E8408A C-Size VXI Mainframe User and Service Manual

image

Contents

1. 440 Hz max 90 Vac min 264 Vac max 47 Hz min 66 Hz max Input Voltage 90 Vac to 264 Vac CAT II Automatic Ranging 280 VA max Inrush Current Power Switch 75A 264 Vac 35A 132 Vac On Standby Cold start Chassis Ground Tap M4 x 0 7 threaded nut insert On rear of mainframe Appendix A E8408A Product Specifications 61 Cooling Specifications This section lists E8408A mainframe cooling specifications including Cooling Table Cooling Chart Cooling Table Area Specification Comments Cooling Modes High or variable cooling mode switchable on the front panel Same fan cools power supply and modules High Speed Fan Mode Maximum airflow all the time Variable Fan Speed Fan speed changes between high and low as a function of ambient Mode and power supply temperatures AirFlow Path Air is routed into the mainframe rear and exhausted out the left side viewed from the front when the mainframe is in horizontal orientation When the mainframe is rack mounted allow 50 mm of clearance at the rear and left side for proper airflow Acoustic Noise High Speed Fan Mode 40 dBA Variable Fan on Low Speed 30 dBA Sound power at bystander position one meter in front of mainframe Cooling Chart Figure A 2 shows E8408A cooling specifications for all E8408A slots per VXI 8 Specification Draft 2 0 and Airflow Test Fixture revision 1 7 with e VXI 8 Standar
2. 34 Installing Mainframe Options Chapter 2 Install Rack Mount Labels Figure 2 14 1 Lay mainframe on its left side 2 Open the Rack Mount Cable Route Adapter Kit E8397A 3 Peel the protective backing from the two rack mount front panel labels in the E8397A kit and carefully place them over the existing labels Press to adhere Lay Mainframe on Side Install Rack Mount Labels NT RM i Figure 2 14 Installing Rack Mount Labels Chapter 2 Installing Mainframe Options 35 Attach Handles to Rack Mount Adapters Figure 2 15 1 Attach a handle to the LH Rack Mount Adapter using two 10 32 X 0 375 flat head screws 2 Attach a handle to the RH Rack Mount Adapter using two 10 32 X 0 375 flat head screws NOTE The rack mount adapter without the opening is the left side LH Rack Mount Adapter and the rack mount adapter with the cable routing opening is the right side RH Rack Mount Adapter Right Hand Rack Left Hand Rack Mount Adapter Mount Adapter T Flat Head Screws J Flat Head Screws Figure 2 15 Attaching Handles to the Rack Mount Adapters 36 Installing Mainframe Options Chapter 2 Attach Rack Mount Adapters to Mainframe Figure 2 16 NOTE The mainframe can be recess mounted up to 207 mm approximately 8 inches in 23 mm approximately 1 inch increments When recessing the mainframe make sure that six screws are
3. Grounding Spring Socket i HI Chassis Shield 4 qr Guide Use a Small Screw Driver to Secure Spring Under SocketTab Figure 2 21 Installing the Grounding Springs and Chassis Shield Installing Backplane Connector Shields This section describes how to install backplane connector shields E8408 80900 in the E8408A mainframe The shields are the implementation of VXI revision 2 0 specification B 7 2 3 which ensures compliance with RFI levels specified in standards EN55011 and CISPR11 A number of VXI instruments require shields for compliance with EN55011 and CISPR11 Refer to the instrument documentation to determine if shields are required WARNING Do not install backplane connector shields while the mainframe is turned on or plugged into an AC power source Parts List The parts included in the E8408 80900 kit are shown in Table 2 5 Table 2 5 E8408 80900 Parts List Quantity Description Part Number 18 Pan Head Screw 4 20 x 25 0624 0702 8 Backplane Connector Shield E1400 80601 Includes two extra screws Chapter 2 Installing Mainframe Options 43 Installation 1 Position Shields Position the shields over the backplane connectors Procedure as shown in Figure 2 22 Two connector shields and four screws are required for each slot
4. I 1 540 mm 21 3 in 133 3 mm 547 6 mm M 5 25 in y 21 56 in D tT Im d 152 4 mm 38 1 mm 38 1mm 6 00 in EE 1 50 in 1 50 in 8 EIA Rack Units Figure A 1 E8408A Mainframe Dimensions Appendix A E8408A Product Specifications 57 Options and Accessories Description Product Number Standard Configuration 4 Slot C Size 175 W VXI Mainframe includes front panel indicators protective bumpers and carrying handle E8408A VXI Instruments Modules Installed with Standard Settings E8404A Option 500 Installed Backplane Connector Shields E8408A Option 918 Standard Configuration except Add Additional Current from 5 2V Source E8408A Option 001 Remanufactured Power Supply Assembly Standard E8408 69005 Remanufactured Power Supply Assembly Option 001 E8408 69006 Airflow Restricter 1 slot E8400 80917 Replacement Fan Assembly E8408 68500 Intermodule Chassis Shield Kit E8400 80919 EMC Filler Panel 1 slot E8400 60202 VXI Slot Filler Panel 1 slot E8400 44305 E8408A User Service Manual if ordered separately E8408 90001 Convert 3 year return to to 1 year On Site Warranty E8408A Option W01 Portable Configuration Carrying Strap and Accessory Pouch E8396A Rack Mount Configuration
5. you must add End Brackets 1494 0061 to the Rack Slide Kit 1494 0413 See the instructions for the 1494 0061 kit to install the rack slides The Rack Slides occupy the center EIA unit of the three EIA units required for the E8408A mainframe 1 Slide four 10 32 Self Retained Nuts on the selected holes in the right hand vertical column of the cabinet See Figure 2 10 for hole locations 2 Attach the RH Rack Slide Stationary Section to the right hand vertical column of the cabinet Use two 10 32 x 0 5 flat head screws at the front of the rack slide and two 10 32 x 0 375 pan head screws at the rear of the rack slide 3 Repeat steps for the LH Rack Slide Stationary Section Right Hand Vertical Columns Pan head Screws Y Right Hand Q Rack Slide Stationary Section Flathead Screws Age ft d Self Retained Nuts Figure 2 10 Installing Rack Slide Stationary Sections 30 Installing Mainframe Options Chapter 2 Installing the Mainframe Install the Mainframe in the Rack Figure 2 11 1 Slide the Rack Slide Intermediate Sections out from the Rack Slide Stationary Sections until they lock into place WARNING To prevent possible injury during rack mounting two people should lift the mainframe into the rack 2 With one person on each side lift the mainframe and slide the Chassis Sections into the Intermediate Sections until the Chassis Sections lock
6. 0 0001 g Hz Spectral Density Survival Swept Sine 5 to 500 Hz Resonance Search 5 minute dwell on resonances at 0 5g Survival Random 0 015 g Hz Spectral Density Altitude Up to 3 000 m Appendix A E8408A Product Specifications 63 Electromagnetic Compliance Safety Specifications Power Supply Protection Over Temperature Protection Over Voltage Protection Short to Ground The E8408A mainframe conforms to the EMC and safety product specifications listed in the Declaration of Conformity see Page 7 EMC performance can be further enhanced with the following shields Front panel gasketing provided per VXI Rev 2 0 B 7 2 3 Backplane connector shields per VXI Rev 2 0 B 7 2 3 Intermodule chassis shields per VXI Rev 2 0 B 7 3 4 The supply is protected from over temperature over voltage short to ground and short to other output Protection modes vary from removing the fault condition to cycling the power on standby switch to removing and reconnecting AC power The front panel Temp temperature indicator flashes when an internal temperature is high enough to cause the mainframe to approach thermal shutdown but the Temp indicator does not cause shutdown When thermal shutdown limit is reached all supplies are automatically turned off Recovery occurs automatically when the mainframe has cooled sufficiently Over voltage will generally cause a global shutdown and recovery occurs when t
7. Started sisi ssssnsonessospensncessnmecenasssssmnansssssnanaraanmenaacsiasenpantaus 9 Chapter es I M 9 Produc OVEN RE 9 Mittal Mamntame SUP iced ose aae ec E EE uS EE pU Ha Fa d zT i 11 AC Power Roguen ents ss ood y EYE rex ed bo d e e p ad T1 Positioning the Mainframe for Adequate Cooling esessss 11 Connecting Permanent Earth Ground scenes ert nre nat nas 12 installing VAI GSP METER ann a dd akkie ocRUECHEUS doa tK Pn pda KEEPER E REP ra UP IG FERIA 13 ILI PORTE uisa ccce cena tco Hence ascen bl ae eA iet pee Ud 13 instalation Guidelines PR TT 13 Installing VAI Instruments Using VIC isses sues see utri tetra cartera panna 14 installing C Size InNStrUMeNtS MNT m T 14 Installing A end B Size Instruments eocuccciiisace cer racer stantia arte 14 xus m lc Memes M 16 Chapter 2 Installing Mainframe Options ss 17 Gapir yi E EC T T ee mee m 17 Converting to Portable Config allglm ssi sosae ns yate Rd xt Ha d d Kad d FR ds at 17 zr et DP 17 ayuu mc n 18 Rack Mounting Using Rack Slide Minime pbi n ER p FI Na AER PP Vete RE REI Ce i 20 PIRS LIS c ea IIE 20 RON PROS OURS MEM 20 Rack Mounting Using Support Rail KIEL esie chocar raa tr in az doeet 31 ges ea Beer ee emir eres D Vnd si EEG Posi dati EC CAR 31 Installation PR
