Home
- Frank`s Hospital Workshop
Contents
1. Model table 1 1 2 3 4 5 Model table 2 Abb 2 1 Model tables Steps for which additional information is required are described hereafter in detail SpaceStation 1 0 gb Alarms and Error Codes SpaceComAlarm indicator red is on Unit Diagnosis Calibration SpaceStation without SpaceCom Neither alarms nor error codes are actually displayed on the SpaceStation without SpaceCom SpaceStation with SpaceCom SpaceCom is still starting SpaceComAlarm indicator red flashes A serious internal fault was detected in the system Carry out a device check see Device Check mp S 2 4 Table 2 1 Alarms on the SpaceCom One of the diodes of the status and alarm display line is on or flashes SpaceCover comfort A message or an alarm of a pump is dis played The status indicator on the right hand side of the On Off button flashes red System not properly configured Check configuration The status indicator on the right hand side of the On Off button lights up red A serious internal fault was detected in the system Carry out a device check see Device Check mp S 2 4 The first LED in the top charge condition display of the operator and status display field of the SpaceCover comfort flashes yellow Battery module of theSpaceCover com fort nearly discharged type pre alarm
2. Visual Inspection Electrical Safety Functional Inspection as per EN 60601 1 or VDE 0750 Part 1 and VDE 0751 Part 1 O Cleanliness completeness damage and O Mains voltage V AC Switch on unit legibility of the labels O Protective conductor resistance Q O Staff call O Pump suspension Unit incl mains connecting O Indicator lamps SPCC O Well running and sealing of the pump line to mains voltage C Audible alarm on SPCC lock output FIBx 0 20 CJ Indicator lamps SPCO O Data and power interface MSC con O Earth leakage currentatNC_____ pA O Functional inspection according to the nector lt 30u A instructions for use and dependent on O Interfaces SPCO O Earth leakage current atSFC_____ pA the configuration FJ Mains cable connector and connect lt 60u A ing line O Housing leakage current SPCO pA Metal collar Ethernet RJ45 lt 25 pA without SL Accessories Used Units Used Perfusor Space serial No O g CJ Infusomat Space serial No O Test result Inspection performed by Defects found which could endanger patients users or third parties Measures to be taken Repair Yes No Special features documentation Unit handed over on To Date Signature Next deadline M690 00 00 46 F04 3891 7548 GB SpaceStation 2 0 gb BIBRAUN SHARING EXPERTISE 7 1 Sheet 1 of 2 Technical Safety Check TSC Index b Master to be added to the documentation Fo
3. aaa 3452 1135 Housing back panel SPS aaa 3452 1089 Device bracket SPS 00000 0020 ee 3452 1127 Power supply SPS 2000002e ee eee ee 3452 1097 Connector holder mains SPS 3452 1143 Interface board SPS 02000 ee eee 3452 1100 with connectors Connector holder data SPS 3452 1160 Module lock SPS a 3452 1119 Housing SPS aaa 3452 1070 incl mains seal F1B 10 3 Spare Parts List SpaceStation with SpaceCom Abb 10 2 Exploded drawing SpaceCom 10 4 SpaceStation 2 0 gb SpaceStation 2 0 gb cC ao N O om 11 12 13 Spare Parts List Battery pack SP NIMH 008 0871 3180 Battery compartment cover SPC 3452 1232 W LAN cover SPCO ssuuuuuuuus 3452 1049 with symbol W LAN cover SPCO aaa 3452 1050 without symbol Tube guide SPS aa 3477 4394 20 pieces Pole clamp guide SPS aaa 3452 1135 Housing back panel SPCO 3452 1052 A hawt eles Res 3452 1048 Device bracket SPS Luuuuuuuuuuuuuu 3452 1127 Insulating hood SPCO aan 3452 1043 SPCO PCBs eid nanaman des 3452 1053 SPCO connecting cable aaa 3452 1042 SPCO insulating piece
4. 1 Replace the module lock UTS 3 2 Check the voltage 12 V 0 6 V 43 5 V DC at the The voltage is measured TS 4 TS 3 interface board plug contacts connection power supply interface board 3 Replace power supply UTS 4 4 Replace interface board UTS 4 Table 2 5 Trouble shooting Part 1 of 2 SpaceStation 2 0 gb Unit Diagnosis Calibration 5 Exchange processor PCB SPCC All luminous fields of the status and alarm display UTS 10 TS 6 line light up briefly during the self test All LEDs of the operator and status display field light up briefly The green LED on the operator and status display field lights up 6 Replace interface board in the SpaceStation UTS 10 7 Replace loudspeaker A short audible alarm sounds UTS 14 TS 8 8 Exchange processor PCB SPCC UTS 14 9 Replace interface board in the SpaceStation The charge state of the battery module is displayed UTS 15 TS 10 on the operator and status display field 10 Exchange the housing upper part SPCC UTS 15 11 Exchange the housing upper part SPCC The loudness changes UTS 16 TS 12 12 Exchange processor PCB SPCC UTS 16 13 Exchange processor PCB SPCC The infusion is displayed green in the status and UTS 18 TS 14 alarm display line 14 Replace interface board in the SpaceStation UTS 18 15 Exchange processor PCB SPCC The alarm is displayed red in the status and alarm UTS 20 TS 16 d
5. Abb 3 26 Legende zu Abb 3 26 ItemDesignation 1 Screw EJOT 30x9 WN 5452 TORX 10IP A2 2 Data connector holder SpaceStation 1 0 gb Disassembly Assembly SpaceStation 4 Tighten the data connector holder screw with a torque of 0 25 0 05 Nm 5 Before fitting the M 2K seals check them for damage and re place if necessary Release Button Note Check the release buttons for damage especially within the notch area before fitting the buttons 3 23 Disassembly Assembly SpaceStation Housing Back Panel 1 Insert the housing back panel on the notches at the two con nector coverings into the housing and rotate the back panel in its final position 2 No screw is inserted in the third mounting hole from the top on the right hand side of the back panel The hole must be covered with a cover cap Abb 3 27 Legende zu Abb 3 27 ItemDesignation 1 Housing notch 2 Empty mounting hole 3 11 Checks after Repair 1 Check the device to ensure safe functionality of the unit see Device Check S 2 4 2 Depending on the work carried out the specific steps of the TSC must be performed see Technical Safety Check TSC S 7 1 3 The activities carried out must be recorded in the following check list 3 24 SpaceStation 1 0 gb Disassembly Assembly SpaceStation Check List for Checks after Repair Visual Inspection Electrical Safety ac
6. Note Is used for additional or special notes concerning information and working steps CAUTION Is used for working steps which may result in damage to the unit system or to a connected device WARNING IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL INJURY References to chapters are shown as follows see Setting Off pg O 8 References to figures and tables are shown as follows Fig 2 3 or Table 2 1 Important Preliminary Remarks References to item numbers in figures are shown as follows Fig 1 1 Item 1 In this case Fig 1 1 is the figure number and Item 1 the item number within the figure When the Service Manual is stored as pdf file these references are displayed green Click with the mouse button on a reference to jump to the corresponding source Menu commands are described as Menu File SpaceStation 2 0 gb List of Abbreviations SpaceStation 2 0 gb Important Preliminary Remarks Abbreviations which are not generally known but are used in this manual are listed below CAN Controller Area Network CE Communaut Europ enne CS Calibration Step DIN Deutsche Industrie Norm German Industrial Norm EN European Norm ESD Electrostatic Discharge FuP Function Microprocessor IEC International Electrotechnical Commission ISO International Organization for Standardization ISP Infusomat Space ISPS Infusomat Space Silicon ISPP Infusomat Spac
7. 00000 3452 1044 SPCO fastening sleeve aa 3452 1047 10 5 Spare Parts List SpaceCover Abb 10 3 Exploded drawing SpaceCover 10 6 SpaceStation 1 0 gb SpaceStation 1 0 gb BR WN Spare Parts List Battery compartment cover SPC 3452 1232 Battery pack SP NIMH 008 0871 3180 Handle SPC ina a ade ds 3452 1224 Housing upper part SPCS 3452 1194 Housing upper part SPCC Luuu 3452 1178 Housing back panel SPC a 3452 1216 Connector board SPCS a 3452 1240 only for SpaceCover standard Loudspeaker SPCC 00000000 00 3452 1267 Processor PCB SPCC 00200000 00s 3452 1259 Housing bottom part SPCS 3452 1208 Housing bottom part SPCC 3452 1186 10 7 Spare Parts List SpaceStation 1 0 gb Revision Documentation 11 Description of Version Version 1 0 Base First version of this Service Manual Release date September 20 2005 Version 2 0 Insertion of the SpaceStation version with SpaceCom incl new spare parts Changed TSC for the SpaceStation Texts corrected Release date December 14 2006 Version List of the Individual Pages SpaceStation 2 0 gb 1 1 Revision Documentation 1 2 SpaceStation 1 0 gb Title and Leading Pages Page 0 1 AA tii Revision 2 0 EDI EE Revision 2 0 Pagg0 3 aka nanana kaba Re ende Revision 2
8. 4 2 Item 2 out of the holder SpaceStation 2 0 gb 4 4 Tube Guide Abb 4 3 Legende zu Abb 4 3 ItemDesignation 1 Tube guide SpaceStation 2 0 gb Disassembly Assembly SpaceCom Designation Ord No Tube guide SPS aaa 3477 4394 20 pieces Disassembly 1 Press the rotary knob of the tube guide out and pull the tube guide Abb 4 3 Item 1 out of the housing Note When the tube guide is broken remove the remaining parts from the housing Disassembly Assembly SpaceCom 4 5 Pole Clamp Guide Abb 4 4 Legende zu Abb 4 4 ItemDesignation 1 Pole clamp guide 2 Countersunk screw M5x12 TORX 25 3 Pole clamp spring 4 Pole clamp lever 4 6 Ord No Pole clamp guide SPS aan 3452 1135 Designation Pole clamp lever see Service Parts and Screw Kit S 4 2 Pole clamp spring see Service Parts and Screw Kit S 4 2 Screws see Service Parts and Screw Kit S 4 2 Disassembly 1 Loosen three screws Abb 4 4 Item 2 and remove screws together with the pole clamp guide Abb 4 4 Item 1 2 Remove pole clamp lever Abb 4 4 Item 4 and pole clamp spring Abb 4 4 Item 3 SpaceStation 2 0 gb 4 6 Housing Back Panel Abb 4 5 Legende zu Abb 4 5 ItemDesignation 1 Housing notch 2 Screw EJOT 30x16 WN 5452 TORX 101P A2 3 Cover cap 4 Blind plug 5
9. BBRAUN B Braun SpaceOnline SHARING EXPERTISE v1 2 0 en Language English M Login name lt No login gt Network connection SpaceOnline O O 2006 11 02 15 30 Status Service Information P saki a Welcome to the SpaceOnline login Configuration Legend Perfusor amp Space Infusomat Space C P Please enter Username and Password E Pre alarm Pump is switched off on Ba Pump is in stand by service Username A Pump is running Password Pump is selected after mouse click on symbol Terms of Use Imprint Login Copyright 2005 2006 by B Braun Melsungen AG spaceOnline B Braun Melsungen AG MD Internet Abb 1 17 Program Version The program version is displayed in the header of the start screen Operation The instructions for use contain a more detailed description of op eration Exit the SpaceOnline Program 1 Close the Internet Browser 1 18 SpaceStation 2 0 gb BCCshow SpaceStation 2 0 gb System Overview Approved Version Note Please note that text and or functions of the BCCshow Program may change depending on the software version The following screen illustrations are only examples and represent the state when the manual was printed 3 26 Connection Before starting the BCCshow Program the PC is to be connected to the device via the RS232 SPCO crossed interface cable Connection Settings 1
10. Connect device to voltage supply The first LED in the bottom charge condi tion display of the operator and status display field of the SpaceCover comfort flashes yellow Battery module of theSpaceCom nearly discharged type pre alarm Connect device to voltage supply The first LED in the top charge condition display of the operator and status display field of the SpaceCover comfort flashes red Battery module of the SpaceCover com fort discharged Connect device to voltage supply The first LED in the bottom charge condi tion display of the operator and status display field of the SpaceCover comfort flashes red Battery module of the SpaceCom dis charged Connect device to voltage supply SpaceStation 2 0 gb Table2 2 Alarms of the SpaceCover comfort Part 1 of 2 Unit Diagnosis Calibration tion display of the operator and status tive or too old display field of the SpaceCover comfort lights up red 8 The first LED in the top charge condition Battery module of the SpaceCover com Replace battery module display of the operator and status display fort defective or too old field of the SpaceCover comfort lights up red 9 The first LED in the bottom charge condi Battery module of the SpaceCom defec Replace battery module Table 2 2 Alarms of the SpaceCover comfort Part 2 of 2 The Most Important Error Modes The following list specifies the most imp
