Home

User Manual - Alpha Modalities

image

Contents

1. NOTE The lift unit and carriage must be installed and inspected by authorized personnel according to instructions issued by Handicare AB Periodic inspection must be performed at least once yearly Use only original spare parts Service and maintenance must be performed according to instructions in the manual The lift must be charged for at least 3 hours before it is used for the first time Check to ensure that the charging points are set for the charging alternative chosen As standard the lift is factory pre set for charging at a charger station with installed spring pins When installing the lift unit for in rail charging on the rail system the system must be fitted with charger rails The charger plate must be set for in rail charging Exercise caution when installing the lift unit so that the charger wheels do not come in con tact with the rail since this may cause short circuiting fy Alteration of the charging alternative must be performed by authorized personnel according to instructions issued by Handicare AB Spring pin SystemRoMedic MANUAL Charger station 1 Check to ensure that the charging points are set for the charger station the charger plate is secured with the spring pins see figure 1 2 Check to ensure that the charger plate springs in and out when pressure is applied to the charger plate see figure 2 3 Guide the lift unit onto the rail so that the plus on the charging point is on
2. point 10 C to 40 C not exceeding normal relative humidity 30 96 to 75 The air pressure shall be between 700 and 1060hPa Service agreements Handicare offers the possibility of service agreements for maintenance and regular testing of your lift Contact your local Handicare representative SystemRoMedic MANUAL 19 Technical information Lift motor Lifting speed Charger IN Charger OUT Batteries Motor cover Hand control Emergency lowering Lift weight Lifting range SWL RiseAtlas450 SWL RiseAtlas625 Sound level Pressure hand control Lift measurements Intermittent operation IP class lift unit IP class hand control Expected service life 24VDC 3 9 cm s 5 0 cm s 1 5 inch s 2 0 imch s with without load Mascot 2240 100 240VAC 50 60Hz Max 0 9A 41 VDC 0 3 Max 0 9A 24 VDC 20st 1 2VDC 3 2 Ah NIMH20XA3200 Flame resistant ABS plastic Electronic Mechanical and electrical 7 7 5kg 15 4 16 5 lbs 200 240 cm 78 7 94 5 inch 205 kg 450 Ibs 285 kg 625 lbs Unload 55 dB Max load 57 dB Buttons on hand control 4 N 23 3 15 1 x 14 6 cm 9 2 x 5 9 x 5 7 inch Op 10 90 active op max 2 min Out of a time of 100 active time must be less than 10 though not more than 2 min IP X3 IP X4 10 years a The device is intended for indoor use Type according to the degree of protection against electric shock Class equipment The product fulfills
3. the demands stated in the MDD 93 42 EEC Detailed description A 25 1 Outlet for handset ABA 2 Handset 3 Lift strap 4 Emergency stop and electrical emer B C 1 lowering 5 Manual emergenoy lowering raising 6 Control panel 4 64 160mm C 9 1 B Lift unit overall height160mm 6 C Installation height 230mm 16 MANUAL SystemRoMedic SystemRoMedic MANUAL 17 MANUAL SystemRoMedic SystemRoMedic MANUAL 19 oystemRoMedic For 25 years we have applied ourselves wholeheartedly to developing smart and easy to use assistive devices for easy transfers and to making life and work easier for both patients and personnel in the care sector Experience innovation and training are the basis for SystemRoMedic a total solution for every imaginable transfer situation Transfer products for moving patients between locations Positioning products for repositioning in the same location Support products that provide mobility support Lifting products adapted for lifting The philosophy behind SystemRoMedic is to prevent occupational injuries while improving the patient s sense of inde pendence and dignity Through a combination of training and a complete range of transfer assistive devices SystemRoMedic offers the means to improving both the work environment and the quality of care while enab
4. CE RiseAtlas450 inn d RiseAtlas625 HiseAtlas art no 50100042 50100043 50100044 50100045 Max RiseAtlas450 205 kg 450 Ibs RiseAtlas625 285 kg 625 Ibs The RiseAtlas lift unit has been developed for use in combination with the appropriate lifting accessories for transferring patients conveniently between two locations RiseAtlas is installed on the appropriate rail system to cover lifting needs in the given situation The rail system may consist of a straight rail with or without rail curves or a traverse system It is also possible to transfer patients between two rooms using RiseAtlas RiseAtlas450 and RiseAtlas625 are available in two models RiseAtlas450 625M without a built in transfer motor HiseAtlas450 625M with a built in transfer motor Handicare s SystemRoMedic product series includes a range of lifts slings and other accessories SystemRoMedic adopts a holistic approach to patient transfers and is organized in four categories transfer positioning support and lifting Functional inspection Visual inspection Inspect lift functions regularly Check to ensure that material is free from damage Before use Ensure that the product is correctly assembled Check rail end stops Check slingbar connection and safety latch function Check lifting movement 3 Always read the manual Always read the manuals for all assistive devices used during a transfer Keep the m