8. The fan operates in Variable Speed or Full Speed mode as controlled by the front panel Fan Speed switch Full Speed mode is recommended for maximum cooling when acoustic noise is not a factor Variable Speed mode allows quietest operation while providing sufficient power supply and module cooling In Variable Speed mode the fan speed is controlled automatically as a function of temperature rise in the mainframe and the ambient temperature Airflow enters the E8408A mainframe rear panel flows through the power supply fan and modules and exhausts out the left side as viewed from the front in horizontal orientation This allows mainframes to be stacked or rack mounted directly on top of one another When placing the mainframe on a work bench or if the mainframe is rack mounted provide at least a one inch clearance at the back and sides to allow proper air flow Air filters are not necessary on E8408A mainframes Getting Started 11 Connecting Permanent Earth Ground WARNING The mainframe can be connected to a permanent earth ground This connection is made on the back of the mainframe see Figure 1 2 For protection from electrical shock when operating at frequencies greater than 66 Hz connect the chassis ground terminal to permanent earth ground 1 Connect a 16 AWG 1 3 mm or larger wire to the PEM nut shown in Figure 1 2 The wire must be green with a yellow stripe or bare no insulation Use a M4 x 10 screw
9. and workmanship when properly installed and used If Agilent receives notice of such defects during the warranty period Agilent will replace software media which does not execute its programming instructions due to such defects 3 Agilent does not warrant that the operation of Agilent products will be interrupted or error free If Agilent is unable within a reasonable time to repair or replace any product to a condition as warranted customer will be entitled to a refund of the purchase price upon prompt return of the product 4 Agilent products may contain remanufactured parts equivalent to new in performance or may have been subject to incidental use 5 The warranty period begins on the date of delivery or on the date of installation if installed by Agilent If customer schedules or delays Agilent installation more than 30 days after delivery warranty begins on the 31st day from delivery 6 Warranty does not apply to defects resulting from a improper or inadequate maintenance or calibration b software interfacing parts or supplies not supplied by Agilent c unauthorized modification or misuse d operation outside of the published environmental specifications for the product or e improper site preparation or maintenance 7 TO THE EXTENT ALLOWED BY LOCAL LAW THE ABOVE WARRANTIES ARE EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED AND AGILENT SPECIFICALLY DISCLAIMS ANY IMPLIED WARRAN
10. excessive levels or to protect VXI measurement modules from noise sources WARNING Do not install intermodule chassis shields while the mainframe is turned on or plugged into an AC power source Parts List rhe parts included in the E8400 80919 kit are shown in Table 2 4 Table 2 4 E8408 80919 Parts List Quantity Description Part Number 1 Chassis Shield E8400 00600 8 Grounding Spring E8400 09101 Includes 4 extra springs Installation Each mainframe slot has top and bottom chassis shield guides Figure Procedure 2 21 Chassis shields should be installed on both sides of an instrument generating interference noise or installed on both sides of a sensitive instrument 1 Insert Grounding Springs Insert grounding springs into the four sockets two top two bottom along the shield guides Use a small flat blade screwdriver to secure each spring under the socket tab NOTE Use the chassis shield to cover the vent holes inside the mainframe when installing grounding springs in the top guide sockets This prevents the springs from falling into the mainframe if they are accidently dropped 2 Install the Chassis Shield Align the chassis shield with the shield guides and gently slide the shield into the mainframe Be careful not to crimp the front grounding springs with the edge of the shield Slide the shield in until it reaches the end of the shield guide 42 Installing Mainframe Options Chapter 2
11. grounding lug and toothed washers or toothed lug as shown in Figure 1 2 2 Attach the other end of the wire to a permanent earth ground using toothed washers or a toothed lug Green Wire Permanent Earth With Yellow Ground Connection Stripe Location Wire Lug Toothed Washers Back Cover Figure 1 2 Connecting the E8408A Mainframe to a Permanent Earth Ground 12 Getting Started Chapter 1 Installing VXI Instruments Slot 0 Functions Chapter 1 Installation Guidelines WARNING Note This section shows how to install A B and C Size VXI instruments modules in an E8408A mainframe including Slot 0 Functions nstallation Guidelines nstalling C Size Instruments nstalling A and B Size Instruments The E8408A mainframe has 4 slots labeled 0 through 3 Any VXI instrument can be installed in slots 1 through 3 However slot 0 is reserved for devices capable of providing the system s slot 0 functionality such as the E1406A Command Module GPIB or E8491A IEEE 1394 This functionality includes locating instruments installed in the mainframe managing arbitrating data flow across the backplane providing the system clock SYSCLK 16 MHz 1 Multiple instruments that combine to create a virtual instrument e g a scanning multimeter and instruments which access the backplane local bus should be installed in adjacent slots 2 To
12. prevent damage to VXI instruments do not install instruments unless the Power On Standby switch is in Standby position 3 Insert the instrument into the mainframe by aligning the instrument with the card guides inside the mainframe Slowly push the instrument into the slot until it seats in the backplane connectors The front panel of the instrument should be even with the front edges of the mainframe When installing or removing an instrument in Slot 0 you will need to move the front panel protective bumper aside Restore the bumper to its original position after the instrument is installed or removed Tighten the retaining screws on the top and bottom of the module All instruments within the VXI mainframe are grounded through the mainframe chassis During installation tighten the instrument s retaining screws to secure the instrument to the mainframe and to make the ground connection The E8408A is shipped with 1 slot ES408 44305 and or 3 slot E8400 44306 plastic filler panels that do not provide EMC protection Getting Started 13 Installing VXI Instruments Using VIC Installing C Size Instruments 14 Getting Started Note To comply with EMC requirements order 1 slot metal filler panels E8400 60202 to cover empty mainframe slots If you use an E1406A Command Module or an E8491A PC Link to VXI Interconnect in Slot 0 of the mainframe you can use the Agilent Technologies VXI Installation
13. two rack mount front panel labels in the E8397A kit and carefully place them over the existing labels Press to adhere Lay Mainframe on Side EIRA GR J Install Rack Mount Labels Ss C a P Figure 2 4 Installing Rack Mount Labels 24 Installing Mainframe Options Chapter 2 Attach Handles to Rack Mount Adapters Figure 2 5 1 Attach a handle to the LH Rack Mount Adapter using two 10 32 X 0 375 flat head screws 2 Attach a handle to the RH Rack Mount Adapter using two 10 32 X 0 375 flat head screws NOTE The rack mount adapter without the opening is the left side LH Rack Mount Adapter and the rack mount adapter with the cable routing opening is the right side RH Rack Mount Adapter Right Hand Rack Left Hand Rack Mount Adapter Mount Adapter Y Flat Head Screws Flat Head Screws Figure 2 5 Attaching Handles to the Rack Mount Adapters Chapter 2 Installing Mainframe Options 25 Attach Rack Mount Adapters to Mainframe Figure 2 6 NOTE The mainframe can be recess mounted up to 207 mm approximately 8 inches in 23 mm approximately 1 inch increments When recessing the mainframe make sure that six screws are used to attach the adapters on each side For mass interconnect systems the mainframe must generally be recessed Check the mass interconnect system manufacturer s instructions BEFORE attaching the
14. 0 32 x 0 375 pan head screws 2 Connect the RH Rack Slide chassis section to the RH Rack Mount Adapter using four 10 32 x 0 375 pan head screws l Left Hand Chassis Section Figure 2 8 Connecting Chassis Section to Rack Mount Adapters 28 Installing Mainframe Options Chapter 2 Chapter 2 Install Channel Nuts for Rack Mount Adapters Figure 2 9 1 Determine the location for the mainframe in the cabinet The E8408A mainframe requires 3 EIA Units and should be placed between the EIA markers 2 Selectthe holes for the LH and RH Rack Mount Adapters See Figure 2 9 for the RH Rack Mount Adapter hole locations LH Rack Mount Adapter hole locations are the same 3 Slide two 10 32 Self Retained Nuts provided in the Rack Mount Cable Route Adapter Kit over the selected holes in the right hand front vertical column of the cabinet These nuts will be used to connect the RH Rack Mount Adapter 4 Slide two 10 32 Self Retained Nuts over the selected holes in the left hand front vertical column of the cabinet These nuts will be used to connect the LH Rack Mount Adapter Right Hand Front Vertical Column of the Cabinet Right Hand Rack Mount Adapter UU 10 32 2 d Self Retained Nuts NM T W uy Units Figure 2 9 Installing Rack Mount Adapter Channel Nuts Installing Mainframe Options 29 Install Rack Slide Stationary Sections Figure 2 10 NOTE Fornon cabinets