11. 10 ItemDesignation 1 Loudspeaker connecting leads 5 10 Checks after Repair 1 Check the device to ensure safe functionality of the unit see Device Check S 2 4 2 Depending on the work carried out the specific steps of the TSC must be performed see Technical Safety Check TSC S 7 1 3 The activities carried out must be recorded in the following check list 5 12 SpaceStation 1 0 gb Disassembly Assembly SpaceCover Check List for Checks after Repair Visual Inspection Electrical Safety Functional Inspection according to IEC EN 60601 1 or VDE 0750 and VDE 0751 O Cleanliness O Locking with SPS O Completeness CJ Battery compartment cover O Damage and faults affecting safety C Damage to and readability of the label General functional check with the SpaceS O Cover caps tation O Connectors C Self test O Alarm O Loudness setting O Battery charge state O Switch over mains battery C Dimming function of the status and alarm display line SpaceStation 2 0 gb 5 13 Disassembly Assembly SpaceCover For your information 5 14 SpaceStation 1 0 gb Cleaning Servicing the Battery SpaceStation 2 0 gb Servicing the Unit 6 Clean and disinfect the SpaceStation at regular intervals To clean the system we recommend mild soap suds WARNING WHILE CLEANING AND DISINFECTING THE SPACESTATION DIS CONNEC
12. Abb 4 7 Item 2 and remove the con tact strip Abb 4 7 Item 1 carefully out of the housing back panel Note If necessary the spring mounted contact pins must be carefully inserted when the contact strip is dismounted SpaceStation 2 0 gb Abb 4 8 Disassembly Assembly SpaceCom Disassembly Designation Ord No Housing back panel SPCO 222200000 3452 1052 Fan SPECO iu icu det id 3452 1048 Legende zu Abb 4 8 ItemDesignation 1 ad A CO N Screw EJOT 30x9 WN 5452 TORX 101P A2 Housing back panel SPCO Fan connecting cable Fan grid SPCO Fan SPCO SpaceStation 2 0 gb Cover caps and screws see Service Parts and Screw Kit S 4 2 Note and outline the cable layout of the fan connecting cable Abb 4 8 Item 3 Unscrew four screws Abb 4 8 Item 1 Remove fan grid Abb 4 8 Item 4 out of the housing back panel Abb 4 8 Item 2 Remove fan Abb 4 8 Item 5 out of the housing back panel Disassembly Assembly SpaceCom 4 7 Device Bracket Abb 4 9 Legende zu Abb 4 9 ItemDesignation 1 Serrated lock washer M5 2 Screw M5x45 A2 DIN912 TORX 25 3 Device bracket 4 Insulating hood 4 10 Designation Ord No Device bracket SPS aa 3452 1127 Insulating hood SPCO aaa 3452 1043 Screws see Service Parts and Screw Kit S 4 2 Disassembly 1 Loosen five screws
13. 8 6 AN A Revision 1 0 Page 9 1 estirada cid e bite RR Re Revision 2 0 Page 92 icon sitgcedo e reb da Pa anda Revision 1 0 Page 10 1 isle od aspe tnr dte PA ERE RE Revision 2 0 Page 10 92 iii ro teen edes Revision 2 0 Page 10 5d esso e ias Revision 1 0 Page 10 4 clues a Revision 2 0 Page 10 5 vic emu ex pet bae Ed eds Revision 2 0 Page 10 6 saaan hah inn URRRPxRAR een Las Revision 1 0 Page 10 Ti du deseo pdt ER a edis Revision 1 0 A ER E EE Revision 1 0 Chapter 11 Revision Documentation Page til aeneo tecti ka Revision 2 0 Page 11 2 span eese its Pa weak Revision 1 0 Page 11 5 3 isi nn nb ARE NARE hm KAT E ane Revision 2 0 Page 11 9 4 duas caes Saeed eee Ven Ce Ros Revision 2 0 Pagg ING 5 Lube rx be E Rex ds Revision 2 0 Page 11 6 iul pen had a es Revision 1 0 SpaceStation 2 0 gb Chapter 12 Index Page 1231 A hn AR La Revision 2 0 Pagel 22 2 AA Revision 2 0 SpaceStation 2 0 gb 1 5 1 6 SpaceStation 1 0 gb A ACCESSONES scere DN heeds km GANAN 1 21 NAA 2 3 Assembly 3 1 3 21 4 1 4 13 5 1 5 11 B Battery compartment aaa 4 3 Battery compartment cover 4 3 5 3 Battery module aaa 4 3 5 4 Blind plug isses exces terr ra iaa 4 7 Block diagram suus 1 7 1 9 1 11 C Checks after repair 3 24 4 14 5 12 SpaceCover acatar AD 5 12 SpaceStation aaa 3 24 4 14 Cleanihg iens AA AA 6 1 ContactistH p sedet se pedi ke Nri den
14. Abb 4 9 Item 2 and remove screws together with the serrated lock washers Abb 4 9 Item 1 and the device bracket Abb 4 9 Item 3 2 Remove insulating hood Abb 4 9 Item 4 from the power supply SpaceStation 2 0 gb 4 8 SPCO PCB Abb 4 10 Legende zu Abb 4 10 ItemDesignation 1 SPCO connecting cable 2 SPCO PCB 3 Screw EJOT 30x9 WN 5452 TORX 10IP A2 SpaceStation 2 0 gb Disassembly Assembly SpaceCom Designation Ord No SPCO PEB a Aa Na 3452 1053 SPCO connecting cable 0 eee eee eee 3452 1042 SPCO insulating piece eee eee eee eee 3452 1044 SPCO fastening sleeve cece ee eee eee 3452 1047 Screws see Service Parts and Screw Kit S 4 2 Disassembly 1 Unscrew two screws Abb 4 10 Item 3 2 Remove SPCO PCB Abb 4 10 Item 2 carefully out of the housing and fold aside pay attention to the length of the SPCO connecting cable Abb 4 10 Item 1 3 Pull off the SPCO connecting cable on the SPCO PCB and re move the board out of the housing 4 Loosen two screws of the interface board and lift the inter face board carefully until the SPCO connecting cable can be pulled off 5 Pull off the SPCO connecting cable at the interface board 4 11 Disassembly Assembly SpaceCom Abb 4 11 Legende zu Abb 4 11 ItemDesignation 1 SPCO insulating piece 2 SPCO fastening sleeve 4 9 Fu
15. Basic software 690A00SS02 Improved functions Approved Software Versions for SpaceCom 695C000001 695C000002 Basic software 695C010003 Adapted to hardware changes 695D010001 Adapted to the pump software 695E010001 Adapted to the pump software Approved Software Versions for SpaceCover comfort 692A000001 Basic software 692A000002 Improved functions 692A000003 Expanded functions 692C000001 Adapted to the pump software 692E000001 Adapted to the pump software SpaceStation 2 0 gb SpaceStation 1 0 gb System Overview Software Update The instructions for updating the software are supplied with the software itself CAUTION If the device is disconnected while the software is being updated or the device or PC is switched off a component of the software may be seriously damaged so that repairs are no longer possible In such a case the software can no longer be updated via the PC and the device must be returned to B Braun 1 13 System Overview Service Program HiBaSeD security advice Pumps or other devices must not be connected to a patient while the equipment is being serviced The user must ensure the correctness of all device settings after The software may only be used for its intended purpose The connection to hardware components e g Perfusor or Infusomat may only be made using the prescribed interfaces and leads These products may not be simultan
16. Oo A W N W LAN slot Status indicator green Alarm indicator red Ethernet RJ 45 connection PS 2 connection for barcode reader RS232 connection USB connection slave USB connection master SpaceStation with SpaceCom The SpaceCom provides further interfaces and connections e g for a barcode scanner These interfaces are depicted in Abb 1 7 The battery module option serves to maintain the functions of the SpaceStation in case of a voltage failure An independent cir cuit in the battery module monitors the battery cells and controls their charge condition The SpaceStation must be covered with a SpaceCover comfort with inserted battery module to guarantee function of the battery module in the SpaceStation with Space Com SpaceStation 2 0 gb System Overview 398 1931 wogeoeds wo e0eds A OV A0 Z Function processor f RXTX 12VDC RX TXF F 12VDC CAN PCA RX TX 12VDC Abb 1 8 Block diagram SpaceStation with SpaceCom SpaceStation 2 0 gb 1 10 System Overview SpaceCover The SpaceCover is connected to a SpaceStation with connector p1 With the SpaceCover standard this connector is used to connect the CAN bus and a serial interface In the SpaceCover comfort this connector provides the voltage su
17. a 4 8 Current versions saut ed vere dcir daran 0 5 D Description iio stp VER Lace Ea ma t iaa 1 1 Device bracket sisse hn 3 6 4 10 Device check i d dex e e aoa en 2 4 Disassembly aaa 3 1 4 1 5 1 E Error codes 2 aisle de cece RI KA e dead 2 3 Error modes 4 lll 2 4 F ELDER MR PEU IHE 4 9 Fan connecting cable aaa 4 8 A hace db en Redi an 4 9 Fastening sleeve 0 eee eee eee eee 4 12 F netlohi csset ra aa rx dene alee as 1 6 H LED AA 5 5 SpaceStation 2 0 gb HOUSING aaa cee a ade 3 20 Housing back panel aaa 4 7 Assembly 0 ccc cece eee eee ees 3 24 4 13 Disassembly 000 cece eee eens 3 5 4 7 Housing back panel SPC aaa 5 8 Housing upper part ANA 5 12 Disassembly a 5 6 I Installation esee 3 21 4 13 5 11 Insulating hood 4 10 Insulating piece eee eee eee A 4 12 Interface board Disassembly aaa 3 12 3 18 Installation 0 0 ccc cece eee 3 21 K Kontakte nois iaa a ara 0 11 L List of abbreviations csse 0 9 Loudspeaket ainda adas ea 5 9 M Module lock Disassembly aaa 3 12 3 18 Installation 5 ute GD te Lahad 3 21 0 OPTIONS A NG Damn te PG 1 21 P Pole clamp guide wana sarsa bie RA KANG 3 4 4 6 Power supply pusang rad dad 3 7 Processor PCBs bait cU de Ea da diia 5 10 R Release button coccion ri ek 3 11 3 23 12 1 Index Responsibilities eee eee ees 0 6 REVISION S
18. button Abb 5 9 Item 4 in the housing bottom part carefully to the rear and remove the processor PCB Abb 5 9 Item 2 out of the housing bottom part SpaceStation 1 0 gb 5 9 Assembly Installation SpaceStation 1 0 gb Disassembly Assembly SpaceCover Assembly or installation of the modules and subsystems is done in reverse order of disassembly Special steps to be observed are de scribed hereafter in detail Only new cover caps are to be used Special Screws Special screws for plastic housings are used in this unit The screws are not self cutting but produce a thread in the plastic of the housing through deformation when fitted in for the first time If the beginning of the thread is not engaged when the screw is fitted a new thread is produced and the old thread is destroyed so that the security of the fixing can no longer be guaranteed Proceed as follows to fit the special screws 1 Put the screw on the thread 2 Rotate screw anti clockwise loosen until a faint click can be heard This click is produced when the screw thread drops into the existing thread 3 Screw in the screw and tighten with the defined torque 5 11 Disassembly Assembly SpaceCover Housing Upper Part Note Pay attention to the correct position of the loudspeaker connect ing leads when mounting the housing upper part please see Abb 5 10 The leads must point upwards Abb 5 10 Legende zu Abb 5
19. screw driver and pull caps out 10 Loosen two screws Abb 3 15 Item 1 and remove them out of the data connector holder 11 Loosen two screws Abb 3 16 Item 2 and remove them from the connector board F5 Abb 3 16 Item 1 12 Press the connector board F5 carefully out of the housing Note The connector board F5 is integrated in the interface board 3 15 Disassembly Assembly SpaceStation 13 Loosen one screw Abb 3 17 Item 1 and remove it out of the data connector holder Abb 3 17 Item 2 Abb 3 17 Legende zu Abb 3 17 ItemDesignation 1 Screw EJOT 30x9 WN 5452 TORX 10IP A2 2 Data connector holder 3 16 SpaceStation 1 0 gb Abb 3 18 Legende zu Abb 3 18 ItemDesignation 1 BR 0 M Module lock Interface board Data connector holder Screw EJOT 30x9 WN 5452 TORX 10IP A2 SpaceStation 1 0 gb Disassembly Assembly SpaceStation 14 Loosen three screws Abb 3 18 Item 4 and remove screws together with the module lock Abb 3 18 Item 1 the interface board Abb 3 18 Item 2 and the data con nector holder Abb 3 18 Item 3 out of the housing Note Note the ribbon cable length to the interface board when dis mounting the data connector holder The ribbon cable must not be kinked 3 17 Disassembly Assembly SpaceStation Disassembly 1 Loosen fours screws Abb 3 19 Item 1 and rem
