TP702 Operators Manual - English
Contents
1. connectors raised above the level of the pad and pump Coil the hose and attach the Clik Tite connectors together on the hose See Figure 4 Page 9 where applicable Secure the hose to the T Pump using the tube set strap Wrap the power cord around the unit www gaymar com OPERATOR S MANUAL TP702 T PUMP OPERATING INSTRUCTIONS THERAPY CYCLES Therapy Cycles PROFESSIONAL ONLY 1 www gaymar com Turn the unit on and place it into 20 or 30 minute therapy cycle The unit will start to heat up Select warming period for 20 or 30 minute cycle LED should be solid and the selected temperature LED is flashing Once the T Pump reaches the desired temperature the Warming LED Red flashes with a short audible beep This is to let the operator user know the unit is at temperature and the timed therapy is starting The warm up period will take about 5 to 10 minutes depending on the selected Setpoint a Note If this is not the first time the unit has cycled the unit is already warmed up the warm up period is about 2 minutes Once the 20 or 30 minute period is up the heater and pump shuts off This stops any flow to the pad The Setpoint and Therapy time LED s will be blinking during the off Time period This is to inform the operator that the unit is not just in standby During this time the pad temperature will slowly drop to the patients body temperature Once the Off Time period is up the pump and
2. Functional Check with an empty reservoir Damage to the T Pump may result Read through and understand each step before performing the test To assure optimum performace dependability and safety the following Functional Check should be performed once per year or as specified in the facility s preventive maintenance program e TFC1 or equivalent Thermometer with 30 F to 125 F 2 C to 52 C range with 2 F 1 C accuracy e TPT9 is used to measure the temperature and the flow of the water entering the pad e Stop Watch for testing Over Temperature Safety Circuit e Mul T Pad Note To order a TFC1 TPT9 and MuleT Pad contact your dealer or Gaymar s International Department Connect the ii to the TPT9 and Mul T Pad as shown in Figure 6 S SSS a SSS Thermometer Mul T Pad TPT9 Flow Indicator Figure 6 Functional Check Setup www gaymar com OPERATOR S MANUAL TP702 T PUMP FUNCTIONAL CHECK Test Procedure Note If at any time the Function test needs to be stopped press the On Standby button www gaymar com Note Stay present for the full duration of the test to observe all readings and to prevent damage to the T Pump Attach a pad to the connector hose See Figure 6 on page 18 Then unkink the pad and hose Open the hose clamps Open the fill cap on top of the pump Fill with room temperature water to the Heating water line LU Plug the T Pump into a proper
3. be filed with the delivering carrier by the purchaser Do not return pumps that were damaged in shipment to GAYMAR without contacting our Technical Service Department for advice see phone numbers below If damaged goods are returned to GAYMAR without notifying the carrier GAYMAR will assume the repairs will be made at the customer s expense TO RETURN PUMPS TO FACTORY FOR REPAIR OR EXCHANGE Merchandise returned to GAYMAR must be accompanied by a Return Goods Number RG issued by GAYMAR authorizing goods to be returned Contact your local dealer or Gaymar s International Department at International 1 716 662 8636 Fax 1 716 662 0730 Provide the model serial number and detailed nature of the problem You will be given a Return Goods Number RG The serial number can be found on the bottom of the T Pump T PUMP T PAD Mul T Pad Clik Tite and Gaymar are registered trademarks of Gaymar Industries Inc U S PATENT 4 068 870 2008 Gaymar Industries Inc All rights reserved OPERATOR S MANUAL TP702 T PUMP SAFETY PRECAUTIONS PRECAUTIONS Risk of explosion Do not use in the presence of flammable anesthetics Risk of electric shock Disconnect power before servicing the A WARNING This device pumps temperature controlled water through a pad Set the pad temperature only as directed by and under the guidance of the prescriber Check the skin integrity of the body surface to which therapy is applied Eva
