Home

Model 8552/8554 Q-Trak Plus Manual

image

Contents

1. Figure 4 3 CO Zero Calibration Step 2 Screen 34 Chapter 3 Figure 4 4 CO Calibration 5 The Q TRAK Plus monitor will indicate the status of the calibration process After the instrument has averaged the zero gas the display will highlight ZERO CAL COMPLETE see Figure 4 5 CO2 ZERO CAL STEP 3 DETECTING ZERO GAS AVERAGING ZERO GAS ZERO CAL COMPLETE TURN OFF ZERO GAS NEXT Figure 4 5 CO Zero Calibration Step 3 Screen Calibration and Maintenance 35 6 Once the zero calibration is complete turn the gas off Press the or V key to continue to the span gas calibration step 7 Detach the regulator from the zero gas cylinder and install the regulator onto the span calibration gas cylinder Press the or V key to continue Figure 4 6 C02 SPAN CAL STEP 1 DISCONNECT ZERO GAS ATTACH SPAN GAS TO CAL COLLAR NEXT Figure 4 6 CO Span Calibration Step 1 Screen 8 Adjust the span value of the calibration gas being used by pressing 4 or gt keys to move the cursor to the required field and either the A or VY keys to increment or decrement the value Press J to accept the calibration gas Figure 4 7 CO2 SPAN CAL STEP 2 ADJUST SPAN VALUE TO CAL GAS VALUE CO2 1000 PPM gt ACCEPT J EXIT 4 Figure 4 7 CO Span Calibration Step 2 Screen 9 Turn on the span gas 0 3 L min and press the J key to begin calibrating the span Figure 4 8 CO2 SPAN CAL STEP 3 TURN ON SPAN
2. Sample Protocol for Log Mode 2 or 3 The following steps describe how to program a sample protocol for Log Mode 2 or 3 Log Mode 2 sample protocol features The sample protocol for Log Mode 2 is set to take unattended readings for one day Logging takes place on 02 04 2006 The logging sample begins at 8 00 a m Logging continues for eight hours Log Mode 3 sample protocol features The sample protocol for Log Mode 3 is set up to take unattended readings for four days Logging first takes place on 02 04 2006 The logging sample begins at 8 00 a m Logging continues for eight hours The instrument is off for 16 hours and then repeats an eight hour test each day for the next three days The following graphic gives the appearance of the dialog box displayed in the TRAKPRO software with above described logging parameters Operation Start Date MM dd pyyy Start Time hh mm Log Interval mm ss Test Length dd hh mm Number of Tests Time Between Tests dd hh mm Percent Memory Required Log 2 Q Trak Plus Logging Protocols Serial Number 8554 04011016 Number Tests Logged 4 Available Memory 100 The Models 8552 and 8554 automatically record all channels including any auxiliary parameters that are installed 0204 2006 agas EE oy 7 r Log3 JV 02 04 2006 foo foo for f 1 mema o ES Cancel Figure 3 10 Q TRAK Plus I
3. 1 1 4 Percent Memory Required 0 0 Figure 3 9 Q TRAK Plus Monitor Logging Protocols Chapter 3 The following table summarizes the information displayed in the Q Trax Plus Logging Protocols dialog box ltem Description o Displays the serial number of the Q TRAK Plus IAQ Monitor Logged in the instrument Available Displays the percent of available memory in the Q TRAK Memory Plus monitor data logging only single point memory is not included Number of Displays the number of tests for LOG 2 and LOG 3 Tests modes Selecting more than one test will cause the instrument to cycle through the logging test more than once separated by the Time Between Tests Time Between Displays the time between tests for LOG 2 and LOG 3 Tests modes Percent Displays the percent of logger memory required to Memory perform a LOG 2 or a LOG 3 mode sample To store the Required results of a LOG 2 or LOG 3 mode sample the Available Memory must be equal to or greater than the percent memory required Operation 17 To program the Q TRAK Plus monitor for Log Mode 2 or 3 enter the following Item Description Start Date Enter the date and time to begin the sample Start Time e Ifthe start date field is left blank the date defaults to the current date The test starts at the start time e f both of the start date and time fields are left blank the test begins as soon as you press the key in the
4. Indoor Air Quality Model 8552 8554 Q TRAK Plus IAQ Monitor TRAKPRO Data Analysis Software enclosed Operation and Service Manual 1980417 Revision June 2006 7 lil Model 8552 8554 Q TRAK Plus IAQ Monitor Operation and Service Manual June 2006 P N 1980417 Rev I SHIP MAIL TO E mail address TSI Incorporated answers tsi com 500 Cardigan Road Website Shoreview MN 55126 3996 http www tsi com USA INTERNATIONAL U S Customer Service Customer Service 001 651 490 2811 800 874 281 1 651 490 2811 Fax Fax 001 651 490 3824 651 490 3824 TSI Copyright TSI Incorporated 2000 2006 All rights reserved Address TSI Incorporated 500 Cardigan Road Shoreview MN 55126 USA Fax No 651 490 3824 Limitation of Warranty and Liability effective July 2000 Seller warrants the goods sold hereunder under normal use and service as described in the operator s manual shall be free from defects in workmanship and material for twenty four 24 months or the length of time specified in the operator s manual from the date of shipment to the customer This warranty period is inclusive of any statutory warranty This limited warranty is subject to the following exclusions a Hot wire or hot film sensors used with research anemometers and certain other components when indicated in specifications are warranted for 90 days from the date of shipment b Parts repaired or
5. Data Analysis Software ccssccssssseeseeseesees 6 Connecting the Q TRAK Plus IAQ Monitor to the Computer 6 Setting up the Communications POrt ceceesssecseeeeceseeeeeseeeeeeeenees 6 Using the Optional Instrument Stand eee eeeeeeceeeeeceeeeeceeeeeeeeeeaees 7 Using the Optional Probe Stand eee eeeseeecsseceeeeecneeeecsecseeseeeseeeeaees 7 Connecting the Optional Portable Printer 0 0 eect eeeseeseeeeeecreeeeenees 8 OPERATION iacneecednuin ace eines eerie 9 OVELVIEW i 2 iS aee ae n a a N r eae r 9 Q TRAK Plus IAQ Monitor Keypad ce eeeecesecseeeecneeeecsseeeeseeeseeeeenees 9 Turning the Instrument OD ee eeeecseeeeceseeeceseceeesecaeeseeeecaeeseeneseeeeaees 9 Survey Mode 22ers aie eG Bie ats eee 10 MIDIS nina Silo ioanssernaih bveeyet evi didn sbeeesetes R A SE 11 Sample Mod ne ta cehetin dite wel en Alle Ral aie ie 11 Recording and Saving a Single Data Point eee eeeseceseeeeeeecneeeeeee 12 Printing Saved Single Data Points 0 0 0 ce eeeseseeeeeeeeeeeeeeeeenees 12 M n Mein usesctess eo cisiese tateepeetel sees eoitesaereacsacvant eg E r a i Time Constant LOS MOE sere erated i r a poti teen teen e Eaa eens Log Mode Display of Real Time Data cece eeeeeeeeneeeeeeeees 13 Log Mode 1 Log Interval oo eee eeeceencecseeceseeeceeeeeeeeceaeeeenees 14 Eog Mode 2 and 3 23 46 cS ddan Raul intely ahalale kien 16 Programming Log Mode 2 or 3 Using TRAKPRO Data Analysis SoftWare ienes siririca ii aiei nsii aie
6. Display of Real Time Data To begin a data logging session select a log mode from the Main Menu and press the A key The Q TRAK Plus monitor will begin recording logged data into the instrument memory and display the real time values see Figure 3 5 1 LOGMODE1 TEST 1 2 751 F TEMP 2 805 PPM C02 3 0 0 PPM CO 75 1 F TEMP 35 1 RH 4 NEXT lt 4 gt TO STOP 5 Figure 3 5 Log Mode Statistical Data Ref Description The Q TRAK Plus monitor is in Log Mode 1 2 The Test ID number is 1 This number is automatically incremented for each new test The real time values for each parameter are displayed Use the lt or P key to view the statistics for the current test Pressing the key will terminate the current test Operation 13 While logging you are able to view statistical summary data of the current test Pressing the lt or key will display the statistics for each parameter The CO summary screen is shown in Figure 3 6 PPM C02 PPM CO2 02 04 05 4 PPM CO2 02 04 05 NEXT lt gt JTOSTOP Figure 3 6 Log Mode CO Summary Screen Ret Description SCS Log Mode 1 Log Interval The log interval is both a frequency and an averaging period For example when the log interval is set to 5 minutes readings will be recorded at 5 minute intervals Each reading will be the average value measured over that 5 minute interval The Select Log Interval menu allows you to view the current logging interval used du
7. Figure 5 3 f TSI Q Trak Plus Flash Memory Programmer z alx Looking For Instrument Looking For Image File Initializing Flash Memory Programming Flash Memon Close Figure 5 3 Q TRAK Plus Flash Memory Programmer Menu Upon Starting From this menu click once on the Start button to begin the firmware upgrade process During this upgrade process a new version of firmware will be written into the instrument and a status bar will be shown to illustrate progress This process will last about 7 minutes Once the upgrade process has successfully completed the Q TRAK Plus Flash Memory Programmer menu will be displayed as shown in Figure 5 4 From here click once on the Close button to terminate the program Your instrument is now ready to communicate with the Model 8592 Air Velocity Probe accessory If you use TRAKPRO Data Analysis Software you will need to upgrade to a new revision 3 41 or newer This TRAKPRO upgrade can be done from the product CD main menu or by visiting www tsi com Chapter 5 f TSI Q Trak Plus Flash Memory Programmer as al xj Found Trak Plus SN 8552 01000000 on serial port COM1 4 Found File qtrak207 a33 7 Initializing Flash Memory Programming Flash Memory Congratulations Your instrument has been reprogrammed successfully Coe Figure 5 4 Q TRAK Plus Flash Memory Programmer Menu Upon Completion Using the Telescoping Probe To e
