Home
Keysight Technologies
Contents
1. 205 g Additional Error Type of Error Impedance Proportional Error 5 Xx 7 10 f MHz Description This test fixture is designed for impedance evaluation of axial radial lead type devices The 16047A employs Kelvin con tacts which realize a wide impedance measurement range The con tact tip can be changed according to the device shape Applicable Instruments 4263B 4268A 4279A 4288A E4980A E4981A 4284A 4285A 4294A denotes the instrument is obsolete applicable in a limited frequency range Frequency DC to 13 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size See figure with module sizes Radai Lead Short Radial Lead EEIN i mm 16047A module sizes Furnished Accessories Description P N Qty Module For Axial Lead 16061 70022 Module For Radial Lead mounting on fixture 16061 70021 Module For Short Radial Lead 16047 65001 Operating Note 16047 90011 NINI BN Each module size for the 16047A is shown above Option 16047A 701 Add Shorting Plate P N 16047 00640 Compensation and Measurement Select one of these modules suitable for the DUT s shape Open and short compensations are recommended before measurement Short compensation Is performed by shorting the contacts of the test fixture with a shorting plate After performing open and short compensations the DUT is connected to the test fixture P N 16047 00640 Material Brass Ni
2. Error Ea fs ur kan Frequency Hz Typical Loss Tangent tan 6 Measurement Accuracy h In c b 10 dataTec Ferdinand Lassalle Str 52 a 72770 Reutlingen Tel 07121 51 50 50 Description The 16454A is designed for accurate permeability mea surements of toroidal shaped magnetic materials Since the con struction of this fixture creates one turn around the toroid with no magnetic flux leakage the need of winding a wire around the toroid is unnecessary The following figure shows the one turn mechanism and how complex permeability is calculated from it ee ea Sw E io dane hini iL itlive pena beili AN rnensure anpere yrdd inne narn Pn HU Ha MH cee Oe EA ot man AU dd EDD A Base HV tts hiai rin tata EET Hi pewnmspain of ep seage ho flesh MIT aterial Uder Tes C ujer diameter oF MJT To naerdiameter o MUT Permeability measurement method of 16454A Complex permeability is calculated from the inductance with and without the toroid When E4991A with option E4991A 002 is used as the measurement instrument direct readouts of complex permeabili ty are possible In addition it is furnished with a small and a large fixture to adapt to a wide range of sizes Applicable Instruments 4294A 42942A E4991A with Opt E4991A 002 4291B with Opt 4291B 002 denotes the instrument is obsolete applicable in a limited frequency range Frequency 1 kHz to 1 GHz DC
3. Fixture Can Magrifymo ens ELF insertar Habe aaler Aseerehey a Fixture overview 37 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Description This test fixture is designed for impedance evaluations of parallel electrode SMDs It achieves stable frequency characteris tics up to 3 GHz and provides highly repeatable measurements The applicable SMD size code is 0603 inch 1608 mm Applicable Instrument 4297A E4992A 4294A 42942A E5061B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A denotes the instrument is obsolete Frequency DC to 3 GHz Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 85 C DUT size The applicable SMD size is 0603 inch 1608 mm For details see the figure below DT Insertion Hole Geeneber THF Be sA Dd uan Sm OH Jd o Oe Fete OCH nsulaior Assembly The 16196A is furnished with three different insulator assemblies since any gaps between the DUT and the cylindrical insulator will result in improper positioning and subseguent measurement errors Select an insulator assembly that reduces the gap the most See the table below for dimensions of the insulator assemblies Hole Diameter of Insulator SMD case size examples assembly mm Length Width Height mm 16196A Q 1 34 1 6x0 8x0 8 0 1 14 1 6x0 8x0 6 p 1 08 1 6x0 8x0 5 Furnished Acces
4. 42 V peak max AC DC Operating Temperature 20 C to 5 C DUT Size See figure below with 16047E s electrode size Gaard Form Ef 4 See Fon View Side View Shorting Plate Furnished IA Furnished Accessories Description P N Oty 16047 01221 16047 01222 0515 1229 16047 00621 16047 90040 Angle right side Angle left side Screws Shorting Plate Operating and Service Manual Compensation and Measurement Open and short compensations are recommended before measurement Short compensation is performed by shorting the contacts of the test fixture with a shorting plate After performing open and short compensations the DUT is connected to the test fixture The following figures show how compensation and measurement are performed Shoting Petr Cala termina Ten tomb en CLIT Measuring 3 Terminal device www datatec de Up to 110 MHz 4 Terminal Pair Lead Components 16060A Transformer Test Fixture Description This test fixture provides a convenient means of mea suring a transformer s self inductance mutual inductance turns ofl ratio and DC resistance in the frequency range of DC to 100 kHz as te N appropriate for each measurement N Applicable Instruments 4263B with Option 4263B 001 Freguency DC to 100 kHz mM Operating Temperature 0 C to 55 C Q ed DUT Size The lead wire of the transformer should not have a diame a ter greater than
5. 16048A Test Leads 16048D Test Leads 16048E Test Leads 16048G Test Leads 16048H Test Leads Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax07121 51 50 10 co uy oonan A A BP BW HO WHY DY 10 11 12 13 14 15 16 17 18 18 ig 19 20 20 21 22 22 info datatec de www datatec de Table of Contents Probes 42941 A Impedance Probe Kit 23 DC Bias Accessories 16065A 200 Vdc External Voltage Bias Fixture 24 16065C 40 Vdc External Voltage Bias Adapter 24 Material 16451B Dielectric Test Fixture 25 28 16452A Liquid Dielectric Test Fixture 29 30 Up to 3 GHz mm 31 Lead Components 16092A Spring Clip Test Fixture 32 SMD 16192A Parallel Electrode SMD Test Fixture 33 34 16194A High Temperature Component Test Fixture 35 36 16196A Parallel Electrode SMD Test Fixture 37 39 16196B Parallel Electrode SMD Test Fixture 40 41 16196C Parallel Electrode SMD Test Fixture 42 43 16196D Parallel Electrode SMD Test Fixture 44 45 16197A Bottom Electrode SMD Test Fixture 46 48 DC Bias Accessories 16200B External DC Bias Adapter 49 Material 16453A Dielectric Material Test Fixture 50 16454A Magnetic Material Test Fixture 51 52 DC High Resistance 09 SMD amp Lead Components 16339A Component Test Fixture 54 SMD 16118A Tweezers Test Fixture 55 iii E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec d
6. 4284A 4288A 4338B E4980A E4981A 4285A 4294A denotes the instrument is obsolete applicable in a limited frequency range Frequency DC to 10 MHz Terminal GOEOD AL Pair BNC Maximum Voltage 40 V peak max AC DC DUT Connection 4 Terminal Operating Temperature 0 C to 55 C DUT Size See figure below Dimensions approx 160 W x 70 H x 98 D mm Weight approx 550 g Additional Error WE ei j Type of Error Impedance peer Proportional Error 2 x f 10 Open Repeatability 15 200x f 10 nS Electrode f MHz Furnished Accessories Description P N Qty Cleaning Rod 5182 7586 To maintain the measurement precision it is recommended that con tact pins be replaced approximately every 50 000 times supplemen tary value Open plate Ze plate High block Low block knob Rod locking knob Position ng rod High block knob Test fixture overview 15 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair SMD Compensation and Measurement Open and short compensations are recommended before measurement Short compensation is performed by bringing down the shorting plate which is already on the fixture to short all 4 terminals Open compensation Is performed by bringing down both the open plate and the shorting plate to separate the high terminals fr
7. O ring for Liguid Outlet 0905 1277 1 D Spacer 1 3 mm thickness 16452 00601 1 F Spacer 1 5 mm thickness 16452 00602 1 F Spacer 2 0 mm thickness 16452 00603 1 F Spacer 3 0 mm thickness 16452 00604 1 F Lid of Liquid Outlet 16452 24002 1 G SMA BNC Adapter 1250 1200 4 H Waterproof Cap for BNC Connector 1252 5821 4 Carrying Case 16452 60111 1 Operation and Service Manual 16452 90020 1 Angle Iron of Stand Body for Fixture Stand 16452 01201 2 Screw of Stand Body or Fixture Stand 0515 0914 4 Screw for Fixture Stand 0515 0914 4 Stand Foot 16452 00611 1 Electrode High and Low NA 2 A B Requires the following interface cables to connect to a measurement instru ment Select accordingly to the required temperature conditions Temperature Model or P N Cable length approx OC 50 C 16048A 0 94 m 20 C to 150 C 16048G for 4294A only Im 20 610130 E 16048H for 4294A only 2m Four BNC m to BNC m adapters P N 1250 0216 are needed to connect the 16048G H and 16452A dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 5010 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Material Compensation and Measurement Short compensation is recommended in combination with the cable length compensation before measurement First set the instrument s cable length compensation function to m Then short compensation is performed by using the furnished shorting pl
8. i Furnished Accessories Frequency Hz Description P N Gty Typical Permittivity er Measurement Accuracy Fixture Holder 16453 01213 1 thickness 1 mm Load 16453 60021 1 FE yon pdiini Porta Tweezers 8710 2081 epee ee n eee ry Carrying Case 1645320071 1 Operation and Service Manual 16453 90010 Compensation and Measurement Open short and load compensa tions are recommended before measurement Open compensation Is performed by separating the high and the low electrodes from each other Short compensation is performed by connecting the high and 1M MG low electrodes together Load compensation is performed by using en the furnished load material After performing open short and load compensations the material under test is inserted into the test fix ture Typical Loss Tangent tan Measurement Accuracy thickness 1 mm E4991A with 16453A 50 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm Material 16454A Magnetic Material Test Fixture Terminal Connector 7 mm Dimensions approx Large Test Fixture Small Test Fixture Weight approx Large Test Fixture 140 g Small Test Fixture 120g Measurement Accuracy typical 30 D x 35 H mm 24 D x 30 H mm NER Hi iM MOD TLC 1 Praguensy bz Typical Permeability ur Measurement Accuracy h In c b 10
9. ior SAG for leaded device Set the ground plate Exchanging the device holder above 500 MHz load compensation is also recommended The fix ture s electrical length must be entered into the electrical length compensation function of the measurement instrument first Then open compensation is performed by separating the high and the low electrodes from each other The separation should be eguivalent in size to the DUT s width Short compensation is performed by using the option 16194A 010 701 shorting bar set Load compensation is performed by using the furnished 50 O SMD chip resistor After per forming open short and load compensations in combination with the electrical length compensation the DUT is inserted into the test fixture The following figures show how measurement is performed ie Leaded device SHORTLISAD campinas iion Placing the device 36 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 16196A Parallel Electrode SMD Test Fixture se N ms Terminal Connector 7 mm DUT Connection 2 Terminal Electrical Length 26 2 mm Dimensions approx 140 W x 48 H x 78 D mm Weight approx 250 g Additional Error Type of Error Impedance Proportional Error 10 xio Open Repeatability 5 40xf uS Short Repeatability 30 125 x f mO f frequency GHz E4982A with 16196A
10. tor board to allow the attachment of user fabricated test fixtures Applicable Instrument 4294A Frequency DC to 110 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 20 C to 150 C Furnished Accessories Description P N Qty Mounting Plate NA Operating and Service Manual 16048 90050 1 Options 16048G 001 Add BNC Bracket P N 16048 60003 Here the BNC Bracket refers to the terminal board with four BNC m connectors Compensation and Measurement Adapter setup is recommended before measurement In the adapter setup menu select 4TP IM Then use the 100 O resistor furnished with the 4294A to perform phase compensation and load data measurement y rui Pi ome 1 instrument Side mem ro tion Ebook Gone Block Gerd Bere Ta 4 DUFF Side Cenrecgthor Black 5 Abceentigg Fre Test fixture overview Description The test leads extend the measurement port with a 4 Terminal Pair configuration It is provided with a BNC male connec tor board to allow the attachment of user fabricated test fixtures Applicable Instrument 4294A Freguency DC to 110 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 20 C to 150 C Furnished Accessories Description P N Qty Mounting Plate NA Operating and Service Manual 16048 90050 1 Options 16048H 001 Add BNC Bracket P N 16048 60003 Here the BNC Bracket refers to the terminal board with four BNC m connectors Compensatio
11. wa hd ee 1 Nas Txo aoegbrpo ds Ditton Ssrtroce Top mU Lore electrode Compatibility with ASTM D257 16008B electrode sizes Opt 16008B 001 Opt 16008B 001 002 A 6mm 26 mm B 88mm 38mm C 98mm 48mm D 110mm 110 mm Fax 07121 51 50 10 info datatec de www datatec de Other Accessories 16190B Performance Test Kit Terminal Connector 7 mm Dimensions approx 350 W x 100 H x 270 D mm Weight approx 2 0 kg co Ms dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de Description The 16190B is a performance test kit designed to verity the impedance measurement accuracy of LCR meters or impedance analyzers that have a 7 mm measurement terminal Refer to the instrument s operation service manual for the method of using these standards Applicable Instrument E4991A E4982A 4287A 4291B 4294A 42942A denotes the instrument is obsolete Furnished Accessories Description Airline 50 O 7 mm Cap Protection 50 O Termination Open Termination ohort Termination Cap Termination Protection Wrench 1 2 and 8 15 Open End Carrying Case Floppy Diskette for Calibration Data Calibration Report Operating Note P N N A 1401 0123 N A N A N A 16190 25011 8710 1770 N A N A N A 16190 90020 Qty www datatec de 61 Mn Other Accessories 16380A Standard Capacitor Set Terminal Connector 4 Te
12. 1540 0692 1000 Chip Resistor 0699 2488 10 Operating Manual 16034 90012 1 Compensation and Measurement Open and short compensations are recommended before measurement When measuring above 3 MHz load compensation is also recommended Open compensation is performed by separating the high and the low electrodes from each other The separation should be eguivalent in size to the DUT s width Short compensation is performed by placing the high and low electrodes in contact together Load compensation is performed by using the furnished 100 Q SMD chip resistor After performing open short and load compensations the DUT is inserted into the test fix ture Refer to the 16034G figures to see how compensation and measurement are performed Device holder top view Vieasurement electrode side view 0 5 AS Mi Unite im Electrode dimensions 14 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair SMD 16044A Test Fixture Description This test fixture is designed for impedance evaluations of low impedance SMD The minimum SMD size that this fixture is adapted to evaluate is 1 6 L x 0 8 W mm The 16044A has a Kelvin 4 Terminal contact which ensures repeatable measurements It is also equipped with a mechanism for easily performing open and short compensation Applicable Instruments 4263B 4268A 4279A
13. 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A E5061B 3L5 w Option 005 refer to 16034E refer to 16034E 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A refer to 16034E 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A 4263B 4284A E4980A 4294A 4294A 4263B w Option 001 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A 4263B 4268A 4288A E4981A 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A 4287A 4291B 4294A 42942A E5061B 3L5 w Option 005 16201A E4982A E4991A 4339B 4339B 4339B E4991A 4291B 4294A 42942A 4287A E4982A 4287A 4291B 4294A 42942A E5061B 3L5 w Option 005 16201A E4982A E4991A refer to 16192A refer to 16192A refer to 16192A 4287A 4291B E5061B 3L5 w Option 005 16201A E4982A E4991A 4263B 4268A 4279A 4294A 4285A 4288A 4294A E4980A E4981A 4339B 4263B 4268A 4279A 4294A 4295A 4288A 4294A E4980A E4981A 4263B 4268A 4284A 4288A 4294A E4980A E4981A 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A 4263B 4284A 4285A 4294A E4980A 4291B w Option 002 E4991A w Option 002 4291B w Option 002 4294A 42942A E4991A w Option 002 4263B 4268A 4284A 4294A 4339B E4980A E4981A 4263B 4268A 4279A 4284A 4285A 4294A 4288A E4980A E4981A