15. 1451A E1452A I O Modules E1476A B SONET Modules E1740A Time Interval Analyzer Chapter 1 Installing A and B Size Instruments Tighten the Two Retaining Screws Extraction Levers gt Sil Slide the Module into the Mainframe Until it Plugs into the Backplane Connectors Seat the Module by Pushing in the Extraction Levers Figure 1 4 shows typical installation of A and B size instruments E1403B A B size Module Carrier extends the P1 connector on the VXIbus backplane and mounts the A B size modules flush with C size modules This carrier is recommended for Agilent B size slave only devices which have the P1 connector E1407A A B Module Carrier extends the P1 and P2 connectors on the VXIbus backplane This carrier is recommended for B Size slave only devices which have the P1 P2 connectors Figure 1 3 Installing C Size Instruments in the E8408A Mainframe Chapter 1 Getting Started 15 Slide the Adapter Module into the Slide the B Size Module into the Mainframe Until it Plugs into Adapter Module Until it Connects the Backplane Connector Figure 1 4 Installing A and B Size Instruments in the E8408A Mainframe Front Panel Features Figure 1 5 shows the E8408A Front Panel features and the table describes the features When the mainframe is connected to an AC supply and the Power Switc
16. 2 Install Screws To install the screws firmly press the screw onto a driver This prevents the screw from falling off as you reach into the mainframe Placing a sheet of paper under the backplane connectors will catch screws if they fall 3 Tighten Screws Tighten the screws by turning them clockwise The screws are thread forming and will go in slowly when you install them the first time Backplane Backplane Backplane Shields N Shields Red f gt 1 Backplane Figure 2 22 Positioning the Backplane Shields on the Connectors 44 installing Mainframe Options Chapter 2 Installing the Airflow Restricter Kit This section shows how to install an optional airflow restricter kit E8400 80917 in the E8408A mainframe To install the kit 1 Disconnect AC power from the mainframe and place mainframe on bench 2 As required remove filler panels and or installed VXI instruments from the mainframe 3 Follow the steps in Figure 2 23 to install the Airflow Restricter Loosen the K gt C Place the Airflow Restricter Fastening Screw JO Panel Clips into the Opening NY of the Slot to be Covered Slide the Airflow Restricter Panel Clip as Shown Slide the Airflow Restricter Panel Clip as Shown K Te NM L Tighten the LES TO GA Fastening Screw vy KK Ze Figure 2 23 Installing the Airflow Restricter Chapter 2 Install
17. 2A 0 8A 0 8A 24V 2A 2A 1 0A 1 0A 246 2A 0 5A 1 0A 0 25A 5 2V5 2A 5 8A 1 0A 1 0A 2V 2A 2A 1 0A 1 0A Specifications apply at the mainframe backplane 0 55 C Imp Rated mainframe peak DC output current as defined by the VXlbus Specification lup Rated mainframe peak to peak dynamic current as defined in the VXIbus Specification by a current vs frequency curve 5V must have 1 0A minimum load for other supplies to be in regulation specification Standard Only Maximum 5V current can be increased if 5 2V load is less than 2 0A maximum Maximum 17 0A 1 2 2 0A Load on 5 2V Option 001 Only Maximum 24V current can be increased if 5 2V load is less than 5 8A maximum Maximum 0 5A 0 24 5 8A Load on 5 2V 7 Maximum 12V current can be increased if 12V load is less than 2 0A maximum Maximum 2 7A 1 2 2 0A Load on 12V 60 E8408A Product Specifications Appendix A Output Voltage Voltage Allowed Variation Ripple Noise DC Load Induced Ripple Noise 5V 0 25V 0 125V 50mV 50mV 12V 0 60V 0 36V 50mV 50mV 12V 0 60V 0 36V 50mV 50mV 24V 1 20V 0 72V 150mV 150mV 24V 1 20V 0 72V 150mV 150mV 5 2V 0 26V 0 156V 50mV 50mV 2V 0 10V 0 10V 50mV 50mV Specifications apply at the backplane 0 55 C Power Input Area Specification Comments Input Voltage 90 Vac min 140 Vac max Frequency 47 Hz min
18. 75000 Series C Agilent E8408A C Size VXI Mainframe pm 1 User and Service Manual Where to Find it Online and Printed Information This manual is distributed as follows Agilent Universal Instrument Drivers CD ROM included with the mainframe World Wide Web at http www agilent com find products For related information VXIbus Configuration Guide System installation hardware software Supplied with Agilent Command Module Embedded Controllers and VXLink ee 00 e zu Agilent Technologies Manual Part Number E8408 90001 Printed in Malaysia E0912 NOTICE In August 2014 Agilent Technologies former Test and Measurement business became Keysight Technologies This document is provided as a courtesy but is no longer kept current and thus will contain historical references to Agilent For more information goto www keysight com KEYSIGHT TECHNOLOGIES Contents E8408A C Size VXI Mainframe User and Service Manual CE UE LACS RSR Eee ER RAR Re 5 Agilent Technologies Warranty Statement 5 U S Govermment Restricted RIMIS rame RR e irt SU e nn RR a 5 Te MIO UTER 5 Documentation Holo cis irent ben pr bid err as beue t dao e d e rbd ta kevin Pda ttre 6 DoE So UVM PRENNE RE m LU 6 lip m 6 Be RS EST OI PST Um 7 Chapter 1 Getting
19. Consultant VIC to assist you in installing VXI instruments To use VIC follow these steps For actual installation of instruments into the mainframe see Installing C Size Instruments or Installing A or B Size Instruments 1 Insert the Agilent Technologies Universal Instrument Drivers CD received with your Slot 0 card into your CD ROM drive and wait a few seconds for the setup instructions to appear 2 If the setup screen does not appear in a few seconds click Start Run and type lt drive gt setup exe in the Run box where drive is your CD ROM drive 3 When the main menu appears click Launch Installer or View Manuals Click Next on the next screen and Yes on the third screen to go to the Choose Type of Installation screen 4 Click the RUN VIC icon and then click Next to start VIC Follow the onscreen instructions to install VXI instruments in the mainframe Figure 1 3 shows typical installation of C Size instruments Power requirements for the following C Size instruments are NOT supported in the Standard version of the E8408A and should NOT be installed in an E8408A mainframe The E1430A and E1437A digitizers can be installed if you have Option 001 E1426A 500 MHz 4 Channel VXI Oscilloscope E1428A 1 GSa s VXI Oscilloscope E1429A B 20 MS s A Ds E1430A 10 MS s 23bit A D with Filter and Memory E1437A 20 MSa s Digitizer with DSP E1445A Arbitrary Function Generator E1450A Timing Module E
20. Nuts over the selected holes in the left hand front vertical column of the cabinet These nuts will be used to connect the LH Rack Mount Adapter Right Hand Front Vertical Column of the Cabinet Right Hand Rack Mount y Adapter P 10 32 74 Self Retained Nuts Figure 2 17 Installing Rack Mount Adapter Channel Nuts 38 Installing Mainframe Options Chapter 2 WARNING Pan Head Screw Right Vertical Column Front Chapter 2 Install Support Rails Figure 2 18 1 Open the Support Rail Kit E3663A 2 Hook the Support Rails in position in the Rack Cabinet The Support Rails occupy the bottom EIA unit of the three units required for the E8408A mainframe 3 Slide a 10 32 Self Retained Nut provided in the Support Rail Kit over the appropriate holes in the front and rear right hand vertical columns in the cabinet Repeat for the left hand vertical columns 4 Secure each Support Rail to the columns as shown in Figure 2 18 To avoid possible injury or equipment damage always secure the Support Rails with mounting screws before installing equipment Right Hand Support Rail Pu LSJ Self Retained p Nuts Right Vertical Column Rear Figure 2 18 Installing Support Rails Installing Mainframe Options 39 Installing the Mainframe Install the Mainframe in the Rack Cabinet Figure 2 19 1 With one person on each side li
