ARGUS 600 Infusion Pump Service Manual
Contents
1. 189 9 Bloc schematic Argus 08 12 04 PJ 30 14 135 E A600D en SM V1 51 14 135 E A600D en SM V1 51 Argus 08 12 04 PJ 31
2. Switch the pump on while keeping the key 1 pressed Multiply the value in the right display by 12 this gives the capacity of the battery in minutes Multiply this time by 0 8 and enter the result on address 368 in the configuration mode This time defines from now on the running time of the pump including a 15 minutes pre alarm valid after a full charge Standard battery 6V 1 2 Ah If this time is less than 2 hrs you should replace the battery part 12 032 If the specified time gt 2 hrs is not required the battery has to be changed only if the time less than 1 5 hrs to respect to environmental pollution High energie battery 6V 4 Ah If this time is less than 8 hrs you should replace the battery part 12 026 If the specified time gt 8 hrs is not required the battery has to be changed only if the time less than 5 hrs to respect to environmental pollution 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ 22 5 Safety standard check Safety standard check Serial no Hospital Dept Customer ARGUS 600 ARGUS Medical AG The safety standard check has to be performed at least every 24 months or after 10000 hours of operation The check has to be done in accordance to the operation and service manuals 1 Visual check for damage cleanness and completeness 2 Keep key MODE pressed vvhile svvitching on the pump 3 Hold the barrel svvitch pressed and move the lever for clamp bottom
3. FrES 1n50 FrES PE50 IVAC 50 JMS 50 Mono 50 PIC 50 PIC PE50 ryco 50 tEru 50 dIPr 50 OnCE 50 brn Pr50 Adress Write Default Function prot value 200 0 Medication enable value 0 8191 201 0 Medication enable value 0 8191 202 0 Medication enable value 0 8191 203 0 Medication enable value 0 8191 204 0 Medication enable value 0 8191 205 0 Medication enable value 0 8191 206 0 Medication enable value 0 8191 207 0 User medication enable value 0 8191 208 0 User medication enable value 0 7 220 5383 Display digit 1 amp 2 user medication 1 0 9999 221 6982 Display digit 3 amp 4 user medication 1 0 9999 222 4500 Display digit 5 amp 6 user medication 1 0 9999 223 17 Display digit 7 amp 8 user medication 1 0 9999 224 5383 Display digit 1 amp 2 user medication 2 0 9999 225 6982 Display digit 3 amp 4 user medication 2 0 9999 226 4500 Display digit 5 amp 6 user medication 2 0 9999 227 18 Display digit 7 amp 8 user medication 2 0 9999 228 5383 Display digit 1 amp 2 user medication 3 0 9999 229 6982 Display digit 3 amp 4 user medication 3 0 9999 230 4500 Display digit 5 amp 6 user medication 3 0 9999 231 19 Display digit 7 amp 8 user medication 3 0 9999 232 5383 Display digit 1 amp 2 user medication 4 0 9999 233 6982 Display digit 3 z 4 user medication 4 0 9999 234 4500 Display digit 5 z 6 user medication 4 0 9999 23
4. 30 008 14 135 E A600D en SM V1 51 VViring diagram Argus 08 12 04 PJ oh fuoneg bu A seoN dua uolsniooO azis duel p UOUMS von I azis uosod B6uu s p s nssald B5uu sS O 011U0 AS p HOA neg 011U0 A0Z p Aouanbau4 m d MOS a a om gt NOHd33 NOHda yore Janug a Ae ds q HO UO OAM z BBBB BBBE ero m gt gt Pressure gt gt gt oo oe D Bolus Start Stop Mode 1 1 10 100 041U09 a eyon Alsyeg 0141109 au 21607 OG NG 4 Weq qd A9 oq Jozzng 1010 oq AZ lt oq oq OV OG PI 159 Jamod 09 UIS Jenug JOJON 18594 uollg os a urew OV Janug Ke o m o Key wey f Jenud Jezzng
5. 600mbar 0x0000 Syringe barrel pump stop Syringe clutch pump stop Syringe empty pump stop Syringe near empty Syringe switch pump stop Timer alarm pump stop KVO Total volume reached pump stop KVO Datalock on Datalock off Pump start ext changed parameter Argus 08 12 04 PJ 14 24 Monitoring of the ARGUS 600 Caution The monitoring of the infusion pump ARGUS 600 over the serial interface of a PC is intended for demonstration purposes only any connection with patients has not been tested under the conditions of EN 60601 1 and are not allowed Switch the pump on with an inserted filled infusion set Enter one of the following command directly in your terminal window or transmit the corresponding ASCH code over your own monitoring program A short sample of a monitoring session is mentioned at the end of this chapter Command Keystrokes in terminal ASCII code Description ENQ Ctrl E 05H Send status see format below SO Ctrl N OEH Sets pump in remote mode STX Ctrl B 02H Start of data entry see format below data Data Data ETX Ctrl C 03H End of data entry DC2 Ctrl R 12H Start infusion DC4 Ctrl T 14H Stop infusion SI Ctrl 0 OFH Sets pump in local mode ACK Ctrl F 06H Alarm suppression 2min FS Ctrl ICH Toggle KVO mode BEL Ctrl G 07H Toggle Buzzer at start mode CAN Ctrl X 18H No start test s at next start ESC Ctrl 1BH Next character following ESC Ctrl will select
6. 14 135 E A600D en SM V1 51 Carrier 10 066 10 074 11 160 11 161 11 162 11 164 11 172 11 173 11 174 12 018 12 019 12 020 14 135 E A600D en SM V1 51 Spindle nut compl Flex board with cable Sleeve in front Sleeve in back Wing Tension spring Housing Carrier Cover Carrier Cog wheel 23 c PCB micro switch Flex cable Potentiometer 10 turns 10K Argus 08 12 04 PJ 27 ured Buds SUOISJO 161711 Z9L HL SEGL LL ver LL peay 49109 peay SuisnoH Buuds dul ews J A UOINSOd red Buuds uoIsJop 1919 UOM syeag JO 18197 sddn 1uawBas 60 Jepun jusw6as 60 woyoq dweid dor durero buueag ad d adid uonoc uuoo elo Ole LL 60c LI 800 LL Lek 18 981 YEL ES c8 i II 841 19 E 9 89 LL LL LL LL LL LL LL LL LL LL LL peay 28 Argus 08 12 04 PJ 14 135 E A600D en SM V1 51 AC Input w X4 qa DMS L N Ai x2 1 X5 X16 30 010 Barrel 30 011 X3 Nurse Ext DC ir p X12 X15 1 Mm 4 EPROM X10 30 007