5. anual where it is accessible to users of the product The user attendant shall be trained professional with thorough understanding of the equipment and lifting of disabled persons e handicare Manual nr 00732 En Ver 1 120124 Table of contents Mounting Unpacking and check Transport and storage Installing the lift unit Different rail systems Final inspection Using the product Important information Before use Safe working load Description of functions and symbols Charging the lift unit Hand control Control panel Preventive service maintenance Emergency stop Electrical emergency lowering Manual emergency lowering raising Trouble shooting Accessories Maintenance Technical information Detailed description 3 6 10 11 11 12 12 12 13 14 15 16 16 MANUAL SystemRoMedic Mounting Unpacking and check With RiseAtlas the following is included in the package 1 RiseAtlas lift unit 1 Handset 1 Manual 1 Charger wall mounted or rail mounted 1 set Charger adaptor for charger station and contact for charger 1 Key for manual emergency lowering inserted in cover Check to ensure that no components are left in the packaging Inspect the lift unit for signs of damage Transport and storage RiseAtlas is shipped with the emergency stop activated to prevent battery discharge during transport or long term storage Installing the lift unit
6. e safe working loads The lowest allowable safe working load always determines the safe working load of the assembled system Always check the safe working loads for the lift and accessories before use Contact your dealer if you have any questions Description of functions and symbols HiseAtlas has many built in functions Each function is described in the respective section in the manual or in the manual for the respective accessory Below is a brief description of the functions that are available for caregivers and patients e Electronic soft start and soft stop e Electrical and mechanical emergency lowering Emergency stop button on cover Mechanical fall protection e Overload protection e ON OFF on handset ED symbol indicates when it is time for service or maintenance ED symbol for low battery charge LED symbol for battery charging Symbols and functions Wi Dy Q A 1 Symbol indicates battery status 2 Symbol indicates when the lift requires service or maintenance 3 Symbol for overload 5 Symbol for emergency stop 4 Symbol for read manual 6 Symbol for directional indication transfer motor 7 Symbol for warning Read the manual 8 MANUAL SystemRoMedic Charging the lift unit RiseAtlas can be used with two different charging systems 1 End of rail charging at a charger station 2 In rail charging For both alternatives the lift unit is charged with a charger
7. g option is used 2 Information concerning service and maintenance This symbol is activated once the lift unit has been in service for 12 months 3 RiseAtlas is equipped with overload protection which prevents the lift from lifting more than the maximum rated load A symbol for overload and a warning triangle light up and a signal is heard If this symbol is activated the patient is too heavy 4 The book symbol is activated prompting the user to read the manual Often the book sym bol will light up at the same time as another symbol lights up indicating that the user should read the manual before using the lift again 5 RiseAtlas is equipped with night lighting which is activated by pressing and holding the ON button on the handset for 3 seconds The LED lamp lights for 4 min if it is not inactivated via the OFF button on the handset 6 Some HiseAtlas models are equipped with an extra accessory that causes a symbol to light 9 x BE up for 30 sec if the emergenoy stop function is activated This lamp will light up each time the lift unit is switched on via the handset 7 The RiseAtlas T models are equipped with a directional indicator When lateral transfer is gt gt gt activated via the handset arrows indicate the direction of travel 8 The warning triangle is activated prompting the user to read the manual Often the warning triangle will light up at the same time as another symbol lights up indicating that
