1. Figure 6 1 RR 30000 SD Current Probe 2011 DynAmp LLC Page 13 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 6 2 BATTERY REPLACEMENT SAFETY WARNING Before removing the battery cover make sure that the Transducer is removed from around any active conductor The battery compartment may be accessed by pressing down on the arrow molded into the battery compartment cover to release the latch then sliding the cover off Remove the old batteries and install the new batteries in the position as shown in Figure 6 2 Slide the cover onto the enclosure until the latch engages The batteries used in the RR 30000 SD Current Probe are type AA alkaline cells or equivalent Replacement with other than the specified battery will invalidate the warranty TO TRANSDUCER Exe 3Vdc 100mA 3Vac INPUT OUTPUT omer PROBE RJ MODEL RR 30000 SD 24 ITEM 041322 OUTPUT 3 0VRMS MAX SCALING 1V 1kA 10kA SERIAL amp DynAmp LLC U S A Figure 6 2 Battery Replacement 2011 DynAmp LLC Page 14 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 6 3 DYNAMP LLC CUSTOMER SUPPORT For further assistance contact DynAmp Customer Support at Americas Telephone 1 614 871 6900 Fax 1 614 871 6910 8 00 AM to 5 00 PM USA Eastern Time From first Sunday in November to second Sunday in March 13 00 GMT to 22 00 GMT
2. CABLE ASSEMBLY BNC MALE TO BNC MALE E Figure 4 2 Output Signal Cable and Adapter 4 2 AC CURRENT TRANSDUCER The RR 30000 SD design utilizes the lightweight and flexibility of the AC Current Transducer This transducer is a versatile current probe that may be wrapped around most conductors Three standard lengths are available They are 24 36 and 48 inches It s application versatility and high voltage isolation rating clearly distinguishes the RR 30000 SD from other current measuring methods The measuring transducer is constructed from non ferrous materials minimizing any circuit loading The frequency response of the RR 30000 SD is wide compared to conventional CTs This allows the user to monitor a much wider range of line harmonic components than conventional CTs allow The RR 30000 SD was designed to be very flexible larger in aperture and smaller in cross section than many conventional CTs This allows measurement in tight places as never before possible Figure 4 3 Type RS5 Flexible Transducer 2011 DynAmp LLC Page 8 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE The output voltage of the transducer is proportional to the product of the current and the frequency At higher frequencies the output voltage of the transducer may damage the input circuits of the electronics package Refer to Chart 4 1 and Chart 4 2 for the safe operating areas of the 3kA and the 30kA range respec
3. I D Closed 12 0 in 305 mm O D Closed 10 9 in 277 mm Weight 0 6 Ibs 0 27 kg 50 48 Length Open 48 in 1220 mm I D Closed 15 9 404 mm O D Closed 14 7 372 2 mm Weight 0 8 Ibs 0 36 kg Safety Rating Double Insulated Installation Category Ill Pollution Degree 2 600VAC to Earth Working Voltage 600VAC to Earth Head Cable Test 5550VAC for 1 minute surface to output ACCESSORIES Standard Output Cable BNC Male to BNC Male 12 in 304 8 mm Adapter BNC Female to Dual Banana Plug 2011 DynAmp LLC 041143 H Page 6 HIGH CURRENT PROBE Installation Operation and Service Manual HIGH CURRENT PROBE 4 DESCRIPTION 4 1 OVERVIEW AND DESCRIPTION The DynAmp Flexible High Current Probe Model RR 30000 SD is an assembly similar in purpose to a CT or current transformer It may be used to measure ac currents from as low as several amps to a maximum of 30 0kA rms The device output is an analog voltage that is proportional to the current in the conductor The output signal is isolated from the hazardous conductor potential and is an exact replica of the ac current waveform in the conductor The output signal is available via a BNC connector and is quite easily connected to many different measurement devices Output adapters have been included with the RR 30000 SD for your convenience These adapters allow quick connection to an oscilloscope vi
