1.     O Battery inert  Note that it is essential that the battery is put to this condition  before international transportation     The menu is accessed by a three second press of the unlock key  The arrow key  scrolls to the functions shown above  and the mode select key allows the function to  be toggled on or off  or selected  For language selection  press and hold the mode  key until the required language is displayed  Releasing the mode key will then set the  display for the language     Pressing the unlock key returns the device to standby mode     IV POLE ATTACHMENT  VENTURI    COMPACT only     The VENTURI    COMPACT is supplied with an IV pole bracket  to provide safe  deployment and additional mobility     e To fit the IV pole bracket  engage the three screws on the bracket into the  corresponding three holes on the underside of the VENTURI    COMPACT  The  screws can be hand tightened  and do not require high torque or a tool    O The fit the bracket with VENTURI    COMPACT attached  use the black hand knob  to secure it to the vertical IV pole  The bracket will fit to IV poles between 19mm  and 26mm diameter    e When dismounting the IV pole bracket from VENTURI    COMPACT  note that the  three screws are captive  and will remain on the bracket when unfastened to  prevent their loss     An IV pole attachment bracket is available as an optional accessory for the VENTURI     AVANTI  Instructions for its use are included with the part     Battery Information      A full
2.   10  C   14  F      e Do not dispose of the battery pack in a fire  or in municipal waste  It should be  recycled  disposed of properly or returned to your dealer for disposal in accordance  with the WEEE directive     O The battery pack is not serviceable and should be replaced if faulty  indicated by   0 C Batt  fault displayed in place of battery charge status icon      CHARGING THE BATTERY    If not fully charged  the battery will automatically charge when the power unit is  plugged into mains power  The battery will charge at the optimum level depending on  history of previous mains power usage and charging over the previous week     REPLACING THE BATTERY  for VENTURI    AVANTI only     Should it be necessary to replace the battery  unfasten the battery cover screw and  detach the cover  The battery module will drop out from its installed position  Replace  the battery pack  note it can only be engaged in only one orientation  Replace the  cover and re fasten the screw  Do not over tighten     NB  For VENTURI    COMPACT  the battery is installed inside the unit and is not    accessible by users  Battery replacement should only be carried out by qualified  service personnel     10    Care and Maintenance    POWER UNIT    Always disconnect the VENTURI    AVANTI  VENTURI    COMPACT power unit from the  electricity supply  if applicable  before cleaning  The power unit can be wiped down  with a damp cloth or alcohol wipe  Do not use solvents  CAUTION  Do not immerse  in wat
3.  default to continuous therapy mode  To switch between therapy modes   press and hold the THERAPY MODE button until power unit beeps to confirm change  of mode  When in intermittent therapy mode  the power unit will provide vacuum  therapy for periods of 5 minutes followed by a 2 minute rest period  The selected  therapy mode is displayed on the power unit screen and can be altered in either  stand by or  unlocked  run modes     MUTE  Press to silence alarm and to clear alarm display from display screen     UNLOCK   LIGHT     The VENTURI    AVANTI   VENTURI    COMPACT power unit will automatically  lock 1 minute after last button operation when running to prevent inadvertent    operation of button functions  except MUTE   as indicated by  L  on the  display screen  Press and hold the UNLOCK button until power unit beeps if further  button operation is needed  i e  change of therapy mode or vacuum level  or returning  power unit to stand by mode   The power unit will lock again 1 minute after last  button operation  Pressing this button will also illuminate the display screen for 10  seconds if power unit is in battery operation  screen will always be illuminated in  mains operation      USER SELECTABLE FUNCTIONS  VENTURI    COMPACT only     The following functions are selectable in a user menu    O Language selection  e Tilt alarm  switch on or off    User Information  continued     O Internal contamination sensor  switch on or off  e Tube constriction alarm  switch on or off