8. harder MHS2500 are used together with stationary or mobile lifts for weighing of users Article no 70100002 ng Article no 70100003 Assistive devices for positioning SystemRoMedic includes a wide range of functional comfortable high quality assistive devices for positioning that can be adapted for different types of lifting and for patients with different needs Ready for life Sling bar B The Ready for life four point slingbar Sling bar RFL X4 is designed to provide more space in the sling for example for obese and or pain sensitive users Article no 70200017 LM 14 MANUAL SystemRoMedic Maintenance The lift must undergo thorough inspection at least once per year Inspection must be performed by authorized personnel and in accordance with Handicare s service manual Repairs and maintenance may only be done by authorized personnel using original spare parts Spent batteries are to be left at the nearest recycling station Spent batteries can also be returned to Handicare a Handicare dealer for recycling Cleaning disinfection Clean the lift with warm water or rubbing alcohol Do not use cleaning agents containing phenol or chlorine as this may damage the material Storage To ensure maximum battery life charge regularly We recommend that the charger is plugged into the lift unit when not in use Do not allow battery charge to fall below 2596 Store the lift at a temperature above freezing
9. ifting of disabled persons It is important that proper accessories and techniques are used by the user attendant and it is also very important that a patient never is left unattended in a lifting situation Warranty does not apply if repairs or alterations are made to the lift by personnel who are not authorized HiseAtlas must not come in direct contact with water e o ensure optimal function the lift must be inspected regularly See section on Maintenance e Under no circumstances may safe working load be exceeded See section on technical information and the product label on the lift ifting accessories must be properly trial fitted and tested in relation to the patient s needs and functional ability e Particular care must be taken when using the equipment near strong sources of interference such as diathermy appa ratus so that e g diathermy cables and similar equipment are not placed near the device When in doubt consult your equipment technician or the supplier Before use Inspect the lift for signs of wear and damage e Check the hand control cable for signs of wear Check to ensure that hand control markings coincide with lift functions Check battery charge level SystemRoMedic MANUAL Safe working load SWL RiseAtlas450 205 kg 450 Ibs RiseAtlas625 285 kg 625 5 Different products on the same lift system lift unit sling and other lifting accessories may have different allowabl
10. ling significant cost savings Our mission to help people has always been and will continue to be the driving force of innovation We love easy transfers Contact your local distributor if you have any questions about the product and its use See www handicare com for a complete list of distributors Always make sure that you have the right version of the manual The most recent editions of manuals are available for downloading from our website www handicare com handicare Handicare AB Veddestav 15 Box 640 SE 175 27 Jarfalla SWEDEN Tel 46 0 8 557 62 200 Fax 46 0 8 557 62 299 E mail info handicare se Internet www handicare com
11. o stop check to ensure that the overload symbol is not activated that the battery symbol does not indicate low battery charge level If the lift cannot be run down check to ensure that load is applied to the lift strap that the battery symbol does not indicate low battery charge level that the directional indicators on the buttons arrow up and arrow down are correct If unusual sounds are heard try to locate the source of the sound Take the lift unit out of service and contact your Handicare dealer SystemRoMedic MANUAL Us Accessories SystemRoMedic Slings SystemRoMedic offers a wide variety of functional and comfortable high quality lifting slings adapted to meet the different requirements of all kinds of lifting situations and users The lifting slings are available in four different materials and in sizes ranging from XXS to XXL All models are safe and very easy to use Sling bars SlingBar is SystemRoMedic s aluminium sling bar It s available in three different widths and is p therefore suitable in most lifting situations and for users of all sizes P SlingBar S article no 70200001 SlingBar M article no 70200002 SlingBar L article no 70200003 ec DANA elas SwiftHook for sling bar article no 70200008 MAU StretcherBar article no 70200006 and StretcherSling article no 46502007 for lifting in a supine position Scale SystemRoMedic s scales C
12. operly e Check the entire length of the strap for signs of damage and inspect seams for wear Press the up button and run the strap all the way up Then press the down button and run the strap all the way down e Test lift function by lifting a person not a patient using an approved sling At the same time check both the manual and electrical emergency lowering function with someone in the lift See section on Emergency lowering Check to ensure that the rail system is equipped with end stops the lift unit to the charger station at the end of the rail and check to ensure that the battery symbol and control panel light up e Check the hand control cable for signs of wear Check to ensure that hand control markings coincide with lift functions e he user attendant shall always pay attention to and report any unusual noises or vibration of the equipment NOTE Before the lift is used for the first time it must be charged for at least 3 hours See section on charging the lift HiseAtlas is shipped with the emergency stop activated to prevent battery discharge during transport or long term storage Always keep the key for the emergenoy lowering device close to the lift All personnel who use the lift must be informed of where the key is kept 6 MANUAL SystemRoMedic Using the product Important information e The user attendant shall be trained professional with thorough understanding of the equipment and l
