Aspect Medical Systems A-2000 XP PLATFORM
Contents
1. ACTIONS SECTION IF ANY FAILURES ARE NOTED SEE MESSAGE AND OPERATOR 1 Connect Sensor Simulator or Test Sensor to PIC cable 2 Connect PIC cable to DSC pigtail e Verify that the Sensor Check begins and each element displays a number then says PASS Expected Impedance Values Sensor Simulator Values Test Sensor Values Electrode 8 Typical Range Electrode Typical Range 1 5 K Ohms 4 6 K Ohms 1 1 K Ohms 1 2 K Ohms 2 10 K Ohms 8 17 K Ohms 2 1 K Ohms 1 3 K Ohms 4 4 K Ohms 3 5 K Ohms 4 1 K Ohms 1 2 K Ohms 3 3 K Ohms 2 4 K Ohms 3 1 K Ohms 1 2 K Ohms 3 Press Menu Exit then highlight and select Sensor Check As Sensor Check executes this time flex the cables and connectors from the DSC up to the sensor to check for intermittent connections or conductors e Verify that the Sensor Check begins and each element displays a number then says PASS Expected Impedance Values Sensor Simulator Values Test Sensor Values Electrode 8 Typical Range Electrode Typical Range 1 5 K Ohms 4 6 K Ohms 1 1 K Ohms 1 2 K Ohms 2 10 K Ohms 8 17 K Ohms 2 1 K Ohms 1 3 K Ohms 4 4 K Ohms 3 5 K Ohms 4 1 K Ohms 1 2 K Ohms 3 3 K Ohms 2 4 K Ohms 3 1 K Ohms 1 2 K Ohms 4 End of procedure 6 5 075 0014 SECTION DIAGNOSTICS AND TROUBLESHOOTING 6 2 4 DSC Cable Problem Isolation Note This A 2000 Service Information Manual contains the maintenance
2. COMM Monitor Figure 3 2 The A 2000 Signal Flow Diagram After passing through input protection circuits in the DSC the EEG signals are differentially amplified and filtered to remove DC and high frequency components The signals are digitized by separate one bit sigma delta analog to digital converters The outputs from the converters are multiplexed onto the DSC communications line and de multiplexed in the monitor prior to being fed to the Digital Signal Processor DSP The DSP filters the signals and computes the processed variables The results are passed to the host CPU for display 3 3 075 0014 SECTION III PRINCIPLES OF OPERATION 3 2 1 The Digital Signal Converter DSC The DSC contains the inputs amplifiers and digitizers for two channels of EEG It has a single point connection that connects via a Patient Interface Cable PIC to a BIS Sensor The sensor and PIC contain circuits for identifying them to the monitor This permits the monitor to configure automatically The DSC constantly monitors the combined source impedance at its inputs and also has facility for measuring the individual impedance of the channel and ground electrodes There are circuits for injecting self test voltages into the amplifier inputs 3 2 1 1 The DSC Preamplifier Circuits The preamplifier circuits contain the input protection circuits montage mode switches differential amplifiers filter and gain amplifiers impedance test circuits self te
3. 5 Monitor 10 1 075 0014 SECTION X APPENDIXII 10 2 A 2000 BLOCK DIAGRAM PREAMP BOARD Patient Power Printer Connector Port J3 J27 INTERCONNECT BOARD POWER SUPPLY BOARD MAIN PROCESSOR BOARD BATTERY KEYBOARD These connectors are internal to the board and are mated as shown DISPLAY MONITOR 10 2 075 0014 SECTION X APPENDIXII 10 3 DSC WIRING DIAGRAM DSC CONNECTOR wire side Pin Wire Color Signal Name Pin 1 RED HBCTLH Pin 2 GREEN HBCTLL Pin 3 YELLOW HBIN Pin 4 BLACK GND 5 TEFLON CHAS GND 10 3 075 0014 SECTION X 10 4 WIRING DIAGRAM MONITOR CABLE 075 0014 LEMO Pin Wire Color Stocko Pin Red Green Yellow Black OA 0 N gt Drain Teflon O14 c0 NI gt 10 4 APPENDIXII PIN 1 SECTION X 10 5 WIRING DIAGRAM PIGTAIL CABLE 075 0014 APPENDIXII Stocko Connector Wire Color Pigtail Connector Drain Teflon 10 Brown Red Orange Yellow Green ABONO N C Blue Purple Gray White N C z lt 01 41 0 Contact Information for Aspect Medical Systems Inc 141 Needham St Newton MA 02464 U S A Main Business Phone 617 559 7000 Main Business Fax 617 559 7400 Customer Service 888
4. THE SYSTEM LEAKAGE CURRENT MUST BE CHECKED AND MUST BE LESS THAN THE IEC601 1 1 LIMIT NOTE When software is upgraded all previously recorded data and monitor configuration settings will be lost Therefore configuration settings should be recorded before the software update is performed 3 3 6 Printed Reports A printer port connector allows connection to the Aspect thermal printer 3 9 075 0014 SECTION III PRINCIPLES OF OPERATION 3 10 075 0014 SECTION IV PREPARATION FOR USE AND INSTALLATION SECTION IV 4 PREPARATION FOR USE AND INSTALLATION INTRODUCTION This section provides an overview of installation information for service personnel working with the Aspect A 2000 BIS Monitor System Please see the A 2000 BIS Monitoring System Operating Manual for full installation instructions e Environment e Instrument connections e Installation and verification procedure 4 1 ENVIRONMENT 4 1 1 Shipping and Storage Environment The monitor and its accessories can be stored or shipped within the following environmental limits Note that these limits apply to non operational storage and shipping situations Temperature 20 C to 60 C Humidity 15 96 to 95 non condensing Pressure 360 mmHg to 800 mmHg Protect the monitor from sudden temperature changes that can lead to condensation within the instrument To minimize condensation avoid moving the system between heated buildings and outside storage Once moved inside all
5. e Plug devices into separate outlet circuit branches Consult Aspect Medical Systems Technical Support for assistance All work involving opening the instrument case must be performed in a static safe environment to prevent damage to electronic components and assemblies This environment includes the operator work area and tools and any other test or storage items that might touch the DSC assemblies Use only the parts and tools specified Use of any others may damage the instrument Handle PCBs only by board edges to avoid contamination moisture skin oils etc on PCB and component surfaces When connecting or disconnecting Digital Signal Converter take care not to touch the exposed contacts of either connector Damage due to electrostatic discharge may result A 2000 is a trademark of Aspect Medical Systems Inc Bispectral Index BIS the BIS logo and Zipprep are trademarks of Aspect Medical Systems Inc and are registered in the U S A E U and other countries 1 4 075 0014 SECTION 13 KEY TOSYMBOLS c c N gt 075 0014 SAFETY PRECAUTIONS ON power connection to the mains OFF power disconnection from the mains Audible Alarm Silenced Operating on Battery Type BF Equipment Type BF Equipment Defibrillator proof Attention Consult Accompanying Documents Attention J1 RS 232 Serial Port Consult Accompanying Documents Attention J2 Printer Port Consult Acc
6. Aspect will bear the costs of shipping the repaired or replacement product back to the user All other shipping costs shall be paid by the customer 7 4 REPACKAGING FOR SHIPPING AND STORAGE If it becomes necessary to return the monitor to the factory use the original shipping container to protect the product Seal the package with reinforced packing tape rather than plastic or masking tape Mark shipping container FRAGILE 7 3 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM 7 5 DISASSEMBLING AND REASSEMBLING THE DSC This section provides instructions for removing and replacing parts of the Digital Signal Converter DSC It describes e Required tools and supplies e Opening and closing the DSC case e Removing and installing DSC cables monitor and pigtail e DSC Checkout and safety tests Note that a fully functional A 2000 BIS Monitoring System and accessories PIC cable Sensor Simulator or Test Sensor and shorting bar must be available to test any DSC that has been opened Do not attempt these repairs unless all required equipment and tools are available WARNING ANY PROCEDURES THAT REQUIRE THE REMOVAL OF THE DSC s COVER AND INTERNAL PARTS SHOULD BE PERFORMED BY A QUALIFIED BIOMEDICAL ENGINEERING TECHNICIAN POWER OFF THE UNIT BY DISCONNECTING FROM THE MONITOR WARNING GROUND WIRE LEAKAGE CURRENT MUST BE CHECKED BY A QUALIFIED BIOMEDICAL ENGINEERING TECHNICIAN WHENEVER THE INSTRUMENT CASE IS OPENED FOR INSPECT
7. BIS INDE X press 6 OR 800 442 7688 press 6 Technical Service 800 442 2051 Email bis_info aspectms com Web www aspectmedical com Aspect Medical Systems International B V Rijnzathe 7d2 3454 PV De Meern The Netherlands Main Business Phone 31 30 662 9140 Main Business Fax 31 30 662 9150 Email amsint aspectms com
8. Defective monitor Defective DSC 2 Defective DSC cable 1 Defective DSC 2 Defective monitor 1 Electro cautery equipment used during self test 2 Defective DSC 3 Defective monitor 1 Incorrect DSC in use 2 Defective DSC 3 Defective monitor 6 7 Corrective Actions 1 Connect DSC Verify all cable connections 2 Inspect repair cable connector end 3 Replace the DSC 4 Replace monitor 1 Replace the DSC 2 Replace the DSC 3 Replace Monitor Replace DSC 1 Replace the DSC 2 Replace the monitor Note This error requires monitor power down to reset error before each corrective action is performed 1 Isolate BIS system from possible noise sources or restart monitor when electro cautery equipment not in use Note that self test runs when the DSC is first connected 2 Replace the DSC 3 Replace monitor 1 Connect the correct DSC 2 Replace the DSC 3 Replace Monitor SECTION Status Messages E11 DSC Failed Self Test E12 Illegal Sensor 13 E15 Re prep Sensor E14 Sensor Not Connected E15 See E13 E16 Last Sensor Check Failed 075 0014 DIAGNOSTICS AND TROUBLESHOOTING Causes 1 Electro cautery equipment used during self test 2 Defective DSC 3 Defective monitor 1 Poor or contaminated connection between sensor and PIC Defective sensor Defective PIC Defective DSC PON 1 Incorrect sens
9. ELECTRODE CONNECTION THE SENSOR OR ELECTRODES SHOULD NOT BE LOCATED BETWEEN THE SURGICAL SITE AND THE ELECTRO SURGICAL UNIT RETURN ELECTRODE THE SENSOR MUST NOT BE LOCATED BETWEEN DEFIBRILLATOR PADS WHEN A DEFIBRILLATOR IS USED ON A PATIENT CONNECTED TO THE A 2000 TO MINIMIZE THE RISK OF PATIENT STRANGULATION THE PATIENT INTERFACE CABLE PIC MUST BE CAREFULLY PLACED AND SECURED WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS OCCURS RE TEST GROUND LEAKAGE CURRENT BEFORE FURTHER USE BE SURE THE MONITOR IS MOUNTED SECURELY IN PLACE TO AVOID PERSONAL INJURY UNIVERSAL PRECAUTIONS SHALL BE OBSERVED TO PREVENT CONTACT WITH BLOOD OR OTHER POTENTIALLY INFECTIOUS MATERIALS PLACE CONTAMINATED MATERIALS IN REGULATED WASTE CONTAINER DO NOT MIX DISINFECTING SOLUTIONS e g BLEACH AND AMMONIA AS HAZARDOUS GASES MAY RESULT WHEN CONNECTING EXTERNAL EQUIPMENT e g DATA CAPTURE COMPUTER THE SYSTEM LEAKAGE CURRENT MUST BE CHECKED AND MUST BE LESS THAN THE 601 1 1 LIMIT THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE EQUIVALENT SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING SYSTEM CONSIDERATION RELATING TO THE CHOICE SHALL INCLUDE e USE OF THE ACCESSORY IN THE PATIENT VICINITY e EVIDENCE THAT THE SAFETY CERTIFICATION OF THE ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE TO THE APPROPRIATE IEC 601 1 AND OR IEC 601 1 1 HARMONIZED NATIONAL STANDARD 075 0014 SECTION
10. Monitor Mounting Solutions booklet part number 070 0031 4 2 INSTRUMENT CONNECTIONS Detailed connection instructions are provided in the A 2000 Operating Manual 4 3 075 0014 SECTION IV PREPARATION FOR USE AND INSTALLATION 4 2 1 Digital Signal Converter Connections The long flexible cable from the A 2000 Digital Signal Converter connects to the electrically isolated digital signal converter connector on the front panel of the monitor Once connected the DSC need not be disconnected again However if you wish to disconnect the DSC cable from the monitor carefully pull on the blue rubber overlay DO NOT twist or pull on the gray cable 4 2 2 Power Cord Connections The 2000 is designed to use only 3 conductor hospital grade power cords Cords must be type SJE SJT or SJO Check for a firm connection 4 2 3 Printer Connector This printer port connector is designed to connect only to the Aspect A 2000 printer for screen prints To connect 1 Turn OFF the monitor 2 Ifa cover is present on the printer connector gently remove cover by prying loose one end and pulling free 3 Align the printer connectors located on the bottom of the monitor and top of the printer so that they mate correctly 4 Secure the printer to the monitor with the mounting screw located in the base of the printer Use caution not to cross thread the mounting screw 5 Turn the monitor 4 4 075 0014 SECTION IV PREPARATION F
11. Ohms 1 2 K Ohms 3 3 K Ohms 2 4 K Ohms 3 1 K Ohms 1 2 K Ohms 6 Repeat the Sensor Check Press the MENU EXIT key then press ARROW key to highlight the SENSOR CHECK line on the menu Press SELECT key to initiate the Sensor Check routine again During the Test sequence flex the cable and connections at the PIC SENSOR and PIC DSC and also the DSC box to Patient end cable Note that gentle flexing of these cables and connectors should not cause the Sensor Check to fail 5 2 6 Physical Integrity Inspection Periodically check the system A 2000 Monitor DSC PIC cable for physical damage to cases and associated cables and connectors 1 Inspect the cases of the monitor and the DSC to insure plastic is not cracked or broken 2 Inspect the gasket seal around the case joining surfaces to insure the integrity of the splash resistance seal 3 Inspect the cables and strain relief mechanisms 4 Inspect the connectors for damage faulty strain relief or contamination 075 0014 5 5 SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 6 075 0014 SECTION DIAGNOSTICS AND TROUBLESHOOTING SECTION VI 6 DIAGNOSTICS AND TROUBLESHOOTING INTRODUCTION This section explains e General troubleshooting using built in diagnostic tools e BIS Monitor System troubleshooting procedure e Status messages causes and corrective actions 6 1 GENERAL TROUBLESHOOTING The A 2000 BIS Monitor System has three built in diagnostic