4. cap on top of the pump To fill for cooling a Fill with cold water to the Cooling water line b Fill with ice to the full capacity of the reservoir AuAuw To fill for heating fill with room temperature T water to the Heating water line KAJA Plug the T Pump into a properly grounded Hospital Grade receptacle Press the On Standby button The light next to the selected temperature begins to flash Use the keypad to set the temperature as directed by the prescriber After setting the temperature press and hold the lock Temperature Setpoint button for 2 seconds to lock the setpoint Note If you toggle past the desired setpoint keep toggling to start at the beginning of the setpoint column If warming the selected water temperature will be reached in approximately 11 minutes and the light next to the selected temperature becomes steady Check the water level If it drops below the operating level add water Apply the Mul T Pad to the patient as prescribed Follow the Mul T Pad instructions Position the pump at or above the level of the pad Note If the pump is placed below the pad s water will drain into the pump when it is shut off If the pump has been overfilled or if multiple pads are connected excess water can leak Press the On Standby button so that the Standby light is lit Unplug the T Pump Close all hose clamps Disconnect pad s from pump To prevent water spillage always disconnect pad from pump with
5. Indicates no flow Turns off heater if pump is tipped Warm Cool Four temperature setpoints on the TP702 Delivery Therapy Cycles Choose from 20 minute 30 minute or Continuous cycles On Standby Indicates power is supplied to the unit Button Over Temp Safety Limit thermostat shuts off pump and heater if the high Thermostat temperature limit is exceeded Self Check Automatic system check at startup Hose cord Convenient and easy storage areas for hose and cord Management Comfortable Designed for a more comfortable grip when moving the Handle Design pump Dual Micro Two electronic circuits one over temperature sensing Processor circuit Tethered Easy Prevents misplacing the cap Only 1 4 of a turn is needed Open Cap to remove or secure the cap Handle Vents The vents in the handle allow air flow to keep the motor and heater inside the unit cool On Standby Button Vents in Handle Easy to Use Keypad Flow Indicator Temperature Setpoints for warm or cool delivery Therapy Time Cycles Hose storage Figure 2 T Pump Features www gaymar com 5 OPERATOR S MANUAL FEATURES TP702 T PUMP KEYPAD FEATURES TP702 T Pump co Professional Figure 3 TP702 Keypad Indicator Warning Light See the Troubleshooting section on page 12 Water Flow Check hoses or clamps for kinks or occlusions g Water Level Check water level Setpoints Press the button at the bottom of the setpoint in
6. OR S MANUAL TP702 T PUMP Note If the system does not shut down in an over temperature condition in 30 seconds press the On Standby button then unplug the T Pump Contact your dealer or Gaymar s Technical Service Department for assistance Telephone 716 662 2551 10 11 12 13 Start the stop watch to make sure the Over Temperature condition is detected in 30 seconds WARNING If this step is not followed damage to the T Pump may result e TheT Pump gives a long beep and flash of the Warning light e The pump stops No water will be flowing to the pad e The Heater is turned on at 100 power with no flow e The 107 F 42 C Setpoint light starts flashing e The 20 minute Cycle light starts flashing e When the T Pump detects an Over Temperature Condition temperature trip point of 115 F 5 F 46 C 2 C the T Pump gives an audible beep and flashes the Warning light The system then cycles and goes into Standby mode with the Warning light on When the Warning light turns on stop the Stop Watch Restart the Stop Watch to make sure the T Pump cools down to a point where the thermostat resets and the Warning light turns off in 60 minutes Or to speed up the cool down go to step 13 To speed up the cool down a Unplug the T Pump b Refill the reservoir with room temperature tap water c Restart the unit in Cooling mode The system starts cooling down in Standby mode The thermostat resets wi
7. SS Gaymar Your patients Our commitment T PUMP TP702 Localized Temperature Therapy System CE 0086 www gaymar com P N 100963000 Rev A 8 08 BEFORE YOU BEGIN OPERATOR S MANUAL TP702 T PUMP CONTENTS SAFETY PRECAUTIONS INTRODUCTION INDICATIONS FOR USE CLINICAL CONTRAINDICATIONS PUMP FEATURES KEYPAD FEATURES TP702 CLIK TITE CONNECTORS START UP PROCEDURE SHUTDOWN PROCEDURE THERAPY CYCLES PROFESSIONAL ONLY caitsssteisctectitirectassne 9 TROUBLESHOOTING STORAGE CLEANING PREVENTIVE MAINTENANCE SERVICE FUNCTIONAL CHECK SPECIFICATIONS WARRANTIES SYMBOLS USED WITHIN THIS MANUAL AND ON THE PRODUCT Warm Water Fill Line tt Cold Water Fill Line a Attention consult Operator s L Manual eae Dangerous Voltage Type BF Applied Part CA Do not penetrate with sharp object Length of cycle time oo ener BEFORE YOU BEGIN Read and understand this T PUMP OPERATOR S MANUAL and all SAFETY PRECAUTIONS see page 3 prior to using the T Pump Only qualified medical service personnel should repair or perform function tests on the T Pump Contact your dealer or Gaymar s Technical Service Department for assistance Telephone 716 662 2551 RECEIVING INSPECTION Check the shipping carton for damage immediately upon receipt If package damage is discovered the device should be unpacked with the carrier s agent present Any claims for shortage or damage must