7. 340 370 10 361 500 362 2 363 10 365 9 367 3 368 120 369 5 370 371 372 374 375 376 380 381 382 383 384 385 386 387 388 389 PG PG PP PAPE PEPE 390 0 391 392 393 394 395 396 397 399 X 600 xx 56 6 o 14 135 E A600D en SM V1 51 Key ON OFF delay in ms additional key STOP if 4 1 Display brightness Buzzer alarm volume Pressure alarm level n 100mbar Time for near empty alarm Battery discharge time min Automatic menu fall back delay time Clock seconds Clock minutes Clock hours Clock days Clock months Clock years Last failure number F XX Last infusion rate at failure last failure number F XX last infusion rate at failure last failure number F XX last infusion rate at failure last failure number F XX last infusion rate at failure last failure number F XX last infusion rate at failure u O O LNS Last Service date year and week 2 last service date 3 last service date Service interval in months 1 24 0 disabled Service interval in hours 1 9999 0 disabled Own adress for SCI 0 no adress or 1 127 Inventory no of the pump xxxx yyyy Inventory no of the pump xxxx yyyy Data xxxx gt clears protection key Argus 08 12 04 PJ 0 3000 123 5 10 2 12 1 15 min 45 615 5 305 00 59 00 59 00 23 01 31 01 12 2000 2099 yyww 0 127 xxxx sand 1 yyyyl 11 1 7
8. amp 8 user medication 10 0 9999 260 5383 Display digit 1 amp 2 user medication 11 0 9999 261 6982 Display digit 3 amp 4 user medication 11 0 9999 262 4500 Display digit 5 amp 6 user medication 11 0 9999 263 1717 Display digit 7 amp 8 user medication 11 0 9999 264 5383 Display digit 1 amp 2 user medication 12 0 9999 265 6982 Display digit 3 amp 4 user medication 12 0 9999 266 4500 Display digit 5 amp 6 user medication 12 0 9999 267 1718 Display digit 7 amp 8 user medication 12 0 9999 268 5383 Display digit 1 amp 2 user medication 13 0 9999 269 6982 Display digit 3 amp 4 user medication 13 0 9999 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ Adress 270 271 272 273 274 275 276 277 278 219 280 281 282 283 306 307 308 309 310 311 312 313 315 316 317 318 320 321 322 323 325 326 327 328 330 331 332 333 335 336 337 338 340 341 342 343 345 346 347 348 Write Default prot gt gt lt gt lt 10 ml 20 ml 30 ml 50 ml value 4500 1719 5383 6982 4500 1720 5383 6982 4500 1721 5383 6982 4500 1722 3000 3000 3000 10 610 160 162 187 5000 5000 5000 20 695 168 214 238 5000 5000 5000 30 820 160 240 260 9999 9999 9999 50 900 165 294 310 14 135 E A600D en SM V1 51 Function Display digit 5 amp 6 user medication 13 0 9999 Display digit 7 amp 8 user medicatio
9. bar Note To speed up the process increase the infusion rate in steps by pressing the key 1 Svvitch the pump off and on again Remove the pressure in the system by opening the line Make a control measurement for a pressure of 0 9 bar To do this select 900 mbar in the function PrL The pressure alarm should be released within amp 200 mbar Calibration of the battery capacity Each battery is subject to a chemical process with a slowly decreasing running time After many charge and discharge cycles the battery may not have the capacity to provide the running time shown in the menu CAP To adjust the running time of the used battery please follow the steps mentioned below Go in the configuration mode of the pump see chapters 1 3 or 1 5 Select address 368 in the left display Enter the data 615 in the right display and press the START STOP key to accept the data This will set the battery discharge time to the maximum of gt 10 hours Switch the pump off Be sure you have unplugged the line connection Switch the pump on and run the pump on battery until it switches off Charge the battery for more than 16 hours Switch on the pump and start an infusion with a rate of 5 ml h The infused sum at this rate multiplied by 12 is now equal to battery operating time in minutes Leave the pump running on battery until it switches off again Connect pump to the AC line 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ 21
10. strain gauge must be executed to guarantee a perfect pressure monitoring Replacement of the EPROM or main board Software updates may reset automatically the configured values in the EEPROM You are reguested to upload and save or write down the present contents of the none write protected addresses before you replace the EPROM or the main board Afterwards you have to re enter these values in the program mode or download your saved file If a PIN code has been used before the same code has to be programmed again Charge the batterie for more than 16 hours Note A pressure sensor calibration is necessary when replacing the pressure sensor a pressure sensor calibration and a volume calibration are necessary when replacing the EEPROM Check the pump accuracy and the pressure alarm level a Check the pump accuracy Select the syringe type Cod 50 Insert a new syringe Codan 50ml filled with distilled water and pump into a cup Pump settings set rate at 200 ml h set total at 20 ml Net weight result 20 g 2 Check the pressure alarm level Adjust the pressure alarm limit to 900mbar 90kPa Insert a syringe filled with water and connect the tube on the patient side to a manometer 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ 18 Set rate of 100 ml h start and pump till the occlusion alarm goes on and immediately read the pressure on the manometer 900mbar 90kPa 200mbar 4 3 Rough alignments Plunger length P Go i
