1.   410 autosampler   420 autosampler   430 autosampler   440 LC fraction collector  500 column valve module  510 column thermostat  530 fluidics module  Personal computer  typical     Printer  typical     Required supply voltage    115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  230 V AC  50 60 Hz  120 230 V AC  50 60 Hz  120 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115 230 V AC  50 60 Hz  115  120  220  240 VAC  115  120  220  240 VAC    Nominal rating  270 VA 550 VA  350 VA   1000 VA   130 VA   130 VA   480 VA   150 VA   200 VA   250 VA   250 VA   100 VA   250 VA   400 VA   150 VA   300 VA   85 VA       Publication no  8510253100  Issue 1  July 2009       Varian Liquid Chromatography Modules       3 1 2 LC System Power Connections    Table 4  power connections for Varian liquid chromatography  module    Cable rating  220 240 V countries  5 A minimum  Cable rating  100 120 V countries  12 A minimum    i ll Plug supplied Required wall socket    Complies with AS3112 To comply with AS3112   HPM 787 or Clipsal  2015  250 V  10 A     Complies with Nema 5 15P   To comply with Nema  5 15R  Hubbell 5262   125 V  15 A   Canada Complies with Nema 5 15P   To comply with Nema  Complies with CS 22 2 No   5 15R  Hubbell 5262  42  125 V  15 A     Complies with CEE 7 To comply with CEE 7  Sheet VII Sheet VII  250 V  10 16  
2.  Varian modular liquid chromatographs are inherently robust  and  the packaging is designed to prevent damage  It must be  remembered that the contents form part of a precision measuring  system and all packages should be handled accordingly  In transit   sharp jolts and shocks must be avoided and the packages must not  be inverted or tilted unnecessarily  Markings on the shipping  cartons generally indicate the required orientation of the carton     After all boxes have been checked for damage  check the checklist  box  All unopened boxes have been checked for damage     Publication no  8510251300  Issue 1  July 2009 Page 25 of 30    Varian Liquid Chromatography Modules       4 4 Unpacking the Instrument    Do not unpack the instrument under any circumstances without the  presence of a Varian field support representative     Place all boxes near the final instrument location     After all boxes have been placed near the final instrument location   check the checklist box  Instrument has been placed near its final  location     Warning   AN The Varian liquid chromatograph module can weigh between  11 34 kg  25 75 Ib   To avoid injury to personnel or damage to  equipment  always use suitable lifting procedures when moving  the modules  Not all options may be present  so consult your  purchase order for more information     Page 26 of 30 Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       5  Materials and Solvents       The Varian field 
3.  software  consult the Varian Modular HPLC System  Installation Overview  publication number 8510251200  which is  supplied with each instrument     For networking Galaxie    software  consult the Galaxie CDS  Installation Guide     Page 22 of 30 Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       4  Instrument Shipping  Information    Generally  Varian modular liquid chromatographs are sold Free On  Board shipping point  with the transportation from this point at the  customer s expense     4 1 Insurance After Delivery    As the carrier   s liability ceases when the equipment is delivered   Varian recommends that the instrument owner arranges separate  insurance that will cover transportation from the delivery point to the  installation site  The delivery point will vary according to the carrier   the shipping method  and in some cases the terms of sale  Some  carriers will only deliver to their own distribution center  while others  may deliver to the actual installation site     4 2 Weights and Dimensions    Table 5  Weights and dimensions    System une Jhen  Wie  on  weon                         Solvent Delivery Modules   210 218 pump 19 7cm 29 2 cm 46 4cm  22 0 kg  7 75 in  11 5 in  18 3in    48 0 Ib   SD 1 pump 27 0cm 41 0 cm 56 0cm   34 0 kg  10 5 in  16 0 in  22 0in   75 0 lb   SD 2 pump 19 7 cm 29 2 cm 57 5cm  27 0kg  7 6 in  11 5 in  20 3 in   60 0 Ib   Detectors   325 UV Vis detector 21 2 cm 29 6 cm 47 5cm  15 5kg  8 3 in  1
4.  temperature maintained between 10 and 35   C  50 and 95   F      Cleanliness    Limit dust levels to less than 36 000 000 particles  0 5 microns or  larger  per cubic meter of air  This is equivalent to a clean office     After the cleanliness requirements have been met  check the checklist  box  Laboratory is free of excessive particulate matter     Instrument Cooling Air Supply    The Varian modular liquid chromatograph requires clean  dry   non corrosive air for cooling purposes  This is supplied to the  instrument through air supply vents in the instrument     The air supply is used to cool the electronics of the instrument   Several of these assemblies contain parts prone to corrosion  The  introduction of cooling air contaminated with high levels of acid  vapor or other corrosive substances may cause damage to the  instrument     Due to the corrosive nature of some analytical work  it is  recommended that in applications demanding high usage of  corrosive materials  an external cooling air supply system be  provided  It is strongly recommended that the cooling air be  supplied from an environmentally controlled area that is away from    Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       the instrument and any other area where corrosive materials are  stored or used     2 3 Workbench    The Varian modular liquid chromatograph is a precision instrument   The workbench must be free from vibration and must be stable and  strong eno