8. Figure 5 8 Velocity Averaging Menu With highlighted as shown in Figure 5 8 a sample is acquired using the key When all of the desired samples have been taken use the W key to select SAMPLING COMPLETED and press the key to terminate sampling VELOCITY AVERAGING 2 15 M S MIN VEL 6 30 M S MAX VEL 4 25 M S AVG VEL VIEW SAMPLES J Figure 5 9 Average Velocity Statistics Page Statistics for the sample set will then be displayed as shown in Figure 5 9 Pressing the escape key will return to the main menu 52 Chapter 5 Individual sample points can be reviewed by pressing the key to select VIEW SAMPLES VELOCITY SAMPLE 2 lt gt 4 21 M S VEL Figure 5 10 Reviewing Individual Samples from the Velocity Sample Set Using the lt or gt key allows you to view individual samples from the velocity sample set The lt key will display a previous or lower numbered sample and the gt key will display the next or higher numbered sample The sample set is displayed as a continuous loop When the last sample is reached pressing the key will take you back to the first sample The maximum number of samples is 99 Pressing the t key will return you to average velocity statistics page Measuring Volumetric Flow Rate MENU TIME CONSTANT LOG MODE 1 SETUP VENTILATION RATES STATISTICS ELOCITY PROBER Figure 5 11 Display Menu when the AVP is Attached to the Q TRak Plus From the main menu may
9. GAS PRESS J TO BEGIN SPAN GAS SAMPLE Figure 4 8 CO Span Calibration Step 3 Screen 36 Chapter 3 10 The Q TRAK Plus monitor will indicate the status of the calibration process The process will end successfully when SPAN CAL COMPLETE message highlights Figure 4 9 CO2 SPAN CAL STEP 4 DETECTING SPAN GAS AVERAGING SPAN GAS SPAN CAL COMPLETE TURN OFF SPAN GAS BACKA DONE J Figure 4 9 CO Span Calibration Step 4 Screen 11 Turn off the span gas press the key to complete CO user calibration and remove the calibration collar If an error is detected during the zero or span calibration process an error message will be displayed The previous calibration values are kept Press the or J key to return to the Calibration Menu ERROR DETECTED OR PROCESS ABORTED DURING CALIBRATION VALUES UNCHANGED Figure 4 10 Error Detected or Process Aborted Screen Calibrating the CO Sensor The CO sensor in some models of TSI IAQ monitors may be adversely affected when exposed to isopropyl alcohol vapors released from TSI instruments utilizing condensation particle counting CPC technology For maximum performance keep CO sensors away from CPC based instruments utilizing isopropyl alcohol TSI recommends calibrating the Q TRAK Plus monitor s CO measurement monthly to help ensure accurate readings The CO concentration measurement is affected by changes in temperature and atmospheric Calibration and Maint
10. Memory cannot be cleared through the TRAKPRO software Calibration See Chapter 4 Calibration and Maintenance later in this manual Printer The PRINTER option allows you to enable or disable the printing function of the single sample test results To enable or disable the printing option scroll and highlight using the A or V key on the Setup menu and press J Once in the printer submenu use either the A or W key to enable or disable the printer output see Figure 3 18 PRINTER PRINTER DISABLED A V Figure 3 18 Printer Output Enable Disable Screen Ventilation Rates Ventilation rates can be measured with the Q TRAK Plus monitor in terms of Outside Air To select one of these options choose VENTILATION RATES by using the A or V key while in the Main Menu and press the key 28 Chapter 3 VENTILATION RATES OUTSIDE AIR Figure 3 19 Ventilation Rates Select Screen Outside Air Percentage Outside Air may be calculated by using a temperature or CO based method Percentage Outside Air is calculated from the ratio between the differences of the supply air return air and outside air by using the following formula RA SA x 100 RA OA Outside Air Where RA PPM of CO or temperature in return air SA PPM of CO or temperature in supply air OA PPM of CO or temperature in outside air The following variables must be entered to perform the calculation OUTSIDE AIR S
11. STEP 3 DETECTING ZERO GAS AVERAGING ZERO GAS ZERO CAL COMPLETE TURN OFF ZERO GAS BACK A DONE J Figure 4 12 CO Zero Calibration Step 4 4 The Q Trak Plus monitor will indicate the status of the calibration process After the instrument has averaged the zero gas the display will indicate ZERO CAL COMPLETE Calibration and Maintenance 39 5 Turn gas off Detach the regulator from the zero CO calibration gas cylinder Press the key to continue to the span gas calibration step 6 Install the regulator onto the span calibration gas cylinder Press the J or V key to continue 7 Adjust the span value of the calibration gas being used by pressing the A or V key Press the key to accept the calibration gas 8 Turn on the span gas 0 3 L min and press the J key to begin calibrating the span 9 The Q TRAK Plus monitor will indicate the status of the calibration process The process will end successfully when SPAN CAL COMPLETE message blinks Figure 4 13 CO SPAN CAL STEP 4 IDETECTING SPAN GAS AVERAGING SPAN GAS SPAN CAL COMPLETE TURN OFF SPAN GAS BACK A DONE J Figure 4 13 CO Span Calibration Step 4 10 Turn off the span gas remove the calibration collar and press the J key to complete CO user calibration If an error is detected during the zero or span calibration process an error message will be displayed If an error is detected during the zero or span calibration process an error message
12. TRAK Plus monitor communications port Setting up the Communications Port To communicate with the Q TRAK Plus monitor the software must be configured for the proper COM port The TRAKPRO software can be manually set to operate on a specific COM port or it can automatically find a Q TRAK Plus monitor that is attached to any COM port To set up the COM port do the following 1 Turn on the Q TRAK Plus monitor 2 Start the TRAKPRO software 3 Select Instrument Setup Communications in the TRAKPRO software The following dialog is displayed Chapter 2 Communications Port Setup r Serial Pot com OK Cancel Find Port Test Baud 9600 Parity N Data 8 Stop 1 HEE Figure 2 1 Instrument Setup Communications Dialog Box 4 Select the correct serial port such as COM2 then select Test The software will verify that you have set up the communications port correctly and that it is communicating with the Q TRAK Plus monitor The system displays an information message indicating whether it was able to establish communications 5 Asan alternative select Find Port to have the monitor search the available COM ports looking for an attached Q TRAK Plus monitor 6 Select OK to accept the setup or Cancel to discard the changes Using the Optional Instrument Stand The Q TRAK Plus IAQ Monitor can be used with an instrument stand available as an accessory When placed on a level surface the instr
13. have been set that are in the future the test start date and or time has passed no test length is set was pressed during sampling When a pre programmed test sequence ends Operation the Q TRAK Plus monitor begins logging immediately after the key is pressed This happens no matter what is entered for the start date the Q TRAK Plus monitor begins at the next specified time regardless of the date the Q TRAK Plus monitor will display the time and date of the next test and the next test ID pressing the key has no impact The program will never execute no data will be logged the Q TRAK Plus monitor samples continuously until the key is pressed or until the memory is full the program stops data logging unless the keypad is locked out the Q TRAK Plus monitor automatically returns to the Survey Mode 21 Memory Considerations The Q TRAK Plus IAQ Monitor has a great deal of memory and will not normally run out The instrument contains separate memory for the single points and for the logged tests filling up the memory for one type of data will have no affect on the other type The Q TRAK Plus monitor can store approximately 440 separate single points in addition to 48 671 logged data points separated by as many as 110 tests That is the instrument can also store approximately 811 hours 33 days of logged tests when recording at 1 minute logging interv
14. the CO Factory service is required to sensor correct this fault condition 59 Symptom Possible Cause Corrective Action TEMP FLT There is a fault in the Factory service is required to temperature sensor correct this fault HUMID There is a fault in the Factory service is required to FLT humidity sensor correct this fault 60 Chapter 6 Appendix A Specifications Specifications are subject to change without notice CO Sensor TY pessier Non Dispersive Infrared NDIR Raperin ra a e 0 to 5000 ppm ACCUTACY orionas 3 of reading 50 ppm at 25 C Add uncertainty of 0 36 of reading per C 0 2 of reading per F for change in temperature Resolution ccceesseesceeseeereeeeeeeees 1 ppm Temperature Sensor TYPO srir teased aieciies Thermistor R esonera a n 0 to 50 C 32 to 122 F ACCULACY cetro Bonesi Daata 0 6 C 1 0 F Resolution 0 0 eee eeeseceseceeeeeenteeeeee 0 1 C 0 1 F Response time s es 30 seconds 90 of final value air velocity at 2 m s Display units 2 0 0 eee eeeees C or F user selectable Humidity Sensor CYP perisa Thin film capacitive Rante curee penen an a 5 to 95 RH ACCURACY eie ir ER 3 RH includes 1 hysteresis Resolution oeer arite 0 1 RH Response time csceeeseeseereeeeees 20 seconds for 63 of final value CO Sensor Sensor LY PO ieie a Electro chemical Ran BO teste Bid elt ieee 0 to 500 ppm ACCULACY