12. SAFETY PRECAUTIONS REPLACE FUSES ONLY WITH ONE OF THE FOLLOWING PARTS Aspect P N 430 0006 1 25 Amps 250V 5x20mm Littelfuse 217 Series 1 25 Amps 250V 5x20mm Wickmann 193 Series 1 25 Amps 250V 5x20mm ALWAYS REPLACE BOTH FUSES TOGETHER EVEN IF ONLY ONE HAS FAILED 1 2 CAUTIONS Caution U S Federal law restricts this device to sale by or on the order of a physician Read the entire A 2000 Operating Manual carefully before using the monitor in a clinical setting Do not autoclave the Digital Signal Converter or Monitor Autoclaving will seriously damage both components Do not block fan output Keep at least three 3 inches of unobstructed space between rear of instrument and wall and other instruments Do not allow other instruments to block airflow Do not block ventilation inlet holes on the underside of monitor Do not open Digital Signal Converter The seal to prevent liquids from entering the Digital Signal Converter may be damaged if opened Service or repairs must be performed only by qualified biomedical technicians Mixing of system configurations will increase sensor and hardware inventory requirements and may create user and support issues Each BIS sensor type has its own placement method Be sure to carefully follow the appropriate package directions The A 2000 has been designed to operate with a disposable BIS Sensor The sensor is a silver silver chloride electrode array that utilizes Aspect s patented Zip
13. SOFTWARE UPDATES 3 8 3 9 0 PRINTED REPORTS ode un ey er eg ede e ha 3 9 4 PREPARATION FOR USE AND INSTALLATION 4 1 4 1 ENVIRONMENT L au uua Vau p cu iia eia uui aaa PN ER Su ERR 4 1 4 1 1 SHIPPING AND STORAGE ENVIRONMENT 0 nnne enne nennen 4 1 4 1 2 OPERATING ENVIRONMENT 9 4 1 4 1 3 POWER REQUIREMENTS AND SYSTEM GROUNDING 4 2 4 1 4 SITE PREPARATION MOUNTING THE MONITOR 4 2 4 2 INSTRUMENT CONNECTIONS u uu 4 3 4 2 1 DIGITAL SIGNAL CONVERTER CONNECTIONS 2 9999 mnn 4 4 4 2 2 POWER CORD CONNECTIONS 4 4 4 2 3 PRINTER 9999 4 4 4 3 INSTALLATION AND VERIFICATION PROCEDURE 4 5 5 PREVENTIVE MAINTENANCE CARE AND CLEANING 5 1 5 4 CARE AND CLEANING bis uuu uuu araara aeai 5 1 5 1 1 CLEANING THE MONITOR AND DIGITAL SIGNAL CONVERTER 5 1 5 1 2 DISINFECTING THE MONITOR AND DIGITAL SIGNAL CONVERTER 5 1 5 1 3 CLEANING THE MONITOR 5 1 5 2 PREVENTIVE
14. Sensor to the PIC 2 1 2 How The A 2000 Works A detailed description of how the A 2000 works is contained within Section V of the A 2000 Operating Manual and will not be discussed in this Service Manual Please refer to the A 2000 Operating Manual 070 0015 for additional information 2 2 INSTRUMENT IDENTIFICATION 2 2 1 A 2000 Monitor Monitor identification information is permanently marked on the rear panel This information includes instrument model and serial numbers power ratings cautions and the Aspect Medical Systems shipping address 2 2 2 A 2000 Digital Signal Converter The A 2000 Digital Signal Converter identification information is permanently marked on the rear panel of the Digital Signal Converter This information includes instrument model and serial numbers and cautions 2 2 3 Software Revision Numbers Software revision numbers are displayed in the Diagnostic Menu 23 PROPRIETARY INFORMATION AND DEVICES Information and descriptions contained in this guide are the property of Aspect Medical Systems and may not be copied reproduced or distributed without prior written permission Portions of the A 2000 design are proprietary and are the subject of patents and patents pending 075 0014 SECTION Il A 2000 SYSTEM OVERVIEW BIS Monitor Monitor Interface Cable Patient Interface Cable PIC Figure 2 1 The A 2000 BIS Monitoring System 2 3 075 0014 2000 SYSTEM OVERVIEW SECTION II Fan Fu
15. The following procedures should be performed periodically to ensure that the A 2000 is functioning properly 5 2 1 Checking the Battery The battery must be tested periodically to verify that the A 2000 will continue to operate during power outages To test 1 Charge the A 2000 by leaving it plugged in with the power ON for 4 hours Note The monitor must be powered ON to recharge the battery 2 Disconnect the A C cord from wall supply 3 Verify that the A 2000 operates reliably for a minimum of 20 minutes 4 Recharge the battery Caution Periodically check the battery by operating an A 2000 that has been disconnected from the wall socket and which has been charging the battery for at least four hours After long periods of storage e g gt 1 month it may be necessary to cycle discharge then charge the battery a few times to get full charge capacity If the A 2000 fails to operate reliably from the battery for twenty minutes battery replacement is required The A 2000 contains an internal Nickel Metal Hydride battery The battery must be removed by a qualified service technician and disposed of or recycled in accordance with the national laws of the country Contact Aspect Medical Systems Inc or the local distributor for servicing of battery WARNING ELECTRICAL SHOCK HAZARD DO NOT REMOVE MONITOR COVERS DURING OPERATION OR WHILE POWER IS CONNECTED TO MONITOR GROUND WIRE LEAKAGE CURRENT MUST BE CHECKED WHENEVER INS
16. a secure vertical pole 1 2 172 in diameter 1 Place pole within clamp bracket and tighten screw using the black finger knob Make sure that there is enough space above the clamp so that you have a few inches to slide the monitor in from above 2 Line up the clamp shoe on back of monitor with the slot on pole clamp and slide monitor down to fit The bottom of the clamp shoe should be seen well below the bottom of the pole clamp and the monitor should snap securely into place 4 2 075 0014 SECTION IV PREPARATION FOR USE AND INSTALLATION Figure 4 1 Pole Clamp To remove the monitor press tab on top of clamp shoe before sliding monitor up The pole clamp may be locked onto the monitor so that the two do not get separated To do this 1 Line up the clamp shoe on back of monitor with the slot on pole clamp and slide monitor down to fit The bottom of the clamp shoe should be seen well below the bottom of the pole clamp and the monitor should snap securely into place Make sure that set screw hole on pole clamp aligns with corresponding hole on clamp shoe Remove black knob screw from pole clamp Using the Allen wrench supplied secure pole clamp to monitor with the set screw provided Replace black knob screw To attach to pole place pole within clamp bracket and tighten screw using the black finger knob 4 1 4 3 Optional Mounting Accessories For information on optional mounting accessories request Aspect s
17. and diagnostic troubleshooting information necessary for qualified technical personnel to test and replace those parts of the equipment that are replaceable by the customer Aspect does not authorize nor provide information to service or repair the internal electronic components of the DSC or the A 2000 Monitor The qualified user may replace the DSC cables by following the procedures in Chapter 7 In cases of obvious damage pulled connectors or crushed cables the need to replace the cables is known If cables are suspect the following may be of aid in determining which cable is defective Monitor cable monitor connector to DSC box is a 5 conductor cable that handles all communications to from the DSC It uses two twisted pair and a ground for digital data transmission and for the clock transitions that are used to generate power in the DSC housing Therefore failure in this cable is usually seen as failure to recognize the DSC E01 DSC Not Connected or E02 DSC Overcurrent messages Pigtail cable DSC box to PIC connector is a 10 lead conductor that brings in the patient s EEG signal and also provides information about the sensor connected If this cable fails the system may indicate that the sensor is not connected or is illegal or the Sensor Check may fail or restart on its own Please note that all of the Pigtail conductors are also continued through the PIC cable to the sensor It is important that the PIC cable be elimin
18. checkout 5 3 075 0014 SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 9 a Press SILENCE key Verify icon shows at BIS banner b Press ARROW BACK key Verify REVIEW MODE screen displays c Press MENU EXIT key Verify REVIEW MODE screen exits d Press ARROW AHEAD key Verify REVIEW MODE screen displays e Press MENU EXIT key Verify REVIEW MODE screen exits f Press MENU EXIT key Verify Setup menu displays g Press DOWN ARROW key Verify highlight bar moves down with each press h Press UP ARROW key Verify highlight bar moves up with each press i Press SELECT key Verify that highlighted menu line is selected 1 Press MENU EXIT until BIS banner is displayed Check real time e Press MENU EXIT key Use the arrows to highlight Advanced Setup then press SELECT key e Verify that the time and date are correct for the local area If incorrect follow the instructions in Section of the A 2000 Operating Manual to reset 10 End of checkout 5 2 4 DSC Checkout Procedure Periodically the DSC and associated cables and connectors should be inspected for physical damage and verification that the self test will pass N DARY IF ANY FAILURES ARE NOTED SEE ERROR MESSAGE AND OPERATOR ACTIONS SECTION Using a known good A 2000 monitor see section 6 2 1 power up the A 2000 and verify the screen message Connect DSC Cable to Front of Monitor displays Connect the DSC that you ar
19. contamination moisture skin oils etc on PCB and component surfaces Unplug the DSC from the monitor Remove the two screws from the rear of the DSC case Separate the case halves by lifting off the top BIS Emblem from the bottom half Remove the ground shield securing screw See Figure 7 2 Carefully remove PCB assembly remove the shield by sliding it over the pigtail cable See Figure 7 3 Caution All work involving opening the instrument case must be performed in a static safe environment to prevent damage to electronic components and assemblies This environment includes the operator work area and tools and any other test or storage items that might touch the DSC assemblies Caution Handle PCBs only by board edges to avoid contamination moisture skin oils etc on PCB and component surfaces Carefully separate the two printed circuit boards to gain access to cable connections Be careful to not flex the PCBs or permanent damage to the connectors may result See Figure 7 4 Perform necessary service e g replacement of cable 7 5 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM Monit Ground shield securing screw Pigtail Shield cable Figure 7 2 Remove the ground shield securing screw Figure 7 3 Remove shield by sliding along pigtail cable 7 6 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM Pre amplifier PCB Pigtail cable Monitor cable im Loc
20. features Power Up e Automatic e Manual These features check the BIS Monitoring System s operability status and also report software and hardware malfunctions IF ANY FAILURES ARE NOTED SEE MESSAGE AND OPERATOR ACTIONS SECTION The power up diagnostics run automatically each time that you turn on the Monitor unit These procedures check the software system memory speaker display serial ports timer counters BIS Engine real time clock Data Memory and Setup Memory This level of diagnostic is described in detail in the monitor s operating manual The automatic diagnostics run continuously in the background while the unit operates These procedures check the Digital Signal Converter for the following conditions interface faults disconnect lead off and power faults This level of diagnostic also checks other monitor functions please reference this detail in the monitor s operating manual The manual diagnostics are operator initiated using the monitor keys and menu choices These procedures check for the proper functioning of the Digital Signal Converter This level of diagnostic also can be used to check other monitor functions please reference this detail in the monitor s operating manual NOTE The manual diagnostics can be run safely while the patient is connected to the BIS Monitor unit however running the diagnostics will temporarily disrupt monitoring Do not run DSC self test during electro caut
21. have occurred in transit and will be the responsibility of Purchaser This warranty does not extend to any Warranted Products or part thereof that have been subject to misuse neglect or accident that have been damaged by causes external to the Warranted Product including but not limited to failure of or faulty electrical power that have been used in violation of Aspect s instructions that have been affixed to any nonstandard accessory attachment on which the serial number has been removed or made illegible that have been modified by anyone other than Aspect or that have been disassembled serviced or reassembled by anyone other than Aspect unless authorized by Aspect Aspect shall have no obligation to make repairs replacements or corrections which result in whole or in part from normal wear and tear Aspect makes no warranty a with respect to any products that are not Warranted Products b with respect to any products purchased from a person other than Aspect or an Aspect authorized distributor or c with respect to any product sold under a brand name other than Aspect Medical Systems Inc 8 4 075 0014 SECTION VIII A 2000 SPECIFICATIONS THIS WARRANTY IS THE SOLE AND EXCLUSIVE WARRANTY FOR ASPECT S PRODUCTS EXTENDS ONLY TO THE PURCHASER AND IS EXPRESSLY IN LIEU OF ANY OTHER EXPRESS OR IMPLIED WARRANTIES INCLUDING WITHOUT LIMITATION ANY WARRANTY AS TO MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ASPECT S MAXIMUM LIAB
