmHg Measurement Modes Auto Automatic measurements at intervals of 1 3 5 10 30 60 and 90 minutes Manual Single measurement initiated by Start Stop button
48. ance Tests To verify your monitor works properly perform the following test Step Action Using the supplied AC power cord connect the monitor to the AC power source Verify that the AC LED is lit Do not connect any patient monitoring input connectors or cables to the monitor If there are any such connections disconnect them If the monitor is Off press the On Standby button The monitor must perform the following sequence a The screen backlight illuminates b Three consecutively higher pitched chimes sound while the version numbers of the boot and operational software display c After successful completion the main monitoring screen displays No vital sign numeric values or waveforms display If any error codes display or the screen remains blank refer to Chapter 6 Troubleshooting After you first press the On Standby button the monitor displays a startup screen and conducts a set of self diagnostic test routines Testing the Monitor 4 11 Functionality Assurance Tests Alarm Test Tools Needed SpO adapter cable M1943A e Clinical Dynamics Corp SmartSat simulator with Nellcor simulator cable Step Action 1 Connect SpO simulator cable to the SpO adapter cable Connect the cable to the SpO patient monitoring input connector Set the simulator as follows Item Setting Oximeter Nellcor SpO0 7 81 BPM 36 Pulse Mod 0 50 Press t
49. art 2 Step Action Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet to the CO module gas outlet Leave the FilterLine connected to the CO module inlet Block the inlet of the FilterLine using your fingertip and observe the flowmeter display The value on the flowmeter X3 should decrease to between 0 and 4 ml minute If the value is within the tolerance limits there are no leakages and the leakage check is completed You can proceed to the Pump Check If the value is not within the tolerance limits there is a leakage between the FilterLine inlet and the pump inlet Check the FilterLine connections and open the monitor to check the tubing connections at the pump inlet and the CO module gas inlet If the connections are good try replacing the FilterLine and repeating the leakage check If the situation remains there is a leakage in the tubing and the CO module must be exchanged 4 32 Testing the Monitor Pump Check Flow Rate Check and Calibration Performance Procedures Step Action Connect the flowmeter inlet to the CO module gas outlet Connect the FilterLine to the CO module inlet Block the inlet of the FilterLine using your fingertip and observe the cell pressure to the right of the slash symbol on the top row of the CO Test display screen The cell pressure X4 should be more than 120 mmHg below the ambient pres
50. as follows Item Setting Oximeter Nellcor SpO0 81 BPM 70 Pulse Mod 5 00 a Power the monitor On Verify that the values are within the following tolerances Oxygen Saturation Range 79 to 83 Heart Rate Range 67 to 73 bpm Press the Alarm Silence button to temporarily silence the audible alarm Verify that the heart rate tone source found in the Heart Rate Menu is set to SpO Press the Volume button on the monitor s front panel Within 3 seconds of having pressed the button rotate the navigation wheel clockwise to verify that the beeping heart rate tone sound level increases Wait 3 seconds Press the Volume button and rotate the wheel counter clockwise and verify that the beeping heart rate tone decreases until it is no longer audible Rotate the wheel clockwise to return the beep volume to a comfortable level After 3 seconds with no wheel activity the volume adjust function terminates Testing the Monitor 4 13 Preventative Maintenance Tests Preventative Maintenance Tests Preventative maintenance refers specifically to the service tests required to make sure the monitor measurement results accurate In cases where the performance of NBP is in question or could have been configured during repair the complete set of NBP tests described in this service manual should be used The tests in this section verify the functionality of the monitor s pneumatic system All of these
51. ations Technical Data Sheet Training 11 13 Monitor Applications and Algorithms 11 14 Training 12 Specifications This chapter includes all hardware regulatory and measuring specifications for the monitor Specifications 12 1 Hardware Specifications Hardware Specifications Parameter Specification Size excluding handle Width 13 50 in 34 3 cm Height 10 25 in 26 0 cm Depth 7 50 in 19 0 cm Weight excluding accessories options cables 14 0 Ib 6 3 kg Display cold cathode fluorescent backlit Screen Type TFT color Active Screen Size 8 3 x 6 2 in 210 8 x 157 5 mm 10 4 in 264 2 mm diagonally Safety Standards 12 2 Specifications Resolution 640 x 480 pixels Recorder Type Thermal Weight 0 9 lb 41 kg Paper Width 50 mm Speeds 12 5 25 and 50 mm s Parameter Specification United States Federal Law restricts this device to sale by or on the order of a physician CE Marking Medical Device Directive 93 42 EEC IEC 60601 1 UL 2601 Can CSA C22 2 601 1 M 90 Protection Class Class L internally and externally powered equipment per IEC 60601 1 clause 2 2 4 Degree of Protection Type CF IPXI per IEC 60601 1 Mode of Operation Continuous Electrical Electrical Parameter Specification Power Sources Internal Battery
52. be calculated accurately and the low battery warning given at the right time Battery conditioning should be performed Every 200 deep discharge cycles We recommend two full charge and discharge cycles to ensure proper conditioning independent of the initial condition of the battery If you need to continue battery powered monitoring replace the battery in the monitor with a different charged and conditioned battery Y ou cannot condition a battery in a monitor that is being used Step Action 1 Insert the battery that is to be conditioned into a monitor that is not currently being used Disconnect the monitor from the AC power supply Turn on the monitor and leave it on until it switches off automatically Turn the monitor Off using the On Standby button located on the front panel Reconnect the monitor the AC power supply Nin RR Ww bd Charge the battery until it is full When you turn the monitor On the 5 bars in the battery gauge on the main monitoring screen will be illuminated T Repeat steps 2 to 6 3 10 Maintaining the Monitor Battery Maintenance Battery The following battery related INOP messages are issued by the monitor All Battery INOPs INOP continue until the monitor is plugged into the AC power or the INOP condition is fixed Messages e Low Battery This is an INOP which indicates that the remaining battery operating time is less than approximately 30
53. cusing on the pulsatile arterial blood In addition to the oximetry function the input signals may be used to calculate heart rate The Microstream CO board consists of an 80C552 Controller the memory system Flash ROM RAM PLA etc the Flow system FilterLine recognition system Inlet solenoid valve Measurement Cell Exciter IR Source Detectors and Temp Sensor and an analog section with ADC The gas inlet allows the connection of Microstream FilterLines The FilterLines are detected by the Optical Code Recognition Measurement of patient temperature is accomplished by processing the signal from a probe containing a resistor whose impedance is temperature dependent The class of such components is called thermistor The C3 is designed to accept the signals from electrically isolated Series 400 probes manufactured by Yellow Springs Incorporated Interchangeable probes in this series may be used for esophageal rectal skin or surface or airway temperature measurement Probes are furnished with a standard 10 feet lead Extension leads are available The signal from the probe is conditioned by the monitor s input circuitry processed and used to drive the numeric display Troubleshooting Troubleshooting Service personnel concerned with troubleshooting should review the following topics in this manual Error Codes see Error Codes on 6 3 Boot Sequence see Performance Assurance Test on 4 11 Per
54. d a defibrillator The ECG sync performance test is required once every year and when the monitor is repaired or when the monitor s parts are replaced Tools Needed Dynatech Nevada MedSim 300 or equivalent simulator Defib Sync Cable M4820A or Switchcraft 850 e Switchcraft 750 Two sections are needed to complete the connection from the monitor to the simulator 1 Ifa Defib Sync Cable is available connect it the ECG Sync output on the monitor then to the Switchcraft 750 or If a Switchcraft 850 is available connect it the ECG Sync output on the monitor then to the Switchcraft 750 2 Connect the Switchcraft 750 to the SYNC A PACE input on the simulator Connect to the monitor Connect to the simulator Defib Sync Cable Switchcraft 750 Switchcraft 850 Switchcraft 750 Step Action 1 Connect the ECG leads and trunk cables between the monitor and the simulator as described above 2 Turn on the monitor and the simulator 3 On the simulator main menu select DEFB then CARD to set the simulator to Defibrillator Tests Cardioversion 4 Press Start to begin the test If the monitor is working properly the following sequence of text messages will appear on the simulator over a period of approximately 3 seconds cardioversion afib cardioversion sync ok and cardioversion converted this message will persist on the simulator 5 If the monitor isn t working properly messages such
55. del QD Imin 85 93 Waveform mmHg Frames Numeric 183 107 Frames 149 180 C min 14 o Message Tr c 37 8 and Icon 7 Adult 01 06 03 01 09 17 Frames Overview 1 7 Additional Documentation Sedation Model O min 585 mmHg 183 107 149 so SpO Frames Numeric Frames mm etCO mmHg 33 m 23 37 8 Message and icon Adult 01 06 03 01 09 17 Frames Additional Documentation To perform test and troubleshooting procedures and to understand the principles of operation and circuit analysis sections of this manual you must know how to operate the monitor Refer to the C3 Instructions For Use guide to understand the various sensors ECG leads blood pressure cuffs CO parameters and accessories and temperature probes that work with the monitor Instructions for cleaning and caring for the accessories can also be found in the individual Directions for Use that accompany these accessories 1 8 Overview 2 Site Preparations This chapter describes how to perform site preparation and how to comply with safety guidelines and requirements Site Preparations 2 1 General Site Preparation Safety Information General Site Preparation Safety Information Warning To avoid contaminating or infecting personnel the service environment or other equipment make sure that equipment w
56. driver to remove the 3 screws fastening the speaker Pull the speaker up The backlight tube is located within the LCD screen However you do not need to disassemble the LCD screen display to replace the backlight tube Step Action Insert a small thin flathead screwdriver into the crevice between the LCD screen and where the backlight tube slides in Using the screwdriver grab the corner of the backlight tube and gently pull the tube out Re insert a new backlight tube Disassembly 7 9 Disassembly Procedures Rear Case Disassembly 7 10 Disassembly You must disassemble the rear case to remove the following modules Main PCB Patient Monitoring I O SpO CO NBP Communications Power Supply Caution Before you remove any module make sure you are wearing ESD protection and you are working in a grounded environment To disassembly the rear case Step Action 1 Verify that the battery has been removed If not see Removing the Battery on page 7 4 2 Verify that the recorder has been removed If not see Removing the Optional External Recorder on page 7 6 If there is no recorder unsnap the recorder door using a flathead screwdriver see figure below Disassembly Procedures Step Action 3 Use a Philips no 2 screwdriver to remove the 9 screws fastening the rear case Screws 1 5 Screw
57. dures Step Action 3 Disconnect both battery lug terminals to release the battery 4 When you re insert the battery verify that the battery is inserted the correct way and the polarity is correct see figure below Step Action 1 Firmly grasp both sides of the wheel and pull straight back from the monitor The wheel should slip off the encoder shaft If the wheel does not slip off the encoder shaft use the needle nose pliers and pull the wheel 2 For further disassembly use the 10 mm socket wrench to unscrew the brass hex nut Disassembly 7 5 Disassembly Procedures Removing the Keypad on the Front Panel Removing the Optional External Recorder 7 6 Disassembly Step Action 1 The keypad is attached with an adhesive to the front panel To remove it carefully lift up one corner of the keypad with a flathead screwdriver 2 Peel the adhesive away from the front panel Step Action 1 Press the external Paper Eject button on the right hand side of the recorder The door drops forward 2 Remove the paper roll Two secure screws are visible on the back panel of the recorder 3 Use a Philips screwdriver to loosen the two captive screws 4 Pull the recorder straight out of the side of the monitor by placing your index and middle fingers in the inside center of the recorder and firmly pulling the recorder towards you Separati
58. dust vibration corrosive or explosive gases and is within specified temperature and humidity ranges See Chapter 12 Specifications Position the monitor so that the display is clearly visible and the controls are easy to reach Warning Cautions and Safety Precautions Patient To better secure patient safety become familiar with the details in Safety Chapter 12 Specifications Patient The patient leakage current is lt SO4A 250 V The monitor has patient input connectors Leakage Type CF that are protected against the effects of defibrillation and electrosurgery Current You will find the following heart symbol on the side of the monitor where the patient monitoring input connectors are located The symbol indicates that the monitor is Type CF and is designed to have special protection against electric shocks particularly regarding allowable leakage currents having a F Type applied part according to the standards IEC 60601 1 EN60601 1 CSAC22 2 601 1 UL 2601 1 and is defibrillator proof 19 Site Preparations 2 3 Preparing to Use the Monitor Preparing to Use the Monitor Power Source Requirements Protecting Against Electrical Shock Equipotential Grounding Combining Equipment 2 4 Site Preparations Warning To avoid contaminating or infecting personnel the service environment or other equipment make sure that equipment which has been used before has been appropriately disinfected
59. e 10 1 General RS 232 Safety Information General RS 232 Safety Information Warning If you connect the monitor to any instrument verify proper operation before clinical use Refer to the device s manual for full instructions Accessory equipment connected to the monitor s data interface must be certified according to IEC Standard 950 for data processing equipment or IEC Standard 60601 1 for electromedical equipment All combinations of equipment must be in compliance with IEC Standard 60601 1 system requirements Anyone who connects additional equipment to the signal import port or signal output port configures a medical system and is therefore responsible to ensure that the system complies with the requirements of system standard IEC Standard 60601 1 1 If in doubt consult the Philips Response Center or your local Philips representative Caution Connection to a battery operated PC should only be allowed by the end user Connection to a line powered device with an ITE Class II power supply is acceptable provided it passes all 60601 1 1 tests without a problem If there is a line powered device with a Class I power supply such as a line powered PC or a laser printer in the system there must be either optical isolation of the data cable or the Class I devices must have isolated transformers Note The monitor and its accessories must be tested by qualified service personnel at regular intervals to verify pr
