        P7240 4 GHz Active Probe Service Manual
         Contents
1.        Calibration  step generator    A    SMA Cable             Trigger input       Figure 8  Test system and probe rise time setup    10  Set the step generator control switch to on     11  On the TDS8000  expand the horizontal scale to locate the step edge  Set the  vertical scale to 10 mV div  and adjust horizontal range to 100 ps div while  maintaining the edge view  For a more stable measurement display  turn  averaging on     P7240    4 GHz Active Probe Service Manual    Peformance Verification       12  Adjust the TDS8000 horizontal and vertical position controls to display a  signal similar to that shown in Figure 8        NOTE  Do not touch the probe tip adapter when making calibration measure   ments  Measurement accuracy is degraded when the probe tip adapter is  handled        13  Use the TDS8000 measurement capability to display rise time  Rise time is  determined from the 10  and 90  amplitude points on the waveform   Record the rise time as f 4     14  Calculate the probe only rise time using the following formula     2 2  to   vlgip     ts    15  Check that the calculated rise time meets the probe specification  The rise  time  t  must be  lt  120 ps     16  Record the results on the test record     P7240 4 GHz Active Probe Service Manual 13    Peformance Verification       Test record    Probe Model   Serial Number        Certificate Number        Temperature   RH          Date of Calibration        Technician        Performance test    Output zero voltage  
2.        Symbols and Terms    A  ZN    Terms in this Manual  These terms may appear in this manual        WARNING  Warning statements identify conditions or practices that could result  in injury or loss of life           CAUTION  Caution statements identify conditions or practices that could result in  damage to this product or other property        Terms on the Product  These terms may appear on the product     DANGER indicates an injury hazard immediately accessible as you read the  marking     WARNING indicates an injury hazard not immediately accessible as you read the  marking     CAUTION indicates a hazard to property including the product     Symbols on the Product  The following symbols may appear on the product     A A ao 8        CAUTION WARNING Double Protective G round Not suitable for  Refer to Manual High Voltage Insulated  Earth  Terminal connection to    the public telecom   munications network    P7240 4 GHz Active Probe Service Manual    UN Z ON 2   Preface    This is the service manual for the P7240 4 GHz Active Probe  Read this preface  to learn how this manual is structured and where you can find other information  related to servicing this product     Manual Structure    This manual contains three sections     Description  Performance Verification   and Adjustment Procedures  and is intended to be used by qualified service  personnel  Replaceable parts are limited to accessories and adapters  and are  described in the Instruction Manual that is shipped wi
3.     16 P7240 4 GHz Active Probe Service Manual    Adjustment Procedures       If an error message appears  click OK  correct the condition that caused the  error message  and then click Continue     4  Proceed to the Ofjset Zero procedure        NOTE  To avoid corrupted test results  do not disconnect the probe under test or  the probe calibration adapter until you have completed the service procedures  lf  the probe or adapter are disconnected before completing the procedures  you  must reconnect the probe and adapter  and exit and restart the application        Offset Zero    Use this procedure to adjust the probe offset zero     1  Connect the test eguipment as shown in Figure 10           Digital TDS7404 Oscilloscope  multimeter  KlipChip  Setofiset switch adapters  to GND connected  together             BNC to Dual  Banana Adapter            Y lead  adapter   TekConnect   Calibration adapter    50 Q Precision  terminator                                              f    BNC SMA adapter             P 7240 Probe    Figure 10  Setup for offset zero    2  Set the offset switch on the calibration adapter to GND     3  Click Offset Zero in the application window  See Figure 11 on page 18     P7240 4 GHz Active Probe Service Manual 17    Adjustment Procedures       ia  P7240 Adjustments v1 0    Circuit to Adjust                Control    Adjust slider to set Q Offset Zero  offset voltage on Store  DMM to 0 0000 V Setting    3969    Index value covers a i Channel Oliset       42