21. OCQAUING 31 Installing Intermodule Chassis Shields issus 40 Parts GIS e 40 prc FOC US c Em 40 Contents 3 4 Installing Backplane Connector Shields nn 41 I 2E o 41 aS PMNS POO ue ddetaxesavauser tint disaidewauconedssrdaunasiscsunbeiedisemcedlasisituwa est 42 installing the Airflow Resiricter dj emm 43 Chapter 3 Servicing the Mainframe ss 45 rre HR ci e 45 Replaceable Farts ak merum 45 Problem ISOLA si uiid RE S itti dd a at d a PR ad 46 Replacing Mainframe Componefttis rrr E ePEPPRIe SS SEFEEPIEREEERE PHAR S Roy AUN EF EDU PEEAR URS 46 Removing Mainframe Accessory Items issus 47 Removing the Mainframe Cover Luise esu eterna ket aaa na 44 Xu E RM A aa ERA Rad tud 48 Replacing the Power Supply Assembly ui 49 Replacing the Monitor Board MEN p 50 Replacing the Fan Assembly iii cii esce bol ea inii 51 Replacing the Front Panel Tapped Strip esses 52 Replacement Power Gorda sn GS ERE CER Rai O a PARE HERR ICH ANM dE 53 eanna MTU e 53 Appendix A E8408A Product Specifications eese 55 Produc DeSEHDUGDL stunt wad we ee ae 55 Produet DIISIESIENES acs ps er o ceti ndi a t en ka Ub d ra ead an ode Rd ieee 55 LPNS and AGGSSSOBS tess cn o IH diabetes decimi em m E I ER e
22. Rack Mount Cable Route Adapter Kit Must also use either E3663A Support E8397A Rail Kit or 1494 0413 Rack Slide Kit Field Installation Kit for Rack Mount Cable Route Adapter Kit E8408 80923 Support Rail Kit Used with E8397A E3663A Rack Slides to mount in rack cabinet Used with E8397A 1494 0413 For non Agilent cabinets must also use 1494 0061 End Brackets 58 E8408A Product Specifications Appendix A General VXI Specifications Category Specifications VXI Device Type Mainframe Data Transfer Bus All per VXIbus Specification Rev 1 4 Size C Size Slots 4 available Connectors P1 P2 Shared Memory n a VXI Buses per VXIbus Spec Rev 1 4 Monitor Specifications Interface Power Switch Voltages Indicator Description Stdby AC power is applied to the mainframe but not to installed VXI instruments On AC power applied to the mainframe and DC to installed VXI instruments Green all voltages on the VXI backplane are present Does not necessarily indicate that voltages are within the VXI specification Flashing Amber one or more voltages is missing or incorrect The backplane voltages and allowed variations are listed below Voltage Allowed Variation 5V 4 875V to 5 125V 12V 11 64V to 12 6V 12V 12 6V to 11 64V 24V 23 28V to 25 2V 24V 25 2V to 23 28V 5 2V 5 46V to 5 044V 2V 2 1V to 1 9V Temp Indicator Gre
23. TY OR CONDITIONS OF MERCHANTABILITY SATISFACTORY QUALITY AND FITNESS FOR A PARTICULAR PURPOSE 8 Agilent will be liable for damage to tangible property per incident up to the greater of 300 000 or the actual amount paid for the product that is the subject of the claim and for damages for bodily injury or death to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Agilent product 9 TO THE EXTENT ALLOWED BY LOCAL LAW THE REMEDIES IN THIS WARRANTY STATEMENT ARE CUSTOMER S SOLE AND EXLUSIVE REMEDIES EXCEPT AS INDICATED ABOVE IN NO EVENT WILL AGILENT OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT SPECIAL INCIDENTAL CONSEQUENTIAL INCLUDING LOST PROFIT OR DATA OR OTHER DAMAGE WHETHER BASED IN CONTRACT TORT OR OTHERWISE FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND THE WARRANTY TERMS CONTAINED IN THIS STATEMENT EXCEPT TO THE EXTENT LAWFULLY PERMITTED DO NOT EXCLUDE RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU U S Government Restricted Rights The Software and Documentation have been developed entirely at private expense They are delivered and licensed as commercial computer software as defined in DFARS 252 227 7013 Oct 1988 DFARS 252 211 7015 May 1991 or DFARS 252 227 7014 Jun 1995 as a commercial item as defined in FAR 2 101 a or as Restricted
24. able DO NOT operate the product in an explosive atmosphere or in the presence of flammable gases or fumes For continued protection against fire replace the line fuse s only with fuse s of the same voltage and current rating and type DO NOT use repaired fuses or short circuited fuse holders Keep away from live circuits Operating personnel must not remove equipment covers or shields Procedures involving the removal of covers or shields are for use by service trained personnel only Under certain conditions dangerous voltages may exist even with the equipment switched off To avoid dangerous electrical shock DO NOT perform procedures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features built into this product have been impaired either through physical damage excessive moisture or any other reason REMOVE POWER and do not use the product until safe operation can be verified by service trained personnel If necessary return the product to Agilent for service and repair to ensure that safety features are maintained DO NOT service or adjust alone Do not attempt internal service or adjustment unless another person capable of rendering first aid and resuscitation is present DO NOT substitute parts or modify equipment Because of the danger of introducing additional hazards do not install substitute parts or perform any unauthoriz
25. adapters to the mainframe For easier installation place the screws in the openings before you begin attaching the adapters 1 Attach the LH Rack Mount Adapter to the left side of the mainframe using six M3 X 0 5 pan head screws 2 Attach the RH Rack Mount Adapter to the right side of the mainframe using six M3 X 0 5 pan head screws Pan Head Screws Left T Rack Mount Adapter Right Hand Rack Mount Adapter Figure 2 6 Attaching the Rack Mount Adapters to the Mainframe 26 Installing Mainframe Options Chapter 2 Set Up Rack Cabinet Remove Rack Slide Chassis Sections Figure 2 7 1 Open the Rack Slide Kit 1494 0413 2 Remove the Left Side LH Rack Slide Chassis Section from the Intermediate Section 3 Slide the Intermediate Section back into the Stationary Section 4 Remove the Right Side RH Rack Slide Chassis Section from the Intermediate Section 5 Slide the Intermediate Section back into the Stationary Section NOTE Be sure to mark the chassis sections as RH or LH so you can install them on the proper side of the Rack Mount Adapters The Rack Slide is marked with RH or LH on the label to identify it Right Rack Slide Shown Intermediate Sedion Figure 2 7 Removing Rack Slide Chassis Section Chapter 2 Installing Mainframe Options 27 Connect Chassis Section to Rack Mount Adapters Figure 2 8 1 Connect the LH Rack Slide chassis section to the LH Rack Mount Adapter using four 1
26. also includes Option 001 that provides more current from the 5 2V source This option allows the mainframe to be used for VXI modules that require more current such as the E1430 E1437 E1438 and E1439 digitizers The Portable Configuration consists of the Standard Configuration plus a carrying strap and an accessory pouch to carry a laptop PC cables etc when transporting the mainframe The Rack Mount Configuration consists of the Standard Configuration with the protective bumpers and carrying handle removed For this configuration the mainframe can be rack mounted using the Rack Mount Cable Route Adapter Kit with either the Rack Slide Kit or the Support Rail Kit Figure 1 1 shows the options and accessories available for the E8408A mainframe See Appendix A E8408A Product Specifications for description and part numbers for mainframe and accessories options Chapter 1 Getting Started 9 Standard Configuration Portable Configuration Protective Bumpers Carrying A nes Handle ET Carrying Strap And Accessory Pouch E8396A a Protective 4 Bumpers I Protective Bumpers gt Rack Mount Configuration Rack Slide Kit Support Rail Kit 1494 0413 OR E3663A Right Hand Shown Right Hand Shown gt i 1 Rack Mount Cable _ Adapter Kit E8397A Figure 1 1 E8408A Mainframe Configurations Op
27. computer software as defined in FAR 52 227 19 Jun 1987 or any equivalent agency regulation or contract clause whichever is applicable You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Agilent standard software agreement for the product involved Trademark Information Microsoft Windows amp and Windows NT are registered trademarks of the Microsoft Corporation All other brand and product names are trademarks or registered trademarks of their respective companies Agilent Technologies E8408A C Size VXI Mainframe User and Service Manual Edition 2 Copyright 1999 2000 2015 Agilent Technologies Inc All rights reserved Documentation History All Editions and Updates of this manual and their creation date are listed below The first Edition of the manual is Edition 1 The Edition number increments by 1 whenever the manual is revised Updates which are issued between Editions contain replacement pages to correct or add additional information to the current Edition of the manual Whenever a new Edition is created it will contain all of the Update information for the previous Edition Each new Edition or Update also includes a revised copy of this documentation history page Edition 5 od ete Sr EC b E EN cut March 1999 Editi012 5 5 5 o tto dut tye Spice Gack nae ue ed February 2000 Edition 2 Rev lis 2 0 3 dis RR m Ren February 2012 Edition R