5. 1 7 in  18 7 in    34 0 Ib                   Publication no  8510251300  Issue 1  July 2009 Page 23 of 30    Varian Liquid Chromatography Modules       Page 24 of 30                                                             System Unit Height Width Depth Weight  335 photodiode array 21 2 cm 29 6 cm 47 5cm  15 5 kg  detector 8 3 in  11 7 in  18 7in   34 0 Ib  356 LC RI detector 21 2 cm 29 6 cm 47 5cm  11 0 kg  9 0 in  12 0 in  19 0 in   25 0 Ib  380 385 LC ELS detector 41 5 cm 20 0 cm 45 0cm  11 0 kg  17 0 in  8 0 in  18 0 in   25 0 lb  Auto Samplers Fraction Collectors  410 autosampler 34 0 cm 30 0 cm 50 0cm   19 0 kg  13 4 in  11 8 in  19 7in    42 0 Ib  with cooling option 21 0 kg  46 0 Ib  420 autosampler 44 0 cm 28 0 cm 54 0cm  22 0 kg  17 3 in  11 0 in  21 3 in    48 0 Ib  with cooling option 26 0 kg  66 0 Ib  430 autosampler 44 0 cm 28 0 cm 40 0cm   21 0 kg  17 3 in  11 0 in  15 7in    46 0 Ib  with cooling option 26 0 kg  58 0 Ib  440 LC fraction collector 51 0 cm 49 0 cm 28 5cm  18kg  21 0 in  20 0 in  12 0in   34 lb  Peripheral Components  500 column valve module 50 8 cm 13 7 cm 38 1cm  13 6 kg  20 0 in  5 4 in  15 0in    30 0 Ib  510 column thermostat 52 0 cm 17 0 cm 33 0cm  16 0 kg  20 5 in  6 8 in  13 0in    35 2 Ib  530 fluidics module 19 7 cm 29 2 cm 47 0cm  15 0 kg  7 6 in 11 5 in  18 5in    32 0 Ib  Ancillary equipment  Mobile   46 0 cm 31 0 cm 46 0cm       phase reservoirs  column 18 0 in  12 0 in  18 0in        mount assembly  dynamic  mixer  manual inject
6. A        After the electrical requirements have been met  check the checklist box   Specified electrical supply and power outlets installed     3 1 3 Fuses    Information on fuses can be found in each of the individual  manuals     Note For safety reasons  any other internal fuse is not operator  accessible  and should only be replaced by Varian authorized  personnel  Fuse information on the rear of the instrument is the  most up to date     Publication no  8510251300  Issue 1  July 2009 Page 19 of 30    Varian Liquid Chromatography Modules       3 2    3 2 1    Page 20 of 30    Gas Supplies    Except for the ELS detector  unless your system is equipped with a  helium sparging unit  there are no gas requirements for the Varian  modular LC system  For gas requirements on these other devices   consult the operation manual or other documentation enclosed with  the particular device     The installation of compressed or liquid gas supplies must comply  with the rules and or regulations imposed by the local authorities  responsible for such use in the workplace     The ELS detector should be supplied with clean  dry nitrogen gas    gt 98  purity  filtered to 0 2 um   between 414 689 KPa    60 100 PSI  in pressure  For optimum performance a gas  pressure of 414 KPa  60 PSI  is recommended  Air can only be  used for non flammable solvents  The instrument is not calibrated  for use with gasses other than air or nitrogen  For operation with  other inert gases contact your Varian repre
7. Varian Liquid  Chromatography  Modules    Pre installation Manual    Installation Category II  Pollution Degree 2  Safety Class 1  EN 61010 1        NOTICE  This document contains references to Varian  a  Please note that Varian  Inc  is now part of Agilent fea no  9910293100  Technologies  For more information  go to     July 2009  www agilent com chem        pe Agilent Technologies       Varian Liquid Chromatography Modules       Varian  Inc      Serving Industries Worldwide  Biosciences   Pharmaceuticals   Clinical Research and Forensics   Food and Agriculture   Chemical Analysis   Environmental   Fuels and Energy    Material Sciences    Sales and Service Contact Details   North America  800 926 3000  925 939 2400  Europe  The Netherlands  31 118 67 1000  Asia Pacific  Australia  613 9560 7133   Latin America  Brazil  55 11 3238 0400    Varian  Inc  Web Site  www varianinc com    Varian  Inc  is the owner of copyright on this document and any  associated software  Under law  the written permission of Varian   Inc  must be obtained before the documentation or the software is  copied  reproduced  translated or converted to electronic or other  machine readable form  in whole  or in part     First published July 2009     Galaxie  Varian and the Varian logo are trademarks or registered  trademarks of Varian  Inc  in the U S  and other countries        2009 Varian Inc     Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       Request for In
8. des information on the laboratory facilities  required to support the operation of the LC system  These include   e Electrical power supply  e Gas supplies  e Waste fluid container  e Network requirements  e Guidelines for software installation    3 1 Electrical Power Supplies    The installation of electrical power supplies must comply with the  rules and or regulations imposed by the local authorities  responsible for the use of electrical energy in the workplace     All power supplies for the Varian modular liquid chromatograph and  its accessories should be single phase  AC  3 wire systems  active   neutral  ground  or two active and ground   Each connection should  be terminated at an appropriate receptacle within reach of each  assembly   s power cable  Use of power boards or extension cables  is not recommended     Avoid using power supplies from a source that may be subject to  electrical interference from other services  such as large electric  motors  elevators  welders and air conditioning units      Varian modular liquid chromatographs are supplied with a 2 meter  long  6 feet and 6 in   mains supply cable     Publication no  8510251300  Issue 1  July 2009 Page 17 of 30    Varian Liquid Chromatography Modules       3 1 1    Page 18 of 30    Electrical Requirements  Table 3  System electrical specifications    System unit       210 218 pump   SD 1 pump   SD 2 pump   325 UV Vis detector    335 Photodiode Array detector    380 385 LC ELS detector  356 LC RI detector 