4.     The probe rise  separately  and    time is calculated using rise times measured from the test system  the test system including the probe     1  Connect the test equipment as shown in Figure 7 on page 11        CAUTION  To prevent damage  use care when working with SMA connectors   support equipment to avoid mechanical strain on the connectors  and use a  torque wrench when tightening connections        10    P7240 4 GHz Active Probe Service Manual    Peformance Verification       TDS7404 Oscilloscope                         0    vomm  com 0 0    00000  000 00  ool   on on0 0  o o MODO    Setofiset switch       o  0  o  0  o  0  O                                 Seno   to GND  SMA TekConnect  Sampling head M to M Calibration adapter  extender cable ee  adapter TekConnect to  Calibration  au Adapter step generator  HEH                                  X    Generator remote  head       CSA8000 TDS8000 Trigger input             SMA cable                Internal  clock output          Figure 7  Test system rise time setup    2  Set the oscilloscope trigger to internal clock     3  Select the channel you have connected to on the 80E0X sampling head  and  then set the oscilloscope vertical scale to 50 mV div        NOTE  The output of the step generator rises from a    250 mV level to ground        4  Adjust the oscilloscope horizontal and vertical position controls to display a  signal similar to that shown in Figure 7     5  Set the oscilloscope horizontal scale to 50 ps div 
5.  adapter the same way as you do the    P7240 Probe        Figure 1  TekConnect to SMA adapter    Probe Tip Adapter The probe tip adapter  Tektronix part number 015 0678 00  provides a low noise  method for connecting the P7240 to signals present on SMA cables  The adapter  has a test point for the probe and two SMA connectors  A 50 ohm termination is  included with the adapter  and is connected to the SMA connector nearest the  probe test point to minimize relections  Connect the cable from the test circuit to  the other SMA connector  See Figure 2        NOTE  When taking measurements  do not touch the probe tip adapter  Measure   ment accuracy is degraded when the probe tip adapter is handled        SMA connector for test circuit       Probe test point    Terminator    Figure 2  Probe tip adapter    P7240 4 GHz Active Probe Service Manual    Description       TekConnect Interface  Calibration Adapter    The TekConnect Interface Calibration Adapter  Tektronix part number  067 0422 00  connects between the host instrument and the probe under test  and  provides connectors for probe signal and offset voltage measurements    See Figure 3        Figure 3  TekConnect Interface Calibration Adapter    When the adapter is connected to the oscilloscope  the adapter is identified as a  valid calibration device  However  additional power supplies necessary to power  the probe are not enabled until a TekConnect probe is connected to the adapter  and identified by the oscilloscope  When 
6.  at probe output       Minimum   Incoming   Outgoing   Maximum                 10 mV  20   C to 30   C   10 mV  DC attenuation accuracy    error   lt 2    Rise time  lt 120 ps  14    P7240 4 GHz Active Probe Service Manual    r  Adjustment Procedures    These instructions are for personnel who are familiar with servicing the product   If you need further details for disassembling or reassembling the product  refer to  the appropriate product manual  Contact your nearest Tektronix  Inc   Service  Center or Tektronix Factory Service for installation assistance        CAUTION  To prevent static discharge damage  service the product only in a  A static free environment  Observe standard handling precautions for static sensi    tive devices when servicing the probe  Always wear a grounded wrist strap    grounded foot strap  and static resistant apparel when servicing the probe        Equipment Setup    Use this procedure to set up the equipment to adjust the probe     1     2  3   4    m    Turn on the oscilloscope and enable the channel you intend to use   Connect the probe calibration adapter to the oscilloscope   Connect the probe to the probe calibration adapter     Verify that the Volts div readout on the oscilloscope channel displays      the oscilloscope recognizes the probe through the adapter      Set the multimeters to read DC volts   Allow 30 minutes for the equipment to warm up     While the equipment is warming  verify that the application has been loaded  onto the os
