Home

PROVECTA 70 - Air Techniques

image

Contents

1. er e b 21 Digital Film Mode Select Button 5252 22 dues Ro ERE ERES S SEES 21 Default Exposure Time Ud ee da Oe 21 TABLE OF 5 Keypad Operation Custom Programming 22 Changing Default Settings to Customized Values 22 Changing Display to Pulses ete tsm 22 Changing Display to Seconds Ou hae Mies oat Mo Met Mos tt afe 22 Changing Default Tooth Settings xo o Sues Reno rerom mede runc m Somos Ren uod 23 Changing Default Child Multiplier 23 Changing Film Multiplier Ratio Settings sese proa rnkGGO RES rem Rd 24 Changing Digital Multiplier Ratio Settings lle 24 IG Re Stl hia ds arg SENTIS e 24 System Operation Creqducbedeteoeielbiubeltedebbidetecqebtezxetbedet be 24 25 Seting Quis tiae E 25 Automatic Exposure Time Selection 25 PROVECTA 70 Available Preset Exposure Time Setting Combinations 25 Manual Exposure Time Selection sad bue des Ge hese 26 Exposure Emission Procedure 4 3 he eh RU 27 Warming a TERES EE OTT TRIER 27 Redding the Ex
2. Test with Tubehead Simulator Reset power by turning the Main Power Switch OFF 0 and ON I If E01 error repeats replace the X ray Control Panel IMMEDIATELY REMOVE PowER TURN THE POWER SWITCH ro OFF 0 Test with Tubehead Simulator Reset power by turning the Main Power Switch OFF 0 and then ON 1 If E02 error repeats replace the X ray Control Panel Check that the switches are not depressed Reset power by turning the Main Power switch OFF 0 and then ON I If E05 error repeats check the switches and cables for defects or short circuits Replace Switch assembly Check that the Control Panel switch is not depressed Reset power by turning the Main Power switch OFF 0 and then ON I If error repeats replace Membrane Assembly Check that the Control Panel switch A is not depressed Reset power by turning the Main Power switch OFF 0 and then ON I If E08 error repeats replace Membrane Assembly Check that the Control Panel switch is not depressed Reset power by turning the Main Power switch OFF 0 and then ON I If E09 error repeats replace Membrane Assembly Check that the Control Panel switch is not depressed Reset power by turning the Main Power switch OFF 0 and then ON I If E10 error repeats replace Membrane Assembly SYSTEM ERROR MESSAGES ERROR 941AUDIBLE SOURCE OF ANOMALY ACTION REQUIRED CODE SIGNAL Check that the Control Panel switch is The
3. The PROVECTA 70 is normally operated in the automatic operation mode where the selection combinations of the TOOTH SELECT ADULT CHILD and FILM DIGITAL Mode Select Buttons provide 20 separate default exposure settings for specific exposure applications These preset exposure time values seconds or pulses associated with the mode select button setting combinations are displayed as shown in the table below The suggested exposure times are base on D speed film exposure Note Keypad button selection is confirmed by the sounding of an audible signal and illumination of the corresponding indicator lamp Automatic Exposure Time Selection Perform the following procedure to set the PROVECTA 70 to operate in the automatic mode using preset exposure time values Press any Mode Selection Button Tooth Select Film Digital Adult Child to initiate the automatic exposure mode of operation 2 Press the ADULT CHILD Mode Select Button to select the patient size parameter Verity that the corresponding Adult Mode or h Child Mode indicator lamp illuminates and audible signal sounds for the desired preset exposure setting selection 3 Press the DIGITAL FILM Mode Select Button to select the type of media processing Verity that the corresponding indicator lamp Digital Mode or Film Mode illuminates and audible signal sounds for the desired preset exposure setting selection 4 Press the CBE TOOTH SELECT Mode Select Button to sequ
4. a a E E BRUM NM tH ma aa ATE EE exposure 15 X RAY TUBEHEAD SPECIFICATIONS Cooling 150 W 120W Cooling Curve of Tubehead 160 E 140 120 80 60 40 20 0 50 100 150 200 250 300 350 min 400 SYSTEM ACTIVATION PROVECTA 70 TURN ON PROCEDURE Turn on the PROVECTA 70 by performing the following procedure 1 Place the Main Power Switch located on the bottom of the X Ray Control Panel in the ON 1 position 2 Observe that the system self check function initiates and the digital display displays following sequence a Atl b Display Test 8 8 8 Firmware Version e g d Last setting e g e g 0 63 Note See System Error Messages pages 31 through 34 for a listing and description of possible start up error codes 3 Once the self check function completes successfully the system starts in the Upper Molar Adult amp Film Processor modes and is ready for operation Verify associated indicator illumination and exposure time display Note When power is turned off the system retains the last selected manual and preset modes Upon turn on the system resumes operation as set prior to turning off X Ray Control Panel MAIN POWER SWITCH X RAY CONTROL PULSE TIME EXPOSURE MODE Indicator Lamp SECONDS TIME EXPOSURE MODE Indicator Lamp TOOTH SELECT Mode Select Button LOWER INCISOR Indicator Lamp LOWE
5. A change in the AC line voltage affects the peak voltage applied to the X Ray tube and the value of high voltage significantly affects the spectrum of the radiation that ultimately affects the optical density of the imaging media The purpose of the X ray Control timer correction is to obtain the same nominal optical density on the imaging media independent of the AC line voltage variations from 108 132 Volts The automatic correction of the exposure time works as follows The internal voltage measuring circuit in the X ray Control Timer continuously monitors the AC line voltage When the user selects the desired exposure time and presses the X ray Exposure button the exposure time will be automatically adjusted based upon the measured AC line voltage The corrected actual exposure time will be displayed until the X ray Exposure button is released NOTE Electromagnetic interference between the equipment and other devices can occur Do not use the equipment in close conjunction with sensitive devices or devices creating high electromagnetic disturbances TECHNICAL SPECIFICATIONS The corrected exposure is calculated applying a correction factor to the selected exposure time based on an empirical law that correlates the dose with the high voltage peak and consequently with the line voltage The qualitative relationship between the multiplying factor and the line voltage is shown in the following graph EXPOSURE TIME CORRECTION FACTORS VS
6. x dd 10 11 X Ray Tube Features eruere E TTE EET IR IT 11 Environmental ossa um eub eue i 11 Apparatus and Detachable Parts Weight 541 otc RIDERE 11 Selected Exposure Time and Corrected Actual Exposure Time 11 Exposure Time Correction Factors vs Line Voltage 12 Technical Factors Measuring Method eeu oes a e eu UAR EU RC EE 13 X Ray TUBE CUNS Sasa a er DU 13 X Ray Tubehead Components 14 X Ray Tubehead Specifications LL ao EE 15 PIGMENT saara owe ee wis ee ewes 15 Filament ue ved arcu n 15 X Ray Fubehede Cooling sso vi tektites Bee he e 16 Cooling uve Tub ehegd eig 16 System ACIIVONON 560 Pe oie ott Dor a see Po pen E 17 X Ray Control Panel Functions LR rro b REP RIS 18 19 Keypad bor quate qo teet 20 Select BUON 66 2650 ol ste 004404 et E 20 Decrease select ran mud edid eund Leu d eu S E ded 20 Tooth select Mode Select Button a td NAT ANSA SEATS 20 Adult Child Mode Select Button rece eoe