9. e AM Radeon AGP   e ATI Mobility 128 AGP   e ATI Rage Mobility 128 AGP   e ATI mobility M3   e ATI3D Rage PRO AGP 2X   e ATI Rage 128 Pro Ultra GL AGP   e ATI Rage XL PCI  B41        Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules          e ATI Rage 128 Ultra  16MB   e Neomagic Magicgraph 128XD  e NeoMagic MagicMedia256ZX  e Mga mil   chip  MGA 2064WB4R2  CNA  TI TVP3026 220MHZ   e Nvidia Riva TNT  with last driver version  chip  NV4  memory  16Mo   Files  nv4_mini sys  nv4_disp dll  Version   4 00 4 0 0   e NVIDIA GeForce4 MX 420  64MB DDR  e NVidia GEForce4 440 MX  e Matrox G200 AGP  List of known non compatible graphic cards   e Intel815 Video Accelerator  4Mo    e S3 Savage graphics adapter works only with 256 colours  e ATI Radeon X1300 pci     Recommended Server  Redundancy  e Redundant Power Supply  e Redundant Network Card  e Mirroring System Disk  RAID 1   e RAID 1 or RAID 5 Storage Disks  e    Hot Plug    Hard Drives  e Backup Unit  DLT type with automated backup software     Client or Standalone system  Windows   XP Pro  SP1or SP2   or Windows  Vista    Enterprise and Business    Server   Windows   2000 server or Windows   Server 2003  with the applicable  Service Packs    NTFS Partitions only   Server must be either part of a Domain or be a Domain Server            For Galaxie    CDS server specifications  please contact your  local Varian sales representative     Varian recommends and supplies Dell    PCs  For a complete li
10. e information required to successfully  prepare a site for a Varian modular liquid chromatograph  installation     On completion of the site preparation  fill in the check list on page iii  and send this pre installation check list to your local Varian agent  or Varian sales and service office  As soon as it is received  Varian  or your Varian agent will contact you to arrange a convenient time  for installation     If you have difficulty in preparing for the installation  and for details  of operator training courses  please contact your Varian sales or  field support representative     1 1 Installation Guidelines    Allow a minimum of 1 day for the installation of the Varian modular  liquid chromatograph by a Varian field support representative     The installation will include the following    e Instrument installation   e instrument software installation and registration  e Accessory installation   e Basic customer training    e Maintenance overview    Publication no  8510251300  Issue 1  July 2009 Page 9 of 30    Varian Liquid Chromatography Modules       This page is intentionally left blank    Page 10 of 30 Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       2  Laboratory Environment    This section includes information on laboratory requirements for the  Varian modular liquid chromatograph including    e Suitability   e Environmental conditions   e Workbench   e PC requirements    2 1 Suitability    The instrument is suitable o
11. ed spares with your  instrument     Warning and Caution Messages    A Warning message is used in the text when failure to observe  instructions or precautions could result in death or injury  The list of  symbols that appear in conjunction with warnings are detailed in the  next section     Hot surfaces    A Caution message is used when failure to observe instructions  could result in damage to equipment  Varian supplied and or other  associated equipment      The following symbol may be used on warning labels attached to  the instrument  When you see this symbol  refer to the relevant  operation or service manual for the correct procedure referred to by  that warning label     Publication no  8510251300  Issue 1  July 2009 Page 5 of 30    Varian Liquid Chromatography Modules       Warning Symbols    The following is a list of symbols that appear in conjunction with  warnings in this manual or on the liquid chromatograph  The hazard  they describe is also shown     A triangular symbol indicates a warning  The meanings of the  symbols that may appear alongside warnings in the documentation  or on the instrument itself are as follows      gt    gt    gt     Broken glass     gt     Fire hazard     gt     Moving parts     gt     FF radiation    Page 6 of 30    Corrosive liquid Electrical shock     gt    gt     Noxious gas Heavy weight   danger to hands      gt     T  Toxic hazard UV radiation   Eye hazard     gt    gt     Heavy weight Explosion hazard   danger to feet     Publica
12. iquid  pressures can be generated by the pump causing a high velocity  stream of volatile and or toxic liquids     e Wear face protection in accordance with local safety  regulations when you inject samples or perform routine  maintenance    e Never open a solvent line or valve under pressure  Stop the  pump first and let the pressure drop to zero    e Use shatter proof reservoirs capable of operating at  50 60 psi  3 4 bar      e Read and adhere to all Notes  Cautions  and Warnings in the  manual     Publication no  8510251300  Issue 1  July 2009 Page 3 of 30    Varian Liquid Chromatography Modules       Page 4 of 30    Ultraviolet Radiation    Varian Liquid Chromatography Modules that use an ultraviolet light  source have shielding to prevent radiation exposure to personnel   For continued protection     e Ensure that protective lamp covers of detectors are in place  during operation     e Do not look directly into detector fluid cells or at the UV light  source  When inspecting the light source or fluid cell  always  use appropriate protective eye covering     e Ozone can be generated by radiation from the source lamps   Exposure to ozone can result in severe irritation to the skin   eyes  and upper respiratory system  The maximum  permissible exposure level is 0 1 ppm  0 2 mg m   Always  ventilate the area surrounding the instrument such that the  concentration of ozone does not exceed the maximum  permissible level  All venting must be to outside air  never  within the b
