Home
        HP1047A RID
         Contents
1.                                                                                                                                                                                                Figure 23 Recorder Control Electronics  Refer to REMOTE CONTROL FUNCTION  ELECTRONICS  The MARKER indicator LED is  ON for the duration of the offset   Control Board eis Board  ae  Recorder a scorder  Contro   Output  Circuits 5 Circuits  O  Con 105  Con 105 Con 115  Con 1101 Key Con 601  Switch    Range   Board     Select   so oO oO   Switeh    Polarity Recorder Marker  Range Switch Board ae  Con 602  O O O  Polarity Recorder Marker  LED LED LED  LED Board  1                54 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information       Summary of Switches and Potentiometers    There are many potentiometers throughout the detector boards  Some of  these are used during normal service calibrations  but many are not     These are pre set at the factory  and should not be misadjusted  The following  information  board by board  is meant to provide information on  potentiometers  switches and certain test points  This information may be  used in the event that a user has misadjusted a potentiometer or changed a  switch position  If  for some reason  a particular circuit cannot be re adjusted  correctly  anew board must be ordered     Any test points which are not mentioned in this manual are used by the  factory in testing and have no practical value in board l
2.     3 Replace the KEY SWITCH board     If no     Is the flush command applying   12 VDC as needed to the  drive wires of the valve     If yes  replace the valve     If no    1 Deactivate the valve from the rear panel  then check  for correct front panel operation    2 Replace the CONTROL board    3 Replace the POWER SUPPLY board        Service Handbook for 1047 Refractive Index Detector   12 2001 117    Troubleshooting Information       Table 11 Flush Valve Problems       Possible Problem Actions       Valve is stuck in the activated position  The manifestation Is the   12 VDC being continuously applied to the drive  of this problem will be that each movement on the wires of the valve   chromatogram will now have a counterpart of the   opposite polarity  For example  a negative solvent front   will be followed by a large positive offset  and all positive   peaks will be immediately followed by negative peaks    Note that the negative peaks will be broad due to the   large post sample cell volume of the detector  In the case   of a chromatogram with peaks close together  it will be   difficult to determine that this is what is actually   happening     If no  replace the valve   If yes     1 Try to deactivate the valve from the rear panel  then  from the front panel    2 Replace the CONTROL board    3 Replace the POWER SUPPLY board           118 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Table 12    Troubleshooting Information    Temperature 
3.    Has the photo diode assembly been replaced recently  It is possible that  the new photo diode assembly requires a re adjustment of the mirror   Refer to    Mirror Adjustment    on page 85 for the mirror adjustment  procedure     Replace the detector left side cover and reconnect the tubes of the flush  valve     If the flush valve is incorrectly plumbed  the detector flow cell may be  destroyed  Make sure the flush valve is plumbed according to Figure 29 on  page 71     10 It is suggested to verify the program of electrical calibrations described in       Electrical Calibrations    on page 93        76    Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Procedures       Power Supply Component Replacement    Refer to Figure 43 on page 138     With the exception of the main power transformer and the POWER SUPPL Y  board  all the components of the power supply are shown here  Refer to  Figure 12 on page 37  Using this wiring diagram  replace the defective  component     If you are replacing an old style rotary voltage selector with the current slide  type  re wire as shown in Figure 12 on page 37        Service Handbook for 1047 Refractive Index Detector   12 2001 77    Procedures       Optical Unit Replacement    The replacement optical unit is available either with AB Motor assembly or  without AB Motor assembly  This procedure assumes that the replacement  optical unit DOES NOT include the AB motor assembly     1 Refer to Figure 41 on page 134
4.   Lens Assembly Screws a    7 Mirror Adjust Glass Access Hole   se  8 Mirror Adjust Glass  9 Optical Balance Glass  10 Lamp Holding Collar Set Screw  1 5 mm HEX   11 Lamp Assembly  12 Lamp Holding Collar   lt    F  PEN  2  8  as A  SY                                                                                                                  12 d 10                                                             Service Handbook for 1047 Refractive Index Detector   12 2001 81    Procedures    Figure 33 Light on Slit                6 When the lamp is correctly adjusted  carefully tighten the holding collar so  that there is no danger of the lamp changing position  Tighten the set screw  in the collar     7 Replace the optical unit insulation foam and top covers  If all else is  correct  mirror adjustment ok  no air in flow cell  etc    display TP1 2 can  be used to determine the light intensity  The average light intensity  TP1  and TP2 both pressed  MUST be between  4 5 and  6 0     8 Perform the complete program of electrical calibrations described in     Electrical Calibrations    on page 93        NOTE It sometimes happens that a new lamp assembly cannot be correctly adjusted   This is due to the fact that the orientation of the bulb in the bulb holder does  not provide even one position where correct adjustment can be achieved  The  bulb is positioned in the holder by three set screws  Loosen two of these  screws  and introduce a small change in the orientation of the bul
5.   M3 6 mm long 0515 0924  Lock washer 2190 0108  Flat washer 3050 0891   24 Nut  M3 0535 0025  Lock washer 2190 0108   25 Standoff  7 mm OD x 8 mm height 0380 0005   26 Lock washer  int tooth 10 5 mm ID  18 mm OD no p n          136 Service Handbook for 1047 Refractive Index Detector   12 2001    Parts Information    Front Panel Assemblies    Figure 42          137    Service Handbook for 1047 Refractive Index Detector   12 2001    Parts Information       Rear Panel Assembly    The following table identifies the individual parts of the Rear Panel Assembly  as shown in Figure 43     Figure 43 Rear Panel Assemblies                         Table 17 Rear Panel Assembly     Description ERC number Part number   1 Ribbon Cable  4  from Output Board CON703 to 01047 61606  Motherboard CON114   2 Output Board  includes two 9 pin terminal strips    ERC75007B 01047 66515   not shown    3 Output Board Holding Clamp 01047 86103   4 Response Time Switch Assembly 01047 60006    5 Ground Terminal Cable Assembly no p n   6 Rear Panel no p n       138 Service Handbook for 1047 Refractive Index Detector   12 2001    Parts Information                      Table 17 Rear Panel Assembly     Description ERC number Part number   7 Fuse Holder Assembly 01047 42301    8 Voltage Selector 01047 81902   g Power Socket 01047 85201   10 Temperature Power Switch 01047 81901   11  Heat Sink 01047 81101    12 Ground Terminal 01047 86101    13  Power Regulator Set 01047 42401   14  Power Regulator Insulator 0104
6.   Remove the detector top cover  Remove  the left side cover  item 15  by removing screws  item 4   Disconnect the  tubes from the IN OUT block  item 17  so that the side cover can be  completely removed  Remove the right side cover  item 3   also by  removing screws  item 4      2 Again referring to Figure 41 on page 134  remove the Preamplifier board  safety cover  item 34   Carefully un solder from the preamplifier board the  three wires  white  green  clear  of the photo diode assembly cable  This  cable can be seen coming from the front of the optical unit beneath the  lamp  Make a note of the position of each wire before un soldering     Disconnect the AB motor assembly cable     4 Disconnect the lamp wires from their positions on the VALVE LAMP  terminal block     5 Disconnect the remaining flow cell capillary tubes  one from the three way  fitting and two from the flush valve     6 At the rear of the optical unit  disconnect the leak drain tubing from the  optical unit leak drain fitting     7 Atthe Temp and Motor Control board  disconnect the optical unit heater  wires from CON201 and the optical unit temp sensor from CON203     8 Carefully raise the detector to gain access to the optical unit holding  screws under the detector  Again refer to Figure 41 on page 134   Completely remove both of the locking screws  item 32   Then  carefully  remove the four nuts which secure the optical unit support feet  item 7      9 Carefully pull the optical unit free of the detecto
7.   SENSOR  about 5 2 KQ cold  about 2 8 KQ at 40  C     Overall circuit performance is best determined by measuring HEATER  voltage power directly at CON201 of the TEMP AND MOTOR CONTROL  board  and by observing the state of the front panel temperature indicator  LED     e TEMP SWITCH off  0 V AC with indicator LED off     e While heating up to setpoint  full power  about 100 V AC   indicator LED  solidly on     e While regulating at the setpoint  about 1 2 power  indicator LED flashing     e Actual temperature greater than setpoint  nearly 0 V AC  indicator LED off        Service Handbook for 1047 Refractive Index Detector   12 2001 43    Hardware Information    Figure 16 Heater Control Electronics    Rear Panel Temp                                                                                                                    ON OFF Switch OR Optical Unit  ne U  oe OR Heater  lemp     Temp  E    from   e WHT Fuse   Sensor  LEG   Transformer  1OOVAC GRAY PINK  2x   secondary    as a BRN BRN i  CON 113  Fl CON 202 CON 201  z    a Temperature 3  YD of  OAMP Regulator aed              Front Panel e IP201    TP203    Temp  Select   nar SS Rog  Driver    108                                                                 cto Z Temp  nN  Switch S Control X  Zz  oO    OF  Mother Temperature and Motor Control Board  Board          44 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    A broken connection from either the TEMPERATURE SELECT SWITC
8.   The WARNING sign denotes a hazard  It calls attention to a procedure   practice or the like  which  if not correctly performed or adhered to   could result in injury or loss of life  Do not proceed beyond a  WARNING sign until the indicated conditions are fully understood and  met        Service Handbook for 1047 Refractive Index Detector   12 2001 11       CAUTION          WARNING          WARNING          WARNING          WARNING          WARNING       General Information    The CA UTION sign denotes a hazard  It calls attention to an operating  procedure  practice or the like  which  if not correctly performed or adhered  to  could result in damage to or destruction of part or all of the equipment  Do  not proceed beyond a WARNING sign until the indicated conditions are fully  understood and met     1047A related warnings    Always take care that the LC system does not have excessive  restriction downstream of the 1047A  Downstream restriction of  greater than 2 bar  29 psi  will damage the 1047A     Always disconnect the line power cord before removing the top cover  of the detector  Line power should also be disconnected before  removing replacing any electrical component     Certain alignment procedures require that the top cover be off while  power is connected to the instrument  Beware of the hazardous  voltages that exist     Always observe proper anti static precautions     Solvents or samples used in detector applications may be very toxic  and or acidic  Take sp
9.   information about the Agilent 1047A  refractive index detector     This manual is available as electronic  version  Adobe Acrobat Reader file   only     
10.   removed  Remove the eight screws indicated in the figure  Carefully pull  the front panel assembly forward to gain easy access to the front panel  components  It will probably be more convenient to disconnect the flush  valve wires from the VALVE LAMP terminal block     Top view with front cover removed                or Zs                                                                                                                                                                                                                                                                            68    Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures    Figure 28 identifies the various components of the front panel assembly  The  figure also indicates those screws which secure a board  and provides  cabling information     Figure 28 Front Panel Assembly                                                                                                                                              Table 4 Legend Front Panel Assembly   Term Description Term Description   1 Digital Display Board 12 Integrator Adjust Assembly   2 Key Switch Board 13 LED Board  2   3 Ribbon Cable  to Motherboard CON115 14 Ribbon Cable  to Motherboard CON109  4 Holding Screws for LED Board  1 15 Holding Screws for LED Board  2   5 LED Board  1 16 Display Switch Board   6 Flush Valve 17 Ribbon Cable  to Motherboard CON111   7 Temperature Select Switch Assembly 18 Holding Screws for Digi
11.  0100 0900      The inlet capillary tube MUST be plumbed as shown  Failure to do so will rupture the flow cell  The  inlet capillary has an ID of 0 25 mm  All other tubes have an ID of 0 8 mm  The inlet capillary is  usually on the bottom  as shown   but always check the ID of the tube to be certain  The flow cell has  a back pressure rating of only 7 bar     C Both the Flush Valve and the 3 Way Fitting can accept SWAGELOK nuts and ferrules  Swage depth is  3 5 mm     D Internally  the 3 Way Fitting is a    T     Internal diameters are large to minimize restriction  DO NOT  substitute a zero dead volume    T    for the 3 Way Fitting     E This is a WIDE ID 1 16 inch union  DO NOT substitute a standard zero dead volume union for this part   excessive restriction might rupture the Flush Valve or the Flow Cell     F The Flush Valve has a back pressure rating of only 2 bar           72 Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures                                                                                  L   _L_L_L_L____E  AB Motor Assembly Replacement and  Adjustment  Replacement  e Remove the detector top cover  Refer to Figure 41 on page 134  and remove  the left side cover  item 15  by removing screws  item 4   Disconnect the  tubes attached to the IN OUT block   e Figure 30 and Figure 31 on page 74 identify the important details of the AB  Motor assembly   Figure 30 AB Motor Assembly Screws    Remove these screws only     4 Refer to Figure
12.  30  Proceed as follows   Disconnect the AB motor assembly cable     6 Remove the four screws  marked    which hold the AB motor assembly to  the Optical Unit  Pull the AB motor assembly away from the Optical Unit  and out of the detector        Service Handbook for 1047 Refractive Index Detector   12 2001 73    Procedures    7 With power still off  re connect the cable of the replacement AB motor  assembly     DO NOT yet connect the replacement AB motor assembly to the optical  unit  Instead  place the motor assembly outside the detector in such away  that the motor will be able to turn freely and safely                                               Figure 31 AB Motor Assembly Parts   ioe  e_ _ i  TA                                 oo m              JH  gt   of                                                                                                                                                                                                                                                                                                                                                                                                             Legend        A Optical Balance Glass Control Shaft  B Sleeve Coupling  C Manual Shaft Turning Knob          74 Service Handbook for 1047 Refractive Index Detector   12 2001    To check for the  correct relationship  between the AB motor  and the optical  balance glass   proceed as follows     Procedures    Adjustment  A Turn on 
13.  3394 6     Verifying the Wander Specification    Wander Specification   1 x 10 8 RIU equal to 2 x 10 8 RIU  equivalent to  20 x 10   RIU      1 For HP 3390 2 3 at ATTENUATION 0     20x 10     RIU    Wander   E F O a  4 2 x10       RIU   mm     Allowable Wander   approx  5 mm for HP 3390 2 3   2 For HP 3394 6 at ATTENUATION 0     20x 10    RIU    Wander   Be Fp aac Nit T  2  1 x 10       RIU   mm     Allowable Wander   approx  10 mm for HP 3394 6        104 Service Handbook for 1047 Refractive Index Detector   12 2001    The Range Dial and  Recorder Outputs    The Recorder Full  Scale Deflection  Setting       NOTE       Performance Verification    For the Recorder Outputs    Verifying 1047A specifications using recorder outputs is a question of  converting the specifications  noise  drift  wander  to an allowable  RECORDER DEFLECTION at a particular 1047A RANGE setting  For this  it  is necessary to understand the relationship between the 1047A RECORDER  OUTPUTS  1 mV and 10 mV   the 1047A RANGE dial and the full scale  deflection setting of the recorder     Assume that the HP 1047A RANGE dial is set to 2 x 10    RIU  At this RANGE  setting  2 x 10 5 RIU of detector response will produce a signal of 10 mV at  the 10 mV RECORDER OUTPUT and a signal of 1 mV at the 1 mV  RECORDER OUTPUT     The least sensitive RANGE setting is 10 mV and 1 mV of signal output for  32 x 10 5 RIU increment of detector response     The most sensitive RANGE setting is 10 mV and 1 mV of signal output
14.  3A Integrator 01040 60105  HP 3394 6A B Integrator or HP 35900A B C 35900 60630  A D Converter  Universal  strip chart recorders usually require this 01037 60105   Unique 1047A Cables  Table 20 Unique 1047A Cables  Figure Description Part number  Figure 45 From 1090 LC EVENTS to 1047A inputs 01047 60202    Figure 46 From 1047A Balance NR output to 1090 LC   01047 60201  Remote Control Port          Service Handbook for 1047 Refractive Index Detector   12 2001 143    Parts Information                                                                            Figure 45 Cable from 1090 LC  ee oo   ae oe a   RG        o  a  eo eee     Se  5 Se  SSS r  le Mi E   Figure 46 Cable from 1047A NR output to 1090 LC                                                                      BLK  RED       144 Service Handbook for 1047 Refractive Index Detector   12 2001    Parts Information                Inlet Outlet Tubes   Table 21 Inlet Outlet Tubes  Description Part number  Insulation for inlet tube no p n  Inlet Capillary Assembly 1 16  0 D   88 cm by 01040 87602  0 12 mm I D   Inlet Capillary Assembly 1 16    0 D   88 cm by 01047 87301  0 25 mm I D     Outlet Tubing Assembly 1 8  wide bore teflon 01040 67602  tube plus 1 16    SST adapter fitting          Service Handbook for 1047 Refractive Index Detector   12 2001 145    Parts Information       146 Service Handbook for 1047 Refractive Index Detector   12 2001    oe Agilent Technologies    In This Book    This manual contains technical
