1. 8 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 UPOZORENJE SR UPOZORNENIE SK ATENCION ES VARNING SV Direction 5330620 1EN OVO SERVISNO UPUTSTVO JE DOSTUPNO SAMO NA ENGLESKOM JEZIKU AKO KLIJENTOV SERVISER ZAHTEVA NEKI DRUGI JEZIK KLIJENT JE DUZAN DA OBEZBEDI PREVODILACKE USLUGE NE POKUSAVAJTE DA OPRAVITE UREDAJ AKO NISTE PROCITALI RAZUMELI OVO SERVISNO UPUTSTVO ZANEMARIVANJE OVOG UPOZORENJA MOZE DOVESTI DO POVREBIVANJA SERVISERA RUKOVAOCA ILI PACIJENTA USLED STRUJNOG UDARA ILI MEHANICKIH DRUGIH OPASNOSTI TENTO N VOD NA OBSLUHU JE K DISPOZ CII LEN V ANGLICTINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK AKO ANGLI TINU POSKYTNUTIE PREKLADATE SK CH SLU IEB JE ZODPOVEDNOS OU Z KAZN KA NEPOK AJTE SA O OBSLUHU ZARIADENIA SK R AKO SI NEPRE TATE N VOD NA OBLUHU A NEPOROZUMIETE MU ZANEDBANIE TOHTO UPOZORNENIA M E VY STI DO ZRANENIA POSKYTOVATE A SLU IEB OBSLUHUJ CEJ OSOBY ALEBO P
2. gt Intervention Drug Dose 0018 0028 3 DS 1 Not Used Radionuclide 0054 0300 UI Selection Code Value Code Meaning 0008 0100 0008 0104 18F C 111A1 18 Fluorine 13N C 107A1 13 Nitrogen 11C C 105A1 11 Carbon 150 C B1038 15 Oxygen 68Ga C 131A3 68 Gallium 68Ge C 128A2 68 Germanium 62Cu C 127A5 62 Copper 82Rb C 159A2 82 Rubidium 22Na C 155A1 22 Sodium 75Br C 113A1 75 76Br C 113A2 T6 Bromine 64Cu C 127A2 64 Copper 1241 C 114A5 124 lodine 140 C 1018C 14 Oxygen 60Cu C 127A4 60 Copper 61Cu C 127A1 61 Copper 66Ga C 131A1 66 Gallium 38K C 135A4 38 Potassium 52Mn C 149A1 52 Manganese 94mTc C 163AA 494m4Technetium 45Ti C 166A2 45 Titanium 86Y C 162A3 86 Yttrium 90Y C 162A7 90 Yttrium Radiopharmaceuticals 0054 0304 UI Selection Code Value Code Meaning 0008 0100 0008 0104 FDG fluorodeoxyglucose C B1031 Fluorodeoxyglucose F 18 FDOPA fluoroDOPA C B1034 Fluoro L dopa F 18 F Fluorine Y X1745 F Fluorine NH3 ammonia C B103C Ammonia N 13 20 water C B1039 Oxygen water O 15 163 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN O2 oxygen C B1038 Oxygen O 15 150 CO carbon monoxide C B103A
3. eae ajaj gt GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN PET trans_scan_id PET norm cal id PET blnk cal id PET cac edge threshold PET cac skull offset PET emiss sub id PET radial filter PET radial cutoff 3d PET axial filter 3d PET axial cutoff 3d PET axial start PET axial spacing PET axial angles used 1 1 1 1 1 1 1 1 1 1 1 1 1 PET ir num iterations 1 PET ir num subsets 1 PET ir recon fov 1 PET ir corr model 1 PET ir loop filter 1 PET pre filt parm 1 PET ir loop filt parm 1 PET response filt parm 1 PET post filter 1 post_filt_parm 1 PET ir_regularize 1 PET regularize_parm 1 PET ac bp filter 1 PET ac bp filt cut off 1 PET ac bp filt order 1 PET img smooth 1 PET ac img smooth parm 1 PET scatter method 1 PET scatter num iter 1 PET scatter parm 0009 10C5 1 PET ctac conv scale 0009 10D4 1 186 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN PET loop filter parm Recon Protocol 0009 10E4 PET prompt gamma 0009 10F1 PET tracerlnjectionUID 0009 10F2 1 1 1 1 PET Q Static ScanMode 0009 10F6 3 SL 1 1 1 1 1 1 PET Q Static Frame Blood Glucose Level IsPatientDiabetic Date of Last Treatment PET QStati
4. 143 A 2 1 SC Image IOD 2 6002 110000000600 00050100000 ates testen nnn nana nnns nnns 143 A 2 2 CT Dose Report SC Image Details sss 144 A 2 2 1 Implementation Specific 145 DOC0550678 19 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement Direction 5330620 1EN REV 11 AS OCOMMONIMODBUEES LS u L petet erm pet enira odie 145 A 3 1 ef 145 3 2 General Study 146 3 3 Patient Study sisse seen 146 A 3 4 General Series 147 5 Frame of Reference Module 2 00000000000000000000 147 A 3 6 General Equipment Module u nnne 148 A 3 7 General Image Module nain tht niani niin ntn 148 JA CT IMAGE MODU LES iei HR hereto RE tate cote Ub ERR TRE a a 149 A 4 1 Image PlaneMod le pei a haba tisstamale quna 149 A 4 2 Image Pixel Module d i t Pe eoa teint 150 A 4 3 Contrast Bolus Module esanari oara e e e AAEE E E
5. eaea ee PET radionuclide_name PET half life 0009 103F 1 PET positron fraction 0009 1040 1 PET source1 holder 0009 1041 180 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN PET source1 activity 0009 1042 PET source1 meas dt 0009 1043 PET source1 radnuclide 0009 1044 PET source1 half life PET source2 holder PET source2 activity PET source2 meas dt PET source2 radnuclide 0009 1049 1 1 1 Tl source2 half life 1 PET source_location 1 PET emission_present 0009 104D 1 PET lower axial acc 0009 104E 1 PET upper axial acc 0009 104F 1 PET lower coinc limit 0009 1050 1 PET upper coinc limit 0009 1051 1 PET coinc delay offset 0009 1052 1 PET coinc output mode 0009 1053 1 PET upper energy limit 0009 1054 1 PET lower energy limit 0009 1055 1 PET normal cal id 0009 1056 1 PET normal 2d cal id 0009 1057 1 PET blank cal id 0009 1058 1 PET wc cal id 0009 1059 1 PET contrast agent 1 PET vqc x axis trans 0009 10CB 1 PET vqc x axis tilt 0009 10CC 1 PET vqc y axis trans 3 1 PET vqc y axis swivel 0009 10CE 1 PET vqc z axis trans 0009 10CF 1 1 PET vqc z axis roll 0009 10D0 181 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Con
6. 22 ectecbesdanactueteeeeievend siweevsecyencndtcvnaee ecteadsnadineteceereveetes 3 LANGUAGES atii q Seah eme iu 3 1 TABLE OF CONTENTS u uu ea a aa aaea a pa a aar a Aea TAAA eini aa Kaata Kida maaka Eain 14 17 INTROD U CTION ST ULU dessvorsesecueescnsszvensadeadhietacetskececsesuauseasdscteeeiecsssisteedanchteceaenisixeesves 23 1 1 OVERVIEW q iocos N N et a N e da a u E 23 1 2 OVERALL CONFORMANCE STATEMENT DOCUMENTATION 25 1 3 INTENDED AUDIENCE 0 uN ite aig Sistas 26 1 4 SCOPE AND FIELD OF 7 40 26 1 5 IMPORTANT REMARKS n shade sta vs toe e e dun e State etae qt case teen ds lads dates 27 1 6 REFERENCES Hn 28 1 7 DEFINITIONS eT 28 1 8 SYMBOLS AND 85 2 2 414 0 1 0000 24 4 60000000000000000 000000000 30 2 NETWORK CONFORMANCE STATEMENT 31 2 1 INTRODUCTION pittore gia peto hn asia eign Beds Pesos 31 2 2 IMPLEMENTATION MODEL 0 20ccccsneescecsseeccaadessnnscccssceccansceseaeaedaeeeecdaadecssunsecanecescaaecesseaceaees 31 2 2 1 Application Data Flow 99 99 32 2 2 2 Functional Definition
7. 174 B 1 3 Private Creator Identification GEMS 01 175 B 1 4 Private Creator Identification GEMS STDY 01 175 B 1 5 Private Creator Identification GEMS IMAQG 01 175 B 1 6 Private Creator Identification GEMS 0039 175 B 1 7 Private Creator Identification GEMS 0039 175 B 1 8 Private Creator Identification GEMS PARM 01 175 B 1 9 Private Creator Identification GEMS 01 176 B 1 10 Private Creator Identification GEMS CT CARDIAC 001 177 DOC0530678 20 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN B 1 11 Private Creator Identification SEHC_CT_ADVAPP 001 177 B 1 12 Private Creator Identification GEMS PETD O1 sees 178 B 2 PET IMAGE IOD PRIVATE DATA ELEMENTS DEFINITION esee eem emm enne 178 B 2 1 Private Creator Identification GEMS PETD 01 178 B 2 2 Discovery PET Patient 178 B
8. 112 7 4 COMMUNICATION 112 7 4 1 Supported Communication Stacks PS 3 8 112 7 4 2 112 7 4 3 Gd dts ooh EMO 112 TAS EE JEDER 113 7 4 3 2 Physical Media Support u L nennen nnne sensn teneis seit 113 7 4 4 Point to Point Stack ue Wi e e Re yi adero erre eee ego duo 113 7 5 5 2 02 113 7 5 1 Standard Extended Specialized Private SOPS 113 7 5 2 Private Transfer SyhfaX688 uu ua eene nnne nna nnne 113 1 0 GONEIGU RATIONs ihi Idi d aliii 113 7 6 1 AE Title Presentation address Mapping sss eene 114 7 6 2 Configurable 9 nnn nennen nnns 114 7 6 3 Configurable DICOM Data nnne nnns nennen nn nnns 114 7 7 SUPPORT OF EXTENDED CHARACTER 5 115 7 8 N CREATE amp N SET REQUEST nnne 115 7 9 ERROR HANDLING AND 116 7 10 USE OF SPECIFIC DICOM DATA 117 7 10 1 acis 120 7 10 2 ER S ERRARE 120
9. 61 3 3 1 2 2 2 Media Storage Application Profile for the RWA Display Directory 61 3 3 1 2 3 Real World Activity 61 3 3 1 2 3 1 Application Profiles for the RWA Local Storage RESTORE CD USB 61 3 4 AUGMENTED AND PRIVATE APPLICATION PROFILES 62 3 4 1 Augmented Application Profiles sss 62 3 4 2 Private Application Profiles 1 12 0 0 0 00000 0 6 62 3 5 EXTENSIONS SPECIALIZATIONS AND PRIVATIZATION OF SOP CLASSES AND TRANSFER SYNTAX 62 3 5 1 Extensions Specializations and Privatizations Of SOP Classes 62 3 5 1 1 SOP Specific Conformance Statement for CT SOP 62 3 5 1 2 SOP Specific Conformance Statement for MR SOP 62 3 5 1 3 SOP Specific Conformance Statement for PET SOP 63 3 5 2 Private Transfer Syntax 200000000000 63 3 0 CONFIGURATION iieri se di te iss ONE Liceul Lo RE E added 63 3 7 SUPPORT OF EXTENDED CHARACTER SETS 63 4 PRINT SCU CONFORMANGE ern meter nete RR 63 4 1 INTRODUCTION
10. rO Ir r St st St st 205 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN D 2 GE Private PET List Data Information Object Implementation This section specifies the implementation of GE private defined Information Object namely GE Private PET List Data IOD used in the GE List Data SOP Class 1 2 840 113619 4 30 The implementation is detailed to the extent that vendor applications can recognize an instance of the object This IOD is not supported for any network Send Receive operations and Archive Restore operations DVD MOD Though it is having same SOP class UID has GE PET Raw Data the PET List Data IOD is handled by using modality 0008 0060 which is having different value GEMS PET LST The exam containing PET List Series list objects will skip the same while transferring by exam Push Series is not supported for PET List Series Push Image is not supported for PET List Frame The IOD contains data that are not published in this document No implementation shall assume any meaning to such data and its meaning and definition is subject to change without notice D 2 1 Discovery PET LIST Data Patient Entity Module This section specifies the attributes of the Patient that describe and identify the Patient who is the subject of a diagnostic Study This Module contains Attributes of the patient that are
11. dereud M 76 5 2 IMPLEMENTATION MODE z cereo ee entes des Z Q huayu saa eue testet eade n exec toda 76 5 2 1 Application data flow diagram sees 76 5 2 2 Functional definitions 20 0 224 4099099922 77 5 2 3 Sequencing of real world activities eene 77 5 3 IAE SPEGIFICATIONS c rece cssc ota s roa t wee ufo tee me vede etes e a ha saa ase TT 5 3 1 DICOM Storage Commitment SCU 77 5 3 1 1 Association establishment policies 78 5 3 1 1 1 a tacere ee ORO e RIPE RE re 78 5 3 1 1 2 Numiber of associations eee ee ree rte eene TI i n ln c dE 78 5 3 1 1 3 Asynchronous NANE EET 78 5 3 1 1 4 Implementation identifying m 78 5 3 1 2 Association initiation policy nr 78 5 3 1 2 1 R al World Activity eode PER te eo ER Re RICH eet 78 5 3 1 2 1 1 Associated Real World Activity Choose Archive Save option 78 5 3 1 2 1 2 Associated Real World Activity Auto Archive 79 5 3 1 2 2 Proposed Presentation 0 00000 eene nnns 79 5 3 1 2 3 Storage Commitment Push Model SOP Class Request Processing 80 5
12. enne 68 4 3 1 2 2 2 Basic Film box SOP 69 4 3 1 2 2 3 Basic Grayscale Image Box SOP 70 4 3 1 2 2 4 Color Image Box SOP Class sse nennen nennen ennt nnn 71 4 3 1 2 2 5 Printer SOP CIass n horrea e e eee SS e dedu 73 4 3 1 2 2 6 3 5 rem den Seren hi ei ei ree ei e ER 74 4 3 1 3 Association Acceptance Policy eene 74 4 4 COMMUNICATION PROFILES u ret drehte eee dra n 74 4 4 1 Supported Communication Stacks parts 8 9 74 WA WT TGPTIPSlIaSgk au EET 74 4 4 1 1 1 A oth nationi tutam 74 4 4 1 1 2 Physical Media Suppott niter deter eee dt e ihe venus edere 74 4 5 STANDARD EXTENDED SPECIALIZED PRIVATE 5 74 46 CONFIGURATION 2 32 25 50 edes hoes AMEE ee 74 4 6 1 AE Title Presentation Address Mapping sss 74 4 6 2 Configurable kaypa 75 4 7 SUPPORT OF EXTENDED CHARACTER SET 75 5 DICOM STORAGE COMMITMENT SCU 76 5 1 INTRODUGTION z
13. 143 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN SC Image Not sent consists entirely of M type 3 element A 2 2 CT Dose Report SC Image Details The product is configured by default to generate a CT X Ray Radiation Dose Report SC Image summarizing the study dose The pixel data contains a textual viewable report of the dose information The next section identifies the additional standard DICOM attributes added to capture the same information as reported in the CT X Ray Radiation Dose Record described in Section 9 The Discovery 710 610 and Optima 560 products will write these additional attributes 144 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN A 2 2 1 gill ac A details Total Number of 0040 Total number fees exposures made during Exposures this Performed Procedure Step Exposure dose 0040 030E Exposure Dose Sequence will contain sequence Total Number of Exposures 0040 0301 items 0018 0060 Peak kilo voltage output gt X Ray Tube Current 0018 8151 3 X Ray tube current in microA NOTE the X Ray Radiation Dose SR records this in mA 0018 1150 3 Time of x ray exposure msec gt Acquisition Type 0018 9302 gt CTDIlvol 0018 9345 Computed Tomography Dose Index in mGy according to IEC 60601 2 44 gt CTDI Phantom 0018 9346 The typ
14. 5 0 2 34 2 2 3 Sequencing of Real World Aclivities 9 35 2 3 AE SPECIFICATIONS desni edana MA 35 2 3 1 DICOM Server AE 5 22 35 2 3 1 1 X Association Establishment Policy esee enne em ee enemies 36 2 3 1 1 1 GONG ral aaa 36 2 3 1 1 2 Number of Associations 36 2 3 1 1 3 Asytichronous N t fe 9 x oid re hana n i ete qe tede 36 2 3 1 1 4 Implementation Identifying Information 20029 9 36 2 3 1 2 Association Initiation by Real World Activity 37 2 3 1 2 1 Push Image s to Remote 37 2 3 1 2 1 1 Associated Real World 37 2 3 1 2 1 2 Proposed Presentation Contexts 37 2 3 1 2 2 Query Remote AE es dee den P deci d ane eade dec vad loe Te diste 39 2 3 1 2 2 1 Associated Real World 39 2 3 1 2 2 2 Proposed Presentation Contexts 40 2 3 1 2 3 Real World Activity Manual Retrieve 42 2 3 1 2 3 1 Associated Real World 42 2 3 1 2 3 2 Proposed Presentation 42 2 3 1 2 4 Real Worl
15. A 7 21 Discovery PET Image Module Refer to Section B 2 7 for details A 7 22 Discovery PET Frame Module Refer to Section B 2 5 for details 172 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN A 7 23 SOP Common This section defines the Attributes which are required for proper functioning and identification of the associated SOP Instances They do not specify any semantics about the Real World Object represented by the IOD Attribute Name Tag Type VR VM Notes SOP Class UID 0008 0016 1 UI 1 SOP Instance UID 0008 0018 Specific Character Set 0008 0005 Always sent as ISO_IR 100 Instance Creation Date 0008 0012 Current Date Instance Creation Time 0008 0013 Current Time Instance Creator UID 0008 0014 3 Ul 1 173 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN APPENDIX B Private Data Elements Enclosed is a listing of private data elements used in this implementation for CT Image IOD definition B 1 CT Image IOD Private Data Elements Definition B 1 1 Private Creator Identification GEMS IDEN 01 Attribute Name Tag H Product id 0009 1004 Image actual date 0009 1027 Equipment UID 0009 10E3 Note Forall images created by Discovery system 0009 xx01 element will have the value CT DISCOVERY B 1 2 Private Creator Identific
16. 63 4 2 IMPLEMENTATION MODEL PRINT 5 63 4 2 1 Application Data Flow 63 4 2 2 Functional Definition 1 1 4 01000000 64 4 2 3 Sequencing of Real World Activities esses 64 4 9 JAE SPEGIFICATIONS colt terree deett t x reet etu Brad cid ae usya 66 4 3 1 DICOM Print SCU AE Specification 66 4 3 1 1 Association Establishment 66 4 3 1 1 1 eee 66 DOC0550678 15 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 4 3 1 1 2 Number of 66 4 3 1 1 3 Asytnichronous Nat re u 1 o xr eee et ete rte deir dele dade eder dc cee 66 4 3 1 1 4 Implementation identifying nens 66 4 3 1 2 Association Establishment emen emnes 67 4 3 1 2 1 Real World Activity iit pde e n e ed Ra I ed Hu eb 67 4 3 1 2 1 1 Associated Real World Activity Print a 67 4 3 1 2 1 2 Proposed Presentation Contexts 2 1 2 000000000010 00022 67 4 3 1 2 2 SOP Specific Conformance Statement 68 4 3 1 2 2 1 Basic Film Session SOP
17. 5 4 Communication Profiles 5 4 1 Supported Communication Stacks parts 8 9 DICOM Upper Layer Part 8 is supported using TCP IP 5 4 2 TCP IP Stack The TCP IP stack is inherited from a UNIX Operating System 5 4 2 1 Physical Media Support Ethernet v2 0 IEEE 802 3 5 4 3 Point to Point Stack A 50 pin ACR NEMA connection is not applicable to this product 5 5 EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS None 5 6 CONFIGURATION Ethernet v2 0 IEEE 802 3 5 6 1 AE Title Presentation Address Mapping The system allows the user to Remove or Update the mapping of remote AE Titles to IP Addresses and Ports These options can be selected from the Remote Host Selection menu displayed by choosing Select Remote Host from the Network pull down menu from the local database manager A remote DICOM host can be designated as a remote archive device by setting the Remote archive node radio button on the remote host configuration panel Once a DICOM host is designated as a remote archive device the host will be available in the Archive device selection menu and can be chosen for any subsequent archive operation 5 6 2 Configurable Parameters The following fields are configurable for the DICOM Server AE Local AE Title the machine hostname Local IP Address Local IP Netmask Max PDU length DOC0550678 83 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11
18. Direction 5330620 1EN For Discovery 710 610 and Optima 560 Products SW Version pet mfk xx Time outs which are set for all hosts are configurable in dcs cfg Note denotes any SOP class time in sec Association time out bi assoc tio nactivity time out bi store tio Session time out bi session tio For Discovery 710 610 Products SW Version pet coreload xx timeouts are set in network cfg xml and mergecom pro files Note All configurations should be performed by a GE Field Service Engineer 5 7 Support of Extended Character Sets This does not support any Extended Character set DOC0550678 84 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 Modality Work List Information Model Definition 6 1 Introduction This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the relevant Networking features on this GEHC product Note that the format of this section strictly follows the format defined in DICOM Standard PS 3 2 Conformance Please refer to that part of the standard while reading this section The Modality Worklist option for Discovery allows a user to query for and display DICOM modality worklist information from a remote hospital or radiology department information system computer For example a user may wish to query for all procedures scheduled to be performed on the scanner In th
19. set to and Days After Today YYYYMMDD YYYYMMDD where this date range represents the specified number of days before today and or after today Note that number of days both before and after can be specified in the same query and that each always includes today All Days Scheduled Procedure Step Start Date 0040 0002 zero length universal matching 6 3 1 2 1 2 Proposed Presentation Context Table The following table shows the proposed presentation contexts for the Worklist Server AE after real world activity Worklist Query has been initiated Presentation Context Table Proposed Abstract Syntax Transfer Syntax Extended Nametit UDList Negotiation Modality Worklist 1 2 840 10008 5 1 4 31 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None Information Model FIND DOC0550678 88 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 3 1 2 1 2 1 SOP Specific DICOM Conformance Statement for the Worklist SOP Class If the remote AE does not support the proposed Presentation Context an appropriate error is logged and the operator is notified This implementation can receive multiple C FIND results over a single association Only one association is opened at a time Each C FIND response received from the remote AE is parsed to verify the length type of the items in the response Upon detecting any error in the response data the
20. 0008 0104 F 10450 recumbent Patient Orientation Modifier 165 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Code Value 0008 0100 Code Meaning 0008 0104 F 10310 prone F 10317 right lateral decubitus F 10319 left lateral decubitus F 10340 supine Patient Gantry Relationship Code Value Code Meaning 0008 0100 0008 0104 F 10470 headfirst F 10480 feet first A 7 11 Discovery PET Image Set Module Refer to Section B 2 6 for details A 7 12 Discovery PET Scan Module Refer to Section B 2 4 for details A 7 13 Frame of Reference Module This section specifies the Attributes necessary to uniquely identify a frame of reference which insures the spatial relationship of Images within a Series It also allows Images across multiple Series to share the same Frame Of Reference This Frame Of Reference or coordinate system shall be constant for all Images related to a specific Frame Of Reference Attribute Name Tag Type VR VM Notes Frame of Reference UID Position Reference Indicator 002000052 1 Juil 1 0020 1040 2 Scan Landmark A 7 14 General Equipment Module This section specifies the Attributes which identify and describe the piece of equipment which produced a Series of Images Attribute Name Tag Type V
21. 158 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Attribute Name Tag Type VR VM Notes Procedure Step Sequence gt Referenced SOP Class UID 0008 1150 gt Referenced SOP Instance 0008 1155 UID Body Part Examined Not used Patient Position Smallest Pixel Value in Series Not used Largest Pixel Value in Series Not used Request Attributes Sequence gt Accession Number gt Reference Study Sequence gt gt Referenced SOP Class 0008 1150 3 Ul 1 UID Referenced SOP Instance 0008 1155 UID Study Instance UID 0020 000D gt Requested Procedure 0032 1060 3 LO 1 Description gt Requested Procedure Code 0032 1064 SQ 1 Sequence gt gt Code Value 0008 0100 gt gt Coding Scheme Designator 0008 0102 Code Meaning 0008 0104 gt Scheduled Procedure 0040 0007 3 LO 1 Description gt Scheduled Procedure Step 0040 0009 ID 3 SH 1 gt Requested Procedure ID 0040 1001 A 7 7 PET Series Module Attribute Name Tag Type VR VM Notes Series Date 0008 0021 1 DA 1 Always shows Acquisition Date Series Time 0008 0031 1 TM 1 Always shows Acquisition 159 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Time Counts S
22. Code Meaning 0008 0104 EBEN Not Used gt Radiopharmaceutical 0018 1071 Volume gt Radiopharmaceutical Start 0018 1072 Time gt Radiopharmaceutical Stop 0018 1073 3 TM 1 Not Used Time Value depends on tracer activity post injection activity half life measure date time admin date time post injection date time Radionuclide Total Dose 0018 1074 Radionuclide Half Life RE gt Radionuclide Positron US Fraction gt Radiopharmaceutical 0018 1077 Specific Activity gt Radiopharmaceutical 0018 0031 gt Radiopharmaceutical Code 0054 0304 Sequence Value is based on gt gt Code Value 0008 0100 1C SH Tracer name Coding Scheme Designator 0008 0102 SRT Code Meaning 0008 0104 EON Tracer name Intervention Drug Information 0018 0026 SQ Not Used Sequence gt Intervention Drug Name 0018 0034 a EIE Not Used gt Intervention Drug Code 0018 0029 Not Used Sequence Code Value 0008 0100 1C SH 1 gt gt Coding Scheme Designator 0008 0102 1C SH 1 Not Used Not Used Tracer name Single item sequence 162 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Code Meaning 0008 0104 Not Used gt Intervention Drug Start Time 0018 0035 Not Used gt Intervention Drug Stop Time 0018 0027 Not Used
23. GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 Grayscale Presentation State 8 1 8 2 Introduction This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the Grayscale Presentation State GSPS referred to as the Exam Split option feature on this GEHC product Note that the format of this section strictly follows the format defined in DICOM Standard PS 3 2 Conformance Please refer to that part of the standard while reading this section The Exam Split option for Discovery allows a GSPS object to be sent to any remote host that supports the display of images using GSPS The Exam Split feature provides the DICOM Modality GSPS service as a service class user SCU This feature works in conjunction with DICOM Modality Work list feature and Modality Performed Procedure Step if installed However the conformance of this feature is dependent on Modality work list feature For information on conformance of Modality Work List feature to DICOM standard please refer to the appropriate section or document Implementation Model Exam Split option have the two options i e Virtual Exam Split and Hard Exam Split e Virtual Exam Split will create Gray Scale Presentation State and Performed Presentation State n Hard Exam Split application will send the user selected images from grouped Images that acquired during acquisitio
24. LightSpeed All prior 1 2 840 113619 6 55 Discovery BrighSpeed Discovery CT 750 HD 09MW29 14 1 2 840 113619 6 267 BrightSpeed Elite 09HW30 4 1 2 840 113619 6 270 Discovery CT 590 RT 09BW35 11 1 2 840 113619 6 278 Optima CT 580 RT Optima CT 580 W BrightSpeed Elite 09BW35 11 1 2 840 113619 6 278 BrightSpeed Elite 09BW35 11 1 2 840 113619 6 278 Select LightSpeed VCT 10 06 3 1 2 840 113619 6 277 Optima 660 10HW202 xx 1 2 840 113619 6 271 Discovery 610 pet coreload xx 1 2 840 113619 2 290 pet mfk xx Discovery 710 pet coreload xx 1 2 840 113619 2 290 pet mfk xx Optima 560 pet mfk xx 1 2 840 113619 2 290 213 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN This page intentionally left blank 214 of 215 GE Healthcare a General Electric Company 3000 N Grandview Boulevard Waukesha Wisconsin 53188 USA www gehealthcare com Imagination at work
25. Remote GSPS AE IP Address IP Address used to contact the remote GSPS AE Remote GSPS IP Port IP Port used to contact the remote GSPS AE Remote GSPS AE Title Application Entity Title used to contact the remote GSPS AE Remote HES AE IP Address IP Address used to contact the remote HES AE Remote HES IP Port IP Port used to contact the remote HES AE Remote HES AE Title Entity Title used to contact the remote 8 6 2 Configurable Parameters The Following are configurable for the GSPS Remote Host IP address Remote Host port Remote AE Title VES_ORIENTATION environment variable for enable disable Flip Rotate functionality in VES Changesplit_mode lt HES VES gt 8 7 SUPPORT OF EXTENDED CHARACTER SETS GSPS feature supports ISO IR 100 character set 8 8 ERROR HANDLING AND RECOVERY Exam Application has an extended error codes and the standard error codes are also handled If the sending GSPS PPS objects request has failed or response is not received before the association timeout the operation is deemed Failed and the system updates the state and indicates the same on the user interface If the operation is Failed detailed message is logged into system log file and system provides an alternative mechanism to retry the failed operation through the user interface to ensure that transient failures do not affect the feature performance 128 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance
26. SCP O O SCP Direction 5330620 1EN Negotiati on None None None None None None None None GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Explicit VR Little 1 2 840 10008 1 Endian 2 1 Explicit VR Big 1 2 840 10008 1 Endian 2 2 2 3 1 3 1 3 Transfer Syntax Selection Policy Sequence of the transfer syntax depends on the dcs cfg file present on the host machine By default sequence is 1 2 840 10008 1 2 1 ELE 1 2 840 10008 1 2 2 EBE 1 2 840 10008 1 2 ILE 1 2 840 113619 5 2 GEIBE Based the above sequence a transfer syntax table is created The transfer syntax table is iterated until it gets a match for the proposed presentation context 2 3 1 3 1 3 1 SOP Specific Conformance to C_STORE SCP The DICOM Server AE conforms to the SOP s of the Storage Service Class at level 1 base Private elements will be discarded from the image when receiving images containing non GE private data elements All of the standard type elements 1 1 2 2 3 will be retained Each C STORE SCP operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 900 seconds and 60 minutes respectively Association Timer duration for SCP to respond to an association request Session Timer duration from association to first command sent by SCU Inactivity Timer
27. The Discovery system allows the user to add Remove or Update the mapping of remote AE Titles to IP Addresses and Ports These options can be selected from the Remote Host Selection menu displayed by choosing Select Remote Host from the Network pull down menu from the local database manager 2 6 2 Configurable Parameters The following fields are configurable for the DICOM Server AE Local AE Title the machine hostname Local IP Address Local IP Netmask Max PDU length For the Discovery 710 610 and Optima 560 product SW Version pet mfk xx time outs which are set for all hosts are configurable in dcs cfg Association time out bi assoc tio Session timeout bi session tio Create timeout bi ncreate tio DOC0550678 52 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN For the Discovery 710 610 and Optima 560 products SW Version pet_coreload xx timeouts are set in network cfg xml and mergecom pro files Note All configurations should be performed by a GE Field Service Engineer Note The local port on which the Discovery system receives DICOM incoming TCP connections is port 4006 2 7 Support of Extended Character Sets In addition to the DICOM default character set Discovery supports the ISO IR 100 Latin alphabet 1 supplementary set for the purpose of interchange As a Storage SCP the product will accept SOP Insta
28. The Following Attribute values are supported Attribute DICOM Tag Valid Range Default Value Image Display Format 2010 0010 STANDARD C R Set in User Interface Printer Dependent Reference Film 2010 0500 Session Sequence Referenced Image 2010 0510 NA NA Box Sequence Film Orientation 2010 0040 PORTRAIT Set in Configuration File Film Size ID 2010 0050 Sent zero length Sent zero length Magnification type 2010 0060 BILINEAR Set in Configuration File CUBIC REPLICATE NONE Max Density 2010 0130 0 4095 Set in Configuration File i 0405 Configuration 2010 0150 Printer Dependent Set in Configuration File Information Smoothing type 2010 0080 Printer Dependent Set in Configuration File Border density 2010 0100 BLACK Set in Configuration File WHITE image density 2010 0110 BLACK Set in Configuration File WHITE Min density 2010 0120 0 4094 Set in Configuration File 2010 0140 Note Attributes sent zero length use the camera default values Note denotes that the attribute is optional for the SCU However we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value DOC0550678 69 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN If
29. The Point to Point Stack is not supported by this implementation 7 5 EXTENSIONS SPECIALIZATION PRIVATIZATION 7 5 4 Standard Extended Specialized Private SOPs PPS for Discovery does not implement any private SOP classes 7 5 2 Private Transfer Syntaxes PPS for Discovery does not implement any private Transfer Syntaxes 7 6 CONFIGURATION The PPS feature is configured by GEHC Field Service Engineers The DICOM configuration items below are configurable or re configurable by the Field Service Engineer and are not accessible by users through the Discovery s user interface 113 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 7 6 1 AE Title Presentation address Mapping PPS allows for the configuration of the following parameters that pertain to the remote AE Remote MPPS AE IP Address IP Address used to contact the remote MPPS AE Remote AE HIS RIS IP Port IP Port used to contact the remote MPPS AE Remote MPPS AE Title Application Entity Title used to contact the remote MPPS AE These parameters define where the MPPS requests will be directed Configuration of these parameters is performed by GEHC Field Service engineers using the installation facilities 7 6 2 Configurable Parameters The following are configurable for the PPS Server AE Local PPS Server AE AE Title The default is host name of the scanner appended with string PPS If th
30. 7 10 3 Series Level o up e ERE RR eden d de te pe 120 8 GRAYSCALE PRESENTATION 86 2 4 1112 22 21 122 8 1 INTRODUGTION rm 122 82 IMPLEMENTATION MODEL uiae d densi en hao ite 122 DOC0550678 18 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 8 2 1 Application Data Flow 123 8 2 2 Functional Definition of 5 123 8 2 3 Sequencing of Real World Activities esses 123 8 2 3 1 GSPS Acquisition System with MWL data a 123 8 2 4 GSPS Server AE Specification sss 124 8 2 4 1 Association Establishment Policies I nn 124 8 2 4 1 1 Gen ral te POI aut ai o d leis PUE geet qha Y 124 8 2 4 1 2 Number of Associations cire ere eee 124 8 2 4 1 3 Asynchronous Nature iicet e e eet e 124 8 2 4 1 4 Implementation Identifying information sss 124 8 2 4 2 Association Initiation Policy 124 8 2 4 2 1 Real World Activifyz GSPS inni eere iten e eda 125 8 2 4 2 1 1 Associated Real World 125 8 2 4 2 1
31. ANSE TT SEA AWS AST TY EO AE HERRE ee i HL K E o TENTO PROVOZN N VOD EXISTUJE POUZE V ANGLICK M JAZYCE V P PAD E EXTERN SLU BA Z KAZN K M POT EBUJE N VOD V JIN M JAZYCE JE ZAJI T N P EKLADU DO ODPOV DAJ C HO JAZYKA KOLEM Z KAZN KA NESNA TE SE O DR BU TOHOTO ZA ZEN ANI BYSTE SI P E ETLI TENTO PROVOZN N VOD A POCHOPILI JEHO OBSAH V P PAD NEDODR OV N T TO V STRAHY M E DOJ T K PORAN N PRACOVN KA PRODEJN HO SERVISU OBSLU N HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM MECHANICK CH I JIN CH RIZIK DENNE SERVICEMANUAL FINDES KUN PA ENGELSK HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK ER DET KUNDENS ANSVAR AT S RGE FOR OVERS TTELSE FORS G IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL HAR V RET KONSULTERET OG ER FORST ET MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDF RE SKADE P GRUND AF ELEKTRISK MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN OPERAT REN ELLER PATIENTEN 3 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 WAARSCHUWING NL WARNING EN HOIATUS ET VAROITUS Fl Direction 5330620 1EN DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS
32. CD DVD USB Storage Interchange AE Specification The DICOM CD DVD USB Storage Media AE provides standard conformance to DICOM Media Interchange Option of the Media Storage Service Class The supported Application Profiles and roles are listed below Table 3 3 1 1 Application Profile Activities and Roles for CD DVD R CD RW DVD RW USB Update Supported Application Real World Activity Profile 3 3 1 1 File Meta Information for Implementation Class and Version The File Meta Information for this implementation is File Meta Information Version 1 Implementation Version Name Software Revision See Appendix E The table in Appendix E identifies the Implementation UID for this product version 3 3 1 2 Real World Activities for the DICOM Media Interchange AE 3 3 1 2 1 Real World Activity RWA Create Generate and Write CD DVD USB The DICOM Media Interchange AE acts as an FSC using the interchange option when requested to copy SOP Instances from the local database to a CD R DVD R CD RW DVD RW USB The user selects the entries in the local database that he she wants the DICOM Media Interchange AE to copy onto Interchange Media The graphic interface allows the user to select the entries patients studies series or images in the local database to be copied onto to the selected Interchange Media DOC0550678 58 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance
33. DCM Comment Not sent Note Target region is filled in per the following table Protocol Category Head SRT 01100 Head Value stored in Target Region Orbit SRT T D0801 Orbital region Neck 7 5 T D1600 Neck 141 of 215 Direction 5330620 1EN GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Upper Extremity SRT T 02220 Shoulder SRT T D3000 Chest SRT 04000 Abdomen SRT T 11503 Lumbar spine SRT T D6000 Pelvis Lower Extremity SRT T D0300 Extremity 9 5 Configuration The product is configured by default to not generate the CT X Ray Radiation Dose SR Record Creation of the CT Dose Record can be enabled via reconfig The product implements the CT Radiation Dose Report template TID 10011 per DICOM The DICOM standard says that the SOP class shall be the X Ray Radiation Dose SR SOP Class 1 2 840 10008 5 1 4 1 1 88 67 However since many products may not support this SOP class the product supports the option via reconfig to create the CT Radiation Dose Report as an Enhanced SR SOP Class 1 2 840 10008 5 1 4 1 1 88 22 reconfig Dose Report set to Full will produce the X Ray Radiation Dose SR SOP class and Yes will produce the Enhanced SR SR setting can be configured in Dose Report in reconfig menu as below Full will produce the X Ray Radiation Dose SR SOP class will produce
34. Nosent sent 135 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN gt Verifying Organization 0040 A027 gt Verification DateTime 0040 A030 Author Observer Sequence 0040 A078 Participant Sequence 0040 A07A gt Participation Type 0040 A080 gt Participation DateTime 0040 A082 gt Include Identified Person or Device Macro Table C 17 3b Custodial Organization 0040 A07C 3 Not sent Sequence gt Institution Name 0008 0080 gt Institution Code Sequence 0008 0082 gt gt Include Code Sequence Macro Table 8 8 1 ee Predecessor Documents 0040 A360 Not sent per condition Sequence Include Hierarchical SOP Instance Reference Macro Table C 17 3 Identical Documents 0040 A525 1C Not sent per condition Sequence gt Include Hierarchical SOP Instance Reference Macro Table C 17 3 Referenced Request 0040 A370 1C Not sent per condition Sequence Study Instance UID 0020 000D Not Sent gt Referenced Study 0008 1110 2 Not Sent Sequence gt gt Include SOP Instance Reference Macro Table 10 11 gt Placer Order 0040 2016 2 Not Sent Number Imaging Service Request gt Filler Order 0040 2017 2 Not Sent Number Imaging Service Request gt Requested Procedure ID 0040 1001 Not Sent gt Requested Procedure 0032 1060 2 Not Sent Description gt Requested Procedure 0032 1064 2 Not Sent Code Sequen
35. REV 7 0020 0052 in conjunction with the Table Height to Direction 5330620 1EN determine if two images are spatially related A 3 4 General Series Module Attribute Name Type Notes 0020 001 1 Generated sequentially always sent Laterality 0020 0060 Always sent zero length sent sent 0018 1030 0008 103E Patient Position 0018 5100 Series Instance UID 0020 000E 9100 Performed Procedure 0040 0244 Step Start Date 0040 0245 0253 2 2 2 FFDL Feet First Decubitus Left FFP Feet First Prone FFS Feet First Supine Always sent Step Start Time Step ID El d M Step Description A 3 5 Frame of Reference Module Attribute Name Tag Type Notes O Frame of Reference UID 0020 0052 See Caution below Position Reference Indicator 0020 1040 2 Value as entered at the user interface p ue 7 220 1 _ 3 3 3 Sent As selected by operator when patient is positioned Defined terms are Head First Prone HFS Head First Supine HFDR Head First Decubitus Right HFDL Head First Decubitus Left FFDR Feet First Decubitus Right o C 3 3 1 3 3 3 C 1 3 3 3 3 147 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN A 3 6 General Equipment Module Manufacturers Model 0008 1090 3 Name Device Serial 0018 1000 Sent if present in ima