13. lication no  8510251300  Issue 1  July 2009    Varian Liquid Chromatography Modules       vi    3 2    3 3  3 4  3 5    4 1  4 2  4 3  4 4    5 1  5 2    3 1 3 Fuses   Gas Supplies   3 2 1 Storage Cylinder Instructions  Waste Fluid Container   Network Requirements   Guidelines for Software Installation  Instrument Shipping Information  Insurance After Delivery   Weights and Dimensions   Transit Damage   Unpacking the Instrument  Materials and Solvents   Materials   Solvents    Operator Familiarization    19  20  20  21  21  22  23  23  23  25  26  27  27  27  29    Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       Safety Practices and Hazards       Your Varian instrument and accessories have been carefully  designed so that when used properly  you have an accurate  fast   flexible and safe scientific instrument     If the equipment is used in a manner not specified by the  manufacturer  the protection provided by the equipment may be  impaired     Information about safety practices appears throughout the  documentation  both hard copy and online  provided with your  instrument and accessories  Before using the instrument or  accessories  you must thoroughly read these safety practices     Observe all relevant safety practices at all times     General    Operation of a Varian liquid chromatography module involves the  use of high pressure liquids  UV light  compressed gas cylinders  and hazardous materials including corrosive fluids a
14. nd flammable  liquids  Careless  improper or unskilled use of this liquid  chromatograph can cause death or serious injury to personnel   and or severe damage to equipment and property     The Varian liquid chromatography module incorporates interlocks  and covers that are designed to prevent inadvertent contact with  any potential hazards  If the instrument is used in any manner not  specified by Varian  this protection may be impaired  It is good  practice to develop safe working habits that do not depend upon  the correct operation of the interlocks for safe operation  It is  essential that no interlock or cover is bypassed  damaged or  removed     The safety practices described below are provided to help the user  operate the instrument safely  Read each safety topic thoroughly  before attempting to operate the instrument and ALWAYS operate  the liquid chromatograph in accordance with these safety practices     Publication no  8510251300  Issue 1  July 2009 Page 1 of 30    Varian Liquid Chromatography Modules       Electrical Hazards    Exposure to high voltages and UV energy can cause severe skin  damage and cataracts of the eyes  while close contact with the  electrical components can result in severe heat burns to the skin   and an electrical discharge which may cause death  severe electric  shock or sub surface skin burns     The Varian liquid chromatography module has been carefully  designed to operate safely and effectively when using components  that conform to Va
15. necessary to correct the interference  at his or her expense     Publication no  8510251300  Issue 1  July 2009 Page 7 of 30    Varian Liquid Chromatography Modules       CE Compliance    Your Varian modular liquid chromatograph has been designed to  comply with the requirements of the Electromagnetic Compatibility   EMC  Directive and the Low Voltage  electrical safety  Directive   commonly referred to as the LVD  of the European Union  Varian  has confirmed that each product complies with the relevant  Directives by testing a prototype against the prescribed EN   European Norm  standards     Proof that a product complies with these directives is indicated by     e the CE Marking appearing on the rear of the product  and    e the documentation package that accompanies the product  containing a copy of the Declaration of Conformity  The  Declaration of Conformity is the legal declaration by Varian  that the product complies with the directives listed above   and shows the EN standards to which the product was tested  to demonstrate compliance  It is also signed by Varian   s  Authorized Representative in the EU  and by the  representative of the manufacturing plant    After all safety regulations have been met  check the checklist box   Principal installation area is in compliance with all relevant safety  regulations     Page 8 of 30 Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       1  Introduction       This publication contains th
16. net mask and gateway if using a fixed IP  for a Varian 325 335 detector     After the appropriate network and work station access are available   check the checklist box  Appropriate network and work station access     Publication no  8510251300  Issue 1  July 2009 Page 21 of 30    Varian Liquid Chromatography Modules       3 5 Guidelines for Software Installation    Varian recommends that you purchase a PC with the Varian  liquid chromatography modules  The supplied PC will come with  the appropriate operating system as loaded by the PC supplier     For instructions on installing the Windows   operating system   please refer to the appropriate manuals supplied with the software   It is the responsibility of the customer to ensure that the operating  system software has been installed and is functional  if the  customer is supplying the PC or the operating system     Note Varian will not assume responsibility for loss of data     The Varian liquid chromatography module requires a version of  Galaxie    software in order to be operated     A Varian field support representative will have advised you what  version of the Galaxie software is required and what options you  have in terms of upgrading     The Varian field support representative will connect the PC to the  HPLC system and any factory approved accessories purchased at  installation  Initial instrument software installation is also included  as part of the system installation  For information on installing the  Galaxie