20. the SpaceCover comfort After start up the battery charge condi tion is displayed in the bottom charge condition display of the operator and status display field of the SpaceCover com fort Remove the battery module out of the SpaceCover comfort Disconnect the SpaceStation from mains when the status in dicator green on the SpaceCom SpaceCom is operative lights up The status indicator green on the SpaceCom remains on for at least 3 seconds The alarm indicator red must not light up or flash Remove the optional battery modules which are not part of the unit out of the device for Procedural Instructions on the TSC For your information 8 6 SpaceStation 1 0 gb Test Equipment and Special Tools Test Equipment Special Tools SpaceStation 2 0 gb Designation For Device Check Service CD namaga dhahigbebivea d eezb dnb Service connector Space aaa Interface cable RS232 SP a Interface cable RS232 SPCO crossed Interface cable RJ45 cross over For Repairs No special test equipment required For the TSC Measuring instrument for electrical safety Service connector Space see above Designation TORX screwdriver kit standard 5 10 25 TORX screwdriver kit Plus 5 10 25 Screwdriver 6IPx60 TORX plus Screwdriver 8IPx60 TORX plus Screwdriver 10IPx60 TORX plus Scr
21. the special screws 1 Put the screw on the thread 2 Rotate screw anti clockwise loosen until a faint click can be heard This click is produced when the screw thread drops into the existing thread 3 Screw in the screw and tighten with the defined torque Interface Board Module Lock Note Before the interface board is installed make sure that the two housing feet are inserted in the housing Note When installing the interface board observe the mounting se quence i e first attach the interface board to the module lock and then install the complete group in the housing 1 Tighten the connector board screws with a torque of 0 25 to 0 05 Nm 3 21 Disassembly Assembly SpaceStation e Abb 3 24 Legende zu Abb 3 24 ItemDesignation 1 Screw EJOT 30x9 WN 5452 TORX 10IP A2S 2 Insulating washer M3 Abb 3 25 Legende zu Abb 3 25 ItemDesignation 1 Rubber cable 3 22 2 Tighten the data connector holder screws to connectors F3 and F4 with a torque of 0 25 0 05 Nm 3 Install the subsystem interface board module lock in the housing in such a way that the rubber cable lies in the hous ing bottom CAUTION When mounting the data connector holder make sure that the rubber cable is not squeezed between the housing and the con nector holder This would result in an incorrect sealing of the housing and damage to the rubber cable SpaceStation 1 0 gb
22. tion with SpaceCom for proper operation Check proper functioning of the lock to another SpaceStation or a SpaceCover Check the battery compartment cover lock of the SpaceCover comfort for proper operation Functional Check 1 2 Connect the SpaceStation to the mains In case of a SpaceStation with SpaceCom the following func tions are to be checked a Fan in the housing back panel starts running for appr 1 sec b Status indicator green on the SpaceCom lights up c Alarm indicator red on the SpaceCom lights up for ap prox 1 minute Trigger an alarm with a pump at each slot The red LED on the service connector SP must light up When a SpaceCover comfort is fitted the alarm must also be displayed in the status and alarm display line of the Space Cover comfort an audible alarm must sound Carry out a self test of the SpaceCover comfort actuate the On Off button on theSpaceCover comfort please see in structions for use Battery Check of the SpaceCover comfort 1 Insert a battery module in the SpaceCover comfort After start up the battery charge condition is displayed in the top charge condition display of the operator and status display field of the SpaceCover comfort Disconnect the SpaceStation from the mains An audible alarm sounds and the status display colour on the operator and status display field of the SpaceCover comfort changes from green to yellow Possible displays are shown on t
23. 0 Page 4 7 is in Revision 2 0 Page 4 8 ica a bean Revision 2 0 Page 4 O Namamahala der d hbri RE ER Revision 2 0 Page 4 10 ace nae aca ee Gada bcn i rcnt Revision 2 0 Page 4 11 sioe tune REA PRE Revision 2 0 Page ds N2 nana hea cnet ode ear Revision 2 0 Page 4 19 xus end id Revision 2 0 Page 4 5 14 Kaka whet a ERG Revision 2 0 Page 4 15 Lions ser e dir er en Revision 2 0 Page 4 16 escri riadas Revision 2 0 Chapter 5 Disassembly Assembly SpaceCover Page as 4804 iet neha dae dei Revision 2 0 Page 322 sone dd thine dander dea Revision 2 0 A uel od Ux ea Revision 1 0 Page b e iia ed PAIRE EIE P aa n Revision 1 0 Pagebi b angina haah an NAL taa Revision 2 0 Page 5 6a dre era dnt on fei NA Revision 1 0 Page 5 7 rra Revision 1 0 Page 5 8 aia ca kid beheld RAE Revision 1 0 Page 5 O AA Revision 1 0 Page 5 70 agama maka ra Th BATA RD AA Revision 1 0 Page Mica dt errem Revision 1 0 Page 6 A ba boas ended Revision 1 0 Page bs Taiana eek ni RR Ka Revision 2 0 Page 5 14 disks e aer VER adi NAA Revision 1 0 1 4 Chapter 6 Servicing the Unit Page 6 IO em oet ence sa HR Gne Revision 2 0 A ket Re ERR ea Revision 1 0 Page 7 Hl oa duree ie Revision 2 0 Page 7 2 11 22 4 bec gra ERE ue RR Edd Revision 1 0 Page B T a n Revision 2 0 PAGES Z culos dades x faoc RR Riad Revision 2 0 Page 8 3 ia ii miu eL Det dd d Ra A pd Revision 2 0 Page E qp ETT Revision 2 0 Page 8 5 cias elec ead te cte Ed ees Revision 2 0 Page
24. 0 Page O aa naei naaa E ER ie e is Revision 2 0 PAGO D i socle ped da ad dex Rel nina Revision 2 0 Page 0 B cecus rra ar Revision 2 0 Page Q 7 ii e noe d Perd Revision 2 0 Page Q B c Revision 2 0 Pagg029 Spheres ahah soin e pn Revision 2 0 Paye 0 10 Ana cee e ame ER RA Re ds Revision 2 0 Page O 11 oet usted A LGA Revision 2 0 Page 0 12 u decena tad uiian ete eee Revision 2 0 Chapter 1 System Overview Page 1 1 iss ens rre ho weed ead NA Revision 2 0 ICD EE Revision 2 0 Page T 3 ted heck tex adn ted ER EE Revision 2 0 Page 1 4 wicca nein eiui it ERR mee Revision 2 0 Paggiba agam nap HG naa hp eT PPRIessPL Revision 1 0 Page 1 0 nisiseme dier ERA APTA Revision 2 0 A pi pLAISaaw NANAY Revision 2 0 Page TEB a a Revision 2 0 A A ber tate ne ew tender fo ets Revision 2 0 Page T 10 cin Khan tete haa Revision 1 0 Page TATT pia E tet Revision 1 0 Page To ri al ide Revision 2 0 A EE Revision 1 0 Page ME us dai ow RANG Revision 2 0 Pages 15 sony cp iaa Revision 2 0 Page SOS m Revision 2 0 Pagel 17 ehh rmn a d Revision 2 0 Page 1518 oeste Revision 2 0 Pagel 219 aio ria in tada Revision 2 0 Page A A NA REG Revision 2 0 Page 1 21 iman ie AGA Revision 2 0 SpaceStation 2 0 gb Page 1 22 2 ii ALA A ed Revision 1 0 Chapter 2 Unit Diagnosis Calibration A ass kaanak hana kA DAAN Revision 2 0 Page 2 2 wid aa td GG Revision 1 0 Page2 3 rta dena Gng a Revision 2 0 o ET Revision 2 0
25. 3 Item 1 and remove screws together with the insulating washers Abb 3 13 Item 3 from the interface board Abb 3 13 Item 2 6 Put the interface board carefully aside for all further activi ties Note Take care not to squeeze the interface board ribbon cable Abb 3 13 Legende zu Abb 3 13 ItemDesignation 1 Screw EJOT 30x9 WN 5452 TORX 101P A2 2 Interface board 3 Insulating washer M3 SpaceStation 1 0 gb 3 13 Disassembly Assembly SpaceStation 7 Release all rotary knobs Abb 3 14 Item 1 of the module lock Note Put a release button on the release shaft without engaging it to unlock the rotary knobs To lock the rotary knobs press them si multaneously in by hand per slot 8 Press 8 seals M 2K Abb 3 14 Item 2 carefully out of the housing press the rotary knobs in the locking position and pull the seals M 2K off the rotary knobs lL RA A Abb 3 14 Legende zu Abb 3 14 ItemDesignation 1 Release button 2 Seal M 2K 3 14 SpaceStation 1 0 gb 2 S Gg d Abb 3 15 Legende zu Abb 3 15 ItemDesignation 1 Screw EJOT 30x9 WN 5452 TORX 10IP A2 2 Cover cap 2 Abb 3 16 Legende zu Abb 3 16 ItemDesignation 1 Connector board F5 2 Screw EJOT 30x9 WN 5452 TORX 10IP A2 SpaceStation 1 0 gb Disassembly Assembly SpaceStation 9 Pierce two cover caps Abb 3 15 Item 2 with a
26. 5452 TORX 10IP A2 4 Loosen the nut Abb 3 7 Item 4 and remove nut together with the serrated lock washer Abb 3 7 Item 3 washers Abb 3 7 Item 1 and the protective conductor cable to connector F1A Abb 3 7 Item 2 from the grounding bolt 5 Carefully pull off the connector of the power supply wires from the interface board Note Note or mark the polarity 6 Loosen two screws Abb 3 8 Item 1 and take them out of the power supply housing connector SpaceStation 2 0 gb Abb 3 9 Legende zu Abb 3 9 ItemDesignation 1 Screw EJOT 30x12 WN 5451 TORX 10IP A2 2 Power supply SpaceStation 1 0 gb Disassembly Assembly SpaceStation 7 Loosen four screws Abb 3 9 Item 1 and take the power supply Abb 3 9 Item 2 with the mains connector holder out of the housing Disassembly Assembly SpaceStation 1 LES Rd PI Ya AU xt SN y NG 2 Abb 3 10 Legende zu Abb 3 10 ItemDesignation 1 Mains connector holder 2 Mains seal FIA 3 Screw EJOT 30x9 WN 5452 TORX 10IP A2 3 10 Disassembly 1 Loosen three screws Abb 3 10 Item 3 and remove screws together with the mains connector holder Abb 3 10 Item 1 and the mains seal F1A Abb 3 10 Item 2 from the power supply SpaceStation 1 0 gb 3 7 Release Button Abb 3 11 Legende zu Abb 3 11 ItemDesignation 1 Release bu
27. All necessary steps to disassemble or dismount the complete unit with all its subsystems and spare parts of the SpaceCom are de tailed in the following description All further disassembly and assembly steps correspond to the SpaceStation without Space Com and are decribed in Chapter Disassembly Assembly Space Station S 3 1 Steps that are not necessary can be skipped Note Special screws for plastic housings are used in this unit Pay at tention to the corresponding notes when you fit the screws Disassembly Assembly SpaceCom Service Parts and Screw Kit All small parts such as cover caps are contained in a SpaceSta tion service part kit Designation Ord No Service part kit SpaceStation 3477 4335 with Release button Release seal Seal M 2K Cover cap for housing Cover cap for handle Housing foot SPS Mains seal FIA Pole clamp lever Pole clamp spring O ring 6 07 x 1 78 O ring 12 0 x 2 0 All screws used in the device are included in a SpaceStation screw kit Designation Ord No Screw kit SpaceStation aa 3477 4343 Screw EJOT 30x9 WN 5452 TORX 10IP A2 Screw EJOT 30x12 WN 5452 TORX 10IP A2 Screw EJOT 30x16 WN 5451 TORX 10IP A2 Screw M5x45 A2 DIN912 TORX 25 Countersunk screw M5x12 TORX 25 Fillister head screw M5x12 EN IS014583 A2 Nut M4 Insulating washer M3 Washer M4 Washer M5 Serrated lock washer M4 Serrated lock washer M5 SpaceStation 2 0 gb 4 2 Battery Mod
28. ENICE zu ede decore be heeded dead nda ite 0 5 S Service Program ira 1 14 Servicing the UNIt aaah eR e aaa ERE ERR RR 6 1 Software cuca cse ed Er hPa das 1 12 Software update aaa 1 13 SPACOCOM lui ida 4 1 SpaceStation SPaceloM as iccad Lakakawabhs RP ride 4 1 Spare parts liSt cir ta lan 10 1 SPCO connecting cable aaa 4 11 SPCO PCB nakka ee tae Ese edes kde banda 4 11 12 2 Special tools Lanka nG ka nn eile bd Ea 9 1 System Overview iusso EUER Fen RU ESQ EA cated ee oa 1 1 T Technical data ss sssi ka ede bea 1 21 Technical Safety Checks ooooooooooommoo 0 5 7 1 Test equipment Jui maa a ad eee ed gei es 9 1 Trouble shooting aa 2 6 GA NANANA ANAN 0 5 7 1 Procedural instructions 8 1 Test equipment aaa 9 1 Tube guid eri e ete or tee onn 3 3 4 5 WwW W LAN module 00 0c c cece eee iei eiee 4 4 SpaceStation 2 0 gb
29. Open the BCCshow ini file after having installed the pro gram 2 Check the line ComPort X 3 Substitute the X for the current ComPort number e g 1 Starting the BCCshow Program 1 Start the BCCshow program The start screen appears 2 Invoke the display of the connected system via System gt Start The SpaceStation with all pumps inserted and the set data is displayed 1 19 System Overview Display for FM system data SpaceCOM 1 1 gt VERSION V3 25 gt Space System Record Info Window a B Braun Space PILLAR 1 Free Place PERFUSOR SPACE Free Place PERFUSOR SPACE 22222 MEM GET Port COM4 not recordin Abb 1 18 Program Version 1 Open the Information window via nfo Version The cur rent program version is displayed in this window Informationen E 2 Close the window by clicking OK o MedKom BCCshow SPACE 1 version 3 30 050929 Abb 1 19 1 20 SpaceStation 2 0 gb Technical Data Options Accessories SpaceStation 2 0 gb System Overview All technical data is indicated in the instructions for use The functions of the individual options are detailed in the instruc tions for use SpaceStation with SpaceCom Designation Part No SpaceCom retrofit Kit oooooooooooomomo o 0871 3160 The optional SpaceCom provides additional interfaces such as Ethernet RJ45 RS232 A SpaceStati