36. S _ PETWawedonUD 3 U 1 PET corector eaa Goma 3 U 1 _ PETcompatbe verson 0017102 3 5 1 E sr s E 24 e PET hosp Meme PETacwd gt O O m ea na 1 rU PET norm 2d cal id 0017 100B PET hosp identifier 0017 100C SH PET archived 0017 100D D 2 13 Discovery PET Well Counter Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET scan type 0009 1018 3 SL PET trame unisiedscan 3 sc 1 promptgamma 3 _ PETWawedonUD 00940 3 U 1 PETwscd ooon 3 U 1 gt compatible_version 0019 1002 3 SH 1 PET software_version 0019 1003 C T 211 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN eaae 3 oT 1 3 1 _ PET cal descrpion 3 to 1 _ hardware 3 to 1 2 rO Ir r St st St st 212 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN APPENDIX E Implementation UID for Product Versions Product Model Software Implementation UID Revision
37. This Module contains Attributes of the patient that are needed for diagnostic interpretation of the Image and are common for all studies performed on the patient 156 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Attribute Name Tag Type VR VM Notes Patient s Name 0010 0010 2 32 1 Patient ID 0010 0020 2 ILO 16 1 Patient s Birth Date 0010 0030 2 IDA 26 1 Patient s Sex 0010 0040 2 CS 16 1 Referenced Patient Sequence 0008 1120 3 SQ 1 Not used gt Referenced SOP Class UID 0008 1150 1C JUI 1 Not used gt Referenced SOP Instance 0008 1155 1C JUI 1 Not used UID Patient s Birth Time 0010 0032 3 ITM 1 Not used Other Patient IDs 0010 1000 3 LO 1 n Not used Other Patient Names 0010 1001 3 PN 1 n Not used Ethnic Group 0010 2160 3 SH 1 Not used Patient Comments 0010 4000 3 LT 1 Not used A 7 3 General Study Module Attribute Name Tag Type VR VM Notes Study Instance UID 0020 000D 1 UI 1 Study Date 0008 0020 2 DA 1 Same as Exam Date Study Time 0008 0030 2 TM 1 Same as Exam Time Referring Physician s Name 0008 0090 2 PN 1 Study ID 0020 0010 2 SH 1 Accession Number 0008 0050 2 SH 1 Study Description 0008 1030 3 LO 1 Physician s of Record 0008 1048 3 PN 1 n Not used Name of Physician s Reading 0008 1060 3 PN 1 n Diagnostician Study Referenced Study Sequence 0008 1110 3 SQ 1 Value copied from SPS gt Re
38. Up to 5 Scheduled Action Item Code Sequence Items are mapped into the image if they system is configured to support GSPS User can enter the value for Requested Procedure ld prior to query If user has entered the value then that value will be sent as part of the query The value returned in the response shall be mapped to the image Truncated to 22 characters GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Attribute Name Expected Expected Mapped Matching Returned into the Key Type Key Type Image Study Instance UID 0020 000D 1 Yes User Modality will be able to Based configure the following If the on User user has set the option to Use option MWL Study UID then study Instance UID will be copied into the final DICOM image header If this option is not set then a new Study instance UID is generated locally on the scanner The default setting is to generate a new study instance uid if a worklist entry is re used however it can be configured to reuse the same study instance uid Referenced Study Sequence 0008 1110 2 Yes Only 1 Referenced Study Sequence is mapped into the image Requested Procedure Priory Transport onde oon Arrangements 6 8 11 Imaging Service Request Entity 6 8 11 1 Imaging Service Request Module Table 6 8 11 1 1 Imaging Service Request Module Attributes Attribute Name Expected Expected Mapped Matc
39. VERKRIJGBAAR ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST DAN IS DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD ZOU HET ONDERHOUDSPERSONEEL DE OPERATOR OF EEN PATIENT GEWOND KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK MECHANISCHE OF ANDERE GEVAREN THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY IF ACUSTOMER S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH IT IS THE CUSTOMER S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER OPERATOR OR PATIENT FROM ELECTRIC SHOCK OR FROM MECHANICAL OR OTHER HAZARDS KAESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES KUI KLIENDITEENINDUSE OSUTAJA NOUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES VASTUTAB KLIENT TOLKETEENUSE OSUTAMISE EEST ARGE URITAGE SEADMEID TEENINDADA ENNE EELNEVALT KAESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST KAESOLEVA HOIATUSE EIRAMINE VOIB POHJUSTADA TEENUSEOSUTAJA OPERAATORI VOI PATSIENDI VIGASTAMIST ELEKTRILOOGI MEHAANILISE VOI MUU OHU TAGAJARJEL TAMA HUOLTO OHJE ON SAATAVILLA VAIN ENGLANNIKSI JOS ASIAKKAAN HUOLTOHENKILOSTO VAATII MUUTA KUIN ENGLANNINKIELISTA MAT
40. duration between two commands after the association Image Reception If the DICOM Server AE returns one of the following status codes then the C STORE operation was unsuccessful and no image will be installed 0110 Processing Failure Indicates that an internal system call has failed while processing an image A711 Out of Resources Indicates that probably there was not enough disk space to store the image The user should attempt recovery by removing some images from the Discovery system A712 Out of Resources Indicates that there was not enough resource such as memory to store the image A800 SOP Class not supported DOC0550678 46 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN In the event of a successful C STORE operation the image has successfully been written to disk The image will then be accessed in the same manner as any other image by the applications on the Discovery system Images may be deleted when instructed to do so by the user Thus the duration of the storage of the image is determined by the users of the Discovery system Attention This Discovery Optima system does not support previous non Discovery based PET CT Raw Data Receiving Pulling Raw Data series acquired from a non Discovery based product onto the Discovery Optima will cause the network process to fail Furthermore Discovery 710 610 and Optima 560 Raw Data format has
41. 0050 Special Needs only displayed on Not available the More Info screen Requested Contrast Agent Requested Contrast Agent only Not available 0032 1070 displayed on the More Info screen Current Patient Location Current Patient Location only Not available 0038 0300 displayed on the More Info screen Additional Patient History Additional Patient History only Patient History 0010 21B0 displayed on the More Info Screen 105 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Note that the display of a specific data item on the More Info screen is contingent on the item being enabled for display Depending on the preferences of each specific site data can either be displayed or not A GE field service engineer can assist in setting these site preferences 6 12 Setting User Preferences 6 12 1 Setting Use MWL Study UID Option Setting this option to Yes copies the Study Instance UID from work list into the final DICOM Image header If this option is set to No then a new study instance uid will be generated locally 1 Click on Patient Schedule 2 Click on Preferences button 3 Set the option Use Study Instance UID to either Yes or No 6 12 2 Setting Custom Query Option This option allows the user to enter values for Accession Number and or Requested Procedure Id which are used for Custom Quer
42. 2 3 Discovery PET Exam Module sss 179 B 2 4 Discovery PET Scan Module u y aa nnne nnns 179 B 2 5 Discovery PET Frame 9 182 2 6 Discovery PET ImageSet 1 1 2 000000000000000000 185 2 7 Discovery PET Image 2 0 nnne nentes 189 APPENDIX DICOMDIR DIRECTORY enm enn 191 CA BASIC DIRECTORY IOD E nns 191 C2 JFILESET IDENTIFICATION MODULE Z u cha nidi 191 DIRECTORY INFORMATION 191 MEDIA DIRECTORY INFORMATION 192 C 5 DIRECTORY RECORD SELECTION 8 1 4 sa sasa 193 5 1 193 5 2 Study Keys scit od ulla en Pase ea e ERO RATE EXT Pen ou RE OR 193 C 5 3 SOMOS Cu mm 194 C 5 4 Image Rd ERE ERE apaspan pha 194 C 5 5 PRESENTATION Keys ener tro up eni EXER A Hox P HRS 196 C 5 6 STRUCTURED REPORT Keys eene nennen enarrant nna ness 197 C 5 7 PRIVATE GEMS PET RAW 22 1 02000001000000000000000 enne nenne 197 APPENDIX D GE DISCOVERY PRIVATE PET
43. 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for association initiated by the DICOM Storage Commitment SCU AE is Maximum Length PDU 10Kbyts 5 3 1 1 2 Number of associations A single association is used to send the commitment request This commitment request association is closed after the request is successfully acknowledged It accepts an association for receiving the response from commitment provider at any time 5 3 1 1 3 Asynchronous nature Asynchronous window negotiation is not supported 5 3 1 1 4 Implementation identifying information The table in Appendix E identifies the Implementation UID for this product version 5 3 1 2 Association initiation policy An association is initiated by the Storage SCU to the provider to send the images to be committed A separate association is established to send the commitment request 5 3 1 2 1 Real World Activity 5 3 1 2 1 1 Associated Real World Activity Choose Archive Save option Behavior for Discovery 710 610 and Optima Products SW Version pet mfk xx The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname Select the Update option if the hostname is already present Set the Archive Node option to Yes If the remote host is not present add the remote host in the local
44. 3 1 3 Association Acceptance 81 5 3 1 3 1 Real World Activity Storage Commitment Response and Set Archive State 81 5 3 1 3 1 1 Accepted Presentation Context 81 5 3 1 3 2 Storage Commitment Push Model SOP Class Response Processing 81 5 3 1 3 2 1 Commit response with SUCCESS status 82 5 3 1 3 2 2 Commit response with FAILURE 82 5 4 COMMUNICATION PROFILES LA ec 83 5 4 1 Supported Communication Stacks parts 8 9 83 5 4 2 Stack 5 is eects 83 5 4 2 1 Physical Media Support n tn nnne e 83 5 4 3 Point tosPoint Stack s aci AR ERR ae 83 5 5 J EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS sse 83 5 62 CONEIGURATION rite s GEM 83 5 6 1 AE Title Presentation Address Mapping sess 83 5 6 2 Configurable Parameters 83 DOC0550678 16 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 5 7 SUPPORT OF EXTENDED CHARACTER SETS ssssssesesececececcececcecceeecesesesaeaeaeaessaeaeeseseseceseseseeeees 84 6 MODALITY WORK LIST INFORMATION MODEL 404 4 442
45. 9 90 6 4 2 OSLSIack ii e e x 90 6 4 3 LCP P SIQCk sit tions tu een 91 6434 91 6 4 3 2 Physical Media Support uuu uuu upa gaa anam aaa asas aaa Susa 91 6 4 4 Point to Point SIaCk aai Bled edhe 91 6 5 EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS T I I 91 6 5 1 Standard Extended Specialized Private 5 91 6 5 2 Private Transfer 91 6 6 CONFIGURNRTION itr rb E nae NER ia eee 91 6 6 1 AE Title Presentation Address Mapping 22 2 2 2 10 001000000004066601000000 91 6 6 2 Configurable Parameters uuu sisse esiste eain tnnt 91 6 7 SUPPORT OF EXTENDED CHARACTER 5 5 92 6 8 MODALITY WORKLIST INFORMATION MODEL DEFINITION 92 6 8 1 Introduction aoa ecole mita Nata tei 92 6 8 2 Modality Worklist Information Model Description sse 92 6 8 3 Modality Worklist Information Model Entity Relationship Model 93 6 8 4 Entity Descriptions i3 eie dde dei REM cd aeta 93 6 8 4 1 Scheduled Procedure 93 6 8 4 2 Requested P
46. Carbon monoxide O 15 150 CO2 carbon dioxide C B103B Carbon dioxide O 15 OAc Acetate C B1043 Acetate C 11 Palmitate C B1044 Palmitate C 11 11C CO carbon monoxide C B1045 Carbon monoxide C 11 11C CO2 carbon dioxide Y X1754 CO2 carbon dioxide Rubidium cation Y X1755 Rubidium cation FluoroSpiperone C B1033 Spiperone F 18 L 2 Fluorotyrosine Y X1757 L 2 Fluorotyrosine Misonidazole C BO7E1 Fluormisonidazole F 18 11C Butanol Y X1759 11C Butanol Deoxyglucose Y X1760 Deoxyglucose Glucose Y X1761 Glucose Methionine C B103E Methionine C 11 N MethylSpiperone Y X1763 N MethylSpiperone Raclopride C B1042 Raclopride C 11 Thymidine FLT C B1036 Thymidine FLT F 18 L 1 Tyrosine Y X1766 L 1 Tyrosine 150 Butanol C B07DC Butanol O 15 EDTA C BO7DD EDTA Ga 68 PTSM C BO7E7 PTSM Cu 62 Choline choline Choline Rubidium chloride C B1037 Rubidium chloride Rb 82 Sodium C B1047 Sodium Na 22 62CuJATSM C B07DB ATSM Cu 64 11C Carfentanil C B103F Carfentanil C 11 11C Flumazenil C BO7DE Flumazenil C 11 18F Flumazenil C BO7DF Flumazenil F 18 Fluorethyltyrosin 07 0 Fluorethyltyrosin F 18 Fluoromethane 07 2 Fluoromethane F 18 Fluorouracil C B07E3 Fluorouracil F 18 Fluorobenzothiazole 07 4 Fluorobenzothiazole F 18 Germanium C B1046 Germanium Ge 68 13N Glutamate C B103D Glutamate N 13 1124 Monoclonal Antibody C BO7E6 Monoclonal antibody 1 124 Sodium Flouride C B1032 Sodium flouride F 18 I124 S0dium lod
47. Conformance Statement REV 11 DOC0550678 Direction 5330620 1EN 3 Real World Activity Choose Get Option will be available once the Query Remote activity is performed The operator can now select one or more study series or image and ask the DICOM Server AE to retrieve the selected image s from the Remote DICOM AE by choosing Get Examination Get Series or Get Images 4 Real World Activity Choose Ping DICOM host Option consists of an operator selecting Ping DICOM host from the Network pull down menu This will cause the DICOM Server AE to initiate a DICOM Verification Request to the remote AE to verify the remote system activeness Real world Activities for Discovery 710 and Discovery 610 SW Version pet coreload xx Push Images This Real World Activity consists of an operator selecting one or more studies series or images in the local database browser The operator then clicks on the destination button in the network panel at the bottom of the local database browser Real World Activity Query Remote causes the DICOM Server AE to initiate an association to the Remote DICOM AE and request the list of all studies Once the DICOM Server AE receives the list of studies the operator will have to choose the study and the local database browser will list the series of the study chosen After receiving the list of series the DICOM Server AE will ask for the list of images for the series chosen by the operato
48. Image SOP instance is described in the Modality Work list Conformance statement Use of the information retrieved in MPPS SOP instances is described later in this document After Post processing Exam Split application presents associated requested Procedure s along with selected acquired Images Exam Split application includes the necessary information related to Requested Procedure scheduled Procedure Steps and the Performed Procedure Step of the images acquired during acquisition After User modifications on Image s user has to click on Send button to create Gray Scale Presentation State and Performed Procedure Step objects to transmit to remote host If there is any Image Orientation user will be notified about the orientation before sending to Remote Host Exam Split will provide Host Selection user interface to select remote host to transmit of Gray Scale Presentation State object Performed Procedure Step object will be sent to default Performed Procedure Step host After successful transmission of Gray Scale Presentation State and Performed Procedure Step objects to remote host user will be notified with success 123 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 2 4 GSPS Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU SOP Class Name SOP Class UID Grayscale Softco
49. Implementation UIDs for different product versions DOC0550678 24 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 1 2 Overall Conformance Statement Documentation Structure The Documentation Structure of the GEHC DICOM Conformance Statements is shown in the Illustration below GEHC DICOM Conformance Statements http Awww ge com DICOM APPLICATION ENTITY SPECIFICATION SERVICE CLASSES INFORMATION OBJECTS MESSAGE EXCHANGES ETC AR Advanta PET Discovery Prod uct 710 610 and Conformance Optima 560 Statement Imp leme ntatio n Conformance Direction Statement Direction 5330620 1EN DICOM STANDARD e DIC OM e Part 16 Standard Specification DOC0550678 25 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN This DICOM Conformance Statement documents the DICOM Conformance Statement and Technical Specification required to interoperate with the GEHC network interface The GEHC Conformance Statement contained in this document also specifies the Lower Layer communications which it supports e g TCP IP However the Technical Specifications are defined in the DICOM Part 8 standard For more information regarding DICOM copies of the Standard may be obtained on the Internet at http medical nema org Comments on the Standard may be addressed to DICOM Se
50. M Sent DCM CT Dose Length Product Total 4 gt CONTAINS NUM EV 113814 1 U Not sent DCM CT Effective Dose Total 5 gt gt HAS TEXT EV 1 MC Not sent PROPERTI 121406 DCM ES Reference Authority 6 gt gt HAS CODE 1 MC Not sent PROPERTI 121406 DCM ES Reference Authority 138 of 215 Direction 5330620 1EN GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 7 gt gt HAS CODE EV G 1 M Not sent CONCEPT C036 SRT MOD Measurement Method 8 gt gt HAS TEXT EV 1 MC Not sent PROPERTI 113815 DCM ES Patient Model 9 gt gt HAS CONT 113816 1 MC Not sent PROPERTI AINE IDCM ES R Condition Effective Dose measured 10 gt gt gt CONTAINS EV 1 M Not sent 113817 DCM Effective Dose Phantom Type 11 gt gt gt CONTAINS 113818 1 M Not sent DCM Dosimeter 12 gt CONTAINS TEXT EV 121106 1 U Not sent DCM Comment 9 4 3 TID 10013 CT Irradiation Event Data NL Rel with VT Concept VM Req Notes Parent Type 1 CONT EV 113819 1 M Sent AINE DOM CT R Acquisition 2 gt CONTAINS TEXT EV 125203 1 U Not sent DCM Acquisition Protocol CONTAINS ICODE EV 123014 1 M Sent See note below for DCM Target specifics Regio
51. Material Code Module The attributes referenced here are not supported For the MPPS associated with an acquisition the following attributes are copied from the Modality Work list SCU into the MPPS request Message if procedure performed corresponds to the SPS information retrieved through the Modality Work list Discovery supports the selection of single or multiple SPS for a scan The following are applicable Single SPS selection results single PPS message Multiple SPS selection results in single PPS message Multiple SPS selection is allowed only if they all correspond to same patient id maximum of 15 SPS s can be selected e Referenced Study Sequence a maximum of ten Item s is supported This attribute will be present only if SPS information is available from Modality Work list SCU At the end of acquisition the user might choose to Defer PPS and later choose to Complete PPS or Discontinue PPS from the user interface provided in the system In this case the date and time when user chooses to Complete PPS or Discontinue PPS is taken as the Performed Procedure Step End Date and Performed Procedure Step End Time respectively Not the actual end date and end time of acquisition Mapping of SPS data to MPPS SOP instance is explained in section 7 10 Mapping of specific SPS data to CT DICOM IMAGE HEADER for PPS is explained in section 7 11 For the MPPS associated with a post processing t
52. Medicine Image Storage 1 2 840 10008 5 1 4 1 1 20 Ultrasound Image Storage 1 2 840 10008 5 1 4 1 1 6 X Ray Angiographic Image Storage 1 2 840 10008 5 1 4 1 1 12 1 Secondary Capture Image Storage 1 2 840 10008 5 1 4 1 1 7 Grayscale Softcopy Presentation State 1 2 840 10008 5 1 4 1 1 11 1 Storage Enhanced SR Storage 1 2 840 10008 5 1 4 1 1 88 22 GEMS PET Raw Information Storage 1 2 840 113619 4 30 3 3 1 2 1 1 2 FSC Directory Options DOC0550678 59 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Refer to Appendix C for a complete listing of all modules and attributes used in the DICOMDIR definition NOTE Modality attribute 0008 0060 in the DICOMDIR Series records should be CT for Secondary Capture SOP Class Images Otherwise the GEHC CT system s DICOM media display browser will not list up the series for contents rendering the retrieval to local storage impossible For USB only one DICOM file set shall be stored in the first partition of a partitioned device If the device is not partitioned only one DICOM file set shall be stored on the device Attention The Discovery 710 610 supports type of raw data format called PET List Data with SOP class name GE Private PET List data and SOP class UID 1 2 840 113619 4 30 Though it is having same SOP class UID as PET Raw data PET List data won t be supporting any Archive Restore operations Exam cont
53. Module This section specifies the attributes of the Patient that describe and identify the Patient who is the subject of a diagnostic Study This Module contains Attributes of the patient that are needed for diagnostic interpretation of the Image and are common for all studies performed on the patient Attribute Name Tag Type VR VM Notes Patient s Name 0010 0010 2 PN 64 1 Patient ID 0010 0020 Patient s Birth Date 0010 0030 DA 26 Patient s Sex 0010 0040 N 4 a 2 D 1 2 Discovery PET RAW Data Exam Module This section specifies the attributes which describe and identify the Study performed upon the Patient Attribute Name Tag Type VR VM Notes Study Instance UID 0020 000D 1 Ul 1 Study Date 0008 0020 2 DA 1 Study Time 0008 0030 2 TM 1 Referring Physician s Name 0008 aia 1 Accession Number 0008 199 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN D 1 3 Discovery PET RAW Data Scan Entity Modules This section specifies the attributes which identify and describe general information about the Scan within a Study Attribute Name Tag VR VM Notes Modality 0008 0060 Series Instance UID 0020 000E Series Number Operators Name Patient Position 0018 5100 2C CS 1 D 1 4 Discovery PET RAW Data Equipme
54. Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Bits Stored 0028 0101 Referenced SOP Class UID in File 0004 1510 If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL Referenced SOP Instance UID in File 0004 1511 Pixel Spacing C 5 5 PRESENTATION Keys If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL Key Tag Type Type Attribute Description for CD for DVD USB Specific Character Set 0008 0005 1C 1C ISO IR 100 Insta
55. RAW menn 199 0 1 GE PRIVATE PET RAW DATA INFORMATION OBJECT IMPLEMENTATION 199 D 1 1 Discovery PET RAW Data Patient Entity Module 199 D 1 2 Discovery PET RAW Data Exam Module seen 199 D 1 3 Discovery PET RAW Data Scan Entity Modules 200 D 1 4 Discovery PET RAW Data Equipment Entity Modules 200 D 1 5 Private Creator Identification Information sees 200 D 1 6 Discovery PET Patient Module sse 200 D 1 7 Discovery PET Exam Module sse 200 D 1 8 Discovery PET Scan Module sss ennt 200 D 1 9 Discovery PET Frame Module einen tren ni ndn dn 200 D 1 10 General Modules i eere den sd a i Pe doe a nea dames 201 0 1 10 1 SOP Common Module 201 D 1 11 Discovery PET Raw Data 2 2 6 201 D 1 12 Discovery PET Correction Cal Module 204 D 1 13 Discovery PET Well Counter Module sss nnne 204 0 2 GE PRIVATE PET LIST DATA INFORMATION OBJECT IMPLEMENTATION 206 D 2 1 Discovery PET LIST Data Patient Entity 206 D 2 2
56. Remote Host entry is selected from the Network pull down menu Once a list of Study Series Image is retrieved the operator can invoke the Get operation by choosing Get Exam or Get Series or Get Image from the Network pull down menu Note Get Exam will not pull the PET List Series list objects if the remote exam contains the same Remote Host Exam containing PET List Series will not show the same during the Network Pull and search operations Behavior for Discovery 710 610 Products SW Version pet coreload xx The operator must select the Remote DICOM AE from the Source pull down on the local database browser and select the hostname of the source from where the images are to be retrieved Provided that the operator had set Custom search to Off when setting the Remote host parameters the Query operation will cause the DICOM Server AE to initiate an association with zero length Patient name Patient id Study date Accession number and Study id to the selected Remote AE when the Hostname entry is selected from the Source pull down menu DOC0550678 59 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Otherwise if the operator had set Custom search to On when setting the Remote host parameters the Query operation will cause a Customize search parameters menu to appear The operator can enter values for Patient name Pati
57. Request Presentation Context Table Abstract Syntax Transfer Syntax Extended Negotiation Name UDUs Study Root 1 2 840 10008 5 1 4 1 2 2 2 Implicit VR 1 2 840 10008 1 2 Query Retrieve Little Endian DOC0550678 50 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN _ f 2 3 1 3 4 2 1 SOP Specific Conformance to C MOVE SCP The DICOM Server AE provides standard conformance to the baseline Study root C MOVE Service Class SCP Each C MOVE SCP operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 300 seconds and 60 minutes respectively These time outs are configurable in dcs cfg as bi assoc tio bi move tio and bi session tio respectively All images requested in a C MOVE RQ will be sent over a single association A C MOVE RSP with a pending status will be returned to the requester every five images The C MOVE SCP will invoke C STORE requests for the following SOP classes 1 2 840 10008 5 1 4 1 1 2 1 2 840 10008 5 1 4 1 1 4 1 2 840 10008 5 1 4 1 1 88 67 In addition to the C MOVE response status values defined in DICOM part 4 the following status values will be returned e C000 Indicates that an error occurred while retrieving records from the local database C001 Indicates all other processing error C011 If the Destination AE retu
58. Requests the Print SCP to create an instance of Basic Film Session The following Attribute values are supported by the N CREATE Attribute DICOM Tag Valid Range Default Value Number of Copies 2000 0010 Print Priority 2000 0020 HIGH MED LOW Set in Configuration file Default value is HIGH Medium Type 2000 0030 CLEAR FILM Set in Configuration File PAPER PROCESSOR Note denotes that the attribute is optional for the SCU However we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value If Failure status is returned during N CREATE operation of Film session the following action will be taken by Print SCU DOC0550678 68 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 0x213 Resource Limitation message will be logged All other status Failure message will be logged In all the cases the print session will be terminated 4 3 1 2 2 2 Basic Film box SOP Class The Print SCU supports the following DIMSE Service Elements for the Basic Film Box SOP Class N CREATE Requests the Print SCP to create an instance of Film Box N ACTION Requests the Print SCP to print the Film Box onto Printer N DELETE Requests the Print SCP to delete the Film Box Instance
59. Stack The Point to Point Communication Stack is not supported by this implementation 6 5 Extensions Specializations Privatizations 6 5 1 Standard Extended Specialized Private SOPs Modality Worklist for Discovery does not implement any private transfer SOP classes 6 5 2 Private Transfer Syntaxes Modality Worklist for Discovery does not implement any private transfer syntaxes 6 6 CONFIGURATION The Modality Worklist feature is configured by GEHC Field Service engineers The DICOM configuration items below are configurable or re configurable by a Field Service Engineer and are not accessible by users through the Discovery user interface 6 6 1 AE Title Presentation Address Mapping ModalityWorklist allows for the configuration of the following parameters that pertain to the remote AE Remote AE HIS RIS IP address IP address used to contact the remote AE Remote AE HIS RIS IP port IP port used to contact the remote AE These parameters define where worklist queries will be directed Configuration of these parameters is performed by GEHC Field Service engineers using the Modality Worklist installation facilities 6 6 2 Configurable Parameters The following parameters are configurable for the DICOM Worklist Server AE DOC0550678 91 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Local Worklist Server AE Title automatically set to hos
60. a list of Study Series Image Once the list of Study Series Image is retrieved the operator can invoke the Get operation by choosing Get Exam or Get Series or Get Image from the Network pull down menu Behavior for Discovery 710 610 Products SW Version pet coreload xx The operator must select the Remote DICOM AE from the Source pull down on the local database browser and select the hostname of the source from where the images are to be retrieved Once a list of Study Series Image is retrieved the operator can invoke the Retrieve operation by clicking on the Local DB button at the bottom of the local database browser 2 3 1 2 3 2 Proposed Presentation Contexts When the Real World activity Get is initiated all presentation contexts shown in table 2 2 1 2 1 2 1 are proposed during association establishment but only the Query Retrieve MOVE related contexts are applicable to this activity DOC0550678 42 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 2 3 1 2 3 2 1 SOP Specific Conformance Statement for C MOVE SCU When the operator starts a Get operation at any level Study Series Image the DICOM Server AE will initiate a C MOVE RQ request to the Remote AE with the DICOM Server AE as the Destination AE The Storage SCP will handle the incoming images as described in section 2 3 1 3 1 A user attention pop up will be posted if a failure s
61. a person s name or a code DICOM information objects can be transmitted with either explicit identification of the type of each data element Explicit VR or without explicit identification Implicit VR with Implicit VR the receiving application must use a DICOM data dictionary to look up the format of each data 1 8 Symbols and Abbreviations AE Application Entity CD R Compact Disk Recordable CT Computed Tomography DICOM Digital Imaging and Communications in Medicine FSC File Set Creator FSU File Set Updater FSR File Set Reader GSPS Grayscale Softcopy Presentation State HIS Hospital Information System IHE Integrating the Healthcare Enterprise IOD Information Object Definition JPEG Joint Photographic Experts Group MPPS Modality Performed Procedure Step MSPS Modality Scheduled Procedure Step MTU Maximum Transmission Unit IP MWL Modality Worklist NTP Network Time Protocol Optional Key Attribute PACS Picture Archiving and Communication System PET Positron Emission Tomography R Required Key Attribute RIS Radiology Information System SC Secondary Capture SCP Service Class Provider SCU Service Class User SOP Service Object Pair SR Structured Reporting TCP IP Transmission Control Protocol Internet Protocol U Unique Key Attribute VR Value Representation DOC0550678 50 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 2 NETWORK CONFORMANCE STATEMENT 2 1 I
62. changed and is no longer backward compatible with previous PET CT systems Sending Pushing Discovery 710 610 and Optima 560 Raw Data Series onto a non Discovery 710 610 or Optima 560 system will also cause the network process to fail This applies to the GE proprietary PET Raw Data IOD Image Installation If the image installation is unsuccessful a message will appear in the Message Log informing the user of the failure and the image will be removed If the image installation process finds that an element is not encoded according to the DICOM standard it will fail to install the image and the file will be removed Image Installation of non GE Created CT MR or PET Images Images received from non GE products are installed as appropriate image object without any of their private data elements Also if some critical fields mandatory are missing then the image will not be installed 2 3 1 3 2 Verification Request from Remote AE This AE is indefinitely listening for associations No operator action is required to respond to a verification message 2 3 1 3 2 1 Associated Real World Activity The Real World Activity associated with the verification request is to send a C ECHO response message with a status of success to the requesting AE 2 3 1 3 2 2 Presentation Context Table Table 2 3 1 3 2 2 1 Acceptable Presentation Contexts for DICOM Server AE and Real World Activity Verification Request Presentation Context Table Abst
63. changes and GE Products are enhanced to support these changes Interaction It is the sole responsibility of the non GE provider to ensure that communication with the interfaced equipment does not 2701215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN cause degradation of GE imaging equipment performance and or function 1 6 References NEMA PS3 Digital Imaging and Communications in Medicine DICOM Standard available free at http medical nema org 1 7 Definitions Informal definitions are provided for the following terms used in this Conformance Statement The DICOM Standard is the authoritative source for formal definitons of these terms Abstract Syntax the information agreed to be exchanged between applications generally equivalent to a Service Object Pair SOP Class Examples Verification SOP Class Modality Worklist Information Model Find SOP Class Computed Radiography Image Storage SOP Class Application Entity AE an end point of a DICOM information exchange including the DICOM network or media interface software i e the software that sends or receives DICOM information objects or messages A single device may have multiple Application Entities Application Entity Title the externally known name of an Application Entity used to identify a DICOM application to other DICOM applications on the network Application Context the sp
64. database with Archive Node option set to Yes Note The remote node should be a Storage Commitment SCP DOC0550678 78 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN The operator must then select Archive pull down menu on the local database manager Select Select Media option menu Select Remote Node in the menu Note Only if you configure the remote node as a archive node will you see the remote node option in the archive menu The user selects the exam series to be committed All the images currently in the selected exam series will be sent to the selected remote archive node which is also the Storage commitment SCP using DICOM C STORE operations Once all the images are transferred the commitment request will be sent on a separate association Behavior for Discovery 710 610 Products SW Version pet_coreload xx The operator must first set remote hosts from Network Configuration from Tools pull down menu If the remote node is already present select one remote host click Edit on the right side bar and make a check on Archive Node check box If the remote host is not present add the remote host with check Note The remote node should be a Storage Commitment SCP STC AE Title can be different from STORE AE title The user selects the exam series to be committed and click on the archival destination from the archive list at the bottom
65. from any other DICOM Compliant system to another DICOM station Discovery also allows query and retrieve of data stored in its local database from a remote station and can query and retrieve images stored in a remote DICOM station In this situation Discovery is providing the DICOM C FIND and C MOVE services as a service class provider SCP and that of a DICOM C FIND and C MOVE service class user SCU Connect Pro option provides additional capabilities Modality Worklist MWL and Performed Procedure Step MPPS which provide integration of this GEHC imaging equipment with Hospital Information System Note the format of this section follows the format of the DICOM Standard Part 2 conformance Annex A hence the paragraph numbering scheme Please refer to that part of the standard while reading this section 2 2 Implementation Model All DICOM functionality on the Discovery product is handled by the DICOM Server Application Entity AE The DICOM Server AE is commanded to perform DICOM services through the buttons and menu selections on the main user interface panel The DICOM Server AE is also listening to a pre defined port for incoming connections DOC0550678 51 of 215 GE Healthcare Discovery 70 QMnd Optima 560 DICOM Conformance Statement REV 11 Standard Choose Pus 2 1 Application Data Flow Diagram ILLUS N 2 0 IMPLE ji ATION MODEL DATA FLOW DIAGRAM Choose Query Remote Option Choose Get Option DICO
66. however the 26 of 215 GE Healthcare REV 11 Discovery 710 610 and Optima 560 DICOM Conformance Statement Direction 5330620 1EN device is acting as a full fidelity storage device it should retain and re transmit all of the private data elements which are sent by GEHC devices 1 5 Important Remarks The use of these DICOM Conformance Statements in conjunction with the DICOM Standards is intended to facilitate communication with GE imaging equipment However by itself it is not sufficient to ensure that inter operation will be successful The user or user s agent needs to proceed with caution and address at least four issues DOC0550678 Integration The integration of any device into an overall system of interconnected devices goes beyond the scope of standards DICOM and of this introduction and associated DICOM Conformance Statements when interoperability with non GE equipment is desired The responsibility to analyze the applications requirements and to design a solution that integrates GE imaging equipment with non GE systems is the user s responsibility and should not be underestimated The user is strongly advised to ensure that such an integration analysis is correctly performed Validation Testing the complete range of possible interactions between any GE device and non GE devices before the connection is declared operational should not be overlooked Therefore the user should ensure that any non GE pro
67. looks for following attributes in NLEVENT REPORT data received from Print SCP If Print SCU does not receive N EVENT_REPORT it requests the Print SCP to retrieve the following set of attributes using N GET 4 3 1 3 Association Acceptance Policy The Print SCU does not accept associations 4 4 Communication Profiles 4 4 4 Supported Communication Stacks parts 8 9 DICOM Upper Layer Part 8 is supported using TCP IP 4 4 1 1 TCP IP Stack The TCP IP stack is inherited from the UNIX Operating System 4 4 1 1 1 Not Applicable 4 4 1 1 2 Physical Media Support Ethernet v2 0 IEEE 802 3 4 5 Standard Extended Specialized Private SOPs None 4 6 Configuration 4 6 1 AE Title Presentation Address Mapping The Local AE title is hostname DCP where hostname is the unique hostname within the network DOC0550678 74 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN The Discovery system allows the user to Remove or Update the mapping of remote DICOM Printer AE Titles to IP Addresses and Ports These options can be selected from the Service Desktop camera installation utility Note If hostname DCP is greater than 16 characters then the name of the local AE title is limited to first 16 characters 4 6 2 Configurable Parameters The following fields are configurable for the DICOM Print SCU AE Note All configurations should be per