22. into an appropriate 3 wire receptacle WARNING FOR PROPER GROUNDING THE POWER RECEPTACLE MUST BE A THREE WIRE GROUNDED OUTLET A HOSPITAL GRADE OUTLET IS REQUIRED NEVER ADAPT THE THREE PRONG PLUG FROM THE MONITOR TO FIT A TWO SLOT OUTLET IF THE OUTLET HAS ONLY TWO SLOTS MAKE SURE THAT IT IS REPLACED WITH A THREE SLOT GROUNDED OUTLET BEFORE ATTEMPTING TO OPERATE THE MONITOR IF THE INTEGRITY OF THE EXTERNAL PROTECTIVE EARTH GROUND IS IN DOUBT THE A 2000 SHALL BE OPERATED FROM ITS INTERNAL BATTERY POWER SOURCE ONLY FOR A 2000s USED OUTSIDE OF NORTH AMERICA A HARMONIZED LINE CORD WITH CONDUCTORS HAVING A CROSS SECTIONAL AREA GREATER THAN 0 75 MM MUST BE USED 4 1 4 Site Preparation Mounting the Monitor Aspect Medical Systems Inc strongly recommends permanent mounting of the A 2000 monitor to the anesthesia machine to enhance safety and facilitate ease of use Please contact your local representative or Aspect to discuss mounting options WARNING BE SURE THE MONITOR IS MOUNTED SECURELY IN PLACE TO AVOID PERSONAL OR PATIENT INJURY 4 1 4 1 Direct Shelf Mounting The A 2000 has a centrally located screw hole on its underside in which a 10 24 machine screw can be used to secure the monitor to a shelf via through hole Engagement length of this screw hole is 23 inches 58 cm If a printer is attached use the holes on the underside of the printer 4 1 4 2 Mounting the Monitor using the Pole Clamp To mount the monitor to
23. power outlet wall outlet is supplying AC Volts of 110vac to 240vac at a frequency of 50hz to 60hz Move power cord to known good outlet e Verify AC power cord is good Swap power cord with known good one e Verify A2000 fuses are OK See section 7 2 for removal replacement procedure e failure continues after above actions monitor component will need to be serviced see section 7 3 6 Press each keyswitch on front panel e Verify appropriate response to keyswitch press See Installation Procedure for details switch by switch 7 Press Menu Exit key then highlight and select ADVANCED SETUP When the Advanced Setup Menu appears highlight and Select DIAGNOSTIC MENU From the Diagnostic Menu highlight and select Display Self Test Press Select to initiate the test Note This test will not be performed when the unit is on battery backup e Verify that all Pixels illuminate If there is a problem see section 7 for instructions on replacing the monitor e Press Menu Exit Menu Exit 8 Proceed to DSC section 6 3 075 0014 SECTION DIAGNOSTICS AND TROUBLESHOOTING 6 2 2 DSC Troubleshooting Procedure Note The successful completion of this test will verify the DSC is recognized and communicates with the monitor It also verifies the handling circuits are functioning properly to and from the DSC IF ANY FAILURES ARE NOTED SEE MESSAGE AND OPERATOR ACTIONS SECTION 1 With no Sensor conn
24. proper impedance values being detected by the BIS Monitoring System during the Sensor Check This test is part of the initial test that each monitor performs The simulator also allows for safety testing of BIS monitors in the field by allowing connection of the test equipment to the monitor via the patient interface cable NOTE The life expectancy of the Sensor Simulator is 25 connect disconnect cycles 1 S4 connect to 4 input signal pins on the patient interface cable The Inputs and are where one connects the test signals to test the BIS Monitor Figure 9 1 Schematic of Sensor Simulator Circuit Test types allowed Sensor Check This checks and verifies that the monitor is reporting the proper impedances that it sees from the Sensor Simulator This procedure verifies the proper functionality of the BIS monitoring system Connect the sensor plus simulator to the BIS Monitor at the patient interface cable The monitor should recognize that a sensor has been connected and report the proper impedance values Electrode 1 2 4 3 Acceptance 4 6 8 17 3 5 2 4 Range in Kohms The monitor then proceeds to display the BIS screen BIS is blank and after a few seconds SQI 100 9 3 075 0014 SECTION IX APPENDIX I Safety Testing Leakage Current Leakage Current testing should be performed by a qualified Biomedical Engineering Technician or authorized personnel only gt Connect Sensor
25. software control An overcurrent detector circuit monitors current to the DSC If the current exceeds the expected value the power is shut off to the DSC by the hardware and the DSP the Host and the user are notified 3 2 2 5 The Interconnect Board The A 2000 monitor is designed with a minimum of internal cabling The Interconnect board provides the physical mounting and electrical connections for the major subassemblies of the A 2000 It has mounted to it the power input module connector for the AC power cord and associated fuses the equipotential stud the power switch and connectors for the printer external the serial port external and the various internal connectors for the Main PCB Switch pad battery and fan 3 6 075 0014 SECTION III PRINCIPLES OF OPERATION 3 2 2 6 The Power Supply The power supply runs from AC power from 100 240 VAC 50 60 Hz It provides 5 V at 6 A and 12 V at 0 72 A outputs and charges the battery 6 cell 7 2 V nominal 1800 mA Signals are provided to the processor to indicate AC FAIL RESET and LOW BATTERY 3 2 2 7 The Battery The battery is for backup use only The battery is a six cell arrangement including temperature and current control elements and has a nominal output of 7 2 volts DC The battery is charged only when the A 2000 Monitor is turned on and running from AC power It is capable of supporting monitor operation for a minimum of 20 minutes Printer use is not enabl
26. that the Purchaser has given notice of such warranty claim within the Warranty Period and the Warranted Product is returned to the factory with freight prepaid Repair or replacement of Products under this warranty does not extend the Warranty Period To request repair or replacement under this warranty Purchaser should contact Aspect directly see contact information on the back cover of this manual Aspect will authorize Purchaser to return the Warranted Product or part thereof to Aspect Aspect shall determine whether to repair or replace Products and parts covered by this warranty and all Products or parts replaced shall become Aspect s property In the course of warranty service Aspect may but shall not be required to make engineering improvements to the Warranted Product or part thereof If Aspect reasonably determines that a repair or replacement is covered by the warranty Aspect shall bear the costs of shipping the repaired or replacement Product to Purchaser All other shipping costs shall be paid by Purchaser Risk of loss or damage during shipments under this warranty shall be borne by the party shipping the Product Products shipped by Purchaser under this warranty shall be packaged in the original shipping container or equivalent packaging to protect the Product If Purchaser ships a Product to Aspect in unsuitable packaging any physical damage present in the Product on receipt by Aspect and not previously reported will be presumed to
27. the Case section 7 4 2 2 above 7 9 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM 7 5 4 DSC Checkout And Safety Tests 1 Perform DSC Self Test See section 5 2 4 or refer to your monitor s operating manual e Verify that all test sections PASS Install a known good PIC cable Connect a Sensor Simulator or Test Sensor tool as described in Appendix I Perform a Patient Interface Cable PIC Checkout Procedure as described in section 5 2 5 of this manual For more detail refer to your monitor s Operating Manual 5 Perform a leakage electrical safety test according to the appropriate institution requirements 6 If appropriate perform a Hipot test according to institution requirements 00 End of Procedure 7 10 075 0014 SECTION VIII SECTION VIII 8 A 2000 SPECIFICATIONS 8 1 INTRODUCTION A 2000 SPECIFICATIONS This section lists specifications for the Aspect A 2000 BIS Monitor GENERAL SPECIFICATIONS Product Description Monitor Weight Monitor Dimensions Digital Signal Converter Weight Dimensions Cable Length Display Size Digital Output Power Requirements Electrical Safety Battery Backup Software Updates Processors 075 0014 BIS Bispectral Index monitor for display of processed data and real time EEG waveforms 3 1 Ibs 1 4 kg 7 0 in wide x 6 8 in high x 4 0 in deep 17 5 cm x 16 9 cm x 10 cm 10 0 oz 0 284 kg including integral cable 2 60
28. 000 Mounting Options Manual 070 0015 A 2000 Operating Manual 075 0013 A 2000 Service Information Manual 085 0017 Clinical Reference Manual ACCESSORIES 186 0109 Paper Printer 150 0037 Pole Clamp Assembly 817 0009 GCX 8 Pivot Support Arm w 6 Up Post 817 0010 GCX 15 Pivot Support Arm w 6 Up Post 817 0013 GCX 6 C Clamp Down Post 817 0011 GCX 7 Channel amp Adapter for Draeger Narkomed 817 0012 GCX 7 Channel w Dovetail Attachment for Ohmeda PARTS 675 0015 Restraining loop 430 0006 Fuse 1 25a 250V 5X20MM 175 0021 Monitor Cable 12 length monitor to DSC box 175 0022 Pigtail Cable for DSC s n D2 D3 3 length DSC to PIC connector 175 0045 Pigtail cable for DSC S N DX D4 9 length DSC to PIC connector 150 0036 Case DSC blue 150 0027 Shield 150 0026 Gasket 675 0004 Tie wrap 605 0024 Screw case 605 0022 Screw shield 800 0007 Hanging strap clip 675 0022 Cable wrap Blue hook amp loop 194 0039 Label DO NOT DISCARD 9 1 075 0014 SECTION IX APPENDIX I MISCELLANEOUS 180 0006 Empty carton with foam 2000 186 0105 Sensor Simulator 186 0137 Sensor Simulator for DSC serial numbers DXnnnnn D4nnnnn and PIC Adaptor 536 0044 RS 232 Cable 9 2 075 0014 SECTION IX APPENDIX 9 2 SENSOR SIMULATOR P N 186 0137 INSTRUCTIONS FOR USE Description of device The Sensor Simulator is a service tool that allows for the verification of
29. 251 A2000 MONITOR rote rubr tet tras hua kapaka 2 2 2 2 2 A 2000 DIGITAL SIGNAL 0000 00 2 2 2 2 3 SOFTWARE REVISION 8 2 2 2 3 PROPRIETARY INFORMATION AND DEVICES 2 2 3 PRINCIPLES OF OPERATION En rbi po uu ki uu uu 3 1 3 1 HOW THE A 2000 MONITORING SYSTEM 3 1 3 2 SYSTEM ARCHITECTURE uu 3 1 3 2 1 THE DIGITAL SIGNAL CONVERTER 5 3 4 32 2 THE 2000 MONITOR 55 vei bere nta te Hog etium p e incor a Dari e AER uad 3 6 2 2 3 A 2000 PRINTER tu ode eost tarea nere n egi x deuil dedo Se 3 7 3 3 SYSTEM 8 a u aulia ca E RYE PY KRLRR ERE GE iranan 3 7 3 3 1 SYSTEM SELF CHECKS 2 inr RE RYE a ek E ke daa ed edad 3 7 3 3 2 DATA MEMORY xs m oleae eee eed ede Aa Ta T Re RR ERR ERR 3 8 9 9 9 SAVED 3 8 3 39 4 BATTERY OPERATION It Per Aa p Rede oria 3 8 3 3 5 DATA TRANSFER AND
30. 4 and 1 with 2 5 Connect this Test Sensor to the PIC All impedance tests should complete successfully with low impedance values Typical values using this alternative Test Sensor are less than 5 K ohms 6 The following chart shows expected Test Sensor values Test Sensor Values Electrode Typical Range 1 1 K ohm 1 2 K ohms 2 1 K ohm 1 3 K ohms 4 1 K ohm 1 2 K ohms 3 1 K ohm 1 2 K ohms 9 5 075 0014 SECTION IX APPENDIX I 9 4 SAFETY TESTER CONNECTION WITH PIC This procedure describes a method to connect an A 2000 BIS Monitor System to a safety tester It uses a current date sensor to provide a contact point for the safety tester leads that correspond to the patient contact points of the BIS System Connect DSC Digital Signal Converter cable to A 2000 monitor Connect PIC Patient Interface Cable sensor cable to other end of DSC Verify sensor for test is of current date code Connect sensor to PIC cable Remove gel and green pad from each sensor contact point Connect safety tester leads per tester instructions to sensor contact point metallized pads i e via alligator clips Apply power to the A 2000 per safety tester instructions Make required leakage measurements 7 Disconnect and dispose of sensor NOTE DO NOT discard PIC or DSC These are re usable components ore ON D Contact points for safety tester See steps 4 and 5 Figure 9 6 Safety Tester Contact
31. 7 10 APPENDIX racio ian hri npe 10 1 10 1 A 2000 SIGNAL FLOW DIAGRAM eene nnne nennt 10 1 10 2 A 2000 BLOCK DIAGRAM eese nennen nnn 10 2 10 3 DSC WIRING DIAGRAM DSC CONNECTOR WIRE SIDE 10 3 10 4 WIRING DIAGRAM MONITOR CABLE u u 10 4 10 5 WIRING DIAGRAM PIGTAIL CABLE u u 10 5 TABLE FIGURES Figure 2 1 The A 2000 BIS Monitoring System 2 3 Figure 2 2 Rear View of u a oe e eet U l rh as ulus 2 4 Figure 3 1 The A 2000 System Block eene 3 2 Figure 3 2 The A 2000 Signal Flow Diagram essent eene 3 3 Figure g l Pole Clamp isa ed ue 4 3 Figure 7 1 Placement of PIC Securing Strap on BIS Monitoring System sse 7 1 Figure 7 2 Remove the ground shield securing screw eese eene 7 6 Figure 7 3 Remove shield by sliding along pigtail 7 6 Figure 7 4 Carefully separate the two printed circuit boards see 7 7 Figure 7 5 For a new gasket cable passages need to be carefully cut esse 7 8 Figure 7 6 Verify that the strain relief is fully secured in its recess sse 7 9 Figure 9 1 Schematic of Sensor Simulator Circuit eee eene 9 3 Figure 9 2 Sensor Simulator ge
32. ANDARDS AS AMATTER OF SAFE PRACTICE THE INSTITUTION SHOULD CONDUCT PERIODIC TESTS TO VERIFY THESE CURRENTS WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS OCCURS RE TEST BEFORE FURTHER USE SHOCK HAZARD DO NOT ATTEMPT TO DISCONNECT THE POWER CORD WITH WET HANDS MAKE CERTAIN THAT YOUR HANDS ARE CLEAN AND DRY BEFORE TOUCHING THE POWER CORD FOR PROPER GROUNDING THE POWER RECEPTACLE MUST BE A THREE WIRE GROUNDED OUTLET A HOSPITAL GRADE OUTLET IS REQUIRED NEVER ADAPT THE THREE PRONG PLUG FROM THE MONITOR TO FIT A TWO SLOT OUTLET IF THE OUTLET HAS ONLY TWO SLOTS MAKE SURE THAT IT IS REPLACED WITH A THREE SLOT GROUNDED OUTLET BEFORE ATTEMPTING TO OPERATE THE MONITOR 1 1 075 0014 SECTION SAFETY PRECAUTIONS IF THE INTEGRITY OF THE EXTERNAL PROTECTIVE EARTH GROUND IS IN DOUBT THE A 2000 SHALL BE OPERATED FROM ITS INTERNAL BATTERY POWER SOURCE ONLY FOR A 2000s USED OUTSIDE OF NORTH AMERICA A HARMONIZED LINE CORD WITH CONDUCTORS HAVING A CROSS SECTIONAL AREA GREATER THAN 0 75 mm MUST USED THE CONDUCTIVE PARTS OF ELECTRODES OR SENSOR AND CONNECTORS INCLUDING THE NEUTRAL ELECTRODE SHOULD NOT CONTACT OTHER CONDUCTIVE PARTS INCLUDING EARTH ANY PROCEDURES THAT REQUIRE THE REMOVAL OF THE DSC s COVER AND INTERNAL PARTS SHOULD BE PERFORMED BY A QUALIFIED BIOMEDICAL ENGINEERING TECHNICIAN POWER OFF THE UNIT BY DISCONNECTING FROM THE MONITOR TO REDUCE THE HAZARD OF BURNS IN THE HIGH FREQUENCY SURGICAL NEUTRAL
