Home
        Accutorr 7 Operators Manual
         Contents
1.                          Alarm Clearable  Alarm Cause Solution   Message  Yes No  Level   Temp Over   No High The patient s   Reduce the patient s   High Limit temperature is too   temperature  or  high  or an error   replace the module   occurs    Temp Over   No High The patient s   Raise the patient s   Low Limit temperature is too   temperature  or  low  or an error   replace the module   occurs    Temp No   Yes Med The probe is   Reconnect the probe    Probe disconnected    Temp Probe   Yes Med The temperature   1  Check if the probe   Misplaced probe is not well   well is installed   placed  or not   2  Properly re insert  inserted into probe   the probe into probe  well  well    Temp Yes Med The measuring time   Return the sensor to   Measuring is over 5 minutes in   the probe well  and   Timeout Monitor mode  take a temperature   again    D 2 4 Recorder Alarm Messages   Alarm Clearable  Alarm Cause Solution   Message  Yes No  Level   Recorder Init   yes Low An error occurred to   Restart the monitor  If   Error the module  or there   the problem still   Recorder Yes Low is a problem with   exists  contact   Selftest Error the communication   Mindray or your  between the   service personnel    Recorder Yes Low module and the   Unavailable monitor    Recorder Yes Low   Comm Error   Recorder Vit   No Low There is a problem   If the message is   High with the system   prompted for several  power supply    times  contact Mindray   a Vit   No Low  Restart the monitor    or y
2.                Measurement accuracy verification  In adult and pediatric modes  the blood pressure  measurements measured with this device are in compliance with the American National  Standard for Electronic or Automated Sphymomanometers  ANSI AAMI SP10  in terms of  mean error and standard deviation by comparing with intra arterial or auscultatory  measurements  depending on the configuration  in a typical patient population  For  auscultatory reference  the 5  Korotkoff sound was used to determine the diastolic    pressure     In neonatal mode  the blood pressure measurements measured with this device are in  compliance with the American National Standard for Electronic or Automated  Sphymomanometers  ANSI AAMI SP10  in terms of mean error and standard deviation by  comparing with intra arterial measurements  depending on the configuration  in a typical    patient population        A 8    Accutorr 7 Operator s Manual       A 6 4 Temp       Standards    Meets standards of ASTM E1112  ASTM E1104  ISO 80601 2 56       Technique    Thermal resistance  use thermistor to measure temperature         Measurement range    Monitor mode 25 to 44  C  77 to 111 2   F   Predictive mode  35 to 43  C  95 to 109 4   F        Accuracy  Monitor mode     25  C to 32  C  not include 32  C     0 2  C  32  C to 44   C  include 32  C    0 1  C   or   77  F to 89 6  F  not include 89 6  F     0 4  F  89 6  F to 111 2  F  include 89 6  F     0 2  F       Resolution     0 1  C       Response Time   
3.           Alarm delay time from  monitor to remote  equipment       The alarm delay time form the monitor to remote equipment is   lt 2 seconds  measured at the monitor s signal output  connector        A 5 5 Data Storage          Trends       5000 groups          Accutorr 7 Operator s Manual                   A 5 6 Wireless Network  Optional        Standards    IEEE 802 11b g n  support Wi Fi       Code error of wireless layer     lt  100 ppm       Priority    All communication data types shall have the same priority        Transmission delay    Total delay of data transmission from the monitor to the CMS    lt 2s    Delay of configuration settings transmission from CMS to  Accutorr 7   lt  2s  measured from configuring on CMS to the  settings take effect on the monitor         Roaming    Network switchover is automatically executed when the  Accutorr 7 moves from the coverage area of one AP to that of  another AP        System capacity    Number of the Accutorr 7 supported by a single AP is  lt  16   Each Accutorr 7 can communicate with the CMS        Dynamic networking  stability    The Accutorr 7 meets its functional requirements when it  moves at a rate of no more than 3 75 m s within a 15 m  non blocking linear distance           Network interruption alarm       When a network interruption occurs  the equipment shows a  disconnection icon on the screen and presents the related  alarms     When the network interruption is resolved  wireless  connection recovers automatic
4.       Alarm Alarm Cause Solution  Message Priority  XX Too High Med  XX value exceeds the upper alarm Check the patient s  limit  condition and check if    the patient category     XX Too Low Med a sere K lower than the lower and alarm limit  aN  settings are correct   SpO2 Desat High The SpO  value has fallen below Check the patient s  the desaturation alarm limit  condition and check if  the alarm limit  settings are correct   No Pulse High The pulse signal was so weak that Check the patient s          the monitor cannot perform pulse  analysis        condition  SpO2 sensor  and measurement site        D 2 Technical Alarm Messages  D 2 1 NIBP Alarm Messages                                              Alarm Clearable  Alarm Cause Solution  Message  Yes No  Level  NIBP Sys Limit   No Low The parameter alarm   Contact Mindray or  Err limit is inadvertently   your service  NIBP Dia Limit   No Low changed  personnel   Err  NIBP Mean No Low  Limit Err  NIBP Sys Over   Yes Low The measured   Check the patient s  Upper Limit pressure is greater   condition and keep    than the specified   the patient relaxed   Peer A res Low NIBP measurement   and still  If the error  Pee upper limit  remains  contact  NIBP Mean Yes Low Mindray or your  Over Upper service personnel   Limit   Accutorr 7 Operator s Manual D 1                                                             Alarm Clearable  Alarm Cause Solution  Message  Yes No  Level  NIBP Sys Over   Yes Low The measured Restart the moni
5.       Nurse Call Cable    009 003116 00       Serial Port Cable    009 003436 00       Recorder Paper    0683 00 0505 02       U S  Power Cord    0012 25 0001       USB disk    023 000217 00       Accessories Kit    Component    Part No           Barcode Scanner Kit       1D Bar Code Scanners       023 000254 00          Accutorr 7 Operator s Manual    15 3             Material    Part No         115 008393 00     USB Cable of ASYMBOL Scanner    009 001397 00       Accutorr 7 Welch Allyn  Temp Support Kit     115 025041 00     Accutorr 7 Welch Allyn Temp Support  Assembly    115 022908 00       Accutorr 7 Welch Allyn Temp Support  Installation Guide    046 006016 00          Li ion Battery Kit   115 018012 00        Li ion Battery  LI23S002A 11 1V 4500mAh 022 000008 00             15 4    Accutorr 7 Operator s Manual       A Product Specifications    A 1 Classifications    The equipment is classified  according to IEC60601 1        Type of protection against  electrical shock    CLASS   EQUIPMENT  equipment energized from an  external and internal electrical power source        Degree of protection against  electrical shock    DEFIBRILLATION PROOF TYPE CF AAPPLIED PART for    SpO2  NIBP  and TEMP       Mode of operation    Continuous       Degree of protection against  harmful ingress of water    IPX1  protection against vertically falling water drops        degree of safety of application in  the presence of a FLAMMABLE  ANAESTHETIC MIXTURE WITH AIR  or WITH OXYGEN OR NITROU
6.       Type Model Patient Category Part No   LNCS Pdtx Pediatric 0600 00 0122  LNCS Adtx Adult 0600 00 0121  Disposable LNCS NeoPt Neonate   lt 1 kg  0600 00 0156  LNCS Neo Adult and Pediatric   gt 40 Kg   0600 00 0157  neonate  lt 3 Kg   LNCS Inf Pediatric and Neonate  3 to 20 Kg  0600 00 0158  LNCS DCI Adult  Finger  0600 00 0126  Reusable  LNCS DCIP Pediatric  Finger  0600 00 0127                Wavelength emited by the sensors is between 600 nm and 1000 nm        Accutorr 7 Operator s Manual 15 1       The maximum photic output consumption of the sensor is less than 18 mW     The information about the wavelength range and maximum photic output consumption  can be especially useful to clinicians  for example  when photodynamic therapy is    performed      15 2 NIBP Accessories                                                                                                                                  Tubing  Type Patient Category Part No   Adult  Small Adult  Child 6200 30 09688  Reusable  Neonate 6200 30 11560  Reusable Cuff  Patent Measurement Limb Circumference  cm  Part No   Category Site  Child 10 to 19 0683 15 0001 01  Small Adult 18 to 26 0683 15 0002 01  Adult r 24 to 35 0683 15 0003 01  rm  Adult Long 27 5 to 36 5 0683 15 0006 01  Large Adult 33 to 47 0683 15 0004 01  Large Adult Long 35 5 to 46 0683 15 0007 01  Adult Thigh 44 to 66 0683 15 0005 01  Disposable Cuff  paver Measurement Limb Circumference  cm  Part No   Category Site  3 1 to 5 7 001B 30 70692  4 3 
7.      Company Contact  Manufacturer  Shenzhen Mindray Bio Medical Electronics Co   Ltd     d  ress Mindray Building  Keji 12th Road South  Hi tech industrial  park  Nanshan  Shenzhen 518057  P R China  Website www mindray com  E mail Address  service mindray com cn  Tel   86 755 81888998  Fax   86 755 26582680  Distributor  Mindray DS USA  Inc   Address  800 MacArthur Boulevard  Mahwah  New Jersey 07430 USA  Tel  1 800 288 2121  1 201 995 8000  Website  www mindray com    EC Representative     Address   Tel     Fax     Shanghai International Holding Corp  GmbH  Europe   EiffestraBe 80  20537 Hamburg  Germany  0049 40 2513175   0049 40 255726       Accutorr 7 Operator s Manual    Preface    Manual Purpose  This manual contains the instructions necessary to operate the product safely and in  accordance with its function and intended use  Observance of this manual is a  prerequisite for proper product performance and correct operation and ensures patient  and operator safety     This manual is based on the maximum configuration and therefore some contents may  not apply to your product  If you have any question  please contact Mindray     This manual is an integral part of the product  It should always be kept close to the  equipment so that it can be conveniently obtained when needed     Intended Audience  This manual is intended for clinical professionals who are expected to have corresponding  working knowledge of medical procedures  practices and terminology as required for t
8.     A    When the alarm sound is switched off  the equipment will give no audible alarm    tones even if a new alarm occurs  Therefore the user should be very careful  about whether to switch off the alarm sound or not     Do not rely exclusively on the audible alarm system for patient monitoring     Adjustment of alarm volume to a low level may result in a hazard to the patient   Always keep the patient under close surveillance           6 5 4 Adjusting Alarm Limits Automatically  The monitor can automatically adjust alarm limits according to the measured vital signs   using the auto limits function  When auto limits are selected  the monitor calculates safe  auto limits based on the latest measured values     To get accurate auto alarm limits  you need to collect a set of measured vital signs as a  baseline  Then  in the main menu  select  Main     Alarm Setup  gt  gt      Auto Set    Select   Ok  in the pop up window  The monitor will create new alarm limits based on the  measured values     Before applying these automatically created alarm limits  confirm if they are appropriate  for your patient  If not  you can adjust them manually     The monitor calculates the auto limits based on the following rules              Parameter  Low alarm Limit High alarm Limit Auto set alarm  limit  Adult Pediatric  Neonate Adult Pediatric  Neonate  Range  SpO2 Same as the Sameasthe  Sameas the Same asthe  Same as the  default alarm default alarm  default alarm default alarm measurement  
9.     At least 20 minutes  after a low battery alarm first occurs        A 4 Physical Specifications             Size 178mmx150mmx260mm  70 08 inch x 59 05 inch x 102 36  inch   Weight  lt 2 5 kg  5 51 Ib  with SpO2 module  NIBP module  recorder       module and a battery        A 5 Hardware Specifications                      Display   Screen type Color TFT LCD  Screen Size  diagonal  8 4    Resolution 800x600 pixels  Recorder   Method Thermal dot array  Paper speed 25 mm s          Number of waveform  channels       1          Accutorr 7 Operator s Manual          A 5 1 LEDs                   Alarm lamp 1  two color coded  yellow and red   Power on LED 1  green   AC power LED 1  green   Battery LED 1  green           A 5 2 Audio Indicator          Speaker       Give alarm tones  45 to 85 dB   key tones  pulse tone  power on  self check tone  support PITCH TONE and multi level tone  modulation  alarm tones comply with IEC 60601 1 8        A 5 3 Monitor Interface Specifications       Power    1 AC power input connector       Wired network    1 RJ45 connector       USB    2 standard connectors  USB 2 0       Equipotential Grounding  Terminal    1          Multi Functions Connector       1 MINI D RIBBON connector       A 5 4 Outputs       Nurse Call Signal       Amplitude    High level   gt 3V  providing a maximum of 3 mA output current   Low level   lt 0 5 V  receiving a maximum of 5 mA input current       Rising and falling time     lt 1ms       Alarm Output  Network connector 
10.     Do not connect non approved devices to the wireless network      Where the monitor is located  the signal strength of other Wi Fi devices on the  same channel should be no greater than  85 dBm      Where the monitor is located  the signal strength of other Wi Fi devices on  adjacent channels should be no greater than  50 dBm       The recommended distance between the patient monitor and other non Wi Fi    wireless devices  including wireless devices at the frequency of 2 4GHz  e g   cellular mobile communication networks  microwave ovens  interphones   cordless phones and electro surgical units  is no less than 20 cm     12 2 4 ADT Communication Setup    1     Select  Main      Maintenance  gt  gt      User Settings  gt  gt     Enter the required  password    Select  Ok  to access  User Settings  menu        2  Select  Network gt  gt       ADT Communication gt  gt   to access  ADT Communication  Setup  menu   3  Set IP address and port  and switch on off ADT query   4  Select  Ok  to exit the menu   12 4 Accutorr 7 Operator s Manual    12 2 5 EMR Communication Setup    1     Select  Main      Maintenance  gt  gt      User Settings  gt  gt     Enter the required  password   Select  Ok  to access User Settings menu     Select  Network gt  gt       EMR Communication gt  gt   to access  EMR Communication  Setup  menu    Set the IP address and port    Select  Ok  to exit the menu     12 2 6 DIAP Communication Setup    1     Select  Main      Maintenance  gt  gt      User Setti
11.   3 Basic Operation  coscssscscicscnsssacscccctecassedsicssseacssesdecetsesececestensseucscesseedsoessadonsssessceonscesstess  susse 3 1    EATA ILEITO  oNN E EE EEEE E EEE E E EE AEE  3 1 1 Unpacking and Checking  3 1 2 Environmental Requirements   32 General O pea O E N AEE EN TAE E ETENN  3 2 1 Connection to AC Power  3 2 2 Using a Battery  3 2 3 Connecting ACCeSSOrieS       sssssssssseseeeececccceeeeeeeessesessessssssssosseeteerceeeeeseeeeeeeeeesssesesssssssssssesteet   3 3 Turning On Off Power  3 3 1 Turning Power On  3 3 2 T  rning off The Monitors neia RA ANE AANEREN   3 4 Using Key  Knob  Touchscreen  3 4 1 Using Keys  34 2 Using the KHO Disina a a a iat aar  3 4 3 Using the Touchscreen   3 5 Changing General Settings  3 5 1  Setting upa Monito aeann an eE ONE NANAS                          Accutorr 7 Operator s Manual       3 5 2 Changing Language  3 5 3 Configuring the Timeout of Clinician ID  3 5 4 Adjusting Alarm VOlUME        scsssssscssssecsssccsseecssssccsssecsssscesueccsssecesssecesscecssucecessecssaecesaseesaes  3 5 5 Adjusting Key Volume   3 5 6 Adjusting the Screen Brightness  3 5 7 SCUING SCLEON  sccsscicssosssasssatscassstiescasseciescbaceaasistatonsseadianaistazesasnscntsaivteasssenaciannsadagaccantauieanatss  3 5 8 Configuring the Timeout of Measured Value  3 5 9 Configuring Measurement Colors  3 5 10 Setting the Date and Time        essssscssssccssssessssccsssecssusccsssecssseeccssecessseessueccssseecsseecesseeessaes  3 5 11 Configuring Unit  3 5 12 Co
12.   Auto mode repetition  intervals    1  2  2 5  3  5  10  15  20  30  60  90 or 120 min                         STAT mode cycle time 5 min  Max measurement time AFMI REARS i   Neonate  90s  Heart rate range 40 to 240 bpm   Adult Pediatric Neonate   Measurement ranges Systolic  40 to 270 40 to 200 40 to 135   mmHg  Diastolic  10 to 210 10 to 150 10 to 100   Mean  20 to 230 20 to 165 20 to 110                Accuracy    Max mean error   5 mmHg  Max standard deviation  8 mmHg       Static pressure                0 mmHg to 300 mmHg  measurement range  Static pressure     measurement accuracy Sang  Resolution 1mmHg          Accutorr 7 Operator s Manual A 7                      Adult  80 to 280  Initial cuff inflation Pediatric  80 to 210  sate al oA Aaa Neonate 60 to 140  ENAA Adult  160  efault initial cu ein    inflation pressure  mmHg  oo  eonate   Adult  297 3 mmHg  Software overpressure Pediatric  240 3 mmHg  protection i    Neonate  147 3 mmHg  Adult   lt 330 mmHg  SAE OS Pediatric   lt 330 mmHg  Neonate   lt 165 mmHg  Alarm limit Range  mmHg  Step  mmHg        Adult   low limit 5  to 270                      Sys High Pediatric   low limit 5  to 200   Neonate   low limit 5  to 135  Sys Low 40 to  high limit 5    Adult   low limit 5  to 230  Mean High Pediatric   low limit 5  to 165   Neonate   low limit 5  to 110  Mean Low 20 to  high limit 5    Adult   low limit 5  to 210  Dia High Pediatric   low limit 5  to 150   Neonate   low limit 5  to 100  Dia Low 10 to  high limit 5 
13.   Interval 30s  Graphic  Trends Zoom on  C 2 3 Record    Saved at Power Affected by  Name Default Config Failure Defaults  SpO2 wave Selected  Yes No  Manual Inputs Selected                   Accutorr 7 Operator s Manual             C 2 4 Others                                                                                                    Name Default Config ane power as aie by  Brightness 5  Alarm Volume 2  Key Volume 2 Yes Yes  Display Setup All Parameters  Parameter Time Out 15 min  Height cm  Weight kg  Pressure mmHg  Tene T Yes No  Glucose mg dl  1 0 Fluid ml  Date Current date      Time Current time      Date Format yyyy mm dd Yes No  Time Format 24h Yes No  C 3 User Maintenance Items   Name Default Config ated power aie by  Spot Check Off  SPO2 Tone Mode 1  Clinician ID Time Out  10 min Yes No  Language English  Network Type LAN  Clear CMS IP at Startup   Off          Accutorr 7 Operator s Manual          FOR YOUR NOTES       C 6 Accutorr 7 Operator s Manual    D Alarm Messages    This chapter lists only the most important physiological and technical alarm messages   Some messages appearing on your monitor may not be included  In the    Solution     column  corresponding solutions are given instructing you to troubleshoot problems  If  the problem persists  contact your service personnel   m       means the alarm level is user adjustable     mM XX represents a measurement or parameter label  such as NIBP  PR  etc     D 1 Physiological Alarm Messages                
14.   Pe NS    A    Insert a USB drive into the USB connector on the monitor   Select  Main     Transfer Data to USB     Select  Period  and select your desired time    Select  Transfer Data      CAUTION    The USB drive you use may have write protect function  In this case  please    make sure the USB drive for data transfer is in read write mode     Do not remove the storage medium during data transfer process  Otherwise   data files may be damaged     The normal monitoring function will be affected during data export  Do not  perform any monitoring activity during data export        Accutorr 7 Operator s Manual 4 9    FOR YOUR NOTES       Accutorr 7 Operator s Manual    5 Managing Configurations    5 1 Overview  When performing continuous patient monitoring  the clinical professional often needs to  adjust the monitor s settings according to the patient s condition  The collection of all  these settings is called a configuration   The monitor provides different sets of  configurations to accommodate the varying patient categories and departments  You can  change the settings from a default configuration and then save it as a user configuration           A WARNING         The configuration management function is password protected  The  configuration management tasks must be performed by clinical professionals           The system configuration items can be classified as   E Parameter configuration items   These items relate to parameters  e g  alarm switch  alarm limits   H   
15.   Plug the power cord into the AC power source  If you run the monitor on battery  power  ensure that the battery is sufficiently charged     4  Press the power on off switch on the monitor s front panel     The monitor will perform alarm system self test during start up  After pressing the power  on off button  the system sounds a beep  and the alarm lamp simultaneously turns yellow   then red  and then turns off  with the start up screen being shown  Then the start up  screen disappears  The alarm system self test succeeds  The monitor enters the normal  monitoring screen           Vi Swsinnic          Do not use the monitor on a patient if you suspect it is not working properly  or  if it is mechanically damaged  Contact your service personnel or Mindray        Check that visual and auditory alarm signals are presented correctly when the  equipment is powered on  Do not use the equipment for any monitoring  procedure on a patient if you suspect it is not working properly  or if it is  mechanically damaged  Contact your service personnel or Mindray           NOTE       Carefully check if the system performs the self test as described above  Contact  your service personnel or Mindray if the self test is abnormal        3 4 Accutorr 7 Operator s Manual    3 3 2 Turning off the Monitor    Before turning off the monitor     1   2   3     Ensure that monitoring of the patient has been completed   Disconnect cables and sensors from the patient   Make sure to save or clear the pat
16.   Pulse  and  Continuous         Pulse   the nurse call signal is a pulse signal and each pulse lasts 1 second   When multiple alarms occur simultaneously  only one pulse signal is output  If  an alarm occurs but the previous one is not cleared yet  a new pulse signal will  also be outputted        Continuous   the nurse call signal lasts until the alarm ends  i e  the duration of  a nurse call signal equals to that of the alarm condition     4  Select  Contact Type  and toggle between  Normally Open  and  Normally    Closed       Normally Open   Select if your hospital s nurse call relay contact is normally  open        Normally Closed   Select if your hospital   s nurse call relay contact is normally  closed     5  Select  Alarm Level  and set the alarm level for nurse call triggering alarms     Select  Alarm Category  and then select the category to which the nurse  call triggering alarms belong     Alarm conditions are indicated to nurses only when   1  The nurse call system is enabled       6 10 Accutorr 7 Operator s Manual    2  Analarm that meets your preset requirements occurs  and  3  The monitor is not in the alarm paused or silence status     If no setting is selected from  Alarm Level  or  Alarm Category   no nurse call signal will  be triggered if alarms occur              A WARNING  e    To obtain the nurse call signal  use the nurse call cable  PN  009 003116 00  we  supply  Otherwise the nurse call function will not work and the monitor may be  damaged        
17.   loose  the cuff application  site  If the error  persists  replace the  cuff   D 2 Accutorr 7 Operator s Manual                                           Alarm Clearable  Alarm Cause Solution   Message  Yes No  Level   NIBP Yes Low The measured NIBP   Contact Mindray or   Overrange value is not within your service  the specified range    Personnel    NIBP Yes Low Patient s arm moves   Check the patient   s   Excessive too much  condition and reduce   Motion the patient motion    NIBP Cuff   Yes Low The NIBP airway may   Check the airway and   Overpress  be occluded  measure again    NIBP Signal   Yes Low The NIBP signal is   Check the patient s   Saturated saturated  condition and reduce   the patient motion or  other sources    NIBP Air Leak   Yes Low The NIBP airway may   The NIBP cuff is not  leak air  properly connected  or   there is a leak in the  airway    NIBP Equip Err   Yes High System error  or   Check the patient s  pump  A D sampling condition and NIBP  or pressure connections  or  transducer error  or   replace the cuff   pointer error during  software running    NIBP Timeout   Yes Low Time is out  In  Adult Pediatric  mode  the  measurement time  is over 120 seconds   in neonate mode   the time is over 90  seconds    NIBP Cuff   Yes Low The cuff type   Verify the patient   Type Wrong applied mismatches   category and replace  the patient category    the cuff    NIBP Illegally   Yes Low An illegal reset   Check if the airway is   Reset occurred during   occlud
18.  0 38 0 73  1 1 20 1 20 2 30  10 3 80 3 80 7 30  100 12 00 12 00 23 00                   For transmitters rated at a maximum output power not listed above  the recommended  separation distance d in metres  m  can be estimated using the equation applicable to the  frequency of the transmitter  where P is the maximum output power rating of the transmitter  in watts  W  according to the transmitter manufacturer     Note 1  At 80 MHz and 800 MHz  the separation distance for the higher frequency range    applies     Note 2  These guidelines may not apply in all situations  Electromagnetic propagation is  affected by absorption and reflection from structures  objects and people           B 4    Accutorr 7 Operator s Manual       B 2 Radio Regulatory Compliance    RF parameters       Protocol    IEEE 802 11b g n       Modulation mode    DSSS and OFDM       Operating  frequency    2412 2472 MHz       Channel spacing    5 MHz       Wireless baud rate    IEEE 802 11b  1 11 Mbps  IEEE 802 119  6 54 Mbps  IEEE 802 1 1n  6 5 65 Mbps       Output power     lt  20 dBm  CE requirement  detection mode   RMS     lt  30 dBm  FCC requirement  detection mode     peak power         Operating mode    Infrastructure          Data security       WPA2 PSK encryption supported           The device  when configured with the WM1010BGN wireless module  complies with part  15 of the FCC Rules  Operation is subject to the condition that this device does not cause  harmful interference     Operation of this 
