1. RESMED ResMed Power Station II RPS II Launch PRODUCT BULLETIN Bulletin Number PB 678 Month Year April 2011 Distribution Asia Pacific Date of implementation Immediate General The ResMed Power Station II RPS II is ready for launch The RPS Il is an external lithium ion battery pack It is used to provide convenient power and backup power to ResMed ventilators and sleep devices running on 24V and 26V DC The RPS II is ideal for travel or when additional mobility and freedom are required such as on the back of a wheelchair and during travel The power capacity and operation time can be doubled by using two 158 batteries with an automatic transition from the primary to the secondary battery Q How the RPS Il works ResMed Power Station Il is an external long life lithium ion battery that simply works alongside ResMed ventilators or sleep therapy devices Charging the battery You can charge one battery from 0 to greater than 95 in less than 4 hours Power or backup power Using the appropriate DC cable you can connect the battery to a ResMed therapy device to provide power or backup power When providing backup power to the DC inlet the AC power cord connects the therapy device to the mains power outlet available on devices with both AC and DC inputs The voltage output selector on the RPS II rear panel lets you select the correct output voltage for your device 24V or 26V Using two batteries doubling capacit
2. attery device compatibility list Product Run Time hours VPAPIII ST A and VPAPIII ST A with QuickNav 6 VPAP S VPAP ST VPAP IV VPAP IV ST 9 S9 Series 13 Stellar 8 VSIII A PCV Mode 7 VSIII A CV Mode 4 5 Elis e 150 A PCV Mode 6 Elis e 150 A CV Mode 4 5 Technical specifications Battery technology Capacity UN classification Power supply Battery output voltage Power supply unit output current rated Battery output current rated Standby current Protection Dimensions L x W x H Battery weight System weight Minimum life cycle Operating conditions Transport storage conditions Operating storage air pressure Recharge time IEC 60601 1 classification Electromagnetic compatibility Lithium ion 97 Wh UN3481 lithium ion batteries packed with equipment Input range 100 240V 50 60HZz 1 0 1 5A Nominal for aircraft use 110V 400Hz 24V 26V 0 5V 90W continuous 3 75A 3 75A 3 46A lt 100 pA Overcharge over discharge excess current short circuit high temperature 230 x 126 x 26 mm 0 9 kg 2 3 kg 2 500 cycles at 23 C to 60 capacity Charging 5 C to 40 C 0 85 maximum humidity Discharging 5 C to 40 C 0 85 maximum humidity 20 C to 45 C 85 maximum humidity 600 hPa to 1060 hPa lt 4hrs to gt 95 of full level Class II double insulation and or Internally Powered Equipment IPX1 IPX0O while charging See further User manual or www resmed com Product complie
3. g Product codes RPS II kits One kit contains 1 x battery 1 x User Guide 1 x PSU Adapter 1 x Carry Bag 24921 ResMed Power Station II Battery Pack ROW Additional items and accessories 24926 RPSII Battery Only 24952 RPSII Carry Bag 24959 RPSII DC Cable VPAPIII ST A Stellar 150 24981 RPSII DC Cable Stellar 0 5m for use in Stellar Mobility bag 24960 RPSII DC Cable VPAP IV 24961 RPSII DC Cable S9 Series 24962 RPSII DC Cable Elise 150 24963 RPSII DC Cable VSIII 24964 RPSIl Battery Coupler Kit 36824 S9 PSU 90W PLUS ANZ PWR CORD Ordering information 1 Order the ResMed Power Station II Battery Pack kit 24921 2 Order the DC cable based on the specific ResMed ventilator or sleep device to be connected to the RPS Il 3 For devices other than S9 or if you want an additional PSU order the correct S9 Power Supply Unit PSU from the list above Optional If two batteries are required order an additional battery only 24926 and a battery coupler kit 24964 ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Tel 61 2 88841000 Fax 61 2 8884 2000 ABN 30 003 765 142 Global leaders in sleep and respiratory medicine www resmed com RESMED Compatibility and run times RPS Il is compatible with e VPAP III ST A VPAP III ST A with QuickNav VPAP S VPAP ST VPAP IV VPAP IV ST S9 series Stellar VS III and Elis e 150 Run Times as stated in the B
