Home

Print The Above Information

image

Contents

1. Re install the ventilator retaining screw into the bottom of the ventilator through the bottom of the machine frame B Re install the ventilator power tubing into the Gas Selector Switch receptacle by pushing the tubing firmly into the receptacle then pulling the tubing backwards firmly D k Push the ventilator bellows system back into its normal position D Re install the two ventilator base knob screws from the backside of the machine frame 3 Mandatory Replacement of Gas Manifold High Pressure Color Coded Tubing Yellow Tubing Positions From Gas Manifold to Every five years or at any necessary time each of the 10 color coded tubes that attaches Air Pressure Gauge to from the Gas Manifold should be replaced using the following procedure B Air Positiononihe Ventilator Gas Selector Toggle Switch NOTE The end of the color coded tubing must be squarely cut angled ends may not seat C properly A Green oxygen and Yellow Air Tubing The connectors from the Gas Manifold to each of II ROUTINE FUNCTIONAL TEST The functional test may be done at annual intervals or at the following components are of the push to insert and lock type any time that internal componenets have been replaced B To dismount push in on the outer ring and pull the tubing out NOTE The Routine and Annual Functional Test may be carried out at any interval This test will indicate any malfunction of the machine or its components a
2. Outlet 3 each Hose Connector Locking Straight 2 each Nylon Screw Plug Ventilator Container Main Body Container Front Plate Container Back Plate Container Top Plate Bellows Body Attachment Part Semi Locking Control Knobs 3 ech Insp Flo Insp Time Expir Time Bellows Assembly Complete Clear Plastic Bellows Cover Latex Free Rubber Bellows Knob Screws to hold Bellows Assembly to Ventilator Cover Plate 4 each w nuts Bag Vent Switch Pressure Relief Valve Scavenger Outlet B V Switch Support Arm w Connectors for Reservoir Bag CO2 Canister Tube Pressure Gauge Tubing Connector w one way valve CO2 Absorber Support Arm 1 8 x 1 2 Holding Knob and Screw Scavenger Waste Gas Scavenger Tubing 3 ea and TEE Adapter Reservoir Bag 1 Liter PAGE NO 20 20 20 20 18 19 8 Ops 8 Ops 8 Ops 11 Ops 11 Ops 11 Ops PART NO 17 0 18 0 18 1 182 18 3 184 18 5 19 0 19 1 182 18 5 23 0 24 0 24 1 242 24 3 25 0 25 3 26 0 DESCRIPTION Bulb Squeeze Hi Lo Pressure Check w CGO Adapter and Tubing Alarm Low Pressure Mains O2 Container 30 psi Electric Switch Switch Cover Alarm Beeper Alarm Plunger 9 Volt Battery Alarm 2 Concentrator Low Pressure Container 1 psi Low Pressure Electric Switch Switch Cover 9 Volt Battery Vaporizer Retainer Aluminum Retainer Holding Knob Screws 2 ea Vaporizer Agent Specific E
3. A 108 6 1 16 Allen Wrench A 108 7 Squeeze Bulb with Common Gas Outlet Adapter A 108 8 Tubing Color Coded A 108 9 Scavenger Tubing with wye and Bellows Adaptor A 108 10 Service Manual CD A 108 11 Krytox Lubricant Tube 1 2 oz A 108 12 Yoke type Regulator Washers 12 OTHER COMPONENT SERVICE 1 Vaporizer Refer to provided Penlon Ltd Vaporizer Manual 2 Air Compressors Refer to provided Vortran Inc Air Compressor Manual APPENDIX INDIVIDUAL PAGES Subject SPECIFICATIONS FOR MAGELLAN 2200 MODEL 2 PROBLEM SOLVING COMPLETE PARTS LISTING OF ENTIRE ANESTHESIA MACHINE GAS FLOW DIAGRAM Page Number 13 15 16 21 SPECIFICATIONS MACHINE DIMENSIONS Height Depth Width Weight Machine Materials CASE DIMENSIONS Height Depth Width Weight Case Materials Operating Temperature Range Relative Humidity in use Relative Humidity storage Required Gas Supply Sources e O2 Main and Cylinders Air and or Air Compressor Oxygen Concentrator Flowmeter Fresh Gas Flow 13 53 4 cm 23 inches 53 4 cm 23 inches 43 2 cm 17 inches 20 4 kg 45 bs Aluminum brass and plastic 66 0 cm 23 inches 60 96 cm 24 inches 53 34 cm 17 inches 121 2 kg 55 bs empty 200 5 kg 100 Ibs machine with accessories Polyethylene dyed green 1 6 C to 43 3 35 F to 160 F 30 to 9096 non condensing 40 to 9096 non condensing 38 to 70 psi 50 psi is optimal 38 to 70 psi 50 psi is optim