60. e CO tests and summarizes how to document the test results CO Test Results Test Expected Test Results What to record on service record Barometric Pressure Check X1 difference between the reference pressure and the measured ambient pressure displayed on the monitor 12 mmHg X1 12 mmHg Leakage Check parts 1 and 2 X2 value of part 1 leakage check on flowmeter X2 4 0 ml min X3 value of part 2 leakage check on flowmeter X3 lt 4 0 ml min Pump Check X4 gt 120 mmHg below the ambient pressure Flow Rate Check X5 CO flow rate 42 5 ml min X5 65 ml min CO Gas Measurement Calibration Check X6 difference between measured CO value and calculated value based on 5 CO cal gas X7 lt 2 6 mmHg or 0 35 kPa Calibration Verification X7 difference between measured CO value and calculated value based on 10 CO cal gas X74 0 07 x value calculated PCO P X1 X2 X3 X4 X5 X6 X7 or PCO F X1 X2 X3 X4 X5 X6 X7 P passed F failed XI xx two digits X2 xx X3 x x X4 xxx X5 x x X6 x x X7 x x 4 30 Testing the Monitor Barometric Pressure Check and Calibration Performance Procedures Check the barometric pressure value in the CO module as follows Step Action Enter the Power Up Defaults Menu see Power Up Defaults Menu on page 5 3 for instructions on accessing this menu Select the E
61. e memory of the last 10 error codes encountered See Chapter 5 Diagnostic mode for further details on Diagnostic Mode Each error code corresponds to a particular problem in the monitor Recommended actions to take when an error code is encountered are listed below Listed below are error codes that correspond to hardware problems and the recommended actions to take should such an error be encountered Hex Code Explanation Recommended Action 1 Improper shutdown 1 Power cycle 2 If this error persists return the monitor to your local Philips representative for service 4 The measure valve of the 3 3V 1 Check the power supply power supply is low 2 Replace the power supply module 3 The measured valve of the 3 3V 1 Check the power supply power supply is high 2 Replace the power supply module 8 The measured value of the 5V 1 Check the power supply power supply is low 2 Replace the power supply module 9 The measured value for the 5V 1 Check the power supply power supply is high 2 Replace the power supply module E A checksum error is detected on the 1 Turn power Off power up settings region of flash 2 Turn power On while pressing memory both the Contrast and Volume buttons simultaneously See Restoring Factory Settings on page 5 14 3 All user selections must be restored 4 If problem persists replace main PCB module Troubleshooting 6 11 Part 3 Using Supp
62. e monitor displays a heart rate of 240 5 bpm Verify that the audible alarm sounds and the Heart Rate frame flashes indicating that the heart rate is above the default upper alarm limit Press the Alarm Silence button twice on the monitor s front panel to invoke the Silence Reset mode 10 Decrease the heart rate setting on the ECG simulator to 120 bpm After at least five heartbeats verify that the monitor displays a heart rate of 120 5 bpm 11 Disconnect the LL lead from the ECG simulator Verify that the Leads Off alarm message displays displays in the Heart Rate frame and the low priority Leads Off INOP sounds 12 Reconnect the LL lead to the ECG simulator Verify that the Leads Off alarm message no longer displays and the audible alarm is silenced 13 Repeat steps 11 and 12 for LA and RA leads then turn your monitor Off Testing the Monitor 4 25 Performance Procedures Respiration The accuracy of the monitor s respiration measurements is 3 breaths per minute In the Test procedure below add the tolerance of the simulator to the acceptable range of readings Step Action 1 Verify that the monitor is turned Off Connect the ECG leads to the appropriate jacks on the ECG simulator Connect the ECG leads to the ECG trunk cable Connect the cable to the ECG patient monitoring input connector Set the simulator for a respiration rate of 120 breaths per minu
63. e that the LED intensity increases 6 Repeat Step 5 and the intensity again decreases This variation is an indication that the micro processor is in proper control of LED intensity 7 Turn the monitor Off 4 28 Testing the Monitor CO Performance Performance Procedures This test checks the performance of your CO measurement for the monitor This test uses calibration equipment that can be ordered contact your Philips representative Refer to the documentation accompanying the equipment for detailed instructions The procedure is summarized in the following steps e Barometric Pressure Check and Calibration if required Leakage Check Pump Check e Flow Check and Calibration if required e CO Cal Check and Calibration if required e CO Cal Verification using Cal 2 gas Note Allow 5 seconds between individual service procedures in order to ensure stable equipment conditions Tools Needed for CO Tests e Screwdriver e Tweezers e Cal 1 gas 5 CO e Cal 2 gas 10 CO3 e Cal gas flow regulator e Cal tube You also need a local barometric pressure rating received from a reliable local source such as an airport regional weather station or hospital weather station The pressure rating must be located at the same altitude as the hospital Note All steps must be performed in the same session Testing the Monitor 4 29 Performance Procedures Documenting The following table lists th
64. e the monitor is power on After you enter the Power up Defaults Menu physiological monitoring is terminated The screen layouts do not display any information associated with normal monitoring operation After configuring the main monitoring screen and connecting the appropriate accessories as not in the C3 Information for Use guide use the following procedure to configure the power up default settings for the monitor Step Action From the main monitoring screen use the navigation wheel to highlight the Setup icon Press the wheel to select the icon From the Set Up Menu select Enter Power up Default Menu A popup box displays Use the wheel to enter the passcode The passcode is 2 1 5 Note This passcode is configured at the factory and CANNOT be changed The Power Up Defaults Menu displays POWER UP DEFAULT MENU Accept Current Settings Selected Audio Off Make Available Audio Off Make Available Auto Set Limits Make Available Audio Pause Period 180s Enter Diagnostic Mode Language Enter Demo Mode Return The available menu items are described in the next section You can configure these menu items as desired After making the desired change highlight Access Current Settings Press the wheel to select this option From the popup menu select Yes Select Return After selecting Return a notice displays instructing you to power down the monitor Next time
65. easures patient leakage current in accordance with EN60601 1 Clause 19 for Leakage Class I type CF equipment In this test 110 of mains voltage is applied between each Current with patient connection and earth power ground Patient leakage current is then measured from Mains Voltage any individual patient connection to earth on the Applied Part Warning AC mains voltage is present on the applied part terminals during this test Exercise caution to avoid electrical shock hazard Table 4 7 Safety Tests Patient Leakage Current with Mains Voltage on the Applied Part Test or Inspection to Perform Expected Test Results Patient Leakage Current AC Maximum leakage current x See Safety Test Diagram Patient lt 50 uA 250V IEC601 1 or UL2601 1 Leakage Current AC on page 4 44 Test at 110 of the nominal line voltage Figure 4 2 Safety Test Diagram Patient Leakage Current AC Test or Inspection to Perform Not presentin Class 2 LN ANO 1 pplied part MD Nw a o PE d Seagal A in and or Output Instrument under test OO gt 4 2 Lo Oman O Am px E HA Ct a Insulating pad ET 1l _ IE NL gt Il Measures patient leakage current from Applied Part to earth caused by external mains voltage on Applied Part with switch S5 open and closed Each polarity combination poss
66. eeds charging connect the monitor to the AC outlet as described Charging the Battery on page 3 9 If the monitor operates for less than one hour on battery power before the low battery alarm occurs the battery should be conditioned See Electrical on page 12 3 for typical battery operating times and conditions If the same symptom persists after the battery is conditioned and indicating a full charge the battery should be replaced see Removing the Battery on page 7 4 for more information Maintaining the Monitor 3 7 Battery Maintenance Battery To get the most out of the battery observe the following guidelines Guidelines The shelf life of a fully charged battery is about 6 months After a battery has been activated keep it charged If it is not in use recharge it every 3 to 4 months Condition the battery approximately every 200 deep discharges Remove the battery from the monitor if it is not being used regularly Leaving the battery in a monitor that is not in regular use using the battery will shorten the life of the battery Battery Disposal We recommend replacing the monitor s sealed lead acid battery at 2 year intervals Follow local governing ordinances and recycling plans regarding disposal or recycling of batteries and other device components Do not dispose of the battery in normal waste containers Warning If the battery shows signs of damage or signs of leakage replace it immediatel
67. en the Power Up Defaults Menu is existed and the Diagnostic Menu displays For more information see Diagnostic Mode on page 5 5 Language English French German All text shown on the screen is in the Italian Japanese Portuguese selected language The selected language Spanish Russian Chinese is effective the next time the monitor is Dutch Turkish Arabic powered up Norwegian Swedish Finnish Polish Czech Return When selected the Power Up Defaults Menu is immediately exited and you are instructed to power down the monitor a Although the language choices are shown here in English they display on the screen in their respective language and script on the Boot Up screen and in the Power Up Defaults Menu 5 4 Configuring the Power up Defaults Menu Diagnostic Mode Diagnostic Mode The purpose of the Diagnostic Mode Menu option is to allow factory field service and hospital biomedical technicians access to a series of test and system related information screens for the purpose of verifying monitor performance or troubleshooting problems To access Diagnostic Mode Step Action 1 Enter the Power Up Defaults Menu see Power Up Defaults Menu on page 5 3 for instructions on accessing this menu 2 Select the Enter Diagnostic Mode option 3 From the popup menu select Yes The Diagnostic Menu displays DIAGNOSTIC MENU Error Codes System Infor
68. en highlight and press the Start Calibration option To end calibration highlight and press Abort Calibration When calibration is complete highlight and press Return to exit pop up menu Pump Op Time Displays how long the pump has been operational If Make No Change is chosen Pump Op Time remains at current value If Reset to Zero is chosen Pump Op Time has a zero value Return Exits the CO Test and returns to the Diagnostic Menu Configuring the Power up Defaults Menu 5 13 Restoring Factory Settings Restoring Factory Settings Caution In addition to restoring factory defaults this procedure clears the contents of trend memory The following technique can be used to restore the monitor s power up default settings which were originally configured in the factory Note Before performing the first step read this procedure COMPLETELY turn the monitor On Step Action 1 Ensure that the monitor is powered Off 2 Simultaneously press the Volume and Contrast buttons on the monitor s front panel 3 Still pressing the Volume and Contrast buttons press the On Standby button to release the two buttons 4 Continue to press the Volume and Contrast buttons until the power up diagnostic sequence is complete When the main monitoring screen displays screen displays y X Fran ais Deutsch Italiano BAGS Portugu s Espa ol Pycckun English
69. ent error 510xxx internal user interface error 511xxx error handling error 513xxx serial driver error Part 3 Using Support Functions Hexadecimal Code Explanation 514xxx system error 515xxx CO error Other Error If an error code occurs that is not listed in the above sections take the following actions Codes Step Action Power cycle the monitor If the error code still displays take the monitor out of service and contact the Philips Response Center or your local Philips representative for advice on remedial action If the monitor powers up and the error code does not reoccur enter the Diagnostic Mode and select the Error Code option Examine the record of the last 10 error codes and determine if the same error code occurred previously If the Error Code screen indicates that the same error code has occurred previously take the monitor out of service and contact the Philips Response Center or your local Philips representative for advice on remedial action If the Error Code screen indicates no previous occurrences of this error the monitor can be returned to service Troubleshooting 6 13 Part 3 Using Support Functions 6 14 Troubleshooting 7 Disassembly This chapter provides step by step procedures that are used to access replaceable parts of the monitor The sections in this chapter describe and photographically illustrate procedures for disa
70. formance Verification see Performance Procedures on 4 22 This is intended to be a practical hands on part of the training program If possible therefore use a monitor to help you learn about troubleshooting the monitor When faced with a call regarding the monitor there are some simple steps that are recommended to solve the failures described below These failures account for the majority of known failure modes Symptom Cause of Failure Remedy Monitor screen appears dim Brightness not properly adjusted Display backlight tube worn Adjust brightness using Contrast button Replace backlight tube according to instructions in Replacing the Backlight Tube on page 9 Monitor screen is blank Display backlight tube worn or loose connection Main PCB module malfunction 1 Check the backlight tube connector 2 Replace the backlight tube Replace the Main PCB module according to instructions in Removing the Main PCB Module on 7 12 Monitor fails to power up when the On Standby button is pressed Battery is dead or monitor is not plugged in Keypad malfunction or keypad connector is loose Make sure the monitor is powered on by either battery or AC power Check that the battery is adequately charged and AC power fuses are securely connected 1 Check the keypad connector 2 Replace the keypad according to the instructions in Removing the Keypad on the Front Panel on 7 6