7. 1 32 Incorporated in Tubehead Assembly non removable 5 Hx7 5 Wx15 D 10 TECHNICAL SPECIFICATIONS Minimum source to skin distance Maximum symmetrical radiation field 20 cm X ray beam diameter Radiation leakage at 1 m Technical factors for radiation leakage X ray tube Features Manufacturer Type Focal spot Inherent filtration Anode thermal capacity Environmental Conditions Operating temperature range Operating relative humidity range Temperature range for transport and storage Max relative humidity for transport and storage Min atmospheric pressure for storage and transport Apparatus and Detachable Parts Weight Gross weight including packaging Net apparatus weight in standard configuration 16 fixed arm 24 fixed arm 35 fixed arm Scissor arm X ray Control Panel Tubehead 16 on center plate 20 cm 8 2 3 8 6 cm 50 mR h duty cycle 1 32 70 kVp 8mA 3 seconds CEI Bologna Italy OCX 70 G 0 8 mm IEC 336 0 5 mm Al equivalent 6kJ 50 F 104 F 10 C 40 C 30 75 4 158 F 20 C 70 C lt 95 non condensing 0 6 atmospheres 100 Ibs 95 Ibs 5 Ibs 7 lbs 9 Ibs 24 Ibs 11 9 Ibs 18 5 lbs 17 2 lbs NOTE Selected Exposure Time and Corrected Actual Exposure Time The Air Techniques X ray Control Timer incorporates a special feature which automatically corrects the selected exposure time when the line voltage varies from 120 V
8. KEY FEATURES l Microprocessor controlled 2 Exposure compensation over input AC voltage range 3 User selectable exposure time in seconds pulses 4 20 preset exposure times 5 Manually selectable exposure times 6 Programmable Adult Child ratio 7 Programmable Film setting 8 Programmable Digital setting 9 Two memorized exposure settings 10 Stable positioning 11 Optional Remote Mounted X ray Control Panel 12 Optional Remote Exposure Switch that is also settable to use in a location where two switch es are required to be activated simultaneously 13 Single 4 x 4 post mount and 16 on center plate mounting 14 Designed for ease of installation and use Printed circuit boards are hinged to allow unit to be easily installed 15 Built in tubehead simulator for training and testing the X Ray Control operation SAFETY NOTICE ATTENTION USERS MANUFACTURING DATE CODE ON SERIAL NUMBER LABEL IS IN THE FORMAT MONTH YYYY Alerts users to important Operating and Maintenance instructions Read carefully to avoid any problems ully to avoid any ks 5 Warns users that uninsulated voltage C UL US within the unit may be of sufficient magnitude to cause electric shock MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK FIRE MECHANICAL AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 CAN CSA C22 2 NO 601 1 66CA Indicates type BF equipment in accordance with IEC 601 1 DO NOT
9. TOOTH SELECT button is not depressed depressed when power is turned ON l Reset power by turning the Main 2 9 Power switch OFF 0 and then X RAY GENERATION IS CONTINUOUS If E12 error repeats replace Membrane Assembly The X ray relay is not closing within the Test with Tubehead Simulator required time after the Exposure Switch Reset power by turning the Main None is pressed Power Switch OFF 0 and ON If E20 error repeats replace the X ray X RAY GENERATION IS INHIBITED Control Panel IMMEDIATELY REMOVE PowER TURN THE POWER SWITCH ro OFF 0 Test with Tubehead Simulator Reset power by turning the Main X RAY GENERATION IS TERMINATED BY A FOWE SANGA ORE sane NIGRO DN UO SAFETY TIMER If E21 error repeats replace the X ray Control Panel The X ray relay is not opening within the required time after the exposure is Continuously completed Pulsating The X ray triac is not closing within the Test with Tubehead Simulator required time after the Exposure Switch Reset power by turning the Main is pressed Power Switch OFF 0 and ON I If E22 error repeats replace the X ray X RAY GENERATION IS INHIBITED Control Panel IMMEDIATELY REMOVE PowER TURN THE POWER SWITCH OFF 0 Test with Tubehead Simulator Reset power by turning the Main Power switch OFF 0 and then ON I If E23 error repeats replace the X ray Control Panel The X ray triac is not opening within the require
10. only to the optional Remote X Ray exposure button This is not a telephone connection Before performing any maintenance the equipment must be disconnected from the input line voltage by means of the relevant circuit breaker at the main panel Do not allow water or other liquids to penetrate inside the equipment to avoid short circuits and corrosion Although the X ray doses provided by modern equipment are reduced as low as reasonably possible during exposure the operator must take all the necessary precautions and or protection measures for the patient and themselves in compliance with existing regulations PROVECTA 70 must be switched off during the use of High Frequency surgical devices or similar instruments placed near the equipment Do Not place ANY additional weight on the extension arms ENVIRONMENTAL RISK AND DISPOSAL The equipment contains in some of its parts solid and liquid substances which must be disposed of at appropriate recycling centers conforming to all local state and federal regulations In particular the equipment contains the following materials and or components Tubehead non biodegradable plastic materials metals glass dielectric oil lead tungsten Other parts of the equipment non biodegradable plastics metals printed circuits and electronic components NOTE Air Techniques Inc is not responsible for disposal of the apparatus or parts thereof and for the related expenses 7 CLEANING amp D
11. problem is detected follow the identified Action Required the ERROR CODE table below and call your Air Techniques Authorized Dealer for service as required ERROR For proper maintenance and safety service MUST be CODE performed only by an Air Techniques Authorized Dealer Service Technician Warning Immediately remove power if error code E02 appears on the display and a pulsating audible tone is heard X Ray generation is continuous and the power switch must bet set to OFF 0 position to end emissions Call your Air Techniques authorized dealer for service 31 SYSTEM ERROR MESSAGES SOURCE OF ANOMALY ACTION REQUIRED ERROR CODE AUDIBLE SIGNAL The X ray relay is closed when power is turned on X RAY GENERATION IS INHIBITED Voltage is present at the Tubehead when power is turned ON 1 Continuously This is a result of multiple component Pulsating failures X RAY GENERATION IS CONTINUOUS The main or Remote Exposure Switch is depressed when power is turned ON 1 X RAY GENERATION IS INHIBITED The DIGITAL FILM button is depressed when power is turned ON I X RAY GENERATION IS INHIBITED The INCREASE button A is depressed when power is turned ON 1 X RAY GENERATION IS INHIBITED The DECREASE W button is depressed when power is turned ON I X RAY GENERATION IS INHIBITED The ADULT CHILD button is depressed when power is turned ON I X RAY GENERATION IS INHIBITED