7. 0 mV range    Click Store Setting ve Lo      after voltage is set    Channel Offset is Overview   disabled in this step    Status  Ready to modify Offset Zero values       Figure 11  Offset Zero window    4  Set the offset zero voltage on the DMM to 0 0000 V by adjusting the Offset  Zero Control  The index value in the Offset Zero Control varies from 0 to  3969  and corresponds to approximately  20 mV of offset voltage range     Use one of the following methods to make the adjustment   m Use your finger or the mouse to drag the on screen slider up or down     m Click or touch the up down arrow keys above and below the slider   Each click adjusts the offset voltage by about 10 uV     m To increment by ten clicks or about 100 uV  click in the bar above or  below the slider     18 P7240 4 GHz Active Probe Service Manual    Adjustment Procedures       5  When the offset zero voltage displayed on the voltmeter is 0 0000 V   0 0001 V  click Store Setting to save the offset zero voltage level        NOTE  In the Summary Info box  the After field updates with the control value  from step 5  This is a temporary value until Accept is selected later  after all  adjustments have been made     If you select Accept now  the application requests confirmation  If you click Yes   the application exits  if you click No  you are able to make further adjustments   You must complete all adjustments before selecting Accept        6  Proceed to the Offset Gain procedure     Offset Gain    Use this 
8. NC SMA       gt i adapter  adapter    Digital multimeter       Power supply                           zi    P7240  Probe       KlipChip  adapters                      Y lead adapter    Figure 6  Setup for DC attenuation accuracy    Record the results of steps 2 through 5 separately  Use the results to calculate the  DC accuracy of the probe in step 6     2  Set the power supply to  1 000 V  Use the multimeter connected to the  power supply to verify that the DC voltage is as close to  1 000 V as  possible  Record this measurement as Vmax     3  Measure and record the reading of the other multimeter as M1     4  Set the power supply to    1 000 V  Use the multimeter connected to the  power supply to verify that the DC voltage is as close to    1 000 V as  possible  Record this measurement as Vmin     5  Measure and record the reading of the other multimeter as M2     P7240 4 GHz Active Probe Service Manual 9    Peformance Verification       6  Determine the percent error by using the formula below                               MiM2 N  PET    Vmax Vmin  x 2 i 3 100   Example    T 0 2028       02021     S oa ea a         0 4049     gajo  Z lm    1 0097 1 x 100       0 97     The calculated error should be  lt 2      7  Record the results on the test record        NOTE  An unacceptable error value may result if a low tolerance termination is    substituted for    the recommended termination        Rise Time    This procedure verifies that the probe meets rise time specifications 
9. Service Manual    Tektronix  pa    P7240 4 GHz  Active Probe    071 1056 00    Warning   The servicing instructions are for use by gualified  personnel only  To avoid personal injury  do not  perform any servicing unless you are gualified to  do so  Refer to all safety summaries prior to  performing service     www tektronix com    ONO  071105600    Copyright    Tektronix  Inc  All rights reserved  Licensed software products are owned by Tektronix or its suppliers and  are protected by United States copyright laws and international treaty provisions     Use  duplication  or disclosure by the Government is subject to restrictions as set forth in subparagraph  c  1  ii  of the  Rights in Technical Data and Computer Software clause at DFARS 252 227 7013  or subparagraphs  c  1  and  2  of the  Commercial Computer Software     Restricted Rights clause at FAR 52 227 19  as applicable     Tektronix products are covered by U S  and foreign patents  issued and pending  Information in this publication supercedes  that in all previously published material  Specifications and price change privileges reserved     Tektronix  Inc   P O  Box 500  Beaverton  OR 97077    TEKTRONIX  TEK  TekConnect  and KlipChip are registered trademarks of Tektronix  Inc     r  General Safety Summary    Review the following safety precautions to avoid injury and prevent damage to  this product or any products connected to it  To avoid potential hazards  use this  product only as specified     Only qualified p