14. 4263B 4268A 4279A 4284A 4285A 4294A 4338B E4980A E4981A 4294A 4294A Page s 58 59 12 13 14 15 16 10 20 20 21 22 22 24 24 18 19 O o6 o 30 60 33 34 35 36 37 45 46 48 49 17 54 61 62 25 28 29 30 50 51 52 63 64 64 23 www datatec de Index Test Fixture Selection By SMD Size Pines 70402 la e 1608 0603 nr 01x05x05 len Sun erste pase ee ent oaos eens SMD General 160346 Parata 110MHz e e ale 20127 0005 y 3216 1208 fe 2235 1210 e SMO General MG resi noww elelelelele loixcexos sxt5sx3 Su en join Ja ele s eie s annu 08 x 08 E L 10 al ee wad Pu heme e ba AD a bee Then SMD Coaxial 161960 Pale ace Te ml SS ee ee ee ee eee et cE 001 Em oe lsisialsiaiol hunu benen ee ee i I I II IL DD IL HR EE SMO Lead 163394 Paraliet foc Jej ejc e c c c ejni 5x05 jax ox n Notes 1 LF ATP denotes test ituros hor use wath four terman el pair type LCR meters and LF enpedance anafyners in low frequency region 110 MHE d AF T mm denotes test fixtures for use with RF enpedance measurement instramenis which have 7 mm coal fost port 3 DC high R denotes test fixtunes tor use with OC high resistance meter 4 Four terrmnal Rolin contact test fet suitable For moasuring bow impedance devices 5 Thons type test forte with metes test loads 6 Test fixture for medraurng SMD and be
15. 8710 2404 Black 8710 2405 Red 56 W dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de Floating DUT Measurement Description The 16117B is designed to operate specifically with 4339B With 16117B 003 wide jaw clips it is capable of holding DUTs with large terminals The clips can be replaced with probes 16117B 001 for measurements of small DUTs such as PC boards or IC sockets 16117B 002 enables the construction of simple custom made test leads Electrical noise effects are reduced by the employ ment of shielded cables A built in interlocking circuit enables safe high voltage measurements Applicable Instruments 4339B denotes the instrument is obsolete Frequency DC Maximum Voltage 1000 V Maximum Current 0 5 mA Resistance Measurement Range 1 x 10 to 1 x 10 0 Operating Temperature 0 C to 55 C DUT Size For clips 16117B 003 lead diameter lt 30 mm For probes and sockets see figure below Jack Option 16117B G02 am EF Probe Option 164 17B O01 Probe Fip lnsaeiator Insulatcr is removakle qe lt a Else RWYT a x 75 Furnished Accessories Operation and Service Manual 16117 90060 EN Options 16117B 001 Add Pin Probes 2 ea 16117B 002 Add Soldering Sockets 2 ea 16117B 003 Add Alligator Clips 2 ea Compensation and Measurement Open compensation is recom mended before measurement Separate the test
16. Adapters are used to adapt the dedicated circuits between the instrument and the test fixtures The 42942A is a terminal conversion adapter that can convert a 4 Terminal Pair configuration to a 7 mm configuration The 16065C is an external DC bias adapter that can apply DC bias to the DUT from an external DC bias source Others Also available are DC bias accessories and performance test equipment 1 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 2 Introduction o The Benefits of Keysight Accessories Each accessory is designed to ensure highly accurate measurements without degrading the performance of the measure ment instrument Minimum residual error preserves the accuracy of the measurement instruments Clearly defined error compensation allows easy calculation of error corrections Strict measurement specifications such as test freguencies and signal levels provide safe and accurate measure ments This document introduces a group of Keysight accessories that are well suited for the following measurement instruments LCR Meters 4263B 100 Hz 120 Hz 1 kHz 10 kHz 100 kHz LCR Meter 4284A 20 Hz 1 MHz Precision LCR Meter E4980A 20 Hz 2 MHz Precision LCR Meter 4285A 75 kHz 30 MHz Precision LCR Meter 4297A 1 MHz 3 GHz RF LCR Meter E4982A 1 MHz 3 GHz LCR Meter Capacitance Meters 426
17. Applicable Instrument 4294A Freguency 40 Hz to 110 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 20 C to 5 C probe only Furnished Accessories Description P N Qty Pin Probe 42941 60002 Adapter BNC SMA 1250 2375 Spare Pin Set 3 ea 42941 60004 1 3 5mm SHORT 1250 2840 1 3 5 mm LOAD 0955 1105 1 Clip lead 8121 0003 1 Ground lead 04193 61679 1 Carrying case 42941 60011 1 Operating and Service Manual 42941 90010 1 Compensation and Measurement Adapter setup and compensation is required before measurement In the Adapter setup menu select PROBE 42941A Use the furnished 3 5 mm short and load standards The open condition can be created by not connecting the probe to anything Perform phase compensation short and load data mea surements For compensation open and short compensation is rec ommended Short compensation is performed by shorting the probe To short the probe it is recommended to use a shorting device with gold plated surfacing which provides stable contact resistance In circuit measurement Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair DC Bias Accessories 16065A 200Vdc External Voltage Bias Fixture Terminal Connector 4 Terminal Pair BNC DUT Connection 4 Terminal External Bias Input connector High Voltage BNC f Dimensions approx 180 W x 120 H x 200 D mm Cable Length approx 40 cm We
18. Bias 500 mA to 500 mA max Operating Temperature 55 C to 200 C When Option E4991A 007 temperature characteristic test kit is used with E4991A the operating temperature range is between 55 C and 150 C The temperature characteristic test kit is unavailable for the 4294A Material Size See figure below Small size Large size Material size Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm Material Furnished Accessories Description P N Qty Fixture Holder 16454 00601 Tweezers 8710 2081 Screw Hex Recess 0515 1050 Holder A 16454 25002 Holder B 16454 25001 Holder C Without hole 16454 25003 1 Holder D Without hole 16454 25004 1 EEO Holder Case 1540 0622 Hex Key for replacing fixtures 8710 1181 1 Carrying Case 16454 60101 Operation and Service Manual 16454 90020 Compensation and Measurement Short compensation is required before measurement Short compensation is performed by only inserting the MUT holder into the test fixture After performing short compensation the MUT is inserted into the fixture as shown below Short Compensation sy ae Remove the cap of the fixture Place aMUT holder only in the fixture Replace the cap by screwing tightly Placing the MUT on the Test Fixture as follows Remove the cap of the fixture Place aMUT onto the MUT holder and insert it Replace the cap by screwing tightly into the fixture 52 a m dataTec Ferd
19. He nn Pe Oe ml i Erpa Sone Cen Erriet r p i F i il i i rr i i i j i i i i rr rr rrr rm yy yny rien Relationship between proportional error short offset error and freguency when measuring low impedance dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 5010 info datatec de www datatec de Appendix Additional Error 3 2 DUT Connection Configuration In order to make short repeatability small there are test fixtures that use the 4T connection configuration for example Keysight 16044A By employing this technigue the effect of contact resistance is reduced and short repeatability is significantly improved As a result the range of accurate low impedance measurements is expanded down to a low mil liohm region Figure below shows the difference between the 2T connection and the 4T connection In a 2T connection the contact resistance that exists between the fixture s contact electrodes and the DUT is measured together with the DUT s impedance Contact resistance cannot be eliminated by compensation because the contact resistance value changes each time the DUT is contacted d Poel Kteus iter Pixiure ANSEHEN TEN YErasperemenl Fax Lare lasteysment tr Bnainumneni F DUL E 2 Ferminal Connection 4 Terminai Connection 2 Terminal and 4 Terminal connector techniques In a 4T connection the v
20. Surface Resistivity Measurement Range up to 4 0 x 107 O Operating Temperature 30 C to 100 C excluding selector switch Material Size Select an electrode so that outer diameter of guard ing electrode is smaller than the DUT s diameter See figure and table below for more details D1 Main Electrode 26 mm 50 mm 6 mm Outer Diameter of Guard Electrode 2 mm D2 Guard Electrode Inner Diameter 38mm 70 mm 88 mm Thickness 10um to 10 mm dataTec 58 Min D3 Guard Electrode Inner Diameter 48mm 80 mm 98 mm Ordering information supplied with 16008B 001 002 standard equipped Supplied with 16008B 001 Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de D Material Size 50 mm to 125 mm 82 mm to 125 mm 100 mm to 125 mm www datatec de DC High resistance Material Note The main electrodes are orderable as a part by using following part numbers Replace the main electrode when the conductive elas tomer on the electrode comes off 26 mm electrode 16008 60083 50 mm electrode 16008 60081 76 mm electrode 16008 60085 ASTM D257 s recommended electrode sizes Choice 1 Choice 2 A 7 6 mm 25 mm B 99 mm 38mm G 100 mm 50 mm D 100 mm 50 mm 59 W dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Furnished Accessories Description P N Oty 16008 1033 1 16008 90011 1 Acryli
21. circuit to malfunction or create an unexpected measurement error Such cir cuit malfunctions and measurement errors can be avoided with cable length compensation Cable length compensation should therefore be performed prior to the open short compensation When the following measurement instruments and test leads are used in combination the undesired effects described above can be eliminated by performing the cable length compensation the values in the table represent cable length settings Model 16048A 16048D 16048E 16048G 16048H 4263B Im 2m 4m 4269A 1m 2m 4279A 1m 2m 4284A 1m 2 m 4 m 4285A 1m 2m 4299A 1m 2m 4294A 1m 2 m E4980A 1m 2m 4m E4981A 1m 2m N A 1 Available for 4284A with Option 006 denotes the instrument is obsolete Measurement Repeatability Dirty electrodes on the test fixture cause an increase in the contact resistance This increase in resistance can cause poor measurement repeatability Therefore the electrode of the test fixture must be kept clean when measurements are per formed 73 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 7 W Index Model Number 16008B 16034E 16034G 16034H 16044A 16047A 16047E 16048A 16048D 16048E 16048G 16048H 16060A 16065A 16065C 16089A B C D 16092A 16117B 16117C 16118A 16190B 16192A 16194A 1
22. clips or probes from each other After performing open compensation the DUT is connected to the clips or probes The DUT can be measured in two configurations floating and grounded The connections are shown in the figure below ma 1 H E T R DAU OTT 1 Grounded DUT Measurement Floating and grounded DUT measurement www datatec de DC High resistance Other Components 1611 C Low Noise Test Leads Description The 1611 C is designed to operate specifically with 4339B It comes with a separate triaxial special screw type female connector and a high voltage BNC special type female connector so that the terminal configuration of the 4339B can be converted to any other configuration Therefore custom made test fixtures can easily be constructed Applicable Instruments 4339B denotes the instrument is obsolete Frequency DC Maximum Voltage 1000 V Maximum Current 10 mA Resistance Measurement Range 1 x 10 to 1 x 101 Q Operating Temperature 0 C to 55 C Furnished Accessories Terminal Connector Type Connector Input Triaxial special screw type Description P N Qty Genel ec cU High Voltage BNC special type female connector 1250 2317 1 Operation and Service Manual 16117 90060 1 Interlock connector enables and disables the application of source voltage from the measurement instrument DUT Connection 2 Terminal with triaxial cable Cable Length approx 1 m co
23. in combination with the elec tr cal length compensation the DUT is inserted into the device inser tion hole Once the measurement of the DUT is complete remove the DUT from the fixture The following figures show how compensation and measurement is performed Short Condition Short Bar DUT Measurement Compensation and measurement 48 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm DC Bias Accessories 16200B External DC Bias Adapter Description This test fixture is designed to measure a DUT with DC bias By connecting an external DC current source to the 16200B it can supply a bias current across the DUT of up to 5 Adc through a j 7 mm port 3 Applicable Instruments 4291B 4294A 42942A 4287A E4982A ll E5061B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4279A 4284A 4285A 4288A E4980A en H E4981A denotes the instrument is obsolete applicable in a limited frequency range Frequency 1 MHz to 1 GHz Terminal Connector 7 mm DC Bias Up to BA 40 V Input a ne a Operating Temperature 0 C to 55 C oltage ivionitor connector a i x x Dimensions approx 170 W x 70 H x 130 D mm Applicable Fixtures 16191A 16192A 16193A 16194 A 16196A B C D 16197A 16092A 16093A B Weight approx 900 g i denotes the instrument is obsolete Furni
24. mA Resistance Measurement Range 1 x 10 to 2x 10 Q Operating Temperature 0 C to 55 C DUT Size For Alligator Clip and Flat Table lead diameter lt 5 mm Terminal Connector Type Connector Input Triaxial special screw type Be Output High Voltage BNC special type LOG ime Ay AM AUU T me _ Z u Dos 11112 Fae Grmi Di Interlock Connector enables and disables Ja the application of source voltage from the SMD module measurement instrument Furnished Accessories l aor Description P N Qty DUT Connection 2 Terminal with triaxial cable Dimensions approx 200 W x 140 H x 230 D mm Axial Lead Module 16339 60102 1 Cable Length approx 0 8 m SMD Module 16339 60101 1 Weight approx 2200 9 Flat Table 16339 6004 1 Miniature Banana Plug cable 16339 61621 2 u sfoooggoudd 100 KO Output Resistor 16339 61001 1 ATT 1 MO Output Resistor 16339 61002 1 i ii e 10 MQ Output Resistor 16339 61003 1 UN 100 MO Output Resistor 16339 61004 1 Alligator Clip 8710 1984 2 a Carrying Case 16339 600201 1 4339B with 16339A Operation and Service Manual 16339 90010 1 Compensation and Measurement Open compensation is recom mended before measurement When using the axial lead module have no DUT connected to the electrodes When using the SMD module separate the high and low electrodes from each other The separation should be equivalent in size to the DUT s width When using the alligator clip and flat tabl
25. simple and reliable For example a mechanically and electrically precise test fixture is required to measure the impedance of SMD components For this measurement Keysight offers dedicated SMD fixtures for impedance measurement instruments that minimize the measurement errors Also specially designed fixtures for other specific applications such as DC bias test dielectric material test and others are available Keysight accessories facilitate a shorter time to market with increased confidence by providing accurate and repeatable measurements 2 Types of Accessories Keysight accessories can be divided into the following five categories Test Fixtures A test fixture is used to hold the electronic components or materials physically and electrically for the measurements Keysight offers vari ous kinds of 4 Terminal Pair test fixtures and 7 mm test fixtures Some of them connect directly to the measurement instrument while oth ers require adapters Test Leads Test leads are used to extend the measurement ports from the UNKNOWN terminals of the instrument to the DUT Using a flexible test lead a DUT that cannot be held with test fixtures can be measured regardless of its size or shape The test leads can also be used as cable extensions when the test sample is located away from the measurement instrument Probes Probes are helpful in measuring components which are already connected to PC boards or have one terminal grounded Adapters
26. than 100 In essence short repeatability is made up of a resistance and an inductance part which become larger as the freguency becomes higher Definition of short repeatability 66 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Additional Error 67 Mn 2 3 Open Offset Error The term Yo x Zx x 100 is called open offset error If Zx is multiplied to this term then AY Yo This term affects the absolute admittance error by adding an offset Open repeatability Yo is determined from the variations in multiple admittance measurements of the test fixture in open condition After performing open compensation the measured val ues of the open condition will distribute around 0 S in the complex admittance plane As shown in the figure below the maximum value of the admittance vector in the complex admittance plane is defined as open repeatability The larger open repeatability is the more difficult it is to measure large impedance values Open repeatability is made up of a stray conductance and stray capacitance part which become larger as the freguency becomes higher Definition of open repeatability 3 New Market Trends and the Additional Error for Test Fixtures 3 1 New Devices Recently the debut of extremely low ESR capacitors and the trend to use capacitors at much higher freguencies have increased demand for low impedanc