12. ray short time Poor film processing technique Insufficient development time Deteriorated developer contaminated developer Developer temperature below recommended value Incorrect developing fluid dilution Expired film Dark radiographs Possible causes Excessive exposure to X ray long time Film exposed to light secondary X ray exposure Poor film processing techniques Contaminated developer Excessive development time Developer temperature above recommended value Incorrect developing fluid dilution Expired film Blurred radiographs details not visible Possible causes The patient moved The tubehead moved Radiographs with herringbone marks Some intra oral films are provided with a thin lead foil having a herringbone type mark on its lower side These films may be exposed to radiation on one side only If the film is exposed on the wrong side the lead foil will absorb a large amount of radiation during exposure The result will be a lighter radiograph and the film will show herringbone marks 29 RADIOGRAPH TROUBLESHOOTING PARTIALLY EXPOSED RADIOGRAPHS Possible causes X ray Tubehead positioned incorrectly with respect to the imaging media Low developer fluid level with resulting partial film development Two or more films placed against each other during development CLOUDED RADIOGRAPHS Possible causes Outdated film check expiration date Fil
13. the manual exposure selection mode exposure time is increased by pressing the INCREASE Select Button and decreased by pressing the DECREASE W Select Button The resultant exposure setting is shown on the Digital Display panel in seconds or pulses 2 Press either the INCREASE A or DECREASE W Select Buttons once to manually change the time setting in one increment or decrement Rapid advancement through the entire range of 44 time values is accomplished by keeping the button depressed Release button to set desired exposure time 3 Verify that the desired exposure time value is shown on the Digital Display panel TIME VALUES IN SECONDS 0 02 26 SYSTEM OPERATION EXPOSURE EMISSION PROCEDURE l Introduce the film plate or sensor into the patient s mouth according to the chosen technique bisecting or parallel Move the Tubehead beam limiter near the patient and direct it exactly towards the tooth under examination Arrange the Tubehead with an angle suitable for the required exposure and positioning Select exposure time value via either the automatic or manual exposure time operating mode as follows Set automatic exposure settings for specific patient size media type and tooth type per procedure of page 25 b Set manual exposure settings as necessary per procedure of page 26 Move as far away as the hand control cable allows in a direction opposite to the X ray beam emission while maintaining visual conta
14. 0 1 A2090 A2210 PROVECTA 70 X Ray System with 16 Straight Arm PROVECTA 70 X Ray System with Remote Control Assembly and 16 Straight Arm PROVECTA 70 X Ray System with 24 Straight Arm PROVECTA 70 X Ray System with Remote Control Assembly and 24 Straight Arm PROVECTA 70 X Ray System with 35 Straight Arm PROVECTA 70 X Ray System with Remote Control Assembly and 35 Straight Arm Kit X Ray Remote Control with Wall Mount amp Cover for outside examination room Kit X Ray Remote wall plate exposure switch for outside examination room Mount Plate Assembly 16 center to center exterior dimensions PROVECTA 70 System Parts Identification TUBEHEAD WITH BEAM LIMITING DEVICE X RAY CONTROL PANEL WALL PLATE 16 INCH ON CENTER STRAIGHT ARM OPTIONAL REMOTE X RAY EXPOSURE BUTTON P N A2090 TECHNICAL 5 5 Technical Features Equipment Manvfacturer IEC 60601 1 Classification Degree of protection against ingress of water TECHNICAL SPECIFICATIONS Air Techniques Inc Class with type B applied parts Ordinary Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide Mode of operation FDA 21 CFR Classification Line voltage Line frequency Rated current Power consumption Line Voltage Regulation Apparent Resistance of Supply Mains Circuit Breakers Circuit Board Fuses Preset exposure times Anatomic selection Expos
15. ATTEMPT INTERNAL SERVICE The interior of the Main Assembly is only accessible by removing hardware with tools and should only be opened and serviced by an authorized Air Techniques Dealer Service representative Call your authorized Air Techniques dealer for service Since the interior of the unit contains high voltage components failure to heed this warning may result in equipment damage personal injury and or death 5 SAFETY GENERAL SAFETY PRECAUTIONS This equipment has been designed and manvfactured in compliance with applicable safety requirements moreover this manual provides all the necessary information for correct and safe utilization It provides warnings related to risks associated with the PROVECTA 70 and should be read very carefully Air Techniques Inc shall not be responsible for any use of PROVECTA 70 equipment different from that for which it was been designed any damage to the equipment the operator or the patient caused either by incorrect installation or maintenance not compliant with the procedures contained in the relevant user s and installation manuals provided with the equipment or by incorrect operation techniques any mechanical and or electrical changes made during or after installation different from those specified in the service manual Only qualified service personnel authorized by Air Techniques Inc or its dealers are allowed to perform repairs on the equipment PROTEC
16. Button Verify that the Digital Display panel shows DIG Press the TOOTH SELECT Mode Select Button to confirm selection Verify that the Digital Display panel shows the factory setting of 0 50 Within 10 seconds press the INCREASE A or DECREASE W Select Button to set the desired multiplier ratio with respect to D speed film exposure time 8 Within 1O seconds press the TOOTH SELECT Mode Select Button to select FACTORY RESET Performing factory reset returns all operating values and settings of the PROVECTA 70 to the values set at the factory during manufacture Refer to the PROVECTA 70 Available Preset Exposure Time Setting Combinations table on page 25 for a listing of the factory set default exposure settings Make sure to write down all manually customized settings prior to performing the factory reset sequence Any settings changed during operation will be lost upon completion of reset Reset all operating values and settings by per forming the following procedure 1 Turn the Power Switch OFF 0 and then to ON 1 again 2 When the initials Atl appear in the Digital Display panel press and hold the DECREASE W Select Button until FAC Factory Reset appears in the display 3 Verify that the display shows 0 63 seconds and the SECONDS UPPER MOLAR ADULT AND FILM LED indicator lamps are illuminated The PROVECTA 70 is now reset to operate with the factory set default settings 24 SYSTEM OPERATION SETTING THE EXPOSURE