71. ge 4 14 Performance Procedures Battery Performance Test page 4 22 Temperature Test page 4 23 ECG Performance page 4 25 Respiration Performance page 4 25 SpO Performance page 4 27 CO Performance page 4 29 Serial Interface and Nurse Call Signal Test page 4 36 ECG Sync Performance Test page 4 38 Patient Safety Tests S 1 Ground Integrity page 4 39 S 2 Electrical Leakage page 4 41 S 3 Earth Leakage Current page 4 41 S 4 Enclosure Leakage Current page 4 42 S 5 Patient Leakage Current page 4 43 S 6 Patient Leakage Current with Mains page 4 44 4 4 Testing the Monitor Test Reporting Test Reporting The following table shows what must be recorded on the Service Record after completing the tests in this chapter Test What to Record Functionality Assurance Tests Visual V P or V F Power On PO P or PO F Preventive Maintenance Tests P NBP PN P X1 X2 X3 X4 X5 or PN F X1 X2 X3 X4 X5 See page 4 15 for details Performance Procedures CO Performance PCO P X1 X2 X3 X4 X5 X6 X7 or PCO F X1 X2 X3 X4 X5 X6 X7 See page 4 30 for details Patient Safety Tests Safety 1 Ground Integrity S 1 P or S 1 F Safety 2 Earth Leakage Current S 2 P or S 2 F Safety 3 Enclosure Leakage Current S 3 P or S 3 F Safety 4 Patient Leakage Current S 4 P or S 4 F Safety 5 Patient Leakage Curre
72. hat at a basic level communication between the SpO module and the main monitor processor is working correctly None of the displayed values can be changed or reset in this screen When in the System A D screen the Return option is always highlighted Press the wheel to return to the Diagnostic Menu Rotating the wheel while in the System A D screen has no effect The A D channel designators are shown in the following table 5 8 Configuring the Power up Defaults Menu Table 5 1 A D Channel Designators 1 ECG 2 RWAVE 3 PACEMAKER 4 RESPIRATION 5 PRESSURE XDUCER 1 6 PRESSURE XDUCER 2 7 NBP OSCILLATORY 8 ECG LEAD OFF 9 TEMPERATURE 10 ISOLATED VOLTAGE REF 11 ISOLATED VOLTAGE ZERO 12 SpO S1 S018 13 NOT USED 14 43 3 VDC POWER SUPPLY 15 12 VDC POWER SUPPLY x 0 33 16 NBP VOLTAGE REF x 0 8 17 GROUND REFERENCE 18 5 VDC POWER SUPPLY x 0 8 19 ADC MID SCALE VALUE 20 ADC ZERO SCALE VALUE 21 ADC FULL SCALE VALUE Diagnostic Mode Configuring the Power up Defaults Menu 5 9 Diagnostic Mode NBP Test A NBP Test screen is provided to facilitate troubleshooting problems and performing verification testing for the NBP subsystem Typically when these tests are performed the pneumatic system is connected to an external pressure reading device and a closed reference volume The NBP Test screen provides a real time numeric display of the pres
73. have only those rights provided for such Software and any accompanying documentation by the applicable FAR or DFARS clause or the Philips standard software agreement for the product involved Text Conventions Explanation of Symbols vi The following conventions for Notes Cautions and Warnings are used in this manual Warning A Warning calls attention to a condition or possible situation that could cause injury to the user and or patient Caution A Caution calls attention to a condition or possible situation that could damage or destroy the product or the user s work Note A Note calls attention to an important point in the text Symbols on products and packaging mean the following All symbols found in this section are used this Instructions For Use guide REF SN CE Reference Number Serial Number 0123 CE Marking o yA 5 to 95 RH Humidity 48 c 4 Temperature Limits Ry hy a as x E a 9 1 tu 0 0 Me sol e g Keep out of Sun Keep Dry Fragile Keep Upright AC LED Battery LED NBP Volume Contrast On Standby Snapshot Continuous Defibrillator proof type CF equipment vii viii ECG Temperature SpO NBP Carbon Dioxide Alarm Alarm Limits Menu Audio Off Silence Reset NBP Automatic Interval Mode Big Numbers Clock Up Alarm Arrow Limit Down Alarm Arrow Limit ll co co co p
74. he On Standby button to turn the monitor On After the normal power up sequence verify that the SpO display initially indicates zero or is blank Verify that the following monitor reaction occurs a After approximately 45 seconds the monitor displays saturation and heart rate as specified by the tester Verify that the values are within the following tolerances Oxygen Saturation Range 79 to 83 Heart Rate Range 33 to 39 bpm b The audible alarm sounds and both the SpO and Heart Rate HR displays flash indicating both parameters have violated the default alarm limits c The HR tone is heard For this test the HR tone source must be set to SpO from the Heart Rate Menu Press the Alarm Silence button to temporarily silence the audible alarm Verify the following a The audible alarm remains silenced b The crossed out bell icon displays in each numeric frame on the screen c The SpO and HR displays continue flashing d The HR tone remains audible e The audible alarm returns in approximately 120 seconds Volume Tools Needed Control Test SpO adapter cable M1943A e Clinical Dynamics Corp SmartSat simulator with Nellcor simulator cable 4 12 Testing the Monitor Functionality Assurance Tests Step Action Connect SpO simulator cable to the SpO adapter cable Connect the cable to the SpO patient monitoring input connector Set the simulator
75. he Volume button is pressed the safety valve and proportional valve opens and remains at maximum Deflate As long as the Alarm Silence button is pressed the proportional valve opens and bleeds off pressure at the rate of 3 1 5 mmHg s It is useful to control the bleed rate to 3mmHg s to facilitate certain AAMI SP10 tests Any time the bleed rate falls below 3 mmHg s the valve opens and remains at maximum as long as the button is pressed Offset Adjustment Momentarily press the Contrast Adjust button invokes the Zero Calibration routine that is performed immediately prior to each blood pressure measurement This routine looks at the pressure in the system and if the pressure is non zero an offset is applied which causes the system pressure to display as Zero When in the NBP Test screen the Return option is always highlighted Press the wheel to return to the Diagnostic Menu Rotating the wheel while in the NBP Test screen has no effect Configuring the Power up Defaults Menu 5 11 Diagnostic Mode CO Test The CO Test Screen is provided to facilitate troubleshooting problems and performing verification testing to ensure the measuring results are accurate See CO2 Performance on page 4 29 for information on how to perform these tests Warning A FilterLine connected to the monitor should never be applied to a human being while the monitor is in Diagnostic Mode Injury could result DIAGNOSTIC MENU
76. he original shipping carton is not available use another suitable carton Prior to shipping the monitor contact the Philips Response Center for a RGA Returned Goods Authorization number Mark the shipping carton and any shipping documents with the RGA number Repacking the Original Carton If available use the original carton and packing materials Pack the monitor as follows Step Action 1 Place the monitor and if necessary accessory items in original packaging 2 Place the shipping carton and seal carton with packaging table 3 Label carton with shipping address return address and RGA number if applicable Repacking in a Different Carton If the original carton is not available use the following procedure to pack the monitor 9 2 Packing for Shipment Step Action 1 Place the monitor in a plastic bag 2 Locate a corrugated cardboard shipping carton with at least 200 psi pounds per square inch bursting strength 3 Fill the bottom of the carton with at least 2 inches of packing materials 4 Place the bagged unit on the layer of packing material and fill the box completely with packing material 3 Seal the carton with packing tape 6 Label the carton with the shipping address return address and RGA number if applicable 10 RS 232 Interface This chapter explains how to properly use and connect the RS 232 interface and cables RS 232 Interfac
77. hich has been used before has been appropriately disinfected and decontaminated Disconnect the monitor from the AC source by unplugging the power cable from the AC power connector located on the rear of the monitor The On Standby button does not disconnect the monitor from the AC mains supply Accessory equipment connected to the monitor s data interface must be certified according to IEC Standard 60950 for data processing equipment or IEC Standard 60601 1 for electromedical equipment All combinations of equipment must be in compliance with IEC Standard 60601 1 1 systems requirements Explosion Hazard Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide Make sure that you have read all applicable instructions before attempting to install the wall mount Wall mounts that are intended to support monitors must be capable of supporting 4 times the weight of the monitor when properly installed Do not mount any portion of a monitoring instrument over a patient s bed Do not exceed the maximum rated load specified for each wall mount Ensure that no electrical wiring utilities or piping interfere with the selected wall mounting location Note Do not return sensors patient cables NBP tubing and cuff or the power cord 2 2 Site Preparations Site Preparation Guidelines Site Preparation Guidelines e Check that the environment is reasonably free from
78. ible is tested using S2 and S6 Safety test according to EN60601 1 Clause 19 4 h Report largest value What to record on the service record S 5 P or S 5 F 4 44 Testing the Monitor Patient Safety Tests Testing the Monitor 4 45 Patient Safety Tests 4 46 Testing the Monitor 5 Configuring the Power up Defaults Menu This chapter discusses the use of the Power up Defaults Menu to configure power on default settings It also explains how to use Diagnostic Mode to obtain service related information about the monitor Configuring the Power up Defaults Menu 5 1 General Safety Information General Safety Information Warning A blood pressure cuff connected to the monitor should never be applied to a human being while the monitor is in Diagnostic Mode Injury could result A FilterLine connected to the monitor should never be applied to a human being while the monitor is in Diagnostic Mode Injury could result Caution In addition to restoring factory defaults this procedure clears the contents of trend memory Note Before performing the first step read this procedure COMPLETELY 5 2 Configuring the Power up Defaults Menu Power Up Defaults Menu Power Up Defaults Menu The purpose of the Power up Defaults Menu is to allow the authorized user to create a power up default for each setting in the monitor Power up defaults are the settings in effect each tim
79. ibration Calibration Check 4 34 Testing the Monitor Step Action Check that the 5 calibration gas and flow regulator are connected Calculate the expected measurement value in mmHg as follows 0 05 x ambient pressure 1 03 value mmHg i e 0 05 x 736 mmHg 1 03 35 7 mmHg with an ambient pressure of 736 mmHg Note Dividing by 1 03 compensates for the dry calibration gas at room temperature relative to breath gases containing water vapor and at body temperature Select CO Calibration Wait for the status message CO module reset in progress to disappear Open the valve on the flow regulator to allow 5 CO gas to flow into the monitor Allow the displayed CO value third row of CO Test menu to stabilize Check that the CO value on the monitor matches the calculated mmHg value 2 6 mmHg If the value is outside the tolerance calibrate as described in Step 9 to 13 Disconnect the 5 calibration gas and connect the 10 calibration gas Calculate the expected measurement value and tolerance in mmHg as follows 0 1 x ambient pressure 1 03 value mmHg 0 07 x value mmHg tolerance i e 0 1 x 737 mmHg 1 03 71 6 mmHg with an ambient pressure of 737 mmHg 0 07 x 71 6 mmHg 5 01 mmHg tolerance Open the valve on the flow regulator to allow 10 CO gas to flow into the monitor Allow the value to stabilize Check that the value on the monitor X6 ma
80. igh voltage is generated by the LCD backlight driver Exercise caution when operating the monitor with the covers open Caution Observe ESD electrostatic discharge precautions when working within the unit Before you remove any module make sure you are wearing ESD protection and you are working in a grounded environment If the internal battery cable has been disconnected pay particular attention to the polarity of the cable before reattaching If the battery cable polarity is reversed it is likely that circuit damage will occur Note If you want to disassemble the RS 232 connector located on the rear panel you will need a 3 16 inch socket wrench Disassembly 7 3 Disassembly Procedures Disassembly Procedures This section provides instructions for closed case front case and rear case disassembly procedures Closed Case This section describes the items that can be removed without disassembling the main case of Disassembly the monitor Procedures Removing the Battery Warning Before you begin any disassembly procedure you must remove the battery Step Action 1 With the rear of the monitor facing you disconnect the battery cover from the rear panel using a Philips no 1 screwdriver to loosen the 4 captive screws fastening the battery cover 2 Slide the battery out by tilting the monitor towards you 7 4 Disassembly Removing the Navigation Wheel Disassembly Proce
81. irs if the monitor has been dropped or as needed Patient Safety Tests on page 4 39 Functional testing procedures When functional defects in the measurements are suspected Functionality Assurance on page 4 3 Performance Assurance At least once every year or as needed if you suspect defects in the measurements Performance Assurance Test on page 4 11 Replace backlight 20 000 hours about 4 years of continuous use or as needed Replacing the Backlight Tube on page 7 9 NBP Maintenance Every year or as needed if the monitor has been opened or you suspect defects in the measurement Preventative Maintenance Tests on page 4 14 Battery Maintenance Every two years Battery Maintenance on page 3 7 Battery Conditioning Approximately every 200 deep discharges Conditioning a Battery on page 3 10 3 4 Maintaining the Monitor Maintenance Checklist Maintenance Checklist Topics See Page Check Here If Date Completed Completed Inspect Cables Cords and Housing 3 6 Replacing the Backlight Tube 7 9 Assembly NBP Preventative Maintenance 4 14 Cleaning Guidelines 3 7 Checking the Battery Status 3 8 Charging the Battery 3 9 Conditioning the Battery 3 10 Testing the Monitor 4 1 Maintaining the Monitor 3 5 Inspecting the Monitor Inspecting the Monitor If you discover a prob
82. itches off If the new battery is partly or fully charged the monitor starts operating A complete battery recharge gt 90 requires 8 hours in standby mode or 14 hours in operational mode To charge the battery Step Action 1 Connect the monitor to the AC power source using the proper power cord 2 Verify that the AC Input Indicator LED is lit 3 Charge the battery for at least 8 hours The battery may require a full discharge charge cycle to restore it to its normal capacity 4 To check for a full charge perform the procedure described in Battery Performance Test on page 4 22 Maintaining the Monitor 3 9 Battery Maintenance Conditioning a Battery Why is Battery Conditioning Necessary When Should Battery Conditioning be Performed How to Condition a Battery Battery conditioning re calibrates the battery to ensure that it has accurate information on the actual battery capacity Warning Never use a monitor that is monitoring a patient to perform battery conditioning The capacity of a battery decreases gradually over the lifetime of the battery Each time a battery is charged its capacity decreases slightly Therefore the operating time of a monitor running off the battery also decreases with each charge cycle Battery conditioning ensures that the value storied in the battery for its fully capacity takes account of this decrease so that the remaining battery charge can