27. tion and measurement are performed anni ee ern PS pane a n ar ny i UR H CHER r HL amp ae CE MD a oe ee y RY Fn anw HIKE TA TRL ray Device holder lion view Measurernernt e ecirode side view Units pem Dimensions www datatec de Up to 110 MHZ 4 Terminal Pair SMD 16034H Test Fixture Description This test fixture is designed for impedance evaluations of array type SMD The minimum SMD size that this fixture is adapted to evaluate is 1 6 L x 0 8 W mm Since the tip of the measurement electrodes are very thin and the device holder is extremely flat the device can be shifted and the measurement electrodes can contact the each elements of the array type component Applicable Instruments 4263B 4269A 4279A 4284A 4285A 4288A 4294A E4980A E4981A E5061B 3L5 with Opt E5061B 005 denotes the instrument is obsolete Terminal Connector 4 Terminal Pair BNC Frequency DC to 110 MHz DUT Connection 2 Terminal Maximum Voltage 42 V peak max AC DC Dimensions approx Operating Temperature 0 C to 55 C 120 W x 50 H x 70 D mm DUT Size See figure below Weight approx 200 g Additional Error Type of Error Impedance W 15 0mm J Proportional Error 0 5 x 10 06 roa a rcl je Open Repeatability 5 500 x f 10 nS Short Repeatablity 10 13 x f 10 mQ OO Furnished Accessories rea Description P N Qty Case for 100 SMD Resistance
28. to see how compensation and measure ment is performed 45 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 16197A Bottom Electrode SMD Test Fixture Terminal Connector 7 mm DUT Connection 2 Terminal Electrical Length 14 mm Dimensions approx 160 W x 70 H x 86 D mm Weight approx 300 g Additional Error Type of Error Proportional Error Open Repeatability Short Repeatability f frequency GHz E4991A with 16197A Frosture und Pressure adjusting nut Ecirode uni Description This test fixture is designed for impedance evaluations of bottom electrode SMDs It achieves stable frequency characteris tics up to 3 GHz and provides highly repeatable measurements This test fixture supports various SMD sizes as small as 1005 mm 0402 inch and as large as 3225 mm 1210 inch Accommodation of the 0603 mm 0201 inch size is available with option 001 Applicable Instrument 4287A E4982A 4294A 42942A E5061B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A denotes the instrument is obsolete Frequency DC to 3 GHz Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 85 C DUT size See figure and table below Test fixture overview 46 a dataTec Ferdinan
29. 07 1 Electrode Plate 16197 00604 1 16197A 001 Shorting Bar Size P N Qty 0 6 x 0 3 x 0 3 mm 16197 29001 4 Options 16197A 001 Add 0603 mm 0201 inch Device Guide Set The 16197As electrode spaces are 0 5 mm 1 mm 1 5 mm and 2 mm and the device guide matches these spaces with appropriate inser tion holes for the applicable SMD The 0 2 mm electrode spacing Is available with option 001 which includes a device guide with 0603 mm 0201inch insertion holes an electrode plate and 4 shorting devices u LY S Y Sor 7 Electrodes configuration and SMD size Compensation and Measurement First of all select the appropriate device insertion hole If the device insertion hole is not positioned in the pressure arm s contact range reposition the device guide and the electrode plate Once this is prepared perform compensation Open and short compensations are recommended in combination with the electrical length compensation before measurement The fixture s electrical length must be entered into the electrical length Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD compensation function of the measurement instrument first Next open compensation is performed by not placing anything in the device insertion hole Short compensation is performed by placing the furnished shorting device in the device insertion hole After per forming open and short compensations
30. 10 info datatec de www datatec de 29 Mn Up to 110 MHz 4 Terminal Pair Material 16452A Liquid Dielectric Test Fixture Terminal Connector 4 Terminal Pair SMA Dimensions approx 85 H x 85 W x 37 D mm Weight approx 1400 g Measurement Accuracy A B C Error A x ag J i Error B Frequency Hz Error C Measurement Error of Instrument LCR meter with 16452A Description This test fixture provides accurate dielectric constant and impedance measurements of liquid materials The 16452A employs the parallel plate method which sandwiches the liquid material between two electrodes to form a capacitor A LCR meter or an impedance analyzer is then used to measure the capacitance created from the fixture Applicable Instruments 4263B 4284A 4285A E4980A 4294A denotes the instrument is obsolete applicable in a limited frequency range Frequency 20 Hz to 30 MHz Operating Temperature 20 C to 125 C Maximum Voltage 30 Vrms Material Capacity Required sample liquid capacity depends on the gap of the electrodes Gap of electrodes 0 3 mm 0 5 mm 1 mm 2 mm Air capacitance 34 9 pF 21 2 pF 10 9 pF 0 9 PF 25 15 10 10 sample liquid capacity 3 4 ml 3 8 ml 4 9 ml 6 8 ml Applicable frequency 20 Hz 30 MHz 1 f 5 ad YT i C F F Furnished Accessories Description P N Oty Shorting Plate 16092 08010 1 E
31. 16334A 5 15M 16089A B C D 5 100 k 16048A D 30M 16048E 2M 16048G H 110M 42941A 40 110M 16065A 50 2M 16065C 100 1M 16451B 30M 16452A 20 30M 16092A De a F 16192A 2 16194A 2G 16196A B C D I l 3G 16197A i 3G 16200B 1M 1G 16453A 1M 1G 16454A 1k 1G 16339A DC For 4339A B only 16118A DC For 4339A B only 16117B C DC For 4339A B only 16008B DC For 4339A B only a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Accessories Catalogue Accessories Organization This document is organized by measurement freguency and DUT to enable guick selection of an appropriate test fixture for a particular measurement application The following tables show the various categories in each primary group Up to 110 MHz Terminal Configuration 4 Terminal Pair Lead components 16047A E 16060A SMD components 16034E G H 16044A 16334A Other components 16089A B C D Port Cable extension 16048A D E G H DC bias accessories 16065A C Material 16451B 16452A Up to 3 GHz Terminal Configuration 7 mm Connector Lead components 16092A 16194A SMD components 16092A 16192A 16194A 16196A B C D 16197A DC bias accessories 16200B Material 16453A 16454A DC High Resistance Measurement Lead components 16339A SMD components 16118A 16339A Other componen
32. 3 GHz Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 85 C DUT size The applicable SMD size is 01005 inch 0402 mm For details see the figure below The 16196D is furnished with two different insulator assemblies since any gaps between the DUT and the cylindrical insulator will result in improper positioning and subsequent measurement errors Select an insulator assembly that reduces the gap the most See the DJT mention Hole nanietes Insulator Asser EF table below for dimensions of the insulator assemblies Hole Diameter of Insulator assembly mm SMD case size examples Length Width Height mm 16196D 0 34 0 4x0 2x0 2 0 30 0 4 x 0 2 x 0 13 0 2 Furnished Accessories Description P N Qty 16196 90040 16196 60412 16196 60414 Operation and Service Manual Insulator Assembly 0 34 mm Insulator Assembly 0 30 mm 1 1 1 Open Plate 16196 29002 Short Plate 16196 29030 1 Push Ring 16196 24004 1 Magnifying Lens 16193 60002 1 Tweezers 8710 2081 Wrench 8710 0909 1 Cleaning Rod 5182 7586 Carrying Case 16196 60450 1 Opt 16196D 710 only 2 16196 15101 as replacement part number www datatec de Up to 3 GHz 7 mm SMD Options 16196D 710 Add the magnifying lens and tweezers To maintain adeguate measurement performance keep the elec trodes and the short plate in good condition Contaminants and abrasion on these parts considerably affect measurement re
33. 4 mm otherwise the alligator clip will not be able to clamp onto it properly Terminal Connector 4 Terminal Pair BNC Furnished Accessories DUT Connection Description P N Oty 2 Terminal for L measurement l l ID Test Leads black Alligator clip to BNC m 16060 61601 2 See figure below for more information Test Leads red Alligator clip to BNC m 16060 61602 2 Operation and Service Manual 16060 90000 1 130504 Compensation and Measurement Open compensation is recommended before measurement Open compensation is performed by connecting the alligator clips of A and B terminals together n i l and separating them from the likewise connected alligator clips of the COMMON terminals After performing open compensation the transformer is connected to the test fixture The A and B Dimensions approx terminals are connected to the high terminals of the transformer u LLU The COMMON terminals are connected to the low terminals of the LU i transformer The following figures show how compensation and Weight approx 300 g f d Additional Error The additional error is negligible when compared to the MSASULEIDENE ale penance instrument s accuracy T Be pe Short here r SS J l gt tr i s Fa separate hem ib en nn i Short here Open compensation Connecting a transformer 11 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121
34. 4288A 4294A 4294A with 42942A more reliable and repeatable measurement E5061B 3L5 w Option E5061B 005 and 16201A E4980A E4981A E4991A E4982A You will improve your measurement results with the proper test dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de 2 E MESSBAR MEHR a a lhr Spezialist fur m ale Mess und Pr fger te d a ta e C KEYSIGHT TECHNOLOGIES Autorisierter Technologie Partner Keysight Web Resources LCR Meters Impedance Analyzers Accessories for Impedance Measurements RF and microwave test accessories Anderungen und Irrtiimer vorbehalten dataTec 30 10 2014 Update Keysight Technologies August 4 2014 5965 4792E E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de
35. 42941A 42942A SMD chip test fixture SMD chip test fixture small SMD chip test fixture general SMD chip test fixture Kelvin contacts 10 MHz Axial and radial test fixture 4 terminal Axial and radial test fixture 110MHz One meter test leads BNC Two meter test leads BNC Four meter test leads BNC One meter test leads BNC 110 MHz Two meter test leads BNC 110 MHz Transformer test fixture Ext voltage bias with safety cover lt 200 vdc External bias adapter lt 40 vdc Kelvin clip leads RF spring clip axial radial and SMD Parallel electrode SMD test fixture High temperature component test fixture Parallel electrode SMD test fixture Bottom electrode SMD test fixture External DC bias adapter SMD chip tweezer Dielectric material test fixture Liquid test fixture Dielectric material test fixture Magnetic material test fixture Impedance probe kit Four terminal pair to mm adapter 4263B 4268A 4279A 4284A 4285A 4287A DC 40 MHz DC 110 MHz DC 110 MHz DC 10 MHz DC 13 MHz DC 110 MHz DC 30 MHz DC 30 MHz DC 1 MHz DC 110 MHz DC 110 MHz DC 100 kHz 50 Hz 2 MHz 100 Hz 1 MHz 5 Hz 100 kHz DC 500 MHz a DC 2 GHz a DC 2 GHz a DC 3 GHz A DC 3 GHz H 1 MHz 1 GHz j 9Hz 15 MHz DC 30 MHz 20 30 MHz 1 MHz 1 GHz 1 kHz 1 GHz 40Hz 110 MHz 40Hz 110 MHz Note Refer to the accessory descriptions for frequency and operational limits 1 3 5 mm M to 7 mm adapter is required 76 W
36. 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair SMD 16034E Test Fixture Description This test fixture is designed for impedance evaluations of SMD The minimum SMD size that this fixture is adapted to evalu ate is 1 6 L x 0 8 W mm Applicable Instruments 4263B 4269A 4279A 4284A 4285A 4299A E4980A E4991A E5061B 3L5 with Opt E5061B 005 4294A denotes the instrument is obsolete applicable in a limited frequency range Frequency DC to 40 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size See figure below Terminal Connector 4 Terminal Pair BNC DUT Connection 2 Terminal Dimensions approx 128 W x 60 H x 71 D mm u Weight approx 270 g Additional Error Type of Error Impedance ER Proportional Error 1 5 x f 10 2 F MHz Furnished Accessories Description P N Oty Operating Manual 16034 90041 1 Compensation and Measurement Open and short compensations are recommended before measurement Open compensation Is per formed by separating the high and low electrodes from each other The separation should be eguivalent in size to the DUT s width Short compensation is performed by contacting the high and low electrodes together After performing open and short compensa tions the DUT is inserted into the test fixture The following figures show how compensation and measurement are performed Pan the BLP w
37. 6196A B C D 16197A 162008 16334A 16339A 16380A 16380C 16451B 16452A 16453A 16454A 42030A 42090A 42091A 42941A 42942A dataTec Name Resistivity Cell SMD Chip Test Fixture SMD Chip Test Fixture Small SMD Chip Test Fixture General SMD Chip Test Fixture Four Terminal 10 MHz Axial and Radial Test Fixture Axial and Radial Test Fixture 110 MHZ One Meter Test Leads BNC Two Meter Test Leads BNC Four Meter Test Leads BNC One Meter Test Leads BNC 110 MHz Two Meter Test Leads BNC 110 MHz Transformer Test Fixture Ext Voltage Bias with Safety Cover lt 200 vdc External Bias Adapter lt 40 vdc Kelvin Clip Leads RF Spring Clip Axial Radial and SMD Low Noise Test Leads Low Noise Test Leads SMD Chip Tweezers Performance Test Kit Parallel Electrode SMD Test Fixture High Temperature Component Test Fixture Parallel Electrode SMD Test Fixture Bottom Electrode SMD Test Fixture External DC Bias Adapter SMD Chip Tweezers Component Test Fixture for 4339A B C Standards Set C Standards Set Dielectric Material Test Fixture Liquid Test Fixture Dielectric Material Test Fixture Magnetic Material Test Fixture Four Terminal Pair Standard Resistor Set Open Termination Short Termination Impedance Probe Kit Four Terminal Pair to 7 mm Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de Applicable Measurement Instrument s 4339B
38. 7121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm Freguency Range 1G Hz 2G Hz 3G Hz 16092A 16192A 16193A 16194A 16196A B C D 16197A 16200B 16453A 16454A Applicable Instrument Up to 3 GHz Terminal Configuration 7 mm Freguency Up to 110 MHz Range Terminal Configuration 4 Terminal Pair Measurement 4263B 4268A 4279A 4284A 4291B 4294A 42942A E5061B 3L5 w Opt 005 16201A Instruments 4285A 4288A 4294A E4980A E4981A 4297A E4982A E4991A Any 4TP instruments excluding 4294A 16085B Freguency DC High Resistance Measurement Measurement 4339B 4349B Instruments 31 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm Lead Components 16092A Spring Clip Fixture Description This test fixture is designed for impedance evaluation of both lead and SMD It is furnished with two modules that can be readily screwed onto the plate to measure either lead or SMD Applicable Instrument 4294A 42942A 4291B 4287A E4982A E5051B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A denotes the instrument is obsolete applicable in a limited frequency range Frequency DC to 500 MHz Maximum Voltage 42 V peak max AC DC Terminal Connecto
39. 8A 120 Hz 1 kHz Capacitance Meter 4279A 1 MHz C V Meter 4288A 1 kHz 1 MHz Capacitance Meter E4981A Capacitance Meter Resistance Meters 4339B DC High Resistance Meter 4349B DC 4 ch High Resistance Meter Impedance Analyzers 4291B 1 MHz 1 8 GHz RF Impedance Material Analyzer 4294A 40 Hz 110 MHz Precision Impedance Analyzer E4991A 1 MHz 3 GHz RF Impedance Material Analyzer Network Analyzer E5061B 3L5 LF RF Network Analyzer denotes the instrument is obsolete 4 ISO 9000 Quality Management ISO 9000 is a set of international standards for guality management and guality assurance These standards were devel oped with the goal of documenting and implementing effective quality systems within companies ISO standards are con sistent with Keysight s quality system in fact the standards within Keysight Technologies Quality Maturity System QMS exceed the intent of ISO 9000 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Tips for Selecting Appropriate Accessories The following topics comprise a helpful guideline for selecting an appropriate accessory for the measurement instrument to be used 1 Selection by Measurement Application Keysight accessories can be used in a wide variety of measurement applications These applications range from basic mea surements such as impedance measur
40. D Cp G A284A 5 same as measurement parameters 4285A i a same as measurement parameters 4287A y 7 4299A Cp D Cp Q Cp G Cp Rp Cs Q Cs Rs 4291B Rs Ls 4294A a Rs Ls 4338B 4339B 4349B E4980A 5 n same as measurement parameters E4981A a Cp D Cp O Cp G Cp Rp Cs D Cs Q Cs Rs E4982A E4991A N A Available denotes the instrument is obsolete Note For more details of the compensation functions refer to the instruction manuals supplied with the measurement instrument 71 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Compensation 3 Electrical Length Compensation In a single port Two Terminal impedance measurement at higher freguencies RF region or higher the wavelengths of the electrical signal are so short that the length of the signal transmission line including the test fixture and cables can cause an undesirable phase shift error The phase shift error is corrected by the electrical length compensation which should be performed in combination with the open short compensation UU parent T EL IR Fest Mixture 7 i i sarplined ir CYFAN BETH Teg Fixing cl RR sed i SEELEN a yn HE a RO mAn Electric Length a TEHE i TMM connector w THI EN ET eect 5 4 Electrical length of a test fixture Residual parameters in the meas