4.  mains outlet in wall    NB  The battery will charge when the unit is connected to mains power   indicated by battery charge status icon on display screen scrolling  from left to right  and provides automatic power back up if mains  power fails  It is recommended to use mains power when convenient  to do so as this will ensure the battery is fully charged when needed     Press RUN STOP button to invoke and display STAND BY mode  the power  unit will beep and therapy mode  operating pressure and battery charge  status will be displayed      Power unit will default to continuous therapy mode at 80mmHg  To switch  between continuous and intermittent therapy modes  press and hold the  THERAPY MODE button until power units beeps to confirm change of mode   Adjust vacuum level if required using the UP and DOWN arrow buttons     Press RUN STOP button again to initialise and run the power unit     NB  Vacuum level and therapy mode can be adjusted when in stand by mode  and for up to 1 minute after power unit is running  Power unit will automatically  lock 1 minute after last button operation when running to prevent inadvertent  operation of button functions  except MUTE   as indicated by  L  on the display  screen  Press and hold the UNLOCK button until power unit beeps if further  button operation is needed  Display screen is only illuminated for a short  period after button operation in battery mode          o       a    6  Once power unit is running  observe the wound site  The dr
5. V NTURI Tait ey    NEGATIVE PRESSURE WOUND THERAPY ME ES    VENTURI    AVANTI   VENTURI    COMPACT  Negative Pressure Wound Therapy Systems       User Manual    Contents    Introduction  List of Components  Cautions and Warnings Information    How to Apply NPWT      Operating the Vacuum Power Unit  User Information  Battery Information  Care and Maintenance  Warning and Fault Alarms    Specification    TALLEY    MEDICAL    Page    11  11    13    Introduction    Thank you for choosing to use the VENTURI    AVANTI   VENTURI    COMPACT  Negative Pressure Wound Therapy  NPWT  system from Talley Medical  In doing  so you have selected an efficient  competitively priced product for the treatment  of many wounds including pressure ulcers  dehisced surgical wounds  diabetic   neuropathic ulcers  venous leg ulcers  post surgical wounds  sinus drainage and  management  traumatic wounds and pre  and post op flaps grafts     The VENTURI    AVANTI   VENTURI    COMPACT NPWT system offers a choice of  continuous or intermittent therapy modes and features a lightweight  versatile  vacuum power unit which benefits from dual power technology  offering a  seamless choice of mains or battery operation  The integral battery is charge   optimised  and provides long lasting power back up when needed  The  battery operation option allows the system to function away from a mains  power supply for extended periods of time  allowing the patient full mobility  during therapy  if required  optional carr
6. acket  The VENTURI    COMPACT is supplied with a bracket for IV  pole attachment  To allow patient full mobility during therapy  both power units  can also be placed in the carry cases provided  In all instances the power unit   canister assembly should be kept upright during use to ensure correct operation   The VENTURI    COMPACT will display a warning if the unit is over tilted     User Information  continued     BUTTON FUNCTIONS    RUN STOP     Press to invoke stand by mode prior to running power unit  Press again to run  power unit  Press and hold whilst power unit is running  and unlocked  to cease  operation and return to stand by mode  In battery operation the power unit  will then power off after one minute of inactivity  If using mains power  switch off by  disconnecting power cable from power unit or turning off mains power     VACUUM LEVEL   y   a  Vacuum level can be adjusted when in stand by mode and for up  to 1 minute after last button operation when power unit is running  using the UP and DOWN arrow buttons  Pressure can be adjusted  in 5mmHg increments between 10mmHg and 120mmHg according to treatment  requirements  the most widely used being between 60 80mmHg  The power unit will  begin operation at the default pressure of 80mmHg  The selected vacuum level is    shown on the display screen     THERAPY MODE  The VENTURI    AVANTI   VENTURI    COMPACT power unit offers a choice of    continuous or intermittent therapy modes  On power on  the power unit will  
7. afety information     How to Apply NPWT    NB  A wound care set must be used with the VENTURI    AVANTI  VENTURI    COMPACT  system to carry out NPWT  see page 4 for available options   Please refer to the  VENTURI    Clinical Guidelines Manual for information on application of wound care  sets and other clinical advice and safety information     CAUTION  The medical professional is responsible for using his her best medical  judgment when using this system  Prior to use  the medical professional s  treating  the wound must assess how to best use the system for an individual wound     1  Remove all packaging from the power unit     How to Apply NPWT  continued     2  If not already in place  attach canister to flat face of power unit by matching up  the rear location pegs and rotating locking knob 1 4 turn clockwise to secure   Ensure canister is correctly located and secured otherwise NO CANISTER alarm  will appear and power unit will not operate     Prepare and seal wound as described in wound care set instructions     4  Attach wound care set connection tubing to the VENTURI    AVANTI  VENTURI     COMPACT power unit canister by lining up locator stud on tubing connector with  notch on canister tubing receptacle located on top corner of canister  twisting  clockwise to lock     5  OPERATING THE VACUUM POWER UNIT     a  If using mains power  plug the smaller end of the power cable into the side  of the VENTURI    AVANTI  VENTURI    COMPACT power unit  and the other end  into