19.  Clearable  Alarm Cause Solution  Message  Yes No  Level  Temp Init Err Yes High An error occurred to   Restart the monitor  If  the Temp module  or   the error remains   there is a problem   contact your service  with the   personnel   communications  between the  module and the  monitor  or Temp  calibration error   Temp Comm   No High An error occurred to   Restart the monitor  If  Err the Temp module  or   the error remains   there is a problem   contact Mindray or  with the   service personnel   communications  between the  module and the  monitor   Temp Alarm   No High The alarm limit of   Contact Mindray or  Limit Err Temp is changed   your service  inadvertently  personnel   Warmup Yes Med The initial probe   Cool the probe and  Timed Out temperature in   retry   measurement is too  high   Warming No Med The thermal resistor   Replace temperature  Resistor Err on the temperature   probe   probe has an error   Can not work  properly    Ambient Yes Med The environmental   Change an  Temp temperature is out   environment and  Overrange the range of the   retry   monitor s  measurement   Temp Voltage   Yes Med The voltage is too   Check the power  Err high or too low  supply   Temp Yes Low The measuring   Retry the  Prediction Err operation is   measurement   improper   Temp SelfTest   No High An error occurs   Replace the module   Err during Temp  module  initialization                       D 6    Accutorr 7 Operator s Manual                                             
20.  Male   DOB  ryyymmd   1963 4 a1    Obtain Patient Information  gt  gt     Ok Cancel    2  Enter the demographic details     If the monitor is connected to ADT database through the eGateway  when you input  patient MRN  the patient information in the same MRN stored in ADT database will  automatically synchronize to patient demographics of the monitor     3  Select  Ok      You can configure the  Patient Demographics  menu     1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt     enter the required  password   select  Ok  to access  User Settings  menu     2  Select  Patient Demographics  gt  gt          Accutorr 7 Operator s Manual 4 1    3  Enter information in  Patient Demographics  menu      Required Information   these items are required and must be entered when  you select the  Save  QuickKey to manually save patient information  and     Optional Information   this information is optional and not required when  admitting a patient   4  Select   to save the configuration and exit the menu   NOTE    In  User Settings  gt  gt      Patient Demographics  gt  gt      gt  Patient Demographics  Setup  menu  the     symbol indicates a required field  The patient demographic  information cannot be saved until all the required data is entered      Ifaclinician ID is entered  but there is no activity on the monitor fora    configured period of time  the monitor clears the ID  Refer to Section 3 5 3  Configuring the Timeout of Clinician ID     4 1 2 Admitting a Pati
21.  Monitor mode   lt 60 s  Predictive mode   lt 20 s  typical test   lt  12s                  Alarm limit Range Step  z  low limit  1    C to 44  C  Temp High ARS    low limit  1 8    C to 111 2  F 0 1  C  25  C to  high limit  1    C 0 2  F  Temp Low Sra aaa  77  F to  high limit  1 8    F             Statistical Results of Clinical Investigation Data                   Clinical BIAS  Acb  Limits of Agreement  LA    Clinical Repeatability  or   Oral 0 02 0 33 0 1  Axilla   0 06 0 38 0 13  Rectum    0 05 0 48 0 14                   Accutorr 7 Operator s Manual A 9          FOR YOUR NOTES       Accutorr 7 Operator s Manual    B EMC and Radio Regulatory Compliance    B 1 EMC    The device meets the requirements of IEC 60601 1 2  All the accessories listed in Chapter  15 also meet the requirements of IEC 60601 1 2 when in use with this device     NOTE      Using accessories  transducers and cables other than those specified may result  in increased electromagnetic emission or decreased electromagnetic immunity  of the equipment       The device or its components should not be used adjacent to or stacked with  other equipment  If adjacent or stacked use is necessary  the device or its  components should be observed to verify normal operation in the  configuration in which it will be used       The device needs special precautions regarding EMC and needs to be installed  and put into service according to the EMC information provided below      Other devices may interfere with this de
22.  acetone based cleaners         Accutorr 7 Operator s Manual 14 1          ne       Be sure to turn off the monitor and disconnect all power cables from the outlets  before cleaning the equipment     Use only Mindray approved cleaners and disinfectants and methods listed in  this chapter to clean or disinfect your equipment  Warranty does not cover  damage caused by unapproved substances or methods     We make no claims regarding the efficacy of the listed chemicals or methods as  a means for controlling infection  For infection control methods  consult your  hospital   s Infection Control Officer or Epidemiologist           Deacon    If you spill liquid on the equipment or accessories  contact Mindray or your  service personnel     14 1 1 Cleaning    Clean your equipment on a regular basis  Consult your hospital s regulations before  cleaning the equipment     The recommended cleaning agents include     Sodium hypochlorite bleach  10    Isopropanol  70      To clean your equipment  follow these rules     1   2     3     4   5     Shut down the monitor and disconnect it from AC power   Clean the display screen using a soft  clean cloth dampened with a cleaning agent     Clean the exterior surface of the equipment using a soft cloth dampened with a  cleaning agent     Wipe off all the cleaning solution with a dry cloth after cleaning if necessary   Dry your equipment in a ventilated  cool place     14 1 2 Disinfecting    If disinfecting is required because of hospital policy  
23.  alarms of different levels occur simultaneously  the monitor will  select the alarm of the highest level  light the alarm lamp and give alarm  sounds accordingly  Multiple alarm messages scroll on the screen      Some physiological alarms  such as the Desat alarm  are exclusive  They have  identical alarm tones and alarm lights as normal high level physiological  alarms  but their alarm messages are displayed exclusively  When an exclusive  physiological alarm and a normal high level physiological alarm are triggered  simultaneously  only the exclusive physiological alarm message is displayed     6 3 3 Alarm Messages    When an alarm occurs  an alarm message will appear in the technical or physiological  alarm area     The alarm message has different background color which matches the alarm level       High level alarms  red  E Medium level alarms  yellow  m Low level alarms  yellow    For physiological alarms  the asterisk symbols     before the alarm message match the  alarm level as follows       High level alarms  ee  E Medium level alarms  eS  E Low level alarms  X    When there are multiple messages  the messages scroll     6 3 4 Flashing Numerics  If an alarm triggered by an alarm limit violation occurs  the numeric of the measurement  in alarm will flash every second  and the corresponding alarm limit will also flash at the  same frequency indicating the high or low alarm limit is violated     6 3 5 Alarm Status Symbols    Apart from the aforementioned alarm indicato
24.  and the remaining alarm pause time are displayed in    For physiological alarms  alarm indicators are not shown  New physiological alarms    will not be presented  If the alarm condition still exists once the pause time expires   the alarms will be presented     remain presented     For technical alarms  alarm sounds are paused  but alarm lamps and alarm messages    The default alarm pause time is 2 minutes  When the alarm pause time expires  or the low    cancel the alarm paused status by pressing and holding the    2 seconds        battery alarm occurs  the alarm paused status is automatically deactivated  You can also    hardkey for more than       Accutorr 7 Operator s Manual       6 7 Switching Off Alarm Sound    When alarm volume is set to 0  the alarm sound is turned off  In the audio alarm off state     E The Audio Off symbol is displayed       Audio indication of all alarms is suspended     You can cancel the alarm sound off status by setting alarm volume to a value from 1 to 10           A WARNING          Pausing or switching off alarms may result in a hazard to the patient  Please be  very careful           6 8 Resetting Alarms    By pressing the hardkey  you can reset the alarm system to acknowledging the  on going alarms and enable the alarm system to respond to a subsequent alarm  condition     For physiological alarms  except the NIBP related alarms  when the alarm system is reset       The alarm sound is silenced   A V appears before the alarm message  indic
25.  between 0 3 and 1 is acceptable  Below 0 3  indicates low perfusion  reposition the SpO2 sensor or find a better site  If low  perfusion persists  choose another method to measure oxygen saturation if possible   Pulse rate  derived from pleth wave   detected pulsations per minute  Obtain PR  through the SpO   or NIBP measurement  When simultaneously measuring NIBP and  SpO2  the PR source is from SpO2     w    A    ni       Accutorr 7 Operator s Manual 7 1    NOTE       If the message  SpO  Low Perf   or  SpO2 Weak Pulse  displays  check sensor  application  re apply or remove the sensor if necessary  to obtain a better  signal       A function tester or SpO  simulator can be used to verify the sensor functions       A function tester or SpO  simulator can be used to determine the pulse rate  accuracy    e A function tester or SpO2 simulator cannot be used to assess the accuracy of an  SpO2 module or an SpO  sensor    7 2 Safety          WARNING       A    Only use SpO  sensors specified in this manual  Follow the SpO2 sensor   s  instructions for use and adhere to all warnings and cautions     When a trend toward patient deoxygenation is indicated  blood samples should  be analyzed by a laboratory co oximeter to completely understand the patient s  condition     Prolonged monitoring may increase the risk of undesirable changes in skin  characteristics  such as irritation  reddening  blistering or burns  Inspect the  sensor site every two hours and move the sensor if the ski
26.  high priority alarms and medium priority alarms are latched     Only the unacknowledged physiological alarm can be latched  The latched alarms will be  cleared when the monitor enters the alarm reset state        Accutorr 7 Operator s Manual 6 9    NOTE       Changing of alarm priority may affect the latching status of corresponding  alarm  Determine if you need to reset the latching status for the specific alarm  after changing its alarm priority     6 11 Actions for Alarm Occurrence  When an alarm occurs  observe the following steps and take proper actions   1  Check the patient s condition    Confirm the alarming parameter or alarm category    Identify the source of the alarm    Take proper action to eliminate the alarm condition    Make sure the alarm condition is corrected     BO NS    For details about how to deal with specific alarms  refer to Appendix D Alarm Messages     6 12 Nurse Call  The monitor provides a multi function connector to output nurse call signals when a  user defined alarm occurs  To obtain a nurse call signal  use the nurse call cable  P N   009 003116 00  we supply to connect the hospital nurse call system to the  multi function connector of the monitor and then follow this procedure   1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt      Enter required  password   Select  Ok  and then access  User Settings  menu     2  Select  Nurse Call  gt  gt   to access the  Nurse Call Setup  menu   3  Select  Signal Type  and toggle between
27.  may fluctuate rather than remaining steady for a period of several  seconds  Often  the patient   SpO2 may fluctuate above and below an alarm limit   re entering the non alarm range several times  During such fluctuation  the monitor  integrates the number of  SpO  points  both positive and negative  until either the  Sat Seconds limit is reached  or the patient  SpO gt 2 re enters the non alarm range and  remains there        Accutorr 7 Operator s Manual 7 5    7 5 7 Changing the Speed of Pleth Wave    In the  SpO2 Setup  menu  select  Sweep  and then select the appropriate setting  The  faster the waveform sweeps  the wider the waveform is     7 5 8 Setting the Alarm Level for SpO2 Sensor Off Alarm    In the  SpO2 Setup  menu  select  SpO2SensorOff Lev   and then select the appropriate  setting     7 6 Measurement Limitations    If the SoO2 measurement seems out of range or inaccurate  check the patient s vital signs   Then check the equipment and SpO  sensor  The following factors may influence the  accuracy of measurement     Ambient light   Physical movement   Low perfusion   Electromagnetic interference  such as MRI environment  Electrosurgical units    Dysfunctional hemoglobin  such as carboxyhemoglobin  COHb  and methemoglobin   MetHb     Presence of certain dyes  such as methylene and indigo carmine    Inappropriate positioning of the SpO2 sensor  or use of incorrect SpO2 sensor      E Drop of arterial blood flow to immeasurable levels caused by shock  anemia  low  
28.  min    1  h    1 5 h    2 h   The monitor automatically measures NIBP based on the specified  time interval     E    Program 1  and  Program 2   The monitor automatically measures NIBP based ona  program configured by user     9 7 2 Setting the Initial Cuff Inflation Pressure  You can set the initial cuff inflation pressure manually  In the  NIBP Setup  menu  select                               Initial Pressure  and then select the appropriate setting   The initial inflation pressure range is as follows   Patient Category Range  mmHg  Default  mmHg   Adult 80   280 160  Pediatric 80 210 140  Neonate 60   140 90  NOTE      For known hypertensive patients  you need to set the initial cuff pressure to a  higher value to reduce the measurement time     9 7 3 Setting NIBP End Tone  The monitor can issue a reminder tone at the completion of NIBP measurement  The NIBP  End Tone is off by default  You can switch it on by accessing the  NIBP Setup  menu     9 7 4 Switching On Off Measurement on Clock  In auto measuring mode  if the clock is enabled  the NIBP automatic measurement  interval will be synchronized with the real time clock     For example  if  Clock  is set to  On   and  Interval  is set to  20min   and then you start an  NIBP auto measurement at 14  03  the next measurement will be taken at 14  20  and the  following measurement time will be 14 40  15 00 and so on        Accutorr 7 Operator s Manual 9 7    9 7 5 Configuring a Custom Program  In the  NIBP Setup  menu  yo
29.  on a heart lung machine     The equipment may fail to measure or produce inaccurate blood pressure measurements  under the following conditions     m    farterial pressure pulses are hard to detect    E Inthe presence of excessive and continuous patient movement such as shivering or  convulsions    During certain cardiac arrhythmias   For pregnant or pre eclamptic patients   Rapid blood pressure changes   Severe shock or hypothermia that reduces blood flow to the peripheries    Obesity  where a thick layer of fat surrounding a limb dampens the oscillations  emanating from the artery    9 4 NIBP Measurement Mode    There are the following modes of measuring NIBP   m    Manual  measurement on demand     Auto  continually repeated measurements at set intervals       E STAT  continually rapid series of measurements over a five minute period  then return  to the previous mode     To set NIBP measurement mode  select NIBP parameter area to access  NIBP Setup   menu     E Select  Interval  to select manual mode or auto NIBP measurement interval   E Select  NIBP STAT  to start a continuous NIBP measurement     If the monitor is in Spot Check mode  you can only manually measure the NIBP  If the  monitor is in Monitor mode  you can measure auto  continuous and manual NIBP     To enable or disable Spot Check  refer to 2 6 2Spot Check Mode        Accutorr 7 Operator s Manual 9 3    9 5 Measuring NIBP  9 5 1 Preparing the Patient    In order to minimize NIBP measurement measurement error
30.  power failure  the equipment stores the configuration in real time  The saved  configuration is the latest configuration     The monitor restores the latest configuration if it restarts within 60 seconds after a power  failure  And it restores the default configuration  rather than the latest configuration  if it  restarts 120 seconds after a power failure  The monitor loads either the latest  configuration or the default configuration if it restarts from 60 120 seconds after a power  failure        Accutorr 7 Operator s Manual 5 3    FOR YOUR NOTES       5 4 Accutorr 7 Operator s Manual    6 Alarms    Alarms  triggered by an abnormal vital sign or technical issue with the monitor  are  visually and audibly indicated to the user           WARNING       A    A potential hazard can exist if different alarm presets are used for the same or  similar equipment in any single area  e g  an intensive care unit or cardiac  operating room    If your equipment is connected to a CMS  remote suspension  inhibition  and  reset of monitor alarms via the CMS may cause a potential hazard  For details   refer to the CMS s instructions for use           6 1 Alarm Categories  The equipment   s alarms can be classified into two categories  physiological alarms and  technical alarms     1     Physiological alarms   Physiological alarms  also called patient status alarms  are triggered by a monitored  parameter value that violates set alarm limits or an abnormal patient condition  The  physiologic
31.  protocol     12 2 Network    12 2 1 Network Connection  The monitor can be connected to the central station through the network     E The Accutorr 7 transmits waveforms and numerics of parameters  Temp  SpO2  NIBP    related alarms and alarm settings  patient information  and operating mode to the  central station  The waveforms  numerics  alarms  alarm settings  patient information   and operating mode displayed on central station are consistent with Accutorr 7     E The central station transmits alarm settings  parameter settings  patient information  and operating mode settings  enter or exit the Standby mode  to the Accutorr 7  The  alarm settings  parameter settings  patients    information and operating mode  settings of the Accutorr 7 are consistent with central station     The monitor can be connected to the ADT system and the EMR system through the  eGateway         The Accutorr 7 transmits the query command message to the ADT system  When the  ADT system receives the query command message  the ADT system will transmit the  patients    information to the Accutorr 7     E The Accutorr 7 transmits numerics of Temp  SpO2  NIBP  related alarms and alarm  settings  patient information  operating mode  and historical data  including trends  and events  to the EMR  When the EMR receives the data  the EMR will send a success  message to the Accutorr 7        CAUTION          Disconnecting from the network may result in data loss  including parameter  waveforms and measureme
32.  s Manual 2 13    FOR YOUR NOTES       2 14 Accutorr 7 Operator s Manual    3 Basic Operation    3 1 Installation          Winne       The equipment should be installed by authorized Mindray personnel     Do not open the equipment housings  All servicing and future upgrades must  be carried out by Mindray trained and authorized personnel     The software copyright of the equipment is solely owned by us  No organization  or individual shall resort to altering  copying  or exchanging it or to any other  infringement on it in any form or by any means without due permission     Connect only approved devices to this equipment  Devices connected to the  equipment must meet the requirements of the applicable IEC standards  e g   IEC 60950 safety standards for information technology equipment and IEC  60601 1 safety standards for medical electrical equipment   The system  configuration must meet the requirements of the IEC 60601 1 medical electrical  systems standard  Any personnel who connect devices to the equipment   s  signal input output port is responsible for providing evidence that the safety  certification of the devices has been performed in accordance to the IEC  60601 1  If you have any questions  please contact Mindray     If it is not evident from the equipment specifications whether a particular  combination with other devices is hazardous  for example  due to summation of  leakage currents  please consult the manufacturers or else an expert in the  field  to ensure th
33.  sent out when the monitor is connected to Electronic    Medical Record system  EMR         4 6    Accutorr 7 Operator s Manual    In the spot check tabular trends     Parameter value triggering high level alarm has a red background  parameter value  triggering medium or low level alarm has a yellow background     Patient data successfully sent to the EMR displays as v   Patient data that failed to transmit to the EMR displays as x     Patient data that is cached on the device but not transmitted displays as Q    4 4 2 Continuous Trends  Select  Continuous Trends  tab to access the Continuous Trends screen     1     Spot Check Trends Continuous Trends Graphic Trends    Delete All 2       Tabular trend 2  Button area 3  Parameter value triggering alarm    In this tab  you can     Select  MRN  or  Name  to select the desired patient   Select  Interval  to set the interval for the data to be displayed     Select the page key  for example       _    and then use up or down arrow  beside the key to turn the page   Select  Print  to print the trend data of selected patient     Select  Delete All  to delete the trend data of selected patient     Parameter value triggering high level alarm has a red background  parameter value  triggering medium or low level alarm has a yellow background        Accutorr 7 Operator s Manual 4 7    4 4 3 Graphic Trends    Select  Graphic Trends  tab to access the Graphic Trends screen  The Graphic Trend  screen displays the current patient s physiologi
34.  system time  Reset with the button cell    and restart the  monitor  If the  problem still exists   contact Mindray or  your service  personnel   PWR Yes Low Power supply to the   Check the  interrupted  monitor was   measurements when  Check meas  interupted  the monitor restarts  If    State                 the problem still  exists  contact your  service personnel or  Mindray           Accutorr 7 Operator s Manual          D 2 6 System Alarm Messages             Alarm Clearable  Alarm Cause Solution  Message  Yes No  Level  No CMS Yes Low The monitor is Check network             disconnected from  the CMS        connection           Accutorr 7 Operator s Manual       FOR YOUR NOTES       D 10 Accutorr 7 Operator s Manual    E Symbols and Abbreviations    E 1 Symbols    pA  uV  Us  A  Ah  bpm  bps    C  cm  dB  DS  oF    GHz    min   ml  mm  mmHg    ms    microampere  microvolt  Microsecond  ampere  ampere hour  beat per minute  bit per second  centigrade  centimeter  decibel   dyne second  fahrenheit  gram   gigahertz   hour   hertz   inch   kilo   kilogram  kilopascal   litre   pound   meter  milliampere hour  mega byte  milligram  minute   milliliter  millimeter  millimeters of mercury    millisecond       Accutorr 7 Operator s Manual       mV millivolt   mW milliwatt   MQ megaohm   nm nanometer   rpm breaths per minute  s second   V volt   VA volt ampere   Q ohm   WwW watt     minus     negative     percent     per  divide  or   ae to     plus     equal to    lt  less 
35.  the monitor is powered on  the    battery status symbol on the monitor screen displays is  when the charging is  complete     13 3 Replacing a Battery  1  Power off the monitor   Open the battery compartment door   Push aside the battery latch and remove the battery   Place a new battery into the slot with its contact point inward     E ae la a    Close the battery compartment door     13 4 Battery Guidelines    Life expectancy of a battery depends on how frequently and how long it is used  Fora  properly maintained and stored lithium ion battery  its life expectancy is about 3 years   For more aggressive use models  life expectancy can be less  We recommend replacing  lithium  ion batteries after 500 full charge discharge cycles or every 3 years from first use   whichever occurs first     To get the most out of the battery  observe the following guidelines     E The battery performance test must be performed once a year  before monitor repairs   or whenever battery performance is suspect     H Condition the batteries every three months  or when their run time becomes  noticeably shorter     E Remove the battery before shipping the monitor or if it will not be used for an  extended period of time     E Remove the battery from the monitor if it is not being used regularly   Leaving the  battery in a monitor that is not in regular use will shorten the battery life      m  When storing batteries  make sure that the battery terminals do not come into  contact with metallic objec
36. 1  D 2 Technical Alarm M SSaQe S       sssssssssssssssssssssssssssssesssssssesssssssecsssssseessssnscessssnseessssnseesssssseesssssseesssssess D 1  E Symbols and Abbreviations              cccccsssssscsssssssscsssssscesssssessssssessessscssessscssessssssessossees E 1  E  Fe SWIMS sisastscssssassasccsssaassassssscossnsssaadasscarsacsscausdegepnsusiasssebsconsisastsosnachsistaas ebsssshasssassecgongasasiasoscssnoasssedtaies E 1  E22 Ab  reviatio NS aisean esai Eoia E n a ATTER EO ans ESE AER NEEESE E 3  F I R A T T E EE S EE F 1  F 1 NIBP Smart Inflation ANOMAly         sssssssscccccsssssssscseccsssssseeeccesesssssessceeeeessssssesseeeeesnsneeseeseeesnnnseeeseeee F 1       6 Accutorr 7 Operator s Manual    1 Safety    1 1 Safety Information          Miani      Indicates a potential hazard situation or unsafe practice that  if not avoided   could result in death or serious injury           ee      Indicates a potential hazard or unsafe practice that  if not avoided  could result  in minor personal injury or product property damage     NOTE       Provides application tips or other useful information to ensure that you get the  most from your product        Accutorr 7 Operator s Manual 1 1    1 1 1 Warnings          Because    This equipment is restricted to one patient at a time     Before putting the system into operation  the operator must verify that the  equipment  connecting cables and accessories are in correct working order and  operating condition    To avoid risk of elec