15. 1 3 Auxiliary Sensor Probe Stand Air Velocity Probe Ref Item Description Part Model Probe Stand 800679 2 Auxiliary Temperature Sensor 8590 Air Velocity Probe 8592 Unpacking and Parts Identification 3 Figure 1 4 Portable Printer Kit Item Description Part Model Portable printer 8930 4 Chapter 1 Chapter 2 Setting Up Supplying Power to the Q TRAK Plus IAQ Monitor The Model 8552 8554 Q TRAK Plus IAQ Monitor may be powered with four AA alkaline batteries or the optional AC adapter Installing the Batteries Insert four AA size batteries as indicated by the diagram located on the inside of the battery compartment TSI ships the Q TRAK Plus monitor with alkaline batteries Rechargeable batteries may also be used battery charger and batteries are not included and are not available from TSI Using the AC Adapter The AC adapter allows you to power the Q TRAK Plus monitor from an AC wall outlet When using the AC adapter the batteries if installed will be bypassed The AC adapter is not a battery charger and will not charge rechargeable batteries The Q TRAK Plus monitor has an internal non user accessible battery that is used to keep memory intact when power is turned off Changing the AA size batteries or disconnecting the AC adapter will not cause data to be lost This battery is designed to last for years TSI will install a new battery if necessary when the unit is returned to the factory for
16. AQ Monitor Logging Example To program this logging example do the following 1 Make sure the Q TRAK Plus monitor is connected to the computer and turned on 2 Select Logging Setup from the Instrument Setup menu TRAKPRO software retrieves the current settings for Log Modes 2 and 3 from the Q TRAK Plus monitor and displays them as shown in the previous dialog box 3 Enter the following for Log Mode 2 and 3 2 and mode 3 protocols Start Time five minutes five minutes from now from now 4 Select Send The logging instrument is programmed for the mode Chapter 3 5 Note in the example that the LOG 2 test requires 4 of the available memory and LOG 3 requires 4 of the memory A total of 100 of the memory is available for use 6 Disconnect the Q TRAK Plus monitor from the computer Refer to other sections of this Operation and Service Manual for details on making measurements using LOG 2 and LOG 3 modes After programming the Q TRAK Plus monitor with the TRAKPRO Data Analysis Software take the monitor to the desired location and turn it on From the Main Menu put the monitor into LOG MODE 2 or LOG MODE 3 whatever was programmed by pressing the lt or gt key Press the key to initiate the Log Mode The display will show the current time and date The following will happen depending upon the protocol no start time was set a test start time was set but no start date a start time and start date
17. AX VEL 4 25 M S AVG VEL 1 53 M3 H FLOW RATE VIEW SAMPLES Figure 5 17 Volumetric Flow Rate Measurement Statistics Page The minimum average and maximum velocities of the sample set along with the calculated volumetric flow rate will be displayed as shown in Figure 5 17 Individual sample points can be reviewed by pressing the J key to select VIEW SAMPLES Using the lt or P key allows you to step through the velocity sample set by The lt key will display a previous or lower numbered sample and the key will display the next or higher numbered sample The sample set is displayed as a continuous loop When the last sample is reached pressing the key will take you back to the first sample The maximum number of samples is 99 Pressing the f key will return you to the Volumetric Flow Rate statistics page VELOCITY SAMPLE 1 lt gt 4 21 M S VEL 1 91 M3 H FLOW RATE Figure 5 18 Sample 1 Of The Velocity Sample Set Chapter 5 Maintenance Periodically inspect the AVP sensor window to ensure that it is clean Dust and oil deposits on the sensor will affect measurement accuracy The telescoping probe should also be kept clean Dirt or oil accumulation will affect the ability to extend and retract the probe Use a soft cloth dampened with water to remove accumulation of unwanted material Dry the probe with a gentle stream of air or allow to air dry If oil or grease is present a small amount of high grade isopro
18. Figure 3 2 10 Chapter 3 2 SAMPLE 805 PPM 1 co2 20 PPM CO 3 75 1 F TEMP 35 1 RH 70 1 F TEMP 2 Figure 3 2 Survey Mode Screen sample Ref Description Primary Parameter CO concentration is shown Sample mode Used for capturing single data points Remaining parameters Status message area Battery status Units To change units displayed in the Survey mode use the A or V key to select the parameter Press either the lt or key and the units of the selected parameter will change accordingly The units displayed on the instrument will be the same units that are stored during the log mode Sample Mode The Sample mode is the default selection in the Survey Mode the reverse highlight indicates that it is selected The Sample mode is used to capture a single 10 second averaged data point of all measured parameters Operation 11 Recording and Saving a Single Data Point With SAWYISHD selected and the key pressed the Q TRAK Plus monitor begins a countdown During these 10 seconds it is taking an average reading of each parameter At the end of the averaging period the following sample screen is displayed SAMPLE 1 75 1 F TEMP 2 805 PPM C02 2 0 PPM CO 75 1 F TEMP 35 1 RH SAVE SAMPLE NO Figure 3 3 Record Single Data Point The selection is highlighted by default If you press the key the single data point will be stored in memory Press the gt key to highlight NO to disca
19. International 001 651 490 2811 Fax 001 651 490 3824 These Application Notes can also be found under TSI s web site http jwww tsi com For current information Iso a 9001 2000 www tsi com Chapter 1 Unpacking and Parts Identification Unpacking the Q TRAK Plus IAQ Monitor Carefully unpack the Q TRAK Plus IAQ Monitor from the shipping container Use the tables and illustrations below to make certain that there are no missing components Contact TSI immediately if anything is missing or damaged Figure 1 1 Models 8552 8554 Q TRak Plus IAQ Monitor with Calibration Collar and Batteries Ref Qty Item Description Part Model 1 1 Q TRAK Plus IAQ Monitor 8552 or 8554 Calibration Collar 801679 Batteries AA alkaline Computer Interface Cable RS 232 800563 5 1 AC Adapter 115 V NEMA 5 2613033 230 V Eur CEE 7 16 2613078 230 V Great Britain 800169 240 V Australian 2613105 16 1 Probe Stand not shown 800679 SII SydVVyvVIVTTT ny t a 07 z g Figure 1 2 Calibration Certificate Operation and Service Manual TRAKPRO Data Analysis Software Ref Qty Item Description Part Model Calibration Certificate Operation and Service Manual 1980417 TRAKPRO Data Analysis Software Chapter 1 Optional Accessories The following pages contain component identification information for optional accessories available for the Q TRAK Plus IAQ Monitor Figure
20. Note None of the Outside Air data is stored in memory It is recommended that data be recorded before you exit the screen If unrealistic data is entered e g supply CO2 value greater than return CO gt the instrument will return an error message Statistics The STATISTICS menu allows you to review either Single Point Statistics or Logged Test Statistics Single Point Statistics The Single Point Statistics screen displays the results of each recorded single data point that was taken in the Survey Mode When the screen is first displayed it shows the information for the most recent single data point Previous samples can be viewed by using the A or V key A sample screen is shown below SAMPLE 3 02 04 05 F TEMP 2 CO2 CO TEMP RH Figure 3 23 Single Point Statistics Logged Test Statistics The Logged Test Statistics screen displays the statistical summary for all valid data for each logged test When the screen is first displayed it shows a summary of the most recent test A sample screen is shown in Figure 3 24 Operation 31 LOGGED TEST STATS 1 TEST 3 START 08 23 02 04 05 2 3 STOP 08 23 02 04 05 LOG INTERVAL 00 05 4 5 DURATION 0 12 30 35 NEXT TEST VALUE lt gt Figure 3 24 Logged Test Statistics Summary Ret Description OO ooy Test number not user adjustable To view the actual parameter statistics for each logged test statistics use the lt or key To view the previo
21. Q Monitor when not in use Auxiliary Sensors 57 Chapter 6 Troubleshooting The table below lists the symptoms possible causes and recommended solutions for the Model 8552 8554 Q TRAK Plus IAQ Monitor Symptom Possible Cause Corrective Action No display Unit not switched on Switch unit on Low or dead batteries Replace the batteries or plug in the AC adapter Dirty battery contacts Clean the battery contacts by rubbing a rubber eraser on each contact LO BAT Low battery charge Replace batteries or use AC message adapter appears Incorrect AC adapter Replace with the correct AC adapter Low AC line voltage Correct the AC line voltage use batteries Dirty battery contacts Clean the battery contacts 0 MEM Memory is full Download the logged and single displays point memory to a PC using TRAKPRO software if desired Clear memory Humidity Probe exposed to intense Shade the probe while taking reading near light samples zero or not believable No keypad Display Lockout switch in Slide lockout switch to normal response lockout position position FACTORY Instrument is near the end Press the key to continue For SERVICE of the annual calibration optimum performance TSI reminder is period recommends annual factory displayed servicing CO CAL The CO sensor requires See CO sensor specifications calibration approximately page 121 CO negative readings CO2 FLT There is a fault in