4. 2 1 5V AA alkaline cells or by mini power connector 2 1mm on end of enclosure Battery life full scale input current 60 Hz 25 30 Hrs Controls Switch Range and Power LED Flashes once per 3 seconds if the Current Probe is on and the battery voltage is good 3kKA OFF 30kA Material ABS Plastic Dimensions 1 0H x 2 4W x 3 8D inches 25 4H x 71 0W x 96 5D mm Weight 0 1 lbs 0 045 kg 2011 DynAmp LLC 041143 H Page 5 Installation Operation and Service Manual TABLE 3 1 MODEL RR 30000 SD SPECIFICATIONS CONTINUED TRANSDUCER Transducer Jacket TPE rubber double insulated flame retardant UL 94 V 0 rated Output Cable Jacket FEP Teflon brown color Couplings Polypropylene flame retardant UL 94 V 0 rated black color Transducer Cover Silicone impregnated fiberglass silver gray color Minimum Bending 1 5 38 1 mm Bandiic Cable OD 0 625 in 15 87 mm Coupling O D 0 875 in 22 20 mm Connecting Cable Length 78 7 in 2 0m shielded Temperature Operational 20 C to100 C 4 F to 212 F Position Sensitivity lt 2 0 with measured bus gt 1 from head External Magnetic Field 1 096 with external bus gt 8 from head Dimensions 50 24 Length Open 24 610 mm I D Closed 8 3 in 209 9 mm O D Closed 7 0 in 178 2 mm Weight 0 4 Ibs 0 18 kg SD 36 Length Open 36 in 915 mm
5. From second Sunday in March to first Sunday in November 12 00 GMT to 21 00 GMT Europe Telephone 41 22 706 1446 Fax 441 22 706 1311 8 30 AM to 5 00 PM Central European Time From last Sunday in October to last Sunday in March 7 30 GMT to 16 00 GMT From last Sunday in March to last Sunday in October 6 30 GMT to 15 00 GMT After Hours Critical Service Emergency Telephone 1 614 871 6906 5 00 PM to 8 00 AM USA Eastern Time From first Sunday in November to second Sunday in March 22 00 GMT to 13 00 GMT From second Sunday in March to first Sunday in November 21 00 GMT to 12 00 GMT Central e mail hel dynamp com DynAmp web www dynamp com O 2011 DynAmp LLC Page 15 041143 H
6. arrow on the latch see Figure 4 3 pointing in the direction of conventional current flow Conventional current flow is defined as current flowing from the positive to the negative potential The Flexible Transducer must be installed with the interconnection cable on the outside of the loop when the latch is engaged The polarity arrow the double insulation and the warning symbols will all be on the outside of the loop It should also be noted that the current probe will produce twice the output voltage if you wrap the transducer around the conductor twice There is minimal shock hazard using the RR 30000 SD as the transducer does not generate high voltages at low frequencies Each transducer has been Hi Pot tested to several thousand volts with no voltage breakdown This particular characteristic allows high current measurement with a wide frequency bandwidth of conductors at less than 600V ac potential to earth WARNING Do not use until you have read this Hazardous potentials may exist in the vicinity of the desired current measurements Use locally approved safety procedures when working near these hazardous potentials It is recommended not to install the DynAmp Flexible High Current Probe around a live bus that is at a hazardous potential lf installation is not possible when the bus is inactive or turned off always use appropriate gloves and or equipment that are approved when working around hazardous potentials when insta
7. 41143 H Installation Operation and Service Manual HIGH CURRENT PROBE 1 SAFETY 1 1 SAFETY SUMMARY The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARNINGS given elsewhere in this manual violates safety standards of design manufacture and intended use of the instrument The DynAmp LLC assumes no liability for the customer s failure to comply with these requirements WARNING Do not use until you have read this Hazardous potentials may exist in the vicinity of the desired current measurements Use locally approved safety procedures when working near these hazardous potentials It is recommended not to install the DynAmp Flexible Current Probe around a live bus that is at a hazardous potential If installation is not possible when the bus is inactive or turned off always use appropriate gloves and or equipment that are approved when working around hazardous potentials when installing the DynAmp Flexible Current Probe in the vicinity of these hazardous potentials The ac transducer and interconnection cable uses double insulation to protect the operator from possible high voltage potentials of the bus The ac transducer and interconnection cable are rated for Installation Category IIl Pollution Degree 2 The maximum voltage to earth rating for the transducer and cable is 600V4c The elec