11. the pole clamp at the rear side Remove 7 screws at the bottom side 6x M4 and 1 x M3 the 2 screws at the left side cannot be removed completely Place the upper part behind the bottom Remove the main board Remove the battery connector and all other cables of the main board Remove the syringe drive Remove the fixing plate and syringe guide Unsolder the connecting leads of the strain gauge DMS on the power board Move the drive head to the left and remove the syringe drive out from the housing Remove the driving head Remove the 3 screws on the cover Important To disassemble the unit open the cogs by hand one third or put a coin between the cogs then pull the cover with the levers out of the housing Remove the power board Important Removing the power board reguires a new basic alignment Unsolder the connecting leads of the strain gauge on the power board Remove all cables from the board Insert the power board Syringe holder must be in the closed position no syringe inserted Loosen the lock screw of the cogwheel on the syringe holder potentiometer axle Fix the power board with the 4 screws Note that the lock screw is accessible from above Solder the connecting leads of the strain gauge and connect the other cables Make sure to remove the AC power cord and operate the pump with open housing on battery power Replace the sidewall motor After each disassembly or replacing of the sidewall the basic alignment of the
12. the pump with address addr if more than one is connected to the serial interface DC1 Ctrl Q 11H Enguiry of inventory number address 396 397 addr 0 127 Address must be the same as in the pump configuration on address 395 Only valid in remote mode Format of data entry STX izi H 4 1 2 Rate 1 1 F Rate 1 0 N 1 1 Total 1 2 Total 181 N Total LEO Total 1 1 Med hex hi digit Med hex lo digit Format of status which will be returned by the pump after typing Ctrl E in the terminal STX 1 0 0 o o o N o o o m o izi Pc Total 1E2 Total 1E1 Total 1 0 Total 1E 1 PrL 1E3 PrL IE2 PrL 1E1 PrL 1 0 Med hex hi digit Infused sum 1E2 Rate 1E0 Infused sum 1E1 Rate 1E2 Rate 1 1 Rate 1E 1 Infused sum 1E0 Med hex lo digit O Statusbyte 1 Statusbyte 2 Statusbyte 3 2 Statusbyte 4 Statusbyte 5 E Infused sum 1E 1 14 135 E A600D en SM V1 51 Argus 08 12 04 PJ Format statusbyte 2 Format statusbyte I P 1 X X X X X X P 1 X X X X X X Aduro seou Suu doys dund Adwa dos dund yoynjo Suri do s dund durep 98uu 5 do s dund y rreq Sun S do s dund uo snp9O ysty SARATV MO SAEMIY LIe g dos dund Aroneg unejeoid mo Awg An r Alone g RIS je Jozzng 009V 1 dA dung ysy SAPA V MO SAEMIY Format statusbyte 4 Forma
13. 5 20 Display digit 7 z 8 user medication 4 0 9999 236 5383 Display digit 1 amp 2 user medication 5 0 9999 237 6982 Display digit 3 amp 4 user medication 5 0 9999 238 4500 Display digit 5 amp 6 user medication 5 0 9999 239 21 Display digit 7 amp 8 user medication 5 0 9999 240 5383 Display digit 1 amp 2 user medication 6 0 9999 241 6982 Display digit 3 amp 4 user medication 6 0 9999 242 4500 Display digit 5 amp 6 user medication 6 0 9999 243 22 Display digit 7 amp 8 user medication 6 0 9999 244 5383 Display digit 1 amp 2 user medication 7 0 9999 245 6982 Display digit 3 amp 4 user medication 7 0 9999 246 4500 Display digit 5 amp 6 user medication 7 0 9999 247 23 Display digit 7 amp 8 user medication 7 0 9999 248 5383 Display digit 1 amp 2 user medication 8 0 9999 249 6982 Display digit 3 amp 4 user medication 8 0 9999 250 4500 Display digit 5 amp 6 user medication 8 0 9999 251 24 Display digit 7 amp 8 user medication 8 0 9999 252 5383 Display digit 1 amp 2 user medication 9 0 9999 253 6982 Display digit 3 amp 4 user medication 9 0 9999 254 4500 Display digit 5 amp 6 user medication 9 0 9999 255 25 Display digit 7 amp 8 user medication 9 0 9999 256 5383 Display digit 1 amp 2 user medication 10 0 9999 257 6982 Display digit 3 amp 4 user medication 10 0 9999 258 4500 Display digit 5 amp 6 user medication 10 0 9999 259 1716 Display digit 7
14. 8 RTC real time clock data invalid Mainboard F 29 Stepper motor power test Mainboard F 30 Plunger position calculation failed F 31 Check for near empty F 32 5Volt supply out of range Mainboard F 33 20Volt supply out of range Mainboard F 34 Pressure reference out of range LM385 2 5V Mainboard F 35 Pressure signal out of range F 36 Pressure result invalid Mainboard F 37 Pressure sensor test failed F 38 Barrel diameter signal test failed F 39 Barrel diameter signal out of range F 40 Clamp diameter signal out of range F 44 45 Address invalid for EEPROM Mainboard F 46 Frequency from uC or RTC out of range Mainboard F 47 Displayboard not present F 48 Key s too long active Displayboard F 54 Movement result invalid F 55 Freguency calculation Mainboard F 56 59 Volume control Mainboard Exceptionally a fault code may appear which is not included in this list In this case we recommend to change the main board 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ 41 42 Replacement of parts Disassembly of the pump NOTE The exploded views in the appendix show the individual assembling steps CAUTION Switch the unit off and disconnect the mains cable from the power outlet before opening the housing Observe the antistatic protection rules when disassembling the ARGUS infusion pump the use of an anti static table mat and a grounded clip are recommended for example Mind the battery voltage Disassembly of the housing Remove