15.  Amplifier Board    Figure 24 is a simplified diagram describing the potentiometers on the  AMPLIFIER BOARD  This diagram describes the current board  8 pots   If  the old board  14 pots  is present  it should be discarded and a new board  ordered  On the board  test points extend out to the rear for convenience     e VR612 613   See    Signal Path Overview    on page 21 and    Electrical  Calibrations    on page 93 for details     e The remaining six potentiometers should be adjusted together  The  detector should be equilibrated at 40  C for at least 90 minutes  All DC  measurements are with respect to chassis GND     Amplifier Board Potentiometers    VR614                                                    VR613 Pos TP605  Polarity  Amp  Span P Polarity ox to  Stag YR605 Control Output  Neg     Int t Polarity  gt          _    __  Me eae Amp TP604  Path  Main ran  stag VR607  Recorder  S      e    VR612 knee  Span VR61D Polarity to  a  Control Output  ge   Polarity  Amp TP608  VR608       Service Handbook for 1047 Refractive Index Detector   12 2001 57    Hardware Information    When the detector is equilibrated  adjust the potentiometers in the order  and  by the procedures  which follow     1    NJ OD OF Pp Ww    Set the front panel FINE ZERO dial to it   s mid position   five turns  Set the  front panel INTEGRATOR dial fully clockwise     Set the Response Time switch  back panal  to its mid position  0 8 s      On the ELECTRICAL AUTOBALANCE BOARD  set both poles of SW40
16.  ERC75011A    ERC751  ERC751  ERC751       04A  05A  06A    01040 22301  0100 0900  01047 04110   01047 60009   2110 0002  01047 66508  01047 66503  01047 66504  01047 66501  01047 66505  01047 66506  01047 66507  01047 24103    see Table 17  on page 138    0515 0761  01047 43701   0515 0094   3050 0225  01047 04101   01047 44101   01047 03701  01047 23201  0515 0915  2190 0586             132 Service Handbook for 1047 Refractive Index Detector   12 2001    Table 15    Parts Information          ERC number Part number       Major Assemblies     Description   38 Screw pan pozi  M4 8 mm long  Lock washer  Flat washer   39 Screw pan pozi  M3 6 mm long  Lock washer  Flat washer   40 Screw pan pozi  M4 10 mm long  Lock washer  Flat washer   41 Screw pan pozi  M4 12 mm long  Lock washer   42 Nut hex M6  Flat washer   43 Screw hex  M4 4 mm long  2 mm key   44 Standoff  7 mm dia x 7 mm height    Extender board for items 22  23  24  25  26  and 27  not shown in figure     0515 0910  2190 0586  3050 0893  0515 0924  2190 0108  3050 0891  0515 0926  2190 0586  3050 0893  0515 0909  2190 0586  0535 0038   3050 0225  0515 0529        no p n  01047 66516          Service Handbook for 1047 Refractive Index Detector   12 2001    133    Parts Information    Major Assemblies    Figure 41                   Service Handbook for 1047 Refractive Index Detector   12 2001    134    Table 16    Parts Information       Front Panel Assembly    The following table identifies the individual parts of the 
17.  Service Handbook for 1047 Refractive Index Detector   12 2001    Contents    Hardware Troubleshooting 111  Detector Turn On Problems 111  Light Intensity Problems 111  Lamp Turn Problems 113  Limit Alarm  AB Motor Drive and Balance Problems 113  Flush Valve Problems 116  Temperature Control Problems 119  Flow Cell Problems 121  Loss of Sensitivity 121  Chromatographic Troubleshooting Guide 122  General Characteristics of RI Detector 122  Optimizing your 1047A 123  Solvent Delivery System 123  Inlet Tubes 123  Outlet Tubes 123  Solvents 123  Troubleshooting the Chromatogram 125  Troubleshooting an Instable Baseline 126    6 Parts Information    This chapter provides parts information for the 1047 Refrac   tive Index Detector 127    Major Components 129   Major Assemblies 131   Front Panel Assembly 135   Rear Panel Assembly 138   Optical Unit 140   Cables 143  General 143  Signal Cables for Hewlett Packard Data Handling Devices 143  Unique 1047A Cables 143       Service Handbook for 1047 Refractive Index Detector   12 2001    Contents    Inlet Outlet Tubes 145       8 Service Handbook for 1047 Refractive Index Detector   12 2001       General Information    This chapter provides general information for the  1047 Refractive Index Detector       General Information       Safety Summary    The following general safety precautions must be observed during all phases  of operation  service  and repair of this instrument  Failure to comply with  these precautions or with specific war
18.  attenuation  then calculate the signal required for 1 mm deflection at  that attenuation with this formula          _ SignalFS  Signal   NTFS  where   signal signal for 1 mm deflection  signal FS signal for full scale deflection  INT FS integrator full scale deflection in mm          Service Handbook for 1047 Refractive Index Detector   12 2001 101    Example of RIU mm  Scaling Factor  Calculation    Table 6    Performance Verification    The 1047A integrator output is calibrated such that 32 mV   1 x 10  5 RIU   Calculate the RIU mm SCALING FACTOR for the integrator at the specific  integrator attenuation     Assume that the integrator needs a signal input of 32 mV for a full scale  deflection at A TTENUA TION 5  This means a 1047A signal of 1 x 10  gt  RIU  will cause a FULL SCALE DEFLECTION of the integrator at  ATTENUATION 5  Assume that the integrator has a full scale deflection of  150 mm     1x 10   RIU     8  oam   6 6x10 RIU    Signal      At attenuation 5  the scaling factor for this integrator is 1 x 10  gt  RIU FS  or  6 6 x 108 RIU mm     RIU Scaling Factors for HP Integrators    For the table below the full scale deflection on the HP 3390 2 3 is a 75 mm  and on the HP 3394 6 is 150 mm        RIU Scaling Factors for HP Integrators          ATTN 339xA RIU FS 3390 2 3 RIU mm 3394 6 RIU mm   3 1  00x10   13x100 65x10   4 197x108 26x10 1 3x 10710   3 391x108 52x107 2 6 x107   2 781x10 10x10  5 2x 101   1 1 56x107 28x109 14x10   0 3 13x107 42x10  21x109  1 6 25 x107 
19.  for  1 64 x 10   RIU increment of detector response     Most recorders have a variable FULL SCALE DEFLECTION setting  1 mV   10 mV and 1 V are most popular     Ideally  the RECORDER FULL SCALE DEFLECTION setting should be  matched to the detector recorder output  For example  if the 10 mV  RECORDER OUTPUT of the1047A is connected to a recorder  the recorder   s  FULL SCALE DEFLECTION should be set to 10 mV  In this wa y  the detector  response needed for a FULL SCALE DEFLECTION of the recorder is  determined by the detector RANGE setting  Peak height can be converted to  detector response by it   s percentage of FULL SCALE DEFLECTION     Example     The 1047A 1 mV RECORDER OUTPUT is connected to the recorder  The  recorder FULL SCALE DEFLECTION is set for 1 mV  The 1047A RANGE dial  is set to 16 x 10 5 RIU  The chromatogram contains one peak  which is 50   FULL SCALE DEFLECTION  This peak then represents a detector response  of 8x 10    RIU     If it is not possible to properly match the detector recorder output and the  recorder full scale deflection setting  then the error will have to be taken into  account during quantization or specifications verification        Service Handbook for 1047 Refractive Index Detector   12 2001 105    Performance Verification    Calculating Recorder The formula for calculating allowable deflection in mm at a specific detector       Deflection RANGE setting is   Recdefl   SPES x RecFS  Range   where    Recdef recorder deflection in mm   Spec 
20.  minutes prevents any outgassing caused by mixing  but is not  usually long enough to change the solvent composition     e The effectiveness of continuously degassing the mobile phase during the  analysis depends on the solvents used and the length of the analysis  Continuous degassing will prevent a degassed mobile phase from  re absorbing air  but there are two side effects to consider     1 Long term retention time changes with complex mobile phases  If the mobile phase comprises a mixture of solvents and one of these  solvents is more volatile than the others  then continuous degassing can  cause changes in retention times related to changing elution strength     2 Long term drift   If the solvent trapped in the reference compartment  may begin to degas at a fixed rate  This will cause a continuously  increasing refractive index difference between the solvent in the  reference compartment and the continuously degassed  flowing solvent  in the sample compartment  This effect will cause long term drift and is  very noticeable when using  for example  THF     The overall performance of the 1047A will vary depending on the solvent  used  Solvent properties such as viscosity and volatility are important factors  to baseline stability        124 Service Handbook for 1047 Refractive Index Detector   12 2001    Troubleshooting Information    Troubleshooting the Chromatogram    1 Rule number one  Turn off the flow  and determine what portion of the  problem  noise  drift  wander e
21.  motor can be driven to it   s true center  mid way  position by  simultaneously pressing the  COURSE and  COURSE buttons on the front  panel  The LIMIT alarm indicator flashes during this movement  This  centering of the motor does NOT cause a balance to occur  If the relationship  between the AB motor and the optical unit balance glass control shaft is  correct  this centering of the AB motor will result in a good optical null  That  is  TP1 and TP2 will be similar in value  A BALANCE command will then  correctly establish the baseline        Service Handbook for 1047 Refractive Index Detector   12 2001 113    Troubleshooting Information       Table 10    Limit alarm  AB motor and balance problems       Possible Problem    Actions       Limit alarm is on  but the AB motor does not respond to  the centering command     OR    The AB motor does not respond to movement commands  generally     TP1 TP2 values do not respond to the movement of the  AB motor    Note  This symptom often occurs if the detector is  transported without tightening the locking screws     The AB motor can be correctly centered from the LIMIT  ALARM state  but the LIMIT ALARM error occurs after a  BALANCE is done     BALANCE command initiated from the rear panel remote  input does not work     BALANCE command initiated from the front panel does  not work     1 Check that all cables are connected correctly and  tight  check that all boards are seated securely   2 Replace the AB MOTOR assembly   3 Try the fo
22.  the REC display  and use  the FINE ZERO dial to make the REC display read 0 00     4 Select the INT display  Adjust VR701 so that the INT display reads 0 00     The FINE ZERO dial should now be able to simultaneously zero the REC  and INT displays     The INTEGRATOR dial should add  10 mV to the existing INT display        Service Handbook for 1047 Refractive Index Detector   12 2001 97    Procedures    Verify that the detector has adequate  RANGE    1 For proper operation  the detector must have adequate range in both the    and   directions     Refer to    AB Motor Assembly Replacement and Adjustment    on page 73   The procedure for checking the relationship between the AB Motor and the  optical balance glass actually does make sure that the detector has adequate  range in both directions        NOTE If the detector does not have adequate range in both directions  the most likely  cause is incorrect adjustment of the mirror   particularly  incorrect use of the  mirror adjustment glass  Refer to    Mirror Adjustment    on page 85           98 Service Handbook for 1047 Refractive Index Detector   12 2001       Performance Verification    This chapter provides information on  performance verification for the 1047 Refractive  Index Detector       Performance Verification    This section is meant to assist in determining the actual performance of the  1047A     Specification Verification  e For the integrator output  LJ calculating integrator deflection  M verifying the noise 
23.  turns      BALANCE the detector  then use the FINE ZERO dial to make the INT  display read 0 00     Press and hold the   COURSE ZERO button     The AB motor should be moving in the counter clockwise direction and  the display reading must be increasing in the positive direction  Make  sure the sleeve coupling is firmly holding both the AB motor shaft and the  optical balance glass control shaft     Watch the INT display  The display must reach 2 volts  and go blank   BEFORE the LIMIT ALARM indicator comes on     While the detector is in the LIMIT ALARM state  center the AB motor   step C above   When the motor has reached it   s center position   BALANCE the detector  The INT display should return to the value  established in step E 2 above        Service Handbook for 1047 Refractive Index Detector   12 2001 75    If this test is NOT  successful        NOTE       Procedures    9    Press and hold the   COURSE ZERO button  The AB motor should be  moving in the clockwise direction and the display reading must be  increasing in the negative direction     Again watch the INT display  The display must reach  2 volts  and go  blank  BEFORE the LIMIT ALARM indicator comes on     Finally  again center the AB motor and balance the detector   Repeat the AB motor adjustment procedure  or    Has the optical unit mirror been adjusted recently  Incorrect use of the  mirror adjust glass may be a cause  Refer to    Mirror Adjustment    on  page 85 for the mirror adjustment procedure     or 
24.  unit     2 Refer to Figure 44 on page 140  Remove the optical unit top cover  item  13   the insulation  item 9  and the optics cover  item 12      3 To identify the mirror  refer to Figure 32 on page 81 preceding   Do not touch the pitch screws     4 Remove the two mirror holding screws  BE CAREFUL NOT TO TOUCH  THE SURF ACE OF THE MIRROR  Carefully remove the mirror and put it  in a safe place  If the detector is turned on now  the LIGHT INTENSITY  ERROR will come on     5 Carefully install the replacement mirror and insert the two mirror holding  screws  Do not tighten the mirror holding screws at this time   instead   proceed with the mirror adjustment procedure described in the next steps     Mirror Adjustment    This is a tedious adjustment that requires patience  The mirror must be  correctly adjusted to ensure proper operation of the detector     1 Activate the flush valve and flush the flow cell with pure isopropanol at a  flow rate of between 4 and 5 ml min  This is to make sure that there are no  air bubbles in the flow cell  Remember  since the mirror is now  mis adjusted  display TP1 2 are not functional     Make sure that there is no restriction downstream of the detector  The flow  cell can be destroyed by excessive back pressure        Service Handbook for 1047 Refractive Index Detector   12 2001 85    Procedures    2    Fill both halves of the flow cell with pure  well degassed HPLC grade  water  Set the detector temperature to 40  C and allow the detector t
25.  width 0 04        Specifications of 1047A RID       Signals    Noise    Wander  Drift    Response Time  Flow Cell    Recommended pH range    Temperature Control    Display    Analog Outputs    Communications    Environment    RI range 1 to 1 75  measurement range    1 64 x 10    DELTA RIU to 32 x 10       DELTA RIU  full  scale       lt  2 5x 109 DELTA RIU peak to peak  flowing water  at 1 ml min  0 8 second response time  10 90        lt   1 x10 DELTA RIU     lt 2 5x 107 DELTA RIU hour  after warm up   0 3  0 8  or 5 seconds  10 90     7 ul volume with 7 bar  103 psi  maximum pressure     2 3 to 9 5 standard  Solvents with a pH between 1 0  and 2 3 must not contain acids which attack stainless  steel     30  35  40  45  50  C     Single line  4 1 2 digit display with real time display  of output voltage  recorder or integrator     One output integrator output 2 V maximum     32 mV 10 DELTA RIU   two recorder outputs   0 1 mV and 0 10 mV      Four control inputs  marker  flush  balance not ready      Temperature range  0 to 40  C       Service Handbook for 1047 Refractive Index Detector   12 2001 15    General Information          Table 1 Specifications of 1047A RID  Humidity  lt  85   non condensing   Power Requirements Line voltage  100  120  220  240 V AC   5    10      Line frequency  48 66 Hz   Power consumption  120 VA  410 BTU hour   Dimensions Height  150 mm  5 9 inch    Width  430 mm  17 0 inch    Depth  370 mm  14 6 inch    Weight  19 kg  41 6 Ib            16 Serv
26.  z O REF     9             No electrical calibrations are needed  However  the polarity of the  Preamplifier board output should be checked  as follows     Turn on the detector  turn on should be normal  Then  BALANCE the  detector        66    Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Procedures    Select INT at the display  The voltage should represent a normal balance   based on the position of the front panel FINE ZERO and INTEGRA TOR  dials     Press and hold the   COURSE ZERO button  The display voltage must  increase in the positive direction  If the voltage increases in the negative  direction  re check the position of the white and green wires on the  Preamplifier board     Output Board    DO NOT TOUCH VR701     Refer to Figure 43 on page 138  and proceed as follows     1    Disconnect the large ribbon cable  item 1  from the OUTPUT board  the  other end of this cable goes to Motherboard CON114      Disconnect the 4 wire ribbon cable from the OUTPUT board  this cable is  part of the Response Time Switch assembly  item 4      Working from the outboard side of the rear panel  remove the board  holding clamps  item 3   These clamps are located on each side of both the  INPUTS and OUTPUTS terminal blocks     The OUTPUT board may now be removed  To remove the board  It will be  necessary to bend the rear panel  item 6  slightly outward   You may also  decide that it is more convenient to remove the six boards from the card   cage     I
27. 1 to  ON     Using VR604  adjust TP604  visible  to measure 0 VDC  1 mV   Using VR605  adjust TP605  visible  to measure 0 VDC  1 mV   Using VR607  adjust TP607  under VR605  to measure 0 VDC  1 mV   Using VR608  adjust TP608  under VR607  to measure 0 VDC  1 mV     On the ELECTRICAL AUTOBALANCE BOARD  set both poles of SW401 to  OFF     Adjust VR615 as follows     Set the front panel RANGE dial to 32 and make sure the front panel  recorder POLARITY button is OFF  set to positive recorder polarity     Press BALANCE     Measure TP607  under VR605   Use the front panel  COURSE ZERO  button to make TP607 measure  2 5 VDC     Now  measure TP608  under VR607   Using VR615  adjust TP608 to  measure  2 5 VDC     Test this calibration by using the front panel recorder POLARITY button  and observing the front panel REC display  The  REC value should equal  the  REC value  If needed  use VR615 to fine tune the  REC value     Adjust VR614 as follows     Make sure the front panel integrator POLARITY leds indicates positive  integrator polarity  integrator polarity is switchable only at the rear  panel      The front panel FINE ZERO and INTEGRATOR dials should be set as  described in step 1 above     At the front panel  select the INT display     Press BALANCE  Use the FINE ZERO and INTEGRATOR dials to make  the INT display read 0 00 and press BALANCE again        58    Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Measure TP604  visible   Use th