4. G Vent Switch Reservoir Bag 24 GAS FLOW DIAGRAM IV Air O From Flow Meters to Common Gas Outlet BAG Vent Switch on Vent Position Flush Button Vaporizer Gas Manifold Vent Power Supply Toggle Switch Check n From Patient mid is 25 GAS FLOW DIAGRAM V Gas Flow From O Concentrator to O Flow Meter High Low O Pressure Toggle Switch Inlet Nipple lt Flow Meter Back of L Frame ud Low O Concentrator Flow Meter Outlet Alarm Concentrator
5. Magellan 2200 Model 2 Anesthesia Machine Service Manual gt Oceanic Medical Products Inc www oceanicmedical com 8005 Shannon Industrial Park Lane Atchison Kansas 66002 Tel 913 874 2000 Fax 913 874 2005 Email oceanicmedical lvnworth com SUBJECT Factory Address and Communications Purpose General Poicy Factory Testing Equipment and Tools Needed for Service Routine Testing Annual Maintenance Parts Needed and Testing Exploded View of Ventilator Control Knobs Testing After Storage Re Supply Component Kits and Installation Intervals Maintenance Components Kits and Installation Intervals APPENDIX Specifications of Magellan 2200 Model 2 Problem Solving Complete List of Part Numbers and Nomenclature Exploded View Diagrams Gas Flow Diagrams PAGE 11 12 12 12 13 13 15 15 19 21 FACTORY ADDRESS AND COMMUNICATION INFORMATION Address 8005 Shannon Industrial Park Lane Atchison KS 66005 Phone 913 874 2000 Fax 913 874 2005 E mail oceanicmedical lvnworth com Website www oceanicmedical com Office Hours Monday through Friday 08 00 to 16 00 CDT PURPOSE OF THIS DOCUMENT This Service Manual provides the reader with important information concerning the general service maintenance and general repair of the Magellan 2200 Model 2 Anesthesia Machine This manual will allow for the service of the Magel
6. al 3 to 10 psi 1 to above 20 each flowmeter Oxygen Flush Valve CO2 Absorber System CO2 Absorber Canister Capacity Directional Valves CO2 Absorber Holding Bracket Bag Ventilator Switch PRV and Scavenger Outlet Port Bellows Bellows Pressure Relief Common Gas Outlet Tubing Circuit Gas Pressure Hoses Gas Inlet Manifold Gas Inlet Manifold Filters Gas Inlet Pressure Regulators Main supply cylinder Safety back up cylinder Oxygen and Air Supply Gauges Oxygen Supply Alarms Main and safety back up 02 Concentrator Alarm power source Alarm on off Air and O2 Flowmeters Oxygen Concentrator Air Compressor Recessed self closing push button color coded and labeled provides 45 55 lpm constant flow while push button is depressed may be used sparingly with O2 concentrator as O2 power source King Systems KAB 9 re fillable or KAB 1 pre filled disposable 400 grams soda lime Built in the CO2 Absorber Plastic secured with knob to main frame of machine Hand operated selector switch and rotating knob for PRV and scavenging outlet Latex free upward inflating range from 0 to 1 6 L Pre set at 60 cmH20 Quick connect size indexed King Systems F 360 61 or any standard anesthesia circle circuit DISS and thread indexed female connectors at both ends DISS and thread indexed male connectors with one way valves Located behind Manifold Air and O2 inlet male connectors DISS thread inde
7. c bellows canister cover counter clockwise until the four plastic retainer parts are clear of the black plastic retainers located on the top of the base of the bellows assembly and gently remove the clear plastic bellows canister cover using an upwards motion 2 Remove the rubber bellows from the Bellows Base am Exhalation Diaphragm Valve Assembly Bellows Base 3 Each time the bellows assemblies are opened for inspection and cleaning all visible components must be carefully inspected and damaged parts must be replaced Bellows The bellows is an elastomer material and deteriorates over time Inspect the bellows