13. s in rail charging The battery symbol lights green when battery capacity exceeds 5096 When the lift unit is activated the battery symbol lights green for 4 min If the battery symbol and the book symbol blink orange when the lift unit is started the circuit for the charging system has been broken and must be checked by a technician NOTE Use only the charger that is intended for the lift unit For maximum service life charge batteries regularly Do not allow the battery charge level to fall below 2596 orange symbol on control panel Emergenoy stop must not be activated during charging Batteries may only be replaced by authorized personnel SystemRoMedic MANUAL Hand control Connection of hand control to lift unit The outlet for the hand control is on the underside if the lift unit figure 1 Press the handset in securely Raising lowering The buttons with the black arrows raise and lower the lift strap Raising lowering is interrupted as soon as pressure on the buttons is j released figure 2 B Transfer along rail applies to RiseAtlas450T 625T The buttons with the black arrows in the lateral direction transfer the lift unit along the rail figure 3 ON OFF via hand control The hand control is equipped with a green ON button for activating the lift unit and a button for switching it off OFF figure 3 The lift unit switches off automatically after about 4 minutes of inactivity Normally the lift mu
14. see figure 2 Manual emergency lowering raising Remove the plastic cover on the same side as the emergency stop strap figure 3 On the inside of the cover there is an Allen key Remove the Allen key and insert it into the hole as indicated figure 4 Turn it clockwise to lower and counter clockwise to raise the slingbar Always ensure that emergency lowering is done in an appropriate location NOTE Lowering is very slow Re setting manual emergency lowering raising Contact an authorized Handicare service technician Do not use the lift unit Allen key MANUAL SystemRoMedic Trouble shooting If lifting motion cannot be activated check to ensure that the emergency stop is not activated on certain models the emergency stop symbol lights up that the lift unit is switched on ON via the handset a signal is heard that the handset is properly connected and pressed in securely disconnect and then replace it and press it in securely that the lift unit is not in the charging position the lift unit is blocked for use when it is in the charging position that the battery is charged that the charger has a power supply the charging lamp on the charger unit should light up when the lift unit is placed in the charging position that load is applied to the lift strap If the lift unit is not working properly contact your Handicare dealer If the lift cannot be run all the way up t
15. st be activated via the green ON button before use Activate start the lift unit The lift is switched on via the green ON button on the hand control A battery symbol lights up for about 2 seconds to indicate battery status If the lamp does not light check to ensure that the emergency stop is not activated Active safety HiseAtlas has a built in strap guard which prevents the lift from being manoeuvred up or down unless the lift is loaded with at least 2 kg The purpose of this safety feature is to prevent the strap from being run in and out without load which may lead to shorter drops when the lift is subsequently used HiseAtlas stops automatically when the lift strap is not loaded e g when the slingbar is lowered to a bed If this happens the strap must be stretched in order to be run down or up If the lift strap becomes twisted during lifting RiseAtlas switches off auto matically and can only be run down To save battery charge the lift unit switches off automatically after about 4 minutes of inactivity 10 MANUAL SystemRoMedic Control panel On the control panel on the underside of the lift unit there are several symbols The symbols light up when a function is activated More information on each function is given below and under Trouble shooting 1 Battery status is indicated for 2 sec when a lift with end of rail charging is activated and for 4 min if the in rail chargin