28. 1047A Refractive Index  Detector    Service Handbook                                              o FEC   BEOORDE  REOOROER REFERENCE MTO REORDER om CHAE       aano     L a a om mogo                                                                FINE ZERO    TEMPERAURE    C  RANGE  x10 RIU F S           i  HP 1047A OC  C     Q   Teorann  RI DETECTOR    O UMIT ALARM     LIGHT THROUGHPUT ERROR                oe Agilent Technologies          Copyright Agilent  Technologies 2001    All rights reserved   Reproduction  adaption   or translation without  prior written permission  is prohibited  except as  allowed under the  copyright laws     Part No  NONE  12 2001    Printed in Germany    Warranty    The information  contained in this  document is subject to  change without notice     Agilent Technologies  makes no warranty of  any kind with regard to  this material   including  but not  limited to  the implied  warranties or  merchantability and  fitness for a particular    purpose     Agilent Technologies  shall not be liable for  errors contained herein  or for incidental or  consequential damages  in connection with the  furnishing  performance   or use of this material     IMPORTANT NOTE    This version of the 1047  service manual includes  all sections from the  original manual 01047   90100 edition 1  May  1991      Part numbers have been  updated as of 12 2001   Some part numbers are  marked bold with     These may not be  available  Contact your  local Agilent support  off
29. 7 85401    15 Screw pan pozi  M3 12mm long 0515 0911  Lock washer 2190 0108   16 Screw pan pozi  M3 6 mm long 0515 0924  Lock washer 2190 0108   17 Screw pan pozi  M3 16 mm long 0515 0920  Lock washer 2190 0108   18 Screw flat pozi  M3 8 mm long 0515 0907   19 Screw pan pozi  M4 8 mm long 0515 0910  Lock washer 2190 0586  Flat washer 3050 0893   20 Nut 2950 0072  Washer  int tooth 2190 0679    21 Nut 0535 0025  Flat washer 3050 0002   22 Screw pan pozi  M3 6 mm long 0515 0924  Lock washer 2190 0108   23 Nut  M3 0535 0004         Heat Sink Assembly  01047 60005  consisting of items 11  13 and 14       Service Handbook for 1047 Refractive Index Detector   12 2001 139    Parts Information       Optical Unit    The following table identifies the individual parts of the Optical Unit as  shown in Figure 44     Figure 44 Optical Unit             140 Service Handbook for 1047 Refractive Index Detector   12 2001    Table 18    Parts Information                Optical Unit    Description Part number  N A Complete Optical Unit see page 129  1 Lamp Holding Collar 01047 22501  2 Lamp Assembly 01037 60002  3 Lamp Support Plate Assembly 01047 04104  4 Lens Slit Assembly 01047 60014  5 Mirror Assembly 01047 60015  6 Cell Assembly 01037 60003  7 Photo diode Lens Assembly 01047 60013  8 Insulation  20 mm thick  in 2 feet by 2 feet pieces  01047 45403   9 Insulation  15 mm thick  in 2 feet by 2 feet pieces  01047 45402   10 Insulation  10 mm thick  in 2 feet by 2 feet pieces  01047 45402   11 O
30. 8 3 x109 42x10   2 1 25 x10 17x108 8 5 x10  3 250x106 3 3x10 17x108  4 5 00 x10    6 7 x10 2 34x108                102 Service Handbook for 1047 Refractive Index Detector   12 2001    Table 6    Performance Verification       RIU Scaling Factors for HP Integrators          ATTN 339xA RIU FS 3390 2 3 RIU mm 3394 6 RIU mm  5 1 00 x10    13x107 6 5 x 10 2  6 2 00x 10   26x107 14x107  7 4 00 x10    53x107 27x107  8 8 00 x10     13x10    53x107  3 1 60x10 21x108 10x10     10 3 20x10  43x10    22x10                   Verifying the Noise Specification  Noise Specification    2 5 x 10     RIU P P  equivalent to 5 x 10  RIUpp    2 For HP 3390 2 3 at ATTENUA TION   2     5x10   RIUpp    Noise   Se Ob ee eI ee  1x10   RIU   mm     Allowable Noise at attenuation   2   5 mm peak to peak  pp  for  HP 3390 2 3     3 For HP 3394 6 at ATTENUATION   2     5x10     RIUpp    Noises aee  0  5x 10       RIU   mm     Allowable Noise at attenuation   2   10 mm P P for HP 3394 6        Service Handbook for 1047 Refractive Index Detector   12 2001 103    Performance Verification    Verifying the Drift Specification  Drift Specification  2 5 x 10 7 RIU hr equal to 250 x 10    RIU hr   1 For HP 3390 2 3 at ATTENUATION 1     250 x 10     RIU    Drift          2  8  3 x 10       RIU    mm     Allowable drift at attenuation 1   30 mm hr for HP 3390 2 3   2 For HP 3394 6 at ATTENUATION 1     250 x 10     RIU    Drift       lt  L  4 2 x 10       RIU   mm     Allowable drift at attenuation 1   60 mm hr for HP
31. 92  Optical Unit Heater Temperature Sensor 92  Electrical Calibrations 93  NULL the detector output 93  Calibrate the LIGHT INTENSITY ERROR THRESHOLD 94  Calibrate the Integrator Output Span 95  Calibrate the Recorder Output Span 95  Calibrate the Autobalance Decision Threshold 96  Verify the Operation of the Baseline Controls and Autobalance 96       Service Handbook for 1047 Refractive Index Detector   12 2001 5    Contents    Calibration of OUTPUT board VR701 97  Verify that the detector has adequate  RANGE 98    4 Performance Verification    This chapter provides information on performance verifica   tion for the 1047 Refractive Index Detector 99    1047A Specifications Verification 101  For the Integrator Output 101  Calculating Allowable Integrator Deflection 101  Example of RIU mm Scaling Factor Calculation 102  RIU Scaling Factors for HP Integrators 102  Verifying the Noise Specification 103  Verifying the Drift Specification 104  Verifying the Wander Specification 104  For the Recorder Outputs 105  The Range Dial and Recorder Outputs 105  The Recorder Full Scale Deflection Setting 105  Calculating Recorder Deflection 106  Verifying the Noise Specification 106  Verifying the Drift Specification 106  Verifying the Wander Specification 107  For the ChemStation 107  Example  Calculating the noise 107  Example  Calculating the Drift 107    5 Troubleshooting Information    This chapter provides troubleshooting information for the 1047  Refractive Index Detector 109       6
32. Board 56  Output Board 56  Amplifier Board 57  Digital Display Switch Board 60  Digital Display Board 60    3 Procedures    This chapter provides procedures for the 1047 Refractive Index  Detector 61    General Information   Read this first  62  Board and Front Panel Controls Replacement 64  Temperature and Motor Control Board 64  Power Supply Board 64  Optical Auto Balance Board 64  Auto Balance Board 65  Control Board 65  Amplifier Board 65  Preamplifier Board 65  Output Board 67  Motherboard 68       4 Service Handbook for 1047 Refractive Index Detector   12 2001    Contents    Boards and Controls located in the front panel 68  TEMPERATURE SELECT SWITCH assembly  RANGE SWITCH BOARD  assembly  FINE ZERO ADJUST assembly or INTEGRATOR ADJUST  assembly 70  LED BOARD 1  KEY SWITCH BOARD  LED BOARD 2  DIGITAL DISPLAY  BOARD or DISPLAY SWITCH BOARD 70  Flush Valve Replacement 71  AB Motor Assembly Replacement and Adjustment 73  Replacement 73  Adjustment 75  To check for the correct relationship between the AB motor and the optical balance  glass  proceed as follows  75  If this test is NOT successful  76  Power Supply Component Replacement 77  Optical Unit Replacement 78  Inside the Optical Unit 80  Lamp Replacement Intensity Adjustment 80  Flow Cell Replacement 83  Mirror Replacement 85  Mirror Adjustment 85  The basic task 88  Helpful hints 88  If it seems not possible to adjust the mirror 89  Photo Diode Replacement Adjustment 90  Replacement 90  Adjustments and Calibrations 
33. Calibrations    on page 93        Service Handbook for 1047 Refractive Index Detector   12 2001 59    Hardware Information    Digital Display Switch Board  On this board  the INT display and REC display offsets are calibrated     Proceed as follows     1    Set the front panel FINE ZERO dial to it   s mid position  5 turns  Set the  front panel INTEGRATOR dial fully clockwise  Both INTEGRATOR and  RECORDER polarity should be set to positive     With the detector on  disconnect one of the lamp wires from the VALVE  LAMP terminal block  This will cause a LIGHT INTENSITY ERROR     At the front panel  select the INT display and press BALANCE  Using  VR1302  closest potentiometer to the optical unit   adjust the INT display  to read 0 VDC  1 mV     At the front panel  select the REC display and press BALANCE  Using  VR1301  closest potentiometer to the display  adjust the REC display to  read 0 VDC  0 02 mV     Re connect the lamp wire  The LIGHT INTENSITY ERROR should clear  and the baseline controls should function normally     Digital Display Board    On this board  the display span is calibrated  Proceed as follows     1    Set the front panel FINE ZERO dial to it   s mid position 5 turns  Set the front  panel INTEGRATOR dial fully clockwise  Both INTEGRATOR and  RECORDER polarity should be set to positive     At the front panel  select the INT display and BALANCE     Observe the DISPLAY SWITCH BOARD  The circuit side of the switch  board is viewed from the top  There is only 
34. Control Problems    When the circuit is working normally  the front panel TEMPERATURE  indicator led is     e Solidly ON when both rear panel switches are first turned on     e FLASHING when the actual temperature  as measured by the Optical Unit  temp sensor  equals the temperature setpoint     e OFF if the actual temperature exceeds the temperature setpoint     Moderately sensitive work can be done after 90 minutes of temperature  equilibration  However  sensitive work  or specification verification  requires  at least 24 hours of temperature equilibration        Temperature Control Problems       Possible Problem    Possible Causes and Actions       The Temperature Control circuit seems to be functioning 1 The ambient environment is too erratic for good  normally  but the no flow detector baseline is temperature control  Consult page 3 7 of the 1047A    unacceptable     Operator s Handbook  Make sure the detector  environment is acceptable    2 The detector temperature setting is not high enough   Make sure that the temperature setting is at least  10  C above ambient temperature    3 The TEMP AND MOTOR CONTROL board is defective   Replace the board    4 The Optical Unit heater or temperature sensor is  defective  Replace the Optical Unit     The front panel TEMPERATURE indicator led does not turn 1 Check the following connections for tightness   on when the rear panel TEMP switch is turned on  a CON203 on the TEMP AND MOTOR CONTROL board     This is the temp sensor connec
35. ER SUPPLY board   OPTICAL AUTOBALANCE board   ELECTRICAL AUTOBALANCE board   CONTROL board   AMPLIFIER board   To DISPLAY SELECT SWITCH board  CON2    To OUTPUT board  CON703    To front panel INTEGRATOR dial          ro mMm  mMm Oo Dd wo YS    v O FTA C       38 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information       Table 3 Cables termination       Term Description       Q To front panel TEMPERATURE select switch   R To RANGE SWITCH board  CON1001    S To LED BOARD  2  CON903    T To rear panel TEMP switch                                                                                                                                                                                Figure 13 Motherboard Layout  A A A  bo CON CON con   con CON CON b  CON  01 102 119 103 104 105 CON  118 108  57 1 ON 1  o  M B1 M BI Al B1 Al B1 Al B1 Al g1 1 G    6   D M    CON  10    G H   J K L R  AA  12     1 CON  115  5 E CON  ar 109  ON 1  06  1 N ws  F e  i NFT  5 SD    A  1 i 1  A A A       Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information       AB Motor Control    Figure 14 shows the relationship of the Optical Balance Motor to the Optical  Unit  The motor shaft and the optical balance glass control shaft are joined  together by a sleeve coupling  The motor can be commanded to go to it   s  mechanical center position  Ideally  the mechanical center position of the of  the motor should cause the optical balance glass t
36. Front Panel  Assembly as shown in Figure 42 on page 137     Front Panel Assembly       Front Panel Assembly               Description ERC number Part number  1 Rotary Knob 01047 47401  2 Set Screw M3 hex  3 mm long 2 mm key 0515 0154  3 Front Panel Cosmetic Plug 01047 47601  4 Front Panel  only  01047 00201  5 Solenoid  flush  Valve 01047 60008  Fasteners for solenoid valve Same as item 22  6 Temperature Select Switch Assembly 01047 60001  7 LED Board 1 01047 66510  8 Push Button Cap 01047 47403  9 Key Switch Board ERC75006B 01047 66509  10 Ribbon Cable  1  from Key Switch Board 01047 61603  CON601 to Motherboard CON115  11 Ribbon Cable  3  from Range Switch Board 01047 61605  CON1001 to Motherboard CON116  12 Range Switch board Assembly 01047 66512  13 Fine Zero Adjust Assembly 01047 60003  14 Integrator Adjust Assembly 01047 60004   15 Digital Display Board ERC75012 01047 66513  16 Display Push Button Cap 01047 47402          Service Handbook for 1047 Refractive Index Detector   12 2001 135    Parts Information                Table 16 Front Panel Assembly     Description ERC number Part number   17 LED Board 2 ERC75009B 01047 66511   18 Display Switch Board ERC75013 01047 66514   19 Ribbon Cable  2  from LED Board 2 CON901 01047 61604   to Motherboard CON109   20 Ribbon Cable  5  from Display Switch Board 01047 61607  CON1302 to Motherboard CON111   21 Nut 2950 0043   22 Screw pan pozi  M4 8 mm long 0515 0910  Lock washer 2190 0586  Flat washer 3050 0893   23 Screw pan pozi
37. H or  the TEMP SENSOR will immediately disable power to CON201 and turn off  the temperature indicator LED  In addition  these TEMPERATURE AND  MOTOR CONTROL board test points are useful   TP201  upper right corner   e Pulsing  approx  2 5 VDC  while temperature is controlling at the setpoint   e  14 VDC while actual temp is less than setpoint   e  14 VDC while actual temp is greater than setpoint or TEMP SELECT  SWITCH is disconnected   TP203  under CON201   e HIGH   5 VDC  during any normal operation   e LOW if TEMP SENSOR is disconnected        Service Handbook for 1047 Refractive Index Detector   12 2001 45    Hardware Information       Flush Valve Control Electronics    Figure 17 outlines both the manual and remote operation of the FLUSH  VALVE      12 VDC is applied to the LAMP VALVE terminal  valve wires  for valve  turn on   12 VDC is removed for valve turn off     The front panel FLUSH button must be off if the rear panel remote input is to  be used  The remote input needs a contact closure  short  for activation  The  first contact turns on the valve  the next contact closure turns off the valve     The front panel FLUSH button will be disabled if the remote input is held in  the constantly short state        CONTROL BOARD TP506 LOW for valve off  HIGH for valve on   POWER SUPPLY BOARD TP305  to TP306    12 VDC       The front panel FLUSH light comes on when the valve is turned on either  from the front panel or the rear panel  The light only indicates that the valv
38. IFF AMP is first taken to the AB   autobalance  DECISION threshold circuit  Measuring TP3  VR503 is  adjusted such that the AB DECISION threshold is 16 x 10 RIU  The output  of the AB DECISION threshold circuit is then taken from the CONTROL  board to the input of the OPTICAL AUTOBALANCE board  Figure 5   The  DIFF AMP output is also taken from the CONTROL board to the input of the  ELECTRICAL AUTOBALANCE board   s  Figure 6 on page 26  ELECTRICAL  AB CIRCUITS     Optical Balance Board    from Control                                                 Board  AB  Decision Level to Electrica  Output AB Board  wo   Con 119  Ultimately to the  AB Motor Assembly Optical AB  N Motor Drive Sensing  Pulse Circuit and Control  Circuit  Optical Autobalance Board             When a BALANCE is initiated  the OPTICAL AUTOBALANCE board judges  the signal from the CONTROL board AB DECISION threshold circuit  If the  signal is below the AB threshold  then only an electrical balance is done        Service Handbook for 1047 Refractive Index Detector   12 2001 25    Figure 6       NOTE       Hardware Information    Electrical Autobalance Board    from Control to Output Board   Board Diff  Response Time  Output Controller                                                       S  Con 103  VR401 TP401  Sw401 Electrical  AB Circuit  Electrical AB aye  Controller TP40  Electrical Autobalance Board          During an electrical autobalance  the ELECTRICAL AUTOBALANCE BOARD  sets the output of it   s ELEC
39. NOTE       Replacement    Procedures    e Replace the optical unit covers and insulation     e Perform the complete program of electrical calibrations described in     Electrical Calibrations    on page 93     Photo Diode Replacement Adjustment    The Photo Diode Assembly  01047 60013  actually consists of the photoed  package and the 1st lens slit group     Never attempt to further disassemble the photo diode assembly  The  relationship of the internal parts are critical  and these are pre adjusted at the  factory     1 Refer to    Parts Information    on page 127     2 Remove the detector top cover  Refer to Figure 44 on page 140  Remove  the optical unit top cover  item 13   the insulation  item 9  and the optics  cover  item 12      Locate the Preamplifier board  Refer to Figure 40 on page 130 if needed      4 Carefully un solder from the Preamplifier board the three wires  white   green  clear  from the photo diode assembly cable  This cable can be seen  coming from the optical unit  directly under the lamp  Before un soldering   note the position of the three wires     5 Refer to    Boards and Controls located in the front panel    on page 68 and  Figure 27 on page 68  This figure shows the holding screws for the detector  front panel assembly  Remove these screws  then carefully allow the front  panel assembly to fall away   it is not necessary to remove any cables or  tubes  but make sure that nothing is damaged     6 Figure 38 on page 91 shows a side view of the opti
40. Refer to    Optical Unit  Replacement    on page 78        92 Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures       Electrical Calibrations    Correct electrical calibration is essential to the proper operation of the  1047A  These calibrations must be performed if        The original  factory set light intensity is altered  e g   by changing the lamp  or re adjusting the mirror     A critical board which is replaced  e g   the AMPLIFIER or CONTROL  board     The calibrations MUST BE PERFORMED IN THE ORDER IN WHICH  THEY ARE GIVEN BELOW     Pure  well degassed HPLC grade water should be in both halves of the flow  cell during calibrations  The detector should be equilibrated at 40  C for  ninety minutes     Average light in tensity  TP1 2 select buttons pressed together  MUST be  no less than  3 5     NULL the detector output    1    Turn on the detector and make sure that the LIGHT INTENSITY ERROR is  not present     At the front panel  turn the INTEGRATOR dial fully clockwise and turn the  FINE ZERO dial to it s mid position  5 turns      Press and hold the front panel   COURSE ZERO button until the LIMIT  ALARM indicator comes on     With the LIMIT ALARM on  set the optical balance glass to it s null position  by pressing the   and  COURSE ZERO buttons simultaneously  The LIMIT  ALARM indicator will flash as the optical balance glass moves to it s null   position     Press the BALANCE button  After the balance is done  select the INT  display  U