83. itors are clinicians and nursing staffs within hospital out patient and ambulatory settings This monitor 1s not intended for helicopter transport or home use The physical and operational characteristics of the monitor are described in the C3 Instructions for Use guide PHILIPS 1 Carbon Dioxide monitoring is available within the C3 Sedation monitors only Overview 1 3 Features and Options Features and Options The measurement parameters and features for each model are indicated below ian Description Display Measurements Printout Color ECG NBP Temp SpO CO Recorder 862474 Standard Color optional 862478 Sedation Color optional Front Panel Description 1 4 Overview Callout Button LED A AC LED B Battery LED C Audio Alarm Control button D NBP button E Volume button F Contrast button G On Standby button The following diagram illustrates the controls located on the front panel of your C3 Front of Monitor Features and Options Use the following diagram to familiarize yourself with the main features of your C3 monitor The recorder module and all patient input connectors are located on the monitor s side panels 01 06 03 01 09 17 Callout Description A Patient Monitoring Input Connectors Main Monit
84. k List e Accuracy Calibration and Performance Procedures e Patient Safety Checks For inspection procedures preventative maintenance procedures cleaning procedures and battery maintenance refer to this chapter and Chapter 2 Site Preparations This refers to the combined Performance Assurance Test and Functionality Testing Procedures found in this chapter These tests verify correct monitor functionality in general terms Preventative Maintenance refers specifically to the service calibration tests required to ensure the monitor measurement results are accurate When authorized Philips personnel service the monitor they report these results back to Philips The collected data forms a database to be used in product development These specific tests are required for the NBP parameter It is not necessary for hospital personnel to report results This concept refers to all the remaining accuracy and performance tests available on the monitor including safety tests and checks for the monitor Testing the Monitor 4 3 Testing Checklist Testing Checklist The tests described in this chapter are listed in the table below Use this table as a checklist Check Here Date Topics See if Completed Completed Functionality Assurance Tests Performance Assurance Test page 4 11 Power On Self Test page 4 11 Alarm Test page 4 12 Volume Control Test page 4 12 Preventive Maintenance Tests NBP pa
85. lem while inspecting monitor contact your biomedical department or Philips Medical Systems With the monitor turned Off Step Action Examine the exterior of the monitor for cleanliness and general physical condition Verify that the housing is not cracked or broken that everything is present that there are no spilled liquids and that there are no signs of abuse Inspect all accessories cables and sensors external to the monitor referring to the accessories documentation Switch the monitor On and verify that the backlight is bright enough Check that screen is at its full brightness If the brightness is not adequate contact your biomedical department Philips Medical Systems recommends replacing the backlight about every 4 years If you are operating the monitor from battery verify that the battery is adequately charged before you begin use Inspecting the Cables and Cords If you discover a problem while inspecting the cables and cords replace the cable or contact your biomedical department 3 6 Maintaining the Monitor Step Action Examine the power plug and cord for damage Make sure that the prongs of the plug do not move in the casing If damaged replace the entire cord with the appropriate Philips power cord Inspect the patient cables and leads and their strain reliefs for general condition Make sure there are no breaks in the insulation Make sure that the connecto
86. ll Signal Test 00 ccc cece eee eens 4 36 ECG Syne NES yA A A PTS pints Aang EA 4 38 Patient Safety Tests ci he See Sia SO te ee SRS 4 39 Ground Integrity ei A HER ald ave x ENS 4 39 Electrical L akage o Sor Ee RED RUE Cie Ges Chee Cafe eR cR etd 4 41 Earth Leakage Current ose re Wg ONAL E 4 41 Enclosure Leakage Current 00 000 ccc cee eee 4 42 Patient Leakage Current esra 0 cette tenets 4 43 Contents 3 Patient Leakage Current with Mains Voltage on the Applied Part 4 44 5 Configuring the Power up Defaults Menu eee eee S I General Safety Information serisi eron ne a p ran cette m a E 5 2 Power Up Defaults Men veritas tr pra 5 3 Menu Options ci GR ICD GU A AR e ee 5 4 Diagnostic Mode 2 2 ubt acr COEUR OCDE NUAGE IS REY De A CER UE 5 5 Error Codes dere A AE ROGA MRE Ph ORE UR DR E CREE 5 6 System Information ccurrsbneete eee 4I 44 Mode D SR pode hows OLAS OES 5 6 System A D Values cio a em e e e re pie een 5 8 NBP TEA VIDE MEC RURSUM RUPEM NES 5 10 COD Test PR Pm 5 12 Restoring Factory Settings score irrita a 5 14 6 Troubleshooting o uo og o ds O51 General Troubleshooting Safety Information 00 ccc eee eee 6 2 Objectives oye arash eiue SYNE ASUS NEN HONO SI UR CSI ET EN Ss 6 3 Concepts egt ute ACE WAR ENDS U ACA or ORC URS E eie qr CR RE 6 3 Trouble shooting tt ouscoipbcehePU PUOI ele Se e D EORR POS Aon 6 3 Error Codes rye Ep RR TER C RU reU OCHO
87. lues listed Voltage for pin 9 is that listed from the No Alarm condition 5 Connect SpO simulator cable to the SpO adapter cable Connect the cable to the SpO patient monitoring input connector 6 Set the simulator switches as follows Item Setting Oximeter Nellcor SpO0 7 81 BPM 36 Pulse Mod 0 50 7 Verify that the monitor is responding to the SpO simulator signal and the audible alarm is sounding If desired press the Alarm Silence button to temporarily silence the audible alarm 8 Connect the DMM positive lead to pin 9 and verify the voltage value listed in Table 4 2 Serial Interface Voltages The voltage for pin 9 is that listed for the Alarm Underway condition Table 4 2 Serial Interface Voltages Pin Signal Direction Measurement V Min Typical Max 1 not used 0 4 0 0 0 4 2 RXD lt lt lt input 0 4 0 0 0 4 3 TXD lt lt lt output 5 0 7 0 15 0 4 DTR lt lt lt output 5 0 7 0 15 0 5 GND 0 4 0 0 0 4 6 DSR lt lt lt input 0 4 0 0 4 0 7 RTS gt gt gt output 5 0 7 0 15 0 4 36 Testing the Monitor Performance Procedures Pin Signal Direction Measurement V CTS lt lt lt input 0 4 0 0 0 4 Alarm Out gt gt gt output 5 0 7 0 15 0 no alarm Alarm Out gt gt gt output 5 0 7 0 15 0 alarm underway Testing the Monitor 4 37 Performance Procedures ECG Sync Test This test checks the performance of ECG synchronization between the monitor an
88. mation System A D Values NBP Test CO2 Test The Diagnostic Menu lists the test and system related information screens Select an item in the menu to invoke a test or information screen The test and information screens that are in the diagnostic menu listed below An explanation of each menu item is given in the sections that follow Error Codes System Information e System A D Values NBP Test d CO Test Configuring the Power up Defaults Menu 5 5 Diagnostic Mode Error Codes DIAGNOSTIC MENU ERROR CODES Error Codes Date Time og goudoonrom 2 Az 0 c Z 3 This screen displays the 10 most recent error code types logged by the monitor The date and time that the error code appears displays to the right of the error code Error codes cannot be changed or reset in this screen When in the Error Code screen the Return option is always highlighted Press the wheel to return to the main monitoring screen Rotating the navigation wheel while in the Error Code screen has no effect Refer to Error Code Categories on 6 12 for more details on error codes System Information DIAGNOSTIC MENU SYSTEM INFORMATION Monitor On Time 0 Backlight On Time 0 Recorder On Time 0 Battery Deep Discharges 0 System Software Version Application V1 00 CO2 Software Version 00 01 NL V01 11 09 05 2000 SN 10971 SpO2 Software Version SW 3 11 ALG 2 57 HW 0 0 DSPFW 0 16 NBP Software Version AG 2 27 SM
89. minutes Maintaining the Monitor 3 11 Battery Maintenance 3 12 Maintaining the Monitor 4 Testing the Monitor This chapter provides a checklist of testing procedures used to verify performance following repairs or during routine maintenance All tests can be performed without removing the monitor s covers If the monitor fails to perform as specified in any test repairs must be done to correct the problem before the monitor is returned to the user Testing the Monitor 4 1 Testing Safety Information Testing Safety Information Warning Perform all leakage tests any time the unit is opened AC mains voltage is present on the applied part terminals during this test Exercise caution to avoid electrical shock hazard Note The monitor must be placed in Diagnostic Mode with the NBP test screen active for each of the NBP tests If the test fails ensure the integrity of the cuff and tubing then test again If the test fails again verify the integrity of all the pneumatic tubing inside the monitor Before beginning the Battery Performance test ensure that the monitor is disconnected from the AC power source 4 2 Testing the Monitor Objectives Concepts Functionality Assurance Preventative Maintenance Performance and Safety Tests Objectives In order to meet this chapter s objectives you should be able to test the monitor through the following types of tests Performance Assurance Chec
90. n 2 Perform the following sequence a Press the Home button b Press the Pressure Tests button c Press the Pressure Leak Test button d Set the Cuff to Internal Confirm that the simulator is active and displays Leak Test 4 Press the Volume button to ensure that both valves are closed Perform an offset adjustment so that the simulator and monitor both display a pressure of 0 mmHg or kPa by doing the following a Press the Contrast button on the monitor s front panel b If needed zero the simulator see To Zero the Simulator on page 4 16 6 Note Complete steps 6 9 within 150 seconds Press and hold the NBP button on the monitor s front panel and inflate to 250 10 Don t go over 270 or safety deflation may occur Slow the inflation rate by pulsing the button when the pressure is over 200 7 Allow 15 20 seconds for the pressure to stabilize Record the pressure displayed on the monitor P1 8 Press and hold the Alarm off button on the monitor to release the pressure in steps of 3 mmHg 4 kPa Simultaneously start the timer 9 Stop the timer when the pressure drops below 150 mmHg 20 kPa Calculate the deflation rate X5 starting pressure 150 of seconds The deflation rate should be 3 3 mmHg s 1 5 mmHg s 0 44 kPa s 0 2 kPa s 10 Press and hold the Volume button until the monitor displays a pressure of 0 mmHg or kPa 11 If no further NBP tests are to be
91. n level is also momentarily shown on the LCD as part of the Copyright screen The value CANNOT be changed CO Software Version Displays the revision level of the software of the Oridion CO module This value CANNOT be changed SpO Software Version Displays the revision level of the software of the MP 205 SpO module This value CANNOT be changed NBP Software Version Displays the revision level of the software for the NBP module This value CANNOT be changed Return Exits the System Information Menu and returns to the Diagnostic Menu Configuring the Power up Defaults Menu 5 7 Diagnostic Mode System A D The System A D screen displays the current value of each analog to digital A D channel in Values volts Some of the channels are for AC coupled signals such as ECG input so the numbers on the screen are constantly changing when an input signal is present These AC coupled values are shown to give an indication as to whether basic functionality of the channel is present but no significance can be derived from the values of the numbers displayed DIAGNOSTIC MENU SYSTEM A D VALUES 12 13 14 15 16 17 18 19 20 21 1 2 3 4 5 6 7 8 9 SpO2 S1 SpO2 Alive SpO2 S2 No Sensor Return The Primary and Secondary Status messages from the SpO module are displayed and updated at the rate of about once per second Presence of the correct SpO message indicates t
92. ned there are specific steps regarding reassembly described in the disassembly instructions in Chapter 7 that must be followed If these instructions are not followed various errors may result For monitors that are defective on arrival first verify that the problems are not related or cannot be remedied by following the steps in the above troubleshooting table If these are not successful and a replacement monitor is necessary contact a Philips Sales Representative to arrange a replacement Disassembly Disassembly procedures are described in the disassembly guide in Chapter 7 Read the whole chapter Whenever the monitor has been opened there are specific steps regarding reassembly described in the disassembly instructions in Chapter 7 that must be followed If these instructions are not followed errors may result If the monitor fails the required performance tests described in Chapter4 then please review the reassembly 11 12 Training Monitor Applications and Algorithms Monitor Applications and Algorithms Information about Found in Supplies http shop medical philips com or Chapter 16 Accessories in the Instructions for Use guide Technical Data Sheet Basic monitor specifications Chapter 12 Specifications Technical Data Sheet Regulatory Compliance Chapter 12 Specifications Technical Data Sheet Other data about algorithms Chapter 12 Specific
93. ng the Front from the Rear Case Disassembly Procedures Step Action 1 Verify that the battery has been removed If not see Removing the Battery on page 7 4 2 Verify that the navigation wheel and brass hex nut have been removed If not see Removing the Navigation Wheel on page 7 5 3 Use a Philips no 2 screwdriver to remove the 6 screws fastening the Rear Case Assembly to the Front Case Assembly 4 Facing the front of the monitor shift the front case slightly to the right making sure that you do not break off the navigation wheel shaft Disassembly 7 7 Repeater Disassembly Procedures Step Action 5 Disconnect the display cable located on your left hand side by pulling the connector towards you Disconnect the backlight tube front panel and speaker cables located on your right hand side Front Panel Backlight Tube Cables Cables Speaker Cable Display Cable 6 Separate the front from the rear case Front Case After you have separated the front from the rear case you can replace the following from the Disassembly front case assembly Speaker e Backlight Tube Caution Before you remove any module make sure you are wearing ESD protection and you are working in a grounded environment 7 8 Disassembly Replacing the Speaker Replacing the Backlight Tube Disassembly Procedures Step Action Use a Philips no 2 screw
94. nitor Off Table 4 1 Settings and Monitor Indications Simulator Settings Monitor Indications Ambient Ambient Sp02 Pulse Rate Light Ac Modulation SpO Pulse Rate Light Freq Level 81 36 120 200 0 50 79 83 33 39 81 112 120 200 5 0096 79 83 109 115 81 201 120 10 0 50 79 83 195 207 81 201 120 200 5 0096 79 83 195 207 specification on the monitor due to the performance characteristics of the simulator An indicates values that produce an alarm Press the Alarm Silence button to temporarily silence the audible alarm a For the pulse rate setting of 201 bpm the pulse rate tolerance of 195 to 207 bpm is greater than 3 bpm accuracy Testing the Monitor 4 27 Performance Procedures LED Excitation This procedure uses normal system components to test circuit operation A SpO2 adult Test finger reusable sensor is used to examine LED intensity control The red LED is used to verify intensity modulation caused by the LED intensity control circuit Step Action 1 Connect an SpO sensor to the monitor 2 Press the On Standby button to turn the monitor On 3 After the monitor completes its normal power up sequence verify that the sensor LED is brightly hit 4 Slowly move the sensor LED in proximity to the photo detector element of the sensor Verify as the LED approaches the optical sensor that the LED intensity decreases 5 Open the sensor and take not