33. AS SPILLAGE OF BLOOD OR SOLUTIONS OCCURS RE TEST BEFORE FURTHER USE 5 2 3 A 2000 Monitor System Checkout Procedure The following test procedure should be performed periodically to ensure that the A 2000 is functioning properly IF ANY FAILURES ARE NOTED SEE ERROR MESSAGE AND OPERATOR ACTIONS SECTION 1 Secure all components e Monitor with power cable e DSC e PIC Patient interface cable connects DSC to patient e BIS Sensor or test sensor or Sensor Simulator 2 Connect power cable to monitor plug power plug into appropriate wall outlet 3 Power up monitor by pressing power switch right lower side of case e Verify beep tone as power switch is activated e Verify fan rear case wall moves air outward e Verify all power on self tests complete successfully approx 30 seconds e Verify next screen says CONNECT DSC CABLE TO FRONT OF MONITOR 4 Connect DSC with PIC and Sensor Test Sensor or Sensor Simulator see Appendix 1 e Verify DSC test completes e Verify SENSOR CHECK screen displays 5 Exit from SENSOR CHECK screen by passing impedance check or by exiting with MENU EXIT key 6 Disconnect power cord from rear of monitor e Verify OPERATING ON BATTERY BACKUP 33 is displayed e Verify battery icon displays in BIS banner 7 Reconnect power cord e Verify battery icon is not displayed in BIS banner e Verify no OPERATING ON BATTERY BACKUP E33 is displayed 8 Perform keyboard
34. Aspect Medical Systems A 2000 XP PLATFORM BISPECTRAL INDEX BIS MONITORING SYSTEM PLATFORM SERVICE INFORMATION MANUAL Rx Only Copyright 2004 Aspect Medical Systems All rights reserved Copying or other reproduction of this document is prohibited without prior written consent of Aspect Medical Systems Aspect Medical Systems Inc Aspect Medical Systems International B V 141 Needham St Rijnzathe 7d2 Newton MA 02464 3454 PV De Meern U S A The Netherlands Tel 617 559 7000 Tel 31 30 662 9140 0123 Tel 888 BIS INDE X U S only Fax 31 30 662 9150 Fax 617 559 7400 amsint aspectms com bis_info aspectms com www aspectmedical com 075 0014 1 02 TABLE OF CONTENTS 1 SAFETY Fidungs rar Sana uno uk Ya E Sari iin nu x 1 1 1 1 1 2 CAUTIONS crated dress n 1 3 1 3 i2dickd eR 1 5 2 A 2000 SYSTEM OVERVIEW U u RE FE za SL A Y 2 1 2 1 INTRODUCING THE A 2000 BIS MONITORING SYSTEM 2 1 2 1 1 PRINCIPAL 2 1 2 1 2 How THE A 2000 WORKS 2 2 2 2 INSTRUMENT IDENTIFICATION ioco exon dian 2 2 2
35. E67 E86 Printing Disabled E87 E88 E89 See E35 E90 Printer Connected E91 Event Marking Disabled E92 E93 Sensor Ground Fault E94 Sensor Overcurrent 075 0014 DIAGNOSTICS AND TROUBLESHOOTING Causes The print head on the printer has been lifted The printer is no longer connected to the A 2000 The A 2000 is running on battery backup The printer has been successfully connected to the A 2000 Data extraction is in progress Not used Problem is detected relating to sensor ground element Sensor is using too much current 6 14 Corrective Actions 1 Return print head to the closed position 2 Ifthe error persists the printer requires service 1 Check the connection If the error persists the printer requires service Note Printer functions only on AC power When power is restored printing will again be available Note This is a message rather than an error No action is necessary Note Event marking will be re enabled when event data extraction has been completed less than 10 minutes 1 Disconnect and examine sensor connection Clean any contamination present 2 Replace sensor if necessary 3 Replace PIC 4 Replace DSC 1 Disconnect and examine sensor connection Clean any contamination 2 Replace sensor if necessary 3 Replace PIC 4 Replace DSC SECTION Status Messages E95 E108 Invalid Sensor 075 0014 DIAG
36. ILITY ARISING OUT OF THE SALE OF THE PRODUCTS OR THEIR USE WHETHER BASED ON WARRANTY CONTRACT TORT OR OTHERWISE SHALL NOT EXCEED THE ACTUAL PAYMENTS RECEIVED BY ASPECT IN CONNECTION THEREWITH ASPECT SHALL NOT BE LIABLE FOR ANY INCIDENTAL SPECIAL OR CONSEQUENTIAL LOSS DAMAGE OR EXPENSE INCLUDING WITHOUT LIMITATION LOST PROFITS DIRECTLY OR INDIRECTLY ARISING FROM THE SALE INABILITY TO SELL USE OR LOSS OF USE OF ANY PRODUCT EXCEPT AS SET FORTH HEREIN ALL PRODUCTS ARE SUPPLIED AS IS WITHOUT WARRANTY OF ANY KIND EITHER EXPRESS OR IMPLIED 8 5 075 0014 SECTION VIII A 2000 SPECIFICATIONS 8 6 075 0014 SECTION IX APPENDIX I SECTION IX 9 APPENDIX I 9 1 SPARES AND ACCESSORIES LIST MAJOR COMPONENTS 185 0070 A 2000 Monitor 185 0071 DSC 2 Assembly serial number D3nnnnn 185 0095 DSC 3 serial number D4nnnnn 185 0124 DSC XP serial number 186 0067 PIC S Cable 186 0107 PIC for DSC serial number DXnnnnn 4 186 0126 PIC Adaptor for DSC S N D3 with Plus or XP compatible sensors 186 0100 BIS Standard sensor 186 0106 BIS Quatro sensor 186 0110 BIS Pediatric sensor 186 0160 BIS Extend sensor 175 0039 Power Cord 10 North America 175 0038 Power Cord 15 North America 186 0031 A 2000 Printer 150 0058 A 2000 Stand MANUALS 070 0031 A 2
37. INST EFFECTS OF CARDIAC DEFIBRILLATION The A 2000 system provides protection for the operator and patient during cardiac defibrillation DEGREE OF PROTECTION AGAINST THE INGRESS OF WATER Monitor degree of protection rating IPX2 ingress of water dripping DSC degree of protection rating IPX4 splash proof MODE OF OPERATION OF THE SYSTEM Continuous Operation under normal load for a normal period without exceeding the specified limits of temperature CLASSIFICATION MEDICAL ELECTRONIC EQUIPMENT CLASSIFIED BY UNDERWRITERS LABORATORIES INC WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 2601 1 IEC601 2 26 CAN CSA C22 2 NO 601 1 CAN CSA 601 2 26 42SA 8 3 075 0014 SECTION VIII A 2000 SPECIFICATIONS 82 WARRANTY Aspect warrants to the initial Purchaser that the A 2000 BIS monitor and the Digital Signal Converter Warranted Product will be free from defects in workmanship or materials when given normal proper and intended usage for a period of one year Warranty Period from the date of its initial shipment to Purchaser Excluded from this warranty are expendable components and supply items such as but not limited to electrodes cables and prep solutions Aspect s obligations under this warranty are to repair or replace any Warranted Product or part thereof that Aspect reasonably determines to be covered by this warranty and to be defective in workmanship or materials provided
38. ION OR SERVICE Caution All work involving opening the instrument case must be performed in a static safe environment to prevent damage to electronic components and assemblies This environment includes the operator work area and tools and any other test or storage items that might touch the DSC assemblies 7 5 1 Required Tools And Supplies Table 7 1 Tools and Supplies to Assemble Disassemble the DSC Tool or Supply Part Required for 1 Phillips screwdriver __ Removing the case cover screws Small flat blade screwdriver Removing the PCB board Needle nose pliers Releasing standoff tabs to separate boards Wire cutters Trim remove tie wraps A 2000 Monitor To perform DSC Self Test PIC cable To perform Sensor Check A 2000 Sensor Simulator To perform Sensor Check XP compatible sensor To perform Sensor Check DSC XP 185 0124 To perform DSC Self Test Caution Use only the parts and tools specified Use of any others may damage the instrument 7 4 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM 7 5 2 The DSC Case 7 5 2 1 Opening the Case Caution All work involving opening the instrument case must be performed a static safe environment to prevent damage to electronic components and assemblies This environment includes the operator work area and tools and any other test or storage items that might touch the DSC assemblies Caution Handle PCBs only by board edges to avoid
39. MAINTENANCE u u u 5 2 5 2 1 CHECKING THE r 5 2 5 2 2 CHECKING LEAKAGE CURRENT 5 2 5 2 3 A 2000 MONITOR SYSTEM CHECKOUT 5 3 52 24 DSC CHECKOUT PROCED RE a e D 5 4 5 2 5 PATIENT INTERFACE CABLE PIC CHECKOUT PROCEDURE 5 4 5 2 6 PHYSICAL INTEGRITY 5 5 5 6 DIAGNOSTICS AND TROUBLESHOOTING 6 1 6 1 GENERAL TROUBLESHOOTING U 6 1 6 2 A 2000 BIS MONITOR SYSTEM TROUBLESHOOTING PROCEDURE 6 2 6 2 1 MONITOR TROUBLESHOOTING 6 2 6 2 2 DSC TROUBLESHOOTING 6 4 6 2 3 PIC AND SENSOR TROUBLESHOOTING 6 5 6 2 4 DSC CABLE PROBLEM 5 6 6 6 3 STATUS MESSAGES AND OPERATOR ACTIONS 6 7 7 SERVICING THE A 2000 5 4 40 7 1 7 1 REPLACING MAJOR 7 1 T end Seinen 7 1 71 2 REPLACE THE 5 e E RM i ERR ER RE EE E 7 2 7 1 3
40. NOSTICS AND TROUBLESHOOTING Causes Corrective Actions The sensor is unusable for 1 Replace sensor one of the following 2 Replace PIC reasons 3 Replace DSC E95 The sensor has been connected too many times E96 E102 E107 The sensor contains invalid information E103 E106 E108 The sensor information could not be read 6 15 SECTION DIAGNOSTICS AND TROUBLESHOOTING 6 16 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM SECTION VII 7 SERVICING THE A 2000 SYSTEM INTRODUCTION This section provides instructions for replacing the PIC DSC and the Monitor and for removing and replacing parts of the Digital Signal Converter DSC If a component needs to be serviced please consult sections 7 2 and 7 3 for instructions on packaging and shipping The A 2000 BIS Monitoring System is designed to be easily serviced by using the built in diagnostic routines see Section 6 of this Manual and the major component swapping techniques described below 7 1 REPLACING MAJOR COMPONENTS 7 1 1 To replace the PIC 1 Unplug the PIC cable from the patient end of the DSC by grasping the connectors NOT the cable and firmly pulling the two sections apart The restraining loop will need to be released by clipping the nylon tie wrap to release the loop 2 attach the replacement PIC cable to the DSC pigtail align and press PIC and DSC connectors together firmly Reconnect the restraining loop by fastening to c
41. OR USE AND INSTALLATION 43 INSTALLATION AND VERIFICATION PROCEDURE 1 Open packages and inspect for all components e Monitor with power cable e DSC Digital Signal Converter e PIC Patient interface cable You will also need a BIS Sensor or Sensor Simulator 2 Connect power cable to monitor plug power connector into appropriate wall outlet 3 Power up monitor by pressing power button lower right side of case e Verify beep tone as power button is activated e Verify fan rear case wall moves air outward e Verify that self test procedure completes successfully approx 30 seconds e Verify next screen says CONNECT DSC CABLE TO FRONT OF MONITOR 4 Connect DSC with PIC and Sensor Refer to the A 2000 Operating Manual for detailed instructions or with PIC and Sensor Simulator See this manual Section 9 2 e Verify that DSC test completes e Verify SENSOR CHECK screen displays 5 Exit from SENSOR CHECK screen by passing impedance check or by exiting with MENU EXIT key 6 Disconnect power cord e Verify OPERATING ON BATTERY BACKUP 33 is displayed e Verify battery icon displays in BIS banner top left corner of screen 7 Reconnect power cord e Verify battery icon is not displayed in BIS banner e Verify no OPERATING IN BATTERY BACKUP E33 DISPLAY 8 Do keyboard checkout e Press SILENCE key Verify icon shows at BIS banner e Press Review ARROW BACK key Verify REVIEW MODE Menu displays e Press MENU EXIT key Verify RE
42. PERATING ON BATTERY BACKUP or E34 BATTERY LOW message SN CE NOTE Since the A2000 monitor has a built in battery backup the monitor will power up with no AC power applied This step assures that this part of the test is performed in AC operation IF ANY FAILURES ARE NOTED e Verify AC power outlet wall outlet is supplying AC Volts of 110vac to 240vac at a frequency of 50hz to 60hz Move power cord to known good outlet e Verify AC power cord is good Swap power cord with known good one e Verify A2000 fuses are OK See section 7 2 for fuse removal replacement procedure e failure continues after above actions monitor component will need to be serviced see section 7 3 6 2 075 0014 SECTION VI DIAGNOSTICS AND TROUBLESHOOTING 4 Unplug the AC power cord from monitor NOTE Since the A2000 monitor has a built in battery backup the monitor will power up with or without AC power applied This step assures that this part of the test is performed in Battery operation e Verify monitor continues to operate time depends on state of battery charge e Verify message E33 OPERATING BATTERY POWER or E34 BATTERY POWER LOW display e Verify battery icon displays in BIS banner border 5 Reconnect AC power cord to monitor Verify Monitor continues to operate e Verify E33 and E34 messages do not display e Verify battery icon does not display in BIS banner border IF ANY FAILURES ARE NOTED e Verify AC
43. Points 9 6 075 0014 SECTION IX APPENDIX I 9 5 PLACEMENT OF PIC SECURING STRAP ON BIS MONITORING SYSTEM INN 1 While the PIC is connected to the DSC place the loop strap over the strain relief as indicated 2 Insert each cable tie through the hole at the end of the loop and secure the tie firmly by pulling it through its own securing latch Cut the excess tie so that the strap is flush to the latch and no excess material is exposed 3 Repeat for the other side of the PIC mated cable If cable has to be removed do not cut the cable tie as it is intended to keep the system together If the loop strap is cut for whatever reason please replace as soon as possible The PIC part number 186 0107 PIC with Loop Strap Tiewraps and Label includes the following 1 each 675 0015 Tie Loop strap 2 each 675 0004 Tie Wrap 1 each 194 0039 Label do not discard 1 each 071 0002 Document instructions 9 7 075 0014 SECTION IX APPENDIX I 9 8 075 0014 SECTION X APPENDIXII SECTION X 10 APPENDIX Il 10 1 A 2000 SIGNAL FLOW DIAGRAM Input H Filter SIGMA DELTA Protect Modulator gt SIGMA DELTA CH2 AMP Modulator Patient Connecti Input Protect DC CN er ah otaa Su DSC Digital HOST Signal CPU DISPLAY i COMM Processor