17. nly for indoor use and is classified  suitable under the following categories  EN 61010 1     e Installation category II   e Pollution degree 2   e Safety class 1  EN 61010 1     Table 1  Suitable environmental conditions for the Varian Modular HPLC  instruments       Condition Altitude Temperature   Humidity   RH   non condensing    Non operating 0 3050 m 5 45   C   Transport   0 10000 ft   41 113   F     Non operating and Sea level 5 45   C  meeting dielectric  41 113   F   strength tests    Operating within 0 3050 m 10 35   C  specifications  0 10000 ft   50 95   F           Publication no  8510251300  Issue 1  July 2009 Page 11 of 30    Varian Liquid Chromatography Modules       2 2    2 2 1    2 2 2    Page 12 of 30    Environmental Conditions    M  M    The area selected for the operation of a Varian modular liquid  chromatograph must be free from drafts  corrosive  atmospheres and vibration  Sample preparation areas and  materials storage facilities should be located in a separate room     The area should be a dust free  low humidity environment  Air  conditioning is strongly recommended for control of the  environment  For optimum analytical performance it is  recommended that the ambient temperature of the laboratory be  between 20 and 25   C  68 and 77   F      After the humidity requirements have been met  check the checklist box   Relative humidity maintained between  8 and 80   RH     After the temperature requirements have been met  check the checklist  box  Lab
18. odules       2 4    Page 14 of 30    PC Requirements    The recommended configuration represents the minimum on which  you can run the software  This PC configuration may be out of  manufacture  or you may want to use a PC you already have     Table 2  Varian modular liquid chromatograph PC requirements       Recommended       Intel Processor minimum 2 8GHz  2 2 GHz for Dual core processors    If the computer will have National Instruments GPIB card s  installed  the  processor should not be a multiprocessor and hyper threading must be  disabled  Intel Core 2Duo Processor is supported     Client or Standalone system  gt  1 GB recommended with Windows XP SP2 or  Windows Vista   operating system   gt   2 GB RAM required Server   gt 2 GB  recommended        Client or Standalone system  gt  10GB recommended Server   gt 160 GB  recommended  Details to be defined with customer according to the yearly  amount of data generated  and the desired retention of data    on line     i e   without off line archiving  NTFS Partitions only        100 1000BaseT Ethernet NIC       Slots available and appropriate for any required cards for instrument interfaces   GPIB     CD ROM drive  gt 16x required  DVD ROM drive recommended  Windows    compatible keyboard and mouse    Minimum 1024x768 screen resolution  1280x1024 screen resolution  recommended           For the use of 335 LC Photo Diode Array detector features  the graphic card  should be in the list below    e INTEL integrated video card   
19. on  using your own instrument  To  check this possibility  please contact your local Varian Sales and  Service Office     When waiting for tests or instrument warm up to complete  the  Varian field support representative will demonstrate some of the  basic system operating procedures     If your system computer was purchased from Varian  it will be  configured  formatted  partitioned and loaded with its operating  system  The Galaxie    Chromatography software will be loaded  during installation     Please note that you must have a working knowledge of the  computer operating system  as this type of instruction is not  provided by Varian  The Online Help supplied with the HPLC  system provides step by step instructions for setting up the system  and detailed operating instructions for the analysis procedures  It  does not include instructions for operation of the computer     Tl After the principal operator has been identified and will be available  during the installation  check the checklist box  Principal operator will  be available during the installation and certification period     Publication no  8510251300  Issue 1  July 2009 Page 29 of 30    Varian Liquid Chromatography Modules       This page is intentionally left blank     Page 30 of 30 Publication no  8510253100  Issue 1  July 2009    
20. or  etc                        Publication no  8510253100  Issue 1  July 2009       Varian Liquid Chromatography Modules          Warning   AN The Varian liquid chromatograph module can weighs between  11 34 kg  25 75 Ib   To avoid injury to personnel or damage to  equipment  always use suitable lifting procedures when moving  the modules         Vv  After the bench space requirements have been met  check the  checklist box  Sufficient bench space is available for all components      Vv  After the bench support requirements have been met  check the  checklist box  Bench can support system weight     4 3 Transit Damage    Transit damage will only be admitted by the carrier if it is reported  as agreed in the terms of his agreement  For any claims against  damage in transit  the following general rules apply     e Before accepting delivery  you should inspect the packages  for signs of obvious damage  The nature of any obvious  damage must be noted on the carrier   s waybill  and then  must be countersigned by a representative of the carrier     e Within the time limits stated in the terms of conditions of  carriage  a further inspection must be made for concealed  damage  If any damage is found at this stage  the carrier  must be notified in writing and all packaging material must be  retained for subsequent inspection by a representative of the  carrier     e A copy of any damage report must be forwarded to the  Varian sales office dealing with the supply of the equipment    