41. SPAN  POLARITY  CONTROL and INT OFFSET stages  INT OUTPUT SPAN is adjusted by  VR613  POLARITY can  only  be changed from the detector rear panel  INT  OFFSET is influenced by the front panel INTEGRATOR dial  and by VR701 on  the OUTPUT board  Figure 9 on page 30   The output of the INT OFFSET  stage is taken to both the digital display and the INT analog output at the rear  panel  The INT analog output is calibrated  OUTPUT SPAN  such that 32 mV  equals 1 x 10   RIU        Service Handbook for 1047 Refractive Index Detector   12 2001 27    Hardware Information    Figure 7 Amplifier Board    from Output  Board  Response to Output Board  Time Controller Int Offset             Path    Int  Polarity    Control                   Activated  from Rear  Panel only     Integrator    Path               pan M a   D arker     Range   z Ra Offset  E Control n          Circuit  ad       Marker    A  o  Also controllable  from Rear Panel           Front Panel       28 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Figure 8 Amplifier Board    to Output Board  to Output Board   ImV Output to Display 10mV Output              MB Can 114 MB Con 111 MB Con 114          103    ae   Von    Amplifier Board           1mV       Rec   Shorting               Polarity    Control    FS Output    Select          Ornv   Control       Rec  Circuits    Control          Polarity Recorder       Service Handbook for 1047 Refractive Index Detector   12 2001    29    Hard
42. TRICAL AB CIRCUITS to zero  An electrical  balance does not zero the DIFF AMP output  If the AB decision signal  exceeds threshold  an optical balance is also done  This means that the  OPTICAL AUTOBALANCE board  indirectly  moves the Optical Balance  Glass until the output of the DIFF AMP is zero  The optical balance is then  followed by an electrical balance  Refer to the AB MOTOR CONTROL for  details of the optical unit and autobalance motor control     On the ELECTRICAL AUTOBALANCE board  both poles of SW401 must be  OFF  If these switches are pushed ON  the ELECTRICAL AB CIRCUITS output  will be held constantly to zero  The ON position of these switches is for  adjusting the offset of the ELECTRICAL AB CIRCUIT output  VR401 is  adjusted so that TP401 is 0 VDC 0 3 mV     When SW401 is OFF     TP402   HIGH normally  LOW during balance  TP403   HIGH normally  LOW when balance command initiated       26 Service Handbook for 1047 Refractive Index Detector   12 2001    Integrator Path    Hardware Information    The detector signal path continues at the output of the ELECTRICAL  AUTOBALANCE board  The signal goes through the RESPONSE TIME  circuit  and is then taken to the input of the AMPLIFIER board  Figure 7 on  page 28 and Figure 8 on page 29   The first stage of the AMPLIFIER board is  influenced by the front panel FINE ZERO dial  After the first stage  the signal  divides into two separate paths  INTEGRATOR AND RECORDER     The INTEGRA TOR path consists of the INT OUTPUT 
43. and samples used in Liquid Chromatography are  often TOXIC and DANGEROUS  Always determine if the detector is filled  with any harmful liquid before beginning a repair     e ALWAYS USE ANTI STATIC PRECAUTIONS when working on the 1047A     e Certain adjustmen t and calibration procedures require that the detector  cover be removed while the detector is powered on  BE VERY CAREFUL  WHILE PERFORMING THESE PROCEDURES  Hazardous voltages are  present  This warning will be assumed for all written procedures        Service Handbook for 1047 Refractive Index Detector   12 2001 63       CAUTION          CAUTION          NOTE       Procedures       Board and Front Panel Controls  Replacement    Refer to Figure 40 on page 130  for board locations  Each board is keyed to  it   s particular location  The key is inserted in to the particular Motherboard  connector  Be careful not to remove this key     Do not touch any potentiometer or switch on any board that is not described  in the Electrical Calibration procedures  These potentiometers and switches  are pre set at the vendor   s factory and may not be touched without risk to the  proper operation of the board     Always disconnect power before removing replacing any board     Temperature and Motor Control Board    Remove the defective board and insert the replacement board  Make sure  that the three connections for the board   CON201  202 and 203  are correctly  attached  Each connector is unique and cannot be put in to an incorrect  p
44. ard     Install the replacement board and re connect the associated cables     No calibrations are needed  Re assemble the detector and check for  proper operation        70    Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures       Flush Valve Replacement    1 Disconnect the flush valve wires from their locations at the VALVE LAMP  terminal block  Remove the two screws that secure the valve and remove  it    2 Install the replacement flush valve     3 Connect the valve wires to their locations at the VALVE LAMP terminal  block  There is no polarity  either wire can be attached to either location        WARNING If the flush valve is incorrectly plumbed  the detector flow cell may be  destroyed  Make sure that the flush valve is plumbed according to the  diagram in Figure 29        Figure 29 Flush Valve Plumbing                                      Detector  Outlet FA       3   Way  Fitting                      Detector N    Inlet ji       a       D                                                                      Optical Unit View  Flush Valve from Front    FAN B         i                                      gt              O  O    a       3  Z       Service Handbook for 1047 Refractive Index Detector   12 2001 71    Procedures       Table 5 Legend Flush Valve Plumbing       Term Description       See the 1047A Operator s Handbook for details of Flow System Operation  Read these notes first   A This is a standard zero dead volume 1 16 inch union 
45. b holder as  shown in Figure 34 on page 83  Repeat this procedure until the replacement  lamp can be correctly adjusted in the detector           82 Service Handbook for 1047 Refractive Index Detector   12 2001    Figure 34       NOTE       Procedures    Orientation of Bulb Holder    V p    WJ  G uf    Possible original Changed orientation  orientation                      Flow Cell Replacement    The Flow Cell Assembly  01037 60003  may be used in either the 1047A or the  obsolete 1037A  The four capillary tubes of the flow cell are not fitted with  screws ferrules at the time of shipment  Included with the flow cell assembly  is a bag which contains a set of fittings for the 1047A and a set of fittings for  the 1037A  Select the fittings required  see Figure 29 on page 71      The optical surfaces of the flow cell itself are protected by a piece of clear  plastic tape  Remove this tape  and clean the optical surfaces with alcohol  and a cotton swab     Failure to remove this protective tape will result in insufficient light intensity     The lamp should be correctly adjusted at this time  If not  refer to    Lamp  Replacement Intensity Adjustment    on page 80 directly preceding     1 Refer to    Parts Information    on page 127     2 Remove the detector top cover  If needed  refer to Figure 40 on page 130  to identify the optical unit     3 Refer to Figure 44 on page 140  Remove the optical unit top cover  item  13   insulation  item 9  and optics cover  item 12      4 Ref
46. cal unit as viewed from  the front of the detector  The figure also shows how the photo diode  assembly will be removed from the optical unit        90 Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures    Figure 38 Optical Block         Photo diode Assembly Setscrew  Access Holes  1 5 mm HEX     2 Mirror Adjust Glass Access Hole  3 Optical Unit Outer Case   4 Optics Block   5 Foam Insulation   6 Photo diode Assembly                         7 Atthis time  observe how the photo diode assembly is seated in the optical  unit     8 The photo diode assembly is secured by two 1 5 mm hex set screws  These  screws can be loosened through the access holes with a 1 5 mm hex  wrench  Push a hex wrench through the foam insulation into the access  holes in the optics block  Loosen both set screws      9 The complete photo diode assembly can now be pulled out of the optics  block as shown in Figure 38     10 Install the replacement photo diode assembly  Make sure to guide the  photo diode cable through the correct hole in the front plate   the hole  directly under the lamp  Make sure the new photo diode assembly is  adjusted correctly   it should be pushed against the front plate  and sitting  flatly on the floor of the optics block  Then  tighten the set screws        Service Handbook for 1047 Refractive Index Detector   12 2001 91    Adjustments and  Calibrations    Procedures    11 Solder the three wires of the photo diode cable to their correct positions  on th
47. d    ALARM does not turn on  The motor sits vibrating at it s 2 Replace the AB MOTOR assembly    mechanical stop  In all other ways  the AB motor and NOTE  The vibrations may have disturbed the AB   BALANCE functions are normal  MOTOR OPTICAL BALANCE GLASS relationship  Verify  this relationship  on page 75  step B            Service Handbook for 1047 Refractive Index Detector   12 2001 115    Troubleshooting Information    Flush Valve Problems       NOTE    Normally  a quiet    click    can be heard when the valve activates     The FLUSH indicator led on the front panel only indicates that a flush  command has been made  The indicator DOES NOT sense the actual position  of the valve  Therefore  the indicator cannot be taken as proof that the valve  is actually in one position or the other     If the valve is activated from the front panel  the rear panel remote input is  disabled  If the valve is activated from the rear panel remote input  the front  panel FLUSH button is disabled           Table 11 Flush Valve Problems       Possible Problem    Actions       Valve leaks liquid externally   WARNING  The pressure rating of the valve is only  2 bar     Valve has an internal cross port leak  This symptom will  most likely be noticed in the form of a baseline drift while  the valve is in the normal position  The drift stops when  the valve is activated     Valve cannot be activated from the rear panel remote  input     Replace the valve  This is most often caused by excessive  r
48. detector power  Turn on should be normal and the LIGHT    N    3    INTENSITY ERROR should not be on  By hanad  adjust the optical balance  glass control shaft so that TP1 and TP2  display  are equal  This means that  the optical balance glass is in it   s null position     At the front panel  press and hold the   COURSE ZERO button until the AB  motor reaches it   s limit  counter clockwise  and the LIMIT ALARM  indicator is on     While the LIMIT ALARM indicator is on  press the   and   COURSE ZERO  buttons simultaneously  The AB motor will now seek it   s mechanical  center position  The LIMIT ALARM indicator will be flashing during this  movement     At this time  the optical balance glass is in it   s null position and the AB  motor is in it   s mechanical center position  Now  the balance glass and the  AB motor may be joined together  As carefully as possible  push the AB  motor assembly onto the optical unit  allowing the sleeve coupling to join  the optical balance glass control shaft to the AB motor shaft  Make sure  that the sleeve coupling has a firm hold on both sides  Fasten the AB motor  to the optical unit with the original four screws     This question is often asked  After step D above  how close must TP1 and  TP2 be to each other  In other words  how may the relationship between  the AB motor and the optical balance glass be checked     Select INT at the display  Turn the front panel INTEGRATOR dial fully  clockwise  Center the front panel FINE ZERO dial  5
49. dicator LED turns on  TP1102  normally HIGH  LOW for LIMIT ALARM     From the LIMIT ALARM state  the AB Motor can be commanded to move to  it   s mechanical center position by pressing the front panel   and   COURSE  ZERO buttons simultaneously  The motor will then move in the appropriate  direction until it   s center position is sensed  A BALANCE is NOT  automatically done after this movement        42 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Hardware Information       Temperature Control Electronics    Figure 16 on page 44 is a functional block diagram of the 1047A heater  control electronics     Actual heating power is provided by the main power transformer 100 V AC  secondary  which can be measured at CON202 of the TEMP AND MOTOR  CONTROL board  A TEMPERATURE REGULATOR delivers power to the  optical unit HEATER  The optical unit TEMP SENSOR reports the actual  optical unit temperature to the TEMPERATURE CONTROL circuit  Required  heater voltage is determined by a comparison between the actual optical unit  temperature and the temperature setpoint     Both poles of switch SW201 must be OFF for correct operation of the heater  control circuitry  The ON position of these poles is used by the factory and will  disable operation of the heater circuit     The HEATER is protected by a thermal fuse which opens at about 76  C  The  heater circuit is protected by a 2 A fuse     Resistance of the HEATER  about 190 Q cold  Resistance of the TEMP
50. e  A D Converter     The Noise Specification is 5 x 10  9 RIU P P  5x10 9   0 0005 x 10 5   The allow able Noise output is 16 pV P P    See  1  above    The Drift Specification is 2 5 x 10 7 RIU hr  2 5 x 107   0 025 x 10 5    The allowable Drift output is 800 pV hr        Service Handbook for 1047 Refractive Index Detector   12 2001 107    Performance Verification       108 Service Handbook for 1047 Refractive Index Detector   12 2001       Troubleshooting Information    This chapter provides troubleshooting  information for the 1047 Refractive Index  Detector       Troubleshooting Information    The following sub sections are guides for troubleshooting general categories  of hardware problems  These guides are meant to be used in conjunction  with the detailed descriptions found in    Hardware Information    on page 19     e Hardware Troubleshooting    1 Detector turn on problems   I Light intensity problems    1 Lamp turn on problems    1 Limit alarm  AB motor and balance problems   J Flush valve problems    1 Temperature control problems    d Flow cell problems       LI Loss of sensitivity    e Chromatographic Troubleshooting       110 Service Handbook for 1047 Refractive Index Detector   12 2001    Troubleshooting Information       Hardware Troubleshooting    Detector Turn On Problems             Table 7 Detector turn on problems  Problem Check  Rear line switch turned on  no response or incorrect Is the line switch functional     detector turn on    Is the line power se
51. e  has been commanded to turn on  it DOES NOT guarantee that the valve itself  is functioning correctly        46 Service Handbook for 1047 Refractive Index Detector   12 2001    Figure 17    Hardware Information    Flush Valve Control Electronics                Output Board    LED Board  1  O Flush LED         Input Row     gt     a    Can 602           Key  Switch  Board       Flush            Regulator Set     12VD0  5VDC    con  104          Valve Driver  Circuits    e TPSO06                 Control Board               Lamp Valve    Terminal    Flush      Vane         Service Handbook for 1047 Refractive Index Detector   12 2001    47    Figure 18    Hardware Information       Remote Control Function Electronics  The detector rear panel is arranged into an INPUT row and an OUTPUT row     The Outputs    Light Intensity Error and Balance Not Ready    For both of these functions  the appropriate pin pair is normally open  The  Light Intensity Error pair is closed as long as the error exists  and opens  again when the error clears  The Balance Not Ready pair is closed for the  duration of the BALANCE     Block Diagram Light Intensity Error                Output Board            pg       Output Row     gt   4 l                            Light Intensity  Error                               n  CON 703  kid                          Front Panel  Light Intensity  Error LED                         other CON  109 CON  114  Board                         Control Board             Ligh
52. e Preamplifier board  If you did not note their original positions  refer  to Figure 26 on page 66     12 Replace the optical unit top covers and insulation  Replace the detector  front panel assembly     How much work needs to be done depends up on how closely the new photo  diode assembly matches the original photo diode assembly  Proceed as  follows     1 Turn on the detector  The INTENSITY ERROR should not be on  The  average intensity  display TP1 2 pressed together  MUST be no less than   3 5  However  remember that typical intensity values range from  4 0 to   5 5  TP 1 2 must also be within 0 2 of each other     2 In your best judgement       If step 1 above is satisfactory  immediately perform the complete  calibration program described in    Electrical Calibrations    on page 93    If step 1 above is NOT satisfactory then  a Is there an air bubble in the flow cell  Try flushing the flow cell   or    b Is the new photo diode assembly correctly installed  Is it adjusted  correctly  Are the photo diode wires correctly soldered to the Preamplifier  board     or    c The new photo diode assembly requires a re adjustment of the mirror   Refer to  Mirror Replacement  and    Mirror Adjustment    on page 85 and  re adjust the mirror     Optical Unit Heater Temperature Sensor    The optical unit heater  or it   s thermal fuse  and temperature sensor ARE  NOT replaceable on the component level  If these components become  defective  the complete optical unit must be replaced  
53. e front panel  COURSE ZERO button to  make TP604 read  2 5 VDC  This is near the maximum possible output for  TP604  The front panel INT display will usually blank out before  2 5 VDC  is reached  However  TP604 must reach  2 5 VDC before the front panel  LIMIT ALARM is set     Now measure TP605  visible   Using VR614  adjust TP605 to measure   2 5 VDC     Press BALANCE  the INT display should return to 0 00  To test this  calibration     Locate the manual shaft turning knob on the OPTICAL BALANCE  MOTOR  If needed  see Figure 31 on page 74  item C     Carefully turn the manual shaft turning knob exactly two full turns in the  counter clockwise direction  The value of the INT display should be  increasing in the positive direction     Note the value of the INT display at the    two turns    position     At the rear panel  change the integrator POLARITY to negative by  shorting the two POLARITY pins in the INPUT row     This will initiate a balance  and the INT display should return to 0 00  The  front panel integrator POLARITY leds should now indicate negative  integrator polarity     Again carefully turn the manual shaft turning knob exactly two full turns  in the counter clockwise direction  The value of the INT display should  now be increasing in the negative direction     In value of the INT display should now be the same as the value in step 3  above  but with a negative polarity     10 Perform the complete program of electrical calibrations described in     Electrical 
54. ecial care when handling the detector        12 Service Handbook for 1047 Refractive Index Detector   12 2001    General Information       About this Manual    This handbook provides service information for the 1047A Refractive Index  Detector  The handbook has been divided into the informational sections  outlined in the table of contents        Service Handbook for 1047 Refractive Index Detector   12 2001 13    General Information       Repair Policy    All printed circuit boards are replaced on the board level  Certain other  electronic components may be replaced on the component level     The Optical Unit may be replaced as an assembly or repaired on the  component level  depending on the failure mode     Certain replacements require subsequent mechanical and electrical  adjustments  These adjustments are documented in this handbook     The components and assemblies available for replacement parts are  described in    Parts Information    on page 127        Warranty    Agilent Technologies warranty policies and procedures apply        14 Service Handbook for 1047 Refractive Index Detector   12 2001    Table 1    General Information       Specifications    1047A specifications for noise  drift and wander are made under the    following conditions     Flow of 1 ml min of adequately degassed HPLC grade water  24 hour  warm up time at detector temperature of 40  C and ambient temperature   lt  25  C  Detector response time 0 8 seconds  For Hewlett Packard integrators     peak