30. PAGE 2 5 Kan Ea hb te e RARO ERN ane oud Revision 2 0 Page 2 6 ooicsebbRRMRPR xx qud RM ES Revision 2 0 CD P Arm Revision 2 0 Page 2 8 aos maana Ere RARE OPERE Revision 2 0 Pages via pane uate Mieke edite eater x duda Revision 2 0 Page irradia Revision 2 0 Page 3 ao dati a eat a need Revision 1 0 Page 3 4 AP Revision 1 0 Page 5 ass nc dated vex eave ae aed pepe are Revision 1 0 Page 3 6 ccd adit ced au hind do nes Revision 2 0 A I4 Sed Revision 2 0 A ewe admet Pate ed uet adu Revision 2 0 Page Vias rar EE DNE Revision 1 0 Page 3 Tirar dere pe A abc ue Revision 1 0 Pages Moca te dein darn ded Revision 1 0 Page Ti Revision 1 0 Page 3 13 cura verd io NG Revision 1 0 Page 3 Third mad Revision 1 0 A AA matte mes Revision 1 0 A AKNG ka BRIAN NAKA tee iedies Revision 1 0 Page 3 A i rra mh geal d Revision 1 0 Pago Bites Revision 1 0 Page di Taiana Revision 1 0 Page 20 cid ve d ed dea Revision 1 0 Page 3 21 cosita bibi EE Revision 1 0 Pag 22 Sis ai bibas napa aa Ga Revision 1 0 Page Li ii Revision 1 0 Page 3 24 ais pde de Revision 1 0 1 3 Page 3 25 iria cr NA IE Revision 1 0 Page 3 26 vesci esie dbpErrP boe deus Revision 1 0 Chapter 4 Disassembly Assembly SpaceCom Page 4 1 neha nka DE pm EK Pee pear Revision 2 0 Page4 2 an dre Revision 2 0 Page 4 3 uio odiosa pe E Re En d Revision 2 0 Page AA durs e AM ii Revision 2 0 Page4 35 genes cant eur EDEEM Dre Abe Revision 2 0 Page 4 6 a antigen ae da Revision 2
31. PAS GmbH Br ckner GmbH Germany 0 11 KI Contact Persons For your information 0 12 SpaceStation 2 0 gb Description System Overview Abb 1 1 Space system Legende zu Abb 1 1 ItemDesignation 1 2 3 4 5 SpaceCover Infusion pump Infusomat Space Infusion syringe pump Perfusor Space SpaceControl SpaceStation SpaceStation 2 0 gb System Overview 1 The modular Space system is destined for treatment of an individ ual patient It is designed for application on intensive care units and in the surgery The Space system is operated in hospitals mainly by doctors and medically trained personnel The SpaceStation forms together with the SpaceCover an organi zation system The Space system infusion and infusion syringe pumps are attached and connected to the SpaceStation The Space system is a modular design of modern infusion technol ogy for stationary mobile or private use The key modules and their connection to the peripheral devices are shown in Abb 1 1 All the pump types Perfusor Space Infusomat Space and Infu somat Space P as well as the other devices of the system are of modular design Up to three pumps can be connected together us ing L rails on the bottom of the unit and grooves on the top They can then be fastened to a drip stand or appropriate rail using the pole clamp The SpaceControl module can be used to exte
32. Screw EJOT 30x9 WN 5452 TORX 10IP A2 6 Housing back panel SPCO SpaceStation 2 0 gb Disassembly Assembly SpaceCom Designation Ord No Cover cap and screws see Service Parts and Screw Kit S 4 2 Disassembly Note The third cover cap on the right side of the housing back panel when viewed from above is a blind plug Abb 4 5 Item 4 which does not need to be removed 1 Pierce ten cover caps Abb 4 5 Item 3 with a screwdriver and pull caps out 2 Unscrew eight screws Abb 4 5 Item 5 3 Unscrew two screws Abb 4 5 Item 2 CAUTION When you remove the housing back panel pay attention to the connecting cables to the unit They might get off 4 Remove the back panel Abb 4 5 Item 6 carefully from the housing and fold aside pay attention to the cable con nections Note Lift the housing back panel carefully at the bottom over the seat of the pole clamp guide push the back panel over the housing notch Abb 4 5 Item 1 and fold aside Disassembly Assembly SpaceCom Abb 4 6 Legende zu Abb 4 6 ItemDesignation 1 Fan connecting cable presented without back panel 2 Connector Abb 4 7 Legende zu Abb 4 7 ItemDesignation 1 Contact strip 2 Screw A2 WN 5452 30x9 5 Pull off connector Abb 4 6 Item 2 of the fan connecting cable Abb 4 6 Item 1 at the SPCO board 6 Unscrew one screw
33. Service Parts and Screw Kit mp S 5 1 Disassembly 1 Unscrew two screws Abb 5 7 Item 3 on the housing back panel and remove screws together with the connector board Abb 5 7 Item 1 Abb 5 7 Item 2 Note The connector board in the SpaceCover comfort is integrated in the SpaceCover comfort processor PCB 2 Take the housing back panel out of the housing bottom part SpaceStation 1 0 gb Disassembly Assembly SpaceCover 5 7 Loudspeaker Designation Ord No Only SpaceCover comfort Loudspea ker SPCC o 3452 1267 Disassembly 1 Pull the loudspeaker connector Abb 5 8 Item 1 off the processor PCB 2 Take the loudspeaker out of the housing bottom part Abb 5 8 Legende zu Abb 5 8 ItemDesignation 1 Loudspeaker SpaceStation 1 0 gb 5 9 Disassembly Assembly SpaceCover 5 8 Processor PCB Only SpaceCover comfort Abb 5 9 Legende zu Abb 5 9 ItemDesignation 1 Contact strip 2 Processor PCB 3 Housing bottom part 4 Release button 5 10 Designation Ord No Processor PCB SPCC aaa 3452 1259 Housing bottom part SPCS aaa 3452 1208 Housing bottom part SPCC aaa 3452 1186 Disassembly 1 Push the contact strip Abb 5 9 Item 1 to the battery module to the rear and out of the housing bottom part Abb 5 9 Item 3 Note The contact strip is integrated in the processor PCB 2 Press the release
34. SpaceCover comfort Page 5 4 Handle Page 5 5 Housing Upper Part Page 5 6 Housing Back Panel Page 5 8 Loudspeaker Page 5 9 Only SpaceCover comfort Page 5 9 Processor PCB Page 5 10 Only SpaceCover comfort Page 5 10 Assembly Installation Page 5 11 Checks after Repair Page 5 12 Servicing the Unit Cleaning Page 6 1 Servicing the Battery Page 6 1 Technical Safety Check TSC Page 7 1 Procedural Instructions on the TSC Visual Inspection Page 8 1 Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 Page 8 1 Functional Inspection Page 8 4 Test Equipment and Special Tools Test Equipment Page 9 1 Special Tools Page 9 1 Spare Parts List Page 10 1 Revision Documentation Description of Version Page 11 1 Version List of the Individual Pages Page 11 1 Index Page 12 1 0 4 SpaceStation 2 0 gb Service Work Technical Safety Checks Current Versions Revision Service SpaceStation 2 0 gb Important Preliminary Remarks The present manual is for your information only The possession of this manual does not authorize the performance of service work Service tasks may only be executed by persons who have received appropriate training on the system from B Braun are included in the revision service possess the necessary test equipment and mechanical aids and fulfill the personal requirements training and knowledge The user is obliged to perform or to have performed the Technical Safety Checks on those m
35. SpaceStation SpaceCover SpaceCom Service Manual Revision 2 0 English BIBRAUN SHARING EXPERTISE This Service Manual is valid for This Service Manual is available under the following part number Languages of this Manual Designation Part No SpaceStation without SpaceCom 0871 3140 SpaceStation with SpaceCom 0 0871 3142 SpaceCover standard 020000e eee eeeee 0871 3147 SpaceCover comfort 00000 eee eee eee 0871 3145 Designation Part No Service Manual SpaceStation English 8713 9220 The Service Manual for this unit can be supplied in the following languages Designation Part No GORMAN errar ra NOR DUE EG 8713 9210 English US niet is 8713 9220U AP 8713 9230 SpaceStation 2 0 gb Important Preliminary Remarks Contact Persons System Overview Unit Diagnosis Calibration Disassembly Assembly SpaceStation SpaceStation 2 0 gb Table of Contents Service Work Technical Safety Checks Current Versions Revision Service Responsibility of the Manufacturer Quality Management Checks and Repair Notes on ESD Spare Parts and Test Equipment Setting Off List of Abbreviations Technical Training Entry for Technical Training Ordering of Spare Parts and Test Equipment Service Hotline Return of Spare Parts and Test Equipment Safety Officer 8 30 MPG Translation Description System Overview Physical construction Functio
36. T THE UNIT FROM THE MAINS SUPPLY CAUTION Take care that no water or liquid enters the device between the housing front and back panel For disinfection by wiping you should use for example Meliseptol from B Braun Allow the unit to dry for at least one minute When you disinfect the device by spraying make sure not to spray in the system openings such as loudspeaker opening interface connec tor The fan in the housing back panel of the SpaceStation with SpaceCom is to be vacuumed regularly If the fan is badly soiled dismount the housing back panel and clean the fan and the area around the fan using a brush The instructions for use contain a detailed description on how to service the battery E Servicing the Unit For your information 6 27 SpaceStation 1 0 gb Technical Safety Check TSC Checklist for Technical Safety Check Every 24 Months Unit SpaceStation SPS SpaceStation with SpaceCom SPCO SpaceCover standard SPCS SpaceCover comfort SPCC Manufacturer B Braun Melsungen AG Observe the Service Manual and the instructions for use All measured values are to be documented Accessories used should be included in testing Make excl use of calibrated measuring equipment Index b Master to be added to the documentation User Article No Unit No Year of Procurement
37. WORK WITH VOLTAGES UP TO 115 230 V AC THESE VOLTAGES MAY CAUSE INJURIES WHICH ARE DANGEROUS TO LIFE AND LIMB THE NATIONAL AND INTERNATIONAL SAFETY REGULATIONS ARE TO BE ADHERED TO THE MAINS CABLE IS TO BE REMOVED Before disassembling the unit the system must be checked see Device Check S 2 4 to isolate the part to be exchanged All necessary steps to disassemble or dismount the complete unit with all its subsystems and spare parts are detailed in the follow ing description Steps that are not necessary can be skipped Note Special screws for plastic housings are used in this unit Pay at tention to the corresponding notes when you fit the screws Disassembly Assembly SpaceStation Service Parts and Screw Kit All small parts such as cover caps are contained in a SpaceSta tion service part kit Designation Ord No Service part kit SpaceStation 3477 4335 with Release button Release seal Seal M 2K Cover cap for housing Cover cap for handle Housing foot SPS Seal mains FIA Pole clamp lever Pole clamp spring O ring 6 07 x 1 78 O ring 12 0 x 2 0 All screws used in the device are included in a SpaceStation screw kit Designation Ord No Screw kit SpaceStation aa 3477 4343 Screw EJOT 30x9 WN 5452 TORX 10IP A2 Screw EJOT 30x12 WN 5452 TORX 10IP A2 Screw EJOT 30x16 WN 5451 TORX 10IP A2 Screw M5x45 A2 DIN912 TORX 25 Countersunk screw M5x12 TORX 25 Fillister head scre
38. and Screw Kit All small parts such as cover caps are contained in a SpaceSta tion service part kit Designation Ord No Service part kit SpaceStation 00 3477 4335 with Release button Release seal Seal M 2K Cover cap for housing Cover cap for handle Housing foot SPS Mains seal FIA Pole clamp lever Pole clamp spring O ring 6 07 x 1 78 O ring 12 0 x 2 0 Disassembly Assembly SpaceCover All screws used in the device are included in a SpaceStation screw kit Designation Ord No Screw kit SpaceStation a 3452 1075 Screw EJOT 30x9 WN 5452 TORX 101P A2 Screw EJOT 30x12 WN 5452 TORX 101P A2 Screw EJOT 30x16 WN 5451 TORX 101P A2 Screw M5x45 A2 DIN912 TORX 25 Countersunk screw M5x12 TORX 25 Fillister head screw M5x12 EN IS014583 A2 Nut M4 Insulating washer M3 Washer M4 Washer M5 Serrated lock washer M4 Serrated lock washer M5 SpaceStation 2 0 gb 5 2 Battery Compartment Cover Abb 5 1 Legende zu Abb 5 1 ItemDesignation 1 Battery compartment cover lock Abb 5 2 Legende zu Abb 5 2 ItemDesignation 1 Battery compartment cover SpaceStation 1 0 gb Disassembly Assembly SpaceCover Designation Ord No Battery compartment cover SPC 3452 1232 Disassembly 1 Rotate the lock Abb 5 1 Item 1 on the battery compart ment cover Abb 5 2 Item 1 and remove the battery compartment cover The battery modul