10. a probe is detected through the adapter   the Volts div readout on the oscilloscope displays        Refer to Table 2 on page 5 for features of the calibration adapter     P7240 4 GHz Active Probe Service Manual    Description       Table 2  TekConnect Interface Calibration Adapter features    Feature    Latch button    Latch    QO    Description    Latch button  The spring loaded latch mechanically retains the adapter to the oscilloscope  To  release the adapter  grasp the adapter housing  depress the latch button  and pull the adapter  straight out of the oscilloscope        Offset AS  switch   2S   K  EA VAR         Offset output select switch  The offset output switch selects between ground and the offset  voltage level from the oscilloscope     Leave the switch in the ground position for the performance verification procedures     Move the switch to the variable position for the adjustment procedures              PS Offset  voltage    output    Signal  out       Theory of Operation    P7240 4 GHz Active Probe Service Manual    Offset voltage  The offset voltage of the probe is accessed through the BNC connector     Measure the offset voltage using a DVM  BNC coaxial cable  and BNC to dual banana jack     Signal out  The SMA connector on the rear of the box allows for direct monitoring of the probe  signal     Refer to Figure 4 as you read the following information     There are two adjustments in the P7240 adjustment procedure  Offset Zero and  Offset Gain  These adjustme
11. and center the waveform     6  Use the oscilloscope measurement capability to display rise time  Increase  the stability of the pulse edge measurement by using averaging  if available   Rise time is determined from the 10  and 90  amplitude points on the  waveform  Record the rise time as ts    The system rise time       that you measured in step 6 represents the rise time of  the test system without the probe     P7240 4 GHz Active Probe Service Manual 11    Peformance Verification       12    After you assemble the test setup that includes the probe  as shown in Figure 8   you will measure the system and probe rise time     4   in step 13  This is used to  calculate the probe rise time  t  in step 14     7  Set the step generator control switch to standby     8  Remove the TekConnect SMA adapter from the test setup     9  Connect the test setup as shown in Figure 8 on page 12        CAUTION  To prevent damaging the SMA connectors  use a 7 32 inch wrench  when connecting and disconnecting the female to female SMA adapter        TDS7404 Oscilloscope                    Sampling head  extender cable    80E 0X       SMA  M to M  adapter TekConnect  calibration adapter          50 Q Terminator  ea Probe tip adapter             Jie                            SMA F to F adapter                    ko   P7240 Probe    CSA8000 TDS 8000                                                                   Internal  clock output          FOO   o FOOT     ee Generator remote head            
12. cilloscope     a  Minimize the scope application window by going to File gt Minimize   b  Go to  Start gt Programs   Tektronix   TekConnect Adjust  c  Ifthe application is not loaded on the oscilloscope  get the application     Retreive the application CD that came with the oscilloscope  insert the  CD in the drive  select Start    Run  and then enter D  setup exe    Or  download the application from the Tektronix website at   www tektronix com Software and Drivers Accessories    When the application starts  the display appears as shown in Figure 9     P7240 4 GHz Active Probe Service Manual 15    Adjustment Procedures           TekConnect Adjustment Application v1 0    Tektronix    Instructions     Channel to Use    Task to Perform  Adjustment       Figure 9  TekConnect probe adjustment application    Application Setup  1  Make the following selections in the application dialog box   a  Select P7240 in the Probe Type to Adjust list     b  Select the oscilloscope channel that the adapter is connected to in the  Channel to Use list     c  Select Adjustment in the Task to Perform list   2  Click Continue     3  The P7240 Probe Adjustments Dialog Box appears  See Figure 11 on  page 18        NOTE  An error message appears if one of the following conditions exists        m A probe is not connected to the adapter   m A probe model other than a P7240 is connected to the adapter     m The P7240 connected to the adapter is defective  unable to properly  communicate with the adapter  