8. WITH RESPECT TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 AND ASTM F 2196 IEC 60601 1 2 www gaymar com 19 WARRANTIES OPERATOR S MANUAL TP702 T PUMP WARRANTIES PUMP PAD SINGLE PATIENT USE PAD REUSABLE PARTS GAYMAR equipment and products are warranted against defects in material and workmanship under normal use and operation from the date of purchase for the time periods listed below for the respective equipment and products Except for such warranty GAYMAR disclaims all other expressed and or implied warranties including but not limited to the implied warranties of merchantability and of fitness for a particular purpose All labor performed and parts provided free of charge for a period of one 1 full year from the date of purchase provided the equipment is returned with prior authorization prepaid to GAYMAR Industries Free replacement of product where defects in materials and or workmanship are evident at time of delivery provided the product is returned with prior authorization prepaid to GAYMAR Industries Free replacement of product where defects in materials and or workmanship occur within 90 days from date of delivery provided the product is returned with prior authorization prepaid to GAYMAR Industries Defective parts will be exchanged free of charge where defects in materials and or workmanship occur within 90 days from date of delivery provided t
9. dicator to toggle through the four setpoints Temperatures are identified in C and F 42 P 107 So 50 CD Setpoint Prevents tampering Lock Press and hold for 2 seconds to lock or unlock the setpoint Therapy Cycles Continuous cycle 30 minute cycle 20 minute cycle Green indicates the unit is on Yellow indicates power is supplied to the unit but the unit is not on 6 www gaymar com OPERATOR S MANUAL TOTU OPERATING INSTRUCTIONS CLIK TITE CONNECTORS The TP702 T Pump is supplied with Clik Tite connectors To connect and disconnect Clik Tite connectors from hose to pad 1 Insert male fittings into female fittings with a twisting motion Figures 4A and 4B Note Refer to Figure 1B on page 4 when connecting multiple pads 2 When fittings are fully inserted snap locking ring into place Figures 4C and 4D 3 To disconnect reverse the procedure 4 To open or close the hose pinch clamps e Open the clamp by pushing the serrated end Figure 5B e Close the clamp by pressing the clamp together Figure 5C Figure 4 Clik Tite Connectors www gaymar com 7 OPERATING INSTRUCTIONS OPERATOR S MANUAL TP702 T PUMP START UP PROCEDURE SHUTDOWN PROCEDURE 10 i p Before filling the pump always attach a pad to the connector hose or close the clamps on the connector hose ends Make sure that there are no kinks in the hose or pad Open the hose clamps Open the fill
10. he electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the TP702 is used exceeds the applicable RF compliance level above the TP702 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the TP702 b Over the frequency range 150 kHz to 80 MHz field strengths are less than 3 V m Guidance and manufacturer s declaration electromagnetic emissions The TP702 is intended for use in the electromagnetic environment specified below The customer or the user of the TP702 should assure that it is used in such an environment RF emissions CISPR 11 The TP702 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The TP702 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions Class A IEC 61000 3 2 Voltage fluctuations Flicker Complies emissions IEC 61000 3 3 www gaymar com 23 T PUMPS AND T PADS ARE MADE IN THE USA ae SS GaymMmar Your patients Our commitment GAYMAR INDUSTRIES INC 10 Centre Dri
11. he parts are returned with prior authorization prepaid to GAYMAR Industries 20 www gaymar com OPERATOR S MANUAL TP702 T PUMP EMC INFORMATION ENGLISH The TP702 complies with EN60601 1 2 2001 Second Edition CISPR Classified as Class B Group 1 ISM equipment Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following pages Portable and mobile RF communications equipment can effect Medical Electrical Equipment CAUTION The TP702 should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the TP702 should be observed to verify normal operation in the configuration in which it will be used www gaymar com 21 OPERATOR S MANUAL EMC INFORMATION 707 BUN EMC INFORMATION Guidance and manufacturer s declaration electromagnetic immunity The TP702 is suitable for use in the electromagnetic environment specified below The customer or the user of the TP702 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic Discharge ESD IEC 61000 4 2 Electrical fast Transient burst IEC61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power frequency 50 60 Hz magnet