22. UPPLY AIR and RETURN AIR Values for each variable may be entered manually or by capturing a sample at the appropriate location Calculating the Outside Air 1 While in the VENTILATION RATES menu use the A or V key to scroll and select 7 OUTSIDE AIR and press the J key 2 Select either CO2 or TEMPERATURE using the A or V key and press the key Press the W key to continue Operation 29 OUTSIDE AIR CALCULATE OUTSIDE AIR USING TEMPERATURE Figure 3 20 Outside Air Screen The following examples assume that CO2 was selected as the measured parameter The OUTSIDE AIR info screen will display TO CALCULATE OUTSIDE AIR MAKE THREE MEASURMENTS OUTSIDE AIR C02 SUPPLY AIR CO2 RETURN AIR CO2 NEXT V Figure 3 21 Outside Air Info Screen 3 Use the A or W key to select an appropriate variable OUTSIDE AIR SUPPLY AIR or RETURN AIR SAMPLE J EDIT gt OUTSIDE AIR 0 OUTSIDE PPM CO2 PPM CO2 PPM C02 SUPPLY 0000 RETURN 0000 CALCULATE Figure 3 22 Sample or Edit CO Parameters 4 Screen selections allow you to press the or J key to either edit the value or take a sample 5 Repeat steps 2 through 4 until all three variables have a valid value 30 Chapter 3 6 Use the A or W key to select CALCULATE and press the J key The OUTSIDE AIR value will display at the top of the screen 7 For repeated calculations repeat steps 3 through 6 8 To exit press the T key
23. age velocity from a series of sample points the minimum and maximum velocities will also be displayed identifying areas that may require corrective action Chapter 5 Q TRAK Plus LCD Display During AVP Operation The following figures show the screens pages displayed on the Q TRAK Plus LCD during operation with the AVP connected to the instrument MENU TIME CONSTANT LOG MODE 1 SETUP VENTILATION RATES STATISTICS Figure 5 5 Default Menu of the Q TRAK Plus no accessory probe attached MENU TIME CONSTANT LOG MODE 1 SETUP VENTILATION RATES STATISTICS VELOCITY PROBER Figure 5 6 Default Menu of the Q TRAK Plus with AVP attached You may highlight gy Gere YAE 2 10 using the A or V key Pressing the enter key will select the VELOCITY PROBE option and display the VELOCITY MENU shown in Figure 5 7 From the velocity menu you can proceed with a velocity averaging measurement such as a fume hood face velocity measurement or you can proceed with a volumetric flow rate measurement such as a supply duct traverse measurement Auxiliary Sensors 51 VELOCITY MENU VELOCITY AVERAGING VOLUMETRIC FLOW RATE Figure 5 7 Velocity Menu Options Measuring Average Velocity AAD OLS UA ENI AENA may be selected by using the A or V key and pressing the J key will display the VELOCITY AVERAGING menu VELOCITY AVERAGING 4 25 M S VEL SAMPLES TAKEN 0 CAPTURE SAMPLER SAMPLING COMPLETED
24. als the number of hours of logged tests may be more or less depending upon the logging interval The total test time should be considered when selecting a logging interval Shorter logging intervals use memory more quickly than longer intervals The logging interval and the available memory determine the maximum possible duration of a data logging session Maximum logging duration Ymemory available X 48 671 X logging interval To prevent loss of any logged data it is recommended you download your data to the computer using the TRAKPRO software as often as possible Auxiliary Sensor For each of the logging modes the auxiliary sensor parameter is logged along with the standard parameters as long as the auxiliary probe is connected to the instrument prior to the start of logging the auxiliary sensor is connected the Q TRAK Plus monitor will before the start of logging record the sensor data during logging the auxiliary sensor is connected the Q TRAK Plus monitor will after the start of logging NOT record the sensor data during logging the auxiliary sensor is disconnected the Q TRAK Plus monitor will during logging display Invalid Data for the 22 sensor for any time while the sensor is disconnected Chapter 3 Keypad Lockout Switch Recording data over extended time periods in Log Mode 1 2 or 3 often requires leaving the Q TRAK Plus monitor unattended To reduce the risk of having an unauthor
25. appropriate log mode Log Interval Enter the log interval to use for the test Test Length Enter the length for the sample e The instrument automatically turns off when the last test is complete at the end of the test length e If the field is left blank you must manually stop the sample by pressing the key in the logging screen Number of Enter the number of tests to perform Tests Time between Enter the time between tests if there is more than one tests test If 0 or blank is entered the next test is started immediately after the previous test While entering values for Log Mode 2 or 3 the Percent Memory Required is dynamically updated to show the amount of logger memory required to take the programmed sample If the protocol defined requires more than 100 of memory decrease the amount of memory required by changing one of the following parameters e Increase the logging interval e Decrease the length for the test e Decrease the number of tests The settings for each separate Log mode must not require more than 100 of the logger memory Note Ifthe percent memory required is greater than the available memory the TrakPro software will not allow you to program Log Mode 2 or 3 3 When finished defining parameters for LOG 2 and LOG 3 modes select Send 4 Disconnect the Q TRAK Plus monitor and cable from the computer 5 The instrument is ready for data logging 18 Chapter 3
26. as 16 Sample Protocol for Log Mode 2 orf 3 sssr 19 Memory Considerations ccsecssessceecseceeeseceeeeeceseeeesseeeeeseeners 22 Auxiliary Sensor seai t e arepa a ee 22 Keypad Lockout Switch oo eeceeeeseeecseceeeceeeecneseeesaeeeesaeenee 23 Setup MENU sonin ied tae deen seeds E Seas EE 24 Fime Dates 233 es Bist cil BR he Sa RRO Ree ld 24 lo EE ADT Di S AEAEE teeters leeseve Bind nated terres 26 Clear Memory r E R E ees 27 Calibra onea a a r a A ari ae rie 28 Printer ehea ar A a E ara a a 28 Ventilation Rates ccccccccccccccccccccccccccccccecccccececececseeceseceeeseseseseseseeeees 28 W Outside Alt isiin a es Ae eae 29 SLALISTICS ogni decsaidysgesdinnss odessucsescescaaecsunemetusndiousseursengeazavescesseranegeizeey 31 Single Point Statistics 0 eee ee ceeeesecesecseecssecaeecaeeeseeeeeeeeeeeeees 31 Logged Test Statistics sisedececsdecdsssacatecesssssecs sctasucstidsaceavsssiataniaasessis 31 CALIBRATION AND MAINTENANCE 00 occ eeeeceeecsecseeeneeeneeees 33 Calibrating the CO2 SeMSot es eeesesseeeecseeeeesseeeceseceeeseeneeseesaeenseaecaees 33 Calibrating the CO Semsor escesesessseessessesenencnsesosenersnsetverenenenenses 37 Calibrating the Temperature or Relative Humidity Measurement 41 Restore Factory Calibration cceessesseseeecsseeeeeeceeeceeeeecsaeeeeeaeeaees 42 Storage PRECAUTIONS lt ce sicicinseivodvsciesdacdoctebece ms seesvheuns NEEE TE 43 AUXILIARY SENSORS circai e
27. at keeps track of the time of day and the date It is very important to have the time and date correctly set for the monitor otherwise time and date stamping of recorded data will not be correct The time clock is set correctly at the factory for U S Central Standard Time If it requires adjustment you may change it by selecting TIME DATE from the Setup menu The following screen is displayed TIME DATE HH MM SS 12 12 31 MM DD YY 02 04 05 CHANGE FORMAT MM DD YY Figure 3 13 Time Date Chapter 3 When first entering this screen the first character in the time field will have a cursor underneath it Use the A or W key to increment and or decrement values Use the lt or key to move to the required field When all adjustments are complete press the key to accept the changes and return to the Setup menu The Q TRAK Plus monitor offers three different formats for displaying the date U S European and ISO 8601 Press lt or gt key to select and highlight QsG W e os YOY then press either the A or V key to cycle through the date formats MM DD YY DD MM YY YYYY MM DD Once the desired format is selected press the J key to confirm the choice Programming Time Date Using TRAKPRo Data Analysis Software The time and date may also be adjusted using the TRAKPRO software To program the date and time 1 Make sure the Q TRAK Plus monitor is connected to the computer and turned on 2 Using the TRAKPRO softwar
28. city of the air Actual velocity is the velocity at which a microscopic particle of dust would be traveling if it were in the air stream with zero acceleration In some instances actual air velocity may be of interest rather than standard air velocity To obtain the actual velocity multiply the standard velocity by the density correction factor 460 T 14 7 Actual Velocity Standard Velocity 460 70 P where T is ambient temperature in F and P is the ambient pressure in psia The conversion in metric units is 273 T 101 4 273 21 1 2 Actual Velocity Standard Velocity where Tm is the ambient temperature in C and P is the ambient pressure in kPa 65 TSI Incorporated 500 Cardigan Road Shoreview MN 55126 U S A For current information Web www tsi com www tsi com P N 1980417 Rev Copyright 2006 by TSI Incorporated Printed in U S A
29. e select Parameters Instrument Setup Clock The software retrieves the current date and time settings from the Q TRAK Plus monitor and displays them in the dialog box displayed in Figure 3 14 Change Instrument Clock x Enter date and time as mm dd ypypy or transfer System Time System Date Instrument Date 02 04 2006 3 12 31 2005 Cancel System Time Instrument Time 15 12 08 00 01 22 Figure 3 14 Time Date Programming with TRAKPRO Data Analysis Software 3 The system date and time from the computer can be transferred to the monitor using the arrow keys Operation 25 Alternately the date and time can be manually entered into the dialog box 4 Select Send to reprogram the Q TRAK Plus monitor Backlight The BACKLIGHT DELAY option on the Setup menu adjusts how long the display backlight stays on after any key is pressed After the delay has elapsed the backlight will automatically turn off To change the backlight delay use the A or W key The options are No Backlight 10 second 30 seconds or 10 minutes Press the J key to select the value and return to the Setup Menu BACKLIGHT DELAY NO BACKLIGHT Figure 3 15 Backlight Screen Note Prolonged use of the backlight may severely reduce the overall battery life The backlight delay may also be adjusted using TRAKPRO Software 1 Make sure the Q TRAK Plus monitor is connected to the computer and turned on 2 In TRAKPRO Softwa