13. el Offset control to display 0 5 V on DMM  2   Use one of the following methods     m Use your finger or the mouse to drag the on screen slider up or down     m Click or touch the up down arrow keys above and below the slider   Each click adjusts the offset voltage by about 1 mV     m To increment by ten clicks or 10 mV  click in the bar above or below the  slider     Record this voltage as M2     20 P7240 4 GHz Active Probe Service Manual    Adjustment Procedures       6  Determine the offset gain target value by using the formula below     Offset gain target value   Vi   M2    7  The Offset Gain voltage is displayed on DMM  1  Adjust the Offset Gain  Control to the offset gain target value   1 mV  using one of the following  methods     m Use your finger or the mouse to drag the on screen slider up or down     m Click or touch the up down arrow keys above and below the slider   Each click adjusts the offset voltage by about 1 mV     m To increment by ten clicks or 10 mV  click in the bar above or below the  slider     8  When the offset gain displayed on DMM  1 is within  1 mV of the offset  gain target value  click Store Setting to save the offset gain level     In the Summary Info box  the AFTER field updates with the control value from  step 8  This is a temporary value until Accept is selected     9  Click Accept to save all the values that you set in these procedures   10  Click Yes to update the contents     This completes the adjustments to the probe     Verify Opera
14. ersonnel should perform service procedures     Do Not Service Alone  Do not perform internal service or adjustments of this  product unless another person capable of rendering first aid and resuscitation is  present     While using this product  you may need to access other parts of the system  Read  the General Safety Summary in other system manuals for warnings and cautions  related to operating the system     To Avoid Fire or Connect and Disconnect Properly  Do not connect or disconnect probes or test  Personal Injury leads while they are connected to a voltage source     Observe All Terminal Ratings  To avoid fire or shock hazard  observe all ratings  and markings on the product  Consult the product manual for further ratings  information before making connections to the product     Do not apply a potential to any terminal  including the common terminal  that  exceeds the maximum rating of that terminal     Do Not Operate Without Covers  Do not operate this product with covers or panels  removed     Do Not Operate With Suspected Failures  If you suspect there is damage to this  product  have it inspected by qualified service personnel     Do Not Operate in Wet Damp Conditions   Do Not Operate in an Explosive Atmosphere   Keep Product Surfaces Clean and Dry     Provide Proper Ventilation  Refer to the manual s installation instructions for  details on installing the product so it has proper ventilation     P7240 4 GHz Active Probe Service Manual i    General Safety Summary
15. nizes the probe through the adapter      Set the multimeters to read DC volts     Allow 30 minutes for the equipment to warm up     P7240 4 GHz Active Probe Service Manual 7    Peformance Verification       Output Zero  Use this procedure to verify the probe output zero   1  Connect the test eguipment as shown in Figure 5     2  Set the offset switch on the calibration adapter to GND        NOTE  Leave the offset switch in the ground position for all of the performance  verification checks        3  Ground the probe tip by connecting the probe tip to the probe ground socket    Connecting two KlipChip adapters together is recommended      4  Observe the multimeter display  and record the results on the test record   The displayed DC level should be 0 000 V  10 mV     Digital TDS7404 Oscilloscope  multimeter                            BNC to Dual KlipChip  Banana Adapter 50 O Precision adapters ae  terminator TekConnect    z    Calibration adapter  BNC cable               BNC SMA adapter                               P7240 Probe          Figure 5  Setup for output zero    8 P7240 4 GHz Active Probe Service Manual    Peformance Verification       DC Attenuation Accuracy    1  Disconnect the KlipChip adapters  and connect the setup as shown in  Figure 6     Digital TDS 7404 Oscilloscope  multimeter              Setoffset switch  to GND                 BNC to Dual  Banana Adapter                                  BNC cable  50 Q Precision  terminator TekConnect       calibration  B
