1.   ij  Accellix                  Operator Printer     Language  Offset to GMT       Deutsch i EAE     12 53 PM  hal    ft       Figure 4  Settings Page    10  In the Settings page  perform the following     e Enter the Operator name in the Operator field  Note that the operator name appears in  the Results page  so we recommend you change the operator name if the operator  changes     e Select a language from the Language list     e Set the GMT Offset to define the date and time settings  The current time is displayed  under the GMT Offset fields     e Define a Printer for print jobs  This defines where printouts are sent     11  Tap the Home icon tt to return to the home page     The device is now ready for use     P  LeukoDX spc 014 Accellix User Manual Rev 02 Jan 2015 Page 5 of 19    ij  Accellix    Using the Accellix Diagnostic Device    The Accellix device diagnoses cells in a sample which could be blood drawn from a patient or growth  media     the sample      The user inserts the sample into a cartridge which is inserted into the device   please see Instructions For Use  IFU  for the specific cartridge being run      1  Inthe Home page  tap           of              Figure 5  Home Page    The process steps  2   7  are displayed by the highlighted gray dots in the bottom of all  screens     2  Enter the Patient ID by using the bar code reader  or tapping it in on screen  The  Patient ID will be displayed on all screens until cartridge is removed from device     P  LeUKOU
2.  and logistics delays  resulting in  quicker action and more satisfied patients     Intended Use    The Accellix platform may be used for a flowcytometric measurement of analytes in biological  samples  using test specific cartridges intended for use with this instrument     Principles of Operation    The Accellix platform  relying on the general principles of flow cytometry  combines fluidics  optics  and electronics systems in order to measure and analyze optical and fluorescence characteristics of  single cells  The Flow Cytometer enables complete automation of a 3 step process  biochemical  sample preparation  i e   the assay   flow cytometry reading and analytical data processing     The Accellix platform is designed to enable automatic workflow  thus it requires the user to perform  only few and simple operational steps   Following is a brief description of the mode of operation of the Accellix system     e The user selects the assay from the menu displayed on the touchscreen   e The user enters the patient ID  e The user places a drop of sample in the Assay Cartridge  closes the Cartridge Cap  inserts the    Cartridge to the Reader Instrument  and closes the Door     P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 1 of 19    Lj  Accellix    After closing the Instrument Door  the complete assay is automatically executed and the data are  analyzed  1 e   the system brings the Cartridge into place  runs the assay  analyzes the results and  displays the CD64 i
3.  negative  controls and data quality checks  Automated data  analysis providing the user with a final results        P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 17 of 19    Appendix B     Symbols    1  Index of Symbols     Ce    CE Mark    DO NOT REUSE    z    LOT    USE BY    BATCH CODE       MANUFACTURER    REP    EU Representative    g    IN VITRO DIAGNOSTIC MEDICAL DEVICE    UPPER AND LOWER TEMPERATURE LIMITS    CONSULT INSTRUCTIONS FOR USE    eE    BIOLOGICAL RISKS    P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015    Lj  Accellix    Page 18 of 19    Lj  Accellix    Appendix C     Safety Information    e Do not open the device door while the device is processing a sample as it will abort the test   This is done to protect the user while the laser is active     e For cartridge and assay specific safety issues please refer to the cartridge IFU   e Dispose of the cartridge as a biohazard as per local regulations   e Electrical safety  The instrument meets    e IEC EN 61010 2 101  2002  ed 1     Safety requirements for electrical equipment for measurement   control  and laboratory use   Part 2 101  Particular requirements for  in vitro diagnostic  IVD  medical equipment    e JEC EN61010 1  2010    e EN61326 EMC emission and immunity requirements    P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 19 of 19    