15. Clear and continue 15 0 User 10ml B D Plastipak 10ml Braun Omnifix 10ml Codan 10ml Fresenius Injectomat 10ml Sheerwood Monoject 10ml ONCE 10ml PIC Indolor 10ml Rymco 10ml Terumo 10ml Braun Injekt 10ml 43 0 Chirana Prema 10 ml User 20ml B D Plastipak 20ml Braun Omnifix 20ml Codan 20ml Sheerwood Monoject 20ml ONCE 20ml Braun Perfusor 20ml Braun Injekt 20ml Chirana Prema 20ml Terumo 20 ml User 30ml B D Plastipak 30ml Codan 30ml ONCE 30ml Braun Omnifix 30ml Terumo 30 ml User 50ml B D Perfusion 50ml B D Plastipak 50 60ml Braun Omnifix 50 60ml Braun Perfusor 50ml Chirana Prema 50 60ml Codan 50ml Codan Perfusion 50ml Dispomed 50 60ml Dispomed Perfusion 50ml Fresenius Injectomat 50 60ml Fresenius Perfusion 50 60ml Ivac 50 60ml JMS 50 60ml Sheerwood Monoject 50 60ml PIC Indolor 50ml PIC Indolor Perfusion 50ml Rymco 50ml Terumo 50 60ml Disoprivan 50ml ZENECA ONCE 50ml Braun Proinjekt 50ml Argus 08 12 04 PJ USEr 10 b d PL10 brn OF101 Cod 10 FrES In10 Mono 10 OnCE 10 PIC 10 ryco 10 tEru 10 brn In10 Chir 10 USEr 20 b d PL20 brn OF20 Cod 20 Mono 20 OnCE 20 brn PE20 brn In20 Chir 20 tEru 20 USEr 30 b d PL30 Cod 30 OnCE 30 brn OF30 tEru 30 USEr 50 b d PE50 b d PL50 brn OF501 brn PE50 Chir 50 Cod 50 Cod PE501 dISP 50 dISP PE50
16. Medication list To display medication names address 29 menu MEd must be set to 1 Using the special function MEd the following medication names can be displayed Refer to the EXCEL file supplied by ARGUS or your local distributor to determine medication enable values adress 200 208 DEC medication number o kb SS A wN Pop 950995 F 00 aw 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 HEX medication number mn 02 xo Oo 1 Oy Ot Pbb bb bb p 0 OT S Q N O T H Medication name Med N Actilyse Adrena0 1 Adrena0 2 Ajmalin Alfentan Alupent Ambroxol Amiodaro Amphoter Aprotini Atracuri Bretyliu Bupivaci Ceruleti Clonidin Diltiaze Dobutami Dopamine Dopexami Esmolol Fentanyl Flecaini Fluimuci Flumazen Furosemi Glucos30 DEC medication UT db be b s s ils BR qs WAN O O gt Q N PF Q WN F medication number ulus Qu 30 31 32 33 34 35 Medication name Glucos 5 Heparin Hydrocor Insulin Isoprena KCL b Ketamin Labetalo Lidocain Liothyro agnesiu Midazola ilrinon Morphin Nacl 0 9 albuphi aloxone eostigm i iz a La A cardip fedipi Nimodipi Nitropru Noradren Omeprazo
17. Pancuron Pentoxit Phentola DEC medication HEX medication SY number WWW to oO I w w w w Up gt D HD D B D BD D B Sr WB WwW UJ O oo A Ot is Q N OF Fi 4E 4F 50 Medication name Phenylep Procaina Propafen Propofol Rapilysi Remifent Risordan Ropivaca Salbutam Somatost Streptok Sufentan Terbutal Theopyli Thiopent Tirofiba rrinitri Urapidil Urokinas Vasopres Vecuroni Verapami DEC medication HEX medication number O UU On Q KO OO 1 O Q Q N P anu au O O O O O O O O O O u O 0 3 Ox Q N O MH AY Medication name 16 user medication names can be custom defined UserM 1 UserM 16 Refer to the EXCEL file for easy definition 1 8 Configuration of a reminder alarm for the safety standard check Special configuration options First the service interval has to be configured either in months or in hours of operation or both addresses 393 394 Next the last service date has to be entered on address 390 Any value greater than 0 entered at the adresses 393 and or 394 will release the reminder alarm after the set service interval has elapsed check also the correct settings of the internal clock PC configuration tool AConfig VVith this additional softvvare the pump may be configur
18. Service Manual for Syringe pump green stream SY P ARGUS 600 Made in Switzerland 0120 IMPORTANT This service manual is intended for the exclusive use of authorized persons vvho have been trained by ARGUS Medical AG in the maintenance and repair of the infusion apparatus mentioned above ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the unit by a non authorized person ARGUS Medical A CH 3627 Heimberg Svvitzerland A member of the CODAN group 14 135 E A600D en SM V1 51 Argus 08 12 04 PJ Display Operation mode Display Infusion rate Display Total Inf Y etc LED bar Pressure Display Alarm Display Alarm mode Key ON OFF Key 100 Key 10 Key 1 Key 0 1 Key MODE Key START STOP QO O Q WN 14 135 E_A600D en SM V1 51 oem VA p 14 Syringe barrel holder 15 Key BOLUS 16 Syringe guide 17 Beak 18 Drive unit 19 Lever for clamp 20 Clutch lever 21 Combination clamp 22 Ext 12VDC and interface RS 232 23 Staff alert 24 Line plug 25 Equipotential plug I 22 21 25 23 24 Argus 08 12 04 PJ Table of contents Page 1 Special key inputs and configurations 4 2 History and pump configurations printout 13 3 Fault finding 17 4 Replacement of parts 18 5 Safety standard check 23 Mechanical drawings 24 Wiring diagram 29 Blo
19. V paeog Aejdsig 009V P eog lqeo oeH lul 222 S6L LL POL LL OG FLL 68 LL 120701 19001 090701 89001 SBuimeap ynouyM syed s eds jesauay 990 OL Argus 08 12 04 PJ 24 14 135 E A600D en SM V1 51 UVZ 1 A9 u heq pive peo1 220 21 uvy M9 Kaye pide pea1 9207 I 21 ajoejdadai feul GC0 Cl IE ajpejdada1 ajewWaj coc Bnid suew 220121 youms UONNAUSNd 22021 MOL Jejewonuejodg0d 12021 260121 Ou Hed 10 Japjoy Aioneg 20211 2 aoz IL woddns 76111 OSXHIN M JOS 96L LL Buuds ainssdiq Z6L LL sBunyy jnoyym ped woyog Buse9 06111 abuuAs 8811 9 pgz eeum Bog 9411 4 plou YMS IZV L yeys 609 691711 Japjoy OBUHAS 91711 Sulxl GOL LL Jepjoy dwe SGK LI Jan09 eneg S LL lPUEH ESKIL dwe uoneuquoo ZSE LL Jep oy JOJOWONUSIOd 0911 10 ajpulds 00711 1 89L LL zz0 21 YOUMS SSL LL 202 LL 26031 92031 25 Argus 08 12 04 PJ 14 135 E A600D en SM V1 51 y 60 Buue q buueag jeg ayejd 5ulx yoq 19 Hoddns uen I uA 602 lpulds 6307 I 82021 cOc LL S6L LL 6601 SQL LL LL VEL 98111 9 9g Jeaym 60 SING 1041 10 0W 9 puids pesu xV 12 609 eys buueag Iduuo5 1886 pue 1010 SZ LL OLE LL 99111 ESL LL ZSV LL 9311 890701 ESL Hl 26 Argus 08 12 04 PJ