17. CHILD reduction is 0 5 50 change in exposure time Each time the ADULT CHILD Mode Select Button is pressed the associated LED lamp illuminates and an audible signal sounds to indicate selection DIGITAL FILM MODE SELECT BUTTON A membrane keypad switch that toggles between automatic o exposure settings for specific media types commonly used with the PROVECTA 70 The user can select between D speed film and digital radiology processing When the FILM mode is G selected the system is set to process film at 100 exposure Digital E time of D speed film When set in the DIGITAL mode the RO CRIT Mia exposure time is reduced to values suitable for the use of i D speed film digital phosphor plate scanners such as A T ScanX or sensors An audible signal is sounded each time the DIGITAL FILM Mode Select Button is depressed and the associated LED is illuminated to indicate selection DEFAULT EXPOSURE TIME SETTINGS The following table lists the factory set default setting combinations available by the operation of the TOOTH SELECT ADULT CHILD and FILM DIGITAL Mode Select Buttons PROVECTA 70 Available Preset Exposure Time Setting Combinations Lamm e SECONDS PULSES SECONDS PULSES SECONDS ri 9 o 1 ow s ors 10 rpe a est e ODE m joel fene om e jen KEYPAD OPERATION CUSTOM PROGRAMMING NOTE When power is turned off
18. ISINFECTING CLEANING AND DISINFECTING PROCEDURES In order to ensure proper hygiene and cleaning of the equipment the following procedures must be followed thoroughly Warning Disconnect the branch circuit breaker or unplug from the electrical outlet before performing cleaning procedures Active internal circuits remain connected to main voltage even when main power switch is turned off Exterior Surfaces Wipe the outside surfaces with a paper towel dampened with a disinfectant solution or household non abrasive cleaner DO NOT SPRAY SOLVENTS OR LIQUID DIRECTLY ON THE X RAY UNIT BE CAREFUL NOT TO ALLOW SOLVENTS TO RUN OR DRIP into the PROVECTA 70 This could cause damage to the PROVECTA 70 Allow to air dry before plugging in or turning back on Parts in Contact with the Patient s Skin To ensure proper cleaning of these parts periodic disinfection with an approved EPA Non Corrosive Surface Disinfectant such as Birex se is recommended PRODUCT DESCRIPTION KEY COMPONENTS The PROVECTA 70 is comprised of the following components a STRAIGHT ARM AND SCISSOR ARM Horizontal straight arms are available in 16 24 or 35 inch lengths to meet all possible reach requirements The scissor arm has a double joint enabling linear and upward extension The tubehead remains balanced in all positions b TUBEHEAD WITH BEAM LIMITING DEVICE 70 kVp voltage and 8 mA current reduce exposure times and the amount of radiation absorbed by the pati
19. LINE VOLTAGE 1 50 1 40 1 30 0 90 EXPOSURE CORRECTION FACTOR 0 80 0 70 e e 19 8 e e T INPUT LINE VOLTAGE VOLTS TECHNICAL SPECIFICATIONS TECHNICAL FACTORS MEASURING METHOD kVp The kVp is defined as the average of the peak high voltage after pre heating time and settling The kVp is assessed with a non invasive instrument having a 1 accuracy at the nominal AC input line voltage mA output current is defined as the average value of the current after pre heating time and settling The output current can be measured by attaching a DC voltmeter 0 10 VDC to the connections behind the circular cap on the side of the Tubehead yoke The correlating relationship is 1 volt per mA of tube current t The exposure time is defined as the time measured with a non invasive meter Accuracy is verified by using an Innovision 8000 instrument X RAY TUBE CURVES OCX 70 G Emission characteristics mA anode current anode current filament voltage V filament voltage V X RAY TUBEHEAD KEY COMPONENTS X ray Tubehead Assembly X Ray Beam Axis Guide Focal Point Guide Yoke Angular Alignment Gauge 14 anode current X RAY TUBEHEAD SPECIFICATIONS Filament Characteristics A filament current filament voltage V Loading PEE eo 1 TRITT Sere ip eS
20. PROVECIA 70 Intraoral X Ray System Operator s Manual AIR TECHNIQUES TABLE OF 5 oft merta m E ore ntn Ent hio ens mote 4 Waranty seo ce ear voto tee el oops ope opor 4 On Line Warranty Registration epp s eC de Aog s ae I CAPS 4 Key Feat res i ara Mote qu UI ENERGIE NUAGE ue pe es 5 SotetyNotice 47 24 5862 co totos ducis Dru opere Btn eons Duis dba ora 5 G neral Sater Precautions et etes santosh eo RU vb Ven DS 6 Protection Against Radiation Warning se 6 Symbols fcc Nee er ode inb Veo dr eee RU GE ER ear 6 K owledge ot Warnings ean 7 Waring sesend Mars qu eai maru eot pO a EO 7 Environmental Risk and Disposal aaa 7 Cleaning and Disinfecting 5231 more OA e rede A eb bee oed 8 Product Description x edite I Roe e XO UC Se EUER 8 Key Components us ELI Casio S QUOS i EE S 8 Available CODO de eost i es dy hee ot ice eso ce cd rd 9 PROVECTA 70 System Parts Ideniticgiotie ats bau eben se bees baa 9 Technical Specifications em Sorores te meris en urea ta mcer tuts e 10 Technica Peau c bis orbs x bee rcm arc bes e rb De ES 10 TubebeGd
21. R PRE MOLAR Indicator Lamp ADULT MODE Indicator Lamp ADULT CHILD Mode Select Button DIGITAL Processing Indicator Lamp DIGITAL FILM Mode Select Button PANEL FUNCTIONS O PULSES O SECONDS a CAUTION X RAYS 8mA Warning This x ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed DIGITAL DISPLAY INCREASE Select Button DECREASE Select Button UPPER MOLAR Indicator Lamp UPPER PRE MOLAR UPPER CANINE LOWER MOLAR Indicator Lamp BITE WING amp UPPER INCISOR Indicator Lamp CHILD MODE Indicator Lamp FILM Processing Indicator Lamp ACTIVE X RAY EMISSION CAUTION Indicator Lamp Note All automatic settings are confirmed by the corresponding LED illumination 18 X RAY CONTROL PANEL FUNCTIONS X Ray Control Panel All functional operating control of the PROVECTA 70 X Ray is provided the X ray Control Panel It allows both automatic and manual selections of exposure times The 5 automatic selection functions allow the operator to further choose among 20 pre set times according to the patient type adult or child tooth type and media type digital or film An audible signal sounds during any X ray emission In addition this signal is heard to confirm keypad button selection and when certain errors occur The location of the panel controls and indicators are shown
22. Star VistaCam Omni VistaCam Omni ic4 Air TECHNIQUES Air Techniques Inc 2005 P N A2051 Rev D CORPORATE HEADQUARTERS 70 Cantiague Rock Road P O Box 870 Hicksville New York 11802 0870 1 800 AIR TECH 516 433 7676 FAX 516 433 7684 1 800 247 8324 WESTERN FACILITY 291 Bonnie Lane Suite 101 Corona CA 92880 1 800 822 2899 951 898 8555 951 898 7646 www airtechniques com
23. TION AGAINST RADIATION WARNING Radiation protection is generally prescribed or regulated by local State or Federal regulatory agencies These regulations are beyond the scope of this manual Operation and use of the device should be limited to trained personnel in accordance with local state and federal regulations The film phosphor plate or sensor must be introduced in the patient s mouth either manually or by means of appropriate holders it must never be held by the operator and only the patient may hold it if required b During radiation exposure the operator must not be in contact with the tubehead or the collimator c During radiation exposure the operator must keep at a certain distance from the X ray source at least 6 1 2 feet in a direction opposite to that of X ray emission d During radiation exposure no one other than the operator and the patient must be present in the room e To reduce the unwanted effects of secondary radiation on the patient we suggest using the relevant leaded protection aprons SYMBOLS USED The following symbols are used on the PROVECTA 70 and in this manual AC Connection to Neutral Conductor Connection to Line Conductor Protection Grounding Functional Grounding OFF Equipment not Connected to Power Line ON Equipment Connected to Power Line cs Ot Or X ray Emission SAFETY KNOWLEDGE WARNINGS Users must exercise every precaution to ensure p