4. X spc 014 Accellix User Manual Rev 02 Jan 2015 Page 6 of 19    ij  Accellix    Enter Patient ID   51436987    4 is     2       Figure 6  Entering Patient ID    3  Tap Next     4  Extract the cartridge from its bag and scan in the cartridge ID using the bar code  reader     5  When the following screen appears  insert the sample into the cartridge  for further  details regarding the sample and the sample loading instructions for a specific test  please refer to the Instructions For Use  IFU  for the test being run   and then insert  the cartridge into the device     ID  51436987    Insert cartridge  amp  Close Door       Figure 6  Instructions How to Insert the Sample and the Cartridge    P  LeukoDX spc 014 Accellix User Manual Rev 02 Jan 2015 Page 7 of 19    ij  Accellix    6  Wait while the screen displays the following message     Please wait while cartridge    QR code is read and confirmed        Please wait while cartidge QR  code is read          Figure 7  Instruction to Wait While Cartridge Code is Internally Scanned    Accellix    Confirm Patient ID  amp  Assay by  pressing Next to start test    Patient ID  51436987    Assay  CD64       Figure 9  Instruction to confirm Patient ID and Assay    P  LeukoDX spc 014 Accellix User Manual Rev 02 Jan 2015 Page 8 of 19    E S  Accellix  7  Tap Next     While the device processes and analyzes the sample  the following message is displayed   Take care not to open the door during processing  opening the door will abort the  testin
5. cceli g  u  Are you sure you want to shut  down the system     Shut Down       Figure 15  Shut Down Page    2  Tap Shut Down to shut down     To turn the device back on  turn the power switch on the rear of the unit to the OFF position and then  turn it back to the ON position     P  LeukoDX sSPC 014 Accellix User Manual Rev 02 Jan 2015 Page 11 of 19         A ccellix    Troubleshooting and Maintenance    Troubleshooting    Computer Issues    o User should contact service if the system does not start  o User should contact service if the system is out of memory  Touchscreen Issues    o If touchscreen does not respond the user can operate the system temporarily using a  USB connected mouse and keyboard and contact service     Internal Instrument Controls and Traceability and Changes in Analytical Performance    The Instrument Controls and Traceability are in place via a set of checks performed upon  powering up the instrument  once a day on a fixed schedule if the instrument is not powered  down and up once a day and upon the occurrence of a group 2 error  see Error section below    This set of self checks includes checking the components required to operate the fully  automated system  If any such self check detects a component no functioning properly the  user will receive an error message that provides information about the failure and instructions  to call service     Internal assay controls are assay specific and are addressed in the cartridge IFU  They  include data cont
6. f check   This self check program is the same automated procedure which is performed by the  instrument automatically upon power up or by automatic scheduling every 24 hours 1f the  system is not powered down once a day  This self check procedure check a significant  number of critical components  such as laser  detectors  cartridge actuation motors and  communication between internal components     Example of cases where the instrument will request the user to reboot the system to perform self   check include     e Data controls fail   e Laser power is if outside 20  of set point value   e Average value of image pixel intensity is below control limit   e Acquired file size after acquisition is complete is within    10  of control criteria    If the self check is completed successfully  the system will request from the user to run another  cartridge  with the same or a new sample   The system may still provide an error message but  it could be related to a specific sample that cannot be measured for the test being run  Please  refer to cartridge IFU for the specific assays being run which provides details about Test  Limitations and Interferences such as sample types and medical conditions that may lead to  assay failure      If the self check fails  an error message will provide information about which component is  failing the self check test and request the user to contact the manufacturer     Specific instructions for each error type are provided on screen as in the examp