44. See E35 E66 Printer Out of Paper E67 E82 E85 Printer Error E68 E69 E73 See E35 E74 See E47 E75 See E35 E76 Illegal Sensor for Demo E77 Simulator Connected 075 0014 DIAGNOSTICS AND TROUBLESHOOTING Causes No data received from BIS Engine in last 10 seconds BIS Engine has not responded to a command within 12 seconds Error in erasing or writing Data Memory portion of FLASH memory Printer is out of paper Printer is off line Print head is up Printer is disconnected Print head temperature out of range 5 Print head voltages out of range SNS 6 Printer hardware failure Not used A Demo Device has been connected to the serial port but the PIC is connected to a sensor A Demo Device has been connected to the monitor 6 13 Corrective Actions 1 2 3 Turn monitor off then on again Verify the software level Replace monitor Turn monitor off then on again Verify the software level Replace monitor Replace monitor Install the paper Put printer on line Close print head Connect printer Replace printer Disconnect the sensor and connect the PIC to the Demo Device or Disconnect Demo Device Note This is a message rather than an error No action is necessary SECTION Status Messages E78 E80 See E35 E81 See E47 E82 See E67 E83 Printer Head is Up E84 Printer Disconnected E85 See
45. Simulator to the patient interface cable of the BIS monitor as if it was sensor connection gt Short the two circular terminals at the end of the simulator using conventional methods such as jumpers or alligator clips Wire attached with screws will work also gt Connect the test signal to the shorted terminals Make sure that you are not touching the simulator beyond this point Proceed to test instrument for Leakage current as per established facility protocols and procedure for safety testing of medical devices 731 1 E al From Safety Tester cu Figure 9 2 Sensor Simulator 9 3 TEST SENSOR Use the following procedure to make a Test Sensor 1 Remove a new sensor from its plastic carrier sheet and place on flat surface with the adhesive facing up NOTE Be careful that gel does not leak onto hands or connector during this procedure Figure 9 3 BIS Sensor 9 4 075 0014 SECTION IX APPENDIX I 2 Place the end of a small paper clip at the midpoint of electrode 2 then lay it across electrode 4 See Figures 9 3 and 9 4 Figure 9 4 Connecting electrodes 2 and 4 3 Fold electrode 3 over onto electrode 4 pressing adhesive surfaces together and making sure the paper clip remains in place 4 Fold electrode 1 over onto electrode 2 Figure 9 5 Connecting electrode 3 with
46. TO REPLACE THE 7 2 7 2 FUSE REMOVAL AND 0 0 7 2 7 3 WHAT TO DO WITH A COMPONENT THAT REQUIRES SERVICE 7 3 1 4 REPACKAGING FOR SHIPPING AND 7 3 7 5 DISASSEMBLING AND REASSEMBLING THE 7 4 7 5 1 REQUIRED TOOLS AND 5 7 4 f THE DSC CASE mhia ua 7 5 7 5 3 THE DSC CABLES MONITOR AND 7 9 7 5 4 DSC CHECKOUT AND SAFETY 8 8 7 10 8 A 2000 SPECIFICATIONS u u ou etn eo in wa Errem 8 1 INTRODU C TO N zu aan saan 8 1 82 WARRANTY iani d pari a th tp 8 4 DEP EIN DMs LS ts Sai Eae EIAS 9 1 9 1 SPARES AND ACCESSORIES 1 9 1 9 2 SENSOR SIMULATOR P N 186 0137 INSTRUCTIONS FOR USE 9 3 9 3 2 00 aaa REN 9 4 9 4 SAFETY TESTER 9 6 9 5 PLACEMENT OF PIC SECURING STRAP ON BIS MONITORING SYSTEM 9
47. TRUMENT CASE IS OPENED BY A QUALIFIED BIOMEDICAL ENGINEERING TECHNICIAN 5 2 2 Checking Leakage Current Leakage current is a primary indicator of electrical shock hazard to personnel making contact with any exposed outer surface of the equipment Each A 2000 BIS Monitor is carefully checked at the factory to verify that leakage current meets the UL2601 and IEC601 1 safety standards The A 2000 should be routinely checked for leakage current at least once a year See Appendix 1 Always check the leakage current after a saline or blood spill or immediately after a major surge in the house electrical system and after every time the monitor case has been opened Keep in mind that liquids such as saline and Ringer s as well as blood are all excellent conductors of electricity Avoid touching any part of the system with wet hands Always work with clean dry hands 5 2 075 0014 SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING WARNINGS SHOCK HAZARD DO NOT ATTEMPT TO DISCONNECT THE POWER CORD WITH WET HANDS MAKE CERTAIN THAT YOUR HANDS ARE CLEAN AND DRY BEFORE TOUCHING THE POWER CORD ELECTRICAL SHOCK HAZARD THE MANUFACTURER S INSPECTION OF THIS APPARATUS VERIFIED THAT THE GROUND LEAKAGE CURRENT AND THE PATIENT SAFETY CURRENT WERE LESS THAN THE SPECIFIED LIMITS ESTABLISHED BY THE APPLICABLE SAFETY STANDARDS ASA MATTER OF SAFE PRACTICE THE INSTITUTION SHOULD CONDUCT PERIODIC TESTS TO VERIFY THESE CURRENTS WHENEVER AN EVENT SUCH
48. VIEW MODE MENU exits e Press Review ARROW AHEAD key Verify REVIEW MODE Menu displays e Press MENU EXIT key Verify REVIEW MODE MENU exits e Press MENU EXIT key Verify Setup Menu displays e Press DOWN ARROW key Verify highlight bar moves down with each press e Press UP ARROW key Verify highlight bar moves up with each press e Press SELECT key Verify that highlighted menu line is selected e Press MENU EXIT Verify that Setup Menu exits 9 End of procedure 4 5 075 0014 SECTION IV PREPARATION FOR USE AND INSTALLATION 4 6 075 0014 SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING SECTION V 5 PREVENTIVE MAINTENANCE CARE AND CLEANING INTRODUCTION This section describes e Care and cleaning procedures e Preventive maintenance 51 CARE AND CLEANING WARNING UNIVERSAL PRECAUTIONS SHALL BE OBSERVED TO PREVENT CONTACT WITH BLOOD OR OTHER POTENTIALLY INFECTIOUS MATERIALS PLACE CONTAMINATED MATERIALS IN REGULATED WASTE CONTAINER 5 1 1 Cleaning the Monitor and Digital Signal Converter Clean any spillage of blood or solutions on either the monitor or Digital Signal Converter as soon as possible Dried blood is very difficult to remove Use lint free absorbent towels for spill cleanups Dampen the towel with detergent and lukewarm water to aid in cleaning After cleaning dry all areas with a lint free absorbent paper towel Wipe Pigtail to PIC cable connector with alcohol and allow to dry completely Residual moisture in
49. able strain relief with the tie wrap Figure 7 1 Placement of PIC Securing Strap on BIS Monitoring System While the PIC is connected to the DSC place the loop strap over the strain relief as indicated 2 Insert each cable tie through the hole at the end of the loop and secure the tie firmly by pulling it through its own securing latch Cut the excess tie so that the strap is flush to the latch and no excess material is exposed 3 Repeat for the other side of the PIC mated cable If cable has to be removed do not cut the loop strap as it is intended to keep the system together If the loop strap is cut for whatever reason please replace as soon as possible Order number 675 0015 7 1 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM 7 1 2 To replace the DSC Disconnect the DSC from the front of the monitor To do this pull on the blue rubber overlay DO NOT twist or pull on the gray cable If necessary unplug the PIC cable from the patient end of the DSC by grasping the connectors NOT the wires and firmly pulling the two sections apart The restraining loop will need to be released by clipping the nylon tie wrap to release the loop To install the replacement DSC hold the cylindrical connector with the flat side up at 12 o clock position and insert firmly into DSC port on front of monitor To re attach the PIC cable align the PIC and DSC connectors and press together firmly Reconnect the restraini
50. ated as a potential problem before opening the DSC to change the pigtail cable They both can generate the same types of faults Procedure Once a problem is isolated to the DSC use the following sequence to determine the most probable cable 1 With no sensor connected run DSC Self test see section 6 2 2 fails or will not start or the DSC is not recognized the Monitor cable is likely suspect This failure may also be related to a PCB problem in the DSC in which case the DSC must be returned to Aspect for Service e If test indicates PASS continue to step 2 2 Connect Sensor Simulator or Test Simulator Run Sensor Check see section 6 2 3 e ffailure is noted swap PIC with known good PIC cable and run again e ffailure repeats the Pigtail cable is likely suspect This failure may also be related to a PCB problem in the DSC in which case the DSC must be returned to Aspect for service Note A continuity test can be used to verify conductor integrity use diagrams in Appendix Il 6 6 075 0014 SECTION DIAGNOSTICS AND TROUBLESHOOTING 6 3 STATUS MESSAGES AND OPERATOR ACTIONS Status Messages E01 DSC Not Connected E02 DSC Overcurrent E03 DSC Power Regulation E04 E05 E06 E07 E08 DSC Error E09 DSC Shut Down E10 Illegal DSC Type 075 0014 Causes 1 DSC disconnected 2 Defective DSC cable 3 Defective DSC 4 Defective monitor Defective DSC cable 2 Defective DSC 3
51. e testing and verify the screen message Connect Sensor to DSC displays Perform DSC Self Test See section 6 2 2 Repeat DSC Self test while flexing cables to see if intermittent opens or shorts exist If PIC and Sensor are available perform PIC Checkout Procedure See section 6 2 3 Repeat Sensor check while flexing cable to see if intermittent opens or shorts exist 5 2 5 Patient Interface Cable PIC Checkout Procedure The Patient Interface Cable PIC checkout is done using the A2000 monitor s Sensor Check routine and a Sensor Simulator see Appendix 9 2 or Test Sensor see Appendix 9 3 IF ANY FAILURES ARE NOTED SEE ERROR MESSAGE AND OPERATOR ACTIONS SECTION Using a known good A 2000 monitor see section 6 2 1 and a known good DSC see section 6 2 2 connect the PIC to be tested to the DSC and the DSC to the A 2000 monitor Power ON the A 2000 monitor and DSC At the completion of the power up self test verify that the screen message Connect Sensor to DSC displays Connect a Sensor Simulator see Appendix 9 2 or Test Sensor see Appendix 9 3 to the PIC cable Press firmly together 5 4 075 0014 SECTION V 4 PREVENTIVE MAINTENANCE CARE AND CLEANING A sensor check is initiated automatically when the sensor and PIC are connected to the DSC It may also be initiated by the user in the Setup Menu The impedance value for each electrode in kilo ohms appears on the screen along with its
52. ected plug DSC into the front of Monitor case e Verify monitor display reads Connect Sensor to DSC 2 Press Menu Exit key then highlight and Select ADVANCED SETUP then highlight and Select DIAGNOSTIC MENU then highlight and Select DSC SELF TEST Verify display shows DSC Test Results PASS this takes approximately 20 seconds Note that test results are posted in 4 tests for 2 channels The numbers are not important but if the test fails or the word FAIL will indicate which test is at fault 3 Press Menu Exit key and re run DSC SELF TEST this time moving cables flexing near connectors to check for intermittent connections or conductors To execute highlight and select DSC SELF TEST e Verify display shows DSC Test Results PASS this takes approximately 20 seconds 4 Press Menu Exit key until Connect Sensor to DSC screen displays IF ANY FAILURES ARE NOTED SEE MESSAGE AND OPERATOR ACTIONS SECTION 5 Proceed to PIC and Sensor Check 6 4 075 0014 SECTION 6 2 3 PIC AND SENSOR Troubleshooting Procedure DIAGNOSTICS AND TROUBLESHOOTING Notes The successful completion of this test will verify the BIS system from the DSC circuits to the patient connector Since the conductors used are located both in the DSC and the PIC use a swapping technique to isolate the faulty component Use a Sensor Simulator or make a Test Sensor for this test See Appendix for details
53. ed during battery operation The battery s life expectancy is approximately 5 years or 200 charge and discharge cycles whichever occurs first 3 2 2 8 The Fan A fan is located on the rear panel of the chassis The fan keeps the temperature rise inside the chassis to within approximately 10 degrees of the ambient temperature 3 2 3 A 2000 Printer The printer is fastened to the monitor bottom and is connected mechanically and electrically no cable required The printer interface is compatible only with the optional A 2000 printer See the A 2000 Printer Operating Manual for details 3 3 SYSTEM FEATURES 3 3 1 System Self Checks The A 2000 monitor has several self checking features to ensure that the system is operating properly These include System Check Initiated when the unit is powered on this test makes certain that system software and hardware components are functioning properly Equipment and Connection Checks The system checks continuously to be sure that the DSC the PIC and patient sensors are operating properly and have not become disconnected DSC Self Test The Digital Signal Converter self test is a thorough test of the entire signal processing chain including the signal processing computers The DSC self test may be initiated in the Diagnostic Menu see Section 6 2 2 Display Self Test This self test also found in the Diagnostic Menu see Section performs a thorough test of the screen display function