95. ns the next time the monitor is turned On Inflation Rate This test verifies the inflation rate of the monitor Test Step Action 1 Turn the simulator on 2 Perform the following sequence a Press the Home button b Press the Pressure Tests button c Press the Pressure Leak Test button d Press the Setup button e Press 3 0 0 then Enter f Set the Cuff to Internal Press the Volume button to ensure that both valves are closed 4 Perform an offset adjustment so that the simulator and monitor both display a pressure of 0 mmHg or kPa by doing the following a Press the Contrast button on the monitor s front panel b If needed zero the simulator see To Zero the Simulator on page 4 16 5 Press the NBP button on the monitor s front panel to activate the pump and simultaneously start the timer Hold the button until the monitor displays a pressure of 250 mmHg 33 3 kPa then stop the timer The inflation rate should be between 1 and 6 seconds Document the number of seconds X3 6 Press and hold the Volume button until the monitor displays a pressure of 0 mmHg or kPa 7 If no further NBP tests are to be conducted turn the monitor Off Normal monitoring operation returns the next time the monitor is turned On Note The over pressure relief may activate and automatically deflate the cuff For more information on the NBP safety functions see page 4 16 4 1
96. nspection of the monitor Cleaning the monitor Battery Maintenance Light maintenance tasks can be defined as those tasks required to maintain the monitor in clean functional working order These include inspection cleaning practice and battery maintenance as well as the continuous observation of replaceable parts for wear A rechargeable battery must be conditioned at regular intervals to prepare it for further charge discharge cycles Conditioning a battery refers to the complete discharge of a charged battery by allowing it to die out while in non critical use The empty battery may then be recharged and put back into use Maintaining the Monitor 3 3 Recommendations for Maintenance Frequency Recommendations for Maintenance Frequency The following table is the recommended maintenance schedule for your C3 monitor The maintenance checklist appears in the next section of this chapter The checklist may be photocopied and should be completed by the maintainer It should be filed for future reference The owner of the equipment is responsible for the performance of the maintenance activities with the schedule below Maintenance Frequency Procedure Inspect the monitor cables Daily Inspecting the Monitor on page 3 6 and cords and Inspecting the Cables and Cords on page 3 6 Cleaning As needed Cleaning on page 3 7 Safety checks according to IEC 60601 1 At least once every 2 years after any repa
97. nt with Mains S 5 P or S 5 F Where P Pass F Fail and X is the measured value as defined in the tests described in this chapter Testing the Monitor 4 5 Recommendations for Testing Frequency Recommendations for Testing Frequency The testing checklist appears in the next section of this chapter Perform the procedure as indicated in the suggested testing timetable These timetable recommendations do not supersede local requirements Suggested Testing Frequency Functionality Assurance Performance Assurance Test System Self Test When functional defects in the measurements are suspected after any repairs if the monitor has been dropped or opened Preventative Maintenance Tests NBP Every year or as needed Performance Tests Battery Performance Temperature Accuracy ECG Performance Resp Performance SpO Performance CO Performance Serial Interface and Nurse Call Signal Test ECG Sync Performance The battery performance test must be performed every two years before monitor repairs or whenever the battery is suspected as being the source of the problems All other test must be performed at least once every year if the monitor has been opened or if you suspect defects in the measurements Safety Checks in accordance with IEC 60601 1 Ground Integrity Electrical Leakage Earth Leakage Current Enclosure Leakage Current Patient
98. nter Diagnostic Mode option From the popup menu select Yes The Diagnostic Menu displays Select the CO Test option nA BR wl ry Connect a FilterLine to the CO patient monitoring input connector This activates the pump in the CO module Check the status line at the top of the screen It will display CO pressure in mmHg ambient cell xxx yyy where xxx is the ambient pressure and yyy is the measured cell pressure The values are displayed with a resolution of 2 mmHg 0 3 kPa up to 475 mmHg 63 2 kPa and a resolution of 1 mmHg 0 1 kPa from 475 mmHg 63 2 kPa to 825 mmHg 109 7 kPa Check whether the ambient pressure X1 matches within the acceptable tolerance of 12 mmHg the reference value you have received If so proceed to the Leakage Check If the value is not correct calibrate as follows Select Set Barometric Pressure mmHg An adjustable value in mmHg is activated Select the value that matches the reference value received from a reliable local source such as an airport regional weather station or hospital weather station If the selected value is not with 12mmHg 1 6 kPa of the current measured ambient pressure verify the reference value by getting another reading from a different source If the ambient pressure displayed also differs from the new reference by more than 12mmHg 1 6 kPa the CO module should be replaced 10 Confirm the barometric pressure se
99. nu selection If pressed twice within 10 seconds the alarm sounding for a specific parameter is silenced temporarily for a preset interval determined by the menu selection Other alarms not related to this parameter can still sound during this period If pressed and held for 2 seconds or more the Audio Off condition is initiated The NBP button is connected to the processor Response of the processor depends upon the state of NBP operation at the time and the action in pressing this button If momentarily pressed less than 2 seconds a single NBP measurement is obtained If pressed for 2 seconds or more the processor initiates a STAT monitoring sequence Pressing the NBP switch at any time a pressure measurement is in effect causes the processor to terminate the measurement and to deflate the cuff Training 11 7 Theory of Operation and System Architecture Navigation Wheel RS 232 1 0 Defib Sync Pulse Speaker 11 8 Training The Contrast button operates in conjunction with the navigation wheel to determine the apparent contrast setting in the display Changing the contrast is a change in the viewing angle Outputs of the button and wheel are connected to the processor Momentarily pressing the button sets the contrast to mid range factory default value Momentarily pressing of the button followed within 3 seconds by a rotation of the wheel are processed to vary the contrast of the display When there has been no wheel rota
100. onnected to the monitor Isolating the You can use the following table to isolate and solve problems which may occur in the Defective monitor Component Symptom Cause of Failure Remedy The battery symbol is not displayed A battery is not present in the monitor the battery is defective or there is a bad connection Install a charged battery see Removing the Battery on page 7 4 If a battery is already present remove it and refit the battery making sure to push is completely into position Check the connection wires to make sure wires are secured to battery Some or all numerics or waves are not displayed Parameters are switched off No accessories are connected Switch parameters on Connect the required accessories If connections are secure replace the suspect accessory Monitor screen appears dim Brightness not properly adjusted Display backlight tube worn Adjust brightness using Contrast button Replace backlight tube according to instructions in Replacing the Backlight Tube on page 7 9 Monitor screen is blank Display backlight tube worn or loose connection Main PCB module malfunction 1 Check the backlight tube connector 2 Replace the backlight tube Replace the Main PCB module according to instructions in Removing the Main PCB Module on 7 12 6 8 Troubleshooting Part 2 Isolating and Solving Monitor Problems Sym
101. oper operation according to the procedures of the user s institution Additional important safety information is located in the C3 Instructions for Use guide 10 2 RS 232 Interface About the RS 232 Interface About the RS 232 Interface The RS 232 interface allows you to send a nurse call signal export trend data to an external PC Warning Anyone who connects additional equipment to the signal input port or signal output port configures a medical system and is therefore responsible to ensure that the system complies with the requirements of system standard IEC Standard 60601 1 1 If in doubt consult Philips Medical Systems Response Center or your local Philips representative Cable A 9 pin connector mounted on the rear panel provides the access port for the serial RS 232 Connections interface to a suitably configured personal computer Or alternatively qualified service personnel can use the connector to send a Nurse Call signal The figure below shows you where the RS 232 interface is located on the monitor TEn so so Hs 1 08 tear ata E ja Mi sa mo i PHILIPS IPS C3 PATIENT MONITOR Manufactured for Philips for Philips 3000 Mi And RS 232 Interface 10 3 About the RS 232 Interface Pin connections for the 9 pin connector are as follows Table 10 1 RS 232 Serial Interface Connections
102. or psss ceio tena cetin cent e 2 5 Unpacking the Monitor 2 0 0 tenet re 2 5 Checking the Shipment 00 2 0 56 ca eiee ee ee x id e E en 2 5 Returning System Components sssr aE E UE E O e re 2 6 Mo a a thd E EE 2 7 Warnings Cautions and Safety Precautions Relating to Wall Mount Installation 2 8 Mounting the GCX Wall Channel 0 00 cect e 2 9 Attaching the Mounting Plate to the Monitor 00 00 e eee ro 2 9 Attaching the Monitor to the Wall Channel 22 0 0 ce eee cee I 2 10 Contents 1 3 Maintaining the Monitor 2 2 02 2506 0 eee e Pies cece eed o dl Maintenance Safety Information 00 cece mh 3 2 Objecttves t E E EAR EE AE E AC he M 3 3 Concepts EA tte ets uc ali E E Att dd 3 3 Tight Maintenance aei ve a A ETE A A A AR 3 3 Battery Conditioning oro gad Se SS ed usen qa veia 3 3 Recommendations for Maintenance Frequency 00 0 cece eee teens 3 4 Maintenance Checklist ue bd dt t eet ees 3 5 Inspecting the Monitor i eoe AA QU Ded oe A 3 6 Inspecting the Cables and Cords 0 ccc cette e 3 6 Cleanings 2 53455 6S ee RS MEME BP em eH CE uM 3 7 Battery Maintenance 3 7 Aboutthe Battery xoc tent the Be Be eee ao ee toot ted SL 3 7 Battery Guidelines ora A gle eut inn ee A e 3 8 Checking the Battery Status e penn ect e 3 8 Identifying Battery Strength llli 3 9 Charguig the Battery a cy une hc a oe UL EU wa ERI ate 3 9 Conditioning a Battery seresa r en
103. oring Screen Handle Recorder optional Not available in all models m ol al s Navigation Wheel Overview 1 5 Features and Options Rear of The following diagram is of the rear panel For an explanation of the symbols located on this Monitor panel see Explanation of Symbols on page vi The C3 has four connections on its rear panel The diagram below is of the rear panel It shows you how to make the four possible connections Defib Sync Equipotential AC Input and RS 232 I O Tr amp eo 100 230V 100 230V 5060 Hz 10A TIALZS0V Callout Connector Connector Type Label A AC Input 3 line connector IEC 320 receptacle 100 240V 50 60Hz 1A B RS 232 I O DB 9 male C C Equipotential Equipotential Grounding Post The equipotential grounding post may be used by facilities as required by their procedures D Defib Sync 2 5mm subminiature phone jack 4 E 1 6 Overview Features and Options Screen The LCD displays parameter values real time waveforms alarm messages and screen Display selection icons Numeric frames displays icons and numeric values of real time physiologic parameter e Waveform frames displays real time waveforms graphical trend or tabular trend data e Message and Icon frames displays alarm messages and screen selection icons Standard Mo
104. ort Functions Error Code Categories 6 12 Troubleshooting Hex Code Explanation Recommended Action 64 The SpO module is sending an 1 Ensure PicoSat module is error message to the host CPU properly connected 2 Replace the PicoSat module 3 If problem persists replace main PCB module The CO module is sending an 1 Ensure Oridion module is error message to the host CPU properly connected 2 Rplace the Oridion module 3 If problem persists replace main PCB module 65 66 PicoSat SpO module has detected 1 Replace PicoSat module an error during initialization 2 If problem persists replace main PCB module 6E 71 PicoSat SpO module has detected 1 Replace PicoSat module an error on its serial port 2 If problem persists replace main PCB module As a reference the following table lists the general categories for other error codes The error code categories are shown only in hexadecimal format Hexadecimal Code Explanation 500xxx internal user interface error 501xxx remove serial port error 502xxx date and time error 503xxx NBP error 504xxx front end error 505xxx alarm error 506xxx audio error 507xxx recorder error 508xxx trend error 509xxx flash memory data error 50axxx SpO error SObxxx ECG error 50cxxx power down task error SOdxxx on board diagnostic error 50exxx power monitor error 50fxxx temperature measurem
105. ort or servicing of the monitor you must review the following Support or sections of this document Service 11 2 Training Overview Support Strategies Theory of Operation and System Architecture Troubleshooting Monitor Applications and Algorithms Disassembly Preventative Maintenance Training Materials Training Materials Essential Materials Optional Materials Overview The essential materials to complete this training are the following The C3 Service Guide 989803129451 For Philips Service Personnel a Training Completion Form must be signed and returned to your supervisor referencing the course number MWPMD C3SUPP Optional materials that can assist you in this training are the following e aC3 monitor C3 Instructions for Use guide Tools needed for performance verification of the monitor Refer to the section titled Equipment on page 4 8 for a listing of these tools This section provides a high level overview of the monitors and their options Order ere F Number Description Display Measurements Printout Color ECG NBP Temp SpO CO Recorder 862474 Standard Color optional 862478 Sedation Color optional Note The monitor is designed for non invasive use only Common use for this monitor includes Physicians offices Transport within a care facility Low acuity monitoring in any clinical envi