20. al connections 14 Electrical test according to EN60601 1 Measurements attached all measurements made with a power cable 2 5 m CODAN Running smooth See service manual Final calibration Other aars Preset level mbar Measured mbar See service manual Check pump accuracy Measured volume ml Green LED battery light Battery prealarm after typ 1 hr 45 min or 7 hrs 45 min Red LED battery alarm acoustical alarm Battery alarm 15 min after prealarm Red LED battery alarm ALARM alarm acoustically After 6 min the pump switches off Staff alerting system Computer interface RS232 External DC The pump has passed the safety standard check and is safe for use Date iene Name Woo 14 135 A600D en SM V1 51 Argus 08 12 04 PJ 23 TAWOD Inu u samep peay Idwod lpulds peasy L dwoo 15 5 95 wy Jeq 562 0 J91IWOUEN 2 2 nid ley 9 Bnid se jes an un jauRdjuo14 Zelds p uondalo1d uononAsul Yous uoneounuep Z0 01 c20 OL 69001 490101 s npoN 960 c1 GEO GL LEO GL OE0 GL ELT LL 902 LL LOG LL 007 LL 2001 z Wed uollelq leo Hed uolle qileo sbumu 1noylm aseq Pulse sum noujim 18109 Buiseo O x pul 009V pieog uen 009
21. c schematic 30 IMPORTANT This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 600 syringe pump The service manual is meant to be used together with the user manual IMPORTANT ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 600 syringe pump by a non authorized person This manual contains the latest data available It is subject to further modifications in accordance with technical improvements 14 135 E A600D en SM V1 51 Argus 08 12 04 PJ 1 Special key inputs and configurations 1 1 General Caution The configuration possibilities mentioned below constitute a modification of the pump and may only be carried out by authorized persons If the decimal points are flashing in a display this display is ready to accept an input by means of the keys 100 10 1 Note Use the interrogation mode to check the present configuration without doing any modification 1 2 Interrogation mode START MODE Hold both keys pressed O Important Do not release the keys MODE and START STOP Frag IE START STOP mi total h h min Enter the requested address eg 52 mi total h h min The programmed data of address 52 are displayed Press the key at least 2 seconds 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ 1 3 Configuration m
22. d START STOP pressed and go in the configu ration mode Select address 399 on the left display of the pump Start recording text received over the serial interface e g by selecting Recording text in the menu of the Hyper terminal A text file which contains the history printout will now be generated Enter the data 4567 on the right display of the pump Press the START STOP key The pump will now transfer the last events registered on the pump in the format mentioned below Stop the recording of the text received over the serial interface this will also close the generated text file generated text file can be opened and printed out by any text program History printout sample KK KKK History KR KKK Pump off Rate 123 0ml h Total 50 0ml Infsum 5 4ml Pump on Rate 10 0ml h Total 50 0ml Infsum 9 0ml etc The possible messages are Battery defective Battery lov prealarm Battery lov pump stop Bolus start Bolus stop External pover off External pover on Occlusion pump stop PC configuration done PC configuration failure 14 135 E A600D en SM V1 51 Mon 06 Mar 2000 08 42 44 Syringe PrLimit Status Syringe PrLimit Status PrLimit change Pump has detected failure Pump off Pump on Pump start Pump stop EVO Rate change Enter setup mode Exit setup mode Infsum cleared Wed 23 Feb 2000 11 54 38 156 900mbar 0x0000 Wed 09 Feb 2000 15 01 58 156
23. ed from a PC over the serial port This softvvare may be available from your local distributor or our service department After changing the configuration a function check and a control measurement has to be made 14 135 A600D en SM V1 51 Argus 08 12 04 PJ 12 2 1 22 History and pump configuration printout Connecting of the ARGUS 600 to the serial interface Caution The infusion pump must be disconnected from the patient before any connection over the serial interface is madel A connection of the ARGUS 600 to a computer is useful to the read the present configuration or history of the pump Even a simple monitoring of the pump can be done over the serial interface RS 232 The connection of the infusion pump vvith your computer over the interface can be done by connecting the interface cable part 10 058 and the follovving steps Connect the RS 232 interface cable to outlet 22 of the infusion pump and to the serial port of your PC Note in vvhich port COM1 or COM2 you have pluged in Start your terminal program on your computer A simple terminal program e g Hyper Terminal is included in every MS Windows 9x and Windows NT systems but must be installed Be sure that you have selected the right serial port COM1 or COM2 and set the following communication parameters Bits per second 4800 Baud Data bits 8 bits Parity None Stop bits 1 bit Protocol None Go to the next step in one of the furthe