54. elet at the corner to allow the user to use a strap to hang the DSC or pin it to the bed sheets The patient connection is accomplished with an Aspect BIS Sensor for low and balanced impedance The Patient Interface Cable pigtail and the monitor cables are strain relieved and permanently attached to the case A proprietary connector is used for the patient interface cable connection and a high quality plastic connector is used for connection to the monitor There are no ventilation holes in the DSC case It will not leak when splashed with liquids The case is electrically shielded both to prevent spurious emissions from the DSC and to prevent externally caused interference with the DSC circuits 3 5 075 0014 SECTION III PRINCIPLES OF OPERATION 3 2 2 The A 2000 Monitor The A 2000 monitor contains the circuits for digitally processing the EEG data computing the processed parameters and displaying the waveforms and processed parameters The monitor also contains the circuits for powering the monitor DSC and companion printer Printer function is not available on battery backup A block diagram depicting the monitor subassemblies appears in Figure 3 1 A signal flow diagram appears in Figure 3 2 The signals are acquired and digitized by the DSC The DSC multiplexes the signals onto the DSC communications line The EEG signals are de multiplexed in the monitor and fed to the Digital Signal Processor DSP The processor filters and down sam
55. elf test The signal is a 2 Hz square wave of approximately 50uV It is applied to the inputs of the differential amplifiers resulting in a test of the entire signal path except for the input connections and protection circuits During self test noise gain and frequency response are checked Interface to the Monitor Patented technology The outputs from the two channels are multiplexed in a field programmable gate array FPGA Multiplexed with the EEG data is status information such as DSC identification mode bipolar or referential lead off indication and power supply faults The status information is only transmitted on command from the host The output is transformer coupled onto a balanced twisted pair line The transformer provides the required patient isolation The communications board decodes the control information coming from the host via a command line Commands such as block amplifier saturation and conduct impedance tests are transmitted DSC Power Supply Patented technology The DSC derives power from the command line The line is balanced twisted pair coupled via a transformer on the communications board The isolated signal is rectified and linearly regulated to provide DSC power 3 2 1 5 The DSC Mechanicals The DSC is contained in a small custom designed plastic case about the size of a computer mouse see Figure 2 1 It is connected to the monitor via a narrow highly flexible cable The case has one large ey
56. ery as it may erroneously indicate a failure 6 1 075 0014 SECTION DIAGNOSTICS AND TROUBLESHOOTING 6 2 A 2000 BIS MONITOR SYSTEM TROUBLESHOOTING PROCEDURE The A 2000 System consists of three major components Monitor DSC and PIC Sensor By using this three step procedure and a component swapping technique the component at fault can easily be determined The steps of this test procedure are Monitor e DSC recognition and self test e PIC and Sensor Following the test procedure is a MESSAGE AND OPERATOR ACTION section that describes appropriate actions for messages that may occur See Section 7 for directions on replacing components and handling components that require service 6 2 1 MONITOR Troubleshooting Procedure Note The successful completion of this test will verify the monitor is functioning and all operator input keyswitch and output display and audio alarm are OK IF ANY FAILURES ARE NOTED SEE MESSAGE AND OPERATOR ACTIONS SECTION Disconnect DSC from Front of Monitor Connect power cord to rear of monitor and attach to AC power source Press POWER ON switch e Verify beep tone as POWER SWITCH is actvated e Verify fan rear case wall is moving air outward e Verify display shows all self tests complete successfully approx 30 seconds e Verify that at completion of self tests display shows CONNECT DSC TO FRONT OF MONITOR e Verify no battery icon in BIS banner border and no E33 O
57. in wide x 1 00 in high x 4 25 in deep 6 6 cm x 2 5 cm x 10 8 cm 12 ft 3 7 m Integral DSC Cable 4 1 2 ft 1 4 m Patient Interface Cable 3 4 in high x 4 5 in wide 8 5 cm x 11 25 cm RS232 serial port printer port isolated from ground 100 240 VAC 50 60 Hz 1 ampere max Conforms to UL 2601 CSA 22 2 No 601 1 and IEC 601 1 20 minutes at full operation Recharge Time 4 hours User via RS 232 serial port Texas Instruments TMS 320C32 50 MHz Sharp LH77790 Risc 25 MHz 8 1 SECTION VIII EEG SPECIFICATIONS e Epoch Duration e Artifact Rejection e EEG Scales e EEG Sweep Speeds e Computed Parameters e User defined Displays e Update Rate e Event Markers e Alarms e Filters e Mode A 2000 SPECIFICATIONS 2 seconds Automatic 25 uV div 1 mV Full Scale Display Type EEG 5 10 25 50 100 uV div 25 mm sec Display Type EEG 15 25 30 mm sec Bispectral Index 95 Spectral Edge Frequency Suppression Ratio EMG and Signal Quality Index TREND DSA BIS Log and real time EEG waveforms 1 second for BIS Index 10 seconds for Trend DSA User selected Auditory and visual user adjustable limits ON 2 70 Hz with notch or OFF 25 100 Hz Sensor automatically selects mode DIGITAL SIGNAL CONVERTER SPECIFICATIONS e Analog to Digital Converter e Sampling Rate e Resolution Input Impedance Noise e Common Mode Rejection Isolation mode e Ba
58. indicates electrical activity that may be interfering with EEG recognition 3 PIC is defective 4 DSC is defective Note This message may occur as the results of artifact non EEG signal such as those generated from motion patient movement or eyeblinks or the presence of electro cautery or other devices The BIS has fallen below the low alarm limit set by the user The numeric display flashes The BIS has risen above the high alarm limit set by the user The numeric display flashes Corrective Actions 1 Re prep the sensor 2 Check impedance 1 If ARTIFACT label appears in EEG waveform box attempt to identify and eliminate artifact source 2 If EMG bar is illuminated attempt to determine and eliminate cause 3 Verify Sensor Check passes If not replace PIC 4 Replace DSC Take note of BIS at limit set by user Take note of BIS at limit set by user SECTION Status Messages E31 Isoelectric EEG Detected E32 Review Data Lost E33 Operating On Battery Backup E34 Battery Power Low E35 E39 E43 E44 E48 E65 E69 E73 E75 E78 E80 E87 E89 Software Error E40 Setup Memory Write Error E41 Real Time Clock Write Error E42 Can t Save Setup E43 E44 See E35 075 0014 DIAGNOSTICS AND TROUBLESHOOTING Causes No discernible EEG activity is detected for several minutes SR 100 Note This message notifies user of a flatline EEG This i
59. ined impedance is checked The DSC can also measure the individual electrode impedance by injecting current into the REF electrode only Individual electrode impedance is derived by subtracting the resulting value from the combined value BIS monitoring is interrupted while individual impedance is checked The ground electrode impedance is also measured while injecting current into the REF electrode BIS monitoring is interrupted while the ground impedance is checked The impedance check signal can occasionally interfere with other monitoring equipment connected to the patient Evoked potential monitors are particularly susceptible because they use a wide bandwidth The automatic impedance check feature can be turned off by selecting Impedance Checking OFF in the Diagnostics Menu See Operating Manual for specific instructions 3 4 075 0014 SECTION III PRINCIPLES OF OPERATION 3 2 1 4 The DSC Communications Circuits The communications circuits contain the analog to digital A D converter for each channel the monitor interface the sensor interface and the power supply circuits crystal controlled DSC master clock is on this board This clock is the system s BIS processing clock Conversion Patented technology The communications board contains sigma delta modulators for the two channels These run at 16384 samples per second Test Signal A calibrated test signal is generated on the communications board during DSC s
60. king tabs Figure 7 4 Carefully separate two printed circuit boards 7 7 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM 7 5 2 2 Closing the Case Caution All work involving opening the instrument case must be performed static safe environment to prevent damage to electronic components and assemblies This environment includes the operator work area and tools and any other test or storage items that might touch the DSC assemblies Caution Handle PCBs only by board edges to avoid contamination moisture skin oils etc on PCB and component surfaces 1 Re connect communications and pre amplifier PC boards together carefully align pins and jacks of the two printed circuit boards together press pins firmly into jacks e Verify that no pins are bent or missing their respective jacks 2 Orient and position shield around pigtail cable fasten shield tabs to retain shield shape 3 Carefully slide shield into position over PCB assembly and secure with ground shield securing screw 4 Prepare bottom case by inserting gasket into retaining groove NOTE If gasket is to be reused note that the cable passages have an access cut in them For a new gasket these cable passages need to be carefully cut in one place to allow cable entry See Figure 7 5 5 Position and align PCB assembly with shield into bottom case assembly 6 Align and position both cables into gasket cable passage and strain reliefs into their cu
61. l activity that may be interfering with EEG recognition 3 PIC is defective 4 DSC is defective Note This message may occur as the result of artifact non EEG signal such as those generated from motion patient movement or the presence of electro cautery or other devices 6 10 Corrective Actions 1 If ARTIFACT label appears in EEG waveform box attempt to identify and eliminate artifact source 2 If EMG is present try to determine and eliminate cause 3 Verify Sensor Check passes If not replace PIC 4 Replace DSC 1 If ARTIFACT label appears in EEG waveform box attempt to identify and eliminate artifact source 2 If EMG is present try to determine and eliminate cause 3 Verify Sensor Check passes If not replace PIC 4 Replace DSC SECTION Status Messages E27 SQI Below 50 E28 SQI Too Low for BIS E29 BIS lt Low Alarm Limit E30 BIS gt High Alarm Limit 075 0014 DIAGNOSTICS AND TROUBLESHOOTING Causes The Signal Quality Index is less than 50 and the numeric display changes from a solid to an outlined number This may occur as the result of artifact such as those generated from motion or eyeblinks The Signal Quality Index is unacceptable The Primary Trend variable cannot be calculated and therefore the numeric display is blanked 1 Artifact such as those generated by motion or eyeblinks is causing loss of EEG recognition 2 EMG Bar
62. n and off The left and right gt arrow keys are used to enter Review mode and to scroll data backward and forward in time In the BIS Log they show the previous lt or next gt screen of data 9 MENU EXIT key is used to enter and exit the Setup Menu In Review mode it will return the user to the main screen It is also used to halt certain procedures such as the sensor check and to answer no to a question A x The T and down J arrows are used to move from one menu selection to another They are also used to increment and decrement values In the BIS Log they are used to zoom time out or in increase or decrease reporting intervals 2 1 075 0014 SECTION II A 2000 SYSTEM OVERVIEW The SELECT key is used to move between options within a menu selection to confirm an entry and to answer yes to a question 2 1 1 2 Digital Signal Converter DSC The Digital Signal Converter is about the size of a computer mouse It contains the EEG amplifiers and analog filters The DSC digitizes EEG waveforms for transmission to and processing by the monitor The DSC s long flexible Monitor Interface Cable connects to the front of the monitor and the shorter DSC pigtail cable connects to the Patient Interface Cable PIC The attachment clip on the DSC is used to secure it in a convenient location near the patient s head The patient connection is accomplished by attaching the Aspect BIS
63. ndwidth 075 0014 Noise shaped sigma delta 16 384 samples second 16 Bits at 256 samples second 50 Mohms minimum 0 3 uV RMS 2 0 uV peak to peak 0 25 Hz to 50 Hz 110 dB at 50 60 Hz to earth ground DSC 2 0 16 800 Hz DSC XP 0 16 450 Hz 8 2 SECTION VIII A 2000 SPECIFICATIONS OPTIONS AND ACCESSORIES BIS sensor Patient Interface Cable Plus PIC Plus Pole Clamp for Monitor Monitor Stand Printer TYPE OF PROTECTION AGAINST ELECTRIC SHOCK OF THE SYSTEM Class 1 Equipment in which protection against electric shock does not rely on basic insulation only but which includes an additional safety precaution Means are provided for the connection of the equipment to the protective earth conductor in the fixed wiring of the installation in such a way that accessible metal parts cannot become live in the event of a failure of the basic insulation DEGREE OF PROTECTION AGAINST ELECTRIC SHOCK OF THE SYSTEM Type BF Equipment providing a degree of protection against electric shock regarding allowable leakage currents and reliability of the protective earth ground connection with an F type applied part An F type applied part is isolated from all other parts of the equipment to such a degree that the patient leakage current allowable in single fault condition is not exceeded when a voltage equal to 1 1 times the highest rated AC supply voltage is applied between the applied part and earth DEGREE OF PROTECTION AGA