106. porate Parkway 300 Sunrise FL 33323 Telephone 954 835 2600 Fax 954 835 2626 ASIA PACIFIC HEADQUARTERS Philips Medical Systems 30 F Hopewell Centre 17 Kennedy Road Wanchai Hong Kong Tel 852 2821 5888 Fax 852 2527 6727 Contents T OVervIeW ie LoL General Safety Information 0 0 0 00 a ttt e nent t ete e 1 2 Introduction zo eet ete ah A 1 3 C3 Patient Monitor Description 2 1 3 Features and Options onte ERIS uen ROG aene a ues ante o nA equa 1 4 Front Panel Description ii CHE Puer eU PU UC eee ue 1 4 Front of Monitor us esL da llas 1 5 Rear of Monitor creset Se A RR S PA as eee een PANE 1 6 Screen Display lt x cnx o A ee euros 1 7 Additional Documentation 0 0c ee s 1 8 2 Site Preparations 2 A E EEOS CEN EERE sl General Site Preparation Safety Information 0 0000 e cece cette eens 2 2 Site Preparation Guidelines EX sub ety eee X MPS eee glee oe Ree RNV 2 3 Warning Cautions and Safety Precautions 0 cece ereere 2 3 Patient Safety t usos rs Khe olde Dai 2 3 Patient Leakage Current ic a RS 2 3 Preparing to Us the Monitor s22 eos Gy Se Res Sake ODER ORDRE e e 2 4 Power Source Requirements 1 2 0 ccc ccc cen teen enn eens 2 4 Protecting Against Electrical Shock 1 0 0 0 0c ccc etn teen ees 2 4 Equipotential Grounding 0 cece cette nett n teens 2 4 Combining Equipment sesten ara Seek a a Ai A 2 4 Disposing of the Monit
107. ptom Cause of Failure Remedy Monitor fails to power up when the On Standby button is pressed Battery is dead or monitor is not plugged in Keypad malfunction or keypad connector is loose Make sure the monitor is powered on by either battery or AC power Check that the battery is adequately charged and AC power fuses are securely connected 1 Check the keypad connector 2 Replace the keypad according to the instructions in Removing the Keypad on the Front Panel on 7 6 No response when pressing the buttons on the front panel Keypad malfunction or keypad connector is loose 1 If only one button does not work verify that the monitor is On 2 Check the keypad connector 3 Replace the keypad according to the instructions in Removing the Keypad on the Front Panel on 7 6 Audible alarm does not sound Speaker malfunction or speaker has a loose connector Main PCB module malfunction 1 Check speaker connector 2 Replace speaker according to the instructions in Replacing the Speaker on 7 9 Replace the Main PCB module according to the instructions in Removing the Main PCB Module on 7 12 All patient data is lost or corrupted indicated in the error log Occurs when more than 15 minutes has elapsed when replacing the battery If this occurs replace the Main PCB module according to the instructions in Removing the Main PCB Module on 7 12 Recorder is no
108. r is based on four times 4x the maximum rated load GCX Wall Channel Always reposition the adjustable channel safety stop under the adapter plate when the height of the monitor is changed The channel safety stop prevents the mounting device from inadvertently falling or being pulled out of the wall channel Do not remove this stop while the mount is supporting a monitor Do not mount any portion of the monitor over a patient s bed Do not exceed the maximum rated load specified for each wall mount Ensure that no electrical wiring utilities or piping interface with the selected wall mounting location Do not install wall channel onto solid brick or brick veneer walls Do not attempt to install wall mounts onto crumbly wall material Check the mounting hardware holding the wall channel or bracket to the wall every 12 months Tighten if necessary Mounting the GCX Wall Channel Mounting the GCX Wall Channel To install the GCX Wall Channel and Tilt Swivel mount follow the documentation packaged with the channel Attaching the Mounting Plate to the Monitor Step Action 1 Unpack the mounting plate and check all parts are present 2 Unscrew the 5 screws from the mounting standoffs located on the rear of the monitor Place the Wall Plate over the mounting standoffs 4 Re insert and tighten the screws to secure the wallplate to the monitor Site Preparations 2 9 Attaching the
109. r limited disassembly or decompilation Upon request you will provide Philips with reasonably detailed information regarding any disassembly or decompilation You may not decrypt the Software unless decryption is a necessary part of the operation of the Software Transfer Your license will automatically terminate upon any transfer of the Software Upon transfer you must deliver the Software including any copies and related documentation to the transferee The transferee must accept these License Terms as a condition to the transfer Termination Philips Medical Systems may terminate your license upon notice for failure to comply with any of these License Terms Upon termination you must immediately destroy the Software together with all copies adaptations and merged portions in any form Export Requirements You may not export or re export the Software or any copy or adaptation in violation of any applicable laws or regulations U S Government Restricted Rights The Software and any accompanying documentation have been developed entirely at private expense They are delivered and licensed as commercial computer software as defined in DFARS 252 227 7013 Oct 1988 DFARS 252 211 7015 May 1991 or DFARS 252 227 7014 Jun 1995 as a commercial item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 Jun 1987 or any equivalent agency regulation or contract clause whichever is applicable You
110. reventative Maintenance Tests Pneumatic The following tests must be performed to verify pneumatic system functionality Perform System these tests in the following order Functionality 1 Pressure Transducer Accuracy 2 Pneumatic Leakage 3 Inflation Rate 4 Over Pressure 5 Deflation Rate Note The pneumatic system includes an over pressure safety limit function and a safety period time out function These safety functions may interfere with NBP tests described in this section In order to avoid activating these safety functions do not pressurize the system above 270 mmHg 36 kPa and do not pressurize the system for time periods that exceed 150 seconds To Zero the If the simulator does not display a 0 before starting a test use the following procedure to Simulator zero the simulator hd Gora ON onc de exc pei ec Disconnect the hose from the simulator Press Home Press SETUP Press MORE Press MORE again Press ZERO PRESSURE Press ZERO Press Home to return to main menu Reconnect the hose Before beginning the above test sequence perform the following three steps 4 16 Testing the Monitor Step Action 1 Turn the NBP simulator On 2 Perform the following sequence a Press the Home button b Press the Pressure Tests button c Press the Pressure Leak Test button d Press the Setup button e Press 2 5 0 then Enter f Set the Cuff to Internal Connect the simulator tubing to the
111. riptions of the theory of operation of each block 1 Carbon Dioxide CO2 monitoring is available only with C3 Sedation models Training 11 5 Theory of Operation and System Architecture O I 49p4029M Joxyeods Oll Z 2 SU ou s queq seyoyms ped ay ynon 101 u09 pue JOUI9IN d J9MOd 9 ueneg ul9jJs S 19MOd einpow OD pug u014 9INPON daN pug 1uo44 Aano ods pue du y dsey 993 pu3 ju014 pajejos u sun OV 209 dan ods dwoe 923 peste Figure 11 1 System Block Diagram 11 6 Training Isolated Front End NBP Front End Power System Micro processor Memory and Control Display Keypad Theory of Operation and System Architecture The Isolated Front End section includes all of the circuitry to convert ECG SpO and Temperature measurements to digital format and to connect this information to the processor The Respiration detection is obtained from two of the three electrodes of the ECG connections The NBP section contains the pumps valves pressure measurement circuitry and control circuitry for the non invasive blood pressure measurement Pressure data is converted to digital format and conveyed to the processor section The power system section contains a power supply capable of operating the monitor and charging the battery from an AC source of 100 to 240 VAC at 50 to 60 Hz This section also contains the circuitry The ba
112. ronment Training 11 3 Support Strategies Support Strategies Unit Exchange Bench Repair 11 4 Training This section provides a summary of how the monitor was designed to be supported and how Philips provides that support There are two main methods of repairing the monitor e Unit exchange Benchrepair The primary repair method used by Philips service personnel is unit exchange The current design of the monitor requires testing if the case is opened The secondary repair method is bench repair Special tools are required and are listed in Equipment on page 4 6 If the case is opened specific tests must be performed These tests are listed in detail along with the Test Map in Chapter 4 At the Philips Parts Center we will stock the board level assemblies and mechanical parts based on existing failure data and customer demand These parts are listed in Chapter 8 The availability of additional spare parts will be notified by service notes The troubleshooting guide in Chapter 6 is sufficient for assembly level bench repairs Before a repair is performed consider whether a repair or a replacement is most suitable Theory of Operation and System Architecture Theory of Operation and System Architecture System Overview Block Diagram This section contains a high level overview of the theory of operation of the major functional modules of the monitor Philips service providers perform Bench Repair at
113. rs are properly engaged at each end to prevent rotation or other strain With the sensor or electrodes applied to the patient and the monitor turned On flex the patient cables near each end to make sure that there are no intermittent faults Cleaning Cleaning Caution Do not immerse the monitor in liquid or use caustic or abrasive cleaners Do not spray or pour any liquid on the monitor or its accessories Do not allow any liquid to penetrate connectors or openings in the monitor s chassis To clean the monitor dampen a cloth with a commercial nonabrasive cleaner and wipe the top bottom and front surfaces lightly For cables sensors and cuffs follow the cleaning instructions in the Directions for Use shipped with those components If liquid is accidentally spilled on the monitor clean and dry it thoroughly before reuse If in doubt about monitor safety refer the unit to qualified service personnel Battery Maintenance About the Battery This section provides some information on how to handle and maintain the battery Warning Follow local governing ordinances and recycling plans regarding disposal or recycling batteries and other device components Storing the monitor for a long period of time without charging the battery can degrade the battery capacity A complete battery recharge gt 90 requires 8 hours in standby mode or 14 hours in operational mode If the battery n
114. rt numbers are listed in the tables that follow Item No corresponds to the callout number in Figure 8 1 through 8 6 the Page Ref for Disassembly indicates the page number of the relevant disassembly procedures in Chapter 7 Disassembly Spare Parts 8 1 Small Parts Kit Small Parts Kit Most small parts and hardware screws hex nuts washers standoffs 2 fuses CO spring and door and 2 housing feet can be found in the Small Parts Kit Part Number 453563480201 The picture below calls out a few of the less commonly known parts within this kit CO Door CO Spring noa Housing Feet Hex Nut for Navigation Wheel Power entry CO Coupler module fuses Retaining Clip 8 2 Spare Parts C3 Top Level Assembly Front Panel Assembly Front Panel Assembly Figure 8 2 1 of 2 Page Ref for Item No Part Number Description E Disassembly 1 453563480151 Main PC Board page 7 12 2 453563480111 Navigation Wheel and Retainer page 7 5 Spring 3 453563480201 Hex Nut page 7 5 Spare Parts Kit 8 4 Spare Parts Figure 8 3 2 of 2 Front Panel Assembly Page Ref for Item No Part Number Description Disassembly 3 453563480021 Backlight Tube page 7 9 4 453563480101 Keypad page 7 6 5 453563480221 Speaker including speaker cable page 7 9 6 453563480081 Display Cables page 7 7 7 453563480051 Bezel Protective Window Display No Disassembly
115. s Along with the waveforms the recorder prints the values of the vital signs being displayed The printing continues until the user presses either recorder button a second time The Snapshot button initiates a snapshot printout for 20 seconds of the same information recorded by the continuous control If scrolling is enabled in a display frame containing trend data when the Continuous button is operated then the trend record for that vital sign is printed If the Snapshot button is operated then only the trend data on the display is printed The recorder may be programmed via the monitor menu display to print a snapshot recording when an alarm condition occurs Printing is accomplished on 50 mm wide thermal paper at recorder speeds programmable up to 55 mm s The technique used in ECG senses the varying potential difference between two points at the skin surface which respond to the electro chemical actions of the muscular activity of the heart Three electrodes are attached to the patient s right arm RA left arm LA and left leg LL The varying potentials at these locations are cable connected to the ECG circuit inputs where they are conditioned and the difference of potential between two selected leads is digitized before transmitting through opto isolators to the processor The processor installed algorithms operate on the signals to develop drivers for the graphic display and to compute the heart rate in beats per minute BPM
116. s 6 9 4 Loi WI Screw 1 Screw 2 5 Gently slide the chassis away from the rear case Disassembly 7 11 Disassembly Procedures Removing the Main PCB Module Step Action 1 Unscrew the PCB module by loosening the following 3 screws Screw 2 2 Gently pull the module towards you 3 After you have removed the Main PCB module you should have access to the CO SpO NBP pump and the Patient Monitoring I O modules CO Module pO Module NBP Pump Patient Monitoring T O Module 7 12 Disassembly Removing the Patient Monitoring I O Module Disassembly Procedures Follow these instructions for removing the I O module from the Main PCB module Step Action Remove the following 2 screws to remove the connector module Screw 1 Screw 2 Note Use a flathead screwdriver to remove Screw 1 Slide the I O module out Disconnect the CO connector and tubing and the NBP tubing Slide the CO exhaust tubing out Pull the I O module away Patient Monitoring VO Module Disassembly 7 13 Disassembly Procedures Removing the SpO Module 7 14 Disassembly After you have removed the Patient Monitoring I O module from the Main PCB module you can remove the SpO module To do this Step Action Remove the 2 screws fastening the SpO2 module to the I O module Screw 1 Remove the connector tha
117. ssembling the monitor in order to remove or replace suspected defective assemblies or components Disassembly 7 1 Introduction Introduction The monitor consists of two main assemblies the front case assembly and the rear case assembly All part numbers and exploded views of some assemblies are located in Chapter 8 Spare Parts on page 8 1 The monitor can be disassembled down to all major component parts including PCBs Battery Cables Function buttons Chassis enclosures Tools Required 7 2 Disassembly You will need the following tools to disassemble the monitor small medium Philips head screwdriver flathead screwdriver needle nose pliers 10 mm socket wrench for navigation wheel 5 16 nut driver Note If you want to disassemble the RS 232 connector located on the rear panel you will need a 3 16 inch socket wrench Disassembly Safety Information Disassembly Safety Information Warning Performance Verification Do not place the monitor into operation after repair or maintenance has been performed until all recommended Performance and Safety Tests listed in Chapter 4 Testing the Monitor of this Service Guide have been performed Failure to perform all tests could result in erroneous monitor readings Before attempting to open or disassemble the monitor disconnect the power supply from the monitor Before you begin any disassembly procedure you must remove the battery H