24. ey MODE The display indicates 120P xxxx Press key START STOP Plunger length for 120mm will be stored Press key MODE The display indicates 20cd xxxx Press key START STOP Clamp diameter for 20mm will be stored Remove the calibration part 1 14 135 E A600D en SM V1 51 Argus 08 12 04 PJ 20 4 5 Put the calibration part 2 no 11 195 d 31mm 1 20mm in place w Press key MODE The display indicates 31bd xxxx Press key START STOP Barrel diameter for 31mm will be stored Press key MODE The display indicates 20P xxxx Press key START STOP Plunger length for 20mm will be stored Press key MODE The display indicates 32cd xxxx Press key START STOP Clamp diameter for 32mm will be stored Remove the calibration part 2 Insert a filled 50 ml syringe and connect the patient line to the pressure measurement system Press key MODE The display indicates 0 2b xxxx and the pump starts to run with a low rate Close the line occlusion Wait until 0 2 bar is reached on the scale At this point press the key START STOP immediately to register the value for 0 2 bar Note To speed up the process increase the infusion rate in steps by pressing the key 1 Press key MODE The display indicates 1 2b xxxx Wait until 1 2 bar is reached on the scale At this point press the key START STOP immediately to register the value for 1 2
25. load Adress VO JAN A Q t 20 23 24 26 27 28 29 30 32 33 34 35 38 39 40 41 42 43 44 45 46 47 48 49 55 Barrel outside B L Length at 50 30 20 10ml P Plunger Length Write Default prot value 0 No 0 No 0 No 0 No 0 No 0 No 0 No 0 No 0 No 0 No 1 Yes 0 No 0 No 0 No 0 No 0 No 1 Yes 1 Yes 0 No 1 Yes 1 Yes 0 No 0 No 0 No 0 No 0 No 1 Yes 1 Yes 1 Yes 0 No 1 Yes 0 No 0 No 0 No 0 No 1 Yes 1 Yes 1 Yes 0 No 0 No 1 Yes 0 No 1 Yes 14 135 E A600D en SM V1 51 Clamp o C kad Function PIN code active Run indication by running decimalpoint Key On Off only at STOP valid Rate change only at STOP valid Key STOP delayed see 361 2nd entry of rate 3 1 9 0 rate caculation disabled Static alarm staff alerting system Display elapsed or remaining time in run mode AE Select remaining time 7 1 Rate change confirmation in stop mode Recall last ml h rate at next power on 9 0 Recall last ml total endvolume at next power on Recall ml inf volume infused at next power on 20 1 SBS Step By Step function Display VTBI Volume To Be Infused Syringe type acknowledge at start KVO KOR mode see 60 Buzzer at start Menu Cir Clear ml inf 15 0 Sk k k k Menu Prl Pressure alarm limit akak akak Men
26. n 1 10mm at end of infusion Syringe barrel diameter outside B in 1 10mm Syringe clamp diameter outside C in 1 10mm Maxima rate in 1 10ml h 1 0 500 0 ml h Prime rate in 1 10ml h 1 0 500 0 ml h Max bolus rate in 1 10ml h 1 0 500 0 ml h Max bolus total in 1 10ml 1 0 25 0 ml Syringe length L in 1 10mm at 30 ml volume Syringe plunger length P in 1 10mm at end of infusion Syringe barrel diameter outside B in 1 10mm Syringe clamp diameter outside C in 1 10mm Maxima rate in 1 10ml h 1 0 999 9 ml h Prime rate in 1 10ml h 1 0 999 9 ml h Max bolus rate in 1 10ml h 1 0 999 9 ml h Max bolus total in 1 10ml 1 0 25 0 ml Syringe length L in 1 10mm at 50 ml volume Syringe plunger length P in 1 10mm at end of infusion Syringe barrel diameter outside B in 1 10mm Syringe clamp diameter outside C in 1 10mm only active for the respective syringe type User 10ml 100 User 20ml 120 User 30ml 140 User 50ml 150 Argus 08 12 04 PJ 10 10 10 10 450 120 150 150 10 10 10 10 500 120 200 200 10 10 10 10 700 120 220 200 10 10 10 10 700 120 280 250 3000 3000 3000 100 800 300 190 250 5000 5000 5000 200 900 300 240 300 5000 5000 5000 250 1000 300 260 300 9999 9999 9999 250 1200 500
27. n 13 0 9999 Display digit 1 amp 2 user medication 14 0 9999 Display digit 3 amp 4 user medication 14 0 9999 Display digit 5 amp 6 user medication 14 0 9999 Display digit 7 amp 8 user medication 14 0 9999 Display digit 1 amp 2 user medication 15 0 9999 Display digit 3 amp 4 user medication 15 0 9999 Display digit 5 amp 6 user medication 15 0 9999 Display digit 7 amp 8 user medication 15 0 9999 Display digit 1 amp 2 user medication 16 0 9999 Display digit 3 amp 4 user medication 16 0 9999 Display digit 5 amp 6 user medication 16 0 9999 Display digit 7 amp 8 user medication 16 0 9999 Infused sum in ml xxxxyyyy xxxx ml Infused sum in ml xxxxyyyy ml Operating time in min xxxxyyyy xxxx ml Operating time in min xxxxyyyy yyyy ml Maxima rate in 1 10ml h 1 0 300 0 ml h Prime rate in 1 10ml h 1 0 300 0 ml h Max bolus rate in 1 10ml h 1 0 300 0 ml h Max bolus total in 1 10ml 1 0 10 0 ml Syringe length L in 1 10mm at 10 ml volume Syringe plunger length P in 1 10mm at end of infusion Syringe barrel diameter outside B in 1 10mm Syringe clamp diameter outside C in 1 10mm Maxima rate in 1 10ml h 1 0 500 0 ml h Prime rate in 1 10ml h 1 0 500 0 ml h Max bolus rate in 1 10ml h 1 0 500 0 ml h Max bolus total in 1 10ml 1 0 20 0 ml Syringe length L in 1 10mm at 20 ml volume Syringe plunger length P i