28. d Modules installed in other slots Measurements taken at 1 500 m altitude Pressure Drop Across Module mm H20 Airflow L S Figure A 2 E8408A Mainframe Cooling Capacity 62 E8408A Product Specifications Appendix A Safety Environmental Specifications This section describes safety and environmental specifications for the E8408A mainframe including Environmental Specifications Electromagnetic Compliance Safety Specifications Power Supply Protection Repair Environmental Specifications Area Specification Operating Indoor Sheltered location where air temperature and humidity is controlled within this Location product s specifications and the equipment is protected against direct exposure to climatic conditions such as direct sunlight wind rain snow sleet and icing water spray or splash hoarfrost or dew typically indoors Pollution degree 2 Temperature Operating Temperature Range 0 C to 55 C Storage Temperature Range 40 C to 70 C Humidity Operating Humidity Range Non condensing Up to 95 RH from 0 C to 40 C Up to 65 RH from 40 C to 55 C Storage Humidity Range Up to 95 RH from 0 C to 55 C Up to 90 RH from 55 C to 65 C Shock End Use Handling Half sine waveform lt 3 msec duration A v 160 cm sec minimum Transportation Trapezoidal waveform A v 605 cm sec 30 g minimum Vibration Operating and Functional 5 to 500 Hz
29. d Screw Handle Screws 2680 0100 12 M3 X 0 5 Pan Head Screw Adapter to Mainframe Screws 0515 0372 4 Adapter Dress Screw 0570 1577 Self Retained Nut for Rack Mount Adapters 0590 0804 Front Panel Label Set Rack Mount Orientation E8408 84305 Rack Slide Kit 1494 0413 1 pair Rack Slides General Devices p n CC1502 99 0016 12 10 32 X 375 Pan Head Screw 0515 0904 4 10 32 X 50 Flat Head Screw 0515 0956 8 10 32 Self Retained Nut 0590 0804 For non Agilent cabinets must add 1494 0061 End Brackets Installation Procedure 22 Installing Mainframe Options There are three main steps to rack mount the E8408A mainframe using the Rack Mount Cable Route Adapter Kit E8397A and the Rack Slide Kit 1494 0413 Configure the Mainframe Set up Rack Cabinet nstall the Mainframe Chapter 2 Configure the Mainframe Remove Mainframe Items as applicable Figure 2 3 1 Remove the carrying handle carrying strap 2 Remove the two protective bumpers 3 Remove the Accessory Pouch Remove Carrying Handle Strap Remove Protective Bumper from Each End Remove Accessory Pouch Figure 2 3 Removing Mainframe Items Chapter 2 Installing Mainframe Options 23 Install Rack Mount Labels Figure 2 4 1 Lay mainframe on its left side 2 Open the Rack Mount Cable Route Adapter Kit E8397A 3 Peel the protective backing from the
30. e cooling 11 portable configuration 19 rack mounting using rack slide kit 21 rack mounting using support rail kit 32 mainframe components replacing 48 monitor board replacing 52 O options and accessories 58 P parts list replaceable 47 peak and dynamic current 60 portable configuration mainframe 19 positioning mainframe for cooling 11 power cords replacement 55 power input 61 power requirements 11 power specifications 60 power supply cooling 11 protection 64 power supply assembly removing 51 problem isolation 48 product description 57 58 product dimensions 57 product overview 9 R rack mounting using rack slide kit 21 using support rail kit 32 removing mainframe cover 50 removing the power supply assembly 51 repair 65 repair mainframe 48 replaceable parts list 47 replacement power cords 55 replacing components removing mainframe cover 50 removing mainframe items 49 removing power supply assembly 51 replacement power cords 55 replacing fan 53 replacing front panel tapped strip 54 replacing monitor board 52 replacing mainframe components 48 replacing the fan 53 Index 67 replacing the front panel tapped strip 54 replacing the monitor board 52 restricted rights statement 5 S safety symbols 6 safety environmental specifications 63 specifications backplane specifications 59 cooling 62 cooling chart 62 cooling table 62 dynamic and peak current 60 e
31. ed modification to the product Return the product to Agilent for service and repair to ensure that safety features are maintained Declaration of Conformity Declarations of Conformity for this product and for other Agilent products may be downloaded from the Internet There are two methods to obtain the Declaration of Conformity e Goto http regulations about keysight com DoC search htm You can then search by product number to find the latest Declaration of Conformity Alternately you can go to the product web page www keysight com find E8408A click on the Document Library tab then scroll down until you find the Declaration of Conformity link Notes Chapter 1 Getting Started Chapter Overview Chapter contents are Product Overview cse peek tam EURE RONA RSS 9 Initial Mainframe Setup 11 Installing VXI Instruments 13 Front Panel Features 16 Product Overview The E8408A VXI mainframe is designed in full compliance with V XIbus specification revision 1 4 and VMEbus system specification revision C 1 The mainframe has three configurations Standard Portable and Rack Mount The Standard Configuration consists of the 4 slot C size mainframe with front panel monitor protective bumpers and a carrying handle Optional filler panels and shields can be added to the Standard Configuration The Standard Configuration
32. em RM 56 General VXI Speciicati NS A 57 Monitor Specii catione RE TT T OO TS 57 Backplane SECOURS sranane nia a iiaa Eie Orie 57 Power OCC nana ten asc pe dio ur AR tte 58 Total LISE POWE e M 58 Peak and Dynamic CHROME a etant ensuite sania 58 snp i iaaa a a 59 Fowor Np cc 59 CooMa SCM el ONS e M 60 COOMO TADO Liaison Mandddiasaaaiess 60 OOT O a E 60 Safety Environmental SpecificatiOnS s sssiiisarssiersinu siisii i 61 Environmental Specifications ccccccrissncedsvsseesnscisstnoncadeessneacdaus REN ER RR RA k andada Eo Nn aaa 61 Electromagnetic Compliance Safety Specifications 62 osa desierit TT o OU T o 62 uilor ee 63 lor aee 65 Contents AGILENT TECHNOLOGIES WARRANTY STATEMENT AGILENT PRODUCT E8408A C Size VXI Mainframe DURATION OF WARRANTY 3 years 1 Agilent Technologies warrants Agilent hardware accessories and supplies against defects in materials and workmanship for the period specified above If Agilent receives notice of such defects during the warranty period Agilent will at its option either repair or replace products which prove to be defective Replacement products may be either new or like new 2 Agilent warrants that Agilent software will not fail to execute its programming instructions for the period specified above due to defects in material
33. en the power supply temperature is acceptable Flashing Amber the power supply is approaching thermal shutdown Fan Indicator Green mainframe cooling fan is operating properly Flashing Amber mainframe cooling fan is not operating Fan Speed Switch Full the mainframe cooling fan is operating at full speed Var iable the mainframe cooling fan is providing adequate cooling default setting Backplane Specifications Appendix A Automatic daisy chain jumpering for BUS GRANT and IACK signals Full differential distribution of CLK10 e ACFAIL and SYSRESET in full compliance with the VMEbus and VXIbus Specifications Surface mount construction and no sockets for maximum reliability E8408A Product Specifications 59 Power Specifications This section describes E8408A mainframe power specifications for the Standard and Option 001 configurations The total power delivered from the Option 001 configuration cannot exceed the total power delivered from the Standard configuration Total Usable Power Temperature Usable Power 90 264 Vac 0 40 C 175W 40 50 C 150 W 50 55 C 125 W 1 Total output before thermal shutdown or safety limitation Peak and Dynamic Current Voltage Peak Current lyp Amps Dynamic Current Ip Amps Standard Option 001 Standard Option 001 Standard Option 001 45y45 17A 19 4A 2 5A 2 5A 12V7 2 7A 2 7A 1 0A 1 0A 12V7 2A
34. enmark 250 VAC 8120 2956 10A Japan 100 VAC 8120 4753 12A India South Africa 250 VAC 8120 4211 10A Power cords supplied by Agilent have polarities matched to the power input socket on the instrument L line or active conductor also called live or hot N neutral or identified conductor E 7 earth or safety ground Cleaning Instructions The only cleaning required for the E8408A mainframe is to wipe the exterior of the mainframe with a clean damp cloth and dry with a clean dry cloth Do not attempt to clean any interior parts of the mainframe Chapter 3 Servicing the Mainframe 55 Notes 56 Servicing the Mainframe Chapter 3 Appendix A E8408A Product Specifications Product Description This section describes the E8408A mainframe including Product Dimensions Options and Accessories General VXI Specifications Mainframe Monitor Specifications Backplane Specifications Product Figure A 1 shows E8408A dimensions Empty mainframe weight is 8 6 Kg Dimensions 19 Ib without protective bumpers and 9 1 Kg 20 Ib with bumpers Weight with four average VXI modules installed is 15 9 Kg 35 Ib Maximum module weight is 3 5 Kg 7 7 lbs per slot to comply with shock and vibration specifications Heavier modules may be installed if the shock and vibration environment is less severe Protective Bumpers TOP i I l f I I i I d l l l l