55. en 100 and 110 mV  this is below the AB decision threshold      5 Press BALANCE  Only an electrical autobalance should be done  the AB  Motor should not move  The INT display should return to 0 00  as originally  set     6 Again press and hold the   COURSE ZERO button  this time until the INT  display reads between 150 and 160 mV  this is above the AB decision  threshold         96 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Procedures    7 Again press BALANCE  This time  the AB Motor should move to it   s null  position as an optical autobalance is done  Then  an electrical autobalance  should return the INT display to 0 00  as originally set     It is important to realize that a BALANCE will return the baseline to the value  set by the FINE ZERO and INTEGRA TOR dials  A BALANCE does not return  the baseline to 0 00 unless  as in the test above  the actual set value is 0 00     Calibration of OUTPUT board VR701  Optional step     Ordinarily  VR701 of the OUTPUT board need not be calibrated  Perform this  procedure only if the relationship between the INT display and the REC  display is not correct  refer to    Verify the Operation of the Baseline Controls  and Autobalance    on page 96  step 2  is not correct     1 The OUTPUT BOARD is located on the detector rear panel assembly   VR701 is at the top of the board  clearly labeled     NULL the detector as described in Null the detector output     Set the front panel RANGE switch to 32  Select
56. er outputs polarity light above the  front panel  recorder  POLARITY button  A change in polarity will also cause  a BALANCE to occur     Block Diagram Integrator Polarity                Output Board                   Input Row     gt f   e       Polarity                                                                                                                                                                                                                                                  CON _703  Front Panel  Integrator  Polarity LED  O   O   CON  114   CON  109 Mother  Board  Optical Auto nedor    Balance e    Polarity  Board A a EAA  Control Circuits  Amplifier Board  Auto  Balance  Board                               Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    MARKER    The MARKER applies a voltage spike to the recorder outputs so that the start  of a chromatogram may be noted on a strip chart recorder  A MARKER can  be initiated from the front panel  or by applying a short to the MARKER pin  pair     Figure 22 Marker            Board       Output  Marker    o         e    Input Row     gt           CON 703                    Front Panel  Marker LED    Panel  Button    Front  Marker       CON 114 CON 115       Mother  Board        Recorder  Qutput  Circuits       Relay and  Driver             Marker  Voltage    Offset          Amplifier Board       Board    Control       52 Service Handbook for 1047 Refractive Index Detecto
57. er to Figure 32 on page 81 to identify the flow cell assembly as it  appears in the uncovered optical unit  Also note the location of the mirror        Service Handbook for 1047 Refractive Index Detector   12 2001 83       NOTE          NOTE          NOTE       Procedures    Do not disturb the mirror pitch screws     5 Remove the t wo mirror holding screws  BE CAREFUL NOT TO TOUCH  THE SURF ACE OF THE MIRROR  Carefully remove the mirror and put it  in a safe place  If the detector is turned on now  the LIGHT INTENSITY  ERROR will come on     6 The two screws which hold the flow cell in place are hidden under the  mirror  With the mirror removed  these screws can now be seen  Remove  the two flow cell holding screws   but do not attempt to remove the flow  cell assembly now     7 The four capillary tubes from the flow cell assembly can be seen coming  from the front of the optical unit  One of the tubes goes to the detector  inlet  Two of the tubes go to the flush valve  The last tube goes to the 3 way  fitting  Now  disconnect these four capillary tubes     8 Refer again to Figure 44 on page 140  Carefully extract the complete flow  cell assembly  item 6  as indicated in the figure  This may be tedious  As  the flow cell assembly is pulled upward  two thin pieces of insulation   covering the capillary tubes  will be displaced     9 Apply asmall amount of thermal compound  part number 6040 0454  to  the replacement flow cell assembly  Usually  the residual thermal  compound fr
58. es an overview of DC power production and distribution  All  test points are clearly labeled across the top of the POWER SUPPLY board     The REGULATOR SET is mounted on the rear panel  These regulators  contribute to the production of the  5 VDC LAMP supply and the  12 VDC  supply  The REGULATOR SET cable is connected to MOTHERBOARD  CON107  To check the performance of the REGULATOR SET  DC  measurements can be made directly at the wires of CON107               5 VDC lamp   TP 302 to BLU  com  almost zero  GND  TP 302 to WHITE  in  8 3 VDC   TP 302 to RED  out  5 VDC    12 VDC   TP 306 to ORN  com  almost zero  GND  TP 306 to YEL  in  19 VDC   TP 306 to BRN  out   12 VDC       Lamp voltage is supplied to the LAMP as soon as the rear panel LINE switch  is turned on  Circuits on the CONTROL board apply  12 VDC to the FLUSH  VALVE when a FLUSH command is initiated        34 Service Handbook for 1047 Refractive Index Detector   12 2001    Figure 11           Mother    Board  x    from  Main  XFMR    secondaries    Power    Von         AC Voltages    Hardware Information    DC Power Distribution                                                                                                                                                                                                                                                                                 Regulator Set   5VDC a     Lamp  mounted on  Rear Panel   2p er  6B    a       12v0C  ab fal    Jg  ov             slia
59. estriction downstream of the valve  Check for  downstream restrictions inside the detector  Check for  excessive restriction downstream of the detector itself   Note   f the 1090 LC is being used  the detector outlet  SHOULD NOT be plumbed to the 1090 injector wash  input     Replace the valve     Can the valve be activated and deactivated normally from  the front panel     If yes    1 The front panel FLUSH button must be off to use the  rear panel remote input    2 Make sure that a contact closure is being applied to  the remote control input  normally open  closed for  activate deactivate    3 Check the OUTPUT board ribbon cable  In this case   other remote control functions will likely also be  defective        116    Service Handbook for 1047 Refractive Index Detector   12 2001    Troubleshooting Information       Table 11 Flush Valve Problems       Possible Problem    Actions       Valve cannot be activated from the front panel FLUSH  button     4 Replace the OUTPUT board     If no   Is the FLUSH command applying   12 VDC as needed to  the valve drive wires     If yes  replace the valve     If no   1 Replace the CONTROL board   2 Replace the POWER SUPPLY board     Can the valve be activated and deactivated normally from  the rear panel remote input     If yes     1 Deactivate the valve from the rear panel  then check  for correct front panel operation    2 Check the KEY SWITCH board ribbon cable  In this  case  other front panel functions are likely to also be  defective
60. evel detector repair     Temperature and Motor Control Board    e SW201   Both poles must be OFF  see    Heater Control Electronics    on  page 44 for details     e TP201  upper right corner    see    Heater Control Electronics    on page 44  for details     e TP203  under CON201    see    Heater Control Electronics    on page 44 for  details     e TP204 205  left of CON201    see    AB Motor Control    on page 40 for  details     Power Supply Board    All test points are clearly labeled across the top of the board  See    Power  Distribution    on page 32 for details     Optical Autobalance Board  VR2  TP1101 1102   See    AB Motor Control    on page 40 for details        Service Handbook for 1047 Refractive Index Detector   12 2001 55    Hardware Information    Electrical Autobalance Board    e SW401   Both poles must be OFF  see    Signal Path Overview    on page 21  for details     e TP401 402 403   see    Signal Path Overview    on page 21 for details     Control Board  Test points extend out to the rear for convenience   e VR501  TP1 4   See    Signal Path Overview    on page 21 for details     e VR502 503  TP2 3   See    Signal Path Overview    on page 21 and    Electrical  Calibrations    on page 93 for details     Output Board    e VR701   See    Signal Path Overview    on page 21 and    Electrical  Calibrations    on page 93 for details        56 Service Handbook for 1047 Refractive Index Detector   12 2001    Figure 24    Zero    Input       Hardware Information   
61. ex Detector   12 2001 21    Hardware Information    Figure 2 Signal Path Overview              Display  Switch  Board             Digital  Display            LED Board         2     Light Intensity  MB Con 109 Error       Preamplifier  Board      gt   R eZ                  MB Con 104 MB Con 119                        Intensity             Error Signal Path AB  Level Starts here Sensing  and           Control        AB  Decision  Level            Diff       AB Motor    Control           Control Board    Optical  Autobalance  Board          22 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Figure 3 Signal Path Overview          A Ink      Adjust    Dial      Time J   gwiteh     10 mVI                     Con  704  1mV    Electrical  AB Circuit               Response  Time  Controller    10mV           INT         Electrical  AB Control              Output  Board    Elec   Autobalance Board          1mV           ME Con 116             MB Con 105    1001  Range Switch  Board    Con            Integrator         Circuits         First    1mV  Stag             Recor       1002    Con    1OmV       Circuits       Amplifier Board  lA    Zer   Front Panel  P Recorder Controls                      Service Handbook for 1047 Refractive Index Detector   12 2001    23    Figure 4    Hardware Information    On the Preamplifier board  S and R each produce a voltage output  proportional to the light reaching it   s photo diode  These voltages are  re
62. g screws while  pressing the mirror     Mirror adjustment can be done with the optical unit completely  uncovered  but stray light will cause some instability of TP1 2  After the  mirror is finally tightened  cover the optical unit  Then the mirror adjust  glass can be used to make TP1 2 equal        88    Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures    If it seems not e Is there an air bubble in the flow cell   possibts i e Has the flow cell been replaced recently  If yes  has the clear protective  mirror  tape been removed from the flow cell   e Did you get fingerprints or dirt on the optical balance or mirror adjust  glass   e Is the mirror a new replacement part  If yes  it could be that an additional  step is needed for mirror adjustment  See Figure 37   Figure 37 Mirror Screws    1  Mirror Holding Screws  2  Mirror Pitch Screws  2 mm HEX        Ordinarily  the pitch screws should not be loosened  However  if the mirror is  a new part  the pitch of the mirror may also need adjustment  Proceed as  follows     a Loosen the mirror holding screws  Establish the mirror position at midway  between full left and full right  Then  tighten the holding screws     b Loosen the pitch screws  Adjust the mirror pitch so that the beam is related  to the photo diodes as shown in Figure 36 on page 88  Then  re tighten the  pitch screws     c Repeat the mirror adjustment procedure        Service Handbook for 1047 Refractive Index Detector   12 2001 89       
63. ht intensity   Correct the intensity loss  then perform the complete program of electrical  calibrations described in    Electrical Calibrations    on page 93     If the OUTPUT SPANS are correct  there is a small possibility that the flow  cell is defective  However  it is most likely that the loss of sensitivity is  related to the chromatographic method  the Liquid Chromatograph or a  worsening detector environment        Service Handbook for 1047 Refractive Index Detector   12 2001 121    Troubleshooting Information       Chromatographic Troubleshooting Guide    General Characteristics of RI Detector    Because they respond to all solutes  RI detectors have excellent versatility   Generally speaking  success depends up on how carefully certain mobile  phase and environmental requirements are observed     For maximum sensitivity  the mobile phase and the sample solute should  have different an RI as possible     However  even under optimum conditions the RI detector has only modest  sensitivity when compared to the UV detector  The RI detector is generally  never used for trace analysis  A limitation to the real sensitivity of RI  detectors is their susceptibility to the environment  temperature changes  being the biggest problem     To even begin to achieve the theoretical sensitivity of the detector  the  ambient environment and the chromatographic system parameters   temperature  flow  etc   must be kept under close control  These factors  most often result in baseline pr
64. ical Unit temp sensor may have resistance   but be defective  Replace the Optical Unit           120 Service Handbook for 1047 Refractive Index Detector   12 2001    Troubleshooting Information    Flow Cell Problems    Flow cell problems can present the following symptoms        Table 13 Flow Cell Problems       Possible Problem Actions       External leaks  Replace the flow cell  Refer to    Flow Cell Replacement     on page 83     This is usually caused by excessive restriction  downstream of the cell  Liquid may be seen coming from  the contingency leak drain on the right side of the  detector  The leak may also cause light intensity  problems  and baseline problems such as noise  drift and  wander    Note  The back pressure rating of the flow cell is only   7 bar     Internal  cross compartment leaks  or u cracks which Same as above   aspirate  In this case  an external leak may not be   observed  This defect often causes light intensity   problems  and baseline problems such as noise  drift and   wander        Loss of Sensitivity    This complaint is usually expressed as an observed diminishment in peak  height under the    identical    chromatographic conditions which previously  produced satisfactory results     When this complaint is made  IMMEDIATELY verify that the INTEGRATOR  and RECORDER OUTPUT SPANS are correctly calibrated     A low OUTPUT SPAN will result in diminished peak height  If the OUTPUT  SPANS are low  determine what has caused the loss in detector lig
65. ice Handbook for 1047 Refractive Index Detector   12 2001    Figure 1    General Information       Pressure Contribution of Detector    Typically the pressure contribution of the detector  NOT SPECIFIED  is no  greater than 35 bar at 5 ml min IPA     The difference between VALVE ON and VALVE OFF mode is no greater than  1 bar     Pressure Contribution   A P    A Px    Figure 1 shows how the pressure is measured     Pressure Diagram          HP 1090  5ml min mus HP 1047A    IPA                     APx               APY                   Service Handbook for 1047 Refractive Index Detector   12 2001 17    General Information       18 Service Handbook for 1047 Refractive Index Detector   12 2001       Hardware Information    This chapter provides hardware information for  the 1047 Refractive Index Detector       Hardware Information       Detection principle  optical system and  detector flow path    Refer to the 1047A Operator   s Handbook  01047 90000   The section  OVERVIEW of the Operator   s Handbook contains detailed information on the  detection principle  optical system and flow path of the 1047A        20 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information       Signal Path Overview    Figure 2 on page 22 and Figure 3 on page 23 show the functional signal path  block diagram for the 1047A  More detailed block diagrams and fundamental  circuit explanations will be found in the following text        Service Handbook for 1047 Refractive Ind
66. ice in case of part  number issues or  upgrades     The latest version of this  manual is available as  Adobe Acrobat Reader   PDF  version only and  can be downloaded from  the Agilent Technolgies  web page  www agilent com        Agilent Technologies    Hewlett Packard Strasse 8    76337 Waldbronn  Germany    Contents    1 General Information    This chapter provides general information for the 1047 Refrac   tive Index Detector 9    Safety Summary 10  General 10  Operation   before applying power 10  Safety Symbols 11  1047A related warnings 12  About this Manual 13  Repair Policy 14  Warranty 14  Specifications 15  Pressure Contribution of Detector 17    2 Hardware Information    This chapter provides hardware information for the 1047 Re   Sractive Index Detector 19    Detection principle  optical system and detector flow path 20  Signal Path Overview 21   Integrator Path 27   Recorder Path 30  Power Distribution 32   AC Power Distribution 32   DC Power Distribution 34   Power Supply Area Wiring Diagram 36  Motherboard 38  AB Motor Control 40  Temperature Control Electronics 43       Service Handbook for 1047 Refractive Index Detector   12 2001 3    Contents    Flush Valve Control Electronics 46  Remote Control Function Electronics 48  The Outputs 48  The Inputs 50  Recorder Control Electronics 53  Summary of Switches and Potentiometers 55  Temperature and Motor Control Board 55  Power Supply Board 55  Optical Autobalance Board 55  Electrical Autobalance Board 56  Control 
67. if contacted  result in personal injury        10 Service Handbook for 1047 Refractive Index Detector   12 2001       WARNING       General Information    Any adjustment  maintenance  and repair of the opened instrument under  voltage should be avoided as much as possible  and when inevitable  should  be carried out by a skilled person who is aware of the hazard involved  Do  not attempt internal service or adjustment unless another person  capable of  rendering first aid and resuscitation is present  Do not replace components  with power cable connected     Do not operate the instrument in the presence of flammable gases or fumes   Operation of any electrical instrument in such an environment constitutes a  definite safety hazard     Do not install substitute parts or perform any unauthorized modification to  the instrument     Capacitors inside the instrument ma y still be charged even if the instrument  has been disconnected from its source of supply     Dangerous voltages  capable of causing serious personal injury  are present  in this instrument  Use extreme caution when handling  testing and adjusting     Safety Symbols    Below you will find some symbols that are used on the instrument and  throughout the documentation        The apparatus will be mark ed with this symbol when it is necessary for the  user to refer to the instruction manual in order to protect the apparatus  against damage     Indicates dangerous voltages     Indicates a protected earth terminal      