4. ith the control buttons and connector The front display of the Rest of World ROW RPS II has symbols only e RPS II has a switchable output between 24V and 26V The switch is placed on the rear panel Elis e 150 and VS III require 26V all other devices require 24V e Different DC cables are available to connect the RPS II to the appropriate device e At launch the User Guide for the ROW version will come with an insert with information about storage and correct DC output voltage amp DC cables for every device platform A new User Guide coming soon will have this information included e Document 1013434 Battery device compatibility list shows the RPS II s compatibility with ResMed therapy devices with the corresponding battery run times This document is available on www resmed com on the Products page under Service amp Support e RPS Il is using the same 90W PSU as the S9 platform users with this 90 W PSU do not need an additional PSU Using two RPS Ils The Battery Coupler connects two RPS Ils to double the run time autonomy and has labeled cables to the two batteries The primary battery is discharged first then the secondary battery takes over and the primary can then be disconnected and recharged Charging two RPS Ils When charging two batteries simultaneously from one PSU using the Coupler Kit 24964 the charging time until both batteries are fully charged might be more than 4 hours Each RPS II battery weighs 0 9 k
5. ries to 50 again after three months of storage When not in use all Lithium lon batteries self discharge over time If not periodically recharged i e every three months the RPS II will eventually self discharge to the point that it can no longer be recharged If this occurs the RPS II is no longer usable and is unrecoverable Note A fully charged battery when left switched on self discharges to less than 50 of full capacity within two weeks of storage Once switched off the battery self discharges to less than 50 of full capacity within one month of storage For further information please contact your nearest ResMed office See www resmed com for details 2011 ResMed Ltd 104552 1 2011 05 ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Tel 61 2 88841000 Fax 61 2 8884 2000 ABN 30 003 765 142 Global leaders in sleep and respiratory medicine www resmed com
6. s with all applicable electromagnetic compatibility requirements EMC according to IEC60601 1 2 See further User manual or www resmed com ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Tel 61 2 88841000 Fax 61 2 8884 2000 ABN 30 003 765 142 Global leaders in sleep and respiratory medicine www resmed com RESMED Service and Maintenance The RPS Il is intended to provide safe and reliable operation when used and maintained in accordance with the instructions provided by ResMed There is no servicing necessary during the lifetime of the battery The service life of the RPS Il exceeds 500 recharge cycles After this cycle the battery will last about 60 of the time from its original condition Cleaning recommendations are provided in the User Guide Some countries require all medical equipment to have a periodic service to confirm that the equipment is in good working order Section 3 of the Service manual General inspection procedures may be used in this situation If the RPS II needs to be returned to a ResMed Service Centre complete the ResMed Service Request Report found in Appendix A of the Service manual Send the completed form along with the RPS II to your ResMed office For more details about service and maintenance see the Service manual document number 248441 Storage Batteries must be charged to 50 indicated by two green LEDs and switched off before storing You must recharge the batte
7. y The two batteries sit neatly on top of each other A Velcro strap is included in the coupler kit to hold them more securely if required Both batteries can be charged together or used simultaneously to power a therapy device via the coupler one battery serving as the primary power source the other as the secondary reserve Provided in a practical bag An RPS II kit 24921 consists of one battery a user guide a PSU adapter to connect the power supply to the battery pack and a convenient carry bag The carry bag also has space for a second RPS II battery pack DC cable and PSU with mains cable to be ordered separately RPS II in its own convenient carry bag ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Tel 61 2 88841000 Fax 61 2 8884 2000 ABN 30 003 765 142 Global leaders in sleep and respiratory medicine www resmed com RESMED Marketing material e RPS II fact sheets are available to order from the Australian Distribution Centre the document number is 1014318 e An RPS II web page is available on line www resmed com rpsll Useful information e RPS II battery capacity is 97 Wh just below 100 Wh which is the present limit for bringing the batteries as hand luggage on board passenger aircrafts without notifying the carrier However we always advise the users to inform the air carrier before travelling e Alerts indicators and power level LEDs are placed on a convenient front display together w