35. ev 2 25 iore Ole ES EAA s August 2015 Safety Symbols Instruction manual symbol affixed to product Indicates that the user must refer to Alternating current AC the manual for specific WARNING or CAUTION information to avoid personal injury or damage to the product se Indicates hazardous voltages Indicates the field wiring terminal that must ae be connected to earth ground before operating the equipment protects against electrical shock in case of fault Direct current DC Calls attention to a procedure practice or condition that could cause bodily injury or death WARNING Calls attention to a procedure practice or Frame or chassis ground terminal typically CAUTION Se thatcol ae or LL connects to the equipment s metal frame m ii ecran E x WARNINGS The following general safety precautions must be observed during all phases of operation service and repair of this product Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design manufacture and intended use of the product Agilent Technologies assumes no liability for the customer s failure to comply with these requirements Ground the equipment For Safety Class 1 equipment equipment having a protective earth terminal an uninterruptible safety earth ground must be provided from the mains power source to the product input wiring terminals or supplied power c
36. ft the mainframe and slide the mainframe fully onto the Support Rails WARNING To prevent possible injury during rack mounting two people should lift the mainframe into the rack 2 Slide the mainframe into the rack until the Rack Mount Adapter flanges are against the rack s front vertical columns 3 Secure the mainframe to the rack using the four adapter dress Screws Support Rail Adapter Dress Screws Figure 2 19 Installing the Mainframe 40 Installing Mainframe Options Chapter 2 Chapter 2 Connect Route VXI Instrument Cables Figure 2 20 1 As required install VXI instruments in the mainframe See Chapter 1 Getting Started 2 Connect cable s to instrument front panel terminals as required 3 Route cables through the right side cable tray RH Rack Mount Adapter as required NOTE Cables should not be routed on the left side of the mainframe as the cables may interfere with proper airflow VXI Instruments Cable Tray Figure 2 20 Routing VXI Instrument Cables Installing Mainframe Options 41 Installing Intermodule Chassis Shields This section describes how to install Intermodule Chassis Shields E8400 80919 in the E8408A mainframe The chassis shield is the implementation of VXI revision 1 4 specification B 7 3 4 that allows grounded shielding between mainframe slots The shield is used to isolate VXI modules that generate electromagnetic interference EMI at
37. g the Mainframe 47 Problem Isolation Table 3 2 lists primary mainframe symptoms possible cause s and suggested actions Table 3 1 shows replacement part numbers Table 3 2 Isolating Problems in the E8408A Mainframe Symptom Possible Cause Action No mainframe power Power Supply fuse blown or defective Power Supply There are NO replaceable power line fuses on mainframe If the mainframe power supply fails to operate replace the Power Supply Assembly Short between some supplies can latch the Power Supply OFF Switch to Standby mode and then remove AC power Recovery can occur when AC power is reapplied and the mainframe Power switch is switched to ON f problem remains replace Power Supply Assembly Flashing Voltages Indicator Backplane voltage s may be incorrect Cycle mainframe Power switch Remove VXI instruments to ensure they are not the cause Check cable connections at Monitor and Backplane Board Replace Power Supply Assembly Flashing Temp Power supply temperature Mainframe is nearing automatic shutdown Indicator is high Check for proper air flow and clearance around the fan Check power requirements of installed VXI instruments Replace Fan Assembly as required Replace Power Supply Assembly as required Flashing Fan Fan has failed or fan Turn off mainframe to avoid possible overheating Indicator connection loose Check fan connectio
38. h is pressed ON the Voltages Temp and Fan lights flash Amber for about one second and then turn Green These lights should remain Green A flashing Amber light indicates an error condition In this 16 Getting Started Chapter 1 case see Chapter 3 Servicing the Mainframe Power at 1 als Voltages Temp Fan Stdb Full y bm I on Var OS CS C2 Figure 1 5 E8408A Mainframe Front Panel Features No Indicator Switch Description 1 Power Switch Stdby AC power is applied to the mainframe but not to installed VXI instruments On AC power applied to the mainframe and DC power to installed VXI instruments 2 Voltages Indicator Green all voltages on the VXI backplane are present Does not necessarily indicate voltages are within VXI specifications Flashing Amber one or more voltages is not present or is incorrect The backplane voltages and allowed variations are listed below Voltage Allowed Variation 5V 4 875V to 5 125V 12V 11 64V to 12 6V 12V 12 6V to 11 64V 24V 23 28V to 25 2V 24V 25 2V to 23 28V 5 2V 5 46V to 5 044V 2V 2 1V to 1 9V 3 Temp Indicator Green the power supply temperature is acceptable Flashing Amber the power supply is approaching thermal shutdown 4 Fan Indicator Green mainframe cooling fan is operating properly Fla
39. he Power On Standby switch is cycled However some over voltage conditions on the 5V and 12V can cause a shutdown that will not respond by cycling the Power On Standby switch In this case the Power On Standby switch should be placed in Standby position and AC power must be removed and reapplied before the mainframe will turn on again Short circuit behavior varies with the specific supply but in general Protection recovery should occur when the short is removed This table summarizes short circuit behavior for each supply voltage Supply Short Circuit Behavior Shorted 5V or Causes global shutdown After about 250 msec the supplies 12V will automatically attempt to turn on Global shutdown and restart attempts will continue until the fault is removed 24V Puts supply in foldback current limiting mode 24V Puts supply in foldback current limiting mode and shuts down the 12V 5 2V and 2V supplies 12V Continuously draws about 7 Amps and shuts down the 5 2V and 2V supplies 5 2V Continuously draws about 7 Amps Standard 5 2V Continuously draws about 12 Amps Option 001 2V Continuously draws about 5 Amps 64 E8408A Product Specifications Appendix A Appendix A Short to Other Protection Repair A short to other output will generally cause a global shutdown When the fault is removed the supplies will typically recover when the Power On Standby switch is cycled However if t
40. hers underneath the handle retainers 3 Attach the Accessory Pouch to the mainframe using four screws Then clip the carrying strap to the D rings Protective Carrying B Protective B nipsr Strap Bumpers Accessory v Pouch Carrying 9 Handle faal X Protective A i d Bumper 7 Screws Figure 2 1 Attaching Carrying Strap and Accessory Pouch Attaching Laptop PC A Laptop PC Platform is enclosed in the Accessory Pouch that you can use Optional tohold your Laptop PC secure when using the PC in an application with the mainframe in horizontal orientation such as on a car seat or lab benchtop The Laptop PC can be attached to the outside of the Accessory Pouch using the four hook and loop squares on the Laptop PC Platform To attach the Laptop PC to the Laptop PC Platform 1 Remove the Laptop PC Platform from the Accessory Pouch 2 Be sure the mainframe is in horizontal orientation Then attach the Laptop PC Platform to the outside of the Accessory Pouch using the two long hook and loop strips on the platform 20 Installing Mainframe Options Chapter 2 NOTE CAUTION Chapter 2 3 Follow the steps in Figure 2 2 to attach the PC to the Laptop PC Platform Once the hook and loop squares are attached to your Laptop PC they will not interfere with the use of the PC for other applications After initial installation you can easily remove the PC and reattach it a
41. his switch is cycled while the fault is still present some supplies may become active or global shutdown may occur again Conditions such as 5V to other output and 12V to 5 2V may cause a shutdown that will not respond to cycling the Power On Standby switch In this case the Power On Standby switch should be placed in Standby position and AC power must be removed and reapplied before the mainframe will turn on again Diagnosis and troubleshooting through the front panel monitor MTTR Mean Time to Repair Item Time MTTR power supply lt 20 min w mainframe and modules fully assembly installed in rack MTTR fan 30 min w mainframe and modules fully installed in rack E8408A Product Specifications 65 Notes 66 E8408A Product Specifications Appendix A Index E8408A C Size VXI Mainframe User and Service Manual A AC power requirements 11 airflow restricter kit installing 44 B backplane connector shields installing 43 C component failures 48 cooling chart 62 cooling specifications 62 cooling table 62 cooling mainframe 11 D display front panel 16 documentation history 6 E electromagnetic compliance specifications 63 64 environmental specifications 63 F fan replacing 53 front panel features 16 front panel tapped strip replacing 54 installing VXI instruments 13 A and B size 15 C size 14 L locating mainframe problems 48 M mainfram