8. 43 H Installation Operation and Service Manual HIGH CURRENT PROBE DynAmp LLC Customer Support For further assistance contact DynAmp Customer Support at Americas Telephone 1 614 871 6900 Fax 1 614 871 6910 8 00 AM to 5 00 PM USA Eastern Time From first Sunday in November to second Sunday in March 13 00 GMT to 22 00 GMT From second Sunday in March to first Sunday in November 12 00 GMT to 21 00 GMT Europe Telephone 41 22 706 1446 Fax 41 22 706 1311 8 30 AM to 5 00 PM Central European Time From last Sunday in October to last Sunday in March 7 30 GMT to 16 00 GMT From last Sunday in March to last Sunday in October 6 30 GMT to 15 00 GMT After Hours Critical Service Emergency Telephone 1 614 871 6906 5 00 PM to 8 00 AM USA Eastern Time From first Sunday in November to second Sunday in March 22 00 GMT to 13 00 GMT From second Sunday in March to first Sunday in November 21 00 GMT to 12 00 GMT Central e mail hel dynamp com DynAmp web www dynamp com 2011 DynAmp LLC Page ix 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page x 041143 Installation Operation and Service Manual HIGH CURRENT PROBE TABLE OF CONTENTS Par Title 1 SAFETY 1 1 SAFETY SUMMARY men nm menemn nennen nenne nnne nnns 1 2 SAFETY SYMB O S eene nennen nnn renis 2 HANDLING AND STORAGE 3 SPECIFIC
9. ATIONS 4 DESCRIPTION 4 1 OVERVIEW AND DESCRIPTION eee 4 2 AG CURRENT TRANSDUGER cete eet tet abies 4 3 EXTERNAL POWER SUPPLY ode mitos ess 5 INSTALLATION 5 1 INSTALLATION INSTRUCTIONS eem 6 MAINTENANCE 6 1 MAINTENANCE i m n ebd 6 2 BATTERY REPLACEMENT 6 3 DYNAMP LLC CUSTOMER LIST OF FIGURES Figure 4 1 RR 30000 SD Current Probe Electronics Package Figure 4 2 Output Signal Cable and Figure 4 3 Type RS5 Flexible Figure 4 4 Input Power Connector Figure 6 1 RR 30000 SD Current Figure 6 2 Battery LIST OF TABLES Table 3 1 Model RR 30000 SD Specifications Table 3 1 Model RR 30000 SD Specifications continued LIST OF CHARTS Chart 4 1 Range Safe Operating Chart 4 2 30 Range Safe Operating 2011 DynAmp LLC 041143 H Page xi Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page xii 0
10. DynAmp DYNAMP FLEXIBLE HIGH CURRENT PROBE MODEL RR 30000 SD 24 36 48 SINGLE PHASE CURRENT PROBE Installation Operation and Service Instructions Manual Item No 041143 Rev H DynAmp 3735 Gantz Road Phone 1 614 871 6900 www dynamp com Grove City Ohio 43123 USA Fax 1 614 871 6910 help ddynamp com Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page ii 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE DynAmp LLC WARRANTY Items and components manufactured by Seller for permanent installation are warranted for two 2 years from the date of shipment Items and components manufactured by Seller for portable and temporary use in more than one location are warranted to be free from defects in material and workmanship for a period of eighteen 18 months from the date of shipment Items and components not manufactured and resold by Seller are warranted by their manufacturer Warranty repair shall be at DynAmp s option in the form of repair or replacement of the defective items or components Concerning warranty repairs DynAmp will be responsible for DynAmp provided time material and transportation costs shipping or travel Actual method of warranty repair correction will be determined by DynAmp at DynAmp s sole option Such warranty repair shall constitute a fulfillment of all DynAmp LLC liabilities in r
11. a a BNC connector or the voltage terminals of a multimeter via banana plugs The RR 30000 SD comes standard with two ranges selectable by the switch mounted on the electronics package The ranges are 3kA and 30kA yielding respectively and 30kA rms ac as the maximum current that can be measured When the RR 30000 SD is turned on the LED mounted behind the translucent ruby panel on the package end slowly blinks to indicate the unit is on and the battery voltage is sufficient to function properly on all ranges This LED indication is characterized by blinking approximately once every three seconds If no light is observed the batteries of the RR 30000 SD should be replaced The RR 30000 SD comes equipped with an external mini jack located on the package This is for those wanting to supply power to the RR 30000 SD for a longer period than the internal battery life will allow External dc adapters are available for this purpose but are sold as an option The power required is 3 0Vpc 100mA Any external dc power supply adapter must be safety rated 0 uum POWER CONNECTOR BNC FEMALE 1 CONNECTOR P SWITCH TO TRANSDUCER Figure 4 1 RR 30000 SD Current Probe Electronics Package 2011 DynAmp LLC Page 7 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE gt ADAPTER BNC FEMALE TO gt DUAL BANANA PLUG