68. name name Performed Station Name 0040 0242 Local system Local system suite id Local system host suite id name Performed Location 0040 0243 Local system Local system suite id Not used suite id Performed Procedure Step 0040 0244 Same as exam Same as exam date Same as exam date Start Date start date Performed Procedure Step 0040 0245 Same as exam Same as exam time Same as exam time Start Time start time Performed Procedure Step 0040 0254 Copied from Not sent as part of Not sent as part of Description SPS if selected image header image header Performed Procedure Step 0040 0252 See Note 1 See Note 1 See Note 1 Performed Procedure Type Description 0040 0255 Not sent as part of image header Not sent as part of image header Not sent as part of image header Procedure Code Sequence 0008 1032 Not sent as part of image header Not sent in MPPS Not sent as part of image header Not sent in MPPS N Not sent as part of image header Not sent in MPPS N N Create Create Create Performed Procedure Step 0040 0250 Date when all Date when all images The date Complete End Date images got got installed PPS or installed Discontinue PPS is invoked not the actual end of post processing Performed Procedure Step End Time 0040 0251 Time when all images got installed Time when all images got installed The time Complete PPS or Discontinue PPS is invoked no
69. needed for diagnostic interpretation of the Image and are common for all studies performed on the patient Attribute Name Tag Type VR VM Notes Patients Name Patient ID Patient s Birth Date Patient s Sex D 2 2 Discovery PET LIST Data Exam Module This section specifies the attributes which describe and identify the Study performed upon the Patient Attribute Name Tag VR VM Notes Study Instance UID 0020 000D 1 UI 1 206 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Study Date 0008 0020 2 DA 1 Study Time 0008 0030 2 TM 1 Referring Physician s Name 0008 a MO 1 Accession Number 0008 D 2 3 Discovery PET LIST Data Scan Entity Modules This section specifies the attributes that identify and describe general information about the Scan within a Study Attribute Name Tag Type VR VM Notes Modality 0008 0060 Series Instance UID Patient Position 0018 5100 2C CS 1 Series Number Operators Name D 2 4 Discovery PET LIST Data Equipment Entity Modules This section specifies the attributes that identify and describe the piece of equipment which produced the instance of the Private SOP class Attribute Name Tag VR VM Notes Manufacturer 0008 0070 2 LO 1 Institution Name 0008 0080 Manufacturer s Model Nam
70. or router protections to ensure that only approved external hosts have network access to the product 2 Firewall or router protections to ensure that the product only has network access to approved external hosts and services DOC0550678 55 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 3 Any communications with external hosts and services outside the locally secured environment use appropriate secure network channels such as a Virtual Private Network VPN DOC0550678 54 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 3 MEDIA STORAGE CONFORMANCE STATEMENT 3 1 Introduction This conformance statement specifies the Discovery conformance to the DICOM Media Interchange It details the DICOM Media Storage Application Profiles and roles that are supported by this product The GEHC PET CT system provides capabilities to DICOM Media Interchange on CD Compact Disc DVD Digital Video Disc Recordable and USB Storage device with VFAT filesystem The system works with most of the IOD s like Computed Tomography CT Magnetic Resonance MR Positron Emission Tomography PET and Digital X Ray images 3 2 Implementation Model Media Server All DICOM functionality on the Discovery and Optima products is handled by the DICOM Server Application Entity AE The DICOM Server AE is commanded to perfor
71. the Enhanced SR SOP class Off will not produce any SOP class 142 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN APPENDIX A CT PET Image and Secondary Capture Modules Attributes Enclosed is a listing of only the Optional Conditional modules attributes used by this implementation for CT SC Note All other modules that are mandatory and attributes that are type 1 amp 2 in nature per the DICOM standard are supported by this implementation but are not explicitly listed here A 1 CT Specific IOD Definition A 1 1 CT Image IOD Modules Entity Name Module Name R f rence uem Pme M Suy Gewmsuy A32 x Paensuy A33 U Series Seres A34 m Frame of Reference Frame of Reference ass M Equipment General Equipment A36 m lt lt lt mae General mage moe M A42 6 x ermae aaa vor A48 sor common A46 M A 2 SC Image IOD lt lt lt lt A 2 1 SC Image IOD Modules Entity Name Module Name Reference Usage Mem Paten Suy A32 Patient Stucy A33 9 Series General Seres A34 M Equipment General Equipment A36 A81 mag General mage A
72. the user must press the Print Button to print the images The Print SCU retrieves the Print SCP status by using the N GET service of Printer SOP Class If the printer returns a FAILURE status the print session will be terminated The printer status is not read The film session is created using the N CREATE service The print session will be terminated if the printer fails to create the film session DOC0550678 64 of 215 GE Healthca re Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 DOC0550678 Direction 5330620 1EN The film box is created using the N CREATE service The print session will be terminated if the printer fails to create the film box The images are placed on the film box by using the N SET If the printer fails to place the image on the film box the print session will be terminated The film will be printed using the N ACTION Only film box printing is supported If the printer fails to print the film the print session will be terminated After the successful N ACTION the film box will be deleted using the N DELETE Any N EVENT REPORTS given to the PRINT SCU will be received but the data is ignored After the film box is deleted the association will be closed Upon the successful completion of the above sequencing and the association is closed the user will be notified of the successful print session 65 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Sta
73. this value as null Referenced Series Sequence 0008 1115 Copied from source image Displayed area selection sequence 0070 005A Exam Split Application will calculate values for Window Width Window Level pan and zoom 132 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Softcopy VOI LUT Sequence 0028 3110 Exam Split Application will calculate and copy from acquired Image s Image Horizontal flip 0070 0041 Exam Split Application will calculate from Image Orientation Image Rotation 0070 0042 Exam Split Application will calculate from Image Orientation Presentation LUT Shape 2050 0020 Exam Split Application will set as IDENTITY SOP Class UID 0008 0016 Exam Application will set as 1 2 840 10008 5 1 4 1 1 11 1 SOP Instance UID 0008 0018 Copied from acquired study Specific Character Set 0008 0005 Copied from acquired study Referenced Patient Sequence 0008 1120 Scheduled Step Attribute Sequence 0040 0270 Copied from source image Originally from SPS sequence gt Accession Number 0008 0050 gt Referenced Study Sequence 0008 1110 gt Study Instance UID 0020 000D gt Requested Procedure Description 0032 1060 gt Requested Procedure Code Seque
74. to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL Trigger Time 0018 1060 Reconstruction Diameter 0018 1100 If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL Samples Per Pixel 0028 0002 2 If present in composite object instances it will be set to same value other wise set to NULL Bits Allocated 0028 0100 2 If present in composite object instances it will be set to same value other wise set to NULL 195 of 215 ili Columns 0028 0011 Echo Numbers 0008 0086 Gantry Detector Tilt 0018 1120 2 If present in composite object instances it will be set to same value other wise set to NULL GE Healthcare Discovery 710 610 and
75. voxel center of the first voxel transmitted of the grid in mm in the registered Frame of Reference The system can be configured to save the upper left hand corner Z coordinate value into 0020 1041 Slice Location instead of the gantry ISO center table location Image Position 0020 0032 1 Always sent System configuration allows images to be flipped and rotated Default behavior is to not allow flip and rotate Image Orientation 0020 0037 1 Always sent System configuration allows images to be flipped and rotated Default behavior is to not allow flip and rotate Pixel Spacing 0028 0030 Always sent 149 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 A 4 2 Image Pixel Module Direction 5330620 1EN Attribute Name Tag Type Notes Samples per Pixel 0028 0002 Always sent with value 1 Photometric 0028 0004 1 Always sent Interpretation 0028 0010 Bits Allocated Always sent Always sent Always sent with value 16 7 00010 0028 0100 A 4 3 Contrast Bolus Module Attribute Name Tag Typ Notes Contrast Bolus Agent 0018 0010 2 Sent if contrast exam as entered in user interface Contrast Bolus Route 0018 1040 3 Sent if contrast exam as entered in user interface A 4 4 CT Image Module Image Type 0008 0008 150 of 215 Always sent See section 2 5 1 Value 3 AXIAL LOCALI
76. 0678 5 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Presentation Context Table Proposal Abstract Syntax Transfer Syntax Extended Negotiation UID Lit 2 1 2 840 10008 5 1 4 1 1 7 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None Secondary Capture Implicit VR Big Endian GE 1 2 840 113619 5 2 Image Storage Private Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 Study Root 1 2 840 10008 5 1 4 1 2 2 Implicit VR Little 1 2 840 10008 1 2 SCU None Query Retrieve 1 Endian FIND Study Root 2 2 840 10008 5 1 4 1 2 2 Implicit VR Little 1 2 840 10008 1 2 None Query Retrieve Endian MOVE Bis UNT RN E Class Endian PET Image 1 2 840 10008 5 1 4 1 1 12 Implicit VR Little Endian 1 2 840 10008 1 2 None Storage 8 Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 Enhanced SR E Implicit VR Little Endian 1 2 840 10008 1 2 None Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 X Ray Radiation E A 840 10008 5 1 4 1 1 88 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None Dose SR Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 GE Private PET 1 2 840 113619 4 30 Implicit VR Little Endian 1 2 840 10008 1 2 SCU None Raw data SOP Explicit VR Little Endian 1 2 840 10008 1 2 1 class Explic
77. 1 Reference Usage Directory Information DICOM Part 3 Annex F 3 C 2 File Set Identification Module Attribute Name Type File Set ID Set by application C 3 Directory Information Module Attribute Name Tag Typ Notes Offset of the First Directory Record of the Root 0004 1200 1 Set by application Directory Entity Offset of the Last Directory Record of the Root 0004 1202 1 Set by application Directory Entity File set Consistency Flag 0004 1212 1 0000H no known inconsistencies Directory Record Sequence 0004 1220 Not supported 2 Offset of the Next Directory Record 0004 1400 Set by application gt Record In use Flag 0004 1410 FFFFH record is in use 2 Offset of Referenced Lower Level Directory 0004 1420 Set by application Entity gt Directory Record Type 0004 1430 1C PATIENT STUDY SERIES IMAGE PRESENTATION SR DOCUMENT and PRIVATE gt Private Record UID 0004 1432 Not supported gt Referenced File ID 0004 1500 1C Generated only for Image Directory Record starting with A A A A ZO1 gt MRDR Directory Record Offset 0004 1504 Not generated gt Referenced SOP Class UID in File 0004 1510 IC Generated for Image Directory Record 191 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN gt Referenced SOP Instance UID in File 0004 1511 1C Generated for Image Directory Record Set to SOP Instance UID 0008 0018 dur
78. 10008 1 2 4 50 Process 1 STD GEN USB JPEG only See Table See Table JPEG Extended Process 2 amp 4 3 3 1 2 1 1 1 1 3 3 1 2 1 1 1 1 Default Transfer Syntax for 1 2 840 10008 1 2 4 51 Lossy JPEG 12 Bit Image Compression Process 4 only STD GEN USB JPEG only 3 4 Augmented and Private Application Profiles 3 4 1 Augmented Application Profiles The CD DVD USB Media Archive Interchange AE does not support any augmented Application Profiles 3 4 2 Private Application Profiles The CD DVD USB Media Interchange AE does not support any private Application Profiles 3 5 Extensions Specializations and Privatization of SOP Classes and Transfer Syntax 3 5 1 Extensions Specializations and Privatizations of SOP Classes The CT MR PET SOP Class Images have definitions extended for Defined Terms and include GE specific Private Data elements The following sections describe the details for these SOP classes 3 5 1 1 SOP Specific Conformance Statement for CT SOP Class Refer to section 2 5 for standard extensions and Appendix B for private data elements 3 5 1 2 SOP Specific Conformance Statement for MR SOP Class Refer to section 2 5 for standard extensions and Appendix B for private data elements DOC0550678 62 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 3 5 1 3 SOP Specific Conformance Statement for PET SOP Class Refer to section 2 5 for standard extensions and A
79. 10008 5 1 1 16 DICOM Implicit 1 2 840 10008 1 2 VR Little Endian Transfer Syntax Print Job SOP 1 2 840 10008 5 1 1 14 DICOM Implicit 1 2 840 10008 1 2 Class VR Little Endian Transfer Syntax None None None Note Certain Presentation Contexts may be negotiated that are not used during the association See the following sections for the DICOM Print services performed by the Discovery DOC0550678 6 01215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 4 3 1 2 2 SOP Specific Conformance Statement The Print SCU supports the following mandatory SOP classes which are defined under the Basic Grayscale Print Management Meta SOP Class The Print SOP Classes supported by Print SCU Optional Print SOP Classes supported by Print SCU NAME Print Job SOP Class 1 2 840 10008 5 1 1 14 Note The asterisk preceding some of the SOP attributes in the following sections indicates optional SCP attributes The SCU is aware of these optional attributes and will not fail on the receipt of a response without the optional attributes the limitation being that this information is not available for the User Some SCP s will terminate the association with a failure upon the receipt of an optional attribute it does not support 4 3 1 2 2 1 Basic Film Session SOP Class The Print SCU supports the following DIMSE Service Elements for the Basic Film Session SOP Class N CREATE
80. 121058 1 M Code sequence P5 08000 CONCEPT DCM SRT Computed Tomography MOD Procedure X ray sent reported 3 gt INCLU DTID 1002 1 M Observer context sent with DE Observer Device context values equal to Context Implementation UID Single item sent 4 gt HAS OBS DATE EV 113809 1 M Sent CONTEXT TIME DCM Start of X ray Irradiation 5 HAS OBS DATE EV 113810 1 M Sent CONTEXT TIME DCM End of X ray 137 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Irradiation 6 gt HAS OBS EV 113705 1 M Code sequence 113014 CONTEXT DCM Scope of DCM Study sent Accumulation 7 gt gt HAS UIDR DCID 10001 1 M Study Instance UID sent PROPERTI EF UID Types ES 8 gt CONTAINS INCLU DTID 10012 1 M See CT Accumulated Dose DE CT Data table below for details Accumulated Dose Data 9 gt CONTAINS INCLU DTID 10013 1 n M See CT Irradiation Event Data DE CT Irradiation table below for details Event Data 10 gt CONTAINS TEXT JEV 121106 1 U Not sent DCM Comment 9 4 2 TID 10012 CT Accumulated Dose Data NL Concept Name Req Notes Parent Type 1 CONT EV 113811 1 M Sent AINE IDCM R Accumulated Dose Data 2 CONTAINS NUM 113812 1 M Sent DCM Total Number of Irradiation Events 3 gt CONTAINS NUM 113813 1
81. 181210 Reconstruction diameter O 00181100 Retumed Data collection diameter 00182090 Retumed ipae _ o 00186134 Reed Echo number O 001860086 fEchotime O 001860081 Remes inversion time O 00182082 Repetition time o 001860080 Tiger tim o 0018060 DovRet P 009 00 Retumed MdscanTime P 00190024 iuh P 001900 INumberofEcho P 0019007E ScotAwet P 0021004A LoctonRAS P 0027040 octo 00270041 Remes Center R Coordinate P 0027042 CenterA Coordinate P 00270043 Table Start Location P 0027060 Retumed Table End Location P 00270081 Retumed RAS Letter for Side P 00270062 Retumed DOC0530678 49 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Description 7 Type Tag Usage RAS Letter for Anterior Posterior P 0027 005 p RAS Letter for Scout Start Location P 0027 0054 Returned RAS Letter for Scout End Location P 0027 0055 Returned ImageDimension X 14 P 00270060 Returned ImageDimension Y 00270061 Returned Note In the above tables the type field h
82. 1EN 1 2 840 10008 5 1 4 1 1 128 Note The Discovery 710 610 supports type of raw data format called PET List Data with SOP class name GE Private PET List data and SOP class UID 1 2 840 113619 4 30 Though it is having same SOP class UID as PET Raw data PET List data won t be supporting any Archive Restore operations Exam containing the PET List Series will skip the same during Save by Exam The Storage Commitment SCU is not supported for GE Private PET List Data o Save by Series on PET List series not supported o Save by Image on PET List frame not supported Note The Optima 560 product does not support PET List Data 5 3 1 2 3 Storage Commitment Push Model SOP Class Request Processing The images in the selected exam series are sent to the remote commitment provider entity using DICOM C STORE operations If there are any failures in the image transfers the Storage commitment request will not be sent The corresponding job will be marked as failed and user will be notified of the status of job If all the images are successfully transferred then the commitment request will be sent on a different association with the list of sop instances If the N ACTION request fails the job will be marked as failed otherwise following sequence of actions will take place The SCU waits for N ACTION RSP from provider on the same association for a configurable amount of time If it does not receive N ACTION RSP during this time it will cl
83. 2 Proposed Presentation Context 125 8 2 4 8 Association Acceptance Policy nn 125 8 37 SAVVE STATE roue ue Mu e S Iu ark teu x dM dE 126 8 3 1 Implementation 20000 000 sisse eene 126 8 3 2 Application Data Flow 126 8 3 3 Sequencing of Real World Activities esses 126 8 8 8 1 GSPS Acquisition System with MWL data 126 84 COMMUNICATION 9 9 22 126 8 4 1 Supported Communication Stacks PS 3 8 sees 126 8 4 2 OSLSIACK RU eere tenido eas 126 8 4 3 TGFP IP Stack SUL n HE E tod eod doa 127 LIEE I E ns 127 8 4 3 2 Physical Media Support 127 8 4 4 Point to Point Stack c eer rd RE RE eo RES sb 5 127 85 5 2 2 127 8 5 1 Standard Extended Specialized Private SOPS 127 8 5 2 Private Transfer Syntaxes isses sassa tha snnt 127 86 CONFIGURATION cg ridotto utter rentre sad nee rte eet va actes Dc 127 8 6 1 AE Title Presentation address Mapping s
84. 22 85 6 1 INTRODUCTION DEC 85 6 2 IMPLEMENTATION MODEL gan tete nea nis ccgueveats S 85 6 2 1 Application Data Flow Diagram 85 6 2 2 Functional na nan 86 6 2 3 Sequencing of Real World Activities sss 86 6 3 AESSPEGIFIGATIONS E tna metuis tt ted Miran ib avia ee 86 6 3 1 Worklist Server AE Specification 9 86 6 3 1 1 Association Establishment eene nnne tein 86 6 3 1 1 1 General usa I 86 6 3 1 1 2 Number of Associations anar 87 6 3 1 1 3 Asynchronous au uuu aq aa aaa Fe aed ie exte da det ays 87 6 3 1 1 4 Implementation Identifying Information 87 6 3 1 2 Association Initiation Policy u eterni reete to erre io a e doe 87 6 3 1 2 1 Real World Activity Worklist 87 6 3 1 2 1 1 Associated Real World 87 6 3 1 2 1 2 Proposed Presentation Context Table nennen 88 6 3 1 3 Association Acceptance 90 6 4 COMMUNICATION 0 0 90 6 4 1 Supported Communication Stacks PS 3 8 PS 3
85. 3 Media Conformance modified to include CD RW DVD RW USB profiles DVD RAM and MOD Profiles are removed Section 10 is added for Security Conformance Section 8 is updated for Save State 8 3 Appendix Section of CT is updated for 8 8 Image Type A3 7 and 4 4 HCSDMO00112251 SPR is addressed for PET Tags Q Static 11 Sep 18 2012 Joe Muth Update for Eagle per GEHC internal review comments See document sheet for specific details 11 Oct 3 2012 Joe Muth Update for Eagle per GEHC internal review comments See R2 document sheet for specific details 11 Oct 9 2012 Joe Muth Update for Eagle per GEHC internal review comments See R3 document sheet for specific details 11 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN CONFORMANCE STATEMENT OVERVIEW This DICOM Conformance Statement captures the DICOM capabilities of the GEHC PET CT scanners identified below This document applies to the following software releases for each product Discovery 710 pet_coreload xx pet_mfk xx Discovery 610 pet_coreload xx pet_mfk xx Optima 560 pet_mfk xx Table 0 1 provides an overview of the network services supported by GEHC PET CT products identified above Table 0 1 NETWORK SERVICES SOP Classes User of Service Provider of SCU Service SCP Transfer CT Image Storage MR Image Storage Secondary Capture Image Storage lt Grayscale Softcopy Presentation Stat
86. 33 LO no value available 0 0 RequestingService 00032 1060 LO no value available 0 0 RequestedProcedureDescription 00032 1064 SQ Sequence with explicit Length 4 1 32 1 RequestedProcedureCodeSequence fffe e000 na Item with explicit Length 3 24 1 Item 0008 0100 SH no value available 0 0 CodeValue 0008 0102 SH no value available 0 0 CodingSchemeDesignator 0008 0104 LO no value available 0 0 CodeMeaning fffe e00d na ItemDelimitationItem for re encoding 0 1 ItemDelimitationItem fffe e0dd na SequenceDelimitationItem for re enc 0 1 SequenceDelimitationItem 0038 0000 UL 32 4 1 VisitGroupLength 0038 0010 LO no value available 0 0 AdmissionID 0038 0050 LO value available 0 0 SpecialNeeds 0038 0300 LO no value available 0 0 CurrentPatientLocation 102 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 0038 0500 LO no value available 0 0 PatientState 0040 0000 UL 192 4 1 ModalityWorklistGroupLength 0040 0100 SQ Sequence with explicit Length 1 136 1 ScheduledProcedureStepSequence fffe e000 na Item with explicit Length 4 12 128 1 Item 0008 0060 CS no value available 0 0 Modality 00032 1070 LO no value available 0 0 RequestedContrastAgent 0040 0001 AE no value available 0 0 ScheduledStationAETitle 0040 0002 DA no value available 0 0 ScheduledProcedur
87. 5330620 1EN Insufficient memory in printer to C605 store the image Combined Print Image size is larger C613 than the Image Box size 4 3 1 2 2 5 Printer SOP Class N GET DIMSE service is supported for the Printer SOP Class If an DIMSE service is received when the association is active Print SCU handles the relevant states but the data received is ignored Print SCU issues the request to retrieve the following attributes The Print SCU issues the N GET service to retrieve the printer status The status is processed as follows If Printer status 2110 0010 is NORMAL the film is printed If Printer status 2110 0010 is FAILURE the print job is terminated The Printer Status Info 2110 0020 attribute is not processed If Printer status 2110 0010 is WARNING one of three things will happen a If the Printer Status Info 2110 0020 is SUPPLY LOW the status is displayed to the user and the print job continues b If the Printer Status Info 2110 0020 is RECEIVER FULL or SUPPLY EMPTY or FILM JAM the status is displayed to the user and the print job is aborted For all other Printer Status Info 2110 0020 values the status is ignored and the print job continues DOC0550678 73 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 4 3 1 2 2 6 Print Job SOP Class Print SCU
88. ACIENTA ELEKTRICK M PR DOM DO MECHANICK HO ALEBO IN HO NEBEZPE ENSTVA ESTE MANUAL DE SERVICIO SOLO EXISTE EN INGLES SI ALGUN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLES ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCION NO SE DEBERA DAR SERVICIO TECNICO AL EQUIPO SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS EL CTRICAS MEC NICAS O DE OTRA NATURALEZA DEN H R SERVICEHANDBOKEN FINNS BARA TILLG NGLIG P ENGELSKA OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRAK N ENGELSKA ANSVARAR KUNDEN FOR ATT TILLHANDAH LLA VERS TTNINGSTJ NSTER F RS K INTE UTF RA SERVICE P UTRUSTNINGEN OM DU INTE HAR L ST OCH F RST R DEN H R SERVICEHANDBOKEN OM DU INTE TAR H NSYN TILL DEN H R VARNINGEN KAN DET RESULTERA SKADOR PA SERVICETEKNIKERN OPERAT REN ELLER PATIENTEN TILL F LJD AV ELEKTRISKA ST TAR MEKANISKA FAROR ELLER ANDRA FAROR 9 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN DIKKAT BU SERVIS KILAVUZUNUN SADECE NG L ZCES TR MEVCUTTUR EGER M TER TEKN SYEN BU KILAVUZU NG L ZCE DI INDA B R BA KA L SANDAN TALEP EDERSE BUNU TERC ME ETT RMEK M TER YE D ER SERVIS KILAVUZUNU O
89. AR ESTES MANUAIS NOUTRO IDIOMA E DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVI OS DE TRADU O N O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSIST NCIA T CNICA O N O CUMPRIMENTO DESTE AVISO PODE COLOCAR EM PERIGO A SEGURAN A DO T CNICO DO OPERADOR OU DO PACIENTE DEVIDO A CHOQUES EL CTRICOS MEC NICOS OU OUTROS ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI N LIMBA ENGLEZA DAC UN FURNIZOR DE SERVICII PENTRU CLIENTI NECESIT O ALT LIMB DEC T CEA ENGLEZ ESTE DE DATORIA CLIENTULUI S FURNIZEZE O TRADUCERE NU S REPARATI ECHIPAMENTUL DEC T ULTERIOR CONSULT RII SI N ELEGERII ACESTUI MANUAL DE SERVICE IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR HA
90. BILINEAR CUBIC REPLICATE NONE 20100120 2020 0030 Not used Referenced Presentation LUT 2050 0500 Not supported Sequence gt Referenced SOP Class UID 0008 1150 Not supported gt Referenced SOP Instance UID 0008 1155 Not supported Note denotes that the attribute is optional for the SCU However we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value If Failure status is returned during N SET operation of Image Box following action will be taken by Print SCU 0xC605 Resources temporarily not available message will be sent All other status Failure message will be sent In all the cases the print session will be terminated 4 3 1 2 2 4 Color Image Box SOP Class The Print SCU supports the following DIMSE Service Elements for Color Image Box SOP Class N SET Requests the Printer to set the image box attributes The Following Attribute values are supported Image Position 2020 0010 Based on Image Display Format gt Photometric Interpretation 0028 0004 RGB DOC0550678 7101215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Pits Alocateds Sais Stes pamm 00280108 i gt Pixel Data 7FE0 0010 Pixel d
91. C0550678 25 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Section 5 DICOM Storage Commitment SCU is the DICOM Conformance Statement related to DICOM Storage Commitment SCU Application This product uses DICOM Storage Commitment Push Model to store the images using remote DICOM entity which is Storage Commitment SCP Section 6 HIS RIS Modality Worklist The Modality Worklist capability for Discovery allows a user to query for and display DICOM modality work list information Modality Worklist is providing the DICOM C FIND service as a service class user SCU Section 7 Performed Procedure Step The PPS option for Discovery allows a Modality Performed Procedure Step to be communicated to the Hospital Radiology information system The PPS feature is providing the DICOM Modality Performed Procedure Step service as a service class user SCU Section 8 Gray Scale Presentation State The Virtual Exam Split capability for Discovery provides the DICOM Modality GSPS service as a service class user SCU Section 9 CT X ray Radiation Dose Structured Report Object Implementation Appendix A specifies the CT PET MR IOD information object Appendix B specifies the private data element definition for CT PET MR IOD Appendix C specifies the DICOMDIR directory information Appendix D specifies the private data element definition for PET Raw and List Data Appendix E specifies
92. DICOM Conformance Statement REV 11 Direction 5330620 1EN 2 4 1 Supported Communication Stacks part 8 51 2 4 2 Stack iint eat tco e Gase atentos 51 2 4 2 1 Physical Media Support rn nnne sn 52 2 4 3 Additional Protocols o eee ipti ta ERR I SERRA de e ERA REFERRI 52 2 4 4 IPV4 and IPv6 Support dai e e s uh 52 2 5 EXTENSIONS SPECIALIZATIONS PRIVATIZATIONS 0 52 2 5 1 Standard Extended Specialized Private SOP Classes 52 2 5 1 1 Standard Extended Specialized Private SOP 52 2 5 1 2 Private SOP Class GE PET Raw 0000 1 52 2 6 CONFIGURATION 52 2 6 1 AE Title Presentation Address Mapping 2 22 2 100 00 000000040400 010000000000 52 2 6 2 Configurable Parameters ur ua miu annee nn nnns 52 2 7 SUPPORT OF EXTENDED CHARACTER 8 53 28 CODES AND CONTROLLED TERMINOLOGY 53 2 8 1 Mapped Coded Terminolog esee 53 2 8 2 PET Isotope and Patient Orientation Module sse 53 2 9
93. DURCH STROMSCHLAGE MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN TO ZEPBIZ AIATIOETAI 2TA MONO EAN TO ATOMO XEPBIZ ENO2 NEAATH ATIAITEI TO ET XEIPIAIO 2E AQ2Z2A EKTO2 TON ATIOTEAEI EYOYNH TOY NA YNHPEZIE2 MHN ETIIXEIPHZETE THN EKTEAEZH EPI AZION gt EPBI gt 2TON E ONAIZMO EKTO2 EAN ZYMBOYAEYTEI KAI EXETE KATANOH EI TO ET XEIPIAIO gt EPBI gt EAN AE AABETE THN lIPOEIAOTIOIHZH AYTH ENAEXETAI NA ATOMO XEIPIZTH H ZTON AZOENH ANO HAEKTPOTIAHZIA MHXANIKOY2 H AAAOY2 KINAYNOY 2 EZEN KARBANTARTASI KEZIKONYV KIZAROLAG ANGOL NYELVEN ERHETO EL HA A VEVO SZOLGALTATOJA ANGOLTOL ELTERO NYELVRE TART IGENYT AKKOR A VEVO FELELOSSEGE A FORDITAS ELKESZITTETESE NE PROBALJA ELKEZDENI HASZNALNI A BERENDEZEST AMIG A KARBANTARTASI KEZIKONYVBEN LEIRTAKAT NEM ERTELMEZTEK EZEN FIGYELMEZTETES FIGYELMEN K V L HAGYASA A SZOLG LTAT MUKODTETO VAGY A BETEG ARAMUTES MECHANIKAI VAGY EGY B VESZ LYHELYZET MIATTI SERULESET EREDM NYEZHETI 5 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 ADVORUN IS AVVERTENZA IT au KO Direction 5330620 1EN ESSI BJ NUSTUHANDB K ER EING NGU F ANLEG EN
94. Discovery PET LIST Data Exam 14 2 1 00000000000000000000 206 D 2 3 Discovery PET LIST Data Scan Entity 207 D 2 4 Discovery PET LIST Data Equipment Entity 207 D 2 5 Private Creator Identification Information sees 207 D 2 6 Discovery PET Patient 2 207 D 2 7 Discovery PET Exam Module 207 D 2 8 Discovery PET Scan Module enne nenne 208 D 2 9 Discovery PET Frame 208 D 2 10 General Modules uu ae le en annee nona Rame Eram daa 208 0 2 10 1 SOP Common Module 208 D 2 11 Discovery PET List Data 208 D 2 12 Discovery PET Correction Cal Module sss 211 D 2 13 Discovery PET Well Counter eene 211 APPENDIX E IMPLEMENTATION UID FOR PRODUCT 213 0 0530678 21 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN DOC0550678 22 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 1 INTRODUCTION Note Thro
95. E If present in composite object instances it will be set to same value otherwise sent empty Completion Flag 0040 A491 1 1 If present in composite object instance it will be set to same value otherwise not sent Content Date 0008 0023 1 1 If present in composite object instances it will be set to same value otherwise not sent Content Time 0008 0033 1 1 If present in composite object instances it will be set to same value otherwise not sent Concept Name Code 0040 A043 1C 1C If present in composite object instances it Sequence will be set to same value otherwise not sent Verifying Observer 0040 A073 E E If present in composite object instances it Sequence will be set to same value otherwise sent empty Verification Flag 0040 A493 1 1 If present in composite object instances it will be set to same value otherwise not sent Verification Date Time 0040 A030 1C 1C Not used Content Sequence 0040 A730 1C 1C Not used C 5 7 PRIVATE GEMS PET RAW Keys Key Tag Type Type Attribute Description for CD for DVD USB Specific Character Set 0008 0005 1C 1C ISO IR 100 Instance Number 0020 0013 1 1 If present in composite object instance it will be set to same value otherwise computed SOP Instance UID 0008 0018 2 2 If present in composite object instances it will be set to same value otherwise sent empty Frame of Reference UID 0020 0052 2 2 If present in composite object instance it will be set to same value oth
96. ERIAALIA TARVITTAVAN KAANNOKSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA ALA YRITA KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMMARTANYT TAMAN HUOLTO OHJEEN MIKALI TATA VAROITUSTA EI NOUDATETA SEURAUKSENA VOI OLLA HUOLTOHENKILOSTON LAITTEISTON KAYTTAJAN TAI POTILAAN VAHINGOITTUMINEN SAHKOISKUN MEKAANISEN VIAN TAI MUUN VAARATILANTEEN VUOKSI 4 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 ATTENTION FR WARNUNG DE NPOEIAONOIHZH EL FIGYELMEZTETES HU Direction 5330620 1EN CE MANUEL DE SERVICE N EST DISPONIBLE QU EN ANGLAIS SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENIR SUR LES EQUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS ETE CONSULTE ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAINER CHEZ LE TECHNICIEN L OPERATEUR OU LE PATIENT DES BLESSURES DUES A DES DANGERS ELECTRIQUES MECANIQUES OU AUTRES DIESE SERVICEANLEITUNG EXISTIERT NUR IN ENGLISCHER SPRACHE FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENOTIGT IST ES AUFGABE DES KUNDEN FUR EINE ENTSPRECHENDE UBERSETZUNG ZU SORGEN VERSUCHEN SIE NICHT DIESE ANLAGE ZU WARTEN OHNE DIESE SERVICEANLEITUNG GELESEN UND VERSTANDEN ZU HABEN WIRD DIESE WARNUNG NICHT BEACHTET SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS DES BEDIENERS ODER DES PATIENTEN
97. ET acq duration 0009 106D uncompressed size PET where is list frame PET unlisted scan 0009 10 1 PET acq bin dur percent 183 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN REV 7 PET tracerInjection UID 0009 10F2 PET Q Static Frame 0009 10F7 UL PET Sharp IR Flag PET Scatter Limit 0015 103E 184 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN B 2 6 Discovery PET ImageSet Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET compatible version 0009 1079 3 SH PET software version SH PET is datetime 0009 107B PET is source PET is contents PET is type 0009 107E PET is reference 0009 107F PET multi patient 0009 1080 PET number of normals PET color id PET window level type 0009 1083 PET rotate 0009 1084 PET flip 0009 1085 PET pan y 0009 1088 PET window level min 0009 1089 PET window level max PET recon method PET attenuation 0009 108C PET atten coefficient 0009 108D PET bp filter 0009 108E PET bp filter cutoff PET bp filter order PET bp center 0009 1091 PET bp center p 0009 1092 PET atten smooth PET atten smooth param PET angle smooth param PET wellcountercal id 0009 1096 185 of 215 ae ON ek
98. EV 7 Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN Recon parameters Attribute Name Tag Type VR VM Notes Recon Matrix Size 0x0015 0x103C 3 US 1 PET Recon 0x0015 0x101F 3 UL 1 Parameters Exists First Packet Number 0 0015 0 1020 3 SL 1 IR Loop Filter Ratio 0 0015 0 1021 3 FL 1 IR Loop Filter 0x0015 0x1022 3 FL 1 Correction BP3d Filter FlagU 0x0015 0x1023 3 UL 1 BP3d Filter CutoffU 0x0015 0x1024 3 FL 1 BP3d Filter OrderU 0 0015 0 1025 3 SL 1 BP3d Filter FlagV 0x0015 0x1026 3 UL 1 BP3d Filter OrderV 0x0015 0x1027 3 FL 1 BP3d Filter CutoffV 0x0015 0x1028 3 SL 1 Decay Flag 0x0015 0x1029 3 UL 1 Image filter 3d flag 0 0015 0 102 3 UL 1 WCC Cal Flag 0x0015 0x102F 3 UL 1 Emission Scatter 0x0015 0x1030 3 UL 1 Flag Recon Algorithm 0x0015 0x1031 3 UL 1 Normalization Flag 0x0015 0x1032 3 UL 1 Emission Deadtime 0x0015 0x1033 3 UL 1 Flag Emission Random 0x0015 0x1034 3 UL 1 Flag Blank Scan Flag 0x0015 0x1035 3 UL 1 210 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN D 2 12 Discovery PET Correction Cal Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET scan type 0009 1018 PET trame unisiedscan Gone 3 sc 1 promptgamma 3
99. Failure status is returned during N CREATE operation of Film box following action will be taken by Print SCU 0x213 Resource Limitation message will be sent 0x106 Unsupported Film Format message will be sent All other status Failure message will be sent In all the cases the print session will be terminated If Failure status is returned during N ACTION operation of Film box following action will be taken by Print SCU 0xC602 Unable to Create Print Job message will be sent All other status Failure message will be sent In all the cases the print session will be terminated If Failure status is returned during N DELETE operation of Film box following action will be taken by Print SCU All the return status Failure message will be sent and the print session will be terminated 4 3 1 2 2 3 Basic Grayscale Image Box SOP Class The Print SCU supports the following DIMSE Service Elements for Grayscale Image Box SOP Class N SET Requests the Printer to set the image box attributes The Following Attribute values are supported gt Bits Allocated 0028 0100 8 if Bits Stored 8 or 16 if Bits Stored 12 gt Bits Stored 0028 0101 8 or 12 gt High Bit 0028 0102 7 if Bits Stored 8 or 11 if Bits Stored 12 DOC0550678 70 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Magnification Type 2010 0060
100. GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement 5330620 1EN Rev 11 2012 General Electric Company All Rights Reserved GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Page Left Intentionally Blank Page 2 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Important Precautions Language BG ZH CN VYSTRAHA CS ADVARSEL DA Direction 5330620 1EN TOBA E CAMO HA AKO HA WR AEE IRA TEE ESE SEAS ATH o ARE SA 2 BB A AS BE EA ZO
101. KUYUP ANLAMADAN EKIPMANLARA M DAHALE ETMEY N Z BUUYARIYA UYULMAMASI ELEKTR K MEKAN K VEYA D ER TEHL KELERDEN DOLAYI TEKN SYEN OPERAT R VEYA HASTANIN YARALANMASINA YOL A AB L R LEGAL NOTES TRADEMARKS All products and their name brands are trademarks of their respective holders OMISSIONS amp ERRORS Customers please contact your GE Sales or Service representatives GE personnel please use the GE Healthcare PQR Process to report all omissions errors and defects in this publication Copyrights All Material Copyright c 2012 by the General Electric Company All rights reserved 10 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 REVISON HISTORY Direction 5330620 1EN Rev Date Author s Change History 1 May 29 2009 Toan T Le Initial release 2 May 29 2009 Toan T Le Update for KH PPR1 3 May 29 2009 Toan T Le Update per review comments 4 May 29 2009 Toan T Le Update per review comments 5 Aug 17 2009 Joe Muth Update for KH PPR2 Software Release 6 Sep 16 2009 Joe Muth Update for KH PPR2 Software Release 7 Oct 06 2010 Rajasekaran Update for Dayton Release 9 Dec20 2011 C Rajasekaran Update for Optima 560 Release and CSE review comments Joe Muth 10 July 2 2012 C Rajasekaran Update for Eagle Release Section 2 Network Conformance Statement is modified for coreload release RWA Section