30. e Dimensions Probe tip diameter 00 Probe base diameter 5 Probe length collapsed Probe length extended Cable Length Cable length relaxed Cable length extended 0 Data Logging Data Points eeceeceeeeeeteeeeeeeee Specifications 0 to 4000 fpm 0 to 20 m s 5 of reading or 5 fpm 0 025 m s whichever is greater 5 to 45 C 41 to 113 F 20 to 60 C 4 to 140 F up to 275 000 cfm up to 467 000 m h up to 130 000 L s 4 to 100 in in 0 1 in increments 10 to 300 cm in 0 1 cm increments Up to 99 sample points 0 25 in 0 6 cm 0 39 in 1 0 cm 11 5 in 29 cm 43 in 109 cm 56 in 142 cm 100 in 254 cm 48 000 approx 33 days of logging at 1 minute intervals 63 Appendix B Standard Velocity vs Actual Velocity Measurements Thermal air velocity sensors are sensitive to changes in both air density and air velocity Thermal anemometers therefore typically indicate velocities with reference to a set of standard conditions For TSI instruments standard conditions are defined as 70 F 21 1 C and 14 7 psia 101 4 kPa Other manufacturers may use a different set of standard conditions Standard velocity is the velocity the air would be moving if temperature and pressure were at these standard conditions It is usually the most useful measure of air flow because it defines the heat carrying capa
31. enance 37 pressure Normal day to day variations due to local weather conditions have little effect However changes in altitude can cause more significant errors For best accuracy calibrate the Q TRAK Plus monitor s CO measurement for your local conditions or if conditions change To calibrate the CO sensor follow the procedure below You can also follow the prompts displayed on the screen while in the user calibration mode You will need a cylinder of pure air for the zero calibration and a cylinder of gas with a known concentration of CO for the span calibration 35 ppm or 200 ppm is recommended Contact your local TSI distributor for available CO calibration kits Figure 4 11 Notes Calibration with CO gas should always be done in a well ventilated area 100 Nitrogen cannot be used as a zero gas for the CO sensor The calibration procedure can be aborted at any time by pressing the f key 1 To put the Q TRAK Plus monitor into CO calibration mode press the A or V keys to select ANA LIYN ON in the Setup Menu 2 Select CO by pressing the A or V keys in the calibration submenu and press the J key Locate the calibration collar and slide it over the main probe Refer to Figure 4 11 38 Chapter 3 Figure 4 11 CO Calibration 3 Turn the gas on 0 3 L min Make sure the cylinder is not empty It should quietly hiss when it is on Press the key to begin zero gas calibration Figure 4 12 CO ZERO CAL
32. g with the calibration 4 Compare the reading on the Q TRAK Plus monitor with the reading from the reference device If they differ use the A or V key to adjust the Q TRAK Plus display as needed so that the displayed measurement matches the measurement indicated by the reference device Press the key to accept and continue Calibration and Maintenance 41 TEMP CAL STEP 2 SET TO CORRECT TEMPERATURE eel F TEMP 1 TO ACCEPT AND CONTINUE Figure 4 16 Temperature Calibration Screen HUMIDITY CAL STEP 2 SET TO CORRECT HUMIDITY VALUE RH TO ACCEPT AND CONTINUE Figure 4 17 Humidity Calibration Screen Note Ifthe two readings are the same there is no need to proceed with the calibration Press the 1 key to terminate the calibration procedure 5 Q TRAK Plus monitor will check to see if the measured parameter is stable The screen displays the parameter in real time to show any fluctuations 6 At the successful conclusion of the calibration the screen will highlight CAL COMPLETE Press the J key to continue If an error is detected during the calibration process an error message will be displayed The previous calibration values are kept Press the f or the J key to return to the User Calibration Menu Restore Factory Calibration On very rare instances it may be necessary to restore the calibration of the instrument to the original factory settings This should only be used when absolutely nece
33. hat displays the instrument s model number serial number firmware version available memory date and current time The display will also show various status messages The startup sequence lasts for 20 seconds before entering Survey mode A reminder may appear when the instrument is due for servicing Escape key Enter key Arrow keys Power on off key Figure 3 1 Q TRAK Plus IAQ Monitor Keypad and Display Survey Mode After the instrument has completed its power up sequence the Q TRAK Plus monitor will automatically go into the Survey mode Note It is not unusual for gas sensors to take some time to become stable If readings appear unusual at start up please allow about one half minute for the sensors to stabilize In the Survey mode the Q TRAK Plus monitor displays all parameters simultaneously The values are updated once per second You may select any of the parameters to be shown in large font at the top of the display This is called the primary parameter Select the primary parameter by using the A or V keys to highlight the parameter of interest then press the key to display that parameter in large font at the top of the screen In addition the units of some parameters may also be changed For example temperature may be shown in either F or C To adjust the units scroll to the desired parameter using the A or W keys then adjust the units using the lt or key refer to the units section in this chapter see
34. highlight KAD KOLO eye gXO39 using the A or V key Pressing the J key will select the VELOCITY PROBE option and display the VELOCITY MENU shown in Figure 5 12 Auxiliary Sensors 53 VELOCITY MENU VELOCITY AVERAGING OLUMETRIC FLOW RATE Figure 5 12 Velocity Menu Options Select ZO AOAI UI COR PROB YN using the A or Y key and press the key to display the VOLUMETRIC FLOW RATE menu shown in Figure 5 12 VOLUMETRIC FLOW RATE SELECT SHAPE RECTANGULAR ROUND Figure 5 13 Selecting the Duct Shape Highlight the shape of the duct using the A or W key and press the key to select ACCEPT J EDIT lt gt ENTER DIMENSIONS RECTANGULAR WIDTH 019 0 CM HEIGHT 010 0 CM FLOW SL S 4 gt mm Figure 5 14 Entering Rectangular Duct Dimensions For square or rectangular ducts use the lt or key to move to a specific digit and use the A or W key to increment or decrement the digit accordingly Press the key when you ve entered the correct width dimension 54 Chapter 5 Repeat these steps to enter the height dimension When both dimensions are set correctly press the enter key J The minimum dimension that can be entered is 4 inches 10 cm and the maximum is 100 inches 300 cm Dimensions can be entered in increments of 0 1 inches or 0 1 cm When velocity measurements are in metric units m s then metric units cm are used for duct dimensions and you can select between m h and L s for fl
35. iiaeiai ea R iE 45 Model 8590 Auxiliary Temperature Sensor ACCESSOTY sseeseseeeseeeee 45 Model 8592 Air Velocity Probe AVP Accessory essceesecrereerseee 46 Basic Applications ss scssssssseessssee sone sbnesoeesbheessosnsseseosseessescterees 46 Upgrading Your Q TRAK Plus IAQ Monitor and TRAKPRO Data Analysis Software c cccsssssessseessessecseesteeseesees 47 Using the Telescoping Probe 0 0 0 eee cee cseereeeeeeeeeeeeeeeeeeseenseenaee 49 Measuring Air Velocity eee cee cseecseeeeeeeeeeeeeeeeeeeeeeeenseeseeesees 49 Measuring Air Flow Rate or Average Velocity e ee ee eeeeeeeeees 50 Q TRAK Plus LCD Display During AVP Operation 0 ee 51 Measuring Average Velocity sssesseseeseesseereeresrrrreresreersserereserreeresee 52 Measuring Volumetric Flow Rate seseeeseeeeeeseeeresesreeresresreeerrreeesre 53 MAIMED ANC sasesana nig enee aria EE EEE EEE EEEIEE E 57 Calibration esiin e E E ee a 57 Special Display Messages nseesssseeesseessseeereerrrresrsreerrsresesrreresreeresene 57 DLOLALE aa er aE schist a R E A E en tloeneeeoaes 57 TROUBLESHOOTING ici se eck cen Bheviceesncesacs skies ca nea or Eri 59 SPECIFICATIONS vscsrusis ceseti co cusetdiguescss snes eosinsuiecvepevisesassesaonerestyentteees 61 STANDARD VELOCITY VS ACTUAL VELOCITY MEASUREMENTS 3 scence Beis i ee 65 To obtain any of the listed Application Notes contact TSI at U S 800 874 281 1 651 490 2811 Fax 651 490 3824
36. in modes of operation Survey Sample and Data Log Survey mode When the Q TRAK Plus monitor is first turned on it will be in the Survey mode This mode is used to display the real time readings of the sensors The instrument will update the displayed readings once each second Sample mode The Sample mode is used to capture a 10 second averaged reading of all the parameters These single sample points may be stored in memory and can be displayed later or downloaded to a computer using the TRAKPRO software Data Log mode The Data Log mode is used to record readings over a period of time and store these readings in the instrument memory Data files may be downloaded to a computer for later analysis using the TRAKPRO software Q TRAK Plus has three different data log modes Q TRAK Plus IAQ Monitor Keypad The Model 8552 8554 Q TRAK Plus IAQ Monitor is controlled through a series of menus that are navigated using the arrow keys A V lt 4 P see Figure 3 1 The Enter key 4 is used to select the desired option while the Escape key 1 is used to go back to the previous menu Each screen contains prompting words and symbols When pressing the keys on the front panel the monitor beeps to confirm each key press The 1 key allows access to the Main Menu from the Survey mode Turning the Instrument On Use the ON OFF key to turn the Q TRAK Plus monitor on and off The instrument s power up sequence consists of a startup screen t