every six months if the system is in use Inspect on the appropriate schedule annual to a period of 3 years if the system is in long term storage Replace the bellows if there is any sign of deterioration Exhalation Valve Diaphragm The exhalation valve diaphragm valve is under the bellow and can be removed by loosening the three thumbscrews The valve seat is now visible WARNING During inspection and cleaning to remove contamination of the valve seat and diaphragm use a soft cloth or brush Abrasive materials may damage the seat and should not be used To preclude damage caused by the patient breathing into the valve you should use a bacterial filter in the expiratory breathing circuit Clean the seat A and the metal disk B attached to the base of the diaphragm valve thoroughl
8. ed time cycled volume constant volume variable Minute Volume Frequency m 1 bimim 30 25 25 20 10 13 Tidal Volume t haz a E k Ti k a 5 T crt E Pre set to maximum of 60 cmH20O located in main vent box Preset to 60 cmH20 40 to 70 psi 50 psi optimal Use toggle switch to select gas source Positive and negative relief valves 1 L reservoir bag vacuum control knob 30 cmH20 0 ml Min 80 cmH20 100 ml min 20 cmH20 1 1 ml cmH20 40 cmH20 1 3 2 1 0 L sec gas flow 4 11 cmH20 0 5 L sec gas flow 1 80 cmH20 3 0 L min gas flow 0 12 cmH20 30 L min gas flow 1 03 cmH20 PROBLEM SOLVING GUIDE PRE USE CHECKLIST SOLUTION See Operators Manual Check main pipeline O2 gas supply and the emergency O2 backup cylinder by observing the gas pressure gauges Resolve by ensuring adequate pipeline and cylinder O2 supplies are adequate and are attached to anesthesia machine Replace soda lime in canister Check Bag Vent Switch for vent position Check PRV Control Knob for closed position Check entire tubing circuit for leaks Check O2 supply and O2 Pressure Gauge to ensure O2 is available Check O2 sensor to ensure proper seating in tee adapter Check that opposing Air or O2 flowmeter is OFF Check Air or O2 gas power supply for 40 70 PSI on pres
9. es are now correct NOTE The results of calibration can vary depending on the test lung utilized Reservoir bag and rubber test lungs may not inflate fully or over inflate causing erroneous results TESTING AFTER STORAGE The following procedures should be used after the Magellan 2200 Model 1 Anesthesia Machine has been stored for 6 MONTHS OR LONGER 1 Set the machine up and prepare for Bench Testing by placing all components needed on the machine as though preparing for actual use 2 Usethe Pre Use Checklist and verify that all components are ready for use 3 Inspect the machine for gas leaks such as the high pressure gas hoses gas connectors and pressure reducing regulators 4 Allow the machine to function for 30 minutes and re inspect the device for any discrepancies 5 If any problems are observed perform the Annual Maintenance and test procedure to re qualify the machine and return to storage or service IV SERVICE FOR MAGELLAN 2200 ANESTHESIA MACHINES Part Number OEM PN with NSN Description Page No or CAGEC Drawing No A 108 Annual Services Component Kit consists of NOTE The A 108 Kit is comprised of the following components Individual components may be utilized at any interval without installing the complete Kit A 108 1 Bellows Latex free A 108 2 Tubing Kit with valves A 108 3 Vaporizer End Caps A 108 4 Manifold Filters and Spacers A 108 5 Tweezers for removing Filters and Spacers