8. eached its capacity and  should be changed  The power unit will cease to run and the visual audible alarm  will continue until the RUN STOP button is pressed or the canister is removed  both  of which cancels the alarm and returns the power unit to stand by mode  Change  canister as detailed on page 7     NB  The audible  Canister Full  alarm can be silenced by pressing the MUTE button   however if the canister is not removed changed the audible alarm will reoccur after  10 minutes     11    Warning and Fault Alarms  continued     Low Vacuum   will alert if vacuum pressure falls below minimum allowable levels   Power unit will continue to run whilst this warning is displayed  Check that wound  dressing is completely sealed and that all tubing connections are secure  Press the  MUTE button to silence clear alarm  Note that alarm will reoccur if fault persists  If  fault re occurs  contact Talley Medical     Tube Constriction   this warning activates when pump demand abruptly ceases   Note that a certain level of flow in the drain is required for the alarm to be active   The warning will occur for example in the case of a blocked tube  kinked tube or ifthe  dressing is obstructed  Note that the tube constriction alarm can be selected on or  off as required  see    User Selectable Functions    on page 8   Default is on     Tilt Alarm  VENTURI    COMPACT only   this warning activates when the device is in  active mode and is placed at an angle that could affect the canister full 
9. er     The unit contains no user serviceable parts  and should only be serviced by a  competent electrical technician  or returned to Talley Medical or your local authorised  dealer  All Talley Medical products should be serviced regularly by Talley Medical or  authorised dealer in order to comply with warranty conditions     NOTE  Before returning equipment to Talley Medical for service  be sure that it has  been properly cleaned and disinfected     CANISTERS    Canisters are disposable and intended for single use only     Warning and Fault Alarms    All audible alarms can be silenced by pressing the MUTE button once  Should any  fault occur  press the MUTE button to reset the power unit  If fault remains re occurs   contact Talley Medical     WARNING ALARMS    No Canister   will alert if canister is missing or is not correctly fitted  The power unit  will fail to operate whilst this warning is displayed  Check that canister is correctly  located and secured  as detailed on page 5     No Battery   will display if battery is not correctly fitted  The power unit will only  operate using mains power whilst this warning is displayed  Check battery installation   NB  battery can only be engaged in one orientation      Low Battery  only appears during battery operation    the power unit will continue  to run whilst this warning is displayed  Press the MUTE button to silence clear alarm   Plug into mains power to charge     Canister Full   will alert to warn that the canister has r
10. essing should  contract noticeably  become firm to the touch and  raisin like   If the dressing  fails to contract  the dressing has not been completely sealed  Reinforce the  dressing seal and or adjust the drain and initiate suction again     7  To change or remove dressing  unlock power unit  press and hold UNLOCK until  power unit beeps   then press and hold the RUN STOP button until power unit  beeps three times to return to stand by mode  Clamp wound care set tubing  and remove by turning anticlockwise and lifting out of tubing receptacle on  canister  Dispose of used wound care set according to local clinical waste policy   If required  apply new wound care set and continue NPWT     6    How to Apply NPWT  continued     10     Canisters should be replaced as required or weekly  To change canister   make sure power unit is in stand by mode  if still running  press and hold the  UNLOCK button  followed by the RUN STOP button   Clamp connection tubing  and remove by turning anticlockwise and lifting out of tubing receptacle   this can be reconnected to new canister and unclamped if wound dressing  is not being changed   Remove sealing plug from its location on top corner  of canister and use to cap tubing receptacle to seal in contents  Rotate  locking knob 1 4 turn anticlockwise and remove canister  Dispose of used  canister according to local clinical waste policy  If continuing NPWT  attach  new canister and connect wound care set tubing as previously described     To 