68. ifier board ribbon cable from CON501 of the  CONTROL board     2 Refer to Figure 41 on page 134  item 33  pre amp support plate   Remove  the two screws which hold the support plate to the card cage  Then  pull  the complete assembly up for access  Be CAREFUL   the Preamplifier  board is still attached to the Optical Unit by the photo diode assembly  cable        Service Handbook for 1047 Refractive Index Detector   12 2001 65    Figure 26    Procedures    Remove the Plexiglas safety cover from the Preamplifier board     The three wires of the photo diode assembly cable  white  clear and green  are soldered to the Preamplifier board  Note the location of these wires on  the board     At the Preamplifier board  carefully un solder the three wires of the photo  diode cable     Remove the standoffs which fix the Preamplifier board to it   s support  plate  then remove the board     Attach the replacement Preamplifier board to the support plate  Carefully  solder the three wires of the photo diode cable to the new Preamplifier  board  If needed  refer to Figure 26     Replace the Plexiglas safety cover  then re attach the complete  Preamplifier board assembly to the card cage  Re connect the Preamplifier  board ribbon cable to CON501 of the CONTROL board     Preamplifier Board                to Optical Unit to CON 501  Photo Diode Assembly Board    7 O SAMP     las  Re ane  Pies Wl    7  VEAR   BLK  ns BRN  lt      Z RED  2 ORN  C    y c  z YEL  zZ    Oo        D GREEN  Ss  N  C 
69. ill usually be near zero     Flow cell defective  leaking internally between  compartments or leaking externally     Flush valve defective  the user has attempted to purge air  bubbles  but half of the cell retains air because the valve  is defective     Flow cell dirty  optical surfaces are coated with an  opaque material  In this case  the TP1 2 intensities will be  abnormally low  but not zero     Lamp does not turn on when the line switch is turned on   In this case  TP1 2 intensities will be zero     Lamp not correctly adjusted  In this case  TP1 2  intensities will be abnormally low  or zero  depending on  the degree of mis adjustment     INTENSITY ERROR THRESHOLD CALIBRATION incorrect   In this case  the actual light intensity is sufficient  but the  intensity error circuit is incorrectly calibrated  The TP1 2  intensities seem correct  but the INTENSITY ERROR  indicator is on     If calibrations seem correct  likely failures are     PREAMPLIFIER board ribbon cable not securely connected  to CONTROL board CON501  In this case  TP1 2 readings  will be 0 00     Activate the flush valve and flush with a flow rate higher  than the analytical flow rate  Many times it will be  necessary to flush with pure methanol  or even  isopropanol  to purge the air bubbles     Replace the flow cell  and perform the alignments and  calibrations described in    Procedures    on page 61     See Flush Valve Problems     1  Perform an appropriate flushing procedure    2  Has the flow cell ju
70. increases  The rate of heat transfer  changes as the flow rate changes  this can cause baseline instability until  thermal equilibrium is reached     The 1047A inlet tube should be insulated using the pipe insulation   0890 1032   this ensures that the thermal stability in the flow system is easier  to achieve     The1047A outlet tube should have a wide internal diameter to avoid high  back pressures which can cause baseline instability or even damage the flow  cell  The tubing should not be coiled as this can cause pressure pulses which  produce reproducible spikes on the baseline     e All solvents must be HPLC grade     e Solvents must not contain any particulate matter and must be thoroughly  degassed before use  Careful filtering removes suspended particulate  matter which might contaminate the cell walls or reduce sensitivity by  light scattering  Degassing prevents bubbles forming in the solvent as it  passes through the flow cell     e Filters should not exceed 2 pm in pore size  The filter material must not  react with the solvent to cause solvent contamination        Service Handbook for 1047 Refractive Index Detector   12 2001 123    Troubleshooting Information    e Ifthe mobile phase is a combination of two or more solvents  these  solvents must be thoroughly mixed  One method for solvent preparation is  to first separately degas the pure solvents and then mix them to form the  desired solvent composition  followed by a further 10 minutes degassing   This extra 10
71. lation  item 9  and optics cover  item 12   Removing these covers  will expose the optics  as shown in Figure 32 on page 81     3 Disconnect the existing lamp wires from their locations at the VALVE  LAMP terminal block  Loosen the lamp holding collar  note that the collar  is secured by set screws  1 5 mm hex  located in the collar  Remove the  lamp holding collar and remove the lamp assembly     4 Fix the replacement lamp assembly into the lamp holding collar  Insert the  lamp into the optical unit  and loosely screw the lamp in to place with the  holding collar  Do not tighten the collar at this time  Connect the lamp  wires to their positions at the VALVE LAMP terminal block  There is no  polarity  either wire can be connected to either position     5 Turn on detector power  Refer to Figure 32 on page 81  The light from the  lamp should be clearly seen in the area of the slit  Now  rotate the lamp  until the slit is completely covered by the light  as shown in Figure 33 on  page 82     If it seems not to be possible to correctly adjust the lamp  see the special  information note below step 8        80 Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures                                                                                                 Figure 32 Inside the Optical Unit  3 4 2 3       1 Mirror Assembly    Fi o  DIP  h  2 Mirror Pitch Screw  2 mm HEX  E    x    J  3 Mirror Holding Screws  4 Cell Assembly 4  5 Slit Lens Assembly z 5  6 Slit
72. ll z  ORN  2   Lamp   gt   ey Mother Valve Lamp  Board Terminal  K  e e Con 107 ORN af  A  ORN S        BRN s    Distributed to general electronics 5     G  BRN  Con 102 ier  i Valve    TP301  5VDC Lamp  N  Rect  R301    TP302  5VDC Lamp GND Con 105  ne R P305  12VDC Flush  for all pC    TP306  12VDC GND    WW        Valve   Voltages  e TP303  5VDC Main Circuit  Value    579  P304 DN  Control Board  Ripple not e TP307  15VDC  greater than  amp  P308 A CND  10mv p p     TPS09    15VDC  Power Supply Board  may be measured here  x   Lamp Voltage measured at the Terminal        29  6    3 0 VDC       Service Handbook for 1047 Refractive Index Detector   12 2001    35    Table 2    Hardware Information    Power Supply Area Wiring Diagram    Figure 12 shows details the wiring of the power supply area  This figure  should be used when a power supply component needs to be replaced        Legend Power Supply Wiring       Term Description   1 To TEMP AND MOTOR CONTROL board CON202   To MOTHERBOARD CON113    To MOTHERBOARD CON110        Orange  to top bottom right posts of TEMP switch   Gray  to transformer    White  to top left post of TEMP switch    Gray  to bottom left post of TEMP switch     Orange  from top post of voltage selector to transformer     oOo ama N oon A WwW N    Black  from middle post of voltage selector to bottom right post of LINE switch     10 Blue  from bottom post of voltage selector to transformer     a  Sai    White  from top left post of LINE switch to line fu
73. llowing boards   a TEMP AND MOTOR CONTROL board  b OPTICAL AB board  c AB board  d POWER SUPPLY board  e KEY SWITCH board    Make sure that the motor is turning the optical unit  balance glass control shaft  Replace or adjust the  coupling sleeve  Then  perform the alignment procedure  for the AB motor assembly     Make sure that the flow cell is free of air bubbles  Make  sure that the INTENSITY ERROR is not present  Purge the  cell of bubbles if needed     If the detector intensity is normal  replace the AB board  and OPTICAL AB board     Does the front panel BALANCE button work correctly     If yes    1 Make sure that the remote signal being applied is a  contact closure  normally open  closed for BALANCE    2 Check the OUTPUT board ribbon cable   in this case   other remote control functions are likely to also be  defective    3 Replace the OUTPUT board     If no   Replace the AB board and the OPTICAL AB board     Does the rear panel BALANCE remote input work  correctly        114    Service Handbook for 1047 Refractive Index Detector   12 2001    Troubleshooting Information             Table 10 Limit alarm  AB motor and balance problems  Possible Problem Actions  If yes     1 Check the KEY SWITCH board ribbon cable  In this  case  other front panel functions are likely to also be  defective    2 Replace the KEY SWITCH board     If no   Replace the AB board and the OPTICAL AB board     The AB motor reaches it s mechanical limit  but the LIMIT 1 Replace the OPTICAL AB boar
74. ment    on page 83     Check for corroded contacts in the temperature selector switch  refer to  Table 15 on page 131   Clean or replace the selector switch     Check for poor quality solder connections of the Photo Diode wires at the  Pre Amplifier board  Carefully re solder the connections        126    Service Handbook for 1047 Refractive Index Detector   12 2001       Parts Information    This chapter provides parts information for the  1047 Refractive Index Detector       Parts Information    This section provides diagrams for parts identification and the complete  parts listings respectively     e Major Components   e Major Assemblies   e Front Panel Assembly  e Rear Panel Assembly  e Optical Unit   e Cables   e Inlet Outlet Tubes    e Screws       NOTE Part numbers have been updated as of 12 2001  Some part numbers are  marked bold with    These may not be available  Contact your local Agilent  support office in case of part number issues or upgrades           128 Service Handbook for 1047 Refractive Index Detector   12 2001    Table 14    Parts Information       Major Components    The following table identifies the major components of the 1047A  with top  cover removed  as shown in Figure 40 on page 130        Major Components                  Description ERC number Part number  N A Optical Unit see Table 15  on page 131  1 LED Board 1 ERC75008A 01047 66510  2 Key Switch Board ERC75006B 01047 66509  3 Display Select Switch Board ERC75013 01047 66514  4 Digital Displa
75. nings elsewhere in this manual violates  safety standards of design  manufacture  and intended use of the instrument   Agilent Technologies  former Hewlett Packard  Company assumes no  liability for the customer   s failure to comply with these requirements     General    This is a Safety Class I instrument  provided with terminal for protective  earthing  and has been manufactured and tested according to international  safety standards     Operation   before applying power    comply with the installation section  Additionally the following shall be  observed     Do not remove instrument covers when operating  Before the instrument is  switched on  all protective earth terminals  extension cords   auto transformers and devices connected to it should be connected to a  protective earth via a ground socket  Any interruption of the protective earth  grounding will cause a potential shock hazard that could result in serious  personal injury  Whenever it is likely that the protection has been impaired   the instrument must be made inoperative and be secured against any  intended operation     Make sure that only fuses with the required rated current and of the specified  type  normal blow  time delay  etc   are used for replacement  The use of  repaired fuses and the short circuiting of fuse holders must be avoided     Adjustments described in the manual are performed with power supplied to  the instrument while protective covers are removed  Energy available at  many points may  
76. nstall the replacement OUTPUT board  Re attach the holding clamps and  re connect both ribbon cables     Turn on the detector  turn on should be normal  Verify that each function  of the OUTPUT board works correctly     No electrical calibrations are necessary  However  refer to    Verify the  Operation of the Baseline Controls and Autobalance    on page 96  step 2   for information on when OUTPUT board VR701 may need adjustment        Service Handbook for 1047 Refractive Index Detector   12 2001 67       NOTE       Figure 27                   Procedures    Motherboard    1  2  3  4    Remove the six boards from the detector card cage   Disconnect all cables from the motherboard   Remove the six screws which secure the motherboard and remove it     Install the replacement motherboard and secure it with it   s six screws   Re connect all cables  refer to    Motherboard Layout    on page 39  if  necessary     Replace the six card cage boards  then check for proper turn on and  operation  No calibrations are needed     Boards and Controls located in the front panel    1  2    Remove the detector top cover     Refer to    Major Assemblies    on page 134  Remove the side panels  items  3 and 15  by removing screws item 4  Disconnect the tubes connected to  the IN OUT block  item 17      Re connecting these tubes incorrectly will destroy the flow cell  Refer to  Figure 29 on page 71 for correct plumbing     Figure 27 represents a top view of the detector with the detector top cover
77. nt Panel Assembly 01047 60002  2 Bottom Casting 01047 24102  3 Top Cover 01047 24103  4 Screw  M5 9 3 mm long 0515 0750  5 Main Power Transformer 01047 61101  6 Optical Unit   with AB motor 01047 60017  without AB Motor 01047 60018  7 Shock Absorbing Foot 4040 2089   8 Three Way Fitting 01047 27601  9 Capillary Tube Assembly 01047 60010  10 Capillary Tube Assembly 01047 60011  11 Preamplifier Board ERC75100H 01047 66502  12 Optical Balance Motor Assembly  includes 01047 60016  complete Assembly plus holding bracket   12a Cable 01047 61608   13 Support Stand Assembly 01047 60012   14 Leak Pan 01047 43201   15 Side cover  left 01047 04103  16 Bulkhead Nut 01046 25701          Service Handbook for 1047 Refractive Index Detector   12 2001 131    Table 15    Parts Information       Major Assemblies         Description    ERC number Part number       17 Capillary Holder   18 1 16  ZDV union  INLET    19 Metal Floor   20 Lamp Valve Terminal Assembly  22a Fuse 2A   21 Motherboard   22 Temperature and AB motor Control Board  23 Power Supply Board   24 Optical Auto Balance Board   25 Auto Balance Board   26 Control Board   27 Amplifier Board   28 Top Cover   29 Rear Panel Assembly    30 Hex Set Screw   31 PC Guide Rail   32 Screw hex  M6 30 mm long  5 mm key  Flat Washer   33 Preamp Support Plate   34 Preamp Plastic Cover   35 Shaft Connection Sleeve   36 Wide bore 1 16    union     outlet      37 Screw pan pozi  M4 6 mm long    Lock washer    ERC75201A    ERC751  ERC751    02A  03A   
78. o  equilibrate at 40  C for ninety minutes     90 minutes yet     Refer to Figure 32 on page 81 preceding  Set both the optical balance glass  and the mirror adjust glass to be perpendicular to the light beam   best  estimation by eye is ok  Note that the mirror adjust glass movement is  accessed through a hole in the side of the optical unit  This hole may be  reached by removing the front panel cosmetic plug  Figure 42 on page 137   item 3   For this purpose  use aa long  flat blade screwdriver with a blade  no more than 4 mm wide     Figure 35 shows the location of the photo diodes inside the optical unit   The photo diodes can be seen by looking from the rear of the optical unit  as shown  The Figure also shows the possible movements of the mirror  during adjustment        86    Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures    Figure 35    Location of Photo Diodes        Glasses    Front View                            Top View  Photodiode Mirror  Holding      Screw He    JA Pitch    R Fis j G    Halding          Screw wy    5 Turn on the detector  and look at the photo diode area  By hand  move the  mirror such that the light beam falls equally on the photo diodes as shown    in Figure 36  Before you start  read the general information below for an  overview of the job  and some helpful hints        Service Handbook for 1047 Refractive Index Detector   12 2001    Figure 36    The basic task    Helpful hints    Procedures    Light beam on Pho
79. o be perpendicular to the  beam  as shown in the figure  The optical balance glass is then in it s    null                                                                      position   Figure 14 Relationship Optical Balance Motor Optical Unit  manual shaft optical balance optical mirror  turning knob glass control balance adjustment  shaft glass glass  tical   01777 c  z  optica l l Pa beam ol    O   shaft  balance   L presne ie    egm   Se  motor ena cig   sleeve  coupling i             Optical Unit Partial Top View    Figure 15 shows the electronic control circuits of the optical balance motor     Legend for Figure 15       A Commands for motor drive circuit  B Motor drive pulses  C Actual motor position signal          40 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Figure 15 Control Circuits of Optical Balance Motor          AB Motor Assy   Temperature and Mator    Control Board                                                                                  C  C  9  Er cs TP205    oter Drive  Mtr  Pos  Sensor 5 Circui  oO   TP204  NAVV  TTR  12                                                                                                                                                                                                             r 12 j  oc  Con 101 p  Na      Rey   Con 119  Con 109  Motor Drive      Motor Sense 5  Con 903 A ane ww  Pulse Ckts   a VRZ  e TF1101  Q Limit Alarm  LED Board  2             Limit Ala
80. oblems which interfere with accurate  quantization  and even with automation     Another severe limitation of RI detectors is that they are impractical to use in  gradient analysis  It is extremely difficult to adequately match the RI of  sample and reference streams during gradient formation        Optimizing your 1047A    on page 123 describes certain key points to  observe when optimizing the 1047A system for best performance        122 Service Handbook for 1047 Refractive Index Detector   12 2001    Solvent Delivery  System    Inlet Tubes    Outlet Tubes    Solvents    Troubleshooting Information    Optimizing your 1047A    In this section we describe some practical hints which can help you optimize  the use of your 1047A     e The solvent delivery system of your liquid chromatograph must produce a  very stable and pulse free flow  Pressure pulses in the flow stream cause  baseline noise and drift     e Ifthe solvent delivery system of your liquid chromatograph has a  multichannel capability  the mobile phase should be pre mixed in a  reservoir and pumped from one channel only     e With the 1090 HPLC  we recommend that the unused solvent delivery  channel lines are flushed with the same solvent for your analysis  This  eliminates the possibility of baseline instability caused by small amounts  of different solvent leaking into the mobile phase stream     e The flow rate can affect baseline stability  Noise caused by eddy currents  in the flow cell increases as flow rate 