28. nto the programming mode see chapters 1 3 or 1 5 Select address 399 Press key MODE Enter data 123 Press key SRART STOP the display indicates 17bd xxxx Press key MODE until 120P xxxx is displayed Loosen the lock screw of the cogwheel on the plunger potentiometer axle Move syringe drive without syringe fully to the left Turn the potentiometer axle in clockwise direction up to the final position and afterwards in the counter clockwise direction until approx 700 is displayed Fix the lock screw Control whether the full stroke can be made Syringe clamp diameter cd Go into the programming mode see chapters 1 3 or 1 5 Select address 399 Press key MODE Enter data 123 Press key SRART STOP the display indicates 17bd xxxx Press key MODE until 20cd xxxx is displayed Remove the syringe and make sure the clamp is fully closed Remove the cover of the driving head 11 212 Remove the clamp spring 11 210 Loosen the lock screw of the position lever 11 208 Turn carefully the potentiometer axle R2 in counter clockwise direction up to the final position Turn position lever 11 208 counter clockwise until it touches the housing see picture below Fix the lock screw make sure the position lever touches the housing Re install the clamp spring then a value of approx 700 is displayed Control whether the clamp stroke can be made Syringe barrel diame
29. ode without PIN code x START es Hold both keys pressed O Important Do not release the keys MODE and START STOP i IH START STOP mi total mi inf h min MODE 73 mi total mi inf h min START STOP mi total ml inf h min 73 mi total mi inf h min START 444 T s T G R t k l 9 1 ie ni ni za mi total mi inf Ol 100 10 G Enter the requested address eg 5 mi total h h min VAN MODE mi total h min 100 10 G Enter the requested data eg 1 mi total ml Lint STOP Confirm data entered acknovvledgement mi total 5 m ml inf ih h min 14 135 A600D en SM V1 51 Argus 08 12 04 PJ 14 15 Configuration mode first input of PIN code IMPORTANT Key 1 MODE amp START STOP 2 START STOP 3 MODE 4 START STOP 5 MODE 6 START STOP 7 MODE 8 100 10 1 9 START STOP 10 ON OFF Remember to make a note of your code and keep it in a safe place Description Keep both keys pressed before switching the unit on Acknovvledgement write protection is inactive Svvitch over to the display 3 Acknovvledgement Svvitch over to the display 3 Acknovvledgement Svvitch over to the display 3 Enter 1 to 4 digit code Acknovvledgement write protection is active Code is never visible End of the pr
30. ogramming mode Programming mode with active write protection code CAUTION Only the code holder can carry out modifications when the write protection is active Key 1 MODE amp START STOP 2 START STOP 3 MODE 4 START STOP 5 100 10 1 6 START STOP 7 100 10 1 8 MODE 9 100 10 1 10 START STOP 1 ON OFF 14 135 E A600D en SM V1 51 Description Keep both keys pressed before svvitching the unit on Acknowledgement write protection is active Switch over to the display 3 Acknowledgement Enter code in the display 3 Acknowledgement Enter reguested address Programmed data appear in the display 3 Switch over to the display 3 Enter reguested data Acknowledgement If the data are accepted entry changes to the display 2 End of the programming mode Argus 08 12 04 PJ Display 2 600 0 gt Display 2 6007 ee aoe Oe 0 Cod Cod 2207 Display 3 1 Display 3 x X X xX XXX OV YA YYYY 1 6 List of the interrogation and configuration functions CAUTION Before you undertake a new configuration or replace the EPROM or the mainboard printout the pump configuration see chap 2 2 Afterwards you can reenter the old not write protected values down
31. r chapters depending on your intention Configuration printout Switch the pump on while keeping the keys MODE and START STOP pressed and go in the configura tion mode Select address 399 on the left hand display Start recording text received over the serial interface e g by selecting Capture text in the menu of the Hyper terminal A text file which contains the present configuration printout will now be generated Enter the data 3456 on the right display of the pump Press the START STOP key The pump will now transfer the present configuration of the pump in the format mentioned below Stop the recording of the text received over the serial interface this will also close the generated text file The generated text file can be opened and printed out by any text program Pump configuration printout sample Configuration profile Wed 19 Jan 2002 11 29 55 Pump type ARGUS600 Inventory number 0000 0000 Software release V0 93 990819 6D5C Infused sum 230ml Operating time 5h32min Last service date 2000 week 12 00 0 50 0 100 0 150 0 200 0400 250 0000 300 0000 350 0000 01 1 51 0 101 0 151 0 201 0000 251 0000 301 0000 351 0000 02 0 52 0 102 0 152 0 202 0000 252 0000 302 0000 352 0000 03 0 53 1 103 1 153 0 203 0000 253 0000 303 0000 353 0000 etc 14 135 E A600D en SM V1 51 Argus 08 12 04 PJ 13 2 3 History printout Switch the pump on while keeping the keys MODE an
32. t statusbyte 3 P 1 X X X X X X P 1 X X X X X X OAOE Were QPULIS OAM dois dund were Jung OAM dois dund paysves 2301 YOO WILA snjog P AT S N ysty SAEMIY MO SABMIV OAM OAM dos dung wpe eqo 8 sun S UHETV eqo D P AT S N IAN JOUY ys y SAPA V MO SAPATV Format statusbyte 5 P 1 X X X X X X JOMO Ga T ydesseg HOMO QIT ydess eg Z 10M0 AI T yUdess egg 1 AO GA T ydeisieg Joddn AYT ydess eg 9p pY Inssad ys y S LMIV MO SAEMIY A sample of a monitoring session Switch the pump on with an inserted filled infusion set Type Ctrl N to set the pump in remote mode Type Ctrl B then 01200200 and then Ctrl C which sets the rate to 12 0 ml h and an infusion total of 200 ml The rate should now be shown in the left display of the pump Type Ctrl R to start the infusion Type Ctrl T to stop the infusion 16 Argus 08 12 04 PJ 14 135 E A600D en SM V1 51 3 Fault finding The fault code in case of a failure is displayed by pressing MODE key 12 F XX appears in display 3 and the source of the failure is listed in the table below Error Error reason Assembly group F 20 Internal watchdog Mainboard F 21 22 ROM test Mainboard F 23 24 RAM test Mainboard F 25 CPU test Mainboard F 26 Invalid function menu F 27 EEPROM data invalid Mainboard F 2