39. cording to IEC EN 60601 1 Functional Inspection or VDE 0750 and VDE 0751 O Cleanliness O Mains voltage according to the TSC O Locking with second SPS O Completeness O Protective conductor resistance ac O Pump locking of all slots O Damage and faults affecting safety cording to the TSC O Damage to and readability of the label O Earth leakage current according to the Connect unit to mains O Cover caps TSC General functional check with SpaceCover C Connectors comfort and pump of every slot O Well running and sealing of the module O Voltage supply of all slots lock O Alarm transmission C Mains cable O Staff call transmission O Accessories SpaceStation 1 0 gb 3 25 Disassembly Assembly SpaceStation For your information 3 26 SpaceStation 1 0 gb Disassembly Assembly SpaceCom 4 4 1 General on the SpaceStation with SpaceCom SpaceStation 2 0 gb Remarks on Disassembly Assembly of the SpaceStation with SpaceCom WARNING DURING DISASSEMBLY AND ASSEMBLY THE OPERATOR SERVICE TECHNICIAN MUST WORK WITH VOLTAGES UP TO 115 230 V AC THESE VOLTAGES MAY CAUSE INJURIES WHICH ARE DANGEROUS TO LIFE AND LIMB THE NATIONAL AND INTERNATIONAL SAFETY REGULATIONS ARE TO BE ADHERED TO THE MAINS CABLE IS TO BE REMOVED Before disassembling the unit the system must be checked see Device Check S 2 4 to isolate the part to be exchanged
40. current SPCO accord comfort and pump of every slot O Well running and sealing of the module ing to the TSC O Fan SPCO approx 1 sec lock O Status indicator green lights up CJ Mains cable CJ Alarm indicator red is on for O Accessories approx 1 min O Voltage supply of all slots O Alarm transmission CJ Staff call transmission O Ethernet interface RJ 45 O RS232 interface O Charge state of the battery module display on the SpaceCover comfort O Switch over mains battery SpaceStation 2 0 gb 4 15 Disassembly Assembly SpaceCom For your information 4 16 SpaceStation 2 0 gb Disassembly Assembly SpaceCover 5 5 1 General SpaceStation 2 0 gb Remarks on Disassembly Assembly Before disassembling the unit the system must be checked see Device Check S 2 4 to isolate the part to be exchanged All necessary steps to disassemble or dismount the complete unit with all its subsystems and spare parts of the SpaceCover stand ard and the SpaceCover comfort are detailed in the following de scription Steps that are not necessary can be skipped Note All figures in this chapter depict the SpaceCover comfort In gen eral these figures also apply to the SpaceCover standard Note Special screws for plastic housings are used in this unit Pay at tention to the corresponding notes when you fit the screws Service Parts
41. ds even without discharge via contact This type of damage is not immediately recognizable Unit malfunctions can even occur after a longer period of operation SpaceStation 2 0 gb Spare Parts and Test Equipment Setting Off SpaceStation 2 0 gb Important Preliminary Remarks Each workstation must be equipped according to the recommen dations with the necessary static protective measures if ESD components or boards are handled Each workstation must be equipped with a conductive table sur face The conductive surface the soldering iron or the soldering stations must be grounded via protective resistors Chairs must be of antistatic design The floor or floor mats should be of electrically conductive material Personnel must wear conductive wristbands which are connected to a central ground potential via protective resistors e g the ground contact of a wall outlet Furthermore it is recommended that personnel wear cotton clothing and electrically conductive shoes to prevent electrostatic charge Only use original spare parts from the manufacturer Do not tamper with assembly groups which can only be exchanged com pletely The spare parts required are listed in Section 9 Service personnel are responsible for the calibration of their test equipment Original test equipment can be calibrated at the works of B Braun Further information is available upon request Additional notes and warnings are set off as follows
42. duced when the screw thread drops into the existing thread 3 Screw in the screw and tighten with the defined torque Housing Back Panel CAUTION When mounting the contact strip see that the springy contacts are not bent or broken 1 Push the contact strip Abb 4 12 Item 2 in the groove Abb 4 12 Item 1 at the battery compartment of the housing back panel and tilt carefully into mounting position 2 Secure the contact strip with a screw Abb 4 12 Item 3 against falling out Note Screw the screw in up to a distance of about 2 mm from the screwed mandrel This is the optimum spacing to clamp the con tact strip 4 13 Disassembly Assembly SpaceCom Abb 4 13 Legende zu Abb 4 13 ItemDesignation 1 Locking tabs Housing back panel Screw A2 WN 5452 30x16 Screw A2 WN 5452 30x9 Housing cover cap W LAN cover Oo c1 A W N Note Before the housing back panel is definitely fitted check the insu lating hood and the insulating piece for proper fit and the correct cable layout of the fan cable and of the connecting cable to the battery contact strip 3 Place on the housing back panel Abb 4 13 Item 2 and fasten it with the two longer screws Abb 4 13 Item 3 and the eight shorter screws Abb 4 13 Item 4 Note Insert the housing back panel on the notches Abb 4 13 Item 1 at the two connector coverings of the housing and tilt the back panel in
43. e PVC KuP Monitoring Microprocessor LCD Liquid Crystal Display MOS Short name of the following company MOS Technology Inc Commodore Semiconductor Group PCA Patient Controlled Analgesia PSP Perfusor Space SP Space System SPC SpaceCover SPCC SpaceCover comfort SPCS SpaceCover standard SPCO SpaceCom SPCT SpaceControl SPS SpaceStation TS Trouble Shooting UTS Unit Test Step TSC Technical Safety Check Important Preliminary Remarks TEMP Temperature VDE Verband der Elektrotechnik Elektronik und Informationstechnik e V German Association of engineers 0 10 SpaceStation 2 0 gb Technical Training Entry for Technical Training Ordering of Spare Parts and Test Equipment Service Hotline Return of Spare Parts and Test Equipment Safety Officer S 30 MPG Translation SpaceStation 2 0 gb Contact Persons Via local representative Application for a technical training course must be made via the responsible representative Please contact your local B Braun subsidary International Technicians Intercompany Nadja Machal Fax 49 5661 75 47 89 e mail 78 nadja machal bbraun com Karl Tippel Tanja Kordel Phone 149 5661 71 35 25 Fax 49 5661 71 35 26 e mail karl tippel bbraun com e mail tanja koerdel bbraun com B Braun Melsungen AG Schwarzenberger Weg 73 79 Wareneingang Werk C 34 212 Melsungen Germany Dr Ludwig Schiitz e mail ludwig schuetz bbraun com
44. e for safety reliability and per formance if mounting enhancements new settings changes or repairs are carried out by duly authorized persons the electrical installation in the corresponding room meets the requirements of the VDE 0107 VDE 0100 part 710 or IEC 60364 7 710 and the national standards the device is used in accordance with the instructions for use and the Service Manual the Technical Safety Checks are performed at regular inter vals a current manual which corresponds to the revision state is used when carrying out maintenance repair and service the service technician takes part in the revision service the technician has participated in a technical training course for the specific B Braun unit B Braun is certified in accordance with DIN EN ISO 9001 and ISO 13485 This certification also includes maintenance and serv ice The unit has the CE label The CE label confirms that the device corresponds to the Directive of the Council for Medical Products 93 42 EC of June 14 1993 Training may only be performed by B Braun The possession of the manual does not authorize the performance of repairs The in structions on electrostatic sensitive components ESD standards must be observed After repair a device check or diagnosis is to be carried out Semiconductors can be destroyed by electrostatic discharge Es pecially MOS components can be damaged by interference from electrostatic fiel
45. e required voltage The voltage is generated by the integrated power supply which is con nected to the supply mains via the connector F1A and a mains cable or an upstream SpaceStation The mains voltage is trans mitted via the internal cabling to connector F1B for connection of a further SpaceStation The pumps are connected with connectors F2A to F2D to the SpaceStation These connectors provide the voltage supply dis tribute the addresses in the Space system via a serial interface transfer data via a bus system CAN bus and transmit a staff call which may be pending Another SpaceStation inserted on top or a SpaceCover is integrat ed in the Space system via connector F5 Connector F4 is used for connecting another SpaceStation or further SpaceStation columns An upstream SpaceStation or an upstream SpaceStation column as well as accessories that may be connected such as a PCA button or a service connector are con nected to connector F3 SpaceStation 2 0 gb System Overview OV AO C Function processor L FE RXTXF 12VDC RX TX RX TXp p p 12VDC a 12VDC CAN PCA 3 3V PCA PCA Abb 1 6 Block diagram SpaceStation without SpaceCom SpaceStation 2 0 gb System Overview Abb 1 1 7 Legende zu Abb 1 7 ItemDesignation 1 0 N
46. e which might be installed in the SpaceCover comfort can drop out when the battery compartment cover is re moved Disassembly Assembly SpaceCover 5 3 Battery Module Only SpaceCover comfort Abb 5 3 Legende zu Abb 5 3 ItemDesignation 1 Battery module option Designation Ord No Battery pack SP NIMH ccc eee eae 0871 3180 Disassembly 1 Remove the battery module option Abb 5 3 Item 1 out of the battery compartment SpaceStation 1 0 gb 5 4 Handle Abb 5 4 Legende zu Abb 5 4 ItemDesignation 1 Handle 2 Washer M5 3 Fillister head screw M5x12 TORX 4 Cover cap for handle SpaceStation 2 0 gb Disassembly Assembly SpaceCover Ord No Handle SPC ccs 3452 1224 Designation Cover caps screws and washers see Service Parts and Screw Kit mp S 5 1 Disassembly 1 Pierce two cover caps Abb 5 4 Item 4 on the housing bottom part with a screwdriver and pull caps out 2 Unscrew two fillister head screws Abb 5 4 Item 3 and remove screws together with washers Abb 5 4 Item 2 3 Pull the handle Abb 5 4 Item 1 upward and out the housing upper part Disassembly Assembly SpaceCover 5 5 Housing Upper Part Designation Ord No Housing upper part SPCS aaa 3452 1194 Housing upper part SPCC aaa 3452 1178 Cover caps screws and release seal see Service Parts and Scre
47. eD Infusomat 6924000001 6924000001 692C000001 SpaceOnline only required with SpaceStation with SpaceCom SpaceStation 2 0 gb System Overview Compatibility List 1 Open the Unit Compatibility window via Help Decice compatibility This window displays the compatibility of the HiBaSeD version and the unit software version 2 Close the window by clicking OK Quit the Service Program 1 Exit the Service Program via File Exit Approved Version Note Please note that text and or functions of the SpaceOnline Pro gram may change depending on the software version The follow ing screen illustrations are only examples and represent the state when the manual was printed 120 Connection Before starting the SpaceOnline Program the PC is to be connect ed to the device via the RJ45 cross over interface cable In addi tion the PC LAN connection must be configured with the correct IP address e g 192 168 100 1 Make sure that the first three digit blocks e g 192 168 100 xxx of PC and the SpaceCom are identi cal Starting the SpaceOnline Program 1 Start the Internet Browser on your PC and enter the IP ad dress of the SpaceCom The SpaceOnline start screen appears 1 17 System Overview Datei Bearbeiten Ansicht Favoriten Extras 7 aW Q anc O x JO suchen she Fevoriten O 2 3 gt LJ Wechselnzu Links Adresse El http 192 168 100 105 spacecom en spacecom html