24. an D Speed film change the film multiplier ratio as recommended by the film manufacturer prior to use CHANGING FILM MULTIPLIER RATIO SETTING The PROVECTA 70 default FILM setting is 10096 of D speed film exposure time Perform the following procedure to change the default film multiplier ratio setting 1 Turn the Power Switch OFF 0 and ON 1 again 2 When the software version is displayed A press the DIGITAL FILM Mode Select Button to enter the DIGITAL FILM multiplier ratio programming mode 3 When FIL is shown on the Digital Display panel press the TOOTH SELECT Mode Select Button to confirm 22 4 Verify that the Digital Display panel shows the factory setting of 1 00 100 of D speed film exposure time 5 Within 10 seconds press the INCREASE A or DECREASE W Select Button to set the desired multiplier ration with respect to D speed film exposure time 6 Within 10 seconds press the SELECT Mode Select Button to select CHANGING DIGITAL MULTIPLIER RATIO SETTING The PROVECTA 70 default DIGITAL setting is 5096 of the FILM setting D speed film exposure time Perform the following procedure to change the default digital media multiplier ratio settings 1 Turn the Power Switch OFF 0 and ON 1 again 2 When the software version is displayed A press the DIGITAL FILM button to enter the DIGITAL FILM multiplier ratio programming mode When FIL is shown on the Digital Display panel press the INCREASE A Select
25. ay is warranted to be free from defects in material and workmanship from the date of installation for a period of twenty four 24 months Any item returned to our factory in Hicksville New York through an Air Techniques Authorized Dealer will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have been defective Dealer labor shipping and handling charges are not covered by this warranty This warranty does not apply to damage due to shipping misuse careless handling or repairs by other than authorized service personnel Warranty void if installed or serviced by other than Authorized Air Techniques Dealer service personnel Air Techniques Inc is not liable for indirect or consequential damages or loss of any nature in connection with this equipment This warranty is in lieu of all other warranties express or implied No representative or person is author ized to assume for us any liability in connection with the sale of our equipment ON LINE WARRANTY REGISTRATION Quickly and easily register your new Provector 70 on line Just have your product model number and serial number available and log onto www airtechniques com When at the Air Techniques web site select the Warranty Registration link and complete the registration form This on line registration ensures a record for the warranty period and helps Air Techniques keep you informed of product updates and other valuable information
26. ct with the X ray Control panel and the patient Press the X ray Exposure button and hold it depressed throughout the entire exposure The X ray emission condition is indicated by the illumination of the ACTIVE X RAY EMISSION CAUTION Indicator Lamp amp A caution and an audible signal sounding At the end of the exposure the system starts the Tubehead cooling cycle which is 32 times the exposure time The remaining cooling time is shown on the Digital Display panel WARNINGS ls The X ray Exposure button is a Dead Man control The button must be kept depressed throughout the entire exposure Releasing the button before the end of the exposure terminates the emission When this happens the the exposure time display flashes while the other panel indicators extinguish The flashing dis play lasts until either the INCREASE A or DECREASE W button is depressed At that time system opera tion resumes in the tubehead cooling cycle and the display shows remaining cooling time If the combination of low AC line voltage and long exposure time results in a corrected exposure time of more than 4 00 seconds 240 pulses the system will inhibit the exposure In this case the display shows the corrected exposure time Reset the unit by pressing the INCREASE A or DECREASE W button and reduce the exposure setting so as not exceed 4 00 seconds 240 pulses If the AC line voltage reaches minimum maximum operating limits the display shows the foll
27. d time after the exposure is completed or the Tubehead is not Continuously connected to the X ray control If E23 error does not repeat turn the Main Power switch OFF 0 and verify X RAY GENERATION IS TERMINATED BY A that the resistance at the wires going SAFETY TIMER to the Tubehead is nominally 1 5 ohms If not remove the Tubehead and verify the resistance at the Tubehead connector is nominally 1 5 ohms If not replace the Tubehead If yes check the wiring in the fixed and scissor arms Pulsating 33 SYSTEM ERROR MESSAGES ERROR CODE AUDIBLE SOURCE OF ANOMALY ACTION REQUIRED SIGNAL The X ray emission exceeds the Test with Tubehead Simulator programmed time and is Reset power by turning the Main terminated by the hardware Power Switch OFF 0 and ON I If E25 error repeats replace the X ray Control Panel The AC input line voltage is less Biel E tth than 106 Volts when the rad SECY DUROS TO LLL Exposure Switch is pressed i If LLL error repeats check the AC line X RAY GENERATION IS INHIBITED VO tage at the input terminals to the X ray Control Panel timer at nominally 5 5 seconds The AC input line voltage is Press the INCREASE A or greater than 134 Volts when the DECREASE W buttons to reset the Exposure Switch is pressed display If HHH error repeats check the AC X RAY GENERATION IS INHIBITED line voltage at the input terminals to the X ray Control Panel The comb
28. ent The tubehead is equipped with a beam limiting device with a 20 cm 8 inches source to skin distance and a 6 cm 2 3 8 inches beam diameter at the output The tubehead is connected to the arm by means of a rotating contact allowing 360 degree horizontal rotation and 285 degree vertical rotation c X RAY CONTROL PANEL The X ray Control Panel provides mounting support for the straight and scissor arms with attached tubehead system electrical connection points and overall operational control for the PROVECTA 70 The X ray Control Panel is a microprocessor controlled digital timer allowing both manual selection and 5 automatic selections of exposure times Automatic selection further allows the operator to choose among 20 pre set times according to the type of patient adult or child the type of tooth and digital or film type 44 fixed times can be selected manually and ranges from 0 02 seconds to 3 20 seconds The timer automatically compensates for input line voltage variations within a range of 120 V 10 d WALL MOUNT PLATE 16 INCH ON CENTER The Wall Mount Plate is designed to span two 2 by 4 wall studs that are spaced 16 inches apart to provide a secure wall mounted surface for the installation of the PROVECTA 70 PRODUCT DESCRIPTION AVAILABLE OPTIONS The following lists additional configurations options available from Air Techniques Inc for the PROVECTA 70 A2000 16 A2000R 16 A2000 24 A2000R 24 A2000 35 A2000R 35 A229