12. heater will restart The patient may feel a cooling in the pad due to the drop in temperature in the hoses Time Therapy LED goes solid and the Setpoint LED will flash until the system temperature reaches the Setpoint Once the Setpoint is reached the Red LED flashes with a short audible beep to let the operator user know the next therapy cycle has started TROUBLESHOOTING TROUBLESHOOTING OPERATOR S MANUAL TP702 T PUMP Possible Cause T Pump will not turn on T Pump will not pump Flow indicator light is on Warning indicator amp Audible alarm Flash Beep 10 The electrical cord is not plugged into a properly grounded Hospital Grade receptacle Water level is low or reservoir is empty Water flow to pad or hose is restricted Clamp is closed Water level is low or reservoir is empty T Pump is filled with water that is too hot A High Heat 107 F 42 C or Cooling Setpoint was selected 50 F 7 10 C Loss of power while unit was in a Therapy mode Possible Power Fail Unit is running after a 20 or 30 minute Off therapy cycle period has reached the desired Setpoint and is now timing the 20 or 30 minute On cycle period The unit just went into or came out of Lock mode Safety Circuit Function Test has started Note This would be followed with the unit starting while still in Standby mode See the Functional Check section on page 18 for details Insert the
13. ic field IEC 61000 4 8 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 Ut 95 dipUt for 0 5 cycle 40 Ut 60 dop in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles lt 5 Ut gt 95 dip in Ut for 5 sec 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 Ut 95 dipUt for 0 5 cycle 40 Ut 60 dop in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles lt 5 Ut gt 95 dip in Ut for 5 sec Note U is the a c mains voltage prior to application of the test level Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality is that of a typical commercial and or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the TP702 requires continued operation during power mains interruptions it is recommended that the device be powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typical locationinatypical commercial and or hospital environment Recommended separation distances betwee
14. it TP22B Mul T Pad 15 w x 22 l 38cm x 56cm All Polymer 10 per carton TP22C Mul T Pad 15 w x 22 1 38cm x 56cm Nonwoven fabric on one side pliable polymer on the other side 1 per carton TP3E Mul T Pad 3 w x 23 l 8cm x 58cm Nonwoven fabric on one side pliable polymer on the other side 10 per carton TP12E Mul T Pad 13 w x 18 l 33cm x 46cm Nonwoven fabric on one side pliable polymer on the other side 20 per carton TP22E Mul T Pad 15 w x 22 l 38cm x 56cm Nonwoven fabric on one side pliable polymer on the other side 20 per carton TP26E Mul T Pad 18 w x 26 46cm x 66cm Nonwoven fabric on one side pliable polymer on the other side 10 per carton TP22G Mul T Pad 15 w x 22 l 38cm x 56cm Heavy Polymer Reuseable 10 per carton To order any of these products contact your dealer or Gaymar s International Department at International 1 716 662 8636 Fax 1 716 662 0730 Or visit our website at www gaymar com www gaymar com 3 OPERATOR S MANUAL INDICATIONS AND CONTRAINDICATIONS 07 UME INDICATIONS FOR USE Localized temperature therapy is recommended in treating the following applications Orthopedic Conditions such as acute injuries chronic pain lower back pain muscle spasm and strains Skin Trauma such as abscesses boils bruises burns and contusions Cold Indications muscle spasm contusions tendonitis pain management Other conditions such as chronic arthr
15. itis neuritis phlebitis tendonitis and I V infiltration infection and localized pain Other applications only as prescribed OK for Use With e Non acute traumatized tissue e Impaired mental status e Insensate body surface e O therapy However if oxygen tent is in use do not use pump inside tent CLINICAL Heating contraindications are CONTRAINDICATIONS e Application to a body surface with compromised blood flow Ischemia area under pressure arterial insufficiency Application to a patient with an increased tendency to bleeding aggravates potential for hemorrhage Application to a body surface with possibility of malignancy tissue metabolism is increased and therefore the growth potential of the malignant tissues Treatment of hematoma within first 24 48 hours potential for re bleeding and hemorrhage Recent sprain or fracture acute inflammatory response e In combination with topical solutions whose toxicity may be affected by the application of heat In combination with other heat sources Cooling contraindications are e Application to a body surface with compromised blood flow or hypersensitivity to cold Ischemia Reynauds Syndrome 4 www gaymar com OPERATOR S MANUAL TP702 T PUMP FEATURES PUMP FEATURES Easy to Use See KEYPAD FEATURES on page 8 Keypad Attached hose 10 ft 305 cm dual hose Connectors allow pads to be connected to the pump Figures 1A and 1B page 4