48. edial products for which these checks have been prescribed by the manufacturer and to carry them out according to the indications of the manufacturer as well as the generally approved technical standards while adhering to the pe riods stated S 6 MP BetreibV B Braun also recommends training on the Technical Safety Checks or to perform at least the steps indicated in the current version of the manual as the TSC requires that the instructions in the manuals are ob served the manuals are a reference for measurements depending on the unit type the Service Program must be called which may lead to a dangerous unit condition in case of inappropriate operation Furthermore a special service connector may be necessary This manual version corresponds to the state when the manual was written B Braun reserves the right to make technical modifi cations The state of the revision is indicated by the index number in the footer of every page The possession of this manual does not automatically mean inclu sion in the revision service You will be included in the revision service after technical training by B Braun Melsungen or awritten order placed with the sales department of B Braun fee required Important Preliminary Remarks Responsibility of the Manufacturer Quality Management Checks and Repair Notes on ESD The manufacturer person who assembles installs or imports the device can only be held responsibl
49. eously used on a patient If pre set values are taken over by the user these and all other entries are made at the user s risk The software may only be used by trained or specially instructed personnel Furthermore the devices are only to be serviced by personnel trained in the repair of this special equipment By pressing YES the user accepts all of these conditions v accept all conditions IV English Abb 1 11 Please observe the instructions for use of the software and the connected hardware components 1 14 Approved Version Note Please note that text and or functions of the Service Program may change depending on the software version The following screen illustrations are only examples and represent the state when the manual was printed 0 0 28 0 1 3 5 1 5 0 2 0 1 Connection Before starting the Service Program the PC is to be connected to the device via the RS232 SP interface cable Starting the Service Program Installation and further operation of the Service Program is de scribed in its separate instructions for use 1 Start the HiBaSeD exe program History Barcode Service Drug list on the PC The Service Program is loaded and start ed and the initial window of the Service Program is displayed 2 Read the notes carefully 3 Mark the field accept all conditions and then the field Yes to confirm that you have read the notes Note Click the fie
50. ewdriver 25IPx60 TORX plus Torque wrench with TORX adapter Ord No 0871 3301 3452 1062 0871 3234 0871 3237 Ord No 4002 4806 4002 4814 4002 4822 4002 4830 sr Test Equipment and Special Tools For your information 9 2 SpaceStation 1 0 gb SpaceStation SpaceCover SpaceStation 2 0 gb Spare Parts List 10 Service part kit SpaceStation 3477 4335 with Release button Release seal Seal M 2K Cover cap for housing Cover cap for handle Housing foot SPS Seal mains FIA Pole clamp lever Pole clamp spring O ring 6 07 x 1 78 O ring 12 0 x 2 0 Screw kit SpaceStation oooocooomomo 3477 4343 Screw EJOT 30x9 WN 5452 TORX 101P A2 Screw EJOT 30x12 WN 5452 TORX 101P A2 Screw EJOT 30x16 WN 5451 TORX 10IP A2 Screw M5x45 A2 DIN912 TORX 25 Countersunk screw M5x12 TORX 25 Fillister head screw M5x12 EN IS014583 A2 Nut M4 Insulating washer M3 Washer M4 Washer M5 Serrated lock washer M4 Serrated lock washer M5 10 1 Spare Parts List SpaceStation without SpaceCom rrr d Abb 10 1 Exploded drawing SpaceStation 10 10 2 SpaceStation 2 0 gb SpaceStation 1 0 gb N DO oO A CO N Spare Parts List Tube guide SPS aaa 3477 4394 20 pieces Pole clamp guide SPS
51. ews seals and housing foot SPS see Service Parts and Screw Kit S 3 2 Disassembly 1 Pull two housing feet Abb 3 22 Item 1 out of the hous ing 2 Remove four release seals Abb 3 22 Item 2 from the housing Note The seals are self adhesive Remove any adhesive residues after disassembly 3 Remove mains seal F1B Abb 3 22 Item 3 out of the housing Note The housing must not be disassembled further The seals behind the corner profiles on the right and left side can only be inserted correctly by means of special tools SpaceStation 1 0 gb 3 10 Assembly Installation Abb 3 23 Legende zu Abb 3 23 ItemDesignation 1 Connector board F5 2 Screw EJOT 30x9 WN 5452 TORX 10IP A2 SpaceStation 1 0 gb Disassembly Assembly SpaceStation Assembly or installation of the modules and subsystems is done in reverse order of disassembly Special steps to be observed are de scribed hereafter in detail Only new cover caps are to be used Special Screws Special screws for plastic housings are used in this unit The screws are not self cutting but produce a thread in the plastic of the housing through deformation when fitted in for the first time If the beginning of the thread is not engaged when the screw is fitted a new thread is produced and the old thread is destroyed so that the security of the fixing can no longer be guaranteed Proceed as follows to fit
52. found on the right side and the rear side Interfaces and a status and alarm LED are located on the right hand side behind the connector cover and the W LAN cover A battery module can be inserted in the battery compartment on the rear side The opening is covered by the battery compartment cover A fan grid with fan is fitted on the left side of the battery com partment SpaceStation 2 0 gb Abb 1 5 SpaceCover comfort Legende zu Abb 1 5 ItemDesignation 1 SpaceCover comfort SpaceStation 1 0 gb System Overview EN SpaceCover standard The SpaceCover standard housing consists mainly of the bottom part and the upper part A SpaceStation can be transported with the carrying handle SpaceCover comfort The physical construction of the SpaceCover comfort corresponds to that of the SpaceCover standard In addition to the SpaceCover standard the following subsystems are installed in the SpaceCover comfort The battery module as option It is inserted from below in the battery compartment of the housing bottom part The opening is covered by the battery compartment cover Status and alarm display line installed at the front of the housing upper part Operating and status display field at the left side of the housing upper part Function System Overview SpaceStation without SpaceCom The SpaceStation supplies the inserted pumps as well as a Space Control connected to the pumps with th
53. gende zu Abb 3 6 ItemDesignation 1 Cable lug for connection to the power supply Cable lug for connection to connector F1B Washer M4 Serrated lock washer M4 Nut M4 a KB CO N SpaceStation 2 0 gb Disassembly Assembly SpaceStation Designation Ord No Power supply SPS aaa 3452 1097 Connector holder mains SPS 22055 3452 1143 Cover cap screws washer serrated lock washers and nuts see Service Parts and Screw Kit S 3 2 Mains seal bottom see Service Parts and Screw Kit S 3 2 Disassembly 1 Pierce two cover caps Abb 3 5 Item 1 with a screwdriver and pull caps out 2 Unscrew two screws Abb 3 5 Item 2 and remove them from the mains connector holder Note Note the position of the protective conductor cable on the grounding bolt 3 Loosen the nut Abb 3 6 Item 5 and remove nut together with the serrated lock washer Abb 3 6 Item 4 washer Abb 3 6 Item 3 the protective conductor cable to the power supply Abb 3 6 Item 1 and the protective con ductor cable to connector F1B Abb 3 6 Item 2 from the grounding bolt Disassembly Assembly SpaceStation Abb 3 7 Legende zu Abb 3 7 ItemDesignation 1 Washer M4 Cable lug for connection to connector F1A Serrated lock washer M4 Nut M4 B W N Abb 3 8 Legende zu Abb 3 8 ItemDesignation 1 Screw EJOT 30x9 WN
54. gnation 1 Housing notch 2 Screw EJOT 30x9 WN 5452 TORX 101P A2 3 Cover cap 4 Housing back panel SpaceStation 1 0 gb Disassembly Assembly SpaceStation Designation Ord No Housing back panel SPS eee eae 3452 1089 Cover cap and screws see Service Parts and Screw Kit S 3 2 Disassembly 1 Pierce ten cover caps Abb 3 3 Item 3 with a screwdriver and pull caps out 2 Loosen nine screws Abb 3 3 Item 2 and remove screws carefully together with the back panel Abb 3 3 Item 4 from the housing Note Lift the housing back panel carefully at the bottom over the seat of the pole clamp guide and push the back panel over the housing notch Abb 3 3 Item 1 Disassembly Assembly SpaceStation 3 5 Device Bracket Abb 3 4 Legende zu Abb 3 4 ItemDesignation 1 Serrated lock washer M5 2 Screw M5x45 A2 DIN912 TORX 25 3 Device bracket Designation Ord No Device bracket SPS 0 0 0 ccc eee eee 3452 1127 Screws see Service Parts and Screw Kit S 3 2 Disassembly 1 Loosen five screws Abb 3 4 Item 2 and remove screws together with the serrated lock washers Abb 3 4 Item 1 and the device bracket Abb 3 4 Item 3 SpaceStation 2 0 gb 3 6 Power Supply Abb 3 5 Legende zu Abb 3 5 ItemDesignation 1 Cover cap 2 Screw EJOT 30x9 WN 5452 TORX 101P A2 5 Abb 3 6 Le
55. he SpaceCover comfort Switch the SpaceStation or the SpaceStation column on and then off via the SpaceCover comfort A self test is carried out SpaceStation 2 0 gb SpaceStation 2 0 gb 5 Procedural Instructions on the TSC Reconnect the SpaceStation to the mains The status display colour on the operator and status display field of the Space Cover comfort changes from yellow to green Interface Check of the SpaceStation with SpaceCom 1 2 Insert a pump in the SpaceStation Connect a PC to the SpaceStation via the interface cable RJ 45 cross over and start the Space Online Program see SpaceOnline S 1 17 Check in the Service Part whether the correct pump type and its On Off status are displayed Note If the IP address of the SpaceStation is unknown all parameters can be reset to the factory settings from software version 695E010001 on via a USB stick with the factorydefault directo ry saved here When the word factorydefault is read automati cally resetting of the parameters in the SpaceStation is started The IP address of the SpaceStation is set to 192 168 100 41 Connect a PC with the SpaceStation via the interface cable RS232 SPCO crossed and start the BCCshow Program see BCCshow gt S 1 19 Check whether the correct pump and its status are displayed on the PC Battery Check of the SpaceStation with SpaceCom 1 Insert one battery module in the SpaceStation and one in
56. isplay line 16 Replace interface board in the SpaceStation UTS 20 17 Pull off the fan connecting cable on the SPCO The fan is running 15 19 TS 18 board and trigger the fan with 12 V DC Connect the pole to the connector pole with red wires and the pole to the connector pole with black wires 18 Exchange fan UTS COM 5 19 The status indicator green of the SpaceCom lights TS 20 TS 21 up 20 Replace the SPCO board UTS COM 5 21 Replace interface board in the SpaceStation UTS COM 5 22 Replace battery module of the SpaceStation The charge condition of the battery module is dis UTS COM 8 TS 23 played in the bottom charge condition display of the operator and status display field of the Space Cover comfort 23 Replace the SPCO board The charge condition of the battery module is dis UTS COM 8 JS 24 played in the bottom charge condition display of the operator and status display field of the Space Cover comfort 24 Replace interface board in the SpaceStation UTS COM 8 Table 2 5 Trouble shooting Part 2 of 2 SpaceStation 2 0 gb 2 7 Unit Diagnosis Calibration For your information 2 8 SpaceStation 2 0 gb Disassembly Assembly SpaceStation 3 3 1 General on the SpaceStation without Space Com SpaceStation 2 0 gb Remarks on Disassembly Assembly of the SpaceStation with out SpaceCom WARNING DURING DISASSEMBLY AND ASSEMBLY THE OPERATOR SERVICE TECHNICIAN MUST