37. ized person either intentionally or inadvertently interrupt the measurements you can lock the keypad When lockout is engaged no values can be seen on the display The keypad lockout switch is located on the side of the Q TRAK Plus monitor between the probe mounting brackets It is a small slide switch and is recessed so that a pointed instrument must be used to move it The instrument case is labeled next to the switch with the symbols I and 0 With the switch in the I or on up position all keypad functions will work normally There are two ways to use the lockout switch the lockout is engaged priorto the the Q TRAK Plus monitor will start of logging switch is in the 0 operate normally until logging position begins the lockout is engaged after the the keypad will ignore any key start of logging presses until the switch is placed in the position Figure 3 11 Keypad Lockout Switch Operation 23 24 Setup Menu The setup menu allows access to a wide range of options for reviewing data examining statistics changing operational parameters performing calibration and more To reach the Setup menu from the Survey Mode press f key To access any of the submenus use the A or V key to highlight the menu and press the J key SETUP MENU TIME DATE BACKLIGHT CLEAR MEMORY CALIBRATION LANGUAGE PRINTER Figure 3 12 Setup Menu Time Date The Q Trak Plus monitor has an internal clock th
38. l only applies to Log 1 mode The log interval for Log 2 and Log 3 modes is adjustable only within TRAKPRO software Log Mode 2 and 3 The Q TRAK Plus monitor may be programmed for more sophisticated data logging modes using the TRAKPRO software and Log Modes 2 and 3 The software programs the start date start time test length logging interval number of tests and the time delay between tests Once the parameters are set the monitor can be placed in the field to begin data logging in either Log Mode 2 or Log Mode 3 If no start time or start date is set the log mode is initiated by pressing the J key Programming Log Mode 2 or 3 Using TRAKPRO Data Analysis Software 1 Make sure the Q TRAK Plus monitor is connected to the computer and turned on 2 Select Logging Setup from the Instrument Setup menu The TRAKPRO software retrieves the current settings for Log Modes 2 and 3 from the Q TRAK Plus monitor and displays them in the following dialog box 16 Q Trak Plus Logging Protocols Serial Number 8554 04011016 Number Tests Logged 4 Available Memory 100 Time Between Tests dd hh mm foo foo foo m foo foo foo E The Models 8552 and 8554 automatically record all channels including any auxiliary parameters that are installed Log 2 Log 3 Start Date MM dd yyyy Start Time hhmm 0353 Log Interval mm ss foo for E Cancel Test Length dd hh mm foo foo oz foo foo for E Number of Tests
39. n concentration of CO for the span calibration Contact your local TSI distributor for available CO calibration kits see Figure 4 4 Note lf using air for zero calibration it must be specified to have less than 10 ppm COs Ambient room air cannot be used see Appendix A The calibration procedure can be aborted at any time by pressing the 1 key Any step in the calibration procedure can be repeated by pressing the A key 1 To put the Q TRAK Plus monitor into CO calibration mode press the A or Y keys to select NA LIAYN ON in the Setup Menu 2 Select CO2 by pressing the A or V keys the in the calibration submenu and press the key see Figure 4 1 33 SELECT CAL ITEM 007 CO TEMPERATURE HUMIDITY AUX RESTORE DEFAULTS Figure 4 1 Calibration Submenu Screen Locate the calibration collar and slide it over the main probe see Figure 4 4 C02 ZERO CAL STEP 1 SLIDE CO2 CAL COLLAR OVER PROBE ATTACH ZERO GAS TO CAL COLLAR NEXT Figure 4 2 CO Zero Calibration Step 1 Screen 3 Install the regulator onto the zero calibration gas cylinder and connect tubing from the cylinder to the fitting marked GAS IN Press the J or Y key to continue 4 Turn the gas on 0 3 L min Make sure the cylinder is not empty It should quietly hiss when it is on Press the J or V key to begin zero gas calibration see Figure 4 3 CO2 ZERO CAL STEP 2 TURN ON ZERO GAS PRESS J TO BEGIN ZERO GAS SAMPLE
40. nsions All data acquired can be evaluated using TRAKPRO Data Analysis Software A picture of the AVP is shown in Figure 5 2 Paa Auxiliary PCB Probe Cable Cable Connector Figure 5 2 Model 8592 Air Velocity Probe Accessory The AVP can be attached to the Q TRAK Plus at anytime and will immediately display a real time air velocity reading as an additional parameter along with the others normally shown on the Q TRAK Plus 46 Basic Applications The AVP adds new functionality to the Q TRAK Plus IAQ Monitor The Q TRAK Plus user can collect velocity based measurements and conduct a greater variety of studies Some applications are e HVAC duct velocity flow rate measurements e Air movement comfort in occupied zones e Volumetric air flow in occupied zones e Fume hood face velocity testing Chapter 5 e Clean room studies e Draft measurements e Wind tunnel testing e Filter velocity checks Upgrading Your Q TRAK Plus IAQ Monitor and TRAKPRO Data Analysis Software Past users of Q TRAK Plus IAQ Monitor instruments will have to upgrade their instrument to be compatible with the Model 8592 Air Velocity Probe accessory Additionally for users of TRAKPRO Data Analysis Software a new revision must be downloaded from the TSI website TSI provides a simple automated procedure for you to easily perform these upgrades These upgrades can readily be done in the field and do not require returning your instrument t
41. o TSI When your instrument is returned to TSI for annual calibration or service TSI will upgrade the firmware for you e Instrument firmware and TRAKPRO upgrades are necessary for compatibility with the Model 8592 Air Velocity Probe accessory TSI recommends uploading all logged data from the Q TRAK Plus instrument to TRAKPRO for permanent storage before upgrading the instrument firmware Not doing so could result in the loss of logged data A product information CD is included with the purchase of a Model 8592 Air Velocity Probe accessory Briefly the steps to upgrade the firmware of your Q TRAK Plus instrument are e Connect the Q TRAK serial cable to both the instrument and your computer e Insert the product CD into your computer s CD disk drive e Select the Upgrade Q TRAK Plus Firmware menu option e Click Start to commence the upgrade process e When the upgrade process is completed click Close to exit The first step in the upgrade procedure is to connect the Q TRAK serial cable to your instrument and your computer just as you would when using TRAKPRO software Auxiliary Sensors 47 48 After inserting the product CD into your computer s CD disk drive the CD will automatically run and its main menu will be displayed From the CD s main menu click once on the Upgrade Q TRAK Plus Firmware menu option The Q TRAK Plus Flash Memory Programmer will be started and its menu will be displayed as shown in
42. ow rate units This selection is made by using the lt or P key ACCEPT J EDIT lt gt ENTER DIMENSIONS ROUND DIAMETER 01 0 CM FLOW SL IS lt gt Figure 5 15 Entering Round Duct Dimensions The dimensioning procedure for round ducts is similar except only the duct diameter needs to be entered as shown in Figure 5 15 Follow a similar key sequence to enter the diameter dimension and flow rate units VOLUMETRIC FLOW RATE 382 5 L S FLOW 4 25 M S VEL SAMPLES TAKEN 2 CAPTURE SAMPLER SAMPLING COMPLETED C Figure 5 16 Sample Capturing Display A duct traverse measurement can now be conducted from which an average velocity and flow rate will be calculated Each velocity sample is captured by pressing the key while is highlighted The word SAMPLING will be blinking while the measurement is being acquired Be careful to keep the probe motionless in the same position while the measurement is being acquired After the measurement is complete CAPTURE SAMPLE will be displayed and the instrument is ready for the next sample Up to 99 sample points may Auxiliary Sensors 55 be taken to comprise a sample set The measured velocity and a calculated volumetric flow rate will be displayed for each sample taken When you have completed acquiring all traverse sample points use the V key to highlight YAYAN eR AANI and press the A key to complete the measurement VOLUMETRIC FLOW RATE 3 59 M S MIN VEL 5 20 M S M
43. poer e hick piers eeens 3 of reading or 3 ppm whichever is greater add 0 5 C 0 28 F away from calibration temperature Resolution csceecseeseeseeereeeeeeeeees 1 ppm Response time sssr lt 60 seconds to 90 of final value 61 Power Requirements Batteries iiae n ied Four AA size alkaline or rechargeable or AC adapter oo eeeeceeeecsseceenreeseeeeee 6 VDC nominal 300 mA Q TRAK Plus monitor mates with 5 5 mm OD x 2 1 mm ID plug center pin positive Approximate battery life Up through 20 hours alkaline Physical External dimensions c000000 107 mm x 183 mm x 38 mm 4 2 in X 7 2 in X 1 7 in Probe length seese 31 2 mm 12 3 in Probe diameter cccccscceseeeee 1 8 cm 0 75 in Welsht wiic ciastetin Qa 0 59 kg 1 3 pounds with batteries Display ress eea e isd evteests eee 128 x 64 Graphics display module with backlight Maintenance Schedule Factory calibration eeee Annually User Calibration cccccccccceceseeeee As needed Serial Interface PY POs se sscesetss tuscspsnasoddeosessbaususeistess RS 232 Baud tat nnnon tomani an nie 9600 Data 0 arraren negnn 8 Stop DSen ainese rea i 1 Handshaking ceescesseeeseeeeeeees None Data format e ceeceeceeeeeeeeees ASCII 62 Appendix A Model 8592 Air Velocity Probe Velocity Temperature Range Operating Range eee Storage Range eens Volumetric Flow Rate Prob