37. 5 hours until the battery is full    4  Remove the AC power and allow the monitor to run from the battery until the  battery is completely depleted and the monitor automatically shuts off    5  Again connect the monitor to the AC power  Fully charge the battery again for use or  charge it to 40   60  for storage    NOTE      The battery charge level indicator does not indicate the capacity or operating  time of the battery  It only indicates the current battery charge level  The actual  battery capacity decreases over time with the use of the battery  For an old  battery  its capacity and operating time may not fulfill battery specifications  even if the battery charge level indicates the battery is in full charge  Please  replace the battery if its operating time is significantly lower than the specified  time    Accutorr 7 Operator s Manual 13 3    13 5 2 Checking a Battery  The performance of a rechargeable battery will deteriorate over time  The battery    performance test must be performed once a year  before monitor repairs  or whenever  battery performance is suspect     To check battery performance    1  Disconnect the monitor from the patient and stop all monitoring or measuring    2  Turn off the monitor  Disconnect the monitor from the AC power    3  Install the battery    4  Connect the monitor to AC power  Allow the battery to be charged uninterrupted for  6 5 hours until the battery is full     5  Note the time  Remove AC power and allow the monitor to run from 
38. 6 1    6 1 Alarm Categories  6 2 Alarm Levels     6 3 Alarm Indicators     6 3 1 Alarm Lamp  6 3 2 Audible Alarm Tones    6 3 3 Alarm Messages  6 3 4 Flashing NUMELICS      c  cssssccssssccsssscssseccssseccsseccssscecsssecesseecssnecesssecesnsecesuscccssccesusecesaeecsasecesseees                2 Accutorr 7 Operator s Manual    6 3 5 Alarm Status SYMbOIS       cssssssscccssssescccsssscecssssecsessnseceessnnsecsessnseceesssssssessnsseessannsesseenseeseste 6 3          GA SECC Ala MS csesscccssascssssassacasecssceasacscuscssssasstasuedessiacasisscoasdecasacesdsicaayasosussscsscacousiscsscataastscassasotayensiasies 6 4   6 5 Selecting Alarm Properties             6 5  6 5 1 Changing the Alarm VOlUME        sssssssscssssscssseccssseccsssecesuscesssecesuecesssecssneecssueecssseessnecesaeees 6 5  6 5 2 Setting the Minimum Alarm VOIUME        ssessssseccsseccssseccsseccssseccsseccssneecssseessueeessneessneeess 6 5  6 5 3 Setting the Interval between Alarm SOUNAS       escssssccsseccsseeccssssccseecsseccssseecsnecesseees 6 5  6 5 4 Adjusting Alarm Limits Automatically    ssssssscccssssesscsssessscsnnsesessnseesecsnssesessnneesseenee 6 6   6 6 Pusing Alarm S esesissssccesssscosssssnssscnsdezenasescsazasonsgsctazsacaissasasasecasvedatenscadscoaisecossschana asaigctessiscosateiobisesaiaencs 6 7   6 7 Switching Off Alarm Sound     cssssssssssssssssssssssssssssssssesssssnsesssssssscsssssuesessssssecssssueecssssnsecssssnseessssnsesss 6 8   6 8 Resetting Alarms            0   6 9 Setting th
39. ADT database through the eGateway  the monitor can  obtain patient information from the ADT database     1   2   3     Select  Patient List  QuickKey   Select  ADT Database   and the ADT database screen displays   Search a patient       Input  Department  name  and then the system searches for the patient within  the department         Input  MRN   and then the system searches for the patient according to the  input medical record number       Input  Last Name  and or  First Name   and the system searches for the patient  by last name and or first name     Select the page key  for example         Y   and then use the up or down arrow  beside the key to turn to another page  or select  Scroll  and then use the up or  down arrow beside the key to select the desired patient     Select  Admit  to access  Patient Demographics  menu   Select  Ok   If necessary  you can modify patient information and then select  Ok      4 2 Manually Input Patient Data  You can choose whether to display the manual input area on the screen and configure the  items to be displayed in the manual input area     1     Select  Main      Maintenance  gt  gt      User Settings  gt  gt     enter the required  password    Select  Ok  to access  User Settings  menu     Select  Manual Inputs gt  gt     In the  Manual Inputs Setup  menu  you can    Configure the items to be displayed in the manual input area       Select  Manual Input Area  and toggle between  On  and  Off  to display or  hide the manual input a
40. Accutorr 7  Vital Signs Monitor    Operator   s Manual    C    0123       Copyright 2013 2015 Shenzhen Mindray Bio Medical Electronics Co   Ltd  All rights  reserved     Release date  January 2015   Revision  3 0       Accutorr 7 Operator s Manual    Intellectual Property Statement    SHENZHEN MINDRAY BIO MEDICAL ELECTRONICS CO   LTD   hereinafter called Mindray   owns the intellectual property rights to this Mindray product and this manual  This  manual may refer to information protected by copyrights or patents and does not convey  any license under the patent rights of Mindray  nor the rights of others     Mindray intends to maintain the contents of this manual as confidential information   Disclosure of the information in this manual in any manner whatsoever without the  written permission of Mindray is strictly forbidden     Release  amendment  reproduction  distribution  rental  adaption and translation of this  manual in any manner whatsoever without the written permission of Mindray is strictly  forbidden           Il    mindray minpray  aNd MINDRAY are the registered trademarks or trademarks    owned by Mindray in China and other countries  All other trademarks that appear in this  manual are used only for editorial purposes without the intention of improperly using  them  They are the property of their respective owners        Manufacturer   s Responsibility    Contents of this manual are subject to changes without prior notice     All information contained in this 
41. Conventional configuration items    These items define how the monitor works  e g   display setup  print and alarm  settings     m User maintenance items  These items relate to user maintenance settings  e g   unit setup  time format     For the important configuration items and their default values and user maintenance  items  see Appendix C Default Configurations     5 2 Accessing  Manage Configuration  Menu    1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt    Enter the required password  and then select  Ok      2  Select  Manage Configuration  gt  gt          Accutorr 7 Operator s Manual 5 1    5 2 1 Setting Default Configuration    The monitor will load the pre set default configuration in the following cases   E The monitor restarts after being switched off for more than 120 seconds   E A patient is admitted    m A patient is discharged       Patient category is changed     The default configuration may come from the latest configuration  the factory default  configuration or the user configuration     To set the default configuration   1  Select  Select Default Config  gt  gt   in the  Mange Configuration  menu     2  Select  Load the Latest Config  or  Load the Specified Config  in  Select Default  Config  menu     When you select  Load the Specified Config   the configuration  adult  pediatric or  neonate  to be restored is subject to the patient category  This configuration can be either  factory configuration or saved user configuration  Take ad
42. Do not rely exclusively on the nurse call system for alarm notification   Remember that the most reliable alarm notification combines audible and  visual alarm indications with the patient s clinical condition              Accutorr 7 Operator s Manual 6 11    FOR YOUR NOTES       Accutorr 7 Operator s Manual    7 Monitoring SpO gt 2    7 1 Overview  SpO2 monitoring is a non invasive technique used to measure the amount of oxygenated  hemoglobin by measuring the absorption of selected wavelengths of light  The light  generated in the probe passes through the tissue and is converted into electrical signals  by the photodetector in the probe  The SpO2 module processes the electrical signal and  displays a waveform and digital values for SpO2 and pulse rate     This device is calibrated to display functional oxygen saturation  It provides the following        3 4 5    1  Pleth waveform  Pleth   visual indication of patient s pulse  The waveform is not  normalized     Y    PI indicator  Graphic Indication of arterial pulse signal strength     Oxygen saturation of arterial blood  SpO2   percentage of oxygenated hemoglobin in  relation to the sum of oxyhemoglobin and deoxyhemoglobin     Perfusion index  PI   Pl is available for Masimo SpO2 module  PI gives the numerical  value for the pulsatile portion of the measured signal caused by arterial pulsation  PI  is an indicator of the pulsatile strength  You can also use it to assess the quality of  SpO2 measurement  Above 1 is optimal 
43. EMP temperature  E 4 Accutorr 7 Operator s Manual    F Anomalies    F 1 NIBP Smart Inflation Anomaly    The NIBP smart inflation function is not fully functional at this time and so may not  perform as expected  The function  accessible only through the  Factory Maintenance   menu  is used for research and development only  Although there is no distinct difference    between smart inflation and normal inflation  it is recommended to keep the function   Off      NOTE      The anomaly would not present a safety hazard to either the operator or the  patient        Accutorr 7 Operator s Manual F 1    FOR YOUR NOTES       F 2 Accutorr 7 Operator s Manual    P N  046 005276 00  3 0   SW Version  03 06 00    
44. IONS sccisscesssscssescocssccsessscscsvessoonssesessostessesescossssccsessoosssecedessseonsseiess svecssssesceees    T2 MEWS  Syste Msns aa RA  EANA O A EAEE A AEE A E ESEE E A EET  12 2 1 Network Connection          4 Accutorr 7 Operator s Manual    12 2 2 Network Type and Settings         ssseessccsssessscssssesecessseseccsssesccessseececssnsecsessnseceessnnesseesnneesees 12 3  12 2 3 Setting up the Wireless Network  Optional           ssssccsssecssseccseccssteccsseeceseeeenees 12 4  12 2 4 ADT Communication Setup  12 2 5 EMR Communication Setup  12 2 6 DIAP COMMUNICATION SETUP    eseescseecseccssccsseccsecessccsseccnsccssecsnecsuesseccuscenseesseeeneeesees 12 5  12 2 7 Central Monitoring System Setup        sssssssccsssseccessssececessnseceessssecesesnseceecsnsessseanseessees 12 5  12 2 8 Setting the Multicast Parameters        esccsssesssseeccsseccssteccsseccssseeesssecesseecssnsecsseecssneeeesnees 12 5       13 Battery          cccccsssssssscsccsssssssccssccsssccsssssssssscsesssssssscscssssssesscssssesssssessssssessssssssesssssssooses 13 1    TIT OVORMICWictssscccssseacssctavsesest aiene ipat es Ee ES ee E RE IEI EE Eer EAEn ii iani  13 2Cha  ging a Battery aeaa N E A RNE EN E RN RAR  13 3 Replacing a Battery  ynstee eent eeskiri enanset  13  Batt    ry Guidelines sy sissccecswcecsscssscccsesceles sits cdessueessscssiccecesccesvastssenss EN Aa  13 5 Battery Maintenance       cessssssssssscssssscsseessssessssscssssscessscesssssessssssssscesssueesssessssseesssseessseesssse
45. If several patients are selected  the recorder will print the data in sequence  Data from  different patients are separated by              11 6 Removing a Paper Jam  If the recorder works incorrectly or produces unusual sound  check if there is a paper jam   If a paper jam is detected  follow this procedure to remove it     1  Open the recorder door   2  Take out the paper and tear off the wrinkled or torn part   3  Reload the paper and close the recorder door     Refer to section 11 2 Loading Paper for additional information     11 7 Cleaning the Recorder Printhead  If the recorder has been used for a long time  deposits of paper debris may collect on the  printhead compromising the print quality and shortening the lifetime of the roller  Follow  this procedure to clean the printhead     1  Take measures against the static electricity such as Disposable Wrist Strap for the  work     Open the recorder door and take out the paper   3  Gently wipe around the printhead using cotton swabs dampened with alcohol     After the alcohol has completely been dried  reload the paper and close the recorder  door     Refer to section 11 2 Loading Paper for additional information     A CAUTION      Do not use anything that may destroy the thermal element      Do not add unnecessary force to the thermal head        11 4 Accutorr 7 Operator s Manual    12 Other Functions    12 1 MEWS System    The MEWS  Modified Early Warning Score  system facilitates a doctor to quickly determine  the degre
46. S  OXIDE    EQUIPMENT not suitable for use in the presence of a  FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH    OXYGEN OR NITROUS OXIDE          Degree of mobility       Portable       A 2 Environmental Specifications    Main Unit       Item    Operating conditions    Storage conditions       0 to 40  without Temp module              Temperature    C  20 to 60  p co 5 to 40  with Temp module   Raive humieiy 15  to 95  10  to 95    noncondensing   Barometric  kPa  57 0 to 107 4 16 0 to 107 4                   AA WARNING         The equipment may not meet the performance specifications given here if  stored or used outside the specified temperature and humidity ranges           NOTE      The environmental specifications of parameter modules are the same as those  of the main unit        Accutorr 7 Operator s Manual          A 3 Power Supply Specifications                                  AC Power   Line voltage 100 to 240 VAC    410     Current 0 9 to 0 5A   Frequency 50 60 Hz   3Hz    Fuse T2AL 250V   Battery   Battery Type Rechargeable lithium ion  LI23S002A   Voltage 11 1 VDC   Capacity 4500 mAh  At least 8 hours when powered by a new fully charged battery   Run time at 25   C 5   C  77  F 41  F  with SpO2 cable connected  and auto  NIBP measurements at an interval of 15 minutes   Monitor power off  less than 5 5 hours to 90   less than 6 5   n hours to 100    Charge time    Monitor power on  less than 10 5 hours to 90   less than 11 5  hours to 100         Shutdown delay
47. al alarms occur only in Monitor mode     Technical alarms   Technical alarms  also called system status alarms  are triggered by a device  malfunction or a patient data distortion due to improper operation or mechanical  problems     Apart from the physiological and technical alarm messages  the monitor shows some  messages telling the system status or patient status  System related messages are  displayed in system message area  parameter related messages are displayed in the  respective parameter message area        Accuto    rr 7 Operator s Manual 6 1    6 2 Alarm Levels  By severity  the equipment s alarms can be classified into three categories  high level   medium level and low level        Physiological Alarms    Technical Alarms                High level Indicate that your patient isin   Indicate a severe device malfunction or an  a life threatening situation improper operation  which could make it  and anemergency treatment   possible that the monitor cannot detect  is demanded  critical patient status and thus threaten the  patient s life  such as low battery   Medium Indicate that your patient   s Indicate a device malfunction or an improper  level vital signs appear abnormal operation  which may not threaten the  and an immediate treatment patient s life  but may compromise the  is required  monitoring of vital physiological parameters   Low level Indicates that your patient   s Indicate a device malfunction or an improper       vital signs appear abnormal  and im
48. all not be liable for any incidental  special  or consequential loss  damage  or  expense directly or indirectly arising from the use of its products  Liability under this  warranty and the buyer   s exclusive remedy under this warranty is limited to servicing or  replacing the affected products  at Mindray option  at the factory or at an authorized  distributor  for any product which shall under normal use and service appear to Mindray  to have been defective in material or workmanship  Recommended preventative  maintenance  as prescribed in the service manual  is the responsibility of the user and is  not covered by this warranty     No agent  employee  or representative of Mindray has any authority to bind Mindray to  any affirmation  representation  or warranty concerning its products  and any affirmation   representation or warranty made by any agent  employee  or representative shall not be  enforceable by buyer or user     THIS WARRANTY IS EXPRESSLY IN LIEU OF  AND MINDRAY EXPRESSLY DISCLAIMS  ANY  OTHER EXPRESS OR IMPLIED WARRANTIES  INCLUDING ANY IMPLIED WARRANTY OF  MERCHANTABILITY  NON INFRINGEMENT  OR FITNESS FOR A PARTICULAR PURPOSE  AND  OF ANY OTHER OBLIGATION ON THE PART OF MINDRAY     Damage to any product or parts through misuse  neglect  accident  or by affixing any  non standard accessory attachments  or by any customer modification voids this warranty   Mindray makes no warranty whatsoever in regard to trade accessories  such being subject  to the warrant
49. ally        A 6 Measurement Specifications                                  A 6 1 SpO   Alarm limit Range     Step       SpO2 High  low limit   2  to 100  Masimo  Desat to  high limit   2    SpO2 Low Nellcor  Desat or 20  whichever is greater  to 1   high limit     2    Desat 0 to  high limit     2    Masimo SpO2 Module   Standards Meet standards of ISO 9919  ISO 80601 2 61   SpO2 Measurement range 1 to 100    PI measurement range 0 02  to 20    SpO  Resolution 1           Accuracy        70 to 100    2   measured without motion in adult pediatric          A 4    Accutorr 7 Operator s Manual             mode    70 to 100    3   measured without motion in neonate mode   70 to 100    3   measured with motion    1  to 69   Not specified              Refreshing rate 1s     lt  20 s  PR75 bpm  average time 8 s  SpO  value rises from  Response time J s  paniei 60  to 95    SpO  averaging time 2 4 s  4 6 s  8 s  10s  12s 14s 16s       Pulse amplitude   gt 0 02     Low perfusion conditions  R Light penetration   gt 5        Low perfusion SpO2       2   accuracy             1The Masimo pulse oximeter with sensors has been validated for no motion accuracy in  human blood studies on healthy adult volunteers in induced hypoxia studies in the range of  70  to 100  SpO2 against a laboratory co oximeter and ECG monitor  This variation equals  plus or minus one standard deviation  Plus or minus one standard deviation encompasses 68   of the population     The Masimo pulse oximeter with senso
50. ameter values displayed on the  screen  as well as SoO2 waveform  patient information and manual inputs as configured in  the  Print Setup  menu     11 5 2 Graphic Trend Recording  Graphic trend strip includes patient information  recording time  and graphic trends of all  the parameters on the screen     In graphic trends strip   m    The mark  A  is shown on the event time     E NIBP measurement taken in Spot Check mode shows as i   E The predictive temperature and PR  from NIBP  measurement show as    m        11 5 3 Continuous Trends Recording  The continuous trends strip includes patient information  recording time  measured value   the saving time for each measurement  and NIBP and Temp complete time     To print the continuous trends    1  Choose one patient  and then select  Record  in the  Continuous Trends  screen   2  Set the start and end time for the recording    3  Select  Record  to start recording     In the continuous trends strip   E The mark     before the time indicates that system time is changed   E The mark     before the parameter value indicates that an alarm is triggerd        Accutorr 7 Operator s Manual 11 3    11 5 4 Spot Check Trends Recording    The spot check trends strip includes patient information  recording time  measured value  and measurement complete time  When  Filter  is set to  Manually Saved  or  All   the  manual input data can be recorded     To print the spot check trends   1  Select the patient and filter   2  Select  Print      
51. any maintenance is required     Make sure to clean and disinfect  if required  the equipment before any test and  maintenance     Z  caution      Care should be taken to change the settings in  User Settings  gt  gt   and  Factory  Maintenance  gt  gt   menus to avoid data loss       Service personnel should acquaint themselves with the test tools and make  sure that test tools and cables are applicable              Accutorr 7 Operator s Manual 14 3       Check Maintenance Item    Recommended Frequency       Performance Test       Visual inspection    When first installed or after reinstalled                    Pressure check  NIBP test   Leakage test  SpO2test  Temp test    1  If the user suspects that the measurement is incorrect   2  Following any repairs or replacement of the module   3  Once a year for NIBP tests     4  Once every two years for SpO2 test and Temp test        Nurse call relay performance test    If user suspects that the nurse call or analog output does  not work well        Electrical safety tests       Electrical safety tests    Once every two years  or as required        Other tests       Power on test    1  When first installed or after reinstalled     2  Following any maintenance or the replacement of any  main unit parts        Touchscreen calibration    1  When the touchscreen appears abnormal   2  After the touchscreen is replaced        Recorder check    Following any repair or replacement of the recorder           Functionality test  Battery che
52. ategory setting for your patient before  measurement  Incorrect patient category selection may present a safety hazard   Do not measure NIBP on patients with sickle cell disease or on the limb where  skin damage has occurred or is expected    Use clinical judgement to determine whether to perform frequent unattended  blood pressure measurements on patients with severe blood clotting disorders  because of the risk of hematoma in the limb fitted with the cuff    Do not use the NIBP cuff on a limb with an intravenous infusion or arterial  catheter in place  This could cause tissue damage around the catheter when the  infusion is slowed or blocked during cuff inflation    NIBP measurements can be affected by the measurement site  the position of  the patient  patient movement  or the patient s physiologic condition  If the  NIBP measurement seems out of range or inaccurate  determine the patient s  vital signs by alternative means and then verify that the monitor is working  correctly    Make sure the hose connecting the NIBP cuff and the monitor is not blocked   twisted  or tangled     Do not apply the cuff on the arm on the side of a mastectomy     Continuous cuff pressure due to connection tubing kinking may cause blood  flow interference resulting injury to the patient              9 2    Accutorr 7 Operator s Manual    9 3 Measurement Limitations  The equipment cannot measure blood pressure when the patient s heart rate is below 40  bpm or above 240 bpm  or if the patient is
53. ating that the alarm is acknowledged     a  E   Theicon appears in the alarm symbol area   E The parameter numeric and alarm limits still flash     The indication of alarm lamp for the physiological alarm depends on the alarm light  setting     E When  Alarm Light on Alarm Reset  is set to  On   the alarm lamp remains flashing   E   When  Alarm Light on Alarm Reset  is set to  Off   the alarm lamp stops flashing     To set  Alarm Light on Alarm Reset      1  Select  Main     Maintenance  gt  gt      gt  User Settings  gt  gt      enter the required  password   Select  Ok  to access  User Settings  menu     2  Select  Alarm Setup  gt  gt   to enter the  Alarm Setup  menu   3  Select  Alarm Light on Alarm Reset   and toggle between  On  and  Off    The default setting for  Alarm Light on Alarm Reset  is  On         6 8 Accutorr 7 Operator s Manual    Technical alarms give different alarm indicators when the alarm system is reset     For some technical alarms  including the NIBP related alarms a v appears before    EN b EE EEA  the alarm message and appears in the alarm symbol area  indicating that the  alarm is acknowledged  The indication of the alarm lamp depends on the alarm light  setting     Some technical alarms are changed to the prompt messages   Some technical alarms are cleared  The monitor gives no alarm indications     For details about the indications of technical alarms when the alarm system is reset  refer  to D 2 Technical Alarm Messages     6 9 Setting the Remi
54. attery for power if AC power  is interrupted     Installing a Battery    The battery compartment cover is on the bottom of the monitor  Refer to 13 3 Replacing  a Battery for additional information regarding battery installation     NOTE       When a battery has been stored for a long time  or the battery is depleted   recharge the battery at once  Otherwise  the low battery may not be sufficient  to power the monitor if the AC power is unavailable     Charging a Battery    The battery is charged whenever the monitor is connected to an AC power source  regardless of whether the monitor is turned on or not     When the battery is charging  the battery indicator is On  The battery charge icon on the  screen dynamically displays the charging status when the monitor is powered on     3 2 3 Connecting Accessories  Insert the hose part of NIBP cuff to the connector on the side of monitor  insert the SoO2  cable into the SpO2 cable connector on the side of the monitor  insert the temperature  probe cable into the TEMP probe connector on the back of the monitor        Accutorr 7 Operator s Manual 3 3    3 3 Turning On Off Power    3 3 1 Turning Power On    Once the monitor is installed  before beginning measurements     1  Check the monitor for any mechanical damage  and make sure that all external  cables  plug ins and accessories are properly connected     2  Check the power supply specification is met if mains power is used  Only use a power  outlet that is properly grounded     3
55. ble        Accutorr 7 Operator s Manual 2 1    2 3 Main unit  2 3 1 Front View    wf wn    mindray Accutorr   7                      Alarm indicator    When a physiological alarm or a technical alarm occurs  this indicator will flash as  defined below       High priority alarm  the lamp quickly flashes red      Medium priority alarm  the lamp slowly flashes yellow      Low priority alarm  the lamp is yellow without flashing    Display screen   Temperature probe well   Probe cover pack holder   AC power indicator     On  indicates that the monitor is connected to the AC power      Off  indicates that the monitor is not connected to the AC power   Power ON OFF switch     Press this key to turn the monitor on      Ifno parameter is being measured  press this key to enter Standby mode         When the monitor is on  press and hold this key for more than 2 seconds to turn  the monitor off     An indicator is built into this switch  It turns on when the monitor is on and turns off  when the monitor is off        Accutorr 7 Operator s Manual    7  Battery indicator    On  indicates that the battery is installed and the AC power is connected       Off  indicates that no battery is installed when AC power is connected  or  indicates that the battery is installed  but no AC power is connected when the  monitor is powered off       Flashing  indicates that the monitor is powered by battery   8  Alarm Reset key    Press this key to disable the audio of present alarms     Press and hold t