41. Q f frequency GHz Description This test fixture is designed for impedance evaluations of parallel electrode SMDs It achieves stable frequency characteris tics up to 3 GHz and provides highly repeatable measurements The applicable SMD size code is 0201 inch 0603 mm Applicable Instrument 4297A E4992A 4294A 42942A E5061B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981 A denotes the instrument is obsolete Frequency DC to 3 GHz Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 85 C DUT size The applicable SMD size is 0201 inch 0603 mm For details see the figure below The 16196C is furnished with one insulator assembly See the table below for the dimensions of the insulator assembly BUT zseton Hole Diameler Sur D TE h Dwy TL Gb O27 iL Tau Electrode Insuiaior Assembly Hole Diameter of Insulator SMD case size examples assembly mm Length Width Height mm 16196C 00 48 05x03 x05 Furnished Accessories Description P N Qty Operation and Service Manual 16196 90040 Insulator Assembly 0 48 mm 16196 60312 Open Plate 16196 29002 Short Plate 16196 29028 Push Ring 16196 24004 Magnifying Lens 16193 60002 1 Tweezers 8710 2081 1 Wrench 8710 0909 Cleaning Rod 5182 586 Carrying Case 16196 60350 Options 1 Opt 16196C 710 only 16196C 710 Add the magn
42. Sure darela kl 7 aln Hwlredi GLa Peel ode Electrode sizes and applicable material sizes Description The 16008B is designed to operate specifically with 4339B It is used to measure surface or volume resistance resistivi ty of insulation materials The following figures show the block dia grams of resistivity measurements Ungar ded Eiactrode u ba i Fi mu ERBEN ER i A i iamurded T ertrodu i inpuuzdeg Fleet ode Wn Guarded E lentrode Volume resistivity measurement Surface resistivity measurement Volume resistivity is measured across the material MUT by the unguarded and guarded electrodes Surface resistivity on the other hand is measured along the surface of the MUT between the guard ed and the unguarded electrodes Notice that the role of the guard and unguarded electrodes switch when measuring volume and sur face resistivity 16008B is provided with three different electrode sizes to meet size requirements by different measurement stan dards The guarded electrode eliminates measurement errors due to the edge effect and arbitrary contact pressure can be applied to the material under test It is also equipped with a high voltage protec tion cover to shut off power when opened Applicable Instruments 4339B denotes the instrument is obsolete Frequency DC Maximum Voltage 1000 V Maximum Current 10 mA Resistance Measurement Range Volume Resistivity Measurement Range up to 4 0 x 1018 Ocm
43. a E MESSBAR MEHR a a lhr Spezialist fur m ale Mess und Pr fger te d d ta e KEYSIGHT TECHNOLOGIES Autorisierter Technologie Partner Keysight Technologies Accessories Selection Guide For Impedance Measurements Selection Guide a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 515050 Fax 07121 51 50 10 info datatec de www datatec de Table of Contents Introduction 1 2 d 4 What are Keysight Accessories Types of Accessories The Benefits of Keysight Accessories ISO 9000 Quality Management Tips for Selecting Appropriate Accessories 1 u PF ee a 7 Selection by Measurement Application Compatibility with Measurement Instruments Frequency DC Bias and Operating Temperature Humidity DUT Device Under Test Dimensions Open amp Short Repeatability and Proportional Error Furnished Accessories Terminal Adapters Accessory Catalogue Applicable Frequency Ranges Accessories Organization Up to 110 MHz 4 Terminal Pair Lead Components 16047A Test Fixture 16047E Test Fixture 16060A Transformer Test Fixture SMD 16034E Test Fixture 16034G Test Fixture 16034H Test Fixture 16044A Test Fixture 16334A Tweezers Contact Test Fixture Other Components Varying in Size or Shape 16089A Large Kelvin Clip Leads 16089B Medium Kelvin Clip Leads 16089C Kelvin IC Clip Leads 16089D Kelvin Alligator Clip Leads Port Cable Extension ii gataTec Ferdinand
44. aceability In an actual measurement there can be an extension of the measurement port with a cable or an adapter conversion to match the test fixture s terminal configuration For this reason cables and conversion adapters are provided for connec tivity with the measurement port These cables and adapters are designed to maintain high accuracy of the measure ment instrument while extending the measurement port Most of the time the measurement accuracy of the instru ment and the cable or adapter are specified together as a whole A test fixture Is an accessory used to connect the DUT to the measurement instrument Many test fixtures are provided to adapt to various shapes and sizes of DUTs A test fixture is either connected directly to the measurement port of the instrument or to the port of the extension cable or conversion adapter as described earlier The test fixture s structure determines the applicable freguency and impedance ranges Hence it is necessary to use the appropriate test fixture for the desired measurement conditions In addition each test fixture has its own inherent characteristic error which is detailed in its operational manual 2 Measurement System Accuracy The eguation for the accuracy of a measurement system is Measurement accuracy Instrument s accuracy Test fixture s error The measurement instrument s accuracy is determined by an eguation with terms that are dependent on freguency measured imped
45. aded components 7 Check marks in the table denote the applicable OUT sures E Applicable to SAD components with thickness 9 65 mm Mol suitable for smaller SMO than IEIS type 9 Minimum SMD sare i nol specihed Mot recommended for smaller SMO than 1668 type because contect repeautnbility degrades wilh the reduction of drrysge size L Parallel electrode Bottom electrode 75 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Index Accessories vs Instruments Matrix Simplify and Improve Your Measurements with Keysight s Test Accessories Selecting a test fixture is as important as selecting the right instru ment Keysight offers a wide range of accessories for axial radial and SMD Chip devices In addition a variety of test leads are avail able to simplify remote testing and systems applications External test fixtures with safety covers are also available For additional product information and literature visit our fixture higher through put fewer handling errors e tighter test limits better measurement accuracy Accessories Web site www keysight com find impedance Test accessories fixtures 16034E 16034G 16034H 16044A 16047A 16047E 16048A 16048D 16048E 160486 16048H 16060A 16065A 16065C 16089A B C D 16092A 16192A 16194A 16196A B C D 16197A 16200B 16334A 16451B 16452A 16453A 16454A
46. ance signal level and measurement time mode By substituting the respective measurement condi tions into the eguation the measurement accuracy is calculated If a cable or a conversion adapter is used then the specified measurement accuracy Is the accuracy of the measurement instrument with the cable or adapter This com bined measurement accuracy Is shown in the instrument s operational manual 65 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Additional Error The eguation for the test fixture s additional error is shown below Ze A Zs Zx Yo Zx x 100 96 De Ze 100 D lt 0 1 Ze Additional Error for lmpedance De Additional Error for Dissipation Factor A Test Fixture s Proportional Error 20 Zs Zx x 100 Short Offset Error Yo Zxx 100 Open Offset Error Zs Test Fixture s short Repeatability O Yo Test Fixture s open Repeatability S Zx Measured Impedance Value of DUT O D Measured D value Proportional error open and short repeatability are mentioned in the test fixture s operational manual and in this acces sory guide By inputting the measurement impedance and frequency proportional error open and short repeatability are usually a function of frequency into the above equation the fixture s additional error can be calculated 2 1 Proportional Error The term proportional err
47. and an operation and service manual 4 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Tips for Selecting Appropriate Accessories 7 Terminal Adapters Terminal Adapters convert the instruments terminal configuration into a 7 mm terminal configuration This means that instruments that do not have a 7 mm terminal connector can use test fixtures with a 7 mm terminal connector The 42942A converts a 4 Terminal Pair configuration into a 7 mm terminal connector which can only be used with the 4294A 42942A Terminal Adapter Applicable Instrument 4294A Frequency 40 Hz to 110 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 40 C Furnished Accessories Description P N Oty Carrying Case 42942 60011 1 Operation and Service Manual 42942 90020 1 Dimensions approx 190 W x 55 H x 140 D mm Options Weight approx 800 g 42942A 700 Add 7mm open short load set Description P N Qty Open Termination 04191 85302 1 Short Termination 04191 85300 1 Load Termination 04291 60043 1 5 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Accessories Catalogue Applicable Freguency Ranges Freguency Range DC 1k 1M 10M 100M 16 Hz 2G Hz 3G Hz 16047A 13M 16047E 110M 16060A 100 k 16034E 40M 16034G H 110M 16044A 10M
48. aptability for a particular measurement will be discussed To see whether a test fixture is adaptable it is important to think about the test fixture s additional error proportional error short repeatability and open repeatability measurement impedance and the test freguency range If the measurement impedance is in the 1 Q to 10 kQ range use only proportional error to calculate the additional error of the test fixture It is fine to assume that this is a close approximation to the fixture s additional error If the measurement impedance is not in this range use proportional error short repeatability and open repeatability to calculate the test fixture s additional error Recent test fixtures have all three terms specified in their operational manu al so use these values for the calculation Some of the recent test fixtures 16044A due to their structure have different performance characteristics with differ ent measurement instruments For these test fixtures refer to their operational manual for more details about the spec ifications If the test fixture is not specified with short and open repeatability how can the test fixture s adaptability be deter mined To measure a test fixture s short repeatability measure the impedance of the short condition after performing short compensation Take the shorting plate out of the fixture and then replace it Measure the short condition again By repeating this process at least 50 times i
49. ate Open com pensation is not performed but its values are used in the dielectric constant eguation as shown below Cp 1 Co Tw CoRp a Correction coefficient e Relative dielectric constant Cp Liquid capacitance measurement data Co Air capacitance measurement data or open compensation data a Eecwode Nepiare Canal Rp Equivalent parallel resistance measurement data Fe 54 Co 17 Ni 29 w Angular frequency w rrf Insulator Alumina AI 0 O ring Viton Fluro rubber The following figures below show how compensation and measurement Is Test fixture overview performed elo prevent bunDies fort artery gour tne feud irrto the intet of the Bxtuire lc Kiectroce until roaches tie dad awal marken hy ihe cotted ine Tha aui level here ihe eine Pa wang Ie i dee e gt i is completely fu u fis with fia uid Insckriess pF ch epee Shoni Plate i High Elecrode Cross Section Short compensation x 1 Fasten Hw id Pouring the liquid into the fixture Note the 16452A is not capable of measuring salt or ionic solutions or other liquids with bulk conductivity due to the electrode polarization phenomenon Keysight is not responsible for any damage e g corrosion smear to the 16452A caused by the reaction between the liguid under test and the 16452A Drain Ligne Must be capped Method of connection 30 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 0
50. ate at a relative humidity of 95 or less at 40 C These same requirements apply to most LCR Meters and Impedance Analyzers When the ambient temperature is not approximately 40 C use an accessory that has no condensation on its surface 4 DUT Device Under Test Dimensions The DUT can vary from chip components axial radial leads or ICs to general electrical materials Select a test fixture lead that is suitable for the shape and size of your components or materials 5 Open amp Short Repeatability and Proportional Error Since a test fixture induces an additional error when measuring the total measurement error is the sum of the measure ment instrument s measurement accuracy and the fixture s additional error Generally a test fixture s additional error con sists of three terms open repeatability short repeatability and proportional error Open and short repeatability exhibit the error factors of the open and short residual impedances which affect the measurements of extremely high and low imped ances respectively Proportional error exhibits the error factor which is proportional to the value of the impedance being measured For more details on this subject please refer to the Appendix 6 Furnished Accessories Each test fixture is shipped with a manual and various other accessories needed for measuring For example the 42941A impedance probe kit is furnished with a pin probe an adapter BNC SMB 3 spare pins a carrying case
51. c Plate Operation and Service Manual Options 16008B 001 Add 26 amp 76 mm diameter electrodes 16008B 002 Add 26 mm diameter electrode Compensation and Measurement Using the selector switch on the 16008B select either the volume resistivity or surface resistivity measurement configuration Open compensation is recommended before measurement and when the measurement configuration Is switched Separate the upper electrode from the main electrode by turning the contact pressure load knob until the distance between the upper electrode and the lower electrode is about 10 mm Then close the top cover and perform open compensation After open compensation the material under test MUT is placed on the main electrode and the upper electrode is placed over it Next turn the load knob to adjust the electrode contact pressure on the MUT Close the top cover and measure the MUT s surface or volume resis tivity It is vital to make measurements which are compatible to a certified test method standard It is shown below that the 16009B can make resistivity measurements which are compatible with ASTM D257 Standard Test Methods for DC Resistance or Conductance of Insulating Materials In the figure and tables shown below the size and shape is compared for two pairs of electrodes one that is speci fied in ASTM D257 and the one that is used with the 16008B Resistivity Cell The similarity implies that the 16008B is compatible with ASTM D257
52. ce tests of Keysight s 4 Terminal Pair LCR meters and impedance ana lyzers Refer to the instrument s operation service manual for the method of using these standards Applicable Instrument 4263B 4268A 4284A 4285A 4288A 4294A E4980A E4981A denotes the instrument is obsolete 16381A 16382A 16383A 16284A Terminal Connector 4 Terminal Pair BNC Capacitance 1 pF 10 pF 100 pF 1000 pF Dimensions approx 142 W x 88 H x 112 D mm capacitors Nominal Accuracy 0 10 Weight approx 7 0 kg including case and 3 capacitors Calibration Stability lt 300 ppm year Dissipation Factor lt 0 0001 Furnished Accessories Description P N Oty 16385A 0 01 uF N A 1 16386A 0 1 uF N A 1 16387A 1 uF N A 1 BNC f f Adapters 1250 0080 4 Calibration Report N A Carrying Case 16380 85104 Operating Note 16380 90221 Option 16380C 001 Add 10 uF standard capacitor 16388A 62 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 63 Min Other Accessories 42030A Four Terminal Pair Standard Resistor Set Terminal Connector 4 Terminal Pair BNC Dimensions approx 94 W x 31 H x 67 D mm resistors Weight approx 3 7 kg including case and 9 resistors dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de Description The 42030A is a s
53. ced by the employment of shielded cables A built in interlocking circuit enables safe high voltage mea surements Applicable Instruments 4339B denotes the instrument is obsolete Frequency DC Maximum Voltage 100 V Maximum Current 0 87 mA Resistance Measurement Range 1 x 10 to 1 x 10 0 Operating Temperature 0 C to 55 C DUT Size See figure below Flectrodes Accessories Description P N Qty 16118 90010 1 Furnished Operation and Service Manual Compensation and Measurement Open compensation is recom mended before measurement Separate the electrodes from each other The separation should be equivalent in size to the DUT s width After performing open compensation the DUT is sandwiched by the tweezers contacts and is measured www datatec de DC High resistance Other Components 16117B Low Noise Test Leads Terminal Connector Type Connector Input Triaxial special screw type Output High Voltage BNC special type Control Interlock connector Interlock connector enables and disables the application of source voltage from the measurement instrument DUT Connection 2 Terminal with triaxial cable Cable Length approx 1 m connector to clips Weight approx 280 g Note Part number corresponding to Probe Socket Clip is as follows am Pin Probes H 9710 2301 Black CH 9710 2302 Red is Soldering Socket n 1200 1903 Black 1200 1904 Red Alligator Clip