57. its final position Note Pay attention to the cable connections when placing on the hous ing back panel These must not be squeezed Note Pay attention to the different screw lengths when mounting the housing back panel The longer screws are provided for the left hand and right hand upper screw opening 4 No screw is inserted in the third mounting hole from the top on the right hand side of the back panel The hole must be covered with a cover cap 4 11 Checks after Repair 4 14 1 Check the device to ensure safe functionality of the unit see Device Check S 2 4 2 Depending on the work carried out the specific steps of the TSC must be performed see Technical Safety Check TSC S 7 1 3 The activities carried out must be recorded in the following check list SpaceStation 2 0 gb Disassembly Assembly SpaceCom Check List for Checks after Repair Visual Inspection Electrical Safety Functional Inspection according to IEC EN 60601 1 or VDE 0750 and VDE 0751 O Cleanliness O Mains voltage according to the TSC O Locking with second SPS C Completeness 3 Protective conductor resistance ac O Pump locking of all slots O Damage and faults affecting safety cording to the TSC O Damage to and readability of the label O Earth leakage current according to the Connect unit to mains O Cover caps TSC General functional check with SpaceCover C Connectors O Housing leakage
58. ld English to switch the language of the notes over to English SpaceStation 2 0 gb ER Programm Auswahl Programm Auswahl C History Viewer C Clone Data Service Program Pabwot Start Abb 1 12 Schnittstellen Scanner seriell Erkenne Ger te Pr fung aktiv com1 Pr fung aktiv com1 Kein Zielgerat gefunden an com Pr fung aktiv com3 Pr fung aktiv com3 Pr fung aktiv com3 Pr fung aktiv com3 Abbrechen Abb 1 13 SpaceStation 2 0 gb System Overview 4 Enterthe password and confirm it by clicking the field Start The Service Program checks the PC interfaces for connected devices of the Space system Units that were found are dis played for a short moment on the screen Note At present a connected SpaceStation is not displayed The work window of the Service Program appears on the screen All devices recognized are listed in the left column 1 15 System Overview HiBaSeD File Extra Tools Update Help Abb 1 14 BIBRAUN zh Version 2 0 1 N www BBraun com Abb 1 15 1 16 Service Program Version 1 Open the HiBaSeD window via He p Info The current version of the Service Program is shown in this window 2 Close the window by clicking OK SpaceStation 2 0 gb Device compatibility List of versions Perfusor Infusomat Cover 1 35 2 0 1 Abb 1 16 Version HiBaS
59. n Unit Software Service Program SpaceOnline only required with SpaceStation with SpaceCom BCCshow Technical Data Options Accessories General Alarms and Error Codes The Most Important Error Modes Device Check Device Check SpaceCom Trouble Shooting General on the SpaceStation without SpaceCom Tube Guide Pole Clamp Guide Housing Back Panel Device Bracket Power Supply Release Button Interface Board Module Lock Housing Assembly Installation Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page O0o0O0OO0O0OO0OOOOOoOOoO 1 NNOO OD 0101 Y C1 0 11 0 11 0 11 0 11 0 11 0 11 0 11 1 1 1 1 1 2 1 6 1 12 1 14 1 17 1 17 1 19 1 21 1 21 1 21 2 1 2 3 2 4 2 4 2 5 2 6 Table of Contents Checks after Repair Page 3 24 Disassembly Assembly SpaceCom General on the SpaceStation with SpaceCom Page 4 1 Battery Module Page 4 3 W LAN Module Page 4 4 Tube Guide Page 4 5 Pole Clamp Guide Page 4 6 Housing Back Panel Page 4 7 Device Bracket Page 4 10 SPCO PCB Page 4 11 Further Disassembly Page 4 12 Assembly Installation Page 4 13 Checks after Repair Page 4 14 Disassembly Assembly SpaceCover General Page 5 1 Battery Compartment Cover Page 5 3 Battery Module Page 5 4 Only
60. nd operation One single pump can be inserted onto this module The pump is then connected via connectors to the module The SpaceStation module allows the set up of a complete pump system with up to 24 pumps Up to four pumps can be installed in every SpaceStation The pumps are supplied with power via the integrated power supply and the built in connectors The pumps are connected to the optional SpaceCom via these connectors SpaceControl can also be integrated into the system Up to six SpaceStations can be set up as a column with a total of 24 pumps SpaceStation placed next to each other can be con nected via special connecting cables if the maximum number of 24 pumps in maximum three columns is not exceeded SpaceCover Standard or SpaceCover Comfort forms the top of each column Alarms are signalled by a row of LEDs and a loud speaker in the SpaceCover Comfort System Overview Physical construction Abb 1 2 SpaceStation front view Legende zu Abb 1 2 ItemDesignation 1 Pump lock Connector F5 Module lock Connector F1B mains output Tube guide Connector F2D ao c1 A CO N SpaceStation without SpaceCom The SpaceStation consists of the housing and the housing back panel The power supply with power connector and power socket for an other SpaceStation is installed on the right side in the housing The pumps in the SpaceStation are locked by 8 locking claws which are built in the ho
61. on without SpaceCom can be upgraded with this retrofit kit Designation Part No Charger SP sisse Wedd Ra ages 0871 3170 battery charging station Connecting cable staff call SP 0871 3232 Power supply cable 12V 0 00 eee eee eee 0871 3231 for ambulance cars Interface cable CAN SP ooo 0871 3230 Interface cable RS232 SP eee eee 0871 3234 Interface cable RS232 SPCO crossed 0871 3237 Converter RS232 SP asesorando 0871 3238 W LAN Gard ii aora eros 0871 3184 Connecting cable SP 60cm ee eee eee eee 0871 3210 for connecting two columns Connecting cable SP 120cm suuses 0871 3215 for connecting two columns PoleClamp SP aa eee kaa 0871 3130 The PoleClamp SP is a holder for one or several Space system pumps Universal clamp aa 3452 1151 1 21 System Overview For your information 1 22 SpaceStation 1 0 gb General SpaceStation 2 0 gb Unit Diagnosis Calibration 2 WARNING WHILE TESTING THE UNIT AND TROUBLE SHOOTING THE OPERA TOR SERVICE TECHNICIAN MUST WORK WITH VOLTAGES UP TO 115 230 V AC THESE VOLTAGES MAY CAUSE INJURIES WHICH ARE DANGEROUS TO LIFE AND LIMB THE NATIONAL AND INTER NATIONAL SAFETY REGULATIONS ARE TO BE ADHERED TO Before each disassembly of a unit subsystem check the connec tors plug contacts and connections for corrosion and tight fit These fault types are no
62. ortant error modes and their clearance Note The error list is in preparation Device Check 1 Detach or loosen all connections and devices from UTS 2 the SpaceStation 2 Connect the SpaceStation via the mains cable to UTS 3 the mains 3 Insert one pump after the other in each slot The pump is locked UTS 4 TS 1 4 The pump can be released at any slot UTS 5 TS 1 5 Plug service connector SP on connector F3 UTS 6 6 Insert the pump in the topmost slot and triggera The red LED of the service connector SP lights up UTS 7 TS 4 staff call Remove the pump while it is switched on The red LED of the service connector SP goes out UTS 8 TS 4 Insert the switched on pump in each other slot The red LED of the service connector SP lights up UTS 9 TS 4 when the lock engages 9 ASpaceCover comfort with SpaceStation is in UTS 10 UTS 20 stalled 10 Mount and lock the SpaceCover comfort on the All luminous fields of the status and alarm display UTS 11 TS 5 SpaceStation line light up briefly during the self test 11 A short audible alarm sounds UTS 12 TS 7 Table 2 3 Device check Part 1 of 2 N Li A5 SpaceStation 2 0 gb Unit Diagnosis Calibration 12 All LEDs of the operator and status display field light UTS 13 TS 5 up briefly 13 The green LED on the operator and status display UTS 14 TS 5 field lights up 14 When the battery module i
63. ove screws together with the insulating washers Abb 3 19 Item 2 Abb 3 19 Legende zu Abb 3 19 ItemDesignation 1 Screw EJOT 30x9 WN 5452 TORX 101P A2S 2 Insulating washer M3 2 Press connectors F3 Abb 3 20 Item 2 and F4 Abb 3 20 Item 3 carefully out of the data connector holder Abb 3 20 Item 1 Note 1 The connectors F3 and F4 are integrated in the interface board Abb 3 20 Legende zu Abb 3 20 ItemDesignation 1 Data connector holder 2 Connector F3 3 Connector F4 3 18 SpaceStation 1 0 gb Abb 3 21 Legende zu Abb 3 21 ItemDesignation 1 Module lock 2 Connector board 3 Connector board lockings SpaceStation 1 0 gb Disassembly Assembly SpaceStation 3 Press the connector board lockings Abb 3 21 Item 3 F2A F2B F2C and F2D carefully aside push the connectors downwards pull them out of the module lock Abb 3 21 Item 1 and remove them with the interface board Abb 3 21 Item 2 from the module lock Note The connectors F2A F2B F2C and F2D are connected to each oth er and integrated in the interface board 3 19 Disassembly Assembly SpaceStation 3 9 Housing Abb 3 22 Legende zu Abb 3 22 ItemDesignation 1 Housing foot SPS 2 Release seal 3 Mains sealF1B Ord No Housing SPS a 3452 1070 incl mains seal F1B Designation Scr
64. paceCom lights UTS COM 7 1520 up for approx 1 minute and goes out Table 2 4 Device check SpaceCom Part 1 of 2 SpaceStation 2 0 gb 2 5 Unit Diagnosis Calibration from the SpaceStation 7 The charge condition of the battery module is dis UTS COM8 T522 played in the bottom charge condition display of the operator and status display field of the Space Cover comfort 8 Connect the SpaceStation with the interface ca UTS COM 9 ble RJ45 cross over via the Ethernet RJ45 con nection of the SpaceCom to a PC 9 Start the SpaceOnline Program see SpaceOnli The data of the SpaceStation and the pumps in UTS COM TS 20 ne S 1 17 serted is displayed on the PC monitor 10 10 Connect the SpaceStation with the interface ca UTS COM ble RS232 SPCO crossed via the RS 232 connec 11 tor of the SpaceCom to a PC 11 Start the BCCshow Program see BCCshow The data of the SpaceStation and the pumps in UTS COM TS 20 S 1 19 serted is displayed on the PC monitor 12 12 Detach or loosen all connections and devices This step terminates the device check Table 2 4 Device check SpaceCom Part 2 of 2 Trouble Shooting Note The following trouble shooting cannot be carried out independ ently It is based on the precise observance of the steps for the de vice check see Device Check m S 2 4 From there reference is made to the corresponding trouble shooting steps
65. pply transmits a unique address in the Space system via a serial interface transfers data via a bus system CAN bus and transmits a staff call which may be pending The internal electronics controls the operator and status display field the loudspeaker as well as the status and alarm display line The status and alarm display line displays the collected messages of the pumps in a column The brightness is controlled dependent on the environment The voltage supplied is converted to the internal voltages required through a voltage transforming and monitoring circuit on the processor PCB An independent circuit in the battery module monitors the battery cells and controls their charge condition SpaceStation 1 0 gb System Overview Loudspeaker Function processor CAN KB LEDs e transceiver Keyboard J Operating unit YH 12 0 Mhz Monitoring 12VDC 12VDC battery Brightness sensor Abb 1 9 Block diagram SpaceCover comfort SpaceStation 1 0 gb 1 11 System Overview Unit Software Position 12 3 4 5 6 7 8 9 10 Digit 6 9 O A O OS S 0 2 Revision level Hardware Software group Device type SpaceStation Abb 1 10 1 12 Approved Software Versions for SpaceStation 690A00SS01
66. r your information B BRAUN M690 00 00 46 F04 3891 7548 GB dis SHARING EXPERTISE SpaceStation 1 0 gb Sheet 2 of 2 Procedural Instructions on the TSC 8 Visual Inspection Electrical Safety according to IEC EN 60601 1 or VDE 0750 and VDE 0751 SpaceStation 2 0 gb 1 Check the unit and accessories for cleanliness 2 Check the unit and accessories for completeness and check configuration 3 Check the unit and its accessories for damage and the labels for readability Pay special attention to the following parts a b c d e Pump suspension Tightness of pump locks Connectors 8 tube guides Mains cable The values to be measured for the electrical safety are stated in the TSC see Technical Safety Check TSC S 7 1 Protective Conductor Resistance Note The protective conductor resistance of each SpaceStation must be measured separately Before carrying out the measurement re move all interface cables to the other units Measure between the protective conductor of the mains cable and the following parts Ground contact in connector F1B middle pin Procedural Instructions on the TSC Earth Leakage Current Note Due to the technical design the patient and housing leakage cur rent is not to be measured on the SpaceStation without Space Com and without pump inserted The earth leakage current is always measured as it i