16. luate patient response to temperature application Check patient s skin for adverse reactions every 30 minutes or as directed by the prescriber Failure to adhere to these warnings could result in patient injury The following Groups Conditions require additional surveillance Group Condition at risk Potential Injury Pediatric patients The portion of an Hyperthermia infant s skin surface in contact with Hypothermia a pad in relationship to their body mass can potentially affect their body temperature Patients with impaired circulation Areas of application under pressure Only qualified medical service personnel should repair the T Pump Improper repair may result in death or serious injury equipment damage or malfunction Use T Pump TP702 controls with Mul T Pads For catalog numbers and descriptions see page 5 Do not place additional heat sources between the patient and pad Skin damage may result CAUTION Federal law restricts this device to sale by or on the order of a physician Do not cover the control unit with blankets pillows or other insulating materials Air flow is required to maintain system performance www gaymar com 1 INTRODUCTION OPERATOR S MANUAL TP702 T PUMP INTRODUCTION Connecting the Pads Heat therapy is effective in the dilation of blood vessels thereby increasing the blood flow to the heated area Heat therapy has a variety of uses the most common being trea
17. ly grounded Hospital Grade receptacle Press the On Standby button The T Pump performs its self test of the lights and audible alarm with light pattern and short audible beep The T Pump goes to Standby mode with only the Standby light on Press and hold the Temperature Setpoint button While still holding the Temperature Setpoint button press the On Standby button Hold both buttons simultaneously for 3 seconds and then release The T Pump gives along audible beep and the Warning light flashes TheT Pump goes into Operating Temperature Function Test mode and the e T Pump starts pumping with the Standby light on with a flow of 9 GPH minimum e 107 F 42 C setting is selected e Continuous light is on e System disables the Wave feature to maintain a steady 107 F supply of water to the pad The T Pump controls to 107 F 42 C at the inlet to the pad for 15 minutes the duration of the test After 12 minutes verify with the TPT9 that the Temperature is at 107 F 2 F 42 C 1 C and flow is 9 GPH minimum Note If the T Pump does not reach the temperature and flow press the On Standby button unplug the unit check the pad and hoses for kinks and start over After 15 minutes observe that the T Pump starts the Over Temperature Safety Circuit Test The T Pump gives a long beep with the Warning light on This signals the start of the Over Temperature Safety Circuit Test FUNCTIONAL CHECK OPERAT
18. n portable and mobile RF communications equipment and the TP702 The TP702 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the TP702 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the TP702 as recommended below according to the maximum output power of the communications equipment Rated maximum output power Separation distance according to frequency of transmitter a imltiey 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 22 www gaymar com OPERATOR S MANUAL TP702 T PUMP EMC INFORMATION TP702 EMC Information The TP702 is suitable for use in the electromagnetic environment specified below The customer or the user of the TP702 should assure that it is used in such an envir
19. nect the hoses together where applicable OPERATOR S MANUAL STORAGE CLEANING PT pUME Institutional Note Clean and change the water monthly or more often depending on Hospital Cleaning aca instructions Clean the outer surfaces of the T Pump with one of the following e A damp cloth and soapy water e A spray cleaner such as Fantastik e A mild abrasive cleanser without bleach 1 Prepare a germicidal solution according to the manufacturer s instructions Use AirKem A 33 available from Ecolabs Inc 370 Wabasha St Paul MN 35102 phone 1 800 247 5362 or from GAYMAR product catalog MTA33 2 Drain the pump 3 Connect hose set together 4 Fill the reservoir to the Heating water line on the back of the reservoir 5 Select the 95 F 35 C temperature setpoint on the keypad 6 Start the T Pump and circulate the solution for one hour 7 Drain the solution and refill the pump with clean water NOTE In a home environment perform only step 2 and the refill instructions in step 7 Pads Accessories Only use Mul T Pads The unique button design allows optimal water flow and provides trouble free operation when the pad is folded This reduces the number of different sizes of pads your facility must keep in inventory The Mul T Pads with Clik Tite connectors can be interconnected to provide therapy to more than one body site at a time Figure 1B page 4 Refer to Catalog Descriptions on page 5 fora list