29. entially advance through the 5 tooth categories until the corresponding indicator lamp illuminates for the desired tooth type selection The resultant exposure setting is shown on the Digital Display in seconds or pulses as required PROVECTA 70 Available Preset Exposure Time Setting Combinations 10 0 of Hororoa wen B MOLAR oan oss T oie o EE we os e fea e sss How os w om ow w Safa at e ef me J fos o os w s Jars s SYSTEM OPERATION Note If using other than D Speed film change the film multiplier ratio as recommended by the film manufacturer prior to use refer to the section on Custom Programming on page 22 The PROVECTA 70 also allows the user to manually set exposure settings to meet the needs of specific patient applications or individual user preference Using the INCREASE A or DECREASE W Select Buttons the manual exposure selection mode enables the user to set exposure time values seconds or pulses as required These user set exposure time values offer 44 specific value settings shown in the tables below Manual Exposure Time Selection Perform the following procedure to set the PROVECTA 70 to operate in the manual exposure selection mode using user set exposure time values l Press either the INCREASE A or DECREASE W Select Buttons to initiate the manual exposure selection mode of operation The Digital Display panel shows the last manually set exposure time Note When set in
30. ersonnel safety and be familiar with the warnings presented throughout this manual and summarized below WARNINGS Federal law restricts this device to sale by or on the order of a dentist This equipment complies with DHHS performance standard 21 CFR Subchapter J This equipment complies with the requirements of IEC 60601 2 7 and UL 60601 1 The X ray Tubehead Assembly P N A2250 complies with IEC 60601 2 28 1993 The equipment must be used in compliance with the procedures contained in this manual and shall not be used for purposes different from those described in this manual To protect the patient from X ray radiation protection accessories such as standard leaded aprons must be used PROVECTA 70 is classified as dental imaging equipment and must therefore be used only under the supervision of qualified dental staff having the necessary knowledge in the field of radiation The user bears legal responsibility related to the possession installation and use of this device Do not remove panels Removing panels would create an electric shock hozard Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide For continued protection against risk of fire replace only with the same type and rating of circuit breakers and fuses Replace components only with authorized replacement parts identified in this manual The RJ11 connector on the internal Circuit Board is for connection
31. he INCREASE A Select Button until PUL pulses appears in the display window Verify that the PULSES indicator illuminates on the control panel 3 Release the INCREASE A Select Button CHANGING DISPLAY TO SECONDS When the PROVECTA 70 was changed to display in pulses and there is a need to change the time value back to seconds perform the following procedure 1 Turn the Power Switch OFF 0 and then to ON 1 again 2 When Atl is shown on the Digital Display panel press and hold the INCREASE A Select Button until SEC seconds appears in the display window Verify that the SECONDS indicator illuminates on the control panel 3 Release the INCREASE A Select Button 22 KEYPAD OPERATION CUSTOM PROGRAMMING CHANGING DEFAULT TOOTH SETTINGS The PROVECTA 70 provides 5 tooth settings and is factory set to start operation in the UPPER MOLAR Perform the following procedure to change one or all of the default tooth settings l 2 3 4 10 Turn the Power Switch OFF 0 and then to ON 1 again When the software version is displayed A will appear along with the software version Press the tooth select button to enter the default tooth setting programming mode When SEL is shown on the Digital Display panel press the INCREASE A Select Button One of the five tooth select indicator lamps will flash and the Digital Display panel will show the current programmed default exposure setting for the selected tooth Wi
32. in the preceding illustration while the function of each is summarized below and described on the following pages DIGITAL DISPLAY Panel A 3 character digital display that shows the selected time value pulses or seconds during operational phases as follows 1 Exposure Preparation Phase displays the selected exposure time in both automatic or manual modes 2 Exposure Phase displays the corrected exposure time adjusted to compensate for variations in the input AC line voltage 3 Cooling Phase displays remaining tubehead cooling time INCREASE Select Button A membrane keypad switch that disables all automatic settings and DECREASE Select Button A membrane keypad switch that disables all automatic settings and allows the user to manually increase exposure time as necessary allows the user to manually decrease exposure time as necessary PULSE TIME EXPOSURE MODE Indicator Lamp A LED indicator lamp that illuminates green to confirm selection of the PULSE time mode for Digital Display operation SECONDS TIME EXPOSURE MODE Indicator Lamp A LED indicator lamp that illuminates green to confirm selection of the SECONDS time mode for Digital Display operation TOOTH SELECT Mode Select Button A membrane keypad switch that allows sequential mode selection of preset exposure time among five different teeth categories as follows UPPER MOLAR Indicator Lamp ee B A LED indicator lamp that illumina
33. ination of low AC input Greater than line voltage and long selected Press the INCREASE or 400 seconds exposure time when the DECREASE W buttons to reset the or Exposure Switch is pressed display 240 pulses Reduce the selected exposure time X RAY GENERATION IS INHIBITED such that the corrected time is less than 4 00 seconds or 240 pulses The Exposure Switch was released prior to the completion of the programmed X ray Press the INCREASE A or iu emission time DECREASE V buttons to reset the display X RAY GENERATION WAS PREMATURELY TERMINATED 34 35 NOTES Air Techniques is a leading manufacturer of fine dental equipment from air and vacuum systems and x ray film processors to an impressive line of new products incorporating the most recent technological advances These new products vital components of the innovative dental practice include intraoral cameras air abrasion systems rapid curing and bleaching systems a digital imaging system which utilizes phosphor plate technolo gy and most recently an intraoral X ray system Air Techniques has been manufacturing quality products for the dental pro fessional since 1962 Acclaim Intraoral Digital Video Camera System AirDent 1 AirDent I CS AirStar Guardian Amalgam Collector A T 2000 XR ARC Light Peri Pro Provecta 70 ScanX SLC STS Vac