64. ng loop by fastening to cable strain relief with the tie wrap 7 1 3 To replace the Monitor 7 Power down Remove power cable from its jack in the rear Disconnect the DSC from the front of the monitor To do this pull on the blue rubber overlay DO NOT twist or pull on the gray cable If required dismount monitor from pole clamp by removing set screw from clamp then depressing top of clamp shoe blue plastic clip at top of aluminum pole clamp and slide the monitor up and off of the clamp To install the replacement Monitor insert power cord to jack on rear of monitor To install DSC hold the cylindrical connector with flat side up at 12 o clock position and insert firmly into DSC port on front of monitor Re mount pole clamp assembly 7 2 FUSE REMOVAL AND INSTALLATION The A 2000 contains two inline fuses 1 25A 250V 5x20mm Aspect P N 430 0006 within the AC power input module located at the rear of the monitor To check or replace a fuse one can access these fuses as follows 1 2 3 4 The fuse holder is part of the AC power input module on the rear of the monitor and is located just above the power cord input connector Disconnect the AC power cord from the wall socket and from the rear of the monitor Locate the plastic locking tabs securing the fuse holder in place Insert the tip of a ball point pen or similar tool into the slots provided to release the plastic locking tabs allowing the fuse holder t
65. ngine Comm Error E18 BIS Engine Not Functional E19 E24 See E17 075 0014 Causes Error in communication between BIS Engine and Host Fatal BIS Engine error detected DIAGNOSTICS AND TROUBLESHOOTING 6 9 Corrective Actions 1 Verify the software level Replace the monitor Turn monitor off then on again Verify the software level Replace the monitor SECTION DIAGNOSTICS AND TROUBLESHOOTING EEG Signal Errors 25 and 26 Note These messages relate to the condition of the EEG signal This signal is often affected by environmental conditions that cause artifact non EEG signal The messages below do not usually indicate a monitor system error instead they serve as a notification of a patient or environmental condition Status Messages E25 Left Side SQI Bad E26 Right Side SQI Bad 075 0014 Causes With 2 channel montage left side signal quality is bad 1 Artifact is causing EEG information to be lost 2 EMG Bar indicates electrical activity that may be interfering with EEG recognition 3 PIC is defective 4 DSC is defective Note This message may occur as the result of artifact non EEG signal such as those generated from motion patient movement or the presence of electro cautery or other devices With 2 channel montage right side signal quality is bad 1 Artifact is causing EEG information to be lost 2 EMG Bar indicates electrica
66. niin niania id eatis de ea a ee e tue 9 4 Figure 9 3 BIS ete e hti t eee eerte pte 9 4 Figure 9 4 Connecting electrodes 2 and H4 sese eere 9 5 Figure 9 5 Connecting electrode 3 with 4 and 1 with 882 essere 9 5 Figure 9 6 Safety Tester Contact Points 9 6 ABOUT THIS MANUAL ABOUT THIS MANUAL This manual contains information necessary for the customer to install maintain service identify and prepare for use the Aspect Medical System BIS Monitoring System Also included are directions to diagnose troubleshoot and repair the system A spare parts and accessories list and system specifications are included This manual is intended to be used in combination with the A 2000 Operating Manual The A 2000 BIS Monitoring System is designed and manufactured using state of the art components and manufacturing processes Field repair or customer repairs are therefore limited by design to replacement of major component assemblies such as the Patient Interface Cable PIC Digital Signal Converter DSC or the A 2000 Monitor itself This manual in conjunction with the A 2000 Operating Manual contains the maintenance and diagnostic troubleshooting information necessary for customer qualified technical personnel to test and replace those parts of the equipment that are replaceable by the custome
67. o spring out of its locked position Remove fuse holder Check fuses with an ohm meter If replacement is needed insert correct fuses see below into fuse holder To re install the fuse holder simply insert the fuse holder and press in with finger until the plastic locking tabs engage securing the fuse holder in place WARNING REPLACE FUSES ONLY WITH ONE OF THE FOLLOWING PARTS Aspect P N 430 0006 1 25 Amps 250V 5x20mm Littelfuse 217 Series 1 25 Amps 250V 5x20mm Wickmann 193 Series 1 25 Amps 250V 5x20mm ALWAYS REPLACE BOTH FUSES TOGETHER EVEN IF ONLY ONE HAS FAILED 7 2 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM 7 3 WHAT DO WITH A COMPONENT THAT REQUIRES SERVICE If it becomes necessary to return a major component to Aspect Medical Systems follow the procedure below Obtain return authorization Contact Aspect s Technical Service Department to obtain a returned materials authorization RMA number The Technical Service phone number is printed on the back cover of this manual The RMA number should appear on the outside of the shipping container Please provide the model number serial number and a brief description of the reason for return Use the original shipping container if available or equivalent packaging to protect the product Seal the package with plastic shipping tape rather than masking tape Mark shipping container FRAGILE If the repair or replacement is covered by the warranty
68. of power during a procedure the monitor automatically switches to back up battery operation During battery operation a battery symbol appears in the BIS numeric region of the display A fully charged battery will provide a minimum 20 minutes of operation The thermal printer is disabled during battery operation The battery symbol blinks to alert the user when only a few minutes of back up power remain The Save Settings feature is disabled when the battery symbol is blinking Battery recharge time is approximately 4 hours NOTE The monitor must be powered ON to recharge the battery 3 3 5 Data Transfer and Software Updates Data may be transferred to and from the A 2000 monitor using the Serial port on the back panel Request the Aspect A 2000 Serial Port Specification P N 070 0017 for a complete description of the protocols This port is also used to easily upgrade system software WARNINGS THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE EQUIVALENT SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING SYSTEM CONSIDERATION RELATING TO THE CHOICE SHALL INCLUDE e USE OF THE ACCESSORY IN THE PATIENT VICINITY e EVIDENCE THAT THE SAFETY CERTIFICATION OF THE ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE TO THE APPROPRIATE IEC 601 1 AND OR IEC 601 1 1 HARMONIZED NATIONAL STANDARD 3 8 075 0014 SECTION III PRINCIPLES OF OPERATION WHEN CONNECTING EXTERNAL EQUIPMENT e g DATA CAPTURE COMPUTER
69. ompanying Documents 1 5 SECTION SAFETY PRECAUTIONS Fuse Replace only with same Type and Rating dy Equipotential Alternating Current Dangerous Voltage Protective Earth ground Storage Temperature Limits 1 6 075 0014 SECTION II A 2000 SYSTEM OVERVIEW SECTION II 2 A 2000 SYSTEM OVERVIEW 2 1 INTRODUCING THE A 2000 BIS MONITORING SYSTEM Aspect Medical Systems A 2000 BIS Monitoring System is intended to monitor the state of the brain by data acquisition of EEG signals in the intensive care unit operating room and for clinical research The BIS a processed EEG variable may be used as an aid in monitoring the effects of certain anesthetic agents The A 2000 processes raw EEG signals to produce a single number called the Bispectral Index or BIS which correlates to the patient s level of hypnosis It operates from an AC power source of 100V to 240V 50 60Hz and provides a minimum 20 minutes of automatic back up battery power The monitor is menu driven with fixed keys for choosing the options available 2 1 1 Principal Components The system is composed of a monitor and a digital signal converter with a patient interface cable PIC and BIS Sensor 2 1 1 1 Monitor The monitor contains the operator control panel an electro luminescent display screen and connectors for the digital signal converter and printer Front Panel Controls LA The SILENCE key toggles audible alarms o
70. or application 2 Poor sensor connections Sensor Check fails Defective PIC Defective DSC Disconnected sensor Poor or contaminated connection between sensor and PIC Disconnected PIC Defective PIC Defective DSC Defective monitor N 1 Atleast one element of sensor has too high impedance and EXIT pressed before sensor check completes 2 Incorrect sensor application 3 Poor sensor connection 4 Defective PIC 5 Defective DSC 6 8 Corrective Actions 1 Isolate BIS system from possible noise sources or restart monitor when electro cautery equipment not in use Note that self test runs when the DSC is first connected 2 Replace the DSC 3 Replace monitor 1 Connect clean connection between sensor and PIC 2 Replace the sensor 3 Replace the PIC 4 Replace the DSC 1 Read Instructions on sensor package and re apply 2 Check sensor connections 3 Re prep replace sensor Verify Sensor Check passes 4 Replace the PIC 5 Replace DSC Connect the Sensor Connect clean connection between sensor and PIC Connect the PIC Replace the PIC Replace the DSC Replace the monitor N 1 Verify Sensor Check passes 2 Read Instructions on sensor package and re apply sensor 3 Check sensor connection 4 Replace the PIC 5 Replace the DSC SECTION Status Messages E17 E19 E20 E21 E22 E23 E24 BIS E
71. ow the monitor to stabilize in the unopened shipping container at the inside ambient temperature before unpacking and placing into service Before operation wipe down all visible condensation and allow the system to reach equilibrium at room temperature 4 1 2 Operating Environment The A 2000 is not designed for use in areas containing flammable gases or vapors WARNING EXPLOSION HAZARD DO NOT USE THE A 2000 IN A FLAMMABLE ATMOSPHERE OR WHERE CONCENTRATIONS OF FLAMMABLE ANESTHETICS MAY OCCUR Temperature The Aspect A 2000 Monitor is designed to operate safely at a room temperature of 5 degrees C to 40 degrees C Conditions that exceed these limits could affect reliability Humidity The monitor is designed to operate within specifications at a relative non condensing humidity of 15 to 95 Pressure The monitor will operate satisfactorily at or above sea level and is unaffected by extremes or changes in altitude within atmospheric pressures of 360 mmHg to 800 mmHg 4 1 075 0014 SECTION IV PREPARATION FOR USE AND INSTALLATION 4 1 3 Power Requirements and System Grounding The A 2000 BIS Monitor requires a power source of 100 240 VAC 50 60Hz Current consumption is 1 ampere maximum including printer load To protect operating personnel and patients the monitor must be properly grounded Accordingly the monitor is equipped with a hospital grade line cord The power cord grounds the system to the power line ground when plugged
72. pare parts and accessories Appendix provides system diagrams vi ABOUT THIS MANUAL SECTION SAFETY PRECAUTIONS SECTION 1 SAFETY PRECAUTIONS INTRODUCTION Caution Carefully read the A 2000 BIS Monitoring System Operating Manual entirely before using the monitor in a clinical setting WARNINGS CAUTIONS AND NOTES The terms warning caution and note have specific meanings in this manual AWARNING advises against certain actions or situations that could result in personal injury or death CAUTION advises against actions or situations that could damage equipment produce inaccurate data or invalidate a procedure although personal injury is unlikely e ANOTE provides useful information regarding a function or procedure KEY TO SYMBOLS A key to the symbols used on the A 2000 appears at the end of this section 11 WARNINGS GROUND WIRE LEAKAGE CURRENT MUST BE CHECKED WHENEVER INSTRUMENT CASE IS OPENED BY A QUALIFIED BIOMEDICAL ENGINEERING TECHNICIAN EXPLOSION HAZARD DO NOT USE THE A 2000 IN FLAMMABLE ATMOSPHERE OR WHERE CONCENTRATIONS OF FLAMMABLE ANESTHETICS MAY OCCUR ELECTRICAL SHOCK HAZARD DO NOT REMOVE MONITOR COVERS DURING OPERATION OR WHILE POWER IS CONNECTED TO MONITOR ELECTRICAL SHOCK HAZARD THE MANUFACTURER S INSPECTION OF THIS APPARATUS VERIFIED THAT THE GROUND LEAKAGE CURRENT AND THE PATIENT SAFETY CURRENT WERE LESS THAN THE SPECIFIED LIMITS ESTABLISHED BY THE APPLICABLE SAFETY ST
73. ples them calculates the processed parameters and outputs the data display 3 2 2 1 The Main Board The Main board contains the Digital Signal Processor the Host processor FPGA field programmable gate array data memory real time clock and DSC interface 3 2 2 2 Digital Signal Processor The functions of the DSP include decimation filters for up to 2 channels of EEG data computation of processed EEG parameters control of the DSC and interface with the host 3 2 2 3 Host Processor The functions of the host processor include control of the display interface with the keypad serial port and printer control of monitor boot up and download control of memory An on board annunciator is included for generating alarm sounds The display interface is a 4 bit LCD type The serial port is standard RS 232 This port is used to communicate between the A 2000 and an attached device The device may be a data capture computer PC or a software upgrade device Software and firmware can be downloaded through this serial port The printer interface is a non standard serial type with extra lines for printer keys and printer control and status It is designed to interface with a Seiko printer interface board 3 2 2 4 The Digital Signal Converter DSC Interface The DSC interface is composed of two unidirectional bi phase encoded serial lines one going to the DSC and another bringing data from the DSC The power to the interface is under