44. pyl alcohol may be applied to clean the AVP Be sure the probe is disconnected from the Q TRaAK Plus before cleaning so the AVP is not powered Do not use strong solvents or high pressure air to clean the probe or sensor permanent damage could result Calibration For the highest degree of accuracy TSI recommends that instruments be returned to the factory annually for calibration For a nominal fee the probe will be checked and recalibrated to original factory specifications and returned to you with a new National Institute of Standards and Technology NIST calibration certificate To have your AVP recalibrated please follow TSI s Returned Material Authorization RMA procedure on our web site at http rma tsi com Special Display Messages In extreme cases temperature or velocity conditions may exceed the operating specifications of the AVP potentially inhibiting the AVP from providing an accurate measurement If the temperature at the sensor exceeds the maximum operating limit the Q TRAK Plus display will read TEMP HI If temperature is below the minimum operating range the instrument display will read TEMP LO If the air velocity exceeds the upper range of the specification the Q TRAK Plus display will read VEL HI Under such conditions the measurement accuracy may be compromised and the AVP might be damaged Storage It is recommended that the Air Velocity Probe accessory be stored unattached from the Q TRAK Plus IA
45. rd the data point Note that the first sample is given the name number Sample 1 The sample number will automatically increment for each subsequent sample Printing Saved Single Data Points A hardcopy of the results of the saved test sample can be generated if the portable printer is attached to the Q TRAK Plus monitor and the Printer Output feature is enabled Main Menu The Main Menu can be accessed through the QUO Qy4 GY 009 by pressing the f key The Main Menu Mode is the highest level in the menu structure It provides a central point where you can select alternate modes of operation MAIN MENU TIME CONSTANT LOG MODE 1 SETUP VENTILATION RATES STATISTICS Figure 3 4 Main Menu Chapter 3 Time Constant Highlighting and selecting UAYISTOO VRS VBI displays the Time Constant Menu This allows you to adjust the dampening time constant that will affect the real time display of parameters in the Survey mode only The values shown in the Survey Mode are updated once per second however the readings shown are averaged over the time constant period Log Mode To select a Log Mode scroll to using the A or W key Adjust the mode using the or key to select LOG MODE 1 2 or 3 Press the key to start logging data Log Mode 1 is programmed and initiated by the keypad Log Mode 2 and Log Mode 3 are programmed using the TRAKPRO software Log Mode will only record data that is visible on the display Log Mode
46. re select Parameters Instrument Setup Other The TRAKPRO software retrieves the current settings from the Q TRAK Plus monitor and displays them in the dialog box displayed in Figure 3 16 Chapter 3 m Backlight Backlight Duration No Backlight 7 Cancel Figure 3 16 Instrument Parameters 3 Click a time duration in the backlight box 4 Click Send to reprogram the Q TRAK Plus monitor Clear Memory The CLEAR MEMORY option on the Setup menu allows you to clear the instrument memory To preserve data integrity the Q TRAK Plus monitor does not automatically erase the older data files It requires a deliberate step on your part to clear the memory Note Performing this step will erase all logged data files or recorded single data points from the instrument s memory There is no reversing this step once it is performed It is recommended that you download all desired logged data or single data points using the TRAKPRO software before clearing the memory To clear the memory select the CLEAR MEMORY option on the Setup menu After pressing the J key the following screen is displayed CLEAR MEMORY SINGLE POINT MEMORY LOGGED TEST MEMORY Figure 3 17 Clear Memory Operation 27 The highlighted default selection is Single Point Memory Pressing erases all Single Point data The Logged Test Memory is unaffected Logged Test Memory is cleared in a similar manner Single Point Memory or Logged Test
47. replaced as a result of repair services are warranted to be free from defects in workmanship and material under normal use for 90 days from the date of shipment c Seller does not provide any warranty on finished goods manufactured by others or on any fuses batteries or other consumable materials Only the original manufacturer s warranty applies d Unless specifically authorized in a separate writing by Seller Seller makes no warranty with respect to and shall have no liability in connection with goods which are incorporated into other products or equipment or which are modified by any person other than Seller The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE TO THE EXTENT PERMITTED BY LAW THE EXCLUSIVE REMEDY OF THE USER OR BUYER AND THE LIMIT OF SELLER S LIABILITY FOR ANY AND ALL LOSSES INJURIES OR DAMAGES CONCERNING THE GOODS INCLUDING CLAIMS BASED ON CONTRACT NEGLIGENCE TORT STRICT LIABILITY OR OTHERWISE SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE OR AT THE OPTION OF SELLER THE REPAIR OR REPLACEMENT OF THE GOODS IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL CONSEQUENTIAL OR INCIDENTAL DAMAGES SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION DISMANTLING OR REINSTALLATION COSTS OR CHARGES No Action regardless of form may be brought again
48. ring LOG MODE 1 see Log Mode 2 and 3 section later in this chapter for more information about data logging The menu screen is shown below in Figure 3 7 SELECT LOG INTERVAL LOG INTERVAL 00 01 MM SS 1 TO ACCEPT LOG INTERVAL AND BEGIN LOGGING Figure 3 7 Log Mode 1 Select Log Interval 14 Chapter 3 The default values for log intervals are set at 1 second 1 minute 5 minutes 15 minutes and 30 minutes To select the logging interval press either the A or V key until the desired interval is shown When the key is pressed the program begins the data logging process The log interval options for LOG MODE 1 may be programmed using TRAKPRO software 1 Make sure the Q TRAK Plus monitor is connected to the computer and turned on 2 Select Instrument Setup Parameters Logging Intervals from the software menu The current settings in the Q TRAK Plus monitor will be retrieved and displayed in the dialog box You may program in new values for each of the five log intervals within defined limits Q Trak with CO Logging Intervals E3 mrm ss 00 01 to 59 59 i 01 00 Ina 05 00 Cancel mea 15 00 l HF 5 30 00 The first logger interval will be the curent value for log mode 1 Figure 3 8 Logging Intervals 3 Select Send to program the Q TRAK Plus monitor with the new values 4 Disconnect the Q TRAK Plus monitor from the computer Operation 15 Note This log interva
49. rs per second m s sm s Measurements are made by holding the probe so the sensor window is positioned in the desired measurement location with the sensor window perpendicular to the air flow stream Be sure the sensor window is fully exposed and the orientation indicator arrow on the probe tip is pointing downstream Hold the probe as still as possible while the measurement is taken When positioning the probe for a measurement be certain the orientation indicator arrow on the probe tip is pointing downstream in the direction of the air flow Measuring Air Flow Rate or Average Velocity Volumetric flow rate can be estimated based on an average velocity measurement across a known area such as a duct You can input the cross sectional dimensions for round square or rectangular ducts and flow rate will be displayed in cfm m h or L s accordingly The measurement units can be changed by simply highlighting the flow displayed with the A or V key and then changing the units displayed using the lt or P key Table 5 2 Flow Rate Measurement Units Flow Rate Actual Standard Measurement Units Flow Rate Units _ Flow Rate Units Cubic feet per minute cfm scfm Cubic meters per hour m h sm h Liters per second L s sL s Other applications where average velocity calculations are important include wind tunnel testing across a filter face and testing fume hood face velocities In addition to calculating the aver
50. service The use of an improper AC adapter will cause severe damage to the instrument Instrument Software and Communications Setup The Q TRAK Plus IAQ Monitor comes with special software called TRAKPRO Data Analysis Software which is designed to provide you with maximum flexibility and power when using the Q TRAK Plus monitor The following sections describe how to install the software and set up the computer Setting up TRAKPRO Data Analysis Software Follow the instructions on the label of the TRAKPRO software to install the software on your computer TRAKPRO software contains a very comprehensive Help Function This utility provides all the necessary information to guide you in all aspects of software operation The software is shipped on a CD ROM It is available on diskette format upon special request Updates are available from the TSI website Connecting the Q TRAK Plus IAQ Monitor to the Computer Each Q TRAK Plus monitor comes equipped with an RS 232 9 pin serial cable One end of the cable is a 9 pin D subminiature connector labeled COMPUTER the other end is an RJ 45 modular connector that connects with the Q TRAK Plus communications port Serial port connectors always have pins male on the computer side 1 Locate an available RS 232 serial port on your computer for example COM1 or COM2 2 Connect the RS 232 cable to the available serial port on your computer 3 Connect the RJ 45 connector to the Q