34. m accidentally exposed to X ray Film accidentally exposed to other natural or artificial light sources Improper development technique RADIOGRAPHS SHOWING A BLACK LINE This line appears when the film is excessively folded RADIOGRAPHS SHOWING SIGNS OF ELECTROSTATIC ELECTRICITY When a film is compressed too much and the air is dry static electricity may occur which displays black marks RADIOGRAPHS WITH CHEMICAL SPOTS Developer or fixer fluid spattered on the film before development and fixing procedures produces spots on the radiograph such spots are dark when caused by developer fluid light when caused by fixer fluid RADIOGRAPHS WITH EMULSION COMING OFF If the film is kept hot water bath too long e g throughout the whole night or in the developer solution too long the emulsion may become softer and partially come off the film base After develop ment the film will show scratches TYPICAL DEFECTS CAUSED BY INCORRECT POSITIONING Radiographs with elongated or shortened image The main beam is not perpendicular to the bisecting angle formed by the tooth longitudinal axis and the film Radiographs with stretched out tip of tooth Probably caused by excessive film folding or bending inside patient s mouth 30 SYSTEM MAINTENANCE To ensure the safe and effective performance of the PROVECTA 70 follow the proper operation as described by the instructions in this manual Periodic maintenance is deter
35. mined by observations made by the operator and or an Air Techniques Authorized Dealer Field Service Technician Refer to the PROVECTA 70 Intraoral X Ray System Installation amp Service Manual for information pertaining to the installation and servicing of the PROVECTA 70 Intraoral X Ray System WARNING PnE OPERATION CHECKS MUST BE PERFORMED BEFORE THE PROVECTA 70 SYSTEM IF ANY PROBLEMS OR ANOMALIES ARE NOTED DO NOT OPERATE THE UNIT AND CONTACT YOUR AIR TECHNIQUES AUTHORIZED DEALER FOR SERVICE The operator should observe before any operating session that the unit is solidly secure on its wall mounting support he scissor arms are correctly balanced and stable the Tubehead is free from oil residues the Exposure switch cable is not damaged there is no external damage on the X Ray unit he scissor arms are correctly balanced and stable he labels and keypad surface are properly secured SYSTEM ERROR MESSAGES ERROR MESSAGES The PROVECTA 70 has been designed with safety features to protect the patient and operator in case of failure of an electrical component The system automatically checks for errors and will report a malfunction by means of an error code on the X ray Control Panel display These error codes are shown in the table on the following pages The table lists the code identifies the potential source of the problem and recommends the corrective action required by the operator to clear the error Warning When a
36. owing Error Code denotes that the voltage is too low under 106 Volts Error Code HHH denotes that the voltage is too high over 134 Volts Contact your Air Techniques authorized Dealer for service when either code is displayed READING THE EXPOSURE COUNTER I 2 Simultaneously press and hold the INCREASE and DECREASE W Select Buttons for more than 3 seconds When P20 is shown on the Digital Display panel release the INCREASE and DECREASE Select Buttons Press either the INCREASE or DECREASE W Select Button to display the number of exposures in thousands e g 004 4 000 Press the INCREASE A or DECREASE W Select Button to display the number of exposures up to 999 e g 354 The total number of exposures is 4 000 354 4 354 Press the TOOTH SELECT Button to exit the exposure counter mode 27 X RAY SYSTEM REACH AND STORAGE DIMENSIONS Straight Arm Length Options 3500 49 25 74 50 80 00 35 50 825 29 00 7 00 12 75 500 47 44 852 52 25 T E E ee E d 16 00 30 50 55 75 61 25 35 50 29 00 12 75 28 69 52 25 Storage Dimensions weigh Ben X Ray Arm Dimension Options 28 RADIOGRAPH TROUBLESHOOTING CHECKS AND CORRECTION OF POSSIBLE FAULTS IN DENTAL RADIOGRAPHS TYPICAL DEFECTS OF INTRA ORAL RADIOLOGY Light radiographs Possible causes Insufficient exposure to X
37. posure Counter Lue Li CEN 27 X Ray System Reach and Storage Dimensions 28 Radiographic Troubleshooting 2 4 S624 e e RE SS 29 30 System MGINIGNONGe suet aou coed nad d eda ad 3 System Error Messages odas dads ad redd od det ds edd 31 34 frd ode icon fcri or ehe et cio s echa 35 INTRODUCTION Congratulations your new PROVECTA 70 Intraoral X ray has been engineered and manufactured by Air Techniques to provide you with years of reliable service The PROVECTA 70 will create radiographs of excellent quality It will perform equally well using digital or film based imaging media Review and follow the guidelines included in this Operator s Manual to thoroughly become famil iar with the operating and safety procedures This will ensure that your PROVECTA 70 gives you the highest level of service Your PROVECTA 70 Intraoral X Ray model A2000 contains the following components ITEM PART NUMBER Straight Arm 16 Inches Long A2200 one only 24 Inches Long A2140 3 35 Inches Long A2140 2 Scissor Arm 2120 7OkVp 8mA Tubehead 2250 16 Inch on Center Wall Mount Plate A2570 Mounting Hardware 2020 Users Manual 2051 Service Manual 2507 Warranty Card N A Form FDA 2579 N A 16 Inch on Center Wall Plate Template 2691 Single Post Mount Plate Template 2077 WARRANTY The PROVECTA 70 Intraoral X r
38. tes green to confirm selection of preset exposure time for the UPPER MOLAR tooth category A LED indicator lamp that illuminates green to confirm selection of UPPER PRE MOLAR UPPER LOWER MOLAR n A preset exposure time for the UPPER PRE MOLAR UPPER Indicator Lamp 6 LOWER MOLAR tooth category A LED indicator lamp that illuminates green to confirm selection of BITE WING amp UPPER INCISOR Indicator Lamp A preset exposure time for the BITE WING amp UPPER INCISOR tooth category LOWER PRE MOLAR Indicator Lamp ALED indicator lamp that illuminates green to confirm selection of preset exposure time for the LOWER PRE MOLAR tooth category LOWER INCISOR Indicator Lamp ALED indicator lamp that illuminates green to confirm selection of preset exposure time for the LOWER INCISOR tooth category ADULT CHILD Mode Select Button membrane keypad switch that toggles between selection of preset exposure time for ADULT and CHILD exposure modes as follows ADULT MODE Indicator Lamp ALED indicator lamp that illuminates green to confirm selection of preset exposure time for the ADULT exposure mode CHILD MODE Indicator Lamp ALED indicator lamp that illuminates green to confirm selection of preset exposure time for the CHILD exposure mode DIGITAL FILM Mode Select Button ALED indicator lamp that illuminates green to confirm selection of DIGITAL Processing Indica