7. formed at  the manufacturer site according to manufacturing and re work procedures   For servicing  see Manufacturer information in the beginning of this document  or see contact  details at www leukodx com     e Shipment for service     In case shipment for service is required please contact the EU  representative or the manufacturer  The original packaging is to be saved and stored in a safe  place to enable its re use for shipping back to the manufacturer for service  In case you  consider the instrument to be contaminated please put the instrument in the provided  biohazard bag inside the original packaging     Cartridge and Assay Specific Issues    Cartridge and Assay Specific issues are specific for each assay run on the instrument and are  provided in the Instructions for Use  IFU  provided which each box of assay specific  cartridges and available online at www leukodx com     The cartridge IFU for each assay provides information about   e Measurement Procedure Details    e Cartridge and Assay Specific Internal Controls and Traceability and Changes in Analytical  Performance    The Cartridge and Assay Specific and Traceability and Changes in Analytical Performance  are also addressed by the instrument by checking that the data generated in a particular run    P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 15 of 19    _  Accellix    meets a certain number of data quality checks  Further information is provided in the  Instructions for Use  IFU  provided w
8. g process     Please wait while assay completes       Time Remaining    1 42       Figure 10  Time to Test End    When processing is complete  a results page appears  similar to the following     CD64 Index    0 68       Figure 81  Results Page    8  Tap Next      or If desired  tap fay to go the page designed print the results  If no printer is connected it will give a  message     P  LeukoDX spc 014 Accellix User Manual Rev 02 Jan 2015 Page 9 of 19     amp   Accellix         Microsoft XPS Document Writer    Fax  East Wing       Figure 92  Choosing Printer    9  Open the door and remove the cartridge when the following message is displayed     Ty    ad    1  Open Door  2  Remove Cartridge  3  Close Door       Figure 13  Instruction to Open Door and Remove Cartridge    10  Dispose of the cartridge as a biohazard according to local regulations     O Aou    a    Please wait while current test  terminates          Figure 14  Shown immediately after an abort or failure     It is displayed while the system returns to initial positions so the cartridge can be removed  this can  take some time depending on how quickly the assay   s script can be aborted and where all the motors  are      LeuUKO SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 10 of 19    ij  Accellix    Shutting Down the Device    The device can be left on and there is no need to shut it down regularly  If you do wish to shut it  down     T  1  In the Home page  tap  0      A confirmation query appears     oe    k 
9. instrument     d  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 14 of 19    Lj  Accellix    e Connection to customer LAN  Wi Fi network  or Printer should be performed by qualified IT  personnel  Connection to customers LAN should be with less than 3 meters cable  Please  contact LeukoDx service department to obtain information regarding how to perform such  operation     e Cleaning the System     The Accellix cytometer may need to be cleaned from time to time  In  order to maintain the integrity of all of the Accellix components the following cleaning  procedure can be used  wipe off the device with a cloth or wipe saturated in isopropyl alcohol   IPA  or an isopropyl alcohol mix  Examples of appropriate cleaning solutions include   CaviWipes and Purisept spray and wipes  The device should be cleaned according to the  directions provided with the cleaning solution     e Decontamination     In case of minor spill from a cartridge  proceed with decontamination  using alcohol according to your institution handling procedure for materials that have been in  contact with blood  All blood products should be treated as if they may potentially contain  pathogens     e Disposal Procedure     The instrument should be disposed of according to local regulations in  place for electronics and medical equipment  Dispose of used cartridges as a biohazard  according to local regulations     e Service Manual     No servicing is to be performs at the user site  All servicing is per
10. ith each box of assay specific cartridges     e Calibration     The instrument self checks ensure that the functionality of the instrument is  tested  If calibration is required for a certain cartridge or assay type  the details of the nature  and frequency of the calibration needed to ensure that the device operates properly and safely  will be described in the IFU for such cartridge     P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 16 of 19    Lj  A ccellix    Appendix A     Technical Specifications    Instrument Description Specification    Use conditions Ambient Temperature  15  C   32  C  Non Condensing  Max  Rel  Humidity 85      Protect the system from direct sunlight  humidity and  dust        Focus and Alignment of Optical platform   Real time focus and excitation alignment utilizing  proprietary image analysis control algorithm          DAQ A D Resolution  Sampling Rate 14 bit  100 KHz    Software LeukoDx custom software running on Windows    7  embedded  64 bit  operating system    Assay automation    Motor control of 7 assay motors  Auto focus  Auto   alignment  Three reagent injection  Sample Mixing   Sample Reading     Sample type  Sample volume  Sample Assay Specific  see relevant cartridge IFU   Limitations    Assay automation Assay Specific  General capabilities are  Three reagent  injection steps  one temperature controlled  Each step  enables mixing and incubation time  automated sample          reading  Assay specific integrated positive and