81. og output at the rear panel   The 10 mV recorder output is taken to both the digital display and the rear  panel  The voltage RIU relationship for the recorder outputs is determined by  the RANGE SWITCH setting     Important points in the signal path block diagram     e When a BALANCE is done  the ELECTRICAL AUTOBALANCE board does  set the output of ELECTRICAL AB CIRCUITS to zero  However  the actual  signal outputs of the detector will be whatever value has been set by the  position of the front panel FINE ZERO and INTEGRATOR dials     e The front panel FINE ZERO dial influences both RECORDER and  INTEGRATOR outputs  The front panel INTEGRATOR dial influences only  the INTEGRATOR output     e At the display  selecting functions TP1 and TP2 displays the actual signal  present at PREAMPLIFIER board TP1 and TP2 respectively     e At the display  selecting functions TP1 and TP2 simultaneously displays  the average of TP1 and TP2        Service Handbook for 1047 Refractive Index Detector   12 2001 31    Hardware Information       Power Distribution    AC Power Distribution    Figure 10 shows details the AC power distribution of the 1047A  AC  secondary voltages may be measured at motherboard CON110 as shown  The  100 V AC secondary for the heater may be measured at CON202 of the TEMP  and MOTOR CONTROL board  The rear panel TEMP switch must be on for  this measurement        NOTE The present AC power supply only employs two primary windings  115 V AC   orn  and 240 V AC  blu  as 
82. om the original flow cell assembly is enough to use on the  new assembly     10 Carefully press the new flow cell assembly into place in the optical unit   Take care to position the flow cell tubes correctly  Take care to replace the  thin pieces of insulation which originally covered the capillary tubes   Secure the flow cell in place with the two flow cell holding screws     Incorrect plumbing will destroy the flow cell     11 Re plumb the flow cell assembly capillary tubes as shown in Figure 29 on  page 71     Correct routing of the stainless steel flow cell capillaries is important for good  baseline stability  Poorly routed capillaries can put strain on the flow cell   causing excessive baseline drift and wander     12 Carefully replace the mirror  Screw the two mirror holding screws in to       84 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE          NOTE       Procedures    place  but do not tighten them down at this time     13 Perform the    Mirror Adjustment    on page 85 procedure     Mirror Replacement    The lamp should be correctly adjusted at this time  If it is not  refer to    Lamp  Replacement Intensity Adjustment    on page 80     If the mirror has not yet been removed  begin with step  1  below  If the  mirror has already been removed  and is now loosely back in place  begin  immediately with    Mirror Adjustment    on page 85     1 Remove the detector top cover  If needed  refer to Figure 40 on page 130  to identify the optical
83. one test pin on the switch  board  labeled TP1  the label may not be present on very old detectors    TP1 is used for the DIGITAL DISPLAY BOARD span calibration     Measure TP1  with respect to chassis GND   At the front panel  use the   COURSE ZERO button to make TP1 read  180 mV     VR1201 is the only potentiometer on the DIGITAL DISPLAY BOARD  By  adjusting VR1201  make the INT display read  1800 mV  In other words  the  INT display should read 10 times the value of TP1  with the opposite  polarity        60    Service Handbook for 1047 Refractive Index Detector   12 2001       Procedures    This chapter provides procedures for the 1047  Refractive Index Detector       Procedures       General Information   Read this first     The figures in    Parts Information    on page 127 may be used as needed as  instructions for the remov al of certain 1047A sub assemblies and parts   Several of these procedures refer to this chapter     The top cover of the 1047A is removed as shown in Figure 25  The top cover  must be removed for most of the procedures described herin                                                           Figure 25 Top Cover  Top Top  Screw Screw  OD  p ee      Screw DcCrew             Front       62 Service Handbook for 1047 Refractive Index Detector   12 2001    Procedures    e ALWAYS DISCONNECT POWER before removing any board or  disconnecting any cable in the 1047A  This basic advice will be assumed  for all written procedures     e Remember that solv ents 
84. or shaft should still be adjusted to two full turns  counter clockwise  with the INT and REC OUTPUT SPANS correctly  calibrated     2 Locate VR503 and TP3 on the CONTROL board  Set up the DC Voltmeter  to measure TP3 with respect to chassis GND     3 Adjust VR503 so that TP3 reads  0 5 VDC     The detector Autobalance circuit is now adjusted so that the  Electrical Optical Autobalance decision threshold is 4 x 10 5 RIU     Verify the Operation of the Baseline Controls and  Autobalance    1 The AB motor shaft should still be adjusted to two full turns  counter clockwise  The front panel INTEGRA TOR dial should be set fully  clockwise and the FINE ZERO dial should be at it   s mid position  5 turns    The RANGE knob should be set to 32     2 Press BALANCE  After the balance  both the REC and the INT displays  should be  theoretically  0 00  The FINE ZERO dial  which affects both  INT and REC  should provide similar baseline range in both the   and    direction  The INTEGRATOR dial  when turned fully counter clockwise   should add  10 mV to the existing value of the INT display  The REC  display is not affected by the INTEGRATOR dial     If the relationship between the INT display and REC display is not correct   refer to the procedure    Calibration of OUTPUT board VR701    on page 97     3 Select the INT display and use the baseline control dials to set the INT  display to a baseline value of 0 00     4 Press and hold the   COURSE ZERO button until the INT display reads  betwe
85. osition  No electrical calibrations are needed  Both poles of SW201 should  be OFF     Power Supply Board    Remove the defective board and insert the replacement board  No electrical  calibrations are needed     Optical Auto Balance Board  DO NOT TOUCH VR2     Remove the defective board and insert the replacement board  No electrical  calibrations are needed        64 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE          NOTE          NOTE          NOTE       Procedures    Auto Balance Board    DO NOT TOUCH VR401     Remove the defective board and insert the replacement board  No electrical  calibrations are needed  Both poles of SW401 should be OFF     Control Board  DO NOT TOUCH VR501     Remove the defective board and insert the replacement board  Make sure to  connect the ribbon cable  from the pre amplifier board  to CON501 correctly   Perform the complete program of electrical calibrations described in     Electrical Calibrations    on page 93     Amplifier Board  DO NOT TOUCH VR615  608  607  614  605 or 604     Remove the defective board and insert the replacement board  Perform the  complete program of electrical calibrations described in    Electrical  Calibrations    on page 93     In older detectors  the original AMPLIFIER board may have 14  potentiometers  The current AMPLIFIER board has only 8 potentiometers   The current AMPLIFIER board is a direct replacement for the original board     Preamplifier Board    1 Disconnect the Preampl
86. presentative of light intensity  and are useful in detector troubleshooting     The intensity voltages are first taken to the CONTROL board  From the  CONTROL board  Figure 4  they are taken directly to the front panel    DIGITAL DISPLAY  where S or R can be selected as TP1 or TP2  respectively   at the display     Control Board    to Electrical to Optical  AB Board AB Board    to Display to LED Board  2          MB Con 111 MB Con 109                         Con 104    Control Board                                                                            Int  TP2  Error                Buffer e    Soca  a AVG  Preamplifier  Board  TF 1      e me      D  TP2 5 p Detector Signal  R   x       Ly Path begins here  VR501  VR503 AR TP3             3  Decision gt              ___ Buffer                               Next  the average intensity  typically between  4 0 and  5 5  is taken to the  CONTROL board INTENSITY ERROR threshhold circuit  Measuring TP2   VR502 is adjusted such that the INTENSITY ERROR is set if the original     factory  intensity changes by  approx   25   TP4 is normally HIGH  and  LOW for error        24 Service Handbook for 1047 Refractive Index Detector   12 2001    Figure 5    Hardware Information    The detector signal path actually begins at the output of the DIFF AMP  With  the detector on and the lamp wires disconnected from the LAMP VALVE  terminal  the offset of the DIFF AMP should be 0  5 mV  measured at TP1   adjusted with VR501  The output of the D
87. ptical Unit Housing 01047 25201  12 Optics Cover 01047 04105  13 Optical Unit Top Cover 01047 04106  14 Teflon Drain Tube 5021 7115  15 Leak Pan Spring 01047 29101   16 Lamp socket 01047 25202  17 Holding screw for top cover and optics housing  total 12   M3 0515 0924   6 mm long   Washer 3050 0891       18 Holding screw for optics cover  lens slit assembly  not shown  0515 0912    and cell assembly  total 10   M3 8 mm long    Washer 3050 0891  19 Mirror holding screw  total 2   M4 15 mm long 0515 0909          Service Handbook for 1047 Refractive Index Detector   12 2001    141    Parts Information             Table 18 Optical Unit    Description Part number  Small washer 2190 0586  Large Washer 3050 0894  20 Support plate holding screw  total 4   M4 18 mm long  0515 1073  flat pozi    21 Mirror pitch set screw  total 2   M4 hex 4 mm long  2 mm key 0515 0529     22 Set screw for photo diode assembly and lamp holding collar 0515 0154   total 3   M3 hex 3 mm long  1 5 mm key          142 Service Handbook for 1047 Refractive Index Detector   12 2001    Parts Information       Cables    General Refer to the 1047A Operator   s Handbook  01047 90000  for specific  information on the connection of cables  At the 1047A  all inputs and outputs  are terminal connectors which accept    spade    type fittings or even bare                         wires   Signal Cables for  Hewlett Packard Data  Handling Devices  Table 19 Signal Cables   1047A to others  Description Part number  HP 3390 2
88. r     10 Remove the AB Motor assembly from the optical unit  It will eventually be  attached to the replacement optical unit  see Figure 30 on page 73         78 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE          NOTE       Procedures    11 Carefully place the new optical unit into the detector  At this time  secure  the optical unit by replacing the four nuts which secure the optical unit  support feet  The two locking screws should be screwed into their holes   but not tightened     12 Carefully re solder to the preamplifier board the three wires of the optical  unit photo diode assembly cable  If you did not make a note of the correct  position of the three wires  refer to Figure 26 on page 66  Then  replace the  safety cover     13 Reconnect the optical unit leak drain tube to the leak drain fitting at the  rear of the optical unit     14 Connect the heater and temp sensor of the replacement optical unit to  their positions on the Temp and Motor control board     15 Connect the lamp wires to the VALVE LAMP terminal block     16 The replacement optical unit is now installed in the detector  At this time   turn on the detector  Turn on should be normal  If the LIGHT INTENSITY  ERROR is on  it will be necessary to re plumb the flow path so that the flow  cell can be flushed to eliminate the LIGHT INTENSITY ERROR  If the  LIGHT INTENSITY ERROR is not on  we will re plumb the flow path later     Incorrect plumbing may destroy the flow cell  Make 
89. r   12 2001       NOTE       Hardware Information       Recorder Control Electronics    Figure 23 on page 54 shows the electronic control circuits of the optical  balance motor     is a functional block diagram of the recorder control electronics     All recorder outputs circuits reside on the AMPLIFIER board  Some of these  circuits are adjustable on the board  The recorder output circuits are  controlled by the recorder control circuits of the CONTROL board     The RANGE switch sets the sensitivity of the detector  signal voltage vs   RIU   Refer to the PERFORMANCE VERIFICATION section for details     The POLARITY control changes the polarity of both recorder outputs from  positive to negative  The POLARITY INDICATOR led is ON for negative  polarity  This control does not affect the integrator output     The RECORDER control  when pushed to    zero     disables  short to GND   both recorder outputs  This allows a strip chart recorder to be mechanically  zeroed without disconnecting the detector signal cable  The RECORDER  indicator LED is ON for this mode     Make sure that the RECORDER indicator LED is OFF for normal operation     The MARKER button applies a 1 mV short duration offset to both recorder  outputs  This produces a deflection on the strip chart to mark the beginning  of a chromatogram  MARKER may also be initiated from the rear panel  remote control inputs        Service Handbook for 1047 Refractive Index Detector   12 2001 53    Hardware Information           
90. rm  Ckt   e TP1102                   Optical AB Board                               Con 115 Con 114                         Con 601                Con 703                   Course Zero O lt     Input Row            Balance          Key Switch Board Output Board             Service Handbook for 1047 Refractive Index Detector   12 2001 41    Hardware Information    The motor can be moved either by the front panel   and   COURSE ZERO  buttons or  under certain conditions  by initiating a BALANCE     The OPTICAL AUTOBALANCE board produces the initial drive command  pulses for the motor  The correct frequency of these pulses  250 Hz  10 Hz  is  critical for proper operation of the motor  Frequency can be measured at  TP1101  normally LOW  pulsing 250 Hz for motor operation  and adjusted  with VR2     The motor driver circuits of The TEMP AND MOTOR CONTROL board  receive the drive command pulses from the OPTICAL AUTOBALANCE  board  On the TEMP AND MOTOR CONTROL board  the operation of the  motor driver circuits can be measured at TP204  normally LOW  pulsing   250 Hz for motor operation  TP205 should be constantly  5 VDC  TP204 205  are near CON201  Finally  the drive command pulses are converted to    12 VDC pulses and delivered to the windings of the motor     The actual mechanical position of the motor is sensed by the motor position  sensor potentiometer circuitry on the OPTICAL AUTOBALANCE board  If  the motor drives too far in either direction  the front panel LIMIT ALARM  in
91. se   White  from line fuse to left post of line filter    Brown  from bottom left post of LINE switch to transformer   Black  from top right post of LINE switch to line fuse    Black  from line fuse to right post of line filter     Grn Yelllow  to ground           a a a a a  N o on A ow N    Red wire and Yellow wire  un used           36 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Power Supply Wiring    Figure 12                            37    Service Handbook for 1047 Refractive Index Detector   12 2001    Table 3    Hardware Information       Motherboard    Figure 13 shows the lay out of the motherboard  Also shown is legend  referred to in the cables termination table     There are no like connectors on the motherboard  This greatly reduces the  possibility of connection errors  Each individual connector is keyed so that  it   s cable may be inserted only in the correct way  The PC board connectors  are also keyed so that a board may not be inserted in the wrong location  Be  careful to note  however  that it is possible to remove these keys from the  motherboard connectors        Cables termination       Term Description       Motherboard holding screws  six total    To the REGULATOR SET mounted on the rear panel    To the main power transformer  secondaries connector   To the AB MOTOR assembly    To the VALVE LAMP terminal    To the GROUND connector cable attached to the rear panel   TEMPERATURE AND MOTOR CONTROL board   POW
92. se the FINE ZERO dial to make the INT display as close to 0 00  as possible  The detector output is now nulled        Service Handbook for 1047 Refractive Index Detector   12 2001 93       NOTE       Figure 39    Front    Procedures    Calibrate the LIGHT INTENSITY ERROR THRESHOLD    1 The detector output should still be nulled     2 Refer to Figure 39 on page 94  In the detector  locate the CONTROL board   Locate VR502 and TP2 on the CONTROL board     On the CONTROL board  the test points extend out from the rear of the board  for easy access  With a DC Voltmeter  prepare to measure TP2 with respect to    chassis GND  The 20 VDC range is sufficient     3 Adjust VR502 so that TP2 reads 7 5 VDC  The detector is now calibrated  such that the LIGHT INTENSITY ERROR will come on if the current light  in tensity  as seen by the photo diodes  deviates by approximately 25     from the original factory in tensity     Location of Boards                                  Lamp    eos                      Amplifier  Board       Control  Board                      VR   VR   VR  502   501  B04       0  Oo oO D  TP2 TP1 TPA TPS                                     2                                                                                                                   Rear       94    Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Procedures    Calibrate the Integrator Output Span    1  2    The detector output should still be nulled     Locate the man
93. shown  In older instruments  a rotary voltage  selector switch was used which employed four primary windings  100 V AC   115 V AC  220 V AC and 240 V AC  In the current wiring scheme  the 100 V AC  wire  red  and the 220 V AC wire  yel  are not used  These wires are capped  and tied out of the way           32 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information                                                                                                                                              Mother    Board  n    a    C  1    Oo  N    Power    Supply  Board                                                                                     Figure 10 AC Power Distribution    Mother  Rear Panel Main r veel 1 Board  Power Line ON OFF Power B 8VAC Tytoletl Con  Receptacle Switch Transf   P 2  10  and Line ransformer yellow  z 0  Filter E 8VAC  yellow 5  white white bro hit  Neutral ak Lown   c wns 5  Line Fuse E E  Eer black       pe white   n  Ground si 75 oe Orage SFOND j  vac    Hot  gt  A BVAC  black black 240 z  VAC     blug     black 8    E BVAC  ae brown 9  Selector    shown set     for 115VAC     100  VAC  x  gray  gray    to CON 202 S  hit  pep ws Temp  Control Board   Part of  l  gt  Temperature  J  orange s to CON 113 Control Circuitry  EE Mother Board  Rear Panel Temp   ON OFF Switch          Service Handbook for 1047 Refractive Index Detector   12 2001    33    Hardware Information    DC Power Distribution    Figure 11 provid