39. the system retains the last selected manual and preset modes Upon turn on the system resumes operation as set prior to turning off CHANGING DEFAULT SETTINGS TO CUSTOMIZE VALUES The PROVECTA 70 factory default settings are based on D speed film exposure with the digital display set to show time values in seconds These defaults allow automatic operation of the system by the selection of the TOOTH SELECT ADULT CHILD and FILM DIGITAL Mode Select Buttons Although this automatic operation is adequate for most applications the PROVECTA 70 may be customized to meet the needs of a specific application or individual preference Customize one or all of the six default settings listed below as necessary by performing the procedures provided by the following paragraphs When all changes have been accomplished resume operation of the PROVECTA 70 per System Operation instructions provided on page 25 1 Changing Display to Pulses Changing Display to Seconds Changing Default Tooth Setting Changing Default Child Multiplier Ratio Changing Film Multiplier Ratio Settings QE co e eeu Changing Digital Multiplier Ratio Settings CHANGING DISPLAY TO PULSES The PROVECTA 70 is set at the factory to display exposure times in seconds Perform the following procedure to change the Digital Display panel to show time values in pulses 1 Turn the Power Switch OFF 0 and then to ON 1 again 2 When Atl is shown in the Digital Display panel press and hold t
40. thin ten seconds repeatedly press the INCREASE Select Button until the desired tooth category is selected Press the TOOTH SELECT Mode Select Button to confirm selection The Digital Display panel will flash together with the indicator lamp of the selected tooth to indicate the default setting can be changed Within ten seconds press the INCREASE A or DECREASE W Select Button to the desired default setting Within ten seconds press the TOOTH SELECT Mode Select Button to exit the default tooth setting program Repeat steps 1 9 until all desired changes have been made CHANGING DEFAULT CHILD MULTIPLIER RATIO The PROVECTA 70 default Child setting is 50 of the adult setting Perform the following procedure to change the default child mode multiplier ratio setting l 2 Turn the Power Switch OFF 0 and ON 1 again When the software version is displayed A press the ADULT CHILD Mode Select Button to enter the CHILD multiplier ratio programming mode When CHI is shown on the Digital Display panel press the TOOTH SELECT Mode Select Button to confirm selection Verify that the Digital Display panel shows the factory setting of 0 50 Within 10 seconds press the INCREASE A or DECREASE W Select Button to set the desired multiplier ratio with respect to D speed film Within 10 seconds press the TOOTH SELECT Mode Select Button to select 23 KEYPAD OPERATION CUSTOM PROGRAMMING NOTE If using film other th
41. time settings by pressing any Mode Selection Button Film Digital Adult Child Tooth Select TOOTH SELECT MODE SELECT BUTTON A membrane keypad switch located in the center of the tooth symbols allows sequential mode selection of preset default exposure times for each of the five indicated teeth types Each time the TOOTH SELECT Mode Select Button shown below is pressed the associated LED lamp illuminates and an audible signal sounds to indicate selection Default exposure times are based on D speed film The five selectable default preset exposure TOOTH SELECT UPPER MOLAR LED 5 Mode Select Button Factory setting 0 63 sec times correlating to teeth categories are as follows UPPER MOLAR 0 63 seconds i cmn UPPER PRE MOLAR AP UPPER PRE MOLAR UPPER CANINE 0 50 seconds LOWER MOLAR LED Factory setting 0 50 sec BITE WING amp 0 32 seconds BITE WING amp UPPER UPPER INCISOR LOWER INCISOR LED INCISOR LED Factory setting 0 30 sec Factory setting 0 32 sec LOWER PRE MOLAR 0 40 seconds LOWER PRE MOLAR LED LOWER INCISOR 0 30 seconds Factory seting 0 40 sec 20 KEYPAD OPERATION ADULT CHILD MODE SELECT BUTTON A membrane keypad switch that toggles between selection of preset exposure time for ADULT and CHILD exposure modes By depressing the button located between the symbols the selection between ADULT factory setting factory setting for the Adult and Child modes are made with the resulting is 1 00 or 100
42. tor Lamp al mL 5 membrane keypad switch that toggles between selection of preset exposure time for DIGITAL and FILM exposure modes as follows preset exposure time for the DIGITAL processing mode FILM Processing Indicator Lamp c5 A LED indicator lamp that illuminates green to confirm selection of V preset exposure time for the FILM processing mode D speed film ACTIVE X RAY EMISSION CAUTION A LED indicator lamp that illuminates yellow to alert the user that x ray RAYS radiation is present during PROVECTA 70 current operation Indicator Lamp 19 KEYPAD OPERATION INCREASE SELECT BUTTON DECREASE SELECT BUTTON ow The INCREASE and DECREASE Select Buttons are membrane keypad switches that allow the user to disable the automatic operating mode default time settings Adult Child Patient Size Tooth Selection and manually increase or decrease exposure time as necessary to any of the 44 values shown below in seconds or pulses When a manual exposure time setting is desired press the associated button Increase or Decrease once to display the last manually set exposure time Manually change the time setting in one increment or decrement by releasing the button and then pressing once Rapid advancement through the 44 time values is accomplished by keeping the button depressed Release button to set desired exposure time Return to the last automatic mode default
43. ure control Timer Accuracy on the selected Exposure Time This is the value of the operator selected timer setting Timer Accuracy on the Corrected Exposure Time This is the actual time of exposure indicated on the Control Panel during emission and determined by the internal algorithm as a function of the line voltage The range of adjustment is 7396 to 14996 as the AC line voltage changes within the rated voltage range See note on next page X ray Control dimensions Tubehead Features Rated output voltage High voltage circuit type Tubehead current Tubehead Power Total filtration Transformer insulation Cooling Pre heating time Interval between exposure duty cycle Beam Limiting Device Tubehead Assembly dimensions less yoke Continuous operation with intermittent loading Class 120V 10 60Hz 8 2 Amps 0 8 120V lt 3 0 43 ohms 10 Amp 10A 250V Ceramic Fast Acting 0 25 x 1 25 IN and 630mA 250V Fast Acting 5 mm x 20 mm from 0 02 to 3 20 seconds in 44 steps 20 pre set times Microprocessor controlled exposure times with automatic compensation for line voltage variations 5 or 33ms whichever is greater 5 or 33ms whichever is greater 12 5 H x 8 50 W x 4 5 8 D 70 kVp 15 over 108 132 V 108 132 VAC Single phase self rectifying 8 mA 2 mA 0 8 kw 2mm Aluminum equivalent 70 kVp Oil bath Convection 100ms 32 times X ray time

Download Pdf Manuals

image

Related Search

Related Contents

業界トップクラスの 小型・軽量2インチ携帯プリンター  VERTIGO WIRELESS  ホイールローダ LX15/LX20/LX30/LX40  ヒメコガネ  Original - Avanti Online  Des couleurs éclatantes, Des imprimés fous, Des mix auDacieux  第37号  Acer Aspire 581-6456  LOI User Manual and Software Guide v1.1  BESCHRIFTUNGSSYSTEM  

Copyright © All rights reserved.
Failed to retrieve file