42. inframe power and remove the AC power cord 2 Follow the steps in Figure 3 3 to remove the Power Supply Assembly 3 Reverse the steps in Figure 3 3 to replace the Power Supply Assembly and to restore the mainframe to its original configuration Remove 7 Screws from Rear Panel Pull Assembly Out to Remove From Mainframe Figure 3 3 Removing the Power Supply Assembly Chapter 3 Servicing the Mainframe 51 Replacing the This section gives guidelines to replace the Monitor Board E8408 66502 Monitor Board See Table 3 1 for a replacement part number 1 2 52 Servicing the Mainframe Turn off the mainframe and remove the AC power cord If the mainframe is rack mounted remove the mainframe from the rack and place the mainframe at a static controlled workstation Remove mainframe option items as required See Removing Mainframe Accessory Items Remove the mainframe cover See Removing the Mainframe Cover Follow the steps in Figure 3 4 to disconnect the ribbon cable and to remove the Monitor Board Reverse the steps in Figure 3 4 to install a replacement Monitor Board and to re connect the ribbon cable Replace the mainframe cover by reversing the steps in Removing the Mainframe Cover Reinstall mainframe options removed in step 3 OR reinstall the mainframe in the rack as required Reconnect the AC power cord 2 Remove Disconnect the T8 Screw Monitor Cable ail i pt pe Slide Board Ou
43. ing Mainframe Options 45 Notes 46 Installing Mainframe Options Chapter 2 Chapter 3 Servicing the Mainframe Chapter Overview This chapter contains information for troubleshooting and replacing selected components of the E8408A VXI mainframe This chapter includes the following information Replaceable Parts List 45 Problem Isolation 46 Replacing Mainframe Components 46 Cleaning Instructions 53 WARNING There are no user serviceable parts inside the mainframe Repair is limited to replacement of the components listed in Table 3 1 Replacement of these components must be performed at a static controlled workstation by trained service personnel only Replaceable Parts List These replacement parts are available from Agilent under the part numbers shown in Table 3 1 Contact Agilent to obtain replacement parts Table 3 1 E8408A Replaceable Part Numbers Component Part Number Power Supply Assembly includes Power Supply E8408 69005 and E8408 66503 Interconnect Board Standard remanufactured Power Supply Assembly includes Power Supply E8408 69006 and E8408 66513 Interconnect Board Opt 001 remanufactured Monitor Board E8408 66502 E8408 66502 Fan Assembly E8408 68500 Protective Bumper E8408 86001 Front Panel Tapped Strip E8408 21201 Chapter 3 Servicin
44. into place 3 Slide the mainframe into the cabinet until the Rack Mount Adapter flanges are against the front vertical columns of the cabinet You will need to depress the Chassis Section Slide Locks twice to fully slide the mainframe to the columns 4 Secure the mainframe to the rack using the four adapter dress Screws SET t if lm Chassis Section Slide Lock D A D ll x Adapter Dress Screws Figure 2 11 Installing the Mainframe Chapter 2 Installing Mainframe Options 31 Connect Route VXI Instrument Cables Figure 2 12 1 As required install VXI instruments in the mainframe See Chapter 1 Getting Started 2 Connect cable s to instrument front panel terminals as required 3 Route cables through the right side cable tray RH Rack Mount Adapter as required NOTE Cables should not be routed on the left side of the mainframe as the cables may interfere with proper airflow VXI Instruments Cable Tray Figure 2 12 Routing VXI Instrument Cables 32 Installing Mainframe Options Chapter 2 Rack Mounting Using Support Rail Kit This section contains instructions to mount the E8408A mainframe in a standard EIA cabinet using the Rack Mount Cable Route Adapter Kit E8397A and the E3663A Support Rail Kit E3663A The E3663A is only compatible with standard EIA cabinets Parts List rabie 2 3 shows the parts for the Standard Rack Mount Cable Routing Kit and the Sup
45. lectromagnetic compliance 63 64 environmental 63 front panel monitor 59 general 59 mainframe 57 options and accessories 58 output voltage 61 peak and dynamic current 60 power 60 power input 61 power supply protection 64 product description 57 58 product dimensions 57 58 repair 65 safety 63 total usable power 60 T total usable power 60 trademark information 5 V VXI instruments installing 13 W WARNINGS 6 47 warranty statement 5 68 Index
46. ns to P3 on Backplane Board Monitorindicators and fan remain OFF when Power switch is pressed Defective Monitor Board E8408 66502 or loose cable to Monitor Board Replace Fan Assembly as required Check cable connections at Monitor and Backplane Board Replace Monitor Board as required Slot 0 Card SYSFAIL Indicator VXI Instrument failure The Slot 0 card E1406A E8491A etc SYSFAIL indicator will turn ON and remain ON when a VXI instrument installed in the mainframe fails Replacing Mainframe Components This section gives guidelines to replace E8408A mainframe components including Removing Mainframe Accessory Items Removing the Mainframe Cover Replacing the Power Supply Assembly Replacing the Monitor Board Replacing the Fan Assembly Replacing the Front Panel Tapped Strip Replacement Power Cords 48 Servicing the Mainframe Chapter 3 Removing For some component removal replacement you may need to remove Mainframe mainframe accessory items as applicable see Figure 3 1 Accessory Items 1 Remove the carrying handle carrying strap 2 Remove the two protective bumpers 3 Remove the Accessory Pouch Remove Carrying Handle Strap Remove Protective Bumper from Each End K 3 lie Remove Accessory Pouch Figure 3 1 Removing Mainframe Accessory Items Chapte
47. port Rails Kit Table 2 3 Rack Mount Kit and Support Rail Kit Parts Lists Quantity Description Part Number Rack Mount Cable Route Adapter Kit E8397A 1 Rack Mount Adapter right E8408 61200 1 Rack Mount Adapter left E8408 61201 2 Handle for Rack Mount Adapters RAF p n 8344 1032 A34 4 10 32 X 0 375 Flat Head Screw Handle Screws 2680 0100 12 M3 X 0 5 Pan Head Screw Adapter to Mainframe Screws 0515 0372 4 Adapter Dress Screw 0570 1577 Self Retained Nut for Rack Mount Adapters 0590 0804 Front Panel Label Set Rack Mount Orientation E8408 84305 Support Rail Kit E3663A 2 Support Rail E3663 00001 4 10 32 Self Retained Nut 0590 0804 4 10 32 X 0 5 Pan Head Screw 2680 0278 Installation There are three main steps to rack mount the E8408A mainframe using the Procedure Rack Mount Cable Route Adapter Kit E8397A and the Support Rail Kit E3663A Configure the Mainframe Set up Cabinet Racks nstall the Mainframe Chapter 2 Installing Mainframe Options 33 Configure the Mainframe Remove Mainframe Items as applicable Figure 2 13 1 Remove the carrying handle carrying strap 2 Remove the two protective bumpers 3 Remove the Accessory Pouch Remove Carrying Handle Strap Remove Protective Bumper from Each End Remove Accessory Pouch Figure 2 13 Removing Mainframe Items
48. r 3 Servicing the Mainframe 49 Removing the For some component removal replacement you may need to remove Mainframe Cover mainframe cover see Figure 3 2 1 2 Turn off the mainframe and remove the AC power cord If mainframe is rack mounted remove the mainframe from the rack and place the mainframe at a static controlled workstation Remove mainframe items as required See Removing Mainframe Items Remove the ten M3x6 flat head screws five on each side of the mainframe Slide the mainframe cover toward the rear of the mainframe and then lift the mainframe cover off the mainframe Om Slide Cover Toward Rear of Mainframe Slide Cover Up to Remove from Mainframe Remove 5 Screws from Each Side Figure 3 2 Removing the Mainframe Cover 50 Servicing the Mainframe Chapter 3 Replacing the This section gives guidelines to replace the Power Supply Assembly Power Supply E8408 69005 Standard or E8408 90006 Option 001 in the E8408A mainframe see Figure 3 3 Assembly NOTE Although the Power Supply Assembly contains the power supply and the Interconnect Board do NOT remove the Power Supply or the Interconnect Board from the Assembly Return the entire Power Supply Assembly to Agilent for a remanufactured Power Supply Assembly See Table 3 1 for replacement part numbers Be sure to observe static free procedures when removing installing the Power Supply Assembly 1 Turn off ma