56. cal trend     Spot Check Trends   Continuous Trends             a Scroll  gt     30 min       1  Event mark area 2 Time scale 3  Graphic area  4  Parameter area 5  Cursor 6  Button area    A timestamp indicating your current position is displayed above the parameter area  The    parameter value corresponding to the cursor time appears in the parameter area  The  measurement value that triggered a high level alarm has red background  The  measurement that triggered the medium low level alarm has a yellow background     Events are marked with colors in the event mark area  Red represents a high level alarm  event  Yellow represents a medium low level alarm event     In the Graphic Trends screen  you can   E Select  Zoom  to set the time length of the trend displayed on the screen     m Select  Scroll  and control    or    key to move the cursor  Select     or  gt   to move the cursor to the previous or next page      lt       Select  Event  and control or    key to quickly locate the event     E Select  Print  to print the selected patient s graphic trend data currently on the  screen        Accutorr 7 Operator s Manual    NOTE    Pausing or switching off alarms will not be recorded as events  The time of these    operations will not be recorded in the system log   A total loss of power has no impact on the saved events   Events recorded earlier might be overwritten by later ones if the storage    memory reaches capacity     4 5 Transferring Data from the Monitor to USB Drive  
57. ce to radio or  television reception  which can be determined by turning the equipment off and on  the  user is encouraged to try to correct the interference by one or more of the following  measures        Accutorr 7 Operator s Manual B 5    Reorient or relocate the receiving antenna   Increase the separation between the equipment and receiver     Connect the equipment into an outlet on a circuit different from that to which the  receiver is connected     Consult the dealer or an experienced radio TV technician for help     The maximum antenna gain permitted complies with the e i r p  limits as stated in  RSS 210    The maximum antenna gain permitted complies with the e i r p  limits specified for  point to point operation  as stated in RSS 210     CE    The radio device used in this product is in compliance with the essential requirements  and other relevant provisions of Directive 1999 5 EC  Radio Equipment and  Telecommunications Terminal Equipment Directive            J  WARNING       This compliance to general radiation exposure limits for an uncontrolled  environment  and minimum of 20 cm separation between monitor and human  body              B 6    Accutorr 7 Operator s Manual    C Default Configurations    This chapter lists some of the most important factory default settings in configuration  management  You cannot change the factory default configuration itself  However  you  can make changes to the settings from the factory default configuration and then save  th
58. ck    1  When first installed   2  Whenever a battery is replaced        Performance test          Once a year or if the battery run time reduced  significantly           14 4    Accutorr 7 Operator s Manual       14 4 Checking Monitor Information  Select  Main     Maintenance  gt  gt      Monitor Information gt  gt    you can view    system software version   copyright information   system configuration by selecting  Monitor Configuration gt  gt    or   status information  such as start time  self test error  and so on by selecting  Monitor    Log gt  gt    Monitor Information   x      System Software Version  V3 0    Internal Version  01 00 00 SVN 3367 Mar 21 2013  10 13 15     C 2005 2013 Mindray Co   Ltd All rights reserved    Monitor Configuration  gt  gt     Monitor Log  gt  gt     You can print out the log information for the convenience of troubleshooting  Select   Recorder  from the  Monitor Log  menu to do recording  The information will not be  saved after system shutdown     14 5 Visual Inspection    Perform an overall inspection on the appearance of the equipment  The test is passed if  the equipment has no obvious signs of damage  Follow these guidelines when inspecting  the equipment     1    Carefully inspect the case  display screen  buttons  and knob for obvious signs of  damage     Inspect all external connections for loose connectors  bent pins or frayed cables   Inspect all connectors on the equipment for loose connectors or bent pins   Make sure that saf
59. cleaning the equipment before  disinfecting is recommended     The recommended disinfectants include        E Ethanol  70    E isopropanol  70    E Perform     classic concentrate OXY  14 2 Accutorr 7 Operator s Manual    14 2 General Inspection  Before first use  follow these guidelines when inspecting the equipment     E Make sure that the environment and power supply meet the specifications   E Inspect the equipment and its accessories for mechanical damage     E Inspect all power cords for damage  and make sure that their insulation is in good  condition       Make sure that only specified accessories are applied    E Inspect if the alarm system functions correctly    E   Make sure that the recorder functions correctly and the recorder paper meets the  requirements    E Make sure that the battery meets the performance requirements    m    Make sure that the monitor is in good working condition    In case of any damage or abnormality  do not use the equipment  Contact the hospital s  biomedical engineers or your service personnel immediately     14 3 Maintenance and Testing Schedule  The following maintenance and tests  except for visual inspection  power on test   touchscreen calibration  and battery check  should be carried out by the service  personnel only  Ensure the monitor is safety and performance tested by qualified service  personnel before initial installation  after repair or upgrade or during regularly scheduled  maintenance  Contact your service personnel if 
60. csseccsssecesseecssueccsssecssnecssueecesseessaees 10 2  10 3 2 Taking a Temperature in Predictive MOCE      sssssssscsssccsseesssssecssecesseecsseeseneeeesnees 10 3  10 3 3 Taking a Temperature in Monitor MOde       eesssssscssescssesccsssccsssecesseecesntecsssecesneecesnees 10 4  10 4 Disinfecting Temperature Probe      cscscssssssssssssssssssssessssssssesssssssessssnsesssssnsesssssnseessssnssessssnseessees 10 5    TT RECOPGING siccssisssicesiccsiicacssessecssccesscessesesstessaescecesacessscetsccscdoessceoseadesestssessiodsscsssssessesssdsess 11 1    TUT Usinga Recorde oansein a i aiia  VAD LOA Gig  PaPe a Eaa A RE E E aE eS  113 Settingth   Recorder asiain ana Ea a a ahei  11 4 Starting and Stopping Recordings    s  11 5 T  016 p PAEA A EEE POAR E E PEE OAE EAE E EOE OEE E  11 5 1 Real time Recording        ssssescsssseccsssccsseecesuteccsseccsssccesseecesnecesssecesssccssuscesssecesaseessuecessseessases  11 5 2 Graphic Trend Re  ording       sssscssssecssssccsseccsseccsssecesseccssseessssecesnecessusccsssecssuseesssecesneessnees  11 5 3 CONTINUOUS Trends RECOFING        ssssccssseccssecccseeccsseccesseecsssecessecessuecesseecesueecsssecesneeeesnees 11 3  11 5 4 Spot Check Trends Recording         ssscsssseccssssccsssccsseccesssecsseccessccesssecesseecesneeessseeesaeeessnees 11 4  11 6 REMOVING a Paper JAM     rsecssescossecneecnssccsscesecenscesssesseesseserssceneesnesseneesnesecnseeneeensssessceneersseeneesse  11 7 Cleaning the Recorder Printhead          12 Other FUNCC
61. de for monitoring patient vital signs     NOTE      In Monitor mode  physiological and technical alarms  and prompt messages are  supported       In Monitor mode  NIBP continuous and auto measuring is enabled        Accutorr 7 Operator s Manual 2 11    2 6 2 Spot Check Mode  The Spot Check mode is intended for on spot measurement in a short time period  When  Spot Check mode is On   Spot Check  displays in the Network and USB connection area at  the top of the screen     To enter Spot Check mode  select  Main     Maintenance gt  gt      gt  User Settings gt  gt      Enter  required password    gt Set  Spot Check  to  On      NOTE      In Spot Check mode  technical alarms and prompt messages are supported  but  no physiological alarms       In Spot Check mode  NIBP continuous and auto measuring is disabled     Monitor Mode vs  Spot Check Mode  The Monitor mode and Spot Check modes have all the features in common except the                                     following    Functions Monitor Mode Spot Check Mode  Configure and use NIBP Interval Yes No  Configure and use NIBP Clock Yes No  Configure and use NIBP Simultaneous Yes No  Use STAT NIBP Yes No  Configure and use Sat Seconds  Nellcor  Yes No  Access  Custom Program  tab  NIBP  Yes No  Access  Alarm Setup  tab Yes No  Access  Continuous Trends  tab Yes No  Access  Graphic Trends  tab Yes No  Connect to the CMS Yes No                      2 12 Accutorr 7 Operator s Manual    2 6 3 Standby Mode    In Standby mode  the patien
62. e Reminder Tone   6 10 LatchiingG Alariniss zsscssssscsictececssteaccetess tecvacteceseetaaeanmniarsatech a epn a aniei ia   6 11 Actions for Alarm OCCUrrenCe weeesesssssssssseessccceesssseesscsceccssssssesecceeeesssusssecseceensnuesesceeesennseeteeeseeeets 6 10       6 12 Nurse Call       7 Monitoring SpO2z            ccccccsssssccscsssscscsssscssesssssscsvesssccsessscssssossscssessssssssesssessossnsssessssees 7 1  TW OVER WIGW anenai EER AAAA NASEN  TEE TAEAE EEE AEE ETE AE EEEE EAT  73 Identifying  SOs MOC Ue  nnn onena ENEE A REANA  7 4 Applying the SCNsor         sssessccsssessscsssssseccssssecssssesccsssseecsssssssecssssseesesssseceesssssesessnsecseennseeseesssesesesnsessee  7 5 CHANGING SPOs SettiNgS isise an is aiina   7 5 1 ACCESSING SPOMENU nissin inani eisein riesia  7 5 2 Adjusting the Desat Alarim       scssssscsssseccsseesssssccsseecsssscesseccssusccsssecesssecesueecssneessneecesneeessnes  73 3 Setting SPOaSENSIIVITY sssini aa a a  7 5 4 Changing Averaging Time        sssscsssssccssseccsseccssscccsssecssssccssseccsssscessecessueessnsecesseecsssecesseeessass  7 5 5 Monitoring SpO2 and NIBP Simultaneously     7 5 6 Sat Seconds Alarm Management       csccssssccssssccsssecssecesseccsssecessecssseecssusceessecssasecsaseessnees  7 5 7 Changing the Speed of Pleth Wave      sssssscsssssssssssssessssssessssssssesssssnsessssnecssssnsesssssnseesses  7 5 8 Setting the Alarm Level for SpO2 Sensor Off Alarm w n    csssssssssssssceccsssseesesceeecsnsneeeeees 7 6  7 6 Measureme
63. e changed configuration as a user configuration     C 1 Parameter Configuration                               C 1 1 NIBP  Name Default Saved at Power Affected by  Config Failure Defaults  Alarm On Off On Yes Yes  Alarm Priority Med Yes Yes  Interval Manual Yes Yes  Clock On Yes Yes  NIBP End Tone 0 Yes Yes  Program Program 1 Yes Yes  Adult 80  Cuff Press   mmHg  in Pediatric 60 Yes Yes    VeniPuncture       Neonate 40       Adult 160       Initial Pressure Pediatric 140 No No       Neonate 90                                        Adult 160  NIBP Sys High Pediatric 120 Yes Yes   mmHg    Neonate 90   Adult 90  NIBE SYS  LOW Pediatric 70 Yes Yes   mmHg    Neonate 40   Adult 110  NIBP Mean High Pediatric 90 Yes Yes   mmHg    Neonate 70   Adult 60  NIBP Meai  Low Pediatric 50 Yes Yes        mmHg                 Neonate 25          Accutorr 7 Operator s Manual                                                                                        hone Default Saved at Power Affected by  Config Failure Defaults  Adult 90  NIBP Dia High Pediatric  70 Yes Yes   mmHg   Neonate 60  Adult 50  NIBE Dlg LOW  Pediatric   40 Yes Yes   mmHg   Neonate 20  C 1 2 Spo   Name Default Config Saved at Power Affected by  Failure Defaults  Alarm On Off On  Alarm Priority Med  Adult pediatric 100  SpO2 High eee tet  Neonate  95  SpO2 Low 90  Desat Low 80 Yes Yes  Sat Seconds 0s   Nellcor   NIBP Simultaneous  Off  Sweep 25 mm sec  Beat Volume 2  Monitor mode  Low  SpOz2 Sensor Off Lev  Yes No  Spot Chec
64. e equipment will  not be used for a long time     NOTE       Itis recommended to always install a fully charged battery in the monitor to  ensure normal monitoring in case of accidental power failure     The on screen battery symbol indicates the battery status as follows     a i  Indicates that the battery is fully charged  The solid portion represents the  current charge level of the battery in proportion to its maximum charge level     a Hi Indicates that the battery has a low charge level and needs to be charged     a i  Indicates that the battery is almost depleted and needs to be charged  immediately  Otherwise  the monitor automatically shuts down     a e lt   Indicates that no battery is installed     The capacity of the battery is limited  When the battery is low  the technical alarm area  displays  Low Battery   the alarm lamp flashes  and monitor produces an alarm sound     If the battery is depleted  the battery symbol on the screen flashes  the technical alarm  area displays  Battery Depleted   the alarm lamp flashes  and the monitor produces  alarm sound  Connect the equipment to AC mains to run the equipment and charge the  battery  Otherwise the equipment will shut down        Accutorr 7 Operator s Manual 13 1    13 2 Charging a Battery  The battery is charged whenever the monitor is connected to an AC power source  regardless of whether or not the monitor is turned on  When the battery is charging  the  AC power indicator and battery indicator are both On  If
65. e necessary safety of patients and all devices concerned will  not be impaired by the proposed combination     Put the equipment in a location where you can easily see the screen  access the  operating controls  and disconnect the equipment from AC power              Accutorr 7 Operator s Manual 3 1    3 1 1 Unpacking and Checking    Before unpacking  examine the packing case carefully for signs of damage  If any damage  is detected  contact the carrier or Mindray     If the packing case is intact  open the package and remove the equipment and  accessories carefully  Check all materials against the packing list and check for any  mechanical damage  Contact Mindray in case of any problem           Ti TE         When disposing of the packaging material  be sure to observe the applicable  waste control regulations and keep it out of children   s reach        Before use  please verify whether the packages are intact  especially the    packages of single use accessories  In case of any damage  do not apply it to  patients           NOTE      Save the packing case and packaging material as they can be used if the  equipment must be reshipped     3 1 2 Environmental Requirements    The operating environment of the equipment must meet the requirements specified in  this manual     The equipment operating environment should be reasonably free from noise  vibration   dust  corrosive  flammable and explosive substances  To maintain good ventilation  the  equipment should be at least 2 i
66. e of illness of a patient based on a calculated score  so that the doctor can take a    preventive measure according to the provided protocol in MEWS     The MEWS system is only applicable to adult patients           ne      The MEWS score and protocol are for reference only and cannot be directly used    for diagnostic interpretation           Select the  MEWS  QuickKey to access the MEWS scoring screen     MEWS Scoring    x       PR 80   RESP 18   NIBP mmHg  166 4 100  Temp  C  38 0   AVPU Alert    Calculate    MEWS Clear  MEWS Protocol  gt  gt     Print       Review    In the menu  you can     E Start MEWS auto scoring by entering parameter values  selecting consciousness    status of a patient  and then selecting  Calculate      Select  Review  to view all the MEWS scores     Select  Clear  to remove the parameter values and score from current screen   Select  MEWS Protocol gt  gt   to view actions to be taken according to MEWS protocol   Select  Print  to print the parameter values and calculated score        Accutorr 7 Operator s Manual    12 1    You can define your own MEWS protocol by using the MEWS Protocol Customize Tool that  came with the monitor  CD P N 047 014049 00      1  Select  MEWS  QuickKey    MEWS Protocol  gt  gt      Customize    Enter the required  password  and then the MEWS Protocol screen displays     2  Select  Import  to import your own protocol     In the  MEWS Protocol  screen  you can also export the current protocol or restore to the  default
67. e user of  variations on Ux  for 5 cycles Ur  for 5 cycles our product requires  power supply continued operation during  input lines IEC 70   Ur  30  dip in   70  Ur 30 dipin   Power mains interruptions   61000 4 11 Ux  for 25 cycles Ux  for 25 cycles it is recommended that our  product be powered from   lt 5 Ur  gt 95 dip    lt 5 Ur  gt 95 dip   an uninterruptible power  in Ur  for 5 s in Ux  for 5 s supply or a battery   Power 3 A m 3 A m Power frequency magnetic  frequency fields should be at levels   50 60 HZ  characteristic of a typical  magnetic field location in a typical  IEC 61000 4 8 commercial or hospital          environment              Note  Ur is the AC mains voltage prior to application of the test level           B 2    Accutorr 7 Operator s Manual          Guidance and Declaration   Electromagnetic Immunity       The device is intended for use in the specified electromagnetic environment  The customer or  the user of the device should assure that it is used in such an environment as described below              Immunity 1EC60601 Compliance Electromagnetic environment    test test level level guidance  Conduced RF 3 Vrms 3Vrms Portable and mobile RF  IEC61000 4 6 150 kHz to 80 communications equipment should be  MHz used no closer to any part of the system   including cables  than the  recommended separation distance  calculated from the equation  appropriate for the frequency of the  transmitter  Recommended separation  distances   d  1 2VP  Radiated RF 3V 
68. easurement  Failure to use a  probe cover can cause inaccurate temperature readings  and patient  cross contamination     Check the disposable probe cover for damage before using  Never use any  probe cover that show signs of damage or contamination for temperature  measurement     Be careful to avoid damaging the temperature probe  Place the temperature  probe in the probe well when not in use    Prior to taking a temperature  instruct the patient not to bite down on the  probe  as patient injury and damage to the probe may result    Ensure that probe covers are disposed of according to local regulations or  hospital s requirements    Accuracy verification of the temperature module is required every two years or    according to your hospital s policy  Please contact Mindray Technical Support  department if accuracy verification is needed              Accutorr 7 Operator s Manual 10 1    NOTE       Patient actions may interfere with oral temperature measurements  Ingesting  hot or cold liquids  eating food  chewing gum  brushing teeth  smoking  or  performing strenuous activities may affect temperature readings for up to 20  minutes after ending the activity       Inthe axillary mode  the probe should directly contact the patient   s skin   Measuring through patient s clothes or long term exposure of patient   s armpit  to the air may result in inaccurate temperature readings       Choose appropriate probe according to measurement site  Using the incorrect  probe may cause pa
69. ecorder may be unable to print  or poor print quality  may result       Never pull the recorder paper with force when a recording is in process   Otherwise  it may cause damage to the recorder        Do not leave the recorder door open unless you are replacing the recorder  paper or correcting an error     11 3 Setting the Recorder    Select  Main      General Setup  gt  gt      Print Setup  gt  gt   to access the  Print Setup  menu  and select the items as you want     11 4 Starting and Stopping Recordings    To manually start a recording  you can either     m Press E  hardkey on the recorder     Select  Record  key in graphic or tabular trend     The monitor will automatically start recording when the  Save  QuickKey is selected to  save the manual input patient data if  Automatically Record on Manual Save  is enabled  from the  User Settings gt  gt   menu     H To manually stop the current recording  you select E  hardkey        11 2 Accutorr 7 Operator s Manual    Recordings stop automatically when    m  A recording is completed    E   The recorder runs out of paper    m    When the recorder has an alarm condition     When a recording is stopped  the following markers will be added     E Automatically stopped recording  print two columns of      at the end of the printout       Manually or abnormally stopped recording  print one column of      at the end of the  printout     11 5 Reports    11 5 1 Real time Recording  Real time recording strip includes recording time  par
70. ed    NIBP measurement   VeniPuncture   Yes Low System deflates the   No operation is    timeout             cuff after a certain  time        required           Accutorr 7 Operator s Manual       D 2 2 SpO  Alarm Messages                                                             Alarm Clearable  Alarm Level Cause Solution  Message  Yes No   SpO2 Sensor   Yes Med  Monitor   The SpO2 sensor   Check the sensor  Off mode  has become   application site and   Masimo  Off   Spot   detached from the   the sensor type  and  Nellcor  Check mode  patient or the   make sure if the  module  or there is   sensor is damaged   SpO2 Sensor   No Low a fault with the   Reconnect the sensor  Fault SpO2 sensor  or an   or use a new sensor    Masimo  unspecified SpO2  has b  SpO2 No   Yes Low rte aoe  used   Sensor   Masimo   Nellcor   SpO2 No Low  Unknown  Sensor   Masimo   SpO2 Too   No Low There is too much   Move the sensor to a  Much Light light on the SpOz   place with lower level   Masimo  sensor  of ambient light or  cover the sensor to  minimize the ambient  light   SpO2 No   No Low SpO2 sensor failed   Move the sensor to a  Pulse to obtain pulse   site with better   Nellcor  signal  perfusion   SpO2 Comm   No High An error occurred   Restart the monitor  If  Stop to the SpO2   the error remains    Masimo  module  or there is   contact your service  Nellcor  a problem with   personnel   the  SpO2 Comm   Yes High communications  Abnormal between the   Masimo  module and the  Nellcor  mo
71. en a system related message is presented  the second line of this area will display  the system prompt message for 30 seconds  The patient information in this area will  be covered temporarily     2  Alarm Information Area    There are three sections in this area  The left side of this area shows the technical  alarm message or prompt message  the middle area shows the physiological alarm  message  and the right side of this area shows the alarm symbol  When there are  multiple messages  they will scroll     IN  indicates alarms are paused     indicates alarms are reset      a  indicates alarm sounds are turned off    3  Parameter and waveform area  displays parameters and waveforms   Manual input area  manually input physiological related values  This area does not  display by default  Refer to 4 2 Manually Input Patient Data for additional  information     5  Menu QuickKeys      Main  Accesses the  Main  screen to configure the monitor  or quickly returns to  the main screen        Accutorr 7 Operator s Manual 2 9              MEWS  Accesses the  MEWS Scoring  screen to evaluate a patient s condition   Refer to 12 1 MEWS System    Patient List  Accesses the  Local Patient List  and  ADT Database  screen to  admit a patient stored in the monitor or ADT database  Refer to 4 1 3 Admitting  a Patient from  Patient List     Review  Displays the spot check trends  continuous trends and graphic trends   Refer to 4 4 Reviewing Patient Data    Save  Accesses the  Results  screen to 
72. ent by Barcode Scanner    1     Connect the barcode scanner to the USB connector on the monitor     2  Aim the barcode scanner at the barcode    3  Select  Ok  on the  Patient Demographics  menu to admit the patient    NOTE       If the patient demographics are obtained from barcode scanner  the patient    demographics cannot be changed     4 1 3 Admitting a Patient from  Patient List     Admit a Patient from  Local Patient List      1   2     Press  Patient List  QuickKey  and then  Local Patient List  screen displays     Select the page key  for example          and then use the up or down arrow  beside the key to turn to another page  or select  Scroll  and then use the up or  down arrow beside the key to select the desired patient     Select  Admit  to access  Patient Demographics  menu   Select  Ok   If necessary  you can modify patient information and then select  Ok      In  Local Patient List   you also can     Select  Add New   and then  Patient Demographics  menu displays  Input patient  information and then select  Ok  to admit the patient     Select  Delete  to remove currently selected patient from the monitor   Select  Delete All  to remove all the patients from the monitor        Accutorr 7 Operator s Manual    NOTE    When a patient s demographics are deleted  the corresponding patient data is  also deleted from the montior     You cannot delete the patient currently being monitored     Admitting a Patient from  ADT Database   When the monitor is connected to 
73. enu by selecting its corresponding  parameter area or waveform area        QuickKeys  QuickKeys are configurable graphical keys  located at the bottom of  the main screen     Hardkeys  A hardkey is a physical key on a monitoring device  such as the Alarm Reset    hardkey   and admit patient hardkey    on the front panel     Pop Up Keys  Pop up keys are task related menu keys that appear automatically on  the monitor screen when required  For example  the confirm pop up key appears  only when you need to confirm a change           Accutorr 7 Operator s Manual 3 5    3 4 2 Using the Knob  Using the knob on the front panel of the monitor can do the following operations     E Rotate the knob clockwise or counterclockwise to move the cursor     E Press the knob to select one item  such as accessing a menu or confirming the  selection    3 4 3 Using the Touchscreen  Select screen items by pressing them directly on the monitor s screen     You can disable touchscreen operation by pressing and holding the  Main  QuickKey for 3    seconds  Then a padlock symbol A on the  Main  QuickKey is displayed     When the screen is locked  you can enable the touchscreen operation by pressing and  holding the  Main  Quickkey for 3 seconds     3 5 Changing General Settings  This section covers only general settings such as language  brightness  date and time  etc   Measurement settings and other settings can be referred to in the respective sections     3 5 1 Setting up a Monitor    To install a m
