1. CAL SCOMM SWAP Model 1310 Indicator Service Manual
2. CAPAC Calibrate SETUP OPTION Calibrate SETUP METHOD 34 Press the SETUP soft key to access the following soft key group CAPAC Press this key to enter the capacity of the scale and the division size OPTION Press this key to enable or disable multi interval use METHOD Press this key to choose which calibration method to use single or multipoint Multipoint is for analog scales only Following are detailed instructions for setting these parameters If you press the CAPAC soft key follow these instructions 1 The display shows the current value for the capacity Press ENTER to accept this value or key in a new capacity and press ENTER The current division size is displayed 2 Press ENTER to accept the division size or key in a new one and press ENTER The display returns to the SETUP display Press the OPTION soft key to enable or disable the multi interval option If you enable the multi interval option the division size you choose under CAPAC applies to weight on the scale from 0 to 2 capacity For weight on the scale from 1 capacity to full capacity the division size will double When multi interval is enabled the division used to check for stability center of zero and AZT is always the smaller division size Overload and underload is always calculated based on the upper division size The displaying and printing division size depends on which interval the scale is in The active interva
3. N AZT FILTER ZERO MORE e eel e pae Set divisions Set divisions Set Push Button and time and time Zero Range for motion BORAT J e aeaa e e e e E A Zero Tracking Set AVG RATE ROC RT TO TRAFF MORE Harmonizer Enable bes eee ee peer eee eee eee e a is oe eee eee Harmonizer Constant Harmonizer Threshold Set display Configure Set Print RTZ Configure update rate rate of change Accum RTZ traffic calculation Accum TimeOut counter Fann eee Rane EE sand cae nfs eng a aed fae each ead in nce Sas ae dl Print TimeOut re arm Zero TimeOut and SCL UNITS KEYS SERIAL VALS Tare TimeOut trigger eats Rien acces pee oa E E aie E e cana points Set the Enable disable Enable disable Setport Enable disable 7 20 0 ieeeeecceeeeeeeeesennessenneeseenneeeeeeneeenntteesnneeennneesstnelecennny number of units of measure front panel baud parity values to active scales keys databits handshake be shown ae _ BEOAD ee OROSII mode and in main l EOM character weight display Enable Enable Enable Log files deadload Ghost Scom event drift analysis Scale disable warn Email on error error CELL SECT Figure 6 Select the type of ghost Individual Cell or Sectional Ghost for pairs of cells Soft key flowchart for Configure menu Model 1310 Indicator Service Man 17 ual Calibrate Menu Only appears if Exit to norma multiple scales operating mode FACTOR View firmware View the View
4. the pienennettiin tennta tenen version calib date calib data PORT 1 PORT 2 PORT 3 PORT 4 downloaded file weight and for current ms ee a fat ee e Serer ed information displayed scale L J etc weight for the Ra current scale Print out calibration Single Point Multi Point data from the selected port e eres SCOMM Appears only if SensorCcomm i is enabled Zero Span calibration calibration ER ey CRNR SECT INFO G CAL SWAP a pean eae Corner balance Section View SCOMM Ghost Change settings CAPAC OPTION METHOD specified scale calibration and sensor calibration for a given engl rr B information sensor Set the Enable disable ff e ee capacity multi interval VERS SETUP VALS and division ees A E eet ETE eee ae er Sre SINGLE View SCOMMs View SCOMM View corner b serial PN and configuration values for Ga ee software revision and cap each sensor ormal Ca p to a mvV V serial w zero amp span points including and span factor zero amp span for all sensors TZERO KEY IN een eee eee E eee ee eens TZERO Temporary zero calibration Key in Reverse counts or calibration millivolts volt SPAN Span calibration Figure 8 Soft key flowchart for Calibration Model 1310 Indicator Service Manual menu 33 Calibrate SETUP Calibrate SETUP
5. Industrial Weighing Systems 9 Richmond St Picton ON Canada KOK 2T0 Ph 613 786 0016 Cell 613 921 0397 Fax 613 476 5293 E mail info iwsystems ca Website www iwsystems ca Industrial Weighing Sytems assumes no responsibility for the accuracy of the calibration instructions extracted from manufactureres documentation Contact us for complete manuals we may have on file info iwsystems ca Industrial weighing systems provides repairs to all insturments and load cells Calibration services available Should you require additional information please do not hesitate to contact WSystems Check out our products at www iswystems ca Configure Menu Exit to normal operating mode CLRPRG SER Clear the Enter the program from indicator memory serial number on display Beeper volume Enable disable Set print Set power up Set Set analog out parameters showing of format ports display mode Excitation of digital to analog channels setpoint state Date format Channel Basis Scale Min Max Default print format and adjust output Enable disable lowercase Enable disable display cycle Enable disable sleep timer MOT
6. ght from the scale then press ENTER After the indicator has calibrated the zero point the display says DONE and asks you to press any key 2 Above the text you will see the weight displayed It should read zero in the increments you ve chosen If not you should perform this step again The display returns to the CAL display If you press the SPAN soft key follow these instructions 1 The current span calibration weight is displayed Press ENTER to accept this weight or key in a new one and press ENTER The display prompts you to apply the test weight load to the scale 2 Apply the test weight load to the scale and press ENTER The indicator determines the span and tells you when it is done Above the text the display should show you the correct test weight If not perform the span again 3 Press any key to return to the CAL display Use the SPEC soft key to access three new softkeys These soft keys are listed below TZERO This stands for temporary zero This calibration procedure is useful when a scale has weight on it that is impractical to remove such as a hopper or bin which is partially full of mate rial This calibration allows you to establish a temporary zero add weights to the scale and do a span without losing the zero reference point KEY IN This calibration procedure allows you to key in calibration data mvV V or counts This means calibration data can be transferred between different indicator