102. L FL 55 FL SS SS 55 55 176 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN B 1 10Private Creator Identification GEMS CT CARDIAC 001 Note Private Group 49 is present only if the appropriate cardiac option is installed on the scanner EkgGatingType 0049 1016 EkgWaveTimeOffFirstDataPoint 0049 101B B 1 11Private Creator Identification GEHC CT ADVAPP 001 Note Private tags 0053 1070 through 0053 109D are only present for multi energy CT images Attribute Name ShuttleFlag 0053 1020 TableSpeedNotReachesTargetFlag 0053 1021 IterativeReconAnnotation 0053 1040 177 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN IterativeReconMode 0053 1041 IterativeReconConfiguration 0053 1042 IterativeReconLevel 0053 1043 reconFlipRotateAnno 0053 1060 HiResMode 0053 1061 RespiratoryFlag 0053 1062 B 1 12Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM Notes Gating Type 0x0015 0x101A 3 US 1 Gating Type Cardiac 1 Respiratory 2 No Gating 3 Total Number of Bins 0x0015 0x101B 3 US 1 The number of selected target phases in the binning protocol use in Adv4D application Phase Value 0x0015 0x101C 3 US 1 The target phase value that the images belong to Note Private Group 15 i
103. M Conformance Statement REV 7 Direction 5330620 1EN D 1 12 Discovery PET Correction Cal Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET scan type 0009 1018 PET trame unisiedscan Gone 3 sc 1 promptgamma 3 S _ PETWawedonUD 3 U 1 PET corector eaa Goma 3 U 1 _ PETcompatbe verson 0017102 3 5 1 E sr s E 24 e PET hosp Meme PETacwd gt O O m ea na 1 rU PET norm 2d cal id 0017 100B PET hosp identifier 0017 100C SH PET archived 0017 100D D 1 13 Discovery PET Well Counter Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET scan type 0009 1018 3 SL PET trame unisiedscan 3 sc 1 promptgamma 3 _ PETWawedonUD 00940 3 U 1 PETwscd ooon 3 U 1 gt compatible_version 0019 1002 3 SH 1 PET software_version 0019 1003 C T 204 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN eaae 3 oT 1 3 1 _ PET cal descrpion 3 to 1 _ hardware 3 to 1 2
104. M SERVER Query AE etrieve SCP Choose ping OF 2 Direction 5330620 1EN Image Remotely Stored Remote system requests image storage Query Remote Q N SCP Retrieve remote QR Remot Reques There are five Real World Activities that will cause the DICOM Server Application Entity DICOM Server AE to initiate a DICOM association to a remote DICOM Application Entity Real world Activities for Discovery 710 Discovery 610 and Optima 560 SW Version pet mfk xx 1 The Choose Push Option Real World activity consists of an operator selecting one or more study series or image in the local database manager and choosing either Push Examination Push Series or Push Image from the Network pull down menu on the local database manager to send the image s to a selected destination 2 Real World Activity Query Remote causes the DICOM Server AE to initiate an association to the Remote DICOM AE and request the list of all studies Once the DICOM Server AE receives the list of studies it will select the first study as determined through the local database manager list sort criterion and request the list of series for that study After receiving the list of series the DICOM Server AE will ask for the list of images for the first series in the list The operator can then select any study in the study list to retrieve the list of series and images DOC0550678 52 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM
105. Module descriptions contain the attributes that are present in a C FIND request message sent by the Worklist Server AE to a remote AE It should be noted that they are the same as those defined in the DICOM Standard PS 3 4 Service Class Specifications and include Name Tag group and element numbers Expected Matching Key Type R required O optional Expected Return Key Type 1 non zero value required 1 conditionally of type 1 2 required to be present possibly with zero length value 3 optional Mapped into The Image whether this data is mapped into subsequently acquired images Notes clarification of this implementation s use treatment of this attribute All data elements in the following Module descriptions are requested by the Worklist Server AE Values of data elements that are not mapped into images and are not otherwise dealt with displayed on the user interface etc are not used and are thus discarded upon receipt See Table B 1 for further information DOC0550678 95 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Data elements for which values can be sent for matching purposes are described as such Data elements for which values are not sent are sent with zero length and universal matching will apply This is the default case if no other description to the contrary is provided 6 8 8 Supported Matching The following are th
106. O A sinn 150 A 4 4 CT Image Module 5 rtr ET ee Ae AT 150 A 4 5 VOLU TMOU is A E etn db m ASTA 154 A 4 6 SOP Gommon Module do 154 AS SCIIMAGE MODUEES LI uu u a A IEEE ARE Te UR Eee etus ap ON A ee pae rimi se daria 154 A 5 1 SC Equipment Module sss 154 A 5 2 Image Pixel Modu le 52 eti ctp et Rec cate eee d ated 155 A 5 3 VOLLUT Mod le npa e 155 A 5 4 SOP Common Module 155 A 6 OTHER ATTRIBUTES eH odd t coe DRE RE ERR re asus aaa 155 PET SPECIFIC IOD COMMON MODULE 8 ener 156 A 7 1 PET Image IOD 1 2 70 00 0000000000000 eene rennen nnne nennen nne 156 A 7 2 ala Er LED SE EFL Ru 156 A 7 3 General Study Module isis iiie eiie enini thanh sinit 157 A 7 4 Patient Study Modula uyu yuna 158 A 7 5 Discovery PET Exam Module Rant 158 A 7 6 General Series Module sisse siiis eiie nnne 158 A 7 7 PET Sones Module pa tine i ree i nad t y e 159 A 7 8 PET Isotope Module u uu due i HER HER apaspan tested 161 A 7 9 PET Multi gated Acquisition Module sene
107. R VM Notes Manufacturer Institution Name Institution Address Station Name 0008 0070 2 LO 1 Institutional Department Name 166 of 215 GE Medical System Hospital name Not Used Not Used Not Used GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Manufacturer s Model Name 0008 1090 BERI Scanner Description Device Serial Number 0018 1000 0018 1020 Software Versions Image software version Frame software version Spatial Resolution 0018 1050 3 Not Used Date of Last Calibration 0018 1200 3 Not Used Time of Last Calibration 0018 1201 3 Not Used Pixel Padding Value 0028 0120 3 05 1 Not Used Or SS A 7 15 General Image Module This section specifies the Attributes which identify and describe an image within a particular series Attribute Name Tag Type VR VM Notes Image Number 0020 0013 2 IS 1 Patient Orientation 0020 0020 Not Used Image Date 0008 0023 2C DA 1 Extract date from image date time Image Time 0008 0033 Extract time from image date time Image Type 0008 0008 3 5 1 Acquisition Number 0020 0012 Not Used Acquisition Date 0008 0022 DA Acquisition Date Acquisition Time 0008 0032 TM 1 Acquisition Time Referenced Image Sequence 0008 1140 SQ 1 Not Used gt Referenced SOP Class UID 0008 1150 1C 1 Not Used gt
108. REN OPERATOREN ELLER PASIENTEN SKADES PA GRUNN AV ELEKTRISK ST T MEKANISKE ELLER ANDRE FARER NINIEJSZY PODRECZNIK SERWISOWY DOSTEPNY JEST JEDYNIE W JEZYKU ANGIELSKIM JESLI DOSTAWCA USEUG KLIENTA WYMAGA JEZYKA INNEGO NIZ ANGIELSKI ZAPEWNIENIE US UGI T UMACZENIA JEST OBOWI ZKIEM KLIENTA NIE PR BOWA SERWISOWA WYPOSA ENIA BEZ ZAPOZNANIA SI ZROZUMIENIA NINIEJSZEGO PODR CZNIKA SERWISOWEGO NIEZASTOSOWANIE SI DO TEGO OSTRZE ENIA MO E SPOWODOWA URAZY DOSTAWCY US UG OPERATORA LUB PACJENTA W WYNIKU PORA ENIA ELEKTRYCZNEGO ZAGRO ENIA MECHANICZNEGO B D INNEGO 7 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 PT BR ATENC O PT PT ATENTIE RO RU Direction 5330620 1EN ESTE MANUAL DE ASSISTENCIA TECNICA SO SE ENCONTRA DISPONIVEL EM INGLES SE QUALQUER OUTRO SERVICO DE ASSISTENCIA TECNICA QUE NAO A GEHC SOLICITAR ESTES MANUAIS NOUTRO IDIOMA E DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVICOS DE TRADUCAO NAO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTENCIA TECNICA O NAO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURAN A DO T CNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES EL TRICOS MECANICOS OU OUTROS ESTE MANUAL DE ASSISTENCIA TECNICA SO SE ENCONTRA DISPONIVEL EM INGLES SE QUALQUER OUTRO SERVICO DE ASSISTENCIA TECNICA QUE NAO A GEHC SOLICIT
109. Referenced SOP Instance 0008 1155 Not Used UID Timezone Offset From UTC 0x0008 0x02 LO Contains the offset from UTC 01 to the time zone for all DA and TM Attributes present in this SOP Instance Encoded as an ASCII string in the format amp ZZZ The components of the string from left to right are amp or and ZZZZ Hours and 167 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN v l al L Minutes of offsets Source Image Sequence 0008 2112 sa 1 Not Used Referenced SOP Class UID 0008 1150 1 1 Not Used gt Referenced SOP Instance 0008 1155 Not Used UID Images in Acquisition 0020 1002 3 15 1 Not Used Image Comments 0020 4000 ERI Not Used Lossy Image Compression 0028 2110 CS A 7 16 Image Plane Module This section specifies the Attributes which define the transmitted pixel array of a two dimensional image plane Attribute Name Tag Type VR VM Notes Pixel Spacing 0028 0030 1 DS 2 Value 1 pixel width Value 2 pixel height Image Orientation Patient 0020 0037 Patient Row inferior Patient Row Posterior Patient Row Superior Patient Column inferior Patient Column Posterior Patient Column Superior Image Position Patient 0020 0032 Patient inferior Patient Posterior Patient Superior Slice Thickness 0018 0050 Slic
110. S 2 ce SIA SS AHE HE 6 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 BRIDINAJUMS LV ISP JIMAS LT ADVARSEL NO OSTRZEZENIE PL Direction 5330620 1EN ST APKALPES ROKASGRAMATA IR PIEEJAMA TIKAI ANGLU VALODA JA KLIENTA APKALPES SNIEDZEJAM NEPIECIESAMA INFORM CIJA CITA VALODA NEVIS ANGLU KLIENTA PIENAKUMS IR NODROSINAT TULKOSANU NEVEICIET APRIKOJUMA APKALPI BEZ APKALPES ROKASGRAMATAS IZLASISANAS UN SAPRASANAS SI BRIDINAJUMA NEIEVEROSANA VAR RADIT ELEKTRISKAS STRAVAS TRIECIENA MEHANISKU VAI CITU RISKU IZRAISITU TRAUMU APKALPES SNIEDZEJAM OPERATORAM VAI PACIENTAM SIS EKSPLOATAVIMO VADOVAS YRA PRIEINAMAS TIK ANGLU KALBA JEI KLIENTO PASLAUGU TIEKEJAS REIKALAUJA VADOVO KITA KALBA NE ANGLU NUMATYTI VERTIMO PASLAUGAS YRA KLIENTO ATSAKOMYBE NEM GINKITE ATLIKTI JRANGOS TECHNIN S PRIEZI ROS NEBENT ATSIZVELGETE EKSPLOATAVIMO VADOVA IR J SUPRATOTE JEI NEATKREIPSITE D MESIO 5 PERSP JIMA GALIMI SUZALOJIMAI DEL ELEKTROS SOKO MECHANINIU AR KITU PAVOJU PASLAUGU TIEKEJUI OPERATORIUI AR PACIENTUI DENNE SERVICEHANDBOKEN FINNES BARE PA ENGELSK HVIS KUNDENS SERVICELEVERANDOR TRENGER ET ANNET SPRAK ER DET KUNDENS ANSVAR A SORGE FOR OVERSETTELSE IKKE FORS K A REPARERE UTSTYRET UTEN AT DENNE SERVICEHANDBOKEN ER LEST OG FORSTATT MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FORE TIL AT SERVICELEVERANDO
111. SECURITY PROFIEES s uit IM DE 53 3 MEDIA STORAGE CONFORMANCE STATEMENT 444422222 55 3 1 INTRODUCTION idit deterret teat edite duree set dei petentes 55 3 2 IMPLEMENTATION MODEL MEDIA SERVER 6 55 3 2 1 Application Data Flow 55 3 2 2 Functional Definitions 5 57 3 2 3 Sequencing of Real World Activities 57 3 2 4 File Meta Information for Implementation Class and Version 57 9 3 AE SPEGIFICATIONS 222 iore b Fla tl ole antl 58 3 3 1 DICOM Storage Media CD DVD USB Storage Interchange AE Specification 58 3 3 1 1 File Meta Information for Implementation Class and Version 58 3 3 1 2 Real World Activities for the DICOM Media Interchange 58 3 3 1 2 1 Real World Activity RWA Create Generate and Write CD DVD USB 58 3 3 1 2 1 1 Application Profiles for the RWA Create CD DVD USB 59 3 3 1 2 2 Real World Activity 60 3 3 1 2 2 1 Application Profiles for the RWA Display
112. SKU EF AD PJONUSTUVEITANDI VIDSKIPTAMANNS ARFNAST ANNAS TUNGUMALS EN ENSKU ER PAD SKYLDA VIDSKIPTAMANNS SKAFFA TUNGUMALAPJONUSTU REYNID EKKI AD AFGREIDA TAKID NEMA PESSI PJONUSTUHANDBOK HEFUR VERI SKO U OG SKILIN BROT SINNA ESSARI A V RUN GETUR LEITT TIL MEI SLA J NUSTUVEITANDA STJ RNANDA EDA SJUKLINGS FR RAFLOSTI VELRAENU EDA RUM H ETTUM IL PRESENTE MANUALE DI MANUTENZIONE E DISPONIBILE SOLTANTO IN INGLESE SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA IL CLIENTE E TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE SI PROCEDA ALLA MANUTENZIONE DELL APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO IL NON RISPETTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL ADDETTO ALLA MANUTENZIONE ALL UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA PER URTI MECCANICI OD ALTRI RISCHI ZOY PAV 1 0529 0 ERA PEA amp FAY ND WEGE LE CMT bO Z lt AG IC HOUR ARTDRVY CR TE SV Oz C Eqo U JE v vAEBMXIASE db DV LIB SAD MILE 00488 00 JES IC 2 E AHJA ZIRE HAE 80 9 29i Sona 80 019181 Clots X SER o nol 0181 AA IA
113. Statement REV 11 Direction 5330620 1EN The DICOM Media Interchange AE creates one File Set per generated Interchange Media user can only create one copy of CD DVD image for a drive at a time any other attempt of creation will wait until the first one is complete Auser cannot create CD DVD USB while restore CD DVD USB is in process o ADICOM Media Viewer is provided along with the selected object instances on the interchange media This viewer can be loaded on a standard PC running Windows XP Windows Vista or Windows 7 Before writing on the Interchange Medium the DICOM Media Interchange AE checks for the following condition The inserted media is blank and write able If the condition is not met an error is displayed and the user needs to replace it with a blank media 3 3 1 2 1 1 Application Profiles for the RWA Create CD DVD USB For the AE Conforms to the Application Profile see the Table 3 3 1 1 in section 3 3 for Create CD DVD USB 3 3 1 2 1 1 1 Options for STD GEN CD STD GEN DVD JPEG and STD GEN USB JPEG Application Profile Following are the SOP Classes supported by the RWA Create CD DVD USB All SOP Instances use the Explicit VR Little Endian Uncompressed Transfer Syntax UID 1 2 840 10008 1 2 1 for creating CD DVD USB TABLE 3 3 1 2 1 SOP Class SOP Class UID MR Image Storage 1 2 840 10008 5 1 4 1 1 4 Positron Emission Tomography 1 2 840 10008 5 1 4 1 1 128 Image Storage Nuclear
114. Statement REV 7 Direction 5330620 1EN 8 9 Use of Specific DICOM Data The following table gives specific usage of some of attributes in the GSPS and splitting PPS SOP instance created with reference to each of the real world scenarios mentioned in 8 2 3 of this document Attribute Name Tag GSPS Splitting PPS Comment Group Length 0002 0000 X Exam Split Application will calculate File Meta Information 0002 0001 X Exam Split Version Application will calculate Media Storage SOP Class 0002 0002 X Exam Split UID Application will calculate Media Storage SOP 0002 0003 X Exam Split Instance UID Application will calculate Transfer Syntax UID 0002 0010 X X Exam Split Application will calculate Implementation Class UID 0002 0012 X Exam Split Application will calculate Implementation Version 0002 0013 X Exam Split Name Application will calculate Patient s Name 0010 0010 X X Copied from acquired image s Patient ID 0010 0020 X X Copied from acquired image s Patient s Birth Date 0010 0030 X X Copied from acquired image s Patient s Sex 0010 0040 X X Copied from acquired image s Instance Creation Date 0008 0012 X Copied from acquired image s Instance Creation Time 0008 0013 X Copied from acquired image s Study Instance UID 0020 000D X Copied from acquired image s Study Date 0008 0020 X Copied from acquired image s Study
115. Storage no size limit blank Interchange Media It initializes the DICOM File set and writes all the specified SOP instances onto the Interchange Media at once The SOP instances written will be limited to instances that match the criteria of one of the Application Profiles that is supported Updating the media is not supported The DICOM Media Interchange AE acts as an FSR when requested to browse the Interchange Media such that user can select the SOP instances that he wants the DICOM Media Interchange AE to copy on the local database by selecting appropriate Study Series Image instances The supported file system during creation and restore are listed below File System Supported Supported during during Media Media Restore Create FSC FSR ISO 9660 YES YES UDF NO YES VFAT YES YES The supported media during creation and restore are listed below Media Supported File System Supported during File System during Media supported Media Restore supported Create FSC FSC FSR FSR CD R YES ISO 9660 YES ISO 9660 CD RW YES ISO 9660 YES ISO 9660 DVD R YES ISO 9660 YES ISO 9660 amp UDF DVD RW YES ISO 9660 YES ISO 9660 amp DOC0530678 56 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN UDF DVD R NO ISO 9660 YES ISO 9660 amp UDF DVD RW NO ISO 9660 YES ISO 9660 a
116. TAINS NUM EV 113833 1 M DCM Maximum X ray Tube Current 140 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 19 gt gt gt CONTAINS NUM EV 113734 DCM Mean X ray Tube Current Sent 20 gt gt gt CONTAINS NUM EV 113834 DCM Exposure Time per Rotation 1 MC Sent per conditional 21 CONTAINS CONT AINE EV 113829 DCM CT Dose 1 MC Sent per conditional 22 gt gt CONTAINS NUM EV 113830 DCM Mean CTDIvol Sent 23 gt gt CONTAINS CODE EV 113835 DCM CTDIw Phantom Type Sent IEC Head Dosimetry Phantom IEC Body Dosimetry Phantom 24 gt gt CONTAINS NUM EV 113836 DCM CTDlfreeair Calculation Factor Not sent 25 gt gt CONTAINS NUM EV 113837 DCM Mean CTDlfreeair Not sent 26 gt gt CONTAINS NUM EV 113838 DCM DLP Sent 27 gt gt CONTAINS NUM EV 113839 DCM Effective Dose Not sent 28 gt gt gt HAS CONCEPT MOD CODE EV G C036 SRT Measurement Method 1 MC Not sent 29 gt gt gt HAS PROPERTI ES NUM EV 113840 DCM Effective Dose Conversion Factor 1 MC Not sent 30 CONTAINS TEXT EV 121106
117. Time 0008 0030 X Copied from acquired image s Accession Number 0008 0050 X Copied from acquired image s 129 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Referring Physician s Name 0008 0090 Copied from acquired image s Study ID 0020 0010 Copied from acquired image s Referenced Study Sequence 0008 1110 Copied from acquired image s Modality 0008 0060 Copied from acquired image s Study Description 0008 1030 This will be set as Requested Procedure Code Meaning If Code Meaning absent it will be Requested Procedure Description Procedure Code Sequence 0008 1032 Copied from acquired image s Protocol Name 0018 1030 Set as Virtual Exam Split Series Description 0008 103E This will be set as Requested Procedure Code Meaning If Code Meaning absent it will be Requested Procedure Description Series Instance UID 0020 000E Exam Split Application will calculate Series Number 0020 0011 Copied from acquired image s Laterality 0020 0060 Set as 0 zero Requested Attribute Sequence 0040 0275 Copied from source image Originally from Request Attributes Sequence gt Accession Number 0008 0050 gt Referenced Study Sequence 0008 1110 gt Study Instance
118. UID 0020 000D gt Requested Procedure Description 0032 1060 gt Requested Procedure Code Sequence 0032 1064 gt Scheduled Procedure Step Description 0040 0007 gt Scheduled Protocol Code Sequence 0040 0008 130 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN gt Scheduled Procedure ID 0040 0009 gt Requested Procedure ID 0040 1001 Performed Station AE Title 0040 0241 will set as Host Exam Application name Performed Station Name 0040 0242 will set as Host Exam Application name Performed Location 0040 0243 will set as Host Exam Application name Performed Procedure Step 0040 0244 X Exam Split Start Date Application will calculate and same time will use in GSPS Performed Procedure Step 0040 0245 X Exam Split Start Time Application will calculate and same time will use in GSPS Performed Procedure Step 0040 0250 Exam Split End Date Application will calculate Performed Procedure Step 0040 0251 Exam Split End Time Application will calculate Performed Procedure Step 0040 0252 Set as In Status Progress Performed Procedure Step 0040 0253 X Will be set as ID Requested Procedure ID 0040 1001 from 0040 0275 Performed Procedure Step 0040 0254 X This will be set a
119. Weight in kg 0010 2000 Medical Alerts 0010 2110 Contrast Allergies 0010 21 0 Pregnancy Status 0032 1032 Requesting Physician 0032 1070 Requested Contrast Agent 0038 0010 Admission ID 0038 0050 Special Needs 0038 0300 Current Patient Location 0038 0500 Patient State 0040 0006 Performing Physician 0040 0010 Scheduled Station Name 0040 0011 Scheduled Procedure Step Location 0040 0012 Pre order Medication 0040 1003 Requested Procedure Priority 0040 1004 Patient Transport Arrangements 0040 3001 Confidentiality Constraint e e e e e e e e e e e e e e e e e e e e e e 6 3 1 3 Association Acceptance Policy The Worklist Server AE does not respond to attempts by a remote AE to open an association 6 4 Communication Profiles 6 4 4 Supported Communication Stacks PS 3 8 PS 3 9 DICOM Upper Layer PS 3 8 is supported using TCP IP 6 4 2 OSI Stack The OSI Communication Stack is not supported by this implementation DOC0550678 90 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 4 3 TCP IP Stack The TCP IP Communication Stack is inherited from the operating system 6 4 3 1 API Not applicable to this product 6 4 3 2 Physical Media Support Ethernet 802 3 provides the physical network layer for this product 6 4 4 Point to Point
120. Worklist Server AE will save worklist entry in binary format for investigative purposes and the Worklist Server AE will continue receiving worklist entries Note All worklist entries valid and invalid can be configured to be saved in binary format On receipt of any error from the remote AE the Worklist Server will issue a C FIND CANCEL and upon receipt of a C FIND RSP or if an applicable timer expires will abort the association All previously received worklist items are retained Warnings received from the remote AE are ignored Each C FIND operation supports a configurable Association Timer This timer starts when the association request is sent or received and stops when the association is established The default time out value is 30 seconds Each C FIND operation supports a configurable Session Timer This timer starts when an association is established and stops when the association is ended The default time out value is 3600 seconds If any of the above timers expires the association is aborted A ABORT and the operation in progress is considered to be failed Any previously received worklist items are discarded 6 3 1 2 1 2 2 Record Acceptance Policy The Discovery implementation adheres to strict value checking of incoming query responses from the remote AE Each response received is examined to verify that all Type 1 attributes are present with non zero length that all Type 2 attributes are present possibly with zero
121. ZER SEGMENTED REFORMATTED PROCESSED COMBINED CTINTERVENTION Value 4 MIN IP MIP AVERAGE VOLREN INTEGRAL HD MIP GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Attribute Name Tag Type jNotes RAYSUM SURFACE MINMAX FLUORO GSI MONO GSI HIGH KV GSI LOW KV GSI MD GSI QC GSI EFF Z GSI CLR OVRLY DIGITALTILT Samples per Pixel 0028 0002 Always sent with value 1 Photometric Interpretation 0028 0004 Always sent Bits Allocated 0028 0100 Always sent with value 16 Bits Stored 0028 0101 Always sent with value 16 High Bit 0028 0102 Always sent with value 15 Reconstruction Diameter 0018 1100 3 Value sent for all images except scouts Exposure 0018 1152 3 Value always sent See Note below Filter Type 0018 1160 3 Sent Defined terms BODY FILTER MEDIUM FILTER 151 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Attribute Name Tag Type Notes 0 0 HEAD FICTER Generator Power 0018 1170 Always sent Focal Spot 0018 1190 Sent Fixed value of 0 7 or 1 2 Convolution Kernel 0018 1210 Not sent for scouts Defined terms SMOOTH SOFT STANDARD STD DETAIL BONE BONEPLUS CHST EDGE SHARP LUNG HD LUNG STANDARD2 DETAIL2 BONE2 BONEPLUS2 EDGE2 STANDARDPLUS2 DETAILPLUS2 HD ULTA HD SOFT PET AC Revolution Time 0018 9305 Sent for
122. aining the PET List Series will skip the same during Save by Exam The Discovery 710 610 uses the GE PET Raw SOP Class for local storage and management of list data Unlike other types of GE PET Raw such as sinogram and spectra data list data is not supported in archive operations such as save and restore and the list data objects will not be visible via Archive browsers List data objects are labeled as List on the database browser and can be identified by tag 0008 0060 see Section D 2 o Save by Series on PET List series not supported o Save by Image on PET List frame not supported PET List Data behavior on PET Applications On saving exams contains PET List Series generated from Anonymous Patient data or Edit Patient Data or Edit Tracer application the exam level archive flag set to Yes after skipping the PET List Series during save Note The Optima 560 product does not support PET List Data 3 3 1 2 2 Real World Activity QUERY The DICOM Media Interchange AE acts as an FSR using the interchange option when requested to browse the Interchange Media When Media Application is requested to provide a directory listing it will read the File set and display the DICOMDIR directory entries for those SOP instances in the File set that correspond to the user selected Application Profile If the media is not blank then Interchange Media gets mounted To remove the media the user has to select the eject button on user interface for t
123. an Date Time plus Image Time Actual Frame Duration 0018 1242 1C IS 1 Image duration Converted from seconds in DB to milliseconds Nominal Interval 0018 1062 IS 1 Not Used Intervals Acquired 0018 1083 Sake Triggers acquired 170 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Triggers rejected Intervals Rejected 0018 1084 IS 1 Primary Counts Prompts 0054 1310 IS 1 Accumulated Total prompts Secondary Counts 0054 1311 IS 1 n Accumulated 3 3 3 Slice Sensitivity Factor 0054 1320 1 Dose Calibration Factor 0054 1322 Scatter Fraction Factor 0054 1323 Dead Time Factor 0054 1324 Decay Factor 0054 1321 S 1 Value 1 total delays Value equals coefficient if coefficient is not NULL Otherwise value equals 1 0 Decay factor Value equals Activity factor times 1 0e 06 Value is converted MBq ml to Bq ml Value equals Scatter subtracted divided by total counts Deadtime factor Referenced Overlay Sequence Not Used Referenced SOP Class UID Not Used gt Referenced SOP Instance 0008 1155 1 Ul 1 Not Used UID Referenced Curve Sequence 0008 1145 3 50 1 Not Used gt Referenced SOP Class UID 0008 1150 1 Ul 1 Not Used gt Referenced SOP Instance Not Used UID Anatomic Region Sequence Not Used gt Code Value Not Used gt Coding Schem
124. ansferred as part of the C FIND request message for a typical query In this particular dump no values are specified for the Scheduled Procedure Step Start and End Dates the attributes are sent with zero length In DICOM this is interpreted as meaning all dates i e universal matching The Modality is also not specified in this particular dump meaning all modalities Note that the user through the use of the Discovery user interface can submit a worklist query that will cause non zero values to be sent for these attributes Table 6 10 1 C FIND Request Message Dump 0008 0000 UL 108 4 1 IdentifyingGroupLength 0008 0005 CS ISO IR 100 12 1 SpecificCharacterSet 0008 0050 SH no value available 05 0 AccessionNumber 0008 0080 LO no value available 0 0 InstitutionName 0008 0090 PN no value available 0 0 ReferringPhysicianName 0008 1110 SQ Sequence with explicit Length 4 1 24 1 ReferencedStudySequence fffe e000 na Item with explicit Length 4 2 16 1 Item 0008 1150 UI no value available 0 0 ReferencedSOPClassUID 0008 1155 UI no value available 0 0 ReferencedSOPInstanceUID fffe e00d na ItemDelimitationItem for re encoding f 0 1 ItemDelimitationItem fffe e0dd na SequenceDelimitationItem for re enc 0 1 SequenceDelimitationItem 00008 1120 SQ Sequence with explicit Length 4 1 4 24 1 ReferencedPatientSequence fffe e000 na Item with explici
125. apping of DICOM Entities Table 6 8 5 1 Mapping of DICOM Entities to Modality Worklist Entities DICOM Modality Worklist Entity Scheduled Procedure Step 6 8 6 Information Model Module Table Within an entity of the DICOM Modality Worklist Information Model attributes are grouped together into related set of attributes called modules A module facilitates the understanding of the semantics concerning the attributes and how the attributes relate to one another A module grouping does not infer any encoding of information into datasets Table 6 8 6 1 identifies the defined modules within the entities that comprise the DICOM Modality Worklist Information Model Modules are identified by Module Name See DICOM PS 3 3 and PS 3 4 for a complete definition of the entities modules and attributes DOC0550678 94 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Table 6 8 6 1 Modality Worklist Information Model Modules 222 o SewduedPmcedueSmp 6892 Pid Stas 8122 VstReoonp 68423 68124 Paenemfcon 68432 gt J Jatin Demographic 68433 27777 68134 6 8 7 Information Model Keys Please refer to DICOM Standard PS 3 3 Information Object Definitions and PS 3 4 Service Class Specifications for a description of each of the Entities contained within the Modality Worklist Information Model The following
126. as chosen Defer PPS the user is provided with an interface to Complete PPS or Discontinue PPS at any later time The user might wish to alter the image set generated through acquisition before invoking these operations Note that the user explicitly uses the user interface provided to invoke this operation as in the case of PPS generated for post processing which is described in the following section PPS messages N CREATE if applicable and N SET will be sent over the same DICOM association The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user 7 2 3 2 PPS from acquisition system without MWL data The system either does not have a Modality Work list Server AE installed or a Modality Work list Server AE installed but no Work List information is obtained from HIS RIS system for the current procedure that is being performed The information required for performing the procedure is supplied through the user interface of the system The use of this information in MPPS SOP instances is described later in this document The system initiates a Start PPS before starting a scan i e when the image acquisition is started PPS Server AE initiates a MPPS Modality Performed Procedure Step N CREATE request to the remote AE MPPS SCP in order to create a MPPS SOP instance at the remote AE The MPPS SCP
127. as the following meaning Required U Unique Optional P Private Only keys with Usage type Matched will be matched against values in the database Values in keys of type Returned will be ignored and will be filled in with data from the database If an optional key is requested that does not appear in any of the tables above that key will be ignored and no corresponding element will be returned If the database does not have a value corresponding to any requested optional key a zero length element will be returned Sequence matching is not supported Range matching is supported for attributes of type date and time Only hierarchical query is supported Special character can be used to match any single character and special character can be used to match any character or set of characters for 0008 0050 Accession Number 0010 0010 Patient s Name 0010 0020 Patient ID and 0020 0010 Study ID 2 3 1 3 4 Retrieve Request From Remote AE This AE is indefinitely listening for associations No operator action is required to respond to a retrieve request 2 3 1 3 4 1 Associated Real World Activity The Real World Activity associated with the Retrieve Request is to send all images corresponding to the C MOVE request to the destination AE through a separate association 2 3 1 3 4 2 Presentation Context Table Table 2 3 1 3 4 2 1 Acceptable Presentation Contexts for DICOM Server AE and Real World Activity Retrieve
128. ata Polarity 2020 0020 NORMAL Magnification Type 2010 0060 BILINEAR CUBIC REPLICATE NONE Smoothing Type 2010 0080 Printer Dependent Requested Decimate Crop Behavior 2020 0040 Note denotes that the attribute is optional for the SCU However we do provide values for all of these optional attributes and if the SCP does not support the requested value it may choose to either return a failure status or ignore the value provided and use its default value If a Normal status is returned during the N SET operation of Image Box a Film Session N Create request is sent If a Failure status is returned during the N SET operation of Image Box the association will be terminated If a Warning status is returned during the N SET operation of the Image Box the association will be terminated The status values which are specific for this SOP Class are defined as follows Status Meaning Error Code Warning Image size larger than image box B604 size the image has been demagnified Image size is larger than the Image B609 Box size The Image has been cropped to fit Image size or Combined Print Image B60A size is larger than the Image Box size The Image or combined Print Image has been decimated to fit Failure Image size is larger than image box C603 size DOC0550678 72 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction
129. ation GEMS ACQU 01 Tag 0019 1002 0019 1003 1 1 1 1 1 1 Table speed 1 Mid scan time 0019 1024 DS 1 Mid scan flag 0019 1025 SS 1 1 1 1 1 1 1 1 1 1 1 1 174 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN B 1 3 Private Creator Identification GEMS RELA 01 Attribute Name Series from which Prescribed 0021 1003 Series Prescribed From 0021 1035 Image Prescribed From 0021 1036 Biopsy position 0021 1091 Biopsy T location 0021 1092 Biopsy ref location 0021 1093 B 1 4 Private Creator Identification GEMS STDY 01 Attribute Name Start time secs in first axial 0023 1070 B 1 5 Private Creator Identification GEMS IMAG 01 Attribute Name Scout Type 0027 1010 0027 101C 0027 101E 0027 101F Smart scan ON OFF flag 0027 1020 Plane Type 0027 1035 Center R coord of plane image 0027 1042 Center A coord of plane image 0027 1043 Center S coord of plane image 0027 1044 Normal R coord 0027 1045 Normal A coord 0027 1046 Normal S coord 0027 1047 Table start location 0027 1050 Table end location 0027 1051 B 1 6 Private Creator Identification GEMS 0039 Attribute Name SR Application Name 0039 1095 B 1 7 Private Creator Identification GEMS 0039 Attribute Name SR Application Name 0039 1095 B 1 8 Private Creator Identification GEMS_PARM_01 Attribute Name 0043 1010 175 of 215 GE Healthcare Disc
130. ationship 100 6 8 12 4 Visit Admission necem t oa patet ted quite ie 100 DOC0550678 17 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 8 13 Patient Entity min EM ae ea ALE eA ei si 100 6 8 13 1 Patient Relationship aaa 100 6 8 13 2 5 Wate ade a ss hu ss 100 6 8 13 3 Patient Demographic r 101 6 8 13 4 Patient Medical u rennen eoe i te nene Fo e ee de Seni ees 101 6 9 PRIVATE DATA 2 555 1 2 IH Hong I e docks Ce toc ot 101 6 10 C FIND REQUEST 55 102 6 11 USE OF SPECIFIC DICOM 105 6 12 SETTING USER PREFERENCES 106 6 12 1 Setting Use MWL Study UID Option essen 106 6 12 2 Setting Custom Query Option terne eeen nne 106 7 PERFORMED PROCEDURE STEP CONFORMANCE STATEMENT 107 7 1 INTRODUCTION au 2 terreat rii sies gli eI been 107 7 2 IMPLEMENTATION MODEL 2 ccc2scececescececsscecaeencetsencesuseeceanecessnencecssceceaencesseaceesaeecsaeseees 107 7 2 1 Application Data Flow Diagram seien nen
131. by this SOP class support an association timer which is configurable The timer is started when a request association request N CREATE request or N SET request is send and stopped when the respective response is received The default time out value is 300 seconds All the operations used by this SOP class support a Session Timer This timer is started when the association is established and stopped when association is ended The default time out value is 3000 seconds If any of the above timers expires the association is aborted and the operation in progress is considered FAILED In any case an operation N CREATE or N SET fails system updates the state to enable operator to manually invoke the operation at any later time 7 3 1 3 Association Acceptance Policy The PPS Server AE does not respond to attempts by remote AE to open an association 7 4 COMMUNICATION PROFILES 7 4 4 Supported Communication Stacks PS 3 8 DICOM Upper Layer PS 3 8 is supported using TCP IP 7 4 2 OSI Stack The OSI Communication stack is not supported by this implementation 7 4 3 TCP IP Stack The TCP IP stack is inherited from the operating system 112 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 7 4 3 1 API Not Applicable to this product 7 4 3 2 Physical Media Support Ethernet 802 3 provides the physical network layer for this product 7 4 4 Point to Point Stack
132. c Recon 0015 103F Privates added for Discovery 710 610 and Optima 560 Attribute Name Tag Type VR VM Phase Matched Series 0015 101D 3 SL 1 CTAC Percent Value 0015 101E 3 SL 1 Recon Matrix Size 0015 103C 3 US 1 PET Recon Parameters Exists 0015 101F 3 UL 1 First Packet Number 0015 1020 3 SL 1 IR Loop Filter Ratio 0015 1021 3 FL 1 IR Loop Filter Correction 0015 1022 3 FL 1 BP3d Filter FlagU 0015 1023 3 UL 1 BP3d Filter CutoffU 0015 1024 3 FL 1 BP3d Filter OrderU 0015 1025 3 SL 1 BP3d Filter FlagV 0015 1026 3 UL 1 BP3d Filter OrderV 0015 1027 3 FL 1 BP3d Filter CutoffV 0015 1028 3 SL 1 Decay Flag 0015 1029 3 UL 1 Image filter 3d flag 0015 102E 3 UL 1 WCC Cal Flag 0015 102F 3 UL 1 Emission Scatter Flag 0015 1030 3 UL 1 Recon Algorithm 0015 1031 3 UL 1 Normalization Flag 0015 1032 3 UL 1 Emission Deadtime Flag 0015 1033 3 UL 1 187 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Emission Random Flag 0015 1034 3 UL 1 Blank Scan Flag 0015 1035 3 UL 1 PET sharp IR Flag 0015 103D 3 UL 1 PET scatter limit 0015 103E 3 UL 1 188 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN B 2 7 Discovery PET Image Module Private Creato