33. ter bd Go into the programming mode see chapters 1 3 or 1 5 Select address 399 Press key MODE Enter data 123 Press key START STOP the display indicates 17bd xxxx Loosen the lock screw of the potentiometer R14 on the power board Turn the potentiometer axle R14 on the power board in the counter clockwise direction up to the final position and afterwards in the clockwise direction until approx 700 appears in the display Fix the lock screw Control whether the syringe barrel can make the full stroke Strain gauge b Important No syringe is inserted and the syringe drive is positioned on the right side Go into the programming mode see chapters 1 3 or 1 5 Select address 399 Press key MODE 14 135 E_A600D en SM V1 51 Argus 08 12 04 PJ 19 Enter data 123 Press key MODE several times until 0 2b xxxx is displayed Adjust the screw of the trimmer R13 on the power board until approx 1500 is displayed LI IG xal 4 81 TI pr Ti 1 Fi m u 2 11 xsi vi c3 T1 xa 1 5 Fa 7 5 ma Lock x17 RM Power Board 44 Final calibration of the assembled pump Go into the program mode see chapters 1 3 or 1 5 Select address 399 Press key MODE Enter data 123 Press key START STOP The display indicates 17bd xxxx Please
34. u CAP Battery capacity akak akak Menu InF ml infused since last power up He HK Menu dLo Data lock k k k k Menu Stb Stand by Menu MEd Medication k k k b Menu tM Timer alarm k k k k Bolus Prime alvvays possible Menu boLr Bolus rate 32 1 Menu tot Bolus total 32 1 Display bolus VTBI instead bolus infused Bolus application automatic 34 1 32 1 Bolus total to be reset after each auto bolus Demo mode all menus enabled RER Clear ml h after infusion completed Clear ml total after inf completed 41 1 Syringe clamp diameter outside control Automatic pressure release after occlusion Pressure display 20 40 60 80 100 Bargraph ON Bargraph with indicator 25 steps 45 1 Standby and battery prealarm low volume Flashing numeric display at alarm Alarm acknowledge only with key MODE Med disp alternate with rate and ml inf 29 1 Argus 08 12 04 PJ Adress 60 65 100 101 102 103 104 105 106 107 108 109 110 111 120 121 122 123 124 125 126 127 128 129 140 141 142 143 144 145 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 Write Default prot value 0 No 0 No I I z Z Z z Z Z Z Z Z z ooo o 66096 ZZ oo 14 135 E A600D en SM V1 51 Function KVO KOR only after infusion completed
35. up and down 4 Hold the lever for clamp bottom in its upper position and actuate the barrel switch 5 Hold the barrel switch pressed and the lever fo clamp bottom in its upper position then actuat the lever for clutch top 6 Check the prime 77 press the key BOLUS twice within 1 sec Housing labels accessories connectors power cable etc Display shows the software release Display of 2 4 7 F ml total ml inf h min Test of the green operation LEDs Run indicator line battery KVO Test of the red alarm LEDs Occlusion near empty empty syringe battery ALARM Alarm Syringe lights on and off Alarm Syringe lights on and off Alarm Syringe lights on and off See service manual Prime 7 Setrate to 111 1 ml h and the total to 0 1 ml END reached the acoustical alarm LED ALARM press START STOP released 8 Test the pump at max rate 999 9 mi h 9 Check of the occlusion alarm pressure 50 ml syringe Pressure increase to21 2 bar Test of the occlusion alarm pressure 10 Check of the pump accuracy Rate 200 ml h Preset volume 20 mi 11 Battery check by setting the rate to 5 ml h disconnect the line and start the pump Running time hrs If the specified typical 2hrs 8 hrs of operation are not reguired the battery has to be changed only if the time is 1 5 hrs 5hrs due to environmental pollution 12 Charge the battery min 16 hrs 13 Check the extern
36. verify that the displayed values in the right hand display are within the following ranges without calibration part Valid ranges in the right hand display Syringe barrel holder unpulled 700 amp 300 Syringe barrel holder pulled 4 200 300 Note Please refer to chapter Rough alignments if the displayed value is out of range Press key MODE The display indicates 120P xxxx Please verify that the displayed values in the right hand display are within the following ranges without calibration part Valid ranges in the right hand display vvithout calibration part Drive unit totally left 600 200 Drive unit totally right 4400 200 Note Please refer to chapter Rough alignments if the displayed value is out of range Press key MODE The display indicates 20cd xxxx Please verify that the displayed values in the right hand display are within the following ranges without calibration part Valid ranges in the right hand display vvithout calibration part Clamp fully closed 700 300 Clamp fully opened clutch amp clamp lever fully pressed 2 000 300 Note Please refer to chapter Rough alignments if the displayed value is out of range Press key MODE until the display indicates 17bd Put the calibration part 1 no 11 194 d 17mm 1 120mm in place s E Press key START STOP Barrel diameter for 17mm will be stored Note Each stored value will be acknowledged by a sound Press k