7. he display prompts for the cal zero mV V value Key in the value from the original indicator and press ENTER The display shows the current cal weight Accept the displayed cal weight if it is the same as the cal weight of the original indicator by pressing ENTER or key in the correct value from the original indicator and press ENTER The display shows the span mV V Accept the displayed span mV if it is the same as the span mV V of the original indicator by pressing ENTER or key in the correct value from the original indicator and press ENTER Display shows the SPEC soft keys If you press the R CAL soft key you can perform a reverse calibration In other words you start with a loaded scale remove all the weight from the scale for spanning then get your zero reading This may require a container on a separate scale into which you discharge the material Press the R CAL soft key Display shows the title Reverse Calibration Span and prompts you to press ENTER Press ENTER The indicator determines span point 1 and shows DONE when finished The display prompts you to press any key to continue Press any key Display prompts you to enter a calibration weight This is the weight that is currently on the scale You will remove all this weight from the scale in step 5 This value is used for spanning the scale Key in the calibration weight value equal to the amount to be removed from the scale in ste
8. l is chosen based on net when a tare weight is active and gross weight when a tare is not active When gross is displayed or printed regardless of the tare value the division size used depends on the interval used After pressing ENTER to accept your choice of enabling or disabling the multi interval option the display returns to the SETUP soft key choices Press the METHOD soft key to choose single or multipoint calibration If you choose Single the choices available under the CAL softkey for calibration allow you to do zero and span calibration If you choose Multi the choices available under the CAL softkey for calibration allow you to calibrate with up to 10 varying weights Model 1310 Indicator Service Manual Calibrate CAL single Calibrate CAL ZERO Press ESC to abort Calibrate CAL SPAN Press ESC to abort Calibrate CAL SPEC Press the CAL soft key and you are presented with the soft key set for the type of calibration method chosen in SETUP Single or Multi If you have chosen Single you will see the softkey set used to set zero and span shown below ZERO Use this key to set the zero reference SPAN Use this key to span the scale SPEC Use this key to access another set of soft keys for specialized calibration Following are detailed instructions for setting these parameters If you press the ZERO soft key follow these instructions 1 The display asks you to remove all wei
9. p 5 and press the ENTER key The indicator prompts you REVERSE CALIBRATION ZERO Remove load Press ENTER Empty the scale and press ENTER The indicator acquires span point 2 and shows DONE when finished Reference zero is acquired at this time Press any key to return to the SPEC display Model 1310 Indicator Service Manual 37 Calibrate CAL SCOMM The SCOMM soft key only appears if the Sensorcomm option is enabled 38 Use the SCOMM soft key to access the corner and section balancing functions and to see information on the SensorComm and associated weight sensors 1 Press SCOMM The following is displayed CALIBRATE CAL SCOMM CRNR sect INFO G_CAL swap The CRNR key is for calibrating the scale using corner weights See Corner Balancing Calibrate CAL SCOMM CRINAR section below The SECT key is for calibrating the scale using section weights See Section Weight Adjustment Calibrate CAL SCOMM SECT section below The INFO key lets you view SensorComm and weight sensor specifications See the section INFO soft key The G CAL key lets you set the ghost calibration See the section Ghost Calibration Calibrate CAL SCOMM GHOST The SWAP key allows you to quickly and easily replace a faulty weight sensor with a new one and be confident your accuracy remains high For instructions on swapping weight sensors see the section called Sensor Swap Procedure Calibrate
10. s if one fails and no new calibration procedure need be done R CAL This stands for reverse calibration Use this procedure when it is impractical to hang weights from a full or partially full hopper or bin You can key in a span weight unload that weight onto another scale perform a zero calibration and your scale is calibrated Model 1310 Indicator Service Manual 35 Calibrate CAL SPEC TZERO Calibration should be verified with certified test weights at your earliest convenience Calibrate CAL SPEC KEY IN Calibrate CAL SPEC KEY IN CNTS 36 Following are detailed instructions for these soft keys If you press the TZERO soft key you are shown two new soft keys TZERO and SPAN To perform the temporary zero spanning procedure press the TZERO soft key The indicator performs a zero function the display says DONE and asks you to press any key Above the text you will see the weight displayed It should read zero in the increments you ve chosen If not you should perform this step again The display returns to the TZERO display Next press the SPAN soft key 1 The current span calibration weight is displayed Press ENTER to accept this weight or key in a new one and press ENTER The display prompts you to apply the test weight load to the scale 2 Add the test weight load to the scale and press ENTER The indicator determines the span and tells you when it is done Above the
11. text the display should show you the correct test weight If not perform the span again 3 Press any key to return to the CAL SPEC TZERO display Press ESC to return to the CAL SPEC display Press the KEY IN soft key if you want to set up a new indicator to replace another indicator and keep the original indicator s calibration settings To do this you must have recorded the zero counts or mV V values of the original indicator in order to transfer that information to the new indicator This information can be found in the Calibrate menu under the INFO and FAC TOR soft keys If you press KEY IN you are asked if you want to set up the calibration using zero counts or mV V Choose one If you press CNTS follow these steps 1 The display shows the current zero counts value Key in the value from the original indicator and press ENTER The display shows the current span weight 2 Key in the span weight from the original indicator and press ENTER The display shows the span counts 3 Key in the span factor from the original indicator and press ENTER Display returns to the SPEC soft keys Model 1310 Indicator Service Manual You may need to remove the weight and weigh it on another scale to determine this value You need to remove all the weight from the scale in this step Calibrate CAL SPEC KEY IN MV V Calibrate CAL SPEC R CAL If you press MV V follow these steps T