20. of various pads and ordering information 14 www gaymar com OPERATOR S MANUAL TP702 T PUMP MAINTENANCE MAINTENANCE SERVICE A WARNING Only qualified medical service personnel should repair the T Pump in accordance with the Service Manual Improper repair mayresultindeathorseriousinjury equipmentdamage ormalfunction Refer to functional check To obtain a copy of the Service Manual contactyour local dealder or Gaymar s International Departmentor visit our web site www gaymar com www gaymar com 15 FUNCTIONAL CHECK FUNCTIONAL CHECK Operating Temperatures amp Over Temperature Safety Circuit Check Interval Required Tools Test Setup 16 OPERATOR S MANUAL TP702 T PUMP This section provides a check of the T Pump s temperature control and over temperature safety circuit Follow the steps in this section carefully paying particular attention to each step and its expected result If at any time the expected result can not be verified press the On Standby button to stop the test then restart the procedure If after a second attempt the expected result can not be verified press the On Standby button to stop the test then unplug the T Pump and call your dealer or Gaymar s Technical Service Department for assistance A WARNING Only qualified service personnel should perform this Functional Check Improperly following the test procedure may result in equipment damage Do not perform this
21. onment Immunity IEC 60601 test Compliance Electromagnetic environment guidance test level level Portable and mobile RF communications equipment should be used no Conducted RE 3 Vrms closer to any part of the TP702 including cables than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter IEC 61000 4 6 150 kHz to 80 MHz Recommended Separation Distance Z J P Radiated RF 3 V m d 1 2 IEC 61000 4 3 80MHzto2 5 d 1 2 4P 80 MHz to 800 MHz GH f d 2 3 4P 800 MHzto 25 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey Should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Q NOTE 1 At 80 MHz and 800 Mhz the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radiobroadcast and TV broadcast cannot be predicted theoretically with accuracy To assess t
22. ow indicator light is on above Ice is depleted Drain excess water to Cooling water line and fill remainder of reservoir with ice Temperature or Therapy Time The buttons have been locked Press and hold the lock button for buttons do not work two seconds Water leaks from hose Damaged O ring Replace Clik Tite connector connectors Locking ring on Clik Tite connector Snap Clik Tite connector shut is not snapped into place See Figure 4 on page 9 12 www gaymar com OPERATOR S MANUAL TP702 T PUMP STORAGE CLEANING STORAGE CLEANING Storage Short term Less than 1 day Storage Long term Draining www gaymar com oa ee PY P Close the hose clamps Disconnect the pad Connect ends of the connector hoses together where applicable Open the hose clamps Leave water in the reservoir Coil and fasten the hose using the tube set strap and wrap the power cord around the unit Drain the pump See instructions below Coil the hose rather than folding it to prevent hose kinks Fasten the hose using the tube set strap and wrap the power cord around the unit Disconnect the T Pump from AC power Clamp the hose clamps Disconnect the pad or hoses from one another keeping hoses at or above the level of the T Pump Open the hose clamps Remove the fill cap and invert the T Pump over a sink When all fluid has drained from the hoses and reservoir replace the fill Cap Con
23. plug fully into the properly grounded Hospital Grade receptacle Refill with room temperature water to proper level Straighten the hose Open the clamp Refill with room temperature water to proper level Refill with room temperature water to proper level Indication only A Setpoint outside body temperature range is selected Insert the plug fully into the receptacle place the unit into Standby mode then unplug the T Pump If power is removed while unit is in On Mode the Power Fail alarm will beep for approximately 10 minutes Indication only to indicate an On Therapy cycle period is timing Indication only Note This should only be performed by medical service personnel If this mode was started in error press the On Standby button to stop the test and go back to Standby mode with the pump off www gaymar com OPERATOR S MANUAL TP702 T PUMP TROUBLESHOOTING Possible Cause T Pump running with the Standby light on Warning indicator on with unit in Standby mode Flow indicator and Standby indicator are on with T Pump not pumping Temperature Setpoint light blinking www gaymar com Safety Circuit Function Test has started Note This would be followed with the unit starting while still in Standby mode See the Functional Check section on page 18 for details Unit shut down in an over temperature condition Safety Circuit Function Test completed successfull