10. eter Pack Air amp Oxygen Flush Button HVLC Pressure Toggle Swi Air Pressure Gauge Oxygen Pressure Gauge Oxygen Auxiliary Flow Selecte L Frame Vaporizer L Frame Securing Screws Bag Vent Switch Absorber Arm Bag Vent Switch Assembly an Dowel Pin Guide for B V Switc Absorber Support Arm Flow Meter Bar Vaporizer Retainer Vaporizer Retainer Screws 19 MAGELLAN 2200 ANESTHESIA MACHINE BACK MAN BODY FRAME Ci 0 MAN LOW PRESSURE OXYGEN 3 245 VAPORIZER HOLCING SPACER 4 SECUF aj 990 GAS MAIN AIR 4 51 wo O1CONCENTRATCR LOW PRESSUR 20 21 GAS FLOW DIAGRAMS GAS FLOW DIAGRAM I Flow from Gas Manifold to L Frame Low Pressure O Alarm Auxiliary Selector 50 PSI O In Pressure Gage Flush Button High O Flow Meter Toggle Switch O Air Ventilator Gas Power Selector Gas Manifold L Frame GAS FLOW DIAGRAM II Air Flow from Gas Manifold to L Frame Pressure Gage 50 PSI Air In _ Air Flow meter ee O Air Ventilator Gas Power Selector Gas Manifold L Frame GAS FLOW DIAGRAM III Air O From Flow Meters to Common Gas Outlet CGO BAG Vent Switch on BAG Position Flush Button A Vent Power Supply i oggle Switch Gas Manifold Vaporizer Flow Meters CO Absorber To Patient m From Patient CO eg gt BA
11. g 90 degree 1 8 Hose Connector Locking straight 2 each Air Pressure Gauge Yellow Dial Female Female Coupling 1 8 NPT PAGE NO 19 19 20 20 20 PART NO 4 3 8 0 7 1 4 3 9 0 7 1 4 3 10 0 10 1 10 2 10 3 2 1 7 1 7 2 11 0 11 1 11 2 11 3 11 4 11 5 11 3 2 2 DESCRIPTION 1 8 Hose Connector Locking 90 degree Oxygen Pressure Gauge Green Dial Female Female Coupling 1 8 NPT 1 8 Hose Connector Locking 90 degree Auxiliary Oxygen Flow Selector Female Female Coupling 1 8 NPT 1 8 Hose Connector Locking 90 degree Main Body Frame Aluminum Support Rods Brass 2 each 1 20 Rod Screws Brass 4 each Dowel Pin for B V Switch Mounting Brass with 20 B rass Screw 1 8 Hose Barb O2 Concentrator Inlet Female Female Coupling 1 8 NPT 1 8 Hose Connector Locking Y Outlet Gas Manifold Body DISS Adapter Air with one way valve Spacer length specific retains bronze filter Bronze Filter cone shaped DISS Adapter Oxygen with one way valve 2 each Spacer length specific retains bronze filter 2 each Bronze Filter cone shaped 2 each Hose Connector Locking Straight 2 each PAGE NO 19 20 7 2 2 2 11 6 12 0 12 1 12 2 12 2 12 30 13 0 13 1 13 2 13 3 14 0 14 1 14 2 15 0 15 1 16 0 16 1 16 2 DESCRIPTION 1 8 Hose Connector Locking
12. h devices may be used for bench testing I Ee ccIxe a S OO zi f m pP 5 Hand Operated Spirometer May be used for gas volume calibration if other mechanical test lungs are not available 6 Rubber Test Lung or Any One Liter Rubber Test Lung or One Liter Gas Reservoir Bag Used to simulate patient delivered gas volumes if mechanical Test Lung is not available 7 Carbon Dioxide Absorber King Systems KAB 9 unit either full of absorbent or empty 8 Patient Tubing Circuit Any approved Patient Tubing Circuit for Anesthesia may be utilized 9 Monitoring Device for Anesthetic Agent Either a specific vaporized agent testing device such as a Bickford Anesthesia Agent Tester Model 3 or an anesthesia monitor capable of measuring vaporized agent may be used 10 Tools The Magellan 2200 Model 2 Anesthesia Machine was designed to be serviced with the following common tools 1 16 Allen s Wrench used with O2 flowmeter control knob and the three ventilator control knobs B Screwdriver regular slotted head Small adjustable crescent wrench or 1 1 2 open head wrench 2 7 16 open head wrench 3 regular pliers D 1 sealing tape roll E Tube of Krytox or other oxygen compliant lubricant F Water based leak testing solution ANNUAL MAINTENANCE PARTS NEEDED AND TESTING Recommended Inspection Replacement of Certain Components A An
13. lan 2200 Model 2 Anesthesia Machine and its accessories This machine may be fully serviced outside of the factory and the reader may contact the factory at any time for assistance GENERAL POLICY STATEMENT The scope of this manual covers general maintenance calibration of controls disinfecting procedures and replacement of certain components at the intervals indicated within this document FACTORY TESTING Each Magellan 2200 Model 2 Anesthesia Machine is tested utilizing certain testing equipment which may or may not be available in the field or in a general biomedical maintenance facility This manual will explain how the Operator and Biomedical Technician may completely and