21. rian   s design criteria  Use of non approved  components may render the system inoperative and or hazardous   It may also invalidate the warranty on the instrument  Use only  related components supplied or authorized by Varian     Page 2 of 30    Disconnect the instrument from all power sources before  removing protective panels to avoid exposure to potentially  dangerous voltages  Panels or covers which are retained by  screws may be opened ONLY by Varian qualified  representatives  Consult the manuals or product labels  supplied with your PC to determine which parts are  operator accessible     When it is necessary to use a non original power cord plug   make sure the replacement cord adheres to the color coding  and polarity described in the manual and all local building  safety codes     Good grounding earthing is essential to avoid a potentially  serious electric shock hazard  Ensure that there is a 3 pin  earth grounded receptacle  Consult the manuals or product  labels supplied with your PC for the relevant grounding  requirements    Replace blown fuses with fuses of the size and rating shown  on the fuse panel or in the manual    Replace faulty or frayed power cords immediately with the  same type and rating    Make sure that voltage sources and line voltage match the  value for which the instrument is wired     Avoid using power supplies from a source that may be  subject to electrical or RF interface from other services  for  example  large electrical motors  eleva
22. sentative for advice     A convenient 4 mm push in connection for the gas source is  provided at the rear of the ELS detector     Storage Cylinder Instructions    Note    Cylinders containing gas under pressure should be firmly secured  to a rigid structure  and the storage area must be adequately  ventilated     Never locate gas cylinders near a source of ignition  or in a position  that is subject to direct heat  Gas storage cylinders often  incorporate a pressure release device which will discharge the gas  at a predetermined temperature  usually around 52   C  125   F      If gases are to be plumbed from a remote storage area to the  instrument site  ensure that the local outlets are fitted with stop  valves  pressure gauges and suitable regulators which are easily  accessible to the instrument operator  The gas outlets must be  provided within 1 5 meters  5 feet  of the instrument     You are responsible for installation of gas supplies or gas cylinders   Varian authorized personnel will not be involved in the setting up or  installation of any gas supplies other than the connections into the  liquid chromatograph      Vv  After the gas requirements have been met  check the checklist box  Gas    supply  at specified purity   regulator  and gas lines are installed     Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       3 3 Waste Fluid Container    Operation of the Varian modular liquid chromatograph requires the  use of a waste con
23. st of  qualified PCs and printers  please contact your local Varian  representative     PCs supplied with Letter of Credit orders will be an international  brand and will be the    Recommended    configuration or better     Higher rated PC components can be substituted for those listed  above e g   processor type  amount of memory  screen size and  resolution  operating system version  etc     Publication no  8510251300  Issue 1  July 2009 Page 15 of 30    Varian Liquid Chromatography Modules       Additionally you need to have a valid copy of Galaxie in order to run  the instrument  This may need to be ordered  Please contact your  local Varian representative if you need more information     2 4 1 Recommended Computer Systems  Varian uses and recommends Dell PCs     To ensure the best performance and reliability when using Varian  software to run your analytical instruments  we recommend using  Dell PCs     Although our software is designed to work on any correctly  configured PC  all testing and evaluation is completed using Dell  PCs only     Details of recommended PC configurations for Varian products can  be found on the Varian Inc  website  www varianinc com     Tl After the PC requirements have been met  check the checklist box   Personal computer with Microsoft Windows XP   or Vista operating  system installed     Page 16 of 30 Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       3  Laboratory Facilities       This section inclu
24. stallation      understand that if the installation site is not prepared in accordance with the enclosed  instructions  additional installation charges may apply     Company name  Name   Company address  Position    Telephone    Preferred installation date   Signed  Date     Pre Installation Checklist    Your site must meet all requirements before you request installation  Do not unpack the boxes   After completing each requirement  place a check in the appropriate checkbox  Fax or mail this  pre installation checklist to your local Varian office  As soon as it is received  a Varian  representative will contact you to arrange a convenient time for installation     Requirements       Principal installation area is in compliance with all relevant safety regulations  Relative humidity maintained between 8     80  RH       Lab temperature maintained between 10 and 35   C  50 and 95   F        Laboratory is free of excessive particulate matter       System bench is free from vibrations       Personal computer with Microsoft Windows   XP or Vista operating system installed       Specified electrical supply and power outlets installed       Gas supply  at specified purity   regulator  and gas lines are installed  if required        IP address  subnet mask and gateway  if fixed IP for a Varian 325  335 detector       Appropriate network and workstation access       Sufficient bench space is available for all components       Bench can support system weight       All unopened boxes ha