16. the user should read the manual before using the lift again Normally a technician must be con tacted when this symbol is activated Preventative service maintenance The symbol for service maintenance lights up automatically when the lift unit has been in service for 12 months Contact an authorized Handicare service technician for maintenance and resetting SystemRoMedic MANUAL 11 Emergency stop To activate emergency stop Pull the red emergency stop strap firmly figure 1 and then release To reset Press the emergency stop strap s button back into place figure 2 If the emergency stop strap is held down constantly electrical emergency lowering will be When the strap is released the lift will resume the emergency stop position Electrical emergency lowering If the handset is not functioning electrical emergency lowering can be activated by holding the emergency stop strap down diagram 1 Lower the patient to a safe height and location Electrical emergency lowering may only used if the handset is malfunctioning The electrical emergency lowering device feeds out the lift strap for as long as the emergency stop strap is held down Ensure that the lowering motion is activated Otherwise electrical emergency lowering must be stopped Always ensure that emergenoy lowering is done in an appropriate location To re set Press the plastic button that holds the red strap in place upward until it clicks
17. the charger plate see figure 2 3 Check to ensure that charger rails have been installed installation of charger rails is described in the installation manual for MilkyWay 4 Guide the lift unit onto the rail so that the plus on the charging point is on the right side according to figure 3 5 Connect the red cable to the charger rail on the right side the charger plate s plus side positive and the black cable on the left side negative according to figure 4 6 Install the end stop on the rail NOTE No spring pin Different rail systems The choice of rail system is determined by lifting needs H system Straight rail curves turntable Freestanding straight rail SystemRoMedic MANUAL 3 Final inspection he HiseAtlas lifting unit and the RiseAtlas trolley must be installed and first time inspected by authorized personnel inspector in accordance with the installation instructions provided by Handicare AB Periodic inspection of the equipment should be undertaken at least once a year Original spare parts should be used and service maintenance of the equipment should be in accordance with the manual e Check to ensure no parts in package Check and enure that the lift has no signs of damage after transportation Check to ensure that the emergency stop is functioning by pulling it out and then pressing the UP or DOWN button If nothing happens the emergency stop is working pr
18. the right side according to figure 3 4 Mount charger contacts according to figure 4 NOTE Push the charger contact A in about 50 mm if only a stop bolt is used When an adjustable end stop is used the charger contact is pushed in about 90 mm 5 Affix the accompanying screws for charger contacts see figure 5 B 6 Connect the red cable to the charger contact on the right side the charger plate s plus side positive and the black cable on the left side negative according to figure 6 7 Install the end stop on the rail Spring pin 4 MANUAL SystemRoMedic In rail charging If the in rail charging option is chosen the rail system must be fitted with charger rails before installation The charger that is included with the lift unit is used It is connected to the charger rails in the same way as the charger station see section Charger station NOTE Alteration of the charging alternative must be performed by authorized personnel according to instructions issued by Handicare AB 1 Set the charging points for in rail charging the charger plate is not secured with the spring pins see figure 1 To set the charging points for in rail charging remove the spring pins that hold the charger plate in place Then the dip switch on the circuit board is set for the desired charging function 2 Check to ensure that the charger plate springs in and out when pressure is applied to
19. unit that is con nected to the charger cables via a DC outlet The charger should be kept in an appropriate location Note that the charger is a separate unit and can be easily mistaken for another charger Ensure that the correct charger is used see Technical information 1 End of rail charging For end of rail charging the lift unit is placed at the charger station at the end of the rail If the rail system is of the traverse type the secondary rail is placed at the end of the primary rail at which the char ger station is located Check to ensure that the battery symbol lights up 2 In rail charging For in rail charging the lift unit is always being charged which means that it can be left anywhere on the rail and will charge auto matically Battery status charger station The battery symbol lights green when battery capacity exceeds 5096 If the battery symbol blinks orange when the lift is started the battery needs to be charged If the battery symbol lights orange and a beep is heard when the lift is run up the battery needs to be charged immediately When the lift unit is placed in the charging position the battery symbol lights green for 4 min The freestanding charger unit also indicates charging status via a coloured lamp When the lift unit is fully charged the charger unit lamp lights green When the lift unit is activated the battery symbol lights up for two seconds to indicate the current battery status Battery statu