49. rom the rack and place the mainframe at a static controlled workstation Remove mainframe accessory items as required See Removing Mainframe Accessory Items Remove the mainframe cover See Removing the Mainframe Cover Remove the Front Panel Tapped Strip as shown in Figure 3 6 Install the replacement Front Panel Tapped Strip by reversing the steps in Figure 3 6 Replace the mainframe cover by reversing the steps in Removing the Mainframe Cover Reinstall mainframe accessory items removed in step 3 OR reinstall the mainframe in the rack as required Reconnect AC power cord 2 D Remove Disconnect the T8 Screw Monitor Cable p moba EF CF re Remove 4 3 Screws Slide the Front E Panel Display Out Slide Front Panel Tapped Strip Out from Front Panel Display Figure 3 6 Removing the Front Panel Tapped Strip Chapter 3 Replacement Power Table 3 3 lists the power cords rated for use with the E8408A mainframe If Cords it is necessary to replace the power cord obtain the appropriate cord listed in the table or use a cord with the same voltage and current ratings Table 3 3 Replacement Power Cords for E8408A VXI Mainframe Country Voltage Part Number Rated Amps U K 250 VAC 8120 1351 10A Australia 250 VAC 8120 1369 10A Europe 250 VAC 8120 1689 10A U S Canada 125 VAC 8120 1378 10A U S Canada 250 VAC 8120 5338 10A Switzerland 250 VAC 8120 2104 10A D
50. s needed 4 As desired connect an interface cable from the PC to the Slot 0 card in the mainframe Use Care When Mainframe is in Vertical Orientation Do not attempt to move or use an attached Laptop PC with the mainframe in vertical orientation as the PC can easily fall off or be detached when the mainframe is in vertical orientation Place Laptop PC Onto the Laptop PC Platform and Press Down to Secure the PC to the Platform Laptop PC P D a Accessory Ta ptop PC Pouch Opened Platformi CT Peel Off the Four Protective Papers from the Foot Pads Figure 2 2 Attaching Laptop PC to Accessory Pouch Installing Mainframe Options 21 Rack Mounting Using Rack Slide Kit Parts List This section contains instructions to mount the E8408A mainframe in a standard EIA cabinet using the Rack Mount Cable Route Adapter Kit E8397A and the Rack Slide Kit p n 1494 0413 The rack slide kit allows you to extend the mainframe from the cabinet for easier access to installed instruments Table 2 2 shows the parts for the Standard Rack Mount Cable Routing Kit and for the Rack Slide Kit Table 2 2 Rack Mount Kit and Rack Slide Kit Parts Lists Quantity Description Part Number Rack Mount Cable Route Adapter Kit E8397A 1 Rack Mount Adapter right E8408 61200 1 Rack Mount Adapter left E8408 61201 2 Handle for Rack Mount Adapters RAF p n 8344 1032 A34 4 10 32 X 0 375 Flat Hea
51. shing Amber mainframe cooling fan is not operating 5 Fan Speed Switch Full the mainframe cooling fan is operating at full speed Var iable the mainframe cooling fan is providing adequate cooling default setting WARNING Hazardous voltages may be present on modules Prevent Chapter 1 electrical shock by covering all unused slots and securely fastening all faceplates Getting Started 17 AVERTISSEMENT Des tensions dangereuses peuvent tre pr sentes sur les modules viter les chocs lectriques en couvrant tous les emplacements inutilis s et fixer solidement tous les plastrons caches 18 Getting Started Chapter 1 Chapter 2 Installing Mainframe Options Chapter Overview NOTE This chapter contains procedures to install hardware options available with the E8408A mainframe including Converting to Portable Configuration 17 Rack Mounting Using Rack Slide Kit 20 Rack Mounting Using Support Rail Kit 31 Installing Intermodule Chassis Shields 40 Installing Backplane Connector Shields 41 Installing the Airflow Restricter Kit 43 Mainframe illustrations in this chapter may show empty mainframes However you may receive integrated pre configured VXI systems with VXI instruments modules already installed in the mainframe In general you will NOT need to remove
52. t to Remove Figure 3 4 Removing the Monitor Board Chapter 3 Replacing the Fan rnis section gives guidelines to replace the Fan Assembly in the E8408A Assembly mainframe see Figure 3 5 See Table 3 1 for a replacement part number 1 Turn off the mainframe and remove the AC power cord NOTE Be sure to observe static free procedures when removing installing the Fan Assembly 2 Remove mainframe accessory items as required See Removing Mainframe Accessory Items 3 Remove the Power Supply Assembly See Replacing the Power Supply Assembly 4 Remove the Fan Assembly by following the steps in Figure 3 5 5 Install the replacement fan assembly by reversing the steps in Figure 3 5 6 Reinstall the Power Supply Assembly by reversing the steps in Replacing the Power Supply Assembly 7 Reinstall mainframe accesssory items removed in step 2 Reconnect AC power cord P3 Connector Backplane Remove Fan Wires from P3 Connector Remove Four Screws and Lift Fan Out of Mainframe Figure 3 5 Removing the Fan Assembly Chapter 3 Servicing the Mainframe 53 Replacing the Front Panel Tapped Strip 54 Servicing the Mainframe This section gives guidelines to replace the Front Panel Tapped Strip See Table 3 1 for a replacement part number 1 2 XI 2s Turn off the mainframe and remove the AC power cord If the mainframe is rack mounted remove the mainframe f
53. the VXI instruments to perform installation procedures in this chapter However to install shields you must first remove any installed instruments To do this reverse the procedure in Installing VXI Instruments in Chapter 1 Getting Started to remove the instruments Converting to Portable Configuration Parts List Chapter 2 The E8408A mainframe can converted to the Portable Configuration by adding a carrying strap and an accessory pouch E8396A to the Standard Configuration see Figure 2 1 As desired you can attach a Laptop PC to the Laptop PC Platform shipped with the E8396A Table 2 1 shows the parts required to convert the E8408A mainframe Standard Configuration to the Portable Configuration Table 2 1 Portable Configuration Parts List Description Part Number Portable Configuration Includes E8396A Carrying Strap w attachment hardware E8408 51201 Accessory Pouch w attachment hardware E8408 54401 Laptop PC Platform E8408 54101 Installing Mainframe Options 19 Installation There are two main steps to convert to Portable Configuration for the Procedure E8408A mainframe Attach Carrying Strap and Accessory Pouch Attach Laptop PC Optional Attaching Carrying Strap 1 Remove the two M5 x 10 flat head screws from the carrying handle and Accessory Pouch 2 Attach Carrying Strap D rings to the Carrying Handle using the two D rings and the two M5 X 16 pan head screws Be sure to replace the was
54. tions 10 Getting Started Chapter 1 Initial Mainframe Setup Note AC Power Requirements WARNING Positioning the Mainframe for Adequate Cooling Chapter 1 The E8408A mainframe is shipped from the factory ready to use This section describes procedures for initial mainframe setup including AC Power Requirements Positioning the Mainframe for Adequate Cooling Connecting Permanent Earth Ground This chapter shows initial setup procedures for the Standard Configuration See Chapter 2 Installing Mainframe Options for information on rack mounting the mainframe and installing mainframe options The E8408A mainframe can be operated at line voltages of 90 140 Vac 47 Hz to 440 Hz or 90 264 Vac 47 66 Hz The mainframe ships with a power cord and with a fuse installed in the power supply The fuse is suitable for all line voltages but is not accessible Appendix A E8408A Product Specifications contains input power specifications The power cord is the only way to disconnect the mainframe from AC power Therefore the power cord must be accessible to the operator at all times When the E8408A mainframe is mounted in a system cabinet the power cord need not be accessible since the cabinet must have its own disconnect device Some fault conditions require removal of AC power for recovery to occur See Chapter 3 Servicing the Mainframe A cooling fan provides air for module and power supply cooling
55. used to attach the adapters on each side For mass interconnect systems the mainframe must generally be recessed Check the mass interconnect system manufacturer s instructions BEFORE attaching the adapters to the mainframe For easier installation place the screws in the openings before you begin attaching the adapters 1 Attach the LH Rack Mount Adapter to the left side of the mainframe using six M3 X 0 5 pan head screws 2 Attach the RH Rack Mount Adapter to the right side of the mainframe using six M3 X 0 5 pan head screws Pan Head Screws es WZ d Mi EZ A 2 i Left Hand Le Rack Mount Adapter Right Hand Rack Mount 4E Adapter Pan TS Screws Figure 2 16 Attaching the Rack Mount Adapters to the Mainframe Chapter 2 Installing Mainframe Options 37 Set Up Rack Cabinet Install Channel Nuts for Rack Mount Adapters Figure 2 17 1 Determine the location for the mainframe in the cabinet The E8408A mainframe requires 3 EIA Units and should be placed between the EIA markers 2 Selectthe holes for the LH and RH Rack Mount Adapters See Figure 2 17 for the RH Rack Mount Adapter hole locations LH Rack Mount Adapter hole locations are the same 3 Slide two 10 32 Self Retained Nuts provided in the Rack Mount Cable Route Adapter Kit over the selected holes in the right hand front vertical column of the cabinet These nuts will be used to connect the RH Rack Mount Adapter 4 Slide two 10 32 Self Retained