74. equipment requires the prior coordination with a frequency coordinator  designated by the FCC for the Wireless Medical Telemetry Service     This Wi Fi device complies with Canadian ICES 001  Cet appareil ISM est conforme a la  norme NMB 001 du Canada     FCC and Industry Canada Radio Compliance of the device configured with WM1010BGN  wireless module  This device complies with Part 15 of the FCC Rules and RSS 210 of  Industry Canada  Operation is subject to the following two conditions   1  this device may  not cause harmful interference  and  2  this device must accept any interference received   including interference that may cause undesired operation  Any changes or modifications  to this equipment not expressly approved by Mindray may cause harmful radio frequency  interference and void your authority to operate this equipment     Federal Communications Commission  FCC  Statement   The wireless module has been tested and found to comply with the limits for a class B  digital device  pursuant to Part 15 of the FCC rules  These limits are designed to provide  reasonable protection against harmful interference in a residential installation  This  equipment generates  uses and can radiate radio frequency energy and  if not installed  and used in accordance with the instructions  may cause harmful interference to radio  communications  However  there is no guarantee that interference will not occur ina  particular installation  If this equipment does cause harmful interferen
75. essure  it  measures cuff pressure oscillation amplitudes  Oscillations are caused by blood pressure  pulses against the cuff  The oscillation with the greatest amplitude is the mean pressure   This is the most accurate parameter measured by the oscillometric method  Once the  mean pressure is determined  the systolic and diastolic pressures are calculated with  reference to the mean     The auscultation determines systolic and diastolic pressures and calculates the mean  pressure  The oscillometric method measures the mean pressure and determines the  systolic and diastolic pressures     As specified by IEC 80601 2 30  NIBP measurements can be performed during  electro surgery and discharge of defibrillator     NIBP diagnostic significance must be decided by the clinician who performs the  measurement     Blood pressure measurements determined with this device are equivalent to those  obtained by a trained observer using the cuff stethoscope auscultatory method or an  intra arterial blood pressure measurement device  within the limits prescribed by the  American National Standard  Manual  electronic  automated sohygmomanometers  or  standards of IEC80601 2 30  EN1060 1  EN1060 3  EN1060 4 and SP10        Accutorr 7 Operator s Manual 9 1    9 2 Safety          WARNING       A    During NIBP measurement  the inflated cuff applies pressure on the application  site  The clinician determines if NIBP measurement is suitable for the patient   Be sure to select the correct patient c
76. ety labels and data plates on the equipment are clearly legible        Accutorr 7 Operator s Manual 14 5    14 6 Power on Test  This test is to verify that the equipment can power up correctly  This test is passed if the    equipment starts up by following this procedure     1  Insert the battery in the battery compartment  and connect the equipment to the AC  mains  The AC mains indicator and battery indicator light up     2  Press the      button on the front panel to turn on the equipment  The work status  indicator lights up inside the Power button     3  The screen lights up   The main interface is displayed  Now the equipment is correctly started     14 7 Battery Check    Refer to 13 5 2 Checking a Battery for battery check instructions     14 8 Calibrating the Touchscreen    Recalibrate the touchscreen whenever the touch interface becomes difficult to maneuver     1  Select  Main     Maintenance gt  gt      Touchscreen Calibration  The symbol  will appear at different positions of the screen     2  Select  in turn the central point of the symbol  After the calibration is  completed  the message  Screen Calibration Completed    is displayed     3  Select  Ok  to confirm the completion of the calibration     14 9 Formatting the Storage Card    The monitor is configured with an SD card for saving data  To format the storage card     1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt      Enter the required  password    Select  Ok  to access the  User Setti
77. external devices when the  Save  QuickKey is selected and confirmed  and or       Automatically Record On Manual Save   the data saves locally and prints to  the recorder when the  Save  QuickKey is selected and confirmed     Select   to save the configuration and exit the menu     4 4 Reviewing Patient Data  Select  Review  QuickKey  and you can review the trends     In Monitor mode  you can view     Spot check trends    E Continuous trends    E Graphic trends    In Spot Check mode  you can only view Spot Check Trends        Accutorr 7 Operator s Manual 4 5    4 4 1 Spot Check Trends  Select the  Spot Check Trends  tab to access the Spot Check Trends screen     Continuous Trends   Graphic Trends    Filter Temp       Send Successful S    Send Fahso    1 201    Delete       1  Tabular trend 2  Button area  3  Data status 4  Parameter value triggering alarm    In this tab  you can       Select  MRN  or  Name  and then select the desired patient    E Select  Filter  to select parameter trends you want to review    m    Select the page key  for example        _   and then use up or down arrow  beside the key to turn the page      E Select  Edit  to edit patient demographics or manual input data of selected patient    E Select  Delete  to delete the trend data of selected patient    E Select  Print  to print the trend data of selected patient      E Select  Send  to transmit the selected patient   s trend data to the EMR through an  eGateway    NOTE      The trend data can only be
78. false alarms caused by motion artifacts  To set the Sat Seconds limit  select   Sat Seconds  in the  SpO2 Setup  menu and then select the appropriate setting     With Sat Seconds alarm management  high and low alarm limits are set in the same way  as traditional alarm management  A Sat Seconds limit is also set  The Sat Seconds limit  controls the amount of time that SpO2 saturation may be outside the set limits before an  alarm sounds  The method of calculation is as follows  the number of percentage points  that the SpO  saturation falls outside the alarm limit is multiplied by the number of  seconds that it remains outside the limit  This can be stated as the equation     Sat Seconds  Points x Seconds    Only when the Sat Seconds limit is reached  the monitor gives a Sat Seconds alarm  For  example  the figure below demonstrates the alarm response time with a Sat Seconds limit  set at 50 and a low SpO  limit set at 90   In this example  the patient   SpO2 drops to 88    2 points  and remains there for 2 seconds  Then it drops to 86   4 points  for 3 seconds   and then to 84   6 points  for 6 seconds  The resulting Sat Seconds are        7 4    Accutorr 7 Operator s Manual      SpO2 Seconds Sat Seconds       2x 2  4   4x 3  12  6x 6  36  Total Sat Seconds  52    bossrasee    reer retire       0123 4 5 6 7 8 9 10 11  Seconds    After approximately 10 9 seconds  a Sat Second alarm would sound  because the limit of  50 Sat Seconds would have been exceeded     Saturation levels
79. ging potential negative consequences to the environment and  human health  For more detailed information with regard to returning and  recycling this product  please consult the distributor from whom you  purchased it      For system products  this label may be attached to the main unit only           ETL CLASSIFIED       Intertek  3191955             The presence of this label indicates the machine was certified by ETL with  the statement     Conforms to AAMI Std ES 60601 1  IEC Std 60601 1 6  IEC Std 60601 1 8  IEC  Std 60601 2 49  IEC Std 80601 2 30  ISO Std 80601 2 56  ISO Std 80601 2 61     Certified to CSA Std C22 2 NO  60601 1  NO  60601 1 6  NO  60601 1 8  NO   60601 2 49  NO  80601 2 30  NO  80601 2 56  NO  80601 2 61           Accutorr 7 Operator s Manual 1 5       FOR YOUR NOTES       1 6 Accutorr 7 Operator s Manual    2 The Basics    2 1 Intended Use  The monitor is intended for monitoring physiologic parameters  including Pulse Oximetry   SpO2   Pulse Rate  PR   Non Invasive Blood Pressure  NIBP  and Temperature  TEMP   on  adult  pediatric  and neonatal patients in healthcare facilities by physicians or appropriate  medical staff under the direction of physicians     This monitor is intended for use only by clinical professionals or under their guidance  It  must only be used by persons who have received adequate training in its use     2 2 Applied Parts    The applied parts of the monitor are SpO2 sensor and cable  NIBP tubing and cuff  and  Temp probes and ca
80. he  monitoring of patients     Illustrations  All illustrations in this manual serve as examples only  They may not necessarily reflect the  setup or data displayed on your equipment     Manual Conventions     italic text is used to quote the referenced chapters or sections   E  Jis used to enclose screen text   m    gt is used to indicate operational procedures        Accutorr 7 Operator s Manual V    FOR YOUR NOTES       VI Accutorr 7 Operator s Manual    Contents    EET a E E E E E E 1 1    Te Ti Safety INformations tias aT iii 1 1  1 1 1 Warnings  1 1 2 Cautions       11 3 NOTES  kenorin o R E A aE  1 2 Equipment Symbols nip ata a Aai ES 1 4  2 THE  Basics ons ssasccdissccccscccvescncesescsucessntaesescecescseneseocesessensecsesaesessncassenccssesoocesencescssseseesesneteesess 2 1    2 1 Intended Use  2 2 Applied Parts  2 3 Main unit  DS VFO VIEW ios csssdcvoss esscerccaseasisdasabcchssedesse utousdoosh cdeescnttedeaghsatabechadces Aa 2 2  2 3 2 Side View  2 3 3 Rear View  23 4 BOttOM Vi CW sasscccccocscsssscssscasscasssciicoaseccsoncoacoassastasonasoacsanasstageonsescuasaiateasessvacsasasndsdostcoaiteassenseains 2 6  2 4 Main Screen     2 5 Menu  2 6  O Peratiing MOGES wicca sccccsciccsssccccchcsacssczctticspe sss daesadeigacuacscadighcuacseauvscensgacoauscnassconagaletdsegebcietpesaatbvebenus  2 6 1 Monitor Mode  2 6 2 Spot Check Mode  26 3  Standby MOE kaion eeaeee ean ara Ea Ea EEEE Eon ea E Saarai Ninia  2 64 Demo Mod  rna asii eiiie iaaii tiai iaaa                  
81. he Selected CMS at Startup    You can clear the selected CMS when the monitor restarts     1  Select  Main      Maintenance  gt  gt       User Settings  gt  gt      Enter the required  password Select  Ok  to access  User Settings  menu     2  Select  Network  gt  gt       Clear CMS IP at Startup  and then select  On  or  Off         On   If configured with more than one CMS  the monitor will clear the selected  CMS when it restarts 2 minutes after power off     The selected CMS will not be cleared when only one CMS is configured  or the  monitor is restarted within 2 minutes        Off   The monitor retains the selected CMS when it restarts        Accutorr 7 Operator s Manual 3 9    FOR YOUR NOTES       3 10 Accutorr 7 Operator s Manual    4 Patient Data Management    4 1 Admitting a Patient    The monitor displays and stores physiological data in the trends as soon as a patient is  connected  This allows you to monitor a patient that is not admitted yet  However  it is  recommended that you fully admit a patient so that you can clearly identify your patient   on recordings  reports and networking devices     NOTE       Ifthe monitor enters and then exits Standby mode during patient monitoring   you must re admit the patient before continuing monitoring     4 1 1 Admitting a Patient by the Admit Patient Hardkey    1  Press the    hardkey to access the  Patient Demographics  menu     Patient Demographics          MRN 12   Last Name Brown   First Name Jack   Bed No  1   Gender
82. his key for more than 2 seconds to pause or restore alarms   9  NIBP Start Stop key    Press to start or stop NIBP measurements   10  Admit patient key    Press this key to admit a new patient     Press this key to return to the main screen   11  Knob    Rotate the knob clockwise or counterclockwise to move the cursor       Press the knob to select one item  such as accessing a menu or confirming the  selection        Accutorr 7 Operator s Manual 2 3    2 3 2 Side View    ON AMR WN  gt     Handle   Recorder indicator  Start stop recording key  Paper outlet   Recorder door  Connector for NIBP cuff  Recorder door latch  Connector for SpO2 cable                Accutorr 7 Operator s Manual    2 3 3 Rear View                                                                                                                      1  Temperature probe well  2  Connector for temperature probe    3  Network connector  It is a standard RJ45 connector used to communicate with  external devices  such as central monitoring system  eGateway  or for upgrading the  system software     4  Multi function connector  connects to the hospital s nurse call system  or connects  external devices through DIAP protocol     5  USB connector  connects to barcode scanner or USB disk   AC power input  7  Equipotential grounding terminal    NOTE      When using the equipment with other devices  their equipotential grounding  terminals should be connected together to eliminate a possible difference in  ground po
83. ient monitoring data as required     Then press and hold the power on off switch for more than 2 seconds to turn off monitor   There is a prompt message when the system is shutting down     T E    Press and hold the power on off switch for 10 seconds to forcibly shut down the  monitor when it could not be shut down normally or under some special  situations  This may cause loss of patient data     The monitor restores the latest configuration if it restarts within 60 seconds  after a power failure  The monitor restores the default configuration  rather  than the latest configuration  if it restarts 120 seconds after a power failure  The  monitor may load either the latest configuration or the default configuration if  it restarts from 60 120 seconds after a power failure     Power failure may cause data corruption on the SD card  It is recommended to  turn off the monitor according to the normal procedures  Do not directly  unplug the power cord  unless there is a charged battery installed  or remove  the battery before shutting down the monitor     NOTE    To completely disconnect the power supply  unplug the power cord     3 4 Using Key  Knob  Touchscreen  3 4 1 Using Keys    The monitor has three types of keys     Softkey  A softkey is a graphic key on the screen  giving you fast access to certain   menus or functions  The monitor has two types of softkeys      Parameter keys  Each parameter area or waveform area can be seen as a softkey   You can enter a parameter setup m
84. inuous  and auto NIBP measurement     9 5 5 Correcting the Measurement when Cuff is not at Heart Level  Apply the cuff to a limb at the same level as the patient s heart  If the cuff is not at the  heart level  do the following to the displayed value     E Add 0 75 mmHg  0 10 kPa  for each centimetre higher  or  E   Deduct 0 75 mmHg  0 10 kPa  for each centimeter lower        Accutorr 7 Operator s Manual 9 5    9 6 Understanding the NIBP Numerics  The NIBP display generally shows numerics as shown below  Your display may be  configured to look slightly different    2 3 4    oN S    an    NIBP    1          125 80  93     00 02 28 Program 1 A 3 min       10 34  5 6 7    Unit of pressure  mmHg or kPa  Systolic pressure  Diastolic pressure    Mean pressure obtained after the measurement and cuff pressure obtained during  the measurement    Time of last measurement  Time remaining to next measurement  Measurement mode    9 7 Setting NIBP    Select NIBP parameter area to access the  NIBP Setup  menu     In Spot Check mode  you can only set     Initial cuff inflation pressure  NIBP end tone  and  Venous Puncture    In Monitor mode  you can additionally set     Interval   Measurement on clock  Custom program  Alarms   STAT NIBP       Accutorr 7 Operator s Manual    9 7 1 Setting Interval  In  NIBP Setup  menu  you can select  Interval  and set to        Manual   NIBP measurement is started manually   B   1 min    2 min    2 5 min    3 min    5 min    10 min    15 min    20 min    30
85. is complete  otherwise  an inaccurate temperature  reading may result    In Predictive mode  if the probe temperature is high due to the environmental  temperature or other causes  cool the probe and then measure the patient s  temperature    The temperature reading displays continuously until the probe is again  removed from the probe well     10 3 3 Taking a Temperature in Monitor Mode  To measure a temperature in the Monitor mode     1     Make sure that the probe is placed in the probe well        2  Make sure that the temperature measurement type and site settings are correct    3  Remove the probe from the probe well and insert it into a cover in the probe cover  pack  Press the probe handle down firmly until the cover engages with the probe    4  Place the probe at the measurement site and then the monitor measures the site  temperature  Refer to Step 4 in 10 3 2 Taking a Temperature in Predictive Mode for  how to place a probe    5  Withdraw the probe  Firmly press the ejection button on the top of the probe to eject  the probe cover  Replace the probe into the probe well    NOTE     In Monitor mode  record the measured value prior to taking the probe away  from the measurement site  The monitor will automatically stop measuring  temperature after 10 minutes from the start of the measurement    10 4 Accutorr 7 Operator s Manual    10 4 Disinfecting Temperature Probe  The recommended disinfectants include  ethanol 70   isopropanol 70    glutaraldehyde type 2  liquid di
86. is within the marked range  If it is not  use a cuff that fits properly     6  Connect the cuff to the NIBP hose  Avoid compression or restriction of NIBP hose  Air  must pass unrestricted through the tubing        9 4 Accutorr 7 Operator s Manual          A warnine          Sustained cuff pressure due to a kinked hose may interfere with blood flow and  could lead to patient injury           NOTE      The use of the equipment is restricted to one patient at a time     9 5 3 Starting NIBP measurement  Start the NIBP measurement in one of the following ways     E Press the hardkey on the monitor s front panel     E Access  NIBP Setup  menu  and then select  Start NIBP  key to start a manual   programmed or automatic NIBP measurement with preset interval       Access  NIBP Setup  menu  and then select  NIBP STAT  to start a continuous NIBP  measurement           A warnine          Continuous non invasive blood pressure measurements may cause purpura   ischemia and neuropathy in the limb with the cuff  Periodically examine the  limb under the cuff to ensure skin color and integrity  If anything abnormal is  seen move the cuff to another site or stop the non invasive blood pressure  measurements immediately           9 5 4 Stopping NIBP Measurement    E Press the hardkey on the monitor s front panel to stop a manual NIBP  measurement  or a continuous NIBP measurement     E Access  NIBP Setup  menu and then select  Stop All  to stop all the NIBP  measurement  including manual  cont
87. issue throughout the measurement         When taking an axillary temperature  lift the patient   s arm to expose the entire  armpit  Apply the probe as high as possible in the armpit  Check that the probe  tip is completely surrounded by the axillary tissue  Lower the patient s arm so  that it is tightly placed at the patient   s side  Keep the patient s arm and the  probe in place throughout the measurement       When taking a rectal temperature  separate the patient s buttocks with one  hand  and gently glide the probe 0 6 inch  1 5 cm  inside the rectum with the  other hand  For pediatric patients  the depth of insertion is less  Tilt the probe so  that it always contacts with patient s tissue  Lubricant can be used in rectal  mode     The monitor sounds a beep when the temperature measurement is complete     Withdraw the probe  Firmly press the ejection button on the top of the probe to eject  the probe cover  Replace the probe into the probe well     In Predictive mode  the monitor automatically enters Monitor mode in the following  cases          Accurate temperature is not reached   m  Neither measurement is taken nor is the probe replaced in the probe well in 60    seconds after the probe is withdrawn from the well     The temperature type automatically changes to Predictive mode when the probe returns  to the probe well        Accutorr 7 Operator s Manual 10 3    NOTE    In Predictive mode  place the temperature probe at the measurement site as  soon as probe warmup 
88. k mode  Off  Sensitivity  Masimo    Normal  Averaging  Masimo    8s                   Accutorr 7 Operator s Manual          C 1 3 PR       Saved at Power    Affected by                                                                                              Name Default Config Failure Defaults  Alarm On Off On  Alarm Priority Med  Adult 120  PR ree  High Pediatric 160  Neonate 200  Yes Yes  Adult 50  PR Pediatric 75  Low  Neonate 100  PR Source SpO2  Beat Volume 2  C 1 4 Temp  Name Default Config Saved at Power Affected by  Failure Defaults  Alarm On Off off Yes Yes  Alarm Priority Med Yes Yes  Temp High 38 0 Yes Yes  Temp Low 35 0 Yes Yes  Temp Type Predictive No No  Oral Axillary probe   Oral for adult and pediatric  Temp Position   No No  Axillary for neonate  Rectal probe  Rectal  Accutorr 7 Operator s Manual C 3          C 2 General Configuration                                                                                                 C 2 1 Alarm  7 Saved at Power Affected by  Name Default Config Failure Defaults  Latching Alarm Off  Minimum Alarm 2  Volume  High Alarm Interval  s    10 s  Med Alarm Interval  s    20 s Yes No  Low Alarm Interval  s    20s  Reminder Tone On  Alarm Light on Alarm   On  Reset  C 2 2 Review  Name Default Saved at Power Affected by  Config Failure Defaults  Name MRN button MRN  Spot Check   Option for All  Trends Name MRN button  Filter All  Name MRN button MRN  Yes No  Continuous   Option for Current patient  Trends Name MRN button
89. l of the equipment   please contact Mindray     Magnetic and electrical fields are capable of interfering with the proper  performance of the equipment  For this reason  make sure that all external  devices operated in the vicinity of the equipment comply with the relevant EMC  requirements  Mobile phones and X ray equipment are a possible source of  interference as they may emit higher levels of electromagnetic radiation   Before connecting the equipment to the power line  check that the voltage and  frequency ratings of the power line are compatible with those indicated on the  equipment    s label or in this manual    Always install or carry the equipment properly to avoid damage caused by a  drop  impact  strong vibration or other mechanical force     If you spill liquid on the equipment or accessories  contact Mindray or your  service personnel        Accutorr 7 Operator s Manual 1 3    1 1 3 Notes    NOTE      Keep this manual in the vicinity of the equipment so that it can be easily located  when needed       The software was developed in compliance with IEC60601 1 4  The possibility  of hazards arising from software errors is minimized      This manual describes all the equipment features and options  Your equipment  may not have all of them       During normal use  the operator is expected to face the front of the equipment      Put the equipment in a location where you can easily view and operate the  equipment      The equipment uses a mains plug asa means of isola
90. les in a way to prevent a  stumbling hazard  Wrap and secure excess cabling to reduce risk of  entanglement by patients or personnel     When disposing of the packaging material  be sure to observe the applicable  waste control regulations and keep it out of children   s reach    When no battery is installed  make sure that the power supply is continuous  A  power interruption will result in patient data loss    Put the equipment in a location where you can easily see the screen  access the  operating controls  and disconnect the equipment from AC power    The equipment uses a mains plug as isolation means to the mains power supply   Please do not position the equipment in a place difficult to operate the mains  plug    The equipment is not intended to be used within the magnetic resonance  MR   environment              1 2    Accutorr 7 Operator s Manual    1 1 2 Cautions    kamion    Only use parts and accessories specified in this manual     Remove the battery before shipping the monitor or if it will not be used for an  extended period of time     Carefully route patient cabling to reduce the possibility of patient  entanglement     Disposable accessories are not designed to be reused  Reuse may cause a risk of  contamination and affect the measurement accuracy     At the end of its service life  the equipment  and its accessories  must be  disposed of in compliance with the guidelines regulating the disposal of such  products  If you have any questions concerning disposa
91. limit limit limit limit range                                  PR  PRx0 8  or 40   PR 30  or 90   PRX1 25  or 240   PR 40  or  Adult Pediatric   bpm  whichever  bpm  bpm  whichever  200 bpm   35 to 240 bpm  is greater whichever is jis smaller whichever is Neanate   greater smaller i  55 to 225 bpm  NIBP S  SYSx0 68  10   SYS 15  or 45  SYSx0 86  38   SYS 15  or  Adult   mmHg mmHg    mmHg 105 mmHg   45 to 270 mmHg  Vesta Is whichever is  pediatric   greater  smaller 45 to 185 mmHg  Neonate   35 to 115 mmHg  6 6 Accutorr 7 Operator s Manual          Parameter    Low alarm Limit    High alarm Limit    Auto set alarm       Adult Pediatric    Neonate    Adult Pediatric    Neonate    limit             Range  NIBP D  Diax0 68  6       Dia 15  or 20  Diax0 86  32   Diat 15  or  Adult   mmHg mmHg  mmHg 80 mmHg   25 to225 mmHg  whichever is whichever is  Pediatric   ti  grea  er smaller 25 to 150 mmHg  Neonate   20 to 90 mmHg  NIBP M  Meanx0 68  8   Mean 15  or   Meanx0 86     Mean  15   Adult   mmHg 35mmHg   35 mmHg jor 95 30 to 245 mmHg  whichever is mmHg  Pediatric   t      greater whichever is 30 to 180 mmHg  smaller  Neonate   25 to 105 mmHg  Temp  T   0 5    C  T   0 5   C  I T 0 5    C  T 0 5   C  Same as the  measurement  range                         6 6 Pausing Alarms    You can temporarily disable alarm sound of all alarms by pressing and holding    hardkey on the panel for above 2 seconds     When alarms are paused     Alarm Information area        The alarms pause symbol A