54. cription P N Oty 16196 90040 16196 60212 16196 60213 16196 60214 Operation and Service Manual Insulator Assembly 0 85 mm Insulator Assembly 0 75 mm Insulator Assembly 0 68 mm 1 1 Open Plate 16196 29002 short Plate 16196 29027 Push Ring 16196 24004 Magnifying Lens 16193 60002 1 Tweezers 8710 2081 1 Wrench 8710 0909 Cleaning Rod 5182 586 Carrying Case 16196 60250 1 Opt 16196B 710 only 40 a dataTfec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD Options 16196B 710 Add the magnifying lens and tweezers To maintain adeguate measurement performance keep the elec trodes and the short plate in good condition Contaminants and abrasion on these parts considerably affect measurement results especially for low value measurements Periodic fixture cleaning and part replacement is recommended to avoid deterioration of measure ment performance The 16196x fixtures are designed with simplicity in mind so that an operator can easily replace parts Spare parts which are likely to be abraded are supplied with the 16196U Maintenance Kit 16196U Maintenance Kit Opt 16196U 010 Upper electrode 5 piece set common to 16196A B C models Opt 16196U 200 Short plate for 0402 inch 1005 mm size 5 piece set for 16196B Opt 16196U 210 Lower electrode 5 piece set for 16196B Compensati
55. cription P N Oty Test Fixture including Electrode A unguarded electrode and cover N A A Electrode B and cover 16451 60013 B Electrode C and cover 16451 60012 1 6 Electrode D and cover 16451 60014 D Attachment for error compensation and cover 16451 60021 E Hex key for replacing electrodes 5188 4452 F Carrying Case 16451 60001 G iS Ie aun nd Blectode iChangoabie H 7 Note signifies diameter Dimensions are in millimeters kple Dimau dg in giiliieters Dimensions of unguarded electrode Dimensions of fixture assembly 27 E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Material Compensation and Measurement There are three measurement methods for the 16451B They are the Contacting Electrode Method used with 16451B s rigid metal electrode without any electrodes on the material under test the Contacting Electrode Method used with thin film electrodes made on the material under test and the Non Contacting Electrode Air Gap method Select the suitable measurement method and the suitable electrode for the material under test according to the following table Summary of Measurement Method Measurement Contacting Electrode Method used Contacting Electrode Method used Non contacting Electrode Method Method with Rigid metal electrode with thin film electrode Accuracy Low gt H
56. d Rigid Noles nls diarmmeset Desernsions Are i rillirselers Sole sigrifees dianveter Sorsorisiors ace ir miillisrieloss Metal Electrode Material size for electrode A Material size for electrode B Eguipped with Electrodes A and B for flat and smooth materials Electrode Type Diameter of MUT Thickness of MUT Diameter of Electrode Max Freguency A 40 mm 56 mm t lt 10mm 38 mm 30 MHz B 10 mm 56 mm t lt 10mm 5 mm 30 MHz my Guarded ebeotroce are y l Guarded thin fim electrode N a i Guard electrod oO ABS H g 7 j N a i gon MUN erg thin fiirt ppa i lectrode E i Ne f l Tss mneenai ta i ta Kam a Test material the qop widia she he a Era ae ital Unguarded thin film Hal ao gets electrode inguartded electrad Electrorde D Flectrode B Hoto niles diagr ereng ng e o Mie ahr Electrodes for contacting electrode method Thin Material size for electrode C Material size for electrode D Film Electrode Equipped with Electrodes C and D for rough or extremely thin materials Electrode Type Diameter of MUT Thickness of MUT Diameter of Electrode Max Frequency U 96 mm t lt 10mm o gt b0 mm 30 MHz D 20 mm 56 mm t lt 10mm 5 14mm 30 MHz diameter of applied thin film electrode 26 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Material Furnished Accessories Des
57. d Lassalle Str 52 Impedance 10x17 96 5 40x f us 30 125 xf mQ Electrades Standard Option Applicable SMD size SMD Size Code Lx W mm 3225 mm 1210 inch 19 220 152 2520 15 3216 mm 1206 inch 3 2 0 15 x 1 6 0 15 2012 mm 0805 inch 2 0 0 15 x 1 25 0 15 1608 mm 0603 inch 1 6 0 15 x 0 8 0 15 1005 mm 0402 inch 1 0 0 1 x 0 5 0 1 Option 16197A 001 Applicable SMD size SMD Size Code Lx W mm 0603 mm 0201 inch 0 6 0 03 x 0 320 03 Furnished Accessories Description P N Ur Operation and Service Manual 16197 90000 Lever Device Guide 16197 25005 Electrode Plate 16197 00603 199 gadng foch Cleaning Rod 5182 7586 Blank Device Guide 16197 25006 r Bs 3 Magnifying Glass 16193 60002 ad Tweezers 8710 2081 Wrench 8710 0909 Carrying Case 16197 60060 Guide hokker H mm H gt 0 4 H gt 0 4 H gt 0 4 H 2 0 4 H gt 0 4 H mm fl 2 0 25 Oty 2 1 3 1 1 1 One is delivered attached to the test fixture 2 Delivered attached to the fixture 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm a7 Wm dataTec Ferdinand Lassalle Str 52 SMD 72770 Reutlingen EIA EIAJ Size Shorting Bar Set Furnished Size P N Qty 1 x 0 5 x 0 5 mm 16191 29005 1 1 6 x 0 8 x 0 8 mm 16191 29006 1 2 0 x 1 2 x 0 8 mm 16191 29007 1 3 2 X 1 6 x 0 8 mm 16191 29008 1 Device Guide 16197 250
58. dipped Thickness 1 0 mm Residual impedance 20nH 1m Shorting plate dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Lead Components 16047E Test Fixture Terminal Connector 4 Terminal Pair DUT Connection 2 Terminal Dimensions approx 135 W x 40 H x 65 D mm Weight approx 200 g Additional Error Type of Error Proportional Error f lt 15 MHz Proportional Error f gt 15 MHz Open Repeatability Short Repeatability f MHz r SIRE BMAF ar A Hi gi Pauh Hi Sram HEH EJIE Elelgente 5 TITHE Jade eRe 3 Fixture leslening screws 2 Elech och mourn Sores BNC Impedance 0 2 x f 10 4 x f 100 2 n 10 ux f 100 IS 2 m 600 m x f 100 O Test fixture overview 10 Ms dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de Connecting a shorting plate Description This test fixture is designed for impedance evaluation of lead type devices up to 110 MHz A guard terminal is available for three terminal devices and a shorting plate comes secured on this fixture Applicable Instruments 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4991A E5061B 3L5 with Opt E5061B 005 denotes the instrument is obsolete Frequency DC to 110 MHz Maximum Voltage
59. e Other Components Varying in Size shape or Grounded 16117B Low Noise Test Leads b6 16117C Low Noise Test Leads 37 Material 16008B Resistivity Cell 58 59 Other Accessories 16190B Performance Test Kit 60 16380A Standard Capacitor Set 61 16380C Standard Capacitor Set 62 42030A Four lerminal Pair Standard Resistor Set 63 42090A Open Termination 64 42091A Short Termination 64 Appendix The Concept of a Test Fixture s Additional Error 65 1 System Configuration for Impedance Measurement 65 2 Measurement System Accuracy 65 67 3 New Market Trends and the Additional Error for Test Fixtures 67 70 Error Compensation 1 1 Open short Compensation 71 2 Open short load Compensation 71 3 Electrical Length Compensation 72 4 Cable Length Compensation 73 Measurement Repeatability 73 Index 74 Test Fixture Selection By SMD Size 75 Accessories vs Instruments Matrix 76 Keysight Web Resources 77 iv E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Introduction When a device under test DUT is measured a test fixture must be used to connect the instrument to the DUT A test fixture is an interface specifically designed to connect the instrument and the contact tips of the DUT 1 What are Keysight Technologies Accessories Keysight Technologies offers a variety of accessories suitable for many applications They are designed to make measurements
60. e www datatec de Up to 3 GHz 7 mm Material 16453A Dielectric Material Test Description The 16453A is designed for accurate dielectric constant Fixture and loss tangent measurements on the E4991A It employs the par allel plate method which sandwiches the material between two electrodes to form a capacitor E4991A measures the capacitance created from the fixture and option E4991A 002 firmware calculates the relative complex permittivity Adjustment to insure parallel elec trodes is required when using the 16451B This adjustment is not required with 16453A because the fixture has a flexible electrode that adjusts automatically to the material surface Applicable Instruments E4991A with Opt E4991A 002 4291B with Opt 4291B 002 denotes the instrument is obsolete applicable in a limited frequency range Dimensions approx 130 H x 50 W x 60 D mm Frequency 1 MHz to 1 GHz Weight approx 600 g Maximum Voltage 42 V peak max AC DC Measurement Accuracy including the E4991A Operating Temperature 55 C to 200 C When Option E4991A 007 temperature characteristic test kit is used with E4991A the operating temperature range is between 55 C and 150 C Material Size a r Terminal Connector 7 mm t RAAR AAAA AA AAAA RA ARRA AARAA ARAARA AAR AAARANRNAARRR ANAR ARR ARKA ARR ARAR ARRR ARKA ANRR RRR RAAR ARRAAKRA RRR RRRA ARKA AKRA NN RARAARAAARAAAARAAARAAAARAAAARAAAARARAAARAAARAAARARAA A
61. e remove the alligator clips from the input terminals of the test fixture After preparing the open con dition of the respective module close the top cover and then per form open compensation Finally the DUT is inserted into the respective module The following figure shows the three component modules SMD module Axial Lead module Alligator clip and Flat table SHD mogul Axlal Lead ynodulg Al sater clip ang Fist lable Component module configurations 54 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de 55 Mn DC High resistance SMD 16118A Tweezers Test Fixture a N Terminal Connector Type Connector Input Triaxial special screw type Output High Voltage BNC special type Control Interlock connector Interlock Connector enables and disables the application of source voltage from the measurement instrument DUT Connection 2 Terminal with triaxial cable Cable Length approx 1 m Weight approx 410 g aso A Lyfrau E mp G d N F Popp _ BILL shunt Connector ee furnished with IN tha 43308 181184 i 4339B with 16118A dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de Description The 16118A is designed to operate specifically with 4339B The tweezer s contacts makes it easy to hold SMD Electrical noise effects are redu
62. e 16191 29001 1 16192A 701 1 x 1x 2 4 mm Shorting Device 16191 29002 1 16192A 701 1 6 x 2 4 x 2 mm Shorting Device 16191 29003 1 16192A 701 2 4 x2 4 x 3 2 mm ohorting Device 16191 29004 1 16192A 701 2 4x 2 4 x 4 5 mm EIA EIAJ Industrial Standard Sized 16191 29005 1 16192A 010 Shorting Device 1 x0 5x0 5 mm 16191 29006 1 16192A 010 Shorting Device 1 6 x 0 8 x 0 8 mm 16191 29007 1 16192A 010 Shorting Device 2 0 x 1 2 x 0 8 mm 16191 29008 1 16192A 010 Shoring Device 3 2 x 1 6 x 0 8 mm Case for Shorting Devices 1540 0692 16192A 010 701 35 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD Options 16194A 010 Add EIA EIAJ industrial standard sizedshorting bar set 16194A 701 Add general sized shorting bar set Compensation and Measurement Before beginning the measure ment the appropriate device holder for a SMD or lead component must be prepared with the text fixture The following figure shows how the device holder is exchanged to match the device type The next step is to perform open and short compensations in combi nation with the electrical length compensation When measuring Remove the ground plate When measuring SMD attach the knob on the device holder Select the device holder suitable for the device type Loosen its knob and insert into the arm Device Ffolder
63. e measurements As a result the test fixture s short repeatability has become increasingly important In the figure below the relationship between proportional error short offset error and freguency are shown when measuring low impedance of 100 mQ and 10 O Notice that when the measured impedance is less than 100 mO short offset error influences the entirety of the test fixture s inherent error As shown in the figure below when the DUT s impedance is 100 mQ and the test fixture s short repeatability is 10 mO the short offset error will be 10 percent Since the proportional error is minimal in low freguencies the additional error will be 10 percent Until recently to allow for additional error in test fixtures tt was common to just specify the proportional error A As shown in the 10 O measurement case if the measured impedance is large in comparison to the test fixture s short repeatability then the short offset error can be ignored completely This is the reason why open and short offset error was not previously specified This is the reason for test fixtures that are only specified with proportional error On the contrary for measured impedance from 1 O to 10 kQ proportional error A alone is sufficient to express the test fix ture s additional error fx gt adam G2 fue my FTE II LID ETAT ELLA oon HULL PE ia B aiD LL AN FEB nl Mar AEE MEN HIL NENN E 16 TODE 134 HM 1104 Ik 10 100k Tht 10M 11014 Frennemdy H Frequerigy
64. ements for discrete devices to advanced measurements such as measurement of resistivities or dielectric constants 2 Compatibility with Measurement Instruments Test fixtures leads are compatible with the measurement instruments when they have the same type of terminal configura tion and useable measurement freguency range The measurement instruments described in this guide are divided into the following three categories based on freguency Freguency Up to 110 MHz Up to 3 GHz Range Terminal Configuration 4 Terminal Pair Terminal Configuration 7 mm TT ce ug m E a i u bn O Measurement 4263B 4268A 4279A 4284A 4291B 4294A 42942A E5061B 3L5 w Opt 005 Instruments 4285A 4288A 4294A E4980A E4981A 16201A 4287A E4982A E4991A Any 4TP instruments excluding 4294A 16085B Frequency DC High Resistance Measurement Measurement 4339B 4349B Instruments 3 W dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Tips for Selecting Appropriate Accessories 3 Frequency DC Bias and Operating Temperature Humidity Each of the Keysight accessories has its own specific operating range Any measurement performed outside this range can increase residual errors and can cause problems Be sure that your measurement environment fits the accessory s specific operating range In the case of humidity Keysight s accessories can oper
65. erdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de Description This test fixture is designed for impedance evaluations of SMD The minimum SMD size that this fixture is adapted to evaluate is 0 6 L x 0 3 W mm Applicable Instruments 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4991A E50613 3L5 with Opt E5061B 005 denotes the instrument is obsolete Freguency DC to 110 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size See figure below LD 1 5 0mm 1 0 3 5 Omen cal Emm reed Bi i 2 Electrode my Di A 1 drai i a Hr H 2 imm3 Furnished Accessories Description P N Oty Case for 100 O SMD Resistance 1540 0692 1 1000 Chip Resistor 0699 2488 10 Operating Manual 16034 90011 1 Compensation and Measurement Open and short compensations are recommended before measurement When measuring above 3 MHz load compensation is also recommended Open compensation is performed by separating the high and the low electrodes from each other The separa tion size should be equivalent to the DUT s width Short compensation is performed placing the high and low electrodes in contact together Load compensation is performed by using the furnished 100 Q SMD chip resis tor After performing open short and load compensations the DUT is inserted into the test fixture The following figures show how compensa
66. fixtures 1 Open short Compensation For a simple measurement system represented by the eguivalent circuit model shown below residual impedance values Rs and Ls and admittance values Co and Go can be corrected by i Measuring open condition for the test fixture s admittance ii Measuring short condition for the test fixture s impedance ii Measuring the test sample then subtracting the admittance Instrument and impedance The above procedure is performed internally by the open short compensation When the test fixture is directly connected to the measurement instrument the open short compensation sufficient ly corrects the measurement error 2 Open short load Compensation DUT When the measurement system is too complicated to be represented as the above equivalent circuit model the open short compensation cannot completely compensate for the residual impedance In this case the open short load com pensation is used rather than the open short compensation The open short load compensation is particularly effective when for example the 16065A is used with the measure ment instrument The following table lists measurement instruments and available compensation functions Model Open compensa Short compensation Load compensation Load s input parameters tion 4263B 5 Q R X Cp D Cp Q Cp G Cp Rp Cs D Cs Q Cs Rs Ls D Ls Q Ls Rs 4269A Cp D Cp Q Cp G Cp Rp Cs D Cs Q Cs Rs 4279A 5 Cp