94. specification   range range setting   RecFS recorder full scale in mm       For the following examples we assume a recorder FULL SCALE  DEFLECTION of 250 mm and a correct detector recorder matching     Verifying the Noise 1  Specification 2    3    Verifying the Drift  Specification    j  i    Noise Specification   2 5 x 10   RIU P P  equivalent to 5 x 10     RIU P P    1047A RANGE setting to 1 64 x 10   RIU  equivalent to 156 25 x 10   RIU      Calculate      5  Noise   oI RIU somm 8mm    156  25 x 10     RIU  Allowable noise at range setting 1 64 is 8 mm P P  Drift Specification  2 5 x 10  T RIU hr   1047A RANGE setting to 1 64 x 10   RIU  equivalent to 1 5625 x 10       Calculate     2 5x 10     RIU    Drift   5  1  5625 x 10      RIU    x  250mm  400mm    Allowable drift at range setting 1 64 is 40 cm hr  or 7 cm over 10 minutes        106 Service Handbook for 1047 Refractive Index Detector   12 2001    Verifying the Wander  Specification    Example  Calculating  the noise    Example  Calculating  the Drift    Performance Verification    Wander Specification   1 x 10   RIU  equivalent to 2 x 10 8 RIU    1047A RANGE setting to 1 64 x 10 5 RIU  equivalent to 15 625 x 10       Calculate      8  Wander   AXU RIUS x  250mm  32mm    15  625 x 10     RIU    Allowable wander at range setting 1 64 is 32 mm     For the ChemStation    In the ChemStation  full scale range is set in pV     1    Oo Ne  WO WD    The 1047A integrator output  1 x 10  gt  RIU   32000 pV  is connected to th
95. specification  M verifying the drift specification  C1 verifying the wander specification  e For the recorder outputs  M the range knob and recorder outputs   1 the recorder full scale deflection setting  m calculating recorder deflection  CI verifying the noise specification  M verifying the drift specification  M verifying the wander specification    e For the ChemStation       m calculating noise and drift       100 Service Handbook for 1047 Refractive Index Detector   12 2001    Calculating  Allowable Integrator  Deflection    Performance Verification       1047A Specifications Verification  The specifications of the 1047A can be found in    Specifications    on page 15     For the Integrator Output    The integrator output of the 1047A is calibrated such that 32 mV equals  1x 10   DELTA RIU  Verifying 1047A specifications is a question of  converting the specifications to an allowable deflection  noise  drift or  wander  on the integrator at a specific integrator attenuation     The formula for calculating allowable deflection in mm at a specific  integrator attenuation          Spec  Deflect      eflection  RIU   mm   where   Spec specification of detector  Deflection allowable deflection in mm  RIU mm scaling factor at a specific attenuation       See Table 6 on page 102 for the RIU scaling factor of Hewlett Packard  integrators  For other integrators  refer to the manual for that integrator   Look up the signal input required for FULL SCALE DEFLECTION at the  specific
96. st been replaced  If so  has the  protective clear tape around the cell been removed  If  not  remove the tape     See    Lamp Turn Problems    on page 113     Has the lamp just been replaced  If so  refer to the    Lamp  Replacement Intensity Adjustment    on page 80  procedure     Have any optical unit repairs just been made  Review the  calibrations that may be needed when repairs are made  to the optical unit     1  CONTROL board   2  DIGITAL DISPLAY board     Connect the ribbon cable securely           112    Service Handbook for 1047 Refractive Index Detector   12 2001    Table 9    Troubleshooting Information    Lamp Turn Problems       Lamp Turn Problems       Possible Problem    Actions       The lamp does not turn on when the line switch is turned Lamp is defective  Replace the lamp and perform the lamp  on  the detector itself turns on normally   intensity adjustment     Lamp voltage is absent  Measure the DC voltage supply to  the lamp wires at the valve lamp terminal  Should be in  the range of 2 5 to 3 0 VDC     If the supply is absent   1  Replace POWER SUPPLY board   2  Replace POWER REGULATOR set        Limit Alarm  AB Motor Drive and Balance Problems    These three areas are very closely related in the detector electronics  The AB  motor position is sensed by a feedback potentiometer which is connected to  the motor shaft  If the motor is driven too far in either direction  the front  panel LIMIT ALARM indicator is turned on  From the LIMIT ALARM state   the
97. sure that the flow path is  plumbed as shown in Figure 29 on page 71     17 The original AB motor assembly may now be installed on the replacement  optical unit  Refer to    AB Motor Assembly Replacement and Adjustment     on page 73  Replace and adjust the AB motor assembly as described     18 Replace the left and right detector covers  If you have not already done so   re plumb the flow path     Incorrect plumbing may destroy the flow cell  Make sure that the flow path is  plumbed as shown in Figure 29 on page 71     Turn on the detector  and set detector temperature to 40  C  After the detector  has equilibrated at 40  C for 90 minutes  perform the complete electrical  calibration program described in    Electrical Calibrations    on page 93        Service Handbook for 1047 Refractive Index Detector   12 2001 79       NOTE       Procedures       Inside the Optical Unit    Lamp Replacement Intensity Adjustment    The replacement lamp assembly  01037 60002  consists of the bulb and it   s  holder  The bulb is pre set in to the holder  but the position of the bulb MAY  need re adjustment  as described in the intensity adjustment procedure  below     The replacement bulb should be cleaned with alcohol and clean cloth  Care  must be taken not to get fingerprints on the bulb     1 Remove the detector top cover  If needed  refer to Figure 40 on page 130   to identify the optical unit and lamp     2 Refer to Figure 44 on page 140  Remove the optical unit top cover  item  13   insu
98. t  Intensity                  Circuit                                                                               48 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Block Diagram Balance Not Ready  Output Board  Output Row      gt     Balance  Not Ready    Mother Board Cot    Auto Balance Board    Relay Driver          Service Handbook for 1047 Refractive Index Detector   12 2001    49    Hardware Information    The Inputs    Balance    A BALANCE is initiated by applying a short to the BALANCE pin pair  or by  the front panel BALANCE button     Figure 20 Block Diagram Balance Input                Output Board    Balance       Input Row lr 7                                                                                                                   CON 703  Front Panel Front Panel  Balance Balance  Button LED  CON 114 CON 115 Mother  Y E Hl Board  C   Auto Balance Optical Auto  Board Balance Board                                                                                                                   50 Service Handbook for 1047 Refractive Index Detector   12 2001       NOTE       Figure 21    Hardware Information    INTEGRATOR POLARITY  The INTEGRATOR POLARITY can only be changed by applying a short to the    POLARITY pin pair  Alternating shorts change INTEGRATOR POLARITY  from plus to minus  The current INTEGRATOR POLARITY is shown under    the front panel digital display     This is not to be confused with the record
99. tal Display Board  8 Range Switch Board Assembly 19 Holding Screws for Key Switch Board   9 Ribbon Cable  to Motherboard CON116 20 Holding Screws for Display Switch Board  10 to Motherboard CON117 21 to Motherboard CON108   11 Fine Zero Adjust Assembly 22 to VALVE LAMP Terminal Block             Service Handbook for 1047 Refractive Index Detector   12 2001 69       NOTE       Procedures    Use Figure 28 on page 69 as a guide for the replacement of a particular front  panel component as described in the two generalized procedures below     TEMPERATURE SELECT SWITCH assembly  RANGE  SWITCH BOARD assembly  FINE ZERO ADJUST assembly  or INTEGRATOR ADJUST assembly    1    Disconnect any cables associated with the assembly to be removed  Refer  to Figure 28 on page 69 for the locations of these cables     Refer to Figure 42 on page 137  Working from the front  remove the  appropriate dial  item 1  and nut  The assembly may now be removed from  the front panel     Install the replacement assembly and re connect any cables associated  with the assembly     No calibrations are needed  Re assemble the detector and check for  proper operation     LED BOARD 1  KEY SWITCH BOARD  LED BOARD 2   DIGITAL DISPLAY BOARD or DISPLAY SWITCH BOARD    Do not touch the position of any potentiometers     Disconnect any cables associated with the board  Refer to Figure 28 on  page 69 for the locations of these cables     Remove the screws which secure the board  Figure 28 on page 69  and  remove the bo
100. tc   is eliminated     Most problems in the RI system are chromatographic in nature  and it is  important to determine what contribution the liquid chromatograph is  making to the problem        NOTE Certain defects in the flush valve can resemble flow problems  Refer to    Flush  Valve Problems    on page 116 to see if your specific symptom is described           NOTE Defects in the flow cell can resemble chromatographic problems  Refer to     Flow Cell Problems    on page 121 to see if your specific symptom is  described        Refer to section 5 of the 1047A Operator s Handbook  01047 90000  for  specific information        Service Handbook for 1047 Refractive Index Detector   12 2001 125    Trou    bleshooting Information    Troubleshooting an Instable Baseline    Excessive Noise    1  2    Make sure the optical unit locking screws are loose     Make sure the laboratory bench is absolutely free of mechanical  vibrations     Make sure that lamp intensity is OK  The average intensity  TP1 and TP2  pressed together  must be between  4 5 and  6 0  Values more negative  than  6 0 can cause excessive noise     Excessive Drift or Wander    1    The environmental temperature must be stable  no open windows  no  excessive air conditioning or ventilation  etc     The column and column capillary must be shielded from temperature  changes and excessive air flow     Make sure that the flow cell stainless steel inlet outlet capillaries are  installed correctly  See    Flow Cell Replace
101. tion    b CON108 on the MOTHERBOARD  This is the  connection from the TEMP SELECT SWITCH  assembly     2 Check the integrity of the rear panel TEMP switch    3 Check resistance through the temperature sensor   if  open  replace the Optical Unit    4 Replace the TEMP AND MOTOR CONTROL board        Service Handbook for 1047 Refractive Index Detector   12 2001 119    Troubleshooting Information       Table 12 Temperature Control Problems       Possible Problem Possible Causes and Actions       5 Replace the TEMP SELECT SWITCH assembly     The front panel TEMPERA TURE indicator led turns on 1 Check the following connections for tightness   normally  but never begins to flash  This means that the a CON202 on the TEMP AND MOTOR CONTROL board   actual temperature  as measured by the temp sensor  This is the input for the 100 V AC heating power   never reaches the setpoint  b CON201 on the TEMP AND MOTOR CONTROL board   This is the connection to the Optical Unit heater   2 Check the 2 A fuse on the TEMP AND MOTOR  CONTROL board  Replace fuse if open   3 Check for 100 V AC heater voltage at CON202 of the  TEMP AND MOTOR CONTROL board  This voltage  should be steadily applied as long as both rear panel  switches are on  If voltage is absent  check the  PRIMARY and AC POWER DISTRIBUTION areas   4 Check resistance through the Optical Unit heater  If  open  replace the Optical Unit   5 Replace the TEMPERATURE AND MOTOR CONTROL  board   6 Replace the TEMP SELECT SWITCH assembly   7 The Opt
102. to Diodes       Light  beam    Photodiodes    As you adjust the mirror by hand  use display TP1 and TP2  The goal of the  mirror adjustment is twofold  TP1 and TP2 should both be as high as  possible  and TP1 and TP2 must be as close together as possible     Average intensity  TP1 2 buttons pressed together  must be no less than  3 5   but you should strive for the maximum intensity  Typical intensities are from   4 0 to  5 5  When TP1 and TP2 are within 0 2 of each other  you may use the  mirror adjust glass  Figure 32 on page 81  to make them equal  DO NOT use  the mirror adjust glass to compensate for a difference greater than 0 2     It could be helpful to slightly tighten the mirror holding screws  then use a  lever  such as a small screwdriver or needle nose pliers  to adjust the  mirror  This provides for a finer resolution of adjustment     The mirror can move in several directions  as shown in Figure 35 on page  87  This means that a very clear procedure for moving the mirror is not  really possible  You must find the mirror position which best meets the  requirements of maximum intensity and photo diode equality     Once the correct position for the mirror has been established  tightening  the holding screws can be a problem  The position of the mirror tends to  change as the screws are tightened  To avoid this  place your finger on top  of the mirror and gently begin to press  If the position of the mirror does  not change  TP1 2 unchanged   then tighten the holdin
103. tting correct   Are all cables in place and tight     Measure test points on POWER SUPPLY board for correct  output     If test point measurements are NOT correct   a  Replace POWER SUPPLY board   b  Replace POWER REGULATOR set        Light Intensity Problems    When operating correctly  the LIGHT INTENSITY ERROR indicator on the  front panel will come on when the light intensity reaching the photo diodes  differs from the original  factory  intensity by about 25   Ordinarily  a LIGHT  INTENSITY ERROR is caused by a reduction in intensity  However  an  intensity increase  as seen by the detector signal electronics  will also cause  the error  The detector also gives a contact closure output  rear panel  while  a LIGHT INTENSITY ERROR is present     Actual light intensity can be displayed by TP1  sample photo diode  and TP2   reference photo diode  on the front panel  Pressing TP1 and TP2 together  will display the average in tensity of the two photo diodes     The average intensity must be at least  3 5  Typical intensity values range  from about  4 to  5 5  depending on optical unit cleanliness  age  and whether  or not certain optical alignments were ever performed on site        Service Handbook for 1047 Refractive Index Detector   12 2001 111    Troubleshooting Information       Table 8 Light Intensity Problems       Possible Problem    Actions       Air in flow cell  this is the most common cause of the  error  In this case  the intensity average as displayed by  TP1 2 w
104. ual shaft turning knob on the front of the AB Motor  assembly  if needed  refer to Figure 31 on page 74      As carefully as possible  turn the AB Motor assembly shaft exactly TWO  FULL TURNS in the COUNTER CLOCKWISE direction  As the motor shaft  is turned  the INT display should be increasing in the positive direction     Two counter clockwise turns of the shaft equals a beam displacement of   16 x 10 5 RIU  Mechanically introducing a known RIU is more reliable and  convenient than the classic method of preparing an RI solution for injection in  to the cell     Use Figure 39 on page 94 to locate the AMPLIFIER board  On the  AMPLIFIER board  locate VR613  VR613 is clearly labeled on the board     Adjust VR613 so that the INT display reads as close to 512 mV as possible   The tolerance for this adjustment is 512 mV  10      The INTEGRA TOR OUTPUT SPAN is now calibrated such that 32 mV  equals 1 x 10   RIU     Calibrate the Recorder Output Span    1    The AB Motor shaft should still be adjusted to two full turns  counter clockwise  with the INT display reading 512 mV     Set the front panel RANGE switch to 32  Select the REC display     Locate VR612 on the AMPLIFIER board  it is clearly labeled  Adjust VR612  so that the REC display reads as close to 5 mV as possible  The tolerance  for this adjustment is 5 mV  10         Service Handbook for 1047 Refractive Index Detector   12 2001 95       NOTE       Procedures    Calibrate the Autobalance Decision Threshold    1 The AB Mot
105. ware Information    Recorder Path The RECORDER signal path consists of the REC OUTPUT SPAN  RANGE    CONTROL  MARKER OFFSET  POLARITY CONTROL  FS OUTPUT SELECT    and REC SHOR TING CONTROL stages  REC OUTPUT SPAN is adjusted by  VR612     Figure 9 Output Board    to Display    from Elec   Autobalance           Board  to Amplifier from Amplifier from Amplifier  Board  Board  Board  Recorder  First Stage Integrator Circuits Circuits    Output Row  INT ImV 1OmvV                                                  Con 703 e    o    Out Z   A 5     peses  NJ GND   lt  2   Resp    Response 5    Time   Time     oO    ou 5 Controler                            Output Board          The RANGE CONTROL is set by the user from the front panel  The MARKER  OFFSET applies a voltage spike to the recorder signal for the purpose of  marking the start of a chromatogram on a strip chart recorder     The MARKER OFFSET can be applied manually from the front panel or  automatically from rear panel        30 Service Handbook for 1047 Refractive Index Detector   12 2001    Hardware Information    Recorder POLARITY can be changed  only  from the front panel  FS  OUTPUT SELECT is the beginning of the separate 1 mV and 10 mV full scale  recorder outputs  The REC SHORTING CONTROL provides an easy means of  shorting  zeroing  the recorder outputs so that a strip chart recorder may be  mechanically zeroed without disconnecting the detector  The 1 mV recorder  output is taken directly to the 1 mV recorder anal
106. y Board ERC75012 01047 66513  5 Preamp Board ERC75100H 01047 66502  6 Mother Board ERC75201A 01047 66508  7 Amplifier Board ERC75106A 01047 66507  8 Control Board ERC75105A 01047 66506  9 Auto Balance Board ERC57104A 01047 66505  10 Optical Auto Balance Board ERC75011A 01047 66501  11 Power Supply Board ERC75103A 01047 66504  12 Temperature and AB Motor Control Board ERC75102A 01047 66503  13 Output Board ERC75007B 01047 66515  14 Main Power Transformer 01047 61101  15 Lamp Assembly 01037 60002  16 Optical Balance Motor Assembly  includes 01047 60016  complete assembly plus mounting bracket   17 Solenoid  flush  Valve Assembly 01047 60008                Service Handbook for 1047 Refractive Index Detector   12 2001    129    Parts Information                                                                                                                                                                                                                                                                                  Figure 40 Major Components  Jeiector Rear  18  6    f  6 870 7 12    14  ww   O  5  P      NU Optical Unit  18  7  2 3    4                                           elector Front          130 Service Handbook for 1047 Refractive Index Detector   12 2001    Parts Information       Major Assemblies    The following table identifies the major assemblies shown in Figure 41 on                   page 134   Table 15 Major Assemblies    Description ERC number Part number  1 Fro
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
S-F402T  Lasure LX515  カタログ(PDF) - 株式会社エフ・エム・アイ  Betriebsanleitung für Stadtroller "SR 59  おじずる省エネルギー対策に最適な提案  Dräger X-am 5600 - Keison Products  ユーザーガイド - コダック アラリス ジャパン株式会社  Furuno FI-50 Series User's Manual  Kongsberg Simrad EA 500  IMS D7214 IBM/NEC PC ANSI C toolset delivery manual    Copyright © All rights reserved. 
   Failed to retrieve file