74. ponents the monitor the digital signal converter DSC and the Patient Interface Cable PIC with BIS Sensor The A 2000 monitor contains the circuits for monitor control digitally processing the EEG data computing the processed parameters and displaying the waveforms and processed parameters The circuits to acquire and digitize the EEG signals reside in the Digital Signal Converter DSC The PIC and Sensor are the patient connection for EEG signal acquisition A block diagram depicting the monitor subassemblies appears in Figure 3 1 A signal flow diagram appears in Figure 3 2 3 1 075 0014 SECTION III PRINCIPLES OF OPERATION PREAMP BOARD Patient Power Printer Connector Port J3 J27 INTERCONNECT BOARD POWER SUPPLY BOARD MAIN PROCESSOR BOARD BATTERY KEYBOARD These connectors are internal to the board and are mated as shown DISPLAY MONITOR Figure 3 1 The A 2000 System Block Diagram 075 0014 SECTION III PRINCIPLES OF OPERATION Posi wiwa um lt lt EE en i ue us KE n m em mine em em eremum eu m ee m Gn min 3 Input d Filter SIGMA DELTA Protect Modulator Input gt Filter SIGMA DELTA 2 Protect AMP Modulator Patient Connecti Digital Signal Processor DSP CPU DISPLAY
75. prep technology and uses a proprietary connector Use of other electrodes is not recommended Continuous impedance checking may need to be disabled if the 1 nanoampere 128 Hz impedance check signal interferes with other equipment e g evoked potential monitors Check the battery periodically by operating an A 2000 that has been disconnected from the wall socket and that has been charging the battery for at least 4 hours After long periods of storage e g more than 1 month it may be necessary to cycle charge then discharge the battery a few times to get full charge capacity If the A 2000 fails to operate reliably from the battery for 20 minutes battery replacement is required The A 2000 contains an internal Nickel Metal Hydride battery The battery must be removed by a qualified service technician and disposed of or recycled in accordance with the national laws of the country Contact Aspect Medical Systems Inc or the local distributor for servicing of battery Avoid liquid ingress to the PIC Contact of fluids with the PIC sensor connectors can interfere with PIC performance 1 3 075 0014 SECTION SAFETY PRECAUTIONS The A 2000 system complies with the electromagnetic compatibility requirements of EN60601 1 2 Operation of this device may affect or be affected by other equipment in the vicinity due to electromagnetic interference If this occurs e Increase separation between devices e Re orient device cabling
76. r Aspect does not authorize nor provide information to service or repair the internal components of the A 2000 Monitor Section discusses important safety precautions Before attempting to set up or service the A 2000 please familiarize yourself with the safety information provided in this section Section provides a functional overview of the A 2000 its principal components and instrument identification Section describes the A 2000 monitor hardware and how it operates Section IV provides preparation for use and installation instructions including environmental considerations instrument connections and system setup and check out Section V describes normal maintenance care and cleaning procedures Section VI describes the A 2000 diagnostic tools and provides tables to aid in troubleshooting the system Section VII describes the removal and replacement procedure for system components and how to return a component for service It also explains how to disassemble and reassemble the DSC to replace cables Note This manual was not designed for repairing the A 2000 BIS system to the level of its electronic circuit boards The manual is intended solely for the troubleshooting and replacement of its major components or DSC cables Section VIII contains the A 2000 specifications and warranty Appendix l provides instructions for use for the Sensor Simulator and describes how to make a test sensor It also contains a list of s
77. s This test will not be performed while the monitor is operating on battery backup 3 7 075 0014 SECTION III PRINCIPLES OF OPERATION Impedance Check Electrode impedance is tested when the DSC and PIC are connected and is monitored continuously unless the user has turned impedance checking off in the Diagnostic Menu Caution Continuous impedance checking may need to be disabled if the 1 nanoampere 128 Hz impedance check signal interferes with other equipment e g evoked potential monitors Diagnostics The A 2000 provides diagnostic codes to assist the user in tracing the source of many problems that may occur Codes are displayed in the right corner of the message region only if the user has selected them in the Diagnostic Menu 3 3 2 Data Memory The monitor stores recorded trend data with time and date of acquisition The duration of trend data stored is approximately 11 12 hours Trend memory can be viewed on the screen in 1 hour segments using the Review arrow keys lt gt to scroll through memory When the memory is full the oldest data are automatically erased as new data are stored Memory will be retained even if the battery has been discharged and remains when the monitor is in the power off condition 3 3 3 Saved Settings The A 2000 monitor will always power up configured to the settings that have been saved in the Advanced Setup Menu 3 3 4 Battery Operation In the event of a power failure or interruption
78. s a normal condition when Sensor Simulator or Test Sensor is connected Data memory has been re initialized The AC power has been lost and the monitor is running on the battery The battery keeps the monitor operating for approximately 20 minutes when the battery is fully charged There are only a few minutes of battery life left A software error has occurred The monitor may stop operating Error in erasing or writing Setup block of FLASH memory Error in writing to Real Time Clock RAM Monitor is unable to write to Setup Memory when battery power is low 6 12 Corrective Actions 1 2 Geese Check the patient Check leads for proper connection and possible shorts Verify Sensor Check passes Verify DSC Self test Verify PIC Use Test Sensor or Sensor Simulator and Sensor Check Turn the monitor off then on again If error persists monitor requires service Restore the AC power Verify power cord Verify AC fuses are good Replace monitor Restore AC power to avoid automatic shutdown 1 2 3 Turn the monitor off then on again Verify the software level Replace monitor 1 Check Setup 2 Replace monitor Replace monitor 2 3 Restore power Save settings again Replace monitor SECTION Status Messages E45 No Updates from BIS Engine E46 BIS Engine Not Responding E47 E74 E81 Data Memory Write Error E48 E65
79. se Holder Potential Equalization Terminal AC Line Printer Connection Port Figure 2 2 Rear View of Monitor WARNING REPLACE FUSES ONLY WITH ONE OF THE FOLLOWING PARTS Aspect P N 430 0006 1 25 Amps 250V 5x20mm Littelfuse 217 Series 1 25 Amps 250V 5x20mm Wickmann 193 Series 1 25 Amps 250V 5x20mm ALWAYS REPLACE BOTH FUSES TOGETHER EVEN IF ONLY ONE HAS FAILED 24 075 0014 SECTION III PRINCIPLES OF OPERATION SECTION Ill 3 PRINCIPLES OF OPERATION INTRODUCTION This section includes e How the A 2000 BIS Monitoring System works e The architecture of the A 2000 monitor and DSC e System Features 3 1 HOW THE A 2000 MONITORING SYSTEM WORKS Sensors placed on the patient s head transmit EEG signals to the Digital Signal Converter The DSC amplifies and digitizes these signals then sends them to the monitor The monitor software filters the data analyzes it for artifact and processes it using digital signal processing techniques The purpose of processing the EEG waveform data is to extract characteristic features from the complex signal in order to provide easier pattern recognition of changes over time during the recording The A 2000 BIS Monitoring System consists of e The BIS monitor with built in battery backup and detachable power cord e The Digital Signal Converter DSC e Aspects BIS Sensor Patient Interface Cable PIC and BIS Sensor 3 2 SYSTEM ARCHITECTURE Hardware is divided into three main com
80. side the connector may affect DSC performance 5 1 2 Disinfecting the Monitor and Digital Signal Converter Use lint free absorbent towels dampened with 10 bleach solution or a commercial disinfectant e g Lysol Professional Disinfectant Foam Cleaner Spray or PDI Germicidal Disposable Wipes After cleaning dry all areas except the monitor display screen see below with a lint free absorbent paper towel Wipe the PIC and DSC pigtail connector ends with alcohol and allow to dry completely WARNING WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS OCCURS RE TEST GROUND LEAKAGE CURRENT BEFORE FURTHER USE DO NOT MIX DISINFECTING SOLUTIONS e g BLEACH AND AMMONIA AS HAZARDOUS GASES MAY RESULT Caution Do not autoclave the Digital Signal Converter or Monitor Autoclaving will seriously damage both components Avoid liquid ingress to the PIC Contact of fluids with the PIC sensor connectors can interfere with PIC performance 5 1 3 Cleaning the Monitor Display Clean the monitor display screen with a mild solution of detergent and warm water or a commercial display screen cleaner available through personal computer dealers Clean with a moist cloth wring out cloth first To avoid scratching the screen never use abrasive cleaners 5 1 075 0014 SECTION V PREVENTIVE MAINTENANCE CARE AND CLEANING 5 2 PREVENTIVE MAINTENANCE The A 2000 monitor is designed so that no periodic adjustment or calibration is required
81. st circuits and parts of the sensor ID circuit 3 2 1 2 DSC Signal Conditioning The input protection circuits are designed to protect the inputs from destruction by electric shock from sources such as electrostatic discharge ESD or defibrillation The protection circuits also reduce the effects of high frequency ambient noise from sources such as electro cautery and other devices The montage mode switches follow the protection circuits These switches configure the inputs appropriately for the sensor The signals are amplified by instrumentation amplifiers which have a fixed gain The amplifiers have DC servos which remove the signals below high pass cutoff frequency In the event of amplifier overload the servos are changed to a higher frequency to facilitate fast recovery blocking under control of the host processor Each channel is further amplified to the level required by the A D converters The amplifiers also serve as filters to prevent aliasing by the converters 3 2 1 3 DSC Impedance Testing In the default state of the DSC the combined channel electrodes impedance is continuously checked A small current approximately 1 nanoampere is injected into each electrode at 128 Hz just above the EEG band The resulting voltages are measured Equal but opposite currents are injected into the and electrodes simultaneously while the digital signal processor measures the resulting voltage BIS monitoring is performed while comb
82. status PASS An electrode passes if the impedance for that electrode is less than 7 5 kilo ohms The ground electrode must be less than 100 kilo ohms to pass HIGH An electrode is labeled HIGH if its impedance value is above 7 5 kilo ohms As long as the combined impedance is less than 15 kilo ohms the Sensor Check will be considered successful If the combined impedance is over the 15 kilo ohms limit you will need to re prep the electrodes and check all connections The monitor will continue to check impedance until it is acceptable NOISE If the signal from the electrode goes beyond the measurable range the label NOISE will appear LDOFF If a lead is too loose or disconnected the label LDOFF will appear NOTE Sensor Check is used in clinical application as an indicator of the patient s skin conductivity When used with a Sensor Simulator or Test Sensor it serves the purpose of verifying the DSC and PIC cable conductors Therefore the status indications called out above need to be interpreted as conditions relating to the DSC and PIC ability to conduct the sensor check signal Values that are too high must be investigated as possible failures to be repaired Expected Impedance Values Sensor Simulator Values Test Sensor Values Electrode 8 Typical Range Electrode Typical Range 1 5 K Ohms 4 6 K Ohms 1 1 K Ohms 1 2 K Ohms 2 10 K Ohms 8 17 K Ohms 2 1 K Ohms 1 3 K Ohms 4 4 K Ohms 3 5 K Ohms 4 1 K
83. touts See Figure 7 6 7 Carefully place the top case and bottom case together e Verify that the sealing gasket is fully into its recess e Verify that no cables are trapped at the joint e Verify that both strain reliefs are fully secured in their recesses 8 Install the two screws into the corners of the case and tighten 9 Perform DSC Checkout and Safety tests See section 7 5 4 Figure 7 5 For a new gasket cable passages need to be carefully cut 7 8 075 0014 SECTION VII SERVICING THE A 2000 SYSTEM Figure 7 6 Verify that the strain relief is fully secured in its recess 7 5 3 The DSC Cables Monitor and Pigtail 7 5 3 1 Removing the Cable Caution Handle PCBs only by board edges to avoid contamination moisture skin oils etc on PCB and component surfaces Open case according to Opening the Case section 7 4 2 1 2 Remove cable by releasing the connector locking tab and pulling straight off of PCB See Figure 7 4 to locate locking tabs e Note that the Monitor cable is connected to the communications PCB e Note that the Pigtail cable is connected to the pre amplifier PCB 7 5 3 2 Installing the Cable Caution Handle PCBs only by board edges to avoid contamination moisture skin oils etc on PCB and component surfaces 1 Install cable by aligning the cable connector to jack and carefully pressing connector into jack Verify that the locking tab engages 2 Close case according to Closing