11. le below     P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 13 of 19    ij  Accellix    CD64  ee  ID  51436987    Accellix  lt   Error  Error  32  Error moving motor 0  A fatal error occurred  You may either     1  Run a System Check to try and clear the error or  2  Shut down the device and Contact manufacturer    System Check      _         Figure 17  Example of Group 2 Error Page    Handling    Handle the instrument with care   Protect the system from direct sunlight  humidity and dust     Environmental operation conditions   Ambient Temperature  15  C   32  C  Non Condensing  Max  Rel  Humidity 85      For information on instrument dimensions and power  refer to Appendix A     Technical  Specifications    Control Method    Control methods and means to verify performance within specifications are in inherent to the different    assay cartridges and respective data processing   Further information is provided in the Instructions for Use  IFU  provided which each box of assay    specific cartridges and available online at www leukodx com    Maintenance    Maintenance     the instrument is maintenance free  the user only needs to follow instructions  provided by errors messages  See Troubleshooting and Error Messages section below     This instrument is classified as a Class 1 Laser product   according to IEC 60825 1 edition 3     The instrument uses internally a 50mW  488nm Class 3B  Laser   The user   customer   technician should never open the                
12. mj  Accellix    p  LeukoDx       Co     A CCCLUX    User Manual    SPC 014 Version 2 0  January 2015       Copyright 2014 LeukoDx Ltd  All rights reserved   Accellix is a trademark of LeukoDx Ltd     All trademarks used herein are the property of their respective owners     P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015    wl    Manufacturer Information    LeukoDx Ltd     Tel   972 2 674 4422  Fax   972 2 674 4455  E mail  office leukodx com    3 HaMarpe St   3rd Floor  AVX Building   Har Hotzvim  Jerusalem 9777403  Israel    EC REP    EU Representative Information    Medpace Medical Device B V   Maastrichterlaan 127 129  NL 6291 EN Vaals  Tel   31 43 306 3320   Fax   31 43 306 3338    Table of Contents    RRC ROC DTE OEI AA P A A OEI TO I A AA  UniS oo US e A E E E  Pinciples OF Operace E  Overview of Operating Elements        sesccccccccccccccccccccccccccecccccececccccccccccccceccecccccccccccceceeee  First Time PMS CAAT ON sineira E aE Oie  Using the Accellix Diagnostic DeviCe         ssssssssssccccccccccccssssssssccccccecccccccossssssscssccecceseosoo  UCI Down the DG VIC 0 ssesscsiissneiai narn E  Troubleshooting and Maintenancce               cccccccccccccssssssssssssssssssssssssssccccccccccssssssssssssees  OTN OU INE O oa E E E S  Appendix A     Technical Specifications              ccccccsssssssssscsscccccccccccccccsssssssssssssssecees  POAC  BD   Sy NOS aa E  Appendix C     Safety MilOr mato nis siceccssssicsccscssessssccdcecsassvsvsssceteccsusssevsessedencsss
13. ndex     e The Cartridge Handling Unit returns the Cartridge to start position and the Cartridge can be  removed  The GUI prompts the user to remove the Cartridge from the Reader Instrument and to  discard it as a biohazard     Overview of Operating Elements    The Accellix platform is a diagnostic system composed of a tabletop Flow Cytometer and a single use  disposable Assay Cartridge    The Accellix Cytometer is mainly operated through the touchscreen and attached barcode reader   Detailed description and functional specifications of the Assay Cartridge are provided in the Cartridge  Instructions For Use  IFU             Cartridge             Figure 1  The Accellix platform  Flow Cytometer and Cartridge  The Assay  Cartridge is shown here upon insertion into the Flow Cytometer    p  LeukoDX spc 014 Accellix User Manual Rev 02 Jan 2015 Page 2 of 19    hj  Accellix    First Time Installation  1  Unpack the package and verify it contains   e The Accellix device  e A power cable  e A bar code reader  e A biohazard bag to enable shipment of a contaminated instrument    The original packaging is to be saved and stored in a safe place to enable its re use for shipping  back to the manufacturer for service     2  Place the device on a clean flat and stable surface     3  Connect one end of the power cable to 100 220V AC 50 60Hz power source with the supplied  power cable  and the other end to the power outlet at the rear of the Accellix unit  In areas  with power disturbances or t