16. nput signal at the probe input and no offset signal  from the host oscilloscope  the Offset Zero is adjusted to produce 0 0000 V at the  probe output  Since the P7240 probe is designed to operate into a host oscillo   scope with a 50 Q load  a precision 50 Q feed thru terminator is used in the  adjustment procedure     The Offset Gain adjustment is used to adjust the offset amplifier gain for a  precise 1X scale factor  Since the P7240 input signal scale factor has been  designed with a 5X scale factor  the Offset Gain is adjusted to null out the  difference between a DC input signal and a 5X scaled offset signal from the host  oscilloscope     6 P7240 4 GHz Active Probe Service Manual    ll A OS  Performance Verification    Use the following procedures to verify the warranted specifications of the P7240  probe  Before beginning these procedures  refer to page 14 and photocopy the  test record and use it to record the performance test results  The recommended  calibration interval is one year  These procedures test the following specifica   tons     m Output zero    m DC attenuation accuracy    um Rise time    Eguipment Setup    Use this procedure to set up the eguipment to test the probe     1     2  3   4    ni    Connect the probe calibration adapter to the oscilloscope   Connect the probe to the probe calibration adapter   Turn on the oscilloscope  and enable the channel     Verify that the Volts div readout on the oscilloscope channel displays      the oscilloscope recog
17. nts are made using digitally controlled potentiome   ters that are controlled by a two wire serial interface from the host oscilloscope   Because the adjustments are digitally controlled  the probe does not need to be  disassembled as part of the adjustment procedure  The software application  provides the control needed to make the adjustments to the probe     Description                                                                               Probe head Compensation box Oscilloscope  Probe tip  b Signal Out  Probe tip  amplifier    EEPROM Probe ID Out    Clock In  Offset ampifier  Ne gt  1 V Offset  Offset to probe Offset     NAA   8 O zero     V     e  se  Ofiset  gain   7V  Ct poe  Linear  regulator  15 V  V   5V  na id Ground  Probe TekConnect  cable Interface    Figure 4  P7240 simplified schematic diagram    The offset signal from the oscilloscope extends the dynamic range of the P7240  probe by compensating for some of the DC voltage that may be part of the signal  applied to the probe  Compensation is accomplished by summing the offset  signal with the probe input signal at the probe input attenuator  as shown in  Figure 4  The host oscilloscope supplies the offset signal  with a range of  1  volt  to the offset amplifier in the probe  Since the offset scale factor of the  P7240 probe is 5X  the effective offset signal range is increased to  5 V     The Offset Zero adjustment is used to null out any amplifier voltage offsets in  the offset signal path  With no i
18. on    Use the performance verification procedures to verify the warranted specifica   tions of the P7240 probe  The recommended calibration interval is one year  No  software is required to complete a performance verification of the probe     The performance verification procedures check the following specifications   m Output zero  m DC attenuation accuracy    um Rise time    Adjustments    The P7240 probe has two electronically adjusted controls that rarely need  adjusting unless a performance verification yields a specification that is out of  tolerance  or the probe fails an oscilloscope probe calibration utility     If you must make adjustments to the probe  you need the TekConnect Adjust   ment Software for the P7240 4 GHz Probe  Refer to Eguipment Setup in the  Adjustment Procedures  step 7  for instructions on finding and loading the  adjustment software on your instrument     The adjustment procedures describe the following adjustments for the probe   um Offset zero    m Offset gain    P7240 4 GHz Active Probe Service Manual 1    Description       Eguipment Reguired    Refer to Table 1 for the eguipment reguired to perform the service procedures     Table 1  Eguipment reguired for performance verification and adjustment procedures    item description  TekConnect adjustment software     Sampling Oscilloscope    Sampling head  with extension cable    Oscilloscope    Performance requirement  P7240 Probe     gt 12 5 GHz bandwidth   gt 12 5 GHz bandwidth  TekC onnect Inte