118. sure in the pneumatic system means for controlling the pump and valves are open or closed Warning A blood pressure cuff connected to the monitor should never be applied to a human being while the monitor is in Diagnostic Mode Injury could result DIAGNOSTIC MENU NBP TEST Pressure mmHg 0 Proportional Valve Open Proportional Valve Open Press NBP to activate pump release to stop pump Press Volume to open both valves release to close valve Press Alarm Silence to open proportional valve and deflate at 3mmHg s release to close valve Press Contrast to perform offset adjustment Return The NBP Test screen elements are described below Menu Item Function Pressure mmHg Displays the real time value of the system pneumatic pressure in mmHg The value is updated at the rate of approximately two times per second Proportional Valve Indicates whether the proportional value is open or closed Safety Valve Indicates whether the safety valve is open or closed 5 10 Configuring the Power up Defaults Menu Diagnostic Mode Menu Item Function Activate Pump As long as the NBP button is pressed the pump runs If system pressure reaches the hardware over pressure protection point 280 to 330 mmHg or 37 3 to 44 0 kPa the safety valve opens and the pump disables until the pressure falls below the safety threshold Open Both Valves As long as t
119. sure shown If the pressure difference is less than 120 mmHg the pump is not strong enough and should be replaced regardless of the Pump Op Time Step Action Connect the flowmeter to the CO FilterLine On the flowmeter check the flow that the CO Pump draws X5 The nominal value is 50 the acceptable limits are 42 5 and 65 ml minute If the value is within these limits proceed to the CO Gas calibration check If not within the limits calibrate as follows Select Flow Rate Check and Calibration Adjust the flow in the monitor by selecting Increase Flow or Decrease Flow until it is as close as possible to 50 ml per minute as indicated on the flowmeter gauge The pump voltage displayed on the second row of the CO Test screen will vary as the flow is changed When you are satisfied that the flow is set as close as possible to 50 ml per minute select Store Flow to confirm the setting Note If the adjusted flow is not stored within 60 seconds of the adjustment the old flow setting 1s restored If the flow cannot be adjusted to within tolerance the pump should be replaced If the flow adjustment still cannot be made this indicates a fault in the CO module which must be replaced Testing the Monitor 4 33 Performance Procedures CO Gas After switching the monitor on or after turning on the etCO On Off setting in the etCO Measurement frame wait at least 20 minutes before checking the cal
120. t indicated below Table 4 4 Earth Leakage Current Values Test Condition Polarity Allowable Leakage Current Normal Normal 300 pA Reversed 300 pA S F C Normal 1000 pA Open Supply Reversed 1000 pA a S F C Single Fault Condition What to record on the service record S 2 Por SQ F Testing the Monitor 4 41 Patient Safety Tests Enclosure This test is compliance with EN60601 1 Enclosure Leakage Current Test at 110 of the Leakage nominal line voltage Current Step Action 1 Connect the AC mains power cord to the analyzer as recommended by the analyzer operating instructions 2 Using the appropriate test cable connect the analyzer to either of the screws on the back of the monitor next to the handle 3 Turn the monitor on 4 Perform the test as recommended by the analyzer operating instructions The analyzer leakage current indication must not exceed the values listed below Table 4 5 Enclosure Leakage Current EN60601 1 1990 A1 A2 A11 A12 A13 and UL2601 24 Ed 1997 US Deviations Test Condition Polarity Allowable Leakage Current Normal Normal 100 pA Reversed 100 pA S F C Open Normal 300 pA Protective Earth Reversed 300 pA What to record on your service record S 3 P or S 3 F 4 42 Testing the Monitor Patient Leakage Current Patient Safety Tests This test measures patient leakage current in accordance
121. t attaches the SpO module to the I O module u Gently pull the SpO module away Disassembly Procedures Removing the CO Module Note This procedure is only for those that have the C3 Sedation model Step Action 1 Remove the 4 screws on the CO module Screw 1 Screw 2 Screw 3 2 Tilt the CO2 module to wards the NBP pump Gently pull the module away from the Main PCB module Note When reassembling the CO module ensure that the CO header matches up with the mating header Disassembly 7 15 Disassembly Procedures Removing the NBP Pump Step Action 1 Remove the connector from the NBP pump Connector 2 Cut the tie wrap 7 16 Disassembly Removing the Power Supply Module Disassembly Procedures Step Action 1 Use the needle nose pliers to unplug the three connectors Battery Cables Power Supply Board Connector Communicati ns Module 2 Remove the two screws that connect the power supply to the rear panel 3 Remove the screws on the side and bottom of the rear panel 4 Cut the tie wrap that holds the battery cables together 3 Use a 5 16 nut driver to unfasten the power supply from the rear panel 6 Pull the power supply module away from the rear panel Disassembly 7 17 Disassembly Procedures 7 18 Disassembly 8 Spare Parts This chapter lists spare parts along with pa
122. t printing Defective recorder Check that the recorder is correctly installed and there is paper in the recorder If problem still exists replace recorder according to the instructions in Removing the Optional External Recorder on page 7 6 Recorder paper will not advance Paper is in incorrectly Defective recorder Replace paper as instructed in the section titled Recorder Paper Replacement in the C3 Instructions for Use guide Check that the recorder is correctly installed and there is paper in the recorder If problem still exists replace recorder according to the instructions in Removing the Optional External Recorder on page 7 6 Troubleshooting 6 9 Part 2 Isolating and Solving Monitor Problems Symptom Cause of Failure Remedy No sound from monitor Tone is switched off or low 1 Turn the volume up by pressing the Volume button 2 Power cycle monitor Tone should sound when monitor boots up 3 Check cable connections to the speaker Incorrect Time Stamps Invalid data stored showing wrong times for data The monitor s real time clock is defective Power the monitor On and then reset the time and date See the C3 Instructions for Use guide for instructions on how to set the date and time Cannot reprogram the time and date correctly Main PCB module malfunction Replace the Main PCB module according to the instructions in Removing
123. t step is to Restore Factory Settings See Restoring Factory Settings on 5 14 for additional information Warning Do not attempt to troubleshoot the monitor if there is any sign of burning or smoking coming from the monitor Call the Philips Response Center or your local Philips representative immediately When first troubleshooting the monitor check for obvious problems by answering the following questions 1 Is the monitor turned On 2 Is the battery adequately charged 3 Ifrunning from mains power supply is the AC power cord connected to the monitor and plugged into an AC outlet You can isolate many problems by observing indicators on the monitor before it is necessary to open the monitor Front Panel LED Description AC LED rE On mains connected battery charging Off mains unplugged Battery LED On DC power on monitor on Off DC power off monitor off Part 1 Troubleshooting Checklist The following two steps are to make sure that the monitor s AC power supply and battery work correctly What To Do If The Monitor Cannot Be Switched On AC Powered Step Action Ensure the battery has had ample time to charge at least 15 minutes A dead battery will prevent the monitor from powering On Check AC fuses in the AC receptacle Check keypad connector or replace keypad See Removing the Keypad on the Front Panel on 7 6 for instructions If monitor s
124. tches the calculated mmHg value within the calculated tolerance If so the CO module is correctly calibrated If the value is outside the tolerance calibrate as follows If not already connected connect the 5 calibration gas 10 Select CO Calibration to enter the Calibration sub menu Wait for the status message CO module reset in progress to disappear 11 Select the value for the calibration gas The default value is 596 12 Select Start Calibration and open the valve on the calibration gas to allow CO gas to flow into the monitor Allow the value to stabilize before the start of the calibration Leave the valve open until the monitor gives a prompt that the gas can be removed 13 The CO module calibrates and prompts when calibration is successful Calibration Verification Reset Pump Operating Time Counters Performance Procedures Step Action Reopen the 5 gas valve and allow the value to stabilize Check that the value displayed on the monitor is correct and within the tolerance see Step 2 in above section Disconnect the 5 calibration gas and connect the 10 calibration gas Open the valve on the flow regulator to allow 10 CO gas to flow into the monitor Allow the value to stabilize Check that the value displayed on the monitor X7 is correct and within the tolerance see Step 6 above If one or both values are not within tolerances
125. te Press the On Standby button to turn the monitor On D Ma BR wy rn After the power up sequence verify the following monitor reactions a The monitor displays a respiration rate of 120 3 breaths per minute b The audible alarm sounds and the Respiration Rate frame flashes indicating the respiration rate is above the default upper alarm limit 7 Press the Alarm Silence button to silence the alarm 8 Decrease the respiration rate setting on the respiration simulator to 20 breaths per minute Verify that the monitor displays a respiration rate of 20 3 breaths per minute 4 26 Testing the Monitor SpOs Performance Dynamic Operating Range Test Dynamic Operating Range LED Excitation Test Tools Needed for SpO Tests SpO testing includes the following tests Performance Procedures e Clinical Dynamics Corp SmartSat simulator with Nellcor simulator cable SpO adapter cable M1943A e SpOsreusable sensor adult finger M1191A The following test sequence verifies proper monitor operation over a range of input signals Step Action Connect SpO simulator cable to the SpO adapter cable Connect the cable to the SpO patient monitoring input connector Set the simulator as indicated in the table below Verify that the monitor readings are within the indicated tolerances Allow the monitor several seconds to stabilize the readings Turn the mo
126. tem Setting Oximeter Nellcor SpO0 7 81 BPM 36 Pulse Mod 0 50 4 Set the NBP simulator to a pressure setting of 120 80 mmHg 16 11 kPa and a heart rate of 80 bpm 4 22 Testing the Monitor Performance Procedures Step Action With the monitor Off press the On Standby button When the monitor is turned on the battery icon may initially indicate a higher charge then it actually holds wait 5 minutes or until after an NBP measurement for an accurate battery reading Verify that the monitor is responding to the SpO simulator signal and that the audible alarm is sounding Use the navigation wheel to select the SpO Menu and permanently silence the SpO audible alarm Select the Heart Rate Menu and silence the Heart Rate alarm Use the navigation wheel to select the NBP Menu and set the Automatic Interval Mode option to 15 minutes Select Return to exit the menu Press the NBP button on the monitor s front panel to start the first NBP measurement Subsequent NBP measurements are taken automatically every 15 minutes Verify that the battery icon displays at the bottom of the display Note the time At least one segment of the battery icon should be filled 10 Keep the monitor On until the low battery warning alarm occurs Note the time Verify that alarm sounds 15 30 minutes before the battery fully discharges 11 Keep the monitor On until it automatically powers down due to lo
127. tests which the exception of the Basic Pneumatic Leakage BPL Test require the use of an NBP simulator Note The monitor must be placed in Diagnostic Mode with the NBP test screen active for each of the NBP tests To place your monitor into this mode see page 4 9 Tools Needed for NBP Testing NBP cuff 40401C NBP tubing M1599B Bio Tek BP Pump 2 simulator or equivalent with an internal test volume of 310 ml 4 14 Testing the Monitor Preventative Maintenance Tests Documenting NBP Test l l Results The following table lists the tests that should be documented and summarizes how to document the NBP test results Test Expected Test Results What to record on service record Pressure Transducer Accuracy Test X1 difference between the pressure on the monitor and on the simulator Difference less than or equal to 5 mmHg 0 7 kPa Pneumatic Leakage Test X2 difference between the pressure at the start P1 and after one minute P2 Difference less than or equal to 6 mmHg 0 8 kPa Inflation Rate Test X3 number of seconds for the monitor to inflate Rate between and 6 seconds Over Pressure Test X4 peak value Peak value between 270 and 330 mmHg 36 kPa and 44 kPa Deflation Rate Test X5 mmHg kPa per second to deflate 1 8 mmHg s to 4 8 mmhg s PN P X1 X2 X3 X4 X5 or PN F X1 X2 X3 X4 X5 P passed F failed Testing the Monitor 4 15 P
128. the CO module must be exchanged If the pump in the CO2 module is replaced the Pump Operating Time counter should be reset to start counting operating time for the new pump Step Action In the CO2 Test menu select Pump Op Time Select Reset to Zero Note When the Pump Op Time has been reset a CO2 Equipment Malfunction INOP will be generated on restarting the monitor To clear this INOP you must perform a flow check and store the flow in Diagnostic Mode select Store Flow Testing the Monitor 4 35 Performance Procedures Serial Perform the following procedure to test the serial port voltages The test is qualitative and Interface only verifies that the serial interface port is powered correctly and that the Nurse Call signal and Nurse is operational The serial connector is a male DB 9 located on the monitor s rear panel Call Signal identified by the RS 232 symbol Test Tools Needed Clinical Dynamics Corp SmartSat simulator with Nellcor simulator cable SpO adapter cable M1943A Step Action 1 Turn the monitor On 2 Set up the DMM with the function set to VDC at a range of 10 volts 3 Connect the DMM negative lead to connector pin 5 GND or the shell of the RS 232 connector 4 Referring to Table 4 2 Serial Interface Voltages connect the DMM positive lead to each pin in turn and verify the voltage va
129. till cannot be switched On replace the Main PCB module See Removing the Main PCB Module on 7 12 for instructions What To Do If The Monitor Cannot Be Switched On Battery Powered Step Action Plug the monitor into an AC power source Ensure the battery has had ample time to charge at least 15 minutes A dead battery will prevent the monitor from powering On Check AC fuses in the AC receptacle Check keypad connector or replace keypad See Removing the Keypad on the Front Panel on 7 6 for instructions If monitor still cannot be switched On replace the Main PCB module See Removing the Main PCB Module on 7 12 for instructions Troubleshooting 6 5 Part 2 Isolating and Solving Monitor Problems Part 2 Isolating and Solving Monitor Problems INOP Messages The following table explains the technical INOP messages that the monitor can issue and suggests a course of actions When actions are numbered always try them in the order given and only proceed to the next action if the current one is not successful in solving the problem incorrect line is attached Message Cause of Failure Remedy CO Equip Malf Malfunction in the CO hardware Verify failure and replace Extension with a known good Extension CO No Tubing The FilterLine is disconnected or an Attach a FilterLine Remember that only Microstream accessories may be used with the monitor