67. for example when additional error is egual to 0 596 If the obtained impedance values are plotted with measurement impedance y axis against freguency x axis a graph similar to the one shown down below can be obtained The shaded area shows the range of impedance that can be measured with an additional error better than 0 596 In the same way other graphs can be drawn with other additional error values to better visualize the accuracy that can be obtained for a given impedance and frequency range The operational manuals of recent test fixtures present such graphs Additional Error of 16034G Type of Error Impedance Proportional Error 0 5 x f 10 90 ee eC ee ee ee Em rir eee EI ai rab FE Se LL LL u Short Repeatability 10 13 x f 10 mO ee F zii i LY TU NU Mut bel ui en 2IL mmm e ARTE or nr Hindley DE ee mer BE ee sai 210 HT OO LLI COL Ll EL UL IN I coe BE 2 Hmm eiii mm iH I 4 zii Ba Hi IE Denen AFEH ee ttt En FFF Sail sti Serie Sete Ser Frecuency ibe Range of impedance measurable with additional error lt 0 5 70 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Compensation Error Compensation Keysight measurement instruments incorporate one of the following four types of error compensation functions to elimi nate residual impedance effects in test
68. g Device 16191 29001 1 16192A 701 1x1x2 4 mm ohorting Device 16191 29002 1 16192A 701 1 6 x 2 4 x 2 mm Shorting Device 16191 29003 1 16192A 701 2 4 x2 4 x3 2 mm ohorting Device 16191 29004 1 16192A 701 2 4 x2 4 x 4 5 mm EIA EIAJ Industrial Standard Sized 16191 29005 1 16192A 010 Shorting Device 1 x 0 5 x 0 5 mm 16191 29006 1 16192A 010 Shorting Device 1 6 x 0 8 x 0 8 mm 16191 29007 1 16192A 010 Shorting Device 2 0 x 1 2 x 0 8 mm 16191 29008 1 16192A 010 Shoring Device 3 2 x 1 6 x 0 8 mm Case for Shorting Devices 1540 0692 1 16192 A 010 701 Magnifying Lens 16193 60002 1 16192A 710 Tweezers 8710 2081 16192A 710 Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD Wr l Options qth positlaning stage Device slage i Length locking knab f 16192A 010 Add EIA EIAJ industrial standard sized shorting bar see i Atagnifying jens set 16192A 701 Add general sized shorting bar set 16192A 710 Add the magnifying lens and tweezers evlce heidar Compensation and Measurement Open and short compensations Widih poaltloning Are recommended in combination with the electrical length compen sation before measurement The fixture s electrical length must be Width pos ltenineg en entered into the electrical length compensation function of the mea ee ea o Y Wn surement instrument first Then open compe
69. here it ig the Di T e mouched by Heal eric at Lie hw wich slactroge A ET nn Loosen the Kaoh en Push the end of the high alecioda niil 8 touches lie DUT 2 Tighten the nob 4 P ull the lever of the tow leclode and remove the DUT l The distance between tne i YSOReE TERT MIXTURE nigh and iaw centact electroctes SHOLIG lhe same as me DUETS Site at 1 cn5ean ths Fon ZF uso he end of ihe high ecrade UGE if touches the hn elect ode 3 Tighten Ee Kren Kuh 2235 TNG width Open compensation Short compensation Device holder Measurement einntrode op vie side view Release fever and psen DIET L One Units nw Inserting a DUT Electrode dimensions 12 E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHZ 4 Terminal Pair SMD 16034G Test Fixture Terminal Connector 4 Terminal Pair DUT Connection 2 Terminal Dimensions approx 120 W x 50 H x 70 D mm Weight approx 200 g Additional Error Type of Error Proportional Error Open Repeatability ohort Repeatablity f MHz BNC Impedance 0 5 x f 10 5 500 x f 10 nS 10 13 x f 10 mQ MAMAPA RATAN PAANI AL wm A CY do oh nel ar ps s T yN HOU LE 4 Fr ern F ins J Der H d iua Ml HEH WM nn r Lzeikans yf kan FOREN DUT measurement 13 Mn dataTec F
70. ifying lens and tweezers To maintain adequate measurement performance keep the elec trodes and the short plate in good condition Contaminants and abrasion on these parts considerably affect measurement results especially for low value measurements Periodic fixture cleaning and part replacement is recommended to avoid deterioration of measure ment performance The 16196x fixtures are designed with simplicity in mind so that an operator can easily replace parts Spare parts which are likely to be abraded are supplied with the 16196U Maintenance Kit 42 E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 16196U Maintenance Kit Opt 16196U 010 Upper electrode 5 piece set common to 16196A B C models Opt 16196U 300 Short plate for 0201 inch 0603 mm size 5 piece set for 16196C Opt 16196U 310 Lower electrode 5 piece set for 16196C Compensation and Measurement Open and short compensations are recommended in combination with the electrical length compen sation before measurement The fixture s electrical length must be entered into the electrical length compensation function of the mea surement instrument first Next open compensation is performed by placing the furnished open plate on top of the insulator assembly Short compensation is performed by placing the furnished shorting plate on top of the ins
71. igh Operation Sol ea Y ret DY DRUD hee EF gt Complex Applicable Thick solid and smooth materials Materials on which thin film can be Thick and soft materials Rough Materials applied without changing its charac materials also teristics Open and short compensations are recommended in combination with the cable length compensation before measure ment When measuring above 5 MHz with the 4285A or the 4294A load compensation is also recommended First set the instrument s cable length compensation function to 1 m Then open and short compensation is performed by using the furnished electrode attachment Load compensation is performed by preparing a working standard After performing open short and load compensations the MUT is sandwiched by the parallel electrodes and the capacitance is measured Relative permittivity is calculated from the measured capacitance in the following manner ta x Cp er d 2 JL X gt x Relative permittivity Capacitance measurement data 8 854 x 10 1 F m Average thickness of test material na MO amp Diameter of guarded electrode For more information on load compensation with the 4294A refer to section 6 calibration of the 4294A operation manual Fixture Compensation when the 16451B is used and section 13 of the 4294A programming manual Measuring Dielectric Materials 28 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50
72. ight approx 1500 g High Voltage BNC f connector for external bias input BNC f connector for voltage monitor output instrument i DDA LCR meter with 16065A 16065C 40Vdc External Voltage Bias Adapter Terminal Connector 4 Terminal Pair BNC External Bias Input connector BNC f Dimensions approx 160 W x 50 H x 150 D mm Cable Length approx 210 mm Weight approx 450 g Instrument LCR meter with 16065C Description This test fixture makes it possible to measure a DUT with up to 200 V DC bias The same modules of 16047A can be used to allow measurements of axial radial lead components Applicable Instruments 4263B 4268A 4279A 4288A E4981A 4284A 4285A 4294A E4980A denotes the instrument is obsolete applicable in a limited frequency range Frequency 50 Hz to 2 MHz Maximum DC Bias 200 V DC max 15 V peak AC max Blocking Capacitor of 5 6 uF is connected in with the Hc terminal Operating Temperature 0 C to 55 C DUT Size See the 16047A figure with module sizes Furnished Accessories Description P N Oty Module For Axial Lead 16061 70022 1 Module For Radial Lead mounting on fixture 16061 70021 1 Module For Short Radial Lead 16047 65001 1 Shorting Bar 16047 00640 1 Operating and Service Manual 16065 90011 1 Compensation and Measurement Open short and load compensa tions are recommended before measurement Short compensation is performed by shorting the c
73. inand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 53 Mn DC High resistance Test Fixtures for DC High Resistance Measurements Applicable Instrument Freguency Range Measurement Instruments Freguency Measurement Instruments Test Fixture Freguency Applicable Instrument 16339A DC 4339B 16118A DC 4339B 16117B C DC 4349B 16008B DC 4339B Up to 110 MHz Up to 3 GHz Terminal Configuration 4 Terminal Pair Terminal Configuration 7 mm pu 4263B 4269A 4279A 4284A 4291B 4294A 42942A E5061B 3L5 w Opt 005 16201A 4285A 4288A 4294A E4980A E4981A 4287A E4982A E4991A Any 4TP instruments excluding 4294A 16085B DC High Resistance Measurement 4339B 4349B dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 5010 info datatec de www datatec de DC High resistance SMD amp Lead Components 16339A Component Test Fixture Description The 16339A is designed to operate specifically with 4339B It is provided with three component modules which are used to hold SMD lead and various type of devices Electrical noise effects are reduced by the employment of a shielded case A built in interlocking circuit enables safe high voltage measurements Applicable Instruments 4339B denotes the instrument is obsolete Frequency DC Maximum Voltage 1000 V Maximum Current 10
74. length com pensation the DUT is inserted into the test fixture Inserting the leaded component 32 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 33 Min Up to 3 GHz 7 mm SMD 16192A Parallel Electrode SMD Test Fixture Description This test fixture is designed for impedance evaluations of parallel electrode SMD The minimum SMD size that this fixture is adapted to evaluate is 1 L mm Applicable Instrument 4291B 4294A 42942A 4287A E4982A E5061B 3L5 with Opt E50616B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4299A E4980A E4981A denotes the instrument is obsolete applicable in a limited frequency range Frequency DC to 2 GHz Terminal Connector 7 mm DUT Connection 2 Terminal Electrical Length 11 mm Dimensions approx 150 W x 70 H x 90 D mm Weight approx 400 g Additional Error Type of Error Impedance Proportional Error 1 5 x f Open Repeatability 2 30 xf us Short Repeatability 30 250 x f mQ f GHz dataTec Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 85 C DUT size 1 mm to 20 mm length WM D B to S eT Hog osam EL to 20 tyst Electrodes Furnished Accessories Description P N Qty Option Operation and Service Manual 16192 90040 1 standard General Sized Shortin
75. lips Applicable Instruments 4263B 4269A 4284A 4285A 4288A 4294A E4980A E4981A denotes the instrument is obsolete applicable in a limited frequency range Freguency 5 Hz to 100 kHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size See figure below approx YH OR a lt mm Furnished Accessories Description P N Oty Operating and Service Manual 16089 90020 1 Compensation and Measurement Open and short compensations are recommended before measurement For open compensation do not connect the Kelvin clips to anything Short compensation is per formed by holding a shorting plate with the Kelvin clips After per forming open and short compensations the DUT is held with the Kelvin clips Description This test fixture makes it possible to measure odd shaped components that cannot be measured with conventional fix tures lt is eguipped with two insulated Kelvin clips Applicable Instruments 4263B 4268A 4294A 4285A 4289A 4294A E4980A E4981A denotes the instrument is obsolete applicable in a limited frequency range Freguency 5 Hz to 100 kHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size See figure below approx 8imm Ji y Furnished Accessories Description P N Oty Operating and Service Manual 16089 90020 1 Compensation and Measurement Open and short compensations are reco
76. m Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size See figure below Furnished Accessories Description P N Qty Operating and Service Manual 16089 90020 1 Compensation and Measurement Open and short compensations are recommended before measurement For open compensation do not connect the Kelvin clips to anything Short compensation is per formed by connecting the Kelvin clips together After performing open and short compensations the DUT is held with the Kelvin clips Description This test fixture makes it possible to measure odd shaped components that cannot be measured with conventional fixtures It is equipped with four alliga tor clips Applicable Instruments 4263B 4268A 4284A 4285A 4289A 4294A E4980A E4981A denotes the instrument is obsolete applicable in a limited frequency range Frequency 5 Hz to 100 kHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size diameter of DUT s leads lt 5 mm Furnished Accessories Description P N Qty Operating and Service Manual 16089 90020 1 Compensation and Measurement Open and short compensations are recommended before measurement For open compensation do not connect the alligator clips to anything Short compensation is performed by holding a shorting plate with the alligator clips Make sure that the alligator clips with the V markers are next to each other After perf
77. mmended before measurement For open compensation do not connect the Kelvin clips to anything Short compensation is per formed by connecting the Kelvin clips together After performing open and short compensations the DUT is held with the Kelvin clips 18 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 19 Mn Up to 110 MHz 4 Terminal Pair Other Components 16089C Kelvin IC Clip Leads be ey E Terminal Connector 4 Terminal Pair BNC DUT Connection 4 Terminal Cable Length approx 1 3m from connector to clip s tip Weight approx 300 g Additional Error The additional error is negligible when compared to the instrument s accuracy 16089D Kelvin Alligator Clip Leads Terminal Connector 4 Terminal Pair BNC DUT Connection 4 Terminal Cable Length approx 0 94m from connector to clip s tip Weight approx 460 g Additional Error The additional error is negligible when compared to the instrument s accuracy Description This test fixture makes it possible to measure odd shaped components that cannot be measured with conventional fix tures lt is equipped with two insulated Kelvin clips Applicable Instruments 4263B 4269A 4284A 4285A 4289A 4294A E4990A E4981A denotes the instrument is obsolete applicable in a limited frequency range Frequency 5 Hz to 100 kHz Maximu
78. n and Measurement Adapter setup is recommended before measurement In the adapter setup menu select 4TP 2M Then use the 100 O resistor furnished with the 4294A to perform phase compensation and load data measurement 22 E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Probes 42941A Impedance Probe Kit qo EH Terminal Connector 4 Terminal Pair BNC Cable Length approx 1 5 m Weight approx 2400 g Basic Measurement Accuracy 1 For detailed information refer to the operation manual or the specifications of 4294A 4294A with 42941A Open compensation Short compensation 23 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Description This impedance probe kit is designed for use with the 4294A It provides the capability to perform in circuit measurements printed circuit patterns the input output impedance of circuits etc with better accuracy and wider impedance coverage from 40 Hz to 110 MHz DUTs can be connected by either using the pin probe the clip lead alligator clip adapter or the BNC adapter All probe adapter can be used from 40 Hz to 110 MHz The pin probe is best for in circuit board mounted components The clip lead is for com ponents too large for the pin probe The BNC adapter is used to con nect circuits or networks eguipped with BNC connectors
79. nnector to connector Weight approx 290 g Adapter features 1 High voltage BNC cable This connector provides the source volt age to the 16117C This is a high voltage BNC connector and is not compatible with standard BNC connectors 2 Triaxial cable The measured signal is carried on the center con ductor of this connector This is a special screw type triaxial con nector and is not compatible with standard triaxial connectors 3 Interlock cable This connector enables the interlock function which enables and disables the application of source voltage from the 4339B when the interlock line is connected and discon nected respectively High voltage BNC special type female connector Triaxial special screw type female connector iz 57 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de DC High resistance Material 16008B Resistivity Cell Terminal Connector Type Connector Input Triaxial special screw type Output High Voltage BNC special type Control Interlock connector Interlock Connector enables and disables the application of source voltage from the measurement instrument DUT Connection 2 Terminal with triaxial cable Dimensions approx 240 W x 180 H x 240 D mm Cable Length approx 1 2 m connector to electrodes Weight approx 7000 g GOLAU Einchoge Size Typical Meigrial Shape