19. procedure to adjust the probe offset gain     1  Set the DC power supply to    2 5 volts  Use a DMM to measure the voltage   Record this voltage as M1     2  Connect the test equipment as shown in Figure 12     Digital   Digital  multimeter TDS 7404 Oscilloscope multimeter     2        1 DC Power Supply                                         BNC to Dual adapters  Banana Adapter Y lead  BNC cable adapter  TekC onnect  Calibration adapter    Set offset switch  to VAR    BNC to Dual  Banana Adapter    Z    BNC cable       50 O Precision  terminator                                                 BNC SMA adapter P7240 Probe  Figure 12  Setup for offset gain    3  Set the offset switch on the calibration adapter to VAR     4  Click OFFSET GAIN in the application window  See Figure 13 on page 20     P7240 4 GHz Active Probe Service Manual 19    Adjustment Procedures       ia  P7240 Adjustments v1 0    Circuit to Adjust Offset Gain Control    a   Min    Adjust slider to set  offset gain on DMM   1 to value calcu   lated in step 6       Offset Zero       Click Store Setting  after gain and offset    are set   Adjust slider to set 2067   2067    DMM  2 to 0 5 V  49   49    Status  Ready to modify Offset Gain values       Figure 13  Offset Gain window       NOTE  Channel offset is multiplied by the 5X attenuator in the probe  The offset  voltage    0 5 V  measured at the BNC connector on the Calibration Adapter  nulls out the    2 5 V applied to the probe input        5  Set the Chann
20. rface    Recommended example    Optional Applications for Windows Based  Oscilloscopes CD    Tektronix TDS8000  Tektronix 80E0X with 012 1568 00 cable  Tektronix TDS 7404          TekConnect calibration adapter TekC onnect Interface 067 0422 00  Calibration Step Generator 250 mV step   lt 30 ps rise time 067 1338 0X  Adapter TekConnect to SMA TCA SMA  Adapter Probe tip with 50 O terminator 015 0678 00  DC Power Supply 1VDC atl mA Tektronix PS280  DMM  2  with leads 0 05  accuracy  0 1 mV resolution Fluke 87 or equivalent  Feedthrough Termination 50 Q  0 05 Q 011 0129 00  BNC to BNC coaxial cable  2  50 O coaxial cable 012 0057 01  Coaxial cable Male to Male SMA 012 0649 00  Adapter SMA Male to Male 015 1011 00  Adapter SMA Female to F emale 015 1012 00  Adapter SMA Male to BNC Female 015 0554 00  Adapter  2  BNC Female to Dual Banana 103 0090 00  Adapter Y lead adapter 196 3456 00  Adapters  2  KlipChip adapter 206 0364 00  SMA torgue wrench 5 16 in  7 in lb    SMA adapter wrench 7 32 in       1 Required for adjustment procedures only    Special Adapters Required    Some of the adapters listed in Table 1 are custom made and available only from  Tektronix  The adapters are described on the following pages     P7240 4 GHz Active Probe Service Manual    Description       TekConnect to SMA The TekConnect to SMA Adapter  Tektronix part number TCA SMA  allows  Adapter signals from an SMA cable or probe to be connected to a TekConnect input   See Figure 1  Connect and disconnect the
21. th the probe     Be sure to read the introductions to all procedures  These introductions provide  important information needed to do the service correctly  safely  and efficiently     Related Documentation    The probe is shipped with the following manual     m P7240 4 GHz Active Probe Instruction Manual  Tektronix part number  071 0759 XX     P7240 4 GHz Active Probe Service Manual iii    Preface       Contacting Tektronix    Phone    Address    Web site    Sales support    Service support    Technical support    1 800 833 9200     Tektronix  Inc    Department or name  if known   14200 SW Karl Braun Drive  P O  Box 500   Beaverton  OR 97077   USA    www tektronix com    1 800 833 9200  select option 1     1 800 833 9200  select option 2     Email  support  tektronix com  1 800 833 9200  select option 3   6 00 a m    5 00 p m  Pacific time      This phone number is toll free in North America  After office hours  please leave a  voice mail message   Outside North America  contact a Tektronix sales office or distributor  see the  Tektronix web site for a list of offices     P7240 4 GHz Active Probe Service Manual    LDS  Description    This document describes the Performance Verification and Adjustment Proce   dures for the P7240 4 GHz Probe  There are no user replaceable parts inside the  probe  If the probe fails the performance verification procedure  and adjusting the  probe cannot compensate for the failure  contact your Tektronix service center     Performance Verificati
22. tion    Do the Performance Verification procedures to verify that the probe meets  specifications     zs Endofdocument zz    P7240 4 GHz Active Probe Service Manual 21    Adjustment Procedures       22 P7240 4 GHz Active Probe Service Manual    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Philips MCD388/12 User's Manual  NAPLES DAY BED - Home Decorators Collection    高精度デジタル温湿度計データロガー付 CENTER-318  Soft-Starter Arrancador Suave Soft-Starter SSW-08    StarTech.com Micro HDMI to VGA Adapter Converter for Smartphones / Ultrabook / Tablet - 1920x1080  EZ-View  Black & Decker PF260 User's Manual  GATE CALLつなぎ方ガイド(PDF) - spaaqs(スパークス)光    Copyright © All rights reserved. 