130. tion for 3 seconds the contrast control function is terminated by the processor The contrast control function is also terminated if the wheel is pressed any time within this 3 second interval Operation of the Volume button accomplishes similar functions for the volume of the heart rate audible tone as the display contrast control button does for the display Pressing this button enables the wheel to vary the tone volume This is a rotating push switch wheel The associated wheel circuitry generates a pulse when pressed and generates a digitally encoded pair of quadrature signals whose relative magnitudes and polarities represent the angular position of the wheel These outputs are connected to the processor where they are interpreted as required for the functions involved Successive angular positions determine the direction of wheel rotation In addition to the functions performed when in conjunction with the keypad as described above the wheel operates in conjunction with the display to select menus and lists of parameter variables This is a rear panel 9 pin connector providing interfaces with other computer systems or equipment The driver for this port is a Universal Asynchronous Receive Transmit UART integrated circuit that interfaces this port with the microprocessor The baud rate for this serial transmission function is programmable from 1200 baud to 38 4 kilobaud Pin number 9 of the RS 232 connector is reserved for a Nurse Call
131. ttery provides the operating power for this power The Microprocessor Memory and Control section contains the system CPU and all digital support circuitry The latter includes the RAM non volatile memory and real time clock This section also contains the display logic keypad switch interface logic RS 232 I O control defibrillator synchronization control and printer logic The display is a TFT color backlit fluorescent LCD unit The active screen size is 8 3 x 6 2 in 210 8 x 157 5 mm 10 4 in 264 2 mm diagonally and the resolution is 640 x 480 pixels The keypad circuit contains five push button membrane switches and two green LEDs The LEDs are driven by the power supply system and indicate the power source Signals from LEDs are returned to the microcomputer for processing and control as required The On Standby button connected directly to the power supply toggles the power between On and Standby modes When in Standby the display is blank and no monitoring is performed However the batteries are charging if an AC source is connected to the rear panel The Audio Alarm Control button is connected directly to the processor and to the system power supply Pressing this button turns off the battery fuse alarm in the system power supply Response of the processor depends upon the action in pressing this button If momentarily pressed less than 2 seconds alarms are paused temporarily for a preset interval determined by the me
132. tting by clicking on the adjusted value so that it is no longer highlighted 11 Check that the ambient pressure displayed in the CO Pressure line at the top of the screen is the same as the value that you selected from the list in Step 8 Testing the Monitor 4 31 Performance Procedures Leakage The Leakage Check consists of two parts Check e Part 1 Checking the tubing between the pump outlet and the CO module outlet e Part 2 Checking the tubing between the pump inlet and the FilterLine inlet Note Check the flowmeter s user guide for details on how to make a correct flow reading Part 1 Step Action Check the ambient and cell pressure shown in the status line on the screen The cell pressure should be approximately 20 mmHg 2 7 kPa lower than the ambient pressure Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube Block the CO module outlet using your fingertip and observe the flowmeter display The value on the flowmeter X2 should decrease to between 0 and 4 ml minute If the value is not within the tolerance limits there is a leakage between the pump outlet and the CO module gas outlet Ifa leakage is found in Step 3 open the CO module and check the tubing connections at the pump outlet and the module CO gas outlet If the connections are good then there is leakage in the module and the CO module must be exchanged P
133. ttom of the page Pages that are rearranged due to changes on a previous page are not considered revised The documentation printing date and part number indicate its current edition The printing date changes when a new edition is printed Minor corrections and updates that are incorporated at reprint do not cause the date to change The document part number changes when extensive technical changes are incorporated First EditiOM oooccncncooncninnnnannnninnnano nono cinannnanocinanonenos March 2003 Use this document in conjunction with the C3 Instructions for Use guide pn 989803228691 Safety Safety Standards CE Marking Europe All warnings precautions and notes are located in the chapters that follow You must read all of this safety information before you begin monitoring with your C3 patient monitor The C3 Patient Monitor is compliant with the following safety standards e UL2601 1 e CAN CSA C22 2 No 601 1 M90 EN60601 1 e EN60601 2 27 e EN60601 2 30 e EN60601 2 49 ENS865 The following products and accessories from Philips Medical Systems carry the CE mark C to Council Directive 93 42 EEC 0123 C3 Products 862474 862478 Accessories 21075A 21076A 21078A 21090A M1837A M4820A M4818A The following accessories from Philips Medical Systems carry one of the following CE marks C or C e to Council Directive 93 42 EEC Accessories M1191A M1192A M1194A MI941A
134. ubing then test again If the test fails again verify the integrity of all the pneumatic tubing inside the monitor Testing the Monitor 4 21 Performance Procedures Performance Procedures The following accuracy and performance procedures are designed to be completed to verify the accuracy and performance of the monitor They must be performed according to the frequency specified in the section titled Recommendations for Testing Frequency on page 4 6 Battery The battery performance test should be performed every two years before monitor repairs or Performance whenever the battery is suspected as being a source of the problems Before performing the Test battery performance test verify that the battery is fully charged see Charging the Battery on page 3 9 Note Before beginning this test ensure that the monitor is disconnected from the AC power source Tools Needed e SpO adapter cable M1943A e Clinical Dynamics Corp SmartSat simulator with Nellcor simulator cable NBP tubing M1599B Bio Tek BP Pump 2 simulator or equivalent with an internal test volume of 310 ml Note The instructions below apply to the Bio Tek BP Pump 2 simulator Step Action 1 Connect SpO simulator cable to the SpO adapter cable Connect the cable to the SpO patient monitoring input connector 2 Connect the NBP simulator to the monitor via the NBP tubing 3 Set the SpO simulator switches as follows I
135. ule replacement Power On Pneumatic Leakage Test All Safety tests Testing the Monitor 4 7 Equipment Equipment The following table lists the equipment required for performance verification 4 8 Testing the Monitor Equipment Description Digital multimeter DMM Fluke model 87 or equivalent Defib Sync Cable and Switchcraft 3 5 mm phone plug M4820A see page 4 38 to use a substitute cable Switchcraft 750 or equivalent SpO adapter cable M1943A SpO reusable sensor adult finger M1191A ECG trunk cable M1540C ICU AAMI or M1550C ICU IEC ECG electrodes Standard ECG lead sets M1605A AAMI or M1615A IEC NBP tubing M1599B NBP reusable cuff adult 40401C Pulse oximeter tester Clinical Dynamics Corp SmartSat simulator with Nellcor simulator cable ECG simulator Dynatech Nevada medSim 300 or equivalent NBP simulator Bio Tek BP Pump 2 or equivalent Temperature simulator MedSim 300 or equivalent Respiration simulator MedSim 300 or equivalent Safety analyzer Bio Tek 601 Pro or equivalent Stopwatch Manual or electronic Screwdriver flathead and philips Tweezers any model Electronic flowmeter M1026 60144 Cal 1 gas 5 CO2 15210 64010 Cal 2 gas 10 CO 15210 64020 Cal gas flow regulator M2267A Cal tube 13907A Serial Numbers Serial Numbers
136. vailable use a suitable carton with appropriate packing material to protect the monitor during shipping Note Do not return sensors patient cables NBP tubing and cuff or the power cord Site Preparations 2 5 Returning System Components Returning System Components 2 6 Site Preparations If it is necessary to return the monitor for service call the Philips Response Center or your local Philips representative for shipping instructions Have all equipment serial numbers available when calling To pack the monitor for return disconnect all cables It is not necessary to return sensors patient cables NBP tubing and cuff or power cord Pack the monitor to be returned in its original shipping carton if available If not use a suitable carton with appropriate packing material to protect the monitor during shipping Mounting Mounting Warning Make sure that you have read all applicable instructions before attempting to install the wall mount Wall mounts that are intended to support monitors must be capable of supporting 4 times the weight of the monitor when properly installed Do not mount any portion of a monitoring instrument over a patient s bed Do not exceed the maximum rated load specified for each wall mount Ensure that no electrical wiring utilities or piping interfere with the selected wall mounting location The rear panel includes threaded standoffs for attaching a mounting system accessory
137. w battery condition Verify that the audible alarm sounds when the monitor automatically shuts down Note the time Press any front panel button on the monitor to terminate this audible alarm 12 If the monitor passes this test immediately recharge the battery If it fails the test replace the battery See Removing the Battery on page 7 4 for more information Temperature The accuracy of the monitor s temperature measurements is 0 1 C 0 2 F In the procedure Test below add the tolerance of the simulator to the acceptable range of readings Tools Needed Dynatech Nevada MedSim 300 or equivalent simulator Temperature cable supplied with the temperature simulator Note The instructions below apply to the Dynatech Nevada MedSim 300 Step Action 1 Verify the monitor is turned Off Connect the temperature cable to the appropriate connector on the temperature simulator Connect the temperature cable to the Temperature patient monitoring input connector Testing the Monitor 4 23 Performance Procedures Step Action 3 Set the temperature simulator as follows a Temperature 37 C 98 6 F b Probe Type YSI 400 Series 4 Press the On Standby button to turn the monitor On After the power up sequence verify the temperature reads 37 C 0 1 C 98 6 F 0 2 F if Fahrenheit is selected as the temperature unit
138. with EN60601 1 Clause 19 for Class I type CF equipment Patient leakage current in this test is measured from any individual patient connection to earth power ground This test requires a sample patient cable for each device parameter These must be configured as recommended by the safety analyzer operating instructions Step Action Configure the electrical safety analyzer as recommended by the analyzer operating instructions Connect the monitor s AC mains power cord to the analyzer as recommended by analyzer operating instructions Connect the ECG test cable between the ECG connector on the monitor and the appropriate input connector on the analyzer Turn the monitor On Perform the test as recommended by the analyzer operating instructions Patient leakage current is measured under various conditions of the AC mains and protective earth conductor For each condition the measured leakage current must not exceed that indicated below Repeat the test for SpO and temperature patient connections using appropriate test cables Table 4 6 Patient Leakage Current Values Allowable Leakage Test Condition Polarity Current Max Type CF Normal Normal 10 pA Reversed 10 uA S F C Open Earth Ground Normal 50 uA Reversed 50 uA What to record on the service record S 4 P or S 4 F Testing the Monitor 4 43 Patient Safety Tests Patient This test m
139. y Do not use a faulty battery in the monitor Checking When the monitor is connected to the AC mains power supply the battery charges the Battery automatically Status Battery status is indicated several ways LEDs on the front panel of the monitor Battery gauge on the monitor s main monitoring screen NOP messages The AC LED is only on when the power cord is connected and the AC power is available to the monitor In this case the battery can be either charging or fully charged 3 8 Maintaining the Monitor Identifying Battery Strength Charging the Battery Battery Maintenance You can determine the monitor s battery strength by the color of the battery icon or by the number of illuminated bars on the battery icon The following table describes the battery icon behavior N f B MDA PESAR Color Used Behavior Illuminated 3 4 or 5 Green Constant display 2 Yellow Constant display 1 Red Constant display 0 Red Battery icon flashes and message displays Ifthe battery operating time is less than 30 minutes the battery icon flashes A low priority audio alarm sounds 1 beep every 4 seconds e When the battery is empty a Low Battery INOP message displays in the Message frame and the audible alarm sounds Pressing any front panel button clears the INOP and silences the audible alarm The battery icon continues to flash until the monitor is powered down or automatically sw
140. you power on the changes you made to the defaults are implemented Configuring the Power up Defaults Menu 5 3 Power Up Defaults Menu Menu Below are the menu options you can configure from the Power up Defaults Menu Options Menu Item Choices Explanation Accept Current Settings Yes If Yes is chosen the current monitor NT settings become the power up defaults O Selected Audio Off Make Available default If Make Available is chosen audible alarms are permanently silenced for a Deny Access particular parameter via the Alarm Limits Menu Some institutions may want to prevent audible alarms from being permanently silenced If so select Deny Access Audio Off Make Available default If Make Available is chosen Alarm Deny Access Suspend Mode can be invoked by pressing and holding the Alarm Silence button for 2 seconds Some institutions may wish to prevent Alarm Suspend from being invoked If so select Deny Access Auto Set Limits Make Available default Deny Access If Make Available is chosen Auto Set Limits are invoked via the Alarm Limits Menu Some institutions may wish to prevent Auto Set Limits from being invoked Is so select Deny Access Silence Period 30 60 90 120 180 s default Time is indicated in seconds Pressing the front panel Alarm Silence button temporarily silences audible alarms for the time indicated Enter Diagnostic Mode Yes No If Yes is chos

Download Pdf Manuals

image

Related Search

Related Contents

OWNER`S MANUAL  User`s Manual for HotPlug LT  Espremedor de Frutas FreshMix™ (66333BZ)  Samsung SC67F0 User Manual (Windows 7)  GE JESO734PIVLRR User's Manual  Recommended Standards of Practice on the Role and Duties  mClicker Application Manual  Owner`s Manual  G-3 Series Gateway User Manual  

Copyright © All rights reserved.
Failed to retrieve file