133. ce gt gt Include Code Sequence Macro Table 8 8 No Baseline Context ID Number is specified 1 gt Reason for the Requested 0040 1002 3 Not Sent Procedure gt Reason for Requested 0040 100A 3 Not Sent Procedure Code Sequence 136 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN gt gt Include Code Sequence Macro Table 8 8 No Baseline Context ID Number is specified Performed Procedure Code 0040 A372 2 Sent empty Sequence gt Include Code Sequence Macro Table 8 8 1 No Baseline Context ID Number is specified Current Requested 0040 A375 1C Not Sent Procedure Evidence Sequence Include Hierarchical SOP Instance Reference Macro Table C 17 3 Pertinent Other Evidence 0040 A385 1C Not sent Sequence gt Include Hierarchical SOP Instance Reference Macro Table C 17 3 Referenced Instance 0008 114A 1C Not sent Sequence gt Include SOP Instance Reference Macro Table 10 11 gt Purpose of Reference 0040 A170 Not sent Code Sequence gt gt Include Code Sequence Macro Table 8 8 1 9 4 SR Document Content Module The tables below capture notes concerning the values stored 9 4 1 TID 10011 CT Radiation Dose NL Relwith VT Concept Name VM Notes Parent Type 1 CONT EV 113701 1 M AINE DOM X ray R Radiation Dose Report 2 gt HAS CODE EV
134. ce UID 0020 000D 1 Value from worklist if present is used Otherwise the scanner creates a unique value for each exam If worklist is reused system can be configured to reuse the study instance uid Default behavior is to generate a new study instance uid if the worklist is reused Study Date 0008 0020 2 Generated for each exam and always sent Study Time 0008 0030 Generated for each exam and always sent 0008 0050 Value from user interface or worklist sent Referring Physician s 0008 0090 2 Value from user interface sent Accepts 32 Name characters Study ID 0020 0010 2 Generated for each exam on the scanner and always sent Study Description 0008 1030 3 Value from user interface or worklist sent Accepts 22 characters by default but is configurable to accept 64 characters Name of Physician s 0008 1060 3 Sent if entered at the user interface Reading Study A 3 3 Patient Study Module Attribute Name ea __ Patient s Age 0010 1010 LM from Date of Birth entered at user interface Patient s Weight 0010 1030 3 Value from user interface or worklist sent Additional Patient s 0010 21b0 3 Value from user interface History 146 of 215 IL ID IVI UIT UI WIDULUVOLy V OYOLOII LU UIICIISC U I CAUTION table height while scanning a series of images Therefore GE Healthcare implementations must use WRB MOP RS DaD QDICOM Conformance Statement
135. cheduled Step Attribute Sequence 0040 0270 Up to 15 items Up to 15 items gt Study Instance UID 0020 000D Copied from SPS if option to copy is selected or else created at the scanner Created at the scanner Copied from source image gt Referenced Study 0008 1110 Copied from Not sent as part of Not sent as part of Sequence SPS if selected image header image header gt Accession Number 0008 0050 Copied from User input on the Copied from source SPS if selected scanner image gt Placer Order Number Imaging Service Request 0040 2016 Not sent as part of image header Not sent as part of image header Not sent as part of image header gt Filler Order Number Imaging Service Request 0040 2017 Not sent as part of image header Not sent as part of image header Not sent as part of image header gt Requested Procedure ID 0040 1001 Copied from User input on the Copied from source SPS if selected scanner image gt Requested Procedure 0032 1060 Copied from Not used Not sent as part of Description SPS if selected image header gt Placer Order 0040 1006 Not sent as part Not used Not sent as part of Number Procedure of image header image header gt Filler Order 0040 1007 Not sent as part Not used Not sent as part of Number Procedure of image header image header gt Scheduled Procedure ID 0040 0009 Copied from Not used Co
136. chronous mode is not supported All operations will be performed synchronously 2 3 1 1 4 Implementation Identifying Information The table in Appendix E identifies the Implementation UID for this product version DOC0550678 56 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 2 3 1 2 Association Initiation by Real World Activity This AE attempts to initiate a new association due to Ping DICOM Host or Push operation initiated by the user A new association is also initiated when the user performs a Query Remote operation or issues a retrieve operation by performing a Get operation at Study Series Image level 2 3 1 2 1 Push Image s to Remote AE 2 3 1 2 1 1 Associated Real World Activity Behavior for Discovery 710 610 and Optima 560 Products SW Version pet_mfk xx The operator must first select a destination by choosing Select Remote Host from the Network pull down menu on the local database manager and then choose a hostname The Push operation will cause the DICOM server AE to initiate an Association when the operator selects one or more study series or images in the local database manager and then chooses either Push Examination Push Series or Push Image from the Network pull down menu on the local database manager Note If multiple study series or images are chosen to be pushed one association w
137. cretariat NEMA 1300 N 17 Street Rosslyn VA 22209 USA Suite 1742 Phone 1 703 841 3200 1 3 Intended Audience The reader of this document is concerned with software design and or system integration issues It is assumed that the reader of this document is familiar with the DICOM Standard and with the terminology and concepts which are used in that Standard 1 4 Scope and Field of Application DOC0550678 It is the intent of this document to provide an unambiguous specification for GEHC implementations This specification called a Conformance Statement includes a DICOM Conformance Statement and is necessary to ensure proper processing and interpretation of GEHC medical data exchanged using DICOM The GEHC Conformance Statements are available to the public The reader of this DICOM Conformance Statement should be aware that different GEHC devices are capable of using different Information Object Definitions For example a GEHC CT Scanner may send images using the CT Information Object MR Information Object Secondary Capture Object etc Included in this DICOM Conformance Statement are the Module Definitions which define all data elements used by this GEHC implementation If the user encounters unspecified private data elements while parsing a GEHC Data Set the user is well advised to ignore those data elements per the DICOM standard Unspecified private data element information is subject to change without notice If
138. d MWL Order Original Modality NM Others 119 of 215 When PPS start N CREATE message is sent this element will have The Value for the Modality tag depends on the original Modality value Modality in the MPPS SOP Gr PT NM PT GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Multiples MWL items selected MWL Order Original Modality Modality in the MPPS SOP 1 All orders have the same Modality Selected Modality value CT or PT or NM CT or PT or NM 2 All orders have the same Modality PT which is not CT nor PT or NM 3 Different Modalities found in orders PT 7 10 1 Patient Level 00100010 00100020 Copied from SPS if selected 0010 0030 Copied from SPS if selected Referenced Patient Sequence 0008 1120 Copied from SPS if selected Ref SOP class uid 008 150 Ref SOP Instance uid 0008 1155 7 10 2 Study Level 0010 0040 Copied from SPS if selected Accession Number 0008 0050 If multiple SPS s are selected then accession number from the first selection determined by the user is used Copied from SPS if selected Referenced Study Sequence 0008 1110 Ref SOP class uid 008 150 Ref SOP Instance uid 0084188 7 10 3 Series Level Attribute Name Usage in CT PET DICOM Image Header Modality 0008 0060 Value CT PT or NM is stored in the image
139. d Activity Verify DICOM Ping enne 43 2 3 1 2 4 1 Associated Real World em emen 43 2 3 1 2 4 2 Proposed Presentation Context 43 2 3 1 3 Association Acceptance Policy cee seceeeeeeeeeeeaeeeeseeaaeeeeeesesaeeeeeesensueeeeeeeenees 44 2 3 1 3 1 Receive Image S ir rimi ionii ka dre enge dete deed edet ee delen ade iiie edad de ed 44 2 3 1 3 1 1 Associated Real World 44 2 3 1 3 1 2 Presentation Context Table 44 2 3 1 3 1 3 Transfer Syntax Selection 46 2 3 1 3 2 Verification Request from Remote 47 2 3 1 3 2 1 Associated Real World ener nennen 47 2 3 1 3 2 2 Presentation Context 47 2 3 1 3 3 Query Request from Remote 48 2 3 1 3 3 1 Associated Real World 48 2 3 1 3 3 2 Presentation Context 48 2 3 1 3 4 Retrieve Request From Remote 50 2 3 1 3 4 1 Associated Real World 50 2 3 1 3 4 2 Presentation Context 50 24 COMMUNICATION PROFILES 32 2 oe a E 51 DOC0550678 14 of 215 GE Healthcare Discovery 710 610 and Optima 560
140. d by the PPS Server DICOM AE The PPS Server DICOM AE is commanded to perform Performed Procedure Step services either automatically or through the user interface 7 2 1 Application Data Flow Diagram The basic Application models for the feature are shown in the following illustration Patient Scan with MWL data automatic Patient Scan without MWL data automatic PPS Server C MPPS SCP Post processing thr user interface DICOM Standard Interface 107 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 7 2 2 Functional Definition of AEs The PPS Server AE is implemented as an application process on the scanner host computer It runs as a daemon serving requests from other applications to send the PPS information to the remote AE and return the results to the requesting application The PPS Server AE initiates the following functions Start PPS Initiates a DICOM association in order to create a DICOM Modality Performed Procedure Step SOP instance in the remote AE If the remote AE accepts a presentation context applicable to Modality performed Procedure Step the PPS Server AE will issue a request to create the SOP instance in the remote AE via the N CREATE service Complete PPS Initiates a DICOM association in order to update a DICOM Modality Performed Step instance that is already created with the remote AE If the remote AE accepts a presentatio
141. d on the response data and is presented to the user 7 3 AE Specification 7 3 1 PPS Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU 110 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN SOP Class Name SOP Class UID Modality Performed Procedure Step 1 2 840 10008 3 1 2 3 3 7 3 1 1 Association Establishment Policies 7 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU from an association initiated by the PPS Server AE is Maximum Length PDU 128 Kbytes The SOP Class Extended negotiation is not supported The maximum number of presentation negotiation items that will be proposed is 1 The user information items sent by this AE are e Maximum PDU Length Implementation UID 7 3 1 1 2 Number of Associations The PPS Server will initiate only one DICOM association at any time to perform a PPS operation to the remote AE 7 3 1 1 3 Asynchronous Nature Asynchronous mode is not supported All operations are performed synchronously 7 3 1 1 4 Implementation Identifying information The table in Appendix E identifies the Implementation UID for this pro
142. duct version 7 3 1 2 Association Initiation Policy The PPS Server AE initiates a new association for every PPS operation initiated 7 3 1 2 1 Real World Activity Performed Procedure Step creation and update 7 3 1 2 1 1 Associated Real World Activity The real world activities are mentioned in section 7 2 3 Sequencing of Real World Activities Each of the real world activity results in either creating a new Performed procedure Step SOP instance at the remote SCP or updating an already created Performed Procedure Step SOP instance as per the DICOM standard 111 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 7 3 1 2 1 2 Proposed Presentation Context Table Direction 5330620 1EN The following table shows the proposed presentation contexts for the PPS Server AE after any of the real world activity listed in section 7 2 3 Sequencing of Real World Activities is initiated Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Modality 1 2 840 10008 3 1 2 3 3 Implicit VR 1 2 840 10008 1 2 SCU None Performed Little Endian Procedure Step 7 3 1 2 1 2 1 SOP Specific DICOM Conformance Statement for MPPS SOP class If the remote AE does not support the proposed Presentation context an appropriate error message logged Only one association is opened at a time All the operations used
143. e 2 GE Private PET 1 2 840 113619 4 3 Implicit VR Little 2 840 10008 1 raw data 0 Endian 2 DOC0550678 1 4 1 UID 2 840 10008 5 1 1 1 2 0 10008 5 1 2 840 10008 5 1 1 1 1 4 1 1 7 1 2 840 10008 5 1 4 1 1 88 67 1 88 22 1 2 840 10008 5 1 4 1 1 2 840 10008 5 1 4 1 1 128 1 2 840 10008 1 1 Implicit VR Little Endian Implicit VR Little Endian Implicit VR Big Endian GE Private Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Implicit VR Big Endian GE Private Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Implicit VR Big Endian GE Private Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Explicit VR Little Endian Explicit VR Big Endian 45 of 215 UID List 1 2 840 10008 1 2 1 2 840 10008 1 2 1 2 840 113619 5 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 113619 5 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 113619 5 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2 2 1 2 840 10008 1 2 1 2 840 10008 1 2 1 1 2 840 10008 1 2
144. e Software Versions 0018 1020 3 LO 1 D 2 5 Private Creator Identification Information Refer to section B 2 1 of this document as GEMS PETD 01 is used for this private IOD D 2 6 Discovery PET Patient Module Refer to section B 2 2 of this document D 2 7 Discovery PET Exam Module Refer to section B 2 3 of this document 207 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN D 2 8 Discovery PET Scan Module Refer to section B 2 4 of this document D 2 9 Discovery PET Frame Module Refer to section B 2 5 of this document D 2 10 General Modules This section defines the attributes which are required for proper functioning and identification of the associated SOP Instances They do not specify any semantics about the Real World Object represented by the IOD D 2 10 1 SOP Common Module Attributes Attribute Name Tag Type VR VM Notes SOP Class UID 0008 0016 1 UI SOP Instance UID Specific Character Set Instance Creation Date Instance Creation Time Instance Creator UID 0008 0014 3 UI a D 2 11 Discovery List Data Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET raw data type 0021 1001 3 us 1 208 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 209 of 215 GE Healthcare R
145. e Commitment Provider Sends the selected images to the remote DICOM AE Closes the association If all the images are transferred without failures the following steps will be executed If there are any failures the job will be marked as failed and the Storage Commitment request will not be sent Establishes a new association for sending the commitment request Here note that a storage commitment request does not imply that the acquisition of images is complete Closes the association for sending the commitment request Receives the response on a different association Updates the archive flag information for successful instances The archive flag is shared with the local archive and the two cannot be distinguished 5 2 3 Sequencing of real world activities The user has to select the remote commitment provider and the exams series to be pushed before clicking Save Exam Series button on the Archive menu 5 3 AE Specifications 5 3 1 DICOM Storage Commitment SCU specifications Provides Standard Conformance to the following DICOM SOP class as an SCU DOC0550678 77 01215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN SOP class name SOP class UID Storage Commitment Push model SOP class 1 2 840 10008 1 20 1 5 3 1 1 Association establishment policies 5 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1
146. e Designator Not Used gt Code Meaning Not Used gt Anatomic Region Modifier 0008 2220 3 SQ 1 Not Used Sequence 2 Code Value Not Used gt gt Coding Scheme Designator 0008 0102 1 SH 1 Not Used gt gt Code Meaning 0008 0104 3 ILO 1 Not Used Primary Anatomic Structure 0008 2228 3 SQ 1 Not Used Sequence gt Code Value 0008 0100 1 SH 1 Not Used gt Coding Scheme Designator Not Used 171 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN gt Code Meaning 0008 0104 Not Used gt Primary Anatomic Structure 0008 2230 3 SQ 1 Not Used Modifier Sequence Code Value 00080400 1 SH 1 Not Used Coding Scheme Designator 0008 0102 KLUB Not Used gt gt Code Meaning 0008 0104 Not Used A 7 19 Overlay Plane Module This section contains Attributes that describe characteristics of an Overlay Plane Overlay Planes are not currently used in GE Discovery ST PET AE A 7 20 VOI LUT Module This section specifies the Attributes that describe the VOI LUT Attribute Name Tag BG E Notes VOI Lut Sequence 0028 3010 Not Used gt LUT Descriptor 0028 3002 1C US US or Not Used SS US gt LUT Explanation r msn sma Not Used gt LUT Data Not Used Window Center soe Y 1 n Window center Window Width 220 ELE 5 1 n Window width Window Center amp Width 0028 1055 3 1 n Not Used Explanation
147. e Location 0020 1041 A 7 17 Image Pixel Module This section specifies the Attributes that describe the pixel data of the image Attribute Name Tag Type VR Notes Samples per Pixel 0028 0002 1 fus Photometric Interpretation Rows 0028 0010 1 US 1 Matrix Size Height Columns 00280011 1 US 1 Matrix Size Width Bits Allocated 0028 0100 1 US 1 168 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Bits Stored 0028 0101 High Bit 0028 0102 Pixel Representation 0028 0103 0001H 2 s complement Pixel Data 7FE0 0010 Pixel data sent as OW OW Planar Configuration 0028 0006 Not Used Pixel Aspect Ratio 0028 0034 Not Used Smallest Image Pixel Value 0028 0106 Not Used Largest Image Pixel Value 0028 0107 Not Used cciocioc 855 8599 Red Palette Color Lookup 0028 1101 1C 3 Not Used Table Descriptor Green Palette Color Lookup 0028 1102 1C 3 Not Used Table Descriptor Blue Palette Color Lookup 0028 1103 1C 3 Not Used Table Descriptor Red Palette Color Lookup 0028 1201 1C 1 n Not Used Table Data Green Palette Color Lookup 0028 1202 1C 1 n Not Used Table Data Blue Palette Color Lookup 0028 1203 1C 1 n Not Used Table Data A 7 18 PET Image Module This section specifies the Attributes that describe the image within a part
148. e Storage Enhanced SR X Ray Radiation Dose SR RT Structure Set Storage Positron Emission Tomography Image Storage GE Private PET Raw Data Storage GE Private PET List Data Storage lt EnhanceaSR 1 Ye 2 2 Query Retrieve Study Root Query Retrieve Information Model FIND Study Root Query Retrieve Information Model MOVE Print Management 2 Basic Grayscale Print Management Meta SOP Class Basic Color Print Management Meta SOP Class Print Job SOP Class Printer SOP Class 2 2 2 Workflow Management Storage Commitment Push Model SOP Class Modality Performed Procedure Step SOP Class Modality Worklist Information Model FIND SOP Class Yes No o Option This means that this service can be purchased separately Table 0 2 provides an overview of the Media Storage Application Profiles supported by the PET CT products identified above 12 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Table 0 2 MEDIA SERVICES Media Storage Application Write Files Read Files Profile FSC or FSU FSR Compact Disk Recordable General Purpose JPEG DVD Yes FSC No General Purpose JPEG USB Yes FSC Yes 13 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 1 TABLE OF CONTENTS
149. e length exceeds 16 characters the AE Title is truncated to a length of 16 Local IP Address Local IP net mask Local IP gateway The following parameters are configurable but need change only in case of a system software upgrade Implementation UID PDU size Association time out period Session time out period 7 6 3 Configurable DICOM Data The following DICOM sequences are currently defaulted to be automatically sent in the PPS N CREATE and PPS N SET however PPS can be configured to not send them 0008 1032 Procedure Code Sequence sent in the PPS N CREATE If not sent the PPS N CREATE will contain an empty Procedure Code Sequence 0040 0260 Performed Protocol Code Sequence sent in the PPS N SET If not sent the PPS N SET will not contain a Performed Protocol Code Sequence 114 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN The configuration is performed by GEHC Field Service Engineers using software environment variables 7 7 SUPPORT OF EXTENDED CHARACTER SETS PPS supports ISO_IR 100 character set 7 8 N CREATE amp N SET REQUEST MESSAGE PPS Feature for Discovery supports all named attributes listed in Table F 7 2 1 in PS3 4 of DICOM standard That is attributes that are not explicitly referenced by name in the table are not supported Example is last row in the table reads All other attributes from Radiation Dose Module and Billing and
150. e of phantom used for CTDI Type Code Seq measurement according to IEC 60601 2 44 gt Single Collimation 0018 9306 3 The width of a single row of acquired data Width in mm gt Total Collimation 0018 9307 3 The width of the total collimation in mm over the area of active x ray detection gt Sprial Pitch Factor 0018 9311 Ratio of the Table Feed per Rotation 0092 9310 to the Total Collimation Width 0018 9307 gt Body Part 0018 0015 Text description of the part of the body Examined examined See the values for target region at the end of section 9 4 3 for the values stored Comments on 0040 0310 Used to record total and per exposure Radiation Dose DLP The format is TotalDLP xxx A 3 Common Modules DLP xxx DLP xxx A 3 1 Patient Module Patients Name 0010 Don As entered at user interface or from 145 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN worklist Supports 5 different components delimited by Supports a maximum length of 32 characters including the delimiter Patient ID 0010 0020 As entered at user interface or from worklist Supports maximum of 16 characters Patient s Birth Date 0010 0030 2 As entered at user interface or from worklist Patient s Sex 0010 0040 2 As entered at user interface or from worklist A 3 2 General Study Module Attribute Type Notes Study Instan
151. e sent along with the response 5 3 1 3 2 2 Commit response with FAILURE status The following attributes are expected as part of DATA Set for N EVENT REPORT from SCP Attribute Value 0008 1198 SOP Class UID 0008 1150 Value received from SCP SOP Instance UID 0008 1155 Value received from SCP 0008 1197 Value received from SCP In case of complete partial failure the user will be notified about the status and the job entry will be paused There is no attempt made to retry automatically the failed sop instances However the user can manually retry the failed jobs Such requests will be treated as new requests This will go through the whole sequence of operations once again The failure reason is ignored Failed SOP instances will have their archive flag information unaltered Note The archive status flag in the browser is a shared flag with local archive When the status is Y it means that the images are archived but doesn t specify whether on local archive device or remote archive device It is left to the user s discretion whether the local sop instances with their archive flag set to Y are to be deleted N EVENT REPORT RSP will be sent on the same association itself No DATA Set will be sent along with the response DOC0550678 82 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Note 0008 1199 Reference SOP Sequence is not handled
152. e types of matching that can be request by the implementation Single Value matching Universal Matching Range of date time 6 8 9 Scheduled Procedure Step Entity 6 8 9 1 SOP Common Module Table 6 8 9 1 SOP Common Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Image Specific Character Set 0008 0005 Matching for this item is supported only for the character set ISO IR 100 SCP is not required to return 0008 0005 if there are no non ASCII characters in the response 6 8 9 2 Scheduled Procedure Step Module Table 6 8 9 2 Scheduled Procedure Step Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Image Scheduled Procedure Step 0040 0100 Sequence Scheduled Station AE 0040 0001 Title Matching is supported as follows either no AE title is supplied universal matching or the scanner s Worklist Server AE title is supplied for matching this is user selectable Scheduled Procedure 0040 0002 1 No Matching is supported as one of the Step Start Date following this is user selectable all days today only today and a number of days before today today and a number of days after today today and a number of days before today and a number of days after today Number of days before after is specified by the user Returned values must be exactly 8 nu
153. eStepStartDate 0040 0003 TM no value available 0 0 ScheduledProcedureStepStartTime 0040 0006 PN no value available 0 0 ScheduledPerformingPhysiciansName 0040 0007 LO no value available 0 0 ScheduledProcedureStepDescription 0040 0008 SQ Seq with explicit Length 4 1 32 1 ScheduledActionItemCodeSequence fffe e000 na Item with explicit Length 4 3 24 1 Item 0008 0100 SH no value available 0 0 CodeValue 0008 0102 SH no value available 0 0 CodingSchemeDesignator 0008 0104 LO no value available 0 0 CodeMeaning fffe e00d na ItemDelimitationItem for re encoding f 05 1 ItemDelimitationItem fffe e0dd na SequenceDelimitationItem for re enc 0 1 SequenceDelimitationItem 0040 0009 SH no value available 0 0 ScheduledProcedureStepID 0040 0010 SH no value available 0 0 ScheduledStationName 0040 0011 SH no value available 0 0 ScheduledProcedureStepLocation 0040 0012 LO no value available 0 0 PreMedication fffe e00d na ItemDelimitationItem for re encoding f 05 1 ItemDelimitationItem fffe e0dd SequenceDelimitationItem for re enc 0 1 SequenceDelimitationItem 0040 1001 SH no value available 0 0 RequestedProcedureID 0040 1003 SH no value available 0 0 RequestedProcedurePriority 0040 1004 LO no value available 0 0 PatientTransportArrangements 0040 1005 LO no value available 0 0 RequestedProcedureLocation 0040 1008 LO no value a
154. ecification of the type of communication used between Application Entities Example DICOM network protocol Association a network communication channel set up between Application Entities Attribute a unit of information in an object definition a data element identified by a fag The information may be a complex data structure Sequence itself composed of lower level data elements Examples Patient ID 0010 0020 Accession Number 0008 0050 Photometric Interpretation 00028 0004 Procedure Code Sequence 0008 1032 Information Object Definition IOD the specified set of Attributes that comprise a type of data object does not represent a specific instance of the data object but rather a class of similar data objects that have the same properties The Attributes may be specified as Mandatory Type 1 Required but possibly unknown Type 2 or Optional Type 3 and there may be conditions associated with the use of an Attribute Types 1C and 2C Examples MR Image IOD CT Image IOD Print Job IOD Joint Photographic Experts Group JPEG a set of standardized image compression techniques available for use by DICOM applications DOC0550678 28 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Media Application Profile the specification of DICOM information objects and encoding exchanged on removable media e g CDs Module a set of Att
155. ed Series Level Keys Modality 00080060 Zero length Seresnumber 1 16 00200011 Zerolengh Series description O 0008 103E Zerolengh Manufacturer 70 00080070 Zerolength _ Imagesinseries O 7 00201002 jZerolengh Table 2 3 1 2 2 2 1 3 Requested Image Level Keys image number R 00200013 Zero eng maget _ O 0006000 Zero length Ros _ o 00280000 Zero length Columns O 0028004 Zero length image postion O 00200032 Zero length image orientation o 00200087 Zerolengh Slice thickness o 00182060 Zero length spacing o 00160088 Zero length 00131120 Zero length Convolution keme o 00181210 Zerolength Reconstruction diameter O 0018 1100 Zerolength Data collection diameter O 00182090 Zerolength Fipane Jo 00184314 Zero length Echo Number 00160086 zerlengh Enim 00180081 zerlengh inversion me O 00160082 zerlengh Repetition ime o 00160080 zerolenth Private Creator Identification P 001900 6 ACQU 01 Dovre Zero Length Mascan tme 001924 Zero Length DOC0550678 41 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Azimuth 2 0019 26 ZeroLength N
156. ed with a new series number 8 4 COMMUNICATION PROFILES 8 4 4 Supported Communication Stacks PS 3 8 DICOM Upper Layer PS 3 8 is supported using TCP IP 8 4 2 OSI Stack The OSI Communication stack is not supported by this implementation 126 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 4 3 TCP IP Stack The TCP IP stack is inherited from the operating system 8 4 3 1 API Not Applicable to this product 8 4 3 2 Physical Media Support Ethernet 802 3 provides the physical network layer for this product 8 4 4 Point to Point Stack The Point to Point Stack is not supported by this implementation 8 5 EXTENSIONS SPECIALIZATION PRIVATIZATION 8 5 1 Standard Extended Specialized Private SOPs GSPS for Discovery does not implement any private SOP classes 8 5 2 Private Transfer Syntaxes GSPS for Discovery does not implement any private Transfer Syntaxes 8 6 CONFIGURATION The Exam Split feature is an option The DICOM configuration items below are configurable or re configurable by the Field Service Engineer and are not accessible by users through the Discovery s user interface 127 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 6 1 AE Title Presentation address Mapping Exam Split allows for the configuration of the following parameters that pertain to the remote AE
157. ence may contain a maximum of 15 sequence items This sequence will only appear in the image header if the VES HES option is installed or PPS is enabled 8 10 2 Accession Number If scheduled procedures are grouped the accession number for each procedure is stored in the Request Attributes Sequence 0040 0275 in the image header If two or more of the accession numbers in the grouped case are different the top level Accession Number 0008 0050 shall contain 1 of the accession numbers 134 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 9 CT X ray Radiation Dose Structured Report Object Implementation 9 1 100 Module Table Entity Name Module Name R f rence Usage em Suey lass Ses 5 Document Series o2 c lt lt 9 2 SR Document Series Module attibute Name Ta Type Referenced Performed 0008 1111 2 Sent as null Procedure Step Sequence 9 3 SR Document General Module Attribute Name Type Notes see LC MEMINI NN eser _ 0905 1 Se Verifying Observer 0040 A073 1C Not sent per conditional Sequence gt Verifying Observer on Oum Not sent Identification Code Sequence gt gt Include Code Sequence Macro Table 8 8 1
158. ent id Study date Accession number and Study id Not entering a value means match on any value for that field Patient name will match on any patient name that contains what the operator entered Patient id Study id and Accession number will match on what the operator enters For Study date the operator selects a range type from the Exam Date pull down menu where the choices are Equals Before Between or After Once a range type is selected the correct number of fields appears and the operator enters dates into those fields Once the desired parameters are entered the operator chooses and that will cause the DICOM Server AE to initiate an association to the selected Remote AE Once a list of Study Series Image is retrieved the operator can invoke the Pull operation by clicking on the Local DB button at the bottom of the local database browser 2 3 1 2 2 2 Proposed Presentation Contexts When the Real World activity Query or Get is initiated all presentation contexts shown in table 2 2 1 2 1 2 1 are proposed during association establishment but only the Query Retrieve FIND related contexts are applicable to this activity 2 3 1 2 2 2 1 SOP Specific Conformance Statement for C FIND SCU After the Query operation is initiated the DICOM Server AE will perform a study root C FIND RQ request at each of the three levels Study Series and Image in succession The Initial Study Level request will ask for all studi
159. equested from the Visit Admission Module 6 8 13 Patient Entity 6 8 13 1 Patient Relationship No data elements are requested from the Patient Relationship Module 6 8 13 2 Patient Identification Table 6 8 13 2 1 Patient Identification Module Attributes Attribute Expected Mapped Name Ta Returned into the Matchi Key Type Image ng Key Type g Patient s 0010 0010 Name Supports up to 32 characters by default however this is configurable to allow 64 characters See also Note1 BINE Patient ID 0010 0020 1 Yes Supports up to 16 characters by default however this is configurable to 64 characters 1 Other 0010 1000 Supports 64 characters Patient ID Modality Worklist server supports 32 characters for patient Name Supports DICOM format for patient Name with as delimiters If patient name in worklist has more than 32 characters then Worklist will be accepted by the server Worklist Browser will display only the first 32 characters More Info screen will display the full patient name Only the FIRST 32 characters is copied into the final DICOM image header by default however this is configurable to allow 64 characters DOC0550678 100 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 8 13 3 Patient Demographic Table 6 8 13 3 1 Patient Demographic Module Attributes Patient s Birth Date 0010 0030 Y
160. er will be able to enter the values for Accession Number prior to the query If value is entered then that value will be sent as part of the query For example if 1234 is entered then 0008 0050 SH 1234 4 1 AccessionNumber User will be able to enter the values for Requested Procedure Id prior to the query If value is entered then that value will be sent as part of the query For example if 3456 is entered then 0040 1001 SH 3456 4 1 RequestedProcedureID 103 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN User will be able to enter the values for Patient Id prior to the query If value is entered then that value will be sent as part of the query For example if 6789 is entered then 0010 0020 LO 6789 4 1 PatientID User will be able to enter the values for Patient Name prior to the query If value is entered then that value will be sent as part of the query For example if Lastname Firstname is entered then 0010 0010 BN Lastname Firstname 18 1 PatientName 104 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 6 11 Use of Specific DICOM Data This section details the use of the DICOM data returned by remote AEs during worklist queries The Discovery user interface fields that display the data along with the data s mapping into resulting acquired and trans
161. erver AE will search the local database for all entries that match the keys requested by the Remote DICOM AE and send back the list of matches The DICOM Server AE will also respond to an incoming retrieval request from a Remote AE by sending the image s to the Destination AE 2 2 2 Functional Definition of AE s DICOM Server Application Entity initiates the following operations Ping DICOM Host Verification The DICOM Server initiates an association and sends a C ECHO RQ message to the remote DICOM AE the remote DICOM Server will send back a C ECHO RSP message with a status of success Push Initiate an association to a Remote AE to send image s If the Remote AE accepts the presentation context applicable to the image s being sent the DICOM Server AE will send the image s by invoking C STORE RQ operation for each image on the same association Query Initiate an association with a Remote AE to query for images on the remote host A Study Root Study Level C FIND RQ request will be sent to the Remote AE once an association has been established After all responses are received DICOM Server AE will issue a Series Level C FIND RQ request to get the series for a study in the list An Image Level C FIND RQ will be issued for the first series in the series list Get Send a C MOVE RQ request to a Remote AE after successful association establishment The DICOM Server AE s Storage SCP will receive the images over a separate associa
162. erwise sent empty Rows 0028 0010 2 2 If present in composite object instances it will be set to same value otherwise sent empty Columns 0028 001 1 2 2 If present in composite object instances it will be set to same value otherwise sent 197 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN empty Private Record UID 0004 1432 1C 1C Not supported 198 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN APPENDIX D GE Discovery Private PET RAW DATA D 1 GE Private PET Raw Data Information Object Implementation This section specifies the implementation of GE private defined Information Object namely GE Private PET Raw Data IOD used in the GE PET Raw Data Storage SOP Class 1 2 840 113619 4 30 The implementation is detailed to the extent that vendor applications can recognize an instance of the object Storage devices that supports full fidelity storage of this IOD can use the data detailed here to optimize the storage and retrieval of the data or display useful information to the user to distinguish among the IOD instances The IOD contains data that are not published in this document No implementation shall assume any meaning to such data and its meaning and definition is subject to change without notice D 1 1 Discovery PET RAW Data Patient Entity
163. es This value is also used to calculate the Patient s Age Patient s Sex 0010 0040 Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Yes Patient s Weight 0010 1030 2 Yes Limited to maximum value of 999 kg Confidentiality constraint 0040 3001 2 No on patient data Patient s Size 0010 1020 Patient s Address 0010 1040 Patient s Telephone 0010 2154 Numbers 6 8 13 4 Patient Medical Table 6 8 13 4 1 Patient Medical Module Attributes Attribute Name Tag Expected Expected Mapped Matching Returned into the Key Type Key Type Patient State 0038 0500 Displayed on More Info Screen Displayed on More Info Screen Displayed on More Info Screen Displayed on More Info Screen Special Needs 0038 0050 Additional Patient History 0010 21B0 Displayed on More Info Screen 6 9 Private Data Dictionary The Modality Worklist implementation does not define any Private Attributes within the Modality Worklist Information Model DOC0550678 101 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 6 10 C FIND Request Message This section provides a detailed description of the C FIND request message data that is provided to the remote AE during a worklist query operation The dump in Table 6 10 1 below lists in exact message order the fields tr