92. m 3V m Recommended separation distances   IEC61000 4 3 80MHz to 80 MHz   800 MHz  2 5GHz             d  1 2 4P    800MHz 2 5GHz  d 2 3VP    Where  P is the maximum output power  rating of the transmitter in watts  W   according to the transmitter    manufacturer and d is the  recommended separation distance in  meters  m       Field strengths from fixed RF  transmitters  as determined by an  electromagnetic site survey     should be  less than the compliance level in each  frequency range       Interference may occur in the vicinity of  equipment marked with the following    Kad     symbol           Note 1  At 80 MHz to 800 MHz  the separation distance for the higher frequency range applies   Note 2  These guidelines may not apply in all situations  Electromagnetic propagation is  affected by absorption and reflection from structures  objects and people    Note 3  The device that intentionally receives RF electromagnetic energy at the exclusion band   2395 825MHz 2487 645MHz  is exempt from the essential performance requirements  but    remains safe           Accutorr 7 Operator s Manual    B 3             a Field strengths from fixed transmitters  such as base stations for radio  cellular cordless   telephones and land mobile radios  amateur radio  AM and FM radio broadcast and TV  broadcast cannot be predicted theoretically with accuracy  To assess the electromagnetic  environment due to fixed RF transmitters  an electromagnetic site survey should be  considered  If the mea
93. manual is believed to be correct  Mindray is not liable for  errors contained herein nor for incidental or consequential damages in connection with  the furnishing  performance  or use of this manual     Mindray is responsible for the effects on safety  reliability and performance of this product    only if    E allinstallation operations  expansions  changes  modifications and repairs of this  product are conducted by Mindray authorized personnel     m  the electrical installation of the relevant room complies with the applicable national  and local requirements     E the product is used in accordance with the instructions for use           ee         Only skilled trained clinical professionals should operate this equipment        Itis important for the hospital or organization that uses this equipment to  perform a reasonable service maintenance plan  Neglect of this may result in  machine breakdown or personal injury              Accutorr 7 Operator s Manual    Warranty    Mindray warrants that components within its products will be free from defects in  workmanship and materials for a period of three years from the date of purchase except  that disposable or one time use products are warranted to be free from defects in  workmanship and materials up to a date one year from the date of purchase or the date of  first use  whichever is sooner     This warranty does not cover consumable items such as  but not limited to  batteries   external cables  and sensors     Mindray sh
94. manually save patient data  Refer to 4 3  Manually Save Patient Data     Network and USB connection area    Display CMS information or Spot Check mode  For details for how to configure CMS   refer to 12 2 7 Central Monitoring System Setup     Indicate the connection of network or USB to this monitor                indicates monitor is successfully connected to a wired network   indicates the wireless function  optional  is working    indicates monitor has failed to connect a wired network   indicates the wireless function  optional  is not working     indicates a USB drive is inserted     Battery status  indicates the status of the battery  For details  refer to Chapter 13  Battery     System time  Tabular trend area  This area displays only in Trend screen mode   NIBP list area  This area displays only in NIBP List screen mode        Accutorr 7 Operator s Manual    2 5 Menu    3  General Setup gt  gt     Alarm Setup  gt  gt     Transfer Data to USB       Load Configuration gt  gt     2 Maintenance  gt  gt           A menu in this monitor is usually composed of   1  Heading  provide a title or description for the current menu     2  Main body  displays options  buttons  prompt messages  etc  A menu button  followed by     gt  gt     opens a secondary window to reveal more options or information     3  x   select to exit the current menu     2 6 Operating Modes  2 6 1 Monitor Mode    The monitor will automatically enter monitor mode after power on  Monitor mode is a  common mo
95. mediate treatment  may be required        operation  which may compromise a certain  monitoring function  but will not threaten the  patient s life        6 3 Alarm Indicators  When an alarm occurs  the equipment will indicate it to the user through visual or audible  alarm indications     E Alarm lamp    Audible alarm tones    E  E Alarm message  E Flashing numerics    6 3 1 Alarm Lamp  If a technical alarm or physiological alarm occurs  the alarm lamp will flash  The color and  flashing frequency match the alarm level as follows        High level alarms   E Medium level alarms   E Low level alarms     NOTE    the lamp quickly flashes red   the lamp slowly flashes yellow   the lamp lights yellow without flashing       When multiple alarms of different levels occur simultaneously  the alarm lamp  flashes and the alarm tone sounds according to the alarm of the highest level   and the alarm messages are displayed circularly     6 3 2 Audible Alarm Tones    The monitor uses different alarm tone patterns to match the alarm priority        6 2    Accutorr 7 Operator s Manual       The alarm tone is distinct from keystroke tone and pulse tone in frequency     The alarm tones identify the alarm levels as follows     m  High level alarms  triple double beeps triple double beep  E Medium level alarms  triple beep  E Low level alarms  single beep    The interval of alarm tone is configurable  Refer to section 6 5 3 Setting the Interval  between Alarm Sounds     NOTE       When multiple
96. monitor uses the wireless  optional   network to obtain data  and the screen displays the icon of wireless  optional   network       The design  installation and maintenance of the wireless  optional  network   s  distribution should be performed by authorized personnel       Ina wireless  optional  network  the existence of obstacles  such as walls  will  affect data transmission or even cause wireless  optional  network interruption       Anaccess point  AP  supports a maximum of 16 monitors through the wireless   optional  network        Accutorr 7 Operator s Manual 12 3    12 2 3 Setting up the Wireless Network  Optional   A Mindray proprietary wireless network  installed by approved Mindray service personnel   is required to support wireless networking     This proprietary network will have the following capabilities     E Support the 802 11g wireless protocol   M Have a channel bandwidth of 20 MHz   E Support WPA2 PSK security     Provide a signal strength at the monitor of no less than  65 dBm   NOTE     Keep network authentication information  e g  password  safe to protect the  network from being accessed by unauthorized users       Authentication and encryption other than WPA2 PSK may expose sensitive  data or allow malicious settings       The total throughput of all the wireless devices connected to the wireless  network should be less than the effective transmitting capability of the wireless  network  The throughput capacity of a single Accutorr 7 is 700 kbps   
97. n quality changes   Change the application site every four hours  For neonates  or patients with  poor peripheral blood circulation or sensitive skin  inspect the sensor site more  frequently           7 3 Identifying SpO  Module    To id    entify which SpO2 module is incorporated into your monitor  see the company logo    located at the side panel  The color of the cable connector matches the company as  shown below     No  Masimo SpO2 module  a purple connector with the Masimo SET logo 9  ASMO D       WORKS    Nellcor SpO2 module  a grey connector with the Nellcor logo        The SpO2 sensor connectors are mutually exclusive        Accutorr 7 Operator s Manual    7 4 Applying the Sensor          WARNING       A    If the sensor is too tight because the application site is too large or becomes too  large due to edema  excessive pressure for prolonged periods may result in  venous congestion distal from the application site  leading to interstitial edema  and tissue ischemia           NOTE     Place the SpO   sensor so that the light source is against the application site      Check if the sensor is in normal condition before monitoring  Do not use the  SpO  sensor if the package or the sensor is found damaged       Do not apply the sensor ona limb with an intravenous infusion or arterial  catheter in place    1  Select an appropriate sensor according to the module type  patient category and  weight   Clean the application site  For example  remove colored nail polish   Appl
98. n the patient   s oxygen saturation level  Contrarily  the longer the averaging time  is  the slower the equipment responds to changes in the patient   s oxygen saturation level   but the measurement accuracy will be improved  For critically ill patients  selecting  shorter averaging time will help understanding the patient s state     To set the Masimo SpO  averaging time  select  Averaging  in the  SpO2 Setup  menu and  then toggle between  2 4 s    4 6 s    8 s    10 s    12 s    14 s  and  16 s      7 5 5 Monitoring SpO   and NIBP Simultaneously    When monitoring SpO2 and NIBP on the same limb simultaneously  you can switch  NIBP  Simultaneous  on in the  SpO2 Setup  menu to lock the SpO2 alarm status until the NIBP  measurement ends  If you switch  NIBP Simultaneous  off  low perfusion caused by NIBP  measurement may lead to inaccurate SpO2 readings and therefore cause false  physiological alarms     7 5 6 Sat Seconds Alarm Management    With traditional alarm management  high and low alarm limits are set for monitoring  oxygen saturation  During monitoring  as soon as an alarm limit is violated  an audible  alarm immediately sounds  When the patient   SpO  fluctuates near an alarm limit  the  alarm sounds each time the limit is violated  Such frequent alarms can be distracting   Nellcor   s Sat Seconds alarm management technique is used to reduce these nuisance  alarms     The Sat Seconds feature is available with the Nellcor SpO2 module to decrease the  likelihood of 
99. nce    settings to be exported  Then select the  Export  button  A status message will  report completion of the transfer     To import the configuration from the USB drive to the monitor    1  Insert a USB drive into the monitor s external device connector    2  Select  Import Config  gt  gt   in the  Manage Configuration  menu    3  In the  Import Config  menu  select the configurations and the user maintenance    settings to be imported  Then select the  Import  button  A status message will  report completion of the transfer     5 6 Loading a Configuration  You may make changes to some settings during operation  However  these changes or the  pre selected configuration may not be appropriate for the newly admitted patient   Therefore  the monitor allows you to load a desired configuration so as to ensure that all  the settings are appropriate for your patient     To load a configuration     1  Select  Main     Load Configuration  gt  gt    The popup menu shows the existing  configurations on the monitor  Selecting  Config on USB drive  gt  gt   will show the  existing configurations on the USB drive     2  Select a desired configuration   3  Select  Load      The current configuration is shown at the top of the  Load Configuration  menu     5 7 Restoring the Latest Configuration Automatically  During operation  you may make changes to some settings  However  these changes may  not be saved as user configurations  To prevent the changes from being lost in case of a  sudden
100. nches  5 cm  away from surrounding objects     When the equipment is moved from one place to another  condensation may occur as a  result of temperature or humidity difference  In this case  never start the system before  the condensation disappears           i ee         Make sure that the equipment operating environment meets the specifications   Otherwise unexpected consequences  e g  damage to the equipment  could  result           NOTE      The equipment uses a mains plug asa means of isolation to the mains power  supply  Do not position the equipment in a place difficult to access the mains  plug        Accutorr 7 Operator s Manual    3 2 General Operation    Read this operator s manual carefully before using this monitor  Familiarize yourself with    the equipment s function and operation  and observe the warnings and cautions included  in the manual     3 2 1 Connection to AC Power    This monitor can be powered by AC power or battery  Connect the power cord to the AC  input on the back of the monitor  and connect the other end of the power cord to the  power outlet           ne      Always use the accompanying power cord with the monitor        The battery is to be used if the integrity of the protective earth conductor or the  protective earthing system in the installation is in doubt           3 2 2 Using a Battery    This monitor can be equipped with rechargeable lithium  ion battery  If a battery is    installed  the monitor system automatically switches to the b
101. ncture  key   Puncture vein and draw blood sample     4  When the puncture is complete  select  VeniPuncture  key  or press the TD  hardkey on the monitor s front panel to deflate the cuff  The cuff deflates  automatically after a set time if you do not deflate it     During puncture  the NIBP display shows the real time inflation pressure of the cuff        9 8 Accutorr 7 Operator s Manual    destination inflation pressure  and the remaining time in venous puncture mode     NIBP      18  80     00 02 41       9 9 Care and Cleaning of Reusable Cuffs    NOTE      Accuracy of cuff pressure transducers indicators is to be verified at intervals  specified by the manufacturer     9 9 1 Reusable Cuffs with Bladders    Take out the bladder before cleaning and disinfecting the cuff     9 9 1 1 Cleaning  Hand or machine washes the cuff in warm water or with mild detergent  Clean the  bladder with a damp cloth  Air dry the cuff thoroughly after washing     NOTE      Machine washing may shorten the service life of the cuff     9 9 1 2 Disinfection    Disinfect the cuff with a cloth dampened in 70  isopropanol or with ultraviolet light  Also  disinfect the bladder with ultraviolet light     NOTE       Prolonged use of disinfectant may cause discoloration of the cuff     Replace the bladder after cleaning and disinfecting the cuff   1  Place the bladder on the top of the cuff  as shown below    2  Roll the bladder lengthwise and insert it into the large opening        Accutorr 7 Operator 
102. nder Tone  When the alarm volume is set to zero  or alarm is reset  the monitor can issue a periodical  reminder tone  The interval of the reminder tone is 1 minute  You can switch on or off the  reminder tone     1     Select  Main      Maintenance  gt  gt      User Settings  gt  gt     Enter required  password   Select  Ok  to access  User Settings  menu     Select  Alarm Setup  gt  gt   to access  Alarm Setup  menu   Set  Reminder Tone  to  On  or  Off      6 10 Latching Alarms  The latching setting for physiological alarms defines how alarm indicators behave when  you do not acknowledge them     If an alarm is latched  alarm indications remain presented even though alarm  conditions end  except that       The parameter reading and violated alarm limit stop flashing       The time when the alarm is last triggered is displayed after the alarm message     If an alarm is not latched  the alarm indications disappear as soon as the alarm  conditions end     When the alarm system is reset  the latched physiological alarms are cleared     To latch a physiological alarm     1     Select  Main      Maintenance  gt  gt      User Settings  gt  gt      Enter required  password   Select  Ok  and then access  User Settings  menu     Select  Alarm Setup  gt  gt      Set  Alarm Latch  to  High only    High amp Medj    All  or  Off       High only   only high priority alarms are latched               AIl   all alarms are latched  and     Off   no alarm will be latched     Hi amp Med   both
103. nds on the setting of  minimum alarm volume  refer to 6 5 2 Setting the Minimum Alarm Volume   and 10  the maximum volume     The alarm tone is switched off when the volume is set to  0            ee          Do not rely exclusively on the audible alarm system for patient monitoring   Adjustment of alarm volume to a low level may result in a hazard to the patient   Always keep the patient under close surveillance           3 5 5 Adjusting Key Volume  1  Select  Main     General Setup gt  gt       2  Select  Key Volume  and then select the appropriate volume  0 10  in which 0 means  off  and 10 the maximum volume     The monitor can provide a tone according to the settings in key volume when you press  the knob or hard key  or touch the screen     3 5 6 Adjusting the Screen Brightness  1  Select  Main     General Setup gt  gt       2  Select  Brightness  and select the appropriate setting for the screen brightness  10 is  the brightest  and 1 is the dimmest     If the monitor operates on battery power  you can set a dimmer screen to prolong the  battery operating time  When the monitor enters Standby mode  the screen automatically  changes to the dimmest setting        Accutorr 7 Operator s Manual 3 7    3 5 7 Setting Screen    You can set the main screen to one of the three pre defined layouts as desired     To set the screen     1   2     Select  Main     General Setup gt  gt       Select  Display Setup  and set the main screen to different layout  All Parameters  display  T
104. nfiguring Printout                   4 Patient Data Management        essessessseesossosscsssessossoseosesessossosecossosesossessossosesossoseosssessossssee 4 1       4 1 Admitting a Patient  4 1 1 Admitting a Patient by the Admit Patient Hardkey  4 1 2 Admitting a Patient by Barcode Scanner         sssssecsessssesseesssecescsssesescsnnseseesnsesssennsesseeese  4 1 3 Admitting a Patient from  Patient List    4 2 Manually Input Patient Data   4 3 Manually Save Patient Data    sssssssssssssessssessccsssessecssseseccssssseecsssseseessssseeeensnseesssnsecesesnseesesesnsessee   4 4 Reviewing Patient Data  4 4 1 Spot Check Trends  4 4 2 CONTINUOUS TLENS   0     sssssessccsssessecssssssecsnsessecssssssecssssecsecsnsecsessnsecescssseescsssessesssneesssensnsesesssees  4 4 3 Graphic Trends   4 5 Transferring Data from the Monitor to USB Drive       5 Managing Configurations  i    scccscsssseesscscsessccessssnscoasecvenascasoscsnsssadonssnsseesecesesesasvansocesoeses 5 1       5 1 Overview  5 2 Accessing  Manage Configuration  Menu  5 2 1 Setting Default Configuration          ssssccssssseesssssseesssssessssssseessssssessssssssesssssnsesssssusecssssnesses  5 3 Saving Current Settings  5 4 Deleting a Configuration  5 5 Transferring a CONFIQUIATION      cssssssccsssssscssssseesssssssecsssssecsssssesssssusecssssuseessssnsecssssnsecsssssnecsssssseeess  5 6 Loading a Configuration  5 7 Restoring the Latest Configuration Automatically          CFAE A EE E AEE OAE EEE doeseveeacs AE E AA E OR 
105. ngs  gt  gt     Enter the required  password    Select  Ok  to access User Settings menu     Select  DIAP Communication gt  gt   to access  DIAP Communication Setup  menu   Set the baud rate   Select  Ok  to exit the menu     12 2 7 Central Monitoring System Setup    You can set up to 30 CMSs for the monitor     1     2     3   4     Select  Main      Maintenance  gt  gt      User Settings  gt  gt     Enter the required  password   Select  Ok  to access User Settings menu     Select  Network gt  gt       Central Station Setup gt  gt   to access  Central Station Setup   menu     Set the name of the CMS and IP addresses   Select  Ok  to exit the menu     12 2 8 Setting the Multicast Parameters  To set the Multicast parameters     1    Select  Main      Maintenance  gt  gt      User Settings  gt  gt     Enter the required  password    Select  Ok  to access User Settings menu     Select  Network  gt  gt       Multicast Setup  gt  gt     Set  Multicast Addr  and  TTL         Accutorr 7 Operator s Manual 12 5    FOR YOUR NOTES       12 6 Accutorr 7 Operator s Manual    13 Battery    13 1 Overview  The monitor is designed to operate from battery power when AC power is not available  In  case of power failure  the equipment automatically runs from the battery     The battery is to be used if the integrity of the protective earth conductor or the  protective earthing system in the installation is in doubt     A CAUTION      Remove the battery before transporting the equipment or if th
106. ngs  menu     2  Select  Format Storage Card   and then select  Ok  in the pop up dialog     14 10 Modifying Password    1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt      Enter the required  password    Select  Ok  to access the  User Settings  menu     Select  Modify Password gt  gt     Enter new password and then select  Ok         14 6 Accutorr 7 Operator s Manual    15 Accessories    The material that patients will come into contact with has passed the bio compatibility  test and is verified to be in compliance with ISO 10993 1           ince      Only use accessories specified in this chapter  Using other accessories may  cause damage to the equipment or not meet the claimed specifications        Single use accessories are not designed to be reused  Reuse may cause a risk of  contamination and affect the measurement accuracy       Check the accessories and their packages for any sign of damage  Do not use  them if any damage is detected        Dispose of accessories according to your hospital regulations       Use the accessories before the expiration date if their expiration date is  indicated       For more details about the accessories  refer to the instructions for use of  corresponding accessories           15 1 SpO  Accessories    Extension Cable             Module type Remarks Part No   Masimo SpO2 Module 8 pins  2 1m 115 020768 00  Nellcor SpO2 Module 8 pins  2 5m 0010 20 42712          SpO2 SensorsMasimo SpO2 module                         
107. nitor   SpOz2 Init Err   Yes High  SpO2 Board   No Low There isa problem   Do not use the  Fault with the SpO2   module and contact   Masimo  measurement your service  board  personnel   SpO2 Low   No Low The SpO2 signal is   Adjust the sensor  Signal too low or too   application site    Masimo  weak   D 4 Accutorr 7 Operator s Manual                               Alarm Clearable  Alarm Level Cause Solution  Message  Yes No   SpO2 Weak   No Low  Signal   Nellcor   SpO2 No Low The SpO2 signal   Check for any possible  Interference has been   sources of signal noise   Masimo  interfered  around the sensor and  check the patient for  great motion   SpO2 Comm   Yes High An error occurred   Restart the monitor  If  Err  Masimo  to the SpO2   the error remains   Nellcor  module  or there is   contact Mindray  a problem with   service personnel   the  communications  between the  module and the  monitor   SpO2 Limit   No Low The alarm limit of   Contact your service  Err SpO2 is changed   personnel    Masimo  inadvertently   Nellcor   PR Limit Err   No Low The alarm limit of   Contact your service   Masimo  PR is changed   personnel   Nellcor  inadvertently   PR No Low The measured PR   Contact your service  Overrange value exceeds the   Personnel    Masimo  measurement  Nellcor  or range   from NIBP                      means the alarm level is user adjustable        Accutorr 7 Operator s Manual    D 5       D 2 3 Temperature Alarm Messages                                     Alarm
108. nsideration do NOT set the high alarm  limit to 100   which is equivalent to switching the alarm off           6 5 Selecting Alarm Properties    6 5 1 C    hanging the Alarm Volume    Select  Main      General Setup  gt  gt      Alarm Volume      The alarm volume range is between X and10  X is the minimum volume  which depends  on the setting of minimum alarm volume  refer to 6 5 2 Setting the Minimum Alarm  Volume   and 10 is the maximum volume     When alarm volume is set to 0  the alarm sound is turned off and a symbol appears  on the screen     6 5 2 Setting the Minimum Alarm Volume  The minimum alarm volume refers to the minimum value you can set for the alarm  volume  which is not affected by user or factory default configurations     Select  Main      Maintenance  gt  gt      User Settings  gt  gt     Enter the required  password    Select  Ok  to access  User Settings  menu     Select  Alarm Setup  gt  gt   to access  Alarm Setup  menu   Select  Minimum Alarm Volume  and then select the appropriate settings     6 5 3 Setting the Interval between Alarm Sounds    1     Select  Main      Maintenance  gt  gt       User Settings  gt  gt     Enter the required  password   Select  Ok  to access  User Settings  menu     Select  Alarm Setup  gt  gt   to access  Alarm Setup  menu     Select  High Alarm Interval  s     Med Alarm Interval  s   and  Low Alarm Interval   s   in turn and then select the appropriate settings        Accutorr 7 Operator s Manual 6 5          WARNING   
109. nt Limitations        sssessssssssssesssseessssecessecssssecssssessssscesssessssseesssscsssseesssseessseessseesssseess 7 6  7 7 Masimo Afora t ON sisses 7 6  7 8 N  llcor Inf  rmati os ssnaisoncnniinnnasannnnenennaninninnn eines 7 7   8 Monitoring PR cssscciiccssscccsceocescsssesescssosessecssoousssusseessscecesescsesssonssoncscesaevessessscesesesesoousseessenes 8 1  8T OVET VIEW nesini paagi t iTi iesse iaaa EEEIEI Raai TE eoat SESAR Na ESRAS S ieia iie 8 1  8 2 PR SOURCE irn EARNAN AE AEE N AONANE ONENEN E 8 1  83 Pulse TONE rayina a A a a 8 1   9  Monitoring NIBP sesscccccscicsccsccesseacssedsscessetccssecstncseadseacasessseansecasosesesadccoaasectsedecscessaccsceesseees 9 1  SRDAN E AA AEE ETE AEE 9 1  DD Safe a a O E A A i 9 2  9 3 M  asurement Limitations ascsssss cssccsssissesscesvsistessescossncossadcosst enussscuvdsesussseuyancoussnzeisesntbcennsaseboacenbiasees 9 3       9 4 NIBP Measurement Mode       Accutorr 7 Operator s Manual    9 5 Measuring  NIBP eea E E TE ANAE  9 5  1 Preparing the Patient asisssssssccscsscsssccassacsasscasssesosascscsacasadassssscsdetaasvsnsscassscsoaassbasassoaauctasasscaains  9 5 2 Preparing to Measure NIBP    i   9 5 3 Starting NIBP measurement         ssessssssssssessooscercoeccceeeeesessssssssssssssoseoeeeeeceeeeeeeeeeeeeeseessesssssss  9 5 4 Stopping NIBP Measurement           esssescsssscsescsescsseecneecssecsneccnsecssecsscecssccnscesscenscesseeeneeesees 9 5  9 5 5 Correcting the Measurement when Cuff is not at Heart Level         