12. e product must be properly installed and operational before the protection is fully functional Therefore adequate protection from humidity shock and temperature must be provided during handling and storage prior to installation The handling and storage of equipment must be sufficient to meet the storage temperature and humidity specifications of the product and to prevent any condensation or contact with water or any other liquid The storage location and container or crate must provide adequate protection from precipitation rain snow ice and direct water contact Adequate shelter must be provided to prevent the accumulation of precipitation rain snow ice and water which can lead to the deterioration or failure of shipping containers or crates and cause water ingress Storage in coastal or industrial areas subject to salt laden or corrosive air or areas of wind driven sand or other abrasive dust must be adequate to prevent the deterioration or failure of shipping containers or crates and cause ingress Frequent inspection of storage areas and storage containers or crates is required to ensure proper storage conditions are being maintained If the shipping container or crate is opened and or the equipment is removed for inspection prior to installation the equipment must be repackaged in the original undamaged container or crate in the same manner as it was shipped to prevent environmental damage or placed in a storage location that meets th
13. e required environmental and storage conditions General product storage temperature and humidity requirements Storage Temperature 40 to 70 C 40 to 158 Storage Humidity 85 non condensing DynAmp LLC does not assume liability for the customer s failure to comply with handling and storage requirements For further assistance contact DynAmp customer support 2011 DynAmp LLC Page 3 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page 4 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 3 SPECIFICATIONS The specifications for the RR 30000 SD are given in table 3 1 TABLE 3 1 MODEL RR 30000 SD SPECIFICATIONS SIGNAL CONDITIONER Input Type RS5 Transducer 3 0VAC or 4 2VPK Output Load gt 500 ohm Output Connector BNC female Switchable Scaling 3kA 30kA 1mV 0 1mV per amp Accuracy 1 0 of full scale Linearity 0 2 of reading 10 100 of full scale Repeatability 0 1 of reading 10 100 of full scale Frequency Response 8Hz to 100kHz 3dB points Phase error lt 0 5 maximum 50 60 Hz Noise 10 0mV ac maximum DC offset lt 5 0mV dc maximum Temperature range Operational 20 C to 85 C 4 F to 185 F Gain Change 0 08 C maximum DC offset 0 3mV C maximum Power 3 0VDC 100mA from
14. espect to said items and components In no event shall DynAmp LLC be liable for consequential damages Information in this document is subject to change without notice 1996 1997 1999 2005 2008 2011 DynAmp LLC All rights reserved Reproduction for purposes other than operation and service without written permission of DynAmp LLC is strictly forbidden This manual includes detailed drawings installation operation service and maintenance Users should evaluate the information in the manual and their particular application DynAmp assumes no liability for any incidental indirect or consequential damages arising fro the use of this documentation While all information presented is believed to be reliable and in accordance with accepted engineering practices DynAmp makes no warranties as to the completeness of the information 2011 DynAmp LLC Page iii 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page iv 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE Caution Refer to manual before using product Model RR 30000 SD Current Probe conforms to the latest European directives and standards concerning safety and electromagnetic compatibility European Low Voltage Directive 73 23 EEC and 93 68 EEC European EMC Directive 89 336 EEC and 93 68 EEC Safety Standards EN 61010 1 Safety requirements for electrical e
15. lling the DynAmp Flexible High Current Probe in the vicinity of these hazardous potentials Do not exceed the minimum bending radius of the AC Current Transducer when installing the transducer around the conductor Exceeding the bending radius will degrade the measurement accuracy Make sure the AC Current Transducer and its output cable are clean before installing them around the conductor If the transducer and cable are not clean the contaminants on them may provide a conductive path for a high voltage breakdown Also check the transducer and output cable for cuts and abrasions Do not use the transducer if damaged 2011 DynAmp LLC Page 11 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page 12 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 6 MAINTENANCE 6 1 MAINTENANCE Preventive maintenance primarily consists of cleaning the transducers and cables to prevent surface contamination Use a mild detergent and a damp cloth to clean the transducers and cables Remove the detergent with a damp cloth then wipe with a clean dry cloth NOTE The use of solvents as cleaners is not recommended unless thoroughly tested and found harmless to all surfaces and parts Do not submerse current probe into water or other fluids TRANSDUCER EXTERNAL POWER SUPPLY 3VDC 350mA ce Me 3kA SERIAL D OFF