164. es in the Remote database The C FIND SCU will not perform any extended negotiation and so will only perform hierarchical query Relational Queries are not supported C CANCEL FIND RQ is not supported Each C FIND SCU operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 300 seconds and 60 minutes respectively If a Cancel or Refused status is returned from the Remote AE the association is closed and the operation terminated The DICOM Server AE will parse each matching C FIND RSP reply and ignore the entries it fails to parse Tables 2 2 1 2 2 2 1 1 2 2 1 2 2 2 1 3 shows the various fields that are requested at the Study Series and Image levels of the C FIND request Query results are filtered based on the Modality field Only CT MR PET Screen Save images are supported Table 2 3 1 2 2 2 1 1 Requested Study Level Keys Study date R X 0008 020 Zero length Study time R X 0008 030 Zero length DOC0530678 40 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Patient s name 7 16 00100010 Zero length PalentiID 16 00100020 Zerolengh Studyid 00200010 Zerolengh Study description 7 0 00081030 Zero length Private Creator Identification 7 000900 GEMS IDEN 01 Suteld 0009 xx02 Zero Length Table 2 3 1 2 2 2 1 2 Request
165. eve option to Yes if the Remote AE needs to Query Retrieve the local system When the DICOM Server AE accepts an association for image storage it will receive any images transmitted on that association and store the images on disk It will also respond to queries from Remote AEs by sending matching entries Any Remote AE can request and receive a list of images on the local database The Remote AE must be configured in the local database managers list of Remote AE for it to be able to retrieve images from DICOM Server AE Any remote AE can open an association to the DICOM Server AE for the purpose of verification 2 3 1 3 1 Receive Image s This AE is indefinitely listening for associations No operator action is required to receive an image 2 3 1 3 1 1 Associated Real World Activity The Real World Activity associated with the Receive Image s operation is the storage of the image on the disk drive of the Discovery 2 3 1 3 1 2 Presentation Context Table Table 2 3 1 3 1 2 1 Accepted Presentation Contexts for DICOM Server AE and Real World Activity Receive Image s Presentation Context Table Accepted Abstract Syntax Transfer Syntax gu taal d DOC0550678 44 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Name Verification CT Image Storage MR Image Storage Secondary Capture Image Storage X ray Radiation Dose SR Enhanced SR PET Image Storag
166. ferenced SOP Class UID 0008 1150 1C UI 1 Value copied from SPS gt Referenced SOP Instance 0008 1155 1C UI 1 Value copied UID from SPS Procedure Code Sequence 0008 1032 3 SQ 1 Not Used Direction 5330620 1EN 157 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN A 7 4 Patient Study Module This section defines Attributes that provide information about the Patient at the time the Study was performed Attribute Name Tag Type VR VM Notes Admitting Diagnosis 0008 1080 3 LO 1 n Not Used Description Patient s Age 0010 1010 3 AS 1 Value equals to Exam date minus Patient Birth date Patient s Size 0010 1020 3 DS 1 Value is always in meter unit Patient s Weight 0010 1030 3 DS 1 Occupation 0010 2180 3 SH 1 Not used Additional Patient s History 0010 21B0 LT 1 A 7 5 Discovery PET Exam Module Refer to Section B 2 3 for details A 7 6 General Series Module Attribute Name Tag Type VR VM Notes Modality 0008 0060 1 CS 1 Series Instance UID Laterality Not used Performing Physician s Name 0008 1050 Not used Protocol Name 0018 1030 LO 1 User selected protocol Series Description 0008 Operators Name 0008 1070 Referenced Performed 0008 1111
167. ferred DICOM images are presented in following table 6 11 1 Table 6 11 1 Specific Data Usage Element Accession Number 0008 0050 Req Number Accession Number Supports maximum of 16 characters Patient ID 0010 0020 Patient ID Patient ID Supports maximum of 16 characters Patient Name 0010 0010 Patient Name Patient Name Supports maximum of 32 characters Patient s Birth Date Patient Age Patient Birth Date user Patient Birth Date emm e feet Patients Sex 0010 0040 Patient s Weight 0010 1030 Weight in Kg Patient s Weight Referring Physician s Name Referring Physician Referring Physician s Name 0008 0090 Requested Procedure Exam Description Study Description Description 0032 1060 Scheduled Procedure Step Date Not available Start Date 0040 0002 Scheduled Procedure Step Not available Start Time 0040 0003 Study Instance UID Study instance UID only displayed Study Instance UID if User has set the 0020 000d on the More Info screen preference to Use MWL Study UID Requested Procedure Id Requested Proc ID Requested Procedure Id 0040 1001 Pregnancy Status 0010 21C0 Pregnancy Status only displayed on Not available the More Info screen Medical Alerts 0010 2000 Medical Alerts only displayed on Not available the More Info screen Contrast Allergies 0010 2110 Contrast Allergies only displayed on Not available the More Info screen Special Needs 0038
168. formance Statement REV 7 PET image set id B 2 5 Discovery PET Frame Module PET image one loc 0009 10D6 PET image index loc 0009 10D7 PET num of rr interval 0009 10DD PET num of time slots 0009 10DE PET num of slices 0009 10DF PET num of time slices 0009 10E0 1 1 PET rest_stress 0009 10E2 1 PET tracerinjection UID 0009 10F2 U 1 PET Q Static Scan Mode 0009 10F6 SL 1 PET Sharp IR Flag 0015 103D UL 1 PET Scatter Limit 0015 103E UL 1 PET Q Static Recon 0015 103F 1 Event histogram Format 0015 1039 L 1 Number of Detector Rows 0015 103A 1 Number of Detector Columns 0015 103B se 1 Physio Gating Type 0015 101A 1 Total Number of Bins 0015 101B 1 Direction 5330620 1EN 1 1 1 0009 10D2 Tro nm dle ete sels pecie s Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET frame id PET trig rej method PET number for reject PET lower reject limit PET upper reject limit PET triggers acquired PET triggers rejected PET scan id 0009 105C 3 men 3 amp 31 3 amp 31 3 amp 31 Gxess 3 1 exse s W 3 182 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN PET compatible version PET software version PET landmark datetime P
169. formed by a GE Field Service Engineer Local IP Address Local IP Netmask Max PDU length The following DICOM print parameters are configurable The valid ranges are shown in earlier sections of this conformance statement Medium Type 2000 0030 Film Destination 2000 0040 Magnification Type 2010 0060 Min Density 2010 0120 Max Density 2010 0130 Empty Image Density 2010 0110 Border Density 2010 0100 Configuration Information 2010 0150 Smoothing Type 2010 0080 The following network timers are supported and are configurable When these timers expire the association is terminated The default values are Association Time out 120 seconds Session Time out 1200 seconds total time for open association N SET Time out 300 seconds N ACTION Time out 300 seconds N CREATE Time out 300 seconds N DELETE Time out 300 seconds N GET Time out 100 seconds Time outs which are set for all hosts are configurable in dprint cfg Association time out bi assoc Session timeout bi session tio Create timeout bi ncreate tio Note All configurations are performed by a GE Field Service Engineer 4 7 Support of Extended Character Set The Print SCU does not support any Extended Character Set DOC0550678 75 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 5 DICOM Storage Commitment SCU Conf
170. ge Number Software Versions 0018 1020 Always sent as 05 for this version does not distinguish individual software releases Spatial Resolution 0018 1050 Sent only for MR images A 3 7 General Image Module Image Type 0008 0008 Always sent Value 3 CT Image IOD specific specializations AXIAL LOCALIZER SEGMENTED REFORMATTED PROCESSED COMBINED CTINTERVENTION Value 4 GE CT Image implementation specific MIN IP MIP AVERAGE VOLREN INTEGRAL 148 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN HD MIP RAYSUM SURFACE MINMAX FLUORO GSI MONO GSI HIGH KV GSI LOW KV GSI MD GSI QC GSI EFF Z GSI CLR OVRLY DIGITALTILT Acquisition Number 0020 0012 3 Generated for each acquisition always sent Acquisition Date 0008 0022 Generated for each acquisition always sent Acquisition Time 0008 0032 3 Generated for each acquisition always sent Format is in fractional seconds as small as 1 millionth of a second A 4 CT Image Modules A 4 1 Image Plane Module Attribute Name Nots o Slice Thickness 0018 0050 Value always sent Image Slice Location 0020 1041 3 Value always sent Note Slice Location reflects the gantry ISO center table location value for the image in mm If the gantry is tilted this value will differ from the image position 0020 0032 Z value which reflects the upper left hand
171. gth Implementation UID 6 3 1 1 2 Number of Associations The Worklist Server AE SCU will initiate only one DICOM association at a time to perform a modality worklist query of a single remote AE 6 3 1 1 3 Asynchronous Nature Asynchronous mode is not supported All operations are performed synchronously 6 3 1 1 4 Implementation Identifying Information The table in Appendix E identifies the Implementation UID for this product version 6 3 1 2 Association Initiation Policy The Worklist Server AE initiates a new association due to an update operation being initiated from the Discovery user interface 6 3 1 2 1 Real World Activity Worklist Query 6 3 1 2 1 1 Associated Real World Activity The operator of the system initiates a query for a modality worklist by either opening the Schedule screen or by opening the Schedule screen and pressing the Update button The choice of which of these two behaviors occurs is user configurable The Worklist Server will then initiate an association with the remote AE in order to query for the worklist A user can configure a number of parameters that directly control the worklist query request The user can request worklist items that are intended for the scanner the user is working at all items that apply to the modality of the scanner the user is working at or all worklist items available These selections and their effects on worklist query parameters are given below Note Only single MWL entry a
172. he appropriate drive DOC0550678 60 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN A user cannot display the directory when create CD DVD USB or restore CD USB are in progress 3 3 1 2 2 1 Application Profiles for the RWA Display Directory For the list of Application Profiles that invoke this AE for the Display Directory CD USB RWA see Table 3 3 1 1 There are no extensions or specializations 3 3 1 2 2 2 Media Storage Application Profile for the RWA Display Directory Following are the SOP Classes supported by the RWA Display Directory of CD USB Information Object Definition Media Storage 1 2 840 10008 1 3 10 Explicit VR Little Endian 1 2 840 10008 1 2 1 Directory Storage SOP Class UID Transfer Syntax Heneroreyntex UID 3 3 1 2 3 Real World Activity RESTORE The DICOM Media CD DVD USB Storage Interchange AE acts as an FSR using the interchange option when requested to copy SOP instances from the CD DVD USB to the local database The user selects the SOP instances that he wants the DICOM Media Interchange AE to copy on the local database by selecting appropriate Study Series Image instances and clicking on the suitable restore buttons Once selected the SOP instances are copied from the media to the local database If the media is not blank then the Interchange Media gets mounted To remove the media the user has to
173. he following restrictions apply on the attributes listed below Referenced Study Sequence The sequence is not sent in the MPPS message Scheduled Step Attribute Sequence a maximum of ONE item is supported The attribute will be send only if SPS information is available in the image instance Referenced Patient Sequence This sequence is not added Scheduled Protocol Code Sequence The sequence is not send in the MPPS message 115 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Performed Procedure Step Start date amp Performed Procedure Step start time The exam date and exam time that is the Start date and Start Time of the associated Study Component Exam is used not the actual time when post processing started Performed Procedure Step end date amp Performed Procedure Step end time The date and time when user chooses to Complete PPS or Discontinue PPS is taken as the Performed Procedure Step End Date and Performed Procedure Step End Time respectively Not the actual end date and end time of post processing Procedure Code Sequence This sequence is sent with ZERO items in the MPPS message Performed Protocol Code Sequence This sequence is sent with ZERO items in the MPPS message Referenced Standalone SOP Instance Sequence The sequence is sent with ZERO items in the MPPS message 7 9 ERROR HANDLING AND RECOVERY PPS Se
174. header depending on the original value of the Modality Tag in the worklist entry or entries 120 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Attribute Name Usage in CT PET DICOM Image Header Please see note above for details Protocol Name 0018 1030 Name of the selected protocol is copied Ta Operator Name 0008 1070 Copied from SPS if selected and present otherwise generated from user input on scanner Referenced Performed 0008 1111 Will be present only if SPS obtained form Procedure Step Sequence HIS RIS is selected for scanning Ref SOP class uid 0008 1150 Value of MPPS SOP class UID Ref SOP Instance uid 0008 1155 Scanner generated unique UID 0 Requested Attribute 040 0275 Will be present only if SPS obtained form Sequence HIS RIS is selected for scanning If multiple SPS s are selected then this will contain multiple items one for each SPS For all other cases this element will not be present Requested Procedure Id 0040 1001 Copied from SPS if selected Scheduled Procedure 0040 0007 Copied from SPS if selected Gems tose Scheduled Protocol Code 0040 0008 Copied from SPS if selected Sequence Created at the scanner The string PPS_ID_ lt examnumber gt is used start date start time description Scheduled Procedure 0040 0009 Copied from SPS if selected Step Id 121 of 215
175. hing Returned into the Key Type Key Type Image Accession Number 0008 0050 2 Yes User will be able to enter value for Accession Number prior to query to be sent as part of C FIND request Supports maximum of 16 characters DOC0550678 98 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Referring Physician s Name 0008 0090 eS w Yes Truncated to 32 characters a INE 6 8 12 Visit Entity 6 8 12 1 Visit Identification Table 6 8 12 1 1 Visit Identification Module Attributes Attribute Name Tag Expected Expected Matching Returned Key Type Key Type Admission ID 0038 0010 Institution Name 0008 0080 6 8 12 2 Visit Status Table 6 8 1 12 2 1 Visit Status Module Attributes Attribute Name Expected Expected Mapped Matching Returned into the Key Type Key Type Image Current Patient Location 0038 0300 ee Displayed on More Info Screen DOC0550678 99 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 8 12 3 Visit Relationship Table 6 8 12 3 1 Relationship Module Attributes Attribute Name Tag Expected Expected Matching Returned Key Type Key Type Referenced Patient Sequence 0008 1120 Referenced SOP Class UID 0008 1150 Referenced SOP Instance UID 0008 1155 6 8 12 4 Visit Admission No data elements are r
176. ickness total coverage exposure time total x ray on time helical total coverage exposure time table velocity axial or cine total coverage macro row thickness no of active rows Eo Slice thickness nominal prospective reconstructed slice selection Note A 4 5 VOI LUT Module Attribute Name Tag Type Notes Rescale Slope 0028 1053 3 C Window Center 0028 1050 Window Center for display Always sent Window Width for display Always sent A 4 6 SOP Common Module Attribute Type 008 008 Window Width 0028 1051 1 SOP Class UID 008 0016 sent SOP Instance UID 008 0018 1 Always sent 1 1 Specific Character Set 0008 0005 ISO IR 100 A 5 SC Image Modules A 5 1 SC Equipment Module Conversion Type 0008 0064 Always sent with value WSD 154 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Modality 0008 0060 Modality of original image A 5 2 Image Pixel Module Samples per Pixel 0028 0002 Always sent with value 1 0028 0004 Photometric 0028 0004 Always sent Interpretation Rows 0028 0010 Always sent Bits Allocated 0028 0100 Bits Stored 0028 0101 Pixel Representation 0028 0103 A 5 3 VOI LUT Module Rescale Slope 0028 1053 3 C Window Center 0028 1050 Window Center for display Always sent Window Width 0028 1051 1 Window W
177. icular PET Series Attribute Name Tag Type VR VM Notes 169 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Image Type 0008 0008 1 CS 1 n Use Image Set Source Samples per Pixel 00280002 1 US 1 1 Photometric Interpretation 0028 0004 1 CS 1 MONOCHROME2 Bits Allocated 0028 0100 Image depth always 16 Bits Stored 0028 0101 Image depth always 16 High Bit 0028 0102 15 0028 1052 0 Rescale Slope Use scale factor Frame Reference Time 0054 1300 1 DS 1 Value equals to Frame Referenced Time if not Zero otherwise value equals to Image Time Trigger Time 0018 1060 1C 05 1 Time interval in msec from the start of the trigger to the beginning of data acquisition for this image Frame Time 0018 1063 DS 1 For Gated Scans value equals to Image bin Duration Otherwise value Rescale Intercept equals to 0 Low R R Value 0018 1081 IS 1 Lower rejection limit High R R Value 0018 1082 IS 1 Upper rejection limit Lossy Image Compression 0028 2110 1C CS 1 Value equals for NO compression Otherwise value equals 01H if compression is lossy Image Index 0054 1330 1 05 1 scan mode image location image time image bin time Acquisition Date 0008 0022 2 DA 1 Value equals Scan Date Time plus Image Time Acquisition Time 0008 0032 2 TM 1 Value equals Sc
178. ide 07 8 Sodium iodide 171247 18F Flutemetamol UNKNOWN Flutemetamol F 18 18F Fluciclatide UNKNOWN Fluciclatide F 18 18F Fluciclovine UNKNOWN Fluciclovine F 18 164 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 A 7 9 PET Multi gated Acquisition Module Direction 5330620 1EN Trigger Source or Type 3 LO 1 Framing Type 0018 1064 Attribute Name Tag Type VR VM Notes Beat Rejection Flag 0018 1080 2 CS 1 Value is Y or N depends on Trigger Rejection Method selected by Users Not Used Not Used Note The Multi Gated Acquisition Module is only sent for Gated respiratory cardiac acquisition A 7 10 NM PET Patient Orientation Module Attribute Name Tag Type VR VM Notes Patient Orientation Code 0054 0410 2 SQ 1 Zero length Sequence sequence gt Code Value 0008 0100 gt Coding Scheme Designator 0008 0102 99SDM gt Code Meaning 0008 0104 gt Patient Orientation Modifier 0054 0412 2C SQ 1 Code Sequence gt gt Code Value 0008 0100 gt gt Coding Scheme Designator 0008 0102 99SDM Code Meaning 0008 0104 1 Patient Gantry Relationship 0054 0414 2 SQ 1 Zero length Code Sequence sequence Code Value 0008 0100 1C SH 1 Coding Scheme Designator 0008 0102 1C SH 1 99SDM Code Meaning 00080104 3 Lo 1 Patient Orientation Code Value Code Meaning 0008 0100
179. idth for display Always sent A 5 4 SOP Common Module A 6 Other Attributes Attribute Name Tag Type Notes Temporal Position Index 0020 9128 3 Filled into CT image for some applications Pass number for shuttle images Columns Nominal Percentage of 0020 9241 3 Filled into CT image for some Cardiac Phase applications Prescribed 155 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN arta of cardiac phase A 7 PET Specific IOD Common Module Definitions A 7 1 PET Image IOD Modules Entity Name Module Name R f rence Patient Patient A 7 2 Study General Study A 7 3 Patient Study 7 4 Discovery Exam A 7 5 Series General Series A 7 6 PET Series A 7 7 PET Isotope A 7 8 PET Multi gated Acquisition A 7 9 NM PET Patient Orientation A 7 10 Discovery PET Imageset A 7 11 Discovery PET Scan A 7 12 Frame of Reference Frame of Reference A 7 13 Equipment General Equipment 7 14 General Image A 7 15 Image Plane A 7 16 Image Pixel A 7 17 PET Image A 7 18 Overlay Plane A 7 19 VOI LUT A 7 20 Discovery PET Image 7 21 Discovery PET Frame A 7 22 General Modules SOP Common A 7 23 A 7 2 Patient Module This section specifies the Attributes of the Patient that describe and identify the Patient who is the subject of a diagnostic Study
180. ill be established for each of the studies series or images Some GE PET Raw SOP instances stored locally in the scanner database are not supported in network operations Refer to Appendix D 2 Behavior for Discovery 710 610 products SW Version pet_coreload xx The operator must first select the exam series image on the local database browser and click on the Remote DICOM AE in the network panel at the bottom of the local database browser to which the operator desires to send the exam series image to 2 3 1 2 1 2 Proposed Presentation Contexts The following table shows the proposed presentation contexts for the DICOM Server AE after Real World Activity Push Operation has been performed Table 2 3 1 2 1 2 1 Proposed Presentation Contexts for DICOM Server AE and Real World activities Push Query Pull Get Images and Verification DICOM Ping Presentation Context Table Proposal Abstract Syntax Transfer Syntax Extended Negotiation UID Lit 1 2 840 10008 5 1 4 1 1 2 Implicit VR Little Endian 1 2 840 10008 1 2 None CT Image Storage Implicit VR Big Endian GE 1 2 840 113619 5 2 Private Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 MR Image Storage 1 2 840 10008 5 1 4 1 1 4 Implicit VR Little Endian 1 2 840 10008 1 2 Implicit VR Big Endian GE 1 2 840 113619 5 2 Private Explicit VR Little Endian 1 2 840 10008 1 2 1 Explicit VR Big Endian 1 2 840 10008 1 2 2 DOC055
181. ing save to media gt Referenced Transfer Syntax UID in 0004 1512 Not supported File C 4 Media Directory Information Module Attribute Name LH M Offset of the First Directory Record of the 0004 1200 Notes s by application Root Directory Entity Offset of the Last Directory Record of the 0004 1202 Set by application Root Directory Entity File set Consistency Flag 0004 1212 0000H no known Directory Record Sequence 0004 1220 Directory Record Type 0004 1430 Referenced File ID 0004 1500 PATIENT EU SERIES IMAGE PRESENTATION SR DOCUMENT and PRIVATE Generated only for Image Presentation Private and SR Document Directory Records starting with AIAIAIAIZOT Generated only for Image Presentation Private and SR Document Directory Records Set to SOP Instance UID 0008 0018 during save to media Referenced SOP Instance UID in File 0004 1511 Referenced Transfer Syntax UID in File 0004 1512 192 of 215 Generated only for Image Presentation Private and SR Document Directory Records Referenced SOP Class UID in file 0004 1510 Generated only for Image Presentation Private and SR Document Directory Records GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN C 5 Directory Record Selection Keys The STD GEN CD and STD GEN DVD Application Profiles will have only the Patient Study Series and Image directory record
182. instances it will be set to same value other wise set to NULL 193 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Accession Number 0008 0050 2 If present in composite object instances it will be set to same value other wise set to NULL C 5 3 Series Keys Attribute Name Specific Character Set 0008 0005 BEEN IR 100 Modality wass 0060 Set to one that is found in composite object instances Series Instance UID 000E This is set to value that is present in the composite object instances otherwise check for presence of Referenced SOP Instance UID in File 0004 1511 if both not present instance is not archived Series Number 0020 0011 If present in composite object instances it will be set to same value other wise set to NULL Manufacturer 0008 0070 2 If present in composite object instances it will be set to same value other wise set to NULL Series Description 0008 103E 2 If present in composite object instances it will be set to same value other wise set to NULL Manufacturer s Model Name 0008 1090 2 If present in composite object instances it will be set to same value other wise set to NULL Image Type 0008 0008 2 If present in composite object instances it will be set to same value other wise set to NULL Series Date 0008 0021 2 If present in composite object instances it will be set to same value other wi
183. ion The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user 7 2 3 3 PPS from post processing system The user initiates post processing on the images generated through acquisition The system creates a Modality performed Procedure Step instance locally in the system If the source image instance has the Scheduled Procedure Step information it is copied into the image instances created Also the system includes the necessary information related to the Modality Performed Procedure Step into the image instance At the end of one or more post processing the user initiates Complete PPS or Discontinue PPS through the user interface provided PPS Server AE initiates a MPPS Modality Performed Procedure Step N CREATE request to the remote AE MPPS SCP in order to create a MPPS SOP instance at the remote AE which is actually a replica of the locally created MPPS SOP instance The remote AE returns response indicating the success failure of the request execution If the response indicates success PPS Server AE initiates a MPPS N SET request to the remote AE in order to update the MPPS SOP instance that is already created with the additional information The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system base
184. is situation Modality Worklist is providing the DICOM C FIND service as a service class user SCU 6 2 Implementation Model All DICOM functionality provided by the Modality Worklist feature is logically provided by the Modality Worklist Server DICOM AE The Modality Worklist Server DICOM AE is commanded to perform DICOM modality worklist query services through the use of the Discovery user interface 6 2 1 Application Data Flow Diagram The Basic and Specific Application models for this device are shown in the following illustration Association Initiation Modality Worklist Modality Worklist SCU Remote Worklist SCP Server AE DICOM Standard Interface DOC0550678 85 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 2 2 Functional definitions The Modality Worklist Server AE is implemented as an application process on the scanner host computer It runs as a daemon serving requests from the user interface to obtain modality worklists query remote AE s and return the results to the user interface The Modality Worklist Server AE initiates the following functions Query Initiates a DICOM association in order to query a remote AE If the remote accepts a presentation context applicable to modality worklist the Modality Worklist Server AE will issue a modality worklist query request via the C FIND service 6 2 3 Sequenci
185. it VR Big Endian 1 2 840 10008 1 2 2 2 3 1 2 1 2 1 SOP Specific Conformance Statement C_STORE SCU This implementation can perform multiple C STORE operations over a single association Upon receiving a C STORE confirmation containing a successful status this implementation will perform the next C STORE operation The association will be maintained if possible Upon receiving a C STORE confirmation containing a Refused status this implementation will terminate the association Upon receiving a C STORE confirmation containing any status that is not Success or Refused this implementation will consider the current request to be a failure but will continue to attempt to send the remaining images in the request on the same association Each C STORE operation supports an Association Timer This timer starts when the association request is sent or received and stops when the association is established The time out is 60 seconds Each C STORE operation also supports an Operation Inactivity Timer This time out starts once the first C STORE request has been issued on association or received and is reset each time a C STORE response has been received or when subsequent C STORES are sent This time out is 300 seconds Each C STORE operation also supports a Session Timer This timer starts when the association is established and stops when the association is ended This time out is 60 minutes DOC0550678 58 of 215 GE Heal
186. length and that the data for all attributes is consistent with respect to the attributes value representation VR Any inconsistencies in the response data with respect to the categories described above are considered errors Upon detecting any such errors in the response data the Worklist Server AE will issue a C FIND CANCEL and upon receipt of a C FIND RSP or if an applicable timer expires will abort the association All previously received worklist items are retained Note that the absence of requested Type 3 attributes is not considered an error Fields considered Type 1 by the Worklist Server include 0010 0010 Patient Name 0010 0020 Patient ID 0020 000D Study Instance UID 0040 0001 Scheduled Station AE Title DOC0550678 89 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 0040 0002 Scheduled Procedure Step Start Date 0040 0003 Scheduled Procedure Step Start Time 0040 0009 Scheduled Procedure Step ID 0040 1001 Requested Procedure ID 1 Start Date must be of the form YYYYMMDD exactly eight numeric characters and Start Time must be of the form HHMMSS exactly six numeric characters Fields considered Type 2 by Worklist Server include 0008 0050 Accession Number 0008 0060 Modality 0008 0090 Referring Physician Name 0010 0030 Patient Date of Birth 0010 0040 Patient Sex 0010 1030 Patient
187. m DICOM services through the buttons and menu selections on the main user interface panel 3 2 1 Application Data Flow Diagram The media intechange application model for the Discovery and Optima systems is shown in the following Illustration Illustration 3 2 SYSTEM MEDIA INTERCHANGE APPLICATION MODEL AND DATA FLOW DIAGRAM Display Directory QUERY Media Storage Media Storage DICOM lt Media Interchange Local Storage RESTORE FSR Record Media RECORD The DICOM Media Interchange Application Entity AE handles the DICOM CREATE RESTORE functionality for the CD DVD and USB storage media The DICOM Media Interchange AE is commanded by the user to perform DICOM services operating on the DOC0550678 55 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement Direction 5330620 1EN DICOM media through the use of buttons and menu selections on the graphical user interface of the platform The DICOM Media Interchange AE has a local storage that may contain various SOP instances These may have been obtained by original creation network DICOM or proprietary or by removable media using other application entities These instances are external to this conformance claim and the origin of SOP instances is outside the scope of this claim The Media Creator initializes Media by acting as an FSC to create a new DICOM File set on a 700MB CD 4 7GB DVD USB
188. meric characters in YYYYMMDD format DOC0550678 96 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Scheduled Procedure Step Start Time 0040 0003 Modality 0008 0060 Scheduled Performing 0040 0006 Physician s Name Scheduled Procedure 0040 0007 Step Description Scheduled Station Name 0040 0010 Scheduled Procedure 0040 0011 Step Location Scheduled Protocol Code 0040 0008 Sequence Code Value 0008 0100 Coding Scheme 0008 0102 Designator Code Meaning 0008 0104 Pre Medication 0040 0012 Scheduled Procedure 0040 0009 Step ID Requested 0032 1070 Contrast Agent 6 8 10 Requested Procedure Entity 6 8 10 1 Requested Procedure Module Table 6 8 10 1 1 Requested Procedure Module Attributes Attribute Name Tag Expected Matching Key Type B Expected Returned Key Type Requested Procedure ID Requested Procedure 0032 1060 Description Requested Procedure Code 0032 1064 Sequence Code Value 0008 0100 Coding Scheme Designator 0008 0102 DOC0550678 97 of 215 Direction 5330620 1EN This attribute is sent with zero length Returned values must be exactly 6 numeric characters in HHMMSS format Matching is supported as follows either no Modality is supplied universal matching or the scanner s Modality is supplied for matching this is user selectable This attribute is sent with zero length
189. mp UDF DVD ROM NO ISO 9660 YES ISO 9660 amp UDF USB Storage YES VFAT YES VFAT VFAT file system 3 2 2 Functional Definitions of AE s 3 2 3 3 2 4 The DICOM Media Interchange AE supports the following functions Generate and write a DICOM File Set FSC in a one shot activity SAVE Read a DICOM File Set FSR on an Interchange Media QUERY It can copy SOP instances from the media onto local storage RESTORE Sequencing of Real World Activities Sequencing of DICOM Media Interchange AE Real World Activities The display function QUERY can only be performed on a piece of media that already has a DICOM File set created With no SOP instances having been added the directory will be displayed empty The save function can only be performed on a blank unused Interchange Media Updates to an already recorded Interchange Medium are not supported There are no other sequencing requirements File Meta Information for Implementation Class and Version The File Meta Information for this implementation is File Meta Information Version 1 Implementation Version Name Software Revision See Appendix E The table in Appendix E identifies the Implementation UID for this product version DOC0550678 57 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 3 3 AE Specifications 3 3 1 DICOM Storage Media
190. n 4 gt CONTAINS CODE EV 113820 1 M Sent DOM CT Acquisition Type 5 gt CONTAINS CODE 232 SRT 1 U Not sent Procedure Context 6 gt CONTAINS UIDR EV 113769 1 M Sent 139 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN EF DCM Irradiation Event UID 7 gt CONTAINS NUM EV 113821 1 U Not sent DOM X ray Filter Aluminium Equivalent 8 gt CONTAINS CONT EV 113822 1 M Sent AINE CT R Acquisition Parameters 9 gt gt CONTAINS NUM EV 113824 1 M Sent DCM Exposure Time 10 gt gt CONTAINS NUM 113825 1 M Sent DCM Scanning Length 11 CONTAINS NUM 113826 1 M Sent DCM Nominal Single Collimation Width 12 gt gt CONTAINS NUM 113827 1 M Sent DCM Nominal Total Collimation Width 13 gt gt CONTAINS EV 113828 1 MC Sent per conditional DCM Pitch Factor 14 gt gt CONTAINS NUM EV 113823 1 M Sent DCM Number of X ray Sources 15 gt gt CONTAINS CONT EV 113831 1 n M Single item sent AINE CT X ray R Source Parameters 16 gt gt gt CONTAINS EV 113832 1 M Sent DCM Identification Number of the X ray Source 17 gt gt gt CONTAINS NUM EV 113733 1 M Sent DCM KVP 18 gt gt gt CON
191. n PPS will not be sent along with these images The DICOM Modality Gray Scale Presentation State service is provided by the Exam Split The CT DICOM AE is commanded to perform transmission of Gray Scale Presentation State object through the user interface The CT DICOM AE is commanded to perform transmission of Hard Split Images through the user interface 122 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 2 1 Application Data Flow Diagram The basic Application models for the feature are shown in the following illustration Operator associates images Scanner receives MWL Requested display settings Window Level Procedures from HIS RIS Window Width Pan Zoom and Image Orientation with a requested procedure Operator selects multiple requested procedures and scans images as a single exam PPS object PPS Manager After post processing Exam Split application presents user selected images and Requested procedures for split to create PPS and GSPS objects GSPS Remote Viewer 8 2 2 Functional Definition of AEs 8 2 3 Sequencing of Real World Activities 8 2 3 1 GSPS Acquisition System with MWL data The system has a Modality Work list Server AE installed Work List information is obtained from HIS RIS system through the use of Basic Work list Management Service Use of the information retrieved in the creation of
192. n context applicable to Modality performed Procedure Step the PPS Server AE will issue a request to update the SOP instance in the remote AE via the N SET service The PPS Status is set to COMPLETED Discontinue PPS Initiates a DICOM association in order to update a DICOM Modality Performed Step instance that is already created with the remote AE If the remote AE accepts a presentation context applicable to Modality performed Procedure Step the PPS Server AE will issue a request to update the SOP instance in the remote AE via the N SET service The PPS Status is set to DISCONTINUED 7 2 3 Sequencing of Real World Activities 7 2 3 1 PPS from Acquisition System with MWL data Note In the Discovery 710 610 and Optima 560 Scanner Products PPS Feature is available for the CT PET NM Modalities Any reference to PPS below and or throughout this Conformance Statement Document are specific both CT and to PET Images from the Discovery 710 610 and Optima 560 systems The system has a Modality Work list Server AE installed Work List information is obtained from HIS RIS system through the use of Basic Work list Management Service Use of the information retrieved in the creation of Image SOP instance is described in the Modality Work list Conformance statement Use of the information retrieved in MPPS SOP instances is described later in this document The system initiates a Start PPS before starting a scan i e when the image acq
193. nce 0032 1064 gt Scheduled Procedure Step Description 0040 0007 gt Scheduled Action Item Code Sequence 0040 0008 gt Scheduled Procedure ID 0040 0009 gt Requested Procedure ID 0040 1001 133 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 10 Image Header Changes Supporting GSPS 8 10 1 Request Attributes Sequence The 0040 0275 Request Attributes Sequence has been expanded to include dicom tags that the VES application requires to complete the VES GSPS and PPS This sequences now contains the following elements 0040 0275 Request Attributes Sequence gt 0008 0050 Accession Number gt 0008 1110 Reference Study Sequence gt gt 0008 1150 Referenced SOP Class UID gt gt 0008 1155 Referenced SOP Instance UID gt 0020 000D Study Instance UID gt 0032 1060 Requested Procedure Description gt 0032 1064 Requested Procedure Code Sequence gt gt 0008 0100 Code Value gt gt 0008 0102 Coding Scheme Designator gt gt 0008 0104 Code Meaning gt 0040 0007 Scheduled Procedure Description gt 0040 0008 Scheduled Action Item Code Sequence gt gt 0008 0100 Code Value gt gt 0008 0102 Coding Scheme Designator gt gt 0008 0104 Code Meaning gt 0040 0009 Scheduled Procedure Step ID gt 0040 1001 Requested Procedure ID The Requested Attributes Sequ