80. nsation is performed by adkatmeni separating the high and the low electrodes from each other The AE ON separation should be equivalent in size to the DUT s width Short compensation is performed by using option 16192A 010 701 short ing bar set After performing open and short compensations in com bination with the electrical length compensation the DUT is inserted into the test fixture The following figures show how compensation and measurement is performed a Place the shoring device DUT se that t contacts the electrode b Adjust the two Knobs and the two device holders l IT ated to hold the shorting device DUT H hh _ dien Tighten knob while pushing the lever Frei x Hi y TH FGM t TEN sa k Sach y j d Release the lever to hold the shorting device DUT i i Perform short compensation or measure the DUT d e Set the DUT on the test fixture just as you set the shoring device in the SHORT compensation f Push the black lever and remove the DUT y Perform open compensaton _ EH Fi EN a Fir z Er I ji Lie For more details please refer to 16192A operation manual Open Short compensation 34 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 16194A High Temperature Component Test Fixture Terminal Connector 7 mm DUT Connection 2 Terminal Electrical Length 50 mm Dime
81. nsions approx 150 W x 40 H x 80 D mm Weight approx 350 g Additional Error SMD Type of Error Proportional Error Open Repeatability Short Repeatability Leaded device Type of Error Proportional Error Open Repeatability Short Repeatability f GHz Impedance 20 x f 80 250 x f us 0 2 2 5 xf O Impedance 20 x f 80 500 x f uS 0 4 12 5 xf Q Description This test fixture is designed for measuring both axial radial leaded devices and SMD within the temperature range from bb to 200 C when used with the E4991A 007 Temperature Characteristic Test Kit bb to 150 C Applicable Instrument 4294A 42942A 4287A E4982A E5061B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4299A E4980A E4981A denotes the instrument is obsolete applicable in a limited frequency range Frequency DC to 500 MHz with open and short compensation DC to 2 GHz with open and short and load compensation Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 200 C DUT size See figure below Furnished Accessories Ele 15mm Sa nym See ridet iii I ateatem T te ibe ae orient Description P N Qty Option Wrench 8710 1181 Standard Tweezers 8710 2081 1 Standard 500 SMD Resistor 0699 2829 10 Standard Operation and Service Manual 16194 90030 1 Standard General Sized Shorting Devic
82. oltage and current terminals are separate Since the voltmeter has high input impedance no current flows into the voltage terminals Hence the voltage that is applied across the DUT can be accurately detected without being affected by the contact resistance Also the current that flows through the DUT flows directly into the current terminal and is accurately detected without being affected by the contact resistance As a result the 4T con nection method can eliminate the effect of contact resistance and realize a small short repeatability By using a 4T test fixture it is possible to measure low impedance with better accuracy than that which can be measured with a 2T test fixture The 2T test fixture can be used up to a higher frequency than the 4T test fixture Since the 2T test fixture has a simple DUT connection configuration the effects of residuals and mutual coupling jwM which cause measurement error to increase with frequency are smaller than those of the 4T test fixture and can be effectively reduced by compensation Thus the 2T connection is incorporated in test fixtures designed for use in the higher freguency region typically up to 40 or 110 MHz 68 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de 3 3 Test Fixture s Adaptability for a Particular Measurement In order to make use of what has been discussed previously the test fixture s ad
83. om the low terminals After performing open and short compensations the DUT is inserted into the test fixture The figures below show how compensation and measurement are performed Open compensation Short compensation Inserting a DUT 1 8 A DYT Mr OO Diameter of probe pin is 0 19 mm Diameter of Blade electrode s contact spot is 0 5 mm Spacing between Blade electrode and probe pin is 0 11 mm SMD can be between 1 6 and 8 0 mm in width Electrode dimensions 16 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair SMD 16334A Tweezers Contact Test Fixture Terminal Connector 4 Terminal Pair BNC DUT Connection 2 Terminal Cable Length approx 1m from BNC connectors to the top of tweezers Weight approx 290 g Additional Error Type of Error Impedance Proportional Error 23 10 f MHz 17 E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de Description This test fixture is designed for impedance evaluations of SMD The minimum SMD size that this fixture is adapted to evalu ate is 1 6 L x 0 8 W mm The tweezers contacts on this fixture makes it easy to hold the DUT Applicable Instruments 4263B 4268A 4279A 4284A 4288A E4980A E4981A 4285A 4294A denotes the instrument is obsolete a
84. on and Measurement First of all install the appropriate insulator assembly into the fixture Then perform compensation Open and short compensations are recommended in combination with the electrical length compensation before measurement The fixture s electrical length must be entered into the electrical length compensation function of the measurement instrument first Next open compensation is performed by placing the furnished open plate on top of the insulator assembly Short compensation is performed by placing the furnished shorting plate on top of the insulator assembly After performing open and short compensations in combi nation with the electrical length compensation the DUT is inserted into the test fixture Once the measurement of the DUT is complete remove the DUT from the fixture by using the furnished push ring Refer to the 16196A figures to see how compensation and measure ment is performed 41 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 16196C Parallel Electrode SMD Test Fixture ke N NS Terminal Connector 7 mm DUT Connection 2 Terminal Electrical Length 27 1 mm Dimensions approx 140 W x 48 H x 78 D mm Weight approx 250 g Additional Error Type of Error Impedance Proportional Error 1 0 x f 96 Open Repeatability 5 40 xf uS Short Repeatability 30 125 x f m
85. on factor ee e a a eae 4 CT z Frequency Er e permittivity of air 8 854x10 F m Zn lee ryd d diameter of electrode A B re as WD t thickness of material mm Az Impedance measurement error of instrument Ad D measurement error of instrument Typical Loss Tangent tan Measurement Accuracy The material is assumed to be ideally flat lection A eye me The above eguation is applicable for electrodes A and i ME E B when using the contacting electrode method Feegndoiy H Friaieiney Hr nun re rrr gyr rr Bl Aed uu He a AH weit ar m i re Eh 4294A Measurement Settings 1 Osc level 500 mV 2 BW 5 3 Adapter setup 1 m 4 4294A with 16451B Compensation Open short and load 25 a dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Material Applicable Instruments 4263B 4268A 4279A 4284A 4285A 4299A E4980A E4981A 4294A denotes the instrument is obsolete applicable in a limited frequency range Freguency DC to 30 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C Material Size sea em Electrode A ail Guarded electrode N CD ge Test materiaf Ten materia Liniguarded Hectrode Elactroda A Electrode B ne A Er st i ee aie Electrodes for contacting electrode metho
86. ontacts of the test fixture with a shorting plate as described for the 16047A Load compensation is performed by inserting a known standard device After performing open short and load compensations the DUT is connected to the test fixture Description This adapter is designed to operate specifically with the 4263B 4269A 4288A and the E4981A By connecting an external DC voltage source to this adapter a bias voltage of up to 40 V can be supplied to a DUT The DUT can be inserted by connecting any direct attachment 4 Terminal Pair test fixture to the adapter Applicable Instruments 4263B 4269A 4288A E4981A Frequency 100 Hz to 1 MHz Maximum DC Bias 42 V peak max AC DC Blocking Capacitor of 100 uF is connected in series with the Hc ter minal Operating Temperature 0 C to 55 C Applicable Fixtures 16034E G H 16044A 16047A D E 16048A B D E 16089A B C D E denotes the instrument is obsolete Furnished Accessories Description P N Qty Operating and Service Manual 16065 90020 1 Compensation and Measurement Open and short compensations are recommended before measurement Short compensation is performed by shorting the contacts of the test fixture that is in use After perform ing open and short compensations the DUT is connected to the test fixture 24 E data Tec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de U
87. or A was derived from the error factor which causes the absolute impedance error to be proportional to the impedance being measured If only the first term is taken out of the above eguation and multiplied by Zx then AZ A Zx O This means that the absolute value of the impedance error will always be A times the mea sured impedance The largeness of proportional error is dependent upon how complicated the test fixture s construction is Conceptually it is dependent upon the stability of each element of the fixture s eguivalent circuit model From previ ous experience proportional error is proportional to the freguency sguared 2 2 Short Offset Error The term Zs Zx x 100 is called short offset error If Zx is multiplied to this term then AZ Zs O It can be concluded that this term affects the absolute impedance error by adding an offset Short repeatability Zs is determined from the variations in multiple measurements of the test fixture in short condition After performing short compensation the measured values of the short condition will distribute around 0 in the complex impedance plane The maximum value of the impedance vector is defined as short repeatability This is shown in the figure below The larger short repeatability is the more difficult it is to measure small impedance values For example if the test fixture s short repeatability is 100 mQ then the additional error of an impedance measurement under 100 mQ will be more
88. orming open and short compensations the DUT is held with the alligator clips Connect the same color test clips to the same terminal of the DUT and have the clips with V markers be closer to the DUT dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 5010 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Port Cable Extension 16048A Test Leads Description The test leads extend the measurement port with a 4 Terminal Pair configuration It is provided with a BNC female con nector board to allow the attachment of user fabricated test fixtures Applicable Instruments 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A denotes the instrument is obsolete Frequency DC to 30 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C Furnished Accessories Terminal Connector 4 Terminal Pair BNC Description P N Qty Cable Length approx Terminal Board with BNC f x4 16032 60001 1 0 94m from connector to cable tip Operating Manual 16089 90001 1 Cable Tip BNC male Weight approx 315 g Compensation and Measurement Cable length compensation is Additional Error For detailed information refer recommended before measurement Set the instrument s cable to the measurement instrument s specifications length compensation function to 1 m 16048D Test Leads Description The test leads extend the measurement port with a 4 Terminal Pair configura
89. otes the instrument is obsolete Terminal Connector 4 Terminal Pair BNC Dimensions approx 94 W x 31 H x 67 D mm Weight approx 120 g 42091A Short Termination Description The 42091A is a short termination and is primarily used for performance tests of Keysight s 4 Terminal Pair LCR meters and impedance analyzers Refer to the instrument s operation service manual for the method of using this standard Applicable Instrument 4263B 4268A 4279A 4284A 4285A 4294A 4339B E4980A E4981A denotes the instrument is obsolete Terminal Connector 4 Terminal Pair BNC Dimensions approx 94 W x 31 H x 67 D mm Weight approx 120 g 64 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Additional Error The Concept of a Test Fixture s Additional Error 1 system Configuration for Impedance Measurement Frequently the system configured for impedance measurements uses the following components 1 Impedance measurement instrument 2 Cables and adapter interfaces 3 Test fixture DU Impedance Measurement Instrument Cable Adapter g fest Fixture System configuration for impedance measurement The impedance measurement instrument s accuracy is defined at the measurement port of the instrument This means that the accuracy at the measurement port is guaranteed and has calibration tr
90. p to 110 MHz 4 Terminal Pair Material 16451B Dielectric Test Fixture Description The 16451B is used to evaluate the dielectric constant of solid dielectric materials accurately and complies with ASTM D150 The 16451B employs the parallel plate method which sand wiches the material between two electrodes to form a capacitor LCR meter or an Impedance Analyzer is then used to measure the capacitance created from the fixture A measurement block diagram of the parallel plate method is shown below Terminal Connector 4 Terminal Pair BNC Dimension approx See page 29 Cable Length approx 0 8 m from connector to electrodes Weight approx 3700 g Measurement Accuracy e accuracy Parallel plate method tan lt 0 1 ae Notice the stray capacitance which is formed on the test material ma 4 MM An 0 04 f 0 S 10 g S shown in the figure above The guard electrode helps to eliminate t an the stray capacitance at the edge of the electrode e Loss Tangent Accuracy A tan Basic Measurement Accuracy including the 4294A tan lt 0 1 Ad Ea Eb n 2 Ea 0 005 0 0004 f2 m 9 a Typical Permittivity e r Measurement Accuracy _ tan A rm Eb 00 erm p 6 gi an supplemental performance characteristics o 7 7 e f measured frequency Hz fs 30MHz 3 a ee ee 1 E measured permittivity o i tan measured dissipati
91. pplicable in a limited frequency range Frequency 5 Hz to 15 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C DUT Size lt 10 mm width See figure below Miectrodes Furnished Accessories Description P N Qty Compensation Block 16334 60001 1 Operating Note 16334 90000 1 Compensation and Measurement Open and short compensations are recommended before measurement Open and short compensa tions are performed by using the furnished compensation block After performing open and short compensations the DUT is sand wiched by the tweezers contacts and is measured www datatec de Up to 110 MHz 4 Terminal Pair Other Components 16089A Large Kelvin Clip Leads Terminal Connector 4 Terminal Pair BNC DUT Connection 4 Terminal Cable Length approx 0 94m from connector to clip s tip Weight approx 300 g Additional Error The additional error is negligible when compared to the instrument s accuracy 16089B Medium Kelvin Clip Leads Terminal Connector 4 Terminal Pair BNC DUT Connection 4 Terminal Cable Length approx 0 94m from connector to clip s tip Weight approx 300 g Additional Error The additional error is negligible when compared to the instrument s accuracy Description This test fixture makes it possible to measure odd shaped components that cannot be measured with conventional fix tures lt is eguipped with two insulated Kelvin c
92. r 7 mm Operating Temperature 0 C to 55 C DUT Connection 2 Terminal DUT Size See figure below Electrical Length 3 4 mm Dimensions approx 150 W x 70 H x 80 D mm Weight approx 180 g MU a Additional Error See figure below Meade 0001A Residue Parameter ae a GE MBWRDAIRU 3 GYF jpt en n us 7 cam le Fa damen i sn i oot me MER ca OE BRACH ze nn a ei ihe A me A Seg j exept N a see ti rab 800 Furnished Accessories TEST TREGUENLY MH ee Pann ee Description P N Qty Shorting Plate 16092 08010 1 Operating Note 16092 90010 1 Compensation and Measurement Open and short compensations are recommended in combination with the electrical length compen sation before measurement The fixture s electrical length must be entered into the electrical length compensation function of the mea surement instrument first When using the SMD module open com pensation is performed by separating the high and the low elec trodes from each other The separation should be equivalent in size to the DUT s width Short compensation is performed by usinf the furnished shorting plate When using the lead component module open compensation is performed by not having the module elec y4 trodes be connected to anything Short compensation is performed by using the furnished shorting plate After performing open and short compensations in combination with the electrical