51. ssary 1 To restore the factory calibration select ININA KO i309 0 FV 0 Bs WS in the Calibration Select page by using the A or V key and press the key 42 Chapter 3 2 The following warning message appears on the display Press the W key to continue or the J key to abort Figure 4 18 RESTORE FACTORY CAL IN SOME CASES IT MAY BE NECESSARY TO RESTORE FACTORY CAL VALUES DO NOT RESTORE UNLESS REQUIRED NEXT V QUIT Figure 4 18 Restore Defaults Screen 3 Another menu will display a selection of parameters in which to restore the factory calibration Use the A or V key to select the desired parameter and press the J key to continue or 1 to abort Figure 4 19 SELECT ITEM TO RESTORE CO TEMPERATURE HUMIDITY Figure 4 19 Select Item to Restore Screen 4 The display will indicate that the factory calibration of the desired parameter has been restored A reminder that a user calibration should be performed is also shown Press either the J or f key to return to the Setup menu Storage Precautions When storing the Q TRAK Plus monitor for more than 30 days remove the batteries This prevents damage due to battery leakage This instrument must be stored in a location where the temperature remains between 20 and 60 C 4 and 140 F and the relative humidity is between 15 and 90 Calibration and Maintenance 43 Chapter 5 Auxiliary Sensors Auxiliary sensors are available for the Q TRAK Pl
52. st Seller more than 12 months after a cause of action has accrued The goods returned under warranty to Seller s factory shall be at Buyer s risk of loss and will be returned if at all at Seller s risk of loss Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND LIABILITY may not be amended modified or its terms waived except by writing signed by an Officer of Seller Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers our service policy is designed to give prompt attention to any problems If any malfunction is discovered please contact your nearest sales office or representative or call TSI s Customer Service department at 800 874 2811 USA or 001 651 490 2811 International 3 CONTENTS UNPACKING AND PARTS IDENTIFICATION o s 1 Unpacking the Q TRAK Plus IAQ MOnitor ccccccssesssesseeseestesesseeesees 1 Optional ACCESSOTIES sooren alre a ei ae ee Ra i 3 SETTING UP harn o a e a A A E anid due 5 Supplying Power to the Q TRAK Plus IAQ Monitor ccssccssesseeeseesees 5 Installing the Battertes 2 22 s s scssceseuassessscaisastessnctsndsssuscnzssoeeotaistyeseess 5 Using the AC Adapters eeni neee E be ine 5 Instrument Software and Communications Setup ceeeseessseeseeeeenees 5 Setting up TRAKPRO
53. ument stand is designed to hold the instrument and the probe at a convenient viewing angle The instrument stand may also be used to hang the monitor on a wall To use the stand simply detach the battery door from the monitor and replace it with the instrument stand see Figure 2 2 Using the Optional Probe Stand A separate probe stand is available as an accessory Use the bracket that is located on the probe and slide it on the probe stand as shown in Figure 2 2 Setting Up 7 Figure 2 2 Optional Probe and Instrument Stands Connecting the Optional Portable Printer To connect the portable printer to the Q TRAK Plus monitor do the following 1 Ensure that the Q TRAK Plus monitor and printer are off 2 Locate the printer interface cable and connect the 9 pin end labeled PRINTER to the printer and the other end to the communications port on the Q TRAK Plus monitor 3 Turn on the Q TRAK Plus monitor then turn on the printer Note Always turn on the Q TRAK Plus monitor before turning on the or random characters reset it by turning it off and then on again If necessary refer to the Portable Printer Manual 8 Chapter 2 Chapter 3 Operation Overview The Model 8552 8554 Q TRAK Plus IAQ Monitor measures CO temperature relative humidity calculates dew point wet bulb and outside air and measures one 1 additional parameter The Model 8554 will also measure CO The instruments have three ma
54. us test use the A or V key Press the J or f key to return to the statistics menu The CO summary screen is shown in the following example LOGGED TEST STATS TEST 3 CO2 MIN 800 PPM MAX 950 PPM AVG 805 PPM NEXT TEST VALUE lt gt Figure 3 25 Logged Test Statistics CO Summary 32 Chapter 3 Chapter 4 Calibration and Maintenance The Model 8552 8554 Q TRAK Plus IAQ Monitor requires very little maintenance to keep it performing well The Q TRAK Plus monitor may be calibrated in the field using the instructions below Even so we recommend that you return your Q TRAK Plus monitor to TSI for calibration annually For a nominal fee we will quickly clean and recalibrate the unit update software and firmware and return it to you in as new working condition along with a Certificate of Calibration and NIST Traceability The factory calibration is more precise than can be accomplished with the procedures below This annual checkup helps ensure that the Q TRAK Plus monitor is always in good operating condition Calibrating the CO Sensor TSI recommends calibrating the Q TRAK Plus monitor CO measurement monthly to help ensure accurate readings To calibrate the CO sensor please follow the procedure below You can also follow the prompts displayed on the screen while in user calibration mode You will need a cylinder of zero CO air or nitrogen for the zero calibration and a cylinder of gas with a know
55. us that enhance the functionality of the instrument If you purchased an auxiliary sensor to measure another parameter you may attach the sensor at any time during the Survey mode and the Q TRAK Plus monitor will display the values instantly The auxiliary sensor port is located at the top of the instrument To attach an auxiliary sensor to the instrument 1 Grasp the connector on the auxiliary sensor 2 Align the black release button with the back of the instrument 3 Press the connector into place Black release N 3 Figure 5 1 Attaching an Auxiliary Sensor To detach an auxiliary sensor from the instrument 1 Grasp the connector on the auxiliary sensor 2 Press the black release button 3 Pull the connector away from the instrument Model 8590 Auxiliary Temperature Sensor Accessory The Model 8590 Auxiliary Temperature Sensor accessory allows you to make and record air temperature measurements The Auxiliary Temperature Sensor accessory can be used to conduct temperature sampling and to log a 45 second temperature measurement All data acquired with this accessory can be evaluated using TRAKPRO Data Analysis Software Model 8592 Air Velocity Probe AVP Accessory The Model 8592 Air Velocity Probe AVP accessory allows you to make and record air velocity and flow rate measurements The AVP can readily be used to conduct velocity sampling and to calculate air flow rate using average velocity and your input duct dime
56. will be displayed The previous calibration values are kept Press the 1 or key to return to the Calibration Menu CALIBRATION ERROR DETECTED VALUES UNCHANGED PRESS 1 TO CONTINUE Figure 4 14 Error Detected or Process Aborted Screen 40 Chapter 3 Calibrating the Temperature or Relative Humidity Measurement To perform a temperature or humidity calibration on the Q TRAK Plus monitor you will need a reference temperature or humidity device one that is more accurate than the Q TRAK Plus monitor To obtain the best accuracy care must be taken so that both the Q TRAK Plus probe and the reference device are able to sense the same air Also it s best to calibrate at a temperature and humidity that is near the range you typically measure Note User calibration of humidity sensor can only be done in units of RH Never submerge the probe or Q TRAK Plus monitor in water or any other liquid The sensor will be damaged 1 Locate the Q TRAK Plus probe and the reference device sensor so that they sense the same air conditions 2 Put the Q TRAK Plus monitor into calibration mode by selecting either temperature or humidity in the Calibration Menu by using the A or V keys and press the J key refer to Figure 4 15 SELECT CAL ITEM CO2 CO TEMPERATURE HUMIDITY AUX RESTORE DEFAULTS Figure 4 15 Calibration Submenu Screen 3 A reminder appears to ensure that the environmental conditions are stable before continuin
57. xtend the probe grasp the handle in one hand while gently pulling on the probe tip with your other hand Be aware that the AVP cable slides through the telescoping probe as it is extended or retracted To retract the probe grasp the handle and push the probe tip toward the handle The rest of the probe sections can then be collapsed using a finger or the palm of your hand Be sure not to block or hold the cable at the handle since the cable should freely pass through the handle while the probe is collapsed The thermistor bead velocity sensor is located in the sensor window of the telescoping probe When using the probe be sure that the last section of the probe is fully extended exposing the entire sensor window Measuring Air Velocity When the AVP is connected to the Q TRAK Plus velocity measurements will automatically be displayed along with the standard Q TRAK Plus parameters The velocity measurement units are user selectable between feet per minute fpm and meters per second m s and between actual and standard units as illustrated in Table 5 1 To select the velocity measurement units highlight the velocity reading Auxiliary Sensors 49 50 with the A or V up down arrow key and then change the units displayed using the lt or gt left right arrow key Table 5 1 Velocity Measurement Units Velocity Actual Standard Velocity Measurement Units Velocity Units Units Feet per minute fpm sfpm Mete

Download Pdf Manuals

image

Related Search

Related Contents

F-01Cカタログダウンロード  Betriebsanleitung Top5 ab SN 902384  Bosch 2609255003    TADC MSRラ蚊  E21A・E18M - NEC Corporation  Cullmann LAGOS Compact 290  Handbuch - U.S. Robotics  La Crosse Technology BC700-CBP battery charger  pompiers.fr : Les extincteurs, types et fonctionnement  

Copyright © All rights reserved.
Failed to retrieve file