24. thin 5 minutes The test is completed 14 Shut down the T Pump as described on page 10 18 www gaymar com OPERATOR S MANUAL TP702 T PUMP SPECIFICATIONS SPECIFICATIONS Classification Size approx Weight Reservoir capacity Flow rate Ambient operating temperature Environmental conditions for transport and storage Tem perature setpoi nts Average temperature accuracy Maximum Contact Surface Temperature High Limit Safety Temperature Class equipment with Type BF applied part suitable for continuous operation Not classified for protection against ingress of liquid Not classified for use in the presence of flammable anesthetics 11 5 x 8 x 8 29 2cm x 20 3cm x 20 3cm 6 5 Ibs 2 9 kg when empty 9 Ibs 4 0 kg with unit filled with water to heating level 93 oz 2 75 1 maximum 9 gph 34 Iph minimum with pad attached 60 F to 90 F 15 6 C to 32 2 C 20 F to 120 F 28 C to 48 C At uncontrolled RH TP702 Professional 107 F 42 C 100 F 38 C 95 F 35 C 50 F 10 C 2 F at 107 F 1 C at 42 C 107 F 42 C 111 F to 118 F 44 C to 48 C Modular adaptable to Country of use Current leakage 100 microamperes maximum Ground resistance Ground resistance Electrical requirements Voltage VAC Frequency Hz Current amps Certifications 05 ohmmax 5 05 ohmmax max 230 10 50 Hz 2 0 amperes MEDICAL ELECTRICAL EQUIPMENT
25. tment of aches and pains in joints and muscles Cooling therapy assists in vasoconstriction decreasing blood flow and decreasing the metabolism in the affected area Cooling therapy is applied in the acute phase of injury minimizing blood loss inflammation of the tissue and can be effective in pain management The GAYMAR T Pump Localized Temperature Therapy System provides therapy by warming or cooling the enclosed water and circulating it through the Gaymar Mul T Pad The pad is connected to the Gaymar T Pump with easy to use Clik Tite connectors The Mul T Pad provides the interface for delivering the temperature therapy The unique button design allows water to flow and provides trouble free operation when the pad is folded to form a customized fit This reduces the number of pads your facility must keep in inventory The pads are applied to the part of the body requiring therapy and the circulating water maintains the pad at the setpoint temperature The setpoint temperature can be locked to prevent tampering Figure 1A Localized Temperature Therapy System with Single Pad with the TP702 The Mul T Pads can be interconnected using Clik Tite connectors to provide therapy to more than one body site at a time Figure 1B Localized Temperature Therapy System with Multiple Pads www gaymar com OPERATOR S MANUAL TP702 T PUMP INTRODUCTION Catalog Descriptions Catalog Product Name TP702 Professional Control Un
26. ve Orchard Park NY 14127 2295 Phone 716 662 8636 Fax 716 662 0730
27. y Unit detected a Flow warning for more than 5 minutes thus goes to standby Unit is warming up to the selected setpoint Unit is in Cooling mode for longer than 40 minutes Important This should only be performed by medical service personnel If this mode was started in error press the On Standby button to stop the test and go back to Standby mode with the pump off Empty the reservoir and refill with room temperature water Make sure all clamps are open Press the On Standby button Verify flow through the pad The Warning light will turn off within 5 minutes Allow the system to cool for 60 minutes or until the Warning light turns off Reference Flow indicator light is on above Correct the problem and press the On Standby to put the unit back into Run mode Indication only Follow the shutdown procedure Drain the water in reservoir to ice fill level and refill with ice Follow the start up procedure OPERATOR S MANUAL TROUBLESHOOTING TP702 T PUMP Possible Cause Both the Temperature and Unit is in Off Therapy cycle time Indication only Therapy Cycle Setpoint lights are blinking T Pump will not heat Reservoir is empty Refill with room temperature water to proper level Flow is blocked Reference Flow indicator light is on above T Pump will not cool Reservoir is empty Refill with room temperature water to proper level Flow is blocked Reference Fl
Download Pdf Manuals
Related Search
Related Contents
OASIS 3D INVERTER SERIES MODE D`EMPLOI DE LA TAXE DE SEJOUR 4600 Gas Plus - User Manual NuTone WS2 User's Manual IOGear GCS1802 User's Manual CENTRON® Meter Technical Reference Guide Hiper HPU-3S350 power supply unit DEIMOS BT H QSCDEIMOS BT H QSC IPI, Plug & Play PC Coprocessor, GFK fiche d`identification ulm de classe 1 b 1 0 1 sf 0 2 8 1 1 e Copyright © All rights reserved.