93. rminal Pair BNC Dimensions approx 142 W x 88 H x 112 D mm capacitors Weight approx 8 0 kg including case and 4 capacitors dataTec Description The 16390A is a standard capacitor set consisting of four precision capacitors 1 pF 16391A 10 pF 16392A 100 pF 16383A 1000 pF 16384A These capacitors are primarily used for performance tests of Keysight s 4 Terminal Pair LCR meters and impedance analyzers Refer to the instrument s operation service manual for the method of using these standards Applicable Instrument 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A denotes the instrument is obsolete Capacitance Nominal Accuracy Calibration Stability Dissipation Factor Furnished Accessories Description 16381A 1 pF 16382A 10 pF 16383A 100 pF 16384A 1000 pF BNC f f Adapters Calibration Report Carrying Case Operating Note 16381A 16382A 1 pF 10 pF 0 10 lt 300 ppm year lt 0 0001 P N N A N A N A N A 1250 0080 N A 16380 85101 16380 90011 Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de 16383 A 100 pF 16284A 1000 pF www datatec de Other Accessories 16380C Standard Capacitor Set Description The 16380C is a standard capacitor set consisting of three precision capacitors 0 01 uF 16385A 0 1 uF 16386A and 1 uF 16387A These capacitors are primarily used for performan
94. shed Accessories Description P N Qty Operation and Service Manual 13200 90011 1 16200B 001 Shorting Device Set Size P N Oty 0 6 x 0 3 x 0 3 mm 16197 29001 2 1 x 0 5 x 0 5 mm 16191 29005 2 1 6 x 0 8 x 0 8 mm 16191 29006 2 2 0 x 1 2 x 0 8 mm 16191 29007 2 Connection example 3 2 x 1 6 x 0 8 mm 16191 29008 2 16200B 001 Load Device Set 0 6 x 0 3 x 0 3 mm 1x 0 5 x 0 5 mm 1 6 x 0 8 x 0 8 mm 2 0 x 1 2 x 0 8 mm 5182 0435 3 2 x 1 6 x 0 8 mm 5182 0436 Options 16200B 001 Add Working Std Set Compensation and Measurement When using the 4291B follow these instructions Perform open short load and low loss calibration at the 7 mm test port of the 4291B Connect the 16200B to the 7 mm test port and connect the test fixture onto the 16200B Open short and load com pensations are recommended before measurement Use the short bars and 51 O SMD resistors furnished with 16200B 001 to perform short and load compensation respectively When using other instruments follow these instructions Connect the 16200B to the 7 mm test port of the measurement instru ment Perform open short load and low loss calibration at the 7 mm test port of the 16200B Then connect the test fixture onto the 16200B and perform open short and electrical length compensations in the usual manner 49 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec d
95. sories Description P N Oty Operation and Service Manual 16196 90040 Insulator Assembly 1 34 mm 16196 60112 Insulator Assembly 1 14 mm 16196 60113 Insulator Assembly 1 08 mm 16196 60114 Open Plate 16196 29002 j Short Plate 16196 29026 1 Push Ring 16196 24004 Magnifying Lens 16193 60002 1 Tweezers 8710 2081 Wrench 8710 0909 Cleaning Rod 5182 7586 Carrying Case 16196 60150 1 Opt 16196A 710 only Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD Options 16196A 710 Add the magnifying lens and tweezers To maintain adeguate measurement performance keep the elec trodes and the short plate in good condition Contaminants and abrasion on these parts considerably affect measurement results especially for low value measurements Periodic fixture cleaning and part replacement is recommended to avoid deterioration of measure ment performance The 16196x fixtures are designed with simplicity in mind so that an operator can easily replace parts Spare parts which are likely to be abraded are supplied with the 16196U Maintenance Kit 16196U Maintenance Kit Opt 16196U 010 Upper electrode 5 piece set common to 16196A B C models Opt 16196U 100 Short plate for 0603 inch 1608 mm size 5 piece set for 16196A Opt 16196U 110 Lower electrode 5 piece set for 16196A Compensation and Measurement First of all install the appropriate in
96. sulator assembly into the fixture Then perform compensation Open and short compensations are recommended in combination with the electrical length compensation before measurement The fixture s electrical length must be entered into the electrical length compensation function of the measurement instrument first Next open compensation is performed by placing the furnished open plate on top of the insulator assembly Short compensation is performed by placing the furnished shorting plate on top of the insulator assembly After performing open and short compensations in combi nation with the electrical length compensation the DUT is inserted into the test fixture Once the measurement of the DUT is complete remove the DUT from the fixture by using the furnished push ring The following figures show how compensation and measurement Is performed 38 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD Compensation and Measurement Removing a DUT 39 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 161968 Parallel Electrode SMD Test Fixture Terminal Connector mm DUT Connection 2 Terminal Electrical Length 26 9 mm Dimensions approx 140 W x 48 H x 78 D mm Weight approx 250 g Addi
97. sults especially for low value measurements Periodic fixture cleaning and part replacement is recommended to avoid deterioration of measure ment performance The 16196x fixtures are designed with simplicity in mind so that an operator can easily replace parts Spare parts which are likely to be abraded are supplied with the 16196U Maintenance Kit 16196U Maintenance Kit Opt 16196U 020 Upper electrode 5 piece set for 16196D Opt 16196U 400 Short plate for 01005 inch 0402 mm size 5 piece set for 16196D Opt 16196U 410 Lower electrode 5 piece set for 16196D Compensation and Measurement First of all install the appropriate insulator assembly into the fixture Then perform compensation Open and short compensations are recommended in combination with the electrical length compensation before measurement The fixture s electrical length must be entered into the electrical length compensation function of the measurement instrument first Next open compensation is performed by placing the furnished open plate on top of the insulator assembly Short compensation is performed by placing the furnished shorting plate on top of the insulator assembly After performing open and short compensations in combi nation with the electrical length compensation the DUT is inserted into the test fixture Once the measurement of the DUT is complete remove the DUT from the fixture by using the furnished push ring Refer to the 16196A figures
98. t will show the variations in the measured impedance of short condition See figure below The final step to determine an approximation of short repeatability is to add a margin to the values obtained For open repeatability measure the admittance of the test fixture s open condition In the same way deter mine open repeatability by measuring at least 50 times Specifications of Short Repeatability Actual Measurement nL of Short Repeatability Measurement of short repeatability 16034G 69 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Additional Error Measurement Settings Measurement Instrument 4294A Measurement Freguency 40 Hz 10 MHz Measurement Parameter Z 0 Compensation Performed short Compensation Bandwidth ad Measurement Method Inserted the shorting plate measured the short condition and then removed the short ing plate Repeated this for 50 times Display Method Overlaying traces by using the Accumulate Mode Lastly of all a method to visually analyze the accurate measurement range of a test fixture is introduced This method is only appropriate when all three error terms proportional error open and short repeatability are known The table below shows the additional error of 16034G The whole eguation with all three terms can be solved for measurement impedance rather than additional error
99. tandard resistor set consisting of nine precision resistor standards which range from 1 mQ to 100 kQ These resistors are primarily used for performance tests of Keysight s 4 Terminal Pair LCR meters and impedance analyzers Refer to the instrument s operation service manual for the method of using these standards Applicable Instrument 4263B 4268A 4284A 4294A 4339B E4980A E4981A denotes the instrument is obsolete Model DC Resistance 42031A 1 mQ 0 2 42032A 10 mQ 0 2 42033A 100 mQ 0 2 42034A 1020295 42035A 100 5 0 196 42036A 100 O 2 0 196 42037A 1 KQ 0 1 42038A 10kO 0 1 42039A 100 kO 2 0 1 Furnished Accessories Description 42031A 1 mQ 42032A 10 mQ 42033A 100 mQ 42034A 1 O 42035A 10 O 42036A 100 O 42037A 1 kO 42038A 10 kO 42039A 100 kQ Calibration Report Carrying Case Operating and Service Manual P N N A N A N A N A N A N A N A N A N A N A 42030 60100 42030 90001 Qty _ www datatec de Other Accessories 42090A Open Termination Description The 42090A is an open termination and is primarily used for performance tests of Keysight s 4 Terminal Pair LCR meters and impedance analyzers Refer to the instrument s operation ser vice manual for the method of using this standard Applicable Instrument 4263B 4268A 4279A 4284A 4285A 4288A 4294A E4980A E4981A den
100. tion It is provided with a BNC female con a nector board to allow the attachment of user fabricated test fixtures Applicable Instruments 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A denotes the instrument is obsolete 1 Requires Option 006 Frequency DC to 30 MHz Maximum Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C Furnished Accessories Terminal Connector 4 Terminal Pair BNC Cable Length approx Description P N Qty 1 89 m from connector to cable tip Terminal Board with BNC f x4 16032 60001 1 Cable Tip BNC male Operating Manual 16048 90031 1 Weight approx 460 g Additional Error For detailed information refer to the measurement Compensation and Measurement Cable length compensation is instrument s specifications recommended before measurement Set the instrument s cable length compensation function to 2 m 20 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Port Cable Extension 16048E Test Leads Description The test leads extend the measurement port with a 4 Terminal Pair configuration It is provided with a BNC female con nector board to allow the attachment of user fabricated test fixtures Applicable Instruments 4263B 4284A E4980A denotes the instrument is obsolete 1 Reguires Option 006 Freguency DC to 2 MHz Maxim
101. tional Error Type of Error Impedance Proportional Error 1 0 x f 90 Open Repeatability 5 40xf uS Short Repeatability 30 125 x f mQ f frequency GHz Description This test fixture is designed for impedance evaluations of parallel electrode SMDs It achieves stable frequency characteris tics up to 3 GHz and provides highly repeatable measurements The applicable SMD size code is 0402 inch 1005 mm Applicable Instrument 4287A E4982A 4294A 42942A E5061B 3L5 with Opt E5061B 005 16201A E4991A When used with 16085B 4263B 4268A 4279A 4284A 4285A 4288A E4980A E4981A denotes the instrument is obsolete Frequency DC to 3 GHz Maximum Voltage 42 V peak max AC DC Operating Temperature 55 C to 85 C DUT size The applicable SMD size is 0402 inch 1005 mm For details see the figure below The 16196B is furnished with three different insulator assemblies since any gaps between the DUT and the cylindrical insulator will CUT Inserti Role Giese DET a Dei O Ee 3109 Eisen Eien Insulator Assembly result in improper positioning and subsequent measurement errors Select an insulator assembly that reduces the gap the most See the table below for dimensions of the insulator assemblies Hole Diameter of Insulator assembly mm SMD case size examples Length Width Height mm 16196B 0 0 85 1 0x0 5x0 5 00 5 LOx0 5x0 35 0 0 68 1 0x0 5x0 35 Furnished Accessories Des
102. ts 16117B C Material 16008B Other Accessories Miscellaneous 16190B 16380A C 42060A 42090 1A 7 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Test Fixtures 4 Terminal Pair for Impedance Measurements up to 110 MHZ Freguency Range DC 1k 1M 10M 100M 16047A 13M 16047E 110M 16060A 100 k 16034E 40M 16034G H 110M 16044A 10M 16334A 5 15M 16089A B C D 5 100 k 16048A D 30M 16048E 2M 16048G H 42941A 40 16065A 50 2M 16065C 100 1M 16451B 16452A 20 Up to 110 MHz Terminal Configuration 4 Terminal Pair Up to 3 GHz Terminal Configuration 7 mm Freguency Range Measurement 4263B 4268A 4279A 4284A Instruments 4285A 4288A 4294A E4980A E4981A 4291B 4294A 42942A E5061B 3L5 w Opt 005 16201A 4287A E4982A E4991A Any 4TP instruments excluding 4294A 16085B Freguency DC High Resistance Measurement Measurement 4339B 4349B Instruments Bm dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de 9 a Up to 110 MHz 4 Terminal Pair Lead Components 16047A Test Fixture Terminal Connector 4 Terminal Pair BNC DUT Connection 4 Terminal Dimensions approx 124 W x 31 H x 62 D mm Weight approx
103. ulator assembly After performing open and short compensations in combination with the electrical length com pensation the DUT is inserted into the test fixture Once the mea surement of the DUT is complete remove the DUT from the fixture by using the furnished push ring Refer to the 16196A figures to see how compensation and measurement is performed 43 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 3 GHz 7 mm SMD 16196D Parallel Electrode SMD Test Fixture Terminal Connector 7 mm DUT Connection 2 Terminal Electrical Length 27 3 mm Dimensions approx 140 W x 48 H x 78 D mm Weight approx 250 g Additional Error Type of Error Proportional Error Open Repeatability ohort Repeatability f frequency GHz 44 a m dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de Impedance 1 0 x f 96 5 40 xf uS 30 125 x f mQ Description This test fixture is designed for impedance evaluations of parallel electrode SMDs It achieves stable frequency characteris tics up to 3 GHz and provides highly repeatable measurements The applicable SMD size code is 01005 inch 0402 mm Applicable Instrument 4287A E4982A 4294A 42942A E5061B 3L5 with Opt E5061B 005 16201A E4981A denotes the instrument is obsolete Frequency DC to
104. um Voltage 42 V peak max AC DC Operating Temperature 0 C to 55 C Furnished Accessories Terminal Connector 4 Terminal Pair BNC Description Hh oh Cable Length approx 3 8 m from connector to cable tip Terminal Board with BNC f x4 16032 60001 1 Cable Tip BNC male Operating Manual 16048 90041 1 Weight approx 690 g po Additional Error For detailed intonation erento the Compensation and Measurement Cable length compensation is messurementinsirureni sspeciheations recommended before measurement Set the instrument s cable length compensation function to 4 m 21 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Up to 110 MHz 4 Terminal Pair Port Cable Extension 16048G Test Leads r aX Terminal Connector 4 Terminal Pair BNC Cable Length approx 1 m Cable Tip BNC female Weight approx 460 g Additional Error For detailed information refer to the operation manual or the specifications of 4294A 16048H Test Leads a Terminal Connector 4 Terminal Pair BNC Cable Length approx 2 m Cable Tip BNC female Weight approx 690 g Additional Error For detailed information refer to the operation manual or the specifications of 4294A Description The test leads extend the measurement port with a 4 Terminal Pair configuration It is provided with a BNC male connec
105. uring circuit The following measurement instruments are equipped with electrical length compensation function Model Compensation Function 4287A Electrical Length Compensation Port Extension 4291B Electrical Length Compensation Port Extension 4294A 42942A Electrical Length Compensation Port Extension E5061B 3L5 Option 005 16201A Electrical Length Compensation Port Extension E4982A Electrical Length Compensation Port Extension E4991A Electrical Length Compensation Port Extension denotes the instrument is obsolete The electrical length is specified for the following test fixtures Model Electrical Length 16092A 3 4 MM 16192A 11 0 mm 16194A 50 0 mm 16196A 26 2 mm 16196B 26 9 mm 16196C 2 1 mm 16196D 27 3 mm 16197A 14 0 mm 72 E dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 5050 Fax 07121 51 50 10 info datatec de www datatec de Appendix Compensation 4 Cable Length Compensation When the test sample is measured with an instrument having a Four Terminal Pair configuration the additional length of the test lead extension between the instrument and the test sample in conjunction with the measurement freguen cy influences the amplitude and phase of the signal being measured This must be taken into account particularly for measurements performed at freguencies of 100 kHz or higher since the modification of the amplitude and phase of the signal can cause the internal measurement
Download Pdf Manuals
Related Search
Related Contents
WD VelociRaptor® Internal Hard Drive Quick Install Bol d`air Jacquier Philips HD 5410/00 Coffeemaker JOURNAL OFFICIEL DE LA REPUBLIQUE ALGERIENNE N° 32 10 Precor C954, C956 Treadmill Operation Manual Télécharger - Rhone Alpes Desinfection bong forte suspensión User`s Manual - MCM Electronics eKワゴン 30俺 Copyright © All rights reserved.