safely maintain this equipment USE AND OPERATION The use and operation of the Magellan 2200 Model 2 anesthesia machine should be performed by a physician anesthesiologist or certified registered nurse anesthetist CRNA trained properly with this device EQUIPMENT NEEDED FOR BENCH TESTING 1 Operators Manual 2 Pre Use Check List Contained in this manual the Operators Manual and Separate Laminated Card 3 Regulated Pressurized Gas Sources Wall outlets for oxygen and air gas cylinders for oxygen and air each with appropriate pressure reducing regulators portable medical grade air compressor and oxygen concentrator 4 Mechanical or Electrical Mechanical Test Lung The factory utilizes the Bio Med VT 2A Test Lung however other brands of suc
14. nd Cap Inlet Port for Tubing from Flowmeters Size Specific End Cap Outlet Port for Tubing from Vaporizer to CGO Cagemount Assembly Backbar and Holding Knobs Manual Operator and Service CD25 1 Penlon Vaporizer Manual Set Up Card Pictorial Pre Use Checklist Card Oxygen Pressure Reducing Regulator Yoke Mount 17 11 Ops 20 20 20 20 20 16 Ops 16 Ops 16 Ops 6 Ops PART NO 26 3 26 4 26 5 A 101 A 102 A 103 A 104 A 105 A 106 A 107 A 108 A 108 A 109 A 110 A 111 DESCRIPTION Oxygen High Pressure Hose Green 24 length Oxygen High Pressure Hose Green 15 length Air High Pressure Hose Yellow 15 length Air Compressor Vortran HP 70 Water Trap for Gas Manifold Air Inlet Connection Carrying Case Air Compressor Carrying Case Magellan 2200 Anesthesia Machine Oxygen Cylinder Mount for Carrying Case Aluminum Trolley for Magellan 2200 Anesthesia Machine Oxygen Analyzer Monitor with Sensor Oxygen Analyzer Sensor Replacement Oxygen Analyzer TEE Adapter PEEP Valve Magnetic Mapleson Jackson Reese CGO Adapter Service Kit Annual Maintenance PAGE NO 6 Ops 6 Ops 18 Ops 7 Ops 5 Ops 5 Ops 11 Ops 11 Ops 11 Ops 20 Ops 18 MAGELLAN 2200 ANESTHESIA MACHINE FRONT DESCRIPTION Main Frame Roos Support Steel Ventilator Bellows Common Gas Outlet Vent Power Toggle Switch Flow M
15. nd the Operator or Biomedical Technician will be able to rectify any component that is not functioning properly C To re insert push the end of the green tubing into the receptacle and then use a pulling action to lock the tube in place Test Procedure 1 Assemble and install all components needed for routine clinical use of the machine as outlined in the Operators Manual or the Laminated Set Up Card 2 Perform the Pre Use Checklist and Laminated Set Up Card with photos in sequence This Checklist is contained in the Operators Manual the separate Pre Use Checklist laminated card and Set Up Card are provided with each machine Green Tubing Positions From Gas Manifold to A Auxiliary O2 Flow Selector B O2 Pressure Gauge C O2 Flush Valve D O2 Flowmeter NOTE If the machine passes the Pre Use Checklist with zero problems proceed to the E O2 Position on Ventilator Gas Selector Toggle Switch Calibration Guide in the next section F O2 Concentrator High Pressure O2 Selector Toggle Switch CAUTION If the machine does not pass the Pre Use Checklist correct the problem s encountered then proceed with the Calibration Guide The Problem Solving Guide on page 13 G O2 Low Pressure Alarm of this Manual will assist the Operator in resolving possible common problem situations H O2 Very Low Pressure O2 Concentrator Pressure Alarm Calibration Guide CAUTION The factory utilizes pressurized Air and Oxygen set at 50 PSI fo