194. nce Number 0020 0013 1 1 If present in composite object instance it will be set to same value otherwise computed SOP Instance UID 0008 0018 E E If present in composite object instances it will be set to same value otherwise sent empty Referenced Series 0008 1115 1C 1C Not used Sequence Content Label 0070 0080 1 1 If present in composite object instance it will be set to same value otherwise not sent Content Description 0070 0081 2 2 If present in composite object instances it will be set to same value otherwise not sent Presentation Creation 0070 0082 1 1 If present in composite object instances it Date will be set to same value otherwise not sent Presentation Creation 0070 0083 1 1 If present in composite object instances it 196 of 215 GE Healthcare REV 7 Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN Time will be set to same value otherwise not sent 0070 0084 Content Creator s Name 2 2 If present in composite object instances it will be set to same value otherwise not sent C 5 6 STRUCTURED REPORT Keys Key Tag Type Type Attribute Description for CD for DVD USB Specific Character Set 0008 0005 1C 1C ISO IR 100 Instance Number 0020 0013 1 1 If present in composite object instance it will be set to same value otherwise computed SOP Instance UID 0008 0018 E
195. nces with any value of Specific Character Set 0008 0005 As a Query SCU it will similarly accept response items with any value of Specific Character Set However it will display in the user interface only characters specified as within ISO IR 6 ASCII or the configured extended character set The product user interface will allow the user to enter characters from the console keyboard that is within ASCII or the configured extended character set If any such extended characters are included in SOP Instances or in query identifier matching fields the product will appropriately specify the extended character set in Specific Character Set 0008 0005 2 8 CODES AND CONTROLLED TERMINOLOGY The product uses coded terminology as defined below 2 8 1 Mapped Coded Terminology The product maps without change coded terminology values supplied in Modality Worklist Scheduled Procedure Steps into Image SOP Instance and Modality Performed Procedure Step attributes as described in Sections 6 and 7 2 8 2 PET Isotope and Patient Orientation Module The encodings used in formatting the sequences in the PET isotope and patient orientation module which are supplied into the Image SOP Instance are described in Appendix A 2 9 SECURITY PROFILES The product does not conform to any defined DICOM Security Profiles It is assumed that the product is used within a secured environment It is assumed that a secured environment includes at a minimum 1 Firewall
196. ned scheduled unit of service that is specified by the Procedure Plan for a Requested Procedure It specifies one or more Action Items events involving equipment i e imaging modality equipment human resources location and time i e start time stop time duration DOC0550678 95 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 6 8 4 2 Requested Procedure Entity Description A Requested Procedure is an instance of a Procedure of a given Procedure Type An instance of a Requested Procedure includes all of the items of information that are specified by an instance of a Procedure Plan that is selected for the Requested Procedure by the imaging service provider 6 8 4 3 Imaging Service Request Entity Description An Imaging Service Request is a set of one or more Requested Procedures selected from a list of Procedure Types An Imaging Service Request is submitted by one authorized imaging service requester to one authorized imaging service provider in the context of one Service Episode 6 8 4 4 Visit Entity Description A Visit is the context in which the treatment or management of an arbitrary subset of a Patient s medical conditions occurs A Visit is limited to the description of a Patient s activities at a single facility 6 8 4 5 Patient Entity Description A Patient is a person receiving or registered to receive healthcare services 6 8 5 Modality Worklist M
197. nen nnns 107 7 2 2 Functional Definition 8 111 00 202000000000009 nnne nnne nain 108 7 2 3 Sequencing of Real World Activities seen 108 7 2 3 1 PPS from Acquisition System with MWL data 108 7 2 3 2 5 from acquisition system without MWL 109 7 2 8 8 PPS from post processing 110 7 3 AE SPECIFICATION en vierte tto tiber cede ok a du Qe Se HAAG tee de qi ve ea 110 7 3 1 PPS Server AE 5 22 9 8 6 110 7 3 1 1 Association Establishment 111 7 3 1 1 1 General nuk S aaa 111 7 3 1 1 2 Number of Associations sett ua a 111 7 3 1 1 3 Asynchronous Nature lu u n U 111 7 3 1 1 4 Implementation Identifying information 111 7 3 1 2 Association Initiation Policy 111 7 3 1 2 1 Real World Activity Performed Procedure Step creation and update 111 7 3 1 2 1 1 Associated Real World 111 7 3 1 2 1 2 Proposed Presentation Context 112 7 3 1 3 Association Acceptance Policy sssssssssssssssssese
198. nformation model are supported Some optional O keys are also supported as described in the following tables Table 2 3 1 3 3 2 1 1 Supported study level keys Description Tag Usage Studydate 1 R 00080020 Matched Studytme 1 R 00080030 Matched Accessionnumber R 00080050 Matched Patient s name R 00100010 Matched Patientid 7 1 R 00100020 Matched Studyid 7 0020000 Matched Studydescripion 771 0 00081030 Returned Suited P 00090002 Returned DOC0550678 48 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Table 2 3 1 3 3 2 1 2 Supported series level keys Description 7 Type Usage Modality R 00080060 Matched Seriesnumber 7 R 0020001 Matched Series description O 0008 103 Manufacturer 0 0008 0070 Images in series 0 00201002 Table 2 3 1 3 3 2 1 3 Supported image level keys Description Type Tag Usage Image number R 90200018 Matched _ o 0006008 Returned O 0028000 Reed Columns o 0026001 Remes image poston o 00200032 Retumed image orientation O 00200037 Retuned Slice thickness O 00150050 Slice spacing o 00160088 Gamy t O 0084120 Remes Convolution keme O 00
199. ng of Real World Activities The user or the system initiates a modality worklist query as a modality worklist SCU to the modality worklist SCP with a given set of query parameters The modality worklist SCP returns responses that match the query parameters Items from the returned worklist responses are presented to the user A subset of the returned worklist responses is included in acquired DICOM images related to the responses 6 3 AE Specifications 6 3 1 Worklist Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU SOP Class Name SOP Class UID Modality Worklist Information Model FIND 1 2 840 10008 5 1 4 31 6 3 1 1 Association Establishment Policies 6 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for an association initiated by the DICOM Worklist Server is Maximum Length PDU 50 Kbytes The SOP Class Extended Negotiation is not supported DOC0550678 86 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN The maximum number of Presentation Context Items that will be proposed is 1 The user information Items sent by this product are e Maximum PDU Len
200. nt Entity Modules This section specifies the attributes which identify and describe the piece of equipment which produced the instance of the Private SOP class Attribute Name Tag Type VR VM Notes Manufacturer 0008 0070 2 LO 1 Software Versions 0018 1020 3 LO 1 Institution Name Manufacturer s Model Name D 1 5 Private Creator Identification Information Refer to section B 2 1 of this document as GEMS PETD 01 is used for this private IOD D 1 6 Discovery PET Patient Module Refer to section B 2 2 of this document D 1 7 Discovery PET Exam Module Refer to section B 2 3 of this document D 1 8 Discovery PET Scan Module Refer to section B 2 4 of this document D 1 9 Discovery PET Frame Module Refer to section B 2 5 of this document 200 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN D 1 10 General Modules This section defines the attributes which are required for proper functioning and identification of the associated SOP Instances They do not specify any semantics about the Real World Object represented by the IOD D 1 10 1 SOP Common Module Attributes Attribute Name Tag Type VR VM Notes SOP Class UID 0008 0016 1 UI SOP Instance UID Specific Character Set Instance Creation Date Instance Creation Time Instance Creator UID 0008 0014 3 UI a D 1 11 Discovery PET Raw Da
201. ntation UID for this product version DOC0550678 66 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 4 3 1 2 Association Establishment Policy Print SCU initiates the association with the Printer 4 3 1 2 1 Real World Activity 4 3 1 2 1 1 Associated Real World Activity Print The Film Composer allows the user to select printers and it also allows the user to drag and drop the images from the viewer application into the film It also allows the user to manipulate some print parameters like film format and number of copies to print When the user presses the Print Button the Film composer communicates this request to Print SCU which then tries to establish the association with the requested printer and transmits the images for printing 4 3 1 2 1 2 Proposed Presentation Contexts The Proposed Presentation Context Table for the Print SCU is as shown in following Table Extended eee Presentation Context Table Proposed Abstract Syntax Transfer Syntax Basic Grayscale TH DICOM Implicit NN NM 2 840 10008 1 2 Print Management VR Little Endian Meta SOP Class Transfer Syntax Basic Color Print 1 2 840 10008 5 1 1 18 DICOM Implicit 1 2 840 10008 1 2 None Management Meta VR Little Endian SOP Class Transfer Syntax Verification SOP 1 2 840 10008 1 1 DICOM Implicit 1 2 840 10008 1 2 Class VR Little Endian Transfer Syntax Printer SOP Class 1 2 840
202. ntroduction This Conformance Statement CS specifies the Discovery and Optima compliance to DICOM It details the DICOM Service Classes and roles that are supported by this product For the remainder of this document the term Discovery will also include the Optima product configurations unless specifically noted The Discovery products use DICOM services to import images for possible further analysis and or processing It also uses DICOM services to export images to other DICOM compliant machines User has the capability to create and display Secondary Capture images and transfer Secondary Capture images via DICOM The Secondary Capture option allows the user to capture and store in DICOM format the presentation state of images as they are displayed on the Image Display viewer This provides the user with a digital alternative to hardcopy filming important portions of the study can be captured and stored on a PACS or analysis workstation rather than captured on film The ability to Receive DICOM CT MR PET images onto the Discovery scanner allows users to co register and fuse the anatomical detail of the CT MR PET images with the Functional PET images through offline registration The PET CT images produced by Discovery are perfectly aligned and fusion registration will be seamless The Discovery DICOM implementation allows the user to send CT MR PET Images Secondary Capture image data generated through the acquisition system or received
203. of the local database browser All the images currently in the selected exam series will be sent to the selected remote archive node which is also the Storage commitment SCP using DICOM C STORE operations Just after all the images are transferred the commitment request will be sent on a separate association 5 3 1 2 1 2 Associated Real World Activity Auto Archive exam series The Auto archive are used to archive the exams series onto local archive media or the remote archive node which shall be a Storage Commitment SCP without manual interface If the default device selected for Auto Archive is a remote Storage Commitment SCP then all the images currently in the specified exam series will be sent to the selected Storage commitment SCP using C STORE operations On successful transfer of all the images the Storage Commitment request will be sent 5 3 1 2 2 Proposed Presentation Contexts The Proposed Presentation Context table for the DICOM Storage Commitment SCU is as shown in following Table Presentation Context Table Abstract Syntax Transfer Syntax Role Extended Storage Commitment 1 2 840 10008 1 20 1 DICOM Implicit VR Little 1 2 840 10008 1 2 SCU None Push Model SOP Class Endian Transfer Syntax The Storage Commitment SCU can send the commitment request for following DICOM SOP classes DOC0550678 79 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620
204. oned in section 8 2 3 Sequencing of Real World Activities Each of the real world activity results in either creating a new GSPS SOP instance at the remote SCP 8 2 4 2 1 2 Proposed Presentation Context Table The following table shows the proposed presentation contexts for the GSPS AE after any of the real world activity listed in section 7 2 3 Sequencing of Real World Activities is initiated Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Grayscale 1 2 840 10008 5 1 4 1 1 Explicit VR 1 2 840 10008 1 2 1 SCU None Presentation 11 1 Little Endian State Grayscale 1 2 840 10008 5 1 4 1 1 Explicit VR 1 2 840 10008 1 2 2 SCU None Presentation 11 1 Big Endian State Grayscale 1 2 840 10008 5 1 4 1 1 Implici plicit VR 4 2 840 10008 1 2 SCU None E Mee 11 4 Little Endian gt ate 8 2 4 2 1 2 1 SOP Specific DICOM Conformance Statement for GSPS SOP class If the remote AE does not support the proposed Presentation context an appropriate error message logged Only one association is opened at a time All the operations used by this SOP class support an association timer which is configurable The timer is started when a request association request N CREATE request or N SET request is send and stopped when the respective response is received The default time out value is 300 seconds All the operati
205. ons used by this SOP class support a Session Timer This timer is started when the association is established and stopped when association is ended The default time out value is 3000 seconds If any of the above timers expires the association is aborted and the operation in progress is considered FAILED 8 2 4 3 Association Acceptance Policy The GSPS AE does not respond to attempts by remote AE to open an association 125 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 3 Save State 8 3 1 Implementation Model Save State will create Grayscale Presentation State GSPS including user annotation 8 3 2 Application Data Flow Diagram The basic Application models for the feature are shown in the following illustration Scan a series Select the scanned series from image browser Associates images display settings Window Level Window Width Pan Zoom and Image Orientation with a requested Press Save State GSPS series is created with new series 8 3 3 Sequencing of Real World Activities 8 3 3 1 GSPS Acquisition System with MWL data e Operator takes some scans After the scan finished images shall be selected from image browser and select Viewer e Operator adds annotation or ROI on images change WW or WL zoom or roam images e After changing settings operator shall select Save State on Viewer menu Then GSPS object shall be creat
206. ormance 5 1 Introduction This section of the DICOM conformance statement specifies the Storage Commitment Push Model SOP class as user the optional attributes and service elements supported 5 2 Implementation Model 5 2 1 Application data flow diagram Illustration 5 2 1 shows the data flow diagram of Storage Commitment SCU There are two Real World activities that will cause the DICOM Server Application Entity DICOM Server AE to initiate an association to a remote DICOM entity that is a Service Class Provider SCP of the Storage Commitment SOP class Real World Activity for Discovery 710 610 and Optima 560 products SW Version pet mfk xx The Choose Archive save Option Real World activity consists of an operator choosing a remote DICOM AE that supports Storage Commitment as provider as the archive device and selecting one or more exam or series from the local database and choosing either Save Exam or Save Series from the archive menu The images to be committed are sent to the remote provider entity first The Commitment request for the transferred image instances is sent after the complete image transfer The Commitment response must come on a different association The Auto Archive Exam Series Real World activity archives the exams series onto a remote DICOM AE that supports Storage Commitment as provider without manual interface The images to be committed are sent to the remote provider entity first On successful transfer of all
207. ose the association It changes the Job state to Wait indicating the job is waiting for the response from commitment provider The job will be marked as failed if the response is not received by stop job time Stop job time is the maximum duration the job can wait for responses DOC0550678 80 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN A New transaction uid will be created for each retry by user The old transaction uid is not applicable for these requests The Following DIMSE service Elements are supported for the Storage Commitment request and processing N ACTION Requests the remote Storage Commitment SCP to commit to storing the image instances The following attributes are sent as part of the DATA Set for the N ACTION request Attribute Tag Transaction UID 0008 1195 Transaction UID Referenced SOP Sequence 0008 1199 SOP Instance UID 008 455 Referenced Study Component sequence attribute is not sent Storage Media File Set ID and Storage Media File Set UID attributes are not supported 5 3 1 3 Association Acceptance Policy 5 3 1 3 1 Real World Activity Storage Commitment Response and Set Archive State The remote storage provider will respond with a commitment success or failure The Set Archive State activity processes that response as described below 5 3 1 3 1 1 Accepted Presentation Context Table The Accepted Pre
208. ource 0054 1002 CS Units 0054 1001 CS Series Type 0054 1000 CS Reprojection Method 0054 1004 CS Number of R R Intervals 0054 0061 US For Gated Scan Number of Time Slots Number of Time Slices Number of Slices Corrected Image Randoms Correction Method Attenuation Correction Method Scatter Correction Method Decay Correction Reconstruction Diameter Convolution Kernel Reconstruction Method Detector Lines of Response Used Acquisition Start Condition Acquisition Start Condition Data Acquisition Termination Condition Acquisition Termination Condition Data Value equals to 1 Provided from 0x0009 0x10dd Number of R R Intervals always 1 if scan is gated For Gated Scan value equal to the number of bin intervals Provided from 0x0009 0x10de Number of Time Slots which is bins 0054 0071 c 0054 0081 0028 0051 0054 1100 0054 1101 0054 1105 3 054 1102 0018 1100 0018 1210 0054 1103 1 0054 1103 3 0054 1104 ips 0018 0073 0018 0074 JEN 0018 0071 e 0018 0075 160 215 e 5 5 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Field of View Shape 0018 1147 EXEC MN Field of View Dimensions 0018 1149 3 is 1 2 Gantry Detector Tilt 0018 1120 Gantry Detector Slew 0018 1121 Type of Detecto
209. overy 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Number of overranges 7 00434016 SS 1 0043 101E DS 0043 101F 55 U S D S 0043 1028 i S D eS US i SH i OB 0043 102B SS 0043 1031 DS 0043 1040 FL r TI 5 L S S S H S S L 0043 1043 L L L L FL FL M Degree of rotation 0043 1041 FL 0043 1044 0043 1045 0043 1046 0043 104D FL 0043 104E FL 0043 1064 B 1 9 Private Creator Identification GEMS_HELIOS_01 Note Dicom elements 0045 1030 1034 and 0045 1036 1039 and 0045 103B are present only if the appropriate cardiac option is Installed on the scanner Attribute Name Number of Macro Rows in Detector 0045 1001 Macro width at ISO Center 0045 1002 0045 1003 0045 1004 VM 0045 1005 SS FL 55 55 55 CFE SS 55 55 0045 100A Z tracking Flag 0045 1000 SS Mean Z error 0045 100E FL 0045 100F FL 0045 1011 Z smoothing Factor 0045 1007 CD O Cm SS Start View 1A 0045 1012 SS Sigma Mode 0045 1013 SS Number of Views 1A 0045 1014 SS 0045 1010 SS R SS FL 55 55 55 5 FL SS 55 F
210. pied from source SPS if selected image gt Scheduled Procedure 0040 0007 Copied from Not used Copied from source Step Description SPS if selected image gt Scheduled Protocol Code 0040 0008 Copied from Not used Not sent as part of Sequence SPS if selected image header Patients Name 0010 0010 Copied from User input on the Copied from source SPS if selected scanner image Patient ID 0010 0020 Copied from User input on the Copied from source SPS if selected scanner image Patient s Birth Date 0010 0030 Copied from User input on the Copied from source SPS if selected scanner image Patient s Sex 0010 0040 Copied from User input on the Copied from source SPS if selected scanner image Referenced Patient 0008 1120 Copied from Not used No item Sequence SPS if selected Performed Procedure Step ID 0040 0253 Created at the scanner Will have the Created at the scanner Will have the following New generated may not be unique 117 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Attribute Name Tag Usage in MPPS Instance status Acquisition Acquisition Post Processing with MWL data without MWL data following PPS_ID_ lt exam PPS_ID_ lt exam number number Performed Station AE Title 0040 0241 Local system Local system host Local system host host name
211. ppendix B for private data elements 3 5 2 Private Transfer Syntax Specification None specified 3 6 Configuration For the CD DVD USB Interchange Media Application the source AE Title encoded in the File Meta Information cannot be modified 3 7 Support of Extended Character Sets The DICOM Media Interchange AE will support copy of SOP instances containing the ISO IR 100 Latin alphabet No 1 supplementary set and DICOM default character sets as defined in PS3 5 Any incoming SOP instances encoded using another extended character set will not be installed in the GEHC PET CT system database 4 PRINT SCU CONFORMANCE 4 1 Introduction This conformance statement specifies the Discovery conformance to the DICOM Print SCU Application Profile The Discovery product uses DICOM Print SCU to print images on DICOM Compliant Printers 4 2 Implementation Model PRINT SCU The DICOM Print SCU is a DICOM print filter that provides the capability to print images to DICOM printers The DICOM Print filter acts as an SCU of the DICOM print management SOP class 4 2 1 Application Data Flow Diagram DICOM print SCU Implementation model DICOM INTERFACE DOC0550678 65 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 4 2 2 4 2 3 Direction 5330620 1EN DICOM Printer The Film Composer is the User interface and this is used to initiate the local real world activity The user issues the p
212. process is displayed to the user If the C ECHO response is received with a success the DICOM Server will post a pop up to the operator indicating that the remote device is alive 2 3 1 2 4 2 Proposed Presentation Context Table Refer to the Table 2 2 1 2 1 2 1 for the Proposed Presentation Contexts for DICOM Server AE and Real World activity Verification 2 3 1 2 4 2 1 SOP Specific DICOM Conformance ECHO SCU The Discovery DICOM Server AE provides standard conformance to the DICOM Verification Service Class DOC0550678 45 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Each ECHO operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 15 seconds and 60 minutes respectively 2 3 1 3 Association Acceptance Policy The DICOM Server AE places limitations on who may connect to it If the Remote AE needs to Push Images or Query Retrieve Images to the local system then it has to be configured in the Local system to do the same To configure a remote AE choose Select Remote Host in the network pull down menu Select the Remote Host and choose Update option if the Remote AE is already present If the Remote AE is not present choose Add option to add the Remote AE Set Push Images option to Yes if the Remote AE needs to push images to the Local system Set Query Retri
213. py Presentation State 1 2 840 10008 5 1 4 1 1 11 1 8 2 4 1 Association Establishment Policies 8 2 4 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU from an association initiated by the GSPS Server AE is Maximum Length PDU 128 Kbytes The SOP Class Extended negotiation is not supported The maximum number of presentation negotiation items that will be proposed is 1 The user information items sent by this AE are e Maximum PDU Length Implementation UID 8 2 4 1 2 Number of Associations The GSPS application will initiate only one DICOM association at any time to perform a GSPS operation to the remote AE 8 2 4 1 3 Asynchronous Nature Asynchronous mode is not supported All operations are performed synchronously 8 2 4 1 4 Implementation Identifying information The table in Appendix E identifies the Implementation UID for this product version 8 2 4 2 Association Initiation Policy The GSPS AE initiates a new association for every GSPS operation initiated 124 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 8 2 4 2 1 Real World Activity GSPS 8 2 4 2 1 1 Associated Real World Activity The real world activities are menti
214. r Manual Query The operator queries one or a set of remote DICOM databases to obtain a list of data at Study Series Image level by clicking on the icon that represents the wanted remote DICOM AE Manual Retrieve This Real World activity will be available once the Manual Query activity is performed The operator can now select one or more studies series or images and ask the DICOM Server AE to retrieve the selected image s from the Remote DICOM AE by clicking on the Local DB button at the bottom of the local database browser Remote Archive This Real World activity consists of an operator choosing a remote DICOM AE that supports Storage Commitment as provider from the archive panel at the bottom of the local database browser The operator chooses the exam or series in the local database browser and clicks on the archival destination from the archive list at the bottom of the local database browser There is no Real World activity required for the DICOM Server AE to respond to an incoming DICOM store query or retrieve The DICOM Server AE is always prepared to respond to a DICOM Store Query or Retrieve by any remote DICOM AE The DICOM Server AE will perform the Real World activity Image Installation after the remote AE sends an image to the Discovery product 55 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Once a Query request is received the DICOM S
215. r Identification GEMS PETD 01 Attribute Name Tag Type d VM PET compatible version 0009 10A4 3 1 PET software version 0009 10A5 1 PET slice number 0009 10A6 1 PET total counts 1 PET other atts 0009 10A8 1 PET other size 1 PET bp center y 0009 10AC 1 PET trans frame id 1 PET tpluse frame id 1 PET profile spacing 1 PET seg qc parm 0009 10C6 1 PET overlap 0009 10C7 1 PET ovlp trans frm id 1 PET ovlp tpulse frm id 1 PET ir z filter flag 0009 10DB 1 PET ir z filter ratio 1 PET posterior shift 1 PET superior shift 1 3D Filter flag 0009 10EA 1 189 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN PET tracerInjectionUID 0009 10F2 1 PET Q Static Scan Mode 0009 10F6 1 PET Q Static Frame 0009 10F7 1 Number of Detector Rows 1 Number of Detector Columns 1 Physio Gating Type 1 1 Total Number of Bins 0015 101B 190 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN APPENDIX C DICOMDIR Directory Information Enclosed here is a listing of only the optional conditional modules and optional attributes used by this implementation in the DICOMDIR definition All standard attributes as defined in Part 3 Addendum Basic Directory Information Object are supported by this implementation but not listed here C 1 Basic Directory IOD Definition Module
216. r Motion 0054 0202 Collimator 0018 1181 2 CS 1 Collimator Grid Name 0018 1180 3 SH 1 Axial Acceptance 00541200 3 DS 1 Axial Mash 00541201 3 Is 2 Transverse Mash 00541202 3 Is 1 Detector Element Size 0054 1203 Not used Coincidence Window Width 0054 1210 DS 1 Difference between Upper and Lower Coincidence limits Energy Window Range 0054 0013 SQ 1 Sequence gt Energy Window Lower Limit 0054 0014 DS 1 gt Energy Window Upper Limit 0054 0015 DS 1 Secondary Counts Type 0054 1220 CS 1 n Value equals DLYD if scan delay events is separate A 7 8 PET Isotope Module Attribute Name Tag Type VR VM Notes Radiopharmaceutical 0054 0016 2 SQ 1 Single item Information Sequence sequence gt Radionuclide Code 0054 0300 SQ Single item Sequence sequence gt gt Code Value 0008 0100 1C SH Code value is based on radionuclide name gt gt Coding Scheme Designator 0008 0102 SRT gt gt Code Meaning 0008 0104 3 LO 1 Radionuclide name 161 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN gt Radiopharmaceutical Route 0018 1070 Not Used gt Administration Route Code 0054 0302 3 SQ 1 Not Used Sequence gt gt Code Value 0008 0100 e Not Used gt gt Coding Scheme Designator 0008 0102 1C SH 1 NotUsed
217. ract Syntax Transfer Syntax Extended Negotiation Name List UID List 1 2 840 10008 1 1 Implicit VR Little Endian 1 2 840 10008 1 2 0 0530678 47 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 2 3 1 3 2 2 1 SOP Specific Conformance to C ECHO SCP The DICOM Server AE provides standard conformance to the DICOM Verification Service Class Each ECHO operation supports an Operation Inactivity Timer with time out value of 15 seconds 2 3 1 3 3 Query Request from Remote AE This AE is indefinitely listening for associations No operator action is required to respond to a query request 2 3 1 3 3 1 Associated Real World Activity The Real World Activity associated with the query request is to search the local database for entries that match the request and send a C FIND response message with a status of pending for each matching entry 2 3 1 3 3 2 Presentation Context Table Table 2 3 1 3 3 2 1 Acceptable Presentation Contexts for DICOM Server AE and Real World Activity Query Request Presentation Context Table Extended Negotiation Abstract Syntax Transfer Syntax UID UIDList__ Study Root 1 2 840 10008 5 1 4 1 2 2 1 Implicit VR 1 2 840 10008 1 2 Query Retrieve Little Endian FIND 2 3 1 3 3 2 1 SOP Specific Conformance to C FIND SCP All Required R and Unique U study series and image level keys for the Study Root Query Retrieve i
218. re sent or if DICOM Server AE receives a C CANCEL MOVE RQ a final C STORE RSP will be sent back with an appropriate status 2 2 3 Sequencing of Real World Activities Real World Activity Query Remote must be performed before Choose Get Option can be performed 2 3 AE Specifications 2 3 1 DICOM Server AE Specification This Application Entity provides Standard Conformance to the following DICOM SOP Classes as an SCU 1 2 840 10008 5 1 4 1 1 4 Note This Application Entity also supports GSPS as an SCU Refer to section 8 for further details The Discovery 710 610 and Optima 560 use the GE PET Raw SOP Class for local storage and management of list data Unlike other types of GE PET Raw such as sinogram and spectra data list data is not supported in network operations such as send and receive and the list data objects will not be visible via Remote AE DICOM Servers Remote hosts List data objects are labeled as List on the database browser and can be identified by tag 0008 0060 see Section D 2 This Application Entity provides Standard Conformance to the following DICOM SOP classes as an SCP Secondary Capture Image Storage 1 2 840 10008 5 1 4 1 1 7 Study Root Query Retrieve FIND 1 2 840 10008 5 1 4 1 2 2 1 Study Root Query Retrieve MOVE 1 2 840 10008 5 1 4 1 2 2 2 DOC0550678 55 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN X Ray Radia
219. returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user System includes the necessary information related to Scheduled Procedure Steps and the Performed procedure Step in the image instances created At the end of image acquisition system initiates a Complete PPS or Discontinue PPS based on the choice selected by the user using the user interface provided The user is also given a choice Defer PPS which is described below PPS Server AE initiates a MPPS 109 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN N SET request to the remote AE in order to update the MPPS SOP instance that is already created The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user At the end of image acquisition if the user has chosen Defer PPS the user is provided with an interface to Complete PPS or Discontinue PPS at any later time The user might wish to alter the image set generated through acquisition before invoking these operations Note that the user explicitly uses the user interface provided to invoke this operation as in the case of PPS generated for post processing which is described in the following sect
220. ributes within an Information Object Definition that are logically related to each other Example Patient Module includes Patient Name Patient ID Patient Birth Date and Patient Sex Negotiation first phase of Association establishment that allows Application Entities to agree on the types of data to be exchanged and how that data will be encoded Presentation Context the set of DICOM network services used over an Association as negotiated between Application Entities includes Abstract Syntaxes and Transfer Syntaxes Protocol Data Unit PDU a packet piece of a DICOM message sent across the network Devices must specify the maximum size packet they can receive for DICOM messages Security Profile a set of mechanisms such as encryption user authentication or digital signatures used by an Application Entity to ensure confidentiality integrity and or availability of exchanged DICOM data Service Class Provider SCP role of an Application Entity that provides a DICOM network service typically a server that performs operations requested by another Application Entity Service Class User Examples Picture Archiving and Communication System image storage SCP and image query retrieve SCP Radiology Information System modality worklist SCP Service Class User SCU role of an Application Entity that uses a DICOM network service typically a client Examples imaging modality image storage SCU and modality worklis
221. rint request using Film Composer Film composer allows printer selection and it composes the pre formatted film file This film file is interpreted by Print SCU and it sends the appropriate messages to DICOM print SCP running on a DICOM printer Functional Definition of AE s DICOM Print SCU Establishes the Association to the requested printer for printing the composed film Sequencing of Real World Activities The DICOM printer is installed through a camera installation process The DICOM printer is selected from the Film Composer Interface for Manual Filming Auto filming is enabled using the camera setup via the camera installation process Images may be acquired for printing in the following manner Images to be printed may be manually dropped from Viewing applications into the manual film composer slots Images to be printed may be automatically dropped from Viewing applications into the manual film composer slots Images to be printed may be manually dropped from Viewing applications into the auto film composer slots Image series to be printed be automatically dropped from the PrintSeries application into the manual film composer slots Images to be printed may be automatically dropped from the Scanning Application into the auto film composer slots Based upon the filming mode and the attributes used to acquire images into the film composer films will be automatically printed when the film session is full or
222. rns a Storage Full condition this status will be returned This status will only be sent if the Destination AE returns a status of A711 and is only applicable if the Destination AE is another GE Healthcare product 2 4 Communication Profiles 2 4 1 2 4 2 Supported Communication Stacks part 8 DICOM Upper Layer Part 8 is supported using TCP IP TCP IP Stack The TCP IP stack is inherited from a UNIX Operating System DOC0550678 51 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN 2 4 2 1 Physical Media Support Ethernet v2 0 IEEE 802 3 Auto senses 10 100 Base T Ethernet connection 2 4 3 Additional Protocols DHCP is not supported in this product 2 4 4 IPv4 and IPv6 Support IPv4 and IPv6 is supported by this product 2 5 Extensions Specializations Privatizations 2 5 4 Standard Extended Specialized Private SOP Classes 2 5 1 1 Standard Extended Specialized Private SOP Classes The product provides Standard Extended Conformance to all supported SOP Classes through the inclusion of additional Type 3 Standard Elements and Private Data Elements The extensions are defined in Sections Please refer section A 4 4 for CT IOD 2 5 1 2 Private SOP Class GE PET Raw Refer to Appendix B amp Appendix D for a complete listing of private data elements used with this implementation 2 6 Configuration 2 6 1 AE Title Presentation Address Mapping
223. rocedure Entity Description 94 6 8 4 3 Imaging Service Request Entity 94 6 8 44 Visit Entity Description eee ie cereo e ge E e are ce e n de 94 6 8 4 5 X Patient Entity Description 4 1000 nennen a 94 6 8 5 Modality Worklist Mapping of DICOM Entities 20 202000000000000 94 6 8 6 Information Model Module Table 2 enne 94 6 8 7 Information Model Keys u al eise nna ann 95 6 8 8 Supported 99 96 6 8 9 Scheduled Procedure Step Entity 96 6 8 9 1 SOP Gommon ModUule 32 4 1 5 ea dori ded dee ese dran RR aede it Bo onde 96 6 8 9 2 Scheduled Procedure Step n nnne nenne 96 6 8 10 Requested Procedure Entity cs ed gestita 97 6 8 10 1 Requested Procedure 2 4 2 1 2 5 10000000000 eene inset nene 97 6 8 11 Imaging Service Request Entity 98 6 8 11 1 Imaging Service Request 98 6 8 12 IT 99 6 8 12 1 Visit Identification nensis sitne ne 99 068 122 Visit Statistiche ie e ie dnd 99 6 8 12 3 Visit Rel
224. rver AE does not define any extended error codes The standard error codes are handled On a response with status success for the N CREATE or N SET request the system updates the state and indicates the same on the user interface On a response with status other than success the operation is deemed Failed and the system updates the state and indicates the same on the user interface If the request has failed or response is not received before the association timeout the operation is deemed Failed and the system updates the state and indicates the same on the user interface If the operation is Failed detailed message is logged into system log file and system provides an alternative mechanism to retry the failed operation through the user interface to ensure that transient failures do not affect the feature performance 116 of 215 GE Healthcare REV 7 Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN 7 10 Use of Specific DICOM Data The following table gives specific usage of some of attributes in the MPPS SOP instance created with reference to each of the real world scenarios mentioned in 7 2 3 of this document Attribute Name Tag Usage in MPPS Instance Acquisition with MWL data Acquisition without MWL data Post Processing Specific Character Set 0008 0005 Not used but copied into image header Not used Not used but copied into image header S