14. o ensure test completion in case of power outage it is  recommended to connect the instrument to the power via an UPS that can provide 300VA or  higher for amount of time of 20 minutes  not supplied with the instrument      RJ 45 port    USB ports    Power switch    Power socket       Figure 1  Rear of Accellix Unit    4  Connect the bar code reader to a USB port at the rear of the Accellix unit     5  Optionally  connect an Ethernet network cable to the RJ 45 port at the rear of the Accellix  unit  This enables connectivity to a network  Alternatively  connect a local printer to one of    P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 3 of 19    ij  Accellix    the USB ports on the rear of the device if Internet is available to the device so the printer   s  driver can be downloaded     See Maintenance section regarding connecting to a network or installing drivers     6  Optionally  connect a keyboard  not supplied  to a USB port on the rear of the unit  This is  not mandatory because you can use the device on screen keyboard     7  Turn on the device by switching the Power switch at the rear of the unit to ON     The following message appears on screen     Performing System Self check    Please Wait       Figure 2  System Self Check    8  Verify that the system completes initialization  and the Home page appears        Figure 3  Home Page    9  Tap Settings to access the Settings page     P  LeukoDX sSPC 014 Accellix User Manual Rev 02 Jan 2015 Page 4 of 19  
15. rols to verify data integrity and maintained analytical performance per each  run     Traceability  The internal control materials were developed based on a fluorescence bead  suspension traceable to the NIST Standard Reference Material SRM 1932  RM 8640     Error Messages  Error Message are classified in two groups     Group 1 errors  Recoverable errors    The system detects that an operation failed and it can be recovered   Examples of recoverable errors are     The system does not recognize the cartridge barcode or a patient ID and asks the user to extract  the cartridge and insert it again or request the user to enter a patient ID manually  If a certain  operation fails repeatedly the system will move to Group 2 error after a certain number of  attempts     Specific instructions for each error type are provided on screen as in the example below     P  LeukoDx SPC 014 Accellix User Manual Rev 02 Jan 2015 Page 12 of 19    ij  Accellix    ID  51436987    Error  1205  Failed to read Cartridge ID    Discard cartridge and repeat with new cartridge       Figure 16  Example of Group 1 Error Page    Group 2 errors  Recoverable errors requiring instrument self check    The system detects that an operation failed but the failure may be connected to a specific  cartridge or a specific patient sample     Before proceeding with repeating the test with another cartridge the system will provide an error  message about which test condition failed and request from the user to perform a sel
16. ssessscesbessss    P  LeukoDX spc 014 Accellix User Manual Rev 02 Jan 2015    Lj  Accellix    Introduction    Accellix is an affordable tabletop diagnostic device with single use assay cartridges that does not  require extensive training nor involve complex operational processes  Accellix dramatically expands  access to flow cytometry capabilities by replacing previously complex testing with a user friendly  automated instrument that can be used by any medical practitioner  nursing staff  or junior lab tech on  a 24 7 basis  Health practitioners get rapid results that previously required sending samples out to a  central lab  Accellix eliminates that transport time  wait times at the lab  and other delays that may  directly lead to degradation in patient health and effective treatment     The Accellix platform s fast and user friendly interface facilitates simple and reliable analysis using  disposable cartridges for sepsis tests     Accellix is mainly intended for     e Hospitals with time critical departments  ICU  ER  OR  neonatal ward    Nurses on  location can run tests  eliminating long waits and queues at a central lab     e External Third party run hospital labs   Simple  automated tests are available 24 7 to all  the lab   s staff  not just senior specialists  This eliminates wait time for batch processing  and  tests can be run and returned quickly  proving especially useful for high priority cases     e Clinics and Doctors   Accellix drastically reduces test travel
    