11. formation        Service due   Contact Talley Medical to arrange service  e Uncalibrated   Contact Talley Medical for recalibration    The power unit will continue to run whilst these warnings are displayed     If you have any queries relating to this system please contact Talley Medical or your  local authorised dealer     12    Specification    POWER UNIT VENTURI    AVANTI  Model Ref  Venturi v Il TG600 07  Construction  Flame retardant ABS  Dimensions  210mm x 205mm x 105mm  Weight  2 2 kg   Mains Cable  3 metres   Mains Input Voltage  230V     10  50Hz  Operating Voltage  6V   Peak Current  850mA RMS   Max  Load Wattage  7W    Continuous  default  therapy  or intermittent  10 to 120mmHg    Vacuum Application     Pressure Range     CANISTER   Construction    Dimensions  212mm x 204mm x 32mm  Capacity  600ml    VENTURI    COMPACT    Venturi v Il TG600 08  Flame retardant ABS  161mm x 155mm x 90mm  1 0 kg  3 metres  230V     10  50Hz  3 7V  450mA RMS  12W  Continuous  default  therapy  or intermittent  10 to 120mmHg    ABS Trans smoked grey  textured  includes 1x 88  dissolvable sachet of desiccant     156mm x 163mm x 45mm  300ml    Talley Medical products are manufactured to comply with BSI  IEC  UL and other  European safety standards  Talley Medical design and manufacture products to  conform to the requirements of ISO9001 2008  IS013485 2003 and Directive     93 42 EEC  Annex Il  excluding Section 4      Talley Medical reserves the right to modify the specification of an
12. indication   The alarm delays for 5 seconds  and is self muting when the unit is returned to  upright  Note that the tilt warning alarm can be selected to be on or off as required   see    User Selectable Functions    on page 8   Default is on     FAULT ALARMS    O C Batt  will appear instead of battery charge status icon    indicates that the  battery cannot be used as it is unable to take a charge  The power unit will only  operate using mains power whilst this warning is displayed  Contact Talley Medical to  order a replacement battery     CCT Fail  will appear instead of battery charge status icon    indicates that the  charging facility is not functioning due to a PCB fault  The power unit will only operate  using mains power whilst this warning is displayed  If fault remains re occurs  contact  Talley Medical     If any of the following faults are displayed the power unit will cease to operate   PUMP  FAULT  TRIAC FAULT  PUMP OPEN  PUMP SHORT  RELEASE KEY  BATTERY FAULT   Should any of these faults occur  press the MUTE button to reset the power unit  If  fault remains re occurs  contact Talley Medical     EMI Fault   indicates that the unit detects the pressure sensor amplifier is adversely  affected by external RF fields  The power unit will continue to run whilst this warning  is displayed  Press the MUTE button to silence alarm  This alarm will clear when  interference ceases     The VENTURI    AVANTI  VENTURI    COMPACT power unit may also display the following  in
13. stop the power unit  press and hold the UNLOCK button until power unit  beeps and  L  clears from display screen  Then press and hold the RUN STOP  button until power unit beeps three times to return to stand by mode  In battery  operation the power unit will then power off after one minute of inactivity  If  using mains power  switch off by disconnecting power cable from power unit or  turning off mains power     Place the user manual in a safe place for future use     User Information    The VENTURI    AVANTI  VENTURI    COMPACT power unit will automatically lock 1  minute after last button operation when running to prevent inadvertent operation  of button functions  except MUTE   as indicated by  L  on the display screen  Press  and hold the UNLOCK button until power unit beeps if further button operation  is needed  i e  change of therapy mode or vacuum level  or returning power unit  to stand by mode   The power unit will lock again 1 minute after last button  operation     Please note that when running  the VENTURI    AVANTI   VENTURI    COMPACT  power unit must be unlocked before it can be returned to stand by mode  i e   the power unit cannot be put into stand by mode when the    L is displayed on the  screen     During operation  the power unit should be placed upright on a horizontal surface  with the display uppermost  The VENTURI    AVANTI power unit can be placed in  the optional metal holder for use at the bedside or attached to an IV pole using  the optional br
14. this product ensure that     O the electricity supply  if applicable  is of the type indicated on the power unit   O the mains lead  if applicable  is free from damage and is positioned so as not to  cause an obstruction   O the system is not used in the presence of flammable anaesthetics    Do not place power unit on or near a heat source     Never use the mains adaptor charging unit whilst placed on top of or near to material  which is flammable or can be damaged by heat     The equipment conforms to IEC 60601 1 for electromagnetic interference  however  should the equipment be subjected to electromagnetic interference outside this  then the unit should be reset  In the unlikely event of interference with other devices  please consider moving the VENTURI    AVANTI  VENTURI    COMPACT power unit     When the canister is not fitted on the VENTURI     AVANTI   VENTURI    COMPACT power unit it is  essential that the aperture on the rear face of the  unit  Fig  1  is not blocked or covered  e g  with a  label   Refer to Service Manual for full details     CAUTION  The medical professional is responsible for  using his her best medical judgement when using  this system  Prior to use  the medical professional s   treating the wound must assess how to best use Figure shows VENTURI    AVANTI   the system for an individual wound  VENTER COMEAG TIS similar       Please see the VENTURI    Clinical Guidelines Manual for indications and  contraindications for use  together with further s