225. s Description Requested Procedure Code Meaning If Code Meaning absent it will be Requested Procedure Description Performed Procedure Type Description 0040 0255 Set as empty Performed Protocol Code Sequence 0040 0260 Set as Empty sequence 131 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Performed Series Sequence 0040 0340 Exam Split Application will calculate and copy from Image s Study and set these sequence values Rescale Type 0028 1054 Exam Split Application will set this value Referenced Performed Procedure Step Sequence 0008 1111 Exam Split Application will calculate and set these sequence values Manufacturer 0008 0070 Exam Split Application will set as GE MEDICAL SYSTEMS Instance Number 0020 0013 Exam Split Application will set as 24 Presentation label 0070 0080 This will be set as Requested Procedure Code Meaning or Requested Procedure Description Presentation Description 0070 0081 Exam Split Application will set as Exam Split Presentation Create Date 0070 0082 Exam Split Application will calculate Presentation Create Time 0070 0083 Exam Split Application will calculate Presentation creators name 0070 0084 Exam Split Application will set
226. s In this Service Class the Information Model plays a role similar to an Information Object Definition of most other DICOM Service Classes DOC0550678 92 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 6 8 3 Modality Worklist Information Model Entity Relationship Model Direction 5330620 1EN The Entity Relationship diagram for the Modality Worklist Information Model schema is shown in Illustration 6 8 3 1 It represents the information that composes a Worklist Item In this figure the following diagrammatic convention is established to represent the information organization Each entity is represented by a rectangular box Each relationship is represented by a diamond shaped box The fact that a relationship exists between two entities is depicted by lines connecting the corresponding entity boxes to the relationship boxes Illustration 6 8 3 1 Modality Worklist Information Model E R DIAGRAM Scheduled Worklist Procedure Step Item contained in Requested Procedure w requested for 1 Imaging Service Request done for 6 8 4 Entity Descriptions Please refer to DICOM Standard PS 3 3 Information Object Definitions and PS 3 4 Service Class Specifications for a description of each of the Entities contained within the Modality Worklist Information Model 6 8 4 1 Scheduled Procedure Step A Scheduled Procedure Step is an arbitrarily defi
227. s 165 A 7 10 NM PET Patient Orientation Module 165 A 7 11 Discovery PET Image Set Module eene 166 A 7 12 Discovery PET Scan Module sisse eee enne enne nnns 166 A 7 13 Frame of Reference Module 1 110200000000000000000000 166 A 7 14 General Equipment Module nnne nnn 166 A 7 15 General Image Module eiii nennen anita saneta 167 A 7 16 Image Plane Module enian 168 A 7 17 Image Pixel Module eie e E eget de caer eoa eom dd 168 A 7 18 PET Image Module eon e te oe enm ans 169 A 7 19 Overlay Plane Module essi senses aine enenatis anna 172 A 7 20 VOLLUT s a eee pe eee 172 7 21 Discovery PET Image 172 7 22 Discovery PET Frame 172 A 7 23 SOP COMMON EE E te iet HERFORD eda be 173 APPENDIX B PRIVATE DATA ELEMENTS U 174 1 CT IMAGE IOD PRIVATE DATA ELEMENTS 4 emnes 174 B 1 1 Private Creator Identification GEMS IDEN 01 174 B 1 2 Private Creator Identification GEMS ACQU 01
228. s only generated images from the Motion Free Application specifically the Advantage4D application version which runs on the CT PET scanner B 2 PET Image IOD Private Data Elements Definition B 2 1 Private Creator Identification GEMS_PETD_01 Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM Private Creator Data Element 0009 0010 1 SH 1 GE Discovery PET Implementation 0009 1001 3 LO 2 Version Name B 2 2 Discovery PET Patient Module Private Creator Identification GEMS PETD 01 Attribute Name Tag S PET Implementation Name 0009 1001 PET patient id 0009 1002 PET compatible version 0009 1003 wam PET patient_datetime 0009 1005 EL 178 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN IsPatientDiabetic 0009 10F4 1 Date of Last Treatment 0009 10F5 1 B 2 3 Discovery PET Exam Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET exam id 0009 1007 UI 1 PET compatible version 0009 1008 PET software version 0009 1009 B 2 4 Discovery PET Scan Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VM PET scan id 0009 100A UI 1 PET compatible version 1 H PET software version 0009 100C 1 PET scan datetime 1 PET scan description 1 PET hospital name 1 PET scanner desc 1 PET man
229. se set to NULL Performing Physician s Name 0008 1050 2 If present in composite object instances it will be set to same value other wise set to NULL Series Type 0054 1000 2 If present in composite object instances it will be set to same value other wise set to NULL C 5 4 Image Keys Character Set 0008 0005 r 100 _ Number 0020 0013 uM present in composite object instances it will be set to same value other wise set to NULL Icon Image Sequence 0088 0200 3 Note supported 022 SOP Class UID 0008 0016 1 If present in composite object instances it will be set to same value other wise set to NULL 194 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Slice Thickness 0018 0050 2 If present in composite object instances it will be set to same value other wise set to NULL Repetition Time 0018 0080 2 If present in composite object instances it will be set to same value other wise set to NULL Echo Time 0018 0081 2 If present in composite object instances it will be set to same value other wise set to NULL Inversion Time 0018 0082 2 If present in composite object instances it will be set to same value other wise set to NULL Number of Averages 0018 0083 2 If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set
230. select the eject button on the User Interface for the appropriate drive A user cannot restore CD DVD USB while create CD DVD USB is in process A user can only restore selected composite objects at a time from a media any other attempt of selections to restore CD DVD USB on media in same drive will wait until the first one is completed 3 3 1 2 3 1 Application Profiles for the RWA Local Storage RESTORE CD USB For the list of Application Profiles that invoke this AE for the Restore RWA see Table 3 3 1 1 For extensions and specialization s see section 3 5 3 3 1 2 3 1 1 Options for STD GEN CD and STD GEN USB JPEG Application Profile Following are the SOP Classes supported by the RWA Local Storage RESTORE CD USB Information Transfer Syntax UID Object SOP Class UID Transfer Syntax DOC0550678 61 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Definition Media Storage 1 2 840 10008 1 3 10 Explicit VR Little Endian 1 2 840 10008 1 2 1 Directory Storage See Table See Table Explicit VR Little Endian 3 3 1 2 1 1 1 1 3 3 1 2 1 1 1 1 1 2 840 10008 1 2 1 See Table See Table JPEG Lossless Process 14 3 3 1 2 1 1 1 1 3 3 1 2 1 1 1 1 selection value 1 1 2 840 10008 1 2 4 70 STD GEN USB JPEG only See Table See Table JPEG Lossy Baseline 3 3 1 2 1 1 1 1 3 3 1 2 1 1 1 1 Sequential with Huffman Coding 1 2 840
231. sentation Context table for the DICOM Storage Commitment SCU is shown in the following table Accepted Presentation Context Table Abstract Syntax Transfer Syntax ii Extended Negotiation Storage Commitment Implicit VR Little Endian 1 2 840 10008 1 2 Push Model 1 2 840 10008 1 20 1 Explicit VR Little Endian 1 2 840 10008 1 2 1 None 5 3 1 3 2 Storage Commitment Push Model SOP Class Response Processing The following DIMSE Service Elements are supported for the Storage Commitment response processing N EVENT REPORT The Response sent by the remote Storage Commitment SCP DOC0550678 81 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Once the N EVENT REPORT is received the following actions will be taken depending on the status of the response 5 3 1 3 2 1 Commit response with SUCCESS status The Archive flag information in the browser for all the successful instances will be updated The status will be changed to Y The job queue entry will be removed The following attributes are expected as part of DATA Set for N Event REPORT from SCP Attribute Tag Transaction UID 0008 1195 Value received from SCP Referenced SOP Sequence 0008 1199 Value received from SCP SOP Class UID 0008 1150 Value received from SCP SOP Instance UID 0008 1155 Value received from SCP N EVENT REPORT RSP will be sent on the same association itself No DATA Set will b
232. spiral scan only Single Collimation Width 0018 9306 Sent for spiral scan only Total Collimation Width Total Collimation Width Scan Options 0018 0022 AXIAL MODE SCOUT MODE AXIAL XRON MODE AXIAL XROFF MODE STATIC XRON MODE STATIC XROFF MODE TUBE HEAT MODE DAS MODE TUBE CAL MODE 152 of 215 10 IQ yyy IKMIV ius VEN ici Vir IVA VAI yg ID Viti IM CAUTION the table height while scanning a series of CT images Therefore GE Healthcare implementations must use the Fr REV7 conjunction with the Table Height MODERATE need peq CoO Statement Direction 5330620 1EN images are spatially related For PET CT hybrid scans scanning the CT Image IOD specific specializations CT Image implementation specific i I 153 of 215 BIOPSY MODE CINE HELICAL ROTGENCAL MODE FLUORO MODE AXIAL LOCALIZER SEGMENTED REFORMATTED PROCESSED COMBINED CTINTERVENTION MIN IP MIP AVERAGE VOLREN INTEGRAL HD MIP RAYSUM SURFACE MINMAX FLUORO GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN Decsciption of how to calculate the Dicom Exposure field 0018 1152 prescribed image coverage lt gt 222 avg mA for image mA image thickness f views to make image Total coverage J exposure exposure time image avg mA slice th
233. ss 128 8 6 2 Configurable 2 128 87 SUPPORT OF EXTENDED CHARACTER 5 5 128 88 ERRORHANDLING AND ee 128 89 USE SPECIFIC DICOM DATA nct ied rt eut Lenta 129 8 10 IMAGE HEADER CHANGES SUPPORTING GSPS sse eere 134 8 10 1 Request Attributes Sequence 0101020020000000404680000200 134 8 10 2 Accession Number e x t ee a RP E EP ADR DR ge anoo 134 9 CT X RAY RADIATION DOSE STRUCTURED REPORT OBJECT IMPLEMENTATION 135 9 1 JOD MODULE TABLE cm 135 9 2 SRDOCUMENT SERIES 135 9 3 SRDOCUMENT GENERAL 135 9 4 CONTENT 137 9 4 1 TID 10011 CT Radiation 2 22 0 000909 137 9 4 2 TID 10012 CT Accumulated Dose 0 138 9 4 3 TID 10013 CT Irradiation Event Data sss 139 9 55 cGONFIGURATION IB Kn ede dE 142 APPENDIX A CT PET IMAGE AND SECONDARY CAPTURE MODULES ATTRIBUTES 143 CT SPEGIFIC IOD DEFINITION s 22 i pde reto e ee cedi eee eee ee nus 143 A 1 1 CT image OD Modules ith nel nS 143 A2 SCOIMAGEIOD iiie qhaqya
234. t SCU imaging workstation image query retrieve SCU Service Object Pair SOP Class the specification of the network or media transfer service of a particular type of data object the fundamental unit of DICOM interoperability specification Examples Ultrasound Image Storage Service Basic Grayscale Print Management Service Object Pair SOP Instance an information object a specific occurrence of information exchanged in a SOP Class Examples a specific x ray image Tag a 32 bit identifier for a data element represented as a pair of four digit hexadecimal numbers the group and the element If the group number is odd the tag is for a private manufacturer specific data element Examples 0010 0020 Patient ID 07 0010 Pixel Data 0019 0210 private data element Transfer Syntax the encoding used for exchange of DICOM information objects and messages Examples JPEG compressed images little endian explicit value representation Unique Identifier UID a globally unique dotted decimal string that identifies a specific object or a class of objects an ISO 8824 Object Identifier Examples Study Instance UID SOP Class UID SOP Instance UID DOC0550678 29 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN Value Representation VR the format type of an individual DICOM data element such as text an integer
235. t Length 4 2 16 1 Item 0008 1150 UI no value available 0 0 ReferencedSOPClassUID 0008 1155 UI no value available 0 0 ReferencedSOPInstanceUID fffe e00d na ItemDelimitationItem for re encoding f 0 1 ItemDelimitationItem 0 na SequenceDelimitationItem for re enc 0 1 SequenceDelimitationItem 0010 0000 UL 96 4 1 PatientGroupLength 0010 0010 PN no value available 0 0 PatientName 0010 0020 LO no value available 0 0 PatientID 0010 0030 DA no value available 0 0 PatientBirthDate 0010 0040 CS no value available 0 0 PatientSex 00100 1000 LO no value available 0 0 Other Patient IDs 0010 1020 DS no value available 0 0 PatientSize 0010 1030 DS no value available 0 0 PatientWeight 0010 1040 LO no value available 0 0 PatientAddress 0010 2000 LO no value available 0 0 MedicalAlerts 00100 2110 LO no value available 05 0 ContrastAllergies 00100 2154 SH no value available 0 0 PatientTelephoneNumber 0010 2160 SH no value available 0 0 Ethinc Group 0010 2150 LT no value available 0 1 AdditionalPatientHistory 0010 21c0 US no value available 0 0 PregnancyStatus 0010 4000 LT no value available 0 0 Patient Comments 10020 0000 UL 8 4 1 ImageGroupLength 0020 000d U no value available 0 0 StudyInstanceUID 0032 0000 UL 64 4 1 StudyGroupLength 0032 1032 PN no value available 0 0 RequestingPhysician 0032 10
236. t name of scanner Local IP address Local IP netmask Local IP gateway The following parameters are configurable by changing their values in the configuration file usr g config WLdcm cfg Note that these parameters typically need not be changed Furthermore no support is provided for retaining changed settings the values will require changing again after a system software upgrade Implementation UID PDU size Association time out period Session time out period C FIND time out period 6 7 Support of Extended Character Sets Modality Worklist will support only the ISO_IR 100 ISO 8859 1 1987 Latin alphabet N 1 supplementary set as extended character sets 6 8 Modality Worklist Information Model Definition 6 8 1 6 8 2 Introduction This section specifies the use of the DICOM Modality Worklist Information Model used to organize data and against which a Modality Worklist Query will be performed The contents of this section are 6 8 2 Information Model Description 6 8 3 Information Model Entity Relationship Model 6 8 4 Information Model Module Table 6 8 5 Information Model Keys Modality Worklist Information Model Description In order to serve as a Service Class Provider SCP of the Modality Worklist Service Class a DICOM Application Entity AE possesses information about the attributes of a number of managed worklist items These items are organized into Modality Worklist Information Module
237. t the actual end of post processing 0040 0008 Modality 0008 0060 Value CT PT Value PT is stored Value CT PT or is stored in image header is stored in the in the image image header header depending on the depending on the original value of the original value of Modality Tag in the the Modality Tag worklist entry or in the worklist entries Please see entry or entries note 2 below for Please see note details 2 below for details Study ID 0020 0010 Same as exam Same as exam Copied from source number number image Performed Protocol Code 0040 0260 Derived from Not used Not used sequence Performed Series 0040 0340 One item for One item for each One item for each Sequence each series series created series created with 118 of 215 GE Healthcare REV 7 Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN Attribute Name Tag Usage in MPPS Instance Acquisition with MWL data Acquisition without MWL data Post Processing created post processing gt Performing Physician s Name 0008 1050 Copied from SPS if selected User input on the scanner Not sent as part of image header gt Protocol Name 0018 1030 The name of the protocol selected on the scanner The name of the protocol selected on the scanner Copied from source image gt Opera
238. ta Module Private Creator Identification GEMS PETD 01 Attribute Name Tag Type VR VM PET raw data type 0021 1001 3 us 1 201 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 202 of 215 GE Healthcare REV 7 Discovery 710 610 and Optima 560DICOM Conformance Statement Direction 5330620 1EN Recon parameters Attribute Name Tag Type VR VM Notes Recon Matrix Size 0x0015 0x103C 3 US 1 PET Recon 0x0015 0x101F 3 UL 1 Parameters Exists First Packet Number 0 0015 0 1020 3 SL 1 IR Loop Filter Ratio 0 0015 0 1021 3 FL 1 IR Loop Filter 0x0015 0x1022 3 FL 1 Correction BP3d Filter FlagU 0x0015 0x1023 3 UL 1 BP3d Filter CutoffU 0x0015 0x1024 3 FL 1 BP3d Filter OrderU 0 0015 0 1025 3 SL 1 BP3d Filter FlagV 0x0015 0x1026 3 UL 1 BP3d Filter OrderV 0x0015 0x1027 3 FL 1 BP3d Filter CutoffV 0x0015 0x1028 3 SL 1 Decay Flag 0x0015 0x1029 3 UL 1 Image filter 3d flag 0 0015 0 102 3 UL 1 WCC Cal Flag 0x0015 0x102F 3 UL 1 Emission Scatter 0x0015 0x1030 3 UL 1 Flag Recon Algorithm 0x0015 0x1031 3 UL 1 Normalization Flag 0x0015 0x1032 3 UL 1 Emission Deadtime 0x0015 0x1033 3 UL 1 Flag Emission Random 0x0015 0x1034 3 UL 1 Flag Blank Scan Flag 0x0015 0x1035 3 UL 1 203 of 215 GE Healthcare Discovery 710 610 and Optima 560DICO
239. tatus is received The specific error message for the failure will be logged Each C MOVE SCU operation supports an Association Timer Operation Inactivity Timer and Session Timer with time out values of 60 seconds 300 seconds and 60 minutes respectively DICOM Server AE will send a C CANCEL MOVE RQ to the Remote AE if the operator Pauses or Clears the job from the local database manager Network queue 2 3 1 2 4 Real World Activity Verify DICOM Ping 2 3 1 2 4 1 Associated Real World Activity Behavior for Discovery 710 610 and Optima 560 Products SW Version pet_mfk xx The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname The operator must then select the Ping DICOM host from Network pull down menu The DICOM server will initiate an association with the remote DICOM AE in order to verify communication at the application level The success or failure of the verification process is displayed to the user Behavior for Discovery 710 610 Products SW Version pet_coreload xx The operator shall select a Remote DICOM AE from the Network Configuration window and click on Ping on the right side bar The DICOM server will initiate an association with the remote DICOM AE in order to verify communication at the application level The success or failure of the verification
240. tement REV 11 Direction 5330620 1EN 4 3 AE Specifications 4 3 1 DICOM Print SCU AE Specification Print SCU provides Standard Conformance to the following DICOM SOP classes as an SCU SOP CLASS Name SOP CLASS UID Basic grayscale print management meta SOP class 1 2 840 10008 5 1 1 9 Basic color print management meta SOP class 1 2 840 10008 5 1 1 18 Print Job SOP class 1 2 840 10008 5 1 1 14 4 3 1 1 Association Establishment Policy 4 3 1 1 1 General The Film Composer provides options to indicate the printer AE title and Printer host name These parameters are used by the Print SCU to establish the association with the desired printer The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for association initiated by the DICOM Print SCU AE is Maximum Length PDU 16 Kbytes 4 3 1 1 2 Number of Associations The Print SCU will initiate only one association with printer Multiple associations are not supported 4 3 1 1 3 Asynchronous Nature The print SCU does not support asynchronous operations All operations will be performed synchronously 4 3 1 1 4 Implementation identifying information The Implementation UID allows unique identification of a set of products that share the same implementation The table in Appendix E identifies the Impleme
241. thcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN If any of the three timers mentioned above expires the connection is closed and the operation in progress is considered failed Note The time outs are configurable When DICOM Server AE initiates an association to issue a C STORE the following will occur 1 The image will be transmitted by the DICOM Server AE with the same elements as was originally received or created locally for the standard elements only 2 3 1 2 2 Query Remote AE 2 3 1 2 2 1 Associated Real World Activity Behavior for Discovery 710 610 and Optima 560 Products SW Version pet_mfk xx The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname To do custom queries select Yes for Custom search option Note Custom queries can be done on the following fields Last Name contains Exam Number Patient Id Accession Number To initiate a Query select Receive from Network pull down menu Note If Custom Search option is set then a small GUI will show up Operator will be able to enter custom query fields Entering values in Last Name contains field will initiate wild card query for patient name The Query operation will cause the DICOM Server AE to initiate an association to the selected Remote AE when the Query
242. the images the Storage Commitment request will be sent The DICOM Server AE to respond to an incoming Storage Commitment response from the remote DICOM AE with the Real World Activity Set Archive State This marks the successfully committed images as archived and permitted for deletion from the local storage Real World Activity for Discovery 710 610 products SW Version pet coreload xx The Choose Archive save Option Real World activity consists of an operator choosing a remote DICOM AE that supports Storage Commitment as provider as the archive device and selecting one or more exam or series from the local database and clicking on the archival destination from the archive list at the bottom of the local database browser The images to be committed are sent to the remote provider entity first The Commitment request for the transferred image instances is sent after the complete image transfer The Commitment response can come on same association or on a different association Illustration 5 2 1 Application Data Flow Diagram DOC0550678 76 of 215 Choose Archive Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement SA y p Opty 11 DICOM Storage Direction 5330620 1EN Auto SERVER Commitment Provider Exam Series Set Archive tandard State 5 2 2 Functional definitions DICOM Server AE initiates the following operations Negotiates and establishes association with remote Storag
243. tion The DICOM Server AE waits for association requests from Remote AEs that wish to perform the following operations Verification If a C ECHO RQ message is received the DICOM Server AE will send back a C ECHO RSP message with a status of success Image Storage If a C STORE RQ message is received the DICOM Server AE will receive the image and try to update the local database If the image is stored successfully on storage media and the database updated a status of success will be returned in a C STORE RSP message Query If a C FIND RQ message is received the DICOM Server AE will search the database for the requested attributes and send back a C FIND RSP message containing a match and a status of pending After all matching records have been sent a status of success will be returned in a C FIND RSP message The Remote AE can terminate the query by sending a C CANCEL FIND RQ message Retrieve If a C MOVE RQ message is received the DICOM Server AE will look up its list of configured Remote AEs for the Destination AE If the Destination AE is configured the DICOM Server AE will open a new association to the Destination AE and use C STORE RQ to send the image s The DICOM Server AE will send a C MOVE RSP message with a DOC0550678 54 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN status of pending after every five images are sent When all images a
244. tion Dose SR 1 2 840 10008 5 1 4 1 1 88 67 Enhanced SR 1 2 840 10008 5 1 4 1 1 88 22 PET Image Information Storage 1 2 840 10008 5 1 4 1 1 128 GE Private PET Raw data Storage 1 2 840 113619 4 30 2 3 1 1 Association Establishment Policy 2 3 1 1 1 General The DICOM Application Context Name ACN which is always proposed is Application Context Name 1 2 840 10008 3 1 1 1 The Maximum Length PDU negotiation is included in all association establishment requests The maximum length PDU for association initiated by the DICOM Server AE is Maximum Length PDU 64 Kbytes SOP class Extended Negotiation is not supported Maximum PDU Length of Zero 0 is not supported The association request will be rejected when a request with Maximum PDU length of zero is received The maximum number of Presentation Context Items that is supported is 60 Note that the same Abstract Syntax may be offered multiple times with different Transfer Syntax The user information items sent by this product are e Maximum PDU Length Implementation Class UID e SCP SCU Role Selection 2 3 1 1 2 Number of Associations The DICOM Server AE SCU will initiate only one DICOM association at a time to perform an image store to a remote host or retrieve image s from a Remote AE The DICOM Server AE SCP can have a maximum of four DICOM associations open simultaneously to receive and store image store or respond to an echo 2 3 1 1 3 Asynchronous Nature Asyn
245. tor s Name 0008 1070 Copied from SPS if selected and present otherwise from user input on the scanner User input on the scanner Not sent as part of image header gt Retrieve AE Title 0008 0054 Local system Local system host host name of the created within the series the series host name name system gt Referenced Image 0008 1140 One item for One item for each One item for each Sequence each image image created within image generated by post processing gt Referenced Standalone SOP Instance Sequence 0040 0220 Not sent as part of image header Not sent as part of image header Not sent as part of image header gt All other attributes from Performed Series Sequence which Table F 7 2 1 of DICOM standard PS3 4 does not explicitly list Not sent as part of image header Not sent as part of image header Not sent as part of image header All other attributes from Radiation Dose Module and Billing and Material Code Module which Table F 7 2 1 of DICOM standard PS3 4 does not explicitly list Not sent as part of image header Not sent as part of image header Not sent as part of image header Note 1 the value IN PROGRESS When PPS end N SET message is sent this element will have either COMPLETED or DISCONTINUED based on user selection Note 2 from the MWL entries as follow Single MWL item selecte
246. types Given below are the attributes supported under each of these records C 5 1 Patient Keys Attribute Name Tag Type Notes Specific Character Set 0008 0005 ISO IR 100 Patient s Name 0010 0010 If present in composite object instances it will be set to same value other wise set to NULL If present in composite object instances it will be set to same value other wise set to NULL If present in composite object Patient Birth Date 0010 0030 Patient Birth Time 0010 0032 instances it will be set to same value other wise set to NULL 0040 instances it will be set to same value other wise set to NULL C 5 2 Study Keys T 1C 2 1 2 2 Patient ID 0010 0020 If present in composite object instances it will be set to same value other wise set to NULL Attribute Name Tag T Study Date 0008 0020 1 If present in composite object instances it will be set to same value other wise set to NULL Study Time 0008 0030 1 If present in composite object instances it will be set to same value other wise set to NULL Study Description 0008 1030 2 If present in composite object instances it will be set to same value other wise set to NULL Study Instance UID 0020 000D 1C This is set to value that is present in the composite object instances otherwise composite object instances are not Archived Study ID 0020 0010 1 If present in composite object
247. ufacturer 0009 1012 1 PET for identifier 0009 1013 1 PET landmark name 1 PET landmark abbrev 1 patient_position 1 scan_perspective 0009 1017 1 scan_type 0009 1018 1 PET scan mode 0009 1019 1 PET start condition 1 179 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN PET start_cond_data PET sel_stop_cond 3 PET sel_stop_cond_data 3 PET collect deadtime 8 PET collect singles S PET collect countrate PET countrate period 009 021 3 SL PET delayed events 0009 1022 NEN PET word size 009 1024 PET axial acceptance 009 1025 amp SG G G G S O O WwW O W C O PET axial_angle 3d 0009 1026 PET theta_compression 0009 1027 PET axial_compression PET gantry_tilt_angle PET collimation PET scan_fov 0009 102B SL PET event_separation 009 102D PET mask_width 009 102E PET binning_mode 009 102 PET tracer_name 0009 1036 PET batch_description 009 1037 PET tracer_activity 009 1038 009 1039 S ss Tl e e meas_datetime pre_inj_volume 0009 103 admin_datetime 0009 103B PET post_inj_activity 009 103C PET post_inj_datetime 0009 103D 0009 103E e A j a lt _
248. ughout this entire document the term Discovery refers to the following Discovery products Discovery 710 610 64 Slice Software Version pet coreload XX Discovery 710 610 16 Slice Software Version pet mfk XX Optima 560 560FX Software Version pet mfk XX Note The Discovery product lines present an Image Works desktop which contains advanced applications offered on the Advantage Workstation The DICOM Conformance Statements related to these applications can be found at the following website http www gehealthcare com usen interoperability dicom products workstation dicom html Advanced Application DICOM Conformance Statement Reference on website CT Perfusion 4 Multi organ and Neuro Volume Viewer 9 x Applications 2087557 Rev 1 Note All references to DICOM refer to year 2009 1 4 Overview Section 1 Introduction provides general information about the content and scope of this document Section 2 Network Conformance Statement is the DICOM Conformance Statement related to this product Conformance Statements define the subset of options selected from those offered by the DICOM standard Section 3 Media Storage Conformance Statement is the DICOM Conformance Statement related to Media Storage Application Profile Section 4 Print SCU Conformance Statement is the DICOM Conformance Statement related to DICOM Print SCU Application This product uses DICOM Print SCU to print the images to DICOM Compliant Printers DO
249. uisition is started The system retrieves necessary information related to the Scheduled Procedure Step from Modality Work list Server PPS Server AE initiates a MPPS Modality Performed Procedure Step N CREATE request to the remote AE MPPS SCP in order to create a MPPS SOP instance at the remote AE The MPPS SCP returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user The DICOM association is closed 108 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN System includes the necessary information related to Scheduled Procedure Steps and the Performed procedure Step in the image instances created At the end of image acquisition system initiates a Complete PPS or Discontinue PPS based on the choice selected by the user using the user interface provided The user is also given a choice Defer PPS which is described below PPS Server AE initiates a MPPS N SET request to the remote AE in order to update the MPPS SOP instance that is already created The N SET is sent over a new DICOM association The remote AE returns response indicating the success failure of the request execution The PPS state information is updated in the system based on the response data and is presented to the user At the end of image acquisition if the user h
250. umber of Echo P 009 7 ZeroLengh Private Creator Identification 10021 00 GEMS_RELA_O1 ScouAnref P 0021 4A ZeroLength Private Creator Identification 7 002700 GEMS IMAG 01 Location RAS 00740 ZeroLength Location P 0027 x41 ZeroLength Center R Coordinate 0027 42 ZeroLength Center A Coordinate 0027 43 Zero Length Table Start Location 7 0027 50 Zero Length End Location 0027 81 ZeroLengh RASLetterforSideoflmage 7 002762 Zero Length RAS Letter for Anterior Posterior 7 0027 63 Zero Length RAS Letter for Scout Start Location P 0027 54 Zero Length RAS Letter for Scout Location 7 0027 65 Zero Length ImageDimension X 10027 60 Zero Length Image Dimension Y 0027 x61 ZeroLength Note1 Refer to section 2 3 1 2 2 1 for Custom Search Query option Note2 Type P refers to a private DICOM element 2 3 1 2 3 Real World Activity Manual Retrieve 2 3 1 2 3 1 Associated Real World Activity Behavior for Discovery 710 610 and Optima 560 Products SW Version pet mfk xx The operator must first select a destination by choosing Select Remote Host from Network pull down menu on the local database manager and then choose a hostname The operator then has to perform the Real World activity Query to get
251. vailable 0 0 ConfidentialityCode 0040 1010 BN no value available 05 0 Names of Intended Recipients 0040 1400 LT no value available 0 0 Requested Procedure Comments 0040 2400 LT no value available 0 0 Imaging Service Request Comm 0040 3001 LO no value available 0 0 ConfidentialityConstraintOnPatientData If the query is for a particular date range the ScheduledProcedureStepStartDate will be filled with a valid date range If either the start or end date are left blank by the user they will simply be blank in the query Below is an example of a date range for August 30 1997 through October 12 1997 0040 0002 DA 19970830 19971012 18 1 ScheduledProcedureStepStartDate Below is an example of a date range for August 30 1997 through the end of time 0040 0002 DA 19970830 18 1 ScheduledProcedureStepStartDate Below is an example of a date range from the beginning of time through August 30 1997 0040 0002 DA 19970830 18 1 ScheduledProcedureStepStartDate If the query is for records for this modality the Modality will be filled in as follows 0008 0060 CS CT 2 1 Modality If the query is for records for this Scanner the Modality will be filled in with CT as above and the Scheduled Station AE Title will be filled in with the value configured for this system For example this station was configured as CTRoom1 0040 0001 AE CTRoom1 8 1 ScheduledStationAETitle Us
252. vailable in MWL for PET CT exam user can select the entry for PET CT procedure DOC0550678 8 of 215 GE Healthcare Discovery 710 610 and Optima 560 DICOM Conformance Statement REV 11 Direction 5330620 1EN This Scanner Modality 0008 0060 set to CT CT Scheduled Station AE Title 0040 0001 set to local AE title All CT Modality 0008 0060 set to CT Systems Scheduled Station AE Title 0040 0001 zero length universal matching This PET Modality 0008 0060 set to PT System Scheduled Station AE Title 0040 0001 set to local AE title All PET Modality 0008 0060 set to PT Systems Scheduled Station AE Title 0040 0001 zero length universal matching All NM System Modality 0008 0060 set to NM Scheduled Station AE Title 0040 0001 zero length universal matching All Scanners Modality 0008 0060 zero length universal matching Scheduled Station AE Title 0040 0001 zero length universal matching The scheduled dates of procedures of interest can be specified for query by selecting a specific date range The date ranges available are Today Days Before Today Days After Today and All Days These selections and their effects on worklist query parameters are given below Today Scheduled Procedure Step Start Date 0040 0002 set to YYYYMMDD where this date is the current date Days Before Today Scheduled Procedure Step Start Date 0040 0002
253. vider accepts full responsibility for all validation required for their connection with GE devices This includes the accuracy of the image data once it has crossed the interface between the GE imaging equipment and the non GE device and the stability of the image data for the intended applications Such a validation is required before any clinical use diagnosis and or treatment is performed It applies when images acquired on GE imaging equipment are processed displayed on a non GE device as well as when images acquired on non GE equipment is processed displayed on a GE console or workstation Future Evolution GE understands that the DICOM Standard will evolve to meet the user s growing requirements GE is actively involved in the development of the DICOM Standard DICOM will incorporate new features and technologies and GE may follow the evolution of the Standard The GEHC protocol is based on DICOM as specified in each DICOM Conformance Statement Evolution of the Standard may require changes to devices which have implemented DICOM In addition GE reserves the right to discontinue or make changes to the support of communications features on its products described by these DICOM Conformance Statements The user should ensure that any non GE provider which connects with GE devices also plans for the future evolution of the DICOM Standard Failure to do so will likely result in the loss of function and or connectivity as the DICOM Standard
254. y 1 Click on Patient Schedule 2 Click on Preferences button 3 Set the option Show Update Parameters to Yes 4 To do a query click on Update button A User Interface appears with provision to enter values for a Accession Number b Requested Procedure Id 106 of 215 GE Healthcare Discovery 710 610 and Optima 560DICOM Conformance Statement REV 7 Direction 5330620 1EN 7 PERFORMED PROCEDURE STEP CONFORMANCE STATEMENT 7 1 Introduction This section of the DICOM Conformance Statement specifies the compliance to DICOM conformance requirements for the Performed Procedure Step feature on this GEHC product Note that the format of this section strictly follows the format defined in DICOM Standard PS 3 2 Conformance Please refer to that part of the standard while reading this section The PPS option for Discovery allows a Modality Performed Procedure Step to be communicated to the Hospital Radiology information system The PPS feature is providing the DICOM Modality Performed Procedure Step service as a service class user SCU This feature works in conjunction with DICOM Modality Work list feature if installed However the conformance of this feature is independent of Modality work list feature For information on conformance of Modality Work List feature to DICOM standard please refer to the appropriate section or document 7 2 Implementation Model The DICOM Performed Procedure Step service is provide

Download Pdf Manuals

image

Related Search

Related Contents

Forza Power Technologies FAV-HD06AP  MANUEL D`UTILISATION Limeuses pneumatiques SH100A  Lector Portátil Inalámbrico (WPR)  Lightweight, High-Resolution Monitoring for Troubleshooting  Yamaha NS-10M STUDIO Owner's Manual  Auto-rádio FM estéreo/AM com toca-discos a laser  Tech air Z0125  Manual de usuario  LG 52LG50 52" Full HD Black  The User Guide for Web  