25. support representative will prepare solutions to  tune and evaluate the Varian liquid chromatography module   Please supply the following materials and solvents     Warning     Chemical Hazard     Noxious Fumes   Danger of skin irritation  burns and inhalation  To avoid injury to  personnel or damage to equipment  always follow appropriate  safety guidelines when using chemicals and always wear    3 appropriate safety equipment and clothing     5 1 Materials  Table 6  Materials       Quantity  Item       2 1 L clean and new Wheaton bottles to use as mobile phase  reservoirs                5 2 Solvents    Table 7  Solvents       Quantity Solvent       2 liters HPLC grade water       Additional tubing is required to complete the installation of your  system  Most of this tubing is included in the accessory kits of LC  modules  Additional tubing  PEEK    or stainless steel  may be  required for installation of special valves or modules  For most  analytical HPLC systems  flow rates up to 1 mL min   1 16 in     Publication no  8510251300  Issue 1  July 2009 Page 27 of 30    Varian Liquid Chromatography Modules       tubing with an internal diameter  ID  of 0 005 in  can be used  throughout  However  use 0 005 in  ID or smaller tubing  downstream of the sample injector or autosampler to prevent peak  broadening  Keep tubing lengths as short as possible to prevent  peak broadening and minimize run time     Semi prep and preparative systems have other interconnect tubing  requiremen
26. tainer for the disposal of excess fluids  Suitable  tubing is supplied with the HPLC system for use with most solvents     A chemically inert container  appropriately sized to hold waste  coming from the instrument must be provided by the instrument  user  It should be located underneath the workbench where it is  protected by the bench and in full view of the operator     3 4 Network Requirements    The Varian 325 or 335 detectors require an Ethernet connection to  the PC via a shielded twisted pair Ethernet crossover cable  Category 5 or better  provided with the detector      If connection is required to an Ethernet network  then a shielded  twisted pair Ethernet non crossover cable Category 5 or better will  be required  not supplied      Most PC   s come pre configured with an Ethernet connection either  built into the motherboard or with an Ethernet network card  installed  Check that the intended PC has this configuration and if  not  a network card will have to be fitted  Refer to the installation  chapter for details     If the user is supplying their own PC  then they are responsible for  installing and configuring the card  They are also responsible for  setting up and maintaining any LAN configuration where a detector  may be used     All network issues are to be dealt with by the users     After the network details have been obtained  IP address  subnet mask  and gateway if using a fixed IP for a Varian 325 335 detector   check the  checklist box   P address  sub
27. tion no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       The following symbols may appear on the instrument for your  information        l Mains power on  0 Mains power off    f  Fuse  YJ Single phase alternating current  Direct current    When attached to the rear of the instrument  indicates  that the product complies with the requirements of  one or more EU directives     When attached to the rear of the product  indicates  SP  that the product has been certified  evaluated  to CSA  1010 1 and UL 3101 1     e Indicates high voltage xenon flash lamp present              The symbols above may be used on warning labels attached to the  instrument  When you see these symbols  refer to the relevant  operation or service manual for the correct procedure referred to by  that warning label     US FCC Advisory Statement    This equipment generates  uses and can radiate radio frequency  energy  and if not installed and operated in accordance with the  instruction manual may cause interference to radio  communications  It has been tested and found to comply with the  limits for a Class A computing device pursuant to Subpart J of Part  15 of Federal Communications Commission  FCC  Rules  which  are designed to provide reasonable protection against such  interference when operated in a commercial environment   Operation of this equipment in a residential area may cause  interference  in which case the user will be required to take  whatever measures may be 
28. tors and welders      Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       Compressed Gas Cylinders    Unless your system is equipped with a helium sparging unit or an  ELSD  there are no gas requirements for Varian liquid  chromatography modules  For gas requirements on these other  devices  consult the operation manual or other documentation  enclosed with the particular device  Compressed gas cylinders  contain highly pressurized gas  If storage conditions are outside of  the recommended suppliers safety codes the cylinders can explode  or rapidly release gas into the environment  This may result in injury  or death     e Store and handle compressed gases carefully and in strict  adherence to safety codes   e Secure cylinders to an immovable structure or wall     e Store and move cylinders in an upright  vertical position   Before transport  remove regulators and install cylinder cap    e Store cylinders in a well ventilated area away from heat   direct sunshine  freezing temperatures  and ignition sources    e Mark cylinders clearly so there is no doubt as to their  contents    e Use only approved regulators and connections    e Use only connector tubing that is chromatographically clean  and has a pressure rating significantly greater than the  highest outlet pressure from the regulator     High Pressure Hazards    If a line ruptures  a relief device opens  or a valve opens  accidentally under pressure  potentially hazardous high l