15. y case with shoulder strap available      NPWT is applied utilising the VENTURI    wound care set which includes gauze   which is applied moistened with saline   a silicone drain and transparent adhesive  film  A choice of flat or channel drains are available  together with a choice of  gauze sizes  according to the wound type being treated     The VENTURI    AVANTI   VENTURI    COMPACT NPWT system will benefit from  careful installation and use  providing a long and effective service life  Please read  this user manual in order to achieve the best possible results     List of Components    Your VENTURI    AVANTI VENTURI    COMPACT NPWT system should comprise the  following items   please ensure you have all of these before installation     O VENTURI    AVANTI VENTURI    COMPACT vacuum power unit  O Mains lead with integral power supply  O Canister  supplied fitted to power unit      capacity 600ml  VENTURI    AVANTI   300ml  VENTURI    COMPACT     ALSO AVAILABLE   O Wound Care Set   flat drain  standard  O Adhesive Gel Patch   O Wound Care Set   flat drain  large  O Y connector   O Wound Care Set   flat drain  abdominal  O Spare battery   O Wound Care Set   channel drain e 12V Vehicle adaptor charger  O 300ml 600ml canister complete with solidifier   Carry case    O Bedside holder  VENTURI    AVANTI only  O  V pole carrier       included as standard with VENTURI    COMPACT    4    Cautions and Warnings    WARNING  Mains adaptor  if applicable  must be earthed   Before using 
16. y charged battery should operate the power unit continuously for at least 24  hours     O Charge status is shown on the display of the power unit when it is in stand by and  run mode       As the batteries are automatically charged as required when the system is  operating on mains power  the battery module should not require removing or  changing in normal use     Battery Information  continued       Use only the mains cable or optional vehicle adaptor supplied with the system       When power unit operation times are noticeably shorter than normal  it is time to  replace the battery pack       Never use any battery pack that is damaged or worn out  Use the battery pack  only for its intended purpose     O For VENTURI    AVANTI only  Because it is removable  the battery pack is short  circuit protected  Accidental short circuiting can occur when a metal object  coin   clip or pen etc   causes direct connection of the gold contacts on the side of  the battery pack   for example when you carry a spare battery in your pocket or  purse  If the battery does become short circuited the internal resettable fuse will  reconnect the circuit after about 30 seconds     O For VENTURI    AVANTI only  Leaving the battery in hot or cold places  such as in  a closed car in summer or winter conditions  will reduce the capacity and lifetime  of the battery pack  Always try to keep the battery between 15  C and 25  C  59  F  and 77  F   The capacity is particularly limited at temperatures below
17. y product without  prior notice in line with a policy of continual product development  Our standard terms    and conditions apply     EXPLANATION OF LABEL SYMBOLS AND STATEMENTS    ATTENTION Consult  accompanying documents    X        Medical Devices Directive  AN 98 42 EEC  WARNING  e  R Electrical Protection Type B  CAUTION    Class II Equipment  Double  Insulated     13    Do not dispose of with the  normal household waste    This is a statement that alerts  the user to the possibility of  serious injury or other adverse  reactions with the use or  misuse of the device    This is a statement that alerts  the user to the possibility of   a problem with the system  associated with its use or  misuse    Notes    14    Notes    15    USER MANUAL PART NUMBER 50 02 15 108 D2    TALLEY GROUP LIMITED  Premier Way  Abbey Park Industrial Estate   Romsey  Hants SO51 9DQ  England  Tel   44 0 1794 503500  e mail  sales talleygroup com          TALLEY MEDICAL USA Kal                      sr    1070 E  Wieland Road Lansing  MI 48906  USA u  TALLEY USA Tel  888 259 9994 a  USA e mail  talleyamerica talleygroup com laa C  01       Healthcare  Tades    MEDICAL  K    gt  Assocation    www talleygroup com    03 2011    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
KOBELCO SL6000J 500t  Asahi Pentax Super Program Instruction Manual  Morsix MT-7 - ILT  Garmin GMR 18 xHD Radome Declaration of Conformity  Samsung 20-50 mm F3.5-5.6 ED Standard Zoom Lens Инструкция по использованию  CB型 ブロア用逆止弁取扱説明書(H-42143) - バルブ事業部  Hotpoint-Ariston LDF 1235 Dishwasher User Guide Manual  Carrera RC Bone Racer    Copyright © All rights reserved. 