110. nts  alarm events  trends and patient data  or cause  functional failure  In the case of network disconnection  check the patient and  solve the network problem as soon as possible           12 2 Accutorr 7 Operator s Manual    12 2 2 Network Type and Settings  The equipment supports both wired and wireless  optional  networking  To set the  network type     1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt      Enter the required  password    Select  Ok  to access the  User Settings  menu     2  Select  Network gt  gt       Monitor Network Setup  gt  gt   to access the  Monitor  Network Setup  menu     In the Monitor Network Setup menu  you can   E Set  Network Type  to  LAN  or  WLAN    E Set  Address Type  to  DHCP  or  Manual        if  Address Type  is set to  DHCP   the monitor can automatically acquire  network parameters        if  Address Type  is set to  Manual   you need to manually input the monitor IP  address  subnet mask and gateway address     If your network is WLAN  in the Monitor Network Setup menu  you can   E Set  Network Name  SSID   and  Password    E Select  WLAN Test  gt  gt   to perform wireless  optional  network connection test     NOTE       The option  Network Type  is active only when the monitor is equipped with an  optional Wi Fi module       When the network type is set to LAN  the monitor uses the wired network to  obtain data  and the screen displays the icon of wired network       When the network type is set to WLAN  the 
111. onitor or change the monitor s location  you need to set it as follows     1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt     Enter the required  password   Select  Ok  and then access  User Settings  menu     2  Set up  Monitor Name    Department  and  Bed No     You can set  Changing Bed No   to     E  Unprotected   enables you to change Bed No  from the  Patient Demographics   menu     E  Protected   prevents you from changing Bed No  from the  Patient Demographics   menu     3 5 2 Changing Language  1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt      Enter the required  password Select  Ok  to access  User Settings  menu   Select  Language  and then select the desired language   Restart the monitor     3 5 3 Configuring the Timeout of Clinician ID  You can configure the retention time of a clinician ID each time it is entered  If a clinician  ID is entered  but there is no activity on the monitor for a configured period of time  the  monitor will clears the ID     1  Select  Main     Maintenance  gt  gt      User Settings  gt  gt      Enter the required  password   Select  Ok  and then access  User Settings  menu     2  Select  Clinician ID Time out  and then set the time        3 6 Accutorr 7 Operator s Manual    3 5 4 Adjusting Alarm Volume    1  Select  Main     General Setup gt  gt      2  Select  Alarm Volume  and then select the appropriate volume  The alarm volume  range is between X to 10  X is the minimum volume  which depe
112. or adult and  pediatric patients     When oral axillary probe is used  the measurement site will automatically be set to  Oral    You can change the site in  Temp Setup  menu        10 2 Accutorr 7 Operator s Manual    10 3 2 Taking a Temperature in Predictive Mode    1   2   3     Make sure that the probe is placed in the probe well   Make sure that the temperature measurement type and site settings are correct     Remove the probe from the probe well and insert it into a cover in the probe cover  pack  Press the probe handle down firmly until the cover engages with the probe     The temperature module starts to warm up when the probe is taken out of the probe  well  The message  Temp Warming Up  displays in Temp parameter area  The  warming up time is about 2 seconds at room temperature  The monitor sounds two  beeps and displays the message  Temp Prediction Ready  on the screen when  warm up is complete  Then you can place the probe at the measurement site     Place the probe at the measurement site and wait until the measurement stabilizes     d  When the dynamic symbol appears  it indicates that the monitor starts to take  the measurement         When taking an oral temperature  apply the probe under the patient s tongue  from either side of the mouth  Make sure that the probe reaches the rear  sublingual pocket  Have the patient close his her lips to hold the probe  Use  your hand to hold the probe in place  Make sure that the probe contacts with  the patient   s oral t
113. ord any needed data prior to changing the date and time        Accutorr 7 Operator s Manual    3 5 11 Configuring Unit    You can configure the measurement unit     To enter the  Unit Setup  menu  select  Main     General Setup gt  gt      gt  Unit Setup gt  gt    and then the  Unit Setup  menu pops up     3 5 12 Configuring Printout    You can select the items to be printed on the strip  By default  all items are selected     To enter the  Print Setup  menu  select  Main     General Setup gt  gt      gt  Print Setup gt  gt    and then the  Print Setup  menu pops up     3 5 13 Selecting a Central Monitoring System  The monitor can be configured with multiple central monitoring systems  CMS   If  Select  CMS  is enabled  you can select one CMS for the current monitoring  In Monitor mode   when a CMS is selected  the Network and USB connection area will display the CMS name   when no CMS is selected  the area displays               To select a central monitoring system  select  Main      General Setup gt  gt       Select  Central Station gt  gt    and then select a CMS in the pop up menu     Refer to section 12 2 7 Central Monitoring System Setup for configuring the name and IP  address of a CMS   To enable the select CMS function     1  Select  Main      Maintenance  gt  gt       User Settings  gt  gt      Enter the required  password  and then select  Ok  to access  User Settings  menu     2  Select  Network  gt  gt       Select CMS  and then select  On      3 5 14 Clearing t
114. our service   OM personnel    Recorder No Low The recorder has   Stop the recording   Head Hot been working for   and resume the  too long time  recording until the                recorder   s print head  cools down           Accutorr 7 Operator s Manual    D 7                      Alarm Clearable  Alarm Cause Solution  Message  Yes No  Level  Rec Head   Yes Low The thermal head of   Restore the control  Wrong Pos  the recorder is in   lever of the recorder  wrong position  to its previous  position   Recorder out   Yes Low The recorder paper   Replace with a new    of paper             is used up        paper roll        D 2 5 Power Alarm Messages                                           Alarm Clearable  Alarm Cause Solution  Message  Yes No  Level  12V Too High No High There is a problem   Restart the monitor  If    with the system   the problem still  Te Too Low No High power supply  exists  contact  5V Too High No High Mindray or your  7 service personnel   5V Too Low No High  3 3V Too High   No High  3 3V Too Low No High  Battery Too   No Med The battery charge is   Connect the monitor  Low too low  to an AC power source          and allow the  Ee J No High ms aa eas iS   atteis to charge  eplete almost depleted  immediately   Power Board   No High No data from power   Restart the monitor  If  Comm Err module has been   the problem still  received for 5   exists  contact  seconds  Mindray or your  service personnel   RT Clock Need   No Low There is a problem   Reset the
115. pensate for the theoretical effect on oximeter  measurements of fetal hemoglobin in neonatal blood        Information of the Test Subjects of the Clinical Study Report                                               Skin color Gender Number Age  years  Health  Black Male 1 28 2 9 19 Healthy  Female 1  Yellow Male 3  Female 9  A 6 2 PR  Module PR High Limit PR Low Limit Step  bpm   Masimo SpO2 Module  low limit  2  to 240 25 to  high limit 2   Nellcor SpO2 Module  low limit  2  to 300 20 to  high limit 2  1  NIBP Module  low limit  2  to 240 40 to  high limit 2              PR from Masimo SpO2 Module       Measurement range    25 to 240 bpm       Resolution    1 bpm       Response time     lt  20 s  PR value sudden change within 25   240 bpm        Accuracy     3 bpm  without motion    5 bpm  with motion           Refreshing rate       1s          Accutorr 7 Operator s Manual       PR from Nellcor SpO2 Module       Measurement range    20 to 300 bpm       Resolution    1 bpm       Response time     lt  30s  PI  gt  0 3  no disturbance  PR value sudden change within  25   250 bpm        20 to 250 bpm   3 bpm                            easily 251 to 300 bpm  not specified   Refreshing rate 1s   PR from NIBP Module   Measurement range 40 to 240 bpm   Resolution 1 bpm   Accuracy  3 bpm or  3   whichever is greater   A 6 3 NIBP   Sener Meet standards of IEC80601 2 30  EN1060 1  EN1060 3   EN1060 4 and SP10   Technique Oscillometry       Mode of operation    Manual  Auto and STAT     
116. rea        Accutorr 7 Operator s Manual 4 3    Manual Inputs    Manual Inputs    Pain Level    Loc    RESP    Patent Poston    NIBP Location    Temp Posifon Ora       Ok    Tere  c   VO Fluid 1 m   Glucose myd   02 Source   02      02 Fow Rato LPM  Cancel    4  Select   tosave the configuration and exit the menu     The following picture shows the manual input area which is located at the bottom of the    screen if turned on        Pain Level RESP    1    Glucose    Temp 1 0 Fluid       After the manual input area is selected  the  Manual Inputs  menu pops up  The    displayed items correspond to the settings in  Manual Inputs Setup      Manual Inputs    as a    Parameter2 LOC    Parameters RESP    Modifier  B Patient Postion    B NIBP Location    O Temp Position    Display  Manual input Area On    ParameterS VO Fluid    Parameter6 Glucose    B 02 Source    BO2     B 02 Flow Rate          4 4    Accutorr 7 Operator s Manual    4 3 Manually Save Patient Data    Select the  Save  QuickKey to save the demographics  measurements and manually  entered data for the current patient     You can configure the data processing mode for the manually saved data     1     Select  Main     Maintenance  gt  gt      User Settings  gt  gt     enter the required  password    gt select  Ok  to access  User Settings  menu     Select     Save    Button Options  gt  gt     In     Save    Button Setup  menu  select       Automatically Send On Manual Save   the data saves locally and is sent to the  
117. rend display or NIBP List display         All Parameters display includes parameter area  waveform area  but no trend       Trend display includes parameter area  tabular trends  but no waveform area     NIBP List display includes parameter area  NIBP list  but no waveform area     Refer to 2 4 Main Screen for additional information     3 5 8 Configuring the Timeout of Measured Value  You can configure the retention time for the digital value of the current NIBP and  temperature measurement displayed on the screen     To set the timeout     1   2     Select  Main      General Setup gt  gt     Select  Parameter Time Out  and select an appropriate setting     The options are 5 min  10 min  15 min  and 30 min and off  When  Off  is selected  the  digital value of the current NIBP and temperature measurement will display on the  screen until a new measured value replaces it     3 5 9 Configuring Measurement Colors  You can set the desired color for SpO2  NIBP  Temp and PR     Select  Main      General Setup gt  gt      Parameter Color Setup gt  gt   and then the  Select  Color  menu pops up     3 5 10 Setting the Date and Time    1     Select  Main     General Setup gt  gt      System Time gt  gt      Set  Date  and  Time     Select  Date Format  and toggle between  yyyy mm dd    mm dd yyyy  and   dd mm yyyy     Select  Time Format  and toggle between  24h  and  12h      Aimon    Changing date and time affects the storage of trends and events and may cause  data loss  Save or rec
118. rs  the equipment also uses the following       Accutorr 7 Operator s Manual 6 3    symbols telling the alarm status     a indicates alarms are paused     PN   a indicates alarms are reset     a indicates the alarm sound is turned off     6 4 Setting Alarms    You can set the switch  limit and level of physiological alarms   Select  Main     Alarm Setup  gt  gt    and then access the  Alarm Setup  screen     Parameter Onoff High Low Level  Spo2 On 100 390 Med  Desat On 80 High  NIBP Sys On 160 30 Med  NIBP Dia On 30 50 Med  NIBP Mean On 110 60 Med  Temp On 38 0 35 0 Med  PR On 120 50 Med  Auto Set Restore Defaults     Auto Set   The monitor will create new alarm limits based on the measured values      Restore Defaults   The restored defaults depend on the settings in  Select Default  Config  screen  If the latest configuration is set as the default configuration  then the  factory configuration will be loaded for the alarm settings  if a specified configuration  is set as the default configuration  then the specified configuration will be loaded for  the alarm settings  Refer to 5 2 1 Setting Default Configuration        6 4    Accutorr 7 Operator s Manual          WARNING       A    Make sure that the alarm limits settings are appropriate for your patient before  monitoring    Setting alarm limits to extreme values may cause the alarm system to become  ineffective  For example  High oxygen levels may predispose a premature infant  to retrolental fibroplasia  If this is a co
119. rs has been validated for motion accuracy in human  blood studies on healthy adult volunteers in induced hypoxia studies while performing  rubbing and tapping motions at 2 to 4 Hz  At an amplitude of 1 to 2 cm and non repetitive  motion between 1 to 5 Hz  At an amplitude of 2 to 3 cm in induced hypoxia studies in the  range of 70  to 100  SpO2 against a laboratory co oximeter and ECG monitor  This variation  equals plus or minus one standard deviation  Plus or minus one standard deviation  encompasses 68  of the population       The Masimo pulse oximeter has been validated for low perfusion accuracy in bench top  testing against a Biotek Index 2 simulator and Masimo   s simulator with signal strengths of  greater than 0 02  and a   transmission of greater than 5  for saturations ranging from 70 to  100   This variation equals plus or minus one standard deviation  Plus or minus one standard  deviation encompasses 68  of the population           Accutorr 7 Operator s Manual A 5       Nellcor SpO2 Module                Standards Meet standards of ISO 9919  ISO 80601 2 61  Measurement range 0 to 100   Resolution 1    70 to 100    2   adult pediatric   Accuracy 70 to 100    3   neonate     0  to 69   Not specified        Refreshing rate    1s       Response time        lt  30s  PI  gt  0 3  no disturbance  SpO2 value sudden change  within 70    100           When the SpO2sensor is applied for neonatal patients as indicated  the specified accuracy  range is increased by  1   to com
120. s  whenever possible check  that the patient       E Is comfortably seated     M Has legs uncrossed    E Has feet flat on the floor    E Has back and arm supported  and    E   The middle of the cuff at the level of the right atrium of the heart   NOTE        It is recommended that the patient relax as much as possible before the NIBP  measurement is performed and that the patient does not talk during  measurement         It is recommended that the patient sit still for 5 min before the first  measurement is taken      The operator should not touch the cuff and tubing during the NIBP  measurement     9 5 2 Preparing to Measure NIBP    1 Power on the monitor     2  Verify that the patient category is correct   If not  select the    hardkey       Patient Demographics     Patient Category  and set the patient category to   Adult    Pediatric  or  Neonatal      3  Connect the NIBP hose to the monitor    Select the appropriate sized cuff by referring to the limb circumference marked on  the cuff     The width of the cuff should be 40   50  for neonates  of the limb circumference  or  2 3 of the upper arm s length  The inflatable part of the cuff should be long enough  to encircle at least 50  to 80  of the limb     5  Apply the cuff to the patient   s upper arm or thigh and make sure the    marking on  the cuff is aligned with the artery  Do not wrap the cuff too tightly around the limb  It  may cause discoloration  and ischemia of the extremities  Make sure that the edge of  cuff 
121. s Manual 9 9          3  Hold the hose and the cuff and shake the complete cuff until the bladder is in  position    4  Thread the hose from inside the cuff  and out through the small hole under the  internal flap     9 9 2 Reusable Bladderless Cuffs  Clean cuffs with warm water and a mild detergent  Do not use a detergent containing  hand conditioners  softeners  or fragrances     Antimicrobial Definition   Bladderless cuffs are treated with an antimicrobial coating  Antimicrobial technology  effectively controls a broad spectrum of bacteria  fungi  algae and yeasts on a wide variety  of treated substrates        9 10 Accutorr 7 Operator s Manual    10 Monitoring Temp    10 1 Overview  The SmarTemp    Temp module is intended for monitoring oral  axillary and rectal  temperature of adult and pediatric patients and axillary temperature of neonatal patients     Temperature can be measured in either Predictive mode or Monitor mode  The default is    Pred    ictive mode           A    WARNING    Do not take oral temperature on the infant  0 3 years    Do not take rectal temperature on the neonate  0 28 days      Use only the specified temperature probe and probe cover  Using other probe  or probe cover  or not using probe cover may cause damage to the monitor or  failure to meet the declared specifications in this manual     The temperature probe cover is disposable  Re use of probe cover may result in  patient cross contamination     Use disposable probe covers for temperature m
122. s that result from the mechanical activity  of the heart  The pulse value can be from SpO gt  or NIBP  The PR parameter area displays its       source   1 3  2  4  1  PR high limit  2  PRlow limit  3  Pulse rate  PR   detected pulsations per minute   4  PR Source  NOTE       A function tester or SpO  simulator can be used to determine the pulse rate  accuracy     8 2 PR Source  The current pulse source is displayed in the PR parameter area  The pulse rate is      stored in the monitor s database and reviewed in the graphic tabular trends      sent via the network to the central monitoring system  if available     8 3 Pulse Tone  You can change the pulse tone volume by adjusting  Beat Volume  in the  SpO2 Setup   menu  When a valid SpO  value exists  the system will adjust the pulse tone according to  the SpO  value        Accutorr 7 Operator s Manual 8 1    FOR YOUR NOTES       8 2 Accutorr 7 Operator s Manual    9 Monitoring NIBP    9 1 Overview  The monitor uses the oscillometric method to measure the non invasive blood pressure   NIBP   This measurement can be used for adults  pediatric and neonatal patients  To  understand how this method works  we will compare it to the auscultative method     With auscultation  Clinicians listen to the Korotkoff sounds to determine blood pressure  when using the auscultatory method  The estimated mean pressure can then be  calculated with reference to these     Since the monitor cannot hear the Korotkoff sounds to determine the blood pr
123. scedenssaseecssiesbdadeusenstacvaessasensboesedeveddeeeseseseesossnceeasee A 1  AYA Classification Seine t E E N ER N Ri A 1  A 2 Environmental Specifications      csssssccsssssscssssssssssssnssesssssssesssssssssssssssessssnseesssssssesssssseessssnseessssss A 1  A 3 Power Supply Specifications        sssssssessssccccsssseesecceecssssnsesesceecenssssssesessssnsseseeseeecennneeeeseeeessnssesess A 2  A 4 Physical Specifications          ssssssssssccssssssssseescecccssssssseceessssnsseeeeceecsnsusssecceeecsssnnsseeseeecennsesecseeesnnneesees A 2  A 5 Hardware Specifications A 2  A 6 Measurement Specifications     sssccsssssssssssssscsssssssssssssssessssssesssssnsessssnseessssnseesssssseessessueessssnseesssses A 4       Accutorr 7 Operator s Manual    Bit EMG E ETE ice sentine arses Ma oni ial Mai neni eae  daa B 1  B 2 Radio Regulatory COMpliance       ssssessssccscssssesssseccssssseeesececsnnssseeseesessnsesseeeeeeeannuesseceesesnnnneseeeeeees B 5  CE Default  CONFI GUPALIONS s nessas sogana rosaa srs nias C 1  C 1 Parameter Configuration         ssscscsssssssssssssssssssssessssssessssssssesssssusesssssnsesssssnseessssneessssnseesssssneesssssseesss C 1       C 2 General Configuration  C 3 User Maintenance Items       D Alarm Messages             ssccscssssesscscccsscssscccscccsscsescccsccsssccssecssccsssccssscsescccsscsssccssccssscssscssees D 1  D 1 Physiological Alarm Messages      sssssssccsssssssessscssccsssssssesceeecsnsssssecesecessssnseeseceeeansueesscesecennseseeeseeee D 
124. sinfectants    To disinfect the temperature probe   1  Disconnect the temperature probe from Temp connector   2  Disinfect the probe with a soft cloth dampened with the recommended disinfectant     3  Wipe off all the remaining disinfectants from the probe with a soft cloth dampened  with water     4  Dry the probe in a cool place           A WARNING       Properly dispose of the used soft cloth              Accutorr 7 Operator s Manual 10 5    FOR YOUR NOTES       10 6 Accutorr 7 Operator s Manual    11 Recording    11 1 Using a Recorder  The thermal recorder records patient information  parameters numerics  measured value  and manual input value   SoO2 waveforms  if configured   and so on            A                   Start Stop key  press to start a recording or stop the current recording    Indicator      On  when the recorder is working properly    Off  when the monitor is switched off      Flash  when a recorder error has occurred  e g   the recorder runs out of  paper    Paper outlet   Recorder door   Latch    11 2 Loading Paper    1     2   3   4    Press the latch in the upper right corner of the recorder door to open the door   Insert a new roll into the compartment as shown below    Close the recorder door    Check if paper is loaded correctly and the paper end is feeding from the top        Accutorr 7 Operator s Manual 11 1    Paper roll          A CAUTION  e    Use only specified thermal paper  Otherwise  it may cause damage to the  recorder   s printhead  the r
125. sssseeess   13 5 1 Conditioning a Battery  oe   1325 2  CHECKING a Battery iseitis iseset iteiten  13 6  Recycling a BALCONY sajenssaactccesveccensnsssscecteueedes ehecensoc ERA A seus a deena       TA  Care and Maintenance  siscssccsscccscscicscccsssescssedessestusisccssosssenscossesetsesesssdeuscedsevossssesessvenseds 14 1    14 1 Cleaning and Disinfection                VA AGT ClO EEEIEE AE EAA  T4A 2 Disinfect gassene A eee eh N E AEEA aa is  14 2 G  ne  ral Inspection ceassssasscosassssnscccnsacsasvscsitucsssascsatsatesasesatasanseadeasaviaasssnisssaascassussarssstansicoaasssiaennatincs  14 3 Maintenance and Testing SCHECUIe         ssssssssssccsssecssesccsssccsssecessscccssccesssecesnecssueccssecssneess 14 3  14 4 Checking Monitor Information        sssscsssssssscsssssssssssssssssssssessssnssesssssssesssssnsssssssnecssssnsesssneessess 14 5  14 7 Batter yGHECK  ninis i scccsia cetecabenecttescsdeacascsectoassustacaatecchsnustichoassphctbea A NE ie a  14 8 Calibrating the Touchscreen       14 9 Formatting the Storage Card     sscssssssssssccssseccsssccsssecssseecssusecssseccsuecessecssasecssuceecaeccsneecssneess  14 10 Modifying Password      csssssssssssssseccesssseesssceccessssesscscecessssussesceeeccsnsuesscseeeesssuneseeceecennneeecseesssnnsees  ERAL E T R 15 1  15 1 SpO2 Accessories   f  15 2 NIBP ACCESSOTIES sitiseni isaisa isisisi iiaiai sepies i ieradas dedii oie ia  15 3 Temp Accessories  TSA OPENS iin aA R i a  A Product Specificati  hS  ascsisisccssecsecsensssc
126. ssssscssssssessessseessesees 9 5   9 6 Understanding the NIBP NUMELICS        sssssecsssssccssscccssecssseccesseceessecesssccesneecsssecesssecesasecesueesenneeese   OF SOCHIMGINIB Pesesssasscssssnasctesacissscasscucsscasesesscavezctstessicascessaevssacoasacssuta O AEAT  ZV SOUCHIAG UNTO Va cs iccascacsasiccosscceags cotnseszctqsenacnighiescsenaveusasgacnavsuadsncdeighcsdvocasusc  echdesigu  stetecobeiebendees  9 7 2 Setting the Initial Cuff Inflation Pressure     9 7 3 Setting NIBP ENA  TONE shssessissssccsascicsiscsscssensasescasvacssagacoasacntsseceucnchetncobincehogh oben cadensdconasterbeted  9 7 4 Switching On Off Measurement ON ClOCK u    cssssssseessccccessssseceseccsssnsseeecceeecnnseseeeeeeeee 9 7  9 7 5 Configuring a CUSTOM PLOGKAIM   ssssscssssssssssssssssssssssesssssnssessssssesssssssssssssuscesssssseessssneeessssses  9 7 6 Setting NIBP Alarm Properties      i  9 7 7 Setting the  Pressure Umit   cesccsccssccssescastasisewscrsassssvossssscezectashccessnccavecssicosteuensscsssuleteeceetectes   9 8 Assisting VENOUS PUNCTULEC        escsescsseccsescnsecsseccseccnsccssccssecenscessecsssccusecsneesssecusecsseesseccuscenscesseesseessees             10 Monitoring Temp                cccsccccscscsccssscscccescccssccescccscccssccsscccsccsssccssccssscssssssssccssccssscsssese 10 1    TOs OVER VIG W ita anean aE ENEE AEON EKN ENES ENEN 10 1  10 2 Setting TeM Prinia i i ii ini 10 2  10 3 Prepatation ngan E 10 2  10 3 1 Selecting Measuring Site ou    sssesssssseccsseccsssecesscc
127. sured field strength in the location in which the  ME EQUIPMENT or ME  SYSTEM  is used exceeds the applicable RF compliance level above  the  ME EQUIPMENT or ME  SYSTEM  should be observed to verify normal operation  If abnormal performance is observed   additional measures may be necessary  such as re orienting or relocating the  ME EQUIPMENT  or ME SYSTEM       gt  Over the frequency range 150 kHz to 80 MHz  field strengths should be less than3V m              A WARNING         The device may be configured with a wireless network module to receive  wireless signals  Other devices may interfere with this device even though they  meet the requirements of CISPR              Recommended separation distances between portable and mobile RF communications  equipment and the device       The device is intended for use in an electromagnetic environment in which radiated RF  disturbances are controlled  The customer or the user of the device can help prevent  electromagnetic interference by maintaining a minimum distance between portable and  mobile RF communications equipment  transmitters  and the device as recommended below   according to the maximum output power of the communications equipment        Rated maximum   Separation distance in meters  m  according to frequency of the       output power of  transmitter  W     transmitter       150 kHz   80 MHz    d  1 24P    80 MHz   800 MHz    d  1 2 4P    800 MHz   2 5 GHz    d   2 34 P                   0 01 0 12 0 12 0 23  0 1 0 38