16. nually The following procedures should be used to ensure proper operation of the machine at least once per year or at any other interval that the Biomedical Technician deems necessary B Parts Needed The following parts will be needed for Annual Service Inlet Filters 2 each P N 11 3 Bellows Latex free P N 13 1 1 Changing the Gas Manifold Inlet Filters Air DISS Inlet Connector A Using a crescent wrench unscrew the DISS Air inlet connector and remove B Remove the filter spacer aluminum ring then remove the cone shaped filter C Inspect the filter for debris If debris is found remove it D Install the filter back into the Gas Manifold with original or new filter with the small end first m Install the filter spacer aluminum ring F Pull open the one way valve located on the end of the threaded portion of the DISS adapter and apply some Krytox or equivalent lubricant Re tape the DISS Air connector threaded portion with sealant tape H Re install the DISS Air connector screwing it closed snugly with the crescent wrench Oxygen DISS Inlet Connectors the first one is located next to the Air Connector the second one is located on the left hand side of the Gas Manifold and the procedure is identical for both Oxygen DISS Inlet connectors A Using a crescent wrench unscrew the DISS Oxygen B Remove the filter spacer aluminum ring then remove the cone shaped filter C In
17. r all testing It is recommended that this pressure be used when possible If higher or lower inlet pressures are routinely utilized the ventilator may be re calibrated at pressure inputs from 40 to 70 PSI Calibration Materials Required Source gas of air or oxygen Appropriate hose set to connect to gas source Pre Use Checklist Operating Ventilator e Test Lung Patient Tubing Circuit Procedure Attach the patient circuit to the equipment and to the test lung Set Minute Volume control knob to 10 LPM Set Frequency conrol knob to 13 BPM Ensure the bellows is fully inflated Pressurize the ventilator and allow the system to cycle several times Verify the following values on the test lung Minute Volume 10 LPM BPM 13 per minute Tidal Volume 770 NOTE Values should be 1096 of settting smiths 10 2 Ventilator Gas Power Toggle Switch Select either Air or Oxygen as the power source NOTE Ensure that the Oxygen system toggle switch is in the high pressure position 11 3 Re Calibration of Individual Control Knobs Tools needed Test Lung A Remove Control Knob by grasping the knob firmaly and pulling it straight off the shaft B With the ventilator still cycling turn the valve shaft until the proper value is shown on the test lung C Replace the Control Knob by facing the pointer to the desired value and pushing it straight on to its full extent D Verify the valu
18. spect the filter for debris If debris is found remove it D Install the filter back into the Gas Manifold with original or new filter Install the filter spacer aluminum ring F Pull open the one way valve located on the end of the threaded portion of the DISS adapter and apply some Krytox or equivalent lubricant G Re tape the DISS Oxygen connector threaded portion with sealant tape H Re install the DISS Oxygen Connector s screwing it closed snugly with the crescent wrench 2 Replacement of Rubber Bellows and Inspection of Exhalation Diaphragm Valve A Removal of Ventilator and Bellows Systems from Main Frame 1 Turn anesthesia machine on the Vaporizer side of the machine b 2 Using a flat head screwdriver remove the ventilator retaining screw 3 Lay the machine down in its normal operating position 4 Remove the two retaining knob screws on the backside of the ventilator nis device i by on fhe physician 5 Disconnect the ventilator gas power tubing from the Ventilator GasPower Toggle Selector Switch by pushing in on the tubing and simultaneously pushing inwards on the retainer ring located on the tubing inlet receptacle 6 Remove the ventilator bellows system by pulling the assembly forward from the main frame B Removal of Rubber Bellows from Bellows Base Inspection of Exhalation Valve Diaphragm Valve and Installing New Bellows 1 Turn the clear plasti