67. rther Disassembly 4 12 6 Remove the fastening sleeve Abb 4 11 Item 2 out of the housing 7 Remove the insulating piece Abb 4 11 Item 1 under the connector board F2A Further disassembly of the SpaceStation is continued with the power supply and corresponds to the disassembly of a SpaceSta tion with SpaceCom This disassembly is described in Chapter Po wer Supply S 3 7 in more detail SpaceStation 2 0 gb 4 10 Assembly Installation Abb 4 12 Legende zu Abb 4 12 ItemDesignation 1 Groove 2 Contact strip 3 Screw A2 WN 5452 30x9 4 Spacing 2 mm SpaceStation 2 0 gb Disassembly Assembly SpaceCom Assembly or installation of the modules and subsystems is done in reverse order of disassembly Special steps to be observed are de scribed hereafter in detail Only new cover caps are to be used Special Screws Special screws for plastic housings are used in this unit The screws are not self cutting but produce a thread in the plastic of the housing through deformation when fitted in for the first time If the beginning of the thread is not engaged when the screw is fitted a new thread is produced and the old thread is destroyed so that the security of the fixing can no longer be guaranteed Proceed as follows to fit the special screws 1 Put the screw on the thread 2 Rotate screw anti clockwise loosen until a faint click can be heard This click is pro
68. s inserted the charge UTS 15 TS 9 condition is displayed in the top charge status dis play of the operator and status display field 15 Change loudness of the alarm tone The loudness changes UTS 16 TS 11 16 Insert pump in the topmost slot and start an infu The infusion is displayed green in the status and UTS 17 TS 13 sion alarm display line 17 Insert the pump in each other slot and start an in The infusion is displayed green in the status and UTS 18 TS 14 fusion alarm display line 18 Trigger an alarm on the pump The alarm is displayed red orange in case of a pre UTS 19 TS 15 alarm in the status and alarm display line 19 If a SpaceCom is integrated in the SpaceStation UTS 20 carry out the device check according to Chapter Device Check SpaceCom sm S 2 5 20 Detach or loosen all connections and devices from This step terminates the SpaceStation the device check Table 2 3 Device check Part 2 of 2 Device Check SpaceCom Disconnect the SpaceStation from the mains and UTS COM 2 switch off a SpaceCover comfort which may be fitted 2 A battery module is installed in the SpaceStation UTS COM 3 UTS COM 4 3 Inserta battery module in the SpaceStation UTS COM 4 4 Connect the SpaceStation to the mains Fan in the housing back panel starts running for UTS COM 5 1S 17 appr 1 sec 5 The status indicator green of the SpaceCom UTS COM 6 TS 20 lights up 6 The alarm indicator red of the S
69. s the upper limit of leakage currents which can be measured The earth leakage current of each SpaceStation must be measured separately Check 1 Before carrying out the measurement remove all interface cables to the other units 2 Remove all pumps out of the SpaceStation 3 Measure the earth leakage current without single fault con dition NC Normal Condition incl mains cable 4 Document value 5 Measure the earth leakage current in single fault condition SFC Single Fault Condition incl mains cable 6 Document value SpaceStation 2 0 gb Abb 8 1 Legende zu Abb 8 1 ItemDesignation 1 Metal frame of the Ethernet RJ 45 connection SpaceStation 2 0 gb Procedural Instructions on the TSC Housing Leakage Current with the SpaceStation with SpaceCom Measurement is to be carried out at nominal voltage 10 in sin gle fault condition SFC Single Fault Condition incl mains ca ble 1 Measure the housing leakage current between mains input at the mains cable and the metal frame of the Ethernet RJ 45 connection in the SpaceCom plug strip Note The metal frames of the connectors in the SpaceCom plug strip are connected with each other 2 Document value Procedural Instructions on the TSC Functional Inspection Mechanical 1 Check proper functioning of the lock of all pump slots with one pump Check the battery compartment cover lock of the SpaceSta
70. t described again in the following trouble shooting list The following equipment is necessary for testing the SpaceStation without SpaceCom and or performing trouble shooting PC Service connector SP Service Program HiBaSeD Interface cable RS232 SP Space system pump SpaceCover comfort for unit diagnosis of the SpaceStation The following additional equipment is necessary for testing the SpaceStation with SpaceCom and or performing trouble shoot ing Interface cable CAN SP Interface cable RS232 SPCO crossed Interface cable RJ45 cross over CAUTION Take special care when carrying out measurements on an open and switched on unit Short circuits and wrong measuring meth ods can cause serious damage to or destroy the subsystems of the device Unit Diagnosis Calibration The unit check calibration and trouble shooting are subdivided into numbered working steps Unit Test Step UTS Calibration Step CS Trouble Shooting TS and are based on each other Beginning with UTS 1 the operation described here has to be ex ecuted The consequences of the steps performed are listed in the Function column If the result corresponds to the consequence the working step must be carried out to which reference is made in the column If yes If the result does not correspond with the function described the working step in column If no is to be ex ecuted One example is given in Abb 2 1
71. tton SpaceStation 1 0 gb Disassembly Assembly SpaceStation Designation Ord No Release button and release seal see Service Parts and Screw Kit S 3 2 Disassembly 1 Press four release buttons Abb 3 11 Item 1 carefully from the inside from the locking shaft using a screwdriver and pull the buttons out of the housing Note Check the release buttons for damage especially within the notch area before fitting the buttons 3 11 Disassembly Assembly SpaceStation 3 8 Interface Board Module Lock Abb 3 12 Legende zu Abb 3 12 ItemDesignation 1 Module lock screw 2 O Ring 6 07 x x1 78 3 Module lock sleeve 3 12 Designation Ord No Interface board SPS 0c cece eee eee 3452 1100 with connectors Connector holder data SPS 3452 1160 Module lock SPS a 3452 1119 Cover cap screws insulating washers O rings see Service Parts and Screw Kit S 3 2 Disassembly 1 Move the lock towards the SpaceCover or another SpaceSta tion in the closed position 2 Loosen one screw Abb 3 12 Item 1 of the module lock and remove it out of the housing 3 Pull the module lock sleeve Abb 3 12 Item 3 out of the housing 4 Check the O ring Abb 3 12 Item 2 for damage and re move it out of the housing if necessary SpaceStation 1 0 gb Disassembly Assembly SpaceStation 5 Loosen two screws Abb 3 1
72. ule TTT Abb 4 1 Legende zu Abb 4 1 ItemDesignation 1 Battery module 2 Battery compartment cover 3 Battery compartment cover lock 4 Battery module lock SpaceStation 2 0 gb Disassembly Assembly SpaceCom Designation Ord No Battery pack SP NIMH oooocococoococoo 0871 3180 Battery compartment cover SPC 3452 1232 Disassembly 1 Rotate the lock Abb 4 1 Item 3 on the battery compart ment cover Abb 4 1 Item 2 and remove the battery compartment cover 2 Push the lock Abb 4 1 Item 4 on the battery module downward and remove the battery module Abb 4 1 Item 1 out of the battery compartment Disassembly Assembly SpaceCom 4 3 W LAN Module U NN Abb 4 2 Legende zu Abb 4 2 ItemDesignation 1 W LAN cover 2 W LAN module 3 Screw EJOT 30x12 WN 5452 TORX 10IP A2 4 Cover cap 4 4 Designation Ord No W LAN module 200000 eee eee eens on request W LAN cover SPCO 2 20020 cee eee eee 3452 1049 with symbol W LAN cover SPCO 222002 eee eee eee 3452 1050 without symbol Cover cap and screw see Service Parts and Screw Kit S 4 2 Disassembly 1 Pierce cover cap Abb 4 2 Item 4 with a screwdriver and pull cap out 2 Unscrew screw Abb 4 2 Item 3 and remove screw to gether with the W LAN cover Abb 4 2 Item 1 3 Pull W LAN module Abb
73. using as subsystem module locking This subsystem also contains the locking for a SpaceCover or another SpaceStation The pump is released with a release button on the left side of the housing The pumps are connected to the system via the connectors F2A to F2D from top to bottom which are integrated in the inter face board while the SpaceStation is connected to other modules via the connectors F3 F4 and F5 The interface board with connectors is installed on the left side in the housing SpaceStation 2 0 gb m i a wc Abb 1 3 SpaceStation without SpaceCom rear view Legende zu Abb 1 3 ItemDesignation 1 Pole clamp guide Release button Connector F3 Connector F4 ao KB CO N Connector F1A power input SpaceStation 2 0 gb System Overview Four tube guides are clipped in on the left and right housing side A pole clamp guide for attachment and locking of the pole clamp is located at the housing back panel System Overview Abb 1 4 SpaceStation with SpaceCom rear view Legende zu Abb 1 4 ItemDesignation 1 Fan grid 2 Battery compartment with battery compartment cover 3 Connector cover 4 W LAN cover for W LAN module SpaceStation with SpaceCom The physical construction of the SpaceStation front side with SpaceCom corresponds exactly to the physical construction of the SpaceStation without SpaceCom The only visible difference is to be
74. w Kit mp S 5 1 Disassembly 1 Pierce five cover caps Abb 5 5 Item 4 with a screwdriver and pull caps out 2 Unscrew five screws Abb 5 5 Item 3 and remove the housing upper part Abb 5 5 Item 1 carefully Note Pay attention to the length of the rubber cable between the op erator panel and the processor PCB SCc when dismounting the housing upper part of the SpaceCover comfort Note Do not remove the housing back panel together with the housing upper part Abb 5 5 Legende zu Abb 5 5 ItemDesignation 1 Housing upper part Seal Screw EJOT 30x9 WN 5452 TORX 10IP A2 BR WN Cover cap for housing 5 6 SpaceStation 1 0 gb Disassembly Assembly SpaceCover 3 Pull the connector lock Abb 5 6 Item 1 on the processor PCB SCc upward 4 Remove the rubber cable to the operator panel 5 Remove two seals Abb 5 5 Item 2 in the housing bottom part Abb 5 6 Legende zu Abb 5 6 ItemDesignation 1 Connector lock SpaceStation 1 0 gb 5 7 Disassembly Assembly SpaceCover 5 6 Housing Back Panel Abb 5 7 Legende zu Abb 5 7 ItemDesignation 1 Connector board 2 Housing back panel 3 Screw EJOT 30x9 WN 5452 TORX 10IP A2 Designation Ord No Housing back panel SPC Luuuuuuuuuu 3452 1216 Connector board SPCS 0000 02 eee eee 3452 1240 only for SpaceCover standard Screws see
75. w M5x12 EN IS014583 A2 Nut M4 Insulating washer M3 Washer M4 Washer M5 Serrated lock washer M4 Serrated lock washer M5 SpaceStation 2 0 gb 3 2 Tube Guide Abb 3 1 Legende zu Abb 3 1 ItemDesignation 1 Tube guide SpaceStation 1 0 gb Disassembly Assembly SpaceStation Designation Ord No Tube guide SPS aaa 3477 4394 20 pieces Disassembly 1 Press the rotary knob of the tube guide out and pull the tube guide Abb 3 1 Item 1 out of the housing Note When the tube guide is broken remove the remaining parts from the housing Disassembly Assembly SpaceStation 3 3 Pole Clamp Guide Abb 3 2 Legende zu Abb 3 2 ItemDesignation 1 Pole clamp guide 2 Countersunk screw M5x12 TORX 25 3 Pole clamp spring 4 Pole clamp lever 3 4 Ord No Pole clamp guide SPS aan 3452 1135 Designation Pole clamp lever see Service Parts and Screw Kit S 3 2 Pole clamp spring see Service Parts and Screw Kit S 3 2 Screws see Service Parts and Screw Kit S 3 2 Disassembly 1 Loosen three screws Abb 3 2 Item 2 and remove the screws together with the pole clamp guide Abb 3 2 Item 1 2 Remove pole clamp lever Abb 3 2 Item 4 and pole clamp spring Abb 3 2 Item 3 SpaceStation 1 0 gb 3 4 Housing Back Panel Abb 3 3 Legende zu Abb 3 3 ItemDesi
Download Pdf Manuals
Related Search
Related Contents
Leica A60 F Manual - Leica Microsystems Manuel d`utilisation - Livre de pièces de rechange Conveyor TR Magnavox MPC500 User's Manual Liberty C080X0C-BR-C7 Installation Guide Détecteur de fuites de gaz combustibles et de StarTech.com 15 ft Beige A to B USB 2.0 Cable - M/M Réseau des bibliothèques de la CAPE Création du dossier CampusFrance Côte d`Ivoire SOMMAIRE Manual de Instrucciones del Juego del Paisaje Copyright © All rights reserved.