128. t failure and possible health hazards     The safety checks or maintenance involving any disassembly of the equipment  should be performed by professional servicing personnel  Otherwise  undue  equipment failure and possible health hazards could result    If you discover a problem with any of the equipment  contact your service  personnel or Mindray     The responsible hospital or institution shall carry out all cleaning and  disinfection procedure specified in this chapter     Do not open the equipment housings  All servicing and future upgrades must  be carried out by the personnel trained and authorized by Mindray only     No modification of this equipment is allowed     The service personnel must be properly qualified and thoroughly familiar with  the operation of the equipment           14 1 Cleaning and Disinfection  This section describes cleaning and disinfection procedure for the monitor only  For the  cleaning and disinfection of other reusable accessories  refer to instructions for use of  corresponding accessories     Keep the equipment and accessories clean  To avoid damage to the equipment  follow  these guidelines     Always dilute cleaners and disinfectants according the manufacturer s instructions or  use lowest possible concentration     Do not immerse any part of the equipment into liquid   Do not pour liquid onto the equipment or accessories   Do not allow liquid to enter the case     Never use abrasive materials  or corrosive cleaners  such as acetone or 
129. t is not being monitored  but the monitor is still powered on     If no parameter is being measured  you can press the power switch to enter Standby  mode  A warning pops up  Select  Yes  to enter the Standby mode     When the monitor is powered by a battery  it will automatically enter the Standby mode  when the following conditions are satisfied     E No key operation within 10 minutes   E No unacknowledged alarms     To exit Standby mode  use any one of the following methods     E Press any hardkey on the front panel   E Rotate the knob     E Connect SpO  sensor  and let the monitor receive SpO2 signal for more than 5  seconds     E Remove the temperature probe from the probe well     NOTE      Ifthe monitor enters and then exits Standby mode during patient monitoring   you must re admit the patient before continuing monitoring     2 6 4 Demo Mode    Demo mode is password protected  and it is used for demonstration purpose only     To enter Demo mode   1  Select  Main     Maintenance gt  gt     2  Select  Demo gt  gt     Enter required password   Select  Ok      To exit Demo mode   1  Select  Main     Maintenance gt  gt     2  Select  Exit Demo            unine      The Demo mode is for demonstration purpose only  To avoid the potential risk  of the simulated data being mistaken for the monitored patient s data  do not  enter the Demo mode while monitoring a patient  Otherwise  improper patient  monitoring and delayed treatment could result              Accutorr 7 Operator
130. temperature or vasoconstrictor     7 7 Masimo Information    YME     E Masimo Patents    This device is covered under one or more the following U S A  patents  5 758 644   6 011 986  6 699 194  7 215 986  7 254 433  7 530 955 and other applicable patents listed  at  www masimo com patents htm     m  Nolmplied License    Possession or purchase of this device does not convey any express or implied license to  use the device with unauthorized sensors or cables which would  alone  or in combination  with this device  fall within the scope of one or more of the patents relating to this device        Accutorr 7 Operator s Manual    7 8 Nellcor Information        NELLCOR        Nellcor Patents    This device may be covered by one or more of the following US patents and foreign  equivalents  5 485 847  5 676 141  5 743 263  6 035 223  6 226 539  6 411 833  6 463 310   6 591 123  6 708 049  7 016 715  7 039 538  7 120 479  7 120 480  7 142 142  7 162 288   7 190 985  7 194 293  7 209 774  7 212 847  7 400 919     mE Nolmplied License    Possession or purchase of this device does not convey any express or implied license to  use the device with unauthorized replacement parts which would  alone  or in    combination with this device  fall within the scope of one or more of the patents relating  to this device        Accutorr 7 Operator s Manual 7 7    FOR YOUR NOTES       7 8 Accutorr 7 Operator s Manual    8 Monitoring PR    8 1 Overview  The pulse numeric counts the arterial pulsation
131. tential        Accutorr 7 Operator s Manual 2 5    2 3 4 Bottom View                                                             1  Battery compartment door  2  Quick release mount latch point       2 6 Accutorr 7 Operator s Manual    2 4 Main Screen  The main screen has three display modes  They are all parameter screen  trend screen and  NIBP list screen     E All Parameter Screen  1 8 7     aean    JACKSON TOM 2013 04 09 13 50 16 4       MRN 213 Centrat   S 6  2  NIBP  mmHg     166 100 122   13 43 Manual  3  Temp P Axillary  Cc  3 8 e 0  Temp Prediction Over  mp V    Fluid Glucose  4   j  5 MEWS Patient List Review       Accutorr 7 Operator s Manual 2 7    E Trend Screen    JACKSON TOM  MRN 213    NIBP  mmHg    2013 04 09 14 12 24 1  Centrali  gt       166 100  122     13 43    Axillary    Temp P    c    38 0    Temp Prediction Over  2013 04 09 14 12 00  14 11  14 11 00  2013 04 0 14 10 30  2013 04 09 14 10 00  14 09 30    2013 04 09    2013 04 09    g Main MEWS Patient List    Manual    Review          2 8    Accutorr 7 Operator s Manual    m  NIBP List Screen 1          8  JACKSON  TOM 2013 04 09 18 33 30     MRN 213 Centrali      6  2  NIBP  mmHg  Manual  3  Temp P Axillary  c   m m  Temp Prediction Ready  2013 04 09  166 100   10    5  a man   MEWS Patient List Review    1  Patient Information System Message Area    This area normally shows patient information  such as patient medical record number   patient name  patient category  room  bed number  clinician ID     Wh
132. than    gt  greater than    lt  less than or equal to   gt  greater than or equal to    plus or minus   x multiply      copyright   E 2 Accutorr 7 Operator s Manual    E 2 Abbreviations    AAMI  AC  ADT  Adu  CE  CISPR  CMOS  CMS  DC  Dia  DIAP  DPI  EEC  EMC  EMI  EMR  ID   IEC  ISO  IEEE    LED  MDD  MetHb  MEWS  MRI  N A  Neo  NIBP  NIBP D  NIBP M  NIBP S    Association for Advancement of Medical Instrumentation  alternating current  Admit Discharge Transfer   adult   Conformit   Europ  enne   International Special Committee on Radio Interference  complementary metal oxide semiconductor  central monitoring system   direct current   diastolic   Datascope Improved ASCII Protocol   dot per inch   European Economic Community  electromagnetic compatibility  electromagnetic interference   Electronic Medical Record   identification   International Electrotechnical Commission  International organization for standardization  Institute of Electrical and Electronic Engineers  internet protocol   light emitting diode   Medical Device Directive   methemoglobin   Modified Early Warning Score   magnetic resonance imaging   not applied   neonate   noninvasive blood pressure   NIBP diastolic pressure   NIBP mean pressure   NIBP systolic pressure    power       Accutorr 7 Operator s Manual       PD photodetector  Ped pediatric  Pleth plethysmogram  PR pulse rate  RAM random access memory  ROM read only memory  SpO2 arterial oxygen saturation from pulse oximetry  TD temperature difference  T
133. the battery until  it shuts off  Note the time again  Calculate the run time by subtracting the start time  from the end time     The operating time of a battery directly reflects its performance  If the operating time of a  battery is noticeably shorter than that stated in the specifications  contact your Mindray  service personnel     NOTE      The battery may be damaged or may have malfunctioned if it only operates for  a short time after being fully charged  The operating time depends on the  configuration and operation  For example  measuring NIBP more frequently will  also shorten the operating time        Replace a battery that has visual signs of damage or no longer holds a charge   Remove the old battery from the monitor and recycle it according to local laws     13 6 Recycling a Battery    Remove the old battery from the monitor and recycle it properly  Follow local laws for  proper battery disposal           A WARNING       Do not disassemble batteries  dispose of them in fire  or cause them to short  circuit  They may ignite  explode  leak or heat up  causing personal injury              13 4 Accutorr 7 Operator s Manual    14 Care and Maintenance    The monitor should be maintained and cleaned on a regular basis  This chapter describes  the basic cleaning  disinfection and test method           Arane       Failure of the responsible individual  hospital or institution using this  equipment to implement a satisfactory maintenance schedule may cause  undue equipmen
134. tient s discomfort and inaccurate measurements        Improper use of probe may also cause patient s discomfort and inaccurate  measurements       Inthe rectal mode  incorrect probe placement may result in bowel perforation        Hospital staff should wash their hands after the patient   s temperature  measurement is taken  This will significantly reduce the risk of cross  contamination and nosocomial contamination     10 2 Setting Temp    Select Temp parameter area to access  Temp Setup  menu  You can set   E Temp type   Predictive  or  Monitor      E Temperature measurement site  the measurement site is related to the probe type   When using oral axillary probe  you can select the site  Oral  and  Axillary   when  using rectal probe  you can select  Rectal      You can select the temperature type and measurement site only when the probe is in the  probe well     10 3 Preparation    10 3 1 Selecting Measuring Site    The temperature module can be configured with 2 types of temperature probe     E  oral axillary probe  blue   and  E rectal probe  red     Use the blue oral axillary probe with blue probe well  and use the red rectal probe with  red well     Be sure to select correct probe according to the measurement site       E  Oral Axillary probe  this probe type is intended for taking oral or axillary temperature  of adult and pediatric patients  or axillary temperature for neonatal patients       E Rectal probe  this probe type is intended for taking rectal temperature f
135. tion to the mains power    supply  Do not position the equipment in a place difficult to access the mains  plug     1 2 Equipment Symbols    Some symbols may not appear on your equipment        f 5 ON OFF for a part of  Caution C O     equipment       Alternating current Battery indicator       Alarm Reset NIBP Start Stop key       ALARM PAUSED Graphical recorder       ORES    Admit patient key Insertion Direction       g    DEFIBRILLATION  PROOF    TYPE CF APPLIED PART Input Output       Network connector Equipotentiality          USB connector MANUFACTURER             Ed   Amg E          1     mPa    Accutorr 7 Operator s Manual                         SN Serial number m  DATE OF MANUAFACTURE  Protection against fluid  CATALOGUE NUMBER IPX1 fae  4 Temperature limit Humidity limitation  T Atmospheric pressure Refer to instruction manual   limitation booklet  AUTHORISED keJ  Interference may occur in  REPRESENTATIVE IN THE the vicinity of equipment  EUROPEAN COMMUNITY marked with this symbol             oo    The product bears CE mark indicating its conformity with the provisions of  the Council Directive 93 42 EEC concerning medical devices and fulfills the  essential requirements of Annex   of this directive        A  M  oO     NO    The following definition of the WEEE label applies to EU member states only   This symbol indicates that this product should not be treated as household  waste  By ensuring that this product is disposed of correctly  you will help  prevent brin
136. to 8 0 001B 30 70693  Neonate  5 8 to 10 9 001B 30 70694  7 1 to 13 1 001B 30 70695  Child 10 to 19 0683 14 0001 01  Arm  Small Adult 18 to 26 0683 14 0002 01  Adult 24 to 35 0683 14 0003 01  Adult Long 27 5 to 36 5 0683 14 0006 01  Large Adult 33 to 47 0683 14 0004 01  Large Adult Long 35 5 to 46 0683 14 0007 01  Adult Thigh 46 to 66 0683 14 0005 01  15 2 Accutorr 7 Operator s Manual       15 3 Temp Accessories                                                          Probe Well  Type Description Part No   Blue  Oral Axillary MO9A 20 62062  Reusable  Red  Rectal MO9A 20 62062 51  Temp Probes  Type Patient Category Measurement Site Part No   Adult  Pediatric  Neonate   Oral Axillary 6006 30 39598  Reusable  Adult  Pediatric Rectal 6006 30 39599  Probe Cover  Type Patient Category Description Part No   Adult  Pediatric  Neonate   Cover  20 pcs pack MO9A 20 62124  Disposable  Adult  Pediatric  Neonate   Cover  2000 pcs pack MO09A 30 62128  15 4 Others  Material Part No        Welch Allyn SureTemp Plus Probe Covers    0198 00 0044       Welch Allyn SureTemp Plus Thermometer Module    0992 00 0198       Welch Allyn SureTemp Plus Oral Probe    0992 00 0213 02       Quick Release Mounting Bracket for Rolling Stand    045 001054 00       Quick Release Mounting Bracket for Wall Mount    045 001055 00       Rolling Stand with Quick Release Mount    045 001057 00       Wall Mount Bracket    045 001059 00       Wall Mount Kit  6  Arm    115 025386 00       Bedrail Clamp    115 020575 00 
137. tor  Lower Limit pressure is lower and retry  If the error  A than the specified remains  contact  NIBP    Dia Over Yes Low NIBP AR Mindray or your  Lower Limit lower limit  service personnel   NIBP Mean Yes Low  Over Lower  Limit  NIBP SelfTest Yes High SelfTest Failed  The Restart the monitor  Err cause may be the and retry  If the error  transducer or A D remains  contact  sampling error  Mindray or your  service personnel   NIBP Init Err Yes Low An error occurred to   Restart the monitor  If  the NIBP module  or   the error remains   there is a problem contact Mindray or  with the your service  communications personnel   between the module  and the monitor   NIBP Comm Yes High An error occurred to   Restart the monitor  If  Err the NIBP module  or   the error remains   there is a problem contact Mindray or  with the your service  communications personnel   between the module  and the monitor   NIBP Loose Yes Low The NIBP cuff is not Check the patient   s  Cuff properly connected    condition and verify  or there is a leak in patient type  Replace  the airway  with an appropriate  cuff and connect it  correctly  If the  problem still exists   contact your service  personnel   NIBP Air   Yes Low An error occurred to   Check the air pressure   Pressure Err the air pressure  Restart the monitor  and retry  If the error  remains  contact your  service personnel   NIBP Weak   Yes Low The patient s pulse is   Check the patient s  Signal weak or the cuff is   condition and change
138. tric shock  this equipment must only be connected toa  properly grounded power outlet  If a properly grounded power outlet is not  available  operate the monitor on battery power    Ensure that the equipment is supplied with continuous electric power during  operation  Sudden power failure may lead to the loss of patient data     To avoid an explosion hazard  do not use the equipment in the presence of  oxygen rich atmospheres  flammable anesthetics  or other flammable agents   such as gasoline      Do not open the equipment housings  All servicing or future upgrades must be  carried out by Mindray trained and authorized personnel     Do not come into contact with patients during defibrillation  Otherwise serious  injury or death could result     Do not touch the equipment   s metal parts or connectors when in contact with  the patient  otherwise patient injury may result     Do not rely exclusively on the audible alarm system for patient monitoring   Adjustment of alarm volume to a low level or off may result in a hazard to the  patient  Remember that alarm settings should be customized according to  different patient situations and always keeping the patient under close  surveillance is the most reliable way for safe patient monitoring    The physiological data and alarm messages displayed on the equipment is not  intended to be directly used for diagnostic interpretation and replace the  competent judgment of a clinician    To avoid inadvertent disconnection  route all cab
139. ts  If batteries are stored for an extended period of time   they should be placed in a cool place with a partial charge of 40  to 60  capacity   Storing batteries at a high temperature for an extended period of time will  significantly shorten the life expectancy of a battery  Do not store the battery at a  temperature beyond  20   C   60   C   4  F   140  F         13 2 Accutorr 7 Operator s Manual          A    WARNING    Keep the battery out of children   s reach             Use only specified batteries      Ifthe battery shows signs of damage or signs of leakage  replace it immediately   Do not use a faulty battery in the monitor      The Lithium ion batteries have a service life of 3 years  Please replace your  battery when it reaches the end of its service life  Failure to replace the battery  may cause serious damage to your device from battery overheating   13 5 Battery Maintenance  13 5 1 Conditioning a Battery    A battery should be conditioned before it is used for the first time  A battery conditioning  cycle is one uninterrupted charge of the battery  followed by an uninterrupted battery  discharge and charge  Batteries should be conditioned regularly to maintain their useful    life     To condition a battery     1     Disconnect the monitor from the patient and stop all monitoring or measuring        2  Turn off the monitor  Disconnect the monitor from the AC power    3  Connect the monitor to the AC power  Allow the battery to be charged uninterrupted  for 6 
140. u can select  Custom Program gt  gt   to configure the duration  of automatic measurement cycle  and the time interval between two NIBP measurements   You can define two programs  respectively program 1 and program 2  Each program can  at most include five cycles  A  B  C  D  and E  In each cycle  the  Duration  and  Interval   can be set individually     You can start the programmed NIBP measurement manually  and then the monitor  automatically performs the measurement based on the cycle and interval you have  defined     When the programmed NIBP measurement is in use  the NIBP parameter area displays as  follows     NIBP  mmHg    125 80  93     10 34 00 02 28 Program 1 A 3 min       In which     Program 1 A  3 min    means   E Program 1  program name   E A cycle name   E   3 min  interval    9 7 6 Setting NIBP Alarm Properties  Select  Alarm Setup gt  gt   from  NIBP Setup  menu  You can set the alarm properties for  this parameter in the popup menu     9 7 7 Setting the Pressure Unit    1  Select  Main     General Setup  gt  gt      gt  Unit Setup  gt  gt     2  Inthe popup menu  select  Pressure  and toggle between  mmHg  and  kPa      9 8 Assisting Venous Puncture  You can use the NIBP cuff to cause sub diastolic pressure to block the venous blood vessel  and therefore help venous puncture     1  Select  VeniPuncture  gt  gt   from the  NIBP Setup  menu  In the pop up menu  verify  that the  Cuff Press   value is appropriate  Change it if necessary     Select the  VeniPu
141. ult as an example  select   Default Adult Config  and toggle between  Defaults  and user configuration s      NOTE      When the monitor powers on  it shows what configuration is restored in the  message area for about 30 seconds     5 3 Saving Current Settings    Current settings can be saved as user configuration  Up to 3 user configurations can be  saved     To save current settings   1  Select  Save Current Settings As gt  gt   in the  Manage Configuration  menu   2  In the popup dialog box  enter the configuration name and then select  Ok      5 4 Deleting a Configuration    To delete a configuration     1  Select  Delete Config  gt  gt   in the  Manage Configuration  menu  The popup menu  shows the existing user configurations on the monitor  Selecting  Config on USB  drive  gt  gt   will show the existing user configurations on the USB drive     Select the user configurations you want to delete and then select  Delete    Select  Yes  in the popup        Accutorr 7 Operator s Manual    5 5 Transferring a Configuration  When installing several monitors with identical user configuration it is not necessary to  set each unit separately  Use a USB drive to transfer the configuration from monitor to  monitor     To export the current monitor s configuration    1  Insert a USB disk to the monitor s external device connector    2  Select  Export Config  gt  gt   in the  Manage Configuration  menu    3  Inthe  Export Config  menu  select the configurations and the user maintena
142. vice even though they meet the  requirements of CISPR      When the input signal is below the minimum amplitude provided in technical  specifications  erroneous measurements could result      Portable and mobile communication equipment may affect the performance of  this device      Other devices that have RF transmitter or source may affect this device  e g  cell  phones  PDAs  and PCs with wireless function         Guidance and Declaration   Electromagnetic Emissions       The device is intended for use in the electromagnetic environment specified below  The  customer or the user of the device should assure that it is used in such an environment           Emission tests Compliance Electromagnetic environment   guidance  Radio frequency  RF  Group 1 The device uses RF energy only for its internal  emissions CISPR 11 function  Therefore  its RF emissions are very    low and are not likely to cause any interference  in nearby electronic equipment        RF emissions CISPR 11 Class A The device is suitable for use in all       establishments other than domestic and those  directly connected to the public low voltage  power supply network that supplies buildings    Harmonic emissions Class A  IEC61000 3 2          Voltage Complies used for domestic purposes  Fluctuations Flicker  Emissions IEC 61000 3 3                Accutorr 7 Operator s Manual B 1             A WARNING       This equipment system is intended for use by healthcare professionals only   This equipment  system ma
143. y cause radio interference or may disrupt the  operation of nearby equipment  It may be necessary to take mitigation  measures  such as re orienting or relocating the  ME EQUIPMENT or ME  SYSTEM  or shielding the location              Guidance and Declaration   Electromagnetic Immunity       The device is intended for use in the electromagnetic environment specified below  The  customer or the user of the device should assure that it is used in such an environment        Immunity test    IEC60601 test  level    Compliance level    Electromagnetic  environment   guidance       Electrostatic  discharge  ESD   IEC 61000 4 2     6 kV contact   8 kV air     6 kV contact   8 kV air    Floors should be wood   concrete or ceramic tile  If  floors are covered with  synthetic material  the  relative humidity should be  at least 30         Electrical fast     2 kV for power     2 kV for power    Mains power quality should             transient burst supply lines supply lines be that of a typical  IEC 61000 4 4  1 kV for  1 kV for commercial or hospital  input output lines input output lines environment   Surge IEC  1 kV line s  to  1 kV line s  to  61000 4 5 line s  line s    2 kV line s  to  2 kV line s  to  earth earth  Voltage dips   lt 5   Ur   gt 95   dip  lt 5   Ur   gt 95   dip Mains power quality should  short in Ur  for 0 5 cycle in Ur  for 0 5 cycle be that of a typical  interruptions commercial or hospital  and voltage 40   Ur  60  dipin   40  Ur 60 dipin   environment  If th
144. y of their respective manufacturers     A condition of this warranty is that the equipment or accessories which are claimed to be  defective be returned when authorized  freight prepaid to Mindray DS USA  Inc   Mahwah   New Jersey 07430 or its authorized representative  Mindray shall not have any  responsibility in the event of loss or damage in transit        Accutorr 7 Operator s Manual Il    Exemptions  Mindray s obligation or liability under this warranty does not include any transportation  or other charges or liability for direct  indirect or consequential damages or delay  resulting from the improper use or application of the product or the use of parts or  accessories not approved by Mindray or repairs by people other than Mindray authorized    personnel     This warranty does not extend to        Malfunction or damage caused by improper use or man made failure   m  Malfunction or damage caused by unstable or out of range power input         Malfunction or damage caused by force majeure events  such as  i  flood  fire and  earthquake or other similar elements of nature or acts of God   ii  riots  war  civil  disorders  rebellions  or revolutions in any country  or  iii  any other cause beyond the  reasonable control of Mindray      Malfunction or damage caused by improper operation or repair by unqualified or  unauthorized service people    E   Malfunction of the instrument or part whose serial number is not legible     E Others not caused by instrument or part itself
145. y the sensor to the patient   Select an appropriate adapter cable according to the connector type and plug this  cable into the SpO2 connector    5  Connect the sensor cable to the adapter cable    7 5 Changing SpO  Settings    7 5 1 Accessing SpO  Menu    By selecting the SpO2 parameter or waveform area  you can access the  SpO2 Setup     men    u     7 5 2 Adjusting the Desat Alarm  The Desat alarm is a high level alarm notifying you of potentially life threatening drops in  oxygen saturation     Select  Alarm Setup  gt  gt   from the  SpO2 Setup  menu  From the pop up menu  you can  set low alarm limit and alarm switch for  Desat   When the SpO gt  value is below the Desat  alarm limit and Desat alarm switch is set to  ON   the message  SpOz Desat  is displayed     7 5 3 Setting SpO  Sensitivity  For Masimo SpO2 module  you can set  Sensitivity  to  Maximum  or  Normal  in the   SpO2 Setup  menu        Accutorr 7 Operator s Manual 7 3    When monitoring neonatal or non critically ill patients who tend to move a lot  noise or  invalid signals may be caused  In this case  it is recommended that the sensitivity is set to   Low  or  Normal  so that the interference caused by motion can be filtered and therefore  the measurement stability can be ensured     7 5 4 Changing Averaging Time    The SpO gt  value displayed on the monitor screen is the average of data collected within a  specific time  The shorter the averaging time is  the quicker the equipment responds to  changes i
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
ダウンロード - コーメイ・エンタープライズ  NGW Aqua P。st 取扱説明書。ー  Samsung HS-SA660GBP User Manual  ADVANCE Live Reporter user manual  取扱説明書 - 三共電子  Sylvania 0027845 fluorescent lamp  ETAdirect ActivityBooking  User manual - Latent variable models handled with optimization  Betriebsanleitung      Copyright © All rights reserved. 
   Failed to retrieve file