19. sure gauges Check Vent Power Selector Switch in Air or O2 position Check expiratory time control to ensure settings If O2 is selected for power check O2 Power Selector Toggle Switch and ensure it is in the High Pressure Position Check top white plastic ring for proper lubrication Inspect relief valve at base of bellows for debris Insure that water trap is instaled on air inlet located on the back of the gas manifold e Remove PEEP valve from system and check for kinks at ET tube patient tubing circuit PART LIST FOR MAGELLAN 2200 MODEL 1 Listed by Assembly Section in Bold Type NOTE Some numbers are repeated as these are used in additional assembly PART NO 1 0 2 0 2 1 2 2 2 3 24 3 0 3 1 4 0 4 1 2 1 4 3 5 0 4 3 6 0 4 3 2 2 7 0 7 1 NSN DESCRIPTION L Frame Aluminum with 8 32 Brass Mounting Screws 4 each Flowmeter Pair Air and Oxygen with Holding Nuts 1 8 Hose Barb Hose Connector Locking Straight Knob Green Fluted for Oxygen Flowmeter Knob Black for Air Flowmeter Common Gas Outlet Body and Holding Nut Common Gas Outlet Coupling Insert Oxygen Flush Button Assembly Gas Actuator Assembly 1 8 Hose Barb 1 8 Hose Connector Locking 90 degree Ventilator On Off Toggle Switch with Holding Nut 1 8 Hose Connector Locking 90 degree 3 each High Pressure 02 02 Concentrator Toggle Switch with Holding Nut 1 8 Hose Connector Lockin
20. xed for O2 Pin indexed yoke mounted for D and E cylinders for O2 0 3000 psi range color coded and clearly labeled Pneumatically actuated when O2 supply falls below 30psi Pneumatically actuated when O2 supply falls below 1 psi 9 volt battery located in body of alarm box Labeled toggle switch located on body of alarm box Calibrated and scaled 0 10 lpm color coded O2 flowmeter has a fluted control knob for easy identification by touch alone To power O2 flowmeter only May be used to power ventilator and air flowmeter Oxygen Analyzer Monitor Oxygen Analyzer Power Source Auxiliary 2 Flow Selector Vaporizer Airway Pressure Gauge Pressure Gauge Tubing Mechanical Ventilator Ventilator Pressure Relief Pressure Relief Valve Ventilator Gas Power Requirements Waste Gas Scavenger Total Machine Gas Leakage Internal System Compliance Internal System Resistance APL Valve Pressure Drop 14 OM 25 ME or equivalent Galvanic cell sensor life expectancy 2 years under normal conditions 2 each AA batteries life expectancy approx 3000 use hours Scaled 0 10 LPM in set increments used for pre post anesthesia Penlon SigmaDelta Series bolt cage mounted temperature compensated very low maintenance Dual scaled in cmH20 and mmHg located on front panel of ventilator May be attached to bag vent switch arm or to a point within the patient breathing circuit operators choice Pneumatically power
21. y and remove any contamination from the surfaces of both components NOTE Any anesthesia breathing circuit filter may be used After cleaning check the small O Ring C located in the bellows base under the diaphragm valve is in place The system will not function if the O Ring is not in place Refit the diaphragm valve assembly to the bellows base and reassemble the bellows assembly 4 Install new bellows to bellows base by placing the bottom bellows crease over the lip of the bellows retainer See photos to the right NOTE Once the bellows is installed turn the vent bellows system upside down and observe that the bellows fully expand and the folds are evenly opened Re adjust the bellows fold where it is attached to the base and re observe the expansion 5 Using Krytox or equivalent lubricant place a small amount of the lubricant on the hard plastic top ring of the bellows 6 Using Krytox or equivalent lubricant place a small amount of the lubricant on the bottom ring of the clear plastic bellows canister cover a ar 7 Re install the clear plastic bellows canister cover by pushing the clear canister down firmly over the bellows base then turning the canister cover clockwise until the retainers are completely covered by the retainer clips 8 Re install the ventilator bellows system E Turn the entire machine onto the vaporizer side of the system A Place the ventilator bellows system partially in place