29. ts and limitations  Typically  a system that uses flow  rates up to 200 mL min requires 1 16 in  OD tubing with an internal  diameter as large as 0 040 in  Use tubing  0 010 in  smaller than  this in the rest of the system  Usually preparative systems use   1 8 in  tubing  0 062 in  ID  throughout  Always consult the module  Operator s Manual for specific recommendations     After the materials and solvent requirements have been met  check the  checklist box  Materials and solvents are on site     Page 28 of 30 Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       6  Operator Familiarization       The Varian liquid chromatography module is installed by a Varian  field support representative  During installation  the Varian field  support representative will demonstrate the basic operating  procedures while conducting the installation performance tests  during the installation procedure  The Varian field support  representative however  is not necessarily experienced in complex  analytical routines and is not authorized to conduct extensive  training     To ensure that your operators benefit the most  operator training  should be completed before your equipment is installed  It is  strongly recommended that you take advantage of the special  training courses that are conducted at various locations by the  Varian sales organization     In some areas it may be possible to arrange for operator training to  be carried out after the installati
30. ugh to support the total weight of the equipment to be  placed on top of the workbench  The bench top should be large  enough to permit free circulation of air through the main instrument  and around each of the accessories     The information provided in the weights and dimensions table will  help make planning easier  Portable or semi permanent trolleys  must not be used as work benches for the HPLC system     To avoid damage through spillage of samples being used  the  instrument bench top should be covered with a material that is  resistant to common organic solvents used in HPLC and  impervious to liquid spillage  For comfortable working conditions  and easy access to the instrument sample introduction system   Varian recommends that the height of the workbench be  approximately 900 mm  36 in   high     After the workbench vibrational requirements have been met  check the  checklist box  System bench is free from vibrations     2 3 1 Workbench Location    The workbench location should permit service access from all  sides  A minimum of 400 mm  16 in   free space at the sides of the  HPLC system and approximately 750 mm  30 in   at the rear of the  instrument is required for maintenance and service access     The Varian modular liquid chromatograph should not be located  close to an access door  window or any other area where drafts  may cause unstable temperature conditions     Publication no  8510251300  Issue 1  July 2009 Page 13 of 30    Varian Liquid Chromatography M
31. uilding     Other Precautions    Use of the Varian liquid chromatography modules and accessories  may involve materials  solvents and solutions which are flammable   corrosive  toxic or otherwise hazardous  Careless  improper or  unskilled use of such materials  solvents and solutions can create  explosion hazards  fire hazards  toxicity and other hazards that can  result in death  serious personal injury or damage to equipment     Always ensure that laboratory safety practices governing the use   handling and disposal of hazardous materials are strictly observed   These safety practices should include wearing appropriate safety  clothing and safety glasses     Air flow to the cooling fans of the Varian liquid chromatography  modules and accessories must be unobstructed  Do not block the  ventilation grills on the Varian liquid chromatography modules and  accessories  Consult the manuals supplied with your PC  monitor  and other peripherals for their specific ventilation requirements     Great care should be taken when working with glass or quartz parts  to prevent breakage and cuts  This is especially important when  removing or replacing lamps     Each Varian liquid chromatography module weighs between  11 34 kg  25 75 Ib   To avoid injury to personnel  or damage to the  instrument or property  always use suitable lifting procedures when    Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       moving the modules  Use only Varian suppli
32. ve been checked for damage       Instrument has been placed near its final location       Materials and solvents are on site  waste container  LC grade water        Principal operator will be available during the installation and certification period       Have any additional criteria been agreed within the contract  If yes  please specify  s   No    O  lt  O  O 0 0 0 0 0  0 0 0 0  0  0  0  0 0 8       Is there any additional equipment that needs to be connected to the system  If yes    Yes   No  please specify  Oo O                   Publication no  8510251300  Issue 1  July 2009 iii    Varian Liquid Chromatography Modules       This page is intentionally left blank    iv Publication no  8510253100  Issue 1  July 2009    Varian Liquid Chromatography Modules       Contents       Safety Practices and Hazards 1  General 1  Electrical Hazards 2  Compressed Gas Cylinders 3  High Pressure Hazards 3  Ultraviolet Radiation 4  Other Precautions 4  Warning and Caution Messages 5  Warning Symbols 6  US FCC Advisory Statement 7  CE Compliance 8  1  Introduction 9  1 1 Installation Guidelines 9  2  Laboratory Environment 11  2 1 Suitability 11  2 2 Environmental Conditions 12   2 2 1 Cleanliness 12   2 2 2 Instrument Cooling Air Supply 12  2 3 Workbench 13   2 3 1 Workbench Location 13  2 4 PC Requirements 14   2 4 1 Recommended Computer Systems 16   Laboratory Facilities 17  3 1 Electrical Power Supplies 17   3 1 1 Electrical Requirements 18   3 1 2 LC System Power Connections 19    Pub
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
User Manuals - CNET Content Solutions  Rentrée sportive 2012 chèque sport region bretagne  PDF 289kB  La chanson de la semaine : Le déserteur de Boris Vian  Turbo-V 250 PCB Controller Description  Elektronisches Schulwörterbuch - Franklin Electronic Publishers, Inc.  LG 22LN4510 Energy Guide : Free Download, Borrow, and Streaming : Internet Archive    TrueGrid ® Brochure  Merchant/400TM    Copyright © All rights reserved. 