16. quipment for measurement control and laboratory use EMC Standards RF Susceptibility EN 50082 1 3V m Residential Commercial and Light Industrial Equipment HF Emissions EN 55011 Group 1 Class B Industrial Scientific and Medical Equipment SAFETY This product is designed to be safe under the following conditions e Indoor use e Altitude up to 2000m e Maximum Ry 80 for temperatures up to 31 C decreasing linearly to 40 Ry at 85 C Use of this product is limited to 600V ACrms or DC between uninsulated conductor and ground and for frequencies below 1kHz 2011 DynAmp LLC 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page vi 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE REVISION HISTORY Page Rev Revision Summary Date all New First Issue 03 96 9 10 A Revise Specifications 10 96 all B Replace Transducer with Type RS5 11 96 Front C Warranty Statement 05 97 10 D Revise Specifications 08 97 Front D Warranty Statement 08 97 ii 1 E Revise CE Statements 08 99 all F Update to DynAmp LLC 06 05 all G ECO 3206 Update drawings to DynAmp logo 11 08 all H PAR 10245 Handling amp Storage New manual 03 11 format 2011 DynAmp LLC Page vii 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page viii 0411
17. tively 3000A Range e m c I o 577 Amps 0 10 1 00 10 00 100 00 Frequency kHz Chart 4 1 3kA Range Safe Operating Area 30kA Range 4 x c un i 2 Oo 1 Frequency kHz Chart 4 2 30kA Range Safe Operating Area 2011 DynAmp LLC Page 9 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 4 3 EXTERNAL POWER SUPPLY The user may choose to purchase a suitable external power supply Figure 4 4 shows the important dimensions of the power connector and the required polarity of the supply voltage The external power supply should have the CE mark and must be rated for 3 0Vpc 340mA INSIDE CONDUCTOR COMMON SHORT TAB OUTSIDE CONDUCTOR VDC d LONG TAB __ INSIDE CONDUCTOR 0 08 2 1mm 0 219 5 5MM Figure 4 4 Input Power Connector 2011 DynAmp LLC Page 10 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 5 INSTALLATION 5 1 INSTALLATION INSTRUCTIONS The RR 30000 SD was made to allow the operator to connect this measurement device around a conductor without disconnecting the conductor as many CTs presently demand Even though the Flexible Transducer output is ac there are instances where the user will want to orient the transducer so that proper polarity will exist at the output terminals This is done by installing the transducer around the conductor with the molded in
18. tronics package is not double insulated or protected against hazardous potentials from connected instruments or mains adapters Make sure the electronics package is well away from the bus and that the inputs of instruments connected to the DynAmp Flexible Current Probe outputs are not hazardous live 1 2 SAFETY SYMBOLS General definitions of safety symbols used on equipment or in manual General definitions of safety symbols used on equipment or in manual Direct current power line Equipment protected by DOUBLE INSULATION or REINFORCED INSULATION Caution refer to accompanying documents 2011 DynAmp LLC Page 1 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE This Page is Intentionally Blank 2011 DynAmp LLC Page 2 041143 H Installation Operation and Service Manual HIGH CURRENT PROBE 2 HANDLING AND STORAGE DynAmp products are engineered and manufactured for use in industrial environments However they contain sensitive electronic and mechanical components which may be damaged and fail if not handled and stored properly All products must be handled and stored with the same care as any precision measurement instrument Severe bumps or jolts may damage internal parts and cause malfunction or premature failure DynAmp products are designed and assembled with conformal coating shock mounting and environmental seals when appropriate or when specified However this protection requires that th